Abstract
Background
Online adaptive stereotactic radiotherapy allows for improved target and organ at risk (OAR) delineation and inter-fraction motion management via daily adaptive planning. The use of adaptive SBRT for the treatment of pancreatic cancer (performed until now using only MRI or CT on rails-guided adaptive radiotherapy), has yielded promising outcomes. Herein we describe the first reported case of cone beam CT-guided stereotactic adaptive radiotherapy (CT-STAR) for the treatment of pancreatic cancer.
Case presentation
A 61-year-old female with metastatic pancreatic cancer presented for durable palliation of a symptomatic primary pancreatic mass. She was prescribed 35 Gy/5 fractions utilizing CT-STAR. The patient was simulated utilizing an end-exhale CT with intravenous and oral bowel contrast. Both initial as well as daily adapted plans were created adhering to a strict isotoxicity approach in which coverage was sacrificed to meet critical luminal gastrointestinal OAR hard constraints. Kilovoltage cone beam CTs were acquired on each day of treatment and the radiation oncologist edited OAR contours to reflect the patient’s anatomy-of-the-day. The initial and adapted plan were compared using dose volume histogram objectives, and the superior plan was delivered. Use of the initial treatment plan would have resulted in nine critical OAR hard constraint violations. The adapted plans achieved hard constraints in all five fractions for all four critical luminal gastrointestinal structures.
Conclusions
We report the successful treatment of a patient with pancreatic cancer treated with CT-STAR. Prior to this treatment, the delivery of ablative adaptive radiotherapy for pancreatic cancer was limited to clinics with MR-guided and CT-on-rails adaptive SBRT technology and workflows. CT-STAR is a promising modality with which to deliver stereotactic adaptive radiotherapy for pancreatic cancer.
Supplementary Information
The online version contains supplementary material available at 10.1186/s13014-022-02125-z.
Keywords: Pancreatic cancer, SBRT, Image guided radiation therapy, CT
Background
Pancreatic cancer is a lethal malignancy with a five-year overall survival rate of 2–10% [1–4]. In recent years, there has been an increased focus on the utilization of stereotactic body radiotherapy (SBRT) for the definitive treatment of pancreatic malignancies [2, 5, 6]. SBRT for pancreatic cancer is also critical in the palliative setting, as SBRT has been demonstrated to elicit durable local control and long-lasting relief of symptoms of local progression such as abdominal pain and gastric outlet obstruction [7–9]. However, the delivery of SBRT for pancreatic tumors is challenging given the close proximity of the mobile and radiosensitive luminal gastrointestinal tract [10]. Magnetic resonance imaging (MRI) guided radiotherapy has been shown to allow precise delineation of daily target and organ at risk (OAR) volumes, improving the efficacy of pancreatic SBRT while minimizing toxicity [11–13]. Recently, the implementation of daily online adaptive planning via stereotactic magnetic resonance guided adaptive radiotherapy (SMART) has yielded promising progression-free and overall survival rates as well as a favorable toxicity profile in the ablation of pancreatic cancer [4,14–16].
Recently, a novel ring gantry computed tomography (CT) based radiotherapy machine has been developed with a high-quality cone-beam CT capable of yielding high resolution on-board volumetric images and an artificial intelligence (AI) enhanced treatment planning system (TPS), which is capable of daily adaptive planning (ETHOS, Varian Medical Systems, Palo Alto, CA) [17–19]. The use of cone beam CT-guided adaptive radiotherapy for the clinical ablation of pancreatic cancer has not yet been described. Herein we describe the first reported treatment of a patient with pancreatic cancer using cone beam CT-guided stereotactic adaptive radiotherapy (CT-STAR), including a discussion of the workflow and dosimetric analysis of the treatment.
Case presentation
Patient presentation
A 61-year-old woman presented following an episode of abdominal pain due to acute pancreatitis. During the patient’s work up, a CT chest/abdomen/pelvis demonstrated a mass in the pancreatic body. Biopsy of the mass confirmed pancreatic adenocarcinoma. The patient met with medical oncology and was recommended neo-adjuvant systemic therapy but declined and pursued alternative therapies. The patient returned to clinic several months later with abdominal pain and interval imaging demonstrating progression of local disease with encasement of the splenic and superior mesenteric veins (Fig. 1) as well as the development of liver metastases. The primary mass measured 4.8 × 3.8 cm. The patient was referred to radiation oncology for consideration of palliative radiotherapy. On interview, the patient reported left upper quadrant and back pain as well as malaise and weight loss. Physical exam was otherwise unremarkable. The patient was recommended SBRT to her primary mass for durable palliation, 35 Gy in 5 fractions, 7 Gy per fraction. Given the high dose per fraction and adjacent critical organs at risk, the treating radiation oncologist elected to use daily online adaptation with cone beam CT-guidance.
Treatment planning and delivery
The patient was simulated utilizing an end-exhale breath-hold CT with intravenous and oral bowel contrast and a 4-dimensional CT. Intravenous contrast was administered at the 45-s delay phase per institutional protocol. The primary image used for planning was the end-exhale breath-hold CT. The 4D-CT is captured in the case that the patient is non-compliant with breath-hold and requires treatment with a different modality and/or dose and fractionation. Of note, as contrast is not delivered with each subsequent daily cone beam CT, the density of the contrast is overrided on the simulation CT to the density of water so that the contrast has no dosimetric impact on the initial plan (PI). The patient was positioned in a custom immobilization device with left arm down and right arm up, per institutional pancreatic SBRT practice. An MRI was obtained at time of simulation and fused to the simulation images for assistance in target delineation. All treatment planning was performed in the ETHOS (v.02.01.00) TPS. The gross tumor volume (GTV) comprised the gross tumor demonstrated on simulation imaging. As the patient was simulated and intended to be treated at end-exhale breath-hold, a internal GTV or internal target volume was not created. No clinical target volume (CTV) was utilized per standard institutional pancreatic SBRT practice. A 0.5 cm uniform volumetric expansion was applied to form a planning target volume (PTV). The relevant organs-at-risk (OARs) were contoured at the axial slices from 3 cm below to 3 cm above the PTV.
A PTV optimization (PTVopt) structure was generated, made from the PTV minus any overlap with critical OARs plus a 5 mm margin on the OARs. The critical OARs were the luminal gastrointestinal structures, namely the stomach, duodenum, small bowel, and large bowel. This PTVopt was used to drive prescription dose to the tumor to drive target coverage, given that areas of direct PTV and OAR overlap are not prioritized for target coverage per our standard adaptive radiotherapy practices [4, 20–22]. Both the PI and adaptive plans (PA) were generated using a strict isotoxicity approach, in which maximum OAR constraints are prioritized over target coverage [21, 23]. However, a minimum dose of 25 Gy was maintained to the PTV to ensure some uncertainty margin coverage. Dose constraints and objectives are in Table 1. Conservative luminal gastrointestinal OAR constraints were used given the palliative nature of the case. We have provided our standard departmental pancreatic adaptive SBRT dose constraints in Additional file 1: Table S1. A beam arrangement of two ¾ co-planar arcs was used, with 30 and 330 degree collimator angles.
Table 1.
Organ-at-risk | Strict constraint | PI mean (std dev) | PI median (range) | PA mean (std dev) | PA median (range) |
---|---|---|---|---|---|
Stomach | V25 Gy < 0.5 cc (cc) | 10.2 (3.6) | 9.7 (5.6–14.8) | 0.0 (0.1) | 0.0 (0.0–0.1) |
Duodenum | V25 Gy < 0.5 cc (cc) | 0.1 (0.1) | 0.1 (0.0–0.2) | 0.0 (0.0) | 0.0 (0.0–0.0) |
Small bowel | V25 Gy < 0.5 cc (cc) | 3.8 (5.1) | 1.2 (0.3–13.8) | 0.0 (0.0) | 0.0 (0.0–0.1) |
Large bowel | V25 Gy < 0.5 cc (cc) | 0.0 (0.0) | 0.0 (0.0–0.1) | 0.1 (0.1) | 0.0 (0.0–0.2) |
Liver | V25 Gy < 33% (%) | 9.5 (1.6) | 10.1 (6.8–11.4) | 12.1 (1.8) | 11.0 (10.3–14.7) |
700 cc < 20 Gy (Gy) | 0.4 (.1) | 0.3 (0.2–0.6) | 0.4 (0.1) | 0.4 (0.3–0.5) | |
Mean < 20 Gy (Gy) | 1.0 (0.2) | 1.0 (0.8–1.2) | 1.2 (0.1) | 1.2 (1.0–1.3) | |
Spinal cord | V25 Gy < 0.5 cc (cc) | 0.0 (0.0) | 0.0 (0.0–0.0) | 0.0 (0.0) | 0.0 (0.0–0.0) |
Kidneys (both) | Mean < 18 Gy (Gy) | 0.8 (0.7) | 0.8 (0.7–0.9) | 0.8 (0.1) | 0.8 (0.7–0.9) |
Target volume | Coverage goal | PI mean (std dev) | PI median (range) | PA mean (std dev) | PA median (range) |
---|---|---|---|---|---|
PTV V100 | N/A (%) | 77.1 (2.1) | 76.0 (74.5–80.0) | 77.8 (5.1) | 80.8 (67.9–81.3) |
PTV D95 | N/A (Gy) | 4.7 (0.1) | 4.6 (4.6–4.9) | 4.8 (0.1) | 4.8 (4.6–5.0) |
PTVopt | 95% (%) | 83.5 (3.8) | 83.5 (78.8–91.4) | 99.6 (0.3) | 99.5 (99.2–100.1) |
GTV V100 | N/A (%) | 89.9 (1.1) | 90.0 (88.5–91.4) | 90.0 (4.0) | 91.8 (82.2–93) |
GTV D95 | N/A (Gy) | 6.0 (0.2) | 6.1 (5.6–6.3) | 6.4 (0.3) | 6.5 (5.7–6.7) |
Mean and median constraint and target metrics for the PI represent the hypothetical use of the PI applied to all five fractions
N/A not applicable, Std Dev standard deviation
Daily PA were created based on the patient’s anatomy-of-the-day. The TPS automatically deformed the OAR and target contours from the PA onto the daily cone beam CT using a vendor supplied elastic deformation algorithm, and the TPS AI auto-adjusted the stomach, duodenum, and liver according to the anatomy-of-the-day. The deformed GTV was then overwritten and the simulation GTV was ridigly copied onto the patient’s anatomy-of-the-day. OARs within a 3-cm contour ring (per standard adaptive protocol [22]) were adjusted by the radiation oncologist in order to confirm accuracy. The initial simulation based treatment plan (PI) was projected on the patient anatomy-of-the-day at the same time that the re-optimized daily adapted plan (PA) was generated. The PI and PA were compared using dose volume histogram (DVH) objectives, and the superior plan that met all dosimetric goals was delivered. Of note, all acquired kV cone beam CTs were considered of sufficient quality for target and OAR delineation as well as daily adaptation per the treating radiation oncologist and medical physicist.
Dosimetric and clinical results
Constraint and coverage metrics for the PI and PA are demonstrated in Table 1. Mean PTV and GTV D95 for all five fractions was 23.25 Gy and 30.20 Gy in the PI and 24.11 Gy and 31.85 Gy in the PA, respectively. Dosimetric parameters, specifically the volume received 25 Gy (V25) and maximum dose (Dmax) for critical luminal gastrointestinal structures, are demonstrated in Figs. 2 and 3. The use of the PI would have resulted in violation of the stomach hard constaint in all five fractions, and violation of the small bowel constraint in four of five fractions (Fig. 2). The PA achieved hard constraints in all five fractions for all four critical luminal gastrointestinal structures. Figure 4 illustrates how the use of daily adaptive planning allowed for a specific radiotherapy fraction to achieve the small bowel hard constraint, where as delivery of the PI would have violated that constraint.
Treatment component times were recorded and are demonstrated in Table 2. Mean (standard deviation) total treatment time was 70 min (68.3–81.7) and treatment time decreased each consecutive fraction. The patient completed all five fractions of CT-STAR without issue. The patient ultimately progressed locally and distantly, and passed away several months after treatment.
Table 2.
Treatment component | Fraction 1 | Fraction 2 | Fraction 3 | Fraction 4 | Fraction 5 | Mean | Standard deviation |
---|---|---|---|---|---|---|---|
Patient setup | 18 | 9 | 17 | 10 | N/R | 13.5 | 4.7 |
CBCT time | 1 | 5 | 1 | 2 | 1 | 2 | 1.7 |
Contouring | 39 | 28 | 17 | 19 | 27 | 26 | 8.7 |
Plan re-optimization | 3 | 7 | 6 | 7 | 6 | 5.8 | 1.6 |
Plan review | < 1 | < 1 | < 1 | 1 | 1 | 0.4 | 0.5 |
Quality assurance | 8 | 3 | 6 | 2 | 2 | 4.2 | 2.7 |
Pre-treatment CBCT | 1 | 2 | 5 | 3 | 4 | 3 | 1.6 |
Beam delivery | 26 | 21 | 8 | 9 | 14 | 15.6 | 7.8 |
Patient exit | 1 | 4 | 1 | 7 | 5 | 3.6 | 2.6 |
Total | 86 | 79 | 63 | 62 | 60 | 70 | 11.7 |
Times are in minutes
N/R not recorded
Discussion and conclusions
Discussion
Herein we describe the first reported use of CT-STAR for the treatment of a patient with pancreatic cancer using a novel ring gantry device. These data demonstrate that the delivery of the PI would have led to nine critical OAR hard constraint violations across all five fractions, and that the daily PA met all critical OAR hard constraints in all five fractions. Furthermore, the use of daily adaptation improved PTVopt, GTV V100, and GTV D95 coverage (Table 1) while alleviating the hard constraint violations. With regards to workflow, the overall treatment times were within the range of previously described treatment times for daily adaptation, and the decreased time per each consecutive fraction suggests that treatment times decrease with increased patient/staff familiarity [22, 24].
The utility of adaptive stereotactic radiotherapy for the treatment of pancreatic cancer can not be understated. The effective ablation of pancreatic cancers requires the delivery of biologic effective dose of at least 100 Gy [10]. However, this is difficult to achieve as the pancreas is adjacent to several mobile and radiosensitive OARs. Initial studies evaluating the use of ablative doses of standard CT-guided stereotactic radiotherapy for the treatment of pancreatic cancer proved efficacious with regards to local control, but also displayed high rates of luminal gastrointestinal organ toxicity [25–27]. Adaptive radiotherapy can improve the therapeutic index of SBRT for pancreatic cancer. Recently, our institution published outcomes for patients with inoperable pancreatic cancer treated with SMART and demonstrated durable progression-free and overall survival rates as well as a favorable toxicity profile [4]. While these data are promising, it’s notable that their application is limited to MR-guided workflows. Prior to the advent of the ETHOS platform, adaptive SBRT for pancreatic cancer was limited to clinics with MR-guided or CT-on-rails workflows [4, 16, 28, 29].
While CT-STAR has the capacity to expand access to adaptive pancreatic SBRT, there are potential limitations of using a CBCT-guided platform instead of a MR-guided platform. The improved soft tissue contrast of MRI can be useful in pancreatic and abdominal contouring, which can be of particular importance when delineating gross organ invasion. In contrast with MR-guidance, gross organ invasion is challenging to delineate on cone beam CT. In our experience [4], approximately 10% of patients with locally advanced pancreatic cancer present with evidence of gross organ, and the use of CBCT-guided adaptive SBRT may be limited in that subset of patients. This may be of consideration when planning to install either a MR- or cone beam CT-guided adaptive platform.
Herein we demonstrate that stereotactic adaptive radiotherapy is able to be delivered on a cone beam CT-guided modality, which promises to increase access to adaptive pancreatic SBRT world wide. This case presentation demonstrates the potential for CT-STAR to provide a additional avenue for radiation oncologists to ablate pancreatic cancer.
Conclusions
CT-STAR is a viable modality for the delivery of adaptive stereotactic radiotherapy for the ablation of pancreatic cancer. Clinical trials are warranted to investigate the impact of this modality on overall and progression-free survival as well as toxicity.
Supplementary Information
Acknowledgements
None.
Abbreviations
- AI
Artificial intelligence
- CBCT
Cone beam computed tomography.
- CT
Computed tomography
- CT-STAR
CT-guided stereotactic adaptive radiotherapy
- CTV
Clinical target volume
- Dmax
Maximum dose recieved
- DVH
Dose volume histogram
- GTV
Gross tumor volume
- MRI
Magnetic resonance imaging
- OAR
Organ at risk
- PA
Adapted plan
- PI
Initial plan
- PTV
Planning target volume
- PTVopt
PTV optimization structure
- SBRT
Stereotactic body radiation therapy
- SMART
Stereotactic magnetic resonance guided adaptive radiotherapy
- TPS
Treatment planning system
- V25
Volume receiving 25 Gy
Author contributions
MK and JPS—the literature researching, writing the manuscript. SNB—preparing the radiotherapy plan. LEH—Conception, realization of the treatment. All authors read and approved the final manuscript.
Funding
Not applicable.
Availability of data and materials
Research data are stored in an institutional repository and will be shared upon request to the corresponding author.
Declarations
Ethics approval and consent to participate
The patient was given ample time to ask questions at time of consultation and she received answers for all questions raised. Consent for radiotherapy was obtained from the patient. All authors read and approved the final manuscript.
Consent for publication
The patient had passed away by the time this manuscript was prepared and therefore consent for publication was unable to be obtained.
Competing interests
Minsol Kim: none. Joshua P Schiff: none. Alex Price: grants: Varian Medical Systems; support for meetings: ViewRay Inc., Sun Nuclear Corporation. Eric Laugeman: honoraria: Varian Medical Systems. Pamela P Samson: none. Hyun Kim: grants: Varian Medical Systems; honoraria: ViewRay Inc., Varian Medical Systems. Shahed N Badiyan: honoraria: Mevion Medical Systems. Lauren E Henke: grants: Varian Medical Systems; consulting fees: Varian Medical Systems, Radialogica; honoraria: ViewRay Inc., Varian Medical Systems; Advisory Board: ViewRay Inc.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Minsol Kim and Joshua P. Schiff should be noted as co-first authors
Contributor Information
Joshua P. Schiff, Email: j.p.schiff@wustl.edu
Lauren E. Henke, Email: henke.lauren@wustl.edu
References
- 1.Rawla P, Sunkara T, Gaduputi V. Epidemiology of pancreatic cancer: global trends, etiology and risk factors. World J Oncol. 2019;10(1):10–27. doi: 10.14740/wjon1166. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Abi Jaoude J, Thunshelle CP, Kouzy R, et al. Stereotactic versus conventional radiation therapy for patients with pancreatic cancer in the modern era. Adv Radiat Oncol. 2021;6(6):100763. doi: 10.1016/j.adro.2021.100763. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.McGuigan A, Kelly P, Turkington RC, Jones C, Coleman HG, McCain RS. Pancreatic cancer: a review of clinical diagnosis, epidemiology, treatment and outcomes. World J Gastroenterol. 2018;24(43):4846–4861. doi: 10.3748/wjg.v24.i43.4846. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Hassanzadeh C, Rudra S, Bommireddy A, et al. Ablative five-fraction stereotactic body radiation therapy for inoperable pancreatic cancer using online MR-guided adaptation. Adv Radiat Oncol. 2021;6(1):100506. doi: 10.1016/j.adro.2020.06.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Bourhis J, Sire C, Graff P, et al. Concomitant chemoradiotherapy versus acceleration of radiotherapy with or without concomitant chemotherapy in locally advanced head and neck carcinoma (GORTEC 99–02): an open-label phase 3 randomised trial. Lancet Oncol. 2012;13(2):145–153. doi: 10.1016/S1470-2045(11)70346-1. [DOI] [PubMed] [Google Scholar]
- 6.Kishan A, Lee P. Having your cake and eating it too: combining SBRT and multi-agent chemotherapy in locally advanced pancreatic cancer. Cureus. 2016;8(7):e686. doi: 10.7759/cureus.686. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Vornhülz M, Anton S, Eross B, et al. Role of stereotactic body radiation in the enhancement of the quality of life in locally advanced pancreatic adenocarcinoma: a systematic review. Radiat Oncol. 2022;17(1):108. doi: 10.1186/s13014-022-02076-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Ryan JF, Rosati LM, Groot VP, et al. Stereotactic body radiation therapy for palliative management of pancreatic adenocarcinoma in elderly and medically inoperable patients. Oncotarget. 2018;9(23):16427–16436. doi: 10.18632/oncotarget.24713. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Buwenge M, Macchia G, Arcelli A, et al. Stereotactic radiotherapy of pancreatic cancer: a systematic review on pain relief. J Pain Res. 2018;11:2169–2178. doi: 10.2147/JPR.S167994. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Reyngold M, Parikh P, Crane CH. Ablative radiation therapy for locally advanced pancreatic cancer: techniques and results. Radiat Oncol. 2019;14:95. doi: 10.1186/s13014-019-1309-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Tyagi N, Liang J, Burleson S, Subashi E, Scripes PG, Tringale KR, Romesser PB, Reyngold M, Crane CH. Feasibility of ablative stereotactic body radiation therapy of pancreas cancer patients on a 1.5 Tesla magnetic resonance-linac system using abdominal compression. Phys Imaging Radiat Oncol. 2021;19:53–59. doi: 10.1016/j.phro.2021.07.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Heerkens HD, van Vulpen M, Erickson B, et al. MRI guided stereotactic radiotherapy for locally advanced pancreatic cancer. Br J Radiol. 2018;91(1091):20170563. doi: 10.1259/bjr.20170563. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Tchelebi LT, Zaorsky NG, Rosenberg JC, et al. Reducing the toxicity of radiotherapy for pancreatic cancer with magnetic resonance-guided radiotherapy. Toxicol Sci. 2020;175(1):19–23. doi: 10.1093/toxsci/kfaa021. [DOI] [PubMed] [Google Scholar]
- 14.Doty DG, Chuong MD, Gomez AG, et al. Stereotactic MR-guided online adaptive radiotherapy reirradiation (SMART reRT) for locally recurrent pancreatic adenocarcinoma: a case report. Med Dosim. 2021;46(4):384–388. doi: 10.1016/j.meddos.2021.04.006. [DOI] [PubMed] [Google Scholar]
- 15.Magallon-Baro A, Milder MTW, Granton PV, Nuyttens JJ, Hoogeman MS. Comparison of daily online plan adaptation strategies for a cohort of pancreatic cancer patients treated with SBRT. Int J Radiat Oncol Biol Phys. 2021;111(1):208–219. doi: 10.1016/j.ijrobp.2021.03.050. [DOI] [PubMed] [Google Scholar]
- 16.Chuong MD, Bryant J, Mittauer KE, et al. Ablative 5-fraction stereotactic magnetic resonance-guided radiation therapy with on-table adaptive replanning and elective nodal irradiation for inoperable pancreas cancer. Pract Radiat Oncol. 2021;11(2):134–147. doi: 10.1016/j.prro.2020.09.005. [DOI] [PubMed] [Google Scholar]
- 17.Pokharel S, Pacheco A, Tanner S. Assessment of efficacy in automated plan generation for Varian Ethos intelligent optimization engine. J Appl Clin Med Phys. 2022 doi: 10.1002/acm2.13539. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Moazzezi M, Rose B, Kisling K, Moore KL, Ray X. Prospects for daily online adaptive radiotherapy via ethos for prostate cancer patients without nodal involvement using unedited CBCT auto-segmentation. J Appl Clin Med Phys. 2021;22(10):82–93. doi: 10.1002/acm2.13399. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Hao Y, Cai B, Green O, et al. Technical Note: An alternative approach to verify 6FFF beam dosimetry for Ethos and MR Linac without using a 3D water tank. Med Phys. 2021;48(4):1533–1539. doi: 10.1002/mp.14757. [DOI] [PubMed] [Google Scholar]
- 20.Henke L, Kashani R, Robinson C, et al. Phase I trial of stereotactic MR-guided online adaptive radiation therapy (SMART) for the treatment of oligometastatic or unresectable primary malignancies of the abdomen. Radiother Oncol. 2018;126(3):519–526. doi: 10.1016/j.radonc.2017.11.032. [DOI] [PubMed] [Google Scholar]
- 21.Green OL, Henke LE, Hugo GD. Practical clinical workflows for online and offline adaptive radiation therapy. Semin Radiat Oncol. 2019;29(3):219–227. doi: 10.1016/j.semradonc.2019.02.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Bohoudi O, Bruynzeel AME, Senan S, et al. Fast and robust online adaptive planning in stereotactic MR-guided adaptive radiation therapy (SMART) for pancreatic cancer. Radiother Oncol. 2017;125(3):439–444. doi: 10.1016/j.radonc.2017.07.028. [DOI] [PubMed] [Google Scholar]
- 23.Henke L, Kashani R, Yang D, et al. Simulated online adaptive magnetic resonance-guided stereotactic body radiation therapy for the treatment of oligometastatic disease of the abdomen and central thorax: characterization of potential advantages. Int J Radiat Oncol Biol Phys. 2016;96(5):1078–1086. doi: 10.1016/j.ijrobp.2016.08.036. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Henke LE, Stanley JA, Robinson C, et al. Phase I trial of stereotactic MRI-guided online adaptive radiation therapy (SMART) for the treatment of oligometastatic ovarian cancer. Int J Radiat Oncol Biol Phys. 2022;112(2):379–389. doi: 10.1016/j.ijrobp.2021.08.033. [DOI] [PubMed] [Google Scholar]
- 25.Pollom EL, Alagappan M, von Eyben R, et al. Single- versus multifraction stereotactic body radiation therapy for pancreatic adenocarcinoma: outcomes and toxicity. Int J Radiat Oncol Biol Phys. 2014;90(4):918–925. doi: 10.1016/j.ijrobp.2014.06.066. [DOI] [PubMed] [Google Scholar]
- 26.Koong AJ, Toesca DAS, Baclay JRM, et al. The utility of stereotactic ablative radiation therapy for palliation of metastatic pancreatic adenocarcinoma. Pract Radiat Oncol. 2020;10(4):274–281. doi: 10.1016/j.prro.2020.02.010. [DOI] [PubMed] [Google Scholar]
- 27.Schellenberg D, Kim J, Christman-Skieller C, et al. Single-fraction stereotactic body radiation therapy and sequential gemcitabine for the treatment of locally advanced pancreatic cancer. Int J Radiat Oncol Biol Phys. 2011;81(1):181–188. doi: 10.1016/j.ijrobp.2010.05.006. [DOI] [PubMed] [Google Scholar]
- 28.Niedzielski JS, Liu Y, Ng SSW, et al. Dosimetric uncertainties resulting from interfractional anatomic variations for patients receiving pancreas stereotactic body radiation therapy and cone beam computed tomography image guidance. Int J Radiat Oncol Biol Phys. 2021;111(5):1298–1309. doi: 10.1016/j.ijrobp.2021.08.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Magallon-Baro A, Milder MTW, Granton PV, den Toom W, Nuyttens JJ, Hoogeman MS. Impact of using unedited CT-based DIR-propagated autocontours on online ART for pancreatic SBRT. Front Oncol. 2022;12:910792. doi: 10.3389/fonc.2022.910792. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Research data are stored in an institutional repository and will be shared upon request to the corresponding author.