Cones and sleeves present good survival and clinical outcome in revision total knee arthroplasty: a meta-analysis

Laura Theresa Fischer 1,, Markus Heinecke 1, Eric Röhner 1, Peter Schlattmann 2, Georg Matziolis 1
PMCID: PMC9309140  PMID: 34389876

Abstract

Purpose

The fixation of revision total knee arthroplasties (rTKA) tends to be difficult, leading to a reduction in implant survival. One option for achieving a more stable anchorage is to use metaphyseal cones and sleeves. The objective of the present paper is to provide a current comparative meta-analysis on survival and clinical results of cones vs. sleeves, with a differentiation between the short- and long-term outcome.

Methods

A search of the literature was conducted systematically to include original papers from 2010 to June 2021. The following parameters were taken into account: revision for aseptic loosening, revision for any reason, periprosthetic joint infections (PJI), KSS as well as KSFS. Studies with a mean follow-up of at least 60 months were defined to be long-term follow-up studies (LT). All other studies were included in the short-term (ST) study analysis. A pooled incidence was used as a summary statistic using a random intercept logistic regression model.

Results

The present meta-analysis included 43 publications with 3008 rTKA. Of these, 23 publications with 1911 cases were allocated to the sleeve group (SG) and 20 papers with 1097 cases to the cone group (CG). CG showed overall numerically higher complication rates in short- and long-term follow-up, compared with SG. Aseptic loosening occurred at a rate of 0.4% in SG (LT) and 4.1% in CG (LT) (p = 0.09). Periprosthetic joint infection (PJI) was more frequent in the cone group (7% in ST and 11.7% in LT) than in the sleeve group (3.4% in ST and 4.9% in LT, p = 0.02 both). The total revision rate was 5.5% in SG (LT) and 14.4% in CG (LT) (p = 0.12). The clinical scores were also comparable between the two groups. Hinged prothesis were used more frequent in the cone group (ST p < 0.001; LT p = 0.10), whereas CC type protheses were used more frequently in the sleeve group (ST p < 0.001; LT p < 0.11).

Conclusions

This meta-analysis takes into account the longest follow-up periods covered to date. Both cones and sleeves represent a reliable fixation method in the case of severe bone loss in rTKA, although the higher rate of PJI after cone fixation remains a source of concern. A metaphyseal fixation of hinged implants should be taken into account.

Level of evidence

II (meta-analysis).

Keywords: Revision total knee arthroplasty, Total knee replacement, Bone defects, Cones, Sleeves, Aseptic loosening

Introduction

Regardless of the reason for revision, implant survival is reduced after revision total knee arthroplasty (rTKA) [23]. One reason for this is the challenging fixation of the implant in the bone stock, which is compromised both in substance and in density [5, 8, 14, 28, 29].

Implant fixation and defect management are oriented according to the estimated bone defect and bone quality [11, 28]. Various techniques are used to achieve the stable anchorage of a revision implant. Apart from cement, allografts, wedges and stem extensions, cones and sleeves have become increasingly popular over the past few years. This is because an additional metaphyseal implant anchorage is superior to a sole diaphyseal fixation [22].

Cones may be regarded as metaphyseally anchored metallic bone grafts, which enable a cementless fixation over their porous surface. Sleeves also follow the principle of cement-free metaphyseal anchorage. In contrast to cones, sleeves are firmly attached to the prosthesis. However, it also bears the risk of fractures during broaching, which represents the most common intraoperative complication when sleeves are used [17, 24, 31].

In numerous individual studies, excellent implant survival has been shown both for cones and for sleeves after a medium-term follow-up [6, 9, 12, 17, 21, 23, 33]. The few studies with long-term follow-up (> 7 years) [1, 3, 6, 13, 15, 25] could not be taken into account in previous meta-analyses [18, 27, 30, 34]. Nevertheless, there is initial evidence of a drop in survival over long-term follow-up [1].

It is thereby obvious, that the revision reasons differ depending on timepoint of failure [2, 3]. Revisions within the first years after implantation of cones or sleeves may result from failing bony integration or persistence of infection (in septic revisions). In contrast, late revisions may result from aseptic loosening of a primarily integrated implant or new infection. Given a different aetiology of failure, specific information about the short- and long-term outcome of cones and sleeves is missing.

Since all published meta-analyses [18, 27, 30, 34] include studies without differentiation between short- and long-term survival, there is lack of evidence about differing results of cones and sleeves depending on length of follow-up.

The objective of the present paper was therefore to conduct a current comparative meta-analysis on the survival and clinical outcome of cones vs. sleeves. Here, in contrast to previous meta-analyses, short- and long-term (ST vs. LT) follow-up were to be differentiated.

Methods

Literature search strategy

The literature search was conducted systematically, following the internationally recognised Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) (Fig. 1). The literature databases PubMed, Ovid Medline, GoogleScholar and Cochrane Library were used. The following search terms were used in combination: (“total knee arthroplasty” OR “revision total knee arthroplasty”) AND (“cones” OR “sleeves”).

Fig. 1.

Fig. 1

PRISMA flow diagram of article selection. PRISMA—Preferred Reporting Items for Systematic Reviews and Meta-Analysis

Original papers published in English between 2010 and June 2021 were included after thorough screening of their relevance with regard to content. All publications with a follow-up of less than 1 year were excluded.

The data were extracted both from the text and from the respective tables and figures. For quality assurance, a second reviewer (M.H.) was consulted in the event of uncertainty and a cross-check was carried out.

For the evaluation of survival, the following parameters were selected: number of implant exchanges in total, implant exchanges due to aseptic loosening and septic reoperations with and without implant exchange e.g. DAIR (debridement, antibiotics and implant retention) in periprosthetic joint infections (PJI). Regarding clinical outcome the KSS (Knee Society Score) and KSFS (Knee Society functional Score) were assessed.

Revisions were defined as implant exchange, excluding PE-exchange only. All complications requiring revision but without prosthesis replacement counted as re-operations.

Descriptive data considered were the mean follow-up of the studies, the reasons for revision (index indication), the level of constraint (non-constrained, condylar constrained (CC) or hinged) as well as the patients’ age and sex.

Statistics

Prevalence represents the ratio of the number of patients with complications to the total of number of patients in that study. A pooled prevalence was used as a summary statistic using a random intercept logistic regression model. Accordingly, prevalence was used to enable a forest plot. The horizontal bars in the plots represent the range of confidence interval (CI). A 95% CI was used in the analysis. Analysis of heterogeneity of prevalence across studies was initially done using a Chi-square test. The degree of heterogeneity was also quantified using I2 values. The I2 statistic describes the percentage of variation across studies that is due to heterogeneity rather than chance. In this meta-analysis the heterogeneity variance tau2 was estimated based on the maximum likelihood estimate according to a random effects logistic regression model. Publication bias was investigated using Egger’s regression test [10]. Statistical analysis was performed using the statistical software R using the package meta [4, 32].

Results

The present meta-analysis included 43 publications with n = 3008 rTKA. Of these, 23 publications with 1911 cases were allocated to the sleeve group (SG) and 20 papers with 1097 cases the cone group (CG). In the subgroup analysis according to the respective follow-up period, a total of 27 studies resulted for the ST group and a total of 16 studies were assigned to the LT group. This meta-analysis takes into account the longest follow-up periods covered to date. (Table 1).

Table 1.

Summary of all studies main results

Author Year FU (months) FU (short term/ long term) Therapy patients knees Age BMI Male Female Hinged CC Unconstrained Septic index indication Aseptic index indication Revision total Aseptic loosening PJI (DAIR and implant exchange)
Abdelaziz 2019 126.5 Long Cone 25 25 65.0 13 12 25 0 0 0 25 13 10 3
Abdelaziz 2019 49.9 Short Cone 72 72 70.0 30.0 41 31 72 0 0 72 0 15 7 8
Bohl 2017 40.5 Short Cone 49 49 69.0 35.7 13 36 5 32 12 12 37 1 1 3
Burastero 2018 43.5 Short Cone 60 60 67.9 27.7 26 34 18 42 0 60 0 5 3 2
Chalmers 2021 24.0 Short Cone 163 163 67.0 33.0 75 88 52 106 5 46 117 6 2 16
De Martino 2015 72.0 Long Cone 18 18 73.0 29.0 6 12 12 6 0 13 5 2 0 2
Derome 2014 33.0 Short Cone 29 29 70.0 19 10 0 18 11 7 22 1 0 2
Erivan 2020 24.0 Short Cone 61 61 60.4 30 31 6 12 43 2 59 5 2 8
Girerd 2016 34.0 Short Cone 51 52 68.0 28.0 25 26 38 14 0 19 33 4 0 4
Hernandez 2021 91.2 Long Cone 59 62 70.0 34.1 21 38 5 41 16 23 39 18 9 11
Howard 2011 33.0 Short Cone 24 24 64.0 11 13 10 11 3 7 17 0 0 0
Jensen 2014 47.0 Short Cone 36 36 69.0 25 11 16 14 6 15 21 4 2 2
Kamath 2015 70.0 Long Cone 63 66 67.0 33.0 27 36 25 33 8 26 40 3 1 7
Lachiewicz 2013 39.0 Short Cone 27 27 64.6 35.9 13 14 3 22 2 13 14 2 1 1
Ohlmeier 2020 22.0 Short Cone 52 52 68.6 30.5 29 23 52 0 0 17 35 2 1 2
Panda 2019 83.0 Long Cone 59 59 69.7 28.7 18 41 0 26 33 2 0 2
Potter 2016 60.0 Long Cone 157 157 64.0 32.7 82 75 75 82 27 10 21
Rao 2013 36.0 Short Cone 26 26 72.0 15 11 26 0 0 7 19 1 0 2
Schmitz 2013 37.0 Short Cone 38 38 72.0 38 0 0 0 38 2 2 0
Villanueva-Martinez 2013 36.0 Short Cone 21 21 73.3 7 14 10 11 0 5 16 1 0 2
Agarwal 2018 95.7 Long Sleeve 103 104 74.7 54 49 0 55 49 31 73 21 7 5
Agarwal 2013 43.0 Short Sleeve 103 104 69.0 54 49 0 55 49 31 73 2 2 2
Alexander 2013 33.0 Short Sleeve 28 30 71.0 14 14 0 30 0 8 22 0 0 1
Algarni 2020 49.0 Short Sleeve 27 27 65.4 37.6 5 22 3 24 0 3 24 1 0 0
Barnett  2014 38.0 Short Sleeve 34 34 60.0 30.7 3 24 7 9 25 4 2 1
Bloch 2019 91.0 Long Sleeve 277 319 70.0 133 144 59 260 0 70 249 4 0 4
Bugler 2015 39.0 Short Sleeve 34 35 72.0 30.2 20 14 0 35 0 0 35 0 0 0
Chalmers 2016 38.0 Short Sleeve 227 227 66.0 34.0 10 166 51 84 143 8 3 12
Dalury 2016 57.6 Short Sleeve 40 40 73.0 32.0 19 21 6 34 0 6 34 1 1 0
Fedorka 2016 58.8 Short Sleeve 50 50 65.6 28 22 0 46 0 25 25 5 3 2
Gill 2020 65.0 Long Sleeve 31 31 8 18 5 0 0 1
Goettsche 2016 24.0 Short Sleeve 67 67 0 67 0 16 51 4 1 1
Graichen  2015 43.2 Short Sleeve 121 121 17 27 77 0 121 12 4 4
Gurel 2021 82.4 Long Sleeve 30 30 69.9 30.2 10 20 0 30 0 8 22 0 0 0
Huang  2014 29.0 Short Sleeve 79 83 63.5 33.0 29 50 10 73 0 20 63 6 3 6
Klim 2018 63.6 Long Sleeve 56 56 73.0 34.0 22 34 56 0 9 0 9
Klim 2020 75.6 Long Sleeve 93 93 68.0 30.0 39 54 0 93 0 52 41 17 0 15
Lai 2020 24.0 Short Sleeve 17 17 68.0 30.4 5 12 0 17 0 8 9 0 0 1
Martin-Hernandez 2016 71.5 Long Sleeve 134 134 75.0 29.8 52 82 0 134 0 12 122 2 0 2
Panesar 2021 91.0 Long Sleeve 99 99 69.7 46 53 99 0 0 32 67 18 2 11
Stefani 2017 37.0 Short Sleeve 47 47 71.0 0 47 0 17 30 1 0 1
Watters  2017 63.0 Long Sleeve 116 116 63.7 30.7 58 58 3 98 13 28 88 3 1 6
Wirries 2019 60.0 Long Sleeve 47 47 67.2 30.6 8 39 23 24 0 19 28 6 3 3

SG and CG were comparable with regard to age, sex distribution (male/female) and index indication for revision (septic/aseptic). (Tables 2 and 3).

Table 2.

Patient’s demographics were comparable and without significant differences between the therapy groups

Follow-up Therapy Age BMI Male (%) Female (%)
short cone 67.8 31.5 49.0 51.0
short sleeve 67.1 33.2 46.0 54.0
long cone 66.8 32.2 43.8 56.2
long sleeve 70.2 30.7 44.2 53.4

Table 3.

The results illustrating the indication for index revision using cone or sleeve, the degree of implant constraint, the rates of implant exchange for any reason, aseptic loosening, and the rates for operative intervention resulting from PJI

short term follow-up (< 5 years) Cones Sleeves p value
Aseptic index RTKA 61.2 (27.8–86.5) 80.8 (66.4–90.0) 0.23
Septic index RTKA 38.9 (13.5–72.2) 19.2 (10.0–33.6) 0.23
Hinged implant 68.2 (20.8–94.6) 1.8 (0.4–7.9) < 0.001
CC implant 20.1 (5.5–52.1) 93.9 (77.0–98.6) < 0.001
Unconstraint implant 2.0 (0.3–12.2) 0.3 (0.01–9.7) 0.37
Implant exchange for any reason 6.1 (3.9–9.3) 4.5 (2.8–7.1) 0.35
Implant exchange for aseptic loosening 4.3 (2.8–6.3) 2.8 (1.8–4.2) 0.16
PJI (with or without implant exchange) 7.0 (4.8–10.0) 3.4 (2.1–5.4) 0.02
Long-term follow-up (> 5 years) Cones Sleeves p value
Aseptic index RTKA 63.9 (39.4–82.8) 60.0 (32.7–82.2) 0.83
Septic index RTKA 36.1 (17.2–60.6) 40.0 (17.8–67.3) 0.83
Hinged implant 64.0 (9.9–96.6) 2.0 (0.02–65.1) 0.10
CC implant 27.7 (5.4–72.1) 89.6 (23.7–99.6) 0.11
Unconstraint implant 1.6 (0.1–31.8) 0.2 (0.2–0.2) 0.23
Implant exchange for any reason 14.4 (5.9–31.0) 5.5 (2.2–12.7) 0.12
Implant exchange for aseptic loosening 4.1 (0.8–19.7) 0.4 (0.1–3.0) 0.09
PJI (with or without implant exchange) 11.7 (8.2–16.3) 4.9 (2.6–9.1) 0.02

Numbers are given in percent with 95% confidence interval in brackets and p value for subgroup differences

With regard to the fixation in groups SG and CG, good short- and long-term prosthesis survival times were seen for both devices. No significant difference was found in relation to prosthesis survival (implant exchange for aseptic loosening, implant exchange for any reason) (Figs. 2 and 3). In contrast to that, periprosthetic joint infections (PJI) were twice frequent in the cone group compared to the sleeve group (Table 3, Fig. 4).

Fig. 2.

Fig. 2

Forest plot illustrating the rates for implant exchange for any reason for cone fixation vs. sleeve fixation with short-term (A) and long-term follow-up (B)

Fig. 3.

Fig. 3

Forest plot illustrating revision rates for aseptic loosening for cone fixation vs. sleeve fixation with short-term (A) and long-term follow-up (B)

Fig. 4.

Fig. 4

Forest plot illustrating revision for PJI with or without implant exchange for cone fixation vs. sleeve fixation with short-term (A) and long-term follow-up (B)

For both devices, significant improvements and good results were seen in all scores (KSS, KSFS) postoperatively (Figs. 5 and 6). A significant superiority of one device over the other could not be demonstrated. Hinged prothesis were used at a higher volume in the CG than in the SG whereas CC type protheses were used more frequently in the SG. (Table 3).

Fig. 5.

Fig. 5

Forest plot illustrating clinical outcome with KSS values in sleeve vs. cone fixation with short-term (A) and long-term follow-up (B)

Fig. 6.

Fig. 6

Forest plot illustrating clinical outcome with KSFS values in sleeve vs. cone fixation with short-term (A) and long-term follow-up (B)

We found publication bias with a bias equal to − 2.86 and p value < 0.01. All papers included were level III (retrospective cohort studies, case–control studies) and IV (case series) studies (Table 4).

Table 4.

Publication bias of the included studies

Author Year Therapy FU (months) Assembly of comparable groups Maintenance of comparable groups High loss to FU (>20%) Measurements: equal, reliable, valid Clear definition of interventions All important outcomes considered Adjustment for potential confounders Overall assessed quality
Abdelaziz 2019 Cone 126.5 No No Yes Yes Yes No scores No Fair
Abdelaziz 2019 Cone 49.9 No No Yes Yes Yes No scores No Fair
Bohl 2017 Cone 40.5 No No No Yes Yes Yes No Fair
Burastero 2018 Cone 43.5 No No No Yes Yes Yes No Fair
Chalmers 2021 Cone 24.0 No No No Yes Yes No scores No Fair
De Martino 2015 Cone 72.0 No No No Yes Yes Yes No Fair
Derome 2014 Cone 33.0 No No No Yes Yes Yes No Fair
Erivan 2020 Cone 24.0 No No No Yes Yes No scores No Fair
Girerd 2016 Cone 34.0 No No No Yes Yes Yes No Fair
Hernandez 2021 Cone 91.2 No No No Yes Yes No scores No Fair
Howard 2011 Cone 33.0 No No No Yes Yes Yes No Fair
Jensen 2014 Cone 47.0 No No No Yes Yes Yes No Fair
Kamath 2015 Cone 70.0 No No No Yes Yes Yes No Fair
Lachiewicz 2013 Cone 39.0 No No No Yes Yes Yes No Fair
Ohlmeier 2020 Cone 22.0 No No No Yes Yes No scores No Fair
Panda 2019 Cone 83.0 No No No Yes Yes Yes No Fair
Potter 2016 Cone 60.0 No No No Yes Yes Yes No Fair
Rao 2013 Cone 36.0 No No No Yes Yes Yes No Fair
Schmitz 2013 Cone 37.0 No No No Yes Yes Yes No Fair
Villanueva-Martinez 2013 Cone 36.0 No No No Yes Yes No scores No Fair
Agarwal 2018 Sleeve 95.7 No No No Yes Yes No scores No Fair
Agarwal 2013 Sleeve 43.0 No No No Yes Yes No scores No Fair
Alexander 2013 Sleeve 33.0 No No Yes Yes Yes Yes No Fair
Algarni 2020 Sleeve 49.0 No No No Yes Yes Yes No Fair
Barnett  2014 Sleeve 38.0 No No No Yes Yes Yes No Fair
Bloch 2019 Sleeve 91.0 No No No Yes Yes No scores No Fair
Bugler 2015 Sleeve 39.0 No No Yes Yes Yes No scores No Fair
Chalmers 2016 Sleeve 38.0 No No No Yes Yes No scores No Fair
Dalury 2016 Sleeve 57.6 No No No Yes Yes Yes No Fair
Fedorka 2016 Sleeve 58.8 No No No Yes Yes No scoreS No Fair
Gill 2020 Sleeve 65.0 No No No Yes Yes No scores No Fair
Goettsche 2016 Sleeve 24.0 No No No Yes Yes Yes No Fair
Graichen  2015 Sleeve 43.2 No No No Yes Yes Yes No Fair
Gurel 2021 Sleeve 82.4 No No No Yes Yes Yes No Fair
Huang  2014 Sleeve 29.0 No No No Yes Yes Yes No Fair
Klim 2020 Sleeve 63.6 No No No Yes Yes Yes No Fair
Klim 2018 Sleeve 75.6 No No No Yes Yes No scores No Fair
Lai 2020 Sleeve 24.0 No No No Yes Yes Yes No Fair
Martin-Hernandez 2016 Sleeve 71.5 No No No Yes Yes Yes No Fair
Panesar 2021 Sleeve 91.0 No No No Yes Yes No scores No Fair
Stefani 2017 Sleeve 37.0 No No No Yes Yes Yes No Fair
Watters  2017 Sleeve 63.0 No No No Yes Yes Yes No Fair
Wirries 2019 Sleeve 60.0 No No Yes Yes Yes Yes No Fair

Discussion

The most important finding of the present study was that no difference regarding overall prosthesis survival and clinical outcome could be determined neither in short-term nor in long-term follow-up between fixation using cones compared with sleeves in rTKA. The subgroup analyses for the endpoints “implant exchange for aseptic loosening” and “implant exchange for any reason” showed no statistically significant difference, again regardless of the follow-up period. This result has to be interpreted taking into account, that cones were used more frequently with higher constrained implants than sleeves. The endpoint “PJI” differed significantly between the groups presenting a higher infection rate in the cone group.

Available data are predominantly with short follow-up. As a result, previous meta-analyses are biased by a disproportionate evaluation of the short-term follow-up and therefore overestimation of reasons for early revisions. These differ between short- and long-term follow-up [2, 3]. Therefore, studies with short- and long-term follow-up were compared separately in the present meta-analysis.

Bone defects and loss of bone substance are frequently encountered in revision total knee arthroplasty and present a challenge to the surgeon, making the implant fixation complicated. However, an optimal implant fixation is essential for a good functional outcome and survival of the prosthesis. Within the context of rTKA implantations, the epiphysis is almost always damaged and cannot be used as the sole fixation zone. In contrast, the metaphysis is usually sufficiently retained and can be used to anchor the implant [12, 22]. Cones and sleeves are two options available for metaphyseal anchorage.

In knee revision arthroplasty, semi-constrained or hinge prostheses are usually used based on the state of ligaments. Numerous studies have shown that the level of constraint influences the survival time and clinical outcome of the prosthesis. Pure hinge knee prostheses and type 3 bone defects are associated with higher numbers of aseptic loosening and worse clinical outcome [1, 7, 26]. A metaphyseal implant fixation seems to reduce that effect because, despite the significant higher volume of hinged prothesis in the CG, our study shows comparable rates of aseptic loosening between SG and CG. Based on these findings, a metaphyseal fixation of hinged implants should therefore be considered.

Regarding the clinical outcome both devices demonstrated postoperatively significant improvements and good results in all scores (KSS, KSFS) without a significant superiority of one device over the other.

In addition to the aforementioned fixation with cones or sleeves, numerous other factors can influence the subsequent outcome. Levent et al. [20] demonstrated smoking, a large femoral canal anteroposterior diameter and right-sided TKA as significant risk factors for aseptic loosening in TKA. Jasper et al. [16] and Klasan et al. [19] showed younger age, higher knee joint activity and male gender as significant risk factors for repeat revision procedures.

Moreover, both tibial and femoral component can get loose, so that the endpoint loosening is influenced by both components’ fixation. Implant geometry, implantation errors or compromises (rotation, anterior overstuffing, reduction of posterior offset, mediolateral overhang) are more frequent on the femoral than on the tibial side. Therefore, it is to be expected that the femoral component influences the clinical outcome more than the tibial component.

The higher rate of PJI in the CG compared to the SG remains a source of concern and needs further investigation. This cannot be explained by an inclusion bias because the number of septic index operations did not differ between the groups. Apart from generally known reasons for higher rates of postoperative infections that were not systematically assessed by most studies (e.g. smoking, diabetes, malnutrition, obesity, ASA classification), a possible explanation for higher numbers of PJI can be found in the different material properties of cones and sleeves. Sleeves have a dense surface that is structured by surface finish or coating. In contrast, the structure of cones is formed by interconnecting pores, resulting in a large total surface area. Given a relation between foreign material surface area and the risk of late infection this may explain the trend toward more PJI-related revisions in the cone group.

The meta-analyses already published on the clinical and radiological outcome of cones and sleeves have failed to show any statistically significant superiority of one anchorage method over the other. However, they do show a good clinical and radiological outcome for both devices in short- and medium-term follow-up. [18, 27, 30, 34]

There are some limitations to our study. One is the heterogeneous data pool, as not all the papers included in the meta-analysis stated means and standard deviations. The considerable heterogeneity of the data may additionally result from the fact, that revision operations per se are heterogenous (e.g. indication, bone defect, soft tissue situation, number of previous operations). Another limitation is the number of patients. It is a decimal power smaller than in studies dealing with primary TKA, so that few outliers have a higher impact on the given standard deviations. This meta-analysis is further limited by a significant publication bias that could not be eliminated by additional literature data after a second search. Only in very few cases, the bone defects were classified consistently, which means that an indication bias cannot be ruled out. In addition, the surgical technique, implant anchorage (with regard to cementation and/or additional stem anchorage), level of constraint of the implanted prostheses and the definition of complications, re-operations and revisions are not uniform.

According to present knowledge, cones and sleeves have not been directly checked against each other in a randomised controlled trial yet. All results and conclusions of the present meta-analysis must be considered with respect to the quality of the individual studies.

Conclusion

In conclusion, both cones and sleeves represent a reliable fixation method for revision knee arthroplasty in the case of severe bone loss. Based on our results, we recommend taking an additional metaphyseal fixation of hinged implants into account. While there is no apparent superiority of one method over the other regarding the overall survival and clinical outcome, the higher rate of PJI after cone fixation remains a source of concern.

Funding

Open Access funding enabled and organized by Projekt DEAL. No funding or grants were obtained for this study.

Declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

Resulting from the study design (meta-analysis) an ethical approval is not applicable. All included studies declared compliance to ethical standards. Table 1 contains a summary of all includes studies. We ourself, did not do any work with human participants and laboratory animals to establish this meta-analysis.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Abdelaziz H, Jaramillo R, Gehrke T, Ohlmeier M, Citak M. Clinical survivorship of aseptic revision total knee arthroplasty using hinged knees and tantalum cones at minimum 10-year follow-up. J Arthroplasty. 2019;34:3018–3022. doi: 10.1016/j.arth.2019.06.057. [DOI] [PubMed] [Google Scholar]
  • 2.Agarwal S, Azam A, Morgan-Jones R. Metal metaphyseal sleeves in revision total knee replacement. Bone Joint J. 2013;95:1640–1644. doi: 10.1302/0301-620X.95B12.31190. [DOI] [PubMed] [Google Scholar]
  • 3.Agarwal S, Neogi DS, Morgan-Jones R. Metaphyseal sleeves in revision total knee arthroplasty: minimum seven-year follow-up study. Knee. 2018;25:1299–1307. doi: 10.1016/j.knee.2018.09.010. [DOI] [PubMed] [Google Scholar]
  • 4.Balduzzi S, Rücker G, Schwarzer G. How to perform a meta-analysis with R: a practical tutorial. Evid Based Ment Health. 2019;22:153–160. doi: 10.1136/ebmental-2019-300117. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Bernatz JT, Brooks AE, Squire MW, Illgen RI, II, Binkley NC, Anderson PA. Osteoporosis is common and undertreated prior to total joint arthroplasty. J Arthroplasty. 2019;34:1347–1353. doi: 10.1016/j.arth.2019.03.044. [DOI] [PubMed] [Google Scholar]
  • 6.Bloch BV, Shannak OA, Palan J, Phillips JR, James PJ. Metaphyseal sleeves in revision total knee arthroplasty provide reliable fixation and excellent medium to long-term implant survivorship. J Arthroplasty. 2020;35:495–499. doi: 10.1016/j.arth.2019.09.027. [DOI] [PubMed] [Google Scholar]
  • 7.Burastero G, Cavagnaro L, Chiarlone F, Alessio-Mazzola M, Carrega G, Felli L. The use of tantalum metaphyseal cones for the management of severe bone defects in septic knee revision. J Arthroplasty. 2018;33:3739–3745. doi: 10.1016/j.arth.2018.08.026. [DOI] [PubMed] [Google Scholar]
  • 8.Chang CB, Kim TK, Kang YG, Seong SC, Kang S-B. Prevalence of osteoporosis in female patients with advanced knee osteoarthritis undergoing total knee arthroplasty. J Kor Med Sci. 2014;29:1425–1431. doi: 10.3346/jkms.2014.29.10.1425. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.De Martino I, De Santis V, Sculco PK, D’Apolito R, Assini JB, Gasparini G. Tantalum cones provide durable mid-term fixation in revision TKA. Clin Orthop Relat Res. 2015;473:3176–3182. doi: 10.1007/s11999-015-4338-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Egger M, Smith GD, Schneider M, Minder C. Bias in meta-analysis detected by a simple, graphical test. BMJ. 1997;315:629–634. doi: 10.1136/bmj.315.7109.629. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Engh GA. Classification of bone defects femur and tibia. Knee arthroplasty handbook: Springer; 2006. pp. 116–132. [Google Scholar]
  • 12.Gill UN, Ahmed N, Noor SS, Memon IA, Memon ZA. Management of the bone loss by metaphyseal sleeves in primary and revision knee arthroplasty: clinical experience and outcome after forty three cases. Int Orthop. 2020 doi: 10.1007/s00264-020-04663-1. [DOI] [PubMed] [Google Scholar]
  • 13.Gurel R, Morgan S, Elbaz E, Ashlenazi I, Snir N, Kadar A, et al. Mid-term clinical and radiographic outcomes of porous-coated metaphyseal sleeves used in revision total knee arthroplasty. Knee Surg Rel Res. 2021;33:1–6. doi: 10.1186/s43019-020-00086-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Ha CW, Park YB. Underestimation and undertreatment of osteoporosis in patients awaiting primary total knee arthroplasty. Archiv Orthop Trauma Surg. 2020 doi: 10.1007/s00402-020-03462-y. [DOI] [PubMed] [Google Scholar]
  • 15.Hernandez NM, Hinton ZW, Wu CJ, Ryan SP, Bolognesi MP. Mid-term results of tibial cones : reasonable survivorship but increased failure in those with significant bone loss and prior infection. Bone Joint J. 2021;103-B:158–164. doi: 10.1302/0301-620X.103B6.BJJ-2020-1934.R1. [DOI] [PubMed] [Google Scholar]
  • 16.Jasper L, Jones C, Mollins J, Pohar S, Beaupre L. Risk factors for revision of total knee arthroplasty: a scoping review. BMC Musc Disord. 2016;17:1–9. doi: 10.1186/s12891-015-0856-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Kamath AF, Lewallen DG, Hanssen AD. Porous tantalum metaphyseal cones for severe tibial bone loss in revision knee arthroplasty: a five to nine-year follow-up. JBJS. 2015;97:216–223. doi: 10.2106/JBJS.N.00540. [DOI] [PubMed] [Google Scholar]
  • 18.Kim HJ, Lee O-S, Lee SH, Lee YS. Comparative analysis between cone and sleeve in managing severe bone defect during revision total knee arthroplasty: a systematic review and meta-analysis. J Knee Surg. 2018;31:677–685. doi: 10.1055/s-0037-1606564. [DOI] [PubMed] [Google Scholar]
  • 19.Klasan A, Magill P, Frampton C, Zhu M, Young SW. Factors predicting repeat revision and outcome after aseptic revision total knee arthroplasty: results from the New Zealand Joint Registry. Knee Surg Sports Traumatol Arthrosc. 2021;29:579–585. doi: 10.1007/s00167-020-05985-8. [DOI] [PubMed] [Google Scholar]
  • 20.Levent A, Suero EM, Gehrke T, Bakhtiari IG, Citak M. Risk factors for aseptic loosening in complex revision total knee arthroplasty using rotating hinge implants. Int Orthop. 2021;45:125–132. doi: 10.1007/s00264-020-04878-2. [DOI] [PubMed] [Google Scholar]
  • 21.Martin-Hernandez C, Floria-Arnal LJ, Muniesa-Herrero MP, Espallargas-Donate T, Blanco-Llorca JA, Guillen-Soriano M, et al. Mid-term results for metaphyseal sleeves in revision knee surgery. Knee Surg Sports Traumatol Arthrosc. 2017;25:3779–3785. doi: 10.1007/s00167-016-4298-4. [DOI] [PubMed] [Google Scholar]
  • 22.Morgan-Jones R, Oussedik S, Graichen H, Haddad F. Zonal fixation in revision total knee arthroplasty. Bone Joint J. 2015;97:147–149. doi: 10.1302/0301-620X.97B2.34144. [DOI] [PubMed] [Google Scholar]
  • 23.Panda I, Wakde O, Singh H, Rajgopal A (2018) Management of large bone defects around the knee using porous tantalum trabecular metal cones during complex primary and revision total knee arthroplasty. Paper presented at: Seminars in Arthroplasty 2018
  • 24.Panegrossi G, Ceretti M, Papalia M, Casella F, Favetti F, Falez F. Bone loss management in total knee revision surgery. Int Orthop. 2014;38:419–427. doi: 10.1007/s00264-013-2262-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Panesar K, Al-Mouazzen L, Nessa L, Jonas SC, Agarwal S, Morgan-Jones R. Revision total knee arthroplasty using an uncemented metaphyseal sleeve, rotating hinge prosthesis: a case series of 99 patients. J Arthroplasty. 2021;36:2121–2125. doi: 10.1016/j.arth.2020.12.047. [DOI] [PubMed] [Google Scholar]
  • 26.Potter GD, III, Abdel MP, Lewallen DG, Hanssen AD. Midterm results of porous tantalum femoral cones in revision total knee arthroplasty. JBJS. 2016;98:1286–1291. doi: 10.2106/JBJS.15.00874. [DOI] [PubMed] [Google Scholar]
  • 27.Roach RP, Clair AJ, Behery OA, Thakkar SC, Iorio R, Deshmukh AJ. Aseptic loosening of porous metaphyseal sleeves and tantalum cones in revision total knee arthroplasty: a systematic review. J Knee Surg. 2020 doi: 10.1055/s-0040-1701434. [DOI] [PubMed] [Google Scholar]
  • 28.Rosso F, Cottino U, Dettoni F, Bruzzone M, Bonasia DE, Rossi R. Revision total knee arthroplasty (TKA): mid-term outcomes and bone loss/quality evaluation and treatment. J Orthop Surg Res. 2019;14:280. doi: 10.1186/s13018-019-1328-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Russell LA. Osteoporosis and orthopedic surgery: effect of bone health on total joint arthroplasty outcome. Curr Rheumatol Reports. 2013;15:371. doi: 10.1007/s11926-013-0371-x. [DOI] [PubMed] [Google Scholar]
  • 30.Scott Kelly L., Abdel Matthew P., Hanssen Arlen D. Metaphyseal sleeves and cones in revision total knee arthroplasty. In: Bono JV, Scott RD, editors. Revision total knee arthroplasty. Cham: Springer International Publishing; 2018. pp. 131–141. [Google Scholar]
  • 31.Sculco PK, Abdel MP, Hanssen AD, Lewallen DG. The management of bone loss in revision total knee arthroplasty: rebuild, reinforce, and augment. Bone Joint J. 2016;98-B:120–124. doi: 10.1302/0301-620X.98B1.36345. [DOI] [PubMed] [Google Scholar]
  • 32.Team RC (2013) R: a language and environment for statistical computing
  • 33.Watters TS, Martin JR, Levy DL, Yang CC, Kim RH, Dennis DA. Porous-coated metaphyseal sleeves for severe femoral and tibial bone loss in revision TKA. J Arthroplasty. 2017;32:3468–3473. doi: 10.1016/j.arth.2017.06.025. [DOI] [PubMed] [Google Scholar]
  • 34.Zanirato A, Formica M, Cavagnaro L, Divano S, Burastero G, Felli L. Metaphyseal cones and sleeves in revision total knee arthroplasty: two sides of the same coin? Complications, clinical and radiological results-a systematic review of the literature. Musculoskelet Surg. 2019 doi: 10.1007/s12306-019-00598-y. [DOI] [PubMed] [Google Scholar]

Articles from Knee Surgery, Sports Traumatology, Arthroscopy are provided here courtesy of Springer

RESOURCES