Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2023 Dec 1.
Published in final edited form as: Magn Reson Med. 2022 Aug 15;88(6):2475–2484. doi: 10.1002/mrm.29409

Repeatability of B1+ Inhomogeneity Correction of Volumetric (3D) Glutamate CEST via High-Permittivity Dielectric Padding at 7T

Paul S Jacobs 1, Blake Benyard 1, Abigail Cember 1, Ravi Prakash Reddy Nanga 1, Quy Cao 2, M Dylan Tisdall 3, Neil Wilson 1, Sandhitsu Das 4, Kathryn A Davis 4, John Detre 1,4, David Roalf 5, Ravinder Reddy 1
PMCID: PMC9529237  NIHMSID: NIHMS1825517  PMID: 36178233

Abstract

Purpose:

Ultra-high field MR imaging lacks B1+ inhomogeneity due to shorter RF wavelengths used at higher field strengths compared to human anatomy. Chemical Exchange Saturation Transfer (CEST) techniques tend to be highly susceptible to B1+ inhomogeneities due to a high and uniform B1+ field being necessary to create the endogenous contrast. High-permittivity dielectric pads have seen increasing usage in MR imaging due to their ability to tailor the spatial distribution of the B1+ field produced. The purpose of this work is to demonstrate that dielectric materials can be used to improve glutamate weighted CEST (gluCEST) at 7T.

Methods:

GluCEST images were acquired on a 7T system on six healthy volunteers. Aqueous calcium titanate pads, with a permittivity of approximately 110, were placed on either side in the subject’s head near the temporal lobes. A post-processing correction algorithm was implemented in combination with dielectric padding to compare contrast improvement. Tissue segmentation was performed to assess the effect of dielectric pads on gray and white matter separately.

Results:

GluCEST images demonstrated contrast enhancement in the lateral temporal lobe regions with dielectric pad placement. Tissue segmentation analysis showed an increase in correction effectiveness within the gray matter tissue compared to white matter tissue. Statistical testing suggested a significant difference in gluCEST contrast when pads were used and showed a difference in the gray matter tissue segment.

Conclusion:

The use of dielectric pads improved the B1+ field homogeneity and enhanced gluCEST contrast for all subjects when compared to data that did not incorporate padding.

Keywords: CEST, Dielectric pads, high field

Introduction

Magnetic Resonance (MR) imaging at 7T field strengths allows for significant increases in spatial resolution and signal-to-noise ratio (SNR) relative to lower field strength counterparts, such as 1.5 or 3T, especially in neuroimaging1,2. However, at 7T, due to the shortened radiofrequency (RF) wavelengths in comparison to human anatomy, inherent B1+ field inhomogeneities are a more significant problem than at lower field strengths. This is primarily due to the wave-like behavior of the RF magnetic field within the tissue volume of interest that disrupts the homogenous field distribution.

Chemical Exchange Saturation Transfer (CEST) techniques can image specific labile protons by saturating the chemical species of interest, with a spectrally selective saturation pulse and allowing that saturation to spontaneously transfer to water over time. This process of saturation transfer is continuously repeated, leading to a buildup of saturation in water magnetization, which can then be directly measured36. However, since this process relies on the B1+ field to create endogenous but transient contrast, CEST imaging is significantly more vulnerable to B1+ inhomogeneities, like those that exist at ultra-high B0 fields, in comparison to other MR imaging sequences7,8. Glutamate-weighted CEST (gluCEST) is particularly challenging in this regard, due to the fast exchange of its protons with surrounding water. It requires both an ultra-high B0 field (≥7T) and a high and uniform B1+ strength across the tissue to generate adequate contrast. As data acquisition has moved from single-slice (two-dimensional: 2D) to volumetric slabs (three-dimensional: 3D), the absolute spread of B1+ amplitude over the field of view increases and, because one now needs to collect more points in k-space per saturation pulse, the minimum B1+ amplitude required to maintain glutamate derived contrast at an acceptable level increase as well9,10.

One of the primary ways in which B1+ inhomogeneity has been mitigated recently has been through the use of parallel transmit (PTX). There have been many studies that have employed PTX based approaches to correct for B1+ inhomogeneity1113, even in the case of CEST imaging, however, it has not been shown that PTX can maintain a high absolute B1+ power over the entire field of view. This would suggest that a multi-facetted correction approach may be necessary to better utilized other techniques in combination with PTX. B1+ correction via post-processing algorithms is another way of compensating for inhomogeneities. In previous work, “calibration” gluCEST data is acquired and fit to a phenomenological function to parametrize correction surfaces over space (B1+, T1)14. In another study, similar “calibration” data is acquired at coarsely sampled B1+ values across measured B1+ maps. In this way separate correction curves could be derived for separate gray and white matter gluCEST segmentations15.

In recent years, there has been an increasing application of high-permittivity dielectric materials in MR imaging to tailor the spatial distribution of the B1+ field and improve SNR within the tissue volume of interest. Neuroimaging studies, both at 3T and 7T, having shown improvement in B1+ homogeneity and SNR include surrounding the head with water pads16, positioning aqueous calcium titanate padding near the temporal lobes and cerebellum17,18, and creating a dielectric helmet for whole cortex enhancement19. Other applications include using dielectric padding composed of aqueous manganese salts to correct for abdominal imaging artifacts20 and the use of barium titanate padding to improve SNR for imaging of the inner ear21, temporomandibular joint22 and femoral arteries23. In the present work, high-permittivity dielectric pads, composed of aqueous calcium titanate (7TNS, Multiwave Imaging, Marseille, France), were used to augment the B1+ field amplitude during the image acquisition process in conjunction with a post-processing image correction algorithm and hence gain enhanced gluCEST contrast. While dielectric pads have been adopted in several 7T imaging studies, it is believed that this is the first time that they have been used to improve brain gluCEST imaging at ultra-high field strengths.

Theory

CEST theory

Chemical exchange saturation transfer (CEST) is an MR imaging technique used to image certain solute compounds of interest via the exchange of protons with the bulk water protons present in the tissue volume3. In the imaging sequence, a frequency-selective RF pulse is used to saturate exchangeable protons in the solute, where there is then a proportional decrease in the water signal because of the saturated protons accumulating in the bulk water pool of the tissue. The amine proton on glutamate, which resonates at 3ppm downfield from water, exchanges at an ideal rate to be imaged using this technique (gluCEST) and can be further improved using 7T and higher field strength MR systems5,7.

Bloch simulations can allow for an understanding of why sufficient B1+ power is necessary for CEST imaging given the proton exchange rate for a metabolite of interest. Figure 1 illustrates that, in contrast to the more widespread technique of amide proton transfer (APT) imaging, accurate gluCEST imaging requires a high and constant B1+ power. Figure 1A shows the simulated CEST contrast from saturation at 3.5ppm as a function of B1+ strength for different concentrations of amide protons. This illustrates that CEST at 3.5ppm will detect differences in [APT] for a broad range of B1+ power, starting at approximately 0.7μT. This sensitivity is quantified in Figure 1C, where the slope of the difference in CEST signal from two concentrations is plotted against the B1+ power used. Figure 1B, analogous to Figure 1A, shows CEST contrast at 3.0ppm (gluCEST) and illustrates its ability to detect differences in the concentration of glutamate. In sharp contrast to APT CEST measurements, B1+ strengths below 1.5μT are insensitive to glutamate; they almost exclusively reflect differences in water and lipid properties. Figure 1D illustrates that optimal gluCEST contrast can be achieved using the highest possible B1+. However, currently, B1+ values above 4μT cannot be used in human experiments due to specific absorption rate (SAR) limitations.

Figure 1.

Figure 1.

a) Simulated CEST contrast from saturation at 3.5ppm (targeting protein amide protons, APT) as a function of B1+ strength for different concentrations of amide. The plot illustrates that CEST at 3.5ppm will detect differences in [APT] comparably for a broad range of B1 power. This sensitivity is quantified in c), where the slope difference between CEST from a 40mM source is plotted against the B1. b) CEST contrast at 3.0ppm (gluCEST). d) Optimal gluCEST signal is only possible with high B1. This figure was adapted from Cember et al14.

Dielectric Padding

Dielectric materials with high permittivity values can augment existing B1+ values via two main mechanisms: 1) The addition of materials to reduce the elliptical shape of the tissue volumes can disperse the B1+ fields more evenly and 2) displacement currents that are created within the dielectric materials create a secondary magnetic field which can add constructively to the primary field from the RF coil24.

In reference to the second mechanism, the displacement current density that exists within the dielectric material can be quantified via the equation below [EQN 1]. This equation shows the total current density (JT) within a material is comprised of conduction currents (JC) and displacement currents (JD). It can be seen the conduction current is dependent on the conductivity (σ) value of the material, while the displacement current depends on the relative permittivity value (εr), frequency (ω), and permittivity of free space (εo). This implies that at a sufficiently high ω or with sufficiently high εr, the displacement current can become a significant contributor to the main magnetic field created by the RF coil25.

JT=JC+JD= σE+iωεoεrE (1)

These high permittivity materials are typically manufactured as a powder, with the internal grain size being an important factor in the macroscopic field altering properties. These powders can be mixed into an aqueous solution (usually with water or deuterium oxide), done here with deuterium oxide, to reduce the background MR signal.

Methods

Dielectric pads

Two 18 × 18 cm2 dielectric pads, having a composition of calcium titanate (CaTiO3) suspended in deuterium oxide (7TNS, Multiwave Imaging, Marseille, France) with an estimated εr of 11018, were utilized in this study.

MR Imaging data acquisition

All images were acquired on a 7T system (MAGNETOM Terra, Siemens Healthcare, Erlangen, Germany) using a single-volume transmit 32-channel receive phased-array head coil (Nova Medical, Wilmington, MA, USA). Six healthy volunteers were scanned with informed consent under an approved institutional regulatory board protocol. All scans were performed with and without dielectric padding placed on either side of the subject’s head near the temporal lobes, as seen in Figure 2. Five repeat imaging experiments were also performed on a single-subject, with and without dielectric padding, to show intra-subject repeatability. 3D gluCEST data was acquired using a modified 3D turbo fast low angle shot (FLASH) sequence, as previously described in Cai et al.5, with TR = 3.5ms, TE = 1.79ms, and shot TR = 6000ms. The 3D slab had a thickness of 24mm (12 slices with each slice being 2mm thick), with an acquired in-plane resolution of 1 × 1mm2 and matrix size of 240 × 196. Each slice was acquired in three shots. Generalized auto calibrating partially parallel acquisitions (GRAPPA) was applied with an acceleration factor of 2. Magnetization preparation was achieved using a B1rms of 3.06μT, and saturation durations of 800ms (eight 99ms Hanning shaped pulses with 1ms interpulse delay) were applied at offset frequencies of (±1.8, 2.1, 2.4, 2.7, 3.0, 3.3, 3.6, 3.9, 4.2ppm) relative to water.

Figure 2.

Figure 2.

Image showing the placement of the two dielectric pads (white arrows) inside the head coil on either side of the subject’s head near the temporal lobes.

Water saturation shift referencing (WASSR) images26 were collected from 0 to ±1.5ppm (step-size of 0.15ppm) with a saturation pulse of B1rms = 0.29μT and a 200ms duration. These images were used to compute B0 maps and had identical imaging parameters as those used for the gluCEST sequence described above. An MP2RAGE sequence, which is a 2D multi-slice Siemens product sequence, was used with the same spatial parameters as described for the gluCEST sequence to generate T1 maps. Relative B1+ maps were generated using square preparation pulses with flip-angles of 20°, 40° and 80°27 and acquired with the same spatial parameters as the 3D gluCEST sequence.

Image post-processing and statistical analysis

GluCEST-weighted images were corrected for the B0 field distribution using the acquired WASSR26 and relative B1+ field maps27. A further post-processing correction strategy, as described in Cember et al.14, was utilized to correct gluCEST contrast and quantitative accuracy in images acquired with and without dielectric padding. This post-processing correction was utilized as a standard step when analyzing gluCEST data in later sections. A tissue segmentation analysis between gray and white matter was also performed to determine the influence of dielectric padding and post-processing on separate tissue types28.

Mixed effect statistical models were performed to estimate the effect of padding on the change in B1+ and gluCEST contrast enhancement with random effect for subject/scans to account for repeated measures.

All image analyses were performed in MATLAB (The Mathworks, Inc., Natick, MA, USA) and statistical analyses were performed in R.

Results

Intersubject reproducibility was established by acquiring B1+ and gluCEST data across six healthy volunteers with and without pads. This can be illustrated in Figure 3 where B1+ maps are presented for each subject showing an increase in signal in the temporal lobes resulting in greater homogeneity across the image. The resulting gluCEST images can be seen in Figure 4 across padding conditions and with and without post-processing. The resulting gluCEST image can be seen in Figure 4 with raw uncorrected data in column A (without dielectric pads) and column B (with dielectric pads); post-processing correction in column C (without dielectric pads) and column D (with dielectric pads). All subjects qualitatively showed an increase in gluCEST contrast, especially in the lateral temporal lobes, when dielectric padding and post-processing were implemented separately. Further enhancement was observed when both dielectric padding and post-processing were used in combination. It should be noted, uncorrected gluCEST data is shown here only as an example to illustrate the effect of only the dielectric pads being present. All remaining gluCEST data presented and analyses performed will only incorporate post-processing across pad conditions. Histograms presenting the distribution and fits of the different subject data can be seen in Figure 6. Figure 6A shows a narrowing B1+ distribution, indicating an increase in homogeneity while Figure 6B shows a slight narrowing of the processed gluCEST contrast values. A mixed effect model to test for statical significance across the different subject data across pad conditions. The results showed no statistically significant differences present in the B1+ values (p = 0.861) across pad conditions. However, suggestive evidence for significance was present for the post-processed gluCEST contrast values (p = 0.064) across pad conditions.

Figure 3.

Figure 3.

Intersubject B1+ reproducibility acquisitions performed on six subjects with and without dielectric padding. These images show an increase in relative B1+ strength in the temporal lobes and a slight reduction in the center magnitude of the image, indicating an overall increase in homogeneity.

Figure 4.

Figure 4.

Reproducibility gluCEST acquisitions performed on multiple subjects. GluCEST data acquired on six subjects without (first column) and with (second column) dielectric pads. A post-processing correction method was implemented on the same data acquired without (third column) and with (fourth column) dielectric padding. These images show substantial contrast improvement in the temporal lobes of all subjects when pads were used compared to data acquired without pads. A similar trend can be observed in the data with post-processing applied, in which temporal lobe contrast improvement can be seen when pads were used.

Figure 6.

Figure 6.

Histogram and fit comparisons for B1+ (a), total gluCEST contrast (b), gray matter segment gluCEST contrast (c), and white matter segment gluCEST contrast (d). Histogram distribution narrowing can be seen in the relative B1+, reflecting an increase in homogenization, and in the total gluCEST and gray matter tissue segment gluCEST, reflecting an enhancement in contrast when pads are used. The histogram comparison for the white matter tissue segmentation shows no major difference when pads are used. These histograms incorporate all data acquired from the intersubject reproducibility scans.

A white/gray matter tissue segmentation analysis of the processed intersubject gluCEST data was performed and can be seen in Figure 5. Gray matter tissue showed a substantial improvement in contrast in the temporal lobes when pads were used in comparison to no pads. In the white matter tissue, dielectric pads did not have as great of an effect as in gray matter. A histogram comparison of the segmented gray and white matter gluCEST contrast distributions can be seen in Figure 6C and 6D respectively across pad conditions. Distribution narrowing can be observed in the gray matter data when pads are used while no major difference can be seen in the white matter distribution. The mixed effect model was used to test for significance and showed a statistically significant difference in the gray matter data across pad conditions (p = 0.006) while in the white matter data no significance was observed (p = 0.223). Average and standard deviation values are presented in Table 1 for relative B1+ strength and gluCEST contrast for all subjects scanned, across padding conditions, and between gray and white matter tissue segmentation.

Figure 5.

Figure 5.

GluCEST contrast maps showing the effect of dielectric padding on segmented gray and white matter tissue. The left-most column shows the gray and white matter tissue segments overlaid on an anatomical image acquired of the same region. The gray matter tissue segment (top row) shows an increase in gluCEST contrast in the temporal lobe regions when pads are used. The white matter tissue segment (bottom row) shows less enhancement than the gray matter tissue segment when pads are used.

Table 1.

Calculated overall and tissue segmented mean and standard deviation values for relative B1 and gluCEST data acquired for all subjects under different padding conditions. Overall values were calculated from the relative B1 and gluCEST contrast values across the whole brain slice while tissue segmented gray matter (GM) and white matter (WM) were calculated separately.

Relative B1 Strength gluCEST Contrast with Post-processing (%)
No Pads 0.63 (0.15) 7.27 (4.37)
No Pads GM 0.71 (0.31) 8.14 (4.04)
No Pads WM 0.60 (0.34) 7.19 (4.33)
Pads 0.63 (0.13) 7.42 (4.29)
Pads GM 0.65 (0.31) 7.85 (4.32)
Pads WM 0.59 (0.33) 7.04 (4.28)

Intrasubject repeatability was established by acquiring B1+ and gluCEST data on the same subject for five scans. Results presented in Figure 7 show B1+ maps (top row) and post-processed gluCEST maps (bottom row) for one of the same subject repeat scans. These results show an increase in B1+ homogeneity and gluCEST contrast in the temporal lobes when pads were used versus absent. Histograms presenting the distribution and fit of B1+ values show a narrowing of the distribution reflecting the increase in homogeneity. However, in comparison, the gluCEST contrast value histogram does not show the same degree of distribution change. Figure 8 shows all gluCEST image slices (apart from first and last slices due to the presence of phase artifacts) from one of the same subject repeat scans when pads were present versus absent. It can be seen qualitatively that when pads were present, the temporal lobes saw an increase in contrast. It should be noted that this contrast enhancement diminished slightly for lower slices. Average and standard deviation values are presented in Table 2 for relative B1+ strength and gluCEST contrast across repeat scans and padding conditions. The same mixed effect model used previously, was performed on the same subject repeat data comparing padding conditions. The results showed no statistically significant differences present in the B1+ values (p = 0.338) across pad conditions. However, a significant difference was present for the post-processed gluCEST contrast values (p = 0.0006) across pad conditions.

Figure 7.

Figure 7.

An example same subject repeatability acquisition showing the effect of dielectric padding on the B1+ homogeneity and gluCEST contrast. The left column shows data acquired with padding showing lack of B1+ homogeneity and gluCEST contrast in the temporal lobes. The center column shows data acquired with padding showing an improvement in homogeneity and contrast. The upper right panel shows a histogram and fit comparison for relative B1+ strength. The lower right panel shows a histogram and fit comparison for gluCEST contrast. These histograms incorporate all data from the intrasubject repeatability scans.

Figure 8.

Figure 8.

An example multi-slice comparison from one intrasubject repeatability acquisition with and without padding. In this example, slice 2 corresponds to areas lower in the brain while slice 11 is 20mm higher. The greatest amount of temporal lobe correction can be seen in the upper slices with a slightly diminished effect for the lower slices. The first and last slices were omitted due to the presence of phasing artifacts.

Table 2.

Calculated mean and standard deviation values for all relative B1 and gluCEST data for five separate scans on a single subject with and without dielectric padding. Mean and standard deviation values were calculated from relative B1 and gluCEST contrast values across the whole brain slice.

Relative B1 Strength CEST Contrast with Post-Processing (%)
No Pads Scan 1 0.63 (0.14) 7.42 (4.06)
Scan 2 0.62 (0.16) 7.29 (4.28)
Scan 3 0.62 (0.16) 7.28 (4.28)
Scan 4 0.62 (0.16) 7.28 (4.29)
Scan 5 0.61 (0.15) 7.35 (4.16)
Pads Scan 1 0.61 (0.13) 7.77 (4.29)
Scan 2 0.63 (0.13) 7.82 (4.31)
Scan 3 0.63 (0.12) 7.54 (4.56)
Scan 4 0.62 (0.13) 7.45 (3.95)
Scan 5 0.61 (0.13) 7.29 (4.06)

Discussion

The aim of this work was to show that dielectric padding can be used to correct B1+ inhomogeneities found in gluCEST images by altering the spatial distribution of the B1+ fields via secondary magnetic field generation. These images were specifically acquired in a slice in line with the hippocampus as there tends to be significant signal drop off in this region. The data presented here have shown that gluCEST contrast can be improved using carefully placed calcium titanate dielectric padding. The relative B1+ maps are reflective of this, showing not only an increase in signal in the temporal lobes but also a decrease in the center B1+ intensity. This dielectric corrective effect was clearly seen across the images acquired across multiple subjects and on repeat scans of the same subject. Images where only the post-processing correction was performed showed that even though the image did show improvement, it could not compensate for all inhomogeneities present, particularly in the temporal lobe regions. Both dielectric pads and the post-processing correction had the effect of shifting the distribution of gluCEST values and narrowing the distribution of B1+ values. The use of post-processing was seen to result in major distribution shifts while the dielectric pads had more of a distribution narrowing effect. When both were used in combination, the dielectric pads allow more of the acquired gluCEST data to fall into a “correctable” regime, leading to better final images. Based on Cember et al.14 the correctable B1+ regime that the post-processing algorithm can accommodate for falls between 1.3 to 3.1μT. With gluCEST contrast being seen at nominal B1+ saturations of 0.7μT (or about 40% nominal value), this post-processing can’t effectively correct for the lower B1+ saturations seen in the uncorrected temporal lobes. Dielectric padding has proven useful in this case by creating a secondary B1+ field to add to regionally lower magnitude primary field. So, while the dielectric pads were of great benefit in correcting inhomogeneities, they do not eliminate the need for adequate post-processing, as clearly illustrated in the work that originally presented this type of post-processing correction method14.

Also, while largely effective across the whole brain, a difference in dielectric pad effectiveness was observed in tissue segmentation analysis between gray and white matter regions. In combination with inherent B1 inhomogeneities present at ultra-high fields, a portion of the present B1 drop off was due to the oblique slice taken in alignment with the hippocampus. In this oblique slice, the anterior brain region was physically closer to the edge of coil where a lower B1 field is present. The gray and white matter regions in this anterior region showed the most contrast improvement when dielectric pads were placed in proximity. However, in this case, the gray matter region showed a much greater improvement in contrast between pad and post-processing conditions than white matter. Within this work, reproducibility was demonstrated between subjects as well as repeatability between subjects with and without padding. This is in strong agreement with work done by Nanga et al.29 showing the reproducibility of 2D gluCEST across six healthy volunteers and furthermore, GM and WM segmentations of this data showed a high level of reproducibility. This is an important point as the addition of dielectric pads in the present work did not alter gluCEST reproducibility that has been previously demonstrated by Nanga et al. Additionally, when evaluating the current gluCEST maps against an optimal standard, comparisons should made between the B1+ maps and not the gluCEST maps themselves. This is due to slight regional variations in glutamate concentration being present across different brain regions and between subjects, whereas the B1+ maps are a result of the RF coil and could in theory be further homogenized to a theoretical constant across the tissue volume. A good example of this can be seen in Figure 8, where the hippocampal region of the brain expectedly shows a much higher gluCEST contrast than some of the immediate surrounding areas.

As B0 strength is increased in the future, the problem of B1 inhomogeneity will only become further exacerbated. Therefore, robust correction methods, like those presented here, will only become more important. One technique that is expected to contribute to more effective correction techniques, but was not explored here, is the use of PTX. With PTX being on the cutting edge of MR technological development, this type of sequence integration would be a valuable addition to dielectric padding and post-processing corrections.

A common criticism of dielectric padding is the lack of flexible form factor brought on by the materials used. With the most used materials being calcium titanate (εr=110) and barium titanate (εr=150), large volume fractions of these materials must be used in aqueous solution to effectively produce secondary magnetic fields17. If a material with a much higher permittivity value was used instead, the volume fraction and overall amount of material, could be reduced while preserving the corrective effect. With a smaller volume fraction of material used thinner pads would be achievable resulting in a potential increase in subject comfort during scans. However, when exploring the usage of any new dielectric material careful consideration must be given to accurate characterization of the potential frequency dependence, especially at the higher frequencies used in ultra-high field imaging. Lastly, while repeatability and reproducibility of dielectric pad usage for gluCEST was established, one difficulty in using these pads is the lack of standard placement. The large size pads used here cover a sizable area of the head and can be used effectively with general placement. However, custom smaller padding with different materials may need more exact positioning. In the future, standard positioning coil inserts could accommodate mounted dielectric material pouches for increased accuracy in placement for not only head studies but also abdominal and skeletal muscle studies.

Conclusion

In this work, high-permittivity dielectric padding, composed of a calcium titanate solution, were used to augment the B1 field amplitude and resulted in gluCEST contrast enhancement. A post-processing correction, implemented in combination with the dielectric padding, showed a further improvement in B1 field uniformity and hence gluCEST contrast. In the future, continued use of dielectric materials with a higher relative permittivity value used in combination with parallel transmit technology and post-processing correction methods, could result in larger improvements in SNR and image quality.

Acknowledgements

Research reported in this work was supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under award Number P41EB029460 and by the National Institute of Aging of the National Institutes of Health under Award Numbers R56AG066656, R01MH120174, and R01AG063869.

References

  • 1.Rutland JW, Delman BN, Gill CM, Zhu C, Shrivastava RK, Balchandani P. Emerging Use of Ultra-High-Field 7T MRI in the Study of Intracranial Vascularity: State of the Field and Future Directions. AJNR Am J Neuroradiol. Jan 2020;41(1):2–9. doi: 10.3174/ajnr.A6344 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Ladd ME, Bachert P, Meyerspeer M, et al. Pros and cons of ultra-high-field MRI/MRS for human application. Prog Nucl Magn Reson Spectrosc. Dec 2018;109:1–50. doi: 10.1016/j.pnmrs.2018.06.001 [DOI] [PubMed] [Google Scholar]
  • 3.Wu B, Warnock G, Zaiss M, et al. An overview of CEST MRI for non-MR physicists. EJNMMI Phys. Dec 2016;3(1):19. doi: 10.1186/s40658-016-0155-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Kogan F, Hariharan H, Reddy R. Chemical Exchange Saturation Transfer (CEST) Imaging: Description of Technique and Potential Clinical Applications. Curr Radiol Rep. Jun 1 2013;1(2):102–114. doi: 10.1007/s40134-013-0010-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Cai K, Haris M, Singh A, et al. Magnetic resonance imaging of glutamate. Nat Med. Jan 22 2012;18(2):302–6. doi: 10.1038/nm.2615 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Haris M, Singh A, Cai K, et al. A technique for in vivo mapping of myocardial creatine kinase metabolism. Nat Med. Feb 2014;20(2):209–14. doi: 10.1038/nm.3436 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Davis KA, Nanga RP, Das S, et al. Glutamate imaging (GluCEST) lateralizes epileptic foci in nonlesional temporal lobe epilepsy. Sci Transl Med. Oct 14 2015;7(309):309ra161. doi: 10.1126/scitranslmed.aaa7095 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Singh A, Haris M, Cai K, et al. Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3 T and 7 T. Magn Reson Med. Aug 2012;68(2):588–94. doi: 10.1002/mrm.23250 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Cember ATJ, Nanga RPR, Reddy R. Glutamate weighted CEST (gluCEST) imaging for mapping neurometabolism: An update on the state of the art and emerging findings from in vivo applications. NMR Biomed. May 31 2022:e4780. doi: 10.1002/nbm.4780 [DOI] [PubMed] [Google Scholar]
  • 10.Nanga RP. Feasibility of Partial 3D GluCEST in Healthy Human Adults at 7.0T. presented at: Proc Annu Meet Int Soc Magn Reson Med 2019; Published online. [Google Scholar]
  • 11.Liebert A, Zaiss M, Gumbrecht R, et al. Multiple interleaved mode saturation (MIMOSA) for B1 (+) inhomogeneity mitigation in chemical exchange saturation transfer. Magn Reson Med. Aug 2019;82(2):693–705. doi: 10.1002/mrm.27762 [DOI] [PubMed] [Google Scholar]
  • 12.Tse DHY, da Silva NA, Poser BA, Shah NJ. B1+ inhomogeneity mitigation in CEST using parallel transmission. Magn Reson Med. Dec 2017;78(6):2216–2225. doi: 10.1002/mrm.26624 [DOI] [PubMed] [Google Scholar]
  • 13.Padormo F, Beqiri A, Hajnal JV, Malik SJ. Parallel transmission for ultrahigh-field imaging. NMR Biomed. Sep 2016;29(9):1145–61. doi: 10.1002/nbm.3313 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Cember ATJ, Hariharan H, Kumar D, Nanga RPR, Reddy R. Improved method for post-processing correction of B1 inhomogeneity in glutamate-weighted CEST images of the human brain. NMR Biomed. Jun 2021;34(6):e4503. doi: 10.1002/nbm.4503 [DOI] [PubMed] [Google Scholar]
  • 15.Singh A, Cai K, Haris M, Hariharan H, Reddy R. On B1 inhomogeneity correction of in vivo human brain glutamate chemical exchange saturation transfer contrast at 7T. Magn Reson Med. Mar 1 2013;69(3):818–24. doi: 10.1002/mrm.24290 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Yang QX, Mao W, Wang J, et al. Manipulation of image intensity distribution at 7.0 T: passive RF shimming and focusing with dielectric materials. J Magn Reson Imaging. Jul 2006;24(1):197–202. doi: 10.1002/jmri.20603 [DOI] [PubMed] [Google Scholar]
  • 17.Haines K, Smith NB, Webb AG. New high dielectric constant materials for tailoring the B1+ distribution at high magnetic fields. J Magn Reson. Apr 2010;203(2):323–7. doi: 10.1016/j.jmr.2010.01.003 [DOI] [PubMed] [Google Scholar]
  • 18.Teeuwisse WM, Brink WM, Webb AG. Quantitative assessment of the effects of high-permittivity pads in 7 Tesla MRI of the brain. Magn Reson Med. May 2012;67(5):1285–93. doi: 10.1002/mrm.23108 [DOI] [PubMed] [Google Scholar]
  • 19.Sica CT, Rupprecht S, Hou RJ, et al. Toward whole-cortex enhancement with an ultrahigh dielectric constant helmet at 3T. Magn Reson Med. Mar 2020;83(3):1123–1134. doi: 10.1002/mrm.27962 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Sreenivas M, Lowry M, Gibbs P, Pickles M, Turnbull LW. A simple solution for reducing artefacts due to conductive and dielectric effects in clinical magnetic resonance imaging at 3T. Eur J Radiol. Apr 2007;62(1):143–6. doi: 10.1016/j.ejrad.2006.11.014 [DOI] [PubMed] [Google Scholar]
  • 21.Brink WM, van der Jagt AM, Versluis MJ, Verbist BM, Webb AG. High permittivity dielectric pads improve high spatial resolution magnetic resonance imaging of the inner ear at 7 T. Invest Radiol. May 2014;49(5):271–7. doi: 10.1097/RLI.0000000000000026 [DOI] [PubMed] [Google Scholar]
  • 22.Manoliu A, Spinner G, Wyss M, et al. Magnetic Resonance Imaging of the Temporomandibular Joint at 7.0 T Using High-Permittivity Dielectric Pads: A Feasibility Study. Invest Radiol. Dec 2015;50(12):843–9. doi: 10.1097/RLI.0000000000000196 [DOI] [PubMed] [Google Scholar]
  • 23.Lindley MD, Kim D, Morrell G, et al. High-permittivity thin dielectric padding improves fresh blood imaging of femoral arteries at 3 T. Invest Radiol. Feb 2015;50(2):101–7. doi: 10.1097/RLI.0000000000000106 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Webb AG. Dielectric materials in magnetic resonance. Concepts in Magnetic Resonance Part A. 2011;38A(4):148–184. doi: 10.1002/cmr.a.20219 [DOI] [Google Scholar]
  • 25.Lakshmanan K, Carluccio G, Walczyk J, et al. Improved whole-brain SNR with an integrated high-permittivity material in a head array at 7T. Magn Reson Med. Aug 2021;86(2):1167–1174. doi: 10.1002/mrm.28780 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Kim M, Gillen J, Landman BA, Zhou J, van Zijl PC. Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magn Reson Med. Jun 2009;61(6):1441–50. doi: 10.1002/mrm.21873 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Volz S, Noth U, Rotarska-Jagiela A, Deichmann R. A fast B1-mapping method for the correction and normalization of magnetization transfer ratio maps at 3 T. Neuroimage. Feb 15 2010;49(4):3015–26. doi: 10.1016/j.neuroimage.2009.11.054 [DOI] [PubMed] [Google Scholar]
  • 28.Ng H, Ong S, Foong K, Goh P-S, Nowinski W. Medical image segmentation using k-means clustering and improved watershed algorithm. IEEE; 2006:61–65. [Google Scholar]
  • 29.Nanga RPR, DeBrosse C, Kumar D, et al. Reproducibility of 2D GluCEST in healthy human volunteers at 7 T. Magn Reson Med. Nov 2018;80(5):2033–2039. doi: 10.1002/mrm.27362 [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES