A review of portable quantitative and semi-quantitative devices for measurement of vitamin A in biological samples

Samantha L Huey a, Jesse T Krisher a, David Morgan b, Penjani Mkambula b, Bryan M Gannon a, Mduduzi NN Mbuya c, Saurabh Mehta a,d,
PMCID: PMC9407042  PMID: 36033130

Graphical abstract

Assessment of vitamin A concentrations in biological specimens or functional indicators of vitamin A status (severity of night blindness or lack thereof) require field-friendly methods for assessment to expand coverage, particularly in low-income settings. In this review, portable devices for measurement were compared with reference methods across key criteria. For example, portable devices to determine vitamin A concentrations in blood or milk samples were catalogued and compared with the reference method of high performance liquid chromatography, and a portable dark adaptometer to assess vision was compared against the reference of a Goldmann-Weekers dark adaptometer.

graphic file with name ga1.jpg

Keywords: Vitamin A, Retinol, Beta carotene, Portable devices, Field devices, Resource-limited settings

Highlights

  • Vitamin A deficiency is a major global health issue leading to poor health outcomes.

  • Vitamin A assessment usually requires a centralized laboratory and equipment.

  • Portable methods to assess vitamin A may overcome the limitations of laboratory-based testing.

  • Blood, milk, and eye function may be used for vitamin A status measurement.

Abstract

Background

We catalog and summarize evidence of the analytical performance of portable quantitative and semi-quantitative devices for the assessment of vitamin A status and vitamin A deficiency (VAD) in various biological samples—including whole blood, plasma, serum, and milk—in addition to VAD determination by functional indicators such as pupillary response.

Methods

We searched the literature for published research articles, patents, and information from manufacturers of mobile devices, particularly those appropriate for low-resource settings. The included devices were required to be portable (lightweight and ideally not needing a power outlet) and to measure vitamin A as well as define VAD. Eligible studies compared a portable device to a reference standard of high-performance liquid chromatography for blood and milk, or a Goldmann-Weekers dark adaptometer for eyes/vision. Where available, identified devices were compared with reference methods across several performance criteria. When possible, we compared the device’s performance reported in published studies against the stated performance criteria from the manufacturers’ websites.

Results

We catalogued 25 portable devices for measuring vitamin A and/or VAD via biological samples. We also identified 18 comparison studies (plus associated reports) assessing nine methods: the iCheck Fluoro, iCheck Carotene, CRAFTi, Tidbit with or without the HYPER filtration system, custom field-friendly immunoassays, and microfluidic assays for blood; the iCheck Fluoro and iCheck Carotene for milk; and the Scotopic Sensitivity Tester-1 for eye function.

Conclusions

The iCheck Fluoro and iCheck Carotene are commercially available for use and are acceptable for measuring vitamin A in blood and milk samples, according to the available validation data. Many of the other identified devices, including other portable fluorometers, photometers, immunoassays, microfluidics-based devices, and dark adaptometers, were proofs of concept and not yet commercially available. Furthermore, none of these other devices included manufacturer-described device performance criteria to compare with descriptions from experimental studies. Several gaps remain, including studies comparing the other portable devices against a reference standard, particularly for functional indicators of vitamin A status/deficiency; available manufacturer-reported device performance criteria against which to compare future results of investigations; and more comprehensive reporting of validation metrics including sensitivity, specificity, precision, and Bland-Altman analysis.

Introduction/background

Vitamin A deficiency (VAD) continues to be a major global health issue leading to poor health outcomes, including night blindness, greater severity of measles infection, and higher mortality risk from infectious diseases (Tanumihardjo et al., 2016). Most existing analytical techniques to assess vitamin A status by measuring serum retinol or retinol binding protein require access to a sophisticated laboratory and equipment such as high performance liquid chromatography (HPLC) (de Pee and Dary, 2002). These methods require extensive sample preparation, are time-consuming, and are potentially prohibitively expensive, depending on the number of samples to be analyzed. Furthermore, VAD is more prevalent in lower income countries, where such laboratory resources may be limited or might not yet exist; in recent vitamin A surveys, <20% of pregnant women at risk have been covered by population surveys globally, possibly partly because of a lack of diagnostics (WHO, 2009).

Portable, field-friendly devices and tools for assessing vitamin A status in populations have the potential to overcome some of the limitations of traditional, laboratory-based testing. These methods may differ in their cost, accuracy, reliability, ease of use, and required consumables/reagents for performing the testing.

A review cataloguing the range of portable tests for vitamin A status and VAD in biological samples, and summarizing these devices’ performance with respect to a reference standard method, is not available. Therefore, the goal of this review was to enable current manufacturers to modify and improve their products according to the gaps identified herein, and to set design goals for new products meeting the current demands of industry, regulators, and other stakeholders.

Materials and methods

In December 2020, we conducted a standardized search of the literature indexed in five databases (MEDLINE, EMBASE, World Health Organization Global Index Medicus, Scopus, and Web of Science) with no restrictions on language, location, or date of publication. We designed a search strategy for MEDLINE (PubMed) (Supplementary Table 1) and translated the search strategy for the remaining databases with guidance from the evidence synthesis specialists at Mann Library, Cornell University. We also used an online search engine to search for other sources such as manufacturers’ websites and patents, and we consulted with subject matter experts within our organizations to gain more information.

We catalogued any portable devices measuring vitamin A or vitamin A deficiency in biological samples, either as reported in studies or provided on manufacturers’ websites. We included both portable devices/methods measuring vitamin A status and devices/methods that indicated VAD. Initially, we considered devices measuring skin carotenoids, as shown in our search strategy; however, because of the lack of established guidance or consensus regarding the conversion of skin carotenoid measurements via Raman resonance spectroscopy (e.g., BioPhotonic Scanner (Pharmanex/Nuskin Enterprises, 2018)) to blood carotenoid measurements and overall vitamin A status (von Lintig, 2020), we determined that these devices were beyond the scope of the review.

The inclusion criteria for our analysis of device performance included certain study designs such as proof-of-concept development studies, method comparison studies, and diagnostic test accuracy studies; studies involving human participants (e.g., observational studies or randomized controlled trials) were considered if the authors described using a portable method for analyzing vitamin A in biological samples. Animal studies were also included. Eligible studies were required to measure vitamin A in any biological sample, including blood, eyes, or breast milk, with a portable device and to compare the device performance with that of a reference method, such as HPLC, depending on the sample type. Studies detailing field friendly methods of sample collection (e.g., dried blood spots) necessitating the use of a non-portable device or a laboratory for analysis were considered beyond the scope of this review.

We contacted the authors to request raw data or more information as needed. We also re-analyzed raw data, when available, as needed.

Results and discussion

Catalog of portable devices

From our search (Fig. 1), we catalogued 25 portable devices, kits, and/or field-friendly assays able to assess a variety of biological sample and vitamin A biomarker types in Table 1a (blood, milk) and Table 1b (eyes/vision assessment).

Fig. 1.

Fig. 1

PRISMA Diagram for study identification and screening. ().

Adapted from Page et al., 2021

Table 1a.

Catalog of all portable devices quantifying vitamin A and vitamin A deficiency: blood, milk.

Device (manufacturer)

Principle/method

# Tests per kit
Vitamin A biomarker Pricing Technical requirements

Sample volume, preparation and setup

Overall time required
Portability

Included in kit

Special storage conditions
Consumables:
Reagents required

Power source

Shelf life
Operational range

Quantification

Outputs
Target setting

Manufacturer support available

Global availability
iCheck Fluoro (BioAnalyt GmBH, Teltow, Germany) (BioAnalyt, 2021)

Fluorescence

100 tests per kit
Retinol, retinyl palmitate, retinyl acetate and other esters Pricing not published








+
1 day training

Whole blood, serum, breast milk: 0.5 mL; no preparation required

<10 min



+++
Compact and lightweight (11 × 4 × 20 cm); 0.45 kg

Device + test kit (iEX MILA reaction vials), syringe

None

+++
Optional:
50 mL conical tubes, weighing dishes, reference samples

Rechargeable battery

12 months at 20–30 °C, no direct sunlight, upright


+++
50–3000 µg RE/L

Quantitative

Sample #, batch #, result, date, time (in transferred data); results (µg RE/L) are stored in the device and transferred to a PC via USB

+++
Lab and field

Yes

>80 countries





+++
iCheck Carotene (BioAnalyt GmBH, Teltow, Germany) (BioAnalyt, 2021)

Photometry

100 tests per kit
Beta-carotene






Pricing not published








+
1 day training

Colostrum, cattle whole blood, cattle serum: 0.4 mL; no preparation required

<10 min



+++
Compact and lightweight (11 × 4 × 20 cm); 0.45 kg

Device + test kit

None



+++
Optional:
50 mL conical tubes, weighing dishes, reference samples

Rechargeable battery

12 months at 20–30 °C, no direct sunlight, upright


+++
0.15–15 mg/L

Quantitative

Sample #, batch #, result, date, time (in transferred data); results (µg RE/L) are stored in the device and transferred to a PC via USB

+++
Lab and field

Yes

>80 countries





+++
CRAFTi (Eurofins CRAFT Technologies Inc., Wilson, NC, USA) (Chaimongkol et al., 2011, Eurofins Craft Technologies, 2020)

Fluorescence

NR
Retinol Pricing not published







+
Minimal training

Serum: 25 µL; requires serum separation

30 min



++
Compact and lightweight (13 × 16.5 × 35 cm); 2.1 kg

NR

NR

++
Fluorometer cuvettes or Durham tubes

Battery (12 V and inverter) or line current (115–230 V)

Fluorescent dye fades; must check periodically

++
0.5–1.5 µmol/L

Quantitative

Fluorescence readings; no detail on data appearance



++
Lab and field

Yes

NR




++
Tidbit (Lu and Erickson, 2017, Lu et al., 2017), ± HYPER filtration system (Lu et al., 2018) (Cornell University, Ithaca, NY, USA)

Fluorescence, multicolor lateral flow

NR
RBP Estimated: $95 manufacturing cost; $1.50 per test; Using HYPER platform, <$1 per test






++
Meant for consumer, clinical, and research use

Serum: 15–20 µL, separated from RBCs with a centrifuge Whole blood: 60 µL, using HYPER (Lu et al., 2018)

Serum: 15 min
Whole blood: 5–20 min for HYPER separation)

+++
NR; meant for field use

Tidbit reader, disposable test strip(s)

None




+++
Lightening-Link Conjugation Kits (Innova Bioscience Ltd., HF180 cards (EMD Millipore); Running buffer (60 µL)

Battery; connect to mobile device, Wi-Fi optional

NR


+++
2.2–20 µg/mL (0.10–0.95 µmol/L)

Quantitative

Result from each sample to smartphone (Nutriphone app) or laptop; stores results internally via 16-GB SD card


+++
Lab and field

Corresponding authors: David Erickson or Saurabh Mehta; not commercially available

NR


++
Electronics-enabled (EE)-µPAD (Diagnostics for All) (Lee et al., 2016)

Paper-based microfluidics for immune detection

NR
RBP Estimated: $20 for prototyping; $0.41 per test, but price expected to decrease below $10 per unit and at $1 per biomarker per unit








++
Meant for clinicians and researchers

Whole blood: 35 µL; no preparation required

13 min








+++
Size of a credit card

NR

NR










++
µPAD

Battery

Stored at room temperature in desiccator box until use








+++
∼10 µg/mL to < 70 µg/mL, according to on graph (Fig. 5)

Quantitative

Measurements are wirelessly transferred to a mobile phone application that geo-tags the data and transmits it to a remote server for real time tracking of micronutrient deficiencies; NFC-enabled smartphone required

+++
Resource limited settings

Yes, but not commercially available

NR







++
RBP-EIA (Scimedx Corp., Dover, NJ, USA) (Hix et al., 2004, Hix et al., 2006, 2020)

Antigen competition assay

8–96 tests per kit
RBP Estimated: <$3.00 per test; pricing not published







++
Meant for health care workers

Serum: 10 µL; portable battery-operated centrifuge to separate whole blood, vortex serum samples and store on ice until assay completed

40 min

++
Size of 96-well plate; requires sink for washing step

NR

NR



++
Well plate, monoclonal anti-RBP antibody, wash buffer, substrate

None required

NR



++
10–40 µg/mL (0.48–1.92 µmol/L)

Quantitative

Read optical densities using EIA plate reader (Revelation, Dynex)


++
Lab and field

Yes

NR





+++
Antigen-antibody reaction based on liquid-semisolid phase (custom) (Ciaiolo et al., 2015)

Custom antigen–antibody reaction based on liquid-semisolid phase + visualization system

NR
RBP Pricing not published









+
Meant for research or diagnostics

Serum: 5 µL; requires serum separation


30 min



++
Petri dishes, pipette, portable viewer, and glass gel holder

NR

NR




++
Dilution: PBS due to high protein concentration; gel; antibodies

None required

NR




++
Depends on time for reaction; range 64 mg/L to 1 mg/L

Qualitative, semi-quantitative

Immunoprecipitates, scored as: “-, +, ++, +++, ++++”

++
Lab and field

Corresponding author: Carlo Ciaiolo; but not commercially available

NR



++
RID plate reader (The Binding Site, San Diego, CA, USA) (Hix et al., 2004, The Binding Site, 2020)

Radial immunodiffusion of antigen–antibody precipitin rings

1–3 plates per kit
RBP Pricing not published
















+
Meant for researchers

Serum: 5 µL; requires serum separation

Incubation for 3 days












+
Compact and lightweight (22 × 14 × 16 cm, 1.14 kg)

User guide and installation CD, USB-A to USB-B cable, power supply, plate reader calibration plate

Indoor use only, altitude < 2000 m, 5–40 °C, relative humidity ≤ 80% at < 31 °C (or ≤ 50% if > 31 °C)

++
Microsoft Windows computer

Power adapter, USB port

NR (warranty: 1 year)












++
Depends on analyte; range for RBP not reported

Quantitative

Precipitin ring diameters, mm











++
Lab

Yes

NR













++
Reference method: HPLC

Chromatography
$20000–$50000 per machine

$50–$100 per test







+
Meant for researchers,

≥500 µL, requires HPLC solvent and other preparation

≥65 min





+
Not portable

n/a

Controlled conditions






+
Can be used for different analyses or when the procurement of vials is difficult

Requires external power source

Requires routine maintenance

+
Depends on analyte

Quantitative

Exact concentration output on attached computer, chromatogram with quantified absorbance for vitamin A concentration


+++
Lab

Yes








+++

Notes: EIA, enzyme immunoassay; HPLC, high-performance liquid chromatography; NR, not reported; RBP, retinol-binding protein; RID, radial immunodiffusion.

+++ = best.

++ = acceptable.

+ = not acceptable.

Table 1b.

Catalog of all portable devices quantifying vitamin A and vitamin A deficiency: eyes.

Device (manufacturer)

Principle/method
Vitamin A biomarker Pricing (estimated, list price from manufacturer website) Technical requirements

Sample site; time for full charge
Portability

Included in kit
Power source

Usage duration per charge
Slit lamps:
Magnification

Dioptic range

Interpupillary range

Slit image width(s)

Filters
Target setting

Manufacturer support available

Global availability
BA 904, BA 904C (Haag-Streit, Harlow, Essex, UK) Haag-Streit, 1900

Slit lamp
Ocular morbidities Pricing not published








+
Meant for researchers and ophthalmologists

Anterior segment; 4–5 h




++
“Lightweight”

BA 904: head and chin rest stand, two energy packs, charger, power supply and large case;
BA 904C: two energy packs, charger, power supply, parking unit and small case

+++
Batteries and chargers

45 min






+++
10×, 16×

−8 to + 8

53–95 mm

NR

Blue, yellow

+++
Field, clinic

Yes

Yes





+++
Hand-held digital slit lamp (HSL-100, HSL-150) Portable slit lamp (Heine®) (Melo et al., 2004, Heine Optotechnik GmbH, 2018)

Slit lamp

Ocular morbidities Pricing not published








+
Meant for researchers and ophthalmologists

Anterior segment; NR




++
70 g

BETA4 SLIM NT rechargeable handle and NT4 table charger (included reducer insert), spare bulb, hard case



+++
Rechargeable or battery handle

NR






+++
10×, 16×

NR

NR

10 × 0.2 to 14 × 4 mm

Cobalt blue

+++
Field, clinic

Yes

Yes





+++
Portable slit lamp
(SL-17) (Kowa Ophthalmic Diagnostic Products, Torrance, CA, USA) (KOWA New Lighter)

Slit lamp
Ocular morbidities Pricing not published








+
Meant for researchers and ophthalmologists

Anterior segment; NR




++
<800 g; 220 × 95 × 220 mm

4 AAA batteries, dust cover, stand, instruction manual; optional: forehead rest, camera connection adapter




+++
4 AAA rechargeable or dry cell batteries

130–140 min






+++
10×, 16×

NR

50–72 mm

1 × 1, 0.15, 0.5, 0.8, 1.6, 12

Cobalt blue

+++
Field, clinic

Yes

Yes





+++
Binocular hand held biomicroscope slit lamp
(PSL One, PSL Classic) (Keeler, Malvern, PA
USA)

Slit lamp
Ocular morbidities Pricing not published











+
Meant for researchers and ophthalmologists

Anterior segment; 2.5 h







+++
900 g; 238 × 116 × 210 mm

Base charger unit, power supply, user instructions, lens cloth








+++
AC-powered

50 min










++
10×, 16×

−7 to + 7

50–72 mm

0.15 mm, 0.5 mm, 0.8 mm and 1.6 mm slits, 12 mm circle and a 1 mm square

Red free, blue, neutral density 0.8 and clear

+++
Field, clinic

Yes

Yes








+++
Handheld Slit Lamp S200 (Digital Eye Center, Miami, FL, USA) (Digital Eye Center, 2021, Digital Eye Center, 2021, Digital Eye Center, 2021)

Slit lamp

Ocular morbidities $2090










++
Meant for researchers and ophthalmologists

Anterior segment; NR





+++
40 g; NR

Universal smartphone adapter, metallic case, accessories.






+++
Rechargeable battery

7 h







++
10×, 16×

Diopter adjustment (not specified)

50–74 mm

Slit width adjustment (not specified)

Red free, green, cobalt blue, heat absorption, clear, neutral density

+++
Field, clinic

Yes

Yes






+++
Handheld Slit Lamp
S2 (Digital Eye Center, Miami, FL, USA) (Digital Eye Center, 2021, Digital Eye Center, 2021, Digital Eye Center, 2021)

Slit lamp

Ocular morbidities $1500










++
Meant for researchers and ophthalmologists

Anterior segment; NR





+++
750 g; 19 × 105 × 230 mm

Smartphone adapter, metallic case, accessories






+++
Rechargeable battery

2 h







++
10×, 16×

−5 to + 5

45–70 mm

0–10 mm

Heat-absorption, gray, red-free, cobalt blue

+++
Field, clinic

Yes

Yes






+++
Digital portable slit lamp Microclear Hyperion (Digital Eye Center, Miami, FL, USA) (Digital Eye Center, 2021, Digital Eye Center, 2021, Digital Eye Center, 2021)

Slit lamp

Ocular morbidities $3800










++
Meant for researchers and ophthalmologists

Anterior segment; NR





+++
600 g; NR

4″ touch screen, 16 GB internal memory, two lithium batteries (4 h each), software and manual





+++
Rechargeable battery

4 h







+++
10×

NR

NR

0–10 mm

Heat-absorption, gray, red-free, cobalt blue

+++
Field, clinic

Yes

Yes






+++
Hand held slit lamp (SL280) (Opticlar, Poole, Dorset) (Optical Visionmed)

Slit lamp

Ocular morbidities $3900











++
Meant for researchers and ophthalmologists

Anterior segment; 2 h






+++
880 g; 163 × 124 × 205 mm

Base plate, aluminum case









+++
Rechargeable battery

6 h








+++
10×, 16×

−7 to + 7

50–78 mm

0.15/0.5/0.8/1.6 mm. Circle 12 mm dia. 1 mm square

Green (red free), cobalt blue, neutral density 0.8, clear

+++
Field, clinic

Yes

Yes







+++
Portable slit lamp (PSL) (Reichert Technologies Inc., Depew, NY, USA) (Reichert Technologies, 2021)

Slit lamp

Ocular morbidities Pricing not published









+
Meant for researchers and ophthalmologists

NR






+
680 g; fits in palm of hand

Two batteries, battery charger








+++
Rechargeable batteries

2 h







+++
10×, 16×

−7 to + 7

50–70 mm

0–11 mm

Cobalt blue, red free, color temperature conversion

+++
Field, clinic

Yes

Yes






+++
Handy Slit Lamp XL-1 (Shin-Nippon by Rexxam Co., Ltd.) (Rexxam, 2021)

Slit lamp
Ocular morbidities Pricing not published








+
Meant for researchers and ophthalmologists

Anterior segment; NR




+++
700 g (195 × 105 × 230 mm)

Carrying case, one battery, battery charger, forehead support, diopter adjustment bar, instruction manual




+++
Rechargeable battery

2 h






+++
10×, 16×

−7 to + 7

50–70 mm

0–11 mm

Cobalt blue, green, conversion

+++
Field, clinic

Yes

Yes





+++
Portable slit lamp S150 (Medi-Works, Shanghai, China) (Mediworks, 2021)

Slit lamp, attachment for phone
Ocular morbidities Pricing not published








+
Meant for researchers and ophthalmologists

Anterior segment; 3.5 h




++
240 g; NR

NR







++
Rechargeable batteries

6 h






+++


NR

NR

0–12 mm

Cobalt blue

+++
Field, clinic

Yes

Yes





+++
SK-LS-1B portable slit lamp (Coburn Technologies, Inc. South Windsor, CT, USA) (Coburn Technologies Inc.)

Slit lamp

Ocular morbidities Pricing not published









+
Meant for researchers and ophthalmologists

Anterior segment; NR





+++
835 g; 320 × 310 × 205 mm

NR, optional iPhone adapter








+++
Rechargeable batteries

≥4 h







+++
10×, 16×

−7 to + 7

49–75 mm

0.1, 0.2, 0.8, 1, 5, 12 mm

Neutral density, red-free, cobalt blue

+++
Field, clinic

Yes

Yes






+++
Device (manufacturer)

Principle/method
Vitamin A biomarker Pricing (estimated, list price from manufacturer website) Technical requirements

Sample site; time for full charge
Portability

Included in kit
Power source

Usage duration per charge
Dark adaptometers and other devices: Other attributes Target setting

Manufacturer support available

Global availability
RetEval (LKC Technologies, Gaithersburg, MD, USA) (Technologies, 2019)

Non-mydriatic flash and flicker ERG/VEP device
Ocular morbidities Pricing not published






+
Meant for researchers and ophthalmologists

Pupils, retina; 4 h full recharge time

+++
240 g; 7 × 10 × 23 cm

Storage case, charging stand





+++
Battery-powered

8 h





+++
Pupil measurements: 1.3–9 mm, <0.1 mm resolution, 28.3 Hz

USB connectivity




+++
Field, clinic

Yes

Yes



+++
Mobile eye testing unit (Agrawal and Sahu, 2020)

Vision drum, trial box, retinoscope, slit-lamp bio-microscope, applanation tonometer, and non-mydriatic fundus camera
Ocular morbidities: conjunctival xerosis with Bitot’s spot (X1B) or keratomalacia (X3B) (World Health, 2014) Pricing not published






+
Meant for researchers, optometrists, ophthalmologists

Various parts of eye

+++
Indicated to be mobile

NR





++
Varies by device in kit






++
See (Agrawal and Sahu, 2020)







+++
Field

N/A; not commercially available

No/custom

++
Scotopic Sensitivity Tester-1 (LKC Technologies, Gaithersburg, MD, USA) (Congdon et al., 1995, Congdon et al., 2000, Sanchez et al., 1997)

Dark adaptometer
Visual score/threshold
Pupillary score/threshold
Pupillary dynamics
Rod function
Pricing not published







+
Meant for researchers and ophthalmologists

Retina; binocular partial bleaching with camera flash (3433 cd-s/m2)

+++
Hand-held

N/A






+++
NR

NR






+
Yellow-green LED light with wavelength at 572 nm, 12 intensity settings, calibrated with EG + G DR 2550 digital radiometer-photometer





+++
Field, clinic

Appears discontinued

No



+
Portable field dark adaptometer (custom)
(Labrique et al., 2015, Palmer et al., 2015, Palmer et al., 2016)

Dark adaptometer
Visual score/threshold
Pupillary score/threshold
Pupillary dynamics
Pricing not published









+
Meant for researchers and ophthalmologists

Retina; binocular partial bleaching with camera flash (>3400 cd-s/m2)



+++
Portable: “Its size and weight allowed it to be carried long distances to areas unreachable by car.”

Digital camera, a retinal bleaching flash, and a Ganzfeld light source inside a pair of light-obscuring goggles

+++
Laptop-powered

10 tests per day








++
See (Labrique et al., 2015)

Assess impaired pupillary responses to a graded series of Ganzfeld light stimuli applied within a pair of “darkroom” goggles with an embedded microcircuit design and regulated by a laptop-powered controller box



++
Field

N/A; not commercially available

No/custom?




+
Emtech A meter V.01 (custom)
(Mehta, 2018, Mehta, 2019b, Mehta, 2019a, Mehta and Mehta, 2018, Banerjee, 2019)

Dark adaptometer
Dark adaptation; identify pictorial representation of objects at low light intensity Pricing not published







+
Meant for researchers and ophthalmologists

Retina; NR




+++
Handheld

Electronic paper module, LCD to display test object, microSD card, keypad




+++
NR

NR






+
Results output to microSD card








+++
Field

N/A; not commercially available

No/custom?


+
Custom-built portable field dark adaptometer
(Steven and Wald, 1941, Wald, 1941)

Dark adaptometer

Visual threshold
Dark adaptation
Pricing not published






+
Meant for researchers and ophthalmologists

Retina; NR



+++
8.4 kg; 21.6 × 21.6 × 29.2 cm; “approximate size and shape of a pocket lamp”

Eyepiece, test unit, cord, cabinet


+++
Three 2-volt discharge storage cells

NR



+++
Results in log units







++
Field, lab

N/A; not commercially available

No/custom?

+
Reference method: Goldmann-Weekers dark adaptometer (Haag-Streit) Visual threshold
Dark adaptation
Not available for purchase















+
Meant for researchers and ophthalmologists; some models require conversion table for a calibration error (Maggiano et al., 1978)

Retina, with pupils dilated; NR

Requires 60–120 min in a dark room

+
Large size and complex; not portable

N/A













+
Requires external power source

N/A













+
Results: luminance in units of log microapostilbs, which requires conversion to the more contemporary unit of luminance, cd/m2












+
Clinic

No

Out of production and not available to order










+

Notes: NR, not reported.

+++ = best.

++ = acceptable.

+ = not acceptable.

Vitamin A deficiency biomarkers

In Table 2, we list definitions of VAD used across studies for a variety of biological sample types, from humans or cattle, including cows, calves, and bulls (Table 2). We also note which studies used particular definitions (e.g., VAD measured as RBP ≤ 0.70 µmol/L was measured by Hix et al. 2004). A previous review by Tanumihardjo (2016) has outlined the utility of biomarkers for vitamin A nutrition status (Tanumihardjo et al., 2016), which we adapted for Table 2. We outline the biomarkers of vitamin A as identified in our literature search below.

Table 2.

Definitions of vitamin A deficiency, by sample type and device.

Blood (whole, serum, plasma)a
Biomarker Typeb Device (studies using) Deficiency or insufficiency definitions usedc
Retinol
Status iCheck Fluoro (BioAnalyt)d (Boateng et al., 2018, Elom et al., 2015, Ghaffari et al., 2019, Raila et al., 2017, Schweigert et al., 2011a, Bechir et al., 2012, Crump et al., 2017, Schweigert et al., 2011b, Whang et al., 2012, Zambo et al., 2012)


Spectrophotometer model 450 (Sequoia-Turner) (Marinovic et al., 1997)

CRAFTI (Craft Technologies) (Chaimongkol et al., 2011)
Severe/clinical deficiency: ≤0.35 µmol/L (10 µg/dL) (WHO, 2011)

Low/subclinical deficiency: ≤0.70 µmol/L (20 µg/dL) (WHO, 2011)

Insufficiency: ≤1.05 µmol/L (30 µg/dL) (de Pee and Dary, 2002)
RBP Status Custom REI (Hix et al., 2004)

EE-µPAD (Lee et al., 2016)

Tidbit (Lu and Erickson, 2017, Lu et al., 2017) ± HYPER filtration (Lu et al., 2018)

Custom Ag-Ab reaction (Ciaiolo et al., 2015)
Deficiency: ≤0.70 µmol/L (Hix et al., 2004)

Deficiency: <16.3 µg/mLe (Lee et al., 2016)

Deficiency: <14.7 µg/mL (correlated with retinol ≤ 0.70 µmol/L) (Lu et al., 2017)

Not defined (Ciaiolo et al., 2015)
Beta-carotene Not defined (indicator of recent dietary intake) iCheck Carotene (BioAnalyt)f (Ghaffari et al., 2019, Hye et al., 2020, Klein et al., 2013, Livingston et al., 2020, Meinke et al., 2016, Raila et al., 2012, Madureira et al., 2020)
Humans:
No official cut-off defined (von Lintig, 2020)

Cattle (Klein et al., 2013, De Ondarza and al., 2009, Schweigert and Immig, 2007):
Deficient: 0.6–1.5 mg/L or < 1.5 mg/L
Marginal: ≥1.5 mg/L to < 3.5 mg/L
Optimal: ≥3.5 mg/L
Milk
Biomarker Typeb Device Deficiency or insufficiency definitions usedc
Retinol Status, exposure iCheck Fluoro (BioAnalyt)d (Jans et al., 2018, Abebe et al., 2019, Engle-Stone et al., 2014, Schweigert et al., 2011a, Schweigert et al., 2011b, Bechir et al., 2012, Crump et al., 2017) Humans:
Inadequate: <1.05 µmol/L (Blaner, 2020)
or milk fat < 8 µg/g (Blaner, 2020)

Cattle: not defined
Beta carotene Not defined (indicator of recent dietary intake) iCheck Carotene (BioAnalyt)d (no studies) Humans: not defined

Cattle: not defined
Eyes
Biomarker Typeb Device (studies using) Deficiency or insufficiency definitions usedc
Visual score/threshold Function Scotopic sensitivity hand-held illuminator (LKC Technologies, Inc.) (Congdon et al., 1995, Sanchez et al., 1997, Reilly et al., 2006)

EmTech A meter V.01g (Mehta, 2018)

Portable visual adaptometer (Wald, 1941, Steven and Wald, 1941)

Scotopic Sensitivity Tester-1TM (SST-1) (Peters et al., 2000)
Abnormal:
≥stimulus #10 (Congdon et al., 1995)
≥-3.76 log cd/m2 h (Congdon et al., 1995)

Highly abnormal:
≥stimulus #11
≥-3.39 log cd/m2 (Congdon et al., 1995)

A decrease of ≥ 0.3 log units after administration of vitamin A supplementation (Wald, 1941, Steven and Wald, 1941)
Dark adaptation: pupillary score/responsiveness [lowest light intensity that stimulated percentage relative change in pupil diameter (Labrique et al., 2015)] Function Scotopic sensitivity hand-held illuminator (LKC Technologies, Inc.) (Congdon et al., 1995, Sanchez et al., 1997, Peters et al., 2000)

Portable field dark adaptometer (PFDA) or digital pupillometer (Labrique et al., 2015, Palmer et al., 2015, Palmer et al., 2016)
Normal:
≥-1.24 log cd/m2 (Congdon and West, 2002)

Abnormal:
≥stimulus #9 (Congdon et al., 1995)
≥-0.575 log cd/m2 (Congdon et al., 1995) i.e.,≥20% (Labrique et al., 2015)
≥-1.11 log cd/m2 (Congdon et al., 2000)
≥-0.9 log cd/m2 (Palmer et al., 2016)
≥15% relative change in diameter (Labrique et al., 2015)
≥10% contraction in pupil size (Palmer et al., 2016)

Threshold:
≥15 cd/m2 (Khan et al., 2019)
Pupillary dynamics [i.e., response time: absolute value of difference in frame numbers from pre- to post-stimulus divided by number of frames per second (Labrique et al., 2015)] Function Portable field dark adaptometer (PFDA) or digital pupillometer (Labrique et al., 2015, Palmer et al., 2015, Palmer et al., 2016) No official cut-off defined
Rod function [dark-adapted rod full-field electro-retinogram responses (Peters et al., 2000)] Function Scotopic Sensitivity Tester-1TM (SST-1) (Peters et al., 2000) No official cut-off defined
Ocular morbidities



Function Mobile eye unit (comprised of vision drum, trial box, retinoscope, slit-lamp bio-microscope, applanation tonometer, non-mydriatic fundus camera) (Agrawal and Sahu, 2020)

Heine HSL-100 biomicroscope equipped with portable slit lamp (Melo et al., 2004)
Night blindness, conjunctival xerosis with Bitot’s spots (X1B), keratomalacia (X3B), ocular lesions (Agrawal and Sahu, 2020) (stages as designated by World Health Organization grading system (World Health, 2014)

Ocular lesions (Melo et al., 2004)

Notes:Ag-Ab, antigenantibody; EE-µPAD, electronics enabled microfluidic paper-based analytical device; HYPER, High-yield paper-based quantitative blood separation system; RBP, retinol-binding protein; REI, rapid enzyme immunoassay.

h cd/m2 is the SI unit of luminance (Congdon et al., 2000).

a

Adapted from reference (Tanumihardjo et al., 2016).

b

Whereas serum or plasma is required to measure circulating vitamin A, some devices can use whole blood as the sample input.

c

Defined by global standards (e.g., World Health Organization) or by study authors.

d

An earlier version of this device is referenced as iCheck Ret 435–1 (Bechir et al., 2012).

e

Correlated with retinol ≤ 0.70 µmol/L, when sandwich ELISA is used for RBP measurement (Erhardt et al., 2004).

f

An earlier version of this device is referenced as iCheck Ret 515–2 (Bechir et al., 2012).

g

Device also referenced as “dark adaptometer” (Banerjee, 2019) or “In-Direct method and system for Vitamin A deficiency detection” (Mehta, 2019b, Mehta, 2019a, Mehta and Mehta, 2018).

Vitamin A liver concentration (µmol vitamin A/g liver) is the gold standard for vitamin A status but requires invasive techniques such as biopsy to be measured (Tanumihardjo et al., 2016). Sampling blood enables the quantification of serum retinol, serum retinol-binding protein (RBP), or provitamin A in the form of beta-carotene; however, each measure has trade-offs. Serum retinol reflects liver stores only at extremes of deficiency (≤0.07 µmol/g liver) or elevation (>1.05 µmol/g liver) (WHO, 2011), because serum retinol is homeostatically regulated by the body. The World Health Organization defines VAD as serum retinol ≤ 0.70 µmol/L (WHO, 2011). RBP is commonly assumed to have a 1:1 ratio with serum retinol, and therefore the same cut-offs are sometimes used for both retinol and RBP. However, this ratio can be affected by the extent of VAD, zinc deficiency, acute phase response, protein-energy malnutrition, liver disease, acutely stressful situations, high fever, antibiotic use, or obesity (Tanumihardjo et al., 2016, de Pee and Dary, 2002). Therefore, previous studies have proposed other deficiency cut-offs, such as 0.69 µmol/L (Semba et al., 2002) or 0.83 µmol/L (Engle-Stone et al., 2011, Gorstein et al., 2008). Recently, the Global Alliance for Vitamin A has recommended analysis of a subsample by HPLC to confirm the cut-off point for VAD; furthermore, given the acute phase response, inflammation markers such as C-reactive protein and alpha-1-acid-glycoprotein must also be measured (Global Alliance for Vitamin A, 2019).

Beta carotene is one of several dietary provitamin A carotenoids, a plant-derived form of vitamin A. The body converts dietary provitamin A carotenoids into retinol with the following conversion factors: 1 µg retinol activity equivalent (RAE) equals 1 retinol equivalent (RE), 1 µg retinol, 2 µg β-carotene in oil, 12 µg β-carotene in mixed foods, or 24 (12–26) µg other provitamin A carotenoids in mixed foods (Institute of Medicine, 2001, Combs and McClung, 2017, Blaner, 2020). The conversion efficiency ratio of beta carotene to RAE is still debated. For example, the European Food Safety Authority suggests that the conversion is 6:1 rather than 12:1 (EFSA Panel on Dietetic Products, Nutrition Allergies, 2015). Carotenoids can be measured in blood, milk, or skin, and several studies have found a positive association of skin carotenoid concentrations with serum or plasma carotenoid status (Zidichouski et al., 2009, Aguilar et al., 2014, Morgan et al., 2019, Hayashi et al., 2020). However, a consensus has not been reached regarding a conversion factor or how the measurements equate to vitamin A status (von Lintig, 2020). Because carotenoids tend to reflect recent dietary intake rather than long-term status, recommended serum carotenoid deficiency cut-offs have not been established in humans (von Lintig, 2020). Deficiency in β-carotene in cow’s blood has been defined as 0.6–1.5 mg/L (Klein et al., 2013, De Ondarza and Al, 2009, Schweigert and Immig, 2007).

In breast milk, retinol may be measured to estimate both the maternal vitamin A status and intake, and the infant intake of vitamin A (Engle-Stone et al., 2014, Tanumihardjo et al., 2016). Additionally, breast milk retinol measurement is influenced by the stage of lactation, time of day, “fullness” of the breast, feeding status if milk from both breasts is analyzed, and whether the milk is hindmilk compared with foremilk (Tanumihardjo et al., 2016). VAD is defined as a milk retinol concentration ≤ 1.05 µmol/L, or ≤ 8 µg/g milk fat (Tanumihardjo et al., 2016). In cows, milk β-carotene levels are often measured and linked to bovine fertility and health.

Because of vitamin A’s role in in producing rhodopsin, the visual pigment of rods in the eyes, VAD can cause ocular manifestations resulting in poor vision (World Health, 2014). These include night blindness, conjunctival xerosis, Bitot’s spots, corneal xerosis, and keratomalacia. Impaired adaption to the dark is among the first symptoms of VAD, and it can be used as a screening tool (World Health, 2014). Tests such as pupillary and visual thresholds can assess dark adaptation by determining the lowest-intensity level of light required to cause pupillary dilation or to visualize an image (World Health, 2014, Labrique et al., 2015).

Comparison studies

From 3230 studies (after de-duplication), we identified 18 studies (19 reports) comparing nine portable methods/devices (index 1) to a reference standard method (Fig. 1); we were unable to retrieve an additional two reports (Craft, 2005, Fujita, 2007). No studies compared two portable methods (i.e., index 1 vs. index 2). Thirteen studies (15 reports) measured human or cattle blood samples (BioAnalyt, 2020, Chaimongkol et al., 2011, Ciaiolo et al., 2015, Elom et al., 2015, Ghaffari et al., 2019, Hix et al., 2004, Hix et al., 2006, Lee et al., 2016, Lu et al., 2017, Lu et al., 2018, Raila et al., 2012, Raila et al., 2017, Schweigert et al., 2011a, Chaimongkol et al., 2008, Lu and Erickson, 2017); four studies measured human or cattle milk samples (Schweigert et al., 2011b, Schweigert et al., 2011a, Engle-Stone et al., 2014, Abebe et al., 2019); and one study measured eye function (Peters et al., 2000). Study details are listed in Table 3. We also re-analyzed the data presented in supplemental Tables S1 and S2 in one publication to calculate the descriptive statistics for plasma and whole blood retinol in samples analyzed by HPLC and iCheck Fluoro (Raila et al., 2017).

Table 3.

Description of included studies comparing a portable method against a reference standard method.

Author
Year
Device Manufacturer Sample tested

Biomarker
Study population Test location (field/laboratory, country) Reference method Ref.
Chaimongkol 2011 CRAFTi Eurofin Craft Technologies Serum

Retinol
Study cohorts Thailand HPLC (Chaimongkol et al., 2011)
Chaimongkol 2008a CRAFTi Eurofin Craft Technologies Serum

Retinol
Study cohorts Thailand HPLC (Chaimongkol et al., 2008)
Ciaiolo 2015 Custom Ab-Ag reaction Custom Serum,

RBP
Patients Italy Nephelometry (Ciaiolo et al., 2015)
Lee 2016 EE-µPAD Custom Serum

RBP
Commercial (ProMedDx) USA ELISA (Lee et al., 2016)
BioAnalyt report (year: NR) iCheck Carotene BioAnalyt Plasma Dairy cows and calves NR HPLC (BioAnalyt, NR)
Raila 2012 iCheck Carotene BioAnalyt Whole blood or plasma

Beta-carotene
Holstein-Friesian cows, local farm Germany, Ireland, France HPLC (Raila et al., 2012)
Ghaffari 2019 iCheck Carotene BioAnalyt Plasma

Beta-carotene
Holstein cows and calves from institutional farms Germany HPLC (Ghaffari et al., 2019)
iCheck Fluoro BioAnalyt Whole blood

Retinol
Raila 2017 iCheck Fluoro BioAnalyt Whole blood, serum

Retinol
Dairy cows and bulls, institutional farms Germany, Japan HPLC (Raila et al., 2017)
Schweigert 2011a iCheck Fluoro BioAnalyt Milk

Retinol
Study cohorts and local cows Germany HPLC (Schweigert et al., 2011b)
Schweigert 2011ba iCheck Fluoro BioAnalyt Plasma or milk

Retinol
Study cohorts Low resource setting HPLC (Schweigert et al., 2011a)
Abebe 2019 ICheck Fluoro BioAnalyt Milk

Retinol
Study cohorts Ethiopia HPLC (Abebe et al., 2019)
Elom 2015 iCheck Fluoro BioAnalyt Serum

Retinol
Study cohorts Morocco HPLC (Elom et al., 2015)
Engle-Stone 2014 iCheck Fluoro BioAnalyt Milk

Retinol
Study cohorts Cameroon HPLC (Engle-Stone et al., 2014)
Hix 2004 RBP-EIA Scimedx Corp Serum

RBP
Study cohorts, commercial Papua New Guinea, Nicaragua HPLC (Hix et al., 2004)
RID The Binding Site Study cohorts Nicaragua HPLC
Hix 2006 RBP-EIA Scimedx Corp Serum

RBP
Study cohorts Cambodia HPLC (Hix et al., 2006)
Peters 2000 SST-1 LKC Technologies Eyes

Dark-adapted final thresholds; rod function
Patients of Retina Foundation of the Southwest USA Goldmann-Weekers Dark Adaptometer (Peters et al., 2000)
Lu 2017 Tidbit, ± HYPER filtration system Custom Serum

RBP
Commercial (Research Blood Components LLC) USA ELISA (Lu and Erickson, 2017, Lu et al., 2017, Lu et al., 2018)

Notes: Ag-Ab, antigenantibody; EE-µPAD, electronics enabled microfluidic paper-based analytical device; EIA, enzyme immunoassay; HYPER, high-yield paper-based quantitative blood separation system; RBP, retinol-binding protein; RID, radial immunodiffusion assay; SST-1, Scotopic Sensitivity Tester-1.

a Meeting abstract, therefore some details are not reported.

b RID may be considered a second index test, because it is not a reference standard; however, in the study, only the first index test, RBP-EIA was the assay undergoing development and validation.

We also identified many studies that used a portable device for assaying samples but did not compare the results to those of a reference method and instead cited previous validation studies. Although the devices used are catalogued and described (Table 1a, Table 1b), these studies are not further detailed in this review.

Study populations were mostly from the US and Germany, in addition to Thailand, Italy, France Ireland, Japan, Ethiopia, Morocco, Cameroon, Papua New Guinea, Nicaragua, Cambodia, and Oman. Portable fluorometers, photometers, enzyme-based assays or immunoassays, microfluidics-based approaches, and a dark adaptometer for eye function were assessed and compared with their respective reference standards.

Table 4 compares the stated performance criteria described by the device manufacturers’ websites to reporting from individual studies using the devices, according to the WHO Affordable, Sensitive, Specific, User-friendly, Rapid and robust, Equipment-free and Deliverable to end-users (ASSURED) criteria for diagnostic tests in resource-limited settings (Kosack et al., 2017). Only the iCheck Fluoro and iCheck Carotene stated performance criteria on the BioAnalyt website and are included in this table. No studies described the cost of the devices. Most devices did not have published cost information, aside from some of the slit lamps showing list prices. Other devices described by the studies were developed as proofs of concept and are either not on the market or are on the market, but lacking performance criteria on the manufacturer’s website.

Table 4.

graphic file with name fx1a.gif
graphic file with name fx1b.gif

We note that the study by Ghaffari et al. (2019) reported both measurements of retinol in whole blood and beta carotene in plasma, but only directly references the iCheck Fluoro device and not the iCheck Carotene (Ghaffari et al., 2019). Whereas the iCheck Carotene requires 0.4 mL of sample, the iCheck Fluoro requires 0.5 mL; the authors stated using only 0.5 mL of sample.

Additionally, the manufacturer (BioAnalyt) lists only “colostrum, cattle whole blood, and serum” as appropriate sample types for the iCheck Carotene; however all studies using the iCheck Carotene, including a BioAnalyt report, also analyzed plasma for beta carotene content (BioAnalyt, 2020, Ghaffari et al., 2019, Raila et al., 2012).

A major gap across all devices is the lack of reporting on sensitivity and specificity compared with a reference standard method.

Most studies compared a portable device to a reference method. Table 5a, Table 5b, Table 5c, Table 5d, Table 5e, Table 5f, Table 5g show the performance of these devices against their reference standards for measuring vitamin A and VAD. Additional analyses conducted with other index (e.g., index 2) tests are described in the text.

Table 5a.

Portable fluorometers: device performance in human blood samples.

Portable device vs. reference iCheck Fluoro vs. HPLC iCheck Fluoro vs. HPLC CRAFTi vs. HPLC CRAFTi vs. HPLC
Vitamin A biomarker Retinola Vitamin Aa Retinola Retinola
Sample type Plasma Serum Serum Serum
Study population 89 children 56 samples 38 women 75 women, 143 children
Concentration difference MD: 0 min: 1.9 µg/L ± 23.2
MD: 15 min: −8.0 µg/L + 22.7
NR MD: −0.07 MD: −0.07
Correlation coefficient 0 min: 0.98
15 min: 0.98
NR NR 0.77
R2 NR >0.95 NR NR
Regression equation NR NR Slope = 0.81 NR
Operational range NR NR NR 0.5–1.5 µmol/L
VAD or VAI (%), index vs. ref Not defined: N = 2/89 vs. NR NR NR <0.7 µmol/L: 9.2% vs. 2.8%
<1.05 µmol/L: 49.5% vs. 43.6%
Precision
Sensitivity NR NR NR VAD ≤ 0.7 µmol/L: 66.7%
VAD ≤ 1.05 µmol/L: 85.3%
Specificity NR NR NR VAD ≤ 0.7 µmol/L: 92.4%
VAD ≤ 1.05 µmol/L: 78.0%
Intra-assay %CV NR 2.5–6.4 %b Agreement noted but not quantified 3.97% vs. 3.45%
Inter-assay %CV NR NR
Inter-observer %CV NR NR
Bland Altman analysis comments No commentary. At 0 min, 3 values fell outside 2 SDs. At 15 min, 4 values fell outside 2 SDs NR No systematic bias No systematic bias; most values within ± 0.5 with normally distributed serum retinol values
Reference (Elom et al., 2015) (Schweigert et al., 2011a)c (Chaimongkol et al., 2008)c (Chaimongkol et al., 2011)

Notes:MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); SD, standard deviation; VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

a

Units: µg/L or µmol/L.

b

Specific %CVs not distinguished.

c

Study abstract, lacking some details.

Table 5b.

Portable fluorometers: device performance in human and bovine milk samples.

Human milk
Bovine milk

Portable device vs. reference
iCheck Fluoro vs. HPLC iCheck Fluoro vs. HPLC iCheck Fluoro vs. HPLC iCheck Fluoro vs. HPLC
Vitamin A biomarker Retinola, milk fatb Retinola, milk fatb Retinola Retinola
Study population 104 women 75 women, 154 samples 1 woman, 16 samples 21 cows
Concentration difference MD: 0.01 µmol/L, 0.03 µg/g fat MD: −0.83 ± 0.14 µmol/L,
−5.6 ± 0.7 µg/g fat
Expressed milk, MD: 103% ± 13 Expressed milk, MD: 105% ± 9
Powdered milk (n = 5), MD: 144% ± 15
Liquid whole milk (n = 5), MD: 118% ± 13
Liquid skim milk (n = 4), MD: 95% ± 10
Correlation coefficient 0.57unadj, 0.59adjc 0.85unadj, 0.79adjc NR NR
R2 0.32unadj, 0.35adjc 0.72unadj, 0.62adjc NR NR
Regression equation NR NR NR NR
Operational range 50–3000 µg RE/L 50–3000 µg RE/L 50–3000 µg RE/Ld NR
VAD or VAI (%), index vs. ref <1.05 µmol/L: 87% vs. 76%
<8 µg/g fat%: 89% vs. 81%
<1.05 µmol/L: 3.9% vs. 2.60%
<8 µg/g fat %: 0% vs. 2%
NR NR
Precision
Sensitivity NR Too few VAD cases to examine NR NR
Specificity NR Too few VAD cases to examine NR NR
Intra-assay %CV 1.1% vs. 1.5–1.6% 0.6 %e NR NR
Inter-assay %CV NR NR NR
Inter-observer %CV NR NR NR
Bland Altman analysis comments Used to present mean difference between measurements; mean difference not significantly different from zero Plotted but no conclusion drawn; appears to show 8 values outside of 2 SDs (µmol/L retinol) and 8 values outside of 2 SDs (µg/g fat) NR NR
Reference (Abebe et al., 2019) (Engle-Stone et al., 2014) (Schweigert et al., 2011b) (Schweigert et al., 2011b)

Notes:MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

a

Units: µmol/L or µg RE/L.

b

Units: µg/g fat%.

c

Adjusted for breast milk fat content.

d

Not reported but based on previous studies using same device.

e

Specific %CVs not distinguished.

Table 5c.

Portable immunoassays: device performance in human blood.

Portable device vs. reference RBP-REI vs. HPLCa RBP-REI vs. HPLCa RID vs. HPLCa Immunoassay vs. nephelometry
Vitamin A biomarker RBPb RBPb RBPb RBPb RBPb
Sample type Serum Serum Serum Serum Serum
Study population 24 children 70 mothers and children 359 children 40 mothers and children 2 healthy adults
(Serum A, B)
Concentration difference MD: 0.22 µmol/L NR NR NR Index (dilution):
Serum A:
(1:10): +
(1:100): -
(1:1000): -
(1:10000): -
Serum B:

(1:10): ++
(1:100): -
(1:1000): -
(1:10000): -

Ref:
Serum A: 46 mg/L
Ref:
Serum B: 42 mg/L
Correlation coefficient 0.93 0.91 0.89 0.84 NR
R2 0.86 0.82 0.79 0.71 NR
Regression equation y = 0.95x + 0.36 y = 0.62x + 0.32 y = 0.65x + 0.27 NR NR
Operational range 10–40 µg RBP/mL 10–40 µg RBP/mL 10–40 µg RBP/mLc NR Immune precipitates:
Neg: -
2 mg/L: +
10 mg/L: ++
100 mg/L: +++
1000–10000 mg/L: ++++
VAD or VAI (%), index vs. ref ≤0.70 µmol/L: 32% vs. 36% NR <0.35 µmol/L: 0.6% vs. 2.2%
≤0.70 µmol/L: 20.9% vs. 22.3%
NR NR
Precision
Sensitivity NR NR 70% NR “Good” (“can detect presence of [RBP] at concentration of few µg/mL”)
Specificity NR NR 93.2% NR NR
Intra-assay %CV 6.7 %d NR NR NR NR
Inter-assay %CV 8.9 %d NR NR NR NR
Inter-observer %CV 13.0 %d NR NR NR NR
Bland Altman analysis comments NR NR NR NR NR
Reference (Hix et al., 2004) (Hix et al., 2006) (Hix et al., 2004) (Ciaiolo et al., 2015)

Notes: MD, mean difference; NR, not reported; RBP, retinol binding protein; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of Medicine, 2001); REI, rapid enzyme immunoassay; VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

a

Reference analyte is retinol.

b

Units: µmol/L or µg RE/L.

c

Not reported but based on previous studies using same device.

d

Reported from separate analysis among unknown total # of samples (“5 adult volunteers and a commercially available source”) analyzing device performance, without reference to HPLC.

Table 5d.

Portable microfluidics-based methods: device performance in human blood.

Portable device vs. reference EE-µPAD, vs. ELISA Tidbit with HYPER platform, vs. ELISAa Tidbit without HYPER platform, vs. ELISA
Vitamin A biomarkerf RBPb RBPb RBPb
Sample type Whole blood Whole blood Serum
Study population 95 adults (commercial) 12 adults 43 adults (commercial)
Concentration difference NR NR NR
Correlation coefficient NR NR 0.75
R2 (index, unless specified) NR Index: 0.81 vs. ref: >0.99 0.56
Regression equation NR Slope = 0.99 Slope = 0.97
RMSE, index vs. ref NR 3.75 vs. 1.3 µg/mL 4.34 µg/mL vs. NR
Operational range ∼10–70 µg/mL (graph) ∼5–20 µg/mL (graph) 2.2–20 µg/mL (0.10–0.95 µmol/L)
VAD or VAI (%), index vs. ref <16.3 µg/mL: AUC = 0.7139 vs. 17.2% NR <14.7 µg/mL (≤0.70 µmol/L): NR vs. 9.3%
Precision
Sensitivity 75% at MFR cutoff, 0.831 NR 100%
Specificity 62.3% at MFR cutoff, 0.831 NR 100%
Intra-assay %CV 10.8% vs. 3.9% 20.3% deviation per test strip, recommend taking average of 3 test strips NR
Inter-assay %CV NR NR NR
Inter-observer %CV NR NR NR
Bland Altman analysis comments NR NR Bias at −0.05 µg/mL (-2.3 nmol/L)
Reference (Lee et al., 2016) (Lu and Erickson, 2017, Lu et al., 2017, Lu et al., 2018) (Lu and Erickson, 2017, Lu et al., 2017)

Notes:MD, mean difference; MFR, multi-faceted ratio i.e., the ratio of the light transmission in the test area to that in the background control area, calculated for RBP for each sample repeat. NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); RMSE, root mean squared error; VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

a

Reference ELISA utilized samples that were filtered using HYPER system.

b

Units: µg/mL, mg/L, or µmol/L.

Table 5e.

Other portable devices: device performance in for assessing eye function (vision).

Eyes
Index 1avs. referenceb,c
Validation of portable device SST-1 vs. Goldmann-Weekers dark adaptometera
Vitamin A biomarker Dark adaptation final thresholdb
Study population 87 patientsc and 24 healthy children and adults
Concentration difference NR
Correlation coefficient 0.88 (adjusted for ceiling effect)
R2 (index, unless specified) 0.77
Regression equation “intercept close to zero”
Operational range 0–30 dB stimulus intensity range (0–3 log units)
VAD or VAI (%), index vs. ref Elevated final thresholds: 75% vs. 82%
Precision
Sensitivity Final threshold elevated: 74.7%
Specificity NR
Intra-assay %CV NR
Inter-assay %CV NR
Inter-observer %CV NR
Bland Altman analysis comments NR
Reference (Peters et al., 2000)

Notes: MD, mean difference; MFR, multi-faceted ratio i.e., the ratio of the light transmission in the test area to that in the background control area, calculated for RBP for each sample repeat. NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); SST-1, Scotopic Sensitivity Tester-1; VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

a

Reference analyte is dark adaptation final threshold.

b

Units: log units.

c

Patients had retinal degeneration with mild to severe loss of rod function from full-field ERG results.

Table 5f.

Portable fluorometers: device performance in bovine blood samples.

Portable device vs. reference iCheck Fluoro vs. HPLCa iCheck Fluoro vs. HPLCa iCheck Fluoro vs. HPLCa iCheck Fluoro vs. HPLCa iCheck Fluoro vs. HPLCb
Vitamin A biomarker Vitamin Ac Vitamin Ac Retinold Vitamin Ac Vitamin Ac Retinold Retinold Retinold Retinold
Sample type Whole blood Whole blood Whole blood Plasma Plasma Plasma Plasma Serum Serum
Study population 28 cows 11 calves 10 cows 28 cows 11 calves 40 cows 92 bulls 29 cows 32 black cattle
Concentration difference Range: 184–336 (see note)e MD: −0.013 ± -0.020 MD: 19.3 MD: 26.5 MD: 0.01 MD: 0.00
Correlation coefficient 0.78 0.90 0.92 0.88 0.96 0.94 0.93
R2 0.61 0.81 0.84 0.77 0.92 0.88 g0.87
0.88 0.94
Regression equation y = 0.77 + 11.26 NR y = 1.18x − 72.64 y = 0.80x + 1.32 NR NR
y = 1.03–30.11
Operational range NR NR NR NR NR
VAD or VAI (%) NR NR NR NR NR NR NR NR NR
Precision
Sensitivity NR NR States test is sensitive and specific but does not quantify these NR NR States test is sensitive and specific but does not quantify these States test is sensitive and specific but does not quantify these States test is sensitive and specific but does not quantify these States test is sensitive and specific but does not quantify these
Specificity NR NR NR NR
Intra-assay %CV NR NR NR NR NR 2.3% vs. 5.3 %f 2.1% vs. 3.3 %f
Inter-assay %CV NR NR NR NR NR
Inter-observer %CV NR NR NR NR NR NR NR NR NR
Bland-Altman analysis comments Good level of agreement and no systematic error. 5% of total values fell outside 95% acceptability limits Good level of agreement and no systematic error; 4% of total values fell outside the 95% acceptability limits Good level of agreement and no systematic error. 1 value fell outside of 95% acceptability limits Good level of agreement and no systematic error. 1 value fell outside of 95% acceptability limits Good level of agreement and no systematic error. 4% of values fell outside the 95% acceptability limits
Reference (Ghaffari et al., 2019) (Raila et al., 2017) (Ghaffari et al., 2019) (Raila et al., 2017)

Notes:MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

a

Reference sample is in plasma.

b

Reference sample is serum.

c

Units: µg RE/L.

d

Units: µmol/L.

e

Reported value from this study appears to be a repeated value for cow whole blood beta carotene content given as 2.09–8.15 mg/L, instead of the calf whole blood vitamin A reported in µg RE/L.

f

Values appear to be an average of intra- and inter-assay %CV.

Table 5g.

Portable photometers: device performance in bovine blood samples.

Portable device vs. reference iCheck Carotene vs. HPLCa iCheck Carotene vs. HPLCa iCheck Carotene vs. HPLCa iCheck Carotene vs. HPLCa iCheck Carotene vs. HPLCa iCheck Carotene vs. HPLCa iCheck Carotene vs. HPLCa
Vitamin A biomarker β-caroteneb β-caroteneb β-caroteneb β-caroteneb β-caroteneb β-caroteneb β-caroteneb
Sample type Whole blood Whole blood Whole blood Plasma Plasma Plasma Plasma
Study population 28 cows 11 calves 23 cows 28 cows 11 calves NR, cows and calves 166 cows
Concentration difference NR NR MD: 0.21 MD: −0.29 MD: 0.02 NR MD: 0.26
Correlation coefficient 0.98 0.98 0.99 0.97 0.98 NR 0.99
R2 0.97 0.96 0.99 0.93 0.96c 0.97 0.98
0.99c 0.98d
Regression equation y = 1.01x + 0.17c NR y = 0.88x + 0.31e
y = 0.97x + 0.40e
y = 1.05x + 0.04e
y = 0.90x + 0.04e
y = 0.90x + 0.17 y = 0.98x + 0.31
Operational range NR 0.4–18 mg/L NR ∼0–9 mg/L (graph) 0.4–18 mg/L
VAD (%) NR NR NR NR NR NR NR
Precision
Sensitivity NR NR NR NR NR NR NR
Specificity NR NR NR NR NR NR NR
Intra-assay %CV NR NR 3.5% vs. 2.3 %f NR NR NR 3.5% vs. 2.3 %f
Inter-assay %CV NR NR NR NR NR NR NR
Inter-observer %CV NR NR NR NR NR NR NR
Bland Altman analysis comments A good level of agreement and no systematic error for β-carotene and vitamin A; “only 5% of the differences in measured values fell outside the 95% acceptability limits for β-carotene in dairy cows” Systematic error did not occur between methods:
4% of differences outside 95% limits
A good level of agreement and no systematic error for β-carotene and vitamin A; “only 5% of the differences in measured values fell outside the 95% acceptability limits for β-carotene in dairy cows” Graph presented, no comment (appears to have good agreement) Systematic error did not occur between methods:
4% of differences outside 95% limits
Reference (Ghaffari et al., 2019) (Raila et al., 2012) (Ghaffari et al., 2019) (BioAnalyt, NR) (Raila et al., 2012)

Notes: MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of Medicine, 2001).

a

Reference sample is in plasma.

b

Units: mg/L.

c

Average from reference analyses done in Germany and Switzerland.

d

Reference analysis done in Germany.

e

Reference analysis done in Switzerland.

f

Appears to represent the average CV for both whole blood and plasma samples.

In human blood samples, both the iCheck Fluoro and the CRAFTi portable fluorometers were used to measure retinol (Table 5a). The iCheck Fluoro studies showed a high correlation (0.98) and an R-squared values over 0.95 with respect to HPLC. Both CRAFTi studies found a mean difference in serum retinol of −0.07 µmol/L, and the 2011 study found moderate sensitivity and specificity in identifying VAD at either ≤ 0.70 µmol/L or ≤ 1.05 µmol/L. Few additional comparative data were available between studies. Bias analysis indicated an acceptable level of agreement (within two SDs or 95% acceptability limits) between these devices’ performance and HPLC.

Of note, we identified an additional report examining “vitamin A… in plasma” as measured by the CRAFTi compared with HPLC (Craft, 2005). The correlation between methods was 0.82. However, these data came from a summary of a poster submitted to a conference, and we were unable to find the full version of the poster; therefore, the report was excluded from our primary results and is not in Table 5a above.

Four studies analyzed either human or cow’s milk samples with the iCheck Fluoro (Table 5b). The device performance varied: some studies reported lower, equivalent, or higher retinol values than those of HPLC. The R2 values for the correlation between the device and HPLC ranged between 0.35 and 0.79 after adjustment for milk fat content.

In one study (Schweigert et al., 2011b), the authors tested increasingly diluted cow’s milk samples with 3.5% fat by using the iCheck, which showed linearity at an R2 of > 0.99 between 100 and 2500 µg RE/L. The same study also showed a positive correlation between percentage milk fat and µg RE/L milk with the iCheck Fluoro. Precision was tested over an operational range of 60 to 600 µg RE/L, and the inter-assay CV was < 3.5% (not shown in table).

RBP was measured in blood with two field-friendly immunoassays reported across three studies comparing a portable device to a reference method (Table 5c). A rapid enzyme immunoassay (RBP-REI), available from Scimedx Corp, was able to detect serum RBP within a range of 10–40 µg/mL, which correlated with the HPLC results (R2 = 0.79 to 0.86) (Hix et al., 2004, Hix et al., 2006). The RBP-REI assay was also compared with another portable, laboratory-based device (index 2), a commercially available radial immunodiffusion plate reader (RID; The Binding Site, San Diego), by measuring RBP in 40 serum samples (Hix et al., 2004). Compared with the higher R2 values in validation against HPLC (R2 = 0.82 and 0.86; Table 5c), the RBP-REI had a lower, but still acceptable, correlation with the RID method (R2 = 0.73; linearity: y = 0.50x + 0.45) (not shown in table). Comparison of RID and HPLC indicated a slightly lower correlation (R2 = 0.71) (Table 5c). Other validity and precision data were not reported for the comparisons of REI vs. RID, or RID vs. HPLC. From the current manufacturer’s website (accessed date: March 15, 2021) (The Binding Site, 2020), RBP was not listed among the human proteins for assessment with the RID plate reader.

A semi-quantitative antigen–antibody binding assay allowed for detection of low concentrations of RBP in serum samples (Ciaiolo et al., 2015). However, because only six samples were used for validation, drawing a conclusion regarding the efficacy of this method is difficult.

We identified two microfluidics-based devices, the EE-µPAD (Lee et al., 2016) and the Tidbit with HYPER filtration (Lu and Erickson, 2017, Lu et al., 2017, Lu et al., 2018), both of which were able to separate whole blood into serum, detect VAD at high sensitivity and specificity with respect to the reference ELISA test, and send results to a mobile device (Table 5d). We note that the ELISA test may not be a suitable reference method for assessing VAD, owing to inherent problems with antibodies to RBP. Neither device is currently on the market.

Although we identified several portable dark adaptometers (Table 1b), we found only one validation study between a portable dark adaptometer, the Scotopic Sensitivity-Tester 1 by LKC Technologies, and a reference standard, the Goldmann-Weekers dark adaptometer used in clinical settings (Peters et al., 2000) (Table 5e). The portable device was comparable to the reference standard in its sensitivity in identifying elevated final thresholds for dark adaptation, with a correlation (R2) of 0.77. However, this study was performed in the US in an eye clinic, and it remains to be tested and compared with the reference standard in field settings.

The iCheck Fluoro, a portable fluorometer, was used to measure bovine blood samples for retinol (Table 5f). Compared with HPLC, mean differences in whole blood, plasma, or serum retinol ranged from −0.01 µmol/L to 26.5 µmol/L, and the iCheck generally displayed higher values than HPLC. The correlation between the iCheck Fluoro and HPLC was positive, ranging in R2 values from 0.61 to 0.96. Weaker correlations were observed in cows (range: 0.78–88) than calves (0.90–0.96).

Raila and et al. (2017) also compared the correlation between bovine whole blood retinol (n = 10) and plasma retinol (n = 10), both measured by the index test iCheck Fluoro, and found a significant positive correlation (R2 = 0.87) (Raila et al., 2017). No studies reported sensitivity and specificity, or distinguished specific %CVs. Bias analysis indicated acceptable agreement between the device performance and HPLC.

The iCheck Carotene, a portable photometer, was used to measure carotenoids in bovine whole blood and plasma. Mean differences in beta-carotene concentration ranged from −0.29 mg/L to 0.26 mg/L in plasma samples in cows and calves (Table 5g). The correlation between iCheck Carotene and HPLC was high, with R2 between 0.93 and 0.99. No studies reported sensitivity, specificity, or specific %CVs. Bias analysis revealed an acceptable level of agreement between the device performance and that of HPLC.

Future perspectives and recommendations

Gaps and recommendations

On the basis of our review of the literature, portable devices fell into five categories:

  • 1.

    Portable fluorometers

  • 2.

    Portable photometers

  • 3.

    Field-friendly immunoassays and/or microfluidics-based devices

  • 4.

    Slit lamps

  • 5.

    Dark adaptometers

We found that, although many portable devices for quantifying vitamin A have been developed and described, only a few devices appear to be currently on the market or commercially available; of these, only two had easily accessible performance criteria information on the manufacturers’ websites related to vitamin A measurement. Studies tended not to report on portable device characteristics.

Some major gaps involve the lack of data reported by studies. Few studies have reported the portable device’s sensitivity and specificity in detecting VAD compared with the reference standard method—a necessary metric for validation and adoption by randomized trials. Furthermore, only the iCheck devices were assessed in more than two studies; other devices should be analyzed further for validation.

Minimal set of criteria for point-of-need devices

See Fig. 2.

Fig. 2.

Fig. 2

Minimal set of criteria for point-of-need devices. Adapted with permission (Huey, 2022).

The device should:

  • 1.

    Be lightweight with a small form factor for easy transport to the necessary location as needed.

  • 2.

    Be standalone without needing additional equipment and self-powered, and should pre-store all the required reagents for the test, and use common reagents that are available on the market.

  • 3.

    Be easy to use with minimal processing steps in the protocol, and should require minimal training effort.

  • 4.

    Have analytical performance (e.g., %CV < 5% or within Bland Altman 95% limits of agreement) comparable to those of the current laboratory standards, with a capability to test various biological samples.

  • 5.

    Be affordable and capable of scaling up with locally available consumables where needed.

  • 6.

    Be able to connect to the internet or an external hard drive with a built-in data management system to allow the test results to be reliably stored and transferred.

  • 7.

    Be able to output test results quickly and present in a format that is easy to interpret.

Conclusions

In this review, we identified 25 portable methods or devices for a variety of biological sample types including those of human (blood, milk, and eye/vision) and animal (blood and milk) origin. These included nine methods measuring biochemical markers of vitamin A or VAD (serum retinol, RBP, milk retinol, retinyl palmitate, and retinyl esters) and 17 portable methods measuring functional biomarkers (measures of eye health, for example dark adaptation).

The iCheck devices, including iCheck Carotene and iCheck Fluoro—for measuring total carotenoids or beta-carotene, or for measuring retinol, retinyl palmitate, retinyl acetate, or other esters, respectively, in blood or milk—were the only devices with manufacturer-reported performance metrics as well as the most information and data available to ascertain the method’s accuracy and precision with respect to those of a gold standard such as HPLC. These methods, in addition to the CRAFTi portable fluorometer, as compared with HPLC, were thus considered acceptable for measuring both blood and milk for biochemical biomarkers of vitamin A and detecting vitamin A deficiency.

In measuring human or cow milk samples’ retinol concentration, the iCheck Fluoro had variable performance across studies, including both lower and higher values than the gold standard HPLC, thus leading to weaker correlation values than those calculated for blood samples. However, the mean differences were<1 µmol/L, and the values were considered to be within the expected variance. Correlation was improved by diluting the samples; dilution may be required for higher accuracy when the portable method is used.

Several portable immunoassays (RBP-REI, RID, general immunoassay) and microfluidics-based methods (EE-µpad, TIDBIT with or without HYPER platform) for measuring RBP in human blood had acceptable correlations with HPLC reference methods and similar detection of VAD. However, these assays appeared not to be commercially available.

One study has measured eye function with a portable dark adaptometer (Scotopic Sensitivity Tester-1), which had comparable results to the gold standard, a Goldmann-Weekers dark adaptometer. However, field studies using this device in comparison to a reference remain to be performed. Given the importance of eye health as a functional indicator of vitamin A deficiency, this gap in the literature is substantial.

Finally, the iCheck Fluoro was used for measuring bovine blood samples for retinol. Generally, the retinol values were higher than those in samples tested by HPLC. Retinol measurements in calves appeared to have stronger correlations than retinol in cow’s blood.

Several studies examined the accuracy of the iCheck Carotene, as compared with HPLC, in determining carotenoid content in cow’s blood. Strong correlations with acceptable levels of agreement were observed between device performance and HPLC performance.

In summary, the iCheck devices are commercially available and are acceptable for measuring vitamin A in blood and milk, on the basis of the available data. Many of the other identified devices were proofs of concept and not yet commercially available. Several gaps remain, including studies comparing the other portable devices against a gold standard, particularly for functional indicators of vitamin A status/deficiency; available manufacturer-reported device performance criteria against which to compare the results of future investigations; and more comprehensive reporting of sensitivity, specificity, precision, and other validation metrics.

CRediT authorship contribution statement

Samantha L. Huey: Data curation, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing. Jesse T. Krisher: Data curation, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing. David Morgan: Project administration, Supervision, Writing – review & editing. Penjani Mkambula: Project administration, Writing – review & editing. Bryan M. Gannon: Conceptualization, Methodology, Writing – review & editing. Mduduzi N.N. Mbuya: Writing – review & editing. Saurabh Mehta: Conceptualization, Funding acquisition, Methodology, Supervision, Writing – review & editing.

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Acknowledgements

We gratefully acknowledge Balaji Srinivasan for designing Fig. 2 and for his inputs on the graphical abstract. Graphical abstract created with BioRender.com. The Cornell authors acknowledge funding support from the Global Alliance for Improved Nutrition and Bill and Melinda Gates Foundation grant number INV-002077 for this systematic review.

Footnotes

Appendix A

Supplementary data to this article can be found online at https://doi.org/10.1016/j.crbiot.2022.04.003.

Appendix A. Supplementary material

The following are the Supplementary data to this article:

Supplementary data 1
mmc1.docx (35.5KB, docx)

References

  1. Abebe Z., Haki G.D., Schweigert F.J., Henkel I.M., Baye K. Low breastmilk vitamin A concentration is prevalent in rural Ethiopia. Eur. J. Clin. Nutr. 2019;73:1110–1116. doi: 10.1038/s41430-018-0334-4. [DOI] [PubMed] [Google Scholar]
  2. Agrawal D., Sahu A. Prevalence of ocular morbidities among school children in Raipur district, India. Indian J. Ophthalmol. 2020;68:340–344. doi: 10.4103/ijo.IJO_1454_19. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Aguilar S.S., Wengreen H.J., Lefevre M., Madden G.J., Gast J. Skin carotenoids: a biomarker of fruit and vegetable intake in children. J Acad Nutr Diet. 2014;114:1174–1180. doi: 10.1016/j.jand.2014.04.026. [DOI] [PubMed] [Google Scholar]
  4. Banerjee, N. 2019. Ahmedabad: No blood test, desi device can detect Vitamin A deficiency for Rs 1. DNA India.
  5. Bechir M., Schelling E., Kraemer K., Schweigert F., Bonfoh B., Crump L., Tanner M., Zinsstag J. Retinol assessment among women and children in sahelian mobile pastoralists. EcoHealth. 2012;9:113–121. doi: 10.1007/s10393-012-0781-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. BioAnalyt . Rapid Measurement of Vitamin E and ß-Carotene at Cow-Side with iCheck. BioAnalyt GmBH; Teltow, Germany: 2020. [Google Scholar]
  7. Blaner W.S. Elsevier; 2020. Vitamin A and Provitamin A Carotenoids. [Google Scholar]
  8. Boateng L., Ashley I., Ohemeng A., Asante M., Steiner-Asiedu M. Improving blood retinol concentrations with complementary foods fortified with Moringa Oleifera leaf powder - A Pilot Study. Yale J. Biol. Med. 2018;91:83–94. [PMC free article] [PubMed] [Google Scholar]
  9. Chaimongkol, L., Winichagoon, P., Craft, N.E., Furr, H.C., 2008. Use of the CRAFTi portable fluorometer to measure serum retinol (vitamin A) concentrations. FASEB J. 22, 1102.4–1102.4.
  10. Chaimongkol L., Pinkaew S., Furr H.C., Estes J., Craft N.E., Wasantwisut E., Winichagoon P. Performance of the CRAFTi portable fluorometer comparing with the HPLC method for determining serum retinol. Clin. Biochem. 2011;44:1030–1032. doi: 10.1016/j.clinbiochem.2011.05.023. [DOI] [PubMed] [Google Scholar]
  11. Ciaiolo C., Genero A., Redoglia V., Omedè P. A simple, effective and inexpensive method to highlight antigen-antibody reaction based on liquid-semisolid phase. J. Biol. Methods. 2015;2(1):2015. [Google Scholar]
  12. Combs, G.F., Mcclung, J.P., 2017. Chapter 6 - Vitamin A. In: Combs, G.F., Mcclung, J.P. (Eds.) The Vitamins (Fifth Edition). Academic Press.
  13. Congdon N., Sommer A., Severns M., Humphrey J., Friedman D., Clement L., Wu L.S., Natadisastra G. Pupillary and visual thresholds in young children as an index of population vitamin A status. Am. J. Clin. Nutr. 1995;61:1076–1082. doi: 10.1093/ajcn/61.4.1076. [DOI] [PubMed] [Google Scholar]
  14. Congdon N.G., West K.P., Jr. Physiologic indicators of vitamin A status. J. Nutr. 2002;132:2889S–2894S. doi: 10.1093/jn/132.9.2889S. [DOI] [PubMed] [Google Scholar]
  15. Congdon N.G., Dreyfuss M.L., Christian P., Navitsky R.C., Sanchez A.M., Wu L.S., Khatry S.K., Thapa M.D., Humphrey J., Hazelwood D., West K.P., Jr. Responsiveness of dark-adaptation threshold to vitamin A and beta-carotene supplementation in pregnant and lactating women in Nepal. Am. J. Clin. Nutr. 2000;72:1004–1009. doi: 10.1093/ajcn/72.4.1004. [DOI] [PubMed] [Google Scholar]
  16. 2020.SciMedX Corporation. 2020. Retinol Binding Protein Test Kit: RBP196 96 Test Kit [Online]. Scimedx Corporation, Dover, NJ, USA. [Accessed 10 February 2021].
  17. Craft, N.E., 2005. Use of the CRAFTi portable fluorometer to measure vitamin A in fortified foods and blood (Poster, text summary). Sight and Life Newsletter. DSM.
  18. Crump L., Béchir M., Ngandolo B.N.R., Daugla D.M., Hattendorf J., Zinsstag J. Seasonal dynamics of human retinol status in mobile pastoralists in Chad. Acta Trop. 2017;166:280–286. doi: 10.1016/j.actatropica.2016.11.040. [DOI] [PubMed] [Google Scholar]
  19. De Ondarza, M., & Al., E., 2009. Production and Reproduction response of Dairy Cows to Supplemental Beta Carotene. Penn State Dairy Cattle Nutrition Workshop.
  20. de Pee S., Dary O. Biochemical indicators of vitamin A deficiency: serum retinol and serum retinol binding protein. J. Nutr. 2002;132:2895S–2901S. doi: 10.1093/jn/132.9.2895S. [DOI] [PubMed] [Google Scholar]
  21. EFSA Panel on Dietetic Products, Nutrition Allergies Scientific opinion on dietary reference values for vitamin A. EFSA J. 2015;13:4028. [Google Scholar]
  22. Elom A.K., Imane E.M., Kaoutar B., Khalid E.K., Asmaa E.H., Mehdi A., Noureddine E.H., Hassan A. Comparison of a fluorometric assay kit with high-performance liquid chromatography for the assessment of serum retinol concentration. African Health Sci. 2015;15:641–646. doi: 10.4314/ahs.v15i2.43. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Engle-Stone R., Haskell M.J., Ndjebayi A.O., Nankap M., Erhardt J.G., Gimou M.M., Brown K.H. Plasma retinol-binding protein predicts plasma retinol concentration in both infected and uninfected Cameroonian women and children. J. Nutr. 2011;141:2233–2241. doi: 10.3945/jn.111.145805. [DOI] [PubMed] [Google Scholar]
  24. Engle-Stone R., Haskell M.J., la Frano M.R., Ndjebayi A.O., Nankap M., Brown K.H. Comparison of breast milk vitamin A concentration measured in fresh milk by a rapid field assay (the iCheck FLUORO) with standard measurement of stored milk by HPLC. Eur. J. Clin. Nutr. 2014;68:938–940. doi: 10.1038/ejcn.2014.63. [DOI] [PubMed] [Google Scholar]
  25. Engle-Stone R., Alex O.N., Kenneth H.B., Martin N., Stephen A.V. Estimating the effective coverage of programs to control vitamin A deficiency and its consequences among women and young children in Cameroon. Food Nutr. Bull. 2015;36:S149–S171. doi: 10.1177/0379572115595888. [DOI] [PubMed] [Google Scholar]
  26. Erhardt J.G., Estes J.E., Pfeiffer C.M., Biesalski H.K., Craft N.E. Combined measurement of ferritin, soluble transferrin receptor, retinol binding protein, and C-reactive protein by an inexpensive, sensitive, and simple sandwich enzyme-linked immunosorbent assay technique. J. Nutr. 2004;134:3127–3132. doi: 10.1093/jn/134.11.3127. [DOI] [PubMed] [Google Scholar]
  27. Fujita M. Micronutrient Initiative; Ottawa, Canada: 2007. Technical Report for a Validation Test of Portable Crafti Fluorometer with Kenyan Samples. [Google Scholar]
  28. Ghaffari M.H., Bernhöft K., Etheve S., Immig I., Hölker M., Sauerwein H., Schweigert F.J. Technical note: rapid field test for the quantification of vitamin E, β-carotene, and vitamin A in whole blood and plasma of dairy cattle. J. Dairy Sci. 2019;102:11744–11750. doi: 10.3168/jds.2019-16755. [DOI] [PubMed] [Google Scholar]
  29. Global Alliance for Vitamin A, 2019. The Global Alliance for Vitamin A (GAVA) Strategic Plan 2016-2020. Global Alliance for Vitamin A.
  30. BioAnalyt GMBH. No date. iCheck Fluoro: iCheck Fluoro measures vitamin A in biological fluids and food [Online]. Teltow, Germany: BioAnalyt. Available from: <https://www.bioanalyt.com/product/icheck-fluoro/> [Accessed 10 February 2021].
  31. Gorstein J.L., Dary O., Pongtorn, Shell-Duncan B., Quick T., Wasanwisut. E. Feasibility of using retinol-binding protein from capillary blood specimens to estimate serum retinol concentrations and the prevalence of vitamin A deficiency in low-resource settings. Public Health Nutr. 2008;11:513–520. doi: 10.1017/S1368980007000821. [DOI] [PubMed] [Google Scholar]
  32. BioAnalyt GmBH. No date. iCheck Carotene: iCheck Carotene measures total carotenoids in food and biological fluids [Online]. Teltow, Germany: BioAnalyt GmBH. Available from: <https://www.bioanalyt.com/product/icheck-carotene/> [Accessed 10 February 2021].
  33. Hayashi H., Sato I., Suganuma H. Cutaneous carotenoid level measured by multiple spatially resolved reflection spectroscopy sensors correlates with vegetable intake and is increased by continual intake of vegetable juice. Diseases. 2020;9 doi: 10.3390/diseases9010004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Hix J., Martinez C., Buchanan I., Morgan J., Tam M., Shankar A. Development of a rapid enzyme immunoassay for the detection of retinol-binding protein. Am. J. Clin. Nutr. 2004;79:93–98. doi: 10.1093/ajcn/79.1.93. [DOI] [PubMed] [Google Scholar]
  35. Hix J., Rasca P., Morgan J., Denna S., Panagides D., Tam M., Shankar A.H. Validation of a rapid enzyme immunoassay for the quantitation of retinol-binding protein to assess vitamin A status within populations. Eur. J. Clin. Nutr. 2006;60:1299–1303. doi: 10.1038/sj.ejcn.1602456. [DOI] [PubMed] [Google Scholar]
  36. Hye N., Klein-Jöbstl D., Blessing A., Burmeister J., Hamann N., Aurich C., Drillich M. Effect of two postpartum intramuscular treatments with β-carotene (Carofertin®) on the blood concentration of β-carotene and on the reproductive performance parameters of dairy cows. Theriogenology. 2020;148:1–7. doi: 10.1016/j.theriogenology.2020.02.042. [DOI] [PubMed] [Google Scholar]
  37. Haag-Streit. 1900. Haag-Streit BA 904 Slit Lamp Instruction Manual. Available from: <https://www.haag-streit.com/fileadmin/Haag-Streit_UK/Downloads/HS_Diagnostics_downloads/Slit_lamps_downloads/BA904_downloads/BA904_IFU__28English_29__1_.pdf> [Accessed 10 Februrary 2021].
  38. Institute of Medicine . The National Academies Press; Washington, DC: 2001. Dietary Reference Intakes for Vitamin A, Vitamin K, Arsenic, Boron, Chromium, Copper, Iodine, Iron, Manganese, Molybdenum, Nickel, Silicon, Vanadium, and Zinc. [PubMed] [Google Scholar]
  39. Jans G., Devlieger R., de Preter V., Ameye L., Roelens K., Lannoo M., van der Schueren B., Verhaeghe J., Matthys C. Bariatric surgery does not appear to affect women's breast-milk composition. J. Nutr. 2018;148:1096–1102. doi: 10.1093/jn/nxy085. [DOI] [PubMed] [Google Scholar]
  40. Coburn Technologies Inc. SK-LS-1B Portable Slit Lamp [Online]. Available from: <https://www.coburntechnologies.com/ophthalmic/digital-slit-lamps/sk-portable-slit-lamp/> [Accessed 11 Feb 2021].
  41. Khan M.J., Usama U., Hanif S., Sadia F. Beta-carotene status are co-related with pupillary diameter changes in obese & normal adults in relation to prediction of heart disease. J. Clin. Res. Ophthalmol. 2019 [Google Scholar]
  42. Klein J., Darvin M.E., Meinke M.C., Schweigert F.J., Müller K.E., Lademann J. Analyses of the correlation between dermal and blood carotenoids in female cattle by optical methods. J. Biomed. Opt. 2013;18 [PubMed] [Google Scholar]
  43. Kosack C.S., Page A.L., Klatser P.R. A guide to aid the selection of diagnostic tests. Bull World Health Organ. 2017;95:639–645. doi: 10.2471/BLT.16.187468. [DOI] [PMC free article] [PubMed] [Google Scholar]
  44. Heine Optotechnik GmbH & Co. KG. 2018. HEINE Hand-held Ophthalmic Instruments: For precise diagnostic examination of the eye.
  45. KOWA New Lighter, Brighter, Whiter Kowa SL-17 Portable Slit Lamp.
  46. Labrique A.B., Palmer A.C., Healy K., Mehra S., Sauer T.C., West K.P., Jr. & sommer, a. A novel device for assessing dark adaptation in field settings. BMC Ophthalmol. 2015;15 doi: 10.1186/s12886-015-0062-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  47. Lee S., Aranyosi A.J., Wong M.D., Hong J.H., Lowe J., Chan C., Garlock D., Shaw S., Beattie P.D., Kratochvil Z., Kubasti N., Seagers K., Ghaffari R., Swanson C.D. Flexible opto-electronics enabled microfluidics systems with cloud connectivity for point-of-care micronutrient analysis. Biosens. Bioelectron. 2016;78:290–299. doi: 10.1016/j.bios.2015.11.060. [DOI] [PubMed] [Google Scholar]
  48. Livingston, M.L., Cowieson, A.J., Crespo, R., Hoang, V., Nogal, B., Browning, M., Livingston, K.A. 2020. Effect of broiler genetics, age, and gender on performance and blood chemistry. Heliyon, 6, e04400–e04400. [DOI] [PMC free article] [PubMed]
  49. Lu, Z., Erickson, D., 2017. TIDBIT: portable diagnostics of multiplexed nutrition deficiencies: iron, vitamin A and inflammation status (Conference Presentation).
  50. Lu Z., O'Dell D., Srinivasan B., Rey E., Wang R., Vemulapati S., Mehta S., Erickson D. Rapid diagnostic testing platform for iron and vitamin A deficiency. Proc. Natl. Acad. Sci. U. S. A. 2017;114:13513–13518. doi: 10.1073/pnas.1711464114. [DOI] [PMC free article] [PubMed] [Google Scholar]
  51. Lu Z., Rey E., Vemulapati S., Srinivasan B., Mehta S., Erickson D. High-yield paper-based quantitative blood separation system. Lab Chip. 2018;18:3865–3871. doi: 10.1039/c8lc00717a. [DOI] [PMC free article] [PubMed] [Google Scholar]
  52. Madureira A.M.L., Pohler K.G., Guida T.G., Wagner S.E., Cerri R.L.A., Vasconcelos J.L.M. Association of concentrations of beta-carotene in plasma on pregnancy per artificial insemination and pregnancy loss in lactating Holstein cows. Theriogenology. 2020;142:216–221. doi: 10.1016/j.theriogenology.2019.10.006. [DOI] [PubMed] [Google Scholar]
  53. Maggiano J.M., Fishman G.A., Evans L.S., Sieving P., Goldbaum M. Calibration error in dark adaptometer. Arch. Ophthalmol. 1978;96:1082–1083. doi: 10.1001/archopht.1978.03910050602024. [DOI] [PubMed] [Google Scholar]
  54. Marinovic A.C., May W.A., Sowell A.L., Khan L.K., Huff D.L., Bowman B.A. Effect of hemolysis on serum retinol as assessed by direct fluorometry. Am. J. Clin. Nutr. 1997;66:1160–1164. doi: 10.1093/ajcn/66.5.1160. [DOI] [PubMed] [Google Scholar]
  55. Mediworks. Portable Slit Lamp S150 [Online]. [Accessed 11 February 2021].
  56. Institute of Medicine, 2001. Dietary Reference Intakes for Vitamin A, Vitamin K, Arsenic, Boron, Chromium, Copper, Iodine, Iron, Manganese, Molybdenum, Nickel, Silicon, Vanadium, and Zinc. The National Academies Press, Washington, DC. [PubMed]
  57. Mehta, P.R., & Mehta, R.V., 2018. In-direct method and system for vitamin a deficiency detection. India patent application 201821013340.
  58. Mehta, P., 2018. Emtech Innovations. YouTube: YouTube.
  59. Mehta, P.R., 2019a. In-direct method and system for vitamin a deficiency detection / PROCEDE ET SYSTEME INDIRECT DE DETECTION D'UNE CARENCE EN VITAMINE A. Canada patent application 3097502.
  60. Mehta, P.R., 2019b. In-direct method and system for vitamin a deficiency detection / PROCÉDÉ ET SYSTÈME INDIRECT DE DÉTECTION D'UNE CARENCE EN VITAMINE A. France patent application PCT/IN2018/050816.
  61. Meinke M.C., Schanzer S., Lohan S.B., Shchatsinin I., Darvin M.E., Vollert H., Magnussen B., Köcher W., Helfmann J., Lademann J. Comparison of different cutaneous carotenoid sensors and influence of age, skin type, and kinetic changes subsequent to intake of a vegetable extract. J. Biomed. Opt. 2016;21 doi: 10.1117/1.JBO.21.10.107002. [DOI] [PubMed] [Google Scholar]
  62. Melo A.M.N., de Carvalho R.A.C., Figueiredo J.F.C., Vannucchi H., Jordão junior A., Rodrigues M.L.V. Serum vitamin A levels in patients with ocular lesions attributable to non-complicated malaria in the Brazilian Amazon region. Trans. R. Soc. Trop. Med. Hyg. 2004;98:485–488. doi: 10.1016/j.trstmh.2003.12.007. [DOI] [PubMed] [Google Scholar]
  63. Morgan E.H., Graham M.L., Marshall G.A., Hanson K.L., Seguin-Fowler R.A. Serum carotenoids are strongly associated with dermal carotenoids but not self-reported fruit and vegetable intake among overweight and obese women. Int. J. Behav. Nutr. Phys. Act. 2019;16:104. doi: 10.1186/s12966-019-0869-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  64. Page M.J., McKenzie J.E., Bossuyt P.M., Boutron I., Hoffmann T.C., Mulrow C.D., Shamseer L., Tetzlaff J.M., Akl E.A., Brennan S.E., Chou R., Glanville J., Grimshaw J.M., Hrobjartsson A., Lalu M.M., Li T., Loder E.W., Mayo-Wilson E., McDonald S., McGuinness L.A., Stewart L.A., Thomas J., Tricco A.C., Welch V.A., Whiting P., Moher D. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021;372 doi: 10.1136/bmj.n71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  65. Palmer, A., Labrique, A., Healy, K., Barffour, M., Siamusantu, W., Chileshe, J., West, K.P., 2015. Impact of Provitamin a Biofortified Maize Flour Consumption on Dark Adaptation in Zambian Children.
  66. Palmer A.C., Healy K., Barffour M.A., Siamusantu W., Chileshe J., Schulze K.J., West K.P., Jr., Labrique A.B. Provitamin A carotenoid-biofortified maize consumption increases pupillary responsiveness among zambian children in a randomized controlled trial. J. Nutr. 2016;146:2551–2558. doi: 10.3945/jn.116.239202. [DOI] [PubMed] [Google Scholar]
  67. Peters A.Y., Locke K.G., Birch D.G. Comparison of the Goldmann-Weekers dark adaptometer and LKC Technologies Scotopic Sensitivity tester-1. Doc Ophthalmol. 2000;101:1–9. doi: 10.1023/a:1002765024774. [DOI] [PubMed] [Google Scholar]
  68. Pharmanex/Nuskin Enterprises, 2018. Raman Measurement of Carotenoids in Living Tissues: A validated method for determining meaningful aspects of human health [Online]. Pharmanex/Nuskin Enterprises. Available from: <https://www.nuskin.com/content/dam/office/n_america/shared/en/scanner/scanner-studies-april-2018.pdf> [Accessed 08 February 2021].
  69. Raila J., Enjalbert F., Mothes R., Hurtienne A., Schweigert F.J. Validation of a new point-of-care assay for determination of β-carotene concentration in bovine whole blood and plasma. Vet. Clin. Pathol. 2012;41:119–122. doi: 10.1111/j.1939-165X.2012.00400.x. [DOI] [PubMed] [Google Scholar]
  70. Raila J., Kawashima C., Sauerwein H., Hülsmann N., Knorr C., Myamoto A., Schweigert F.J. Validation of blood vitamin A concentrations in cattle: comparison of a new cow-side test (iCheck™ FLUORO) with high-performance liquid chromatography (HPLC) BMC Veterinary Res. 2017;13:126. doi: 10.1186/s12917-017-1042-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  71. Reilly, M., G/Egziabher, T., Abebe, Y., Pruitt, A., Chall, A., Stoecker, B.J. 2006. Indicators of vitamin A status in rural villages in Southern Ethiopia. FASEB J., 20, LB99–LB99.
  72. Sanchez A.M., Congdon N.G., Sommer A., Rahmathullah L., Venkataswamy P.G., Chandravathi P.S., Clement L. Pupillary threshold as an index of population vitamin A status among children in India. Am. J. Clin. Nutr. 1997;65:61–66. doi: 10.1093/ajcn/65.1.61. [DOI] [PubMed] [Google Scholar]
  73. Schweigert F.J., Frey S.K., Mothes R., Dary O., Juarez P., Lascano V. Sight and Life; Basel, Switzerland: 2011. A New Test Kit’s Potential for the Rapid Analysis of Vitamin A in Human and Cow Milk. Sight and Life. [Google Scholar]
  74. Schweigert F., Immig I. Rapid assessment of ß-carotene status. International Dairy Topics. 2007;6 [Google Scholar]
  75. Schweigert F., Raila J., Mothes R., Frey S. Point of care measurements of Vitamin A in blood and breast milk for low resource settings. Ann. Nutr. Metab. 2011;58:382. [Google Scholar]
  76. Semba R.D., Yuniar Y., Gamble M.V., Natadisastra G., Muhilal, Assessment of vitamin A status of preschool children in Indonesia using plasma retinol-binding protein. J Trop Pediatr. 2002;48:84–87. doi: 10.1093/tropej/48.2.84. [DOI] [PubMed] [Google Scholar]
  77. Eurofins Craft Technologies, 2020. Specialized Solutions - Vitamins, carotenoids, pigments, antioxidants, and phytochemicals [Online]. [Accessed 10 February 2021 2021].
  78. Rexxam, S.-N. B. HANDY SLIT LAMP XL-1 [Online]. Available from: <http://www.shin-nippon.jp/products/xl1/> [Accessed 11 February 2021].
  79. The Binding Site, 2020. RID Reader [Online]. The Binding Site Group Ltd. Available from <https://www.bindingsite.com/en/our-products/immunoassays/digital-rid-plate-reader/overview> [Accessed 05 March 2021 2021].
  80. Digital Eye Center Ophthalmic Imaging Solutions. Handheld Slit Lamp S200 [Online]. Available from <https://www.digitaleyecenter.com/product/handheld-slit-lamp-s-200/> [Accessed 11 February 2021].
  81. Digital Eye Center Ophthalmic Imaging Solutions. Handheld Slit Lamp [Online]. Available from <https://www.digitaleyecenter.com/product/portable-slit-lamp/> [Accessed 11 February 2021].
  82. Steven D., Wald G. Vitamin A deficiency: A field study in newfoundland and labrador. J. Nutr. 1941;21:461–476. [Google Scholar]
  83. Tanumihardjo S.A., Russell R.M., Stephensen C.B., Gannon B.M., Craft N.E., Haskell M.J., Lietz G., Schulze K., Raiten D.J. Biomarkers of nutrition for development (BOND)-vitamin A review. J. Nutr. 2016;146:1816S–1848S. doi: 10.3945/jn.115.229708. [DOI] [PMC free article] [PubMed] [Google Scholar]
  84. Opticlar VisionMed. New SL280 Hand Held Slit Lamp [Online]. Available from <https://www.opticlar.co.uk/catalogue/vision-med/ophthalmology/slit-lamps/opticlar-sl280-led-hand-held-slit-lamp/> [Accessed 11 February 2021].
  85. Digital Eye Center Ophthalmic Imaging Solutions. Digital Portable Slit Lamp Microclear Hyperion [Online]. Available from <https://www.digitaleyecenter.com/product/digital-portable-slit-lamp-led-16/> [Accessed 11 February 2021].
  86. LKC Technologies, 2019. RETeval(R) Visual Electrodiagnostic System [Online]. Available from <https://lkc.com/wp-content/uploads/2021/01/2019_LKC_RETeval_Brochure.pdf> [Accessed 11 February 2021 2021].
  87. Reichert Technologies, Portable slit lamp PSL [Online]. Available from <https://www.medicalexpo.com/prod/reichert/product-70667-422317.html> [Accessed 11 February 2021].
  88. Von Lintig J. Elsevier; 2020. Carotenoids. [Google Scholar]
  89. Wald G. A portable visual adaptometer. J. Opt. Soc. Am. 1941;31:235–238. [Google Scholar]
  90. Whang C., Bermudez O., Chomat A.-M., Solomons N.W. Vitamin A concentrations in breast milk from rural Mam-Mayan women in the Western Highlands of Guatemala. Center for Studies of Sensory Impairment, Aging, and Metabolism (CeSSIAM) Bulletin of Research Abstracts. 2012;23:6–7. [Google Scholar]
  91. WHO . World Health Organization; Geneva: 2009. Global Prevalence of Vitamin A Deficiency in Populations at Risk 1995–2005, WHO Global Database on Vitamin A Deficiency. [Google Scholar]
  92. WHO, 2011. Serum retinol concentrations for determining the prevalence of vitamin A deficiency in populations. In: Organization, W.H., & System, V.A.M.N.I. (Eds.). WHO Press, Switzerland.
  93. World Health, O., 2014. Xerophthalmia and night blindness for the assessment of clinical vitamin A deficiency in individuals and populations. Unpublished, 1-6.
  94. Zambo Z., Sanchez-Mena J.D., Orozco M., Solomons N.W. Mounting and adaptation of a fluorescent rapid-assay device (iCHECK® FLUORO) for vitamin A in sugar and biological fluids. Center for Studies of Sensory Impairment, Aging, and Metabolism (CeSSIAM) Bulletin of Research Abstracts. 2012;23:4. [Google Scholar]
  95. Zidichouski J.A., Mastaloudis A., Poole S.J., Reading J.C., Smidt C.R. Clinical validation of a noninvasive, Raman spectroscopic method to assess carotenoid nutritional status in humans. J Am. Coll. Nutr. 2009;28:687–693. doi: 10.1080/07315724.2009.10719802. [DOI] [PubMed] [Google Scholar]