Abstract
This study aims to investigate the variability of pulmonary nodule (PN) volumetry on multiphase coronary CT angiograms (CCTA).
Two radiologists reviewed 5973 CCTA scans in this cross-sectional study to detect incidental solid noncalcified PNs measuring between 5 and 8 mm. Each radiologist measured the nodules’ diameters and volume, in systole and diastole, using 2 commercially available software packages to analyze PNs.
Bland-Altman analysis was applied between different observers, software packages, and cardiac phases. Bland-Altman subanalysis for the systolic and diastolic datasets were also performed.
A total of 195 PNs were detected within the inclusion criteria and measured in systole and diastole. Bland-Altman analysis was used to test the variability of volumetry between cardiac phases ([−47.0%; 52.3%]), software packages ([−50.2%; 68.2%]), and observers ([−14.5%; 27.8%]). The inter-observer variability of the systolic and diastolic subsets was [−13.6%; 31.4%] and [−13.9%; 19.7%], respectively.
Using diastolic volume measurements, the variability of PN volumetry on CCTA scans is similar to the reported variability of volumetry on low-dose CT scans. Therefore, growth estimation of PNs on CCTA scans could be feasible.
Keywords: coronary CT angiogram, lung cancer, pulmonary nodule, incidental finding, volumetry, volume doubling time
1. Introduction
Most scientific societies now recognize Coronary CT angiogram (CCTA) as a first-line examination in diagnosing and clinically managing stable coronary artery disease (CAD).[1,2] CCTA has replaced invasive coronary angiography in diagnosing CAD and improved clinical outcomes for these patients.[3–8] However, the increasing number of CCTA scans performed led to an increase in reported incidental extracardiac findings, with pulmonary nodules (PNs) being the most common (up to 28% of scans).[9–11]
The Fleischner Society recommends low-dose chest CT (LD-CT) follow-up of incidental solid, noncalcified PNs measuring between 5 and 8mm since the growth rate is a better marker for malignancy than size, calcifications, and morphological characteristics.[12–14] As such, incidental PNs detected in CCTA scans and between 5 and 8 mm in diameter should have an LD-CT scan as soon as possible.[12] We use dedicated volumetry tools to calculate the volume doubling time (VDT) and estimate the growth rate. When the VDT is shorter than 30 or longer than 400 days, the PN is more likely related to benign pathology (inflammatory causes or hamartomas, respectively). In comparison, a VDT between 30 and 400 days is suspicious for malignancy.[15] However, these tools suffer from numerous known limiting factors, including the specific software package, nodule’s size, contact to other structures, use of contrast agent, slice thickness, slice overlap, “kernel,” degree of chest expansion, motion artifacts, and many others.[15–20] In addition, factors related to cardiac function have also been suggested but not conclusively studied.[21]
Some authors have suggested that CCTA could be useful in lung cancer screening, given the considerable overlap of risk factors with CAD.[22] Compared to LD-CT, the higher temporal resolution and electrocardiogram (ECG)-gating makes CCTA scans more robust to motion and breathing artifacts. Likewise, estimating growth from previous CCTA scans could expedite the diagnosis and decision-making while allowing for better resource management in radiology departments. However, growth estimation by PN volumetry is related to the measurement variability, as it impacts the optimal waiting time between follow-up scans.[23] As such, a higher variability increases the overlap between benign and malignant pathology, reducing the discriminating power of volumetry (Fig. 1). The reported inter-scan variability of PN volumetry on LD-CT is around 25%, meaning that an increase in volume above 25% has a 95% likelihood of actual growth.[24,25] Recently, Bartlett et al published their results on nonmetastatic PNs (28–150 mm3) and proposed that a threshold of 15% of volume increase could be more appropriate and that closer surveillance may be justifiable.[26]
Figure 1.
The importance of measurement variability in growth estimation. All graphs present expected volume measurements of 3 pulmonary nodules with different growth-rates [i.e., VDT = 30 days, as in inflammatory changes (green); VDT = 250 days, as in malignancy (red); and VDT = 600 days, as in benign pathology (blue)]. The shaded areas represent possible volume measurements over time given the different measurement variability values [left (top and bottom)—25%; top right—10%, and bottom right—50%]. Volume measurements in overlapping shaded areas cannot be confidently attributed to 1 growth rate or another. The starting volume of a nodule [top left—80 cc3; bottom left—150 cc3] does not change the optimal waiting time between follow-up scans for a confident discrimination between suspicious and benign pathology (dashed orange line), but the measurement variability can significantly shorten (top right) or extend it (bottom right).
Current guidelines and recommendations regarding the optimal waiting time between follow-up scans assume this variability of 25% for volumetry tools on LD-CT.[23] For PN volumetry to apply to CCTA scans, its variability must be within these limits; otherwise, the current recommended waiting times would not be optimal. We found no published reports related to the variability of PN volumetry on ECG-gated CT scans.
This study aims to investigate the variability of PN volumetry on CCTA scans when applied to incidental solid PNs between 5 and 8 mm in diameter.
2. Methods
The Institutional Research Committee Review Board approved this retrospective cross-sectional study (observational, analytical) and waived the requirement for written informed consent due to the use of existing clinical data.
2.1. Study sample
The study sample includes consecutive patients who underwent a CCTA scan between January 1, 2016, and December 31, 2019, in our institution. All patients had one or more CCTA scans performed on the same equipment (Somatom Definition Flash; Siemens). Images were acquired during inspiration and contrast injection (75–95 mL of Niopam 370, at 5−7 mL/s), with anatomical coverage from the carina through the cardiac apex. Acquisition parameters include peak kilovoltage (kVp) between 100−120 kV; current modulation (CareDose 4D) with 320 mAs as reference; average acquisition time of 1−2 seconds; collimation of 128 × 0.6 mm; and pitch of 3.4. Reconstruction parameters include 2 mm and 0.75 mm overlapping slices through the prescribed range, with a B20f algorithm and field of view (FOV) of 22 cm.
The study’s inclusion criteria comprise scans with one or more incidental solid, noncalcified PN within the FOV in systole and diastole, measuring between 5 and 8 mm in long-axis diameter. Exclusion criteria comprise the absence of systolic (30–40%) or diastolic phase (70–80% of the cardiac cycle) in the archive; or when one of the cardiac phases did not present the PN in its FOV.
2.2. Observers and measurements
Two cardiothoracic radiologists with 10 (observer 1) and 5 years of experience (observer 2) identified nodules fitting the eligibility criteria within the study sample. In addition, a consensus decision between both observers and a third chest radiologist (with >25 years of experience) resolved disagreements regarding whether an appearance met the inclusion criteria.
Each observer measured each nodule using 2 different PN volumetry software packages in systole and diastole. The software packages used were Carestream Vue PACS v 11.4.01.1011 (Carestream Health, Inc, Rochester, NY) and Syngo via VB20 (Siemens Healthineers AG, Erlangen, Germany), as tool 1 and tool2, respectively.
Both tools performed semiautomatic segmentation by placing 1 seed point in the middle of the nodule. The observers did not correct the resulting segmentation. Still, they recorded the adequacy of the segmentation as “inadequate” if 3 consecutive segmentation attempts had failed or when the segmentation did not adequately represent the nodule. Also recorded were the observer’s initials; software package used; cardiac phase; long-axis diameter of the nodule; location of the nodule in axial and coronal planes; and nodule volume calculated using the volumetry tool.
2.3. Statistical analysis
The data were analyzed using SPSS software (ver. 26.0; IBM Corporation).
Descriptive statistical analysis of the included nodules was performed.
All cases with inadequate nodule segmentation were excluded from further analysis.
The association of volume measurements was tested using Kendall tau correlation coefficients (τ) between different cardiac phases, software packages and observers.
The Bland-Altman analysis is a well-known statistical tool for method comparison and has been used previously to study the variability of PN volumetry.[27] Volume measurement agreement was tested by Bland-Altman analysis and plots. In addition, further inter-observer analyses of the systolic and diastolic data subsets were performed.
The estimation of the limits of agreement (LOA) in the Bland-Altman analysis used the nonparametric quartile analysis after the exclusion of normality of the differences of measurements. Possible reasons for this non-normal distribution include the presence of a natural limit (i.e., volume must be a positive number) and the restriction of the nodules’ diameter between 5 and 8mm (i.e., sorted by diameter).
3. Results
Of a total of 5973 candidate CCTA scans performed between 2016 and 2019, 4478 were excluded for not having both a systolic and diastolic phase in the archive. Of the remaining 1495 scans, 1357 were excluded for not having qualifying solid PNs in both systolic and diastolic phases, and 31 scans were excluded after consensus decision (as not representing true nodules).
From the 107 scans included, a total of 195 nodules were identified with a manual long-axis measurement between 5 and 8 mm. The mean age of the patients was 67.8 years, with a male to female ratio of 1.34 (Table 1). Nodules were more frequently identified in the upper thirds of the FOV in the coronal plane (49.2%) and between the middle (39.5%) and posterior (40.5%) thirds in the axial plane.
Table 1.
Demographics of the sample population.
| (n = 195) | ||
|---|---|---|
| Age (yrs): M ± SD | 67.8 ± 11.7 | |
| Gender: n (%) | Male | 112 (57.4) |
| Female | 83 (42.6) |
M = Mean, SD = standard deviation
For each nodule, each set of volume measurements was repeated by each observer, for systole and diastole, and using each software package, resulting in a total of 1560 volume measurements (8 measurements per nodule). Figure 2 provides an example of measurement. The quality of the nodule segmentation was considered inadequate in more cases using 1 software package (tool 1; n = 157; 20.1%) than the other (tool 2; n = 95; 12.2%).
Figure 2.
Example of a nodule included in the study measured in systole (A) and diastole (B). The seed points (*) were placed in the nodule center, and the yellow lines show the semiautomatic segmentation result, which was not corrected and considered adequate by both observers. Notice the small vessel (arrows) approaching the nodule, which was excluded from the segmentation in both instances. The volume measured in diastole was 80% greater than in systole.
The mean volume was close to 50mm3 with a median of 35mm3, ranging from 10.4 to 400 mm3 (Table 2).
Table 2.
Descriptive statistical analysis of volume measurements in the sample population and at different phases of the cardiac cycle, using different software packages and by different observers.
| Volume | n | Mean (mm3) | SD (mm3) | Q1 (mm3) | Median (mm3) | Q3 (mm3) |
|---|---|---|---|---|---|---|
| Sample population | ||||||
| 1237 | 63.282 | 68.039 | 24.5 | 38 | 69 | |
| Between systole and diastole | ||||||
| Systole | 645 | 51.99 | 46.11 | 23.375 | 36 | 62 |
| Diastole | 645 | 50.836 | 44.947 | 23.3 | 35.5 | 60.4 |
| Difference | 645 | −1.154 | 13.13 | −3.6 | −0.5 | 1.7 |
| Between different software packages | ||||||
| Tool 1 | 616 | 47.108 | 38.438 | 23.3 | 34.5 | 59.8 |
| Tool 2 | 616 | 48.807 | 42.257 | 22 | 35 | 59 |
| Difference | 616 | 1.699 | 15.942 | −3.25 | −2.0 | 6.7 |
| Between different observers | ||||||
| Observer 1 | 642 | 53.62 | 46.993 | 24 | 36.6 | 62.9 |
| Observer 2 | 642 | 52.91 | 47.172 | 23.4 | 36.2 | 62 |
| Difference | 642 | −0.701 | 8.723 | 0 | 0 | 0 |
| Between different observers (systolic dataset) | ||||||
| Observer 1 | 318 | 44.668 | 29.936 | 23.4 | 34.9 | 59 |
| Observer 2 | 318 | 43.751 | 29.583 | 23 | 34.05 | 58 |
| Difference | 318 | −0.917 | 9.911 | 0 | 0 | 0 |
| Between different observers (diastolic dataset) | ||||||
| Observer 1 | 322 | 43.862 | 29.316 | 23 | 34.6 | 57.2 |
| Observer 2 | 322 | 43.019 | 28.611 | 23 | 34 | 55 |
| Difference | 322 | −0.842 | 6.806 | 0 | 0 | 0 |
Q1 = first quartile, Q3 = third quartile, SD = standard deviation.
The volume measurements of the solid PN show a very strong association between systole and diastole, with a Kendall tau correlation coefficient (τ) of 0.812 (CI95 = [0.785; 0.838], df = 645, P < .0001); between software packages with a τ = 0.744 (CI95 = [0.714; 0.771], df = 616, P < .0001); and between different observers, with a τ = 0.942 (CI95 = [0.916; 0.957], df = 682, P < .0001) globally; τ = 0.928 (CI95 = [0.874; 0.955], df = 338, P < .0001) for the systolic dataset; and τ = 0.956 (CI95 = [0.931; 0.971], df = 344, P < .0001) for the diastolic dataset.
The Bland-Altman plots show an increase in the differences’ variability as the measurement’s magnitude increases, particularly for different cardiac phases and software packages. The estimated LOA of the percent Bland-Altman plots are [−47.0%; 52.3%], between phases of the cardiac cycle, and [−50.1%; 68.2%] between software packages (Fig. 3 and Table 3), with no significant bias. The estimated LOA of the percent Bland-Altman plots between different observers are [−14.5%; 27.8%] globally, [−13.6%; 31.4%] for the systolic dataset and [−13.9%; 19.7%] for the diastolic dataset.
Figure 3.
Relative (percent) Bland-Altman plots, with estimated limits of agreement, when comparing volume measurements between systole and diastole (top left corner), different software packages (top right corner), and different observers (bottom). Interobserver subanalysis is presented for the systolic (bottom left corner) and diastolic datasets separately (bottom right corner).
Table 3.
Results of the Bland-Altman analysis.
| Bland-Altman (mm3) | Percent Bland-Altman (%) | |||||
|---|---|---|---|---|---|---|
| Bias | lower LOA | Upper LOA | Bias | lower LOA | Upper LOA | |
| Between systole and diastole | 1.15 | −24.91 | 36.76 | 1.73 | −47.02 | 52.29 |
| Between software packages | 1.7 | −39 | 34.08 | 1.35 | −50.16 | 68.21 |
| Between observers | 0.7 | −6.2 | 17.39 | 1.49 | −14.45 | 27.77 |
| Between observers (systolic dataset) | 0.88 | −4.7 | 24.47 | 1.75 | −13.63 | 31.36 |
| Between observers (diastolic dataset) | 0.53 | −8.58 | 15.85 | 1.23 | −13.92 | 19.66 |
LOA = estimated limits of agreement.
4. Discussion
Growth estimation is feasible using PN volumetry on the diastolic phase of CCTA scans. Conversely, comparing volume measurements from different cardiac phases could be as unreliable as using different volumetry tools (similar measurement variability), implying a lower discriminating power between benign and malignant pathology.
For the Bland-Altman analyses, we defined the a priori LOA as a difference of ±25% in volume, according to the previously reported variability of PN volumetry in LD-CT scans, which form the basis of recommendations for waiting time between follow-up scans.[15,23,28]
In the study’s results, the shapes of the Bland-Altman plots for different software tools and cardiac phases show that the difference in volume increases proportionally with their mean.[27] Therefore, the variability is assumed to be relative or proportional (represented as a volume percentage). However, this proportionality is less obvious between different observers with larger nodules. For example, for PNs larger than 65.6 mm3, the estimated LOA between observers vary between −8.6 mm3 and 15.9 mm3, which will be <25% of their volume. The proportionality seems to apply for smaller nodules, so the percentage Bland-Altman analysis is still relevant with the estimated LOA ([−14.45%; 27.77%]) much closer but outside the a priori LOA. However, in the inter-observer subanalysis specific for diastolic measurements, the estimated LOA ([−13.92%; 19.66%]) are well within the a priori LOA. The lower variability seen in diastole is likely related to less cardiac motion when compared with systole and the higher temporal resolution compared to LD-CT scans. In addition, the low interobserver variability is an expected benefit of semiautomatic volumetry tools.[15]
Identified limitations of the present study include (1) the large portion of excluded candidate CCTA scans, which was due to departmental protocol not to archive, per default, both systolic and diastolic phases; (2) the disproportionate representation of smaller nodules in comparison to larger ones; and (3) the non-normal distribution of the differences between measurements, which implied a nonparametric statistical approach.
To the best of our knowledge, this is only the second (and largest) study to focus on PN volumetry in ECG-gated CT scans after Boll et al suggested that factors related to the pulmonary circulation (cardiac phase and cardiac motion) could influence nodule segmentation.[21] It is also the first study to measure the in vivo variability of PN volumetry on CCTA, which is the main requirement for growth estimation. Therefore, our conclusions still need to be reproduced and validated independently, and further research is needed with a larger sample size before volumetry of PN in CCTA scans becomes useful for growth estimation between 2 CCTA scans. Further research should also compare in vivo volume measurements between CCTA in diastole and LD-CT, before growth of PN can be estimated by comparing CCTA and LD-CT scans (as used in Lung cancer screening), since significant protocol differences exist.
The increased variability of volume measurements between cardiac phases could model the effect of hemodynamic changes often seen between follow-up scans, like the acute onset of heart failure or pleural effusion. Such events are likely more significant than the relatively small intrascan changes seen between systole and diastole. Since the actual in vivo volume of a PN is unknown, a common approach in studies on PN volumetry is to use “coffee-break” studies, where the PN is scanned twice in a short time frame (typically minutes), thus excluding any real growth. In this way, measurement variability explains any volume difference.[15,26,27] However, the impact of such hemodynamically significant events is difficult to study directly since the absence of growth cannot be assumed over any substantial period. In addition, the extra radiation and time demands of “coffee-break” studies make them more suited to phantom studies. On the other hand, CCTA scans allow several opportunities to measure the volume of a PN at different cardiac phases and within a single scan (i.e., no growth occurred), thus providing a tool better suited for clinical research.
5. Conclusions
The interobserver variability of PN volumetry on CCTA scans was comparable to the reported for LD-CT only in diastole. As such, growth estimation of PN between 2 CCTA scans could be feasible using diastolic measurements. However, using systolic volume measurements or comparing volume measurements from systole and diastole implies a higher variability, which could lower the discriminating power of PN volumetry.
Acknowledgments
We would like to acknowledge Prof Paulo Pereira for statistical support.
Abbreviations:
- CAD =
- Coronary artery disease
- CCTA =
- Coronary CT angiography
- CI95 =
- 95% Confidence interval
- CT =
- Computerized Tomography
- ECG =
- Electrocardiogram
- FOV =
- Field of view
- kVp =
- Peak kilovoltage
- LD-CT =
- Low-dose CT
- LOA =
- Limits of agreement
- PN =
- Pulmonary nodule
- VDT =
- Volume doubling time.
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.
The authors have no conflicts of interest to disclose.
How to cite this article: Pinto E, Penha D, Hochhegger B, Monaghan C, Marchiori E, Taborda-Barata L, Irion K. Variability of pulmonary nodule volumetry on coronary CT angiograms. Medicine 2022;101:335(e30332).
Contributor Information
Diana Penha, Email: dianapenha@gmail.com.
Bruno Hochhegger, Email: brunohochhegger@gmail.com.
Colin Monaghan, Email: colingrowler@aol.com.
Edson Marchiori, Email: edmarchiori@gmail.com.
Luís Taborda-Barata, Email: tabordabarata@fcsaude.ubi.pt.
Klaus Irion, Email: klaus.irion@btinternet.com.
References
- [1].Al-Mallah MH, Aljizeeri A, Villines TC, et al. Cardiac computed tomography in current cardiology guidelines. J Cardiovasc Comput Tomogr. 2015;9:514–23. [DOI] [PubMed] [Google Scholar]
- [2].Knuuti J, Wijns W, Saraste A, et al. 2019 ESC guidelines for the diagnosis and management of chronic coronary syndromes. Eur Heart J. 2020;41:407–77. [DOI] [PubMed] [Google Scholar]
- [3].Tzolos E, Newby DE. Coronary computed tomography angiography improving outcomes in patients with chest pain. Curr Cardiovasc Imaging Rep. 2019;12:15. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [4].Taron J, Foldyna B, Mayrhofer T, et al. Risk stratification with the use of coronary computed tomographic angiography in patients with nonobstructive coronary artery disease. JACC Cardiovasc Imaging. 2021;14:2186–95. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Vij A, Kassab K, Chawla H, et al. Invasive therapy versus conservative therapy for patients with stable coronary artery disease: an updated meta-analysis. Clin Cardiol. 2021;44:675–82. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Maron DJ, Hochman JS, Reynolds HR, et al. Initial invasive or conservative strategy for stable coronary disease. N Engl J Med. 2020;382:1395–407. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Leipsic JA, Achenbach S. The ISCHEMIA trial: implication for cardiac imaging in 2020 and beyond. Radiol Cardiothorac Imaging. 2020;2:e200021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Newby DE, Adamson PD, Berry C, et al. ; SCOT-HEART Investigators. Coronary CT angiography and 5-year risk of myocardial infarction. N Engl J Med. 2018;379:924–33. [DOI] [PubMed] [Google Scholar]
- [9].Ramanathan S, Ladumor SB, Francis W, et al. Incidental non-cardiac findings in coronary computed tomography angiography: is it worth reporting? J Clin Imaging Sci. 2019;9:40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [10].Kay FU, Canan A, Abbara S. Common incidental findings on cardiac CT: a systematic review. Curr Cardiovasc Imaging Rep. 2019;12. [Google Scholar]
- [11].Williams MC, Hunter A, Shah AS, et al. Impact of noncardiac findings in patients undergoing CT coronary angiography: a substudy of the Scottish computed tomography of the heart (SCOT-HEART) trial. Eur Radiol. 2018;28:2639–46. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Bueno J, Landeras L, Chung JH. Updated Fleischner society guidelines for managing incidental pulmonary nodules: common questions and challenging scenarios. RadioGraphics. 2018;38:1337–50. [DOI] [PubMed] [Google Scholar]
- [13].Walter JE, Heuvelmans MA, de Jong PA, et al. Occurrence and lung cancer probability of new solid nodules at incidence screening with low-dose CT: analysis of data from the randomised, controlled NELSON trial. Lancet Oncol. 2016;17:907–16. [DOI] [PubMed] [Google Scholar]
- [14].MacMahon H, Naidich DP, Goo JM, et al. Guidelines for management of incidental pulmonary nodules detected on CT images: from the Fleischner society 2017. Radiology. 2017;284:228–43. [DOI] [PubMed] [Google Scholar]
- [15].Devaraj A, van Ginneken B, Nair A, et al. Use of volumetry for lung nodule management: theory and practice. Radiology. 2017;284:630–44. [DOI] [PubMed] [Google Scholar]
- [16].Soo E, Edey AJ, Mak SM, et al. Impact of choice of volumetry software and nodule management guidelines on recall rates in lung cancer screening. Eur J Radiol. 2019;120:108646. [DOI] [PubMed] [Google Scholar]
- [17].Honda O, Johkoh T, Sumikawa H, et al. Pulmonary nodules: 3D volumetric measurement with multidetector CT - effect of intravenous contrast medium. Radiology. 2007;245:881–7. [DOI] [PubMed] [Google Scholar]
- [18].Christe A, Bronnimann A, Vock P, et al. Volumetric analysis of lung nodules in computed tomography (CT): comparison of two different segmentation algorithm softwares and two different reconstruction filters on automated volume calculation. Acta Radiol. 2014;55:54–61. [DOI] [PubMed] [Google Scholar]
- [19].Ravenel JG, Leue WM, Nietert PJ, et al. Pulmonary nodule volume: Effects of reconstruction parameters on automated measurements - a phantom study. Radiology. 2008;247:400–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [20].Petrou M, Quint LE, Nan B, et al. Pulmonary nodule volumetric measurement variability as a function of CT slice thickness and nodule morphology. Am J Roentgenol. 2007;188:306–12. [DOI] [PubMed] [Google Scholar]
- [21].Boll DT, Gilkeson RC, Fleiter TR, et al. Volumetric assessment of pulmonary nodules with ECG-Gated MDCT. Am J Roentgenol. 2004;183:1217–23. [DOI] [PubMed] [Google Scholar]
- [22].Robertson J, Nicholls S, Bardin P, et al. Incidental pulmonary nodules are common on CT coronary angiogram and have a significant cost impact. Heart Lung Circ. 2019;28:295–301. [DOI] [PubMed] [Google Scholar]
- [23].Heuvelmans MA, Oudkerk M, De Bock GH, et al. Optimisation of volume-doubling time cutoff for fast-growing lung nodules in CT lung cancer screening reduces false-positive referrals. Eur Radiol. 2013;23:1836–45. [DOI] [PubMed] [Google Scholar]
- [24].Wormanns D, Kohl G, Klotz E, et al. Volumetric measurements of pulmonary nodules at multi-row detector CT: in vivo reproducibility. Eur Radiol. 2004;14:86–92. [DOI] [PubMed] [Google Scholar]
- [25].Gietema HA, Schaefer-Prokop CM, Mali WPTM, et al. Pulmonary nodules: interscan variability of semiautomated volume measurements with multisection CT—influence of inspiration level, nodule size, and segmentation performance. Radiology. 2007;245:888–94. [DOI] [PubMed] [Google Scholar]
- [26].Bartlett EC, Kemp SV, Rawal B, et al. Defining growth in small pulmonary nodules using volumetry: results from a “coffee-break” CT study and implications for current nodule management guidelines. Eur Radiol. 2022;32:1912–20. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [27].Talwar A, Willaime JMY, Pickup LC, et al. Pulmonary nodules: assessing the imaging biomarkers of malignancy in a “coffee-break”. Eur J Radiol. 2018;101:82–6. [DOI] [PubMed] [Google Scholar]
- [28].Giavarina D. Understanding bland Altman analysis. Biochem Medica. 2015;25:141–51. [DOI] [PMC free article] [PubMed] [Google Scholar]