Tracking and Dynamic Tuning of a Wireless Powered Endoscopic Capsule

Lucas Murliky 1,*, Gustavo Oliveira 1, Fernando Rangel de Sousa 2, Valner João Brusamarello 1,*
Editor: Christian Vollaire
PMCID: PMC9506451  PMID: 36146266

Abstract

This work presents an inductive wireless power transfer system for powering an endoscopy capsule supplying energy to power electronic devices allocated inside a capsule of ≈26.1 mm × 9 mm. A receiver with three coils in quadrature with dimensions of ≈9 mm × 9 mm × 10 mm is located inside the capsule, moving freely inside a transmitter coil with 380 mm diameter through translations and revolutions. The proposed system tracks the variations of the equivalent magnetic coupling coefficient compensating misalignments between the transmitter and receiver coils. The power on the load is estimated and optimized from the transmitter, and the tracking control is performed by actuating on a capacitance in the matching network and on the voltage source frequency. The proposed system can prevent load overheating by limiting the power via adjusting of the magnitude of voltage source VS. Experimental results with uncertainties analysis reveal that, even at low magnetic coupling coefficients k ranging from (1.7 × 103, 3.5 × 103), the power on the load can be held within the range of 100–130 mW. These results are achieved with any position of the capsule in the space, limited by the diameter of the transmitter coil and height of 200 mm when adjusting the series capacitance of the transmitter in the range (17.4, 19.4) pF and the frequency of the power source in the range (802.1, 809.5) kHz.

Keywords: endoscopic capsule, wireless power transfer, WPT dynamic tracking

1. Introduction

Capsule endoscopy is a technique that employs a tiny wireless camera embedded in a small device [1,2]. During the procedure, the device is swallowed by the patient, allowing it to take photos while travelling through the digestive tract. The characteristics of the application mean that several restrictions related to the small volume of the capsule are imposed. This constraint primarily impacts the energy autonomy of the electronic circuits, as it limits the size of the batteries. Moreover, high quality images and motion control of the capsule are necessary features for improving diagnostic ability, and cannot be achieved without expending considerable extra power [3]. To overcome these issues, Wireless Power Transfer (WPT) based on low- or mid-range inductive coupling has been considered as a possible solution for powering the electronic systems and recharging the batteries [1,4,5,6].

In recent years, many applications for restricted environments have been implemented via WPT, including sealed and embedded sensors in concrete walls [7] and various devices intended for implantation in the human body [8,9,10,11,12] that are capable of carrying both power and data [13]. In addition, WPT has found applications in consumer electronics for charging the batteries of mobile devices [14] and electrical vehicles [15,16,17,18,19].

In an inductive link, the magnetic coupling coefficient k between the transmitting and the receiving coils is dependent on the distance. Thus, low magnetic coupling coefficients are associated with cases where the transmitter and receiver coils are relatively far each other. A dynamic k is associated with cases in which one or both coils can move. Moreover, the inductive link usually works via tuning to the frequency of the transmitter power source (ETX). The syntony is defined by the equivalent circuit of the coils and a capacitive matching network. The tune of the circuit guarantees the efficiency or the maximum possible power transferred (MPPT) to the load on the receiver (ERX). However, both a the dynamic magnetic coupling coefficient k and dynamic loading require active (dynamic) tuning in order to track changes and maximize the power delivered to the load. Active tuning [20] can be achieved by frequency sweeping [21], using switching capacitors [22], using saturable variable inductors [23], or by a combination of techniques [24,25]. None of the aforementioned works have investigated the performance of an inductive link for very low values of k, as featured by long distances (>radii of the coil). Furthermore, most of the works cited above present a single variable for tuning the resonant circuit, missing the possibility of attaining extra degrees of freedom for controlling the power delivered to a load. Different optimization methods can be used to track the maximization or minimization process of generic objective functions through the variation of controllable parameters [26,27,28,29]. In the present article, an output power Po optimization process is implemented by varying the frequency and the capacitive network simultaneously.

This work is an extension of the work presented in [30], and focuses on the estimate of k between coils by measuring the voltage VS and current IS in the ETX without any intervention with respect to the ERX (where the load is connected). The main innovation presented in this article lies in the optimization procedure used to track and dynamically compensate for variations in distances, and thus for the value of k between the coils, without increasing the volume of the ERX, especially for applications with very low magnetic coupling coefficients (k1×102). Thus, we present a case study of a specific inductive link: ETX coil with a diameter of =380 mm and ERX composed by three coils in quadrature (with =9 mm ×9 mm ×10 mm) for use in a commercially available endoscopic capsule of ≈26.1 mm ×9 mm [31,32]. The geometric setup of this application is characterized by a large asymmetry between the coils. Furthermore, the ETX can move over large distances, resulting very low values of k in the range of (1.7×103,3.5×103). The chosen frequency was centered at 805 kHz. In this case, it is only possible to maintain the power on the load and compensate for the misalignment between coils when the designed MPPT is sensitive to the range of possible variations of the actuating parameters. Thus, the presented procedure controls the MPPT by simultaneously actuating on two variables, namely, the frequency of the power source (ω) the and ETX series capacitance (C1). In addition, the matching capacitive network is customized with an initial value of C1 in order to allow real-time continuous maximization of power on the load. Table 1 summarizes different techniques for dynamically tuning the WPT applied to the endoscopic capsule and compares them with the method proposed in this article.

Table 1.

Tuning methods applied to inductive links for endoscopic capsule.

Reference Output Dynamic Tuning Method
[31] η Frequency scanning
[3] η Capacitor switching network
[33] η Multiple frequency transmitter
[34] Po No dynamic tuning
This work Po Frequency scanning and capacitor switching network

η = Efficiency; Po = Power on load.

The proposed system can both compensate the inductive link and guarantee a range of power values on the load operating on the MPPT given the actual circuit and variables ω,C. The proposed approach allows the capsule to move freely, guaranteeing minimum required power on the load by adjusting the input voltage source magnitude, implying a lower SAR level. The results show the advantages of the presented approach in compensating the WPT with low values of k. The power on the load is maintained above the lower limit of 100 mW on a load of 47 Ω with ETX moving inside the available free space by adjusting the frequency of the power source in the range of (802.1, 809.5) kHz and the compensating capacitance in the range of (17.4, 19.4) pF. Finally, we limited the power to 130 mW by controlling the magnitude of the power source VS in order to avoid overheating on the load.

2. Multicoil Inductive Link

The presented WPT design is asymmetrical, with a large ETX coil and a small ERX applied to the case study of a Swallowable Wireless Endoscopy Capsule (SWEC) of length × diameter 26.1 mm ×9 mm. The ETX is large enough to fit around the patient’s body and the ERX is small enough to fit inside the SWEC (Figure 1) with three coils in quadrature.

Figure 1.

Figure 1

Endoscopy system showing the external coil (ETX) and detail of the capsule with four coils (L1, L2, L3 and L4) in quadrature (ERX).

The ETX–ERX pairing forms a weakly coupled inductive link with a low k (1.7×103<k<3.5×103) because of the large distance between ETX and ERX. In clinucal application, after being swallowed the endoscopy capsule should move freely in the gastrointestinal tract while facing rotations and translation movements. The quadrature coils inside the capsule ensure minimal or critical k in at least one of these coils [31,35]. Thus, the receiver coils can pick up the magnetic field from three different directions and balance the output-induced voltage, providing enough power to the load in any possible position of the capsule in the gastrointestinal tract.

2.1. Design of the Inductive Link to the SWEC

Both the ETX and ERX coils are solenoids (Figure 1). The ETX coil was built with two sections, =380 mm, 43 turns, and a height of 30 mm each, presenting inductance L1. In the capsule, three coils are mounted orthogonally; L2 and L3 have =9 mm with 14 turns, while L4 has =10 mm and 14 turns. Supposing that the ETX coil is perfectly aligned with the z axis (Figure 1) and fully coupled to one of the receiving coils, we can disregard other receiver coils because they are in quadrature. Then, supposing that the ERX is positioned in the center of the ETX solenoid without any angular misalignment, the magnetic coupling coefficient k can be plotted as shown in Figure 2 when the ERX moves through the z axis. The red boxes depict the limits of the ETX solenoid width of 30 mm each. Figure 3 shows the values of k over the xy plane when ERX is aligned and centered with z axis at Z=125 mm.

Figure 2.

Figure 2

The magnetic coupling coefficient k between the ETX and one receiver coil aligned with the z axis when the ERX moves over a range of 250 mm.

Figure 3.

Figure 3

Value of k over xy plane when one coil of the ERX is aligned with the z axis at Z=125 mm (Figure 2).

As stated in [30], by rotating the capsule along the x and y axes with angles θ and ϕ (according Figure 4), the values of the mutual inductances (see Figure 5) vary as well.

Figure 4.

Figure 4

ERX Rotation angles ϕ and θ over x,y,z space (reference ETX and ERX aligned with z axis).

Figure 5.

Figure 5

Model of the inductive link with one transmitter and three receiver coils located inside the endoscopic capsule, which moves freely through translations and rotations in the gastrointestinal tract.

Thus, three magnetic coupling coefficients between the ETX and the coils of the ERX can be described as functions of the angles θ and ϕ (1):

k1=Ck0cosϕk2=k0cosθsinϕk3=k0sinθsinϕ (1)

where k0 is the magnetic coupling coefficient when one of the coils from the ERX is aligned with the z axis and C is a constant describing the effects caused by differences in the diameters of the receiver coils (L2=9 mm, L3=9 mm, and L4=10 mm, respectively; see Figure 1). Figure 4 shows the θ and ϕ angles over the adopted x,y,z space.

The mutual inductance (2) depends on the magnetic coupling coefficient k and on the inductances of the transmitter LETX and receiver LERX:

M=kLETXLERX (2)

The approach presented in this article is based on tracking the tuning in the transmitter circuit ETX and monitoring the effects of the reflected impedance by the ERX. Thus, the design of the coils in the ETX–ERX system must guarantee a mimimum (critical) coupling coefficient kc throughout the whole capsule path in order to allow the implemented circuit to detect impedance variations and track the tuning.

The designed inductive link is able to produce k0 in the range (1.7×103,3.5×103) (see Figure 2 for z250 mm to z125 mm displacement and constant C1). In this specific case, the presented power control system can maintain the required power on the load, as far as the compensation for variations of the reflected impedance detected on the ETX during misalignments or load variations. Although the maximum power under a given pair (ω,C1) is obtained when the reactance on the load is cancelled, maximum power transfer cannot be guaranteed because the resistive part of the load may not match the real part of the link impedance. Only the reactive portion of an impedance can be adjusted by varying both ω and C1. Thus, the tuning and cancellation of the reactive portion cannot guarantee the maximum, only the “maximum possible power transfer” (MPPT) to the load.

2.2. Power Transfer Analysis

The multi-coil inductive link described in [30] is shown in (Figure 5). The mixed-compensation matching network is composed of four capacitors (C1C4) in an SPSP configuration [36,37]. The ETX is powered by a voltage source VS with internal resistance RS and the ERX is connected to an AC–DC converter and to a load RL representing the SWEC. R1R4 and RC1, ..., RC8 are intrinsic resistances of the inductances and ESR of the compensation capacitors, respectively. Finally, M1N=k1NL1LN are the mutual inductances, where k1N represents the magnetic coupling coefficient between L1 and LN, N=2,3,4. The cross mutual inductances of the receivers M23,M24,M34 are disregarded because the coils are orthogonal, and thus we can assume three independent receiving coils.

Figure 6 shows part of the inductive link composed of L1 and a single coil receiver L2, the power source, and the compensation networks, with ZCx=Rx+1jωCx.

Figure 6.

Figure 6

Inductive link with a single coil receiver.

All circuit parameters are known except for the mutual inductance. Furthermore, the presented tracking strategy depends on the estimation of the magnetic coupling factor k. Thus, the voltage and the current and voltage in ETX are measured in order to infer the value of k. As shown in [24], the magnetic coupling factor k can be estimated from the measured source current (Is^) and the circuit parameters as follows:

k^(Is^)=±T1T2T3ωT4L1L2 (3)

where

T1=Rs+ZC1+ZC2T2=Vs(ZC2+ZL1)+Is^T5T3=ZC4(ZC3+ZL2)+RL12(ZC3+ZC4+ZL2)T4=(Rs+ZC1+ZC2)T6(RL12+ZC4)T5=(ZC2ZL1+Rs(ZC2+ZL1)+ZC1(ZC2+ZL1))T6=(Vs+Is^(Rs+ZC1+ZC2)).

Thus, with the estimated value of the magnetic coupling coefficient k^ we can estimate the mutual inductance as defined in (2). In addition, we can describe the power on the load RL12 as Po12=12IL12^(Is^)2RL12, where IL12^(Is^) is the estimate of the current on RL12. However, as ERX is composed by three orthogonal coils, we can generalize this conclusion to the inductive links formed by L1, L3 and by L1, L4. Thus, the three inductive links are connected to rectifiers in parallel with the load [38,39] and the power on the load RL, and the efficiency can be calculated with

Po=maxVL12,VL13,VL142RL (4)
η=Po12·e{VS^·IS*^}. (5)

3. Methodology

The estimated input current and voltage IS^ and VS^ are real-time monitored (Figure 7) and the control system runs an optimizing algorithm in order to maximize the power on the load by adjusting both the voltage source frequency ω and the capacitance C1 of the ETX.

Figure 7.

Figure 7

System architecture of continuous system tuning whenever the load or the relative position of the coils changes.

We can estimate the magnetic coupling coefficient K (with (3), an equivalent mutual inductance Meq=f(M12,M13,M14)) [24], and the the reflected impedance of the ETX. Thus, the control system is able to track the MPPT by running the optimization process and searching ω and C1, reestablishing the tuning of the circuit:

maximizeωR;C1SPo(ω,C1,IS^)=12n=24IL1n^(ω,C1,IS^)2RLsubjecttoPoPmaxmW (6)

where IL1n^ are the electric currents supplied to the load by each coil in the ERX. In addition, a threshold for the dissipated power can be included in (6), limiting the MPPT to Pmax mW, and Pmax can be controlled by the AC power source magnitude of the ETX. The optimization process can be executed using dedicated firmware for the ETX. The proposed optimization method is better detailed in [24].

The system continuously reads VS^ and IS^ and estimates variations in the equivalent mutual inductance Meq=f(M12,M13,M14) caused by variations of the magnetic coupling coefficients k1n and load when the capsule moves or the equivalent load changes. The proposed control architecture detects these variations and continues to track the tuning by finding a new frequency ω and new capacitance C1 in the transmitter.

The k of the presented WPT is in the range (1.7×103,6.5×103) (Figure 3). In order to find the best matching network for this range, we applied the method presented in [36] to find 1000 sets of four compensation capacitors (fixed capacitance of C2,C3,C4 and initial capacitance of C1) for different values of k. Although these sets produce similar outputs for a specific k, they present different responses for the realized dynamic tuning when changing ω and C1 during the variation of k. Figure 8 shows the results of three different sets of capacitors; it can be observed that the best choice is dependent on k.

Figure 8.

Figure 8

Three different sets of capacitors {S1,S2,S3} with output Po×k in the range 1.0×103,7.0×103 during the tuning procedure while optimizing ω and C1; S1 shows low performance at high values of k, while S2 shows the best performance for k>4.7×103 and S3 for k<4.7×103.

4. Results

In order to evaluate the proposed method while taking into account previously mentioned restrictions in terms of dimensions, we designed an inductive link. The restrictions define the inductive link; its inductances L1 and L2 and respective intrinsic resistances R1 and R2 depend on the geometry of the coil structure, including the coil shape, copper wire thickness, wire spacing, coil size, number of turns, etc. Table 2 presents the measured parameters of the inductive link (measured with an Agilent U1730C Series LCR meter) and their respective estimated (standard) uncertainties u(x). A Rigol DG1022 function generator combined with a class-B power amplifier were used as the sinusoidal voltage source of the inductive link. Assuming fixed magnitude and frequency, the standard uncertainties can be represented by 12.0±0.1 V and 805±5×102 kHz, respectively (Table 2). However, this application scans both the magnitude and frequency, and consequently the uncertainties vary with the movement of the capsule.

Table 2.

Design parameters of the inductive link.

Symbol Parameter Value u(x)
f Frequency 805 kHz 5×102 kHz
RS Internal Resist. of VS 700 mΩ 7 mΩ
VS Magnitude of VS 12.0 V 0.1  V
R1 ETX DC Resistance 10.18  Ω 0.05  Ω
R2 ERX DC Resistance 572 mΩ 6 mΩ
R3 ERX DC Resistance 588 mΩ 6 mΩ
R4 ERX DC Resistance 603 mΩ 6 mΩ
C1 Capacitance 18.3  p F 0.3  pF
C2 Capacitance 2.57  p F 0.04  pF
C3=C5 Capacitance 10.22  n F 0.05  nF
C7 Capacitance 7.22  n F 0.04  nF
C4=C6=C8 Capacitance 27.5  n F 0.02  nF
L1 ETX Inductance 1.87 mH 0.09  mH
L2 ERX Inductance 5.04  μ H 0.08  μH
L3 ERX Inductance 5.05  μ H 0.08  μH
L4 ERX Inductance 7.06  μ H 0.11  μH
ZL=RL Load Impedance 47.0  Ω 0.2  Ω

Measurement of the magnitude and phase of the current Is^ were implemented with a gain and phase detector (AD8302) circuit, as presented in [40]. This IC presents a typical nonlinear gain measurement <0.5 dB in the 1.8 V range and sensitivity of 30 mV/dB. The phase measurement has a sensitivity of 10 mV/ in the 1.8 V range. The uncertainty of the current magnitude is estimated by combining the different sources of circuit variability as u(IS)8.5 mA, assuming a specific current IS=500 mA. As the uncertainty u(IS) is dependent on certain variable parameters of the circuit, it can be expected to vary during compensation and maximum possible power tracking in the load. Here, the uncertainty of the measured phase was estimated as u(Is)0.6.

The parameters and their respective uncertainties were substituted in Equations (5), (6), and (8), and the values of the magnetic coefficient k, power on the load, and efficiency were estimated using the Monte Carlo method [41] with 500,000 cycles (N = 500,000). The results for k, Po, and η were obtained with a 95% confidence level, as follows: k=(3.5±0.2)×103, Po=250±21mW, and η = (7.4±0.3)% for this specific point of IS.

The coils were fixed and had the parameters described in Table 2. The spindle was moved to perform different values of k (according Figure 2) and the proposed closed-loop control system was implemented in order to track the MPPT on the load. At each movement of the vertical spindle, the coils misaligned in the z direction and the magnetic coupling coefficient varied as expected. The angular misalignment was implemented by placing the ERX in a movable frame to perform θ and ϕ variations according to Figure 4. The experimental coils of the ETX and ERX as well as the apparatus used to induce movement along the z axis (spindle) and to vary the θ and ϕ angles are shown in Figure 9. At each step, the system detected the unbalanced conditions through measurement of VS and IS, actuating on ω and C1, and re-establishing the MPPT on the load, as depicted by Figure 7.

Figure 9.

Figure 9

Experimental apparatus: (a) ETX coils with a spindle to adjust the movement of ERX on the z axis. (b) ERX coils inside a structure for varying angles ϕ and θ.

Considering the ETX coils aligned with one of ERX coils in axis z, depicted in Figure 4 with the ERX fixed at the spindle in the center of L1), Figure 10 shows the variation of the magnetic coupling coefficient k1n while holding ϕ=90 and varying θ. Rec2 illustrates the variation of the magnetic coupling coefficient k13 between the ETX and ERX coil L3 with =9 mm. Initially, it is completely aligned; as the angle θ increases, k13 decreases to zero. Rec3 illustrates the variation of the magnetic coupling coefficient k14 between the ETX and ERX coil L3 with =9 mm, starting from zero coupling (because the initial position is orthogonal to the ETX coil), increasing the magnetic coupling as angle θ increases, and reaching the maximum at θ=90. Rec1 represents the magnetic coupling coefficient k12 between ETX and ERX coil L2 with =10 mm. The magnetic coupling coefficient k12 remains at zero because this coil remains in an orthogonal position in relation to ETX during the variation of θ. These results show that the magnetic coupling coefficients can be estimated by monitoring the effects on the ETX when only one coil is fully coupled to the transmitting system and the other two coils are in quadrature. Otherwise, there is only one measurable effect in the ETX caused by two or three magnetic coupling coefficients from different circuits.

Figure 10.

Figure 10

Magnetic coupling coefficient of each receiver coil while θ is varying with fixed ϕ=90.

This particular case was simulated with the Finite Element Method (FEM) using COMSOL Multiphysics software, resulting in k0=3.5×103, which satisfies Equation (1). The implemented system can estimate the equivalent magnetic coupling coefficient keq from three misaligned coils. The special case of only one of the receiver coils aligned with the ETX coil produces keq equal to k from a single coil.

Figure 11a–c show the magnetic coupling coefficient from each of the ERX coils when the capsule is fixed at z=125 mm (see Figure 2) during the variation of angles θ and ϕ (again, the ETX and ERX =9 mm coils are aligned with the z axis). Figure 11d shows the combination of k12, k13, and k14 while the θ and ϕ angles vary according to Figure 4.

Figure 11.

Figure 11

Magnetic coupling coefficients from each of the receiver coils while θ and ϕ change: (a) k12, (b) k13, (c) k14, (d) combination of k12, k13, and k14.

Figure 12a,b present the respective simulated and experimental MPPT on the load with the angles θ and ϕ varying in the interval 0,90. Figure 12b shows the results of 49 measured points in steps of 15 of both ϕ and θ repeated three times. referring to Figure 12b, Table 3 shows points of the efficiency and power on the load with varying θ and ϕ. It can be observed that the frequency and the capacitance C1 vary within the intervals 791.3,818.0 kHz and 17.69,19.10 pF for the simulated results and (802.1,809.5) kHz and (17.4,19.4) pF for the experimental results; in addition, there is a correlation of k with Po and η. Referring to Figure 12a, the mean squared error (MSE) can be calculated as MSE = 51.9 mW.

Figure 12.

Figure 12

Power on the load when ω and C1 are used to tune the inductive link: (a) simulated results and (b) experimental results.

Table 3.

Experimental and simulated data of the inductive link when the ERX coils are in the center of the ETX coils and ϕ=θ.

Simulated Experimental
ϕ θ Freq. [kHz] C1 [pF] k Po [mW] η [%] Freq. [kHz] C1 [pF] k Po [W]
0 0 791.3 19.10 3.5×103 427.3 11.51 808.6 18.3 (3.6±0.2)×103 0.37±0.03
15 15 791.7 19.08 3.3×103 406.4 10.50 808.5 18.3 (3.2±0.2)×103 0.33±0.03
30 30 791.7 19.08 3.0×103 346.8 8.86 807.2 17.4 (2.8±0.1)×103 0.30±0.03
45 45 791.7 19.08 2.4×103 252.0 6.13 809.5 17.4 (2.6±0.1)×103 0.21±0.02
60 60 818.0 17.71 2.6×103 224.0 5.53 809.5 17.4 (2.6±0.1)×103 0.19±0.02
75 75 817.8 17.69 3.2×103 316.2 8.13 802.1 19.4 (3.4±0.2)×103 0.29±0.02
90 90 818.0 17.71 3.5×103 350.2 9.10 802.1 19.4 (3.5±0.2)×103 0.34±0.03

Confidence Level of Po and k, CL = 95%.

In our experiments, a power limitation was implemented in order to avoid overheating on the load. Thus, the magnitude of the voltage source voltage was controlled after the inductive link was tuned. Every time Po exceeded a threshold (130 mW in this work), a new magnitude of VS was estimated and adjusted. Figure 13a shows the variations of the magnitude of VS in the ETX used to order to keep Po limited at 130 mW in the ERX under rotation of angles θ and ϕ (Figure 13b). Table 4 presents simulated and experimental points in reference to Figure 13.

Figure 13.

Figure 13

Simulated voltage VS with varying θ and ϕ angles and Po maintained at 130 mW. (a) Simulated Voltage; (b) Power on the load.

Table 4.

Experimental and simulated data when the PDL is limited to 130 mW.

Simulated Experimental
ϕ θ VS Po [mW] VS Po [W]
0 0 7.28 128.7 7.8 0.13±0.01
15 15 7.49 129.6 7.9 0.13±0.01
30 30 8.14 129.6 8.0 0.14±0.01
45 45 9.61 129.5 10.0 0.13±0.01
60 60 8.17 129.6 9.5 0.13±0.01
75 75 6.95 129.8 8.0 0.12±0.01
90 90 6.59 128.9 7.5 0.13±0.01

Confidence Level of Po, CL = 95%.

The results presented in Figure 13b are similar to the results presented in Figure 12b, with the exception that the limiting threshold of 130 mW maintains the power constant even with varying k1n.

Figure 14 shows the simulated and experimental power on the load Po when varying the magnetic coupling coefficient keq. When detecting that the ERX has moved, thereby changing the value of keq, ω and C1 are adjusted to track the tuning of the circuit. The dashed lines show the limits of the interval 100,130 mW. The dotted line indicates Po without any threshold and the solid line indicates Po with the threshold set at 130 mW, resulting in an MSE of =10.6 mW when considering the solid line in Figure 14 as a reference.

Figure 14.

Figure 14

Simulated and experimental power delivered to the load after system tuning and when controlling the voltage source VS to avoid overheating on the load, with varying k.

Figure 15 presents the simulated and experimental variations of the voltage magnitude VS in order to limit the power delivered to the load (130 mW in this case study) when the SWEC is aligned with the center of the ETX and moving in the z axis in the 25,125 mm range (see Figure 2). Each position on the z axis presents a scaling of the variation curve of VS for the plane θ×ϕ, as shown in Figure 13. The experiment in Figure 15 shows 30 positions and their respective k values when the ERX moves along the z axis. This experiment was repeated three times with MSE =0.32 V, considering the simulation (solid line) as a reference.

Figure 15.

Figure 15

Simulated and experimental variations of the voltage magnitude VS, limiting the power delivered to the load (130 mW in this case study) when the ERX is aligned in the center of the ETX and moving in the z axis, with varying k according Figure 2.

5. Discussion

Table 5 shows a comparative summary of articles that have addressed the problem of WPT with the endoscopy procedure. The TX/RX column shows the of the ETX and ERX coils, d is the distance between coils on the ETX (when such a structure is present), Po is the power on the load (mW), and η is the efficiency of the inductive link. Table 5 shows that the dimensions of the coils are somewhat similar. However, there are differences in coil structure and coil size, and several articles do not implement dynamic compensation. Consequently, there are differences in the performance of each ETX–ERX pair. In [33], the authors present an ETX with two coils, one on the floor concatenated with another coil around the patient. The authors of [42] present the implementation of a mobile capsule without any dynamic tuning, while in [3] the authors propose a modified Helmholtz transmitter coil on the ETX with dynamic tuning by switching capacitors (as in [43]). The authors of [44] present optimization of the coil geometry of the inductive link in order to maximize efficiency, and [35] present a comparison of different types of ETX and ERX coils. The authors of [31] propose an inductive link with a pair of planar transmitter coils with dynamic tuning by frequency scanning. It is notable that proposal in the present work has smaller dimensions than [3,33,35,42,43]; this parameter is important for implementation of the SWEC and has strong impact on k. Furthermore, the efficiency of the proposal in the present work is higher than in the other methods presented in Table 5.

Table 5.

Comparative summary of results from the current literature employing WPT for endoscopy procedures.

Ref. Freq. [kHz] TX/RX [mm] d [mm] Po [mW] η [%]
[33] 1356 480/11 24 3.04
[42] 218 500/14 600 2
[3] 246 400/11 150 534 4.93
[44] 16,470 220/9 26 0.02
[43] 250 350/11.5 70 758 8.21
[35] 246 360/12.7 130 238.52 3.26
[31] 1000 202/8.9 100 70 1
TW 802–808 380/10 100 367.8 9.1

TW = This work.

The experimental results presented in this work emulate all possible movements of a capsule within the dimensions of a human torso limited by a diameter of 380 mm and a height of h=200 mm. Furthermore, the experimental results were obtained in an environment different from the final application of the capsule in the human body. Studies [44,45] have shown that the use of biological tissues causes a slight decrease in k and the power on the load. In this case, the same system presented in this article could be implemented, with the only difference being that the magnitude of the voltage source would have to be increased in order to maintain the load power requirements. We intend to carry out such experiments with biological tissues and SARS analysis in our future work.

6. Conclusions

This article presents the design, analysis, and experimental results of an IWPT system to power a capsule (≈26.1 mm × 9 mm) along a path, such as the gastrointestinal tract in an endoscopy exam. The capsule contains an ERX receiver with three solenoid-type coils (≈9 mm ×9 mm ×10 mm) moving freely through rotations and translations. The ETX transmitter is composed of two solenoids with =380 mm and width 30 mm connected in series and physically arranged concentrically at a distance of 130 mm in order to produce an approximately homogeneous internal magnetic field. As the ERX coils move in relation to the ETX coils, the magnetic coupling coefficient k between the emitter and the receivers varies from 0 to 1, and therefore the three coils are in quadrature, ensuring that at least one coil can pick up enough energy to power the capsule. In addition, the ERX is inaccessible, and both the adjustments and the actuation of any control action is performed in the ETX.

Thus, this work presents a procedure for continuously maximizing the power on the load by compensating for the variations caused by the movements of the coils in the ERX. The procedure is performed by measuring the voltage VS and current IS in the transmitter and simultaneously varying a capacitor in the matching network and the frequency of the AC power source in the ETX circuit. The approach of actuating on two variables instead of one has been shown to have advantages when dealing with very low magnetic coupling coefficients, such as in the case of this application (k2×103). In addition, the procedure estimates the power on the load, checking for an upper limit and actuating on the magnitude of the voltage source VS in ETX (if needed), thereby avoiding overheating on the load.

The experimental results were obtained considering a minimum and maximum power at the load within in the 100,130 mW range. The experiments included linear motion along an axis of ≈200 mm and two rotation angles of ERX coils. The power in the load was maintained within the proposed range by moving the ERX in the space limited by the diameter of the ETX coils and a height of up to 160 mm. In the presented case study, the variation of the magnetic coupling coefficient k varied within the range 1.7×103,3.5×103, which required the frequency to be adjusted in the (802.1,809.5) kHz range and the capacitance C1 in the (17.4,19.4) pF range. Compared to similar published works found in the literature, the proposal in this work achieved higher power transfer efficiency.

Author Contributions

Conceptualization, V.J.B. and L.M.; Investigation, L.M.; Methodology, L.M. and V.J.B.; Simulations and development: G.O. and L.M.; Writing, review and editing, L.M., V.J.B. and F.R.d.S.; Funding acquisition: V.J.B. and F.R.d.S. All authors have read and agreed to the published version of the manuscript.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

Funding Statement

This study was financed in part by the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior-Brasil (CAPES)-Finance Code 001 and CNPq. We also had financial support from FAPERGS-Fundação de Amparo a Pesquisa do Estado do Rio Grande do Sul-Brasil-20/2551-0000932-7.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Basar M.R., Ahmad M.Y., Cho J., Ibrahim F. Application of Wireless Power Transmission Systems in Wireless Capsule Endoscopy: An Overview. Sensors. 2014;14:10929–10951. doi: 10.3390/s140610929. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Kim J.H., Nam S.J. Capsule Endoscopy for Gastric Evaluation. Diagnostics. 2021;11:1792. doi: 10.3390/diagnostics11101792. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Basar M.R., Ahmad M.Y., Cho J., Ibrahim F. Stable and High-Efficiency Wireless Power Transfer System for Robotic Capsule Using a Modified Helmholtz Coil. IEEE Trans. Ind. Electron. 2017;64:1113–1122. doi: 10.1109/TIE.2016.2614268. [DOI] [Google Scholar]
  • 4.Khan S.R., Pavuluri S.K., Cummins G., Desmulliez M.P.Y. Wireless Power Transfer Techniques for Implantable Medical Devices: A Review. Sensors. 2020;20:3487. doi: 10.3390/s20123487. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Haerinia M., Shadid R. Wireless Power Transfer Approaches for Medical Implants: A Review. Signals. 2020;1:209–229. doi: 10.3390/signals1020012. [DOI] [Google Scholar]
  • 6.Campi T., Cruciani S., De Santis V., Maradei F., Feliziani M. Near Field Wireless Powering of Deep Medical Implants. Energies. 2019;12:2720. doi: 10.3390/en12142720. [DOI] [Google Scholar]
  • 7.Mayordomo I., Drager T., Alayon J.A., Bernhard J. Wireless power transfer for sensors and systems embedded in fiber composites; Proceedings of the Wireless Power Transfer (WPT); Perugia, Italy. 15–16 May 2013; pp. 107–110. [Google Scholar]
  • 8.Wu R., Li W., Luo H., Sin J.K.O., Yue C.P. Design and Characterization of Wireless Power Links for Brain—Machine Interface Applications. IEEE Trans. Power Electron. 2014;29:5462–5471. doi: 10.1109/TPEL.2014.2301173. [DOI] [Google Scholar]
  • 9.Hui S.Y.R., Zhong W., Lee C.K. A critical review of recent progress in mid-range wireless power transfer. IEEE Trans. Power Electron. 2014;29:4500–4511. doi: 10.1109/TPEL.2013.2249670. [DOI] [Google Scholar]
  • 10.Harrison R. Designing efficient inductive power links for implantable devices; Proceedings of the 2007 IEEE International Symposium on Circuits and Systems (ISCAS 2007); New Orleans, LA, USA. 27–30 May 2007; pp. 2080–2083. [DOI] [Google Scholar]
  • 11.Olivo J., Carrara S., Micheli G.D. IronIC Patch: A Wearable Device for the Remote Powering and Connectivity of Implantable Systems; Proceedings of the Instrumentation and Measurement Technology Conference (I2MTC), 2012 IEEE International; Graz, Austria. 13–16 May 2012; pp. 286–290. [Google Scholar]
  • 12.Li P., Bashirullah R. A wireless power interface for rechargeable battery operated medical implants. IEEE Trans. Circuits Syst. II Exp. Briefs. 2007;54:912–916. doi: 10.1109/TCSII.2007.901613. [DOI] [Google Scholar]
  • 13.Rush A., Troyk P.R. Dual Inductive Link Coil Design for a Neural Recording System; Proceedings of the 33rd Annual International Conference of the IEEE EMBS; Boston, MA, USA. 30 August–3 September 2011; [DOI] [PubMed] [Google Scholar]
  • 14.Waffenschmidt E., Staring T. Limitation of inductive power transfer for consumer applications; Proceedings of the 13th European Conference on Power Electronics and Applications; Barcelona, Spain. 28–10 September 2009; pp. 1–10. [Google Scholar]
  • 15.Budhia M., Covic G.A., Boys J.T. Design and Optimization of Circular Magnetic Structures for Lumped Inductive Power Transfer Systems. IEEE Trans. Power Electron. 2011;26:3096–3108. doi: 10.1109/TPEL.2011.2143730. [DOI] [Google Scholar]
  • 16.Pinuela M., Yates D.C., Lucyszyn S., Mitcheson P.D. Maximizing DC-to-Load Efficiency for Inductive Power Transfer. IEEE Trans. Power Electron. 2013;28:2437–2447. doi: 10.1109/TPEL.2012.2215887. [DOI] [Google Scholar]
  • 17.Wang C., Stielau O.H., Covic G.A. Design Considerations for a Contactless Electric Vehicle Battery Charger. IEEE Trans. Ind. Electron. 2005;52:1308–1314. doi: 10.1109/TIE.2005.855672. [DOI] [Google Scholar]
  • 18.Huh J., Lee S.W., Lee W.Y., Cho G.H., Rim C.T. Narrow-Width Inductive Power Transfer System for Online Electrical Vehicles. IEEE Trans. Power Electron. 2011;26:3666–3679. doi: 10.1109/TPEL.2011.2160972. [DOI] [Google Scholar]
  • 19.Shin S., Shin J., Song B., Lee S., Kim Y., Jung G., Jeon S. Wireless Power Transfer System for High Power Application and a Method of Segmentation; Proceedings of the Wireless Power Transfer (WPT); Perugia, Italy. 15–16 May 2013; pp. 76–78. [Google Scholar]
  • 20.Pantic Z., Lukic S.M. Framework and Topology for Active Tuning of Parallel Compensated Receivers in Power Transfer Systems. IEEE Trans. Power Electron. 2012;27:4503–4513. doi: 10.1109/TPEL.2012.2196055. [DOI] [Google Scholar]
  • 21.Brusamarello V., Blauth Y.B., de Azambuja R., Muller I., de Sousa F.R. Power Transfer With an Inductive Link and Wireless Tuning. IEEE Trans. Instrum. Meas. 2013;62:924–931. doi: 10.1109/TIM.2013.2245041. [DOI] [Google Scholar]
  • 22.Lim Y., Tang H., Lim S., Park J. An Adaptive Impedance-Matching Network Based on a Novel Capacitor Matrix for Wireless Power Transfer. IEEE Trans. Power Electron. 2014;29:4403–4413. doi: 10.1109/TPEL.2013.2292596. [DOI] [Google Scholar]
  • 23.Aldhaher S., Luk P., Whidborne J.F. Tuning Class E Inverters Applied in Inductive Links Using Saturable Reactors. IEEE Trans. Power Electron. 2014;29:2969–2978. doi: 10.1109/TPEL.2013.2272764. [DOI] [Google Scholar]
  • 24.Murliky L., Porto R.W., Brusamarello V.J., Rangel de Sousa F., Triviño-Cabrera A. Active Tuning of Wireless Power Transfer System for compensating coil misalignment and variable load conditions. AEU—Int. J. Electron. Commun. 2020;119:153166. doi: 10.1016/j.aeue.2020.153166. [DOI] [Google Scholar]
  • 25.Lee M., Lee C., Park C. Transceiver for Wireless Power Transfer Using a Cross-Coupled Oscillator for a Wireless On-Wafer Test. IEEE Trans. Instrum. Meas. 2017;66:2097–2105. doi: 10.1109/TIM.2017.2677658. [DOI] [Google Scholar]
  • 26.Zhou X., Ma H., Gu J., Chen H., Deng W. Parameter adaptation-based ant colony optimization with dynamic hybrid mechanism. Eng. Appl. Artif. Intell. 2022;114:105139. doi: 10.1016/j.engappai.2022.105139. [DOI] [Google Scholar]
  • 27.Chen H., Miao F., Chen Y., Xiong Y., Chen T. A Hyperspectral Image Classification Method Using Multifeature Vectors and Optimized KELM. IEEE J. Sel. Top. Appl. Earth Obs. Remote. Sens. 2021;14:2781–2795. doi: 10.1109/JSTARS.2021.3059451. [DOI] [Google Scholar]
  • 28.Wu D., Wu C. Research on the Time-Dependent Split Delivery Green Vehicle Routing Problem for Fresh Agricultural Products with Multiple Time Windows. Agriculture. 2022;12:793. doi: 10.3390/agriculture12060793. [DOI] [Google Scholar]
  • 29.An Z., Wang X., Li B., Xiang Z., Zhang B. Robust visual tracking for UAVs with dynamic feature weight selection. Appl. Intell. 2022 doi: 10.1007/s10489-022-03719-6. [DOI] [Google Scholar]
  • 30.Murliky L., Oliveira G., Gosmann G., Brusamarello V.J., de Sousa F.R. Study of a Wireless Energy Transmission System for an Endoscopy Capsule with Dynamic Tuning; Proceedings of the 2021 IEEE International Instrumentation and Measurement Technology Conference (I2MTC); Glasgow, UK. 17–20 May 2021; pp. 1–6. [DOI] [Google Scholar]
  • 31.Khan S.R., Desmulliez M. Towards a Miniaturized 3D Receiver WPT System for Capsule Endoscopy. Micromachines. 2019;8:545. doi: 10.3390/mi10080545. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Kang T., Kim S., Oh K.I., Hwang J.H., Lee J., Park H., Byun K., Lee W. Evaluation of Human Body Characteristics for Electric Signal Transmission Based on Measured Body Impulse Response. IEEE Trans. Instrum. Meas. 2020;69:6399–6411. doi: 10.1109/TIM.2020.2970870. [DOI] [Google Scholar]
  • 33.Sun T., Xie X., Li G., Gu Y., Deng Y., Wang Z. A Two-Hop Wireless Power Transfer System With an Efficiency-Enhanced Power Receiver for Motion-Free Capsule Endoscopy Inspection. IEEE Trans. Biomed. Eng. 2012;59:3247–3254. doi: 10.1109/TBME.2012.2206809. [DOI] [PubMed] [Google Scholar]
  • 34.Wen F., Chu X., Li Q., Jing F., Zhao W., Chu Z. Receiver Localization and Power Stabilization for Wireless Power Transfer System; Proceedings of the 2020 IEEE International Conference on Applied Superconductivity and Electromagnetic Devices (ASEMD); Tianjin, China. 16–18 October 2020; pp. 1–2. [DOI] [Google Scholar]
  • 35.Sravya D., Bobba P.B., Rani M.N.S. Investigation of Wireless Power Transfer System with Different Coil Structures used in Wireless Capsule Endoscope; Proceedings of the 2021 International Conference on Sustainable Energy and Future Electric Transportation (SEFET); Hyderabad, India. 21–23 January 2021; pp. 1–6. [DOI] [Google Scholar]
  • 36.de Azambuja R., Brusamarello V.J., Haffner S., Porto R.W. Analysis and optimization of an inductive power transfer with a randomized method. IEEE Trans. Instrum. Meas. 2014;63:1145–1152. doi: 10.1109/TIM.2013.2296397. [DOI] [Google Scholar]
  • 37.Koran A., Badran K. Adaptive Frequency Control of a Sensorless-Receiver Inductive Wireless Power Transfer System Based on Mixed-Compensation Topology. IEEE Trans. Power Electron. 2021;36:978–990. doi: 10.1109/TPEL.2020.3000455. [DOI] [Google Scholar]
  • 38.Li H., Li G., Xie X., Huang Y., Wang Z. Omnidirectional wireless power combination harvest for wireless endoscopy; Proceedings of the 2014 IEEE Biomedical Circuits and Systems Conference (BioCAS) Proceedings; Lausanne, Switzerland. 22–24 October 2014; pp. 420–423. [DOI] [Google Scholar]
  • 39.Basar M.R., Ahmad M.Y., Cho J., Ibrahim F. An Improved Wearable Resonant Wireless Power Transfer System for Biomedical Capsule Endoscope. IEEE Trans. Ind. Electron. 2018;65:7772–7781. doi: 10.1109/TIE.2018.2801781. [DOI] [Google Scholar]
  • 40.Porto R.W., Brusamarello V.J., Pereira L.A., Sousa F.R. Fine Tuning of an Inductive Link Through a Voltage-Controlled Capacitance. IEEE Trans. Power Electronics. 2017;32:4115–4124. doi: 10.1109/TPEL.2016.2598284. [DOI] [Google Scholar]
  • 41.Evaluation of Measurement Data—Supplement 1 to the “Guide to the Expression of Uncertainty in Measurement”—Propagation of Distributions Using a Monte Carlo Method. JCGM; Paris, France: 2008. Technical Report, JCGM 101. [Google Scholar]
  • 42.He S., Yan G.Z., Ke Q., Wang Z.W., Chen W.W. A wirelessly powered expanding-extending robotic capsule endoscope for human intestine. Int. J. Precis. Eng. Manuf. 2015;16:1075–1084. doi: 10.1007/s12541-015-0139-5. [DOI] [Google Scholar]
  • 43.Basar M.R., Ahmad M.Y., Cho J., Ibrahim F. A 3-coil wireless power transfer system with fine tuned power amplifier for biomedical capsule; Proceedings of the 2017 IEEE Asia Pacific Microwave Conference (APMC); Kuala Lumpar, Malaysia. 13–16 November 2017; pp. 142–145. [DOI] [Google Scholar]
  • 44.Na K., Jang H., Ma H., Bien F. Tracking Optimal Efficiency of Magnetic Resonance Wireless Power Transfer System for Biomedical Capsule Endoscopy. IEEE Trans. Microw. Theory Tech. 2015;63:295–304. doi: 10.1109/TMTT.2014.2365475. [DOI] [Google Scholar]
  • 45.Khan S.R., Pavuluri S.K., Cummins G., Desmulliez M.P.Y. Miniaturized 3-D Cross-Type Receiver for Wirelessly Powered Capsule Endoscopy. IEEE Trans. Microw. Theory Tech. 2019;67:1985–1993. doi: 10.1109/TMTT.2019.2893204. [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Not applicable.


Articles from Sensors (Basel, Switzerland) are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES