Skip to main content
American Journal of Translational Research logoLink to American Journal of Translational Research
. 2022 Jul 15;14(7):4757–4767.

Subcutaneous continuous glucose monitoring in critically ill patients during insulin therapy: a meta-analysis

Yan Yao 1,*, Yi-He Zhao 1,*, Wen-He Zheng 2, Hui-Bin Huang 1
PMCID: PMC9360883  PMID: 35958452

Abstract

Background: Using continuous glucose monitoring (CGM) in critically ill adult patients requiring insulin therapy has increased with inconsistent results. Thus, we conducted a meta-analysis to assess the effect of CGM and frequent point-of-care (POC) measurements in such a patient population. Methods: We searched PubMed, Embase, Cochrane Library, China national knowledge infrastructure, and Wanfang for relevant articles from inception to Jan 15, 2022. Randomized controlled trials (RCTs) were considered if they focused on critically ill patients who required insulin and were treated with CGM or any POC measurements. We used the Cochrane risk evaluating tool to assess study quality. Subgroup analysis and publication bias were also conducted. Results: We finally included 19 RCTs with 1,852 participants. The quality of the included studies were at a low to moderate levels. Overall, CGM devices significantly reduced hypoglycemia incidence (Risk ratio (RR) 0.35; 95% CI, 0.25-0.49; P<0.00001) than the POC measurement. Further subgroup and sensitivity analyses confirmed this result. The CGM group also had lower overall mortality (RR 0.54; 95% CI, 0.34-0.86; P=0.01), lower glucose variability, and nosocomial infection. The time in, below, or above target blood glucose range, insulin use, and length of stay in the ICU were comparable between the two groups. In addition, few studies provided data in favor of decreased nursing workload and medical costs in the CGM group. Conclusions: The CGM technique could significantly reduce hypoglycemia incidence, overall mortality, and glucose variability compared to POC measurement in critically ill patients. However, further large, well-designed RCTs are required to confirm our results.

Keywords: Subcutaneous continuous glucose monitoring, intensive care, hypoglycemia, meta-analysis, mortality

Introduction

Dysglycemia, including hypoglycemia, hyperglycemia, and high hyperglycemic variability, is a typical concition in intensive care unit (ICU) settings, whether patients have prior diabetes or not [1,2]. It is associated with an increased poor prognosis in these patients [1]. Based on the available evidence, maintaining blood glucose (BG) levels around 8.0 mmoL/L seems preferable for most critically ill patients [3]. However, there is an inherent risk of insulin-induced hypoglycemia in BG regulation, which is related to higher mortality [4]. Therefore, precise BG control is essential.

In most ICU settings, several frequent point-of-care (POC) measurements, including fingertip, venous blood, and blood gas analysis, are commonly used to guide insulin therapy. However, intermittent measurement can detect only one instant BG and does not reflect long-term day-to-week glucose levels [5]. Thus, frequent blood collection is essential, leading to inevitable labor-intensive, time-consuming, and cost-increasing [5,6]. Furthermore, intermittent BG measurement may still overlook the hypoglycemia episodes between two measures [7]. Therefore, continuous glucose monitoring (CGM), which can continuously and automatically provide instant BG values, has become more attractive [8].

Subcutaneous CGM is the most established clinical use among the CGM technologies [5]. CGM can measure glucose in interstitial fluid through minimally invasive subcutaneous sensors [9]. Conceptually, CGM can automatically provide BG values every few minutes, thus making it more readily to identify trends in glucose concentrations. On the other hand, CGM help to reduce the incidence of severe hyperglycemia and hypoglycemia by more rapidly and appropriately adjusting insulin infusions [10]. CGM-derived glucose values have shown higher accuracy than BG measurements in diabetic patients [11]. Recently, CGM has been gradually applied to critically ill patients. Numerous studies have demonstrated that the subcutaneous CGM devices have relatively good accuracy in measuring interstitial glucose levels and are not affected by electrolyte and acid-base imbalance, the severity of illness, and BMI in critically ill patients [12]. In addition, CGM is less invasive with a lower risk of infection, reduced blood loss, and is popular among ICU members for its ease to use [9,13]. However, whether these advantages of CGM translate into improved patient prognosis remains unclear. Several randomized controlled trials (RCTs) compared CGM with POC measurement to guide insulin use in ICU patients with inconsistent results [6,12,14]. This might be related to the different studies’ BG control strategies and the small sample sizes of the RCTs.

Several RCTs on this topic have recently been published [15-18], with some of these studies having a small sample size and inconsistent conclusions. As a result, we aimed to conduct a systematic review and meta-analysis of available RCTs to address the above limitations using the increased power of meta-analytic techniques. We hypothesized that CGM devices might benefit more for glucose control and the prognosis than any POC measurements in critically ill patients requiring insulin therapy.

Methods

This study protocol was registered on the International Platform of Registered Systematic Review and Meta-analysis Protocols database (INPLASY2021120102), and is available at inplasy.com (https://inplasy.com/inplasy-202112-0102). We present our results following the Preferred Reporting Items of Systematic reviews and Meta-Analyses guidance [19] ( Appendix File 1 ). This work did not require ethical approval.

Search strategy and selection criteria

Two investigators (YY and Y-HZ) ran a systematic search without language restrictions in the PubMed, Embase, Cochrane Library, Wanfang, and China national knowledge infrastructure, from inception through Jan 15, 2022, the date of our most recent search. We searched for potentially relevant RCTs using CGM in critically ill patients using Medical Subject Headings and keywords. Our full-search strategy is attached in Appendix File 2 . There were no restrictions based on language. We screened titles and abstracts for eligibility and assessed full texts of the potentially eligible articles for final eligibility. We also reviewed the reference lists of related papers to find relevant studies. Disagreements were solved through discussion by the two review authors.

RCTs were eligible if they compared insulin treatment guided by subcutaneous CGM devices to any frequent point-of-care (POC) measurements in critically ill adult (≥18 years old) patients. We excluded trials enrolling children, breastfeeding women, or pregnant. Studies that reported microdialysis for detecting glucose concentration in the interstitial fluid were excluded since they were techniques that preceded the CGM devices of today. We also removed papers that were only available in abstract form, meeting reports, or included less than ten patients. Disagreements were reviewed by a third author (H-BH), who had a deciding vote.

Data extraction and quality assessment

The aforementioned independent investigators (YY and Y-HZ) undertaken data extraction from included trials on study design, first author, published year, patient characteristics, study interventions, and clinical outcomes of interest.

We used the Cochrane risk-of-bias method to examine RCTs for evidence of bias [20]. We assigned high, unclear, or low values to the following items: sequence generation, allocation concealment, blinding, incomplete outcome data, selective outcome reporting, and other sources of bias. We determined that if there was a high risk of bias in any area, the overall risk of bias in the study was high. As caregivers blinding was difficult in these studies, we evaluated blinding solely at the outcome assessment. Disagreements were recognized and resolved through discussion between the two authors.

Data analysis

The primary outcome was the incidence of hypoglycemia (defined by the author of included RCTs, respectively). Secondary outcomes included time in, below, or above the target BG range (%), glucose variability parameters (i.e., coefficient of variation [CV], or mean amplitude of glucose excursions [MAGE], as defined by authors), length of stay (LOS) in ICU or hospital, mortality (28 days or ICU or hospital), nosocomial infection, nursing workload and medical cost.

Testing the potential influencing factors of our primary outcome, we conducted sensitivity analyses by pooled studies with the following: (1) type of CGM devices; (2) study design (blinding or non-blinding); (3) POC measurement; (4) low limitation of target BG range >6.1 mmol/L or <6.1 mmol/L; and (5) hypoglycemia definition (<2.2 mmol/L, <3.5 mmol/L, or <4 mmol/L), (6) geographic location (China or the other countries) and glucose management protocol (with or without intensive glucose control management), if available.

For dichotomous outcomes, we combined the results to calculate the pooled risk ratio (RR) and associated 95% confidence intervals (CI). Mean differences (MD) and 95% CI were calculated for continuous outcomes. Some studies reported treatment effects measured in the median with interquartile range (IQR). Thus, we calculated the mean from the median and the standard deviations (SD) from IQR [21]. To evaluate heterogeneity, we employed the I 2 statistics. In situations with inconsiderable heterogeneity (I 2<50%), a fixed-effect model was employed, whereas a random-effect model was used in cases of significant heterogeneity (I 2>50%) [22]. Whenever there was heterogeneity, sensitivity analyses were conducted, eliminating one trial in each turn to examine the impact of a single study on the total pooled estimate. Furthermore, where data from at least two RCTs were available, we conducted statistical analyses. Visual inspection of funnel plots was used to determine Publication bias. We performed all used Review Manager (Version 5.3) to conduct all of our analyses.

Results

Trial identification and characteristics

Our electronic search revealed 1689 citations after database searching, 1345 were selected for full-text review, and 19 RCTs were eligible for final analysis (Figure 1) [6,12,14-18,23-34]. The excluded studies based on the full-text review with exclusion reasons were summarized in Appendix File 2 .

Figure 1.

Figure 1

Selection process for studies included in the meta-analysis.

Table 1 describes the key characteristics of the included RCTs, whereas Appendix File 3 describes the BG parameters and outcome data. The final analysis comprised 1852 participants (sample sizes ranging from 24 to 177 individuals), including 885 in the CGM group and 967 in the POC group. These studies were published from 2008 onwards. Most studies enrolled mixed-ICU patients, except five from neurosciences [23,31,34], medical [12], or cardiosurgical ICU patients [24]. Studies varied concerning the target BG range and severe hypoglycemia criteria. As to the POC measurements used, fingertip BG, arterial BG, and peripheral venous BG were used in 17, 5, and 1 studies. CGM system application varied among the included trials, with the Medtronic MiniMed of the most used. The risk of bias was low across the included RCTs ( Appendix File 4 ). A visual examination of a funnel plot revealed no evidence of publication bias ( Appendix File 5: Figure S5).

Table 1.

Characteristics of the included studies

Study Country Design Type of patients APACHEII % Diabetes Sample size (CGM/C) CGM system Control BG measurement Study duration (Hour) Mean age (CGM/C), Year Male (CGM/C) %
Holzinger 2010 [12] Austria RCT; SC; SB MICU NA NA 63/61 Medtronic MiniMed Selective arterial BG 72 58/62 68/57
Huang 2011 [27] China RCT; SC; NB Mixed ICU NA 100 40/80 Medtronic MiniMed Fingertip BG 168 31-72 57
Leelarathna 2013 [23] United Kingdom RCT; SC; NB NICU 12.9 SC 100 12/12 FreeStyle Navigator Arterial BG 48 63/58 75/75
11.2 SC
Kopecky 2013 [24] Prague RCT; SC; NB Cardiac ICU NA 33 12/12 Medtronic MiniMed Arterial BG 24 68/68 50/67
Boom 2014 [6] Netherlands RCT; SC; SB Mixed ICU NA 100 78/78 FreeStyle Navigator I, Abbott Indwelling arterial catheter BG 24 66/67 48/61
De Block 2015 [26] Belgium RCT; MC; SB Mixed ICU 29xed 22.8 16/19 GlucoDay, A. Menarini Diagnostics Arterial BG 96 64/68 50/48
28xe
Qi 2016 [28] China RCT; SC; NB Mixed ICU NA 81.2 48 NA Fingertip BG NA 47-85 54
Sun 2017 [25] China RCT; SC; NB Mixed ICU NA 100 135 NA Fingertip BG 168 59/59 51
Preiser 2018 [14] Belgium RCT; SC; SB Mixed ICU ≥10 100 39/38 Gluco Clear Peripheral venous catheter BG 72 62/60 80/66
Lu 2018 [15] China RCT; SC; SB Mixed ICU 22xed 27 74/70 DGMS, San MediTech Fingertip BG 120 50/49 50/48
22xe
Zhang 2020 [16] China RCT; SC; NB Mixed ICU 20.1d ICU 0 32/32 Medtronic MiniMed Fingertip BG NA 69/68 59/63
28.9d IC
Guan 2017 [30] China RCT; SC; NB Mixed ICU 15-25 0 60/70 CGM-2009 Fingertip BG 72 52 58
Li 2019 [17] China RCT; SC; NB Mixed ICU 16.2±3.2/16.9±3.8 0 37/35 Medtronic MiniMed Fingertip BG 168 59/62 NA
LV 2012 [31] China RCT; SC; SB NICU ≥15 0 59/58 DGMS, San MediTech Fingertip BG 72 61 52
Tian 2019 [32] China RCT; SC; SB Mixed ICU NA 19.2 81/80 RGMS-III Fingertip BG 72 61/65 69/65
Wang 2015 [33] China RCT; SC; NB Mixed ICU NA 100 32/64 Medtronic MiniMed Fingertip BG 72 58 44
Yuan 2008 [34] China RCT; SC; NB NICU 16.6±6.8/15.3±6.9 16.1 36/32 Medtronic MiniMed Fingertip BG 168 65/66 72/69
Fan 2013 [29] China RCT; SC; NB Mixed ICU >15 0 69/79 CGMS-2009 Fingertip BG 72 58/60 56/53
Zhang 2018 [18] China RCT; SC; NB Mixed ICU NA NA 58/58 TouchRulteaTM Fingertip BG NA 45/44 60/62

APACHEII = Acute Physiology, Age and Chronic Health Evaluation II; BG = blood glucose; C = control group; CGM = continuous glucose monitoring; M = male; MC = multiple-center; MICU = medical ICU; NA = not available; NB = not blind; NICU = Neurosciences ICU; RCT = randomized controlled trial; SB = single blind; SC = single-center.

Primary outcome

Data on the incidence of hypoglycemia were available in the 19 RCTs. In the pooled analysis, the use of CGM devices significantly reduced hypoglycemia incidence (19 trials, 1,572 patients; RR 0.35, 95% CI 0.25 to 0.49, P<0.00001; I 2=0%) (Figure 2). Despite the absence of considerable heterogeneity, we conducted stratified analyses based on predefined major research features and clinical variables. In general, all the subgroup studies indicated that the occurrences of hypoglycemia among groups were similar. Sensitivity analyses were then performed, revealing that the results were constant when the analyses were limited to studies that defined hypoglycemia as <2.2 mmol/L or <3.3 mmol/L, or <4 mmol/L. Table 2 shows the findings of subgroup and sensitivity analyses in detail.

Figure 2.

Figure 2

Forest plot showing the incidence of hypoglycemia.

Table 2.

Subgroup analyses on primary outcome of Incidence of hypoglycemia

Subgroup analyses Study number Patient number Hypoglycemia event in CGM croup Hypoglycemia event in control croup Risk ratio (95% CI) I2, % P
CGM devices Medtronic [12,17,24,34] 244 5 of 148 22 of 140 0.23 [0.09, 0.57] 0 0.001
Others* [6,14,15,18,23,26,29-32] 1,101 26 of 542 67 of 559 0.40 [0.26, 0.61] 13 <0.0001
Not reported [25,28] 183 6 of 91 20 of 92 0.32 [0.14, 0.74] 20 0.008
Study design Blinded [6], 12, 14, 15, 31, 32], 730 14 of 367 31 of 363 0.46 [0.25, 0.83] 31 0.01
Unblinded [17,18,23-25,28-30,34] 842 23 of 414 78 of 428 0.30 [0.20, 0.47] 0 <0.00001
Control BG measurement Fingertip [15,17,18,25,28-32,34] 992 25 of 552 76 of 559 0.33 [0.22, 0.51] 2 <0.00001
Arterial [6,12,23,24,26] 384 4 of 190 18 of 194 0.27 [0.10, 0.68] 0 0.006
Venous [14] 77 8 of 39 15 of 38 0.52 [0.25, 1.08] - -
Low limitation of target BG range ≥6.1 mmol/L [12,15,25,28,29,31] 688 13 of 342 47 of 345 0.29 [0.16, 0.51] 0 <0.0001
<6.1 mmol/L [6,14,17,23-26] 461 13 of 203 34 of 206 0.40 [0.23, 0.70] 0 0.001
Not reported [18,30,32,34] 409 11 of 235 28 of 240 0.39 [0.20, 0.78] 34 0.006
Average APACHEII score >20 [15,26,29,31] 475 9 of 205 29 of 211 0.33 [0.16, 0.67] 4 0.002
<20 [17,23,34] 164 4 of 85 13 of 79 0.28 [0.10, 0.83] 0 0.02
Others [6,10,12,18,24,25,28,30,32] 992 4 of 85 13 of 79 0.37 [0.24, 0.57] 28 <0.00001
Country China [15,17,18,25,28,29,31,32,34] 1,111 25 of 552 76 of 559 0.33 [0.22, 0.51] 2 <0.0001
Others [6,12,14,23,24,26,30] 461 12 of 229 33 of 232 0.38 [0.21, 0.68] 0 0.001
% Diabetes >50% [6,14,23,25,33] 461 14 of 229 35 of 232 0.40 [0.23, 0.70] 0 0.001
<50% [15,17,24,26,29,30-32,34] 871 20 of 431 55 of 440 0.38 [0.23, 0.41] 0 <0.0001
Not reported [12,18] 240 3 of 121 19 of 119 0.16 [0.05, 0.51] 0 0.002

BG = blood glucose; CGM = continuous glucose monitoring.

*

FreeStyle Navigator, GlucoDay, Gluco Clear, DGMS, San MediTech, CGM-2009, RGMS-III, and TouchRulteaTM.

Secondary outcomes

The CGM group had a lower overall mortality rate (12 trials, n=1,294; RR 0.54; 95% CI, 0.34 to 0.86; I 2=56%; P=0.01) (Figure 3). The time in target BG range (7 trials, n=584, MD 7.58%; 95% CI, -0.18 to 15.34, I 2=85%; P=0.06), time below target BG range (4 trials, n=422, MD 1.31%; 95% CI, -3.63 to 1.00, I 2=82%; P=0.27) and time above target BG range (3 trials, n=345, MD -7.86%; 95% CI, -20.54 to 4.83, I 2=88%; P=0.22) were also similar. In addition, there was no significantly difference in insulin use (6 trials, n=516, MD 0.01; 95% CI, -0.17 to 0.18, I 2=0%; P=0.93) and the length of stay in ICU between the two groups (9 trials, n=863; MD -2.28 days; 95% CI, -5.99 to 1.39; I 2=98%; P=0.22). As to glucose variability parameters, six trials provided outcome of MAGE and pooled data showed lower MAGE in CGM group (n=767; MD -1.41 mmoL/L; 95% CI, -2.24 to -0.58; I 2=95%; P=0.0009); while four used CV as interest and the pooled data tended to be lower during CGM therapy (n=404; MD -1.41%; 95% CI, -3.50 to 0.46; I 2=88%; P=0.08). Four trials reported outcome of nosocomial infection [16,18,32,34] and the pooled data showed that CGM group had a significantly lower infections (n=378; RR 0.21; 95% CI, 0.10 to 0.44; I 2=9%; P<0.0001). Only two RCTs described the costs between the two strategies, with one [6] showing the CGM group had lower mean daily costs per patient while the other [28] reported no difference between groups. (See Appendix File 6: Figures S6, S7, S8, S9, S10, S11, S12 and S13).

Figure 3.

Figure 3

Forest plot showing the overall mortality rate.

Discussion

In the current meta-analysis, we compared the use of CGM devices with any frequent POC measurements in critically ill adult patients requiring insulin treatment for ICU dysglycemia. We found that CGM technique could significantly reduce hypoglycemia incidence during insulin treatment. Further subgroup and sensitivity analyses confirmed this finding. CGM-guided insulin treatment was associated with lower overall mortality, nosocomial infection, and glucose variability than the POC measurement. In addition, time in, below, or above the targeted glucose range, insulin use, and ICU LOS were comparable between the two groups.

Our findings are consistent with those of a recent meta-analysis published in Chinese [35], which showed that the use of CGM decreased hypoglycemia incidence in critically ill patients. However, the pooled data from ICU and non-ICU research may have contributed to the high heterogeneity of included RCTs. Moreover, the authors included only seven trials of 531 patients and mainly focused on patients receiving intensive insulin therapy, a glucose control strategy that preceded the CGM devices of today and had already not been recommended to apply in critically ill patients. To overcome these limitations, we enlarged the prior meta-analysis to include 19 RCTs with more than 1,800 patients [6,12,14-18,23-34]. Thus, with larger sample size, we had more power to evaluate the effect of CGM in the ICU setting and conducted the subgroup and sensitivity analyses based on various clinical characteristics to confirm our primary outcome. Moreover, we also assessed other important clinical outcomes (e.g., time in, below, above target glucose time, overall mortality, insulin use, and ICU LOS). These findings provided effect and safety evidence of the robustness of our primary outcome.

Our results suggest some important clinical implications of CGM. First, the CGM-guided glucose control strategy significantly reduced the occurrence of hypoglycemia by 67%, suggesting that CGM could help ICU members detect hyperglycemia and hypoglycemia, thus reducing potential complications in critically ill patients. Second, CGM devices help reduce the need for regular blood glucose monitoring. This is a time-consuming and labor-intensive procedure in the ICU setting, especially for high glucose fluctuation, such as steroid-induced hyperglycemia, diabetic ketoacidosis, or hyperglycemic hyperosmolar syndrome. Most CGM devices only require calibration 2-3 times per day and reduce the workload of medical staff in the ICU. In the study by Boom and colleagues [6], the authors found the CGM significantly reduced the daily nursing workload for glucose control (17 versus 36 minutes; P<0.001) compared to the intermittent POC glucose measurements. In two COVID-19 studies, CGM devices were associated with a reduction in POC testing by 60% [36] and 63% [37] in critically ill patients who required continuous insulin infusions. In addition, using CGM can shorten the time the caregivers contact the critically ill patients, thereby reducing the risk of transmission of infectious microorganisms to ICU medical staff, especially from the COVID-19 patients [36,37].

Our results showed that the CGM-guided strategy significantly reduced mortality in critically ill patients. It is consistent with other improved outcomes in the present meta-analysis, such as decreasing the incidence of hypoglycemia and improved glycemic variability. Several studies have demonstrated that more significant glycemic variability in ICU patients was related to higher in-hospital mortality, independent of mean BG levels and hypoglycemia [38,39]. However, we should interpret the mortality outcome with caution since this result has significant heterogeneity. As a secondary outcome, we did not further explore the potential factors for the heterogeneity. However, explaining the heterogeneity is very difficult because the heterogeneity might be due to the different etiologic distributions of the ICU population, disease severity, glycemic control strategies, and organ function supports among the included studies. In addition, most included RCTs with a small sample size are more likely to conclude an overestimation of the treatment effect. Therefore, further clarification of the results by high-quality RCT studies is required.

We found that CGM did not show an advantage in time in the target glucose range, which might be related to the development of glucose management and experienced nursing teams in the ICU. The control group maintained glycemic control well among the included studies, possibly offering little space for improvement by CGM [12]. In addition, patients with varied difficulty in glycemic control, e.g., in the study by Lu et al. [15], about 50% of recruited patients were SAP, which made controlling blood glucose levels and getting them into the target range more challenging than in other studies.

Although our results are encouraging. However, CGM still needs some improvement. First, in the future, more designs and applications of closed-loop systems integrating CGM and automatic insulin infusion systems will be required and ultimately achieve the goal of fully closed-loop blood glucose management. Second, the life span of the biosensor is relatively short, about seven days. Third, the Mean Absolute Relative Difference (MARD), an indicator to quantify the deviation from the reference measurement, was 12.5% in the present meta-analysis, which complied with the clinical application criteria of CGM [40]. However, we should pay more attention to the accuracy of CGM in some situations in the ICU setting. For example, CGM has a lag time due to glucose transport from the blood to the subcutaneous interstitium, which takes approximately 15-20 minutes [9]. Thus, if the patient’s blood glucose level fluctuates widely, the lag time should be considered. There is currently a lack of adequate data to correlate glucose levels in the blood with interstitial fluids in some individuals with severe generalized edema, such as hypoalbuminemia and hepatic failure. In addition, critically ill patients often require high doses of pressors, leading to peripheral vasoconstriction. In such patients, blood circulation to the skin where the CGM is placed may decrease, thus affecting CGM readings’ accuracy. Although a small prospective study focusing on intraoperative CABG showed good accuracy in patients receiving CGM and large amounts of vasoactive drugs (MARD of 12.9 and Clarke error grid analysis showed 98.6% of glucose values falling into zones A and B) [41]. More data from the ICU population are still needed to confirm this.

There are some limitations to our meta-analysis. (1) Few studies provide data related to the accuracy of CGM, such as MARD, Clarke Error Grid (CEG), and ISO criteria. (2) Several secondary outcomes, including the costs, workload, and infections during the glucose management guided by CGM, were reported in only a few studies. (3) Significant heterogeneity was found among the included studies. However, we did not explore the heterogeneity due to fewer related studies. (4) Most included studies were single-center designs, and there may be differences in the underlying treatment among the centers. Also, most included studies were small samples, which could amplify its effect. (5) CGM requires calibration with fingerstick glucose, usually 2-3 times per day. However, only about half of the studies provide data on CGM calibration. Enhanced calibration leads to a decreased MARD, more points in CEG zone A, and better conformity with ISO criteria [42].

Conclusion

More and more physicians focus on CGM during glucose management in the ICU. Generally, our meta-analysis of aggregate data shows that using the CGM technique significantly reduces hypoglycemia incidence, overall mortality, nosocomial infection, and glucose variability compared to POC measurement in critically ill patients. However, further large, well-designed RCTs will be needed to confirm our results.

Disclosure of conflict of interest

None.

Supporting Information

ajtr0014-4757-f4.pdf (1.2MB, pdf)

References

  • 1.Badawi O, Waite MD, Fuhrman SA, Zuckerman IH. Association between intensive care unit-acquired dysglycemia and in-hospital mortality. Crit Care Med. 2012;40:3180–3188. doi: 10.1097/CCM.0b013e3182656ae5. [DOI] [PubMed] [Google Scholar]
  • 2.McCowen KC, Malhotra A, Bistrian BR. Stress-induced hyperglycemia. Crit Care Clin. 2001;17:107–124. doi: 10.1016/s0749-0704(05)70154-8. [DOI] [PubMed] [Google Scholar]
  • 3.Mesotten D, Van den Berghe G. Glycemic targets and approaches to management of the patient with critical illness. Curr Diab Rep. 2012;12:101–107. doi: 10.1007/s11892-011-0241-8. [DOI] [PubMed] [Google Scholar]
  • 4.Hermanides J, Bosman RJ, Vriesendorp TM, Dotsch R, Rosendaal FR, Zandstra DF, Hoekstra JB, DeVries JH. Hypoglycemia is associated with intensive care unit mortality. Crit Care Med. 2010;38:1430–1434. doi: 10.1097/CCM.0b013e3181de562c. [DOI] [PubMed] [Google Scholar]
  • 5.Sun MT, Li IC, Lin WS, Lin GM. Pros and cons of continuous glucose monitoring in the intensive care unit. World J Clin Cases. 2021;9:8666–8670. doi: 10.12998/wjcc.v9.i29.8666. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Boom DT, Sechterberger MK, Rijkenberg S, Kreder S, Bosman RJ, Wester JP, van Stijn I, DeVries JH, van der Voort PH. Insulin treatment guided by subcutaneous continuous glucose monitoring compared to frequent point-of-care measurement in critically ill patients: a randomized controlled trial. Crit Care. 2014;18:453. doi: 10.1186/s13054-014-0453-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Krinsley JS, Bruns DE, Boyd JC. The impact of measurement frequency on the domains of glycemic control in the critically ill--a Monte Carlo simulation. J Diabetes Sci Technol. 2015;9:237–245. doi: 10.1177/1932296814566507. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Krinsley JS, Chase JG, Gunst J, Martensson J, Schultz MJ, Taccone FS, Wernerman J, Bohe J, De Block C, Desaive T, Kalfon P, Preiser JC. Continuous glucose monitoring in the ICU: clinical considerations and consensus. Crit Care. 2017;21:197. doi: 10.1186/s13054-017-1784-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Chen C, Zhao XL, Li ZH, Zhu ZG, Qian SH, Flewitt AJ. Current and emerging technology for continuous glucose monitoring. Sensors (Basel) 2017;17:182. doi: 10.3390/s17010182. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Scrimgeour LA, Potz BA, Sellke FW, Abid MR. Continuous glucose monitoring in the cardiac ICU: current use and future directions. Clin Med Res (N Y) 2017;6:173–176. [PMC free article] [PubMed] [Google Scholar]
  • 11.Perez-Guzman MC, Shang T, Zhang JY, Jornsay D, Klonoff DC. Continuous glucose monitoring in the hospital. Endocrinol Metab (Seoul) 2021;36:240–255. doi: 10.3803/EnM.2021.201. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Holzinger U, Warszawska J, Kitzberger R, Herkner H, Metnitz PG, Madl C. Impact of shock requiring norepinephrine on the accuracy and reliability of subcutaneous continuous glucose monitoring. Intensive Care Med. 2009;35:1383–1389. doi: 10.1007/s00134-009-1471-y. [DOI] [PubMed] [Google Scholar]
  • 13.Kosiborod M, Gottlieb RK, Sekella JA, Peterman D, Grodzinsky A, Kennedy P, Borkon MA. Performance of the Medtronic Sentrino continuous glucose management (CGM) system in the cardiac intensive care unit. BMJ Open Diabetes Res Care. 2014;2:e000037. doi: 10.1136/bmjdrc-2014-000037. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Preiser JC, Lheureux O, Thooft A, Brimioulle S, Goldstein J, Vincent JL. Near-continuous glucose monitoring makes glycemic control safer in ICU patients. Crit Care Med. 2018;46:1224–1229. doi: 10.1097/CCM.0000000000003157. [DOI] [PubMed] [Google Scholar]
  • 15.Lu MZ, Zuo YY, Guo J, Wen XP, Kang Y. Continuous glucose monitoring system can improve the quality of glucose control and glucose variability compared with point-of-care measurement in critically ill patients: a randomized controlled trial. Medicine (Baltimore) 2018;97:e12138. doi: 10.1097/MD.0000000000012138. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Zhang H. Effect of continuous real-time glucose monitoring system on blood glucose level in severe patients. J Shandong Med Coll. 2020;192:69–70. [Google Scholar]
  • 17.Li M, Yao L, Ji XQ, Chen C, Cui J, Zhao JJ, Wu YY. Effect of real-time continuous monitoring system on serum inflammatory factors and prognosis in patients with sepsis. J Endocrine Surg. 2019;13:245–248. [Google Scholar]
  • 18.Zhang K. Accuracy and safety of ambulatory blood glucose monitoring system in ICU critically ill patients. Chin Health Care Nutr. 2018;28:106. [Google Scholar]
  • 19.Moher D, Liberati A, Tetzlaff J, Altman DG PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. BMJ. 2009;339:b2535. doi: 10.1136/bmj.b2535. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Higgins JP, Altman DG, Gøtzsche PC, Jüni P, Moher D, Oxman AD, Savovic J, Schulz KF, Weeks L, Sterne JA Cochrane Bias Methods Group; Cochrane Statistical Methods Group. The cochrane collaboration’s tool for assessing risk of bias in randomised trials. BMJ. 2011;343:d5928. doi: 10.1136/bmj.d5928. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Wan X, Wang WQ, Liu JM, Tong TJ. Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range. BMC Med Res Methodol. 2014;14:135. doi: 10.1186/1471-2288-14-135. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Higgins JP, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in meta-analyses. BMJ. 2003;327:557–560. doi: 10.1136/bmj.327.7414.557. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Leelarathna L, English SW, Thabit H, Caldwell K, Allen JM, Kumareswaran K, Wilinska ME, Nodale M, Mangat J, Evans ML, Burnstein R, Hovorka R. Feasibility of fully automated closed-loop glucose control using continuous subcutaneous glucose measurements in critical illness: a randomized controlled trial. Crit Care. 2013;17:R159. doi: 10.1186/cc12838. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Kopecky P, Mraz M, Blaha J, Lindner J, Svacina S, Hovorka R, Haluzik M. The use of continuous glucose monitoring combined with computer-based eMPC algorithm for tight glucose control in cardiosurgical ICU. Biomed Res Int. 2013;2013:186439. doi: 10.1155/2013/186439. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Sun Y, Wang S. Clinical effect of CGMS and intensive insulin therapy on ICU critical patients with hyperglycemia. Med J NDFNC. 2017;10:688–691. [Google Scholar]
  • 26.De Block CE, Gios J, Verheyen N, Manuel-y-Keenoy B, Rogiers P, Jorens PG, Scuffi C, Van Gaal LF. Randomized evaluation of glycemic control in the medical intensive care unit using real-time continuous glucose monitoring (REGIMEN trial) Diabetes Technol Ther. 2015;17:889–898. doi: 10.1089/dia.2015.0151. [DOI] [PubMed] [Google Scholar]
  • 27.Brunner R, Adelsmayr G, Herkner H, Madl C, Holzinger U. Glycemic variability and glucose complexity in critically ill patients: a retrospective analysis of continuous glucose monitoring data. Crit Care. 2012;16:R175. doi: 10.1186/cc11657. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Qi A, Li H. Efficacy and security of dynamic glucose monitoring combined with continu-ous subcutaneous insulin infusion in treatment of critically ill patients. Diabetes New World. 2016;R587:116–118. [Google Scholar]
  • 29.Fan X, Liu M. Significance of dynamic blood glucose monitoring on glycemic management in patients with stress hyperglycemia. J Intern Intensive Med. 2013;19:282–293. [Google Scholar]
  • 30.Guan Y, Liu L. Accuracy and safety of CGMS in ICU patients. Today Nurse. 2017;8:90–91. [Google Scholar]
  • 31.Lv SY, Ji MS, Kong XR, Cai Y, Jin ZC. Real-time continuous glucose monitoring system in critically craniocerebral trauma patients with hyperglycemia. J Jiangsu Univ. 2012;22:497–499. [Google Scholar]
  • 32.Tian S, Li J, Wang J. Application of dynamic glucose monitoring system in intensive insulin therapy in severe patients. Chin J Prim Med Pharm. 2018;25:3102–3104. [Google Scholar]
  • 33.Wang L. Combined application of continuous glucose monitoring system and insulin pump in ICU diabetic patients. J Taihan Med Coll. 2015;36:1416–1417. [Google Scholar]
  • 34.Yan S, Feng Q, Zhong LH. Therapy of CGMS combined with insulin pump for critical patients with hyperglycemia. Chin J Crit Care Med. 2008;28:707–709. [Google Scholar]
  • 35.Wang C, Zhu Y, Shuai X. The effect of real time continuous blood glucose monitoring system versus intermittent blood glucose monitoring in critically ill patients under the intensive insulin therapy: a meta-analysis. Chin J Emerg Med. 2015;24:320–324. [Google Scholar]
  • 36.Agarwal S, Mathew J, Davis GM, Shephardson A, Levine A, Louard R, Urrutia A, Perez-Guzman C, Umpierrez GE, Peng LM, Pasquel FJ. Continuous glucose monitoring in the intensive care unit during the COVID-19 pandemic. Diabetes Care. 2021;44:847–849. doi: 10.2337/dc20-2219. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Davis GM, Faulds E, Walker T, Vigliotti D, Rabinovich M, Hester J, Peng L, McLean B, Hannon P, Poindexter N, Saunders P, Perez-Guzman C, Tekwani SS, Martin GS, Umpierrez G, Agarwal S, Dungan K, Pasquel FJ. Remote continuous glucose monitoring with a computerized insulin infusion protocol for critically ill patients in a COVID-19 medical ICU: proof of concept. Diabetes Care. 2021;44:1055–1058. doi: 10.2337/dc20-2085. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Egi M, Bellomo R, Stachowski E, French CJ, Hart G. Variability of blood glucose concentration and short-term mortality in critically ill patients. Anesthesiology. 2006;105:244–252. doi: 10.1097/00000542-200608000-00006. [DOI] [PubMed] [Google Scholar]
  • 39.Al-Dorzi HM, Tamim HM, Arabi YM. Glycaemic fluctuation predicts mortality in critically ill patients. Anaesth Intensive Care. 2010;38:695–702. doi: 10.1177/0310057X1003800413. [DOI] [PubMed] [Google Scholar]
  • 40.Wernerman J, Desaive T, Finfer S, Foubert L, Furnary A, Holzinger U, Hovorka R, Joseph J, Kosiborod M, Krinsley J, Mesotten D, Nasraway S, Rooyackers O, Schultz MJ, Van Herpe T, Vigersky RA, Preiser JC. Continuous glucose control in the ICU: report of a 2013 round table meeting. Crit Care. 2014;18:226. doi: 10.1186/cc13921. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Perez-Guzman MC, Duggan E, Gibanica S, Cardona S, Corujo-Rodriguez A, Faloye A, Halkos M, Umpierrez GE, Peng L, Davis GM, Pasquel FJ. Continuous glucose monitoring in the operating room and cardiac intensive care unit. Diabetes Care. 2021;44:e50–e52. doi: 10.2337/dc20-2386. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.De Block C, Manuel-y-Keenoy B, Van Gaal L, Rogiers P. Intensive insulin therapy in the intensive care unit: assessment by continuous glucose monitoring. Diabetes Care. 2006;29:1750–1756. doi: 10.2337/dc05-2353. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ajtr0014-4757-f4.pdf (1.2MB, pdf)

Articles from American Journal of Translational Research are provided here courtesy of e-Century Publishing Corporation

RESOURCES