Abstract
Purpose
To develop a coil‐based method to obtain accurate sensitivity profiles in 13C MRI at 3T from the endogenous 23Na. An eight‐channel array is designed for 13C MR acquisitions. As application examples, the array is used for two‐fold accelerated acquisitions of both hyperpolarized 13C metabolic imaging of pig kidneys and the human brain.
Methods
A flexible coil array was tuned optimally for 13C at 3T (32.1 MHz), with the coil coupling coefficients matched to be nearly identical at the resonance frequency of 23Na (33.8 MHz). This is done by enforcing a high decoupling (obtained through highly mismatched preamplifiers) and adjusting the coupling frequency response. The SNR performance is compared to reference coils.
Results
The measured sensitivity profiles on a phantom showed high spatial similarity for 13C and 23Na resonances, with average noise correlation of 9 and 11%, respectively. For acceleration factors 2, 3, and 4, the obtained maximum g‐factors were 1.0, 1.1, and 2.6, respectively. The 23Na profiles obtained in vivo could be used successfully to perform two‐fold acceleration of hyperpolarized 13C 3D acquisitions of both pig kidneys and a healthy human brain.
Conclusion
A receive array has been developed in such a way that the 13C sensitivity profiles could be accurately obtained from measurements at the 23Na frequency. This technique facilitates accelerated acquisitions for hyperpolarized 13C imaging. The SNR performance obtained at the 13C frequency, compares well to other state‐of‐the‐art coils for the same purpose, showing slightly better superficial and central SNR.
Keywords: flexible RF coil, hyperpolarization, parallel Imaging, X‐Nuclei MRI
1. INTRODUCTION
Parallel imaging 1 has become a standard strategy to accelerate acquisitions and improve encoding efficiency of MR imaging. Standard parallel imaging requires prior information of the receive coil sensitivities to correct for the geometric aliasing artifacts created as data are sampled below the Nyquist rate. The coil sensitivity profiles can be obtained prior to the actual scan (pre‐calibration), or obtained from a fully sampled center of k‐space (auto‐calibration). 2
In hyperpolarized MRI, imaging time is scarce due to the rapid polarization decay of the hyperpolarized sample. Parallel imaging is then an appealing approach, since in this case the SNR penalty is ideally only proportional to the coil g‐factor (for multi‐shot imaging, with flip angles adjusted to spend the same amount of magnetization 3 ). However, the separate acquisition of the sensitivity profiles is not a viable option, since the thermal (non‐hyperpolarized) signal available is prohibitively low. This is especially true for in vivo hyperpolarized 13C imaging, 4 due to the low natural abundance of the 13C isotope (1.11%).
In the hyperpolarized 13C case, several approaches have been proposed to estimate the coil profiles. One solution is to acquire them in advance from phantom measurements and co‐register these spatially to the in vivo measurements in the reconstruction. 5 , 6 Another proposed solution is to add fiducial markers to the coils, such that the position of each element can be calculated and allow for numerical generation of the sensitivity profiles. 7 , 8 Both these methods are possible at the low frequency of 13C (32.1 MHz at 3T), due to the minor changes on B1 field distribution introduced by the sample, which allows to approximate the profiles using Biot‐Savart law (quasi‐static approximation) for the field calculation. However, accurate co‐registration of the individual coil elements from phantom measurements is challenging for flexible arrays, where the conformation of the array is not fixed.
The auto‐calibrated approach has also been applied to hyperpolarized 13C imaging. 9 , 10 However, this approach does not always perform well in the presence of Gibbs ringing and edge artifacts, which can be a problem for hyperpolarized 13C imaging. This is due to the inhomogeneous spatial distribution of the hyperpolarized tracer, with both highly vascularized regions and regions with low metabolic activity, which often creates images with high‐contrast between adjacent voxels.
Finally, a calibrationless method, which does not require estimation of the coil profiles also exists and has been applied to hyperpolarized 13C imaging. 11 When coil profiles cannot be estimated reliably, this method provides a good alternative to acceleration using multi‐channel coil arrays. The method, however, requires a pseudo‐random sampling pattern similar to compressed sensing and is, therefore, not applicable with more standard k‐space trajectories.
To overcome these limitations, a hardware‐based method to obtain highly accurate 13C sensitivity profiles is proposed, taking advantage of the proximity in Larmor frequency of the 23Na nuclei (1.7 MHz difference at 3T). Endogenous 23Na is abundant enough in biological tissue to be imaged with relatively high resolutions within the same scan session as the hyperpolarized 13C experiment. It also shows a rather uniform distribution for different tissues, which avoids Gibbs ringing and edge artifacts. 12 Additionally, obtaining the 13C sensitivity profiles during the same hyperpolarized 13C scan session, facilitates the integration of parallel imaging into the clinical workflow of the MRI scanner.
The idea in this work was to design a 13C receive array (with SNR response optimal at the 13C frequency), with a coupling matrix tailored such that the coupling levels were nearly identical at 13C and 23Na. In this way, the sensitivity profiles are expected to be similar. Some work has already been done to estimate the transmit parameters of the 13C scans from the 23Na naturally present in biological tissue 13 , 14 using commercially available volume coils.
However, for 13C multichannel arrays, the situation can be very different. The low loading that the small surface coils experience means that individual coils have a high quality factor (Q‐factor), which increases the coupling between next neighboring coils. This coupling can then be minimized through preamplifier decoupling, 15 but preamplifier decoupling is also narrowband in frequency for high‐Q coils. Therefore, the similarity of sensitivity profiles between 13C and 23Na depends directly on what is the exact level of preamplifier decoupling obtained at these two frequencies. For this reason, in a typical 13C array, a direct attempt to use the 23Na profiles for 13C reconstructions will most likely result in suboptimal reconstruction (as shown inRef. [13] for a commercial eight‐channel array).
The design proposed here relies on creating a high level of preamplifier decoupling, which can be obtained by matching the low noise amplifier (LNA) to a higher impedance than the noise optimal, in a controlled way. 16 Then, its frequency response can be tuned to the intermediate frequency between 13C and 23Na, such that the coupling levels will still be low (and equal) at both frequencies. As a proof of principle, a flexible eight‐channel array implementing this method is designed and evaluated.
2. METHODS
2.1. Flexible coil design
In a rigid array high decoupling levels can be achieved through geometric overlap and mismatched preamplifiers. 15 , 17 Geometric decoupling is less effective in flexible arrays since coupling coefficients are different for different array conformations. At high frequencies (128 MHz and above), self‐resonant flexible coils are feasible, 18 , 19 , 20 where some sort of transmission line structure is used. Such designs allow some freedom to create magnetic field distributions different from those created by simple loop coils, ultimately allowing reduced mutual coupling based purely on the geometry of the individual coils. At low frequencies, these methods are less practical, since they require very large dimensions of the individual coils. Also, the loading introduced by the biological sample is lower, meaning that any extra loss introduced by the coil materials and electronics has a notable effect on the final coil performance. Therefore, copper loops tuned with high‐Q ceramic capacitors are hard to outperform at these low frequencies.
Based on this rationale, this work follows a different approach for the design of a flexible array, with the use of standard flexible copper wire, tuned using lumped elements. In order to achieve the required decoupling, a high mismatch between the coil and preamplifier is created, as described in Ref. [16], where the coils are matched to a higher impedance (200 Ω, for the sample‐loaded coils) than the optimum for the preamplifier (50 Ω). In this way, a controlled noise penalty is accepted by not matching the coils to the optimal noise impedance of the LNA (WMA32C, WanTCom, USA). This matching enables a level of decoupling above 30 dB (S21 < −30 dB), which ultimately allows the flexible array to perform well when conforming to different subjects.
The essential design feature of the proposed array is that the level of decoupling provided by the mismatched preamplifiers should be as similar as possible for the two frequencies of interest. The Noise Figure (NF) of the LNA is 1.3 dB when matched to 200 Ω, compared to 0.7 dB at the optimal impedance of 50 Ω (data obtained from a private communication with the manufacturer, and confirmed experimentally).
Active decoupling from the transmitter was implemented following the low noise circuit design described in Ref. [21], which provided a level of decoupling above 45 dB (S21 < −45 dB). Its frequency response was also tuned to be the same at the 13C and 23Na frequencies, such that any residual B1 + distortion due to imperfect detuning of the receive coils would be the same for both frequencies. At this point, the noise correlation matrices of an initial four‐channel array were measured, with and without this modification on the preamplifier decoupling response, in order to confirm that the decoupling after the modification was sufficient (see Supporting Information Figure S1, which is available online).
The SNR variation as a function of frequency was measured for one of the array elements by exciting an RF tone through a pickup loop fixed at a distance from the coil element, with the measured spectrum recorded by a spectrum analyzer (PSA E4440A, Keysight, CA, USA). The excited tone was swept over frequency, and the SNR measured as the ratio of the recorded spectrum, divided by the average of a noise spectrum.
The final coil array was made with eight loops of 85 mm diameter each. Each loop was built with standard flexible copper coax (RG‐316), where the outer jacket was used to create the conductive loop. Each individual loop included a trap circuit, which creates a high impedance at the 1H frequency, therefore allowing anatomical 1H imaging in the presence of the 13C array. The trap was made by including a 22 nH inductor (B07T, Coilcraft, USA) in the conductive loop, in parallel with a ceramic capacitor (E‐Series, Johanson Technology, USA). The measured unloaded‐to‐loaded Q‐ratio for the individual elements was QU/QL = 220/80 when loaded with a human head. The inductance of the individual elements was estimated to be 245 nH, leading to an estimate of equivalent losses of 225 mΩ for the coil, and 394 mΩ for the sample. The coil loss was measured including the solder joints, and the tuning, matching, and decoupling circuitry. The fabricated array is shown in Figure 1.
2.2. Array performance: SNR and noise correlation
All MR experiments in this study were performed using a 3T GE MR750 MRI system (GE Healthcare, Milwaukee, USA). The performance of the array was evaluated through MRSI measurements (CSI, FOV = 36 × 36 × 15 cm3, matrix size = 24 × 24, slice‐selective pulse [1.8 ms, minimum‐phase], TE = 1.43 ms, flip angle = 70°). The measurements were performed on a Specific Anthropomorphic Mannequin (SAM) phantom filled with ethylene glycol doped with 17 g/L of NaCl (291 mmoL/L 23Na concentration) to emulate tissue loading. 22 The 13C measurement was acquired with a TR of 1 s (total acquisition time = 9 min 36 s), while the 23Na was acquired with a TR of 219 ms and eight averages (total acquisition time 16 min 40 s). The CSI datasets for each of the two different nuclei were filtered with a 20 Hz line‐broadening filter and the separate coil images were reconstructed based on the first in‐phase time point across the acquired FIDs. Coil images were then normalized to their sum‐of‐squares combined images to remove transmit coil effects and masked to remove the background noise. By normalizing by the combined image, magnitude effects resulting from transmit inhomogeneity cancel out. Masking was done to increase visual clarity. Noise correlation matrices were calculated based on the last 200 points of the raw FIDs. The flip angle calibration was performed using the Bloch‐Siegert shift, as described in Ref. [23].
Finally, the 13C SNR was compared with a transmit‐receive birdcage volume coil (RAPID Biomedical, Rimpar, Germany) and a rigid eight‐channel 13C receive coil (GE Healthcare, Milwaukee, USA), following the QA protocol described in Ref. [22]. The birdcage is double tuned (1H‐13C), has an inner diameter of 255 mm and length of 170 mm, while the rigid array is made of rectangular loops of 50 × 100 mm2 size. Both coils are designed for human head imaging and are representative of the state‐of‐the‐art of 13C coils for brain imaging at 3T. The two paddles of the rigid eight‐channel array were placed directly attached to both sides (left–right) of the head phantom, therefore, placed as close as possible to the phantom.
Measurements done with receive‐only coils were performed using a dedicated 13C transmit coil of the clamshell type (RAPID Biomedical, Rimpar, Germany). The electromagnetic design of this coil is based on a Helmholtz pair. This design provides a reasonably similar B1 + distribution across the frequency range covering the 13C and 23Na frequencies, as shown in Ref. 13].
2.3. Parallel imaging MR experiments
2.3.1. In vitro: Phantom
Parallel imaging reconstruction of 13C data using 23Na sensitivity profiles based on SENSE 1 was first verified in phantom experiments, based on retrospectively undersampled data using the SAM phantom described in the previous section.
Data were acquired with a Cartesian sequence, a 2D 13C GRE with FOV = 28 × 28 cm2, matrix size = 48 × 48, slice thickness = 3 cm, TR = 1000 ms, flip angle = 70°, spectral‐spatial excitation (22 ms pulse length), TE = 9 ms, and total scan time = 32 min. 24 The dataset was retrospectively undersampled by removing lines in k‐space uniformly, with acceleration rates R = 2, 3, and 4. Coil g‐factors were calculated for each acceleration case.
The 23Na data for coil sensitivity mapping were acquired using the same 2D GRE sequence with FOV = 26.6 × 26.6 cm2 (5% smaller due to the 13C‐23Na frequency difference), matrix size = 48 × 48, slice thickness = 2.85 cm, TR = 22.6 ms, TE = 1.43 ms, slice‐selective pulse (1.8 ms, minimum‐phase), total scan time = 21 min 42 s. The flip angle used during this measurement is estimated to be around 20°, because the transmit coil used is not tuned for 23Na, and its efficiency at that frequency is very low. SENSE reconstruction requires low T1‐weighting; therefore, a relatively long TR was chosen to minimize this effect.
Data were reconstructed in MATLAB R2019b. SENSE reconstruction and g‐factor calculation were done using the code made available by Michael S. Hansen. 25 Sensitivity maps were estimated from the 23Na data by first normalizing the reconstructed multi‐channel images with the sum‐of‐squares combined image and then performing edge‐preserving smoothing using the built‐in MATLAB function “imguidedfilter.” Edge‐preserving smoothing was performed separately for real and imaginary parts of the image data with the default degree of smoothing and a window size set relative to the image size (10% of the number of pixels).
A Cartesian acquisition was used at this part of the study (despite its known inefficiency for 23Na), to avoid sequence related artifacts and obtain coil profiles comparable to the Cartesian 13C acquisition. In this way an accurate comparison of the 13C and 23Na coil profiles could be made, with minimal bias from the differences in sequence parameters.
2.3.2. In vivo: Pig kidney and human brain
A 3D 13C blipped stack‐of‐spirals (R = 2) sequence 6 was designed in two versions for 13C MR imaging of pig kidneys and human brain, after hyperpolarized [1‐13C]pyruvate injection. The sequences were designed with different resolution for pyruvate and metabolites, 26 with a 45 ms spiral readout, spectral‐spatial excitation, 24 and with TE = 9 ms. Other sequence parameters can be found in Table 1. Flip angles on the metabolic products were chosen to maximize the total accumulated signal as described previously. 6
TABLE 1.
Pig kidneys | Human brain | |||
---|---|---|---|---|
Acq. parameters | Pyruvate | Products | Pyruvate | Products |
FOV (cm3) | 28 × 28 × 10 | 28 × 28 × 12 | ||
Matrix size | 70 × 70 × 10 | 35 × 35 × 10 | 40 × 40 × 8 | 20 × 20 × 8 |
Nominal spatial resolution (mm3) | 4 × 4 × 10 | 8 × 8 × 10 | 7 × 7 × 15 | 14 × 14 × 15 |
Flip angles (°) | 9 | 34 | 10 | 28 |
TR (ms) | 86.7 | 77 | ||
Frequency acq. order | Pyr‐Lac‐Pyr‐Bic‐Pyr‐Ala | Pyr‐Lac‐Bic‐Ala | ||
Excitations per volume | 5 | 4 | ||
Temporal resolution a (s) | 0.867 | 2.601 | 1.232 |
Abbreviations: Ala, alanine; Bic, bicarbonate; Lac, lactate; Pyr, pyruvate.
Time between full volume acquisitions at the same frequency.
Prior to in vivo acquisition, the 13C blipped stack‐of‐spirals sequences were evaluated in vitro using the SAM phantom with the following parameters: TR = 1000 ms, flip angle = 70°, and total scan time = 25 min (for the kidney sequence) and 20 min (for the brain sequence). For comparison, a fully sampled stack‐of‐spirals adaptation of the kidney sequence was also tested with total scan time = 25 min.
CG‐SENSE parallel imaging reconstruction of the non‐Cartesian blipped stack‐of‐spirals datasets was done as previously described, 6 including 15 Hz filtering on the spiral readouts and B0 off‐resonance correction. 27 But contrary to previous implementation of the off‐resonance correction method, this study did not require any manual shift of the B0 map frequency, since the 13C center frequency was set directly based on prior information of the 1H to 13C‐lactate frequency difference on the scanner. 13
For the in vivo acquisitions, 23Na data for sensitivity mapping were acquired using a 3D cones based ultra short echo time sequence, with FOV = 48 × 48 × 48 cm3, matrix size = 120 × 120 × 120, 197 excitations per volume acquisition, TR = 120 ms, TE = 0.5 ms, non‐selective excitation, four averages, and total scan time 1 min 35 s. The 23Na data were reconstructed using 3D gridding. Sensitivity maps were estimated using the same approach as for the in vitro data, after matching the FOV and resolution to the 13C dataset.
The human experiment was approved by the Danish Medicines Agency and the Ethics Committee of the Central Region of Denmark, while the animal experiments were approved by the Danish Animal Inspectorate. The human experiment was performed with a healthy, male volunteer. The pre‐clinical work was performed on a 40 kg healthy female Danish Landrace pig. The pig was anesthetized using propofol and fentanyl. The eight‐channel array was firmly wrapped at the kidney position and adjusted to be most centered in both kidneys. The cross‐section of the pig was approximately 28 cm on the left–right dimension, and 22 cm on the anterior–posterior dimension. The pig was mechanically ventilated and no respiratory gating was used, since motion artifacts are generally low in the kidneys when using fast read‐out sequences.
For anatomical reference, routine proton images were acquired using the body coil of the scanner. For the pig experiment, proton images were acquired using a multi‐slice axial T2 Propeller sequence (TR/TE = 6654/100 ms, flip angle = 142°, FOV = 34 × 34 cm2, slice thickness = 4 mm, matrix size 256 × 256). For the human experiment, proton images were acquired using a multi‐slice axial fast spin echo sequence (TR/TE = 900/7 ms, flip angle = 111°, FOV = 25 × 25 cm2, slice thickness = 8 mm, matrix size 280 × 280). In addition, B0 field maps were acquired at the 1H frequency for both in vivo studies using the scanner software IDEAL‐IQ (GE Healthcare); a 3D sequence to estimate parameter maps from a single breath‐hold acquisition (TR/TE = 5/2 ms, flip angle = 3°, FOV = 48 × 48 cm2 (pig) 44 × 44 cm2 (human), slice thickness = 15 mm, matrix size 128 × 128).
For MRI with hyperpolarized pyruvate, [1‐13C]pyruvic acid was polarized using dynamic nuclear polarization (SPINlab, GE Healthcare) with AH111501 as an electron paramagnetic agent (EPA), usually yielding >40% polarization. 28 The sample was dissolved with water for injection and buffered to physiological temperature, acidity, and osmolarity. For the human experiment, the pyruvate was filtered to remove the EPA and quality controlled. The hyperpolarized [1‐13C]pyruvate (250 mM) was injected at a dose of 0.43 mL/kg and chased by 20 mL of saline. Imaging was started just after pyruvate injection, before the chaser, to capture the temporal dynamics.
3. RESULTS
3.1. Array performance
The measured preamplifier decoupling response over frequency for one of the individual elements of the array is shown in Figure 2A. The difference between the optimal decoupling (obtained around 33 MHz) and the decoupling at 13C and 23Na frequencies is 6–7 dB. This is the level of coupling sacrificed with this design compared to an array optimized for 13C. A typical frequency response for one of the coil elements is shown in Figure 2B, which shows that the best SNR performance is obtained at the 13C frequency. For a sample loaded coil, the SNR performance at the 23Na using this LNA and coil matching, is about half of that at the 13C frequency.
In Figure 3A,B, the single‐slice coil profiles are shown measured with the SAM phantom for both 13C and 23Na frequencies. The relative error of the magnitude images was calculated as the absolute difference divided by the 13C magnitude images. Both the relative error of the magnitude images (Figure 3A) and the absolute error of the phase images (Figure 3B) are generally low, especially for the regions close to the coil loops with high SNR. The highest errors are found at the edges opposite the loops, which also correspond to low‐SNR regions where the profile measurements were prone to more errors. Figure 3C shows the noise correlation matrices. Note that, in loaded coils (in the presence of the sample), noise correlation is not a pure metric of coil decoupling. This is because part of the noise correlation between individual coils comes from the fact that nearby coils see the same noise from the sample (e.g., overlapped region on neighboring coils). In any case, the measured values are in general low, and showing good agreement of the average correlation levels between 13C and 23Na (9% and 11%, respectively).
Figure 4 shows the 13C SNR level of the array compared to the state‐of‐the‐art volume coil and rigid 8‐channel paddle coil. The SNR was measured as described in Ref. [22] over a slice of 20 mm placed at the center of the coil array in the axial plane. The SNR performance from the flexible array compares well to the traditional array, showing slightly better superficial and central SNR. This is most likely due to the slightly increased distance between coil and sample for the rigid array (due to its thicker casing), and the tighter fitting that the flexible array can provide. The comparison to the birdcage volume coil shows that the central SNR is very similar for both coils (≈ 50), with the superficial SNR being three to four times higher for the array.
3.2. Parallel imaging performance: Phantom
The results from the in vitro study, demonstrating a SENSE reconstruction of 13C data using 23Na acquired sensitivity profiles with the same coil array, are shown in Figure 5 for different accelerations rates (R = 2, 3, and 4). The SENSE reconstructions of the undersampled data are compared to the fully sampled sum‐of‐squares reconstruction. After retrospective undersampling, the total scan time amounts to 16 min for R = 2, 11 min for R = 3, and 8 min for R = 4. To approximately match the scan time for the fully sampled data in the comparison, averages were removed resulting in a total scan time of 16 min for the comparison to R = 2 and R = 3, and a total scan time of 8 min for comparison to R = 4.
The sum‐of‐squares reconstructions of the fully sampled data and the SENSE reconstructions of the undersampled data show minimal differences for acceleration rates R = 2 and R = 3. This is also reflected in the associated g‐factor maps, with g‐factors equal to 1 for R = 2 and close to 1 for R = 3, meaning none or minimal noise amplification in the parallel imaging reconstruction. For acceleration rate R = 4, substantial noise amplification is seen in the top‐center of the phantom with g‐factors >2.
Figure 6 shows the high‐resolution SNR images from the phantom test of the variable resolution blipped stack‐of‐spirals 13C sequence designed for the pig kidney imaging, compared to a fully sampled stack‐of‐spirals. The CG‐SENSE reconstruction with 23Na sensitivity maps is able to remove the majority of the aliasing artifacts seen in the sum‐of‐squares reconstruction in the undersampled dataset while retaining the SNR level. This correlates well with the close to unity g‐factor for the R = 2 accelerated GRE in Figure 5. Notice, however, the superficial “spike‐pattern” artifacts seen in the last two slices in both the fully sampled and undersampled images. These are attributed to B0 distortions caused by some of the electronics on the coil (see Supporting Information Figure S1). Due to the very tight fit to the subject, materials and components disturbing the magnetic field slightly, can introduce visible superficial artifacts. With the thin shell of the SAM phantom (2 mm), and hereby short distance from coil to signal, these images represent an extreme case of these artifacts. The artifacts were, therefore, not considered a concern for the in vivo experiments, but highlight the need to carefully choose the position of the electronic boards in flexible arrays.
3.3. Parallel imaging performance: In vivo pig kidney and human brain
Finally, the performance of the array was evaluated in vivo in animal and in human. First, the 3D cones 23Na data were acquired prior to the hyperpolarization experiments. The reconstructed 23Na images after matching the FOV and resolution to the 13C pyruvate data are shown in Figure 7 (as SNR maps) for both in vivo studies together with the magnitude of the final processed coil sensitivity maps used for reconstruction of the 13C data. The phase images of the coil sensitivity maps are shown in the Supporting Information Figure S1.
The SNR levels are different across the different slices, with a maximum SNR at the central slices and much lower SNR levels at the peripheral slices. This is partly due to the large FOV in the inferior–superior direction (120 mm for the pig and 100 mm for the human study), relative to the size of the coil elements (Ø = 85 mm). From the final sensitivity maps, however, it can be seen that the SNR levels of the 3D 23Na images for both the abdomen of the pig and the head of the human are sufficient to clearly delimitate the individual coil positions and field distribution across the full volume, despite signal variations due to both anatomy and coil geometry.
Figure 8 shows the time‐averaged metabolic maps after hyperpolarized [1‐13C] pyruvate injection for both in vivo experiments. The images are shown as transparent overlays on anatomical 1H images. Metabolite‐only images are shown in Supporting Information Figure S1. Typical metabolic activity is observed both in the healthy pig kidneys and in the healthy human brain. The brain metabolism distribution and the image quality are comparable to previously reported experiments with similar conditions. 26 Alanine images are not shown for the brain, due to insufficient signal.
Supporting Information Videos S1 and S2 show the metabolic dynamics in animations of the pig kidney and human brain data, respectively. Supporting Information Figures S1 show more images from the brain data and provides transparency to the reconstruction pipeline. Supporting Information Figure S1 shows the CG‐SENSE‐reconstructed metabolic maps compared to a 3D gridded, sum‐of‐squares reconstruction without off‐resonance correction, similar to the phantom images in Figure 6. Supporting Information Figure S1 shows the pyruvate and lactate images following different reconstruction procedures. Supporting Information Figure S1 shows the mean time curves.*
4. DISCUSSION
This work investigates the possibility of introducing a hardware modification into an eight‐channel flexible array dedicated for 13C MR, such that its individual sensitivity profiles can be accurately obtained from biologically abundant 23Na. It is shown that for an array where the decoupling between individual elements is very high (S21 < −30 dB), a sacrifice can be made on the tuning of the preamplifier decoupling response, such that the 13C and 23Na coupling matrices are similar. The 13C SNR level measured with this array shows an adequate performance, comparable to two state‐of‐the‐art coils for 13C brain imaging, which have shown good SNR performance in previous studies. 22 , 29 In general, this result confirms that the extra coupling accepted at the 13C frequency with this method does not imply a remarkable effect on the final 13C SNR performance, for the FOV of a human head. It should also be taken into account that the FOV coverage of the coils is different across the z‐axis; therefore, the SNR performance at the central slice does not provide all the information needed to categorically evaluate a coil. In this case, it is used to have an objective metric to benchmark the SNR performance of the proposed array, since a 20 mm slice thickness is relevant for hyperpolarized 13C MRSI.
The in vitro results demonstrate that the 23Na coil profiles are sufficient for sensitivity mapping and parallel imaging reconstruction of 13C data with acceleration factors up to R = 4 for the given undersampling strategy relative to the eight‐channel coil geometry. Geometry factors for acceleration factors up to R = 3 were insignificant. The SAM phantom has a relatively high 23Na concentration (291 mmol/L) compared to in vivo conditions in the brain (gray matter: 36 mmol/L, white matter: 31 mmol/L, cerebrospinal fluid [CSF]: 125 mmol/L). 30 This, together with the homogenous distributions of both 13C and 23Na signal, provided an advantageous starting point for testing the coil setup. However, despite the more challenging conditions in vivo, results from both the pig abdomen and human brain show successful parallel imaging reconstruction of undersampled (R = 2) hyperpolarized 13C data by means of sensitivity profiles from 23Na data obtained with the same coil in the same subject. The results obtained from the human brain are comparable in quality to those reported for similar experiments using arrays with higher channel‐count (e.g., 24‐channel array in Ref. [26]).
Regarding the 23Na imaging performance, this study is only a proof of principle, and significant hardware improvements can be undertaken to obtain better 23Na SNR, especially regarding transmit coil efficiency. In fact, a better design for this purpose has recently been proposed, 14 where a large‐FOV four‐rung birdcage could be tuned optimally at the 23Na frequency, with a power efficiency only 3 dB lower at the 13C frequency.
The results presented here show that it is possible to obtain sufficient 23Na SNR from both the pig abdomen and human brain tissue in a reasonable time (1 min 30 s), even with a suboptimal 23Na coil setup. However, this may be more challenging for other body parts with smaller 23Na concentration, or higher 23Na gradients. 31 For example, there is a 90–95 mmol/L difference between tissue and CSF in the brain. 30 , 32 However, as there is no 13C signal from pyruvate or its metabolites in CSF (due to no metabolism nor direct feeding from the supplying vessels to the brain), this is unlikely to distort image domain unfolding. For thoracic and abdominal imaging, self‐gating acquisitions have already been successfully used for 23Na imaging, to avoid blurring from breathing motion. 33 Additionally, due to the low frequencies, quasi‐static field approximations can be made to overcome this limitation. Such field approximations can also be used to estimate the absolute sensitivity profiles from the 23Na coil images, which would allow for intensity correction of the final metabolic maps. In the current implementation, 23Na coil images were normalized to their sum‐of‐squares combined image to cancel out anatomic signal variations before estimating the relative sensitivity profiles. In cases where intensity correction is not needed, for example, for calculating metabolic rate maps, this is the most straightforward and recommended approach.
The conclusions of this study are not general and will not work equally well for any coil geometry and array channel count. For example, bigger coils or higher Q coils do require higher decoupling to prevent performance degradation; therefore, this method might not be suitable for those. In the specific case of flexible arrays, high decoupling is already a need to minimize detuning of the different elements for the changing configurations and matching conditions. Also in these arrays, the high sample loading experienced due to their better anatomical fitting provides a broader frequency response of their SNR performance (as shown in Figure 3), making them less sensitive to detuning due to coupling. This SNR vs. frequency response is dependent on the preamplifier noise response for different input impedances and can, therefore, also potentially be improved, if one designs one's own preamplifier.
It is important to mention that the method proposed here for receive arrays can only be used if the transmit coil used has a sufficient degree of similarity in its B1 + profiles for the 13C and 23Na frequencies. In this work, a commercial 13C transmit coil is used, where the B1 + similarity had already been verified. 13
In summary, with the proposed method parallel imaging of hyperpolarized 13C can be integrated and automated, with less sources of error compared to pre‐calibration methods that need accurate co‐registration of the profiles and regular coil control to assure that the calibrated profiles match the current state of the coil. Furthermore, this method also allows parallel imaging with flexible coils, which can provide increased SNR through tight patient‐specific fitting.
5. CONCLUSIONS
This work demonstrates a method to obtain accurate sensitivity profiles for hyperpolarized 13C imaging from the 23Na naturally present in biological samples, provided that the receive hardware is specifically designed for that purpose. A flexible eight‐channel receive array for 13C at 3T (32.1 MHz) has been developed based on this method, and its parallel imaging performance has been evaluated in vitro and in vivo in both a hyperpolarized 13C pig kidney and human brain experiment, showing successful results.
This method can be used to obtain accurate 13C sensitivity profiles in cases where the low natural abundance of 13C makes it impossible, facilitating the use of accelerated sequences for hyperpolarized 13C experiments.
ACKNOWLEDMENTS
This work has been partly funded by the Danish Council for Independent Research (DFF – 4005‐00531), the Danish National Research Foundation (DNRF124), Lundbeck Foundation (R272‐2017‐4023 & R278‐2018‐620), and the European Research Council (ERC‐856432, HyperQ). Dr. James Grist was funded by the Oxford BHF Centre of Research Excellence by grant code (RE/18/3/34214), and by a European Commission grant (858 149, AlternativesToGd). Prof. Damian Tyler was funded by a British Heart Foundation Senior Research Fellowship (FS/19/18/34252).
Supporting information
Sanchez‐Heredia JD, Olin RB, Grist JT, et al. RF coil design for accurate parallel imaging on 13C MRSI using 23Na sensitivity profiles. Magn Reson Med. 2022;88(3):1391‐1405. doi: 10.1002/mrm.29259
Christoffer Laustsen and Jan H. Ardenkjær‐Larsen contributed equally to this work.
Funding information British Heart Foundation, Grant/Award Number: FS/19/18/34252; Danish Council for Independent Research, Grant/Award Number: DFF ‐ 4005‐00531; Danish National Research Foundation, Grant/Award Number: DNRF124; European Comission, Grant/Award Number: 858149; H2020 European Research Council, Grant/Award Number: 856432; Lundbeckfonden, Grant/Award Numbers: R272‐2017‐4023; R278‐2018‐620; Oxford BHF Centre of Research Excellence, Grant/Award Number: RE/18/3/34214
Footnotes
We provide access to raw data through Zenodo, https://doi.org/10.5281/zenodo.6359357.
REFERENCES
- 1. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;962:952‐962. [PubMed] [Google Scholar]
- 2. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202‐1210. [DOI] [PubMed] [Google Scholar]
- 3. Lee RF, Johnson G, Grossman RI, Stoeckel B, Trampel R, McGuinness G. Advantages of parallel imaging in conjunction with hyperpolarized helium—a new approach to MRI of the lung. Magn Reson Med. 2006;55:1132‐1141. [DOI] [PubMed] [Google Scholar]
- 4. Ardenkjær‐Larsen JH, Fridlund B, Gram A, et al. Increase in signal‐to‐noise ratio of >10,000 times in liquid‐state NMR. Proc Natl Acad Sci U S A. 2003;100:10158‐10163. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Tropp J, Lupo JM, Chen A, et al. Multi‐channel metabolic imaging, with SENSE reconstruction, of hyperpolarized [1‐13C] pyruvate in a live rat at 3.0tesla on a clinical MR scanner. J Magn Reson. 2011;208:171‐177. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Olin RB, Sánchez‐Heredia JD, Schulte RF, et al. Three‐dimensional accelerated acquisition for hyperpolarized 13 C MR with blipped stack‐of‐spirals and conjugate‐gradient SENSE. Magn Reson Med. 2020;84:519‐534. [DOI] [PubMed] [Google Scholar]
- 7. Dominguez‐Viqueira W, Geraghty BJ, Lau JYC, Robb FJ, Chen AP, Cunningham CH. Intensity correction for multichannel hyperpolarized 13 C imaging of the heart. Magn Reson Med. 2016;75:859‐865. [DOI] [PubMed] [Google Scholar]
- 8. Ohliger MA, Gordon JW, Carvajal L, et al. 55 Mn‐based fiducial markers for rapid and automated RF coil localization for hyperpolarized 13 C MRI. Magn Reson Med. 2021;85:518‐530. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Arunachalam A, Whitt D, Fish K, et al. Accelerated spectroscopic imaging of hyperpolarized C‐13 pyruvate using SENSE parallel imaging. NMR Biomed. 2009;22:867‐873. [DOI] [PubMed] [Google Scholar]
- 10. Ohliger MA, Larson PEZ, Bok RA, et al. Combined parallel and partial fourier MR reconstruction for accelerated 8‐channel hyperpolarized carbon‐13 in vivo magnetic resonance spectroscopic imaging (MRSI). J Magn Reson Imaging. 2013;38:701‐713. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Gordon JW, Hansen RB, Shin PJ, Feng Y, Vigneron DB, Larson PEZ. 3D hyperpolarized C‐13 EPI with calibrationless parallel imaging. J Magn Reson. 2018;289:92‐99. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Grist JT, Riemer F, Hansen ESS, et al. Visualization of sodium dynamics in the kidney by magnetic resonance imaging in a multi‐site study. Kidney Int. 2020;98:1174‐1178. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Grist JT, Hansen ESS, Sánchez‐Heredia JD, et al. Creating a clinical platform for carbon‐13 studies using the sodium‐23 and proton resonances. Magn Reson Med. 2020;84:1817‐1827. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Kaggie JD, Lanz T, McLean MA, et al. Combined 23 Na and 13 C imaging at 3.0 tesla using a single‐tuned large FOV birdcage coil. Magn Reson Med. 2021;86:1734‐1745. [DOI] [PubMed] [Google Scholar]
- 15. Roemer PB, Edelstein WA, Hayes CE. The NMR phased array. Magn Reson Med. 1990;225:192‐225. [DOI] [PubMed] [Google Scholar]
- 16. Sánchez‐Heredia JD, Johansen DH, Hansen RB, et al. Improved decoupling for low frequency MRI arrays using non‐conventional preamplifier impedance. IEEE Trans Biomed Eng. 2019;66:1940‐1948. [DOI] [PubMed] [Google Scholar]
- 17. Reykowski A, Wright SM, Porter JR. Design of Matching Networks for low noise preamplifiers. Magn Reson Med. 1995;3:848‐852. [DOI] [PubMed] [Google Scholar]
- 18. Corea JR, Flynn AM, Lechêne B, et al. Screen‐printed flexible MRI receive coils. Nat Commun. 2016;7:10839. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Zhang B, Sodickson DK, Cloos MA. A high‐impedance detector‐array glove for magnetic resonance imaging of the hand. Nat Biomed Eng. 2018;2:570‐577. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Ruytenberg T, Webb A, Zivkovic I. A flexible five‐channel shielded‐coaxial‐cable (SCC) transceive neck coil for high‐resolution carotid imaging at 7T. Magn Reson Med. 2020;84:1672‐1677. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Sanchez‐Heredia JD, Szocska Hansen ES, Laustsen C, Zhurbenko V, Ardenkjær‐Larsen JH. Low‐noise active decoupling circuit and its application to 13C cryogenic RF coils at 3T. Tomography. 2017;3:60‐66. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Sánchez‐Heredia JD, Olin RB, McLean MA, et al. Multi‐site benchmarking of clinical 13C RF coils at 3T. J Magn Reson. 2020;318:106798. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Schulte RF, Sacolick L, Deppe MH, et al. Transmit gain calibration for nonproton MR using the Bloch‐Siegert shift. NMR Biomed. 2011;24:1068‐1072. [DOI] [PubMed] [Google Scholar]
- 24. Schulte RF, Sperl JI, Weidl E, et al. Saturation‐recovery metabolic‐exchange rate imaging with hyperpolarized [1‐ 13 C] pyruvate using spectral‐spatial excitation. Magn Reson Med. 2013;69:1209‐1216. [DOI] [PubMed] [Google Scholar]
- 25. http://hansenms.github.io/sunrise/sunrise2013/ Accessed 2018‐03‐27.
- 26. Gordon JW, Autry AW, Tang S, et al. A variable resolution approach for improved acquisition of hyperpolarized 13C metabolic MRI. Magn Reson Med. 2020;84:2943‐2952. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Hansen RB, Sánchez‐Heredia JD, Bøgh N, et al. Coil profile estimation strategies for parallel imaging with hyperpolarized 13C MRI. Magn Reson Med. 2019;82:2104‐2117. [DOI] [PubMed] [Google Scholar]
- 28. Ardenkjaer‐Larsen JH, Leach AM, Clarke N, Urbahn J, Anderson D, Skloss TW. Dynamic nuclear polarization polarizer for sterile use intent. NMR Biomed. 2011;24:927‐932. [DOI] [PubMed] [Google Scholar]
- 29. Autry AW, Gordon JW, Carvajal L, et al. Comparison between 8‐ and 32‐channel phased‐array receive coils for in vivo hyperpolarized 13 C imaging of the human brain. Magn Reson Med. 2019;82(2):833‐841. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Mirkes CC, Hoffmann J, Shajan G, Pohmann R, Scheffler K. High‐resolution quantitative sodium imaging at 9.4 tesla. Magn Reson Med. 2015;73:342‐351. [DOI] [PubMed] [Google Scholar]
- 31. Zaric O, Juras V, Szomolanyi P, et al. Frontiers of sodium MRI revisited: from cartilage to brain imaging. J Magn Reson Imaging. 2021;54:58‐75. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Grist JT, Riemer F, McLean MA, et al. Imaging intralesional heterogeneity of sodium concentration in multiple sclerosis: initial evidence from 23 Na‐MRI. J Neurol Sci. 2018;387:111‐114. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Platt T, Umathum R, Fiedler TM, et al. In vivo self‐gated 23 Na MRI at 7 T using an oval‐shaped body resonator. Magn Reson Med. 2018;80:1005‐1019. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.