Computed tomographic pulmonary angiography: Three cases of low-tube-voltage acquisition with a slow injection of contrast medium

Atsushi Urikura 1,2,, Tsukasa Yoshida 1, Masahiro Endo 3, Koiku Asakura 1, Rui Sato 4, Atsushi Saiga 4, Michihisa Moriguchi 5, Kazuaki Nakashima 6, Takeshi Aramaki 1,4
PMCID: PMC9575461  PMID: 36262117

Abstract

Acute pulmonary thromboembolism occurring during cancer treatment has been increasing with the number of cancer patients and chemotherapy cases. Computed tomographic pulmonary angiography (CTPA) for evaluating the pulmonary artery is generally performed using rapid injection of contrast medium. However, intravenous catheters for contrast medium injection might cause extravasation due to rapid injection. This case series describes three patients who underwent contrast-enhanced computed tomography combined with low-tube-voltage imaging and slow injection. Low-tube-voltage slow-injection CTPA can be an effective technique for obtaining high contrast enhancement while accommodating fragile veins and low injection rates.

Keywords: Pulmonary thromboembolism, computed tomography, contrast enhancement, low-tube-voltage, slow injection

Introduction

The recent increase in chemotherapy with molecular-targeted drugs has also promoted an increase in thromboembolism. The diagnosis of acute pulmonary thromboembolism (PTE) via computed tomography (CT) is versatile due to its high spatial resolution and simplicity.1 The important roles of CT diagnosis include both determining the treatment efficacy in patients with carcinoma and diagnosing PTEs that need to be clinically ruled out.

Contrast-enhanced CT is effective in diagnosing subsegmental branches of the pulmonary artery and is the first choice for imaging. Generally, computed tomographic pulmonary angiography (CTPA) has been used to evaluate the pulmonary artery and is performed under rapid contrast medium injection.2 However, certain conditions increase the difficulty of CTPA. Notably, intravenous catheters for contrast medium injection might cause extravasation due to rapid injection. To reduce the risk of extravasation, slow injection of contrast medium can be one solution; however, low fractional doses result in a decrease in the CT values of the pulmonary vessels.

We have some experience with contrast-enhanced CT combined with low-tube-voltage imaging and CTPA under slow injection without a decrease in CT values. We found that the contrast enhancement effect of low-tube-voltage imaging was comparable to that of conventional CTPA. In this case series, we detail our experience with three patients who underwent CTPA using low-tube-voltage CT under slow injection.

Case reports

Detailed patient characteristics and scan conditions are presented in Table 1, whereas contrast enhancement attenuations for the PTE phase are summarized in Table 2. The multi-detector-row CT scanners used were Aquilion Precision (Canon Medical Systems, Tochigi, Japan) in Cases 1 and 3 and Aquilion ONE (Canon Medical Systems, Tochigi, Japan) in Case 2. Synapse Vincent software (Fujifilm Medical, Tokyo, Japan) was used to measure the attenuations in the main trunk of the pulmonary artery, right and left pulmonary arteries, ascending aorta, and superior vena cava (SVC) with a circular region of interest larger than 60 mm2.

Table 1.

Detailed case characteristics and data acquisition.

Case 1 Case 2 Case 3
Age (years) 80 79 52
Sex Female Male Female
Height (cm) 155.3 158.6 161
Body weight (kg) 50.5 53.3 65.3
Body mass index (kg/m2) 20.9 21.2 25.2
eGFR (mL/min/1.73 m2) 32.2 88.7 71.5
Gage of catheter 24 22 22
Iodine concentration (mg/mL) 300 300 370
Contrast volume (mL) 70 96 97
Injection rate (mL/s) 0.7 1.5 1.0
Iodine dose (mg iodine/kg) 420 540 552
Fractional dose (mL/kg/s) 4.2 8.5 5.5
Time to scan for PTE phase (s) 80 50 80
Inhale breath holding Yes No Yes
CT scanner Aquilion Precision Aquilion ONE Aquilion Precision
Tube voltage (kVp) 100 100 100
Tube current (mA) 50–500 50–800 50–500
Collimation (mm) 80 × 0.5 100 × 0.5 80 × 0.5
Pitch 1.39 0.87 1.39
Thickness (mm) 2 2 2
CTDIvol (mGy) 7.0 15.0 7.1
DLP (mGy cm) 250 645 278
Scan time (s) 3.3 2.7 3.6

The tube current values represent the minimum and maximum values for automatic tube current modulation; specifically, 10 Hounsfield unit of standard deviation at 5 mm thick was used.

eGFR: estimated glomerular filtration rate; CTDIvol: volume computed tomography dose index; DLP: dose length product.

Table 2.

Contrast enhancement attenuation for each anatomical location.

Case 1 Case 2 Case 3
Main trunk of pulmonary artery 286.3 321.0 295.8
Right pulmonary artery 274.6 299.8 286.0
Left pulmonary artery 263.8 301.0 290.5
Average attenuation in pulmonary artery 274.9 307.3 290.8
Left atrium 232.9 279.8 285.3
Acceding aorta 239.2 293.4 242.4
Superior vena cava 390.1 446.6 346.8

Values represent CT values (Hounsfield unit). The attenuations were measured with a circular region of interest larger than 60 mm2 using Synapse Vincent software (Fujifilm Medical, Tokyo, Japan).

Case 1

Case 1 was an 80-year-old woman who had undergone surgery for uterine cancer and adenocarcinoma of the lung. Contrast-enhanced chest-to-pelvis CT was performed to evaluate for recurrence and rule out PTE.

A small peripheral catheter (24-gage) was inserted intravenously for a slow contrast medium infusion at 0.7 mL/s (70 mL of contrast agent containing 300 mg/mL of iodine). Scanning was commenced 80 s after the contrast medium was injected. The mean pulmonary artery attenuation was 274.9 Hounsfield units (HU). PTE was ruled out following CTPA examination. The iodine dose was reduced by approximately 25% due to reduced renal function. The use of a low tube voltage (100 kVp) was effective in maintaining contrast enhancement (Figures 1).

Figure 1.

Figure
1.

An 80-year-old woman underwent chest-to-pelvic computed tomography (CT) for evaluating recurrence and ruling out a pulmonary artery thrombus. (a, b) Two different levels of axial images and (c) a coronal image. The iodine dose was reduced (420 mg iodine/kg) due to reduced renal function. CT values were 274.9, 232.9, 239.2, and 390.1 Hounsfield units (HU) at the pulmonary artery, left atrium, acceding aorta, and superior vena cava, respectively. Images were displayed with a window level and width of 60 and 300 HU, respectively.

Case 2

Case 2 was a 79-year-old man who had undergone surgery for a right abscess that had developed as a complication after resection of the lower right lobe of the lung. This case developed acute respiratory distress syndrome with severe pneumonia (Figure 2).

Figure 2.

Figure
2.

A 79-year-old man underwent surgery for a right abscess as a complication of resection of the lower right lobe of the lung. Same-level axial images reconstructed by (a) soft and (b) lung kernel. (c) A coronal image. Diffuse consolidation and ground-grass opacity can be observed in both lungs. CT values were 307.3, 279.8, 293.4, and 446.6 Hounsfield units (HU) at the pulmonary artery, left atrium, acceding aorta, and superior vena cava, respectively. Although the attenuation of SVC was the highest among the three cases, no evidence of streak artifacts was present. Images were displayed with a window level and width of 60 and 300 HU, respectively.

Contrast-enhanced chest-to-pelvis CT was performed to rule out a pulmonary thrombus and determine the source of his fever. Although a dynamic CTPA was attempted under ventilator management, the fragile veins made it difficult to secure a large-gage catheter. A 22-gage catheter was inserted intravenously, after which contrast medium was injected at a rate of 1.5 mL/s (96 mL of contrast agent containing 300 mg/mL of iodine). Scanning was commenced 50 s after the contrast medium was injected. Given that it was impossible to perform the scan under breath-hold, motion artifacts were reduced by high-speed acquisition. The mean CT value of the pulmonary artery was 307 HU, ruling out PTE.

Case 3

Case 3 involved a 52-year-old woman under chemotherapy after primary transverse colon cancer resection with hepatic and peritoneal metastases. Chest-to-pelvis contrast-enhanced CT was performed to determine the efficacy of chemotherapy and exclude PTE. A 22-gage catheter was inserted intravenously, after which contrast medium was injected at a rate of 1.0 mL/s (97 mL of contrast agent containing 370 mg/mL of iodine). Scanning was commenced 80 s after the contrast medium was injected.

Here, thrombi were diagnosed in both lower lobe pulmonary arteries as shown in Figure 3. Moreover, we found that CTPA with slow injection can depict thrombus in the subregional branches or those located more peripherally.

Figure 3.

Figure 3.

A 52-year-old woman received chemotherapy after resection of primary transverse colon cancer with hepatic and peritoneal metastases. Two different levels of (a, b) axial images and (c, d) coronal images. A thrombus can be observed in the peripheral pulmonary artery of the left lower lobe (arrowhead). The mean CT value of the pulmonary artery was 290.8 Hounsfield units (HU), which was sufficient to identify a small thrombus. Images were displayed with a window level and width of 60 and 300 HU, respectively.

Discussion

This series compiles cases in whom diagnosis of PTE was underwent via CTPA under slow contrast medium injection. CTPA is the first-line imaging procedure for PTE diagnosis.3,4 Generally, CTPA requires rapid contrast medium injection at a rate of approximately 3.5–4.0 mL/s.2 A large-gauge peripheral catheter must be placed intravenously to safely perform CTPA. Patients with carcinoma have an increased incidence of PTE5 and often possess fragile veins in the upper extremities due to frequent chemotherapy or other treatments. Dynamic CT with catheters placed in fragile vessels often lead to extravasation. The slow-injection CTPA allows for contrast enhancement compared to that of conventional CTPA in cases wherein placing large-gage catheters prove difficult. The mean CT values of the pulmonary artery in all three cases ranged from 275 to 307 HU. Scanning was commenced when 75%–80% of the contrast volume had been injected. To maintain the contrast effect during CTPA at low fractional doses, a tube voltage of 100 kVp was used. One study showed that conventional CTPA performed under rapid injection was essential in single-helical CT given its insufficient ability to diagnose subsegmental thrombosis.6 However, our proposed imaging protocol provides sufficient diagnostic performance with a high spatial and temporal resolution multi-detector-row CT.

The saline flush often performed in CTPA7 could be omitted via slow injection. In the current cases, the maximum CT value of the SVC was 447 HU, with no strong beam hardening artifact having been observed (as shown in Case 2). The use of a significantly lower fractional dose compared to conventional CTPA allowed for a reduction in the extreme concentration increases and artifacts of the SVC.8 The omission of the saline flush has the advantage of increased throughput during examination in addition to economic benefits.

An inspiratory transient attenuation artifact, which is rarely encountered in CTPA, can be characterized as a temporary decrease in contrast enhancement in the right heart system caused by inspiration.9,10 This artifact decreases the diagnostic performance of PTE. Exhalation CT is effective in preventing artifacts. However, in cases where respiratory control is impossible, such as in Case 2, or when the diagnosis of pneumonia is required, expiration reduces diagnostic performance. Slow-injection CTPA is performed after a long delay time. Given that the injected contrast agent is already circulating throughout the body and that contrast enhancement from the inferior vena cava has also been achieved, it is unlikely that an inspiratory transient attenuation artifact would occur even when imaging is performed under inspiration.

The slow-injection low-tube-voltage protocol proposed herein might cause dose starvation in obese patients due to a maximum tube current limitation. Advanced protocols using dual energy CT and/or deep learning–based reconstruction might solve these problems.11,12

In conclusion, low-tube-voltage slow-injection CTPA is a promising technique for achieving high contrast enhancement while accommodating fragile veins and low injection rates. Although our proposed method does not require a saline flush, artifacts produced by the SVC are minimal.

Footnotes

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

Informed consent statement: This retrospective study was approved by the institutional review board (J2022-17). Given the retrospective, noninvasive nature of this study, the requirement for written informed consent from the patients was waived and information was instead made available on the institutional website.

ORCID iD

Atsushi Urikura https://orcid.org/0000-0002-5799-0944

Kazuaki Nakashima https://orcid.org/0000-0002-1023-0454

References

  • 1.Seon HJ, Kim KH, Lee WS, et al. Usefulness of computed tomographic pulmonary angiography in the risk stratification of acute pulmonary thromboembolism. Comparison with cardiac biomarkers. Circ J 2011; 75: 428–436. [DOI] [PubMed] [Google Scholar]
  • 2.Stein PD, Yaekoub AY, Matta F, et al. Resolution of pulmonary embolism on CT pulmonary angiography. AJR Am J Roentgenol 2010; 194: 1263–1268. [DOI] [PubMed] [Google Scholar]
  • 3.Perrier A, Howarth N, Didier D, et al. Performance of helical computed tomography in unselected outpatients with suspected pulmonary embolism. Ann Intern Med 2001; 135: 88–97. [DOI] [PubMed] [Google Scholar]
  • 4.Schoepf UJ, Costello P. CT angiography for diagnosis of pulmonary embolism: state of the art. Radiology 2004; 230: 329–337. [DOI] [PubMed] [Google Scholar]
  • 5.Prandoni P, Lensing AW, Piccioli A, et al. Recurrent venous thromboembolism and bleeding complications during anticoagulant treatment in patients with cancer and venous thrombosis. Blood 2002; 100: 3484–3488. [DOI] [PubMed] [Google Scholar]
  • 6.Goodman LR, Curtin JJ, Mewissen MW, et al. Detection of pulmonary embolism in patients with unresolved clinical and scintigraphic diagnosis: helical CT versus angiography. AJR Am J Roentgenol 1995; 164: 1369–1374. [DOI] [PubMed] [Google Scholar]
  • 7.Jones SE, Wittram C. The indeterminate CT pulmonary angiogram: imaging characteristics and patient clinical outcome. Radiology 2005; 237: 329–337. [DOI] [PubMed] [Google Scholar]
  • 8.Kalisz K, Buethe J, Saboo SS, et al. Artifacts at Cardiac CT: Physics and Solutions. Radiographics 2016; 36: 2064–2083. [DOI] [PubMed] [Google Scholar]
  • 9.Mortimer AM, Singh RK, Hughes J, et al. Use of expiratory CT pulmonary angiography to reduce inspiration and breath-hold associated artefact: contrast dynamics and implications for scan protocol. Clin Radiol 2011; 66: 1159–1166. [DOI] [PubMed] [Google Scholar]
  • 10.Chen YH, Velayudhan V, Weltman DI, et al. Waiting to exhale: salvaging the nondiagnostic CT pulmonary angiogram by using expiratory imaging to improve contrast dynamics. Emerg Radiology 2008; 15: 161–169. [DOI] [PubMed] [Google Scholar]
  • 11.Ohana M, Jeung MY, Labani A, et al. Thoracic dual energy CT: acquisition protocols, current applications and future developments. Diagn Interv Imaging 2014; 95: 1017–1026. [DOI] [PubMed] [Google Scholar]
  • 12.Urikura A, Yoshida T, Nakaya Y, et al. Deep learning-based reconstruction in ultra-high-resolution computed tomography: Can image noise caused by high definition detector and the miniaturization of matrix element size be improved? Phys Med 2021; 81: 121–129. [DOI] [PubMed] [Google Scholar]

Articles from Acta Radiologica Open are provided here courtesy of SAGE Publications

RESOURCES