Skip to main content
BMC Musculoskeletal Disorders logoLink to BMC Musculoskeletal Disorders
. 2022 Aug 4;23:750. doi: 10.1186/s12891-022-05698-8

Subsidence following cervical discectomy and implant-to-bone ratio

Bartosz Godlewski 1,, Adam Bebenek 2, Maciej Dominiak 1, Grzegorz Karpinski 1, Piotr Cieslik 3, Tomasz Pawelczyk 4
PMCID: PMC9351094  PMID: 35927645

Abstract

Background

Implant subsidence is an undesirable effect after anterior cervical discectomy and fusion (ACDF). We investigated the relation between the rate of implant subsidence and the ratio of the implant surface area to the surface area of the adjacent bone.

Methods

We operated 170 disc spaces in a group of 104 patients. Two types of implants were used: 1) PEEK (polyetheretherketone) cages and 2) titanium-coated (TC) PEEK cages. Patients were randomised to receive a specific implant using a randomisation table. All implants had a surface area of 1.61 cm2. Based on computed tomography images, bone surface areas were calculated for vertebral bodies immediately adjacent to the interbody implants. The implant-to-bone surface ratio was then calculated for each disc space. Implant subsidence was assessed over 12 months of follow-up, and associations between implant subsidence, the type of implant, and the implant-to-bone surface ratio were investigated.

Results

Twelve months after the surgery, computed tomography was performed on 86 patients (144 disc spaces). Furthermore, in 166 disc spaces and 102 patients, conventional radiographs were obtained.

Subsidence was observed in 21% of the examined intervertebral spaces, and it was more frequently associated with higher values of bone surface area and lower values of the implant-to-bone surface ratio. The type of implant (PEEK vs TC-PEEK cages) did not significantly influence the rate of implant subsidence.

Conclusions

Implant subsidence was significantly related to the value of a coefficient representing the ratio of the implant's surface area to the bone surface area of the adjacent vertebral bodies, with subsidence occurring significantly more rarely for coefficient values ≥ 0.37.

Keywords: Subsidence, Radiological measurements, Cage size, Polyetheretherketone (PEEK), Titanium-coated PEEK

Background

Cervical discectomy and fusion is the prevailing procedure performed for degenerative cervical disease. Cervical disc replacement with a stand-alone cage can restore physiologic disc height, provide immediate load-bearing support to the cervical spine and may promote osseous fusion. Implant subsidence after ACDF is an undesirable effect that should be prevented [15]. We investigated the relation between implant subsidence and the ratio of the surface area of the implant to that of the adjacent bone.

Methods

We operated a total of 170 disc spaces in a group of 104 patients (age: 51.2 ± 10.3; female 73.1%). We used two types of implants: 1) PEEK (polyetheretherketone) cages and 2) titanium-coated (TC) PEEK cages. A randomisation table was used to assign patients a specific implant. Either one or two disc spaces were operated on during one surgery. All patients were operated on by the same surgeon and according to the same technique. In each case, the interior of an implant was filled with nanoparticle hydroxyapatite. Bone surface areas were calculated for vertebral bodies immediately adjacent to the interbody implants on the basis of computed tomography images. All implants used had the same surface area of 1.61 cm2 (Fig. 1). The implant-to-bone surface ratio was then calculated for each disc space (Fig. 2). Implant subsidence was assessed over 12 months of follow-up. We measured the height of interbody spaces in the centre of vertebral bodies, determining the distance between the endplates of adjacent vertebral bodies. The measurements were noted down with an accuracy of 1 mm. Measurements were made on radiographs obtained in one X-ray centre, following the same procedure and utilising the same equipment. The radiographic indices were evaluated at the following times: 1) before surgical procedure, 2) one day after surgery, 3) one month following the surgery, 4) six months post-surgery, 5) one year after the surgery. Subsidence was diagnosed if the implant displaced ≥ 3 mm into the adjacent endplates in relation to radiographs collected one day after the surgery (Fig. 3). Associations between implant subsidence and the type of implant as well as the coefficient representing the ratio of the implant surface area to the bone surface area of the adjacent vertebral bodies were investigated. The research was approved by the Bioethics Committee of the Andrzej Frycz Modrzewski University in Cracow (Resolution 4/2019) and was conducted in compliance with the Declaration of Helsinki.

Fig. 1.

Fig. 1

Procedure for measuring bone surface area of vertebral bodies immediately adjacent to implants on the basis of CT images obtained at 12 months post-surgery. A (cm) x B (cm) = surface area (cm2). All implants had the same surface area of 1.61 cm.2

Fig. 2.

Fig. 2

Comparison of the implant/bone surface ratio on the basis of two disc spaces. A) Patient No. 4, space No. 7, bone surface area of 4.62 cm2, implant surface area of 1.61 cm2, implant/bone ratio of 0.34. B) Patient No. 14, space No. 23, bone surface area of 3.68 cm2, implant surface area of 1.61 cm.2, implant/bone ratio of 0.43

Fig. 3.

Fig. 3

Example of cage subsidence. a) before surgery, b) one day after surgery, c) 12 months after surgery

Results

We completed 104 surgical procedures, operating 170 disc spaces. A total of 100 PEEK implants and 70 TC-PEEK implants were used. No association was observed between subsidence and sex (Chi-squared = 0.708, df = 1, p = 0.4) or implant type used (Chi-squared = 0.501, df = 1, p = 0.479). No significant associations were seen between age and subsidence status (Student’s t = 1,28, df = 99, p = 0,203). Significant correlations were observed between age and vertebral surface area (Pearson’s r = 0.229, p = 0.0352) and implant-to-bone surface area ratio (Pearson’s r = -0.222, p = 0.0412). Twelve months following the surgery, CT scans were performed on 86 patients (144 disc spaces), while typical radiographs were done on 102 patients (166 disc spaces). 166 complete sets of radiographic measurement data were used to assess changes in interbody space height and the presence of implant subsidence (98 for PEEK cages and 68 for TC-PEEK cages). Subsidence was identified in 21% of cases (35 disc spaces), which included 21 PEEK cages (21.4%) and 14 TC-PEEK cages (20.6%). Statistical analysis failed to find a significant association between the type of implant (PEEK vs TC-PEEK) and the presence of implant subsidence (Chi-squared 0.017, df = 1, p = 0.89615) (Table 1). Linear regression was used to assess the association between bone surface measurements and subsidence. Since there was a significant correlation between age and bone surface area and implant-to-bone surface area ratio, age was used in analyses as a covariate to control the effect of age on the results. Subsidence was significantly more frequently associated with a) higher values of bone surface area and b) lower values of the implant-to-bone surface area ratio (linear regression: a) R2 = 0.189; B = 0.997; p = 0.0004; b) R2 = 0.15; B = -0.057; p = 0.0025) (Table 2; Fig. 4). Analysis of the Receiver Operating Characteristics (ROC) curve and Youden’s J statistic [6, 7] calculated on its basis served to determine the value of the coefficient representing the ratio of an implant to vertebral body surface area that significantly differentiated the rate of implant subsidence, at ≥ 0.37 (Fig. 5). There was a correlation between the incidence of implant subsidence and the level treated, with subsidence seen most often following surgery at the C6/C7 level (51%), and least often following surgery at the C3/C4 level (3%). Implant subsidence was noted at 12 months post-surgery in 34.6% of all cases of C6/C7 surgery and 11.8% of all cases at the C3/C4 level (Chi-squared 8.502, df = 3, p = 0.0367). These results are summarised in Table 3.

Table 1.

Association between type of implant and subsidence rate at 12 months post-surgery

Type of implant Subsidence χ2 (df) p
Yes
n (%)a
No
n (%)a
Titanium-coated PEEK 14 (40) 54 (41) 0.017 (1) 0.8962
PEEK 21 (60) 77 (59)

Key to abbreviations and symbols:

acolumn percentages;

Χ2 value of the statistic of the Chi2 test for independence, df Degrees of freedom, p Two-tailed test probability for Chi2 test statistics;

Table 2.

Relationships between bone surface and implant-to-bone surface area ratio and subsidence, including age as a covariate to control for the confounding variable

Parameter Subsidence Marginal mean (SE) B SE 95% CI for B t p
No Yes
Bone Surfacea 3.93 (0.065) 4.53 (0.147) 0.597 0.161 (0.278-0.197) 3.72 0.0004
Implant-to-bone surface ratio areab 0.412 (0.0066) 0.361 (0.0149) - 0.05 0.0164 (-0.084 --0.019) -3.12 0.0025

Key to abbreviations and symbols:

aR2 for the model 0.189;

bR2 for the model 0.15;

B linear regression coefficient estimate, SE Standard error of the estimate, CI Confidence interval, t t statistics for the linear regression, p two-tailed test probability for t-test statistic;

Underline marks significant associations for p < α = 0.05

Fig. 4.

Fig. 4

Differences in surface ratio between cases where subsidence was observed after 12 months (Yes) and those without subsidence (No). Bars show means ± 95% confidence Intervals

Fig. 5.

Fig. 5

Receiver Operating Characteristics (ROR) for the surface ratio and the presence of subsidence after 12 months. Area under curve, AUC = 0.71 (95% CI 0.606–0.815). Cut-off point was determined using Youden’s J statistic, which maximises sensitivity and specificity

Table 3.

Implant subsidence rate and treatment level

Treatment level Subsidence χ2 (df) p
Yes
n (column %) row %
No
n (column %) row %
C3/C4 1 (2.9) 11.11 8 (6.11) 88.89 8.502 (1) 0.0367
C4/C5 2 (5.71) 12.5 14 (10.7) 87.5
C5/C6 14 (40.0) 15.73 75 (57.3) 84.27
C6/C7 18 (51.43) 34.62 34 (25.95) 65.38

Key to abbreviations and symbols:

Χ2 value of the statistic of the Chi2 test for independence, df degrees of freedom, p two-tailed test probability for Chi2 test statistics;

Underline marks significant associations for p < α = 0.05; α – level of statistical significance; sample size differences are due to missing data

Discussion

Most cervical spine procedures in patients with degenerative disc disease involve discectomy and removal of osteophytes in posterior vertebral body surfaces followed by interbody stabilisation with an interbody implant (cage). Cages are made of a variety of materials, differing in structural design, shape and surface topography. Implant subsidence is a widely known and recognised phenomenon. A number of definitions of implant subsidence exist. It may be defined as the length of immersion of the implant (in millimetres) beyond the borders of the adjacent endplates or as the percentage reduction in interbody space height [812]. Cage subsidence may influence spinal biomechanics and alignment, cause segmental kyphosis and contribute to adjacent segment disease. The decreased height of the interbody space may lead to foraminal stenosis [24, 1014]. Current literature suggests a subsidence rate ranging from 19.3 to 42.5% [1, 12]. There is currently controversy and debate on the correlation between cage subsidence and clinical outcome [1, 3, 8, 1017]. We investigated implants made of PEEK and TC-PEEK. The elasticity modulus of PEEK is similar to that of bone, which results in minimising subsidence and optimising the interaction of the compressive forces at the graft-host interface. Since PEEK implants do not distort the anatomical image, they are advantageous in subsequent post-operative imaging. The radiolucent property of PEEK allows for the appropriate assessment of bone in-growth. Furthermore, compared to metallic implants, PEEK produces fewer artefacts on MRI and CT scans [220]. TC-PEEK cages preserve the biomechanical and radiographic advantages of PEEK. Improved osseointegration is achieved by adding a plasma-sprayed surface layer of titanium. Our statistical analyses showed that the type of implant (PEEK vs TC-PEEK) did not significantly affect implant subsidence. A significant association was detected between the rate of implant subsidence and the ratio of the implant surface area to the bone surface area of the adjacent vertebral bodies. Some papers show that anterior implant placement within the disk space reduces the risk of subsidence compared to more posteriorly placed implants [13, 21, 22]. Park J-Y et al. found that a distance of ≥ 3 mm between the anterior margin of the vertebral body and that of the cage was a statistically significant risk factor for subsidence (p < 0.001) [2]. Some papers indicate that implant subsidence can be limited with additional fixation with a cervical plate. Instrumentation appears to be helpful, particularly for ACDF involving two or more levels [23]. Dai and Jiang studied radiologic and clinical results after ACDF with and without a plate and reported higher subsidence in the ACDF group without a plate [24]. The rate of subsidence is also influenced by implant size and the size of the adjacent endplates between which the implant is placed. Implant size should match the specific size of the adjacent endplates to reduce the risk of subsidence [1, 3, 21, 25]. Smaller cages may be more prone to subsidence on account of the smaller area to distribute the acting forces, and the absence of support on harder bone found at the edges of vertebral endplates [1, 2528, 29, 30]. Mende KS et al. observed a significant correlation between the cage/endplate ratio and the incidence of subsidence. Their measurements were based on radiographs. Cages covering more than 65% of the sagittal endplate diameter were significantly less frequent to subside than those below 65% (overall: 64.6 vs 35.4%, p < 0.01) [1]. Yang et al. demonstrated that a small anterior–posterior cage diameter (12 mm vs 14 mm) and a large intraoperative distraction significantly increased the risk of subsidence [31]. In our study, we calculated surface areas of the vertebrae immediately adjacent to the interbody implants on the basis of CT images. The ratio of implant surface area to bone surface area was computed for each interbody space analysed. We established that implant subsidence occurred significantly less often when the coefficient was ≥ 0.37. We also observed a correlation between the subsidence rate and the treatment level, with subsidence occurring significantly more frequently after surgery at the level of C5/C6 and C6/C7 than C3/C4 or C4/C5 (Table 3) (Pearson’s Chi^2, Chi-squared 8.501849, df = 3, p = 0.03670). Similar results were obtained by Kao et al., who mentioned that the treatment levels below the C5 level (C5/C6, C6/C7) had more chance of subsidence than treatment levels above the C5 level (C2/C3, C3/C4, C4/C5) [19]. Bartels et al. reported that the incidence of cage subsidence was significantly higher for C6/C7 fusions than for fusions at other levels  [32]. With regard to pathophysiology, some of the endplates should be removed to expose the subchondral bone to facilitate fusion. The use of a high-speed drill can facilitate this, but surgeons need to pay close attention to endplate integrity in order to prevent subsidence. In our experience, in most cases, the disc and osteophytes of posterior vertebral body edges can be removed, and sufficient decompression of neural structures can be performed without disrupting endplate continuity.

Conclusions

The type of implant (PEEK vs titanium-coated PEEK cages) did not significantly affect the rate of implant subsidence. Subsidence was significantly associated with the ratio of the implant surface area to the surface area of the adjacent vertebral bodies, with subsidence being much less frequent for ratios ≥ 0.37. 

Acknowledgements

None.

Abbreviations

CT

Computed Tomography

PEEK

Polyetheretherketone

ROC

Receiver Operating Characteristics

TC-PEEK

Titanium-coated PEEK

Authors’ contributions

BG: research concept and design, performed all surgical procedures in the analysed group of patients, collection and assembly of data, data analysis and interpretation, writing the article. AB: collecting and assembly of data. MD, GK, PC: collection and assembly of data. TP: statistical analysis and interpretation, writing the article. All authors read and approved the final manuscript.

Funding

This study was not supported by external research funding.

Availability of data and materials

The datasets used and /or analysed during the current study are available from the corresponding author on reasonable request.

Declarations

Ethics approval and consent to participate

This study obtained ethical approval from the institutional ethics committee of Andrzej Frycz Modrzewski Cracow University in Cracow (Resolution 4/2019) and was conducted in compliance with the Declaration of Helsinki. Written informed consent was obtained from all individual participants included in the study.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Footnotes

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Mende KC, Eicker SO, Weber F. Cage deviation in the subaxial cervical spine in relation to implant position in the sagittal plane. Neurosurg Rev. 2018;41(1):267–274. doi: 10.1007/s10143-017-0850-z. [DOI] [PubMed] [Google Scholar]
  • 2.Park JY, Choi KY, Moon BJ, Hur H, Jang JW, Lee JK. Subsidence after single-level anterior cervical fusion with a stand-alone cage. J Clin Neurosci. 2016;33:83–88. doi: 10.1016/j.jocn.2016.01.042. [DOI] [PubMed] [Google Scholar]
  • 3.Jang HJ, Chin DK, Kim KH, Park JY. Does Graft Position Affect Subsidence After Anterior Cervical Discectomy and Fusion? Glob spine J Published online. 2020 doi: 10.1177/2192568220963061. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Obermueller T, Wagner A, Kogler L, Joerger AK, Lange N, Lehmberg J, Meyer B, Shiban E. Radiographic measurements of cervical alignment, fusion and subsidence after ACDF surgery and their impact on clinical outcome. Acta Neurochir (Wien) 2020;162(1):89–99. doi: 10.1007/s00701-019-04139-1. [DOI] [PubMed] [Google Scholar]
  • 5.Godlewski B, Stachura K, Czepko RA, Banach M, Czepko R. Analysisis of changes in cervical spinal curvature and intervertebral disk space height following ACDF surgery in a group of 100 patients followed up to 12 months. J Clin Neurosci. 2018;52:92–99. doi: 10.1016/j.jocn.2018.04.005. [DOI] [PubMed] [Google Scholar]
  • 6.Youden WJ. Index for rating diagnosctic tests. Cancer. 1950;3(1):32-5. 10.1002/1097-0142(1950)3:1<32::aid-cncr2820030106>3.0.co;2-3 [DOI] [PubMed]
  • 7.Schisterman EF, Perkins NJ, Liu A, Bondell H. Optimal cut-point and its corresponding Youden Index to discriminate individuals using pooled blood samples. Epidemiology. 2005;16(1):73–81. doi: 10.1097/01.ede.0000147512.81966.ba. [DOI] [PubMed] [Google Scholar]
  • 8.Wu WJ, Jiang LS, Liang Y, Dai LY. Cage subsidence does not, but cervical lordosis improvement does affect the long-term results of anterior cervical fusion with stand-alone cage for degenerative cervical disc disease: a retrospective study. Eur Spine J. 2012;21(7):1374–1382. doi: 10.1007/s00586-011-2131-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Godlewski B, Bebenek A, Dominiak M, Karpinski G, Cieslik P, Pawelczyk T. PEEK versus titanium-coated PEEK cervical cages: fusion rate. Acta Neurochir (Wien) 2022;164(6):1501–1507. doi: 10.1007/s00701-022-05217-7. [DOI] [PubMed] [Google Scholar]
  • 10.Zajonz D, Franke AC, von der Höh N, Voelker A, Moche M, Gulow J, Heyde CE. Is the radiographic subsidence of stand-alone cages associated with adverse clinical outcomes after cervical spine fusion? An observational cohort study with 2-year follow-up outcome scoring. Patient Saf Surg. 2014;8(1):1–9. doi: 10.1186/s13037-014-0043-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Brenke C, Dostal M, Scharf J, Weiß C, Schmieder K, Barth M. Influence of cervical bone mineral density on cage subsidence in patients following stand-alone anterior cervical discectomy and fusion. Eur Spine J. 2015;24(12):2832–2840. doi: 10.1007/s00586-014-3725-9. [DOI] [PubMed] [Google Scholar]
  • 12.Karikari IO, Jain D, Owens TR, Gottfried O, Hodges TR, Nimjee SM, Bagley CA. Impact of subsidence on clinical outcomes and radiographic fusion rates in anterior cervical discectomy and fusion: a systematic review. J Spinal Disord Tech. 2014;27(1):1–10. doi: 10.1097/BSD.0b013e31825bd26d. [DOI] [PubMed] [Google Scholar]
  • 13.Schmieder K, Wolzik-Grossmann M, Pechlivanis I, Engelhardt M, Scholz M, Harders A. Subsidence of the wing titanium cage after anterior cervical interbody fusion: 2-year follow-up study. J Neurosurg Spine. 2006;4(6):447–453. doi: 10.3171/spi.2006.4.6.447. [DOI] [PubMed] [Google Scholar]
  • 14.Gercek E, Arlet V, Delisle J, Marchesi D. Subsidence of stand-alone cervical cages in anterior interbody fusion: warning. Eur Spine J. 2003;12(5):513–516. doi: 10.1007/s00586-003-0539-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Lee CH, Kim KJ, Hyun SJ, Yeom JS, Jahng TA, Kim HJ. Subsidence as of 12 months after single-level anterior cervical inter-body fusion. Is it related to clinical outcomes? Acta Neurochir (Wien) 2015;157(6):1063–1068. doi: 10.1007/s00701-015-2388-6. [DOI] [PubMed] [Google Scholar]
  • 16.Klingler JH, Krüger MT, Sircar R, Kogias E, Scholz C, Volz F, Scheiwe C, Hubbe U. PEEK Cages versus PMMA Spacers in Anterior Cervical Discectomy: Comparison of Fusion, Subsidence, Sagittal Alignment, and Clinical Outcome with a Minimum 1-Year Follow-Up. ScientificWorldJournal. 2014;2014:398396. doi: 10.1155/2014/398396. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Haden N, Latimer M, Seeley HM, Laing RJ. Loss of inter-vertebral disc height after anterior cervical discectomy. Br J Neurosurg. 2005;19(6):469–474. doi: 10.1080/02688690500495109. [DOI] [PubMed] [Google Scholar]
  • 18.Tomé-Bermejo F, Morales-Valencia JA, Moreno-Pérez J, Marfil-Pérez J, Díaz-Domingues E, Piñera AR, Alvarez. Degenerative Cervical Disc Disease: Long-term Changes in Sagittal Alignment and Their Clinical Implications After Cervical Interbody Fusion Cage Subsidence: A Prospective Study With Standalone Lordotic Tantalum Cages. Clin spine Surg. 2017;30(5):E648-55. 10.1097/BSD.0000000000000293. [DOI] [PubMed]
  • 19.Shiban E, Gapon K, Wostrack M, Meyer B, Lehmberg J. Clinical and radiological outcome after anterior cervical discectomy and fusion with stand-alone empty polyetheretherketone (PEEK) cages. Acta Neurochir (Wien). 2016;158(2):349-55. 10.1007/s00701-015-2630-2. [DOI] [PubMed]
  • 20.Kao TH, Wu CH, Chou YC, Chen H Te, Chen WH, Tsou HK. Risk factors for subsidence in anterior cervical fusion with stand-alone polyetheretherketone (PEEK) cages: a review of 82 cases and 182 levels. Arch Orthop Trauma Surg. 2014;134(10):1343-51. 10.1007/s00402-014-2047-z. [DOI] [PMC free article] [PubMed]
  • 21.Godlewski B, Dominiak M. Advantages and disadvantages of the use of varioius types of interbody implants in cervical spine surgery. Critical review of the literature. Ortop Traumatol Rehabil. 2020;22(4):213-20. 10.5604/01.3001.0014.3457. [DOI] [PubMed]
  • 22.Barsa P, Suchomel P. Factors affecting sagittal malalignment due to cage subsidence in stand-alone cage assisted anterior cervical fusion. Eur Spine J. 2007;16(9):1395-400. 10.1007/s00586-006-0284-8. [DOI] [PMC free article] [PubMed]
  • 23.Yamagata T, Takami T, Uda T, Ikeda H, Nagata T, Sakamoto S, Tsuyuguchi N, Ohata K. Outcomes of contemporary use of rectangular titanium stand-alone cages in anterior cervical discectomy and fusion: cage subsidence and cervical alignment. J Clin Neurosci. 2012;19(12):1673-8. 10.1016/j.jocn.2011.11.043. [DOI] [PubMed]
  • 24.Wang JC, McDonough, Endow KK, Delamarter RB. Increased fusion rates withs cervical plating for two-level anterior cervical discectomy and fusion. Spine (Phila Pa 1976). 2000;25(1):41-5. 10.1097/00007632-200001010-00009. [DOI] [PubMed]
  • 25.Dai LY, Jiang LS. Anterior cervical fusion with interbody cage containing beta-tricalcium phosphate augmented with plate fixation: a prospective randomised study with 2-year follow-up. Eur Spine J. 2008;17:698-705. 10.1007/s00586-008-0643-8 [DOI] [PMC free article] [PubMed]
  • 26.Soo AE, Olsson E, Lim M. Prediction of Cervical Endplate Size: One Size Does Not Fit All. Orthopedics. 2016;39(3):e526-31. 10.3928/01477447-20160427-11. [DOI] [PubMed]
  • 27.Lee YS, Kim YB, Park SW. Risk factors for post-operative subsidence of single-level anterior cervical discectomy and fusion: the significance of the preoperative cervical alignment. Spine (Phila Pa 1976). 2014;39(16):1280-7. 10.1097/BRS.0000000000000400. [DOI] [PubMed]
  • 28.Lou J, Liu H, Rong X, Li H, Wang B, Gong Q. Geometry of inferior endplates of the cervical spine. Clin Neurol Neurosurg. 2016;142:132-6. 10.1016/j.clineuro.2016.01.027. [DOI] [PubMed]
  • 29.Cho DY, Lee WY, Sheu PC. Treatment of multilevel cervical fusion with cages. Surg Neurol. 2004;62(5):378-85. 10.1016/j.surneu.2004.01.021. [DOI] [PubMed]
  • 30.Ritzel H, Amling M, Pösl M, Hahn M, Delling G. The thickness of human vertebral cortical bone and its changes in aging and osteoporosis: a histomorphometric analysis of the complete spinal column from thirty-seven autopsy specimens. J Bone Miner Res. 1997;12(1):89-95. 10.1359/jbmr.1997.12.1.89. [DOI] [PubMed]
  • 31.Yang JJ, Yu CH, Chang BS, Yeom JS, Lee JH, Lee CK. Subsidence and nonunion after anterior cervical interbody fusion using a stand-alone polyetheretherketone (PEEK) cage. Clin Orthop Surg. 2011;3(1):16-23. 10.4055/cios.2011.3.1.16. [DOI] [PMC free article] [PubMed]
  • 32.Bartels RHMA, Donk RD, Feuth T. Subsidence of stand-alone cervical carbon fiber cages. Neurosurgery. 2006;58(3):502-7. 10.1227/01.NEU.0000197258.30821.50 [DOI] [PubMed]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets used and /or analysed during the current study are available from the corresponding author on reasonable request.


Articles from BMC Musculoskeletal Disorders are provided here courtesy of BMC

RESOURCES