Portable magnetic resonance imaging of patients indoors, outdoors and at home

Teresa Guallart-Naval 1,#, José M Algarín 2,3,#, Rubén Pellicer-Guridi 4,5, Fernando Galve 2,3, Yolanda Vives-Gilabert 4,6, Rubén Bosch 1, Eduardo Pallás 2,3, José M González 1, Juan P Rigla 1, Pablo Martínez 1, Francisco J Lloris 4, Jose Borreguero 1, Álvaro Marcos-Perucho 7, Vlad Negnevitsky 8, Luis Martí-Bonmatí 9, Alfonso Ríos 1, José M Benlloch 2,3, Joseba Alonso 2,3,
PMCID: PMC9338984  PMID: 35907975

Abstract

Mobile medical imaging devices are invaluable for clinical diagnostic purposes both in and outside healthcare institutions. Among the various imaging modalities, only a few are readily portable. Magnetic resonance imaging (MRI), the gold standard for numerous healthcare conditions, does not traditionally belong to this group. Recently, low-field MRI technology companies have demonstrated the first decisive steps towards portability within medical facilities and vehicles. However, these scanners’ weight and dimensions are incompatible with more demanding use cases such as in remote and developing regions, sports facilities and events, medical and military camps, or home healthcare. Here we present in vivo images taken with a light, small footprint, low-field extremity MRI scanner outside the controlled environment provided by medical facilities. To demonstrate the true portability of the system and benchmark its performance in various relevant scenarios, we have acquired images of a volunteer’s knee in: (i) an MRI physics laboratory; (ii) an office room; (iii) outside a campus building, connected to a nearby power outlet; (iv) in open air, powered from a small fuel-based generator; and (v) at the volunteer’s home. All images have been acquired within clinically viable times, and signal-to-noise ratios and tissue contrast suffice for 2D and 3D reconstructions with diagnostic value. Furthermore, the volunteer carries a fixation metallic implant screwed to the femur, which leads to strong artifacts in standard clinical systems but appears sharp in our low-field acquisitions. Altogether, this work opens a path towards highly accessible MRI under circumstances previously unrealistic.

Subject terms: Magnetic resonance imaging, Imaging techniques

Introduction

Standard clinical MRI scanners make use of powerful superconducting magnets that interact strongly with the vast amount of hydrogen nuclei in the human body1. These magnets enable the high SNR and spatial resolution typical for magnetic resonance images. Regrettably, these magnets also require cryogenic refrigeration, they are bulky, heavy, expensive to build, site, operate and maintain, and they ultimately constitute a formidable barrier to the accessibility and democratization of MRI24. Besides, high-field scanners are subject to patient safety risks, e.g. due to projectile incidents5; they are limited in the imaging pulse sequences that can be played out due to increased specific absorption rates (SAR) of electromagnetic energy in tissues at the corresponding higher excitation radio-frequencies (RF)6; they generate undesirable acoustic noise due to strong magnetic interactions during scans7; and they induce severe image artifacts around metallic implants due to magnetic susceptibility effects810. Low-field systems (<0.3 T) can overcome all of the above and are nowadays gaining traction as affordable complements to standard MRI scanners. Recent achievements with low-field scanners include in vivo brain and extremity imaging11,12, hard-tissue imaging1315 and even quantitative MRI and fingerprinting16,17. The main penalty to pay for operating in this regime is a significant loss in SNR and spatial resolution. However, the diagnostic value of the resulting reconstructions is not necessarily compromised, due to a number of reasons: (i) contrast-to-noise ratio (CNR), a more relevant metric for diagnosis than SNR, does not depend as strongly on field strength for some relevant contrast mechanisms18,19; (ii) multiple health conditions and diseases may be diagnosed without the exquisite detail provided by high-field images2; (iii) SAR constraints are less pronounced at low fields, allowing for efficient pulse sequences which increase the duty cycle to partly compensate the SNR loss2; and (iv) machine learning algorithms can be trained to recover image quality from noise-corrupted low-field data by e.g. transfer learning20,21.

The scope of conceivable applications for MRI technologies widens extraordinarily once the need for large superconducting magnets is removed. For instance, vehicles have been equipped with low-field systems22,23, and point-of-care and bedside neuroimaging have been demonstrated with a 64 mT FDA-cleared scanner24,25. The latter is arguably the most successful attempt at mobile MRI so far. However, it is based on a yoked magnet, which makes it heavy (>600 kg) and too large for standard door clear opening in residential constructions (32” in USA, 80 cm in Europe). Low-cost devices with improved mobility would enable MRI applications beyond clinical environments to home and hospice care, small clinics, rural areas or sports clubs and school facilities. Autonomously powered scanners could even be operated outdoors, e.g. in sports events, field hospitals or NGO and military camps26, making MRI available to a large fraction of the world population with no or insufficient access24.

In this article we present a 72 mT extremity MRI scanner based on a yokeless Halbach magnet mounted on a wheeled structure of width 70 cm, with an overall weight 250 kg and component cost <50 k€. After checking that the system performs as expected for in vivo images under controlled ambient conditions in an MRI physics laboratory, we took images of the right knee of a volunteer in different indoor and outdoor environments, including the living room at the volunteer’s apartment, and in open air connected to a portable gasoline generator. All knee images were acquired with identical 3-dimensional Rapid Acquisition with Relaxation Enhancement (3D-RARE) sequences, in about 12 min each. The electromagnetic interference (EMI) spectrum was different at the various locations, which results in slightly different noise patterns in the reconstructed images. Nevertheless, they all yield valuable anatomical information in clinically acceptable times. The volunteer had undergone a femoral shaft osteotomy and carried a fixation metallic implant screwed to the femur. This hardware is sharply defined in our low-field acquisitions, where previous high-field images suffered from strong susceptibility-induced image distortions.

Results

Scanner performance in laboratory

The system (Fig. 1) is built around a permanent magnet array in a Halbach configuration12, for a field strength of 72 mT. Technical details on the apparatus can be found in the Methods section. The scanner is usually in the controlled environment of an MRI physics laboratory, where the temperature is stabilized at 18.0±0.2 C and the air relative humidity at 45±10%. In these conditions, the Larmor frequency is stable down to the kilo-hertz at 3.076 MHz over weeks, and minor corrections to the frequency tuning and impedance matching electronics suffice to compensate for the different electronic loading of the RF coil by different subjects and body parts. The abundant surrounding electronic equipment and scanners generate substantial EMI in the laboratory at frequencies within our detection bandwidth. For this reason, we conceal the resonant RF coil behind three grounded shields: one is the outermost scanner housing (blue in Fig. 1e), a 1.5 mm-thick copper sheet; another is inside the magnet between the RF coil and the gradient assembly, consisting of a series of 0.1 mm-thick copper tape strips of width 5 cm and tin-soldered along the seams for electrical continuity; and last is an electrically conducting cloth (Holland Shielding Systems, Dordrecht, the Netherlands) that can be wrapped around the subject at both scanner ends, to avoid antenna effects that otherwise couple EMI to the coil from the inside, despite the other two shields.

Figure 1.

Figure 1

Photographs of the low-field extremity scanner: (a) 72 mT Halbach magnet; (b) gradient assembly; (c) RF Tx/Rx coil; (d) view of the scanner inside with a phantom in place; and (e) full system mounted on a transportable structure and in open air.

The images in Figs. 2 and 3 show the scanner performance in the laboratory and correspond to in vivo 3D-RARE acquisitions of different healthy subjects on different days. The images in Fig. 2 show selected slices of a left knee, a right hand and a right wrist, with acquisition times ranging from 10 to 19 min (Methods). All images show sufficient tissue contrast and spatial resolution to identify relevant anatomical features, including muscles, fat, cortical bone, bone marrow, tendons, ligaments, veins, arteries and fascia. In these images we show different contrast mechanisms1, with weightings on T1, T2 and proton density (ρ). The images are unprocessed after Fourier reconstruction and weak EMI effects result in a faint line along the horizontal (phase-encoded) direction in Fig. 2b. Figure 3 includes the complete set of slices of an axial knee acquisition, showing small distortions towards the edges of the field of view due to non-ideal field distributions.

Figure 2.

Figure 2

Single slices of 3D-RARE in vivo acquisitions of different volunteers in the MRI physics laboratory: (a) T1-weighted image of a knee, acquired in 19 min; (b) T1-weighted image of a hand (10 min), with a faint EMI line visible along the phase-encoded direction; (c)–(e) T1, ρ and T2-weighted images of a wrist (12 min).

Figure 3.

Figure 3

Complete set of axial slices of a T1-weighted 3D-RARE knee acquisition (11.5 min), showing small distortions towards the edges of the field of view, and some aliasing between the first (top left) and last (bottom right) images.

In vivo imaging of metallic implants

In a second set of experiments, we demonstrate in vivo MR images in the presence of metallic implants without the strong susceptibility-induced artifacts typical of high-field acquisitions8,10, which often hamper post-operative assessment of orthopedic procedures9. The volunteer for these tests had been diagnosed with lateral gonarthrosis due to cartilage damage in their right knee and had a femoral shaft osteotomy to remove pressure from the damaged tissue. The fixation metallic implant screwed to the femur is cleanly visible in a lateral X-ray computed radiographic image (Fig. 4c), but leads to high intensity fringes around the metallic hardware in high-field MR images due to incorrect spin mapping (see Fig. 4d, taken at 3 T). These effects depend supralinearly on the magnetic field strength and are barely perceptible at fields <0.1 T28. The field dependence is notorious in the images: the SNR and resolution are much higher in the 3 T system, but the metallic implant geometry is accurately defined in our 72 mT 2D and 3D reconstructions, and can be readily segmented with standard data post-processing. The low-field images were taken in 12 min (Fig. 4a,b) and 20 min (Fig. 4e) with T1-weighted 3D-RARE acquisitions (Methods).

Figure 4.

Figure 4

Images of fixation metallic implant attached to the femur, consisting of a plate and seven screws: (a) sagittal view of a raw low-field image acquired with the 72 mT system (9 mm slice from T1-weighted 3D-RARE acquisition with in-plane resolution of 1.3×2 mm2, 12 min scan time, eight years after femoral shaft osteotomy); (b) same, but BM4D-filtered27 and rescaled by ×2 to increase the number of pixels; (c) lateral X-ray computed radiography (two weeks after surgery); (d) sagittal view of the same knee, acquired with a Siemens Skyra 3 T system (T1-weighted 2D-RARE acquisition with slice thickness 3.9 mm and pixel resolution 0.26×0.26 mm2, one year after surgery); and (e) 3D reconstruction from T1-weighted 3D-RARE acquisition with isotropic resolution of 2 mm, 20 min scan time, where selected muscle and fat segments have been removed (eight years after surgery).

Indoor, outdoor and residential MRI

The goal of the last set of experiments is to evaluate the portability of the scanner and its performance under various environments and conditions. To establish comparisons as unbiased as possible, the acquisitions across all five scenarios are of the intervened knee of the same volunteer as in Fig. 4, and all with the same sequence parameters: T1-weighted 3D-RARE with a total scan time 12 min (Methods). The slices in Fig. 4 have been selected to show the third screw from the top (Fig. 4) as it runs through the bone from the top of the image, where the prosthetic plate is implanted. As a general indicator of the image quality, we measure the SNR in a region of interest (ROI) in the femur bone marrow (red boxes in the unprocessed reconstructions in Fig. 5). To this end, we estimate the signal strength as the average voxel brightness in the ROI, and the noise as the average voxel brightness in the background (white boxes). Prior to each acquisition, we measured the spectral noise density as picked up by the detection RF coil with the subject inside the scanner. The average signal strength in these spectra speaks of the white noise amplitude in the Rx chain, which is ideally close to thermal (Johnson) noise in the coil (Methods). Besides, we often encounter stronger peaks, indicating EMI at discrete frequencies. These can be suppressed by covering the subject meticulously with the shielding cloth.

Figure 5.

Figure 5

Photographs during acquisitions (left) and axial slice from 3D-RARE reconstructions (right, no post-processing) at five different locations: (a) in an MRI physics laboratory; (b) in an office room; (c) outside a campus building, connected to a nearby power outlet; (d) in open air, powered from a small fuel-based generator; and (e) at the volunteer’s home.

The first acquisition (Fig. 5a) took place in the same laboratory as above and serves as reference under controlled ambient conditions. For the acquisition in the MRI physics laboratory, the Larmor frequency was 3.076 MHz, the measured noise level (50 nV/Hz1/2) was compatible with Johnson noise (Methods), there is no visible EMI and the femur SNR is 21.

The second scan took place in an office room (Fig. 5b) around 20 m away from the laboratory, in the same building and floor. The Larmor frequency here decreased to 3.064 MHz due to a higher temperature. The noise amplitude is still consistent with Johnson noise levels and EMI is not visible in the reconstruction. The SNR in the marrow ROI is 25 and the overall image quality is comparable to the reference image, perhaps even slightly sharper.

The third image was acquired outdoors, at basement level, just outside the laboratory building (Fig. 5c). The system was powered through a 30 m cable running down three floors from the laboratory. The conducting cloth wrapped around the subject purposely connected the scanner shielding to the concrete floor to improve the otherwise resistive connection between the laboratory ground and earth. During this acquisition the volunteer reported sensing the presence and conversations of bypassers, a light breeze on the grounded cloth, and weak tremors due to vehicles driving through the underground parking. The resulting image quality seems not to be strongly influenced by any of these, with an SNR of 19 in the ROI, and a noise spectrum of comparable amplitude to indoor acquisitions. The Larmor frequency was 3.065 MHz.

The fourth scan was also taken outdoors, in this case in open air in a university campus (Fig. 5d, Larmor frequency 3.063 MHz), far from power outlets and operating autonomously with a portable electricity generator. The latter is based on a low-consumption gasoline engine, weighs <20 kg, costs <600 € and has an autonomy >10 hours with the scanner at continuous operation (Methods). We grounded the system electrically as before, with the conducting cloth offering low-resistance paths between the scanner shielding, the floor concrete and the ground terminal in the generator. The spectrum was significantly more populated in this case, with a mean amplitude roughly twice the expected Johnson limit, presumably due to noise originating at the engine. Consequently, the quality of the resulting image is lower than in the previous acquisitions (SNR11), and an EMI line is visible along the vertical (phase-encoding) direction. Nevertheless, the main anatomic features, different tissues and metallic implants are all still clearly identifiable.

The last image was acquired at the volunteer’s apartment. This is located in a low-density town in the province of Valencia, Spain. The system was transported in a small truck from the university campus to a parking lot approximately 300 m away from the entrance to the apartment block, and pushed along the sidewalk into the building, the elevator, the apartment and ultimately, the living room. Throughout the way, the only wheelchair-adapted elements were the lowered-sidewalks at pedestrian crossings. After transport, the system required re-connecting some of the RF electronics modules we had packed in a separate box, and tightening some screwed connectors that had become loose during transport through the rugged, tiled sidewalks. Other than that, the system was plugged to a wall power outlet, tuned to 3.065 MHz and ready to use. The noise spectrum in the apartment was clean, again compatible with Johnson noise levels. The SNR in the ROI for this acquisition is 19.

Discussion

In conclusion, we have demonstrated the viability of a portable, low-cost system for magnetic resonance imaging indoors, outdoors and at home. In this work, we have focused on healthy volunteers and subjects carrying metallic implants. Nevertheless, the acquired images contain sufficient anatomical information to diagnose a large variety of articular diseases, including effusion, synovial engorgement, tendon disruption or bone fractures.

System portability is a major goal for low-field systems, since this is not expected to be possible with high-field scanners in the near future. Our setup makes use of a Halbach magnet, as do others12,29,30. An important advantage of Halbach configurations is their reduced weight compared to yoked magnets. For instance, the 64 mT system from Hyperfine Inc. weighs >600 kg and the 55 mT system from Liu et al. around 750 kg31. In contrast, the weight of our system is comparable to that of a hospital bed (140 kg) with a patient (80 kg), making it amenable to transport by a single person on a flat surface. Therefore, even if the open design of yoked magnets eases patient handling and comfort (especially for neuroimaging), a Halbach configuration is arguably advantageous in terms of portability. Gradient efficiency is also improved in Halbach configurations in the sense that yoked magnets tend to make use of planar gradients to preserve the overall system openness. Our gradients are on cylindrical surfaces, which means that stronger gradients can be achieved for equivalent currents. Besides, we do not need the full power available from our gradient amplifiers, so one could consider trading efficiency for linearity, which may be useful for certain applications. Regarding the RF circuitry, the antennas employed in other low-field scanners are mostly dedicated head coils for neuroimaging applications. We have not yet explored this, because our scanner is somewhat small for head imaging. Finally, to complete this comparison with other existing low-field systems, we must stress that Hyperfine Inc. is well ahead of any other initiative, including ours, both in terms of having designed a final product and having certified it for clinical use. Nevertheless, future scanners with greatly enhanced portability will probably require the aforementioned benefits of Halbach magnets.

Looking ahead, our 72 mT scanner can be still upgraded in various ways. Machine learning algorithms have been shown to boost the performance in other low-field systems and can be readily incorporated to ours. These can be used, via transfer learning, to increase the spatial resolution of scans a posteriori based on multiple acquisitions, prior knowledge about the sample32, or with networks trained with paired datasets of low and high-field images, to recover from the former features visible otherwise only with the latter21,33. Deep learning and convolutional neural networks can also be employed to increase reconstruction quality through image denoising, artifact detection and active noise cancellation20,31,34. Quantitative MRI, radiomics and fingerprinting16,17,35 show promising potential in situations where subtle texture changes contain potentially valuable information for the patients. Also special-purpose pulse sequences and reconstruction methods can enhance the efficiency of low-field MRI13,17, and hardware developments and contrast agents which are mainstays in clinical high-field MRI (e.g. parallel imaging, optimization of RF detection coils for different body parts, gadolinium contrast enhancement), are seldom used in the still mostly experimental low-field systems available2,36. Finally, for our particular scanner, the GUI and overall system usability can be improved to facilitate operation by non-experts.

All in all, the scanner performance demonstrated in this work, especially if upgraded with the above capabilities, sets a path towards accessible MRI, democratizing its use and benefits, and qualitatively expanding the circumstances where it can provide value.

Methods

Scanner

The scanner employs a Halbach cylinder magnet including almost 4600 N48 NdFeB cubes of side 12 mm to generate B072 mT at the field of view, and another 1100 N42 smaller cuboids (64 mm3) to shim the inhomogeneity from  15,700 down to 3100 ppm over a spherical volume of 20 cm in diameter. This was designed for an inner diameter of 27 cm following methods described in Ref. 37, but we optimized for three magnet layers (rather than two) to increase the field strength from 50 to 72 mT. All in all, the magnet includes 23 rings held in place by 8 external screws traversing the complete setup. The Larmor frequency dependence on air temperature is -6 kHz/C.

The gradient coil geometry is optimized with conventional target-field methods, following the procedures described in Ref. 37. Our x (axial), y (vertical) and z (horizontal) gradients have, respectively: efficiencies 0.53, 0.91 and 0.89 mT/m/A; resistances 0.35, 0.38 and 0.40Ω with a wire of diameter 1.5 mm; inductances 180, 227 and 224μH; and deviations 27.2, 1.1 and 1.0 % from perfect linearity over a 15 cm DSV. These coils are wound on and glued to curved 3D-printed Nylon molds, and the whole assembly is supported by a methacrylate cylinder. We have not encountered the need for water or air cooling. The gradient analog waveforms are generated with an OCRA1 board38, connected to a Red Pitaya Stemlab 122.88-16 SDR39 via Serial Peripheral Interface (SPI), and amplified by AE Techron 7224 power amplifiers (Indiana, USA), which can deliver up to 45 A onto our loads at relevant duty cycles. This corresponds to gradient fields of up to 25 mT/m along x and 40 mT/m along the y and z. Under normal operating conditions, the Larmor frequency decreases by 10 Hz/min due to heating of the permanent magnets from the power dissipated by the gradient coils.

We used two Tx/Rx RF antennas, one for the images in Figs. 2 and 3 (of inner diameter 14 cm), and a larger one for the implanted knee (20 cm). Both are solenoid coils tuned and impedance-matched to the proton Larmor frequency (3 MHz). The RF coil holders were 3D-printed in polylactic acid (PLA), and the wire was fixed with cyanoacrylate adhesive. The coils are inside a grounded copper shield to mitigate noise pick-up and prevent interference between the gradients and the RF system, and a conductive cloth covers the subject during in vivo acquisitions. The RF low-noise (45 dB gain, 50ohm, noise figure <1 dB) and power amplifiers (250 W, maximum duty cycle of 10 % with 10 ms pulses), as well as the passive Tx/Rx switch, were purchased from Barthel HF-Technik GmbH (Aachen, Germany).

The control electronics are based on MaRCoS, an open-source, high-performance Magnetic Resonance Control System4042.

The diameter and length of the scanner are around 53 and 51 cm respectively, excluding electronics and the mobile structure, with a bore opening 24 cm (inner diameter of gradient structure) and a weight of 200 kg. Once on the mobile, open structure and equipped with all the required electronics and the control computer, the overall system dimensions are 70×88×166 cm3 and the weight is 250 kg.

Pulse sequences

The protocols for experiments involving human subjects were approved by the Ethics Committee (CEIm) of La Fe Hospital in Valencia (IIS-F-PG-22-02, research agreement number 2019-139-1).

Some aspects common to all the images presented in this work are: (i) the duration of resonant π/2 and π-pulses in all images are 40μs and 80μs, respectively; (ii) the readout dephasing gradient pulses after the RF excitation pulses are pre-emphasized by a factor 1.008 to place the echoes at the center of the data acquisition windows and mitigate the effects of imperfect gradient waveforms and induced Eddy currents; and (iii) an automatic Larmor-frequency calibration is run before every new scan, i.e. full sequence for an image.

The knee image in Fig. 2a was acquired with a T1-weighted 3D-RARE sequence, with FoV=130×140×180 mm3, a resolution of 1.85×1.75×2 mm3, ETL=5, TE=20 ms, TR=200 ms, BW=17.5 kHz, and 4 averages for a total scan time of 19.2 min. The x, y and z axes correspond to the readout (RO), phase-encoding (PE) and slice-encoding (SE) directions, respectively.

The hand image in Fig. 2b was acquired with a T1-weighted 3D-RARE sequence, with FoV=180×180×50 mm3, a resolution of 1.5×1.5×5 mm3, ETL=10, TE=20 ms, TR=400 ms, BW=30 kHz, and 13 averages for a total scan time of 10.4 min. The x, y and z axes correspond to the RO, SE and PE directions, respectively.

The wrist image in Fig. 2c was acquired with a T1-weighted 3D-RARE sequence, with FoV=180×140×80 mm3, a resolution of 1.5×1.5×10 mm3, ETL=3, TE=20 ms, TR=100 ms, BW=30 kHz, and 30 averages for a total scan time of 12 min. The x, y and z axes correspond to the RO, SE and PE directions, respectively.

The wrist image in Fig. 2d was acquired with a ρ-weighted 3D-RARE sequence, with FoV=180×140×80 mm3, a resolution of 1.5×1.5×10 mm3, ETL=5, TE=20 ms, TR=1000 ms, BW=30 kHz, and 5 averages for a total scan time of 12 min. The x, y and z axes correspond to the RO, SE and PE directions, respectively.

The wrist image in Fig. 2e was acquired with a T2-weighted 3D-RARE sequence, with FoV=180×140×80 mm3, a resolution of 1.5×1.5×10 mm3, ETL=5, echo spacing of 20 ms, effective TE=100 ms, TR=1000 ms, BW=30 kHz, and 5 averages for a total scan time of 12 min. The x, y and z axes correspond to the RO, SE and PE directions, respectively.

The knee images in Fig. 3 were acquired with a T1-weighted 3D-RARE sequence, with FoV=150×150×180 mm3, a resolution of 1.50×1.85×10 mm3, ETL=5, TE=20 ms, TR=200 ms, BW=25 kHz, and 12 averages for a total scan time of 11.5 min. The x, y and z axes correspond to the SE, PE and RO directions, respectively.

The knee images in Fig. 4a,b were acquired with a T1-weighted 3D-RARE sequence, with FoV=200×200×180 mm3, a resolution of 1.3×2×9 mm3, ETL=5, TE=20 ms, TR=200 ms, BW=37.5 kHz, and 9 averages for a total scan time of 12 min. The x, y and z axes correspond to the RO, PE and SE directions, respectively.

The knee image in Fig. 4e was acquired with a T1-weighted 3D-RARE sequence, with FoV=200×200×180 mm3, a resolution of 2×2×2 mm3, ETL=10, TE=20 ms, TR=300 ms, BW=22.5 kHz, and 4 averages for a total scan time of 20 min. The x, y and z axes correspond to the RO, PE and SE directions, respectively.

The knee images in Fig. 5 were acquired with a T1-weighted 3D-RARE sequence, with FoV=180×200×200 mm3, a resolution of 1.2×2×10 mm3, ETL=5, TE=20 ms, TR=200 ms, BW=37.5 kHz, and 9 averages for a total scan time of 12 min. The x, y and z axes correspond to the SE, PE and RO directions, respectively.

Data acquisition

The receive chain consists of an analog stage (RF coil, passive Tx/Rx switch and low-noise amplifier) followed by a digital stage. The digitization is performed at 122.88 Ms/s by an analog-to-digital converter in the Red Pitaya Stemlab board4042. The digital signal is mixed down by complex multiplication with a numerically-controlled oscillator set to the Larmor frequency. The real and imaginary data components pass first a cascaded integrator-comb filter and finally a finite impulse response filter. The resulting data conform the sought in-phase and quadrature components of the magnetic resonance signal. These are sent to the control computer and can be Fourier-transformed for image reconstruction and post-processing.

Image reconstruction and post-processing

All images have been reconstructed directly by an Inverse Fast Fourier Transform protocol implemented in the MaRCoS GUI we have developed in Python40. The presented reconstructions are therefore subject to distortions due to field inhomogeneity and gradient non-linearities. These can be mitigated by reconstruction algorithms which include information on the field maps43, but we have not found this necessary at this stage. The only post-processing operations we have used in this work are BM4D-filtering27 and image rescaling to increase the number of pixels, and only where explicitly indicated in the main text.

Noise

The spectral noise density of the MR data is bounded from below by Johnson noise due to thermal fluctuations of electrons in the resistive elements R in the receive chain (up to the LNA). These are dominated by the coil, with quality factor Q93 (88) and R5 (5.5Ω) in the unloaded (loaded) case. For a given acquisition bandwidth, the integrated noise amplitude is expected to be (4kBR·BW)1/2, with kB the Boltzmann constant. In the controlled environment of the MRI physics laboratory, we measure 50 nV/Hz1/2 after a 45 dB low-noise pre-amplifier, in agreement with the estimated Johnson level. We use this as a reference to evaluate the signal quality and the shielding efficiency of the conductive cloth, both in the laboratory and in the rest of locations.

We have found situations where suppressing noise down to Johnson levels is not trivial, and indeed did not achieve it when the system was powered by the portable generator. The control computer is another significant source of 50 Hz noise and needs to be as far as possible in the rack to reconstruct clean images. We also find it often necessary to ensure the subject is sufficiently covered by the conductive cloth, and extending some of it on the floor helps.

Generator

For the autonomous experiments outdoors we powered the system from a “Limited 2000i” gasoline-fueled generator from Genergy (Calahorra, Spain). This motor delivers up to 2 kW at 230 V and 50 Hz (single phase). It costs <600 €, weighs 19 kg and has a fuel tank capacity of 4 l and an autonomy of 10.8 hours at 25 % load (500 W), which is more than required for continuous operation of the scanner.

Ethical approval

All experiments were carried out in accordance with the guidelines in the Declarations of Helsinki and following Spanish regulations and under the research agreement from La Fe Hospital in Valencia (IIS-F-PG-22-02, agreement number 2019-139-1).

Consent to participate

Written informed consent was obtained from all subjects prior to study commencement.

Consent for publication

Written informed consent for publication was obtained from all subjects prior to study commencement.

Acknowledgements

We thank the four anonymous volunteers for their participation, Andrew Webb and Thomas O’Reilly for discussions on MRI hardware, Benjamin Menküc for contributions to MaRCoS, and Marcus Prier for the development of the OCRA1 board.

Author contributions

Low-field images taken by T.G.N. and J.M.A. with help from R.B., P.M., F.J.L., J.P.R. and J.A. Portable system built by R.P.G., T.G.N., J.M.A., F.G., R.B., E.P., J.M.G. and J.A. with help from J.P.R., P.M. and F.J.L. Data analysis and evaluation by J.M.A., T.G.N., F.G., L.M.B. and J.A. Control electronics and software developed by V.N., Y.V.G., J.M.A., T.G.N. and J.B. Portability experiments conceived by J.A., J.M.A., T.G.N. and A.M.P. Project conceived and supervised by A.R., J.M.B. and J.A.. Paper written by J.A. and L.M.B., with input from all authors.

Funding

This work was supported by the Ministerio de Ciencia e Innovación of Spain through research grant PID2019-111436RB-C21. Action co-financed by the European Union through the Programa Operativo del Fondo Europeo de Desarrollo Regional (FEDER) of the Comunitat Valenciana 2014-2020 (IDIFEDER/2018/022 and IDIFEDER/2021/004). JMG and JB acknowledge support from the Innodocto program of the Agencia Valenciana de la Innovación (INNTA3/2020/22 and INNTA3/2021/17).

Data availability

All anonymized datasets, reconstruction and post-processing methods generated and/or used during the present study are available from the corresponding author upon reasonable request.

Code availability

The codes for MaRCoS and the GUI are publicly available from open-source repositories at https://github.com/vnegnev/marcos_server and https://github.com/yvives/PhysioMRI_GUI, respectively.

Competing interests

PhysioMRI Tech S.L. is a for-profit organization spun off the Institute for Molecular Imaging and Instrumentation and proprietor of the low-field scanner presented in this work. JMA, FG, JB, JMB and JA have patents pending that are licensed to PhysioMRI Tech S.L. JMA, FG, AR, JMB and JA are co-founders of PhysioMRI Tech S.L. All other authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

These authors contributed equally: Teresa Guallart-Naval and José M. Algarín.

References

  • 1.Haacke EM, Brown RW, Thompson MR, Venkatesan R, et al. Magnetic Resonance Imaging: Physical Principles and Sequence Design. New York: Wiley; 1999. [Google Scholar]
  • 2.Marques JP, Simonis FF, Webb AG. Low-field MRI: An MR physics perspective. J. Magn. Reson. Imaging. 2019;49(6):1528–1542. doi: 10.1002/jmri.26637. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Sarracanie M, Salameh N. Low-field MRI: How low can we go? A fresh view on an old debate. Front. Phys. 2020;8:172. doi: 10.3389/fphy.2020.00172. [DOI] [Google Scholar]
  • 4.Wald LL, McDaniel PC, Witzel T, Stockmann JP, Cooley CZ. Low-cost and portable MRI. J. Magn. Reson. Imaging. 2020;52(3):686–696. doi: 10.1002/JMRI.26942. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Watson RE. Lessons learned from MRI safety events. Curr. Radiol. Rep. 2015;3(10):1–7. doi: 10.1007/S40134-015-0122-Z. [DOI] [Google Scholar]
  • 6.Panych LP, Madore B. The physics of MRI safety. J. Magn. Reson. Imaging. 2018;47(1):28–43. doi: 10.1002/JMRI.25761. [DOI] [PubMed] [Google Scholar]
  • 7.Price DL, De Wilde JP, Papadaki AM, Curran JS, Kitney RI. Investigation of acoustic noise on 15 MRI scanners from 0.2 T to 3 T. J. Mag. Reson. Imaging Off. J. Int. Soc. Mag. Reson. Med. 2001 doi: 10.1002/1522-2586. [DOI] [PubMed] [Google Scholar]
  • 8.Lüdeke KM, Röschmann P, Tischler R. Susceptibility artefacts in NMR imaging. Magn. Reson. Imaging. 1985;3(4):329–343. doi: 10.1016/0730-725X(85)90397-2. [DOI] [PubMed] [Google Scholar]
  • 9.Harris CA, White LM. Metal artifact reduction in musculoskeletal magnetic resonance imaging. Orthop. Clin. North Am. 2006;37(3):349–359. doi: 10.1016/J.OCL.2006.04.001. [DOI] [PubMed] [Google Scholar]
  • 10.Stradiotti P, Curti A, Castellazzi G, Zerbi A. Metal-related artifacts in instrumented spine. Techniques for reducing artifacts in CT and MRI: State of the art. Eur. Spine J. 2009;18(SUPPL. 1):102–108. doi: 10.1007/S00586-009-0998-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Cooley CZ, McDaniel PC, Stockmann JP, Srinivas SA, Cauley SF, Śliwiak M, Sappo CR, Vaughn CF, Guerin B, Rosen MS, Lev MH, Wald LL. A portable scanner for magnetic resonance imaging of the brain. Nat. Biomed. Eng. 2020;5(3):229–239. doi: 10.1038/s41551-020-00641-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.O’Reilly T, Teeuwisse WM, Gans D, Koolstra K, Webb AG. In vivo 3D brain and extremity MRI at 50 mT using a permanent magnet Halbach array. Mag. Reson. Med. 2020 doi: 10.1002/mrm.28396. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Algarín JM, Díaz-Caballero E, Borreguero J, Galve F, Grau-Ruiz D, Rigla JP, Bosch R, González JM, Pallás E, Corberán M, Gramage C, Aja-Fernández S, Ríos A, Benlloch JM, Alonso J. Simultaneous imaging of hard and soft biological tissues in a low-field dental MRI scanner. Sci. Rep. 2020;10(1):21–470. doi: 10.1038/s41598-020-78456-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Borreguero, J., González, J. M., Pallás, E., Rigla J. P., Algarín, J. M., Bosch, R., Galve, F., Grau-Ruiz, D., Pellicer, R., Ríos, A., Benlloch, J. M., Alonso, J. Prepolarized MRI of hard tissues and solid-state matter. NMR Biomed. e4737 (2022). 10.1002/NBM.4737 [DOI] [PMC free article] [PubMed]
  • 15.Borreguero, J., Galve, F., Algarín, J. M., Benlloch, J. M., Alonso,J. Slice-selective zero echo time imaging of ultra-short T2 tissues based on spin-locking. arXiv:2201.06305 (2022). [DOI] [PMC free article] [PubMed]
  • 16.O’Reilly T, Webb AG. In vivo T1 and T2 relaxation time maps of brain tissue, skeletal muscle, and lipid measured in healthy volunteers at 50 mT. Magn. Reson. Med. 2021;87(2):884–895. doi: 10.1002/MRM.29009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Sarracanie M. Fast quantitative low-field magnetic resonance imaging with OPTIMUM - optimized magnetic resonance fingerprinting using a stationary steady-state cartesian approach and accelerated acquisition schedules. Invest. Radiol. 2021 doi: 10.1097/RLI.0000000000000836. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Rutt BK, Lee DH. The impact of field strength on image quality in MRI. J. Magn. Reson. Imaging. 1996;6(1):57–62. doi: 10.1002/JMRI.1880060111. [DOI] [PubMed] [Google Scholar]
  • 19.Ghazinoor S, Crues JV, Crowley C. Low-field musculoskeletal MRI. J. Magn. Reson. Imaging. 2007;25(2):234–244. doi: 10.1002/jmri.20854. [DOI] [PubMed] [Google Scholar]
  • 20.Koonjoo N, Zhu B, Bagnall GC, Bhutto D, Rosen MS. Boosting the signal-to-noise of low-field MRI with deep learning image reconstruction. Sci. Rep. 2021;11(1):1–16. doi: 10.1038/s41598-021-87482-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Garcia Hernandez, A., Fau,P., Rapacchi, S., Wojak, J., Mailleux,H., Benkreira, M., Adel, M. Improving Image Quality in Low-Field MRI with Deep Learning 60–263 (2021). 10.1109/ICIP42928.2021.9506659
  • 22.Nakagomi M, Kajiwara M, Matsuzaki J, Tanabe K, Hoshiai S, Okamoto Y, Terada Y. Development of a small car-mounted magnetic resonance imaging system for human elbows using a 0.2 T permanent magnet. J. Mag. Reson. 2019;304:1–6. doi: 10.1016/j.jmr.2019.04.017. [DOI] [PubMed] [Google Scholar]
  • 23.Deoni SC, Deoni AT, Burton P, Beauchemin J, D’Sa V, Boskamp E, Samantha B, McNulty C, Mileski W, Welch BE, Huentelman M. Residential MRI: Development of a mobile anywhere-everywhere MRI lab. Res. Sq. 2021 doi: 10.21203/RS.3.RS-1121934/V1. [DOI] [Google Scholar]
  • 24.Sheth KN, Mazurek MH, Yuen MM, Cahn BA, Shah JT, Ward A, Kim JA, Gilmore EJ, Falcone GJ, Petersen N, Gobeske KT, Kaddouh F, Hwang DY, Schindler J, Sansing L, Matouk C, Rothberg J, Sze G, Siner J, Rosen MS, Spudich S, Kimberly WT. Assessment of brain injury using portable, low-field magnetic resonance imaging at the bedside of critically ill patients. JAMA Neurol. 2021;78(1):41–47. doi: 10.1001/JAMANEUROL.2020.3263. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Mazurek MH, Yuen MM, Cahn BA, Rosen MS, Gobeske KT, Gilmore EJ, Hwang D, Kaddouh F, Kim JA, Falcone G, Petersen N, Siner J, Spudich S, Sze G, Kimberly WT, Sheth KN. Low-field, portable magnetic resonance imaging at the bedside to assess brain injury in patients with severe COVID-19 (1349) Neurology. 2021 doi: 10.7759/CUREUS.15841. [DOI] [Google Scholar]
  • 26.Sarracanie M, LaPierre CD, Salameh N, Waddington DEJ, Witzel T, Rosen MS. Low-cost high-performance MRI. Sci. Rep. 2015;5(1):15–177. doi: 10.1038/srep15177. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Maggioni M, Katkovnik V, Egiazarian K, Foi A. Nonlocal transform-domain filter for volumetric data denoising and reconstruction. IEEE Trans. Image Process. 2013;22(1):119–133. doi: 10.1109/TIP.2012.2210725. [DOI] [PubMed] [Google Scholar]
  • 28.Van Speybroeck C, O’Reilly T, Teeuwisse W, Arnold P, Webb A. Characterization of displacement forces and image artifacts in the presence of passive medical implants in low-field (<100 mT) permanent magnet-based MRI systems, and comparisons with clinical MRI systems. Phys. Med. 2021;84:116–124. doi: 10.1016/j.ejmp.2021.04.003. [DOI] [PubMed] [Google Scholar]
  • 29.ZimmermanCooley C, Haskell MW, Cauley SF, Sappo C, Lapierre CD, Ha CG, Stockmann JP, Wald LL. Design of sparse halbach magnet arrays for portable MRI using a genetic algorithm. IEEE Trans. Mag. 2017 doi: 10.1109/TMAG.2017.2751001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Purchase AR, Vidarsson L, Wachowicz K, Liszkowski P, Sun H, Sarty GE, Sharp JC, Tomanek B. A short and light, sparse dipolar Halbach magnet for MRI. IEEE Access. 2021;9:95294–95303. doi: 10.1109/ACCESS.2021.3093530. [DOI] [Google Scholar]
  • 31.Liu Y, Leong ATL, Zhao Y, Xiao L, Mak HKF, Chun A, Tsang O, Lau GKK, Leung GKK, Wu EX, Linfang X. A low-cost and shielding-free ultra-low-field brain MRI scanner. Nat. Commun. 2021;12(1):1–14. doi: 10.1038/s41467-021-27317-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Van Reeth E, Tham IWK, Tan CH, Poh CL. Super-resolution in magnetic resonance imaging: A review. Concepts Mag. Reson. Part A. 2012;40A(6):306–325. doi: 10.1002/cmr.a.21249. [DOI] [Google Scholar]
  • 33.Iglesias J. E, Schleicher R., Laguna S., Billot B., Schaefer P., McKaig, B., Goldstein,J. N., Sheth, K. N., Rosen, M. S., Kimberly, W. T. Accurate super-resolution low-field brain mri. arXiv preprint arXiv:2202.03564 (2022). [DOI] [PMC free article] [PubMed]
  • 34.Küstner T, Liebgott A, Mauch L, Martirosian P, Bamberg F, Nikolaou K, Yang B, Schick F, Gatidis S. Automated reference-free detection of motion artifacts in magnetic resonance images. Magn. Reson. Mater. Phys. Biol. Med. 2018;31(2):243–256. doi: 10.1007/S10334-017-0650-Z/FIGURES/10. [DOI] [PubMed] [Google Scholar]
  • 35.Simpson G, Spieler B, Dogan N, Portelance L, Mellon EA, Kwon D, Ford JC, Yang F. Predictive value of 0.35 T magnetic resonance imaging radiomic features in stereotactic ablative body radiotherapy of pancreatic cancer: A pilot study. Med. Phys. 2020;47(8):3682–3690. doi: 10.1002/MP.14200. [DOI] [PubMed] [Google Scholar]
  • 36.Waddington DE, Boele T, Maschmeyer R, Kuncic Z, Rosen MS. High-sensitivity in vivo contrast for ultra-low field magnetic resonance imaging using superparamagnetic iron oxide nanoparticles. Sci. Adv. 2020;6(29):998–1015. doi: 10.1126/sciadv.abb0998. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.O’Reilly T, Teeuwisse W, Webb A. Three-dimensional MRI in a homogenous 27 cm diameter bore Halbach array magnet. J. Mag. Reson. 2019;307:106–578. doi: 10.1016/j.jmr.2019.106578. [DOI] [PubMed] [Google Scholar]
  • 38.OCRA1 - SPI controlled 4 channel 18 BIT DAC and RF attenuator. https://zeugmatographix.org/ocra/2020/11/27/ocra1-spi-controlled-4-channel-18bit-dac-and-rf-attenutator/
  • 39.STEMlab 122.88-16 SDR kit basic. https://www.redpitaya.com/p52/stemlab-12288-16-sdr-kit-basic
  • 40.Guallart-Naval T., et al., Benchmarking the performance of a low-cost Magnetic Resonance Control System at multiple sites in the open MaRCoS community. arXiv preprint arXiv:2203.11314 (2022). [DOI] [PMC free article] [PubMed]
  • 41.Negnevitsky, V., O’Reilly, T., Pellicer-Guridi, R., Vives-Gilabert, Y., Craven-Brightman, L., Schote, D., Algarín, J. M., Prier, M., Stockmann, J., Witzel, T., Menküc, B., Alonso, J., Webb, A. in Book of Abstracts ESMRMB 2021 38th Annual Scientific Meeting, Vol. 34, 172. (Springer, New York, 2021). 10.1007/s10334-021-00947-8
  • 42.Craven-Brightman, L., O’Reilly, T., Menküc, B., Prier, M., Pellicer-Guridi, R., Alonso, J., Wald, L. L., Zaitsev, M., Stockmann, J., Witzel, T., Webb, A., Negnevitsky, V. in Proceedings of the 2021 ISMRM & SMRT Annual Meeting and Exhibition, Abstract 0748 (ISMRM, 2021). https://cds.ismrm.org/protected/21MPresentations/abstracts/0748.html
  • 43.Koolstra K, O’Reilly T, Börnert P, Webb A. Image distortion correction for MRI in low field permanent magnet systems with strong B0 inhomogeneity and gradient field nonlinearities. Magn. Reson. Mater. Phys. Biol. Med. 2021;34(4):631–642. doi: 10.1007/S10334-021-00907-2. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

All anonymized datasets, reconstruction and post-processing methods generated and/or used during the present study are available from the corresponding author upon reasonable request.

The codes for MaRCoS and the GUI are publicly available from open-source repositories at https://github.com/vnegnev/marcos_server and https://github.com/yvives/PhysioMRI_GUI, respectively.