Skip to main content
PLOS One logoLink to PLOS One
. 2022 Jul 19;17(7):e0271673. doi: 10.1371/journal.pone.0271673

Short communication: A comparison between two glucose measurement methods in beef steers during a glucose tolerance test

Kirsten R Nickles 1,#, Alejandro E Relling 1,, Alvaro Garcia-Guerra 2,, Francis L Fluharty 3,, Anthony J Parker 1,*,#
Editor: Marcio Duarte4
PMCID: PMC9491837  PMID: 35853033

Abstract

Glucose tolerance tests (GTT) are commonly performed in beef cattle to evaluate the glucose-insulin signaling pathway. Blood samples are obtained via a catheter and then transferred back to the laboratory for further analysis. A hand-held glucometer used chute-side can make performing GTT’s and quantifying blood glucose concentration much easier and faster for research purposes. The purpose of this study was to evaluate the agreement between a hand-held electronic glucometer (Precision Xtra; Abbott Diabetes Care Inc., Mississauga, ON, Canada) for chute-side use in beef cattle compared with a colorimetric assay in the laboratory (Stanbio Glucose LiquiColor; Stanbio Laboratory, Boerne, TX, USA). A GTT was performed on 13 Simmental × Angus steers during the growing phase. Blood samples were obtained via a jugular catheter. Glucometer readings were taken immediately after blood was sampled from the jugular with no preservative, and laboratory measurements were conducted on plasma preserved with sodium fluoride. A paired t-test (P = 0.40), Pearson’s correlation (P < 0.001; r = 0.95), Bland-Altman plot, and Lin’s concordance correlation coefficient (LCCC = 0.90) were completed to evaluate the performance of the glucometer relative to the results from the laboratory assay. Based on the results, we conclude that the glucometer is an acceptable method for measuring blood glucose concentration in beef cattle under field conditions.

Introduction

Glucose tolerance tests (GTT) are commonly used in beef cattle research as an indicator of insulin resistance [1]. This procedure has been modified by several research groups, but generally consists of an intravenous bolus infusion of 50% glucose followed by continuous blood sampling to monitor glucose and insulin concentration. The standard method of measuring glucose in plasma is with a colorimetric assay to quantify glucose concentration. The Stanbio assay used in our laboratory is based on the glucose oxidase methodology adapted by Trinder et al. [2]. In this method, glucose is oxidized in the presence of glucose oxidase. After hydrogen peroxide is formed and reacts with phenol and 4-aminoantipyrine, a red-violet quinone complex forms. The intensity of the red-violet color is directly proportional to the glucose concentration (Stanbio Laboratory, Boerne, TX, USA). Though the laboratory method is the standard, it is more expensive and time consuming compared with the hand-held glucometer.

An alternative method is a hand-held electronic glucose measuring system designed for use in humans. This system uses electrochemical test strips in which blood is applied after the test strip is inserted into the glucometer. The blood is then drawn up the test strip via capillary action. Once it is in the glucometer, it reacts with glucose oxidase and forms gluconic acid. The gluconic acid then reacts with the test strip electrodes and creates an electrical current that is proportional to the concentration of glucose in the blood. This hand-held system has been validated for use in measuring glucose and β-hydroxybutyrate in dairy cows [36], however, this has not been validated for measuring glucose chute-side when performing a GTT in beef cattle. It is possible that beef cattle may have different response curves to a GTT (i.e. peak plasma glucose concentration, baseline plasma glucose concentration) and thus plasma glucose concentrations compared with dairy cattle, as it has been previously reported that plasma glucose and insulin concentrations were different between beef and dairy cows at the same stage of lactation [7].

The objective of the present study was to compare the glucose concentration using whole blood and the hand-held glucometer compared with the standard laboratory assay using plasma from samples obtained while completing a GTT on 13 Simmental × Angus steers during the growing phase. We hypothesized that the two different methods would show acceptable agreeance and that the hand-held meter would be a suitable method for quantifying glucose concentration chute-side.

Materials and methods

All procedures were approved by The Ohio State University Institutional Animal Care and Use Committee (Animal Use Protocol # 2019A00000142).

Animals and procedures

Thirteen Simmental × Angus steer calves were used for this glucose quantification comparison. Steers were trained in the chute 5 days/week for two weeks to allow for the steers to become acclimated to standing in the chute and being touched during the GTT.

Steers were fasted for 24 hours before the GTT. The morning of the GTT, steers were weighed to determine bolus size (0.25 g of glucose/kg BW delivered in a 50% weight/volume dextrose solution. After being weighed, jugular catheters were placed in the steers, and then steers were returned to their pen and allowed a one hour rest period before the GTT began. Blood samples were collected at 5 and 2 minutes before administration of the glucose bolus to determine fasted plasma glucose concentration. Subsequent blood samples were collected immediately after glucose bolus infusion (0 minutes), 5, 10, 15, 20, 30, 60, and 120 minutes after glucose bolus infusion. Before and after each 10 mL blood sample was collected, the catheter line was flushed with 4–5 mL of sterile heparinized saline (9 g/L of NaCl). All blood samples were transferred to a tube containing sodium fluoride and then immediately placed on ice. The sodium fluoride tubes were transferred back to the laboratory and centrifuged for 25 min at 2500 x g and 4°C. The plasma was then further aliquoted into individual microcentrifuge tubes to determine plasma glucose concentrations at a later date.

Glucometer method

As blood was being sampled via the jugular catheter for the GTT, blood was immediately placed on the glucometer test strip to determine whole blood glucose concentration. The glucometer was used according to the label descriptions and directions of the manufacturer and is reported to measure blood glucose concentrations from 20–500 mg/dL (Precision Xtra; Abbott Diabetes Care Inc., Mississauga, ON, Canada). According to the user’s manual, if a sample reads “LO”, the meter has determined that the blood glucose concentration is below 20 mg/dL, and a sample reads “HI”, the meter has determined that the blood glucose concentration is above 500 mg/dL. One sample that was measured with the glucometer returned a “HI” result, and the glucose concentration was recorded as 500 mg/dL as recommended by the glucometer’s user manual. All samples measured with the glucometer were only measured once.

Laboratory trinder method

The laboratory analysis was completed using a colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure, Stanbio Laboratory, Boerne, TX, USA). Any sample that was outside of the linear portion of the standard curve was diluted with a 1:2 dilution. All samples were run in duplicates and the intra- and inter-assay coefficient of variations were 2.93% and 3.00%, respectively.

Statistical analysis

First, using the raw measurements from the glucometer and the average of the duplicate laboratory samples, a paired t-test was completed using the TTEST procedure of SAS (SAS 9.4). Additionally, a Pearson’s Correlation was completed for the two methods using the CORR procedure of SAS. For both of these models, data was assessed for normality using the residuals panel in SAS. When plotted, the differences between the pairs of observations were approximately normally distributed. Next, the Bland-Altman [8] approach was used to plot the difference between the two measurements against their mean to determine the two approaches’ agreement. Bland and Altman [8] and Petrie and Watson [9] recommend first performing a paired t-test to test the null hypothesis that the mean of the differences between the two methods is zero, and that the differences are evenly scattered above and below zero. The paired t-test determined if there was evidence of a systematic difference between the hand-held glucometer and the laboratory assay. The next step is to perform a Pearson correlation, however, Bland and Altman [8] caution that a Pearson correlation coefficient only gives indication of how close the observations in the scatter diagram are to a straight line and do not assess agreement. To assess agreement, one needs to know how close the points are to the line of perfect agreement (the 45° line through the origin). The Bland-Altman plot determines the limits (± 1.96 standard deviations) within which 95% of the differences are expected to lie when the difference between the two measurements in a pair are plotted against their mean. If there is no evidence of a systematic effect, the points should be scattered evenly above and below the line corresponding to a zero difference. If the variability of the differences is not constant (i.e. funnel shape), Petrie and Watson [9] recommend transforming the data and repeating the process. If there is no evidence of a systematic effect in the Bland-Altman plot of either the raw or transformed data, the next step is to complete an index of agreement which can either be the intraclass correlation coefficient (ICC) or the Lin’s concordance correlation coefficient (LCCC). These two indexes are similar and describe the closeness of the points to the line of perfect agreement and can be used to assess agreement because both accuracy and precision are incorporated. A Lin’s concordance correlation coefficient was calculated to assess agreement between the two methods.

Results

The paired t-test indicated no evidence of a systematic difference between the two methods of measurement with a test statistic of -0.86, (P = 0.40). A Pearson correlation coefficient for the two methods of measurement was then completed on the raw data (Fig 1; P < 0.001; r = 0.95), and indicated precision (i.e. the random variation describing the tightness of the points about the best-fitting straight line) between the two methods.

Fig 1. Pearson correlation coefficient (P < 0.001, r = 0.95) of the raw data (N = 130 samples from 13 steers), with the glucose concentration as measured by the glucometer on the y-axis (Precision Xtra; Abbott Diabetes Care, Inc., Mississauga, ON, Canada) plotted against the plasma preserved in sodium fluoride tubes and analyzed in a laboratory with a colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure, Stanbio Laboratory, Boerne, TX, USA) on the x-axis.

Fig 1

A Bland-Altman plot was completed for the raw data between the two measurements (Fig 2), with the mean concentration measured by the glucometer plotted against the difference in mean concentration between the two methods. The raw data demonstrated a funnel shape, indicating that the variability of the differences was not constant as the mean of the two measurements increased. Therefore, the raw data was log transformed as recommended by Petrie and Watson [9]. The transformed Bland-Altman plot is shown in Fig 3. Transforming the data eliminated the funnel effect, and the points are evenly scattered above and below the line representing the mean, corresponding to no systematic difference between the two methods. Additionally, since the scatter of the points is random with no funnel effect, we can conclude that the size of the discrepancy between the two methods is not related to the magnitude of the count. In Figs 2 and 3, the dashed lines represent the upper and lower limits of agreement. We expect 95% of the absolute differences to be less than the upper and lower limits of agreement, which is the case for the transformed data.

Fig 2. Bland-Altman plot of the raw data (N = 130 samples from 13 steers), with the mean glucose concentration as measured by the glucometer (Precision Xtra; Abbott Diabetes Care, Inc., Mississauga, ON, Canada) and the plasma preserved in sodium fluoride tubes and analyzed in a laboratory with a colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure, Stanbio Laboratory, Boerne, TX, USA) on the x-axis, plotted against the difference in glucose concentrations determined by the glucometer and the laboratory on the y-axis.

Fig 2

The mean difference is represented by the solid line (mean = 4.6) and the 95% confidence limits are represented by the dashed lines.

Fig 3.

Fig 3

Bland-Altman plot of the log transformed data (N = 130 samples from 13 steers), with the mean glucose concentration as measured by the glucometer (Precision Xtra; Abbott Diabetes Care, Inc., Mississauga, ON, Canada) and the plasma preserved in sodium fluoride tubes and analyzed in a laboratory with a colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure, Stanbio Laboratory, Boerne, TX, USA) on the x-axis, plotted against the difference in glucose concentrations determined by the glucometer and the laboratory on the y-axis. The mean difference is represented by the solid line (mean = 0.03) and the 95% confidence limits are represented by the dashed lines.

In addition, since there is no evidence of a systematic effect, we can estimate the Lin’s concordance correlation coefficient. For this correlation coefficient, a perfect agreement is achieved when the value is equal to 1, and there is no agreement when it is equal to 0. For the transformed data, the Lin’s concordance correlation coefficient is 0.90 (95% confidence interval = 0.87–092). Petrie and Watson (2013) note that based on a previous article published by McBride et al. [10], a Lin’s concordance correlation coefficient of 0.90 ≤ rc ≤ 0.95 is considered moderate agreement.

Discussion

Taking into consideration the paired t-test, Pearson correlation coefficient, Bland-Altman plot, and Lin’s concordance correlation coefficient, we accept our hypothesis that the hand-held Precision Xtra glucometer moderately agrees with the laboratory method and is acceptable to use for rapid, chute-side measurement of glucose in beef cattle. Several studies have previously evaluated the agreement of the Precision Xtra glucometer with laboratory analysis of glucose in dairy cattle, however, this is the first study to our knowledge that has compared the two methods in beef cattle during a GTT [36].

Our results agree with those of previous reports that indicate the Precision Xtra glucometer has acceptable agreement with laboratory measurement in dairy cattle [3,5]. Both of these papers demonstrated Bland-Altman plots and reported that at least 95% of the observations fell within the 95% confidence intervals, indicating good agreement between the two methods. Neither paper, however, reported a Lin’s concordance correlation coefficient or an intraclass correlation coefficient. Therefore, while there is agreement between the two methods based on the Bland-Altman plots, the assessment of agreement cannot be definitive without one of the indexes being calculated.

Additionally, of the papers that have previously reported unacceptable agreement between the Precision Xtra and laboratory analysis of glucose concentration only Lopes et al. [6] reported an LCCC or ICC while Megahed et al. [4] did not. It is interesting that Lopes et al. [6] showed moderate association between the two methods with a Pearson correlation coefficient (r = 0.71), indicated that their reported LCCC of 0.74 demonstrated strong agreement with the reference method, and reported that the difference between the methods fell within acceptable limits of agreement (± 1.96 standard deviations) at least 95% of the time, however, continued to state that according to the American Society for Veterinary Clinical Pathology guidelines, only 54.6% of the Precision Xtra glucose readings had a total observed error of ≤ 20% and declared the meter inadequate to measure glucose in dairy cows. In the present study, we did not measure total observed error, as Bland and Altman [8] and Petrie and Watson [9] do not discuss calculating this total observed error. Additionally, Petrie and Watson [9] follow the guidelines proposed by McBride [10] that state that a LCCC less than 0.90 is indicative of poor agreement between two methods. Following these guidelines, the LCCC of 0.74 reported to be in support of strong agreement should be observed with caution.

Similar to Zakian et al. [5], we used the glucose oxidase method for the laboratory glucose measurement. Using this method as the gold standard, we obtained similar results to Zakian et al. [5] and found the glucometer to be in moderate agreement to the laboratory measurement. As mentioned by Zakian et al. [5], one possibility for this improvement in agreement between the two methods may be the reference method used, as the other papers that evaluated the Precision Xtra glucometer compared its measurements to a hexokinase reference method [4,6]. Wittrock et al. [3] reported acceptable agreement between the two methods, however, did not mention which reference method was used and only stated that glucose concentrations were determined using a commercial reagent kit.

Of the previous studies that evaluated the agreement of the Precision Xtra glucometer with the laboratory measurements, only Wittrock et al. [3] similarly performed a GTT. Performing this method to assess agreement during a GTT is of great interest, as there are many samples that must be taken during this procedure and the hand-held glucometer could make performing the procedure quicker and cheaper if glucose concentration could be analyzed chute side. The present dextrose infusion that was provided to the beef steers (0.25 g of glucose/kg BW delivered in a 50% weight/volume dextrose solution) was the same as that used by Wittrock et al. [3] in dairy cows, however, the sampling timeline was different. Steers were sampled for blood in the current study at 5 and 2 minutes before glucose bolus infusion, and then subsequent samples were taken immediately after glucose bolus infusion (0 minutes) and then 5, 10, 15, 20, 30, 60, and 120 minutes after infusion. Wittrock et al. [3] sampled for blood immediately before dextrose infusion and then at 10 and 80 minutes after infusion. Wittrock et al. [3] reported that measurements that were within the physiological range of 2.3 to 5.2 mmol/L were slightly lower with the glucometer compared with the laboratory value. Alternatively, the authors found that the high glucose concentrations were generally overestimated by the glucometer, though the authors were not aware of a methodological reason for this difference. In the present study, we found that the glucometer and laboratory measurements were very precise and accurate up to concentrations ≤ 300 mg/dL. At glucose concentrations greater than 300 mg/dL, we observed our greatest differences between the glucometer and the laboratory measurement. Except for one blood sample that was obtained 5 minutes after glucose infusion, these high concentrations of glucose (> 300 mg/dL) all coincided with our time 0 blood sample which was sampled immediately after glucose infusion. Generally, the glucometer overestimated glucose concentration when compared with the laboratory measurement at these high concentrations. This result is similar to that reported by Wittrock et al. [3]. However, when the concentrations greater than 300 mg/dL were removed from the analysis, the LCCC was only improved to 0.92 which still indicates moderate method agreement according to Petrie and Watson [9]. Therefore, we have only presented the statistical analyses including all of the data points from the GTT.

Since the full range of values included in the data set provided acceptable Bland-Altman plots and moderate agreement according to the LCCC, we conclude that the hand-held glucometer is acceptable to use for rapid, chute-side testing of blood glucose concentration in beef cattle. This glucometer was tested during a GTT, and while all of the data was included in the analyses presented, we caution its use at supraphysiological glucose concentrations such as that occurring immediately after glucose infusion during a GTT. However, it seems that under normal physiologic conditions the hand-held glucometer agrees with the laboratory glucose oxidase reference method. Additionally, based on our suppliers and current costs to complete a GTT for 13 steers, the glucometer method was 57% cheaper on a per sample basis compared with the laboratory method.

Supporting information

S1 Table

(XLSX)

Acknowledgments

The authors would like to thank the staff at the Eastern Agricultural Research Station for all of their help in completing this project.

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

The author(s) received no specific funding for this work.

References

  • 1.Hayirli A. The role of exogenous insulin in the complex of hepatic lipidosis and ketosis associated with insulin resistance phenomenon in postpartum dairy cattle. Vet Res Comm. 2006;30:749–774. doi: 10.1007/s11259-006-3320-6 [DOI] [PubMed] [Google Scholar]
  • 2.Trinder P. Determination of blood glucose using 4-aminophenazone. J Clin Path. 1959;22:246. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Wittrock JAM, Duffield TF, LeBlanc SJ. Short communication: Validation of a point-of-care glucometer for use in dairy cows. J Dairy Sci. 2013;96:4514–4518. doi: 10.3168/jds.2012-6533 [DOI] [PubMed] [Google Scholar]
  • 4.Megahed AA, Hiew MWH, Townsend JR, Messick JB, Constable PD. Evaluation of an electrochemical point-of-care meter for measuring glucose concentration in blood from periparturient dairy cattle. J Vet Intern Med. 2015;29:1718–1727. doi: 10.1111/jvim.13608 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Zakian A, Tehrani-Sharif M, Mokhber-Dezfouli MR, Nouri M, Constable PD. Evaluation of a point-of-care electrochemical meter to detect subclinical ketosis and hypoglycaemia in lactating dairy cows. Austr Vet J. 2017;95:123–128. doi: 10.1111/avj.12568 [DOI] [PubMed] [Google Scholar]
  • 6.Lopes RB, Valldecabres A, Silva-del-Río N. Technical note: Glucose concentration in dairy cows measured using 6 handheld meters designed for human use. J Dairy Sci. 2019;102:9401–9408. doi: 10.3168/jds.2018-15688 [DOI] [PubMed] [Google Scholar]
  • 7.Hart IC, Bines JA, Balch CC, Cowie AT. Hormone and metabolite differences between lactating beef and dairy cattle. Life Sci. 1975;16:1285–1291. doi: 10.1016/0024-3205(75)90313-6 [DOI] [PubMed] [Google Scholar]
  • 8.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet. 1986;327:307–310. doi: 10.1016/S0140-6736(86)90837-8 [DOI] [PubMed] [Google Scholar]
  • 9.Petrie A, Watson P. Statistics for veterinary and animal science: 3rd Ed. John Wiley & Sons, Ltd.; 2013. [Google Scholar]
  • 10.McBride GB. A proposal for strength-of-agreement criteria for Lin’s concordance correlation coefficient. NIWA Client Report. 2005;HAM2005-062. [Google Scholar]

Decision Letter 0

Marcio Duarte

23 Mar 2022

PONE-D-21-38258A comparison between two glucose measurement methods in beef steers during a glucose tolerance testPLOS ONE

Dear Dr. Parker,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Because one of the reviewers raised several crucial questions regarding the hypothesis tested, methods and statistical analysis, I invite you to revise the manuscript taken into consideration his comments. Also, please make sure to reply all his comments with detailed answers to allow myself and the reviewer to better understand what was proposed in the manuscript. Because we did not have many comments from one of the reviewers at this time, and was quite difficult to find available reviewers, I am passing the manuscript for you to review with the comments made to avoid any further delay. However, after your revision the manuscript shall be sent to a third reviewer for assessment. 

Please submit your revised manuscript by May 07 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Marcio Duarte, PhD

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf  and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.

Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.

Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.

We will update your Data Availability statement to reflect the information you provide in your cover letter.

3. We note that you have stated that you will provide repository information for your data at acceptance. Should your manuscript be accepted for publication, we will hold it until you provide the relevant accession numbers or DOIs necessary to access your data. If you wish to make changes to your Data Availability statement, please describe these changes in your cover letter and we will update your Data Availability statement to reflect the information you provide.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: No

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: This manuscript aimed to evaluate the effectiveness of a hand-held glucometer in beef cattle The manuscript is within the scope of the Plos One and can be accepted for publication. However, this manuscript should be classified as a short communication.

Reviewer #2: Lines 56-59 - I really didn’t understand why you are pointing out these limitations and possible errors of the laboratory method as a justification for using the hand-held glucometer. Because you are using this method as a gold-standard method for your comparison and recommendation of the alternative method.

Lines 60-62 - This sentence sounds very informal to a paper’s introduction. I believe that should be in a more appropriate place, such as discussion.

Lines 68-70 - What is the real necessity to validate the same method for beef cattle? Why the results should be different for beef cattle when compared with dairy cattle? We are talking about the method.

Line 101 and line 104 – Is the range of reading from 20 to 300 or 20 to 500? There is confusion here.

Lines 105-107 - This is not correct. This sample should be deleted from the database because you are not certain about the value.

Lines 118-120 - Why did you not test for normality?? You need to be certain about distribution because that could be biasing the results.

Lines 120-123 – In a general way, the information concerning statistical analysis and methods for assessing adequacy were poorly described in this section. Some explanations about these methods are dispersed over the results and discussion, which is not the best place for that. I suggest that authors should be providing these details in the current section.

Line 132 - Why did you not include regression to test the adequacy of your new method?

Lines 137-139 - I am really not sure about the confidence to test transformed variables. It is proved that variables transformation to follow Gaussian distribution, do not represent a good estimate of the real value when a back-transformation is applied. I believe that is a similar case. The funnel shape is also present in the transformed data, but with lower intensity. The authors should think about that.

Lines 142-1444 - It is clear that there is a bias over 300 mg/dL (Figures 1 and 2) and I believe that should be analyzed and reported in the manuscript.

Lines 160-174 - This entire paragraph is closer to a material and methods than a discussion.

Lines 215-218 - Which is the necessity to repeat this text?

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Otavio R. Machado Neto

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2022 Jul 19;17(7):e0271673. doi: 10.1371/journal.pone.0271673.r002

Author response to Decision Letter 0


3 May 2022

I have attached a document titled response to reviewers comments that details the author's response to both reviewers

Attachment

Submitted filename: Response to Reviewers PLOSone Glucose steers.docx

Decision Letter 1

Marcio Duarte

6 Jul 2022

Short communication: A comparison between two glucose measurement methods in beef steers during a glucose tolerance test

PONE-D-21-38258R1

Dear Dr. Parker,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Marcio Duarte, PhD

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: The authors have answered all of my concerns. Because of that, I believe the manuscript is now ready for publication.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: Yes: Alex Lopes da Silva

**********

Acceptance letter

Marcio Duarte

11 Jul 2022

PONE-D-21-38258R1

Short communication: A comparison between two glucose measurement methods in beef steers during a glucose tolerance test

Dear Dr. Parker:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Marcio Duarte

Academic Editor

PLOS ONE


Articles from PLoS ONE are provided here courtesy of PLOS

RESOURCES