Skip to main content
Research and Practice in Thrombosis and Haemostasis logoLink to Research and Practice in Thrombosis and Haemostasis
letter
. 2022 Sep 20;6(6):e12772. doi: 10.1002/rth2.12772

Standardization of D‐dimer reporting in the COVID‐19 era

Litao Zhang 1,2, Zhenlu Zhang 1,2,
PMCID: PMC9487443  PMID: 36186103

D‐dimer test, one of the most common tests for coagulation, plays an important role in management of patients with thrombosis, and its demand has risen sharply during the current global COVID‐19 epidemic. However, some main limitations, including inconsistent reporting units, various thresholds for different assays, lack of standardization, and poor harmonization, 1 have followed its widespread application and may cause confusion and misinformation. 2 Recently, a communication from the ISTH Scientific and Standardization Subcommittee (SSC) on Fibrinolysis reported by Bevan and Longstaff described a potential method to generate a stable standard material for D‐dimer. 3 This is quite important for standardization of D‐dimer testing; however, future investigations are still needed for confirmation. What can we do to improve the standardization of D‐dimer results before achieving the D‐dimer assay standardization? Based on recent evidence and the authors' experience, we have two proposals in an unofficial position.

First, reporting the D‐dimer ratio (DDR) would be a step forward in standardizing D‐dimer reporting. DDR means the ratio of the D‐dimer value to the upper limit of the normal range (ULN) for the current D‐dimer assay (DDR = D‐dimer/ULN; e.g., DDR = 2.0 for a D‐dimer = 1000 ng/ml with 0–500 ng/ml of normal range). DDR can show directly a proportional level of D‐dimer elevation, and is independent of type of unit used, and also accounts for the cutoff value used. This is a simple and helpful transformation to improve the comparability among various D‐dimer detection methods. We carried out a pilot study to test D‐dimer values of 12 samples using four common commercial D‐dimer assays simultaneously (Table 1). The original D‐dimer values from the same sample analyzed with four different assays showed a composite of ~500‐fold differences, compared to the differences of no more than 2‐fold in DDR values (D‐dimer value vs. DDR, 513.8 ± 97.4 vs 1.46 ± 0.21; see Table 1 for calculation details). Apparently, Reporting DDR could potentially remove the heterogeneity of the D‐dimer results reported from different assays.

TABLE 1.

D‐dimer and DDR results of 12 samples on four commercial assays

D‐dimer value D‐dimer ratio
Assay 1 a Assay 2 b Assay 3 c Assay 4 d Maximum difference e Assay 1 Assay 2 Assay 3 Assay 4 Maximum difference
ULN 243 0.5 1 0.5
Report unit ng/mL mg/L μg/ml μg/ml
Express unit DDU FEU DDU FEU
Sample 1 130 0.32 0.60 0.27 481.5 0.53 0.64 0.6 0.54 1.21
Sample 2 132 0.25 0.96 0.34 528.0 0.54 0.50 0.96 0.68 1.92
Sample 3 153 0.26 0.78 0.37 588.5 0.63 0.52 0.78 0.74 1.50
Sample 4 254 0.64 1.02 0.64 396.9 1.04 1.28 1.02 1.28 1.25
Sample 5 755 2.05 2.87 1.97 383.2 3.11 4.10 2.87 3.94 1.43
Sample 6 967 2.52 3.83 2.55 383.7 3.98 5.04 3.83 5.10 1.33
Sample 7 1030 1.82 3.70 1.59 647.8 4.24 3.64 3.70 3.18 1.33
Sample 8 1525 3.07 5.33 3.34 496.7 6.28 6.14 5.33 6.68 1.25
Sample 9 2091 4.67 5.97 4.01 521.4 8.61 9.34 5.97 8.02 1.56
Sample 10 2772 6.18 7.40 4.97 557.7 11.41 12.36 7.40 9.94 1.67
Sample 11 2891 5.83 8.07 5.94 495.9 11.90 11.66 8.07 11.88 1.47
Sample 12 5095 7.44 13.38 8.01 684.8 20.97 14.88 13.38 16.02 1.57
Total 513.8 ± 97.4 1.46 ± 0.21

Abbreviations: DDU, D‐dimer unit; FEU, fibrinogen equivalent unit; ULN, the upper limit of the normal range.

a

Assay 1, Werfen ACL‐TOP 750 with HemosIL D‐Dimer HS reagent;

b

Assay 2, Sysmex CS5100 with Siemens Innovance D‐dimer reagent;

c

Assay 3, Sekisui CP3000 with Nanopia D‐dimer reagent;

d

Assay 4, Stago STA‐R MAX with STA Liatest D‐Di reagent.

e

Maximum difference means that the maximum value of four results from the same sample divided by the minimum; e.g., Sample 12: 5095 divided by 7.44 is 684.8 for D‐dimer value, and 20.97 divided by 13.38 is 1.57 for D‐dimer ratio.

More importantly, DDR has been used and accepted by more and more clinicians in clinical practice, especially in the management of COVID‐19. D‐dimer was measured in 28 hospitals from five countries in the RAPID study, 4 a randomized controlled trial that compared the effects of therapeutic heparin with prophylactic heparin among moderately ill patients with COVID‐19. The investigators employed D‐dimer ratios to make the results comparable across sites since they used different types of D‐dimer assays. Similarly, another large multicenter clinical trial (ATTACC, ACTIV‐4a, and REMAP‐CAP) performed at 121 sites in nine countries to evaluate anticoagulation with heparin in patients with COVID‐19 also elected to use the DDR instead of absolute D‐dimer values to stratify the patients in the study cohorts. 5 , 6

Second, confusion exists between the fibrinogen equivalent unit (FEU) and D‐dimer unit (DDU), which the manufacturers use to report D‐dimer levels based on the molecular weight of fibrinogen and D‐dimer, respectively. 7 , 8 The following points further expand on the difference and difficulties pertaining to FEU and DDU.

  1. The only difference between the two units is a difference in molecular weight but has resulted in different normal reference range cutoffs between medical centers with resultant confusion and uncertainty in medical professionals as discussed above. 2 An official communication from the ISTH SSC on fibrinolysis (Thachil J et al.) recommended that “a standardized measuring units should be used for reporting patient results and suggested FEUs either in μg/L or mg/L.” 8

  2. Age‐adjusted D‐dimer cutoffs have been demonstrated to effectively improve the rule‐out performance of the D‐dimer tests in elderly patients with suspected venous thrombosis. Guidelines propose that an age‐adjusted D‐dimer cutoff value instead of fixed cutoff values 9 should be implemented. The age‐adjusted D‐dimer cutoff calculation is “Age × 10 mg/L,” for patients older than 50 years. A couple of studies in this regard used DDUs. 10 , 11

  3. There is evidence that more laboratories are reporting D‐dimers as FEUs versus DDUs,12 which further supports migration toward reporting D‐dimers in FEU units. It is in this background that we propose exclusive use of FEUs instead of DDUs to report D‐dimer results.

Such a change in the units of D‐dimer reporting will not be without stumbling blocks. For example, the manufacturers might be hesitant to change product information sheets that have been cleared or approved for use by regulatory authorities. Lippi et al. 12 proposed that international standardization societies such as the ISTH or the International Federation of Clinical Chemistry and Laboratory Medicine should take the lead in pursuing the challenging undertaking of standardization of the D‐dimer results reporting. 12 The authors are of the opinion that consistent reporting of D‐dimer tests in FEU will be welcomed by clinicians, research units, standardization societies, and medical educators.

In conclusion, it might be the right time to act to improve the standardization of D‐dimer reporting and help laboratories and clinicians better use the D‐dimer tests.

AUTHOR CONTRIBUTIONS

LZ and ZZ drafted the manuscript.

FUNDING INFORMATION

This work was supported by the Hubei Provincial Nature Science Foundation of China (2020CFB865) and 2020 Wuhan Young & Middle‐age Medical Backbone Training Programme.

RELATIONSHIP DISCLOSURE

The authors declare that they have no conflicts of interest regarding this article.

Handling Editor: Dr Johnny Mahlangu

REFERENCES

  • 1. Thachil J, Lippi G, Favaloro EJ. D‐dimer testing: laboratory aspects and current issues. Methods Mol Biol. 2017;1646:91‐104. [DOI] [PubMed] [Google Scholar]
  • 2. Favaloro EJ, Thachil J. Reporting of D‐dimer data in COVID‐19: some confusion and potential for misinformation. Clin Chem Lab Med. 2020;58:1191‐1199. [DOI] [PubMed] [Google Scholar]
  • 3. Bevan S, Longstaff C. Is it possible to make a common reference standard for D‐dimer measurements? Communication from the ISTH SSC subcommittee on fibrinolysis. J Thromb Haemost. 2022;20:498‐507. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Sholzberg M, Tang GH, Rahhal H, et al. Effectiveness of therapeutic heparin versus prophylactic heparin on death, mechanical ventilation, or intensive care unit admission in moderately ill patients with covid‐19 admitted to hospital: RAPID randomised clinical trial. BMJ. 2021;375:n2400. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. REMAP‐CAP Investigators , ACTIV‐4a Investigators , ATTACC Investigators , et al. Therapeutic anticoagulation with heparin in noncritically ill patients with Covid‐19. N Engl J Med. 2021;385:790‐802. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. REMAP‐CAP Investigators , ACTIV‐4a Investigators , ATTACC Investigators , et al. Therapeutic anticoagulation with heparin in critically ill patients with Covid‐19. N Engl J Med. 2021;385:777‐789. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Institute Clinical and Laboratory Standards . Quantitative D‐Dimer for the Exclusion of Venous Thromboembolic Disease: Approved Guideline CLSI Document H59‐A. CLSI; 2011. [Google Scholar]
  • 8. Thachil J, Longstaff C, Favaloro EJ, et al. The need for accurate D‐dimer reporting in COVID‐19: communication from the ISTH SSC on fibrinolysis. J Thromb Haemost. 2020;18:2408‐2411. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Konstantinides SV, Meyer G, Becattini C, et al. 2019 ESC guidelines for the diagnosis and management of acute pulmonary embolism developed in collaboration with the European Respiratory Society (ERS). Eur Heart J. 2020;41:543‐603. [DOI] [PubMed] [Google Scholar]
  • 10. Cini M, Legnani C, Frascaro M, Sartori M, Cosmi B, Palareti G. D‐dimer use for deep venous thrombosis exclusion in elderly patients: a comparative analysis of three different approaches to establish cut‐off values for an assay with results expressed in D‐dimer units. Int J Lab Hematol. 2014;36:541‐547. [DOI] [PubMed] [Google Scholar]
  • 11. Parks C, Bounds R, Davis B, Caplan R, Laughery T, Zeserson E. Investigation of age‐adjusted D‐dimer using an uncommon assay. Am J Emerg Med. 2019;37:1285‐1288. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Lippi G, Tripodi A, Simundic AM, Favaloro EJ. International survey on D‐dimer test reporting: a call for standardization. Semin Thromb Hemost. 2015;41:287‐293. [DOI] [PubMed] [Google Scholar]

Articles from Research and Practice in Thrombosis and Haemostasis are provided here courtesy of Elsevier

RESOURCES