Abstract
Purpose
To investigate the effect of Corneal Visualization Scheimpflug Technology tonometry (CST) on intraocular pressure (IOP).
Design
Cohort study.
Participants
Patients with and without primary open-angle glaucoma (POAG) were included.
Methods
Intraocular pressure was measured using the Icare rebound tonometer (ICRT; Icare Finland Oy) and the biomechanically corrected IOP (bIOP) using the CST. Intraocular pressure was measured at baseline with ICRT, followed by a CST measurement in one eye with the fellow eye acting as a control. Icare measurements were repeated at 10 seconds and 1, 2, 4, 8, 15, 30, and 60 minutes in both eyes. The ratio of test eye IOP to fellow eye IOP was used to control for intrasubject variation.
Main Outcome Measures
Intraocular pressure change following Corneal Visualization Scheimflug Technology tonometry.
Results
Forty participants (mean age, 54.09 ± 20.08 years) were included comprising 20 patients with POAG and 20 patients with no ocular abnormalities other than cataract. Mean central corneal thickness was similar in those without POAG (547.4 ± 55.05 μm) and with POAG (520.22 ± 37.59 μm; P = 0.14). No significant change was found in IOP measured with the ICRT in the fellow eye versus the 1-hour period in either the healthy (P = 0.87) or POAG (P = 0.92) group. Significant changes were found in IOP after CST measurement for both healthy (P < 0.01) and glaucomatous (P < 0.01) eyes. After the CST measurement, the IOP reduced continuously from a mean of 13.75 mmHg to 10.84 mmHg at 4 minutes for healthy eyes and from 13.28 mmHg to 11.11 mmHg at 8 minutes for glaucomatous eyes before approaching (83% for healthy eyes and 92% POAG eyes) the pre-CST measurement at 1 hour.
Conclusions
Corneal Visualization Scheimpflug Technology tonometry causes a significant reduction in IOP in both glaucomatous and healthy eyes that lasts for at least 1 hour afterward.
Keywords: Biomechanics, CST, Glaucoma, Intraocular pressure
Abbreviations and Acronyms: bIOP, biomechanically corrected intraocular pressure; CCT, central corneal thickness; CST, Corneal Visualization Scheimpflug Technology tonometry; GAT, Goldmann applanation tonometry; ICRT, Icare rebound tonometer; IOP, intraocular pressure; POAG, primary open-angle glaucoma
Glaucoma is one of the most common causes of blindness, with more than 60 million people affected worldwide.1,2 Elevated intraocular pressure (IOP) remains the major modifiable risk factor for the development and progression of the disease, making it the most important therapeutic entity.3 Since its invention in 1954, Goldmann applanation tonometry (GAT) has been the gold standard for measuring IOP.4 Goldmann applanation tonometry is a contact method of tonometry based on the Imbert-Fick law of applanation and requires local anesthesia.5 The Icare rebound tonometer (ICRT) has gained popularity recently as a handheld device that reproducibly measures IOP, comparable with GAT, without the need for anesthesia.6 The ICRT calculates the IOP by measuring the rebound acceleration of a probe that is propelled perpendicularly to the patient’s cornea from a distance of 4 to 8 mm.7 Apart from its ease of use, ICRT has an advantage over GAT in that it does not seem to influence IOP on repeated measurement in contrast to GAT, which may induce a reduction in IOP.8
Despite the popularity of GAT and ICRT in clinical practice, they do not compensate for corneal biomechanics, which has led to continued interest in new developments within the field of tonometry.9 The Corneal Visualization Scheimpflug Technology (CST) instrument (Oculus) uses an ultra–high-speed Scheimpflug camera to visualize the corneal response to an air-puff impulse, which deforms the cornea and anterior chamber.10 The changes in the deformation of the cornea induced by the impulse have been used to develop a biomechanically corrected IOP (bIOP).11 The bIOP is thought to be less influenced by age and corneal properties such as thickness when compared with GAT or ICRT.12,13 However, the CST causes a significant deformation of the cornea, and it is unclear whether this leads to an alteration of the anterior chamber and IOP and how long these effects may persist. Therefore, the aim of this study was to evaluate changes in IOP after use of a CST.
Methods
This retrospective study was conducted between April and September 2019 at The Royal Liverpool University Hospital. The study was approved by the institutional review board (identifier, A03058). The study adhered to the tenets of the Declaration of Helsinki. The requirement for informed consent was waived because of the retrospective nature of the study. Inclusion criteria were healthy patients undergoing routine ophthalmic examination and those with a diagnosis of bilateral primary open-angle glaucoma (POAG) of similar severity in both eyes. Primary open-angle glaucoma was defined as a characteristic glaucomatous progressive optic neuropathy in the absence of congenital or other ocular disease with or without raised IOP.14 Exclusion criteria included previous ocular surgery, secondary or angle-closure glaucoma, keratoconus or other corneal pathologic features, dense or hypermature cataract, a history of uveitis, or a combination thereof. Consecutive patients seeking treatment at the outpatient department for cataract assessment were included. The ICRT measurements were obtained 3 times in both eyes at baseline using the Icare ic100 (Icare Finland Oy) and after this, the CST was used to measure IOP once in 1 eye selected at random.
The eye that had not undergone CST served as the control. Measurement with CST was accepted as accurate when the “OK” quality index displayed on the device monitor. After CST, measurements were obtained in both eyes at 10, 60, 120, 240, 480, 900, 1800, and 3600 seconds using ICRT. Single-use probes were changed at each new examination with measurements obtained with the patient in the seated position at 90° to the corneal apex. Local anesthesia, dilating drops, or cycloplegics were not used in any of the examinations. Student t tests were used to compare baseline characteristics, including baseline IOP and central corneal thickness (CCT) measured using the Pachmate 2 (DGH Technology, Inc). The differences and the ratio of the test eye IOP and fellow eye IOP at baseline and at intervals after CST measurement were used to compare the 2 groups using repeated-measures analysis of variance using SPSS software version 25 (SPSS, Inc). A least squares method was used to fit curves to the changes in the ratio between the test and control eye using Maple software version 19 (Maplesoft).
Results
Forty participants were included comprising 20 patients with an established history of POAG and 20 patients with no ocular abnormalities other than cataract. Mean age of the patients was 54.09 ± 20.08 years (range, 21–89 years). Mean CCT was similar in those without glaucoma (547.4 ± 55.05 μm) and in those with glaucoma (520.22 ± 37.59 μm; P = 0.14). The bIOP and IOP were measured in 21 right eyes and 19 left eyes (healthy eyes: 10 right eyes and 10 left eyes; glaucoma group: 11 right eyes and 9 left eyes). No significant difference was found in the mean bIOP and IOP for the healthy control participants (14.06 ± 2.28 mmHg and 14.62 ± 2.55mmHg, respectively; P = 0.14) or those with glaucoma (14.29 ± 3.22 mmHg and 14.93 ± 3.84 mmHg, respectively; P = 0.07).
No significant change was found in the IOP measured with the ICRT in the control eye after measurement of the bIOP in the test eye over the 1-hour period in either the healthy (P = 0.87) or glaucomatous (P = 0.92) group. Significant changes in the IOP were found after CST measurement for both healthy (P < 0.01) and glaucomatous (P < 0.01) eyes (Figs 1 and 2). After the CST measurement, the IOP reduced continuously from a mean of 13.75 mmHg to 10.84 mmHg at 4 minutes for healthy eyes and from 13.28 mmHg to 11.11 mmHg at 8 minutes for glaucomatous eyes, before approaching (83% for healthy eyes and 92% for POAG eyes) the pre-CST measurement at 1 hour (Figs 1 and 2).
The reduction in the IOP after CST is evident as the ratio of the test eye IOP to fellow eye IOP, as shown in Figure 3. A plot of this ratio with time after the CST measurement shows that the greatest reduction in IOP occurred earlier and was greater in healthy eyes compared with glaucomatous eyes. A function of the form a + bxi + cxj + dxk + exl was used to provide a best least squares curve for the ratio of the IOP for healthy and glaucomatous eyes using MapleSoft 2019. This gave
for eyes with glaucoma and
for eyes without glaucoma. Extrapolating from these equations, it is expected that the IOP after CST would return to baseline at approximately 60 minutes in glaucomatous eyes and 90 minutes in the healthy eyes.
Discussion
Noncontact and corneal-compensated tonometry represent an expanding field of ophthalmic development,15 and the CST has enabled the development of a bIOP.16 We found that IOP measured with the ICRT and the bIOP were similar at baseline in both healthy and glaucomatous eyes and is in agreement with previous studies with average hysteresis and corneal thickness.17,18
However, after CST measurement, we found a significant drop in the IOP compared with the fellow eye in both healthy and glaucomatous eyes. Numerous studies have documented a reduction in IOP after GAT.19,20 This is likely to be multifactorial, but one explanation is that mechanical pressure on the anterior chamber caused by applanation alters the angle and increases the aqueous outflow, which subsequently reduces the IOP.20 Compared with GAT, the cornea and anterior chamber are deformed to a much greater extent using the CST. It is likely that this deformation alters IOP and aqueous outflow and accounts for the significant and prolonged reduction in IOP after CST.
However, it is of note that the reduction in IOP was greater and occurred earlier in the healthy versus the glaucomatous eyes. Although the patients were taking a prostaglandin analog, the magnitude of the observed effect is unlikely to be accounted for by the effect of the prostaglandin analog on corneal biomechanics.21 It is more probable that the difference may reflect greater resistance in outflow and or differences in scleral rigidity in eyes with glaucoma.22 Aqueous outflow facility is known to fluctuate and change.23 The relationship between IOP rise and aqueous outflow facility is not a new concept and was demonstrated in vitro in experiments involving filling of the anterior chambers of enucleated specimens.24,25 This relationship also was demonstrated during the same period that a rise in IOP occurs during scleral indentation followed by a reduction in IOP that is thought to be secondary to raised aqueous outflow.26 It is probable that a similar mechanism occurs with the CST.
For both healthy and glaucomatous eyes, the IOP had not yet returned to baseline levels by 1 hour. We did not measure the IOP after 1 hour, so it is difficult to predict when it returns to the pre-CST level. Extrapolating from the best fit curves suggests that healthy eyes take longer to reach baseline (70 minutes) than eyes with POAG (62 minutes), possibly because of the greater reduction in IOP in healthy eyes. Previous studies have demonstrated a good repeatability of bIOP using 3 consecutive CST measurements obtained at 1-minute intervals.12,27 This suggests that the CST induces little or no change in IOP, which is contrary to what we found in our study, or that the bIOP does not reflect the reduction in IOP. Changes in corneal curvature or the presence of corneal edema seem to influence the IOP measured on ICRT.28,29 We did not measure corneal curvature or thickness over the course of the study, and such alterations in these parameters may have contributed to the changes in IOP that we noted.
Because of the short interval between time points for measurement (e.g., 30 seconds, 1 minute) and changing the probe, it was not possible to repeat the ICRT at each interval. However, the Icare ic100 automatically measures the IOP 6 times, providing an average measurement only if the device deemed variation acceptable (standard deviation of measurements, <3.5 mmHg).30 Despite these limitations, it is clear that IOP changes significantly after use of the CST and is unlikely to return to baseline for at least 1 hour.
Acknowledgments
The authors thank Peter Wolfrum, Flavia Pennisi, and Chiara Bonzano.
Manuscript no. D-20-00009.
Footnotes
Disclosure(s): All authors have completed and submitted the ICMJE disclosures form.
The author(s) have no proprietary or commercial interest in any materials discussed in this article.
HUMAN SUBJECTS: Human subjects were included in this study. The human ethics committees at The Royal Liverpool University Hospital approved the study. The requirement for informed consent was waived because of the retrospective nature of the study. All research adhered to the tenets of the Declaration of Helsinki.
No animal subjects were included in this study.
Author Contributions:
Conception and design: Romano, Kaye
Analysis and interpretation: Borroni, Hristova, McLean, Rocha de Lossada, Kaye
Data collection: Borroni, Gadhvi, McLean, Rocha de Lossada
Obtained funding: Study was performed as part of the regular employment duties at the Royal Liverpool University. No additional funding was provided.
Overall responsibility: Borroni, Gadhvi, Hristova, McLean, Rocha de Lossada, Romano, Kaye
References
- 1.Wang Y.X., Xu L., Wei W.B., Jonas J.B. Intraocular pressure and its normal range adjusted for ocular and systemic parameters. The Beijing Eye Study 2011. PLoS One. 2018;13 doi: 10.1371/journal.pone.0196926. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Tham Y.-C., Li X., Wong T.Y., et al. Global prevalence of glaucoma and projections of glaucoma burden through 2040: a systematic review and meta-analysis. Ophthalmology. 2014;121:2081–2090. doi: 10.1016/j.ophtha.2014.05.013. [DOI] [PubMed] [Google Scholar]
- 3.Konstas A.G., Kahook M.Y., Araie M., et al. Diurnal and 24-h intraocular pressures in glaucoma: monitoring strategies and impact on prognosis and treatment. Adv Ther. 2018;35:1775–1804. doi: 10.1007/s12325-018-0812-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Zhang H., Sun Z., Li L., et al. Comparison of intraocular pressure measured by ocular response analyzer and Goldmann applanation tonometer after corneal refractive surgery: a systematic review and meta-analysis. BMC Ophthalmol. 2020;20 doi: 10.1186/s12886-019-1288-6. 23–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Zakrzewska A., Wiącek M.P., Machalińska A. Impact of corneal parameters on intraocular pressure measurements in different tonometry methods. Int J Ophthalmol. 2019;12:1853–1858. doi: 10.18240/ijo.2019.12.06. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Kutzscher A.E., Kumar R.S., Ramgopal B., et al. Reproducibility of 5 methods of ocular tonometry. Ophthalmology Glaucoma. 2019;2:429–434. doi: 10.1016/j.ogla.2019.07.006. [DOI] [PubMed] [Google Scholar]
- 7.Lamparter J., Hoffmann E.M. [Measuring intraocular pressure by different methods] Ophthalmologe. 2009;106:676–682. doi: 10.1007/s00347-009-1971-8. [DOI] [PubMed] [Google Scholar]
- 8.Jóhannesson G., Hallberg P., Eklund A., et al. Effects of topical anaesthetics and repeated tonometry on intraocular pressure. Acta Ophthalmol. 2014;92:111–115. doi: 10.1111/aos.12058. [DOI] [PubMed] [Google Scholar]
- 9.Sedaghat M.-R., Momeni-Moghaddam H., Yekta A., et al. Biomechanically-corrected intraocular pressure compared to pressure measured with commonly used tonometers in normal subjects. Clin Optom (Auckl) 2019;11:127–133. doi: 10.2147/OPTO.S220776. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Tubtimthong A., Chansangpetch S., Ratprasatporn N., et al. Comparison of corneal biomechanical properties among axial myopic, nonaxial myopic, and nonmyopic eyes. Biomed Res Int. 2020;2020:8618615. doi: 10.1155/2020/8618615. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Hirasawa K., Nakakura S., Nakao Y., et al. Changes in corneal biomechanics and intraocular pressure following cataract surgery. Am J Ophthalmol. 2018;195:26–35. doi: 10.1016/j.ajo.2018.07.025. [DOI] [PubMed] [Google Scholar]
- 12.Matsuura M., Murata H., Fujino Y., et al. Repeatability of the novel intraocular pressure measurement from Corvis ST. Transl Vis Sci Technol. 2019;8:48. doi: 10.1167/tvst.8.3.48. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Bao F., Huang Z., Huang J., et al. Clinical evaluation of methods to correct intraocular pressure measurements by the Goldmann applanation tonometer, Ocular Response Analyzer, and Corvis ST tonometer for the effects of corneal stiffness parameters. J Glaucoma. 2016;25:510–519. doi: 10.1097/IJG.0000000000000359. [DOI] [PubMed] [Google Scholar]
- 14.European Glaucoma Society Terminology and Guidelines for Glaucoma, 4th edition—chapter 2: Classification and terminology. Supported by the EGS Foundation: Part 1: Foreword; Introduction; Glossary; Chapter 2 Classification and terminology. Br J Ophthalmol. 2017;101:73–127. doi: 10.1136/bjophthalmol-2016-EGSguideline.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Maklad O., Eliasy A., Chen K.-J., et al. Simulation of air puff tonometry test using arbitrary Lagrangian-Eulerian (ALE) deforming mesh for corneal material characterisation. Int J Environ Res Public Health. 2019;17(1):54. doi: 10.3390/ijerph17010054. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Salvetat M.L., Zeppieri M., Tosoni C., et al. Corneal deformation parameters provided by the Corvis-ST pachy-tonometer in healthy subjects and glaucoma patients. J Glaucoma. 2015;24:568–574. doi: 10.1097/IJG.0000000000000133. [DOI] [PubMed] [Google Scholar]
- 17.Bañeros-Rojas P., Martinez de la Casa J.M., Arribas-Pardo P., et al. [Comparison between Goldmann, Icare Pro and Corvis ST tonometry] Arch Soc Esp Oftalmol. 2014;89:260–264. doi: 10.1016/j.oftal.2014.02.001. [DOI] [PubMed] [Google Scholar]
- 18.Kuebler A.G., Wiecha C., Reznicek L., et al. Comparison of different devices to measure the intraocular pressure in thyroid-associated orbitopathy. Graefes Arch Clin Exp Ophthalmol. 2019;257:2025–2032. doi: 10.1007/s00417-019-04367-2. [DOI] [PubMed] [Google Scholar]
- 19.Yaoeda K., Fukushima A., Shirakashi M., et al. Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma. Clin Ophthalmol. 2018;12:1473–1478. doi: 10.2147/OPTH.S174277. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Gaton D.D., Ehrenberg M., Lusky M., et al. Effect of repeated applanation tonometry on the accuracy of intraocular pressure measurements. Curr Eye Res. 2010;35:475–479. doi: 10.3109/02713681003678824. [DOI] [PubMed] [Google Scholar]
- 21.Amano S., Nejima R., Inoue K., et al. Effect of topical prostaglandins on the biomechanics and shape of the cornea. Graefes Arch Clin Exp Ophthalmol. 2019;257:2213–2219. doi: 10.1007/s00417-019-04435-7. [DOI] [PubMed] [Google Scholar]
- 22.Dastiridou A.I., Tsironi E.E., Tsilimbaris M.K., et al. Ocular rigidity, outflow facility, ocular pulse amplitude, and pulsatile ocular blood flow in open-angle glaucoma: a manometric study. Invest Ophthalmol Vis Sci. 2013;54:4571–4577. doi: 10.1167/iovs.13-12303. [DOI] [PubMed] [Google Scholar]
- 23.Carreon T., van der Merwe E., Fellman R.L., et al. Aqueous outflow—a continuum from trabecular meshwork to episcleral veins. Prog Retin Eye Res. 2017;57:108–133. doi: 10.1016/j.preteyeres.2016.12.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Moses R.A. The effect of intraocular pressure on resistance to outflow. Surv Ophthalmol. 1977;22:88–100. doi: 10.1016/0039-6257(77)90088-1. [DOI] [PubMed] [Google Scholar]
- 25.Brubaker R.F. The effect of intraocular pressure on conventional outflow resistance in the enucleated human eye. Invest Ophthalmol. 1975;14:286–292. [PubMed] [Google Scholar]
- 26.Semes L.P., Blash M.M., Woolley T. Intraocular pressure response to scleral indentation. Optom Vis Sci. 1993;70:729–732. doi: 10.1097/00006324-199309000-00008. [DOI] [PubMed] [Google Scholar]
- 27.Lopes B.T., Roberts C.J., Elsheikh A., et al. Repeatability and reproducibility of intraocular pressure and dynamic corneal response parameters assessed by the Corvis ST. J Ophthalmol. 2017;2017:8515742. doi: 10.1155/2017/8515742. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Salvetat M.L., Zeppieri M., Miani F., et al. Comparison of iCare tonometer and Goldmann applanation tonometry in normal corneas and in eyes with automated lamellar and penetrating keratoplasty. Eye (Lond) 2011;25:642–650. doi: 10.1038/eye.2011.60. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Shimmyo M., Ross A.J., Moy A., Mostafavi R. Intraocular pressure, Goldmann applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians, Hispanics, and African Americans. Am J Ophthalmol. 2003;136:603–613. doi: 10.1016/s0002-9394(03)00424-0. [DOI] [PubMed] [Google Scholar]
- 30.Nakakura S. Icare(®) rebound tonometers: review of their characteristics and ease of use. Clin Ophthalmol. 2018;12:1245–1253. doi: 10.2147/OPTH.S163092. [DOI] [PMC free article] [PubMed] [Google Scholar]