Abstract
Aim
To establish the pharmacokinetic (PK) and pharmacodynamic (PD) equivalence of proposed biosimilar insulin 70/30 (Biocon's Insulin‐70/30) and HUMULIN® 70/30 (HUMULIN‐70/30; Eli Lilly and Company, IN).
Materials and Methods
In this phase 1, automated euglycaemic glucose clamp study, 78 healthy subjects were randomized (1:1) to receive a single dose of 0.4 IU/kg of Biocon's Insulin‐70/30 and HUMULIN‐70/30. Plasma insulin concentrations and glucose infusion rates (GIRs) were assessed over 24 hours. Primary PK endpoints were area under the insulin concentration‐time curve from 0 to 24 hours ‐ AUCins.0−24h ‐ and maximum insulin concentration ‐ Cins.max. Primary PD endpoints were area under the GIR time curve from 0 to 24 hours ‐ AUCGIR.0−24h ‐ and maximum GIR ‐ GIRmax.
Results
Equivalence was shown between Biocon's Insulin‐70/30 and HUMULIN‐70/30 for the primary PK/PD endpoints. The 90% confidence intervals of the treatment ratios were entirely within the acceptance range of 80.00%‐125.00%. The secondary PK/PD profiles were also comparable. There were no clinically relevant differences in the safety profiles of the two treatments and no serious adverse events were reported.
Conclusion
PK/PD equivalence was demonstrated between Biocon's Insulin‐70/30 and HUMULIN‐70/30 in healthy subjects. Treatment with Biocon's Insulin‐70/30 and HUMULIN‐70/30 was well tolerated.
Keywords: basal‐bolus insulin, biosimilar insulin, pharmacodynamics, pharmacokinetics, type 1 diabetes, type 2 diabetes
1. INTRODUCTION
Optimal glycaemic control is required in patients with type 1 diabetes (T1D) and type 2 diabetes (T2D) to minimize the risks of long‐term complications. Several reference‐listed recombinant human insulins (rHIs) are available ‘over‐the‐counter’ in the United States, 1 implying that patients switch freely between these marketed rHIs in the real‐world scenario. The recent update in U.S. Food and Drug Administration (FDA) regulations, that is, switching insulin's status from ‘drug’ to ‘biologic’, could facilitate approval of biosimilar insulins in the United States, enabling a competitive market. 2 , 3 The Endocrine Society also strongly recommends expedited approval of insulin biosimilars as a policy, thereby increasing access to reasonably priced, life‐saving insulin for patients with diabetes in the United States. 4
Management of diabetes frequently requires insulin regimens beyond basal insulin in patients with T2D, 5 and this is an absolute requirement in patients with T1D. 6 Basal‐bolus is an effective, but intensive, multiple daily injection therapy, successful only in patients who comply with at least four daily injections. 5 To improve treatment adherence, premixed‐insulins are recommended as fixed‐component formulations of short‐, intermediate‐ or long‐acting insulins. 7 These increase the convenience of dosing, hence compliance, and lead to long‐term optimal glycaemic control in patients with T2D on a well‐regulated lifestyle with a consistent insulin requirement. 8 Most premixed‐insulins are prescribed for fasting and postprandial glycaemic control, dosed twice‐daily based on premeal glucose levels. 9
Premixed‐insulin analogues offer some advantages over traditional premixed‐insulins; however, premixed rHIs (neutral protamine Hagedorn, regular and premix) retain the advantage of low cost and established efficacy compared with analogues. 10 , 11 , 12 For the commercially available premixed‐rHI, the average price increase has been in the range of 50%‐75% during 2014–2021 in the United States. 13 , 14 This impacts an individual's adherence to clinical treatment, with many resorting to rationing or skipping doses to make their insulin supply last longer, risking their health and lives. 15 , 16 , 17
Biocon's biosimilar Insulin‐70/30 (Biocon's Insulin‐70/30) is a premixed‐insulin produced by recombinant DNA technology utilizing the Pichia pastoris (yeast) cell line and has been developed as a proposed biosimilar to HUMULIN® 70/30 (Eli Lilly and Company, Indianapolis, IN; hereafter referred to as ‐70/30) to improve glycaemic control in adult patients with diabetes. It contains 70% intermediate‐acting isophane insulin and 30% short‐acting human soluble insulin. To ensure equivalence of a proposed biosimilar product to the reference product, stringent regulatory requirements are met, which include multiple, orthogonal analytical methods to evaluate high similarity in structure, function and pharmacokinetic (PK)/pharmacodynamic (PD) equivalence. 3 Additionally, demonstration of similar PK/PD profiles is considered proof of similar efficacy of the biosimilar and the reference insulin. 18 As per the recent FDA guidance for biosimilar insulins, 3 if ‘high similarity’ for a proposed biosimilar is shown by comparative analytical assessment using state‐of‐the‐art technology, there is a possibility of little or no residual uncertainty regarding immunogenicity.
This article is the second in a series of clinical study publications aimed at evaluating the PK/PD equivalence of Biocon's rHIs versus reference biologics in healthy subjects (RHINE studies). Results from RHINE‐1 have been published earlier. 19 This study assesses the PK/PD equivalence and the safety and tolerability of Biocon's Insulin‐70/30 and HUMULIN‐70/30 using the euglycaemic clamp technique in healthy subjects.
2. METHODS
2.1. Study design
This phase 1, randomized, double‐blind, two‐treatment, two‐sequence, crossover, 24‐hour automated euglycaemic glucose clamp study (EudraCT: 2018‐003193‐26; Clinicaltrial.gov: NCT04022291) was conducted at two centres: Profil Institut für Stoffwechselforschung GmbH, Neuss, Germany and Profil Mainz GmbH & Co. KG, Mainz, Germany. Eligible subjects were randomly allocated to a sequence of two single doses: one each of Biocon's Insulin‐70/30 and HUMULIN‐70/30 (Figure 1 ). Based on body weight measured before the first dosing, 0.4 IU/kg of Biocon's Insulin‐70/30 (Biocon Biologics Ltd, India) and HUMULIN‐70/30 (US sourced), both 100 IU/mL, was administered subcutaneously into a lifted skin fold of the abdominal wall into the peri‐umbilical area using a standardized skin‐fold technique. Insulin was administered at the left and right lower abdominal quadrants with a BD Micro‐fine + 0.5 mL U‐100 syringe fitted with a 0.30 (30G) × 8 mm needle. Blood was collected predose and at prespecified intervals until 24 hours postdose to measure blood glucose (BG), insulin and C‐peptide. There was a washout period of 5‐7 days between dose administrations to avoid any carry‐over effect.
FIGURE 1.

Study design
2.2. Study subjects
This study included healthy male and postmenopausal female subjects aged 18‐55 years (both inclusive), with a body mass index (BMI) of 18.5‐29.0 kg/m2 (both inclusive) and fasting plasma glucose concentration (FPG) of 100 mg/dL or less. Primary exclusion criteria were receipt of any medicinal product in clinical development within 30 days or five times its half‐life (whichever was longer) before randomization in this trial; history or presence of clinically relevant co‐morbidity, as judged by the investigator; systolic blood pressure of less than 95 or more than 140 mmHg and/or diastolic blood pressure of less than 50 or more than 90 mmHg after resting for at least 5 minutes in the supine position; and pulse rate at rest outside the range of 50‐90 beats/min.
2.3. Ethics
The study was conducted in accordance with Good Clinical Practice and conformed to the ethical principles of the Declaration of Helsinki and all local and federal laws and regulations. The study was approved by a competent ethics committee (Ethikkommission der Ärztekammer Nordrhein in consultation with Ethik‐Kommission der Landesärztekammer Rheinland‐Pfalz) as well as a competent authority (Bundesinstitut für Arzneimittel und Medizinprodukte, BfArM) before study initiation. Subjects provided written informed consent before initiation of any study procedure.
2.4. Euglycaemic glucose clamping
Euglycaemic glucose clamping was performed using a glucose clamp device (ClampArt®; Profil Institut für Stoffwechselforschung, Neuss, Germany). The quality of the clamp data was reviewed regularly by the investigator and clamp supervisors, and evaluated based on all BG measurements during the clamp procedure 20 where the glucose infusion rate (GIR) was more than 0 mg/kg/min. The GIR necessary to keep the BG concentration at the target level was recorded every minute throughout the glucose clamp duration. BG was analysed at the study site using a Super GL glucose analyser (Dr. Müller Gerätebau GmbH, Freital, Germany) for verification of measurements of ClampArt. The mean clamp coefficient of variation (CV) was required to be less than 15% and the mean deviation from target (DFT) was required to be within the range of ±10 mg/dL after dosing until end of the clamp. Appendix S1 gives further details on the glucose clamp procedure.
2.5. PK sampling
Blood samples were taken for PK (plasma insulin and serum C‐peptide levels) analysis at predefined time intervals.
2.6. Bioanalytical methods
Validated, ultra‐performance, liquid chromatography with tandem mass spectrometry (LC–MS/MS) detection was used to analyse study samples. Insulin concentrations in plasma were measured using LC–MS/MS over a concentration range of 50‐8000 ng/L. C‐peptide in serum was measured using a validated electrochemiluminescence immunoassay (Roche Diagnostics, Switzerland) over a concentration range of 0.2‐32.0 ng/ml. All validation and sample quantification runs met prespecified acceptance criteria, including incurred sample reproducibility.
2.7. PK assessment
Primary PK parameters included AUCins.0−24h (area under insulin concentration‐time curve from 0‐24 hours) and Cins.max (maximum insulin concentration). Other parameters included AUCins.0−2h, AUCins.0−6h, AUCins.0−12h, AUCins.12−24h, AUCins.0−∞ (AUC in indicated time intervals), tins.max (time‐to‐maximum insulin concentration), t50%−ins(early) (time‐to‐half‐maximum insulin concentration before Cins.max), t50%−ins(late) (time‐to‐half‐maximum insulin concentration after Cins.max), t½ (terminal elimination half‐life) and λ z (terminal elimination rate constant). The values of all individual PK parameters were calculated using non‐compartmental methods in WinNonlin® v. 8.1 (Certara, Princeton, NJ). SAS® for Windows (SAS Institute Inc., NC; v. 9.4) was employed for all other statistical calculations (t50%−ins(early), t50%−ins(late), demography, PD parameters and safety) and all statistical analyses.
Primary PK analyses were conducted using Owen's method to correct endogenous insulin secretion by the C‐peptide–based correction formula. 21 Exogenous insulin concentration was calculated as:
where Insulin EXOG is exogenous insulin, Insulin OBS is observed insulin concentration, C‐peptide OBS is observed serum C‐peptide concentration and F is the mean of insulin/C‐peptide concentration ratios at –30, –15 and 0 minutes.
A sensitivity analysis of the primary PK endpoints was performed using the same mixed model as described for the primary analysis with uncorrected (i.e. without applying Owen's correction for C‐peptide) insulin concentrations. C‐peptide–based exclusion rules (described in the next section for the primary PD parameters) were applied to the PK sensitivity analysis.
2.8. PD assessment
Primary PD parameters included AUCGIR.0−24h (area under GIR time curve from 0‐24 hours) and GIRmax (maximum GIR). Other parameters included AUCGIR.0−2h, AUCGIR.0−6h, AUCGIR.0−12h, AUCGIR.12−24h (AUCGIR in indicated time interval), tGIR.max (time‐to‐maximum GIR), t50%−GIR(early) (time‐to‐half‐maximum GIR before GIRmax), t50%−GIR(late) (time‐to‐half‐maximum GIR after GIRmax) and onset of action.
Primary PD analysis was conducted using C‐peptide–based exclusion of profiles. For this, C‐peptide concentration‐time profiles were inspected during the blinded data review meeting to identify and exclude profiles as predefined for the study.
To account for meaningful fluctuations that can reflect changes in endogenous insulin concentration during the clamp period, profiles meeting predefined criteria were excluded from the primary PD analysis set. Sensitivity analysis for PD data was conducted using all profiles without applying any C‐peptide–based exclusion rules.
2.9. Safety assessments
All adverse events (AEs) were evaluated in terms of intensity, duration, severity, outcome and relationship to study medication throughout the study. Other safety parameters included injection site reactions, local tolerability, hypoglycaemic episodes, vital signs, physical examinations, 12‐lead electrocardiograms (ECGs) and standard laboratory safety tests (Table S1 ).
2.10. Statistical analysis
Equivalence between Biocon's Insulin‐70/30 and HUMULIN‐70/30 was considered shown if 90% confidence intervals (CIs) for primary PK endpoints, AUCins.0−24h‐ratio and Cins.max‐ratio, and primary PD endpoints, AUCGIR.0−24h‐ratio and GIRmax‐ratio, lay within an acceptance interval of 80.00%‐125.00%.
2.11. Sample size
Based on a maximum conservative CV estimate of 35% and an assumed ratio of 0.95 between the reference and the test insulins, a sample size of 70 subjects was considered necessary to establish equivalence with sufficient power of at least 90% (sample size calculation based on α = 0.05, 90% CIs in the range 80.00%‐125.00%, ~10% dropout rate). Accounting for potential dropouts, a total of 78 subjects were randomized in the study.
2.12. PK and PD endpoints
Analysis of PK and PD parameters was based on the respective per‐protocol population (PPP). PPP for PK and PD included all randomized subjects who completed the trial without any important protocol deviation. PK endpoints were derived from individual insulin concentration‐time profiles and PD endpoints were derived from individual GIR profiles. Raw GIR and BG profiles were used to calculate GIR AUCs and onset of action. GIRmax and other time‐related PD parameters were calculated from smoothed GIR profiles (SAS procedure PROC LOESS with smoothing factor 0.25). Single profiles of subjects who did not provide evaluable PK data were excluded from the PPP for PK if less than 50% of concentration measurements were above the lower limit of quantification or zero postdosing (i.e. 12/25 measurements). Single profiles of subjects who did not meet clamp quality criteria were excluded from the PPP for PD analysis.
The PK/PD endpoints (insulin and GIR AUCs, Cins.max and GIRmax) were analysed with analysis of variance (ANOVA) using log‐transformed data. The ANOVA was based on a mixed model (proc MIXED) using clinical site, sequence, period, treatment, site by sequence and site by treatment as fixed effects and subject within site by sequence as a random effect. As site by treatment effect was not significant at the 5% level, this fixed effect was dropped from the model.
Within the model, the least square mean (LS‐mean) for each treatment and difference of LS‐means between the treatment groups and corresponding 90% CI were calculated, exponentially back‐transformed, and multiplied by 100 to determine the estimated ratio % of responses between insulin formulations and corresponding 90% CIs.
2.13. Safety
Analysis of safety endpoints was based on the safety analysis set, which included all randomized subjects who received at least one dose of study treatment. Safety data were summarized by visit/treatment using descriptive statistics.
3. RESULTS
3.1. Subject disposition and baseline characteristics
Of the 110 subjects screened, 78 (77 males and one female) were randomized to one of the two treatment sequences. Seventy‐four subjects completed the study. Two subjects voluntarily withdrew from the study after treatment with HUMULIN‐70/30 and two subjects were withdrawn by the investigator because of general discontinuation criteria (low‐dose administration because of wrong body weight documentation; clamp termination because of massive recurring cannula problems) after treatment with Biocon's Insulin‐70/30.
The age, BMI and FPG ranged from 19 to 55 (mean 37.1) years, 19.0 to 29.0 (mean 24.68) kg/m2 and 59 to 100 (mean 87.7) mg/dL, respectively. Demographic characteristics were similar for both treatment sequences. No subjects received concomitant medication before the first administration of the investigational medicinal product. The disposition, demographics and baseline characteristics of the subjects are presented in Table S2. Table S3 lists the medical history and concomitant illnesses.
PPP for PK consisted of 73 subjects for both treatments. No profiles were excluded from the PPP for PK because of non‐evaluable data. One subject was excluded from PPP for PK analysis because of a mix‐up of transfer tubes for PK sample collection (with respect to time point). PPP for PD consisted of 72 subjects for both treatments. No profiles were excluded from PPP for PD because of non‐fulfilment of any of the defined clamp quality criteria. Two profiles each receiving Biocon's Insulin‐70/30 and HUMULIN‐70/30 were excluded from the PPP for PD based on C‐peptide exclusion rules and during the blinded database release meeting (because of extended gaps in continuous glucose monitoring), respectively.
3.2. Pharmacology
3.2.1. PK analyses
For the primary analysis, 90% CIs for geometric mean ratios (Biocon's Insulin‐70/30/HUMULIN‐70/30) were within the 80.00%‐125.00% limits for both primary PK endpoints (Table 1 ). Mean AUCins.0−24h and Cins.max were equivalent for both treatments. Mean C‐peptide–corrected plasma insulin concentration‐time profiles were equivalent between Biocon's Insulin‐70/30 and HUMULIN‐70/30 (Figure 2 ).
TABLE 1.
Primary and secondary PK and PD endpoints (PPP): C‐peptide–corrected data
| Endpoint | Biocon's insulin‐70/30 | HUMULIN‐70/30 | |||
|---|---|---|---|---|---|
| N | LS‐mean | N | LS‐mean | Geometric LS‐mean ratio Biocon's insulin‐70/30/HUMULIN‐70/30 (90% CI) | |
| Primary parameters | |||||
| Primary PK endpoints a | |||||
| AUCins.0−24h (h*ng/L) | 73 | 10 486.41 | 73 | 11 246.57 | 93.24 (89.24%; 97.42%) |
| Cins.max (ng/L) | 73 | 998.76 | 73 | 1074.39 | 92.96 (88.27%; 97.90%) |
| Primary PD endpoints b | |||||
| AUCGIR.0−24h (mg/kg) | 72 | 3453.86 | 72 | 3839.99 | 89.94 (84.17%; 96.12%) |
| GIRmax (mg/kg/min) | 72 | 5.30 | 72 | 5.76 | 92.06 (86.01%; 98.53%) |
| Secondary parameters | |||||
| Secondary PK endpoints a | |||||
| AUCins.0−2h (h*ng/L) | 73 | 1080.59 | 73 | 1193.75 | 90.52 (84.98%; 96.42%) |
| AUCins.0−6h (h*ng/L) | 73 | 4239.20 | 73 | 4627.04 | 91.62 (87.75%; 95.65%) |
| AUCins.0−12h (h*ng/L) | 73 | 7180.13 | 73 | 7862.94 | 91.32 (87.08%; 95.76%) |
| AUCins.12−24h (h*ng/L) | 73 | 3204.25 | 73 | 3251.31 | 98.55 (91.52%; 106.12%) |
| AUCins.0−∞ (h*ng/L) | 69 c | 12 409.45 | 71 c | 12 741.76 | 97.39 (93.66%; 101.27%) |
| t ins.max (h) d | 73 | 3.00 | 73 | 2.50 | ‐ |
| t 50%−ins(early) (h) d | 73 | 0.72 | 73 | 0.73 | ‐ |
| t 50%−ins(late) (h) d | 73 | 7.88 | 73 | 7.95 | ‐ |
| λ z (1/h) d | 69 c | 0.0957 | 71 c | 0.1102 | ‐ |
| t 1/2 (h) d | 69 c | 7.24 | 71 c | 6.29 | ‐ |
| Secondary PD endpoints b | |||||
| AUCGIR.0−2h (mg/kg) | 72 | 183.69 | 72 | 213.08 | 86.20 (77.15%; 96.32%) |
| AUCGIR.0−6h (mg/kg) | 72 | 1257.21 | 72 | 1424.65 | 88.25 (82.31%; 94.62%) |
| AUCGIR.0−12h (mg/kg) | 72 | 2426.82 | 72 | 2760.39 | 87.92 (82.02%; 94.23%) |
| AUCGIR.12−24h (mg/kg) | 72 | 959.46 | 72 | 980.54 | 97.85 (87.28%; 109.70%) |
| t GIR.max (h) d | 72 | 4.33 | 72 | 4.50 | ‐ |
| t 50%−GIR(early) (h) d | 72 | 1.55 | 72 | 1.48 | ‐ |
| t 50%−GIR(late) (h) d | 72 | 10.83 | 72 | 10.32 | ‐ |
| Onset of action (min) d | 72 | 36.0 | 72 | 32.0 | ‐ |
Abbreviations: AUCGIR.0−2h, area under the glucose infusion rate curve from 0 to 2 hours; AUCGIR.0−6h, area under the glucose infusion rate curve from 0 to 6 hours; AUCGIR.0−12h, area under the glucose infusion rate curve from 0 to 12 hours; AUCGIR.0−24h, area under the glucose infusion rate curve from 0 to 24 hours; AUCGIR.12−24h, area under the glucose infusion rate curve from 12 to 24 hours; AUCins.0−2h, area under the insulin concentration curve from 0 to 2 hours; AUCins.0−6h, area under the insulin concentration curve from 0 to 6 hours, AUCins.0−12h, area under the insulin concentration curve from 0 to 12 hours; AUCins.0−24h, area under the insulin concentration curve from 0 to 24 hours; AUCins.12−24h, area under the insulin concentration curve from 12 to 24 hours; AUCins.0−∞, area under the insulin concentration curve from 0 to ∞; CI, confidence interval; Cins.max, maximum observed insulin concentration; GIRmax, maximum glucose infusion rate; λ z, terminal elimination rate constant of insulin; LS‐mean, least square mean; N, number of healthy subjects; PD, pharmacodynamics; PK, pharmacokinetics; PPP, per‐protocol population; t GIR.max, time to maximum glucose infusion rate; t 50%−GIR(early), time from dosing to the first time point where the GIR was ≥GIRmax/2; t 50%−GIR(late), time from dosing to the first time point after t GIR.max where the GIR was ≤GIRmax/2; t 1/2, terminal elimination half‐life; t ins.max, time to maximum observed insulin concentration; t 50%−ins(early), time from dosing to the first time point where the concentration was ≥C ins.max/2; t 50%−ins(late), time from dosing to the first time point after t ins.max where the concentration was ≤C ins.max/2.
One subject was excluded from the PPP for PK because of a mix‐up of transfer tubes for PK sample collection.
Two profiles each receiving Biocon's Insulin‐70/30 and HUMULIN 70/30 were excluded from the PPP for PD based on C‐peptide exclusion rules and during the blinded database release meeting (because of extended gaps in continuous glucose monitoring), respectively.
AUCins.0−∞, λ z, and t ½ were only determined if the adjusted R‐square value of the regression lines was ≥0.7.
Median values are presented.
FIGURE 2.

C‐peptide–corrected mean insulin profiles: linear scale (PPP for PK). One subject was excluded from the PPP (n = 73) for PK analysis because of a mix‐up of transfer tubes for PK sample collection (with respect to time point). PK, pharmacokinetics; PPP, per‐protocol population
Sensitivity analysis assessment showed exclusion of three profiles (Biocon's Insulin‐70/30: two; HUMULIN‐70/30: one) from the primary PK endpoints applying C‐peptide–based exclusion rules. Results of sensitivity analysis, based on uncorrected data applying the C‐peptide–based exclusion rules, were similar to the primary analysis (AUCins.0−24h: 90% CI 91.62%, 98.58%; Cins.max: 90% CI 88.81%, 97.95%; both within the 80.00%‐125.00% limits), thus indicating the robustness of the study.
The secondary endpoint analyses showed mean values of secondary PK endpoints to be comparable between Biocon's Insulin‐70/30 and HUMULIN‐70/30 (Table 1 ). Although secondary endpoints were not expected to meet bioequivalence criteria, AUCins.0−2h, AUCins.0−6h, AUCins.0−12h, AUCins.12−24h and AUCins.0−∞ met the bioequivalence criteria.
3.2.2. PD endpoints
For the primary analysis, the same three profiles that were excluded from the sensitivity analysis of primary PK endpoints were excluded for the PD analysis. The 90% CIs for geometric mean ratios (Biocon's Insulin‐70/30/HUMULIN‐70/30) were within the 80.00%‐125.00% limits for both primary PD endpoints (Table 1 ). Mean GIR profiles were similar between Biocon's Insulin‐70/30 and HUMULIN‐70/30 (Figure 3 ). Mean AUCGIR.0−24h and GIRmax were equivalent for both the treatments.
FIGURE 3.

Mean GIR profiles (PPP for PD). Two profiles each receiving Biocon's Insulin‐70/30 and HUMULIN 70/30 were excluded from the PPP (n = 72) for PD based on C‐peptide exclusion rules and during the blinded database release meeting (because of extended gaps in continuous glucose monitoring), respectively. GIR, glucose infusion rates; PD, pharmacodynamics; PPP, per‐protocol population
Results of sensitivity analysis, without applying any C‐peptide–based exclusion rules, were similar to the primary analysis (AUCGIR.0−24h: 90% CI 84.87%, 96.91%; GIRmax: 90% CI 86.56%, 99.08%; both within the 80.00%‐125.00% limits), thus indicating the robustness of the study.
Mean values of secondary PD endpoints were overall comparable between Biocon's Insulin‐70/30 and HUMULIN‐70/30 (Table 1 ). Although secondary endpoints were not expected to meet the bioequivalence criteria, AUCGIR.0−6h, AUCGIR.0−12h and AUCGIR.12−24h met the criteria.
3.2.3. Clamp performance
No clamps were excluded based on predefined clamp quality criteria. Mean precision variability was less than 5% for both treatments. Mean deviation from the clamp target was 0.142 and 0.216 mg/dL for Biocon's Insulin‐70/30 and HUMULIN‐70/30, respectively. Based on the fulfilment of acceptability criteria, the clamp quality was considered as good and comparable between the treatments.
3.3. Safety
Overall, 43 treatment‐emergent AEs (TEAEs; Biocon's Insulin‐70/30: 21 TEAEs/18 subjects; HUMULIN‐70/30: 22 TEAEs/17 subjects) were reported during the study (Table 2 ). The most frequently reported AEs were headache and haematoma. All TEAEs were ‘resolved’, except for three events that were ‘resolving at follow‐up visit’ (Biocon's Insulin‐70/30: moderate thrombophlebitis; HUMULIN‐70/30: moderate infusion site pain and mild vascular access site inflammation), and all were considered unrelated to the study medication. The number of TEAEs was comparable between the two treatments (Biocon's Insulin‐70/30: injection site reaction [n = 1], headache [n = 6]; HUMULIN‐70/30: nausea [n = 1], injection site reaction [n = 2], headache [n = 6]). No serious AEs, deaths or discontinuations for safety/tolerability reasons occurred in the study. All the AEs were mild to moderate in intensity (Biocon's Insulin‐70/30: mild 18, moderate three; HUMULIN‐70/30: mild 16, moderate six).
TABLE 2.
Summary of treatment‐emergent adverse events (TEAEs)
| Biocon's Insulin 70/30 (N = 76) | HUMULIN 70/30 (N = 77) | |
|---|---|---|
| m, n (%) | m, n (%) | |
| All TEAEs | 21, 18 (23.7%) | 22, 17 (22.1%) |
| Causally related TEAEs | 7, 7 (9.2%) | 9, 7 (9.1%) |
| TEAEs by severity | ||
| Mild | 18, 16 (21.1%) | 16, 13 (16.9%) |
| Moderate | 3, 3 (3.9%) | 6, 6 (7.8%) |
| Severe | ‐ | ‐ |
| TEAEs by outcome | ||
| Recovered/resolved | 20, 17 (22.4%) | 20, 15 (19.5%) |
| Recovering/resolving | 1, 1 (1.3%) | 2, 2 (2.6%) |
| TEAEs by SOC | ||
| Gastrointestinal disorders | 1, 1 (1.3%) | 3, 3 (3.9%) |
| Abdominal pain upper | ‐ | 1, 1 (1.3%) |
| Diarrhoea | ‐ | 1, 1 (1.3%) |
| Nausea | 1, 1 (1.3%) | 1, 1 (1.3%) |
| General disorders and administration site conditions | 3, 2 (2.6%) | 3, 2 (2.6%) |
| Injection site reaction | 1, 1 (1.3%) | 2, 1 (1.3%) |
| Oedema | 1, 1 (1.3%) | ‐ |
| Oedema peripheral | ‐ | 1, 1 (1.3%) |
| Peripheral swelling | 1, 1 (1.3%) | ‐ |
| Infections and infestations | 1, 1 (1.3%) | 1, 1 (1.3%) |
| Gastroenteritis | ‐ | 1, 1 (1.3%) |
| Oral herpes | 1, 1 (1.3%) | ‐ |
| Injury, poisoning and procedural complications | 2, 2 (2.6%) | 6, 5 (6.5%) |
| Infusion site erythema | ‐ | 1, 1 (1.3%) |
| Infusion site haematoma | ‐ | 1, 1 (1.3%) |
| Infusion site pain | 1, 1 (1.3%) | 1, 1 (1.3%) |
| Infusion site swelling | 1, 1 (1.3%) | 1, 1 (1.3%) |
| Vascular access site inflammation | ‐ | 2, 2 (2.6%) |
| Musculoskeletal and connective tissue disorders | 2, 2 (2.6%) | ‐ |
| Pain in extremity | 2, 2 (2.6%) | ‐ |
| Nervous system disorders | 8, 8 (10.5%) | 7, 7 (9.1%) |
| Headache | 8, 8 (10.5%) | 7, 7 (9.1%) |
| Respiratory, thoracic and mediastinal disorders | ‐ | 1, 1 (1.3%) |
| Oropharyngeal pain | ‐ | 1, 1 (1.3%) |
| Vascular disorders | 4, 4 (5.3%) | 1, 1 (1.3%) |
| Haematoma | 3, 3 (3.9%) | 1, 1 (1.3%) |
| Thrombophlebitis | 1, 1 (1.3%) | ‐ |
| Hypoglycaemic TEAEs | ||
| Overall | 14, 11 (14.5%) | 21, 14 (18.2%) |
| Causally related hypoglycaemic TEAEs | 11, 9 (11.8%) | 18, 13 (16.9%) |
| Hypoglycaemic TEAEs by severity | ||
| Mild | 14, 11 (14.5%) | 18, 14 (18.2%) |
| Moderate | ‐ | 3, 1 (1.3%) |
| Hypoglycaemic TEAEs by outcome | ||
| Recovered/resolved | 14, 11 (14.5%) | 21, 14 (18.2%) |
| Clinically significant hypoglycaemia | ‐ | 3, 1 (1.3%) |
| Documented symptomatic hypoglycaemia | ‐ | 1, 1 (1.3%) |
| Asymptomatic hypoglycaemia | 14, 11 (14.5%) | 17, 13 (16.9%) |
Note: N, number of patients in the relevant treatment group; m, number of events; n, number of patients; % = n/N *100.
Abbreviations: SOC, system organ class; TEAEs, treatment‐emergent adverse events.
After treatment with HUMULIN‐70/30, three clinically significant hypoglycaemic events of moderate severity were observed in one subject (two related and one unlikely related) and one documented symptomatic hypoglycaemic event of mild severity was observed in another subject (deemed related). All events had the outcome of being ‘resolved’.
No clinically significant changes in vital signs, physical examinations or ECGs were observed. There were no other clinically significant findings in haematology, biochemistry or urinalysis clinical laboratory tests throughout the study.
4. DISCUSSION
This study showed similarity in rate and extent of absorption and glucose‐lowering activity between Biocon's Insulin‐70/30 and HUMULIN‐70/30. Demonstration of PK/PD equivalence was based on 90% CI, for the ratio of test and reference products, being contained within the predefined acceptance limits of 80.00%‐125.00%. 2 This was supplemented by similar results of the secondary endpoints.
A crossover, double‐blind, euglycaemic clamp trial using single subcutaneous doses of test and reference formulations is considered most suitable by the FDA 2 and European Medicines Agency 18 to simultaneously assess exposure and activity of biosimilar insulin products. 2 , 18 , 22 , 23 The clamp setting, based on an automated glucose clamp technique with continuous BG measurements and minute‐by‐minute adaptations of GIRs, achieves the highest clamp quality possible and reduces potential investigator‐related bias and the risk of any drug‐induced hypoglycaemia. 24 , 25 The quality of clamp performance is critical for interpretation of the data. 18 In this study, based on the fulfilment of acceptability criteria, the clamp quality (both precision and DFT data) was considered good and comparable between the treatments. A clamp duration of 24 hours, considered as a clinically meaningful treatment duration, was chosen to assess the complete single‐dose PK/PD profile of both drugs.
In clinical settings, the treatment dose is individualized based on body weight, metabolic needs, BG monitoring results and glycaemic control goal; and the same was followed in this study. The study included healthy male and postmenopausal female subjects, representing the most sensitive and thus appropriate trial population. Besides better availability, healthy subjects are known to exhibit lower intra‐individual variability. The study population was as per the inclusion and exclusion criteria specified (healthy males and postmenopausal females irrespective of race/ethnicity) in the protocol. However, the actual enrolment was determined by the availability of healthy volunteers at the study site, not excluding any race or gender per se. It is noted that in the healthy subjects, presence of endogenous insulin can potentially interfere with the PK/PD assessments. The following measures ensured suppression of endogenous insulin and/or minimized potential interference with the PK/PD results: (a) using 0.4 IU/kg dose, which is towards the higher end of the recommended range for insulin doses in clamp studies, (b) selecting a clamp target of 81 mg/dL ± 10% to facilitate suppression of endogenous insulin while avoiding induction of hypoglycaemia/counter‐regulatory hormones at the lowest end of the target range, (c) determining C‐peptide levels in parallel with insulin concentrations to identify subjects whose endogenous insulin production potentially interfered with insulin PK/PD measurements and (d) using C‐peptide–based correction methods for primary analyses of PK parameters and C‐peptide–based exclusion rules for primary analyses of PD parameters to further rule out any impact of endogenous insulin on PK/PD outcomes.
Both Biocon's Insulin‐70/30 and HUMULIN‐70/30 were generally well tolerated with no clinically relevant safety issues. Headache was the most prevalent AE in this study, as has been commonly reported in numerous other glucose clamp studies. 24 , 25 , 26 , 27 Clinically significant and documented symptomatic hypoglycaemic events that occurred during the glucose clamp procedure with HUMULIN‐70/30 were transient and resolved with intravenous glucose infusion. No clinically relevant differences were observed in the safety profiles of both the drug formulations with regard to type, frequency and severity of AEs; local tolerability; vital signs; physical examination; ECG; and clinical laboratory results.
Diabetes management is a lifelong process and has a significant economic impact on healthcare systems. 28 , 29 , 30 Rising costs make access to affordable insulin far more difficult for people with diabetes, especially mid‐ to low‐income individuals, those on high deductible health plans, or those who are uninsured in countries like the United States. According to T1 International's 2018 survey with 1478 respondents from 90 countries, insulin rationing is widespread, with 18% (253/1408) of all respondents and nearly 26% (162/627) of the US respondents reportedly having rationed insulin at least once in the previous year. 31 A recent editorial in Nature Biotechnology talks about the rising costs and hence inequitable access to insulin that is forcing one in four patients with diabetes in the United States to ration their insulin. 32
Biocon's biosimilar Insulin‐70/30 can provide reliable and affordable access to patients who are candidates for premixed‐insulins. A biosimilar cost advantage may also enable more patients who currently use vials and syringes to opt for pens. Insulin pens (reusable or disposable) are associated with better patient compliance because of accurate dosing, lower episodes of hypoglycaemia and less injection site pain, etc., and hence better control of diabetes in the long term. 33 , 34 , 35 A report on the coverage of insulin delivery devices by US‐managed care organizations shows that both the prescribers and the patients have a perception of insurance coverage preferring the usage of vials over pens. 33 This is, however, not the case, as the majority of insurance plans cover pens and vials to the same extent. 33 HUMULIN‐70/30 vials and pens have a common approved label from the FDA and the same would apply to Biocon's biosimilar Insulin‐70/30. As such, vials and pens may be used interchangeably as per the requirement of the patient.
In conclusion, this study demonstrated equivalence between Biocon's Insulin‐70/30 and HUMULIN‐70/30, when administered as a single subcutaneous injection, for the primary PK/PD endpoints. The study also showed equivalence for the secondary PK/PD endpoints between the two treatments. Both insulin preparations were well tolerated and had similar safety profiles.
AUTHOR CONTRIBUTIONS
Design: SNA, SL, AM and SMNM. Conduct/data collection: OK, LPM, GS, NS and JP. Analysis: GS, SL, AM, SMNM and GCL. Writing and review of the manuscript: all the authors. All the authors read and approved the final version of the manuscript. SNA, as the guarantor of this work, takes full responsibility for the work, including the study design, access to data and the decision to submit and publish the manuscript.
CONFLICT OF INTEREST
LPM has received speaker honoraria and travel grants from Eli Lilly, Gan & Lee Pharmaceuticals and Novo Nordisk. GS, AM, JP, SL and SNA are employees of Biocon Biologics Ltd and hold stocks in Biocon. SMNM and GCL are employees of Biocon Biologics Ltd. NS was employed with Biocon Biologics Ltd at the time of study conduct and preparation of the manuscript but is no longer employed by Biocon Biologics Ltd. OK is an employee of Profil, Neuss, Germany and has nothing to disclose.
PEER REVIEW
The peer review history for this article is available at https://publons.com/publon/10.1111/dom.14768.
Supporting information
Appendix S1 Glucose clamp procedure
Table S1 Routine laboratory safety tests
Table S2: Participant disposition, demographics and baseline characteristics
Table S3: Medical history and concomitant illness
ACKNOWLEDGEMENTS
Writing assistance under the direction of the authors was provided by Ubhayabharathi Gurunath (previously employed with Biocon Biologics Ltd). Editorial assistance was provided by Pooja Hurkat and Geetanjali Tonpe (both from Biocon Biologics Ltd).
Plum‐Mörschel L, Klein O, Singh G, et al. Pharmacokinetic and pharmacodynamic equivalence of Biocon's biosimilar Insulin 70/30 with US‐licensed HUMULIN® 70/30 formulation in healthy subjects: Results from the RHINE‐3 (Recombinant Human INsulin Equivalence‐3) study. Diabetes Obes Metab. 2022;24(9):1819‐1828. doi: 10.1111/dom.14768
Funding information Biocon SDN BHD.
DATA AVAILABILITY STATEMENT
The data that support the findings of this study are available from the corresponding author upon reasonable request.
REFERENCES
- 1. LaMattina J. Human insulin is available for as little as $24.88—but it's not for everyone. Forbes Magazine; 2021. https://www.forbes.com/sites/johnlamattina/2020/01/30/human-insulin-is-available-for-as-little-as-2488but-its-not-for-everyone/?sh=6ae2e46e5c48. Accessed February 16, 2022.
- 2. U.S. Food and Drug Administration . Clinical pharmacology data to support a demonstration of biosimilarity to a reference product: guidance for industry; 2016. https://www.fda.gov/media/88622/download. Accessed December 12, 2021.
- 3. U.S. Food and Drug Administration . Clinical immunogenicity considerations for biosimilar and interchangeable insulin products: guidance for industry; 2019. https://www.fda.gov/media/133014/download. Accessed February 15, 2022.
- 4. Endocrine Society . Addressing insulin access and affordability: an Endocrine Society position statement. J Clin Endocrinol Metab. 2021;106(4):935‐941. [DOI] [PubMed] [Google Scholar]
- 5. Ampudia‐Blasco FJ, Rossetti P, Ascaso JF. Basal plus basal‐bolus approach in type 2 diabetes. Diabetes Technol Ther. 2011;13(Suppl 1):S75‐S83. [DOI] [PubMed] [Google Scholar]
- 6. Janez A, Guja C, Mitrakou A, et al. Insulin therapy in adults with type 1 diabetes mellitus: a narrative review. Diabetes Ther. 2020;11:387‐409. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Kalra S, Czupryniak L, Kilov G, et al. Expert opinion: patient selection for premixed insulin formulations in diabetes care. Diabetes Ther. 2018;9(6):2185‐2199. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Cooppan R. Design and management of insulin therapy: integrating the new treatment tools. Medscape Diabetes Endocrinol. 2007;51(2):97‐102. [Google Scholar]
- 9. Silver B, Ramaiya K, Andrew SB, et al. EADSG guidelines: insulin therapy in diabetes. Diabetes Ther. 2018;9(2):449‐492. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Neborachko M, Pkhakadze A. Do all evidences say that the insulin analogues are more effective than human insulins. J Diab Meta Syndro. 2020;3:009 https://www.henrypublishinggroups.com/wp-content/uploads/2020/06/do-all-evidences-say-that-the-insulin-analogues-are-more-effective-than-human-insulins.pdf. Accessed February 9, 2022 [Google Scholar]
- 11. Diabetes.co.uk . Cost of diabetes; 2019. https://www.diabetes.co.uk/cost-of-diabetes.html. Accessed February 9, 2022.
- 12. Health Action International . Comment from Health Action International to the Expert Committee on the application to include long‐acting analogue insulins in the WHO Model List of Essential Medicines for type 1 diabetes; 2019. https://www.who.int/selection_medicines/committees/expert/22/applications/s18.5_insulin_analogues_HAI.pdf. Accessed February 9, 2022.
- 13. O'Neill Hayes T, Farmer J. Insulin cost and pricing trends; 2020. https://www.americanactionforum.org/print/?url=https://www.americanactionforum.org/research/insulin-cost-and-pricing-trends/. Accessed February 9, 2022.
- 14. Lee B. How much does insulin cost? Here's how 28 brands and generics compare; 2022. https://www.goodrx.com/healthcare-access/research/how-much-does-insulin-cost-compare-brands. Accessed February 17, 2022.
- 15. Beran D, Ewen M, Lipska K, Hirsch IB, Yudkin JS. Availability and affordability of essential medicines: implications for global diabetes treatment. Curr Diab Rep. 2018;18(8):48. [DOI] [PubMed] [Google Scholar]
- 16. Cefalu WT, Dawes DE, Gavlak G, et al. Insulin Access and Affordability Working Group: conclusions and recommendations. Diabetes Care. 2018;41(6):1299‐1311. [DOI] [PubMed] [Google Scholar]
- 17. Robin Respaut CT. U.S. insulin costs per patient nearly doubled from 2012 to 2016: study; 2019. https://www.reuters.com/article/us‐usa‐healthcare‐diabetes‐cost/u‐s‐insulin‐costs‐per‐patient‐nearly‐doubled‐from‐2012‐to‐2016‐study‐idUSKCN1PG136. Accessed February 9, 2022.
- 18. European Medicines Agency . Guideline on non‐clinical and clinical development of similar biological medicinal products containing recombinant human insulin and insulin analogues; 2015. https://www.ema.europa.eu/en/documents/scientific-guideline/guideline-non-clinical-clinical-development-similar-biological-medicinal-products-containing_en-0.pdf. Accessed February 16, 2022.
- 19. Plum‐Mörschel L, Singh G, Murugesan SMN, et al. Pharmacokinetic and pharmacodynamic equivalence of Biocon's biosimilar Insulin‐R with the US‐licensed Humulin® R formulation in healthy subjects: results from the RHINE‐1 (Recombinant Human INsulin Equivalence‐1) study. Diabetes Obes Metab. 2022;24(4):713‐721. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Benesch C, Heise T, Klein O, Heinemann L, Arnolds S. How to assess the quality of glucose clamps? Evaluation of clamps performed with ClampArt®, a novel automated clamp device. J Diabetes Sci Technol. 2015;9(4):792‐800. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Owens DR. Human Insulin: Clinical Pharmacological Studies in Normal Man. New York, NY: Springer Publishing; 1986. [Google Scholar]
- 22. DeFronzo RA, Tobin JD, Andres R. Glucose clamp technique: a method for quantifying insulin secretion and resistance. Am J Physiol. 1979;237(3):E214‐E223. [DOI] [PubMed] [Google Scholar]
- 23. Heinemann L, Anderson JH Jr. Measurement of insulin absorption and insulin action. Diabetes Technol Ther. 2004;6(5):698‐718. [DOI] [PubMed] [Google Scholar]
- 24. Liu D, Moberg E, Kollind M, Lins PE, Adamson U, Macdonald IA. Arterial, arterialized venous, venous and capillary blood glucose measurements in normal man during hyperinsulinaemic euglycaemia and hypoglycaemia. Diabetologia. 1992;35(3):287‐290. [DOI] [PubMed] [Google Scholar]
- 25. American Diabetes Association Workgroup on Hypoglycemia . Defining and reporting hypoglycemia in diabetes: a report from the American Diabetes Association Workgroup on Hypoglycemia. Diabetes Care. 2005;28(5):1245‐1249. [DOI] [PubMed] [Google Scholar]
- 26. Ocheltree SM, Hompesch M, Wondmagegnehu ET, Morrow L, Win K, Jacober SJ. Comparison of pharmacodynamic intrasubject variability of insulin lispro protamine suspension and insulin glargine in subjects with type 1 diabetes. Eur J Endocrinol. 2010;163(2):217‐223. [DOI] [PubMed] [Google Scholar]
- 27. Wolzt M, de la Peña A, Berclaz PY, Tibaldi FS, Gates JR, Muchmore DB. AIR inhaled insulin versus subcutaneous insulin: pharmacokinetics, glucodynamics, and pulmonary function in asthma. Diabetes Care. 2008;31(4):735‐740. [DOI] [PubMed] [Google Scholar]
- 28. American Diabetes Association . Economic costs of diabetes in the US in 2017. Diabetes Care. 2018;41(5):917‐928. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Alhowaish AK. Economic costs of diabetes in Saudi Arabia. J Family Community Med. 2013;20(1):1‐7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Seuring T, Archangelidi O, Suhrcke M. The economic costs of type 2 diabetes: a global systematic review. Pharmacoeconomics. 2015;33(8):811‐831. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. T1 International . Costs and rationing of insulin and diabetes supplies: findings from the 2018 T1 International patient survey; 2019. https://www.t1international.com/media/assets/file/T1International_Report_-_Costs_and_Rationing_of_Insulin__Diabetes_Supplies_2.pdf. Accessed February 15, 2022.
- 32. Insulin: the new battleground for drug pricing [Editorial]. Nat Biotechnol. 2022;40(1):1. [DOI] [PubMed] [Google Scholar]
- 33. Kimball ES, Aagren M. Coverage of insulin delivery devices and basal insulin analogs by US managed care organizations. J Med Econ. 2011;14(6):720‐728. [DOI] [PubMed] [Google Scholar]
- 34. Bastian MD, Wolters NE, Bright DR. Insulin pens vs. vials and syringes: differences in clinical and economic outcomes. Consult Pharm. 2011;26(6):426‐429. [DOI] [PubMed] [Google Scholar]
- 35. American Diabetes Association . Devices & technology. Insulin pens. https://www.diabetes.org/tools-support/devices-technology/insulin-pens. Accessed April 28, 2022.
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Appendix S1 Glucose clamp procedure
Table S1 Routine laboratory safety tests
Table S2: Participant disposition, demographics and baseline characteristics
Table S3: Medical history and concomitant illness
Data Availability Statement