Abstract
Today's optical neuromodulation techniques are rapidly evolving, benefiting from advances in photonics, genetics and materials science. In this review, we provide an up-to-date overview of the latest optical approaches for neuromodulation. We begin with the physical principles and constraints underlying the interaction between light and neural tissue. We then present advances in optical neurotechnologies in seven modules: conventional optical fibers, multifunctional fibers, optical waveguides, light-emitting diodes, upconversion nanoparticles, optical neuromodulation based on the secondary effects of light, and unconventional light sources facilitated by ultrasound and magnetic fields. We conclude our review with an outlook on new methods and mechanisms that afford optical neuromodulation with minimal invasiveness and footprint.
Keywords: optical neuromodulation, optogenetics, neural interfaces
This review article provides a comprehensive overview of recent advances in optical neuromodulation in the context of addressing various physical constraints of in vivo light delivery.
INTRODUCTION
Understanding neural computation requires causal manipulation of neural activity to dissect the complex circuit connections and the relationship between neural activity and certain behaviors. Arguably, the first demonstration of neuromodulation for deconstructing the causal relationship between structures and functions was realized by the Italian scientist Luigi Galvani in the late 18th century to stimulate the sciatic nerve of a frog. Modern electrical stimulation techniques offer many insights into the cognitive functions of specific circuits and structures in the nervous system. However, there remain two fundamental challenges of electrical stimulation: first, no technique has yet succeeded in spatially confining the electric field down to a single neuron; second, electrical stimulation lacks neuron-type specificity to manipulate a subgroup of neurons with a particular molecular signature. To this end, optical stimulation addresses these two challenges owing to the sophisticated physical approach of confining the light focus, and the molecular biology strategy of genetically modifying specific neurons with light-sensitive proteins.
Scientists have long sought to modulate neural activity with light. Besides naturally light-responsive cells such as photoreceptors and intrinsically photosensitive retinal ganglion cells (ipRGCs) in the retina, it has been known for more than a century that light influences the activity of neurons that are not specialized to be light sensitive [1]. The invention of lasers in the 1960s revitalized the research area of optical stimulation in nerve cells, with the first demonstration of laser stimulation of abdominal ganglion neurons in aplysia in the early 1970s [2]. The development of photochemistry has provided neuroscientists with another toolbox for optical neuromodulation. The light-triggered uncaging of neurotransmitters, such as glutamate, has been used to analyze neural networks since the 1990s [3]. However, all these optical neuromodulation methods lack cell-type specificity, and all neurons within the range of illumination are theoretically subject to the effects thereof. In the early 2000s, the advances of genetics and molecular biology enabled selective photostimulation of genetically modified neurons. Specifically, the photo-uncaging of ligands that bind to ionotropic receptors provides a means to optically stimulate genetically modified neurons that ectopically express these receptors [4,5]. Finally, optogenetics combines many desirable advantages of neuromodulation with light: the illumination can be spatially confined to the diffraction limit; neural activity can be modulated with the temporal resolution of single spikes; and activation is genetically targeted via the expression of opsins in neurons that normally lack them [6,7].
All methods used in biology are limited by their physical constraints. For optogenetics, delivery of light in neural tissues represents one of the biggest challenges, deeply rooted in the intrinsic interaction between light and matter. The brain is a highly scattering and absorbing medium, owing to the presence of lipids in a wide range of size scales (e.g. neuron membranes and the myelin sheath) in an aqueous environment, and endogenous chromophores (e.g. heme and dopamine). To address this challenge, many methods have been developed to deliver light precisely to the targeted region in the nervous system with minimal acute and chronic damage to the tissue and reduced restraint of the animal's native behavior. The motivation behind these different approaches also represents the main incentive for this review paper.
In this review, we present an overview of the latest optical approaches for neuromodulation with a focus on strategies to deliver light in vivo with reduced invasiveness and restraint of the subject. Specifically, we begin with a theoretical framework of the physical principles involved in the development of these methods, citing relevant mathematical formulations to facilitate the understanding of physics constraints in these methods. We then present advances in optical neuromodulation in seven modules: conventional optical fibers, multifunctional fibers, optical waveguides, light-emitting diodes (LEDs), upconversion nanoparticles, optical neuromodulation based on the secondary effects of light, and unconventional light sources facilitated by ultrasound and magnetic fields. We end this review paper by providing our perspective on this field based on the aforementioned physical principles and recent developments in related fields.
PHYSICAL PRINCIPLES BEHIND OPTICAL NEUROMODULATION METHODS
A more complete theoretical framework of physical principles related to optical neuromodulation methods is available in the Supplementary Data.
Physical representations of light propagation
Understanding the physical constraints of optical neuromodulation methods requires fundamental knowledge in the electromagnetic theory of light propagation and light–matter interactions. To this end, we can express light propagation in one dimension (1D) in terms of the instantaneous electric field as a function of position z and time t:
(1) |
where is the amplitude of the electric field in the propagating wave and ω is the angular frequency, while and are the real and imaginary parts of the complex wavevector, , in scalar representation. Here, contains the information of the phase velocity of light propagation, and as a result, the refractive index (RI) of the medium. In addition, reflects the absorption of light in the medium. Based on Equation 1, we can express light intensity, which is defined as the magnitude of the Poynting vector, as follows:
(2) |
in which λ is the wavelength of light, and μ are the permittivity and permeability of the brain tissue, respectively, and is the complex RI that indicates the absorptivity of the brain tissue. Light intensity should have a dimension of energy per unit area and time (i.e. J m–2 s–1 or W m–2 in SI units).
Scattering of light in brain tissue
Scattering represents a common behavior of light–tissue interaction in the brain that contributes to the attenuation of light intensity over depth. Specifically, scattering of light in a medium is due to the heterogeneity of the RI in the medium. For light–brain interactions, this heterogeneity usually occurs on the level of nanometers to microns as the result of the spatial separation of water and lipid molecules with distinct RIs [8]. For example, a myelinated axon comprises an aqueous core with an RI of ca. 1.36 and an optically dense sheath with an RI of ca. 1.44 [9]. In this example, the myelin sheaths act as scatterers in an aqueous medium, much akin to oil droplets in the aqueous phase of milk, thus effectively blocking the transmission of light through the brain.
The scattering (and absorption) of spherical particles in a medium can be solved mathematically by the Mie theory to yield the scattering cross section, Csca. The scattering cross section is defined as the ratio of the power of scattering to the incident light intensity, with the exact solution as well as the Rayleigh approximation available in the Supplementary Data. Understanding the scattering of brain tissue requires knowing both the scattering cross section (how strongly each scatterer scatters light) and the density of scatterers (how many scatterers there are). The scattering coefficient μs is thus defined as the product of the density of scatterers, , and Csca:
(3) |
We can interpret μs as the inverse of the scattering mean free path, i.e. the average distance between two consecutive scattering events. Therefore, light experiences more scattering events within the same path length in a more scattering medium with a large μs. Due to fixed in a specific tissue, the wavelength dependence of μs reflects that of Csca, exhibiting a monotonic decrease with an increasing wavelength (Fig. 1a) [10].
The physical interpretation of μs in the context of the scattering mean free path provides a handy method for modeling scattering with the Monte Carlo method, which simulates a random walk of photon packets through three-dimensional (3D) space [11]. Due to the anisotropy of scattering, photon packets ‘remember’ their original travel direction after each scattering event (Fig. 1a, inset). The scattering coefficient μs and anisotropy factor g (Fig. 1b) can be combined to yield a new quantity, the reduced scattering coefficient, μs’:
(4) |
Therefore, μ′s describes the diffusion of photon packets in a completely random walk with no ‘memory’ of the previous direction of propagation after scattering. The inverse of μ′s is the transport mean free path, the traveled distance of a photon packet after which the ‘memory’ of the original propagation direction is eliminated (Fig. 1c, inset). Similar to μs, μ′s also decreases with an increasing λ (Fig. 1c).
Absorption of light in brain tissue
Besides scattering, absorption represents another mechanism contributing to the attenuation of light in brain tissue. One can define the absorption coefficient α as follows:
(5) |
With this definition, Equation 2 can be reduced to a more familiar form:
(6) |
where I0 combines all pre-exponential terms. The wavelength dependence of α of the brain tissue is shown in Fig. 1d. As can be seen, the absorption spectrum of the brain has two prominent features: first, the absorption peaks below 600 nm are attributed to hemoglobin in the blood; second, several absorption peaks at 980 nm, 1200 nm and 1450 nm correspond to the overtones of water molecules, which are abundant in all soft biological tissues [12]. Apparently, specific brain regions may exhibit distinct absorption properties from the bulk brain due to the accumulation of certain chromophores. For example, neuromelanin gives rise to the black color of the substantia nigra (literally ‘black substance’ in Latin) in the brain, due to a broad absorption spectrum extending from the ultraviolet (UV) to the near-infrared (NIR) region [13]. Therefore, it is necessary to consider the heterogeneity of light absorption in brain tissue when designing optical neuromodulation methods targeting different brain regions.
A direct consequence of light absorption is the elevated temperature in the local brain region. Stujenske and Gordon et al. modeled light-induced heating in brain tissue for optogenetics. Due to the strong absorption of 532-nm light in the brain (Fig. 1d), their model predicts a gradual temperature increase up to 2.2°C within a few hundred microns of the light source at 10 mW output (e.g. an optical fiber) [14]. Owen and Kreitzer et al. later found that this temperature increase alone can activate an inwardly rectifying potassium channel in the striatum, thus biasing behavioral results of optogenetic studies [15]. Since a threshold intensity is usually required for opsin activation (e.g. 1 mW/mm2 for channelrhodopsin-2 (ChR2) and 10 mW/mm2 for halorhodopsin (NpHR)) [16,17] and since long periods of continuous illumination are commonly used for optogenetic inhibition, it is crucial to control for non-specific heating due to tissue absorption in optical neuromodulation methods [18].
Penetration depth of light in brain tissue
The discussion of scattering and absorption above lays the foundation for understanding the penetration depth of light in brain tissue. The effective penetration depth of incident light is given as:
(7) |
The effective penetration depth of light in the brain in the range of 400 to 1800 nm is plotted in Fig. 1e. Furthermore, the inverse of the effective penetration depth is usually referred to as the effective attenuation coefficient:
(8) |
Thus, the incident light intensity decays exponentially with depth as follows:
(9) |
It is noteworthy that the maximum penetration depth is found at 1070 nm, which is close to the operation wavelength of the Nd:YAG laser, 1064 nm. The wide availability of the Nd:YAG laser and the maximum penetration depth of light near its operation wavelength in the brain thus make this wavelength region (1050–1100 nm) attractive for deep-brain neuromodulation [19]. As a comparison, we estimate the percentage of photons reaching a depth of 4 mm in the brain in Fig. 1f. Specifically, we choose four representative wavelengths relevant in optogenetics: 470 nm, which corresponds to the maximum activation of ChR2; 635 nm, which is the activation wavelength of opsins with the greatest red shift to date [20,21]; 980 nm, which is the wavelength used for deep-brain optogenetics with upconversion nanoparticles [22,23]; and 1064 nm, which is a representative NIR wavelength at which the Nd:YAG laser operates. As expected, 1064-nm light attenuates by ∼20 fold over a depth of 4 mm, in stark contrast to the attenuation of 109 fold at 470 nm, 103 fold at 635 nm and 102 fold at 980 nm over the same depth. The attenuation of light at different wavelengths is derived from the spectral properties of brain tissue reported in ref. [19] and [24].
Optical waveguides for light delivery
The attenuation of light in the brain, which represents a combination of both scattering and absorption, prohibits efficient light delivery into the deep brain regions. To address this challenge, optical waveguides, such as optical fibers and dielectric slab waveguides, are commonly used for illuminating a specific brain region with efficient light delivery. A good optical waveguiding material should exhibit minimal attenuation of light traveling therein. In addition, propagating light must be spatially confined within the optical waveguide by satisfying the following inequality for an optical fiber:
(10) |
where n1 and n2 represent the RI of the core and cladding material of the fiber, respectively. We will see how the inequality of (Equation 10) is satisfied in a number of fiber optic devices for optogenetics later.
Besides optical fibers, the planar slab waveguide confines propagating light in the waveguiding medium with an RI of ng between the substrate (ns) and the cover (nc). The longitudinal wave vector, , which is the component of wave vector k in the propagating direction, needs to satisfy the following inequalities:
(11) |
OPTICAL NEUROMODULATION TECHNIQUES
Conventional optical-fiber-based optogenetic interfaces
Implementation of optogenetics in vivo requires delivery of light into the brain of live animals, which is subject to the physics of light–tissue interaction discussed above [25]. Specifically, neurons that are genetically modified with light-sensitive proteins (i.e. opsins), such as ChR2 or NpHR, can be excited or inhibited by blue light (∼470 nm) or yellow light (∼590 nm), respectively. To deliver and express opsin genes, three methods are primarily used: viral transduction, transgenic or knock-in mouse lines and Cre recombination [26]. Specifically, genetically engineered viruses, such as lentiviruses (LVs) and adeno-associated viruses (AAVs), offer fast and efficient expression with high spatial localization [26]. Compared to LVs with larger sizes, AAVs can be packaged into higher titers, thus exhibiting a higher expression level with a larger volume of transduced tissue [27]. Various serotypes of AAVs feature distinguished transduction characteristics in terms of the transduced cell type, transduction level and spatial spread [28]. For example, AAV1 and AAV5 provide a widespread neuronal expression while the expression of AAV2 is more confined in space, with a 5- to 8-fold smaller transduced volume in the hippocampus region [29]. After successful opsin expression, light sources are applied to illuminate the genetically modified neurons.
To deliver light to the superficial regions of the brain, LEDs can be placed on the skull or a cranial window [30]. Despite the success in modulating cortical activity, the strong scattering and absorption of visible light in biological tissue makes it challenging to deliver light to deep-brain regions of interest without a transmitting device or agent. This challenge is visualized quantitatively in Fig. 1e, which reveals a 1/e decay distance of merely 0.1 mm for 470-nm incident light. To this end, low-loss optical fibers are commonly implanted into the brain to create light paths for in vivo optical stimulation in an effective and precise manner.
The earliest fiber-optic neural interface was implemented by Aravanis et al. [31] and was later presented in more detail by Zhang et al. (Fig. 2a) [32]. In general, a cannula is used to guide the insertion of the optical fiber to a specific brain target that harbors genetically modified neurons. This very cannula can also be used as a guide to deliver viral vectors into specific brain regions. Besides light intensity measurements at the tip of the fiber, activity measurements of light-sensitive neurons upon illumination are commonly used to validate successful optogenetic excitation or inhibition [32]. The most straightforward method of neuron activity measurements is via simultaneous electrical recordings of neural activities. To this end, metal microwires have been manually attached to optical fibers to realize a bidirectional interface between the implanted device and the subject [33,34].
To validate the capability of optogenetics in regulating neural circuit activity in vivo, Anikeeva et al. developed a high-throughput electrophysiology platform coupled with optical fibers for simultaneous optogenetics and extracellular neural recordings of light-modulated firing dynamics (Fig. 2b) [35]. In this method, four tetrode bundles composed of a total of 16 polymer-coated NiCr microwires are attached to the shaft of an optical fiber, which acts as both a light source and a structural support for the tetrodes. This assembly thus yields a multichannel device termed the ‘optetrode’. Since the optetrode enables single-unit electrophysiological recordings from distinct locations surrounding the central optical fiber, diverse brain activity patterns are observed under different illumination conditions. The utility of the optetrode in behavioral neuroscience experiments has been confirmed in ChR2-expressing and non-expressing animals engaged in an open field exploration task.
In the optetrode, rigid metal wires exacerbate the mechanical mismatch between the implant and brain tissue, usually leading to chronic tissue damage and an elevated immune response, such as neuron death, glial scarring and astrocyte reactivity [36]. To mitigate these challenges and retain the multichannel recording capability in a fiber-optic system, nanoelectronic coating (NEC) with a thickness of <1 μm has been conformally applied to the surface of the traditional optical fiber, forming a multifunctional neural probe with a diameter of 30 μm (Fig. 2c) [37]. Assisted by surface tension, NEC devices naturally wrap around a curved hosting probe when they are lifted across the water–air interface. The utility of these devices is validated by simultaneous recordings and optogenetic stimulation in the medial prefrontal cortex (mPFC).
Another challenge of the conventional optical fiber arises from its limited illumination volume, which is usually confined to the tip of the fiber. To this end, tapered optical fibers have been engineered to control and adjust their illumination volumes inside the brain (Fig. 2d) [38]. The tapered design permits the selection of optical modes and reduces tissue damage. The focused ion beam (FIB) micromachining technique enables the creation of optical windows along the tapered region, while gold plating is used to prevent light leakage [39]. To examine the illumination profile, simulations, and benchtop and in vivo animal experiments, are conducted under various incident light angles and with different numbers of optical windows. Coupled with a Neuronexus silicon probe, tapered fibers demonstrate the reconfiguration of light delivery at multiple points with immediate neural responses. In a later study, large volumes of brain illumination were accomplished by full numerical aperture (NA) light injection through the tapered fiber, while restricted light illumination could be quickly adjusted by modifying the incident angle of a single input ray [38]. The authors observed divergent behavioral outcomes from the animals when different brain regions were stimulated via a single implanted tapered fiber, thus validating its unique advantages for optogenetics.
Besides new advances in optical fibers, engineered opsins have been developed to afford red-shifted activation spectra [40–42] and ultra-sensitive light responses [21,43,44], thus enabling transcranial optogenetics with optical fibers mounted on the skull without invasive implantation in the brain. A recent demonstration by the Deisseroth lab leverages the high sensitivity and red-shifted activation of the potent channelrhodospin, ChRmine, to achieve transcranial modulation of deep-brain neural circuits. Combined with systemic viral delivery [45], ChRmine enables neuromodulation of behaving mice without any intracranial surgery. A significant advantage of 635-nm responsive ChRmine over 470-nm responsive ChR2 arises from the much greater brain penetration of 635-nm photons than 470-nm photons. Specifically, Fig. 1f reveals that with the same incident photon flux, a million times more 635-nm photons can penetrate through 4-mm brain tissue than their 470-nm counterparts. However, it is worth noting that only 0.1% of 635-nm photons incident on the brain surface can penetrate to a depth of 4 mm. Given the wavelength-dependent penetration depth shown in Fig. 1e, we envision that opsins responsive to longer-wavelength NIR light may offer transcranial and transdermal deep-brain optogenetics with the greatest penetration and least non-specific heating.
Multifunctional fibers as versatile optical interfaces
In the previous section, outfitting conventional optical fibers with existing recording electrode technologies permits a straightforward solution for simultaneous optogenetics and electrophysiology in vivo. However, the large footprint, mechanical rigidity and complicated assembly procedure prevent them from achieving optical manipulation and electrical recordings in long-term neuroscience studies. Micropipette pulling, which represents one of the most classical techniques in neuroscience, offers an elegant strategy for miniaturizing the optical fiber while integrating multiple functions therein [46]. Using a conventional micropipette puller, the De Koninck lab fabricated a dual-core microprobe with a graded index optical fiber for light transmission and a hollow core filled with an electrolyte solution (e.g. 1–3 M NaCl) for electrical recordings [46]. Despite the integrated optical and electrical modalities in the same microprobe, this platform only harbors a single electrode and thus disallows multiplexed recordings.
To address the challenges of conventional optical fibers, the Anikeeva lab has pioneered the development of a series of flexible, multifunctional and biocompatible fibers [47,48]. Unlike the micropipette pulling method, the thermal drawing process (TDP), which is conventionally used for fabricating optical fibers, is employed to incorporate a wide array of functional elements. These functional elements include recording electrodes of different materials (e.g. conductive polyethylene and tin) and diameters, optical waveguides with low loss (1.6–2.6 dB/cm) and microfluidic channels, all incorporated in a single device. In a typical TDP, a bulky template, the ‘preform’, is fabricated via a standard machining and consolidation process, after which hundreds of meters of long thin fibers can be drawn under an applied stress and an elevated temperature. It is noteworthy that only a few thermoplastics or low-melting-point materials comply with the drawing temperature of the surrounding polymer matrix for making the embedded polymer waveguides or electrodes [49]. Specifically, for fabricating an all-polymer waveguide with different refractive indices, polycarbonate (PC) with an RI of 1.58 serves as the core material, while the cyclic olefin copolymer (COC; RI = 1.53) [47], poly(methyl methacrylate) (PMMA; RI = 1.48) [50] or polyvinylidene difluoride (PVDF; RI = 1.426) [51] are used as its pairing cladding material. Note the general requirement of , which is mathematically derived in (Equation 10) above. Additionally, the convergence-fiber drawing method makes it possible to incorporate silica fibers inside the multifunctional fiber probes, thus mitigating the transmission loss in polymer waveguides [52].
In the first demonstration of multifunctional fibers, optical waveguides were realized in both the core and ring structures via the RI contrast between PC and COC. Also present in the multifunctional fibers were electrodes composed of a conductive polymer composite (e.g. conductive polyethylene (CPE)) and microfluidic channels (Fig. 3a) [47]. To validate the multifunctionality of this device, the authors performed electrophysiological monitoring of neural activities and simultaneous optogenetics stimulation in the mPFC of freely moving, transgenic Thy1-ChR2-YFP mice. In addition, the authors leveraged the built-in microfluidic channels to inject synaptic blockers, 6-cyano-7-nitroquinoxaline-2,3-dione (CNQX), into the same brain region. As a result, the optically evoked neural activities were significantly diminished despite light stimulation, followed by slow recovery and restoration of original firing patterns. A salient advantage of multifunctional fiber arises from its reduced dimensions and stiffness, making it comparable to human hair, thus leading to a minimal glial response and the long-term recording capability (over two months after implantation) of single units.
The demonstrated versatility of multifunctional fibers also enables one-step optogenetics [48]. Traditional optogenetic experiments in wild-type mice require a two-step surgery, in which the first surgical step delivers viral vectors for gene transduction, while the second step allows for the implantation of optical fibers or optrodes [32]. This traditional approach, also commonly practiced, imposes repeated tissue damage and often results in the misalignment between transgene delivery and activation. In contrast, multifunctional fibers offer an elegant one-step alternative with lower invasiveness and greater precision. Specifically, the built-in microfluidic channels enable the delivery of viral vectors containing opsin genes to the targeted brain region in the wild-type mouse brain. In addition, the neighboring electrodes and optical waveguide, which are constructed in the same device to naturally ensure the alignment, offer time-dependent monitoring of opsin expression over weeks. Using these devices implanted into multiple locations in the mouse brain, the authors demonstrated opto-electrophysiological studies of projections from the basolateral amygdala (BLA) to the mPFC and ventral hippocampus (vHPC) in behaving animals.
One common challenge in fiber-based optogenetics arises from the limited volume of illumination in the brain. Nonetheless, the intrinsic flexibility of multifunctional fibers enables an elegant solution to address this issue. The Jia group recently invented a fiber-based spatially expandable neural interface (Fig. 3b) [51]. In this approach, a femtosecond laser micromachining technique is used to first create a depth-dependent fiber probe with exposed functional interfacing sites along the fiber length, thus resulting in a layout similar to a single-shank Michigan probe. In addition, a spatially expandable fiber probe is realized by leveraging a unique helical scaffolding fiber to afford 3D coverage in space. After implantation, both depth-dependent and spatially expandable fiber probes enable optogenetic neuromodulation with simultaneous electrical recording, with chronic stability over five months. Specifically, these implanted fiber probes can record neural activity from various brain regions, such as the cortex, CA1 and CA3 fields of the HPC, hypothalamus, thalamus and amygdala in a mouse model of infection-induced epilepsy.
Another challenge of multifunctional fibers arises from the trade-off between the need to incorporate multiple functional elements and the inevitable increase in their bending stiffness. To mitigate this challenge, an adaptive hydrogel hybrid probe has been realized by encapsulating a bundle of fibers with different functions in the hydrogel matrix (Fig. 3c) [53]. Specifically, UV irradiation bonds the pre-gel solution to the surface of functional fibers via a one-step polymerization process, resulting in a thin and uniform hydrogel layer covering the fiber array. This probe has at least two advantages: first, the initial bending stiffness of the probe ensures deterministic implantation into deep brain regions. Second, hydration and swelling of the hydrogel coating after implantation results in a tissue-favorable matrix, leading to a substantial decrease in observed foreign body responses. As a result, the adaptive hydrogel hybrid probe exhibits stable performance in recording optically evoked potentials and single units for up to six months, and a reduced immune response.
Other than applications in the brain, multifunctional fibers have also been used in the spinal cord (Fig. 3d) [54]. To accommodate the deformation of the spinal cord during the natural movement of the subject, stretchable COC elastomer (COCE; RI = 1.51) and polydimethylsiloxane (PDMS; RI = 1.41∼1.47) were chosen as the core and cladding material, respectively. Silver nanowires (AgNWs) were dip-coated onto the surface of the fiber to form a conductive ring electrode, maintaining good electrical conductivity under applied strains. Electrical recordings of the spinal cord with optical stimulation demonstrated the suitability of stretchable fiber probes for studying the peripheral nervous system (PNS) [55,56]. More recently, stretchable optical fibers consisting of styrene-ethylene-butylene-styrene (SEBS) with different refractive indices and helical stretchable polymer electrodes have been reported with multi-axial sensing modalities under extreme deformation conditions [57].
Optical waveguides for individually addressed and multiplexed optogenetics
In the previous section, different materials are used for optical waveguiding and electrical recordings in separate channels. In contrast, some materials afford simultaneous electrical conductivity and optical transparency, thus affording multiplexed waveguiding and recordings through the same channels. For instance, a 4×4 ZnO-based micro-optoelectrode array (MOA) with a geometry of a Utah array was created for simultaneous and multiplexed recordings and optogenetic stimulation (Fig. 4a) [58]. Specifically, ZnO pillars (RI ∼2.0) coated with parylene-C (RI ∼1.6) acted as both the waveguide and the recording electrode. Multiplexity in optical stimulation was realized via a laser-scanning system, which spatially controlled light input to individual optoelectrodes with tunable light intensity during in-vivo optogenetic studies. Finally, multiplexed optical modulation with simultaneous recordings revealed the causal relationship between cortical microcircuit dynamics and evoked limb motions in anesthetized mice.
The back-end optical interfacing used in the MOA required a customized laser-scanning system, thus presenting a technical challenge for its wide adoption. Hence, a monolithically fabricated photonic circuit offers an alternative approach, with efficient input light coupling and compact packaging. The first reported on-chip waveguide measures 11 μm × 20 μm at the distal end and is composed of silicon oxynitride (RI = 1.51) as the core material and silicon dioxide (RI = 1.46) as the cladding. The waveguide assumes the layout of a typical Michigan probe and can be coupled to a laser via an optical fiber [59]. A more compact design with fiberless coupling is accomplished by coupling the waveguide mixer to the side-emitting injection laser diode via a gradient-index (GRIN) lens (Fig. 4b) [60]. Leveraging micropatterned photonic circuits, blue and red light can be directed into a shared waveguide port for optically modulating the same population of pyramidal cells in the mouse brain, that expresses both ChR2 and eArch3. Four of these multicolor optoelectrodes can be further packaged in parallel into a multi-shank optogenetic probe for dissecting complex neural circuits [61].
Although the integration of waveguides with electrodes into a chip-scale device creates new opportunities, the multiplexity of light-emitting sites remains limited. To overcome this challenge, a silicon-based neural probe harboring nanophotonic circuits has been developed to achieve multiple reconfigurable optical paths after implantation (Fig. 4c) [62]. In this device, optical waveguides with a cross-sectional area of 200 nm × 350 nm are made of a SiN core (RI = 1.996) and a SiO2 cladding (RI = 1.446). To afford tunable configuration, a waveguide mixer is equipped with a localized microheater, acting as a switch, which can tune the RI of the mixer via a local temperature increase. These multimode interferometer (MMI)-based switches can thus steer the direction of light propagation into predefined optical output ports. Furthermore, full integration of nanophotonic and nanoelectronic circuits allows recordings via Pt electrodes, thus suggesting the scalability and complementary metal-oxide-semiconductor (CMOS)-compatibility of this technology.
Wired and wireless LED devices for optogenetics
Besides light delivery via waveguiding devices, LEDs provide an alternative approach for in-vivo optogenetics. One advantage of LEDs arises from their lithographic fabrication, compatible with conventional on-chip input/output (I/O) bonding methods for back-end connection. To this end, Wu et al. reported a four-shank silicon probe with each shank containing eight recording sites (Ti/Ir, 11 μm × 13 μm) and three monolithically integrated blue-light-emitting InGaN microscopic LEDs (μLEDs, 10 μm × 15 μm). Owing to photolithographically defined interconnects, all recording electrodes can be independently addressed and all μLEDs independently controlled (Fig. 5a) [63]. Despite these advantages, monolithically fabricated μLEDs exhibited intrinsic artifacts due to electromagnetic interference and photovoltaic effects, which impaired the temporal resolution of recorded neuronal responses to light stimulation [64]. To alleviate these artifacts, a minimal-stimulation-artifact (miniSTAR) μLED was recently developed by introducing a shielding layer to reduce the electromagnetic interference [65]. Moreover, a heavily boron-doped silicon substrate was used to mitigate the photovoltaic effect. In addition, transient pulse-shaping control further diminished lingering stimulation artifacts. Both in-vitro and in-vivo experiments demonstrated reduced artifacts with miniSTAR μLEDs, thus resulting in high-quality recordings under optogenetic control with high spatiotemporal resolution.
All previous examples, such as optical fibers and wired μLEDs, require a tethered interface for modulated animals. Advanced cable management with optical/electrical commutators has reduced the challenges for studies involving freely moving animals [66]. However, these tethered systems still restrict animal mobility and constrain experimental designs especially in complex and ethologically relevant animal models. To eliminate those constraints, Kim et al. developed a tether-free multifunctional optoelectronic system that integrates GaN microscale inorganic LEDs (μ-ILEDs) with Pt microelectrodes, a microscale inorganic photodetector (μ-IPD) and a temperature microsensor or microheater on a removable injection microneedle (Fig. 5b) [67]. The embedded μ-IPD and temperature sensors can provide real-time feedback on the light intensity of the μ-ILEDs and the local temperature changes, respectively. To power the μ-ILEDs, a wireless module was incorporated via a head-mounted printed circuit board for radiofrequency (RF) scavenging. A fully wireless system was implemented in a conditioned place preference (CPP) experiment in a Y maze, where tested animals learned to self-stimulate their dopaminergic neurons by activating the wireless μ-ILEDs. The self-stimulation behavior is a hallmark assay for validating the efficacy of optogenetic interfaces and identifying reward neurons [68].
The externally mounted headstage antenna demands an exposed cranial surface, thus making it challenging to probe the spinal cord or PNS where a stable tissue/skull interface does not exist. Furthermore, the constant movement in the PNS necessitates miniaturized and wireless light sources for optogenetics therein [55]. To address these challenges, a relatively light-weight, wireless and fully internally implantable optogenetic device has been developed and can be powered by a resonant cavity [69]. This device was deployed to three regions of interest, the premotor cortex in the brain, the dorsal horn of the spinal cord and the peripheral nerve ends in the hind paw. The efficacy of this device for optogenetic control was validated by measuring the animals’ circling behavior and running speed as well as quantifying c-fos expression in immunostained tissue slices. Taken together, these experiments demonstrate the utility of wireless μLED devices for optogenetic modulation in the PNS with minimal footprint and reduced invasiveness.
The mechanical mismatch between fully implantable μLED devices and implanted tissue also hinders their compatibility with specific tissue targets, especially in the context of the natural motions of the subject. To reduce the mechanical stress on the spine and sciatic nerve, the Rogers lab developed stretchable wireless LED devices encapsulated in PDMS, which were implanted in rats to improve the dynamic tissue interface (Fig. 5c) [70]. Specifically, the built-in antennas featured a serpentine pattern to harvest RF power with high stretchability. This stretchable design also ensures reliable device activation within a wide bandwidth for RF harvesting since the relevant range of applied strains only induces a relatively small shift in the center frequency. With this device, the authors demonstrated chronic sciatic nerve activation for six months. The utility of this device was tested by studying the spinal pain pathways, in which the device was implanted over the sciatic nerve and the spinal epidural space. Later, a fully implantable and stretchable optogenetic system utilizing a resistive strain gauge was developed by the same group for closed-loop optogenetic control of bladder functions by real-time monitoring and signal processing [71].
While various wireless designs have been created for optogenetics, fine adjustment of the light intensity of μ-ILEDs at multiple illuminating sites still requires full development. To this end, a fully implantable, battery-free optogenetic system was invented, enabling precise regulation of output intensity and independent operation of four spatially distributed light sources [72]. In particular, the integration of multiple active components, such as a purpose-designed rectifier, a microcontroller and a digital-to-analogue converter (DAC), provided a versatile platform for a set of wirelessly powered optogenetic tools. Specifically, the one-way communication protocol allowed for not only individual control of the μ-ILED devices on two bilateral shanks, but also programmable control of multiple devices within a single trial, via a predetermined pulse length combination for each μ-ILED. More recently, an advanced version of this device was applied to flying species such as songbirds [73].
Additional control of neural activities can be achieved by combining the wireless optoelectronic system with pharmacological delivery via integrated microfluidic channels. A compact paradigm of such a multifunctional and wireless platform was presented by Jeong et al. with a battery-powered wireless optofluidic neural probe (Fig. 5d) [74]. Notably, each microfluidic channel was connected to a separated reservoir that was programmatically activated for pumping biologics via thermal expansion of gating microspheres induced by Joule heating. Using this optofluidic device, the authors delivered a μ-opioid receptor agonist to modulate the animal's behavior via pharmacological intervention. A real-time place preference test was used to validate this wireless multimodal neural interface: wireless optogenetic activation of the VTA-NAc (ventral tegmental area to nucleus accumbens) projection drove a robust real-time place preference, while wireless delivery of a dopamine receptor D1 antagonist abolished the place preference. Many efforts have since been made towards the optimization of the wireless optofluidic system. For example, an electrochemical micropump was engineered to induce volume expansion by water electrolysis, realizing on-demand fluidic delivery via a wireless and fully implantable optofluidic cuff system [75].
Upconversion nanoparticles for optogenetics
Besides optical waveguides and light-emitting devices, upconversion nanoparticles (UCNPs) present another mechanism for delivering light into neural tissue for optogenetics. Specifically, upconversion is an anti-Stokes process realized by the absorption of multiple photons of lower energy (longer wavelengths) followed by the emission of a single photon of higher energy (shorter wavelengths). Upconversion luminescence can be generally produced via two processes: excited-state absorption, which is commonly found in lanthanide ions, and triplet-triplet annihilation, which explains upconversion luminescence in some organic chromophores [76]. The advantages of UCNPs can be seen from the wavelength-dependent scattering coefficients in Fig. 1a and c: scattering of light in the brain follows an inverse relationship with wavelength, thus enabling the deeper penetration of NIR light than its visible counterpart.
Lanthanide-doped UCNPs offer a unique mechanism to convert low-energy incident NIR light into high-energy visible light of different wavelengths for activating various opsin variants. UCNP-enabled optogenetics was first proposed by Deisseroth and Anikeeva in 2011, followed by its first experimental implementation in 2015. The first demonstration was credited to the Yawo lab, who synthesized NaYF4:Sc/Yb/Er UCNPs to convert 975-nm NIR light to 543-nm green light for activating red-shifted channelrhodopsins such as C1V1 and mVChR1 [77]. In another early demonstration, the Han group synthesized dye-sensitized core/active shell UCNPs with sufficient upconversion efficiency, modulating the activity of cultured hippocampal neurons with an 800-nm laser [78]. Under 800-nm irradiation, these UCNPs produced green emission near 550 nm, which activated a red-shifted channelrhodopsin, ReaChR, for modulating the firing activity of cultured hippocampal neurons (Fig. 6a).
Since the advantages of UCNPs lie in their unique ability to convert brain-penetrant NIR light into visible emission, several groups have demonstrated deep-brain optogenetics with either UCNP-containing devices implanted or UCNPs injected into the brain region of interest [22,23,79–81]. Specifically, the Shi group packaged UCNPs in a glass micro-optrode, which acted as an implantable device to convert remotely applied 980-nm NIR light to blue or green light emission from the tip of the micro-optrode in the brain (Fig. 6b) [23]. An impressive accomplishment of this work was the tetherless optical interface enabled by a robotic laser projection system that automatically tracks the animal head during behavioral experiments. The authors also demonstrated multiplexed optogenetic stimulation enabled by doping different ions in their UCNPs. As an example, Tm3+ produces 470-nm blue emission for activating ChR2, while Er3+ produces 540-nm green emission for activating a red-shifted channelrhodopsin, C1V1 [23,79]. The same group also demonstrated wireless optogenetic inhibition with 980-nm irradiation in the rat brain, which was implanted with a device containing core–shell–shell UCNPs with enhanced emission in the 540–570-nm range [80]. More recently, lanthanide-doped upconversion microparticles were also demonstrated to enable fiberless optogenetics with a 976-nm laser in freely behaving mice [81].
Previous in-vivo demonstrations of tetherless optogenetics have been limited to a maximum penetration depth of ca. 2 mm in the rodent brain. This penetration limit is partly owing to the relatively low efficiency of upconversion and the lingering attenuation of NIR light through the brain tissue. To mitigate these challenges, the McHugh and Liu labs optimized the blue-light emission intensity of UCNPs with a core/shell structure comprising a NaYF4:Yb,Tm core and an epitaxially grown NaYF4 shell with an improved NIR-to-visible conversion efficiency of ∼2.5% (Fig. 6c, top) [22]. Owing to this high efficiency, the authors demonstrated a local 470-nm emission intensity of 0.063 mW/mm2 upon an incident 980-nm laser power of 2 W. The authors injected these nanotransducers in the VTA, a deep-brain region located at 4.2 mm underneath the skull (Fig. 6c, bottom). Transcranial delivery of 980-nm light via a skull-fixed optical fiber successfully activated ChR2-expressing dopaminergic neurons, which was validated with fast-scan cyclic voltammetry and c-fos immunostaining. In addition, the authors also demonstrated this approach in multiple neural systems, including successful induction of fear memories in the dentate gyrus of freely behaving mice in a fear-conditioning experiment.
Despite recent advances, several challenges remain for UCNP-based optogenetics. First, upconversion is a multi-photon process with the theoretically maximal quantum yield (QY) of 100/n% (n = number of absorbed photons). Furthermore, due to the non-linear nature of the upconversion process, the QYs of UCNPs decrease monotonically as the power density goes down, with the QYs <0.1% under physiologically permissible light power densities [82,83]. Second, the excitation of UCNPs is largely fixed at a handful of wavelengths, such as 980 nm and 808 nm, due to the intrinsic electronic structures of lanthanide ions [82]. The limited choices of excitation wavelengths impose challenges for deep-brain optogenetics due to a water absorption peak at 980 nm and lingering scattering at 808 nm [12,84]. For example, the 980-nm wavelength corresponds to a local minimum of penetration efficiency (Fig. 1f), leading to merely 1% of the incident photons surviving 4-mm brain tissue. Based on the wavelength-dependent effective penetration depths in the brain (Fig. 1e), we envision that opsins and nanotransducers responsive to 1064-nm NIR light offer the greatest penetration depth in the brain, thus potentially enabling transcranial and transdermal deep-brain optogenetics throughout the entire rodent brain.
Photothermal, photoelectrochemical and photoacoustic neuromodulation
Transgene delivery of opsins in neurons represents a challenge for implementing optogenetics in larger mammals, especially humans. In contrast, it has been known for over a century that light, especially high-energy pulsed light, can control the activity of neurons that do not express light-sensitive proteins [1]. Therefore, developing optical methods to modulate the activity of cells without opsin expression provides an alternative to optogenetics for both fundamental studies and clinical applications. To date, many strategies have been applied to achieve optical neuromodulation in the absence of opsins. In particular, four light-based mechanisms are leveraged for neuromodulation: the optocapacitive effect, photothermal activation of transient receptor potential cation channels (TRPs), photoelectrochemical and photovoltaic effects, and mechanical stimulation via the photoacoustic process.
The optocapacitive effect requires efficient light absorbers and high photon flux from strong light sources to produce usable energy for exciting cells via the capacitive change of the cell membrane. In this process, a rapid and transient temperature increase occurs upon light irradiation, modulating the capacitance of the cell membrane, which is dependent on the rate of temperature change. This capacitance change in turn induces a capacitive current across the membrane, leading to depolarization of an electrogenic cell such as the neuron (Fig. 7a) [85]. In 2012, Shapiro et al. leveraged this mechanism to modulate cultured cells based on the non-specific absorption of water in the infrared [86]. Specifically, water exhibits strong infrared absorption near 1880 nm due to overtone stretching modes [12]. As a result, when a pulsed infrared laser strikes cells cultured in an aqueous environment, the local temperature rapidly increases and drives the capacitive current. In their experiments, the authors demonstrated successful production of capacitive currents induced by transient heating in three models: oocytes, cultured mammalian cells (e.g. human embryonic kidney [HEK] cells) and artificial lipid bilayers, all at varied laser pulse energies.
The abundance of water in biological systems limits the spatial specificity of optocapacitive neuromodulation with lasers of water-absorbing wavelengths. To address this challenge, Carvalho-de-Souza et al. leveraged the plasmonic absorption band of gold nanoparticles (AuNPs) near 523 nm, at which wavelength water absorption is minimum. In this work, AuNPs were conjugated to three proteins including a synthetic Ts1 neurotoxin targeting voltage-gated sodium channels, and antibodies targeting TRPV1 and P2X3 ion channels, in dorsal root ganglion (DRG) neurons [87]. Conjugated AuNPs exhibited stable photothermal performance, thereby reproducibly evoking action potentials even after convective washout. In addition, mouse hippocampal brain slices were mixed with AuNPs to show the feasibility of optocapacitive neuromodulation in an ex-vivo neural tissue model. Facilitated by a voltage-sensitive dye (i.e. indocyanine green (ICG) with infrared fluorescence), the authors validated successful neuromodulation with functionalized AuNPs over a relatively large area in the brain slice.
Owing to their demonstrated biocompatibility and biodegradability, silicon nanostructures represent promising candidates for electrical and optical neuromodulation as well [88]. The Tian group pioneered the use of silicon mesostructures for neuromodulation via the optocapacitive mechanism [85,89,90]. Compared to the AuNPs discussed above, silicon mesostructures are micron-sized particles that prevent receptor-mediated endocytosis. In addition, amorphous silicon benefits from enhanced light absorption, reduced rigidity and resistance to laser-induced particle change. Therefore, silicon mesostructures enable efficient and highly localized optocapacitive modulation of DRG neurons with remotely applied 532-nm laser pulses. In comparison with AuNPs, the authors demonstrated an enhanced photothermal efficacy in capacitive current generation by silicon mesostructures, as evidenced by an energy threshold around 30-fold less than that needed for previously reported AuNPs [85].
The optocapacitive mechanism requires a rapid temperature increase on the order of kilokelvins per second to drive a sufficient transmembrane capacitive current [91]. Such a steep slope of temperature increase precludes its application in vivo due to the limited ability to focus a pulsed laser through the scattering tissues. As an alternative to the non-specific optocapacitive mechanism, researchers have developed photothermal agents that target temperature-sensitive ion channels, such as TRPV1 and TRPA1 [92]. As a proof-of-concept in-vitro demonstration, the Pu group developed organic semiconducting polymer nanoparticles (SPNs) with strong absorption of NIR light (e.g. at 808 nm) [93]. In this work, SPNs were functionalized to bind TRPV1, thus potentially affording molecular specificity with less off-target effects. The authors demonstrated the advantages of SPNs over gold nanorods in terms of their better photothermal performance, characterized by the conversion efficiency, heating capability, stability and efficacy of cell modulation.
Another advantage of semiconducting polymers arises from their tunable bandgaps, which result in a broad range of absorption peaks in the first and second near-infrared windows (NIR-I and NIR-II windows, i.e. 700–900 nm and 1000–1700 nm, respectively) [12]. Taking advantage of biological tissue transparency in the NIR-II window, the Hong group recently reported in-vivo photothermal neuromodulation through the scalp and skull of freely behaving mice without any fiber implantations or head tethering (Fig. 7b) [19]. This advantage was achieved via macromolecular infrared nanotransducers for deep-brain stimulation (MINDS), conjugated polymer nanoparticles coated with an amphiphilic shell. Specifically, MINDS consisted of a semiconducting polymer core for absorbing NIR-II light at 1064 nm and a water-soluble and biocompatible polymer shell. As has been seen in Fig. 1e, 1064 nm is near the global maximum of the effective penetration depth of brain tissue, thus affording the deepest neuromodulation in the brain. The authors performed in-vivo studies to demonstrate the feasibility and specificity of NIR-II neuromodulation in three selected brain regions: the hippocampus, motor cortex and VTA. Specifically, an increase in the neural firing rate upon NIR-II illumination in the hippocampus via simultaneous electrophysiological recording was observed. Furthermore, behavior manipulation using MINDS and NIR-II light was also demonstrated in free-moving mice, exhibiting unilateral circling and CPP with photothermal neuromodulation in the secondary motor cortex and VTA, respectively. The sufficiency and necessity of TRPV1, MINDS and 1064-nm light were validated with electrophysiology, immunostaining and behavioral experiments. Notably, this demonstration represents the first time that photothermal neuromodulation has been applied in the deep brain of behaving animals in an implant-free and tether-free manner.
The energy conversion efficiency is a key parameter for all photothermal neuromodulation methods and is defined as the ratio of the generated thermal energy to the absorbed photon energy. Among an array of photothermal materials, the reported efficiency is distributed in a wide range from 19.2% to 98.9% [19,94,95]. However, the reader should be cautioned that a high photothermal efficiency does not imply a significant increase in temperature due to thermal diffusion to the environment where the photothermal agent is located. Specifically, to model the spatial distribution of temperatures with temporal dynamics, the Pennes bioheat equation should be solved numerically in live biological tissues [96]. In addition, since the photothermal efficiency only describes that of an internal process (akin to the ‘internal’ quantum efficiency of photovoltaics), the absorption cross section needs to be included when considering potential photothermal candidates. As a rule of thumb, materials that strongly absorb incident light and efficiently generate heat after absorption should be used for photothermal neuromodulation.
In addition to neuromodulation with the photothermal effect, the photoelectrochemical process, which generates Faradaic current under light illumination of certain materials, has also been leveraged for neuromodulation (Fig. 7c) [97]. For example, coaxial p-type/intrinsic/n-type (PIN) Si nanowires (SiNWs) with atomic Au on the surfaces were synthesized by the Tian group via a chemical vapor deposition process. Upon 532-nm light illumination, these hybrid nanostructures produced cathodic photocurrents sufficient to depolarize a targeted neuron. Specifically, electrons and holes migrated to the n-type shell and p-type core, respectively, thus driving Faradaic currents in a loop in response to light stimulation. Using these devices, the authors successfully induced action potentials in rat DRG neurons, which shared a similar waveform with those triggered by conventional current injection. To elucidate the mechanism of this method, the authors first ruled out the thermal effect as the main contributor by measuring a small temperature increase during laser illumination.
Taking a step further, the Tian group synthesized a series of silicon-based interfaces leveraging photoelectric (i.e. capacitive and Faradaic) and photothermal effects for multiscale interfacing with an array of biological systems including organelles, cultured neurons, brain slices and the live mouse brain [89]. On a subcellular level, SiNWs were found to not only serve as stimulators for modulating intracellular calcium, but also act as transport markers for labeling motor protein kinetics with 592-nm laser illumination. On a cellular level, extracellular photostimulation of 532-nm laser pulses successfully produced action potentials in cultured DRG neurons using a planar PIN junction in a Si diode. Additionally, on a tissue level, 473-nm photostimulation was used to produce excitatory postsynaptic currents (EPSCs) in ex-vivo brain slices via an array of PIN junctions in a mesh architecture. Finally, on the organism level, the authors successfully demonstrated 473-nm photostimulation of neural activity in the live mouse brain via an Au-coated PIN Si mesh array. This work represents a tour-de-force of non-genetic neuromodulation with Si nanostructures as multiscale optical neural interfaces based on a wide array of light-induced physical effects [89,98].
Beyond the mechanisms discussed above, photoacoustic stimulation represents an emerging technique for neuromodulation. Compared with conventional ultrasound neuromodulation methods [99], photoacoustic stimulation benefits from the much higher spatial resolution of light illumination (Fig. 7d) [100]. In one of the early demonstrations, optical fibers were used to deliver laser pulses, which led to efficient heat generation of light absorbing materials (e.g. graphite) and thermal expansion of the surrounding material (e.g. epoxy) with high spatial precision [101]. As a result, acoustic waves were generated to activate neurons, as evidenced by calcium imaging in vitro and electrophysiological recordings in vivo. A behavioral assay was also used to evaluate the evoked motor response and validate the high spatial resolution of this approach.
Besides high spatial resolution, photoacoustic neuromodulation also offers a unique tool for investigating the mechanism underlying ultrasound neurostimulation. Very recently, photoacoustic nanotransducers (PANs) comprising semiconducting polymer nanoparticles were developed for generating localized ultrasound upon NIR-II irradiation [100]. PANs were conjugated with antibodies to specifically target mechanosensitive TRPV4 channels, thus helping elucidate the mechanism of ultrasound modulation via activating these ion channels. In another recent work, a tapered fiber optoacoustic emitter (TFOE) was engineered for single-neuron photoacoustic stimulation [102]. The TFOE was fabricated by coating the fiber tip with carbon nanotubes (CNTs) in a PDMS matrix. Subcellular stimulation of neurites was demonstrated by 1030-nm laser pulses, which produced calcium wave propagation as observed in fluorescence imaging. Owing to the elimination of an ultrasound input, TFOE permitted simultaneous intracellular recording via the standard patch clamp technique, since light induces minimal mechanical disruption of the ‘gigaseal’ between the patch pipette and the cell membrane.
Ultrasound- and magnetic-field-mediated light sources for optogenetics
All approaches discussed above rely either on light directly or on the secondary effects of light (e.g. photothermal, photoelectrochemical and photoacoustic). Nonetheless, a long-standing challenge of in-vivo light delivery remains unresolved; that is, due to the intrinsic scattering and absorption of light in biological tissues, invasive procedures are usually required to remove overlying tissues and implant optical fibers. Although red-shifted opsins (e.g. ChRmine) and NIR-responsive materials (e.g. UCNPs and MINDS) have demonstrated transcranial and even transdermal deep-brain stimulation with reduced invasiveness, transgene and nanoparticle delivery usually still requires invasive intracranial injection [19,21,22]. Addressing these challenges requires two biophotonic innovations: first, a nanotransducer is needed to convert tissue-penetrant energy, such as ultrasound and magnetic fields, into localized light emission; second, these nanotransducers are preferably delivered via intravenous instead of intracranial injection, since intravenous injection represents one of the simplest and least invasive methods for administering pharmacologics into the body.
In the first attempt to produce localized emission with ultrasound, the Hong lab introduced the concept of ‘sono-optogenetics’ and demonstrated the proof-of-concept of this approach in live mice via a minimally invasive ultrasound interface [103]. Implementing this approach requires fulfilling the two biophotonic innovations above. On one hand, a nanotransducer is needed to convert tissue-penetrant ultrasound into localized light emission. To this end, the authors used defect engineering strategies to make zinc sulfide (ZnS) nanoparticles co-doped with trace amounts of silver (Ag) and cobalt (Co) ions (ZnS:Ag,Co) to afford a unique mechanoluminescence property. Upon 400-nm photoexcitation, Co dopant ions trap the excited electrons and store the photoexcitation energy until they are triggered by ultrasound. Ultrasound depletes the stored energy in the trap presumably via a piezoelectricity-induced mechanism and releases the energy as light emission. Finally, Ag dopant ions tune the emission color to 470-nm blue light to activate ChR2. In so doing, ZnS:Ag,Co mechanoluminescent nanoparticles can convert ultrasound, which has deeper tissue penetration (>1 cm for 1.5 MHz ultrasound vs. <1 mm for 470 nm light), to local light emission after photoexcitation, thus allowing us to metaphorically ‘see the sound’ (Fig. 8a, left) [104]. On the other hand, these nanotransducers can be delivered in vivo with minimal invasiveness by leveraging a unique endogenous mechanism offered by the circulatory system in the mammalian body. Specifically, after ZnS:Ag,Co NPs are administered intravenously, they can be recharged by photoexcitation in superficial vessels near the skin, pumped by the heart into deep-tissue vessels, and then gated by focused ultrasound to release the stored energy as visible light, all without exiting blood circulation (Fig. 8a, right). Using this circulation-delivered light source enabled by mechanoluminescent nanoparticles, the authors were able to evoke hindlimb motion synchronized with focused ultrasound (FUS) pulses in mice expressing ChR2 in the motor cortex.
Besides ultrasound, the Wang and Chang groups have demonstrated the production of light emission from mechanoluminescent materials by a rotating magnetic field [105]. A magneto-luminescence microdevice was developed, composed of terbium-doped calcium zinc oxysulfide (CaZnOS:Tb) crystals and a magnet bar (Fig. 8b). Under a rotating magnetic field, the magnet rotates in response and thus exerts mechanical stress on the CaZnOS:Tb crystals, producing light emission at 544 nm. This green emission arises from Tb3+ dopants and matches the activation spectrum of two optogenetic proteins, KillerRed and C1V1. Specifically, KillerRed is a genetically encoded protein that produces reactive oxygen species under green-light irradiation, thus inducing cancer cell death with remotely applied magnetic fields in both in-vitro and in-vivo conditions. Furthermore, C1V1 is a red-shifted channelrhodopsin that can be excited by the magneto-luminescence microdevice, enabling the authors to modulate neuron activity and animal behaviors in a wireless manner.
Since ultrasound- and magnetic-field-mediated light sources rely on energy stored in defects, such as electron and hole traps in the material, the density of these traps determines the theoretical limit on maximum energy that can be released by incident ultrasound or an external magnetic field. Although the exact mechanism of mechanoluminescence and the nature of energy-storing traps are still to be elucidated [106], the energy storage capacity of trap-engineered materials can be determined via time-dependent luminescence measurements in an integrating sphere. In one of the few studies of this kind, it was found that only ∼1.6% of the dopant, Eu2+ ions in a mechanoluminescent material, SrAl2O4:Eu,Dy, participates in energy storage [107]. Therefore, developing brighter light sources gated by ultrasound and magnetic fields depends on advancements that will produce more efficient optical storage materials. Taken together, both sono-optogenetics and magneto-luminescence microdevices demonstrate a new paradigm of in-vivo light delivery in deep tissues for minimally invasive optogenetics. We envision that the same light delivery approaches can be used in conjunction with other optical neuromodulation approaches and any biomedical applications that need a light source deep inside the body.
SUMMARY AND OUTLOOK
Optical neuromodulation methods have transformed neuroscience and are continuing to inform new therapies for neurological and psychiatric diseases. Compared to the neural interfaces of other modalities, optical approaches benefit from diffraction-limited high spatial resolution, excellent temporal precision, the ability to propagate in free space, wavelength multiplexity, and orthogonality with electrical approaches. Owing to the unique ability of optogenetics to target distinct cell populations, spatiotemporally controlled illumination methods offer unique opportunities to dissect the neural circuits driving specific behaviors and disorders by turning neurons on and off with cell-type specificity.
An important consideration for choosing the appropriate optical methods for neuromodulation arises from the complex biological environments of various organs and structures in the central nervous system (CNS) and PNS. To guide the selection and further development of optical neuromodulation approaches, we summarize design considerations for suitable optical approaches in the context of technical challenges specific to each listed organ (Table 1). Additionally, we also point out some exciting new directions in optical neuromodulation methods in the following paragraphs.
Table 1.
Nervous system | Organ | Reported optical approaches | Technical challenges | Design considerations |
---|---|---|---|---|
CNS | Brain | • Conventional optical fibers• Multifunctional fibers• Optical waveguides• Wired and wireless LED devices• Upconversion nanoparticles• Ultrasound- and magnetic-field-mediated light sources• Photothermal, photoelectrochemical and photoacoustic materials | • Dense neuronal density requires optical control with high spatial resolution.• Regions of interest are distributed in 3D, especially in the depth direction. The depth distribution of different brain targets imposes challenges to conventional illumination methods due to limited tissue penetration of visible light.• The interconnection between brain regions requires multisite optical control, ideally via a light source that can be repositioned in operando.• Neuroplasticity in the brain requires a chronically stable optical interface. | • Fine spatial control can be realized via taper optical fibers, spatially expandable fiber probes and reconfigurable nanophotonic silicon probes.• Ultrasound and NIR-II light offer deeper brain penetration for neuromodulation.• Focused ultrasound may facilitate raster scanning of the light emission spot inside the brain.• Flexible optoelectronics with reduced footprint enable a chronically stable optical interface for long-term neuromodulation. |
Spinal cord | • Conventional optical fibers• Multifunctional fibers• Wireless LED devices• Upconversion nanoparticles | • Mechanical motion is associated with natural movement of the subject.• The lack of the skull prevents a securely cemented interface. | • Flexible optoelectronics are required to conform to the curvilinear morphology of the spinal cord.• Stretchable optoelectronics accommodate mechanical motion.• Wireless interfaces are preferred due to the lack of an exposed surface for cementing a wired device. | |
PNS | Heart | • Conventional optical fibers• Wired and wireless LED devices | • Multiple neurons distributed over a large area need to be stimulated simultaneously for optogenetic pacing.• Cardiac motion may lead to an unstable optical interface. | • Stretchable film array with densely packed LEDs can conform to the surface of the myocardium. |
Bladder | • Conventional optical fibers• Wireless LED devices | • The dimensional change of the bladder due to its activity needs to be accommodated by the optical interface. | • Stretchable optoelectronics can accommodate the bladder volume change. | |
Gut | • Wireless μLED devices• External light sources through an abdominal window | • Neurons are sparsely distributed through a large volume of abdominal tissue.• Constant bowel movements exist, thus placing challenges for a stable interface.• Different neuron types are segregated at different locations along the gastrointestinal tract, thus requiring region specificity of illumination. | • An internal light source is needed for a broad illumination area.• A relocatable light source is preferred to screen different neuron types at distinct locations in the gut.• A stretchable and conformal optoelectronic interface is desired to accommodate bowel movements. |
We envision new methods for delivering light into the deep brain and the PNS with reduced invasiveness. Despite recent advances in transcranial optogenetics [19,21,22], it remains challenging to deliver light specifically to a deep-brain region while sparing all underlying brain tissue from irradiation. In addition, many challenges remain for optogenetics in the PNS. Specifically, the sparse distribution of neurons and nerve terminals over large areas, constant movement of the tissue, and the lack of cranial bone all represent technical challenges of optogenetic stimulation in the PNS [55,56]. Therefore, new methods for light delivery in vivo, especially those that allow for repositioning the illuminated region during experiments, will significantly benefit neuroscience studies in the PNS, such as in the gut [108,109]. Besides sono-optogenetics [103,104], which enables a circulation-delivered light source gated by external ultrasound, other internal light sources such as luciferase/luciferin offer another strategy of light delivery for optogenetics without an exogenous device [110]. Furthermore, ultrasound interference offers an elegant approach to modulating the local density of biological tissues, thus realizing a virtual GRIN lens for non-invasive light delivery. New knowledge about the complex interaction between tissue-penetrant energy (ultrasound [104], infrared light [19], X-ray [111] and radiofrequency [112,113]) and matter (endogenous tissues and exogenous materials) thus offers unique opportunities for next-generation in-vivo optical neuromodulation approaches.
We envision new methods will sensitize neurons to incident light, especially tissue-penetrant wavelengths in the NIR-I and NIR-II spectra, via endogenous and genetically encoded mechanisms. Although red-shifted opsin variants have been developed with sensitivity up to 635 nm, multi–vibrational-mode thermal statistics prohibit finding infrared-responsive opsins activated by longer wavelengths, at least in nature [114]. Nonetheless, inspired by the molecular basis of infrared detection by snakes, heat-sensitive transient receptor potential vanilloid (TRPV) ion channels, which can be genetically encoded in specific cell types, offer an alternative mechanism of infrared sensitivity [115]. This approach has already been demonstrated in transdermal NIR-II neuromodulation in the brain and engineered NIR vision in the retina [19,116]. In addition, to sensitize the neural tissue to photothermal stimulation, endogenous chromophores such as dopamine and melanin may be leveraged to produce efficient light absorbers in neurons with cell-type specificity, thus affording selective photothermal neuromodulation therein.
Besides these new possibilities, we also highlight the fundamental physical limitations when developing new materials and approaches for optical neural interfacing. The interaction between light and biological tissues, as discussed at the beginning of this review, dictates the limits in penetration depth and permissible power density for optical neuromodulation. Specifically, Fig. 1e reveals that a global maximum exists near 1070 nm for the effective attenuation length in the brain, which results from the compromise between decreasing scattering and increasing absorption at longer wavelengths. Therefore, the penetration depth of optical neuromodulation methods cannot be further optimized by going beyond the NIR-II window to reduce scattering according to Equation S12 (see Supplementary Data). This fundamental constraint limits the application of light-based approaches in larger-brained subjects such as humans without resorting to an invasive means. In addition, exogenous light-emitting devices cannot be miniaturized indefinitely due to size limitations: for example, as the size of μ-ILEDs shrinks, it becomes more challenging to epitaxially grow the light-emitting material on a smaller mesa, in addition to a higher density of defects that significantly impact the device performance [117]. As another example, mechanoluminescent nanoparticles emit light in response to ultrasound-induced mechanical stress, with the maximum light intensity theoretically limited by the number of trap energy states in the host materials [106]. Therefore, one must consider the physical principles governing neuromodulation when designing new optical neural interfaces with smaller footprints and higher throughput.
Supplementary Material
Contributor Information
Shan Jiang, Department of Materials Science and Engineering, Stanford University, Stanford, CA 94305, USA; Wu Tsai Neurosciences Institute, Stanford University, Stanford, CA 94305, USA.
Xiang Wu, Department of Materials Science and Engineering, Stanford University, Stanford, CA 94305, USA; Wu Tsai Neurosciences Institute, Stanford University, Stanford, CA 94305, USA.
Nicholas J Rommelfanger, Wu Tsai Neurosciences Institute, Stanford University, Stanford, CA 94305, USA; Department of Applied Physics, Stanford University, Stanford, CA 94305, USA.
Zihao Ou, Department of Materials Science and Engineering, Stanford University, Stanford, CA 94305, USA; Wu Tsai Neurosciences Institute, Stanford University, Stanford, CA 94305, USA.
Guosong Hong, Department of Materials Science and Engineering, Stanford University, Stanford, CA 94305, USA; Wu Tsai Neurosciences Institute, Stanford University, Stanford, CA 94305, USA.
FUNDING
G.H. acknowledges the support by a National Institutes of Health (NIH) Pathway to Independence Award (National Institute on Aging 5R00AG056636-04), a National Science Foundation (NSF) CAREER Award (2045120), a Rita Allen Foundation Scholars Award, a Beckman Technology Development Grant, a gift from the Spinal Muscular Atrophy (SMA) Foundation, a gift from the Pinetops Foundation, two seed grants from the Wu Tsai Neurosciences Institute, and a seed grant from the Bio-X Initiative of Stanford University. X.W. acknowledges the support by a Stanford Graduate Fellowship. N.J.R. acknowledges support from the NSF Graduate Research Fellowships Program (GRFP) and a Stanford Bio-X fellowship. Z.O. acknowledges the support by a Wu Tsai Neurosciences Institute Interdisciplinary Scholar Award.
Conflict of interest statement. None declared.
REFERENCES
- 1. d’Arsonval MA. Action physiologique des courants alternatifs. CR Soc Biol 1891; 43: 283–6. [Google Scholar]
- 2. Fork RL. Laser stimulation of nerve cells in Aplysia. Science 1971; 171: 907–8. 10.1126/science.171.3974.907 [DOI] [PubMed] [Google Scholar]
- 3. Callaway EM, Katz LC. Photostimulation using caged glutamate reveals functional circuitry in living brain slices. Proc Natl Acad Sci USA 1993; 90: 7661–5. 10.1073/pnas.90.16.7661 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Lima SQ, Miesenböck G. Remote control of behavior through genetically targeted photostimulation of neurons. Cell 2005; 121: 141–52. 10.1016/j.cell.2005.02.004 [DOI] [PubMed] [Google Scholar]
- 5. Zemelman BV, Nesnas N, Lee GAet al. . Photochemical gating of heterologous ion channels: remote control over genetically designated populations of neurons. Proc Natl Acad Sci USA 2003; 100: 1352–7. 10.1073/pnas.242738899 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Boyden ES, Zhang F, Bamberg Eet al. . Millisecond-timescale, genetically targeted optical control of neural activity. Nat Neurosci 2005; 8: 1263–8. 10.1038/nn1525 [DOI] [PubMed] [Google Scholar]
- 7. Zemelman BV, Lee GA, Ng Met al. . Selective photostimulation of genetically chARGed neurons. Neuron 2002; 33: 15–22. 10.1016/S0896-6273(01)00574-8 [DOI] [PubMed] [Google Scholar]
- 8. Yu T, Zhu J, Li Det al. . Physical and chemical mechanisms of tissue optical clearing. iScience 2021; 24: 102178. 10.1016/j.isci.2021.102178 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Antonov IP, Goroshkov AV, Kalyunov VNet al. . Measurement of the radial distribution of the refractive index of the Schwann's sheath and the axon of a myelinated nerve fiber in vivo. J Appl Spectrosc 1983; 39: 822–4. 10.1007/BF00662830 [DOI] [Google Scholar]
- 10. Yaroslavsky AN, Schulze PC, Yaroslavsky IVet al. . Optical properties of selected native and coagulated human brain tissues in vitro in the visible and near infrared spectral range. Phys Med Biol 2002; 47: 2059–73. 10.1088/0031-9155/47/12/305 [DOI] [PubMed] [Google Scholar]
- 11. Wang L, Jacques SL, Zheng L. MCML—Monte Carlo modeling of light transport in multi-layered tissues. Comput Methods Programs Biomed 1995; 47: 131–46. 10.1016/0169-2607(95)01640-F [DOI] [PubMed] [Google Scholar]
- 12. Hong G, Antaris AL, Dai H. Near-infrared fluorophores for biomedical imaging. Nat Biomed Eng 2017; 1: 0010. 10.1038/s41551-016-0010 [DOI] [Google Scholar]
- 13. Grieco C, Kohl FR, Hanes ATet al. . Probing the heterogeneous structure of eumelanin using ultrafast vibrational fingerprinting. Nat Commun 2020; 11: 4569. 10.1038/s41467-020-18393-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Stujenske JM, Spellman T, Gordon JA. Modeling the spatiotemporal dynamics of light and heat propagation for in vivo optogenetics. Cell Rep 2015; 12: 525–34. 10.1016/j.celrep.2015.06.036 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Owen SF, Liu MH, Kreitzer AC. Thermal constraints on in vivo optogenetic manipulations. Nat Neurosci 2019; 22: 1061–5. 10.1038/s41593-019-0422-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Klapoetke NC, Murata Y, Kim SSet al. . Independent optical excitation of distinct neural populations. Nat Methods 2014; 11: 338–46. 10.1038/nmeth.2836 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Zhao S, Cunha C, Zhang Fet al. . Improved expression of halorhodopsin for light-induced silencing of neuronal activity. Brain Cell Bio 2008; 36: 141–54. 10.1007/s11068-008-9034-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Pinto Cardozo DF, Lammel S. Hot topic in optogenetics: new implications of in vivo tissue heating. Nat Neurosci 2019; 22: 1039–41. 10.1038/s41593-019-0426-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Wu X, Jiang Y, Rommelfanger NJet al. . Through-scalp deep-brain stimulation in tether-free, naturally-behaving mice with widefield NIR-II illumination. bioRxiv: 2020.10.21.348037. [Google Scholar]
- 20. Marshel JH, Kim YS, Machado TAet al. . Cortical layer–specific critical dynamics triggering perception. Science 2019; 365: eaaw5202. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Chen R, Gore F, Nguyen Q-Aet al. . Deep brain optogenetics without intracranial surgery. Nat Biotechnol 2020; 39: 161–4. doi: 10.1038/s41587-020-0679-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Chen S, Weitemier AZ, Zeng Xet al. . Near-infrared deep brain stimulation via upconversion nanoparticle–mediated optogenetics. Science 2018; 359: 679–84. 10.1126/science.aaq1144 [DOI] [PubMed] [Google Scholar]
- 23. Wang Y, Lin X, Chen Xet al. . Tetherless near-infrared control of brain activity in behaving animals using fully implantable upconversion microdevices. Biomaterials 2017; 142: 136–48. 10.1016/j.biomaterials.2017.07.017 [DOI] [PubMed] [Google Scholar]
- 24. Genina EA, Bashkatov AN, Tuchina DKet al. . Optical properties of brain tissues at the different stages of glioma development in rats: pilot study. Biomed Opt Express 2019; 10: 5182–97. 10.1364/BOE.10.005182 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Fenno L, Yizhar O, Deisseroth K. The development and application of optogenetics. Annu Rev Neurosci 2011; 34: 389–412. 10.1146/annurev-neuro-061010-113817 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Guru A, Post RJ, Ho Y-Yet al. . Making sense of optogenetics. Int J Neuropsychopharmacol 2015; 18: yv079. 10.1093/ijnp/pyv079 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Parr-Brownlie LC, Bosch-Bouju C, Schoderboeck Let al. . Lentiviral vectors as tools to understand central nervous system biology in mammalian model organisms. Front Mol Neurosci 2015; 8: 14. 10.3389/fnmol.2015.00014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Haery L, Deverman BE, Matho KSet al. . Adeno-associated virus technologies and methods for targeted neuronal manipulation. Front Neuroanat 2019; 13: 93. 10.3389/fnana.2019.00093 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Burger C, Gorbatyuk OS, Velardo MJet al. . Recombinant AAV viral vectors pseudotyped with viral capsids from serotypes 1, 2, and 5 display differential efficiency and cell tropism after delivery to different regions of the central nervous system. Mol Ther 2004; 10: 302–17. 10.1016/j.ymthe.2004.05.024 [DOI] [PubMed] [Google Scholar]
- 30. Huber D, Petreanu L, Ghitani Net al. . Sparse optical microstimulation in barrel cortex drives learned behaviour in freely moving mice. Nature 2008; 451: 61–4. 10.1038/nature06445 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Aravanis AM, Wang L-P, Zhang Fet al. . An optical neural interface: in vivo control of rodent motor cortex with integrated fiberoptic and optogenetic technology. J Neural Eng 2007; 4: S143–56. 10.1088/1741-2560/4/3/S02 [DOI] [PubMed] [Google Scholar]
- 32. Zhang F, Gradinaru V, Adamantidis ARet al. . Optogenetic interrogation of neural circuits: technology for probing mammalian brain structures. Nat Protoc 2010; 5: 439–56. 10.1038/nprot.2009.226 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Gradinaru V, Thompson KR, Zhang Fet al. . Targeting and readout strategies for fast optical neural control in vitro and in vivo. J Neurosci 2007; 27: 14231–8. 10.1523/JNEUROSCI.3578-07.2007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Armstrong C, Krook-Magnuson E, Oijala Met al. . Closed-loop optogenetic intervention in mice. Nat Protoc 2013; 8: 1475–93. 10.1038/nprot.2013.080 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Anikeeva P, Andalman AS, Witten Iet al. . Optetrode: a multichannel readout for optogenetic control in freely moving mice. Nat Neurosci 2012; 15: 163–70. 10.1038/nn.2992 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Salatino JW, Ludwig KA, Kozai TDYet al. . Glial responses to implanted electrodes in the brain. Nat Biomed Eng 2017; 1: 862–77. 10.1038/s41551-017-0154-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Zhao Z, Luan L, Wei Xet al. . Nanoelectronic coating enabled versatile multifunctional neural probes. Nano Lett 2017; 17: 4588–95. 10.1021/acs.nanolett.7b00956 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Pisanello F, Mandelbaum G, Pisanello Met al. . Dynamic illumination of spatially restricted or large brain volumes via a single tapered optical fiber. Nat Neurosci 2017; 20: 1180–8. 10.1038/nn.4591 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Pisanello F, Sileo L, Oldenburg IAet al. . Multipoint-emitting optical fibers for spatially addressable in vivo optogenetics. Neuron 2014; 82: 1245–54. 10.1016/j.neuron.2014.04.041 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Lin JY, Knutsen PM, Muller Aet al. . ReaChR: a red-shifted variant of channelrhodopsin enables deep transcranial optogenetic excitation. Nat Neurosci 2013; 16: 1499–508. 10.1038/nn.3502 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Chuong AS, Miri ML, Busskamp Vet al. . Noninvasive optical inhibition with a red-shifted microbial rhodopsin. Nat Neurosci 2014; 17: 1123–9. 10.1038/nn.3752 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Zhang F, Prigge M, Beyrière Fet al. . Red-shifted optogenetic excitation: a tool for fast neural control derived from Volvox carteri. Nat Neurosci 2008; 11: 631–3. 10.1038/nn.2120 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Bedbrook CN, Yang KK, Robinson JEet al. . Machine learning-guided channelrhodopsin engineering enables minimally invasive optogenetics. Nat Methods 2019; 16: 1176–84. 10.1038/s41592-019-0583-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Gong X, Mendoza-Halliday D, Ting JTet al. . An ultra-sensitive step-function opsin for minimally invasive optogenetic stimulation in mice and macaques. Neuron 2020; 107: 197. 10.1016/j.neuron.2020.06.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Chan KY, Jang MJ, Yoo BBet al. . Engineered AAVs for efficient noninvasive gene delivery to the central and peripheral nervous systems. Nat Neurosci 2017; 20: 1172–9. 10.1038/nn.4593 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. LeChasseur Y, Dufour S, Lavertu Get al. . A microprobe for parallel optical and electrical recordings from single neurons in vivo. Nat Methods 2011; 8: 319–25. 10.1038/nmeth.1572 [DOI] [PubMed] [Google Scholar]
- 47. Canales A, Jia X, Froriep UPet al. . Multifunctional fibers for simultaneous optical, electrical and chemical interrogation of neural circuits in vivo. Nat Biotechnol 2015; 33: 277–84. 10.1038/nbt.3093 [DOI] [PubMed] [Google Scholar]
- 48. Park S, Guo Y, Jia Xet al. . One-step optogenetics with multifunctional flexible polymer fibers. Nat Neurosci 2017; 20: 612–9. 10.1038/nn.4510 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49. Yan W, Dong C, Xiang Yet al. . Thermally drawn advanced functional fibers: new frontier of flexible electronics. Mater Today 2020; 35: 168–94. 10.1016/j.mattod.2019.11.006 [DOI] [Google Scholar]
- 50. Guo Y, Werner CF, Canales Aet al. . Polymer-fiber-coupled field-effect sensors for label-free deep brain recordings. PLoS One 2020; 15: e0228076. 10.1371/journal.pone.0228076 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Jiang S, Patel DC, Kim Jet al. . Spatially expandable fiber-based probes as a multifunctional deep brain interface. Nat Commun 2020; 11: 6115. 10.1038/s41467-020-19946-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52. Jiang S, Song J, Zhang Yet al. . Nano-optoelectrodes integrated with flexible multifunctional fiber probes by high-throughput scalable fabrication. ACS Appl Mater Interfaces 2021; 13: 9156–65. 10.1021/acsami.0c19187 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53. Park S, Yuk H, Zhao Ret al. . Adaptive and multifunctional hydrogel hybrid probes for long-term sensing and modulation of neural activity. Nat Commun 2021; 12: 3435. 10.1038/s41467-021-23802-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Lu C, Park S, Richner TJet al. . Flexible and stretchable nanowire-coated fibers for optoelectronic probing of spinal cord circuits. Sci Adv 2017; 3: e1600955. 10.1126/sciadv.1600955 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Montgomery KL, Iyer SM, Christensen AJet al. . Beyond the brain: optogenetic control in the spinal cord and peripheral nervous system. Sci Transl Med 2016; 8: 337rv5. doi: 10.1126/scitranslmed.aad7577 [DOI] [PubMed] [Google Scholar]
- 56. Shahriari D, Rosenfeld D, Anikeeva P. Emerging frontier of peripheral nerve and organ interfaces. Neuron 2020; 108: 270–85. 10.1016/j.neuron.2020.09.025 [DOI] [PubMed] [Google Scholar]
- 57. Zhang Y, Li X, Kim Jet al. . Thermally drawn stretchable electrical and optical fiber sensors for multimodal extreme deformation sensing. Adv Optical Mater 2021; 9: 2001815. 10.1002/adom.202001815 [DOI] [Google Scholar]
- 58. Lee J, Ozden I, Song Y-Ket al. . Transparent intracortical microprobe array for simultaneous spatiotemporal optical stimulation and multichannel electrical recording. Nat Methods 2015; 12: 1157–62. 10.1038/nmeth.3620 [DOI] [PubMed] [Google Scholar]
- 59. Wu F, Stark E, Im Met al. . An implantable neural probe with monolithically integrated dielectric waveguide and recording electrodes for optogenetics applications. J Neural Eng 2013; 10: 056012. 10.1088/1741-2560/10/5/056012 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60. Kampasi K, Stark E, Seymour Jet al. . Fiberless multicolor neural optoelectrode for in vivo circuit analysis. Sci Rep 2016; 6: 30961. 10.1038/srep30961 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61. Kampasi K, English DF, Seymour Jet al. . Dual color optogenetic control of neural populations using low-noise, multishank optoelectrodes. Microsyst Nanoeng 2018; 4: 10. 10.1038/s41378-018-0009-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62. Mohanty A, Li Q, Tadayon MAet al. . Reconfigurable nanophotonic silicon probes for sub-millisecond deep-brain optical stimulation. Nat Biomed Eng 2020; 4: 223–31. 10.1038/s41551-020-0516-y [DOI] [PubMed] [Google Scholar]
- 63. Wu F, Stark E, Ku P-Cet al. . Monolithically integrated μLEDs on silicon neural probes for high-resolution optogenetic studies in behaving animals. Neuron 2015; 88: 1136–48. 10.1016/j.neuron.2015.10.032 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64. Hong G, Lieber CM. Novel electrode technologies for neural recordings. Nat Rev Neurosci 2019; 20: 330–45. 10.1038/s41583-019-0140-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65. Kim K, Vöröslakos M, Seymour JPet al. . Artifact-free and high-temporal-resolution in vivo opto-electrophysiology with microLED optoelectrodes. Nat Commun 2020; 11: 2063. 10.1038/s41467-020-15769-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66. Sidor MM, Davidson TJ, Tye KMet al. . in vivo optogenetic stimulation of the rodent central nervous system. J Vis Exp 2015; 95: 51483. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67. Kim T-I, McCall JG, Jung YHet al. . Injectable, cellular-scale optoelectronics with applications for wireless optogenetics. Science 2013; 340: 211–6. 10.1126/science.1232437 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68. Schultz W. Neuronal reward and decision signals: from theories to data. Physiol Rev 2015; 95: 853–951. 10.1152/physrev.00023.2014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69. Montgomery KL, Yeh AJ, Ho JSet al. . Wirelessly powered, fully internal optogenetics for brain, spinal and peripheral circuits in mice. Nat Methods 2015; 12: 969–74. 10.1038/nmeth.3536 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70. Park SI, Brenner DS, Shin Get al. . Soft, stretchable, fully implantable miniaturized optoelectronic systems for wireless optogenetics. Nat Biotechnol 2015; 33: 1280–6. 10.1038/nbt.3415 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71. Mickle AD, Won SM, Noh KNet al. . A wireless closed-loop system for optogenetic peripheral neuromodulation. Nature 2019; 565: 361–5. 10.1038/s41586-018-0823-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72. Gutruf P, Krishnamurthi V, Vázquez-Guardado Aet al. . Fully implantable optoelectronic systems for battery-free, multimodal operation in neuroscience research. Nat Electron 2018; 1: 652–60. 10.1038/s41928-018-0175-0 [DOI] [Google Scholar]
- 73. Ausra J, Munger SJ, Azami Aet al. . Wireless battery free fully implantable multimodal recording and neuromodulation tools for songbirds. Nat Commun 2021; 12: 1968. 10.1038/s41467-021-22138-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74. Jeong J-W, McCall JG, Shin Get al. . Wireless optofluidic systems for programmable in vivo pharmacology and optogenetics. Cell 2015; 162: 662–74. 10.1016/j.cell.2015.06.058 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75. Zhang Y, Mickle AD, Gutruf Pet al. . Battery-free, fully implantable optofluidic cuff system for wireless optogenetic and pharmacological neuromodulation of peripheral nerves. Sci Adv 2019; 5: eaaw5296. 10.1126/sciadv.aaw5296 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76. Zhou J, Liu Q, Feng Wet al. . Upconversion luminescent materials: advances and applications. Chem Rev 2015; 115: 395–465. 10.1021/cr400478f [DOI] [PubMed] [Google Scholar]
- 77. Hososhima S, Yuasa H, Ishizuka Tet al. . Near-infrared (NIR) up-conversion optogenetics. Sci Rep 2015; 5: 16533. 10.1038/srep16533 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78. Wu X, Zhang Y, Takle Ket al. . Dye-sensitized core/active shell upconversion nanoparticles for optogenetics and bioimaging applications. ACS Nano 2016; 10: 1060–6. 10.1021/acsnano.5b06383 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79. Lin X, Wang Y, Chen Xet al. . Multiplexed optogenetic stimulation of neurons with spectrum-selective upconversion nanoparticles. Adv Healthcare Mater 2017; 6: 1700446. 10.1002/adhm.201700446 [DOI] [PubMed] [Google Scholar]
- 80. Lin X, Chen X, Zhang Wet al. . Core–shell–shell upconversion nanoparticles with enhanced emission for wireless optogenetic inhibition. Nano Lett 2018; 18: 948–56. 10.1021/acs.nanolett.7b04339 [DOI] [PubMed] [Google Scholar]
- 81. Miyazaki T, Chowdhury S, Yamashita Tet al. . Large timescale interrogation of neuronal function by fiberless optogenetics using lanthanide micro-particles. Cell Rep 2019; 26: 1033–43. 10.1016/j.celrep.2019.01.001 [DOI] [PubMed] [Google Scholar]
- 82. Zheng W, Huang P, Tu Det al. . Lanthanide-doped upconversion nano-bioprobes: electronic structures, optical properties, and biodetection. Chem Soc Rev 2015; 44: 1379–415. 10.1039/C4CS00178H [DOI] [PubMed] [Google Scholar]
- 83. Kaiser M, Würth C, Kraft Met al. . Power-dependent upconversion quantum yield of NaYF4: Yb3+, Er3+ nano-and micrometer-sized particles–measurements and simulations. Nanoscale 2017; 9: 10051–8. 10.1039/C7NR02449E [DOI] [PubMed] [Google Scholar]
- 84. Ahrar K, Gowda A, Javadi Set al. . Preclinical assessment of a 980-nm diode laser ablation system in a large animal tumor model. J Vasc Interv Radiol 2010; 21: 555–61. 10.1016/j.jvir.2010.01.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85. Jiang Y, Carvalho-de-Souza JL, Wong RCSet al. . Heterogeneous silicon mesostructures for lipid-supported bioelectric interfaces. Nat Mater 2016; 15: 1023–30. 10.1038/nmat4673 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86. Shapiro MG, Homma K, Villarreal Set al. . Infrared light excites cells by changing their electrical capacitance. Nat Commun 2012; 3: 736. 10.1038/ncomms1742 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87. Carvalho-de-Souza JL, Treger JS, Dang Bet al. . Photosensitivity of neurons enabled by cell-targeted gold nanoparticles. Neuron 2015; 86: 207–17. 10.1016/j.neuron.2015.02.033 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88. Tian B, Lieber CM. Synthetic nanoelectronic probes for biological cells and tissues. Annual Rev Anal Chem 2013; 6: 31–51. 10.1146/annurev-anchem-062012-092623 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89. Jiang Y, Li X, Liu Bet al. . Rational design of silicon structures for optically controlled multiscale biointerfaces. Nat Biomed Eng 2018; 2: 508–21. 10.1038/s41551-018-0230-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90. Jiang Y, Tian B. Inorganic semiconductor biointerfaces. Nat Rev Mater 2018; 3: 473–90. 10.1038/s41578-018-0062-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91. Carvalho-de-Souza JL, Pinto BI, Pepperberg DRet al. . Optocapacitive generation of action potentials by microsecond laser pulses of nanojoule energy. Biophys J 2018; 114: 283–8. 10.1016/j.bpj.2017.11.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92. Dhaka A, Viswanath V, Patapoutian A. Trp ion channels and temperature sensation. Annu Rev Neurosci 2006; 29: 135–61. 10.1146/annurev.neuro.29.051605.112958 [DOI] [PubMed] [Google Scholar]
- 93. Lyu Y, Xie C, Chechetka SAet al. . Semiconducting polymer nanobioconjugates for targeted photothermal activation of neurons. J Am Chem Soc 2016; 138: 9049–52. 10.1021/jacs.6b05192 [DOI] [PubMed] [Google Scholar]
- 94. Wu Z-C, Li W-P, Luo C-Het al. . Rattle-type Fe3O4@CuS developed to conduct magnetically guided photoinduced hyperthermia at first and second NIR biological windows. Adv Funct Mater 2015; 25: 6527–37. 10.1002/adfm.201503015 [DOI] [Google Scholar]
- 95. Jiang Y, Zhao X, Huang Jet al. . Transformable hybrid semiconducting polymer nanozyme for second near-infrared photothermal ferrotherapy. Nat Commun 2020; 11: 1857. 10.1038/s41467-020-15730-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96. Chen R, Romero G, Christiansen MGet al. . Wireless magnetothermal deep brain stimulation. Science 2015; 347: 1477–80. 10.1126/science.1261821 [DOI] [PubMed] [Google Scholar]
- 97. Parameswaran R, Carvalho-de-Souza JL, Jiang Yet al. . Photoelectrochemical modulation of neuronal activity with free-standing coaxial silicon nanowires. Nat Nanotechnol 2018; 13: 260–6. 10.1038/s41565-017-0041-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98. Ledesma HA, Li X, Carvalho-de-Souza JLet al. . An atlas of nano-enabled neural interfaces. Nat Nanotechnol 2019; 14: 645–57. 10.1038/s41565-019-0487-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99. Ye PP, Brown JR, Pauly KB. Frequency dependence of ultrasound neurostimulation in the mouse brain. Ultrasound Med Biol 2016; 42: 1512–30. 10.1016/j.ultrasmedbio.2016.02.012 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100. Jiang Y, Huang Y, Luo Xet al. . Neural stimulation in vitro and in vivo by photoacoustic nanotransducers. Matter 2021; 4: 654–74. 10.1016/j.matt.2020.11.019 [DOI] [Google Scholar]
- 101. Jiang Y, Lee HJ, Lan Let al. . Optoacoustic brain stimulation at submillimeter spatial precision. Nat Commun 2020; 11: 881. 10.1038/s41467-020-14706-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102. Shi L, Jiang Y, Fernandez FRet al. . Non-genetic photoacoustic stimulation of single neurons by a tapered fiber optoacoustic emitter. Light: Science & Applications 2021; 10: 143. doi: 10.1038/s41377-021-00580-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103. Wu X, Zhu X, Chong Pet al. . Sono-optogenetics facilitated by a circulation-delivered rechargeable light source for minimally invasive optogenetics. Proc Natl Acad Sci USA 2019: doi: 10.1073/pnas.1914387116. 10.1073/pnas.1914387116 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104. Hong G. Seeing the sound. Science 2020; 369: 638. 10.1126/science.abd3636 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105. Zhang Y, Zhang X, Wang Het al. . Remote regulation of optogenetic proteins by a magneto-luminescence microdevice. Adv Funct Mater 2021; 31: 2006357. 10.1002/adfm.202006357 [DOI] [Google Scholar]
- 106. Zhang J-C, Wang X, Marriott Get al. . Trap-controlled mechanoluminescent materials. Prog Mater Sci 2019; 103: 678–742. 10.1016/j.pmatsci.2019.02.001 [DOI] [Google Scholar]
- 107. Van der Heggen D, Joos JJ, Rodríguez Burbano DCet al. . Counting the photons: determining the absolute storage capacity of persistent phosphors. Materials 2017; 10: 867. 10.3390/ma10080867 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108. Bai L, Mesgarzadeh S, Ramesh KSet al. . Genetic identification of vagal sensory neurons that control feeding. Cell 2019; 179: 1129–43. 10.1016/j.cell.2019.10.031 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109. Han W, Tellez LA, Perkins MHet al. . A neural circuit for gut-induced reward. Cell 2018; 175: 887–8. 10.1016/j.cell.2018.10.018 [DOI] [PubMed] [Google Scholar]
- 110. Tung JK, Gutekunst C-A, Gross RE. Inhibitory luminopsins: genetically-encoded bioluminescent opsins for versatile, scalable, and hardware-independent optogenetic inhibition. Sci Rep 2015; 5: 14366. 10.1038/srep14366 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111. Matsubara T, Yanagida T, Kawaguchi Net al. . Remote control of neural function by X-ray-induced scintillation. Nat Commun 2021; 12: 4478. 10.1101/798702. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112. Rommelfanger NJ, Ou Z, Keck CHCet al. . Differential heating of metal nanostructures at radio frequencies. Phys Rev Applied 2021; 15: 054007. 10.1103/PhysRevApplied.15.054007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 113. Rommelfanger NJ, Hong G. On the feasibility of wireless radio frequency ablation using nanowire antennas. APL Mater 2021; 9: 071103. 10.1063/5.0053189 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114. Luo D-G, Yue WWS, Ala-Laurila Pet al. . Activation of visual pigments by light and heat. Science 2011; 332: 1307–12. 10.1126/science.1200172 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115. Gracheva EO, Ingolia NT, Kelly YMet al. . Molecular basis of infrared detection by snakes. Nature 2010; 464: 1006–11. 10.1038/nature08943 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 116. Nelidova D, Morikawa RK, Cowan CSet al. . Restoring light sensitivity using tunable near-infrared sensors. Science 2020; 368: 1108–13. 10.1126/science.aaz5887 [DOI] [PubMed] [Google Scholar]
- 117. Liu Z, Lin C-H, Hyun B-Ret al. . Micro-light-emitting diodes with quantum dots in display technology. Light Sci Appl 2020; 9: 83. 10.1038/s41377-020-0268-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data