Abstract
Background
The purpose of this study was to evaluate the reproducibility of planned component positioning including tibial rotational alignment in patient-specific instrumentation (PSI) for total knee arthroplasty (TKA).
Methods
A total of 100 knees of 100 patients underwent TKA using PSI (n = 50) or the conventional method (n = 50). Full-length anteroposterior radiographs of the lower limb were taken in the standing position, and the coronal alignments of the femoral and tibial components were measured. Computed tomography (CT) images of the lower limb were obtained preoperatively and postoperatively, and the rotational alignments of the femoral and tibial components were measured. The difference from the preoperative planning in tibial rotational alignment was measured using three-dimensionally merged pre- and postoperative images. The mean values and rates of outliers in each measurement were compared between the PSI group and the conventional group.
Results
There were no significant differences in coronal alignment of the femoral and tibial components and rotational alignment of the femoral component between the two groups. With respect to rotational alignment of the tibial component from the preoperatively planned reference axis, the PSI group showed a lower rate of outliers (internal rotation > 10°) than the conventional group (p < 0.05).
Conclusions
This study demonstrated that the difference from the preoperative planning in tibial rotational positioning was accurately evaluated using novel three-dimensional measurement method, and PSI could reduce outliers in rotational alignment of the tibial component (internal rotation > 10°). PSI is a useful technique for improving the reproducibility of the planned tibial rotational positioning in TKA.
Keywords: Total knee arthroplasty, Patient-specific instrumentation, Conventional method, Tibial rotational alignment
Introduction
Total knee arthroplasty (TKA) is an established surgical procedure to treat end-stage knee osteoarthritis. Accurate bone cutting and restoration of mechanical limb alignment are the important factors affecting long-term survival. Coronal malalignment is the common cause of revision surgery [1–3]. Rotational alignment of the components is also important for successful TKA. Malrotation of either the femoral or tibial component can lead to patellar subluxation or dislocation [4] and unexplained knee pain [5, 6]. Unexpected malalignment and malposition of the components are a problem for orthopedic surgeons performing TKA.
The conventional 2D planning-based technique using intramedullary or extramedullary alignment guides is still popular for bone cutting and component positioning in TKA. A meta-analysis reported that mechanical axis malalignment of greater than 3° occurred in a mean of 31.8% of patients who underwent conventional TKA [7]. Optimal placement of the prosthetic components is a major issue in TKA, and computer-assisted surgery or patient-specific instrumentation (PSI) has been developed to improve limb and prosthesis alignment and optimal positioning of the components.
PSI for TKA is a technique that provides 3D templating and custom-fit cutting guides of the distal femur and proximal tibia for each patient using bone models based on computed tomography (CT) or magnetic resonance imaging. This system helps reduce operative time and the number of instrumentation trays [8, 9]. PSI is expected to increase the accuracy of component positioning and reduce outliers [8, 10, 11]. Many studies have investigated the coronal and sagittal alignment of the components using PSI. However, there are few reports evaluating the effect of PSI on rotational alignment in the transverse plane, especially in tibial rotation [12]. In addition, these previous studies evaluated the rotational alignment of the tibial component using only postoperative CT images [13, 14], and it was unknown whether the component was positioned as preoperatively planned. Thus, the present study investigated the accuracy of the rotational positioning in the tibial component using the same reference axis as the preoperative planning.
The purpose of this study was to evaluate the reproducibility of component positioning including tibial rotational alignment in PSI technique for TKA using the same reference axis as the preoperative planning. It was hypothesized that PSI improves the reproducibility of the planned component positioning and reduces outliers compared to the conventional technique.
Materials and methods
Patients
This retrospective, case–control study was approved by the institutional review board (#B200500006), and written, informed consent was obtained from all patients. Inclusion criteria for this study were patients with osteoarthritis or rheumatoid arthritis of the knee who underwent primary cemented cruciate-retaining TKA (Vanguard™ Complete Knee System; Zimmer Biomet Inc., Warsaw, IN, USA) without resurfacing the patella by several surgeons between 2018 and 2020. Exclusion criteria were patients with a history of osteotomy, valgus knee deformity, or post-traumatic osteoarthritis. A total of 100 knees of 100 patients were enrolled in this study. PSI was used in 50 consecutive knees, and the conventional method was used in 50 consecutive knees.
Preoperative planning and surgical procedure
Preoperative planning and the surgical procedure were performed according to the previously described methods [15, 16]. For the PSI group, CT of the lower limb through the hip and ankle was performed preoperatively using the Signature™ system manufacturing protocol. Virtual 3D models of the femur and tibia were created based on CT data with a specific software program, and the patient-specific cutting jigs were generated based on the surgeon’s predetermined default settings: femur varus/valgus 0° (perpendicular to the mechanical axis), flexion 3°, rotation 0° (parallel to the transepicondylar axis); tibia varus/valgus 0° (perpendicular to the mechanical axis), posterior slope 3°, and rotation 0° (line connecting the medial third of the tibial tuberosity and the center point of the attachment of the posterior cruciate ligament).
A standard anterior midline skin incision was made, and the knee joint was exposed through a medial parapatellar arthrotomy. A measured resection technique was used in all patients. For the PSI group, patient-specific cutting guides were carefully placed over the articular surfaces prior to removing osteophytes, and positioning guide pins were inserted through the pin holes. For the conventional group, traditional jig-based instrumentation was used. Target alignment of component positioning was similar to the setting in the PSI group. The tibial anterior–posterior (AP) axis (rotation 0°) was visually determined with reference to the line connecting the medial third of the tibial tuberosity and the center of the attachment of the posterior cruciate ligament, and as a positioning guide, a Kirschner wire was inserted parallel to this reference line just below the bone cutting level prior to bone resection. The bone cuts were made through the slot on the standard resection blocks using traditional saw blades. Femoral and tibial components were inserted with bone cement.
Measurements of coronal alignment
Full-length anteroposterior radiographs of the lower limb were taken in the standing position, and the α angle and β angle were measured according to the method reported by Ewald [17]. The α angle was defined as the angle between the femoral functional axis and the articular surface of the femoral component. The β angle was defined as the angle between the functional axis of the tibia and the articular surface of the tibial component. Outliers were defined as deviations of more than 3° either varus or valgus from the targeted neutral component positioning [15, 18].
Measurements of rotational alignment using CT-based simulation software
CT images of the lower limb were obtained preoperatively and postoperatively, and measurements of rotational alignment were performed using Orthomap 3D software (Stryker Corporation, Kalamazoo, MI, USA). The measurement of femoral rotational alignment was carried out using the protocol previously described by Berger et al. [4]. A single axial plane of the femur was used at the level of the femoral epicondyles. The surgical transepicondylar axis was drawn from the sulcus of the medial epicondyle to the prominence of the lateral epicondyle. The posterior condylar line of the femoral component was drawn connecting the posterior margins of the medial and lateral posterior component condylar surfaces. The angle between these two lines was measured as the femoral rotational alignment. Outliers of femoral rotational alignment were defined as deviation of more than 3° either internally or externally from the targeted neutral component positioning [18, 19]. For measurement of the difference from the preoperative planning in tibial rotational alignment, the preoperative CT images were superimposed on the postoperative CT images. On the preoperative CT image, the line connecting the center of the knee joint and the center of the ankle joint was set as the z-axis (Fig. 1). The reference axis of tibial rotation was set as the line connecting the medial third of the tibial tuberosity and the center point of the attachment of the posterior cruciate ligament, and this line was also used as the reference axis of rotation during the TKA procedure (Fig. 2). The base of the tibial component was set as the xy plane on the postoperative CT image. The anteroposterior (AP) axis of the component was defined as the line perpendicular to the major axis of the implant. By superimposing the preoperative and postoperative CT images, the angle between the reference axis of tibial rotation and the AP axis of the tibial component was measured as the tibial rotational angle (Fig. 3). Outliers of tibial rotational alignment were defined as deviation of internal rotation of greater than 10° from the targeted neutral component positioning [6].
Statistical analysis
Statistical analysis was carried out using IBM SPSS Statistics (version 27.0; IBM Corporation, Chicago, IL, USA). Normality of the data distributions was assessed using the Shapiro–Wilk test. Student’s t test or the Mann–Whitney U test was used to compare continuous variables between two groups. Pearson’s chi-squared test was used to compare categorical variables. A p value < 0.05 was considered significant. The intra-rater and inter-rater reliabilities of measurements were assessed by calculating intraclass correlation coefficients (ICCs).
Results
Patients’ demographics
There were no significant differences between the PSI group and the conventional group in age, sex, body mass index, and hip–knee–ankle angle (Table 1).
Table 1.
PSI group (n = 50) | Conventional group (n = 50) | p value | |
---|---|---|---|
Gender, male/female | 12/38 | 11/39 | 0.812 |
Mean age, years | 74.7 ± 9.0 | 74.7 ± 8.1 | 0.948 |
Height, cm | 154.8 ± 8.9 | 154.0 ± 8.5 | 0.912 |
Weight, kg | 64.5 ± 13.3 | 63.4 ± 13.2 | 0.924 |
Body mass index, kg/m2 | 26.5 ± 4.6 | 26.4 ± 4.3 | 0.545 |
Preoperative hip–knee–ankle angle, ° | − 10.3 ± 6.1 | − 10.4 ± 7.1 | 0.658 |
Values are number or mean ± standard deviation
Radiographic assessment of coronal alignment
The postoperative values of the α angle and the β angle were 88.8 ± 2.0° and 90.1 ± 2.4° in the PSI group and 89.2 ± 2.1° and 88.9 ± 1.8° in the conventional group, respectively (Table 2). The number of outliers (> ± 3°) for the α angle and β angle was 9 (18%) and 6 (12%) in the PSI group and 6 (12%) and 10 (20%) in the conventional group, respectively (Table 2). No significant differences were found in these values.
Table 2.
PSI group (n = 50) | Conventional group (n = 50) | p value | |
---|---|---|---|
Mean value | |||
α angle | 88.8 ± 2.0 | 89.2 ± 2.1 | 0.16 |
β angle | 90.1 ± 2.4 | 88.9 ± 1.9 | 0.83 |
Outliers out of ± 3° | |||
α angle | 9 (18%) | 6 (12%) | 0.40 |
β angle | 6 (12%) | 10 (20%) | 0.28 |
Values are number (percentage) or mean ± standard deviation
CT-based assessment of rotational alignment
The mean value of femoral rotation was − 1.5 ± 2.5° in the PSI group and − 0.4 ± 3.4° in the conventional group (n.s, Table 3). The number of outliers (> ± 3°) for femoral rotation was 13 (26%) in the PSI group and 15 (30%) in the conventional group (n.s, Table 3). The mean value of tibial rotation from the preoperatively planned reference axis was − 2.6 ± 6.3° in the PSI group and − 4.7 ± 8.1° in the conventional group (n.s, Table 3). The number of outliers (< − 10°) for tibial rotation was 6 (12%) in the PSI group and 15 (30%) in the conventional group (p < 0.05, Table 3).
Table 3.
PSI group (n = 50) | Conventional group (n = 50) | p value | |
---|---|---|---|
Femoral rotation | |||
Mean value | − 1.5 ± 2.5 | − 0.4 ± 3.4 | 0.11 |
Outliers out of ± 3° | 13 (26%) | 15 (30%) | 0.66 |
Tibial rotation | |||
Mean value | − 2.6 ± 6.3 | − 4.7 ± 8.1 | 0.16 |
Outliers < − 10° | 6 (12%) | 15 (30%) | 0.027 |
Values are number (percentage) or mean ± standard deviation
Reproducibility of measurements
The ICCs for inter-rater reliabilities by two independent observers were 0.88 for the α angle, 0.95 for the β angle, 0.83 for femoral rotation, and 0.94 for tibial rotation. The ICCs for intra-rater reliabilities at more than 2-month intervals were 0.84 for the α angle, 0.92 for the β angle, 0.82 for femoral rotation, and 0.93 for tibial rotation.
Discussion
The most important finding of the present study was that PSI enabled the reduction of rotational malalignment of the tibial component, which was defined as internal rotation > 10°. However, no significant differences were found in the other parameters between PSI and the conventional technique. These findings partially confirm the initial hypothesis that PSI improves component positioning and reduces outliers compared to the conventional technique. The major advance in this study was the accurate measurement of the difference from the preoperative planning in tibial rotational alignment using three-dimensionally merged images with preoperative ones.
This study focused on evaluation of tibial rotational positioning, since it has been less reported than rotational alignment of the femoral component. Several studies evaluated tibial rotational alignment in the transverse plane of the bone resection level [4, 20–22], although tibial rotational measurements were changed among the different resection levels [23, 24]. Determination of the tibial AP axis based on the bone resection level seems to be unreliable as the rotational reference. In the present study, the tibial AP axis was determined using the preoperative CT image and then three-dimensionally merged with the postoperative CT image using the simulation software. This method allows the accurate measurement of the difference from the preoperative planning in tibial rotational alignment and may reduce the uncertainty in measurement of tibial rotation, although the measurement has complexity and diversity.
The definition of outliers is an important issue when evaluating the accuracy of component positioning in TKA. An outlier of coronal alignment was defined as a mechanical axis with more than 3° varus or valgus in most studies [7, 25, 26], and it seems to be generally accepted. In contrast, there is no agreed definition of the cutoff value for outliers in rotational alignment, especially tibial rotation. This may be due to wide variation in anatomical landmarks and measurement methodology [6]. Although several studies reported various cutoff values for tibial rotation affecting clinical outcomes [5, 27–29], a systematic review identified internal rotation > 10° in the tibial component as a negative prognostic factor [6]. On the basis of these previous reports, internal rotation > 10° was defined as a clinically relevant cutoff value for outliers for tibial rotation in the present study.
PSI has been becoming increasingly popular and is expected to improve surgical performance, including accuracy of component positioning and reduction of outliers. However, the superiority of PSI compared to the conventional technique has been controversial in several systematic reviews [8, 30, 31]. The majority of the reports comparing PSI and conventional technique showed no evidence of superiority in reducing outliers of component alignment, although there is a lack of information on the assessment of tibial rotational alignment [30, 31]. The present study demonstrated that no significant differences were found in coronal alignment, but there was a significant reduction of clinically relevant outliers (internal rotation > 10°) in tibial rotational alignment. The data support previous reports with regard to coronal alignment, but indicate the effectiveness of PSI in rotational positioning of the tibial component.
There are several types of PSI designs, and the procedures of tibial rotational positioning vary. Some PSIs do not directly guide the tibial component positioning, and the rotational positioning is determined in reference to extramedullary rod, AP marker pin, or rotational marker on the PSI. The other types of PSI directly determine the tibial rotational positioning through guide holes located on the tibial plateau, and this type of PSI was used in the present study. The latter types may reproduce the planned tibial rotational positioning better than the former ones, although there are no studies comparing between them.
This study has several limitations. First, this was not a randomized, controlled study but a retrospective, case–control study with an evidence level of III. Second, the patient population was small. Third, the data included only radiographic and 3D-CT measurements, and the relationship with clinical outcomes was not evaluated.
Conclusion
This study demonstrated that the difference from the preoperative planning in tibial rotational positioning was accurately evaluated using novel three-dimensional measurement method, and PSI could reduce outliers from the planned positioning in rotational alignment of the tibial component (internal rotation > 10°). PSI is a useful technique for improving the reproducibility of the planned tibial rotational positioning in TKA.
Acknowledgements
Not applicable.
Abbreviations
- TKA
Total knee arthroplasty
- PSI
Patient-specific instrumentation
- CT
Computed tomography
- AP
Anteroposterior
- ICCs
Intraclass correlation coefficients
Author contributions
MS and KK contributed to study design. MS, KK, SY, SN, and YI conducted the study. MS, KK, SY, and SN were involved in data collection. MS, KK, SY, SN, and YI contributed to data interpretation. : MS, KK, and YI helped in drafting manuscript. KK takes responsibility for the integrity of the data analysis. All authors read and approved the final manuscript.
Funding
The author(s) received no financial support for the research, authorship, and/or publication of this article.
Availability of data and materials
The data and materials used and/or analyzed during the current study are not publicly available but available from the corresponding author on reasonable request.
Declarations
Ethics approval and consent to participate
This study was approved by the ethics committee at Yokohama City University Hospital (#B B200500006). Informed consent was obtained from all participants included in the study.
Consent for publication
All authors agree to submit and publish the article.
Competing interests
The authors declare that they have no competing interests.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Fang DM, Ritter MA, Davis KE. Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty. 2009;24:39–43. doi: 10.1016/j.arth.2009.04.034. [DOI] [PubMed] [Google Scholar]
- 2.Ritter MA, Faris PM, Keating EM, Meding JB. Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res. 1994;299:153–156. doi: 10.1097/00003086-199402000-00021. [DOI] [PubMed] [Google Scholar]
- 3.Thiele K, Perka C, Matziolis G, Mayr HO, Sostheim M, Hube R. Current failure mechanisms after knee arthroplasty have changed: polyethylene wear is less common in revision surgery. J Bone Jt Surg Am. 2015;97:715–720. doi: 10.2106/JBJS.M.01534. [DOI] [PubMed] [Google Scholar]
- 4.Berger RA, Crossett LS, Jacobs JJ, Rubash HE. Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. 1998;356:144–153. doi: 10.1097/00003086-199811000-00021. [DOI] [PubMed] [Google Scholar]
- 5.Bell SW, Young P, Drury C, Smith J, Anthony I, Jones B, et al. Component rotational alignment in unexplained painful primary total knee arthroplasty. Knee. 2014;21:272–277. doi: 10.1016/j.knee.2012.09.011. [DOI] [PubMed] [Google Scholar]
- 6.Panni AS, Ascione F, Rossini M, Braile A, Corona K, Vasso M, et al. Tibial internal rotation negatively affects clinical outcomes in total knee arthroplasty: a systematic review. Knee Surg Sports Traumatol Arthrosc. 2018;26:1636–1644. doi: 10.1007/s00167-017-4823-0. [DOI] [PubMed] [Google Scholar]
- 7.Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K. Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty. 2007;22:1097–1106. doi: 10.1016/j.arth.2007.08.001. [DOI] [PubMed] [Google Scholar]
- 8.Gong S, Xu W, Wang R, Wang Z, Wang B, Han L, et al. Patient-specific instrumentation improved axial alignment of the femoral component, operative time and perioperative blood loss after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2019;27:1083–1095. doi: 10.1007/s00167-018-5256-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Moerenhout K, Allami B, Gkagkalis G, Guyen O, Jolles BM. Advantages of patient-specific cutting guides with disposable instrumentation in total knee arthroplasty: a case control study. J Orthop Surg Res. 2021;16:188. doi: 10.1186/s13018-021-02310-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Pauzenberger L, Munz M, Brandl G, Frank JK, Heuberer PR, Laky B, et al. Patient-specific instrumentation improved three-dimensional accuracy in total knee arthroplasty: a comparative radiographic analysis of 1257 total knee arthroplasties. J Orthop Surg Res. 2019;14:437. doi: 10.1186/s13018-019-1465-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Schotanus MGM, Thijs E, Heijmans M, Vos R, Kort NP. Favourable alignment outcomes with MRI-based patient-specific instruments in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2018;26:2659–2668. doi: 10.1007/s00167-017-4637-0. [DOI] [PubMed] [Google Scholar]
- 12.Tandogan RN, Kort NP, Ercin E, van Rooij F, Nover L, Saffarini M, et al. Computer-assisted surgery and patient-specific instrumentation improve the accuracy of tibial baseplate rotation in total knee arthroplasty compared to conventional instrumentation: a systematic review and meta-analysis. Knee Surg Sports Traumatol Arthrosc. 2021;30:2654–2665. doi: 10.1007/s00167-021-06495-x. [DOI] [PubMed] [Google Scholar]
- 13.Heyse TJ, Tibesku CO. Improved tibial component rotation in TKA using patient-specific instrumentation. Arch Orthop Trauma Surg. 2015;135:697–701. doi: 10.1007/s00402-015-2157-2. [DOI] [PubMed] [Google Scholar]
- 14.Silva A, Sampaio R, Pinto E. Patient-specific instrumentation improves tibial component rotation in TKA. Knee Surg Sports Traumatol Arthrosc. 2014;22:636–642. doi: 10.1007/s00167-013-2639-0. [DOI] [PubMed] [Google Scholar]
- 15.Boonen B, Schotanus MG, Kort NP. Preliminary experience with the patient-specific templating total knee arthroplasty. Acta Orthop. 2012;83:387–393. doi: 10.3109/17453674.2012.711700. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Kotela A, Kotela I. Patient-specific computed tomography based instrumentation in total knee arthroplasty: a prospective randomized controlled study. Int Orthop. 2014;38:2099–2107. doi: 10.1007/s00264-014-2399-6. [DOI] [PubMed] [Google Scholar]
- 17.Ewald FC. The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res. 1989;248:9–12. doi: 10.1097/00003086-198911000-00003. [DOI] [PubMed] [Google Scholar]
- 18.Roh YW, Kim TW, Lee S, Seong SC, Lee MC. Is TKA using patient-specific instruments comparable to conventional TKA? A randomized controlled study of one system. Clin Orthop Relat Res. 2013;471:3988–3995. doi: 10.1007/s11999-013-3206-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Fehring TK. Rotational malalignment of the femoral component in total knee arthroplasty. Clin Orthop Relat Res. 2000;380:72–79. doi: 10.1097/00003086-200011000-00010. [DOI] [PubMed] [Google Scholar]
- 20.Chanalithichai N, Tammachote N, Jitapunkul C, Kanitnate S. Rotational component alignment in patient-specific total knee arthroplasty compared with conventional cutting instrument. Eur J Orthop Surg Traumatol. 2019;29:1297–1304. doi: 10.1007/s00590-019-02423-5. [DOI] [PubMed] [Google Scholar]
- 21.Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C. A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation. J Bone Jt Surg Am. 2007;89:236–243. doi: 10.2106/JBJS.F.00386. [DOI] [PubMed] [Google Scholar]
- 22.Mizu-uchi H, Matsuda S, Miura H, Okazaki K, Akasaki Y, Iwamoto Y. The evaluation of post-operative alignment in total knee replacement using a CT-based navigation system. J Bone Jt Surg Br. 2008;90:1025–1031. doi: 10.1302/0301-620X.90B8.20265. [DOI] [PubMed] [Google Scholar]
- 23.Graw BP, Harris AH, Tripuraneni KR, Giori NJ. Rotational references for total knee arthroplasty tibial components change with level of resection. Clin Orthop Relat Res. 2010;468:2734–2738. doi: 10.1007/s11999-010-1330-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Sunnassee Y, Zhang H, Southern EP, Wang Y, Shen Y. Reliability of intra-articular rotational axes at standard tibial resection level and effect of resecting distally. J Knee Surg. 2015;28:223–228. doi: 10.1055/s-0034-1378193. [DOI] [PubMed] [Google Scholar]
- 25.Fu Y, Wang M, Liu Y, Fu Q. Alignment outcomes in navigated total knee arthroplasty: a meta-analysis. Knee Surg Sports Traumatol Arthrosc. 2012;20:1075–1082. doi: 10.1007/s00167-011-1695-6. [DOI] [PubMed] [Google Scholar]
- 26.Lee BS, Cho HI, Bin SI, Kim JM, Jo BK. Femoral component varus malposition is associated with tibial aseptic loosening after TKA. Clin Orthop Relat Res. 2018;476:400–407. doi: 10.1007/s11999.0000000000000012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Lutzner J, Kirschner S, Gunther KP, Harman MK. Patients with no functional improvement after total knee arthroplasty show different kinematics. Int Orthop. 2012;36:1841–1847. doi: 10.1007/s00264-012-1584-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Nicoll D, Rowley DI. Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Jt Surg Br. 2010;92:1238–1244. doi: 10.1302/0301-620X.92B9.23516. [DOI] [PubMed] [Google Scholar]
- 29.Thielemann FW, Konstantinids L, Herget GW, Knothe D, Helwig P, Sudkamp NP, et al. Effect of rotational component alignment on clinical outcome 5 to 7 years after TKA with the Columbus Knee System. Orthopedics. 2016;39:S50–S55. doi: 10.3928/01477447-20160509-17. [DOI] [PubMed] [Google Scholar]
- 30.Sassoon A, Nam D, Nunley R, Barrack R. Systematic review of patient-specific instrumentation in total knee arthroplasty: new but not improved. Clin Orthop Relat Res. 2015;473:151–158. doi: 10.1007/s11999-014-3804-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Zhang QM, Chen JY, Li H, Chai W, Ni M, Zhang ZD, et al. No evidence of superiority in reducing outliers of component alignment for patient-specific instrumentation for total knee arthroplasty: a systematic review. Orthop Surg. 2015;7:19–25. doi: 10.1111/os.12150. [DOI] [PMC free article] [PubMed] [Google Scholar]