Hyperpolarized 129Xe imaging of the brain: Achievements and future challenges

Yurii Shepelytskyi 1,2, Vira Grynko 2,3, Madhwesha R Rao 4, Tao Li 1, Martina Agostino 1, Jim M Wild 4,5, Mitchell S Albert 1,2,6,
PMCID: PMC9314594  PMID: 35253919

Abstract

Hyperpolarized (HP) xenon‐129 (129Xe) brain MRI is a promising imaging modality currently under extensive development. HP 129Xe is nontoxic, capable of dissolving in pulmonary blood, and is extremely sensitive to the local environment. After dissolution in the pulmonary blood, HP 129Xe travels with the blood flow to the brain and can be used for functional imaging such as perfusion imaging, hemodynamic response detection, and blood–brain barrier permeability assessment. HP 129Xe MRI imaging of the brain has been performed in animals, healthy human subjects, and in patients with Alzheimer's disease and stroke. In this review, the overall progress in the field of HP 129Xe brain imaging is discussed, along with various imaging approaches and pulse sequences used to optimize HP 129Xe brain MRI. In addition, current challenges and limitations of HP 129Xe brain imaging are discussed, as well as possible methods for their mitigation. Finally, potential pathways for further development are also discussed. HP 129Xe MRI of the brain has the potential to become a valuable novel perfusion imaging technique and has the potential to be used in the clinical setting in the future.

Keywords: brain, HP 129Xe brain imaging, hyperpolarized xenon‐129, magnetic resonance imaging/spectroscopy, perfusion imaging

1. INTRODUCTION

There are multiple brain imaging modalities currently available for clinical diagnostic use, including ultrasound, CT, single‐photon emission CT, PET, and MRI. MRI is a noninvasive technique that uses no ionizing radiation and can produce images with high spatial resolution and contrast‐to‐noise ratio. Despite numerous developments and discoveries since MRI was invented in 1973, 1 the main limitation of MRI remains the same: low sensitivity. 2 , 3 The MRI signal originates from the net magnetization of the sample due to the small population difference between the Zeeman energy levels of nuclei with typically a one‐half spin number. Conventional MRI uses the NMR signal from water protons (1H); numerous contrast agents are being developed to enhance the 1H MRI signal and provide the ability to localize the area of interest. 3 , 4 , 5 Many of these agents, such as gadolinium chelated contrast agents, are focused on decreasing the spin–lattice (T1) and effective spin–spin (T2 *) relaxation of 1H nuclei, which increases the MR contrast in T1‐weighted and T2 *‐weighted images. Despite the wide use of 1H contrast agents, this approach is limited due to the presence of the background signal from surrounding tissues, which limits any increase in contrast‐to‐noise ratio. Additionally, there are a variety of techniques, such as BOLD functional MRI, arterial spin labeling (ASL), and MRA, which require multiple image acquisitions and complicated image postprocessing procedures for accurate data interpretation. 6 , 7

Another fundamentally different method for enhancing the MRI signal involves creating a hyperpolarized (HP) nuclear state. 8 The HP state is a metastable state that can achieve up to a 105 times larger spin population excess, compared with the thermal equilibrium state. Traditional HP MRI techniques work with non‐proton MRI‐sensitive nuclei such as xenon‐129 (129Xe), helium‐3 (3He), and carbon‐13 (13C). 9 , 10 , 11 , 12 The signal from HP nuclei can be enhanced by up to 105 times, and MRI images of HP agents can be acquired with almost no background signal. Due to this signal boost, imaging of low‐concentration HP agents becomes possible. Currently, the main application of HP gas MRI is for lung imaging of healthy individuals and individuals with lung disorders. 2 , 12 HP 129Xe undergoes gas exchange in the lungs, 13 , 14 , 15 easily dissolves in pulmonary blood, 12 , 15 , 16 and then distributes throughout the body. Because HP 129Xe has a sufficiently long T1 relaxation time in the blood (T1 in a range of 3.4–7.8 s), 17 , 18 , 19 , 20 HP 129Xe MRI has the potential to produce functional images of highly perfused organs. 8 , 21 , 22 , 23 Although this idea was originally formulated at the end of the 20th century, 8 HP 129Xe imaging in the brain is only recently under extensive development, and HP 129Xe imaging of the kidneys has just been demonstrated about a year ago.

Despite the intensive development of dissolved‐phase HP 129Xe imaging in brain tissues over the past decade, there have been no dedicated comprehensive reviews for the advances in this area. This review article aims to highlight the current progress and development in the field of HP 129Xe brain imaging, as well as discuss the technical challenges associated with this technology. The imaging approaches currently used are also reviewed and discussed. It is anticipated that insights into the challenges and opportunities of this field can be highlighted and aid in further advancements in the methodology and technique development of this technology with subsequent clinical translation.

2. HP 129XE SPECTROSCOPY AND CSI OF THE BRAIN

Historically, xenon was used in medicine as an anesthetic due to its ability to dissolve in brain tissue. 24 , 25 , 26 , 27 , 28 In addition to its anesthetic applications, xenon was widely used for cerebral blood flow evaluation using Xenon CT (Xe‐CT). 29 , 30 , 31 Implementation of the hyperpolarization process for boosting the 129Xe MRI signal established an entirely new field of brain imaging and investigations with HP 129Xe. 8 One of the properties that is most important for brain research with HP 129Xe dissolved in various brain tissues is its chemical shift. The first in vivo 129Xe brain MR spectrum was obtained by Swanson et al. in 1997 from the rat brain. 32 A single blood‐tissue resonance peak was identified and used to produce an HP 129Xe 2D CSI of the rat brain (Figure 1). Later that year, Mugler et al. performed the first 129Xe MRS study of the human head. 33 In that study, volunteers inhaled between 300 and 500 ml of HP 129Xe in one breath; 15 consecutive spectra of the head were subsequently acquired during and after a 15‐s breath‐hold period. The spectra showed one peak from the gas phase and one peak from the dissolved phase that was shifted 196 ppm from the gas peak. The dissolved phase peak appeared at the end of the inhalation period at approximately 5 s and disappeared 40 s after the start of the breath‐hold. The main limitation for the acquisition of human brain images at that time was the extremely low polarization of HP 129Xe achievable, approximately 2%. 33

FIGURE 1.

MRM-29200-FIG-0001-c

(A) Hyperpolarized (HP) xenon‐129 (129Xe) axial 2D CSI of the rat brain in grayscale. (B) Color‐coded overlay onto a high‐resolution proton image. The SNR of the HP 129Xe image was equal to 20. (C) High‐resolution proton spin‐echo MRI image used for brain localization. Images are reprinted with permission from the publisher 32

Obtaining a spectral peak from HP 129Xe dissolved in the brain allowed the conduction of a dynamic study of the distribution of xenon in the rat brain using 1D and 2D CSI. 34 Swanson et al. detected a signal from the rat brain using 1D CSI with a low flip angle and investigated the time evolution of this signal. The polarization of 129Xe in this study, however, was still low (5%–8%); the study was performed primarily to observe the signal evolution within the body of the rat.

Duhamel et al. used a different approach for observation of the HP 129Xe solubility in the rat brain at 2.35 T. 35 They injected naturally abundant HP 129Xe dissolved in a lipid emulsion, into the carotid artery, and observed two peaks at 199 and 194 ppm. These peaks were identified as 129Xe dissolved in the tissue and the intravascular compartment, respectively. 35 It was clear that the signal intensity was too small to observe peaks from all brain tissues. Therefore, a final conclusion regarding the specific resonance frequencies of all 129Xe compartments, rather than merely just their frequency ranges, was not possible. Wakai et al. was able to observe all the individual 129Xe resonances by averaging 60 acquisitions during continuous breathing of an enriched HP 129Xe gas mixture. They observed five 129Xe spectral peaks in the rat brain that ranged between 189 and 210 ppm. 36 Following this work, Nakamura et al. assigned a spectral peak at 195 ppm to the brain tissue: one at 210 ppm to HP 129Xe dissolved in the blood, and one at 189 ppm to non‐brain tissues (assumed to be muscle). 37 Their conclusions on the resonance frequencies of HP 129Xe in the brain were aided by using a rat model involving an arterial ligation. Kershaw et al. found the peaks at 195 and 192 ppm originated from gray and white matter, respectively. 38 Additionally, the peaks at 189 and 198 ppm were interpreted as signals from the jaw muscle and fat tissue. 38

The ability to distinguish the HP 129Xe peaks in the human brain has become possible with the availability of increased xenon polarization. 1D CSI spectra of the human brain with 129Xe polarized up to 8%, and a 2D‐CSI image using 14% polarized 129Xe dissolved in brain tissue superimposed on a 1H image, were obtained by Kilian et al. in 2002. 39 Two additional peaks at 198 and 195 ppm were observed on the 1D‐CSI spectra beside an already identified peak at 196 ppm from Mugler et al.'s results. 33 The 2D‐CSI measurements (Figure 2) revealed at least three additional peaks at 185, 193, and 200 ppm after spectral averaging, in addition to a peak previously observed at 197 ppm. 39 Following their initial study, Kilian et al. performed an additional 2D CSI using isotopically enriched HP 129Xe to determine the origin of 129Xe peaks in the tissue compartment. 40 The authors observed two dominant peaks from HP 129Xe in the brain region at 196 and 193 ppm, and two additional minor peaks from HP 129Xe in non‐brain tissues located below the brain at 190 and 201 ppm. The origins of the dominant peaks at 196 and 193 ppm were proposed to come from the gray and white matter, respectively.

FIGURE 2.

MRM-29200-FIG-0002-b

Two‐dimensional‐CSI spectra of HP 129Xe dissolved in brain tissue superimposed onto a 1H image. The image was reprinted with permission from the publisher 39

After more than 10 years since these animal studies, during which time the polarization of HP 129Xe was significantly improved, Rao et al. demonstrated the first HP 129Xe human brain CSI, with detailed spectroscopy, at 1.5 T in 2015, where a red blood cell (RBC) 129Xe peak was observed for the first time. 41 During the following year, they published a detailed study on the assignments of all the observed HP 129Xe brain peaks, which was based on high‐resolution spectroscopy and CSI measurements. 42 In the latter study, 3 healthy volunteers each inhaled 1 L of HP 129Xe, followed by a 20‐s breath‐hold, during which the acquisition was performed. CSI was conducted to assign the HP 129Xe peaks obtained from the spectroscopy results (Figure 3D) to various tissue compartments within the head. An HP 129Xe peak at 188 ppm (Figure 3A) was assigned to HP 129Xe dissolved in soft muscular tissue in the cheek and 129Xe in the midbrain. The peak at 192 ppm (Figure 3B) corresponded to HP 129Xe dissolved in white matter; the peak at 196 ppm (Figure 3C) corresponded to HP 129Xe dissolved in gray matter; and the peak at 200 ppm (Figure 3E) was assigned to HP 129Xe dissolved in the plasma, fat tissue outside of the brain, and CSF. The final peak observed at 217 ppm (Figure 3F) showed high signal intensity at the location of the Circle of Willis and corresponded to HP 129Xe dissolved in RBCs. The results of this study mostly agreed with the results obtained from previous studies using animal models. One difference, however, was in frequency of the RBC peak, which was reported at 210 ppm in rats, and at 217 ppm in humans.

FIGURE 3.

MRM-29200-FIG-0003-c

Two‐dimensional CSI of spatially resolved peaks from 129Xe in the human head superimposed onto 1H images. (A) Tissue in the cheek muscle and the midbrain/brainstem. (B) White matter and cartilaginous soft tissue. (C) Gray matter. (D) Spectra of the whole brain with a bandwidth of 136.0 9 ppm and a spectral resolution of 0.33 ppm. (E) Body interstitial fluid/plasma, fat tissue outside of the brain, and CSF. (F) Red blood cells (RBCs). The figure was reprinted with permission from the publisher 42

A spectroscopic study by Li et al 43 evaluated the influence of the pulmonary oxygen concentration on the HP 129Xe brain signal. The authors proposed an “apparent relaxation time” as a parameter that reflects the dependence of the HP 129Xe brain signal on the pulmonary oxygen concentration. The optimal pulmonary oxygen concentration range for maximizing the SNR of 129Xe brain images was reported to be between 25% and 35%, which agrees with previous experimental and theoretical findings. 44

Antonacci et al. raised an important question of the effect of the macroscopic susceptibility gradients on the dissolved‐phase HP 129Xe chemical shift. 45 They pointed out the lack of consistency of the HP 129Xe chemical shift dissolved in the same tissues in the different studies. To solve this problem, they proposed a novel method for mitigation of the effects of the macroscopic susceptibility gradients by referencing the dissolved 129Xe resonances with the chemical shifts of the nearby 1H water protons. This allows the comparison of the chemical shift values from different studies and aids in the correct identification of the origin of the peaks.

3. RELAXATION TIME MEASUREMENTS

Other important characteristics of HP 129Xe are the spin–lattice or longitudinal (T1) and spin–spin or transverse (T2) relaxation times. The image quality depends on the TR and TE, which are set based on the longitudinal and transverse relaxation time values, respectively. The first measurement of the longitudinal relaxation was performed ex vivo in rat brain tissue at 9.4 T by Wilson et al. 46 T1 relaxation times were determined at varying oxygenation levels and were reported to be 18 ± 1 s in the oxygenated state and 22 ± 2 s in the deoxygenated state. Following this ex vivo study, Duhamel et al. measured the longitudinal and transverse magnetization in vivo at 2.35 T. 47 The T1 of HP 129Xe dissolved in brain tissue was calculated to be 14 ± 1 s, and the T2 * was measured to be 8.0 ± 1.2 ms. 47 Spin–lattice relaxation for the white matter was also derived from human brain dynamic spectra by Kilian et al. to be 8 s. 48 The next measurement of T1 relaxation was performed in vivo in the rat brain by Wakai et al, who proposed a method for measuring the longitudinal relaxation without the need for an estimation of the flip angle. 49 Using this approach, the longitudinal relaxation time of 129Xe in the rat brain was found to be 26 ± 4 s. Due to the large discrepancies reported for T1 in these studies, in 2008, Zhou et al. reinvestigated the longitudinal relaxation time of 129Xe dissolved in the rat brain by developing a mathematical description of the HP 129Xe wash‐out process from the brain. 50 The authors determined the longitudinal relaxation time of 129Xe dissolved in the rat brain using a two‐pulse method (T1 = 15.3 ± 1.2 s) and a multipulse protocol (T1 = 16.2 ± 0.9 s). 50

The effective spin–spin relaxation (T2 *) for the 129Xe gray‐matter peak in the rat brain was estimated primarily from the linewidth at half‐height of the peak by Mazzanti et al. to be 5.42 ± 0.3 ms at 4.7 T (observed at 194.7 ppm), 51 and by Rao et al. in the human brain at 1.5 T to be 8.8 ms. 42

In summary, there is a lack of consistency among the measured relaxation values (Table 1). Furthermore, there were no T1 measurements performed for HP 129Xe dissolved in the gray matter in humans. An accurate assessment of HP 129Xe T1 in the gray matter at different magnetic field strengths is vital for the practical implementation of HP 129Xe brain imaging. Indeed, one of the main potential applications of HP 129Xe brain imaging is rapid quantification of cerebral perfusion, as well as blood–brain barrier permeability. According to multiple mathematical models that were developed, 48 , 52 , 53 , 54 however, the HP 129Xe signal dynamics depend equally on both tissue perfusion and the T1 relaxation time. Therefore, until T1 is quantified with high accuracy in healthy individuals as well as in patients with neurodegenerative or cerebrovascular diseases, brain perfusion imaging using HP 129Xe MRI will remain mostly qualitative.

TABLE 1.

Chemical shift, longitudinal, and transverse relaxation of HP 129Xe dissolved in the different brain tissues

Chemical shift relative to gas peak (ppm) T1 (s) T2 * (ms)
Rat brain studies Tissue

194.5 (2.0 T) 32

199 (2.35 T) 35

195 (4.7 T) 37

198 (4.7 T) 38

14.0 ± 1.0 (2.35 T) 47

3.6 ± 2.1 (2.35 T) 55

26 ± 4 (4.7 T) 49

15.3–16.2 (4.7 T) 50

8.0 ± 1.2 (2.35 T) 47

5.42 ± 0.3 (4.7 T) 51

Muscle

189 (4.7 T) 37

189 (4.7 T) 38

Blood 210 (4.7 T) 37
Gray matter 195 (4.7 T) 38
White matter 192 (4.7 T) 38
Arterial blood

18 ± 1 (oxygenated 9.4 T) 46

22 ± 2 (deoxygenated 9.4 T) 46

13.7 ± 1.6 (4.7 T) 17

13–16 (1.5 T) 19

Human brain studies Gray matter

196.5 (2.94 T) 48

196 (3 T) 40

196 (1.5 T) 42

8.8 (1.5 T) 42
White matter

193 (3 T) 40

192 (1.5 T) 42

8 (2.94 T) 48
RBCs

222 (1.5 T) 19

224 (4.7 T) 17

217 (1.5 T) 42

Blood plasma

197 (1.5 T) 19

198 (4.7 T) 17

200 (1.5 T) 42

Muscle tissue 188 (1.5 T) 42

Surprisingly, there were no T2 * measurements conducted for HP 129Xe dissolved in human brain white matter, cerebral blood plasma and RBCs, and soft muscle tissue. These measurements are essential for further HP 129Xe brain imaging pulse‐sequence development because they will allow proper optimization of the imaging TEs.

4. STRUCTURAL BRAIN IMAGING WITH HP 129XE

At the end of the 20th century, the initial discovery of boosting the 129Xe signal with hyperpolarization for structural imaging of the brain using HP 129Xe was extremely promising. 8 After signal‐intensity measurements at the beginning of the 21st century, however, it became clear that the polarization of HP 129Xe needed to be significantly higher than what was possible at that time. Swanson et al. measured an SNR of 20 in the rat brain with 3.1 × 3.1 × 10 mm3 voxels. 32 Due to the limited concentration of HP 129Xe dissolved in brain tissue, CSI was used as the main imaging approach for HP 129Xe brain studies. To address the limitation of relying on CSI for HP 129Xe brain imaging, Nouls et al. developed a fast 3D radial gradient‐echo (GRE) acquisition approach for HP 129Xe brain imaging. 56 They acquired high‐resolution 3D images of the HP 129Xe distribution in the rat brain with an isotropic 32 × 32 × 32 matrix and a voxel size of 3.65 × 3.65 × 3.65 mm3 (Figure 4).

FIGURE 4.

MRM-29200-FIG-0004-c

Three‐dimensional HP 129Xe MR images of rat brains. The dissolved HP 129Xe image (color) is overlaid onto a 1H anatomical image (grayscale). The 129Xe signal largely matches the brain tissue. The 129Xe signal was notably intense in the olfactory bulb and midbrain regions and was largely absent from the cerebellum. The images are reprinted with permission from the publisher 56

Recently, Friedlander et al. demonstrated spectrally resolved HP 129Xe imaging of the rat brain using iterative decomposition with echo asymmetry and least‐square estimation (IDEAL) using a spiral readout at 3.0 T. 57 , 58 Using this approach, images of 129Xe dissolved in brain tissue and RBCs were acquired with an SNR of 31 ± 4 and 16 ± 2, respectively, and a resolution of 0.5 × 0.5 cm2. 57 Using time‐resolved dynamic spiral IDEAL imaging, Friedlander et al. was able to perform, for the first time, 129Xe local blood–brain barrier (BBB) permeability assessment in hypercapnic and normocapnic rats during continuous breathing of HP 129Xe. 58 Successful IDEAL decomposition of the dissolved‐phase HP 129Xe signal will likely be of interest for human brain imaging in future studies.

The first HP 129Xe structural human‐brain image was acquired at 1.5 T by Rao et al. in 2015, once the process of polarization was improved. 41 The image was acquired in an axial projection using a 2D spoiled GRE sequence with a voxel size of 6.88 × 6.88 × 50 mm3. The HP 129Xe brain image correlated with the corresponding anatomical 1H MR image.

Rao et al. recently went on to perform 3D isotropic spectroscopic imaging of HP 129Xe in the human brain. 59 The acquisition matrix was 10 × 10 × 10, yielding a slice thickness of 2 cm and an acquisition voxel size of 8 cm3. The acquired images were interpolated to a voxel size of 0.24 cm3 and a slice thickness of 0.625 cm. This novel approach for HP 129Xe spectroscopic imaging cold be potentially implemented in further brain oxygenation mapping.

The most recent contribution to HP 129Xe structural brain imaging was achieved by Grynko et al. by acquiring 3D multislice images of the human brain at 3 T (Figure 5). 60 Five slices of the human brain were imaged with a slice thickness of 20 mm and an acquisition voxel volume of 1.22 cm3, which is the smallest acquisition voxel volume of HP 129Xe human‐brain imaging currently achieved. The highest SNR was reported to be 18.76 ± 4.95 from the inhalation of 1 L of HP 129Xe polarized to about 50%.

FIGURE 5.

MRM-29200-FIG-0005-c

The first HP 129Xe 3D gradient‐echo (GRE) multislice image of the human brain. (A) 1H T2‐weighted anatomical axial turbo spin‐echo (TSE) images of a representative healthy volunteer. B‐D, Axial anatomical images of gray matter (B), white matter (C), and CSF (D) segmented using high‐resolution TSE 1H T2‐weighted images of a representative healthy volunteer. (E) Three‐dimensional GRE HP 129Xe axial brain slices acquired 10 s into the breath‐hold. (F) Thresholded HP 129Xe axial brain slice images superimposed on top of the corresponding 1H anatomical images from (A). It can be seen that the HP 129Xe signal corresponds well to the gray‐matter distribution in the brain. In addition, a partial correlation has been observed between the white‐matter images and the HP 129Xe images. The image was reprinted with permission from the publisher 60

These two recent studies provided a significant step forward, demonstrating the ability of HP 129Xe brain imaging to produce multiple slices, which will allow the accurate and precise anatomical localization of HP 129Xe dissolved in the human brain.

Despite these recent achievements in structural HP 129Xe imaging of the brain, the imaging voxel size remains approximately two orders in magnitude larger compared to that of conventional anatomical proton MRI (1H voxel size ∼ 10 mm3). The low concentration of 129Xe dissolved in the brain tends to significantly restrict the spatial resolution achievable for imaging. Based on previous uptake models 48 , 61 (detailed in the next section) and the Oswald solubility of 129Xe in pulmonary blood, 62 it is estimated that only about 1%–2% of the amount of inhaled 129Xe actually dissolves in the brain tissues. In spite of the low concentrations of 129Xe achievable in brain tissue, however, it will be seen in the following sections that there is great value for using HP 129Xe for perfusion and other functional studies of the brain.

5. HP 129XE UPTAKE MODELS

The dynamics of HP 129Xe uptake in the brain and its wash‐out are complex and require careful consideration of multiple factors. Therefore, an accurate mathematical model of HP 129Xe signal dynamics is required for appropriate experimental design.

The first attempt to model 129Xe uptake in brain tissues was performed by Peled et al. in 1996. 61 They proposed an uptake model that calculated the time‐dependent build‐up of polarized 129Xe in brain‐tissue compartments based on estimates of the 129Xe relaxation times in tissue, perfusion rates, arterial transmit time, and partition coefficients. The authors considered continuous breathing a dose of 70% enriched HP 129Xe mixed with 30% O2. The model predicted a maximum concentration of 129Xe of 27 uM for gray matter and 8 uM for white matter and myelin, reached at 60 s after inhalation. Martin et al. extended Peled's model by accounting for different breathing protocols and estimated the 129Xe concentration in the brain for a wide range of T1 values for the gas and tissue phases. 63 The key 129Xe T1 parameters used in the model were as follows: 1000 s in the polarizer cell, 6 s in the arterial blood and in the tissue, and 12 s in the mouth and lungs. In this model, the polarization of 129Xe was assumed to be equal to 100%. The lung to brain transit time was estimated to be 5 s. Three different breathing protocols were investigated: continuous breathing, hyperventilation followed by a breath‐hold, and hyperventilation followed by continuous breathing. The maximum HP 129Xe concentration in the gray matter was calculated to be 0.09 mM at 15 s following inhalation for both the hyperventilation with a breath‐hold, and the continuous breathing methods. The 129Xe concentration in the brain was predicted to be 0.04 mM at 50 s after inhalation for the continuous breathing protocol. 63 Although both of these initially proposed models did not account for such factors as chemical shift effects, HP 129Xe passage through biological membranes, and 129Xe exchange between brain compartments, they provided useful information for conducting further research. Kilian et al. went on to propose an improved 129Xe uptake model based on spectroscopic MR measurements at various time points after 129Xe inhalation. 48 The model was in agreement with the quantitative 129Xe spectroscopy experimental data. Kilian also reported that the longitudinal 129Xe relaxation in gray matter was slower than that in white matter (T1g > T1w). 48 This model, however, also had drawbacks. It did not account for the gradient of 129Xe solubility in gray matter and white matter or the exchange of 129Xe between the tissues and the bloodstream. Following this, Shepelytskyi et al. expanded on Kilian's model for the case of HP 129Xe dynamic imaging, and implemented it for perfusion imaging of the human brain. 52

An additional mathematical description of the HP 129Xe wash‐out process was developed by Zhou et al. for better estimation of the longitudinal relaxation time of 129Xe dissolved in the brain tissues. 50 Following this, Kimura et al. developed an HP 129Xe uptake model and investigated 129Xe uptake using chemical shift saturation recovery spectroscopy in the mouse brain. 64 This model allowed improvements to the estimation accuracy of the HP 129Xe longitudinal relaxation time in the mouse brain. Imai et al. also developed a theoretical model of the 129Xe signal dynamics in the mouse brain, and suggested a method for its quantitative measurement under continuous 129Xe ventilation conditions. 65

The most recent kinetic model of HP 129Xe uptake was developed by Rao et al. for the determination of the transfer rate of inhaled xenon from cerebral blood to the gray matter in the human brain. 53 , 66 Using the time course of the HP 129Xe spectroscopic signal, the authors introduced a tracer kinetic model that explains the exchange of 129Xe between these compartments. 53 In this model, the transient ratio of the HP 129Xe concentration from gray matter to the blood was calculated from single‐voxel MRS spectra. The slope of the transient ratio over time was proposed as a physiological marker of BBB permeability. The main advantage of Rao et al.'s model compared with other uptake models is that it considers the forward exchange of 129Xe between the cerebral blood and gray‐matter tissue. However, it only considers the forward transfer of HP 129Xe, and neglects the gradient of the HP 129Xe concentration in the gray matter.

Further development of more complex mathematical models is required to accurately describe the concentration of HP 129Xe in the brain, and consequently, its signal dynamics. Future models must also make a special effort to describe the HP 129Xe diffusion processes in brain tissues. They must consider both forward and retrograde transfer of HP 129Xe as well as properly describe the HP 129Xe concentration gradient within the brain tissues. The diffusion coefficients of HP 129Xe in the cerebral blood and brain tissues currently remains unknown, which reduces the accuracy of the previously established models, as they often use the value of the 129Xe diffusion coefficient in aqueous solutions as a substitute for that in the brain. Dedicated measurements of these important physical parameters for HP 129Xe are required to make further progress on HP 129Xe uptake modeling.

6. PERFUSION IMAGING WITH HP 129XE

Perfusion imaging is used widely in the clinic for the assessment of cerebrovascular physiology, to diagnose brain pathologies such as brain tumors and stroke. Hyperpolarized 129Xe is an exogenous agent that can act as an imaging agent for the evaluation of brain perfusion. This application was first predicted by Duhamel et al. in 2002, but the low polarization values of HP 129Xe (18%–20%) at that time did not permit the researchers to obtain high‐quality results for perfusion assessment in the rat brain. 47 The next attempt for perfusion evaluation was performed by Rao et al. in the human brain in 2018, with HP 129Xe polarized to 35% at 1.5 T. 66 , 67 , 68 Three healthy volunteers inhaled 1 L of 87% enriched 129Xe for a 24‐s breathhold duration. The HP 129Xe uptake images were acquired with time intervals of 8, 16, 24, 32, 40, and 48 after inhalation (Figure 6A–F). The volunteers resumed breathing at 24 s. The resulting images were zero‐padded up to an 80 × 80 in‐plane resolution from 7.81 × 7.81 × 130 mm3 voxels with a 32 × 32 resolution. The first four images were signal averaged for further comparison to ASL imaging. The signal averaged HP 129Xe images exhibited certain advantages over ASL perfusion imaging in that they did not require averaging over a period of several minutes, and they further lacked any undesired signals from blood vessels (Figure 6H). The SNR from the averaged images (Figure 6G) was 31 ± 9, 24 ± 4, and 23 ± 2 from the 3 healthy volunteers. 68 Although this technique is limited by the quantity of 129Xe that is delivered to the brain and also by the loss of 129Xe polarization during transport from the lungs to the brain, it can be used for qualitative perfusion estimates in the human brain.

FIGURE 6.

MRM-29200-FIG-0006-b

Brain perfusion in vivo images of a healthy volunteer. (A–F) HP 129Xe imaging at 1.5 T at 8 s (A), 16 s (B), and 24 s (C) after inhalation during a breath‐hold, and 32 s (D), 40 s (E), 48 s (F) after continuing breathing. (G) Average of the first four images (A–D) with 33‐s total imaging time. (H) Pseudo‐continuous arterial spin‐labeling (ASL) image at 3 T; summation of seven contiguous sections with total imaging time of 10 min. Images are reprinted with permission from the publisher 68

Recently, Rao et al. demonstrated a moderate correlation between the cerebral perfusion values as measured by ASL and the 129Xe uptake in the human brain. 69 To investigate this correlation, the ASL perfusion images were corrected for any depolarization that 129Xe would experience using two exponential terms to account for the T1 polarization decay of HP 129Xe in the blood and during its residency time in gray matter. The authors reported a moderate positive correlation (correlation coefficient range of 0.34–0.63) between the corrected perfusion images obtained using ASL and the HP 129Xe brain images.

The most recent study on HP 129Xe perfusion imaging was conducted by Shepelytskyi et al. in 2020. 52 They demonstrated a novel 129Xe time‐of‐flight (TOF) MRI technique capable of quantitative perfusion measurements. It was a different approach compared with that used by Rao et al. 68 It was based on the time‐resolved depolarization of dissolved HP 129Xe in the brain and the acquisition of dynamic images after subsequent TOF wash‐in delays. It fostered the absence of any background signal and isolated the HP 129Xe delivered by the cerebral blood flow. Cerebral perfusion was recalculated from the dynamic HP 129Xe TOF images using a modified version of Kilian's HP 129Xe uptake model. 48 Three dynamic TOF images were acquired using incremental TOF delay times: 2.5, 6.7, and 7.1 s for axial projections, and 1, 6.5, and 7.1 s for sagittal projections. The images were acquired using a 20 × 20 acquisition matrix with a 12.5 × 12.5 × 70 mm3 voxel volume, and zero‐padded to a 32 × 32 matrix. Figure 7 shows the dynamic HP 129Xe images and resulting perfusion map, overlaid on T2‐weighted 1H brain images.

FIGURE 7.

MRM-29200-FIG-0007-c

Example of perfusion map acquisition. (A,F) High‐resolution, T2‐weighted 1H scans for brain localization. (B–D) Three dynamic HP 129Xe time‐of‐flight (TOF) images acquired 2.5, 6.8, and 7.1 s after the application of a depolarization RF pulse in the axial projection. The gradual SNR increase can be observed with increasing wash‐in time. (E) The perfusion map created by the pixel‐by‐pixel recalculation of the TOF slope was used to calculate the sum of the perfusion rates of gray matter and white matter superimposed on top of a high‐resolution proton brain image. (G–I) Three dynamic TOF images acquired after 1, 6.5, and 8 s in the sagittal view. (J) Perfusion map in the sagittal view. Similar to (E), the intensity values were the net sum of the white‐matter and gray‐matter perfusion rates. Images are reprinted with permission from the publisher 52

In spite of the fact that several major advances have been made in the development of HP 129Xe perfusion imaging, this methodology remains largely in its infancy. There are numerous technical challenges associated with these approaches that originate from the issues highlighted previously. Most applications of HP 129Xe MRI for cerebral perfusion imaging remain, to a large extent, qualitative due to the fact that there have not been any imaging techniques developed that allow an accurate implementation of the existing HP 129Xe dynamic models. Shepelytskyi et al. carried out a quantitative estimation of cerebral perfusion 52 ; however, they used several significant simplifications that somewhat reduce the accuracy of their estimations.

7. FUNCTIONAL MRI WITH HP 129XE

Because HP 129Xe acts as a natural cerebral blood flow tracer, it was suggested that HP 129Xe brain imaging should be capable of detecting physiological activity in the brain via changes in the local hyperpolarized 129Xe density contrast. 8 The fundamental principles behind this mechanism are quite simple compared with the conventional BOLD technique for functional brain MRI (fMRI): Blood flow to areas of the brain that respond to stimulation is higher, and consequently, the local concentration of HP 129Xe in these regions will also be higher. Brain activation maps can therefore be created after subtraction of an HP 129Xe reference image acquired during a resting state. This approach was used by Mazzanti et al. in 2011, who first demonstrated the ability of HP 129Xe brain MRI to detect and map sensory stimulation of the rat brain. 51 Two‐dimensional HP 129Xe CSI images were acquired before and after stimulation (Figure 8B–D) from injection of capsaicin into the fore‐paw. The authors observed an increase of the HP 129Xe signal in the somatosensory brain regions responsible for pain processing. 51

FIGURE 8.

MRM-29200-FIG-0008-c

HP 129Xe fMRI data from three animals. The HP 129Xe signal is shown as a false‐color overlay on the corresponding 1‐mm‐thick coronal proton reference image taken from the same animal. The left panel shows the HP 129Xe signal intensity during baseline, and the right panel shows HP 129Xe signal intensity after injection of capsaicin 20 ul (3 mg/ml) into the right forepaw. The color scale represents SNR, and only signal with SNR above two are shown. Superimposition of a rat brain atlas (18) shows specific areas of the brain: cingulate cortex (Cg), the motor cortex (M), primary somatosensory cortex, SS1 forelimb region (SS1 and SS1 fl), the secondary somatosensory cortex (SS2), and striatum (CPu). The images were reprinted with permission from the publisher 51

Although these early results seemed promising, this methodology was fraught with inherent errors and limitations. The direct subtraction approach for two HP 129Xe brain images is associated with a high level of potential errors caused by the interbreath‐hold variability of the HP 129Xe signal. This signal variability has been estimated to be about 30%, which can cause a large potential for false‐positive and false‐negative results during hemodynamic response mapping. This challenge was overcome by Shepelytskyi et al., who performed fMRI of the human brain using the novel HP 129Xe TOF imaging technique, which had been developed for perfusion assessment. 52 The HP 129Xe TOF pulse sequence substantially reduced the interbreath‐hold signal variability 70 and functioned well for an accurate assessment of the hemodynamic response (HDR). 52 The HDR to visual and motor stimuli (Figure 9) was investigated. The resulting functional brain HDR maps (Figure 9B) correlated well with conventional 1H‐BOLD fMRI (Figure 9D).

FIGURE 9.

MRM-29200-FIG-0009-c

Detection of a hemodynamic response from a colorful visual stimulus using HP 129Xe perfusion mapping validated by BOLD brain functional MRI (fMRI). (A) Experimental design used for hemodynamic response detection. Two separate perfusion maps were acquired during the control (gray screen) and visual stimulation. (B) Hemodynamic response map created by subtracting the control perfusion map from the stimulated perfusion map and overlaid on top of a high‐resolution proton scan. Activation of the occipital lobe, superior parietal lobe, and frontal gyrus was observed. (C) BOLD fMRI experimental design for validation of the HP 129Xe technique. (D) BOLD fMRI 3D activation maps demonstrate a correlation with a 129Xe hemodynamic response map. The activated areas are indicated by colored arrows. Images are reprinted with permission from the publisher 52

Although a spatial correlation between the HP 129Xe HDR maps and conventional 1H‐fMRI images was observed, the HP 129Xe HDR maps had a substantially lower spatial resolution. The HDR maps had a single slice thickness of about 100 mm and an in‐plane pixel size of 7.81 mm2, whereas the conventional proton fMRI images were acquired with a 4‐mm slice thickness and 3.91‐mm2 in‐plane spatial resolution. Despite the significant limitations in spatial resolution, HP 129Xe HDR mapping outperformed conventional fMRI in terms of temporal resolution, as the whole brain was mapped in less than in 20 s.

8. BRAIN DISEASE DETECTION WITH HP 129XE

Despite the low signal intensity of HP 129Xe dissolved in the human brain, it is possible to evaluate various differences in xenon physical properties between healthy subjects and subjects with brain‐related diseases. Zhou et al. demonstrated the first application of HP 129Xe brain CSI for in vivo ischemic stroke imaging 71 in a rat model. The large hypointense region corresponding to the ischemic core (Figure 10C) was observed in an HP 129Xe image (Figure 10B). 71 These results were corroborated by conventional 1H DWI as well as by histology (Figure 10A).

FIGURE 10.

MRM-29200-FIG-0010-c

In vivo evaluation of stroke using 2D 129Xe CSI. (A) Representative 1H apparent diffusion coefficient map image obtained after a right middle cerebral artery occlusion. (B) Corresponding HP 129Xe 2D CSI indicating the large signal void corresponding to the ipsilesional hemisphere. (C) Corresponding 2,3,5‐triphenyltetrazolium chloride (TTC)–stained brain section of the same animal. (D) Tricolor map based on the ADC and TTC images shown in (A) and (C). Green, red, and blue represent nonischemic stroke. The images were reprinted with permission from the publisher 71

Following this initial study in animal models, Rao et al. conducted HP 129Xe brain perfusion imaging in a 52‐year old volunteer who had a stroke 2 years and 3 months before imaging with HP 129Xe. 72 The conventional proton MRI revealed intracranial arterial occlusion with collateralization (Figure 11A). To evaluate perfusion using HP 129Xe, three 32 × 32 images were acquired during a breath‐hold at 8, 16, and 24 s after inhalation of 1 L of HP 129Xe with 35% polarization. The images were reconstructed up to an 80 × 80 in‐plane resolution with subsequent averaging from a 32 × 32 matrix with voxel size 6.875 × 6.875 × 50 mm3. The final image (Figure 11D) revealed a region of signal hypointensity, which indicated poor 129Xe uptake in the stroke area. The regional cerebral blood flow (Figure 11C) calculated from pseudo‐continuous ASL (Figure 11B), however, was higher in the same area, which indicated a delayed hyperperfusion. The lower 129Xe signal can be explained by a shorter mean transit time due to a higher cerebral blood flow. This reduces the transfer of 129Xe to the tissue and delays the delivery of 129Xe to that area, which affects the magnetization because of its T1 decay. Overall, this pioneering study demonstrated proof‐of‐principle contrast for using HP 129Xe imaging for stroke imaging in human subjects.

FIGURE 11.

MRM-29200-FIG-0011-c

Brain MR images acquired in the same session from a subject with established stroke. (A) Axial T1‐weighted image showing infarct in the centrum semiovale of the left cerebral hemisphere (arrow). (B) An axial image from pseudo‐continuous ASL shows hyperintensity in the cerebral cortex adjacent to infarction. (C) Map of CBF estimated from ASL in (B) shows increased perfusion. (D) Hyperpolarized 129Xe brain image shows reduced uptake in the brain tissue supplied by the left internal carotid artery. The 129Xe signal in the region of hypointensity in (D) was 60% lower when compared with the average signal in the healthy region. Images are reprinted with permission from the publisher 72

Another neurological disorder that affects the cerebral blood flow and the brain tissues is Alzheimer's disease (AD). To investigate the possibility of using HP 129Xe imaging for AD detection, Hane et al. conducted an HP 129Xe washout study in 2018. 54 Four participants diagnosed with mild to moderate AD and 4 age‐matched healthy volunteers underwent HP 129Xe gas MRS and MRI during a 20‐s breath‐hold. Sixty dynamic MRS scans were acquired every 2 s starting from initialization of the breath‐hold. Three dynamic balanced SSFP MRI images were acquired at 10, 20, and 30 s after gas inhalation. Five different peaks were observed using MRS that agreed with the spectroscopy results from Rao et al. in 2016. 42 Interestingly, however, in this study, the 129Xe signal from gray matter was 43% lower in AD participants compared with healthy volunteers, and the white‐matter peaks were not statistically different between the two subject cohorts. This reduction in HP 129Xe signal resulted in a decrease in the SNR of images acquired from the AD subjects (Figure 12A). The white‐matter and gray‐matter spectral peaks were monitored over time: The 129Xe washout half‐life for healthy participants was 20 and 16 s for white matter and gray matter, respectively, whereas the 129Xe washout half‐life for participants with AD was 42 and 43 s in white matter (Figure 12C) and gray matter (Figure 12D), respectively. The analysis of the dynamic 129Xe MR images (Figure 12B) revealed that the Xe washout parameters were similar in the caudal brain regions for both cohorts of participants, whereas the prefrontal regions showed a reduction of the localized 129Xe washout parameter in AD volunteers. Therefore, a 129Xe retention parameter was proposed as a potential biomarker for AD detection.

FIGURE 12.

MRM-29200-FIG-0012-c

(A) Axial and sagittal 129Xe MRI of healthy controls and Alzheimer's disease (AD) participants. (B) Xenon washout parameter maps of healthy controls age‐matched to AD patients overlaid onto T2‐weighted anatomical images. MRS SNR of 129Xe‐WM (C) and 129Xe‐GM (D) spectral peaks as a function of time for healthy controls (blue) and AD participants (red). The participants inhaled 500 ml of HP 129Xe and held their breath for 20 s. 129Xe MRS from the brain region was acquired every 2 s. An increase in 129Xe signal after approximately 10 s was noticed as the 129Xe reached the brain. At 20 s, the participant exhaled and the 129Xe signal began to decrease at different rates for AD participants compared with healthy controls for white matter and gray matter. Images are reprinted with permission from the publisher 54

9. DISCUSSION

HP 129Xe MRI of the brain is a promising medical imaging modality that is currently under extensive development. Thirteen articles on HP 129Xe brain imaging were published between the period of the invention of HP 129Xe MRI in 1994, and 2008, while 26 articles were published between 2008 to the present. Furthermore, the number of papers published in the HP129Xe brain imaging field grew steadily over the past decade. The most prominent practical application of HP 129Xe brain imaging so far has been its use for cerebral perfusion imaging. 52 , 54 , 68 , 72 The free dissolution of HP 129Xe in the pulmonary blood renders 129Xe an exogeneous blood‐flow contrast agent. The signal intensity of HP 129Xe brain images is determined primarily by the tissue perfusion, but is further regulated by the level of polarization, the amount of Xe that is inhaled, and the concentration of xenon that is transferred to the brain (Xe solubility is 0.17 in the blood, 0.135 in gray matter, and 0.224 in white matter 48 ). Furthermore, HP 129Xe is an exogeneous perfusion inhalation contrast agent that does not provide any undesired background signal.

Unlike 1H‐ASL perfusion imaging, the lack of background signal and need for intensive signal averaging both provide some of the main advantages of HP 129Xe brain perfusion imaging. With this in mind, the acquisition protocol for HP 129Xe MRI can potentially be simpler compared with that required for ASL MRI for future implementation in the clinic. Because no signal averaging is required, the specific absorption rate of HP 129Xe perfusion imaging scans can also potentially be lower compared with ASL proton scans. Also, in contrast to ASL MRI, HP 129Xe brain perfusion imaging can be performed at low field, due to both the exogeneous nature of HP 129Xe and the fact that its signal has a weak dependence of the Bo magnetic field strength. 73 , 74 , 75 While ASL perfusion imaging is already well developed for clinical use, HP 129Xe perfusion imaging is still in its infancy; further improvements to its method of signal acquisition, and increases to the 129Xe polarization level, will be required to render HP 129Xe perfusion imaging competitive with ASL MRI.

Another advantage of HP 129Xe brain perfusion imaging is its ability for extremely rapid image acquisition. This fact originates from the nonrecoverable nature of the hyperpolarized longitudinal magnetization. Because the HP state is a metastable non‐equilibrium state, spin–lattice relaxation destroys the longitudinal component of its net magnetization over time. Therefore, the use of a short TR is highly beneficial, which also results in a short scan time. HP 129Xe brain image acquisition times are typically on the order of seconds 54 , 72 (Table 2), although further shortening of the scan time is usually possible. Such short image‐acquisition times also reduce the sensitivity of HP 129Xe cerebral perfusion imaging to motion artifacts. In contrast, ASL perfusion imaging scans usually require several minutes due to the need for multiple signal averages, which makes conventional ASL MRI techniques very sensitive to motion artifacts. 76 , 77 , 78

TABLE 2.

Image parameter comparison for human HP 129Xe perfusion imaging, 1H ASL perfusion imaging, and HP 129Xe structural imaging

Imaging parameter HP 129Xe perfusion imaging 1H ASL perfusion imaging HP 129Xe structural imaging
SNR (arb. units)

26 ± 4.36 68 ;

11.2 ± 2.9 (sagittal) 52 ;

9.5 ± 2.9 (axial) 52 ;

12.15 ± 5.45 70

8.74 ± 2.02 (pCASL) 89

16.5 ± 2.2 (pCASL) 90

21.5 ± 3.6 (VSASL) 90

30.7 ± 10.1 (CASPR) 91

18.76 ± 4.95 (axial) 60

19.47 ± 3.25 (sagittal) 60 ;

Acquisition matrix

32 × 32 67 , 70 , 72 ;

20 × 20 52

64 × 64 89 , 90 , 92 , 93 , 94 , 95 , 96 ;

73 × 73 91

128 × 128 97

32 × 32 60
Reconstruction matrix

80 × 80 67 , 72 ;

48 × 48 67 ;

32 × 32 52 , 70 ;

64 × 64 89 , 94 , 96 ;

128 × 128 90 , 98

32 × 32 60
Number of slices 1 52 , 67 , 70 , 72

1 89 ;

8 90 ;

50 91

5 60
Slice thickness (mm)

50 67 , 72 ;

70 (axial) 52 ;

130 (sagittal) 52 ;

3 91

4 90 ;

128 89

20 60
TR (ms)

34 67 , 72 ;

6.1 52 , 70

4000–4600 89 , 90 , 95 , 98

6300 91

6.2 60

Abbreviations: CASPR, Cartesian acquisition with spiral profile reordering; pCASL, pseudo‐continuous ASL; VSASL, velocity‐selective ASL.

Despite the aforementioned advantages of HP 129Xe cerebral perfusion imaging, this methodology currently has several limitations. First, to perform HP 129Xe MRI, the research center or clinical site must possess an MRI scanner capable of performing multinuclear imaging. In addition, a high‐yield 129Xe polarizer (an expensive piece of equipment) is needed, in addition to dedicated MRI coils tuned to the resonance frequency of 129Xe. It is desirable to use a dual‐tuned 1H/129Xe RF head coil, as initial 1H brain localization is required before HP 129Xe brain imaging. Additionally, the use of isotopically enriched 129Xe is often required to achieve acceptable SNR levels, as the concentration of HP 129Xe is relatively low in the brain tissues. The necessity of specialized equipment and isotopically enriched 129Xe gas renders HP 129Xe brain perfusion imaging much more expensive than conventional clinical 1H perfusion MRI techniques.

As previously mentioned, another challenge for HP 129Xe brain imaging is the relatively low concentration of 129Xe dissolved in brain tissues (on the order of μM 48 ). Therefore, the overall HP 129Xe signal level originating from the brain is quite low, resulting in relatively low image SNR, which significantly limits the use of HP 129Xe for anatomical brain imaging. In common practice, to optimize the HP 129Xe brain image SNR, the acquisition matrix is typically kept at a low resolution, and a single slice image is commonly acquired. The most commonly used acquisition matrix is 32 × 32, which is two times smaller compared with the most frequently used acquisition matrix for ASL imaging. In addition, ASL‐based perfusion imaging techniques can acquire images with a slice thickness about 3–4 mm, whereas the minimum slice thickness achieved so far for HP 129Xe imaging is 20 mm. This yields an HP 129Xe voxel size that is at least 20 times larger compared with typical ASL voxel sizes. Recent advances in 3D‐GRE HP 129Xe brain imaging can help to increase the spatial resolution of HP 129Xe cerebral perfusion images and potentially render them comparable to modern clinical ASL standards. Further increases in the HP 129Xe polarization (ideally up to the theoretical limit of 86% 79 ) could potentially facilitate the enhancement in the signal required for use of an acquisition matrix of 64 × 64, which would meet current clinical standards for perfusion imaging. Even if this is accomplished, a 64 × 64 acquisition matrix will not be sufficient for structural brain imaging with HP 129Xe, as conventional 1H MRI is capable of much higher resolution. Therefore, it can be foreseen that HP 129Xe brain MRI does not bode well as a new anatomical MRI imaging modality, but has great potential for applications in the fields of functional imaging, such as perfusion imaging, blood flow detection, and BBB permeability imaging.

Extensive development of both hardware and MR pulse sequences is required to increase the SNR and spatial resolution of HP 129Xe brain MRI. The highest reported level of 129Xe polarization used for brain imaging so far was about 50%, 52 , 60 , 70 which is a significant advancement compared with polarization values used for earlier experiments. 40 , 48 An increase in the polarization level will produce a linear increase in the 129Xe signal level. In addition, an increase in the isotopic enrichment of 129Xe gas used for imaging will also give a linear increase in the image SNR. Finally, the development and implementation of a multichannel phased‐array receiver RF coil could also increase the SNR of HP 129Xe brain imaging. Although preliminary results have been reported using a six‐channel phased array 129Xe brain coil for in vivo single‐voxel spectroscopy, 80 further implementation of the parallel imaging approach for imaging purposes is essential to advance HP 129Xe imaging of the human brain.

Alongside hardware development, future work must also be focused on imaging pulse sequence development and breathing protocol optimization. Due to the short TR requirements, HP 129Xe brain imaging mostly uses GRE imaging pulse sequences. Until recently, the most commonly used MR protocol for 129Xe brain imaging was a thick single‐slice 2D GRE image acquisition with standard sequential k‐space filling. 52 , 54 , 68 , 72 The use of non‐Cartesian k‐space trajectories, which oversample the center of a k‐space (such as radial trajectories) can further increase the image SNR, and may allow the acquisition of thinner slices and higher spatial resolution in HP 129Xe brain images. The downside of using non‐Cartesian k‐space trajectories is that they undersample the outer edges of k‐space, which results in blurriness of the image.

Recent development by Rao et al. allowed progression to 3D multislice isotropic 129Xe brain MRI spectroscopic imaging, which could be further implemented for brain oxygenation mapping and for voxel‐wise quantification of HP 129Xe dissolved in different brain compartments. 59

Another pulse‐sequence approach worth pursuing is translation from 2D‐GRE to 3D‐GRE imaging. 3D‐GRE imaging will allow the image SNR to increase through additional phase encoding in the slice‐selection direction and will allow multislice image acquisition. A proof‐of‐concept demonstration of 3D‐GRE HP 129Xe imaging in humans was recently demonstrated by Grynko et al. 60 In this study, use of a 3D‐GRE sequence allowed reduction of the voxel size by 93% compared with the 2D‐GRE imaging approach. Because all of the spins in the volume of interest get excited simultaneously, 3D‐GRE multislice imaging better uses the hyperpolarized magnetization, compared with 2D‐GRE multislice imaging. Combining 3D‐GRE pulse sequences with non‐Cartesian k‐space trajectories has the potential to further improve the quality of HP 129Xe brain images.

The breathing protocol is another vital factor that should be carefully considered. The most commonly used breathing protocol for HP 129Xe human brain imaging is the inhalation of 1 L of HP gas followed by a subsequent breath‐hold. 52 , 54 , 68 , 72 , 81 , 82 This approach, however, is associated with a high level of signal variability (∼30%). 70 , 82 This signal variability can be caused by numerous factors, such as the exact quantity of HP 129Xe gas dispensed in the bag each time, the T1 relaxation during gas storage before being administered, different concentrations of HP 129Xe in the lungs, cerebral perfusion values, lung–brain arterial transit times, and the amount of time into the breath‐hold when image acquisition begins. All of these factors affect the concentration of HP 129Xe dissolved in the brain at a particular moment in time. A recent study demonstrated that the use of a time‐resolved initial depolarization pulse (TOF technique) reduces the variability of the HP 129Xe signal by up to 2.4 times. 70 Use of an initial depolarization pulse, therefore, is highly beneficial for all further HP 129Xe brain imaging studies. Despite achieving a significant reduction, however, the presence of an initial depolarization pulse did not completely eliminate the interbreath‐hold signal variability issue. A contributing factor that cannot be eliminated originates from variations in blood flow in the cerebral arteries feeding the tissues, which directly affects the HP 129Xe brain signal. The highest level of signal variability was observed to correspond to the brain region supplied by the posterior cerebral artery, whereas the lowest variability corresponded to the region supplied by the anterior cerebral artery. 70

To maximize the image SNR, data acquisition should be performed once the brain tissues are saturated with HP 129Xe. 70 Based on the various HP 129Xe brain uptake models previously developed, 48 , 52 the concentration of HP 129Xe in the brain reaches a maximum at approximately 15 s into the breath‐hold for typical values of cerebral perfusion, arterial transit times, and T1 relaxation times in the blood and brain tissues. To maximize the HP 129Xe SNR, therefore, the image should be acquired at this moment in time. If a subject cannot hold his or her breath for this amount of time (eg, subjects with pulmonary disorders or children), however, or if the imaging purpose is to acquire dynamic images over a breath‐hold (eg, to quantify HP 129Xe uptake), the resulting image SNR can potentially be lower.

It is worth considering other breathing protocols. One that might be of interest for future studies, and that might overcome some of these issues, is continuous breathing using HP 129Xe premixed with oxygen. Continuous breathing protocols have been used for animal lung studies, 83 , 84 and human lung studies with 3He 85 , 86 , 87 and 129Xe. 88 They might be beneficial for HP 129Xe brain imaging, as they can prolong the plateau of the maximum brain 129Xe concentration period. This could allow the conductance of longer scans, signal averaging, and acquisition of HP 129Xe brain images in subjects that are not capable of performing a long breath‐hold.

Currently, the main advantages of ASL over HP 129Xe cerebral perfusion imaging is the higher in‐plane spatial resolution (typically 64 × 64), thinner slices, and commercially available software for image analysis. As indicated in Table 2, the SNR values of HP 129Xe perfusion images are comparable with clinically available pseudo‐continuous ASL perfusion images. The recent implementation of a 3D‐GRE readout has further increased the SNR of HP 129Xe brain images up to the level that is comparable with velocity‐selective ASL. The spatial resolution of HP 129Xe images, however, remains at least four times lower compared with ASL techniques. To bridge this gap, further improvements to the HP 129Xe SNR, as discussed previously, can be converted into increasing the acquisition matrix and reducing the slice thickness.

In addition, the further development of mathematical models for HP 129Xe signal dynamics and its conversion into computer algorithms for HP 129Xe cerebral perfusion image calculation can improve the accuracy of HP 129Xe perfusion imaging. Additional experimental characterization of HP 129Xe in the brain should accompany the optimization of mathematical models, as there still remain multiple fundamental physical properties of HP 129Xe dissolved in cerebral blood and brain tissues that remain unknown. For example, accurate measurements of the T1 relaxation times of each of the HP 129Xe spectral components, as well as HP 129Xe diffusion coefficients, are required. Without these experimental data, it will not be possible to quantify cerebral perfusion accurately.

In spite of these shortcomings, it is inspiring that HP 129Xe brain imaging has already demonstrated its potential to image subjects with AD 54 and stroke. 68 Moreover, because the HP 129Xe brain signal depends on the cerebral perfusion, as well as the permeability of the BBB, it might also be useful for the detection of other diseases associated with cerebral blood flow changes (eg, Parkinson's disease, 99 atherosclerosis 100 ) or those with associated BBB impairment (eg, cerebral small vessel disease, 101 multiple sclerosis 102 ). Additionally, due to the high lipophilicity of xenon, HP 129Xe imaging may also be useful for brain cancer detection. Although there were two preliminary studies on this application, 103 , 104 the validation of HP 129Xe brain imaging for use in cancer detection will require further proof‐of‐concept studies.

10. CONCLUSIONS

HP 129Xe brain imaging is a promising imaging modality that has been developing rapidly over the past several years. With further development, it has the potential to provide rapid and direct imaging of perfusion with an SNR comparable to that of ASL perfusion imaging, even at low field. HP 129Xe perfusion imaging has an extremely fast acquisition time (less than 20 s), has no endogenous background signal, and is much simpler in practice than other MRI techniques from the MR pulse‐sequence design point of view. The rapid acquisition times possible for HP 129Xe perfusion images ensure its insensitivity to motion artifacts. In addition, due to xenon's ability to cross the BBB, assessment of BBB permeability can readily be performed using HP 129Xe MRI. 53 , 58 HP 129Xe perfusion imaging has the potential to become a valuable new perfusion imaging technique that eventually will take its place alongside that of clinical ASL MRI and dynamic contrast‐enhanced perfusion imaging.

ACKNOWLEDGMENT

The authors thank Dilip Balamore for stimulating discussions on the topic of 129Xe brain MR imaging. The authors are also grateful to Vincent Knight for manuscript editing. This research was funded by the Ontario Research Fund (ORF RE 09 029), Northern Ontario Academic Medical Association (A‐18‐05), Mitacs Elevate (IT25574), and Natural Science and Engineering Research Council (NSERC) Discovery grant (RGPIN‐2017‐05359). JMW and MRR were funded by the Medical Research Council (MRC ‐ MR/M008894/1). YS was supported by the Mitacs Elevate Postdoctoral Fellowship (IT25574). VG was supported by an Ontario Trillium Scholarship.

Shepelytskyi Y, Grynko V, Rao MR, et al. Hyperpolarized 129Xe imaging of the brain: Achievements and future challenges. Magn Reson Med. 2022;88:83‐105. doi:10.1002/mrm.29200

Yurii Shepelytskyi and Vira Grynko contributed equally to this work.

Funding information Medical Research Council, MRC ‐ MR/M008894/1; Mitacs Elevate, IT25574; Natural Sciences and Engineering Research Council of Canada, RGPIN‐2017‐05359; Northern Ontario Academic Medicine Association, A‐18‐05; Ontario Research Foundation, ORF RE 09 029

REFERENCES

  • 1. Lauterbur PC. Image formation by induced local interactions: examples employing nuclear magnetic resonance. Nature. 1973;242:190‐191. [PubMed] [Google Scholar]
  • 2. Couch MJ, Blasiak B, Tomanek B, et al. Hyperpolarized and inert gas MRI: the future. Mol Imaging Biol. 2015;17:149‐162. [DOI] [PubMed] [Google Scholar]
  • 3. Caravan P. Strategies for increasing the sensitivity of gadolinium based MRI contrast agents. Chem Soc Rev. 2006;35:512‐523. [DOI] [PubMed] [Google Scholar]
  • 4. Wahsner J, Gale EM, Rodríguez‐Rodríguez A, Caravan P. Chemistry of MRI contrast agents: current challenges and new frontiers. Chem Rev. 2019;119:957‐1057. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Xiao YD, Paudel R, Liu J, Ma C, Zhang ZS, Zhou SK. MRI contrast agents: classification and application (review). Int J Mol Med. 2016;38:1319‐1326. [DOI] [PubMed] [Google Scholar]
  • 6. Buyuksarac B, Ozkan M. Utilization of MR angiography in perfusion imaging for identifying arterial input function. MAGMA. 2017;30:609‐620. [DOI] [PubMed] [Google Scholar]
  • 7. Kim SH, Lee HHS, Kang BJ, et al. Dynamic contrast‐enhanced MRI perfusion parameters as imaging biomarkers of angiogenesis. PLoS One. 2016;11:e0168632. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Albert MS, Cates GD, Driehuys B, et al. Biological magnetic resonance imaging using laser‐polarized 129Xe. Nature. 1994;370:199‐201. [DOI] [PubMed] [Google Scholar]
  • 9. Kurhanewicz J, Vigneron DB, Ardenkjaer‐Larsen JH, et al. Hyperpolarized 13C MRI: path to clinical translation in oncology. Neoplasia. 2019;21:1‐16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Leary D, Svenningsen S, Guo F, Bhatawadekar S, Parraga G, Maksym GN. Hyperpolarized 3He magnetic resonance imaging ventilation defects in asthma: relationship to airway mechanics. Physiol Rep. 2016;4:e12761. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Svenningsen S, Kirby M, Starr D, et al. Hyperpolarized 3He and 129Xe MRI: differences in asthma before bronchodilation. J Magn Reson Imaging. 2013;38:1521‐1530. [DOI] [PubMed] [Google Scholar]
  • 12. Kruger SJ, Nagle SK, Couch MJ, Ohno Y, Albert M, Fain SB. Functional imaging of the lungs with gas agents. J Magn Reson Imaging. 2016;43:295‐315. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Chang YV. MOXE: a model of gas exchange for hyperpolarized 129Xe magnetic resonance of the lung. Magn Reson Med. 2013;69:884‐890. [DOI] [PubMed] [Google Scholar]
  • 14. Chang YV, Quirk JD, Ruset IC, Atkinson JJ, Hersman FW, Woods JC. Quantification of human lung structure and physiology using hyperpolarized 129Xe. Magn Reson Med. 2014;71:339‐344. [DOI] [PubMed] [Google Scholar]
  • 15. Driehuys B, Cofer GP, Pollaro J, Mackel JB, Hedlund LW, Johnson GA. Imaging alveolar‐capillary gas transfer using hyperpolarized 129Xe MRI. Proc Natl Acad Sci USA. 2006;103:18278‐18283. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Stewart NJ, Horn FC, Norquay G, et al. Reproducibility of quantitative indices of lung function and microstructure from 129Xe chemical shift saturation recovery (CSSR) MR spectroscopy. Magn Reson Med. 2017;77:2107‐2113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Albert MS, Balamore D, Kacher DF, Venkatesh AK, Jolesz FA. Hyperpolarized 129XE T1 in oxygenated and deoxygenated blood. NMR Biomed. 2000;13:407‐414. [DOI] [PubMed] [Google Scholar]
  • 18. Albert MS, Kacher DF, Balamore D, Venkatesh AK, Jolesz FA. T1 of 129Xe in blood and the role of oxygenation. J Magn Reson. 1999;140:264‐273. [DOI] [PubMed] [Google Scholar]
  • 19. Wolber J, Cherubini A, Dzik‐Jurasz ASKK, Leach MO, Bifone A. Spin‐lattice relaxation of laser‐polarized xenon in human blood. Proc Natl Acad Sci U S A. 1999;96:3664‐3669. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Norquay G, Leung G, Stewart NJ, Wolber J, Wild JM. 129Xe chemical shift in human blood and pulmonary blood oxygenation measurement in humans using hyperpolarized 129Xe NMR. Magn Reson Med. 2017;77:1399‐1408. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Chacon‐Caldera J, Maunder A, Rao M, et al. Dissolved hyperpolarized xenon‐129 MRI in human kidneys. Magn Reson Med. 2020;83:262‐270. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Albert MS, Balamore D. Development of hyperpolarized noble gas MRI. Nucl Instrum Methods Phys Res A. 1998;402:441‐453. [DOI] [PubMed] [Google Scholar]
  • 23. Ruppert K. Biomedical imaging with hyperpolarized noble gases. Rep Prog Phys. 2014;77:116701. [DOI] [PubMed] [Google Scholar]
  • 24. Cullen SC, Gross EG. The anesthetic properties of xenon in animals and human beings, with additional observations on krypton. Science. 1951;113:580‐582. [DOI] [PubMed] [Google Scholar]
  • 25. Bein B, Hanne P, Hanss R, et al. Effect of xenon anaesthesia on accuracy of cardiac output measurement using partial CO2 rebreathing. Anaesthesia. 2004;59:1104‐1110. [DOI] [PubMed] [Google Scholar]
  • 26. Sanders RD, Franks NP, Maze M. Xenon: no stranger to anaesthesia. Br J Anaesth. 2003;91:709‐717. [DOI] [PubMed] [Google Scholar]
  • 27. Roehl AB, Goetzenich A, Rossaint R, Zoremba N, Hein M. A practical rule for optimal flows for xenon anaesthesia in a semi‐closed anaesthesia circuit. Eur J Anaesthesiol. 2010;27:660‐665. [DOI] [PubMed] [Google Scholar]
  • 28. Martins CF, Neves LA, Chagas R, Ferreira LM, Coelhoso IM, Crespo JG. Removing CO2 from xenon anaesthesia circuits using an amino‐acid ionic liquid solution in a membrane contactor. Sep Purif Technol. 2021;275:119190. [Google Scholar]
  • 29. Yonas H, Darby JM, Marks EC, Durham SR, Maxwell C. CBF measured by Xe‐CT: approach to analysis and normal values. J Cereb Blood Flow Metab. 1991;11:716‐725. [DOI] [PubMed] [Google Scholar]
  • 30. Van Roost D, Hartmann A, Quade G. Changes of cerebral blood flow following dexamethasone treatment in brain tumour patients. A Xe/CT study. Acta Neurochir. 2001;143:37‐44. [DOI] [PubMed] [Google Scholar]
  • 31. Sturnegk P, Mellergård P, Yonas H, Theodorsson A, Hillman J. Potential use of quantitative bedside CBF monitoring (Xe‐CT) for decision making in neurosurgical intensive care. Br J Neurosurg. 2007;21:332‐339. [DOI] [PubMed] [Google Scholar]
  • 32. Swanson SD, Rosen MS, Agranoff BW, Coulter KP, Welsh RC, Chupp TE. Brain MRI with laser‐polarized 129Xe. Magn Reson Med. 1997;38:695‐698. [DOI] [PubMed] [Google Scholar]
  • 33. Mugler JP, Driehuys B, Brookeman JR, et al. MR imaging and spectroscopy using hyperpolarized 129Xe gas: preliminary human results. Magn Reson Med. 1997;37:809‐815. [DOI] [PubMed] [Google Scholar]
  • 34. Swanson SD, Rosen MS, Coulter KP, Welsh RC, Chupp TE. Distribution and dynamics of laser‐polarized 129Xe magnetization in vivo. Magn Reson Med. 1999;42:1137‐1145. [DOI] [PubMed] [Google Scholar]
  • 35. Duhamel G, Choquet P, Leviel J‐L, et al. In vivo 129Xe NMR in rat brain during intra‐arterial injection of hyperpolarized 129Xe dissolved in a lipid emulsion. C R Acad Sci III. 2000;323:529‐536. [DOI] [PubMed] [Google Scholar]
  • 36. Wakai A, Nakamura K, Kershaw J, Kanno I. In vivo MR spectroscopy of hyperpolarized Xe‐129 in rat brain. Int Congr Ser. 2004;1265:139‐143. [Google Scholar]
  • 37. Nakamura K, Kondoh Y, Wakai A, Kershaw J, Wright D, Kanno I. 129Xe spectra from the heads of rats with and without ligation of the external carotid and pterygopalatine arteries. Magn Reson Med. 2005;53:528‐534. [DOI] [PubMed] [Google Scholar]
  • 38. Kershaw J, Nakamura K, Kondoh Y, Wakai A, Suzuki N, Kanno I. Confirming the existence of five peaks in 129Xe rat head spectra. Magn Reson Med. 2007;57:791‐797. [DOI] [PubMed] [Google Scholar]
  • 39. Kilian W, Seifert F, Rinneberg H. Chemical shift imaging of human brain after inhaling hyperpolarized 129Xe‐gas. In: Proceedings of the 10th Annual Meeting of ISMRM, Honolulu, Hawaii, USA, 2002. p 758.
  • 40. Kilian W, Seifert F. 2D chemical shift imaging of hyperpolarized isotopically enriched 129Xe within human brain. In: Proceedings of the 13th Annul Meeting of ISMRM, Miami Beach, Florida, USA, 2005. p 1163.
  • 41. Rao M, Stewart N, Norquay G, Griffiths P, Wild J. Imaging the human brain with dissolved xenon MRI at 1.5T. In: Proceedings of the 23rd Annual Meeting of ISMRM, Toronto, Ontario, Canada, 2015. p 1254.
  • 42. Rao M, Stewart NJ, Norquay G, Griffiths PD, Wild JM. High resolution spectroscopy and chemical shift imaging of hyperpolarized 129Xe dissolved in the human brain in vivo at 1.5 tesla. Magn Reson Med. 2016;75:2227‐2234. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Li H, Zhang Z, Zhong J, et al. Oxygen‐dependent hyperpolarized 129Xe brain MR. NMR Biomed. 2016;29:220‐225. [DOI] [PubMed] [Google Scholar]
  • 44. Zhou X, Tzeng YS, Mansour JK, Mazzanti ML, Sun Y, Albert MS. Optimum ventilation mixture ratio for maximizing hyperpolarized 129Xe MR brain signal. In: Proceedings of the 17th Annual Meeting of ISMRM, Seattlel, Washington, USA, 2007. p 1270.
  • 45. Antonacci MA, Zhang L, Burant A, McCallister D, Branca RT. Simple and robust referencing system enables identification of dissolved‐phase xenon spectral frequencies. Magn Reson Med. 2018;80:431‐441. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Wilson GJ, Santyr GE, Anderson ME, Deluca PM. Longitudinal relaxation times of 129Xe in rat tissue homogenates at 9.4 T. Magn Reson Med. 1999;41:933‐938. [DOI] [PubMed] [Google Scholar]
  • 47. Duhamel G, Choquet P, Grillon E, et al. Global and regional cerebral blood flow measurements using NMR of injected hyperpolarized xenon‐129. Acad Radiol. 2002;9(Suppl 2):S498‐S500. [DOI] [PubMed] [Google Scholar]
  • 48. Kilian W, Seifert F, Rinneberg H. Dynamic NMR spectroscopy of hyperpolarized 129Xe in human brain analyzed by an uptake model. Magn Reson Med. 2004;51:843‐847. [DOI] [PubMed] [Google Scholar]
  • 49. Wakai A, Nakamura K, Kershaw J, Kondoh Y, Wright D, Kanno I. A method for measuring the decay time of hyperpolarized 129Xe magnetization in rat brain without estimation of RF flip angles. Magn Reson Med Sci. 2005;4:19‐25. [DOI] [PubMed] [Google Scholar]
  • 50. Zhou X, Mazzanti ML, Chen JJ, et al. Reinvestigating hyperpolarized 129Xe longitudinal relaxation time in the rat brain with noise considerations. NMR Biomed. 2008;21:217‐225. [DOI] [PubMed] [Google Scholar]
  • 51. Mazzanti ML, Walvick RP, Zhou X, et al. Distribution of hyperpolarized xenon in the brain following sensory stimulation: preliminary MRI findings. PLoS One. 2011;6:e21607. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Shepelytskyi Y, Hane FT, Grynko V, Li T, Hassan A, Albert MS. Hyperpolarized 129Xe time‐of‐flight MR imaging of perfusion and brain function. Diagnostics. 2020;10:630. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53. Rao MR, Norquay G, Stewart NJ, Wild JM. Measuring 129Xe transfer across the blood‐brain barrier using MR spectroscopy. Magn Reson Med. 2021;85:2939‐2949. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Hane FT, Li T, Plata J‐A, Hassan A, Granberg K, Albert MS. Inhaled xenon washout as a biomarker of Alzheimer's disease. Diagnostics. 2018;8:41. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55. Choquet P, Hyacinthe JN, Duhamel G, et al. Method to determine in vivo the relaxation time T1 of hyperpolarized xenon in rat brain. Magn Reson Med. 2003;49:1014‐1018. [DOI] [PubMed] [Google Scholar]
  • 56. Nouls J, Cleveland Z, Freeman M, Moeller H, Hedlund L, Driehuys B. 3D MRI of the hyperpolarized 129Xe distribution in the rat brain. In: Proceedings of the 19th Annual Meeting of ISMRM, Montréal, Quebec, Canada, 2011. p 879.
  • 57. Friedlander Y, Zanette B, Couch M, Kassner A, Santyr G. Spiral‐IDEAL for time‐resolved imaging of hyperpolarized 129Xe kinetics in the rat brain. In: Proceedings of the 27th Annual Meeting of ISMRM, Montréal, Quebec, Canada, 2019. p 4300.
  • 58. Friedlander Y, Zanette B, Lindenmaier A, et al. Hyperpolarized 129Xe MRI of the rat brain with chemical shift saturation recovery and spiral‐IDEAL readout. Magn Reson Med. 2022;87:1971‐1979. [DOI] [PubMed] [Google Scholar]
  • 59. Rao MR, Collier GJ, Norquay G, Schulte RF, Wild JM. 3D isotropic spectroscopic imaging of hyperpolarized 129Xe in the human brain. In: Proceedings of the 29th Annual Meeting of ISMRM [Virtual], 2021. p 3567.
  • 60. Grynko V, Shepelytskyi Y, Li T, Hassan A, Granberg K, Albert MS. Hyperpolarized 129Xe multi‐slice imaging of the human brain using a 3D gradient echo pulse sequence. Magn Reson Med. 2021;86:3175‐3181. [DOI] [PubMed] [Google Scholar]
  • 61. Peled S, Jolesz FA, Tseng C‐H, Nascimben L, Albert MS, Walsworth RL. Determinants of tissue delivery for l29Xe magnetic resonance in humans. Magn Reson Med. 1996;36:340‐344. [DOI] [PubMed] [Google Scholar]
  • 62. Weathersby PK, Homer LD. Solubility of inert gases in biological fluids and tissues: a review. Undersea Biomed Res. 1980;7:277‐296. [PubMed] [Google Scholar]
  • 63. Martin CC, Williams RF, Gao J‐H, Nickerson LDH, Xiong J, Fox PT. The pharmacokinetics of hyperpolarized xenon: implications for cerebral MRI. J Magn Reson Imaging. 1997;7:848‐854. [DOI] [PubMed] [Google Scholar]
  • 64. Kimura A, Imai H, Wakayama T, Fujiwara H. A simple method for quantitative measurement and analysis of hyperpolarized 129Xe uptake dynamics in mouse brain under controlled flow. Magn Reson Med Sci. 2008;7:179‐185. [DOI] [PubMed] [Google Scholar]
  • 65. Imai H, Kimura A, Akiyama K, Ota C, Okimoto K, Fujiwara H. Development of a fast method for quantitative measurement of hyperpolarized 129Xe dynamics in mouse brain. NMR Biomed. 2012;25:210‐217. [DOI] [PubMed] [Google Scholar]
  • 66. Rao M, Stewart N, Norquay G, Wild J. Quantification of perfusion and xenon‐transport across the blood‐brain barrier in humans with hyperpolarized 129Xe brain MR at 1.5T. In: Proceedings of the 23rd Annual Meeting of ISMRM, Toronto, Ontario, Canada, 2015. p 1445.
  • 67. Rao M, Stewart N, Gri P, Norquay G, Wild J. Advances in imaging the human brain with inhaled hyperpolarized xenon‐129 MRI at 1.5 T. In: Proceedings of the 25th Annual Meeting of ISMRM, Honolulu, Hawaii, 2017. p 4543.
  • 68. Rao MR, Stewart NJ, Griffiths PD, Norquay G, Wild JM. Imaging human brain perfusion with inhaled hyperpolarized 129Xe MR imaging. Radiology. 2018;286:659‐665. [DOI] [PubMed] [Google Scholar]
  • 69. Rao M, Norquay G, Wild J. Investigating gas‐exchange and tissue perfusion in the human brain using a combination of proton and hyperpolarized xenon‐129 MRI. In: Proceedings of the 27th Annual Meeting of ISMRM, Montréal, Quebec, Canada, 2019. p 3095.
  • 70. Shepelytskyi Y, Grynko V, Li T, Hassan A, Granberg K, Albert MS. The effects of an initial depolarization pulse on dissolved phase hyperpolarized 129Xe brain MRI. Magn Reson Med. 2021;86:3147‐3155. [DOI] [PubMed] [Google Scholar]
  • 71. Zhou X, Sun Y, Mazzanti M, et al. MRI of stroke using hyperpolarized 129Xe. NMR Biomed. 2011;24:170‐175. [DOI] [PubMed] [Google Scholar]
  • 72. Rao MR, Norquay G, Stewart NJ, Hoggard N, Griffiths PD, Wild JM. Assessment of brain perfusion using hyperpolarized 129Xe MRI in a subject with established stroke. J Magn Reson Imaging. 2019;50:1002‐1004. [DOI] [PubMed] [Google Scholar]
  • 73. Parra‐Robles J, Cross AR, Santyr GE. Theoretical signal‐to‐noise ratio and spatial resolution dependence on the magnetic field strength for hyperpolarized noble gas magnetic resonance imaging of human lungs. Med Phys. 2005;32:221‐229. [DOI] [PubMed] [Google Scholar]
  • 74. Middleton H, Black RD, Saam B, et al. MR imaging with hyperpolarized 3He gas. Magn Reson Med. 1995;33:271‐275. [DOI] [PubMed] [Google Scholar]
  • 75. Venkatesha AK, Zhang AX, Mansour J, et al. MRI of the lung gas‐space at very low‐field using hyperpolarized noble gases. Magn Reson Imaging. 2003;21:773‐776. [DOI] [PubMed] [Google Scholar]
  • 76. Petersen ET, Zimine I, Ho YCL, Golay X. Non‐invasive measurement of perfusion: a critical review of arterial spin labelling techniques. Br J Radiol. 2006;79:688‐701. [DOI] [PubMed] [Google Scholar]
  • 77. Amukotuwa SA, Yu C, Zaharchuk G. 3D pseudocontinuous arterial spin labeling in routine clinical practice: a review of clinically significant artifacts. J Magn Reson Imaging. 2016;43:11‐27. [DOI] [PubMed] [Google Scholar]
  • 78. Alsop DC, Detre JA, Golay X, et al. Recommended implementation of arterial spin‐labeled perfusion MRI for clinical applications: a consensus of the ISMRM perfusion study group and the European consortium for ASL in dementia. Magn Reson Med. 2015;73:102‐116. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79. Norquay G, Parnell SR, Xu X, Parra‐Robles J, Wild JM. Optimized production of hyperpolarized 129Xe at 2 bars for in vivo lung magnetic resonance imaging. J Appl Phys. 2013;113:044908. [Google Scholar]
  • 80. Puddu C, Rao M, Rodgers O, Maunder A, Wild J. 6‐channel RF array for hyperpolarized 129Xe brain MRI at 3T. In: Proceedings of the Annual Meeting of ISMRM [Virtual], 2020. p 4099.
  • 81. Shepelytskyi Y, Hane FT, Grynko V, Li T, Hassan A, Albert MS. Hyperpolarized 129Xe functional brain mapping. In: Proceedings of the Annual Meeting of ISMRM [Virtual], 2020. p 0556.
  • 82. Rao M, Norquay G, Wild J. Assessment of repeatability of imaging inhaled hyperpolarized xenon‐129 in the human brain. In: Proceedings of the 27th Annual Meeting of ISMRM, Montréal, Quebec, Canada, 2019. p 3319.
  • 83. Loza LA, Kadlecek SJ, Pourfathi M, et al. Quantification of ventilation and gas uptake in free‐breathing mice with hyperpolarized 129Xe MRI. IEEE Trans Med Imaging. 2019;38:2081‐2091. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84. Imai H, Matsumoto H, Miyakoshi E, Okumura S, Fujiwara H, Kimura A. Regional fractional ventilation mapping in spontaneously breathing mice using hyperpolarized 129Xe MRI. NMR Biomed. 2015;28:24‐29. [DOI] [PubMed] [Google Scholar]
  • 85. Hamedani H, Emami K, Kadlecek S, et al. Differences in pulmonary function testing and distribution of alveolar oxygen tension in asymptomatic smokers. Am J Respir Crit Care Med. 2012;185:A6046. [Google Scholar]
  • 86. Hamedani H, Clapp JT, Kadlecek SJ, et al. Regional fractional ventilation by using multibreath wash‐in 3He MR imaging. Radiology. 2016;279:917‐924. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 87. Hamedani H, Kadlecek S, Xin Y, et al. A hybrid multibreath wash‐in wash‐out lung function quantification scheme in human subjects using hyperpolarized 3He MRI for simultaneous assessment of specific ventilation, alveolar oxygen tension, oxygen uptake, and air trapping. Magn Reson Med. 2017;78:611‐624. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 88. Hamedani H, Bermudez F, Baron R, et al. A comparison of hyperpolarized gas imaging of aeration and fractional ventilation. Am J Respir Crit Care Med. 2019;199:A5772. [Google Scholar]
  • 89. Vidorreta M, Wang Z, Chang YV, Wolk DA, Fernández‐Seara MA, Detre JA. Whole‐brain background‐suppressed pCASL MRI with 1D‐accelerated 3D RARE stack‐of‐spirals readout. PLoS One. 2017;12:1‐16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90. Liu D, Xu F, Li W, van Zijl PC, Lin DD, Qin Q. Improved velocity‐selective‐inversion arterial spin labeling for cerebral blood flow mapping with 3D acquisition. Magn Reson Med. 2020;84:2512‐2522. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 91. Greer JS, Wang X, Wang Y, et al. Robust pCASL perfusion imaging using a 3D Cartesian acquisition with spiral profile reordering (CASPR). Magn Reson Med. 2019;82:1713‐1724. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 92. Kilroy E, Apostolova L, Liu C, Yan L, Ringman J, Wang DJJ. Reliability of two‐dimensional and three‐dimensional pseudo‐continuous arterial spin labeling perfusion MRI in elderly populations: comparison with 15o‐water positron emission tomography. J Magn Reson Imaging. 2014;39:931‐939. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 93. Lim J, Wu W, Chau WJ, Detre JA, Dinges DF, Rao H. Imaging brain fatigue from sustained mental workload: an ASL perfusion study of the time‐on‐task effect. Neuroimage. 2010;49:3426‐3435. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 94. Schmithorst VJ, Ph D, Hernandez‐garcia L, et al. Optimized simultaneous ASL and BOLD functional imaging of the whole brain. J Magn Reson Imaging. 2015;39:1104‐1117. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 95. White CM, Pope WB, Zaw T, et al. Regional and voxel‐wise comparisons of blood flow measurements between dynamic susceptibility contrast magnetic resonance imaging (DSC‐MRI) and arterial spin labeling (ASL) in brain tumors. J Neuroimaging. 2014;24:23‐30. [DOI] [PubMed] [Google Scholar]
  • 96. Noguchi T, Yoshiura T, Hiwatashi A, et al. Perfusion imaging of brain tumors using arterial spin‐labeling: correlation with histopathologic vascular density. Am J Neuroradiol. 2008;29:688‐693. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 97. Okazawa H, Higashino Y, Tsujikawa T, et al. Noninvasive method for measurement of cerebral blood flow using O‐15 water PET/MRI with ASL correlation. Eur J Radiol. 2018;105:102‐109. [DOI] [PubMed] [Google Scholar]
  • 98. Liu K, Li B, Qian S, et al. Mental fatigue after mild traumatic brain injury: a 3D‐ASL perfusion study. Brain Imaging Behav. 2016;10:857‐868. [DOI] [PubMed] [Google Scholar]
  • 99. Melzer TR, Watts R, MacAskill MR, et al. Arterial spin labelling reveals an abnormal cerebral perfusion pattern in Parkinson's disease. Brain. 2011;134:845‐855. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 100. Sfyroeras GS, Karkos CD, Gerassimidis TS. Cerebral perfusion patterns in patients with extracranial carotid atherosclerosis and the impact of carotid stenting: a review. J Cardiovasc Surg (Torino). 2008;49:497‐502. [PubMed] [Google Scholar]
  • 101. Wong SM, Jansen JFA, Zhang CE, et al. Blood‐brain barrier impairment and hypoperfusion are linked in cerebral small vessel disease. Neurology. 2019;92:e1669‐e1677. [DOI] [PubMed] [Google Scholar]
  • 102. Leech S, Kirk J, Plumb J, McQuaid S. Persistent endothelial abnormalities and blood? Brain barrier leak in primary and secondary progressive multiple sclerosis. Neuropathol Appl Neurobiol. 2007;33:86‐98. [DOI] [PubMed] [Google Scholar]
  • 103. Wolber J, McIntyre DJO, Rodrigues LM, et al. In vivo hyperpolarized 129Xe NMR spectroscopy in tumors. Magn Reson Med. 2001;46:586‐591. [DOI] [PubMed] [Google Scholar]
  • 104. Il'yasov AV, Mazitov RK, Enikeev KM, Panov AN, Il'yasov NA, Khasanov RZ. 129Xe NMR in study of tissues and plants. Appl Magn Reson. 1999;17:77‐84. [Google Scholar]

Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley