Abstract
Radiation dose optimization is particularly important in pediatric radiology, as children are more susceptible to potential harmful effects of ionizing radiation. However, only one narrative review about artificial intelligence (AI) for dose optimization in pediatric computed tomography (CT) has been published yet. The purpose of this systematic review is to answer the question “What are the AI techniques and architectures introduced in pediatric radiology for dose optimization, their specific application areas, and performances?” Literature search with use of electronic databases was conducted on 3 June 2022. Sixteen articles that met selection criteria were included. The included studies showed deep convolutional neural network (CNN) was the most common AI technique and architecture used for dose optimization in pediatric radiology. All but three included studies evaluated AI performance in dose optimization of abdomen, chest, head, neck, and pelvis CT; CT angiography; and dual-energy CT through deep learning image reconstruction. Most studies demonstrated that AI could reduce radiation dose by 36–70% without losing diagnostic information. Despite the dominance of commercially available AI models based on deep CNN with promising outcomes, homegrown models could provide comparable performances. Future exploration of AI value for dose optimization in pediatric radiology is necessary due to small sample sizes and narrow scopes (only three modalities, CT, positron emission tomography/magnetic resonance imaging and mobile radiography, and not all examination types covered) of existing studies.
Keywords: as low as reasonably achievable, computed tomography, convolutional neural network, deep learning, dose reduction, generative adversarial network, image processing, machine learning, medical imaging, noise
1. Introduction
Radiology is an indispensable part of modern healthcare. However, most of the medical imaging modalities, such as computed tomography (CT), positron emission tomography (PET), and general radiography, use ionizing radiation for image production [1,2,3,4,5,6,7,8,9,10,11,12,13,14,15,16]. Although the radiation dose involved in these imaging modalities is low (<100 mSv), and their real risk is unclear, some epidemiologic and biologic studies have demonstrated that these radiological examinations can cause cancers [17,18,19,20,21,22,23]. Hence, “as low as reasonably achievable” (ALARA) has become the fundamental principle of radiology practice [17,24,25]. International Commission on Radiological Protection (ICRP) has introduced the diagnostic reference levels (DRLs) initiative for radiological departments to identify examinations with radiation doses exceeding their corresponding DRLs and trigger the radiation dose-optimization process [26,27,28,29,30,31,32]. As the radiation used in radiological examinations is the source of signal, a reduction of the radiation amount results in a decrease of signal strength and an increase of image noise. Traditionally, the dose-optimization process involves the manipulation of a range of exposure/scan parameters and identification of parameters that deliver the lowest radiation dose but still producing images able to meet minimal diagnostic requirements [33,34,35,36]. Since the introduction of digital medical imaging, image processing has played an important role in the radiation dose optimization [37,38,39]. However, typical image processing techniques are unable to overcome the tradeoff between image noise and spatial resolution [9,10,11,12]. For the last few years, artificial intelligence (AI) has been introduced into radiology for radiation dose optimization. Studies have demonstrated its ability in pushing the limit, i.e., able to further reduce the radiation dose but without sacrificing image quality, such as noise and spatial resolution [1,6,9,10,11,12,15,16].
The dose optimization is particularly important for pediatric patients because they have longer life and more rapid cell proliferation, leading to two to three times more susceptibility to the potential harmful effects of ionizing radiation than the adult counterpart [17,33,36,40]. Nonetheless, dose optimization in pediatric radiology is challenging, as there is a greater variation of body size and composition within and across age groups [4,33]. Despite being an important and challenging topic area, apparently, only one narrative review article on this area has been published yet, and it is about deep learning (a subset of AI) image reconstruction (DLIR) for dose optimization in pediatric CT [17]. Hence, it is timely to conduct a systematic review about the use of AI for dose optimization in pediatric radiology. The purpose of this article is to systematically review published original studies to answer the question “What are the AI techniques and architectures introduced in pediatric radiology for dose optimization, their specific application areas, and performances?”
2. Materials and Methods
This systematic review on the AI for radiation dose optimization in pediatric radiology was conducted as per the PRISMA guidelines and patient/problem, intervention, comparison, and outcome (PICO) model [41,42]. This involved literature search, article selection, and data extraction and synthesis.
2.1. Literature Search
The literature search with use of electronic scholarly publication databases, including Google Scholar, PubMed/Medline, ScienceDirect, Scopus, and Web of Science was conducted on 3 June 2022 to identify articles about the AI for dose optimization in pediatric radiology published between 2017 and 2022. The search statement used was (“Artificial Intelligence” OR “Machine Learning” OR “Deep Learning”) AND (“Dose Optimization” OR “Dose Reduction”) AND (“Pediatric” OR “Children”) AND (“Radiology” OR “Medical Imaging”). The keywords used in the search were based on the review focus. The year range was determined based on a narrative review about current and future applications of AI in radiology, which showed the use of AI for dose optimization in radiology not evident before 2017 [43].
2.2. Article Selection
A reviewer with more than 20 years of experience in conducting literature review was involved in the article selection process [44]. Only peer-reviewed original research articles that were written in English and focused on the use of AI for dose optimization in pediatric radiology were included. Grey literature, conference abstracts, editorials, review, perspective, opinion, commentary, and non-peer-reviewed (e.g., those published via the arXiv research-sharing platform, etc.) articles were excluded because of the following reasons: Well-established methodological guidelines were not available for proper selection of grey literature. Conference abstracts could not provide complete study information. Only secondary information was presented in editorials, review, perspective, opinion, and commentary articles. Non-peer-reviewed articles might provide unsubstantiated information [45,46].
Figure 1 illustrates details of the article selection process [41]. A three-stage screening process through assessing (1) article titles, (2) abstracts, and (3) full texts against the selection criteria was employed after duplicate article removal from results of the database search. Every non-duplicate article within the search results was retained until its exclusion could be decided. Lists of references of the included papers were reviewed for additional, relevant article identification [46].
2.3. Data Extraction and Synthesis
A data extraction form (Table 1) was developed based on a recent systematic review on the use of AI in radiology [45]. The data, including names and countries of authors, publication years, clinical domains (radiology/nuclear medicine), AI techniques (such as machine learning and deep learning (DL)), model architectures (e.g., convolutional neural network (CNN), generative adversarial network (GAN), etc.), specific application areas (i.e., examination types and approaches that AI was used to achieve dose optimization), imaging modalities, details of AI model development (i.e., whether homegrown or commercially available model and arrangement of model training and testing), AI model evaluation approach (e.g., phantom study, clinical study, etc.), and key findings of AI model performance in dose optimization (including figures of dose reduction and diagnostic values and subjective and objective image assessment scores), were extracted from each included paper. To facilitate comparison of the AI model performance, percentage of dose reduction (if not reported) was synthesized based on the available absolute dose figures. When multiple image-quality-related figures were reported in a study, the most clinically relevant figures were presented. Diagnostic values were considered the most clinically relevant performance figures, while the objective image assessment scores were determined least relevant [47,48]. Quality assessment scores were determined for all included articles based on the quality assessment tool for studies with diverse designs (QATSDD) and expressed as percentages [49]. Less than 50%, 50–70%, and greater than 70% were considered low, moderate, and high study quality, respectively [46].
Table 1.
Author, Year, and Country | Clinical Domain | AI Technique and Architecture | Application Area for Dose Optimization | Imaging Modality | AI Model Development | AI Model Evaluation Approach | Key Findings of AI Model Performance |
---|---|---|---|---|---|---|---|
Brady et al. (2021), USA [1] | Radiology | DL-Convolutional neural network | DLIR of contrast-enhanced pediatric chest-abdomen-pelvis CT | CT | Commercially available model (AiCE, Canon Medical Systems, Tochigi, Japan) trained by image pairs of lower-dose CT with HIR and high-dose CT with MBIR and tested with datasets not involved in the training | Retrospective clinical study involving 19 children (mean age: 11 ± 5 y; range: 3–19 y) | With SBIR as reference, 52% dose reduction with noise texture and spatial resolution maintained, highest radiologists’ confidence rating (scale 1–10) among 4 approaches (DLIR: 7 ± 1; SBIR & MBIR: 6.2 ± 1; FBP: 4.6 ± 1), and object detectability improved by 51%, 18%, and 11% when compared with FBP, SBIR, and MBIR, respectively. |
Jeon et al. (2022), Republic of Korea [2] | Radiology | DL-Convolutional neural network | DLIR of non-contrast pediatric abdominal CT | CT | Commercially available model (AiCE, Canon Medical Systems) trained by image pairs of lower-dose CT with HIR and high-dose CT with MBIR and tested with datasets not involved in the training | Phantom study involving phantoms with diameters, 16 (pediatric) and 32 cm (adult) | For 80–120 kV, CTDIvol of DLIR images of pediatric phantom with CNR similar to corresponding FBP images was 5% of counterpart, representing 20-fold dose reduction potential. |
Kim et al. (2017), Republic of Korea [3] | Radiology | DL-Gaussian mixture model | Post-processing of non-contrast pediatric abdominal CT images | CT | Homegrown model without training and testing details disclosed | Phantom study involving PMMA phantoms with diameters 12, 16, 20, 24, and 32 cm | Contrast-to-noise ratio dose increase by 1.7–4.9 times and 1.6–4.2 times for settings of 80–140 kV and fixed-tube current of 200 mA and 50–300 mA and fixed-tube potential of 120 kV, respectively. |
Krueger et al. (2022), Germany [4] | Radiology | DL-Convolutional neural network | Post-processing of pediatric mobile chest and abdominal X-ray images acquired in intensive care units | Mobile radiography | Commercially available model (SimGrid, Samsung Electronics Co., Ltd., Suwon-si, Republic of Korea) trained by 30,000 images | Retrospective clinical study involving 210 images of 134 children (mean age: 4.2 y; range: 0–18 y) | Subjective image quality assessment demonstrated significant image quality improvement for patients with weight greater than 10 kg (odds ratio = 6.68, p < 0.0001), indicating its dose reduction potential. |
Lee et al. (2021), Republic of Korea [5] | Radiology | DL-Convolutional neural network | Post-processing of pediatric abdominal DECT with lower CM concentration and noise-optimized virtual monoenergetic IR | CT | Commercially available model (ClariCT.AI, ClariPI, Seoul, Republic of Korea) trained by 410,000 image pairs of low- and standard-dose CT from 210 patients and tested with datasets not involved in the training | Retrospective clinical study involving 29 children (mean age: 10.1 y; range: 2–19 y) | 19.6% CTDIvol and 14.3% CM concentration reductions in pediatric abdominal DECT with noise-optimized virtual monoenergetic IR when compared with those of standard CT. |
Nagayama et al. (2022), Japan [6] | Radiology | DL-Convolutional neural network | DLIR of contrast-enhanced pediatric abdominal CT | CT | Commercially available model (AiCE Body Sharp, Canon Medical Systems) trained by image pairs of lower-dose CT with HIR and high-dose CT with MBIR and tested with datasets not involved in the training | Phantom and retrospective clinical study involving 20 cm diameter Catphan 700 phantom (The Phantom Laboratory, Greenwich, NY, USA) and 65 children (mean age: 25.0 ± 25.2 months; range: 0–81 months), respectively | In pediatric contrast-enhanced 80 kV abdominal CT, 53.7% SSDE reduction with better image quality (e.g., lower noise, noise texture, and edge sharpness improvements, etc.) when compared with standard-dose HIR. |
Park et al. (2022), Republic of Korea [7] | Radiology | DL-Generative adversarial network | Post-processing of contrast-enhanced pediatric abdominal CT | CT | Homegrown model trained by 840 unpaired low- (42 patients; mean age: 7.2 ± 2.5 y) and standard-dose (42 patients; mean age: 6.2 ± 2.2 y) pediatric abdominal CT images and validated with 41 datasets (820 images; patient mean age: 7.4 ± 2.2 y) not involved in the training | Retrospective clinical study involving 660 images from 33 children | When compared with standard-dose CT, 36.6% CTDIvol reduction with image noise (7.1 ± 2.7) and CNR (portal vein: 21.2 ± 10.1; liver: 8.5± 4.3) similar to those of SAFIRE images (noise: 9.5 ± 4.0; CNR: 21.2 ± 9.8 (portal vein) and 8.5 ± 5.0 (liver)), and visual assessment (standard-dose and DL-processed image differentiation) yielded a sensitivity and specificity of 61.2% and 35.0%, indicating similar image quality. |
Sun et al. (2021), People’s Republic of China and USA [8] | Radiology | DL-Convolutional neural network | DLIR of pediatric neck, chest, and abdominal CT angiography | CT | Commercially available model (TrueFidelity, General Electric Healthcare, Chicago, IL, USA) trained by image pairs of low-dose CT projection (raw) data and higher-dose CT reconstructed by FBP from phantoms and patients | Retrospective clinical study involving 32 children with Takayasu’s arteritis (mean age: 9.1 ± 4.5 y; range: 1–17 y) | High-strength DLIR had highest small artery detection and diagnostic confidence scores based on a 5-point scale (3.53 ± 0.51 and 4.09 ± 0.30) when compared with FBP (2.94 ± 0.25 and 2.91 ± 0.30), ASIR-V 50% (3.03 ± 0.18 and 3.03 ± 0.18), and ASIR-V 100% (2.84 ± 0.37 and 3.00 ± 0.00) groups, respectively, demonstrating its dose reduction potential. |
Sun et al. (2021), People’s Republic of China and USA [9] | Radiology | DL-Convolutional neural network | DLIR of pediatric chest CT angiography | CT | Commercially available model (TrueFidelity, General Electric Healthcare) trained by image pairs of low-dose CT projection (raw) data and higher-dose CT reconstructed by FBP from phantoms and patients | Retrospective clinical study involving 33 children (mean age: 5.9 ± 4.2 y; range: 4 months–13 y) | High-strength DLIR images had highest scores of subjective image assessment with a scale of 1–5 (noise: 4.05 ± 0.21 (little); vascular edge: 4.05 ± 0.58 (clear identification); vascular contrast: 4.14 ± 0.64 (good)) when compared with ASiR-V 100% (3.36 ± 0.58; 2.86 ± 0.56; 4.00 ± 0.62) and ASiR-V 50% (2.27 ± 0.55; 3.77 ± 0.61; 3.14 ± 0.64), respectively, demonstrating its potential for further dose reduction. |
Sun et al. (2021), People’s Republic of China and USA [10] | Radiology | DL-Convolutional neural network | DLIR of pediatric chest CT angiography | CT | Commercially available model (TrueFidelity, General Electric Healthcare) trained by image pairs of low-dose CT projection (raw) data and higher-dose CT reconstructed by FBP from phantoms and patients | Prospective case-control study involving 54 children (control group: n = 27; mean age: 9.5 ± 2.4 y; range: 5–13 y; and study group: n = 27; mean age: 9.3 ± 3.1 y; range: 5–14 y) | High-strength DLIR with 70 kV, NI of 22, and CM injection time of 4 s allowed 36% radiation dose and 53% CM dose reductions with scores of subjective image assessment against a 5-point scale similar to control group, ASiR-V 50% with 80 kV, NI of 19, and CM injection time of 8 s (artery contrast: 4.56 vs. 4.78; image quality: 3.67 vs 3.44; diagnostic confidence: 4.74 vs. 4.74; p > 0.05). |
Sun et al. (2021), People’s Republic of China and USA [11] | Radiology | DL-Convolutional neural network | DLIR of pediatric chest CT angiography | CT | Commercially available model (TrueFidelity, General Electric Healthcare) trained by image pairs of low-dose CT projection (raw) data and higher-dose CT reconstructed by FBP from phantoms and patients | Prospective case-control study involving 92 children (control group: n = 46; mean age: 5.9 ± 4.2 y; range: 4 months–13 y; and study group: n = 46; mean age: 5.9 ± 4.2 y; range: 4 months–13 y) | High-strength DLIR with 70 kV allowed 11% radiation dose and 20% CM dose reductions with higher scores of subjective image assessment against a 5-point scale (noise: 4 (little); vascular contrast: 4 (good); vascular edge: 4 (clear identification)) when compared with control group, ASiR-V 50% with 100 kV (noise: 2 (high); vascular contrast: 3 (fair); vascular edge: 3 (identifiable)). |
Sun et al. (2021), People’s Republic of China and USA [12] | Radiology | DL-Convolutional neural network | DLIR of non-contrast pediatric head CT | CT | Commercially available model (TrueFidelity, General Electric Healthcare) trained by image pairs of low-dose CT projection (raw) data and higher-dose CT reconstructed by FBP from phantoms and patients | Retrospective clinical study involving 50 children (median age: 2 y; range: 0.1–14 y) | High-strength DLIR images with 0.625 mm slice thickness had similar subjective image quality score and measured noise when compared with ASiR-V 50% 5 mm slice thickness images (p > 0.05) but able to reduce radiation dose by 85% and improve lesion detection (69 vs. 65 detected) due to spatial resolution increase. |
Theruvath et al. (2021), USA [13] | Nuclear medicine | DL-2.5 dimensional encoder-decoder U-Net convolutional neural network | Post-processing of pediatric and adult whole-body PET images | PET/MRI | Commercially available model (SubtlePET 1.3, Subtle Medical, Menlo Park, CA, USA) trained by low- and high-count PET image pairs from whole-body PET/CT and PET/MRI studies of pediatric and adult patients and tested with adult brain and whole-body studies | Prospective clinical study involving 20 pediatric and adult lymphoma patients (mean age: 16.0 ± 6.0 y; range: 6–30 y) | Up to 50% 18F-FDG dose reduction with 100% sensitivity and specificity for correct assessment of pediatric and adult lymphoma patients’ treatment response. |
Wang et al. (2021), USA and Germany [14] | Nuclear medicine | DL-Convolutional neural network | Post-processing of pediatric and adult ultra-low-dose whole-body PET/MRI images to synthesize standard-dose PET images | PET/MRI | Homegrown model development based on Lim et al.’s [50] open-source enhanced deep super-resolution network model through transfer learning with 17 standard-dose PET, simulated 6.25% ultra-low-dose PET and MRI training datasets, and 6 independent testing datasets acquired in USA | Prospective clinical study involving 34 pediatric and adult lymphoma patients in USA (n = 23; mean age: 17 ± 7 y, range: 6–30 y) and Germany (n = 11; mean age: 14 ± 5 y; range: 3–18 y) | Expert readers’ agreements of tumor diagnosis between standard and AI-processed 6.25% ultra-low-dose PET images (kappa = 0.942 (USA datasets) and 0.912 (Germany datasets)) were significantly greater than the agreements between standard and 6.25% ultra-low-dose PET images (kappa = 0.650 (USA datasets) and 0.834 (Germany datasets)). Diagnostic accuracy of AI-processed 6.25% ultra-low-dose PET images was adequate, representing 93.75% dose reduction capability. |
Yoon et al. (2021), Republic of Korea [15] | Radiology | DL-Convolutional neural network | DLIR of pediatric contrast enhanced abdominal and non-contrast and contrast enhanced chest CT | CT | Commercially available model (TrueFidelity, General Electric Healthcare) trained by image pairs of low-dose CT projection (raw) data and higher-dose CT reconstructed by FBP from phantoms and patients | Phantom and retrospective clinical study involving The Phantom Laboratory’s 20 cm diameter Catphan 500 phantom and 51 pediatric patients (mean age: 11.5 ± 4.6 y; range: 1–18 y), respectively | When compared with ASiR-V 50%, medium- and high-strength DLIR images of contrast enhanced abdominal (n = 23) and non-contrast (n = 16) and contrast enhanced (n = 12) chest CT had statistically significantly higher subjective image quality score and lower noise (p < 0.001), illustrating its dose reduction potential. |
Zhang et al. (2022), People’s Republic of China [16] | Radiology | DL-Convolutional neural network | DLIR of non-contrast pediatric abdominal and chest CT | CT | Commercially available model (TrueFidelity, General Electric Healthcare) trained by image pairs of low-dose CT projection (raw) data and higher-dose CT reconstructed by FBP from phantoms and patients | Phantom and prospective clinical study involving a pediatric (equivalent to 5-year-old patient) whole body phantom (PBU−70, Kyoto Kagaku Co., Ltd., Kyoto, Japan) and 20 children (mean age: 5.4 ± 1.2 y; range: 4–6 y), respectively | When compared with ASiR-V 70%, high-strength DLIR achieved about 70% and 60% dose reductions for pediatric non-contrast abdominal (n = 10) and chest (n = 10) CT, respectively. However, high-strength DLIR did not statistically significantly improve subjective image assessment score of chest CT. |
18F-FDG, fluorine−18-fluorodeoxyglucose; AI, artificial intelligence; AiCE, Advanced Intelligent Clear-IQ Engine; ASiR-V, adaptive statistical iterative reconstruction-Veo; CM, contrast medium; CNR, contrast-to-noise ratio; CT, computed tomography; CTDIvol, volume computed tomography dose index; DECT, dual-energy computed tomography; DL, deep learning; DLIR, deep learning image reconstruction; FBP, filtered back projection; HIR, hybrid iterative reconstruction; IR, image reconstruction; MBIR, model-based iterative reconstruction; MRI, magnetic resonance imaging; NI, noise index; No., number; PET, positron emission tomography; PMMA, polymethyl methacrylate; SAFIRE, sinogram affirmed iterative reconstruction; SBIR, statistical-based iterative reconstruction; SSDE, size-specific dose estimate; y, year.
3. Results
Sixteen articles met the selection criteria and were included in this review. Table 1 shows these study characteristics [1,2,3,4,5,6,7,8,9,10,11,12,13,14,15,16]. All but one article were published in the last two years, representing that the AI for dose optimization in pediatric radiology has only just become popular [1,2,4,5,6,7,8,9,10,11,12,13,14,15,16]. Nearly all (14) studies were determined high quality [1,4,5,6,7,8,9,10,11,12,13,14,15,16], and the lower quality ones were “pure” phantom studies [2,3]. All studies used the DL technique [1,2,3,4,5,6,7,8,9,10,11,12,13,14,15,16] and were conducted by 12 groups of researchers from USA (n = 8) [1,8,9,10,11,12,13,14], People’s Republic of China (n = 6) [8,9,10,11,12,16], Republic of Korea (n = 5) [2,3,5,7,15], Germany (n = 2) [4,14], and Japan (n = 1) [6]. Only two studies were about nuclear medicine (whole-body PET/magnetic resonance imaging (MRI)) [13,14]. For the 14 radiology-related studies [1,2,3,4,5,6,7,8,9,10,11,12,15,16], all except one were related to CT, covering body parts such as the abdomen (n = 10) [1,2,3,4,5,6,7,8,15,16], chest (n = 8) [1,4,8,9,10,11,15,16], head (n = 1) [12], neck (n = 1) [8], and pelvis (n = 1) [1], and four focused on CT angiography [8,9,10,11] as well as one on dual-energy CT (DECT) [5]. Thirteen studies (81.3%) used commercially available AI models for dose optimization (TrueFidelity, General Electric Healthcare (GE): n = 7 [8,9,10,11,12,15,16]; AiCE, Canon Medical Systems: n = 3 [1,2,6]; ClariCT.AI, ClariPI: n = 1 [5]; SimGrid, Samsung Electronics Co., Ltd.: n = 1 [4]; SubtlePET 1.3, Subtle Medical: n = 1 [13]). Ten studies (62.5%) employed DLIR for CT dose optimization as a result of the dominance of GE TrueFidelity and Canon AiCE with the CNN architecture [1,2,6,8,9,10,11,12,15,16]. Hence, the CNN was the most popular (87.5%) AI architecture among the included studies [1,2,4,5,6,8,9,10,11,12,13,14,15,16].
Clinical evaluation of the AI model performance was conducted in all but two studies [1,4,5,6,7,8,9,10,11,12,13,14,15,16], and the use of phantom for an additional evaluation was also noted in three (21.4%) of the clinical studies [6,15,16]. Collectively, these clinical studies covered pediatric patients aged from 0 to 18 years [1,4,5,6,7,8,9,10,11,12,13,14,15,16]. All except one study recruited less than 100 patients for the model evaluation [1,5,6,7,8,9,10,11,12,13,14,15,16]. The only exception had 134 patients [4]. A retrospective approach was employed in about two-thirds (9 out of 14) of the clinical studies [1,4,5,6,7,8,9,12,15]. About 70% of (11) included studies reported absolute dose figures and/or dose reduction percentages. The performance of dose reduction of the AI models with acceptable image quality ranged from 11% to 95% [1,2,5,6,7,10,11,12,13,14,16]. More than half (6) of these studies reported that their AI models were able to achieve dose reductions between 36% and 70% [1,6,7,10,13,16] although three other studies showed dose reductions between 85% and 95% [2,12,14], and another two showed 11–20% dose reductions [5,11]. For the two most popular commercial AI models, GE TrueFidelity and Canon AiCE, great variations of their dose reduction performances, namely 11–85% and 52–95%, were noted, respectively [1,2,6,10,11,12,16].
4. Discussion
The findings of this systematic review on the AI for radiation dose optimization in pediatric radiology are consistent with several recent narrative reviews about the use of AI in radiology [17,43,51]. For the narrative review about the current and future applications of AI in radiology published in 2018 [43], only one study regarding low-dose CT denoising published in 2017 was cited [52]. However, recently, more studies about the use of AI for dose optimization have been published, resulting in a narrative review about the AI for dose optimization in pediatric CT available in 2021 [17]. This demonstrates that the use of AI for dose optimization in pediatric radiology has attracted the attention of the profession recently. That narrative review indicated the DLIR allowed 30–80% dose reduction in pediatric CT but was still able to produce images with diagnostic quality. This systematic review with inclusion of more studies about dose optimization in pediatric CT and covering other imaging modalities shows that the majority of the AI models were able to reduce the radiation dose by 36–70% [1,6,7,10,13,16]. Nonetheless, three studies included in this systematic review demonstrated that the use of AI could achieve further radiation dose reduction (up to 95%) [2,12,14]. Apparently, the large variation of dose reduction performances is due to the retrospective nature of many included studies [1,4,5,6,7,8,9,12,15], which did not allow further manipulation of examination/scan parameters to obtain ultra-low-dose images for evaluating whether the AI models could restore the quality of these ultra-low-dose images to close to the original [9]. Although there is a greater flexibility for phantom studies to manipulate the examination/scan parameters without any ethical and radiation dose concerns, enabling further exploration of the potential of these AI models, their evaluation outcomes tend to be less clinically relevant [47,48]. For example, Jeon et al. [2] reported that Canon AiCE was able to reduce the CT dose by 95% with the contrast-to-noise ratio values of the DLIR phantom images similar to those reconstructed by filtered back projection, but it is unclear whether these findings could be translated into clinical practice exactly. Nonetheless, Wang et al.’s [14] clinical prospective study showed that their homegrown AI denoising model developed through transfer learning with the use of 17 standard-dose PET simulated 6.25% ultra-low-dose PET, and MRI training datasets were able to reduce the radiation dose by 93.8% for the whole-body PET examinations with adequate diagnostic accuracy. This implies that it is feasible to use the AI denoising to achieve about 90% dose reduction in the clinical practice although all included studies had small sample sizes and/or number of training datasets [1,4,5,6,7,8,9,10,11,12,13,14,15,16], which is a common issue of AI studies in radiology due to limited availability of medical images [53]. Nevertheless, through the use of transfer learning (i.e., retraining an existing AI model using a smaller number of datasets with or without modification of its architecture) to develop an AI model for performing a similar task, such a model could provide a dose-optimization performance comparable to commercially available models (e.g., Canon AiCE, GE TrueFidelity, etc.) trained with more datasets [2,12,14,43].
It is within expectation that all but two studies used the AI models with the deep CNN architecture because the CNN architecture emerged in 1980s, and hence, it has been widely used in radiology, with satisfactory performances well-demonstrated [1,2,4,5,6,8,9,10,11,12,13,14,15,16,37]. However, one included study published in 2022 employed the more recent and advanced DL architecture: GAN, which was designed in 2014 [7,51]. According to a narrative review about the use of GAN in radiology published in 2021 [51], the CNN-based denoising models could cause CT images having a plastic-like appearance, which is similar to those produced by iterative reconstruction due to over-smoothing. In contrast, the GAN is a more complex architecture with a generator and a discriminator, which requires simultaneous training of these two, increasing the complexity of model development [37]. Nonetheless, the GAN-based denoising models are able to preserve texture details and hence produce images with quality matching standard images [51]. The GAN-based dose-optimization study included in this systematic review also demonstrated similar findings that their readers were unable to differentiate between the standard-dose and GAN-processed images although only 36.6% dose reduction was achieved in their study [7]. Another non-CNN-based dose-optimization study included in this review employed the Gaussian mixture model (GMM) architecture [3]. The use of GMM for medical image denoising was reported before the emergence of GAN [54]. However, it is not widely adopted in radiology, and its clinical performance in pediatric radiology dose optimization remains unclear [3,17,43,51].
This paper is the first systematic review on the AI for radiation dose optimization in pediatric radiology covering the imaging modalities of CT, PET/MRI, and mobile radiography and hence advancing the previous narrative review on the AI for dose optimization in pediatric CT published in 2021 [17]. Although it is well-known that radiation dose burden is a significant issue in pediatric CT [1,2,3,4,5,6,7,8,9,10,11,15,16], the dose involved in a PET scan is comparable to that of a CT examination [14]. Furthermore, general radiography is the most common radiological examination type for pediatric patients despite being a low-dose modality [36]. Nonetheless, as per the ALARA principle, the value of AI for dose optimization in other modalities that use ionizing radiation for pediatric examinations should be explored in the future [17,24,25]. Moreover, given the relatively narrow focus and small sample size of the included studies, future studies on this topic area for CT, PET, and general radiography need to have greater scale and wider scope [1,4,5,6,7,8,9,10,11,12,13,14,15,16]. Besides, further exploration of the use of GAN for dose optimization appears warranted [7,51].
This systematic review has two major limitations. Article selection, data extraction, and synthesis were performed by a single author, albeit one with more than 20 years of experience in conducting literature reviews. According to a recent methodological systematic review [44], this is an appropriate arrangement provided that the single reviewer is experienced. Additionally, through adherence to the PRISMA guidelines and the use of the data extraction form (Table 1) devised based on the recent systematic review on AI in radiology and QATSDD, the potential bias should be addressed to certain extent [41,45,46,49]. In addition, only articles written in English and published within last five years were included, potentially affecting comprehensiveness of this systematic review. Nevertheless, this review still has a wider coverage than the previous narrative review on the AI for dose optimization in pediatric CT [17].
5. Conclusions
This systematic review shows that the deep CNN was the most common AI technique and architecture used for radiation dose optimization in pediatric radiology. All but three included studies evaluated the AI performance in dose optimization of abdomen, chest, head, neck, and pelvis CT; CT angiography; and DECT through DLIR. The majority of studies demonstrated that the AI could reduce radiation dose by 36–70% without losing diagnostic information. Despite the dominance of commercially available AI models based on the deep CNN, the homegrown models, including the one with the more recent and advanced architecture, i.e., GAN, could provide comparable performances. Future exploration of the value of AI for dose optimization in pediatric radiology is necessary, as the sample sizes of the included studies appear small, and only three imaging modalities, namely CT, PET/MRI, and mobile radiography, rather than all examination types were covered.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
Not applicable.
Conflicts of Interest
The author declares no conflict of interest.
Funding Statement
This work received no external funding.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Brady S.L., Trout A.T., Somasundaram E., Anton C.G., Li Y., Dillman J.R. Improving image quality and reducing radiation dose for pediatric CT by using deep learning reconstruction. Radiology. 2021;298:180–188. doi: 10.1148/radiol.2020202317. [DOI] [PubMed] [Google Scholar]
- 2.Jeon P.H., Kim D., Chung M.A. Estimates of the image quality in accordance with radiation dose for pediatric imaging using deep learning CT: A phantom study; Proceedings of the 2022 IEEE International Conference on Big Data and Smart Computing (BigComp); Daegu, Korea. 17–22 January 2022; pp. 352–356. [DOI] [Google Scholar]
- 3.Kim S.H., Seo K., Kang S.H., Bae S., Kwak H.J., Hong J.W., Hwang Y., Kang S.M., Choi H.R., Kim G.Y., et al. Study on feasibility for artificial intelligence (AI) noise reduction algorithm with various parameters in pediatric abdominal radio-magnetic computed tomography (CT) J. Magn. 2017;22:570–578. doi: 10.4283/JMAG.2017.22.4.570. [DOI] [Google Scholar]
- 4.Krueger P.C., Ebeling K., Waginger M., Glutig K., Scheithauer M., Schlattmann P., Proquitté H., Mentzel H.J. Evaluation of the post-processing algorithms SimGrid and S-Enhance for paediatric intensive care patients and neonates. Pediatr. Radiol. 2022;52:1029–1037. doi: 10.1007/s00247-021-05279-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Lee S., Choi Y.H., Cho Y.J., Lee S.B., Cheon J.E., Kim W.S., Ahn C.K., Kim J.H. Noise reduction approach in pediatric abdominal CT combining deep learning and dual-energy technique. Eur. Radiol. 2021;31:2218–2226. doi: 10.1007/s00330-020-07349-9. [DOI] [PubMed] [Google Scholar]
- 6.Nagayama Y., Goto M., Sakabe D., Emoto T., Shigematsu S., Oda S., Tanoue S., Kidoh M., Nakaura T., Funama Y., et al. Radiation dose reduction for 80-kVp pediatric CT using deep learning-based reconstruction: A clinical and phantom study. AJR Am. J. Roentgenol. 2022;23:1–10. doi: 10.2214/AJR.21.27255. [DOI] [PubMed] [Google Scholar]
- 7.Park H.S., Jeon K., Lee J., You S.K. Denoising of pediatric low dose abdominal CT using deep learning based algorithm. PLoS ONE. 2022;17:e0260369. doi: 10.1371/journal.pone.0260369. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Sun J., Li H., Li H., Li M., Gao Y., Zhou Z., Peng Y. Application of deep learning image reconstruction algorithm to improve image quality in CT angiography of children with Takayasu arteritis. J. X-ray Sci. Technol. 2022;30:177–184. doi: 10.3233/XST-211033. [DOI] [PubMed] [Google Scholar]
- 9.Sun J., Li H., Li J., Yu T., Li M., Zhou Z., Peng Y. Improving the image quality of pediatric chest CT angiography with low radiation dose and contrast volume using deep learning image reconstruction. Quant. Imaging Med. Surg. 2021;11:3051–3058. doi: 10.21037/qims-20-1158. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Sun J., Li H., Li J., Cao Y., Zhou Z., Li M., Peng Y. Performance evaluation of using shorter contrast injection and 70 kVp with deep learning image reconstruction for reduced contrast medium dose and radiation dose in coronary CT angiography for children: A pilot study. Quant. Imaging Med. Surg. 2021;11:4162–4171. doi: 10.21037/qims-20-1159. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Sun J., Li H., Gao J., Li J., Li M., Zhou Z., Peng Y. Performance evaluation of a deep learning image reconstruction (DLIR) algorithm in “double low” chest CTA in children: A feasibility study. Radiol. Med. 2021;126:1181–1188. doi: 10.1007/s11547-021-01384-2. [DOI] [PubMed] [Google Scholar]
- 12.Sun J., Li H., Wang B., Li J., Li M., Zhou Z., Peng Y. Application of a deep learning image reconstruction (DLIR) algorithm in head CT imaging for children to improve image quality and lesion detection. BMC Med. Imaging. 2021;21:108. doi: 10.1186/s12880-021-00637-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Theruvath A.J., Siedek F., Yerneni K., Muehe A.M., Spunt S.L., Pribnow A., Moseley M., Lu Y., Zhao Q., Gulaka P., et al. Validation of deep learning-based augmentation for reduced 18F-FDG dose for PET/MRI in children and young adults with lymphoma. Radiol. Artif. Intell. 2021;3:e200232. doi: 10.1148/ryai.2021200232. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Wang Y.J., Baratto L., Hawk K.E., Theruvath A.J., Pribnow A., Thakor A.S., Gatidis S., Lu R., Gummidipundi S.E., Garcia-Diaz J., et al. Artificial intelligence enables whole-body positron emission tomography scans with minimal radiation exposure. Eur. J. Nucl. Med. Mol. Imaging. 2021;48:2771–2781. doi: 10.1007/s00259-021-05197-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Yoon H., Kim J., Lim H.J., Lee M.J. Image quality assessment of pediatric chest and abdomen CT by deep learning reconstruction. BMC Med. Imaging. 2021;21:146. doi: 10.1186/s12880-021-00677-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Zhang K., Shi X., Xie S.S., Sun J.H., Liu Z.H., Zhang S., Song J.Y., Shen W. Deep learning image reconstruction in pediatric abdominal and chest computed tomography: A comparison of image quality and radiation dose. Quant. Imaging Med. Surg. 2022;12:3238–3250. doi: 10.21037/qims-21-936. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Nagayama Y., Sakabe D., Goto M., Emoto T., Oda S., Nakaura T., Kidoh M., Uetani H., Funama Y., Hirai T. Deep learning-based reconstruction for lower-dose pediatric CT: Technical principles, image characteristics, and clinical implementations. Radiographics. 2021;41:1936–1953. doi: 10.1148/rg.2021210105. [DOI] [PubMed] [Google Scholar]
- 18.Pearce M.S., Salotti J.A., Little M.P., McHugh K., Lee C., Kim K.P., Howe N.L., Ronckers C.M., Rajaraman P., Sir Craft A.W., et al. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: A retrospective cohort study. Lancet. 2012;380:499–505. doi: 10.1016/S0140-6736(12)60815-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.de Gonzalez A.B., Salotti J.A., McHugh K., Little M.P., Harbron R.W., Lee C., Ntowe E., Braganza M.Z., Parker L., Rajaraman P., et al. Relationship between paediatric CT scans and subsequent risk of leukaemia and brain tumours: Assessment of the impact of underlying conditions. Br. J. Cancer. 2016;114:388–394. doi: 10.1038/bjc.2015.415. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Lee K.H., Lee S., Park J.H., Lee S.S., Kim H.Y., Lee W.J., Cha E.S., Kim K.P., Lee W., Lee J.Y., et al. Risk of hematologic malignant neoplasms from abdominopelvic computed tomographic radiation in patients who underwent appendectomy. JAMA Surg. 2021;156:343–351. doi: 10.1001/jamasurg.2020.6357. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Mathews J.D., Forsythe A.V., Brady Z., Butler M.W., Goergen S.K., Byrnes G.B., Giles G.G., Wallace A.B., Anderson P.R., Guiver T.A., et al. Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: Data linkage study of 11 million Australians. BMJ. 2013;346:f2360. doi: 10.1136/bmj.f2360. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Halm B.M., Franke A.A., Lai J.F., Turner H.C., Brenner D.J., Zohrabian V.M., DiMauro R. γ-H2AX foci are increased in lymphocytes in vivo in young children 1 h after very low-dose X-irradiation: A pilot study. Pediatr. Radiol. 2014;44:1310–1317. doi: 10.1007/s00247-014-2983-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Vandevoorde C., Franck C., Bacher K., Breysem L., Smet M.H., Ernst C., De Backer A., Van De Moortele K., Smeets P., Thierens H. γ-H2AX foci as in vivo effect biomarker in children emphasize the importance to minimize X-ray doses in paediatric CT imaging. Eur. Radiol. 2015;25:800–811. doi: 10.1007/s00330-014-3463-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Ng C.K.C., Sun Z. Development of an online automatic computed radiography dose data mining program: A preliminary study. Comput. Methods Programs Biomed. 2010;97:48–52. doi: 10.1016/j.cmpb.2009.07.001. [DOI] [PubMed] [Google Scholar]
- 25.MacKay M., Hancy C., Crowe A., D’Rozario R., Ng C.K.C. Attitudes of medical imaging technologists on use of gonad shielding in general radiography. Radiographer. 2012;59:35–39. doi: 10.1002/j.2051-3909.2012.tb00172.x. [DOI] [Google Scholar]
- 26.Ng C.K.C., Sun Z. Development of an online automatic diagnostic reference levels management system for digital radiography: A pilot experience. Comput. Methods Programs Biomed. 2011;103:145–150. doi: 10.1016/j.cmpb.2010.07.008. [DOI] [PubMed] [Google Scholar]
- 27.Ng C.K.C., Sun Z., Parry H., Burrage J. Local diagnostic reference levels for x-ray examinations in an Australian tertiary hospital. J. Med. Imaging Health Inform. 2014;4:297–302. doi: 10.1166/jmihi.2014.1250. [DOI] [Google Scholar]
- 28.Sun Z., Ng C.K.C., Wong Y.H., Yeong C.H. 3D-printed coronary plaques to simulate high calcification in the coronary arteries for investigation of blooming artifacts. Biomolecules. 2021;11:1307. doi: 10.3390/biom11091307. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Sun Z., Ng C.K.C., Sá Dos Reis C. Synchrotron radiation computed tomography versus conventional computed tomography for assessment of four types of stent grafts used for endovascular treatment of thoracic and abdominal aortic aneurysms. Quant. Imaging Med. Surg. 2018;8:609–620. doi: 10.21037/qims.2018.07.05. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Sun Z., Ng C.K.C., Squelch A. Synchrotron radiation computed tomography assessment of calcified plaques and coronary stenosis with different slice thicknesses and beam energies on 3D printed coronary models. Quant. Imaging Med. Surg. 2019;9:6–22. doi: 10.21037/qims.2018.09.11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Sun Z., Ng C.K.C. Use of synchrotron radiation to accurately assess cross-sectional area reduction of the aortic branch ostia caused by suprarenal stent wires. J. Endovasc. Ther. 2017;24:870–879. doi: 10.1177/1526602817732315. [DOI] [PubMed] [Google Scholar]
- 32.Sun Z., Ng C.K.C. Synchrotron radiation imaging of aortic stent grafting: An in vitro phantom study. J. Med. Imaging Health Inform. 2017;7:890–896. doi: 10.1166/jmihi.2017.2094. [DOI] [Google Scholar]
- 33.Al Mahrooqi K.M.S., Ng C.K.C., Sun Z. Pediatric computed tomography dose optimization strategies: A literature review. J. Med. Imaging Radiat. Sci. 2015;46:241–249. doi: 10.1016/j.jmir.2015.03.003. [DOI] [PubMed] [Google Scholar]
- 34.Sun Z., Ng C. Dual-source CT angiography in aortic stent grafting: An in vitro aorta phantom study of image noise and radiation dose. Acad. Radiol. 2010;17:884–893. doi: 10.1016/j.acra.2010.03.004. [DOI] [PubMed] [Google Scholar]
- 35.Almutairi A.M., Sun Z., Ng C., Al-Safran Z.A., Al-Mulla A.A., Al-Jamaan A.I. Optimal scanning protocols of 64-slice CT angiography in coronary artery stents: An in vitro phantom study. Eur. J. Radiol. 2010;74:156–160. doi: 10.1016/j.ejrad.2009.01.027. [DOI] [PubMed] [Google Scholar]
- 36.Feghali J.A., Chambers G., Delépierre J., Chapeliere S., Mannes I., Adamsbaum C. New image quality and dose reduction technique for pediatric digital radiography. Diagn. Interv. Imaging. 2021;102:463–470. doi: 10.1016/j.diii.2021.01.009. [DOI] [PubMed] [Google Scholar]
- 37.Sun Z., Ng C.K.C. Artificial intelligence (enhanced super-resolution generative adversarial network) for calcium deblooming in coronary computed tomography angiography: A feasibility study. Diagnostics. 2022;12:991. doi: 10.3390/diagnostics12040991. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Sun Z., Ng C.K.C. High calcium scores in coronary CT angiography: Effects of image post-processing on visualization and measurement of coronary lumen diameter. J. Med. Imaging Health Inform. 2015;5:110–116. doi: 10.1166/jmihi.2015.1366. [DOI] [Google Scholar]
- 39.Sun Z., Ng C.K.C., Xu L., Fan Z., Lei J. Coronary CT angiography in heavily calcified coronary arteries: Improvement of coronary lumen visualization and coronary stenosis assessment with image postprocessing methods. Medicine. 2015;94:e2148. doi: 10.1097/MD.0000000000002148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Christie S., Ng C.K.C., Sá Dos Reis C. Australasian radiographers’ choices of immobilisation strategies for paediatric radiological examinations. Radiography. 2020;26:27–34. doi: 10.1016/j.radi.2019.06.003. [DOI] [PubMed] [Google Scholar]
- 41.PRISMA: Transparent Reporting of Systematic Reviews and Meta-Analyses. [(accessed on 24 June 2022)]. Available online: https://www.prisma-statement.org.
- 42.Eriksen M.B., Frandsen T.F. The impact of patient, intervention, comparison, outcome (PICO) as a search strategy tool on literature search quality: A systematic review. J. Med. Libr. Assoc. 2018;106:420–431. doi: 10.5195/jmla.2018.345. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Choy G., Khalilzadeh O., Michalski M., Do S., Samir A.E., Pianykh O.S., Geis J.R., Pandharipande P.V., Brink J.A., Dreyer K.J. Current applications and future impact of machine learning in radiology. Radiology. 2018;288:318–328. doi: 10.1148/radiol.2018171820. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Waffenschmidt S., Knelangen M., Sieben W., Bühn S., Pieper D. Single screening versus conventional double screening for study selection in systematic reviews: A methodological systematic review. BMC Med. Res. Methodol. 2019;19:132. doi: 10.1186/s12874-019-0782-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Jeong J.J., Tariq A., Adejumo T., Trivedi H., Gichoya J.W., Banerjee I. Systematic review of generative adversarial networks (GANs) for medical image classification and segmentation. J. Digit. Imaging. 2022;35:137–152. doi: 10.1007/s10278-021-00556-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Ng C.K.C. A review of the impact of the COVID-19 pandemic on pre-registration medical radiation science education. Radiography. 2022;28:222–231. doi: 10.1016/j.radi.2021.07.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Petri S.A., Ng C.K.C. Comparison of the performance of computed radiography and direct radiography in glass soft tissue foreign body visualisation. S. Afr. Radiogr. 2018;56:18–25. [Google Scholar]
- 48.Kleinfelder T.R., Ng C.K.C. Effects of image postprocessing in digital radiography to detect wooden, soft tissue foreign bodies. [(accessed on 28 June 2022)];Radiol. Technol. 2022 93:544–554. Available online: https://pubmed.ncbi.nlm.nih.gov/35790309/ [PubMed] [Google Scholar]
- 49.Sirriyeh R., Lawton R., Gardner P., Armitage G. Reviewing studies with diverse designs: The development and evaluation of a new tool. J. Eval. Clin. Pract. 2012;18:746–752. doi: 10.1111/j.1365-2753.2011.01662.x. [DOI] [PubMed] [Google Scholar]
- 50.Lim B., Son S., Kim H., Nah S., Lee K.M. Enhanced deep residual networks for single image super-resolution; Proceedings of the 2017 IEEE Conference on Computer Vision and Pattern Recognition Workshops (CVPRW); Honolulu, HI, USA. 21–26 July 2017; pp. 1132–1140. [DOI] [Google Scholar]
- 51.Wolterink J.M., Mukhopadhyay A., Leiner T., Vogl T.J., Bucher A.M., Išgum I. Generative adversarial networks: A primer for radiologists. Radiographics. 2021;41:840–857. doi: 10.1148/rg.2021200151. [DOI] [PubMed] [Google Scholar]
- 52.Wolterink J.M., Leiner T., Viergever M.A., Isgum I. Generative adversarial networks for noise reduction in low-dose CT. IEEE Trans. Med. Imaging. 2017;36:2536–2545. doi: 10.1109/TMI.2017.2708987. [DOI] [PubMed] [Google Scholar]
- 53.Kimy H.E., Cosa-Linan A., Santhanam N., Jannesari M., Maros M.E., Ganslandt T. Transfer learning for medical image classification: A literature review. BMC Med. Imaging. 2022;22:69. doi: 10.1186/s12880-022-00793-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Garg G., Prasad G., Coyle D. Gaussian mixture model-based noise reduction in resting state fMRI data. J. Neurosci. Methods. 2013;215:71–77. doi: 10.1016/j.jneumeth.2013.02.015. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Not applicable.