Skip to main content
The British Journal of Radiology logoLink to The British Journal of Radiology
. 2022 Jul 8;94(1126):20201350. doi: 10.1259/bjr.20201350

Clinical evaluation of data-driven respiratory gating for PET/CT in an oncological cohort of 149 patients: impact on image quality and patient management

Michael Messerli 1,2,1,2,, Virginia Liberini 1, Hannes Grünig 1,2,1,2, Alexander Maurer 1,2,1,2, Stephan Skawran 1,2,1,2, Niklas Lohaus 2,3,2,3, Lars Husmann 1,2,1,2, Erika Orita 1, Josephine Trinckauf 1, Philipp A Kaufmann 1,2,1,2, Martin W Huellner 1,2,1,2
PMCID: PMC9328056  PMID: 34520673

Abstract

Objectives:

To evaluate the impact of fully automatic motion correction by data-driven respiratory gating (DDG) on positron emission tomography (PET) image quality, lesion detection and patient management.

Materials and Methods:

A total of 149 patients undergoing PET/CT for cancer (re-)staging were retrospectively included. Patients underwent a PET/CT on a digital detector scanner and for every patient a PET data set where DDG was enabled (PETDDG) and as well as where DDG was not enabled (PETnonDDG) was reconstructed. All PET data sets were evaluated by two readers which rated the general image quality, motion effects and organ contours. Further, both readers reviewed all scans on a case-by-case basis and evaluated the impact of PETDDG on additional apparent lesion, change of report, and change of management.

Results:

In 85% (n = 126) of the patients, at least one bed position was acquired using DDG, resulting in mean scan time increase of 4:37 min per patient in the whole study cohort (n = 149). General image quality was not rated differently for PETnonDDG and PETDDG images (p = 1.000) while motion effects (i.e. indicating general blurring) was rated significantly lower in PETDDG images and organ contours, including liver and spleen, were rated significantly sharper using PETDDG as compared to PETnonDDG (all p < 0.001). In 27% of patients, PETDDG resulted in a change of the report and in a total of 12 cases (8%), PETDDG resulted in a change of further clinical management.

Conclusion:

Deviceless DDG provided reliable fully automatic motion correction in clinical routine and increased lesion detectability and changed management in a considerable number of patients.

Advances in knowledge:

DDG enables PET/CT with respiratory gating to be used routinely in clinical practice without external gating equipment needed.

Introduction

Positron emission tomography (PET) images radiotracer uptake in patients, thereby enabling the detection and quantification of physiologic and pathophysiological processes in vivo.1 In oncological patients, 18F-fludeoxyglucose (18F-FDG) PET is used to assess increased glucose metabolism in primary tumors, lymph node or distant metastases,2,3 whereas CT enables a detailed assessment of morphological disease extent with a high spatial resolution.4,5 Therefore, hybrid imaging with 18F-FDG PET/CT has evolved as an important tool for staging of cancer patients in many different etiologies.3 Further substances tracing prostate-specific membrane antigen (PSMA) and DOTA-peptides (DOTATATE) play also an important role in PET imaging, namely for the assessment of prostate cancer and neuroendocrine tumors.6–8

However, PET image quality may be significantly degraded by respiratory motion, among other causes.9 The radiotracer distribution appears blurred on images affected by motion, rendering exams suboptimal for diagnostic purposes and/or quantitative assessment, particularly in the chest and upper abdomen.10 To overcome this issue (i.e. to decrease the effect of motion), respiratory gating can be applied. This requires the acquisition of a gating signal, which is usually done by external devices, such as a pressure belt or a video camera tracing infrared markers, both of which may be inconvenient for the patient, and may also prolong the examination time owing to device mounting and unmounting.11 Also, a single misregistration may spoil motion correction of the exam, in fact an erroneous misregistration of the underlying static attenuation maps may cause an over- or underestimation of activity of the region. A respiratory gating signal may also be extracted from the acquired PET raw data using different methods, generally referred to as data-driven gating (DDG).12,13 Until recently, this method was mainly reserved to academic environments and could only be used retrospectively (i.e. based on list-mode raw data) and not under real-world conditions while scanning patients,14 although nowadays DDG has become available on the market and is becoming more widespread.

Recently, a new digital PET detector system with lutetium-based scintillator crystal arrays and silicon photomultipliers was introduced. A first study carried out in a mixed population of cancer patients showed an improved performance of digital PET/CT with regard to pathologic and physiologic structures.15 This may potentially be further enhanced by the use of DDG.

Accordingly, the purpose of our study was to evaluate the impact of DDG using the latest generation digital PET/CT scanner with regard to PET (a) image quality, (b) lesion detection and (c) patient management in an unselected oncological cohort.

Methods and materials

Patients

This retrospective study included consecutive patients (n = 149) who underwent a clinical PET/CT for (re-)staging of oncological diseases during a pilot evaluation phase of 6 weeks at the Department of Nuclear Medicine of the University Hospital of Zurich who gave general written informed consent for the use of their data for scientific purposes. Overall, 132 patients (89%) underwent an 18F-FDG PET/CT scan and 17 (11%) underwent a non-FDG PET/CT scan (n = 7 PSMA; n = 9 DOTATATE; n = 1 DOPA), Table 1. The local ethics committee approved the study protocol (Trial-No. BASEC 2019-00207).

Table 1.

Demographic data of study subjects (n = 149)

Female/male, n (%) 61 (41%) / 88 (59%)
Age, years 61 ± 15 (24 - 89)
Body weight, kg 76 ± 18 (40 - 142)
Body height, m 1.72 ± 0.1 (1.50–1.95)
BMI, kg/m2 25.6 ± 5.5 (14.9–43.4)
Injected dose, MBq 190 ± 74 (88 - 317)
Type of primary disease
 Head and neck cancer 14 (9.4%)
 Breast cancer 15 (10.1%)
 Lung cancer 19 (12.8%)
 Esophageal cancer 3 (2.0%)
 Small bowel cancer 0 (0%)
 Colon cancer 8 (5.4%)
 Pancreatic cancer 7 (4.7%)
 Cholangiocarcinoma 4 (2.7%)
 Urogenital cancer 18 (12.1%)
 Melanoma 27 (18.1%)
 Lymphoma 12 (8.1%)
 Cancer of unknown origin 3 (2.0%)
 Neuroendocrine tumor 10 (6.7%)
 Paraneoplastic syndrome 9 (6.0%)
Tracer used
 FDG 132 (88.6%)
 PSMA 7 (4.7%)
 DOTATATE 9 (6.0%)
 DOPA 1 (0.7%)

BMI = body mass index; MBq = Mega-Becquerel; PET = positron emission tomography.

Values are given as absolute numbers and percentages in parenthesis or mean ± standard deviation (range).

PET/CT imaging protocol

All patients underwent a PET/CT on a digital detector scanner (GE Discovery Molecular Insights - DMI PET/CT, GE Healthcare, Waukesha, WI) with a high intrinsic system sensitivity of 22 cps/kBq.

132 patients (89%) underwent an 18F-FDG PET with a body mass index (BMI)-adapted 18F-FDG dosage regimen, as previously described16 and presented in Table 2. Participants fasted for at least 4 h prior to the scan, and blood glucose levels were below 160 mg dl−1 at the time of the 18F-FDG injection. A free-breathing CT scan was performed from the vertex of the skull to the mid-thighs (or feet), and used for attenuation correction purposes as well as for anatomic localization of 18F-FDG uptake. The CT scan was performed using automated dose modulation (range 15–100 mA, 120 kV). Immediately after the CT scan, PET images were acquired covering the identical anatomical region. The PET acquisition time was set to 2.5 min per bed position, with 6–11 bed positions per patient (depending on patient size), with an overlap of 23%. All PET images were reconstructed using a Bayesian penalized likelihood reconstruction method (Q.Clear, GE Healthcare, Waukesha, WI). Data sets were reconstructed with a 256 × 256 pixel matrix. Spatiel resolution were radial 1.62 mm, tangential 2.30 mm and axial 3.09 mm, as assessed by full width at half maximum (FWHM) measurements at 1 cm from center of field of view were.17

Table 2.

PET imaging protocols for the tracers used in the study group (n = 149)

Tracer Scan time p.i. Injected activity Bed time Reconstruction setting
FDG 60 min BMI-adapteda 2.5 min Q.Clear (β-value of 450)
PSMA 90 min 3 MBq/Kgb 2.5 min Q.Clear (β-value of 1200)
DOTA 60 min 150 MBq 2.0 min Q.Clear (β-value of 1000)
DOPA 60 min 150 MBq 2.5 min Q.Clear (β-value of 1000)

BMI, body mass index; FDG = Fludeoxyglucose; MBq = Mega-Becquerel; PET = Positron emission tomography; PSMA = prostate-specific membrane antigen.

a

FDG-dose of 1.5 MBq/kg body weight was injected for patients with a BMI of <20 kg/m2, 2 MBq/kg body weight for patients with a BMI of 20–24.5 kg/m2, and 3.1 MBq/kg body weight for patients with a BMI >24.5 kg/m2 (without exceeding a maximum injected FDG dose of 320 MBq).

b

minimum = 200 MBq, maximum = 300 MBq.

17 patients (11%) underwent a PET/CT scan with a tracer other than 18F-FDG. The imaging protocols used for PSMA, DOTATATE and DOPA-imaging were described previously,18 and are presented in Table 2.

Use of DDG for PET acquisition

In our study, we used a software-based prospective DDG algorithm (MotionFree, GE Healthcare, Waukesha, WI) that was recently introduced on DMI PET/CT, in combination with a motion correction algorithm (Q.Static, GE Healthcare, Waukesha, WI) that utilizes the quiescent phase of the respiratory cycle. Unlike externally driven respiratory gating, which usually relies on infrared camera tracking of chest motion, DDG methods solely use PET raw data in combination with dimensionality reduction techniques, in order to extract the respiratory signal. This technique utilizes a principal component analysis to compute the spatiotemporal variation of list mode data. The algorithm measures respiration-like frequencies within the PET data. At the end of base acquisition time, for each bed position for which motion screening was prescribed, an on-the-fly decision is made as whether motion has been detected. The whole body DDG reconstructed image is composed of a non-gated part for bed positions that were not triggered, and a quiescent gated part on the bed positions that were triggered; with the non-gated bed positions being 2.5 min in duration, and the gated positions being 5 min (i.e. doubling the per-bed acquisition time) in duration but with only 50% of coincidences retained, namely those occurring in the defined quiescent periods. The algorithm provides a signal-to-noise measure of respiration-like frequencies within the data, denoted as R-value, that is configurable (R-value threshold). The R-factor (i.e. signal-to-noise ratio of respiration-like frequencies within the data operating as threshold for triggering) was set to R > 10.0.19,20

For further analysis, for every patient a PET data set where DDG was enabled (hereafter referred to as PETDDG) and another data set where DDG was not enabled (hereafter referred to as PETnonDDG) was reconstructed; PETnonDDG was processed using a non-gated reconstruction using the first 2.5 min per bed position, with the additional 2.5 min of data that were collected on DDG-triggered bed positions discarded.

Subjective PET image analysis and assessment of clinical impact of DDG

All PET data sets were evaluated by two readers (M.M. and M.W.H., with 3 and 8 years of experience in PET/CT reading, respectively) blinded to the acquisition mode used. All scans were reviewed independently on a dedicated workstation (Advantage Workstation, v. 4.6; GE Healthcare, Waukesha, WI) and in random order. Readers were blinded to the clinical information. In case of discrepancy of image rating, a final decision was reached by consensus.

The readers first rated the general image quality, motion effects and organ contours. For this purpose, data sets were viewed using maximum intensity projection (MIP) views of PET and axial views with reformatted sections. The two readers evaluated the images according the scores described in Table 3. The frequency of attenuation correction artifacts in PETnonDDG and PETDDG was assessed by noting the presence of curvilinear cold artifacts on dome of liver/diaphragm or at lung base because of respiration mismatch on PET images with CT attenuation correction.

Table 3.

Subjective image quality scores

Overall image quality 1 = poor
2 = acceptable
3 = good
4 = very good
Motion effects 1 = no
2 = slight
3 = moderate
4 = severe
Organ contoursa 1 = ill-defined
2 = moderate blurring
3 = slight blurring
4 = sharp
a

Organ contour assessment included liver, spleen, stomach, kidney, collecting system, bowel, heart.

In a second session performed 8 week after initial scanning, both readers reviewed all scans in consensus on a case-by-case basis. Decisions were made for every patient as (a) whether additional lesions were apparent using PETDDG, (b) whether the use of PETDDG led to a change of report, and (c) whether PETDDG resulted in a change of management according to the institutional protocols that is based on international, interdisciplinary guidelines.

Statistical analyses

Categorical variables are expressed as proportions, and continuous variables are presented as mean ± standard deviation or median (range), depending on the distribution of values. Qualitative image ratings (i.e. overall image quality, motion artefacts, and organ contours) were compared of PETnonDDG and PETDDG reconstruction using the Wilcoxon matched pairs signed rank test. Analyses were carried out using SPSS release 25.0 (IBM Corporation, Armonk, NY) and MedCalc v. 15.8 (MedCalc Software, Ostend, Belgium). A two-tailed p-value of <0.05 was considered to indicate statistical significance.

Results

A total of 149 patients (61 female, 88 male, mean age 61 ± 15 years) referred for the initial staging (n = 41) or restaging (n = 108) of different oncological diseases undergoing PET/CT participated in our study. Detailed demographic information is presented in Table 1.

Technical aspects of DDG in study cohort

The results of the technical aspects using DDG in the study cohort are given in Table 4. Overall, in 85% of the patients at least one bed was acquired using DDG, leading to an average scan time increase of 4:37 min in the whole study cohort (n = 149), and 5:28 min in the subgroup of patients whose exams were triggered by DDG (n = 126).

Table 4.

Technical aspects of DDG in study group

Gating yes / gating no 126 (84.6%) / 23 (15.4%)
 Gating yes FDG 109/132 (83%)
 Gating yes PSMA 7/7 (100%)
 Gating yes DOTATATE 9/9 (100%)
 Gating yes DOPA 1/1 (100%)
Number of bed positions acquired 6 ± 1 (5–11)
Amount of bed positions triggered 1.69 ± 0.1 (1.48–1.94)
 no gating 23 (15%)
 one bed triggered 18 (12%)
 two beds triggered 68 (46%)
 three beds triggered 32 (21%)
 four beds triggered 8 (5%)
Total scan time, min 15:16 ± 3:02 (10:00 - 27:30)
Total time increase per study in whole cohorta, min 4:37 ± 3:02 (0:00 - 10:00)
Total time increase per study in DDG cohortb, min 5:28 ± 1:52 (2:30 - 10:00)

DDG, Data-driven gating.

Values are given as absolute numbers and percentages in parenthesis or mean ± standard deviation (range).

a

n = 149.

b

n = 126.

Impact of DDG on subjective image quality

The results of the subjective image assessment including all study subjects are given in Table 5. General image quality was not rated differently for PETnonDDG and PETDDG images (p = 1.000). Motion effects (i.e. indicating general blurring) were rated significantly lower in PETDDG images and contours of different organs, including liver and spleen, were rated sharper using PETDDG as compared to PETnonDDG (all p < 0.001), Table 5. The improvement of organ contours was most pronounced for the liver, spleen, heart and collecting system, and less pronounced for the stomach, kidney and bowel, see Table 5. Overall, there were nine cases of attenuation correction artefacts in PETDDG as compared to seven cases in PETnonDDG, respectively.

Table 5.

Subjective images analysis comparing

PETnonDDG PETDDG p-value
Overall image qualitya 3.8 3.8 1.000
Motion effectsb 1.2 1.0 <0.001
Organ contoursc
 Liver 3.5 3.8 <0.001
 Spleen 3.6 3.9 <0.001
 Stomach 3.8 3.9 <0.001
 Kidney 3.8 4.0 <0.001
 Collecting system 3.6 4.0 <0.001
 Bowel 3.6 3.8 <0.001
 Heart 3.5 3.8 <0.001
Mean increase in organ contour rating, %
 Liver reference +9.7%
 Spleen reference +10.6%
 Stomach reference +3.1%
 Kidney reference +6.5%
 Collecting system reference +12.8%
 Bowel reference +6.3%
 Heart reference +10.0%

DDG = Data-driven gating; PET = Positron emission tomography.

a

Overall image quality scale: 1, poor; 2, acceptable; 3, good; 4, very good.

b

Motion effects scale: 1, no; 2, slight; 3, moderate; 4, severe.

c

Organ contours: 1, ill-defined; 2, moderate blurring; 3, slight blurring; 4, sharp.

Impact of DDG on lesion detectability and clinical management

Overall, in 40/149 patients (27%) PETDDG resulted in a change of the report, mainly due to additional apparent lesions, Figure 1. When reviewing all cases, we observed that in a total of 12 cases (8%), PETDDG resulted in a change of further clinical management, Table 6.

Figure 1.

Figure 1.

Impact of DDG in the study cohort (n = 149) on lesion detectability and patient management. DDG, data-driven respiratory gating.

Table 6.

Cases (12/149) of patients where DDG led to change in management

Patient Tracer Indication for PET Advantage of DDG Change in management
#01 FDG Paraneoplastic inflammatory syndrome Detection of increased uptake in adrenal gland Initiation of laboratory work-up and hormone substitution
#02 FDG Follow-up of pancreatic carcinoma Liver metastasis detected Upstaging and change of therapy regimen
#03 DOTA Staging of pancreatic neuroendocrine tumor Correct identification of intrapancreatic accessory spleen Biopsy obviated
#04 FDG Staging of pancreatic carcinoma Additional lymph node detected Alteration of surgical approach
#05 DOPA Search for pheochromocytoma Detection of small pheochromocytoma Surgical resection
#06 FDG Follow-up of breast cancer Focal lesion in breast detected Initiation of breast ultrasound and biopsy
#07 FDG Follow-up of chorioncarcinoma Liver metastasis detected Change of therapy regimen
#08 FDG Follow-up of nasal carcinoma Additional lymph node detected Radiotherapy-field adapted
#09 FDG Staging of cerebral lymphoma Focal lesion in liver detected MRI of liver performed
#10 FDG Follow-up of thyroid cancer Homogenous liver uptake (focal lesion in non-DDG images) Refrain from additional work-up
#11 DOTA Staging of gastric neuroendocrine tumor Lymph node metastasis detected Upstaging and change of therapy regimen
#12 DOTA Staging of familial MEN1 Retroperitoneal NET detected Upstaging

DDG = Data-driven gating, FDG = Fludeoxyglucose, PET = Positron emission tomography.

A representative case illustrating an additional apparent lesion with PETDDG is presented in Figure 2. Representative cases with change of clinical management by using PETDDG for 18F-FDG, DOTATATE, and 18F-DOPA, respectively, are presented in Figures 3–5.

Figure 2.

Figure 2.

Representative case of 18F-FDG-PET/CT imaging of a 62-year-old female with breast cancer, using a digital PET/CT system without respiratory gating (a–d) as well as using DDG;(e–h) . An additional apparent lesion in the liver (arrow) was noted in PET images using DDG leading to a change of report but in this case not clinical management as there are multiple other liver metastasis present; (SUVmax in PETnonDDG = 3.2; and SUVmax in PETDDG = 6.1). DDG, data-driven respiratory gating; FDG, fludeoxyglucose; PET, positron emission tomography; SUV standardized uptake value.

Figure 3.

Figure 3.

Representative case of 18F-FDG-PET/CT imaging of a 37-year-old female (corresponding to patient #7 of Table 5) with chorioncarcinoma undergoing PET/CT for follow-up. PET/CT images without respiratory gating (a–c) as well as using (d–f). A tiny focus of increased FDG-uptake in the liver is noted (arrow) only in PET images using DDG compatible with a liver metastasis (SUVmax in PETnonDDG = 3.7; and SUVmax in PETDDG = 4.6). DDG, data-driven respiratory gating; FDG, fludeoxyglucose; PET, positron emission tomography; SUV standardized uptake value.

Figure 4.

Figure 4.

Representative case of 68Ga-DOTATATE-PET/CT imaging of a 60-year-old male (corresponding to patient #11 of Table 5) with neuroendocrine tumor of the stomach undergoing PET/CT for follow-up. PET/CT images without respiratory gating (a–c) as well as using (d–f). A new retroperitoneal lymph node metastasis was detected (arrow) only in PET images using DDG; with magnified images of the retroperitoneal region (asterisk), (SUVmax in PETnonDDG = 5.7; and SUVmax in PETDDG = 8.5). DDG, data-driven respiratory gating; FDG, fludeoxyglucose; PET, positron emission tomography; SUV standardized uptake value.

Figure 5.

Figure 5.

Representative case of 18F-DOPA-PET/CT imaging of a 48-year-old female (corresponding to patient #5 of Table 5) with previous pheochromocytoma on the left side that was surgically resected now undergoing PET/CT search for pheochromocytoma on the right side. PET/CT images without respiratory gating (a–d) as well as using (e–h). A tiny focus of increased DOPA-uptake in the right adrenal gland is noted (arrow) and the patient underwent surgical resection of the adrenal where a small pheochromocytoma was proven, (SUVmax in PETnonDDG = 5.5; and SUVmax in PETDDG = 10.3). DDG, data-driven respiratory gating; FDG, fludeoxyglucose; PET, positron emission tomography; SUV standardized uptake value.

Discussion

This study sought to evaluate the impact of DDG (i.e. software-based respiratory PET gating) on image quality, lesion detection and patient management in an unselected oncological cohort using a latest generation silicon-based digital detector PET/CT scanner.

The major findings of our study are as follows: (1) the real-time use of DDG was feasible in clinical routine during a test period of 6 weeks; (2) DDG was activated in 85% of the patients, resulting in a mean time increase by 4:37 min (mean total scan time 15:16 min); (3) while general image quality was not affected by DDG, its use resulted in a significant reduction of motion artifacts and sharper delineation of organs (particularly liver and spleen); and (4) the use of DDG resulted in additional apparent lesions in 27% of patients and (5) the use of DDG translated into a change of management in 8% of patients. Following the results of our study, we have implemented DDG in our clinical routine and use it in our standard protocols.

PET/CT has evolved as an invaluable tool for staging of cancer patients in many different etiologies.3 Respiratory motion during image acquisition may, however, significantly degrade PET image quality and even render the detectability of small lung or liver lesions impossible.9,10,21 To decrease the effect of motion, respiratory gating can be used. This requires the acquisition of a gating signal, usually by external devices, such as a pressure belt or a video camera, which may be inconvenient for the patient and technically challenging for technicians.11

Contrarily, software-based DDG is able to detect respiratory motion within PET data, using the static phase for reconstruction, without the need for additional hardware or inreased radiation dose.22 Indeed, the DDG software used in our study was introduced to our clinical routine from the beginning without any specific patient preparation or instruction used, and after minimal training of staff. Using an R-factor of >10.0, the fully automated respiratory gating method was activated in the majority of patients (i.e. in 85%). This prolonged the scan time by 4:37 min in the whole cohort (up to 10 min in single subjects). One may argue that this is a relevant time increase which potentially impacts on clinical workflow. However, given the significant reduction of motion artifacts, improved visualization of various organ and particularly the amount of additional apparent lesions, we believe this approach may merit consideration.

Previous studies already reported an advantage of respiratory gating for PET/CT.23–25 Guerra et al, e.g. described in a multicenter, retrospective study the benefits in reporting accuracy and quantification of lung lesions by increasing the overall accuracy in lung lesion detection and characterization.25 In another multicenter, retrospective study Crivellaro et al reported respiratory-gated PET/CT technique as a valuable clinical tool in diagnosing liver lesions by reducing undetermined findings, improving diagnostic accuracy, and confidence in reporting.24 Further, benefits from PET respiratory gating in a prospective study of 74 patients were described by Büther et al.23; improvements in diagnostic image quality, increase in SUV, and decrease in metabolic volumes were reported with gated PET images as compared to non-gated PET images.

More recently, two different prospective studies aimed to directly compare the data-driven method to an external device-based system for the respiratory gating in PET. In a cohort of 56 patients undergoing whole-body 18F-FDG PET, Buther et al26 found that DDG-based motion-correction delivered qualities comparable to hardware-based approaches, even if SUV measurements were significantly higher in both reconstruction compared to static images (p < 0.001). On a larger cohort of 144 whole-body 18F-FDG PET examinations, Walker et al20 found that the use of DDG resulted in an increase in lesions’ SUVmax (i.e. 0.66 ± 0.1 g ml−1, p < 0.0005) and a reduction in lesions’ PET volume compared to the external device-based system. Moreover, while the device-based gated reconstruction failed in 16% of exams, the ones using DDG always provided a clinically acceptable image. In another recent study by Walker et al, the authors concluded that DDG provided superior performances to that of an external device-based system.27 We observed a slightly higher rate of attenuation correction artifacts in DDG compared to non-gated images (9 cases vs to 7 cases), although not statistically significant, this is an issue that should be investigated further.

We acknowledge that our study has some limitations. First, this current evaluation of DDG included an unselected oncological cohort of consecutive patients which resulted in a relatively heterogenous study population. The impact of DDG in different diseases and different tracers used may vary, especially as different radiotracers have different distribution properties and are thus expected to result in more or less successful gating outcomes. We therefore strongly ecourage future studies to systematicelly assess the impact and value of DDG in different diseases and radiotracers. Second, a clinical reader assessment as the one performed in our study, might carry an inherent bias since it is virtually impossible to completely blind readers to the image ‘appearance’ of different reconstruction algorithms. Third, results in our study are based on an R-factor of >10.0. The use of other R-factors leads to a different time consumption of DDG and may ultimately also differently impact lesion detection and patient management. Fourth, we did not compare DDG with an external gating technique in our study. Fifth, we did not compare the impact of prolonged non-gated PET acqusition on image quality nor compared a prolonged non-gated PET with DDG. Sixth, we only evaluated one respiratory gating method in our study using prospective on-console reconstruction, which roughly doubles the acquisition time of gated bed positions, and not includes remote offline reconstruction of 4D respiratory phase-matched PET data. While the DDG approach used in our study involves additional acquisition time in beds where motion was detected (in our cohort on average 1.8 bed positons per patient, if motion was detected), other techniques, being partly not clinically available yet, do not prolong the acquisition times and may thereby be more accepted in clinical environments. Such techniques rely on post-processing of gated PET image datasets of step-and-shoot acquisition, utilizing 100% of coincidence events, or on continuous-bed-motion acquisition28 or other techniques based on a optical flow-based de-blurring of motion effects within image reconstruction.26

In conclusion, the deviceless DDG based on a principal component analysis of PET data was evaluated. It provided reliable fully automatic motion correction in clinical routine, and increased lesion detectability and changed management in a considerable number of patients. Thanks to this new technique, PET/CT with respiratory gating can be used routinely in clinical practice without external gating equipment needed. Future studies should systematically assess the impact of DDG in different disease entities and radiotracers to further improve the clinical value of this technique.

Footnotes

Acknowledgment: Dr. Stephan Skawran is supported by a grant form the Palatin-Foundation, Switzerland. Dr. Michael Messerli received a research grant from the Iten-Kohaut Foundation, Switzerland. Dr. Martin W. Huellner received grants from GE Healthcare and a fund by the Alfred and Annemarie von Sick grant for translational and clinical cardiac and oncological research. The authors would like to thank Corina Weyermann, Michèle Hug, Freya Klein and Juliana Koller for their excellent technical support.

Conflict of interest: Dr. Martin W. Huellner is a recipient of speaker’s fees and research grants by GE Healthcare (unrelated to the current study). Apart from that the authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.

Funding: The University Hospital Zurich holds a research agreement with GE Healthcare (unrelated to the current study). No further specific grants from funding agencies in the public, commercial, or not-for-profit sectors were received for this study.

Ethics approval: The present study was approved by the local ethics committee and was conducted in compliance with ICH-GCP rules and the Declaration of Helsinki.

Informed consent: Only patients with documented willingness to the use of their medical data for research were included.

Contributor Information

Michael Messerli, Email: michael.messerli@usz.ch.

Virginia Liberini, Email: v.liberini@gmail.com.

Hannes Grünig, Email: Hannes.Gruenig@usz.ch.

Alexander Maurer, Email: Alexander.Maurer@usz.ch.

Stephan Skawran, Email: michael.messerli@usz.ch.

Niklas Lohaus, Email: Niklas.Lohaus@usz.ch.

Lars Husmann, Email: Lars.husmann@usz.ch.

Erika Orita, Email: Erika.Orita@usz.ch.

Josephine Trinckauf, Email: Josephine.Trinckauf@usz.ch.

Philipp A. Kaufmann, Email: pak@usz.ch.

Martin W. Huellner, Email: Martin.Huellner@usz.ch.

REFERENCES

  • 1.Hess S, Blomberg BA, Zhu HJ, Høilund-Carlsen PF, Alavi A. The pivotal role of FDG-PET/CT in modern medicine. Acad Radiol 2014; 21: 232–49. doi: 10.1016/j.acra.2013.11.002 [DOI] [PubMed] [Google Scholar]
  • 2.Pieterman RM, van Putten JW, Meuzelaar JJ, Mooyaart EL, Vaalburg W, Koëter GH, et al. Preoperative staging of non-small-cell lung cancer with positron-emission tomography. N Engl J Med 2000; 343: 254–61. doi: 10.1056/NEJM200007273430404 [DOI] [PubMed] [Google Scholar]
  • 3.Lardinois D, Weder W, Hany TF, Kamel EM, Korom S, Seifert B, et al. Staging of non-small-cell lung cancer with integrated positron-emission tomography and computed tomography. N Engl J Med 2003; 348: 2500–7. doi: 10.1056/NEJMoa022136 [DOI] [PubMed] [Google Scholar]
  • 4.Goeckenjan G, Sitter H, Thomas M, Branscheid D, Flentje M, Griesinger F, et al. Prevention, diagnosis, therapy, and follow-up of lung cancer. Pneumologie 2010; 64(Suppl 2): e1–164. doi: 10.1055/s-0029-1243837 [DOI] [PubMed] [Google Scholar]
  • 5.Naidich DP. High-Resolution computed tomography of the pulmonary parenchyma: past, present, and future? J Thorac Imaging 2010; 25: 32–3. doi: 10.1097/RTI.0b013e3181cc4de6 [DOI] [PubMed] [Google Scholar]
  • 6.Müller J, Ferraro DA, Muehlematter UJ, Garcia Schüler HI, Kedzia S, Eberli D, et al. Clinical impact of 68Ga-PSMA-11 PET on patient management and outcome, including all patients referred for an increase in PSA level during the first year after its clinical introduction. Eur J Nucl Med Mol Imaging 2019; 46: 889–900. doi: 10.1007/s00259-018-4203-0 [DOI] [PubMed] [Google Scholar]
  • 7.Singh S, Poon R, Wong R, Metser U. 68Ga PET imaging in patients with neuroendocrine tumors: a systematic review and meta-analysis. Clin Nucl Med 2018; 43: 802–10. doi: 10.1097/RLU.0000000000002276 [DOI] [PubMed] [Google Scholar]
  • 8.Lee DY, Kim Y-I. Prognostic value of maximum standardized uptake value in 68Ga-Somatostatin receptor positron emission tomography for neuroendocrine tumors: a systematic review and meta-analysis. Clin Nucl Med 2019; 44: 777–83. doi: 10.1097/RLU.0000000000002694 [DOI] [PubMed] [Google Scholar]
  • 9.Nehmeh SA, Erdi YE. Respiratory motion in positron emission tomography/computed tomography: a review. Semin Nucl Med 2008; 38: 167–76. doi: 10.1053/j.semnuclmed.2008.01.002 [DOI] [PubMed] [Google Scholar]
  • 10.Pépin A, Daouk J, Bailly P, Hapdey S, Meyer M-E. Management of respiratory motion in PET/computed tomography: the state of the art. Nucl Med Commun 2014; 35: 113–22. doi: 10.1097/MNM.0000000000000048 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Schoder H, Larson SM, et al. Effect of respiratory gating on quantifying PET images of lung cancer. J Nucl Med 2002; 43: 876–81. [PubMed] [Google Scholar]
  • 12.Kesner AL, Chung JH, Lind KE, Kwak JJ, Lynch D, Burckhardt D, et al. Validation of software gating: a practical technology for respiratory motion correction in PET. Radiology 2016; 281: 239–48. doi: 10.1148/radiol.2016152105 [DOI] [PubMed] [Google Scholar]
  • 13.Manber R, Thielemans K, Hutton BF, Barnes A, Ourselin S, Arridge S, et al. Practical PET respiratory motion correction in clinical PET/MR. J Nucl Med 2015; 56: 890–6. doi: 10.2967/jnumed.114.151779 [DOI] [PubMed] [Google Scholar]
  • 14.Büther F, Dawood M, Stegger L, Wübbeling F, Schäfers M, Schober O, et al. List mode-driven cardiac and respiratory gating in PET. J Nucl Med 2009; 50: 674–81. doi: 10.2967/jnumed.108.059204 [DOI] [PubMed] [Google Scholar]
  • 15.Baratto L, Park SY, Hatami N, Davidzon G, Srinivas S, Gambhir SS, et al. 18F-Fdg silicon photomultiplier PET/CT: a pilot study comparing semi-quantitative measurements with standard PET/CT. PLoS One 2017; 12: e0178936. doi: 10.1371/journal.pone.0178936 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Messerli M, Stolzmann P, Egger-Sigg M, Trinckauf J, D'Aguanno S, Burger IA, et al. Impact of a Bayesian penalized likelihood reconstruction algorithm on image quality in novel digital PET/CT: clinical implications for the assessment of lung tumors. EJNMMI Phys 2018; 5: 27. doi: 10.1186/s40658-018-0223-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Pan T, Einstein SA, Kappadath SC, Grogg KS, Lois Gomez C, Alessio AM, et al. Performance evaluation of the 5-ring Ge discovery MI PET/CT system using the National electrical manufacturers association nu 2-2012 standard. Med Phys 2019; 46: 3025–33. doi: 10.1002/mp.13576 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Ferraro DA, Garcia Schüler HI, Muehlematter UJ, Eberli D, Müller J, Müller A. Impact of 68 Impact of PET data driven respiratory motion correction and BSREM reconstruction of 68 Ga-DOTATATE PET/CT for differentiating neuroendocrine tumors (NET) and intrapancreatic accessory spleens (IPAS)Ga-PSMA-11 PET staging on clinical decision-making in patients with intermediate or high-risk prostate cancer. Eur J Nucl Med Mol Imaging 2020; 47: 652–64. doi: 10.1007/s00259-019-04568-1 [DOI] [PubMed] [Google Scholar]
  • 19.Khamis H. A Novel Approach to Seamlessly Integrating Respiratory Motion for all Patients and all Exams. Available from: https://www.gehealthcare.co.uk/ [Accessed Apr 2020].
  • 20.Walker MD, Morgan AJ, Bradley KM, McGowan DR. Evaluation of data-driven respiratory gating waveforms for clinical PET imaging. EJNMMI Res 2019; 9: 1. doi: 10.1186/s13550-018-0470-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Osborne DR, Acuff S, Neveu M, Kaman A, Syed M, Fu Y. 90Y liver radioembolization imaging using Amplitude-Based gated PET/CT. Clin Nucl Med 2017; 42: 373–4. doi: 10.1097/RLU.0000000000001613 [DOI] [PubMed] [Google Scholar]
  • 22.Walker MD, Bradley KM, McGowan DR. Evaluation of principal component analysis-based data-driven respiratory gating for positron emission tomography. Br J Radiol 2018; 91: 20170793. doi: 10.1259/bjr.20170793 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Büther F, Vehren T, Schäfers KP, Schäfers M. Impact of data-driven respiratory gating in clinical PET. Radiology 2016; 281: 229–38. doi: 10.1148/radiol.2016152067 [DOI] [PubMed] [Google Scholar]
  • 24.Crivellaro C, De Ponti E, Elisei F, Morzenti S, Picchio M, Bettinardi V, et al. Added diagnostic value of respiratory-gated 4D 18F-FDG PET/CT in the detection of liver lesions: a multicenter study. Eur J Nucl Med Mol Imaging 2018; 45: 102–9. doi: 10.1007/s00259-017-3795-0 [DOI] [PubMed] [Google Scholar]
  • 25.Guerra L, De Ponti E, Elisei F, Bettinardi V, Landoni C, Picchio M, et al. Respiratory gated PET/CT in a European multicentre retrospective study: added diagnostic value in detection and characterization of lung lesions. Eur J Nucl Med Mol Imaging 2012; 39: 1381–90. doi: 10.1007/s00259-012-2148-2 [DOI] [PubMed] [Google Scholar]
  • 26.Büther F, Jones J, Seifert R, Stegger L, Schleyer P, Schäfers M. Clinical evaluation of a data-driven respiratory gating algorithm for whole-body PET with continuous bed motion. J Nucl Med 2020; 61: 1520–7. doi: 10.2967/jnumed.119.235770 [DOI] [PubMed] [Google Scholar]
  • 27.Walker MD, Morgan AJ, Bradley KM, McGowan DR. Data-Driven Respiratory Gating Outperforms Device-Based Gating for Clinical 18F-FDG PET/CT. J Nucl Med 2020; 61: 1678–83. doi: 10.2967/jnumed.120.242248 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.van der Vos CS, Koopman D, Rijnsdorp S, Arends AJ, Boellaard R, van Dalen JA, et al. Quantification, improvement, and harmonization of small lesion detection with state-of-the-art PET. Eur J Nucl Med Mol Imaging 2017; 44(Suppl 1): 4–16. doi: 10.1007/s00259-017-3727-z [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from The British Journal of Radiology are provided here courtesy of Oxford University Press

RESOURCES