Skip to main content
Journal of Vascular Surgery Cases, Innovations and Techniques logoLink to Journal of Vascular Surgery Cases, Innovations and Techniques
. 2022 Jun 14;8(3):356–357. doi: 10.1016/j.jvscit.2022.04.016

Pararenal aortic aneurysm repair using a physician-modified stent-graft with inner branches

Tsuyoshi Shibata 1, Yutaka Iba 1, Tomohiro Nakajima 1, Itaru Hosaka 1, Nobuyoshi Kawaharada 1
PMCID: PMC9309584  PMID: 35898570

A 90-year-old man had presented with a 55-mm pararenal aortic aneurysm (expansion rate, 6 mm/mo). Gutter endoleaks have limited the use of parallel grafts. Physician-modified fenestration is one of the effective techniques in emergency cases.1 However, its use has posed concerns regarding the development of endoleaks from fenestration sites.2 In addition, company manufactured devices have a time delay. Hence, we assembled inner branched stent grafts to reduce the possibility of a type IIIc endoleak. Sapporo Medical University approved the present study (approval no. 20-005). The patient provided written informed consent for the report of his case details and imaging studies.

A sterile, 58-mm long, 26/26-mm Zenith alpha abdominal cuff (Cook Medical, Inc, Bloomington, IN) was modified. Based on the measurements from the computed tomography scan and the three-dimensional printed aortic model, an 8 × 8-mm fenestration for the superior mesenteric artery and two 6 × 6-mm fenestrations for the bilateral renal arteries were created at predetermined locations of the Dacron fabric via electrocautery. Subsequently, two inner branches were created by cutting the 6 × 6 × 100-mm Viabahn (W.L. Gore & Associates, Flagstaff, AZ) into 15-mm segments. Each inner branch was attached to the 6 × 6-mm fenestrations for the bilateral renal arteries using running Proline sutures inside the stent graft (A/Cover). Finally, the modified stent grafts were manually resheathed.

The stent graft system was inserted into the aorta via the left femoral artery. The fenestrations were confirmed to be in the same positions as the visceral arteries. Device deployment was fully unsheathed. A 7-mm Viabahn balloon expandable endoprosthesis (W.L. Gore & Associates) was deployed via the left brachial artery from the inner branch to the right renal artery. A similar procedure was performed for the left renal artery. Finally, the main body of the graft, an Excluder aortic stent graft (W.L. Gore & Associates), was deployed inside the modified stent grafts. Completion angiography revealed patent visceral arteries and no endoleaks (B). The patient recovered uneventfully.

To the best of our knowledge, three recent studies have reported on physician-modified inner branched stent grafts using an actual thoracic stent graft.3, 4, 5 This technique allows surgeons to achieve a longer landing zone and possibly a better seal and should be considered for urgent and emergency cases. To the best of our knowledge, our case is the first in which an aortic-modified cuff was used, which was advantageous in terms of costs and the minimum necessary extension to the sealing zone.

graphic file with name gr1.jpg

Footnotes

Author conflict of interest: none.

The editors and reviewers of this article have no relevant financial relationships to disclose per the Journal policy that requires reviewers to decline review of any manuscript for which they may have a conflict of interest.

Appendix

Cover image.

Cover image

References

  • 1.Shibata T., Shingaki M., Morishita K. Surgeon-modified endovascular grafts for treating ruptured arch aneurysm due to proximal endograft failure after thoracic endovascular aneurysm repair combined with total debranching. Gen Thorac Cardiovasc Surg. 2021;69:118–121. doi: 10.1007/s11748-020-01414-2. [DOI] [PubMed] [Google Scholar]
  • 2.Jussi M., Emanuel R., Akhilesh J., Bernardo C., Thanila A., Keouna P., et al. Outcomes of target vessel endoleaks after fenestrated-branched endovascular aortic repair. J Vasc Surg. 2020;72:445–455. doi: 10.1016/j.jvs.2019.09.055. [DOI] [PubMed] [Google Scholar]
  • 3.D’Oria M., Mirza A.K., Tenorio E.R., Karkkainen J.M., DeMartino R.R., Oderich G.S. Physician modified endograft with double inner branches for urgent repair of supraceliac para-anastomotic pseudoaneurysm. J Endovasc Ther. 2020;27:124–129. doi: 10.1177/1526602819890108. [DOI] [PubMed] [Google Scholar]
  • 4.Torrealba J., Panuccio G., Kolbel T., Gandet T., Heidemann F., Rohlffs F. Physician-modified endograft with inner branches for the treatment of complex aortic urgencies. https://doi.org/10.1177/15266028211061275 [e-pub ahead of print]. J Endovasc Ther. [DOI] [PubMed]
  • 5.Shibata T., Kawaharada N., Kiyomitsu Y., Syuichi N. Physician-modified stent graft with inner branches for treating ruptured thoracoabdominal aortic aneurysm. Eur J Cardiothorac Surg. 2022;61:952–954. doi: 10.1093/ejcts/ezab543. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Vascular Surgery Cases, Innovations and Techniques are provided here courtesy of Elsevier

RESOURCES