Abstract
Sensors, such as optical, chemical, and electrical sensors, play an important role in our lives. While these sensors already have widespread applications, such as humidity sensors, most are generally incompatible with flexible/inactive substrates and rely on conventional hard materials and complex manufacturing processes. To overcome this, we develop a CNT-based, low-resistance, and flexible humidity sensor. The core–shell structured CNT@CPM is prepared with Chit and PAMAM to achieve reliability, accuracy, consistency, and durability, resulting in a highly sensitive humidity sensor. The average response/recovery time of optimized sensor is only less than 20 s, with high sensitivity, consistent responsiveness, good linearity according to humidity rates, and low hysteresis (− 0.29 to 0.30 %RH). Moreover, it is highly reliable for long-term (at least 1 month), repeated bending (over 15,000 times), and provides accurate humidity measurement results. We apply the sensor to smart-wear, such as masks, that could conduct multi-respiratory monitoring in real-time through automatic ventilation systems. Several multi-respiratory monitoring results demonstrate its high responsiveness (less than 1.2 s) and consistent performance, indicating highly desirable for healthcare monitoring. Finally, these automatic ventilation systems paired with flexible sensors and applied to smart-wear can not only provide comfort but also enable stable and accurate healthcare in all environments.
Supplementary Information
The online version contains supplementary material available at 10.1186/s40580-022-00326-6.
Keywords: Wearable humidity sensor, Flexible sensor, Core–shell structure, Resistive type sensor, Multiple respiratory monitoring
Introduction
Wearing of face masks by the general population has become a common and ubiquitous daily lifestyle worldwide. The daily wearing of personal protective equipment such as masks creates moisture inside owing to sweat and differences in internal and external temperature. Electrostatics-based masks are especially vulnerable to humidity, reducing the performance of the mask, and causing not only discomfort and nervous temperament to the wearer but also harm to the skin [1, 2].
Recently, many researchers have made great efforts to eliminate this inconvenience, including sensors and ventilation systems with additional filters. However, most of the early wearable sensors added to various accessories have been focused on function and performance rather than user convenience [3, 4]. In addition, owing to the lack of batteries, wireless communication technology, and measurement efficiency, these were operated internally via various communication devices, and sensors were also based on existing medical sensors, which caused great inconvenience to users. While the target performance of these state-of-the-art sensors has already been reached, most sensors so far have generally been incompatible with flexible/variable substrates and have relied on conventional inorganic hard materials and complex fabrication processes [5]. Multi-functional sensors with flexibility and scalability are rapidly being developed using various properties of organic materials, and the design of wearable sensors has become a major research topic. Wearable sensors can simplify daily life procedures and provide useful tools for on-site monitoring of internal and external parameters. These next-generation sensors must have low cost, low weight, and low energy to ensure compatibility with wearable technologies [6, 7].
To overcome this bottleneck, many researchers have been striving to develop more flexible and smaller bands, patches, clothing, and contact lenses that are easier for users to wear [8, 9]. Recently, the development of flexible electronic sensors has attracted interest for potential applications in industry and agriculture, as well as in personal healthcare, health assessment, and sports monitoring [10–14]. Meanwhile, flexible humidity sensors are rarely reported compared to other sensors because of a lack of recognition of their importance, despite humidity being one of the most intensively measured variables in our lives, as it is associated with a comfortable living environment, medical facilities, and body health information, among others [5]. However, flexible humidity sensors are promising candidates for potential applications in skin-like electronics, robotic human–machine interfaces, wearable electronic systems, soft robotics, and biomedical applications [15–18]. As a result, many studies on flexible humidity sensors are required. Humidity sensors work on two main operating principles: capacitive and resistive sensing methods [19]. Capacitive sensors are the most commonly used sensors and have the advantage of being quick and simple to react to; however, their accuracy is low owing to their large error range. Resistive sensors can be accurately measured; however, their measurement efficiency and response speed decrease depending on the range of resistance [20]. To compensate for the drawbacks of capacitive and resistive sensors, many studies have been conducted using carbon-based materials such as graphene, carbon nanotubes (CNTs), carbon black, and carbon fibers. However, despite many of these studies, flexibility and efficiency are limited by the rigidity and high resistance of the flexible humidity sensor [21, 22].
Among various materials to overcome the above limitations, the core–shell structural material has the most efficient structure competent of forming organic–inorganic nanocomposites that have attracted attention in various fields, including sensors [23]. Core–shell structured materials can be expected to have many advantages such as enhanced electrical properties, versatility, material stability, dispersibility, and minimized consumption of valuable materials based on a large surface part due to an efficient structure [24]. Therefore, it is the most attractive structure competent of overcoming the limitations of existing humidity sensors.
Here, we aim to develop a flexible humidity sensor that can measure precise humidity intuitively and stably by enabling humidity measurement with very fast and consistent resistance changes depending on the atmospheric humidity. To facilitate this, we employed a complex of chitosan (Chit) and PAMAM dendrimer G3 (PM) as an organic phase that acted as an adhesive to coat individual carbon nanotubes (CNTs) uniformly. The materials formed were characterized by a core–shell structure encased in numerous amines of Chit and PM, and these amines allowed for flexible properties and quick reactions with moisture through self-assembly. When applied to smart wear, such as face masks, we confirmed that real-time multi-respiratory monitoring and ventilation systems showed very accurate and consistent outcomes.
Experimental section
Materials
Pristine multi-walled carbon nanotubes (CNTs, > 95%, 20–30 nm of diameter, 10–30 μm of length) were purchased from Carbon Nanotech Co. (Pohang, Korea). CNTs were refluxed in a 5 N HCl solution for 1 d prior to use and washed in deionized water. All chemicals, chitosan (Chit, low molecular weight, 75–85% degree of deacetylation, 50–190 kDa), PAMAM dendrimer, ethylene core, generation 3.0 solution (PM), and including organic solvents, which were purchased from Sigma-Aldrich (St. Louis, MA, USA), were of analytical grade and used without further purification.
Fabrication of PM-embedded Chit/CNT nanocomposites (CNT@CPM)
In order to fabricate CNT@CPM nanocomposites, each CNT was wrapped with PM-embedded Chit molecules in a core–shell structure that was prepared as follows: First, 45 mg of Chit was dissolved in 10 ml of acetic acid (1 v/v%) to make a Chit solution. 15 mg of CNT and each mol% of PM (Table 1) were then homogenized in the Chit solution using a homogenizer (MN600P-200, Micronox Corp., Seongnam, South Korea). This acidic CNT@CPM solution was slowly neutralized by 1 N ammonia solution, followed by dialysis membrane tube at a molecular weight cut-off of 12,000–14,000 Da (Spectrum Laboratories, Savannah, GA, USA) against deionized water for 3 d to remove any small molecules including organic and inorganic side products. The aqueous CNT@CPM solution showed a high degree of dispersion and stably with high hydrophilicity. (Additional file 1: Fig. S1).
Table 1.
Code name | CNT (mg) | Chit (mg) | PM (mg / wt%) | Free amine ratio (times) |
---|---|---|---|---|
CNT@CPM-1 | 15 | 45 | 0 / 0 | 1 |
CNT@CPM-2 | 1.8 / 3 | 40.5 | ||
CNT@CPM-3 | 3.0 / 5 | 66.8 | ||
CNT@CPM-4 | 6.0 / 10 | 132.7 |
Immobilization of CNT@CPM nanocomposites on the flexible polyimide surface of a gold electrode
CNT@CPM nanocomposites were immobilized on the gold electrode with a flexible polyimide surface. The flexible substrate was first washed ultrasonically in deionized water for 30 s. CNT@CPM nanocomposite solution (20 µL) was dropped on a gold electrode and dried at 60 °C. The resulting CNT@CPM nanocomposite-coated electrodes were thoroughly rinsed with water and absolute ethanol to remove any small molecules, including inorganic side products.
Characterizations
The Chit, CNT, PM, and CNT@CPM nanocomposites were characterized through qualitative analysis using Fourier transform infrared spectroscopy (FT-IR Perkin-Elmer, Spectrum BXII FT-IR spectrometer, USA), which was repeated 16 times under potassium bromide pellet conditions over a spectral range of 400–4000 cm−1 at 4 cm−1 resolution. Raman spectroscopy (RAMAN DXR2xi, ThermoFisher, USA) was repeated 16 times under probing conditions over a spectral range of 50–3500 cm−1 at a resolution of 2 cm−1 using 532 nm laser excitation wavelengths, and X-ray photoelectron spectroscopy (XPS) was performed using an Electron Spectroscopy for Chemical Analysis II system (AXIS SUPRA, Kratos, UK) with a monochromatized aluminum K anode (350 W, 25 mA) at the National Center for Interuniversity Research Facilities (NCIRF) at Seoul National University. Atomic field microscopy (AFM) was used to show these characteristics, and the measurement conditions were as follows. The sample size was 5-μm diameter square, and analysis was conducted in contact mode using a cantilever with a resolution of 0.2 nm for the X/Y axis and 0.01 nm for the Z axis. The Brunauer–Emmett–Teller (BET) surface of the samples (100 mg) was determined from triplicate analysis using the BJH equation N2-desorption method (Quantachrome, PM33, USA). The contact angles were measured by dropping 20 μL of water droplets onto the sample surface. The relative amine number was calculated by calculating the molar amount of PM added in method parts. The morphological analysis of the CNT@CPM nanocomposite samples was performed using high-resolution transmission electron microscopy (HR-TEM; JEOL Ltd, JEM 3010, Japan) with a resolution of 0.17 nm and field emission scanning electron microscopy (FE-SEM; TESCAN, MIRA LMH microscope, Tescan, Czech Republic). The FE-SEM studies were conducted using samples that had been sputter-coated with a 5-nm Pt layer prior to analysis. Energy dispersive X-ray spectroscopy (EDS, BRUKER, Quantax, USA) and line-scan analysis were conducted under the resolution conditions of 125 eV and SDD type.
Humidity sensing properties
A humidity and temperature controller (Model TM-NFM-L, Jeio Tech, Korea) was used to measure relative humidity (RH) at 25 °C. Resistance was measured with an LCR meter (Model EDC-1635, 0.1 Ω–20 MΩ) at 1 kHz, 1 V, and 25 °C. The resistance variation graph for humidity was analyzed every 10 s by measuring the resistance of each humidity every 10 s when the relative humidity was adjusted every 10 percent from 30%RH to 100%RH. The hysteresis curves were repeated 5 times from 30%RH to 100%RH and again to 30%RH, and dynamic transient response curves were measured by changing from 30%RH up to each RH. The long-term stability test of each CNT@CPM flexible sensor was conducted by measuring the resistance once every 3 d for over 1 month while sensor was maintained at 50, 70, and 90%RH. The hysteresis extracted the average of the relative humidity differences (i.e., the average of the x-axis differences) between the adsorption and desorption graph at the same resistance value in range of 30–100%RH [25]. These average values compared in steps of 10% RH for each sample.
Electrochemical analysis
All electrochemical analyses were performed directly using the CNT@CPM flexible sensors. The current–voltage (I-V) curves were measured using a Voltage sag protector (VSP) potentiostat (BioLogic, France), the measurement conditions were increased and decreased by 50 mV per second from − 1.0 V to 1.0 V at 25 °C and 30%RH, and the average value was used after three repeated measurements. A three-electrode system was used with Ag/AgCl (3.0 M KCl) from Cypress systems (Lawrence, KS, USA) as a reference electrode, a 0.5 mm Pt wire counter electrode, and 4 mm diameter CNT@CPM flexible sensors as the working electrode. Cyclic voltammetry (CV) curves were measured in a mixed solution of 2.0 mM K3Fe(CN)6 (0.5 M KCl) and 2.0 mM K4Fe(CN)6 (0.5 M KCl). Impedance measurements (EIS) were performed after each sample in the impedance analyzer (PalmSens BV, Nederland) was for 30 min per humidity range in the 25 °C humidity chamber, and the measurement conditions were increased and decreased by 0.01 V per second from − 1.0 V to 1.0 V at frequencies (61 = 10/dec.).
Statistical analysis
All experiments were repeated at least three times and data are presented as means ± standard deviation (S.D.).
Results and Discussion
A schematic illustration summarizing the overall progress of our present paper is presented in Fig. 1. A flexible sensor based on the CNT@CPM nanocomposites was composed of a CNT core and PM trapped in a Chit-shell structure. As depicted in Fig. 1a, Chit acted as a biocompatible glue on the surface of CNT, an excellent conductive material, as the shell and PM was utilized to increase the humidity efficiency via the augmentation of amine groups that were formed as core–shell structures via self-assembly by adjusting the pH level. More specifically, PM was trapped in the shell of the Chit and immobilized, resulting in an increased surface area, and it had a direct effect on sensitivity and linearity for effective humidity sensing. It is worth noting that in this process, the core–shell fiber could be formed as a flexible nanocomposite film pursuant to strong hydrogen bonds. These advantages are reflected in the flexible humidity sensor for smart wear. The CNT@CPM composites were immobilized on a flexible polyimide (PI) substrate, which was made from an Au/Ni electrode through masking, producing a flexible humidity sensor. To monitor respiration, we fabricated a humidity-controllable mask by integrating a microcontroller that activated the fan when the humidity was above 70% and a high-performance CNT@CPM-3 humidity sensor. It was confirmed that not only did it monitor the respiratory rate, depth, and occurrence of exhalation or inhalation, but it also controlled the humidity level inside the mask, validating its performance for realistic applications (Fig. 1b). This flexible humidity sensor is not limited to masks and could be used for multiple smart wear in everyday life (Fig. 1c).
The morphological changes and properties of the samples formed at different PM concentrations (0–10 wt%) are compared in Fig. 2. Based on HR-TEM images (Fig. 2a), the CPM thickness values of 4.4 ± 0.51 nm for CNT@CPM-1, 7.5 ± 0.68 nm for CNT@CPM-2, 12.2 ± 0.94 nm for CNT@CPM-3, and 17.4 ± 1.77 nm for CNT@CPM-4, as derived from 50 thickness measurements. The thickness values used to acquire the average thickness are shown in the distribution graph in Additional file 1: Fig. S2. The HR-TEM images clearly show an increase in the thickness of the shell layer on the surface of CNTs with increasing concentration of PM, clearly indicating the formation of the CNT core and Chit/PM-shell structured nanocomposites. The PM was very evenly distributed among the polymers of Chit, resulting in an increase in the overall coating thickness as the shell structure became loose (Additional file 1: Fig. S3). These results are described with a consistent nano-rough topology. The FE-SEM images (Fig. 2b) also showed the same results as those of the HR-TEM. Obviously, the nanopores between the nanofiber strands shown in CNT@CPM-1 became thicker and were being blocked owing to the increasing PM ratio. Strong interaction with density changes according to fiber thickness can allow the flexible PI-based sensor to be strongly and reliably attached. The AFM images were measured to determine the change in the hydrophilic properties according to the ratio of PM. The AFM data (Fig. 2c, d) showed that the PM was very evenly attached to the core–shell structure of CNT@CPM-1. The surface morphological topography emerged as peaks and valleys originating from the structure of CNT@CPM. The surface roughness of each substrate, given by the average roughness parameter (Ra) value of 39.7 ± 1.24 nm for CNT@CPM-1, 32.18 ± 2.32 nm for CNT@CPM-2, 76.43 ± 4.11 nm for CNT@CPM-3, and 98.25 ± 6.83 nm for CNT@CPM-4, showing a threefold difference. The CNT@CPM nanocomposites well reflected that the nanocomposites exhibited a consistent morphology with varying shell thickness. The relative height deviation increases because the diameter of CNT@CPM increases according to the amount of PM (Additional file 1: Fig. S4). These innate hydrophilic properties of Chit and PM, as well as the highly nano-rough surface topological features of the nanocomposites, might influence the increased hydrophilicity [26]. The nanocomposites also exhibited a level of water-uptake capacity, which allowed for the rapid appearance of adhesion and reactivity with humidity. The wettability (decrease in the water contact angle, Fig. 2d) increased as the PM ratio increased (each contact angle results are shown in Additional file 1: Fig. S5). These phenomena were due to the increased ratio of primary amines caused by PM, which increased the surface roughness and hydrophilicity [27]. Qualitative analysis was conducted using XPS to confirm whether the CNT, Chit, and PM were efficiently combined with the core–shell structure. The XPS analysis is shown in Fig. 2e for comparison and analysis of the variation in the binding energy between elements by the efficient core–shell structure. The CNTs-related characteristic peaks at 284.68 (C 1s) and 533.08 eV (O 1s) and Chit-related characteristic peaks at 286.68 (C 1s), 533.28 (O 1s), and 399.98 eV (N 1s) were observed. The CNT@CPM-3 nanocomposites spectra of oxygen and nitrogen were slightly shifted (O 1s; 533.28 to 532.28 eV, and N 1s; 399.98 to 398.42 eV). After PM was added, the peaks related to Chit and CNT@CPM-3 were generally shifted to lower positions by 1.0–1.56 eV, indicating intimate physicochemical interactions between the highly polar functional groups of the PM and Chit (C = O, N–H, and O–H). These phenomena indicated efficient complexation due to the large contact area between the CNTs, Chit, and PM. Numerous hydrogen bonds were formed between the Chit shell and the PM functional groups in the core–shell structure, consequently decreasing their binding energy [28, 29] (full spectra in Additional file 1: Fig. S6). The characteristic peaks of CNTs, Chit, and PM [30, 31] are well represented in the FT-IR spectra (Additional file 1: Fig. S7) of the CNT@CPM-3 nanocomposite. The RAMAM analysis results (Additional file 1: Fig. S8) from CNTs, CNT@CPM-1, and CNT@CPM-3 all demonstrated feature D and G bands [32]. The RAMAN spectra of post-composition via the core–shell nanocomposite CNT@CPM-3 simply superimposes the relative strength change for the D and G bands, suggesting a change in the amorphous content of CNT@CPM-3. The ID/IG ratio of the core–shell nanostructure increased from 0.839 to 1.196, as it was synthesized into Chit and PM.
These results were also confirmed using BET measurements of the surface area. The results of measuring the surface area according to the PM ratio of CNT@CPM are summarized in Table 2. As the ratio of PM increased, the surface area and pore volume increased; however, a tendency to decrease again occurred at 10 wt% PM. This appeared to be caused by reduced nanopores as hydrogen bonds became stronger, because the number of primary amines and coating thickness increased as more PM was embedded in Chit.
Table 2.
CNT@CPM-1 | CNT@CPM-2 | CNT@CPM-3 | CNT@CPM-4 | |
---|---|---|---|---|
Surface area (m2 g−1) | 11.09 ± 1.20 | 26.36 ± 4.93 | 25.14 ± 5.13 | 12.11 ± 3.56 |
Pore size (nm) | 3.10 ± 0.16 | 3.80 ± 0.50 | 3.73 ± 0.77 | 3.76 ± 0.56 |
Pore volume (cc g−1) | 0.05 ± 0.01 | 0.06 ± 0.01 | 0.06 ± 0.01 | 0.05 ± 0.02 |
The CNT@CPM nanocomposites were immobilized, as shown in Fig. 3a, on a sensor physically adhered to the Au/Ni electrode on the PI flexible surface, and it can be seen that it was still flexible owing to the efficient structure of the core shell after immobilization (applied using flexible properties of CNT@CPM film). As the conductive sensing film was immobilized, the gold electrode was connected to each other, and the primary amines were greatly increased by the added PM (over 40 times to 132.7 times, Table 1), which increased the reaction and efficiency of water molecules in humidity. A photograph and actual size of a bare flexible sensor is shown in Additional file 1: Fig. S9. EDS mapping and line scanning analysis were conducted to confirm that the materials were evenly coated on the sensor. By showing the distribution of C and Au in the EDS mapping and line scanning images of the humidity sensors, it was shown that gold electrodes were fully covered by individual CNT@CPM core–shell nanocomposites (the FE-SEM images for each ratio of PM are compared in Fig. 2b, all clearly showing core–shell structures). The EDS results (Additional file 1: Fig. S10) and mapping images (Additional file 1: Fig. S11) clearly show that these were evenly coated. Additionally, the FE-SEM image of the cross-sectional flexible humidity sensor (Additional file 1: Fig. S12) shows that CNT@CPM-3 is coated very stably, and thickness of each layer can be confirmed that thickness of Au and Ni are average 3.75 ± 0.04 μm and 1.80 ± 0.01 μm, respectively, and coated CNT@CPM layer is average 2.76 ± 0.16 μm. The humidity sensor showed that the even coating of these core–shell structures not only provides a wide surface area for responsiveness to humidity, but it can also establish a fast and consistent response. The CNT@CPM contained numerous amino groups (− NH2) that can act as proton donors and acceptors, providing a possible proton transport mechanism with humidity [33]. Here, the electrical conductivity could be generated by H+ ions generated through the dissociation of humidity molecules [34] that were first chemically adsorbed on the amine surface and increased by the transfer of the generated H+ ions through the humidity phase that is physiosorbed, as shown in Additional file 1: Fig. S13 (Grotthuss chain reaction) [34, 35]. Consequently, a higher level of humidity would increase the number of physically adsorbed humidity molecule layers, allowing faster proton hopping among adjacent water molecules in a continuous water layer and increasing the resistance [34]. The reason for the increase in the electrical resistance of CNTs, due to adsorption of water molecules, can be the donation of electrons by adsorbed water molecules to the valence band of CNTs, which exhibit a hole transport similar to that of a p-type semiconductor [21]. Electron transfer quickly reduces the concentration of holes in the CNTs, resulting in a rapid increase in resistance. This implies a decrease in the number of holes and an increase in the separation between the Fermi level and valence band [36]. The resistance variation curves of the CNT@CPM nanocomposite humidity sensors exhibited a linear function of RH with a positive slope (dR/dRH > 0). Here, the increasing number of water molecules with an extremely high resistance compared to the CNT@CPM phase caused a positive slope [34]. The response and sensitivity (S) values of the CNT@CPM nanocomposite sensors to humidity were calculated using the following equations [34]:
1 |
2 |
where ΔR/R0 = (RRH − R0)/R0 is the resistance change, R0 is the initial resistance value (30 RH%) of the device, and RRH is the steady-state resistance when exposed to different RH values. The sensor response, curve slope, and linearity of the sensors resulting from the sensor materials with different PM concentrations are shown and summarized in Fig. 3b, c, and Additional file 1: Table S1. The CNT@CPM nanocomposites exhibited high sensitivity and low hysteresis (Fig. 3b). Sensitivity and hysteresis integrated graph, showing sensitivity was 10 to 237 Ω/%RH from 0 to 10 wt%, and hysteresis was (− 0.291 ± 0.079) ~ (0.300 ± 0.001) %. Furthermore, the curve showing hysteresis through repeated testing of sensors (even 5 cycles) according to each humidity showed high reproducibility, repeatability, and low measuring factors (Fig. 3c). These data clearly show that the CNT@CPM-3 nanocomposite sensors have relatively high values of sensor response, sensitivity, and linearity among the tested sensor materials. When the PM concentration ratio increased from 0 to 5 wt%, the linearity of the resistance curves gradually increased, reaching an R2 value of 0.998 (CNT@CPM-3), because the increase in PM concentration caused an improvement in the sensor response by efficient hydrogen bonding due to an efficient reaction even at low humidity. At 10 wt%, the R2 value was reduced to 0.968 because the reactivity was too high in the low humidity range. EIS measurements for our sensors according to humidity showed that %RH was related to the direct resistivity of the sensor itself (Additional file 1: Fig. S14). Nevertheless, the resultant absolute impedance Nyquist plots disclosed no useful information and did not follow the common EIS curves. These plots only indicated that there were differential charge transfers according to the amounts of water molecules in chamber. Higher water molecule content in the cell could increase the charge transfer resistance and the gap between the carbon electrodes, eventually leading to the resistance trends observed in Fig. 3c. As the humidity increased from 30 to 100%, the resistance of the CNT@CPM-3 sensors increased from 150 to 300 Ω. In addition, there is a rapid and direct increase in the resistant responses when the humidity increases from 30 to 100% RH, as shown in the logarithmic representation shown in Additional file 1: Fig. S15, which indicates that moisture absorption is enhanced in these humidity sensors. These results demonstrate that the resistance exhibits excellent logarithmic linearity toward RH as an excellent humidity sensor [37]. Additional file 1: Fig. S16 shows the I-V characteristics of the sensors under different PM concentrations at 25 °C. It is clear from these I-V curves that the current of the sensor decreases with an increase in the core–shell structured CNT@CPM nanocomposites. The curves are linear, proving the ohmic characteristics of the contacts, as the resistance of the sensor increases with the increase in PM nanocomposites. Hence, the CNT@CPM-3 nanocomposite could be used as the maximum concentration to maintain these I-V characteristics. Figure 3c shows that the resistance of the sensors increases significantly with an increase in humidity, which is consistent with the CV curve shown in Additional file 1: Fig. S17, resulting in the stable maintenance of the resistance values of the humidity sensor. CV data indicates that the oxidation and reduction peaks of CNT@CPM nanocomposites are well shown and that compared with those of CNT@CPM-1, the peaks of CNT@CPM-3 modified electrodes increased dramatically according to the increased electrochemical reactions between the electrolytes and the amino groups of CNT@CPM nanocomposites on the electrode surface [38].
The dynamic transient responses were among the most important characteristics for evaluating the performance of the humidity sensors (Fig. 4). The patterns differed depending on the PM ratio. The sensors containing small amounts of PM (CNT@CPM-1; Fig. 4a, and CNT@CPM-2; Fig. 4b) exhibited an exponential trend in the sensitivity graph. When the PM ratio was relatively high, the graph showed a linear trend for CNT@CPM-3 (Fig. 4c) and a logarithmic trend for CNT@CPM-4 (Fig. 4d). These changes in the trends were due to the rapid response to resistance changes in the low humidity range caused by a hydrophilicity that increased with the increasing number of free amines from the added PM. Nevertheless, the response time of the humidity sensor was still shown to change very quickly, with less than 10–40 s, when RH was increased or decreased from 30%RH to each %RH. These characteristics enable efficient adsorption and desorption from water molecules due to influence of a large surface area by CNT@CPM core shell structure. After aggregating these results, we applied the CNT@CPM-3 flexible sensor, which has the most linear responsiveness, for subsequent experiments.
Our sensor could provide fast and stable attainment of sensitivity and long-term stability as an optimal humidity sensor. The CNT@CPM sensors were immobilized on a flexible sensor that could provide good contact with the skin or wearable devices. However, the development of wearable electronics requires long-term and mechanical stability characteristics that require sensor reliability and durability under harsh conditions. The long-term stability results of CNT@CPM-3 showed few resistance changes for each relative humidity over a period of over 1 month (Fig. 5a), indicating that it remains a stable sensor over a long duration at 50, 70, and 90%RH. Different proportions of sensors (CNT@CPM) showed the same long-term stability results and unchanged morphologies (Additional file 1: Fig. S18, S19). The mechanical properties of the flexible sensor were evaluated by monitoring their responses under different bending angles (Fig. 5b). The resistance of the device did not change even after bending at 180°. Additionally, there was no change in resistance after 15,000 bending cycles, demonstrating the superior flexibility and durability of the sensor (Fig. 5c, Additional file 1: Fig. S19, S20). The RH response to humidity changes was evaluated over 15,000 bending cycles as well. These results demonstrated that the response was nearly unchanged even when the device was bent up to 180° over 15,000 bending cycles, which indicates the excellent mechanical durability, stability, and robustness of the CNT@CPM based humidity sensor. When bending was performed about 18,000 times, CNT@CPM was detached, which is a result of bending in a harsher condition than the actual condition. Therefore, the sensor can be effectively applied to practical applications such as skin-attachable or wearable devices. It is clear that the humidity detection performance of the flexible sensors remained almost constant within experimental errors at different bending numbers and angles, demonstrating their potential as a promising flexible wearable sensing device [39]. More interestingly, the sensor can also respond to other volatile polar organic molecules (chloroform, acetone, ethanol, methanol, dichloromethane, tetrahydrofuran, acetic acid, and ammonia) through resistance changes in a different manner, indicating its great potential for application in organic molecule sensing (Fig. 5d).
The humidity sensing performance and characteristics of the CNT@CPM flexible sensors were compared with other previously published resistance-based humidity sensor papers [40–50]. Obtained using various sensor materials, as shown in Table 3. The references showed that sensors with high sensitivity typically exhibited consistent performance due to high sensitivity and linearity but showed limitations, such as high response and recovery times, very high resistance ranges, and high-power requirements (low efficiency). Conversely, references showed that sensors with high response and recovery times can operate at low power (high efficiency) owing to low resistance but have limited operational range and inconsistent limitations owing to low sensitivity and linearity. For CNT@CPM flexible sensors, we demonstrated the advantages of having high sensitivity and linearity over a full range that complements the limitations of other existing references, consistent performance with low power (high efficiency), and high response and recovery times.
Table 3.
Material | Working range (%) | Working resistance (Ω) | Response time (s) | Recovery time (s) | Sensitivity (%) | Reference |
---|---|---|---|---|---|---|
CuO nano wire | 20–80 | 0.5 ~ 3*106 | 120 | 120 | - | [40] |
SnO2 nanowire | 30–85 | 8*105 ~ 2*107 | 120–170 | 30–60 | 35 | [41] |
GO/PDDA film | 11–97 | 0.5 ~ 2.7*106 | 108–147 | 94–133 | 37.43 | [42] |
Black phosphorous | 11.3–97.3 | 0.6 ~ 2*104 | 255–360 | 10–480 | 42.85–99.17 | [43] |
VS2 | 0–50 | 5*103 ~ 3*105 | 30–40 | 12–50 | 300 | [44] |
Epoxy/CNT | 10–90 | 0.8 ~ 1.3*103 | 38 | 60 | 1,990 | [45] |
CNT/PVA fiber | 46–98 | > 0.8*104 | – | – | 140,000 | [46] |
GO/PANI | 11–97 | – | 8 | 5 | 20 | [47] |
Pt/GO fiber | 6.1–99 | > 500 | 2 | 4 | 4.51 | [48] |
GO/PEI | 11–97 | – | 34 | 10 | 27.3 | [49] |
CNT/GO | 10–80 | 13 ~ 96 | 4 | 3 | 9.8 | [50] |
CNT@CPM | 30–100 | 12 ~ 370 | 10–40 | 10–40 | 56.7–111.1 | This work |
Fig. 6 shows the results of adsorption and desorption over real-time and photographs of the attached 3 M mask (6200 dust mask) for electrical signal amplification and sensor processing. Fig. 6a shows an image of the mask that was created by stitching a microcontroller and CNT@CPM-3 flexible sensor directly to an ordinary 3 M mask for real-time respiratory monitoring. The analog value of the CNT@CPM-3 flexible sensor was measured consistently as the resistance; therefore, a standard curve was drawn (Additional file 1: Fig. S21). This linear standard curve enabled us to produce accurate and consistent results. When deep breathing out, the exhaled air reached the humidity sensor, and the RH increased immediately, thus generating an exhalation signal. In contrast, during inhalation, dry air reaches the sensor, and an inhalation signal was generated (Fig. 6b). The actual performance of the sensor is shown in the Additional file 2: Video S1. With this high responsiveness and consistency, normal, fast, and deep breathing could be easily distinguished by the curve of the graph according to the frequency of rapid change in values, which presents a clear difference in current frequency (Fig. 6c). Apnea was simulated during the breathing process and measured to evaluate the performance of the sensor (Fig. 6d). The sensor accurately determined and measured repeated breathing and apnea, indicating that the sensor could be used to detect sleep apnea symptoms. In addition, the CNT@CPM-3 flexible sensor also distinguished breathing through the mouth and nose (Fig. 6e). The two types of breathing have different slopes of moisture exhaled due to the different levels of moisture exhaled from the mouth and nose. These results show that our humidity sensors have high sensitivity and rapid response, which can be used practically to accurately and conveniently monitor human respiration and other respiratory processes. The CNT@CPM-3 flexible sensor with fast response, low power, and consistency can be used as an efficient respiratory monitoring system. Fig. 6f shows that the humidity inside the mask can be adjusted through a ventilation system by applying additional fans to the previously identified mask humidity sensors. The humidity sensor was attached to the inside of the mask to measure humidity in real time and was configured to ventilate through a fan if it exceeded a certain humidity (set to over 70 RH%). To verify that the established system was working, the sensor was inserted into the preset humidity and temperature controller to confirm that the Light emitting diode (LED) lamp was lit by humidity and the fan was turning. The actual performance is shown in Fig. 6f, Additional file 1: S22 and Additional file 3: Video S2. Compared to the conditions without the fan system and with the fan system (Fig. 6g), the humidity gradually increased under medium-intensity conditions. Under harsh conditions, the humidity inside the modalities of the two masks increased rapidly, and the humidity inside was reduced quickly and effectively by the fan system. As a result, if the humidity inside the masks and smart clothes increased due to exercise and activities, it could be detected by sensors and ventilated through fans to create a pleasant internal environment. This fast, consistent, and responsive sensor and reactive fan ventilation system can be applied to masks and clothing, giving the wearer comfort in a partially or completely enclosed environment.
Conclusions
Here we reported that resistance-type CNT@CPM flexible humidity sensors were realized from core–shell structure. The structure of the CNT@CPM nanocomposites was confirmed through qualitative and quantitative analyses. As the ratio of PM increased, the resistance graph over a RH range was transformed from an exponential graph to a logarithmic graph as the shell thickness and hydrophilicity increased. Among them, the CNT@CPM-3 flexible sensor showed a linear graph, and the sensor response hysteresis was only (− 0.152 ± 0.093) in the RH range of 30–100%, which was superior to conventional electric resistance humidity sensors. Moreover, we apply the sensor to smart-wear that could perform accurate human health monitoring of multiple types of respiration (normal, fast, deep, oral, nasal, and apneic breathing). These flexible sensors not only exhibited the ability to detect several organic gases, but it also showed consistent and stable results even after prolonged exposure to physical stress, such as bending. Taken together, novel CNT@CPM flexible sensor, which demonstrated stability and high and accurate response times, is a promising and convenient human healthcare monitoring system.
Supplementary Information
Acknowledgements
This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIT) (No. 2020R1F1A1072210), National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIT) (No. 2022R1C1C2010863), and Priority Research Center Program provided by Ministry of Education (2019R1A6A1A11034536).
Abbreviations
- AFM
Atomic field microscopy
- BET
Brunauer–Emmett–Teller
- Chit
Chitosan
- CNTs
Carbon nanotubes
- PM
PAMAM dendrimer ethylenediamine core, generation 3
- CNT@CPM
PM-embedded Chit/CNT nanocomposites
- CV
Cyclic voltammetry
- EDS
Energy dispersive X-ray spectroscopy
- EIS
Impedance measurements
- FE-SEM
Field emission scanning electron microscopy
- FT-IR
Fourier transform infrared spectroscopy
- HR-TEM
High-resolution transmission electron microscopy
- I-V curves
Current–voltage curves
- PI
Polyimide
- RH
Relative humidity
- XPS
X-ray photoelectron spectroscopy;
- ID/IG
Intensity of D band/Intensity of G band
Author contributions
HSK and JHK contributed equally. HSK carried out most of the experimental studies, conceptualization, writing of original draft preparation, review, editing, and revision. JHK carried out most of the experimental studies, conceptualization, and writing of original draft preparation. JYH carried out supporting experiments and writing of original draft preparation. USS carried out conceptualization, writing of review, editing, revision, and supervision. All authors read and approved the final manuscript.
Funding
This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIT) (No. 2020R1F1A1072210), National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIT) (No. 2022R1C1C2010863), and Priority Research Center Program provided by Ministry of Education (2019R1A6A1A11034536).
Availability of data and materials
The datasets used and/or analyzed during the current study are available from the corresponding author, upon reasonable request.
Declarations
Competing interests
This article has not been accepted for published elsewhere. The authors declare that they have no competing interests.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Han-Sem Kim and Ji-Hye Kang are equally contributed to this work
Contributor Information
Han-Sem Kim, Email: massem@dankook.ac.kr.
Ueon Sang Shin, Email: usshin12@dankook.ac.kr.
References
- 1.Matusiak Ł, Szepietowska M, Krajewski P, Białynicki-Birula R, Szepietowski JC. Inconveniences due to the use of face masks during the COVID-19 pandemic: a survey study of 876 young people. Dermatol. Ther. 33, e13567 (2020) [DOI] [PMC free article] [PubMed]
- 2.Kim H-J, Park SJ, Kim D-I, Lee S, Kwon OS, Kim IK. Moisture effect on particulate matter filtration performance using electro-spun nanofibers including density functional theory analysis. Sci. Rep. 2019;9:1–8. doi: 10.1038/s41598-018-37186-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Chen S, Sun L, Zhou X, Guo Y, Song J, Qian S, Liu Z, Guan Q, Meade Jeffries E, Liu W. Mechanically and biologically skin-like elastomers for bio-integrated electronics. Nat. Commun. 2020;11:1–8. doi: 10.1038/s41467-019-13993-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Gong S, Du S, Kong J, Zhai Q, Lin F, Liu S, Cameron NR, Cheng W. Skin-like stretchable fuel cell based on gold-nanowire-impregnated porous polymer scaffolds. Small. 2020;16:2003269. doi: 10.1002/smll.202003269. [DOI] [PubMed] [Google Scholar]
- 5.Tian L, Li Y, Webb RC, Krishnan S, Bian Z, Song J, Ning X, Crawford K, Kurniawan J, Bonifas A. Flexible and stretchable 3ω sensors for thermal characterization of human skin. Adv. Funct. Mater. 2017;27:1701282. doi: 10.1002/adfm.201701282. [DOI] [Google Scholar]
- 6.Jayathilaka WADM, Qi K, Qin Y, Chinnappan A, Serrano-García W, Baskar C, Wang H, He J, Cui S, Thomas SW. Significance of nanomaterials in wearables: a review on wearable actuators and sensors. Adv. Mater. 2019;31:1805921. doi: 10.1002/adma.201805921. [DOI] [PubMed] [Google Scholar]
- 7.Castaneda D, Esparza A, Ghamari M, Soltanpur C, Nazeran H. A review on wearable photoplethysmography sensors and their potential future applications in health care. Int. J. Biosens. Bioelectron. 2018;4:195. doi: 10.15406/ijbsbe.2018.04.00125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Ma L, Wu R, Patil A, Zhu S, Meng Z, Meng H, Hou C, Zhang Y, Liu Q, Yu R. Full-textile wireless flexible humidity sensor for human physiological monitoring. Adv. Funct. Mater. 2019;29:1904549. doi: 10.1002/adfm.201904549. [DOI] [Google Scholar]
- 9.Kim JJ, Wang Y, Wang H, Lee S, Yokota T, Someya T. Skin electronics: next-generation device platform for virtual and augmented reality. Adv. Funct. Mater. 2021;31:2009602. doi: 10.1002/adfm.202009602. [DOI] [Google Scholar]
- 10.Yao S, Myers A, Malhotra A, Lin F, Bozkurt A, Muth JF, Zhu Y. A wearable hydration sensor with conformal nanowire electrodes. Adv. Healthc. Mater. 2017;6:1601159. doi: 10.1002/adhm.201601159. [DOI] [PubMed] [Google Scholar]
- 11.Wang X, Xiong Z, Liu Z, Zhang T. Exfoliation at the liquid/air interface to assemble reduced graphene oxide ultrathin films for a flexible noncontact sensing device. Adv. Mater. 2015;27:1370–1375. doi: 10.1002/adma.201404069. [DOI] [PubMed] [Google Scholar]
- 12.Wu J, Sun Y-M, Wu Z, Li X, Wang N, Tao K, Wang GP. Carbon nanocoil-based fast-response and flexible humidity sensor for multifunctional applications. ACS. Appl. Mater. Inter. 2019;11:4242–4251. doi: 10.1021/acsami.8b18599. [DOI] [PubMed] [Google Scholar]
- 13.Lee J, Sul H, Lee W, Pyun KR, Ha I, Kim D, Park H, Eom H, Yoon Y, Jung J. Stretchable skin-like cooling/heating device for reconstruction of artificial thermal sensation in virtual reality. Adv. Funct. Mater. 2020;30:1909171. doi: 10.1002/adfm.201909171. [DOI] [Google Scholar]
- 14.Shit A, Heo SB, In I, Park SY. Mineralized soft and elastic polymer dot hydrogel for a flexible self-powered electronic skin sensor. ACS. Appl. Mater. Interfaces. 2020;12:34105–34114. doi: 10.1021/acsami.0c08677. [DOI] [PubMed] [Google Scholar]
- 15.Kim SH, Jung S, Yoon IS, Lee C, Oh Y, Hong JM. Ultrastretchable conductor fabricated on skin-like hydrogel–elastomer hybrid substrates for skin electronics. Adv. Mater. 2018;30:1800109. doi: 10.1002/adma.201800109. [DOI] [PubMed] [Google Scholar]
- 16.Yin R, Wang D, Zhao S, Lou Z, Shen G. Wearable sensors-enabled human–machine interaction systems: from design to application. Adv. Funct. Mater. 2021;31:2008936. doi: 10.1002/adfm.202008936. [DOI] [Google Scholar]
- 17.Sun JY, Keplinger C, Whitesides GM, Suo Z. Ionic skin. Adv. Mater. 2014;26:7608–7614. doi: 10.1002/adma.201403441. [DOI] [PubMed] [Google Scholar]
- 18.Illeperuma WR, Rothemund P, Suo Z, Vlassak JJ. Fire-resistant hydrogel-fabric laminates: a simple concept that may save lives. ACS. Appl. Mater. Interfaces. 2016;8:2071–2077. doi: 10.1021/acsami.5b10538. [DOI] [PubMed] [Google Scholar]
- 19.Alrammouz R, Podlecki J, Vena A, Garcia R, Abboud P, Habchi R, Sorli B. Highly porous and flexible capacitive humidity sensor based on self-assembled graphene oxide sheets on a paper substrate. Sens. Actuators. B Chem. 2019;298:126892. doi: 10.1016/j.snb.2019.126892. [DOI] [Google Scholar]
- 20.Farahani H, Wagiran R, Hamidon MN. Humidity sensors principle, mechanism, and fabrication technologies: a comprehensive review. Sensors. 2014;14:7881–7939. doi: 10.3390/s140507881. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Tulliani J-M, Inserra B, Ziegler D. Carbon-based materials for humidity sensing: a short review. Micromachines. 2019;10:232. doi: 10.3390/mi10040232. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Fennell JF, Jr, Liu SF, Azzarelli JM, Weis JG, Rochat S, Mirica KA, Ravnsbæk JB, Swager TM. Nanowire chemical/biological sensors: status and a roadmap for the future. Angew. Chem. Int. Ed. 2016;55:1266–1281. doi: 10.1002/anie.201505308. [DOI] [PubMed] [Google Scholar]
- 23.Zou H, Luo Z, Yang X, Xie Q, Zhou Y. Toward emerging applications using core–shell nanostructured materials: a review. J. Materials. Sci. 2022;1:31. [Google Scholar]
- 24.Kalambate PK, Huang Z, Li Y, Shen Y, Xie M, Huang Y, Srivastava AK. Core@ shell nanomaterials based sensing devices: a review. TrAC. Trends. Anal. Chem. 2019;115:147–161. doi: 10.1016/j.trac.2019.04.002. [DOI] [Google Scholar]
- 25.Yun S-W, Cha J-R, Gong M-S. Water-resistive humidity sensor prepared from new polyelectrolyte containing both photo-curable 4-styrylpyridinium function and thiol anchor. Sens. Actuators. B Chem. 2014;202:1109–1116. doi: 10.1016/j.snb.2014.06.065. [DOI] [Google Scholar]
- 26.Kim H-S, Lee J-H, Mandakhbayar N, Jin G-Z, Kim S-J, Yoon J-Y, Jo SB, Park J-H, Singh RK, Jang J-H. Therapeutic tissue regenerative nanohybrids self-assembled from bioactive inorganic core/chitosan shell nanounits. Biomaterials. 2021;274:120857. doi: 10.1016/j.biomaterials.2021.120857. [DOI] [PubMed] [Google Scholar]
- 27.Guo W, Chen B, Do VL, ten Brink GH, Kooi BJ, Svetovoy VB, Palasantzas G. Effect of airborne hydrocarbons on the wettability of phase change nanoparticle decorated surfaces. ACS. Nano. 2019;13:13430–13438. doi: 10.1021/acsnano.9b06909. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Tan C, Arshadi M, Lee MC, Godec M, Azizi M, Yan B, Eskandarloo H, Deisenroth TW, Darji RH, Pho TV. A robust aqueous core–shell–shell coconut-like nanostructure for stimuli-responsive delivery of hydrophilic cargo. ACS. Nano. 2019;13:9016–9027. doi: 10.1021/acsnano.9b03049. [DOI] [PubMed] [Google Scholar]
- 29.Tunc I, Suzer S, Correa-Duarte MA, Liz-Marzán LM. XPS characterization of Au (core)/SiO2 (shell) nanoparticles. J. Phys. Chem. B. 2005;109:7597–7600. doi: 10.1021/jp050767j. [DOI] [PubMed] [Google Scholar]
- 30.Kim H-S, Song M, Seo J-W, Shin US. Preparation of electrically conductive bucky-sponge using CNT-cement: conductivity control using room temperature ionic liquids. Synth. Met. 2014;196:92–98. doi: 10.1016/j.synthmet.2014.07.011. [DOI] [Google Scholar]
- 31.Matai I, Gopinath P. Chemically cross-linked hybrid nanogels of alginate and PAMAM dendrimers as efficient anticancer drug delivery vehicles. ACS. Biomater. Sci. Eng. 2016;2:213–223. doi: 10.1021/acsbiomaterials.5b00392. [DOI] [PubMed] [Google Scholar]
- 32.Hwang J-Y, Kim H-S, Kim JH, Shin US, Lee S-H. Carbon nanotube nanocomposites with highly enhanced strength and conductivity for flexible electric circuits. Langmuir. 2015;31:7844–7851. doi: 10.1021/acs.langmuir.5b00845. [DOI] [PubMed] [Google Scholar]
- 33.Bayer T, Cunning BV, Selyanchyn R, Nishihara M, Fujikawa S, Sasaki K, Lyth SM. High temperature proton conduction in nanocellulose membranes: paper fuel cells. Chem. Mater. 2016;28:4805–4814. doi: 10.1021/acs.chemmater.6b01990. [DOI] [Google Scholar]
- 34.Kim H-S, Kim JH, Park S-Y, Kang J-H, Kim S-J, Choi Y-B, Shin US. Carbon nanotubes immobilized on gold electrode as an electrochemical humidity sensor. Sens. Actuators. B Chem. 2019;300:127049. doi: 10.1016/j.snb.2019.127049. [DOI] [Google Scholar]
- 35.Yadav B, Singh R, Singh S. Investigations on humidity sensing of nanostructured tin oxide synthesised via mechanochemical method. J. Exp. Nanosci. 2013;8:670–683. doi: 10.1080/17458080.2011.597442. [DOI] [Google Scholar]
- 36.Cao C, Hu C, Fang L, Wang S, Tian Y, Pan C. Humidity sensor based on multi-walled carbon nanotube thin films. J. Nanomater. 2011;2011:1. [Google Scholar]
- 37.Jin L, Jiang Y, Zhang M, Li H, Xiao L, Li M, Ao Y. Oriented polyaniline nanowire arrays grown on dendrimer (PAMAM) functionalized multiwalled carbon nanotubes as supercapacitor electrode materials. Sci. Rep. 2018;8:1–10. doi: 10.1038/s41598-018-24265-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Chen Q, Ai S, Fan H, Cai J, Ma Q, Zhu X, Yin H. The immobilization of Cytochrome c on MWNT–PAMAM–Chit nanocomposite incorporated with DNA biocomposite film modified glassy carbon electrode for the determination of nitrite. J. Solid. State. Electrochem. 2010;14:1681–1688. doi: 10.1007/s10008-010-1010-3. [DOI] [Google Scholar]
- 39.He J, Xiao P, Shi J, Liang Y, Lu W, Chen Y, Wang W, Théato P, Kuo S-W, Chen T. High performance humidity fluctuation sensor for wearable devices via a bioinspired atomic-precise tunable graphene-polymer heterogeneous sensing junction. Chem. Mater. 2018;30:4343–4354. doi: 10.1021/acs.chemmater.8b01587. [DOI] [Google Scholar]
- 40.Hsueh H, Hsueh T, Chang S, Hung F, Tsai T, Weng W, Hsu C, Dai B. CuO nanowire-based humidity sensors prepared on glass substrate. Sens. Actuators. B Chem. 2011;156:906–911. doi: 10.1016/j.snb.2011.03.004. [DOI] [Google Scholar]
- 41.Kuang Q, Lao C, Wang ZL, Xie Z, Zheng L. High-sensitivity humidity sensor based on a single SnO2 nanowire. J. Am. Chem. Soc. 2007;129:6070–6071. doi: 10.1021/ja070788m. [DOI] [PubMed] [Google Scholar]
- 42.Zhang D, Tong J, Xia B, Xue Q. Ultrahigh performance humidity sensor based on layer-by-layer self-assembly of graphene oxide/polyelectrolyte nanocomposite film. Sens. Actuators. B Chem. 2014;203:263–270. doi: 10.1016/j.snb.2014.06.116. [DOI] [Google Scholar]
- 43.Late DJ. Liquid exfoliation of black phosphorus nanosheets and its application as humidity sensor. Microporous. Mesoporous. Mater. 2016;225:494–503. doi: 10.1016/j.micromeso.2016.01.031. [DOI] [Google Scholar]
- 44.Feng J, Peng L, Wu C, Sun X, Hu S, Lin C, Dai J, Yang J, Xie Y. Ultrathin nanosheets: giant moisture responsiveness of VS2 ultrathin nanosheets for novel touchless positioning interface. Adv. Mater. 2012;24:1917–1917. doi: 10.1002/adma.201290085. [DOI] [PubMed] [Google Scholar]
- 45.Arunachalam S, Izquierdo R, Nabki F. Low-hysteresis and fast response time humidity sensors using suspended functionalized carbon nanotubes. Sensors. 2019;19:680. doi: 10.3390/s19030680. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Ramos PG, Morales NJ, Goyanes S, Candal RJ, Rodríguez J. Moisture-sensitive properties of multi-walled carbon nanotubes/polyvinyl alcohol nanofibers prepared by electrospinning electrostatically modified method. Mater. Lett. 2016;185:278–281. doi: 10.1016/j.matlet.2016.08.125. [DOI] [Google Scholar]
- 47.Zhang D, Wang D, Li P, Zhou X, Zong X, Dong G. Facile fabrication of high-performance QCM humidity sensor based on layer-by-layer self-assembled polyaniline/graphene oxide nanocomposite film. Sens. Actuators. B Chem. 2018;255:1869–1877. doi: 10.1016/j.snb.2017.08.212. [DOI] [Google Scholar]
- 48.Choi SJ, Yu H, Jang JS, Kim MH, Kim SJ, Jeong HS, Kim ID. Nitrogen-doped single graphene fiber with platinum water dissociation catalyst for wearable humidity sensor. Small. 2018;14:1703934. doi: 10.1002/smll.201703934. [DOI] [PubMed] [Google Scholar]
- 49.Yuan Z, Tai H, Ye Z, Liu C, Xie G, Du X, Jiang Y. Novel highly sensitive QCM humidity sensor with low hysteresis based on graphene oxide (GO)/poly (ethyleneimine) layered film. Sens. Actuators. B Chem. 2016;234:145–154. doi: 10.1016/j.snb.2016.04.070. [DOI] [Google Scholar]
- 50.Li X, Chen X, Yao Y, Li N, Chen X, Bi X. Multi-walled carbon nanotubes/graphene oxide composites for humidity sensing. IEEE. Sens. J. 2013;13:4749–4756. doi: 10.1109/JSEN.2013.2273615. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets used and/or analyzed during the current study are available from the corresponding author, upon reasonable request.