Abstract
Background: Chronic lung allograft dysfunction (CLAD) is the major cause of death beyond 2 years after lung transplantation and develops in 50% of all patients by 5 years post-transplant. CLAD is diagnosed on the basis of a sustained drop of 20% for at least 3 months in the forced expiratory volume (FEV1), compared to the best baseline value achieved post-transplant. CLAD presents as two main phenotypes: bronchiolitis obliterans syndrome (BOS) is more common and has better prognosis than restrictive allograft syndrome (RAS). Respiratory oscillometry is a different modality of lung function testing that is highly sensitive to lung mechanics. The current study investigated whether spectral and intrabreath oscillometry can differentiate between CLAD-free, BOS- and RAS-CLAD at CLAD onset, i.e., at the time of the initial 20% drop in the FEV1.
Methods: A retrospective, cross-sectional analysis of 263 double lung transplant recipients who underwent paired testing with oscillometry and spirometry at the Toronto General Pulmonary Function Laboratory from 2017 to 2022 was conducted. All pulmonary function testing and CLAD diagnostics were performed following international guidelines. Statistical analysis was conducted using multiple comparisons.
Findings: The RAS (n = 6) spectral oscillometry pattern differs from CLAD-free (n = 225) by right-ward shift of reactance curve similar to idiopathic pulmonary fibrosis whereas BOS (n = 32) has a pattern similar to obstructive lung disease. Significant differences were found in most spectral and intrabreath parameters between BOS, RAS, and time-matched CLAD-free patients. Post-hoc analysis revealed these differences were primarily driven by BOS instead of RAS. While no differences were found between CLAD-free and RAS patients with regards to spectral oscillometry, the intrabreath metric of reactance at end-inspiration (XeI) was significantly different (p < 0.05). BOS and RAS were differentiated by spectral oscillometry measure R5, and intrabreath resistance at end expiration, ReE (p < 0.05 for both).
Conclusion: Both spectral and intrabreath oscillometry can differentiate BOS-CLAD from CLAD-free states while intrabreath oscillometry, specifically XeI, can uniquely distinguish RAS-CLAD from CLAD-free. Spectral and intrabreath oscillometry offer complementary information regarding lung mechanics in CLAD patients to help distinguish the two phenotypes and could prove useful in prognostication.
Keywords: oscillometry, lung transplant, CLAD, pulmonary function test (PFT), graft dysfunction
Introduction
Survival following lung transplant has steadily improved. However, chronic lung allograft dysfunction (CLAD) remains the main barrier to long-term survival (Parulekar and Kao, 2019; Thabut and Mal, 2017). CLAD is defined by an irreversible decline in forced expiratory volume in 1 second (FEV1) to below 80% of the best baseline value achieved post-transplant and sustained for at least 3 months after other reversible causes have been ruled out (Glanville et al., 2019; Sato et al., 2011). CLAD develops in 50% of patients by 5 years after lung transplant, and once established, is associated with a poor prognosis.
CLAD manifests as two main phenotypes, bronchiolitis obliterans syndrome (BOS) and restrictive allograft syndrome (RAS) (Kotecha et al., 2020; Verleden et al., 2014; Verleden et al., 2015; Verleden et al., 2019). BOS presents with an obstructive pattern on pulmonary function testing and histopathological lesions of obliterative bronchiolitis in the small airways but no concurrent interstitial fibrosis (Meyer et al., 2014; Parulekar and Kao, 2019) whereas RAS has a restrictive pattern with peripheral lung fibrosis on histology and pleuroparenchymal opacities on computed tomography (CT) imaging (Glanville et al., 2019; Sato et al., 2011; Verleden et al., 2014). While it is less common, RAS has a worse prognosis with a survival of 6–18 months (Sato et al., 2011) compared to 3–5 years in BOS (Kulkarni et al., 2019; Parulekar and Kao, 2019). Rarely, patients can present with a mixed phenotype. Patients with BOS can also evolve to RAS, with such transitions portending worse survival (Parulekar and Kao, 2019).
Currently, there is no specific treatment for CLAD. Clinical trials of potential therapies are hampered by the fact that the diagnosis of CLAD can only be confirmed 3 months after onset (i.e., the first 20% drop in FEV1). Thus, the benefit of early treatment cannot be assessed. Furthermore, CLAD is a progressive disease where the decline in allograft function begins before the 80% FEV1 threshold is reached. While spirometry provides useful information on the CLAD development, it is insensitive and non-specific to changes in the smaller airways, where the initial pathophysiological processes of CLAD manifest (Kouri et al., 2021; Stockley et al., 2017). Studies have shown that FEV1 remains unchanged until 75% of the small airways are obliterated (Burgel et al., 2013; Cosio et al., 1978), by which time CLAD is well advanced. If the diagnosis of CLAD could be confirmed early using sensitive markers that can predict progression to CLAD, there is potential for earlier intervention (i.e. before or at the time of the 20% drop in FEV1) and the possibility of preventing the establishment of CLAD.
Oscillometry is an increasingly employed pulmonary function test (PFT) modality that is highly sensitive to changes in respiratory mechanics (Bates et al., 2011). It measures the total respiratory impedance (Zrs) during normal quiet (or tidal) breathing, expressed as the respiratory resistance (Rrs) and reactance (Xrs). Rrs reflects the viscous losses in the respiratory system (with the resistance of the airways as the main determinant) while Xrs, at the lower frequencies, reflects the elasticity of the lungs and the chest wall. Standard (also known as spectral) oscillometry provides the mean values of Zrs at different frequencies over entire breath cycles. Intrabreath oscillometry is a novel modality that tracks changes in lung mechanics continuously during inspiration and expiration, and focuses on the zero-flow instants of breathing, such as end expiration and end inspiration, where the upper airway nonlinearities are minimal (Gray et al., 2019; Chiabai et al., 2021; Hantos, 2021; Makan et al., 2022). Spectral oscillometry has been found to be more sensitive than spirometry in detecting chronic obstructive pulmonary disease (COPD) and asthma in the small airways, and is able to distinguish COPD from interstitial lung disease (Dellacà et al., 2004; Fujii et al., 2015; Paredi et al., 2010; Sugiyama et al., 2013). Our group showed that spectral oscillometry follows changes associated with biopsy-proven acute lung allograft rejection when spirometry could not (Cho et al., 2020) and intrabreath oscillometry, specifically the reactance at end inspiration (XeI) is highly correlated with two independent markers of disease severity in idiopathic pulmonary fibrosis (Wu et al., 2022).
We hypothesized that oscillometry can provide additional information at time of CLAD onset (i.e., at initial time of 20% drop in the FEV1) to help in the early identification of CLAD. The current study investigated whether spectral and intrabreath oscillometry can differentiate between CLAD-free, BOS, or RAS at the time of CLAD onset.
Methods
The study was approved by the University Health Network (UHN) Research Ethics Board (REB# 17-5652). All double lung transplant recipients were eligible for enrollment. Single lung recipients, those who remained hospitalized at 3 months post-transplant, and patients who died before enrollment were excluded. Written informed consent were obtained prior to oscillometry. Oscillometry was conducted according to European Respiratory Society guidelines (King et al., 2020; Wu et al., 2020; Chang et al., 2022) using the tremoflo C-100 device (Thorasys, Montreal, QC, Canada). Spectral oscillometry was measured with the standard 5–37 Hz multifrequency signal, and intrabreath oscillometry was measured at 10 Hz frequency as previously described (Wu et al., 2022; Chang et al., 2022). A minimum of three recordings with a coefficient of variation of Rrs at 5 Hz of ≤10% was needed to pass quality control. Oscillometry was completed prior to spirometry, which were conducted without/with plethysmography as part of routine care. All testing occurred at the Toronto General Hospital (TGH) Pulmonary Function Laboratory. Lung transplant patients are followed at the TGH Pulmonary Function Laboratory weekly 3 months, then at 6, 9, 12, 18 and 24 months post-lung transplant and annually thereafter. Oscillometry is also performed prior to any additional spirometry that is requested for clinical indications.
CLAD was diagnosed and phenotyped according to International Society for Heart and Lung Transplantation guidelines (Glanville et al., 2019). Phenotyping was conducted by an experienced lung transplant physician with review of the patient’s electronic medical records, including pulmonary function tests, lung CT imaging, bronchoscopy findings and when available, histology of the lung obtained at time of retransplant or autopsy. Demographic data and clinical parameters known to affect lung function and graft rejection, including primary lung disease, donor–recipient human leukocyte antigen (HLA) match status, and cytomegalovirus (CMV) donor–recipient status, were prospectively collected and maintained in the Toronto Lung Transplant Database.
Statistical analyses were performed using Prism 6.0 (GraphPad Software) and RStudio version 4.1.1 (The R Foundation). Comparisons among groups were performed using one-way ANOVA for normal distributed variables and Kruskal–Wallis one-way analysis of variance for non-normal variables. Pearson’s chi-squared test was used for categorical variables. The data are shown as mean ± standard deviation (SD) or median and interquartile range (IQR), as appropriate. Pairwise comparisons between non-normally distributed variables were analyzed using unpaired Wilcoxon signed rank test with Bonferroni corrections.
Results
Demographics
Between December 2017 to July 2022, 798 patients underwent lung transplantation. Single lung transplants (n = 91) were excluded (Figure 1). Of the 707 double lung transplant recipients, we excluded patients due to early death post-transplant (n = 27), ongoing hospitalization at 3 months (n = 91) or declined participation (n = 77). Of the 512 patients enrolled, 101 patients were excluded from our analysis due to early dropout/early death (n = 97) or re-transplantation (n = 4). CLAD developed in 85 of the remaining 411 patients; 47 participants were excluded from the current analysis due to lack of oscillometry data within 4 weeks of CLAD onset (n = 20), other CLAD phenotypes (n = 27), or insufficient data to complete CLAD phenotyping (n = 13). Of the 38 CLAD patients with complete data for analysis, 32 had BOS and six had RAS. Each of these 38 CLAD patients were time-matched to at least two CLAD-free lung transplant patients who had oscillometry completed within 2 weeks of the CLAD-onset dates to create a CLAD-free control group (n = 225).
Comparison of the CLAD-free, BOS, and RAS groups revealed their baseline clinical characteristics were similar (Table 1). The primary indications for transplant were pulmonary fibrosis, COPD and cystic fibrosis. The CLAD-free and CLAD groups were similar with respect to sex, age and height at transplant except for the body mass index which was significantly different at time of CLAD-onset or time-matched date (CLAD-free). The three groups were also similar with respect to CMV match status and immunologic risk at time of transplant (Table 1). CLAD patients were time-matched to CLAD-free patients who had an oscillometry measurement within 2 weeks of CLAD onset to control for duration of follow-up post-transplant. This duration of follow-up post-transplant was similar among CLAD-free, BOS, and RAS groups (Table 1).
TABLE 1.
CLAD-free (n = 225) | CLAD (n = 38) | p-value | ||
---|---|---|---|---|
BOS (n = 32) | RAS (n = 6) | |||
Male, n (%) | 136 (60.4) | 17 (53.1) | 4 (66.7) | 0.688 |
Age, years | 60.0 [49.0, 66.0] | 54.5 [40.5, 61.5] | 63.5 [56.8, 68.8] | 0.094 |
Height, m | 1.70 ± 0.1 | 1.67 ± 0.1) | 1.75 ± 0.1 | 0.107 |
Weight, kg | 74.5 ± 16.7 | 81.0 ± 16.5 | 87.0 ± 14.1 | 0.030 |
BMI, kg/m2 | 26.4 ± 4.9 | 29.0 ± 6.0 | 28.3 ± 4.6 | 0.019 |
PRA Positive, n (%) | 108 (48.0) | 21 (65.6) | 3 (50.0) | 0.175 |
VCM Positive, n (%) | 56 (24.9) | 9 (28.1) | 0 (0.0) | 0.337 |
ACM Positive, n (%) | 22 (9.8) | 6 (18.8) | 0 (0.0) | 0.212 |
CMV match status, n (%) | 0.685 | |||
R−/D+ (mismatch) | 102 (45.3) | 18 (56.2) | 2 (33.3) | |
R−/D− (negative) | 45 (20.0) | 6 (18.8) | 2 (33.3) | |
R+/D+ (positive) | 78 (34.7) | 8 (25.0) | 2 (33.3) | |
Primary disease, n (%) | ||||
Pulmonary Fibrosis | 85 (37.8) | 10 (31.1) | 1 (16.7) | 0.165 |
COPD/Emphysema | 69 (30.7) | 7 (21.9) | 3 (50.0) | |
Cystic Fibrosis/Bronchiectasis | 24 (10.7) | 6 (18.8) | 1 (16.7) | |
Other | 47 (20.9) | 9 (28.1) | 1 (16.7) | |
Follow-up, days | 265.0 [168.0, 438.0] | 346.0 [195.8, 545.0] | 178.5 [163.3, 319.0] | 0.186 |
Values at CLAD-onset or time-matched date | ||||
FVC (L) | 3.04 ± 0.92 | 2.73 ± 1.16 | 2.58 ± 0.97 | 0.138 |
%predicted | 75.47 ± 18.77 | 65.62 ± 21.44 | 56.77 ± 14.70 | 0.002 |
FEV1 (L) | 2.34 ± 0.78 | 1.57 ± 0.67 | 2.04 ± 0.75 | <0.001 |
%predicted | 74.21 ± 22.30 | 48.28 ± 20.11 | 57.73 ± 11.80 | <0.001 |
FEV1/FVC (%) | 77.18 ± 12.89 | 58.23 ± 11.43 | 79.78 ± 6.31 | <0.001 |
%predicted | 98.16 ± 16.40 | 72.88 ± 14.20 | 103.13 ± 9.09 | <0.001 |
TLC (L) | 4.80 ± 1.17 | 4.78 ± 1.52 | 4.46 ± 1.03 | 0.825 |
%predicted | 78.63 ± 16.47 | 77.09 ± 16.79 | 63.74 ± 15.13 | 0.129 |
RV (L) | 1.72 ± 0.55 | 1.90 ± 0.49 | 1.65 ± 0.33 | 0.242 |
%predicted | 88.32 ± 33.63 | 104.51 ± 37.28 | 75.46 ± 22.64 | 0.039 |
RV/TLC (%) | 36.25 ± 8.71 | 41.24 ± 9.01 | 37.72 ± 8.36 | 0.018 |
%predicted | 99.98 ± 29.71 | 122.81 ± 39.01 | 103.38 ± 17.22 | 0.001 |
DLCO (ml/min/mmHg) | 15.35 ± 4.35 | 17.60 ± 5.50 | 11.17 ± 3.30 | 0.071 |
%predicted | 71.57 ± 13.73 | 73.61 ± 16.58 | 65.57 ± 17.81 | 0.679 |
ACM, actual cross match; BMI, body mass index; CMV, cytomegalovirus; COPD, chronic obstructive lung disease; D, donor; DLCO, diffusing capacity for carbon monoxide; FEV1, forced expiratory volume in one second; FVC, forced vital capacity; PRA, panel of reactive antibodies; R, recipient; RV, residual volume; TLC, total lung capacity; VCM, virtual cross match. Continuous normal variables are reported as mean (SD); non-normal variables as median [IQR].
Spectral oscillometry of the CLAD phenotypes
Figure 2 illustrates the impedance vs. frequency graphs of a CLAD-free, a BOS and a RAS patient (left panels). The pattern of the spectral oscillogram in BOS was characteristic of obstructive lung disease (Eddy et al., 2019), manifested in enhanced frequency dependence of Rrs (R5-R19) as an index of peripheral inhomogeneity with low values of X5 (reactance at 5 Hz) and high values of Ax (area of reactance), which are measures of lung elastance and ventilatory inhomogeneity, while RAS presents with a restrictive pattern that resembles idiopathic pulmonary fibrosis (Wu et al., 2022) with a primarily rightward shift in the reactance curve and moderately high Ax but normal resistance values.
Comparison of the spectral oscillometry parameters revealed significant differences in most parameters between CLAD phenotypes at CLAD-onset and those of time-matched CLAD-free patients, as shown in Table 2. Post-hoc analysis between BOS vs. CLAD-free and RAS vs. CLAD-free revealed that these differences were primarily observed in the 32 BOS patients (Bonferroni adjusted p < 0.0001 for all, Table 3) as the spectral oscillometry parameters in the six RAS patients revealed that they were similar to the CLAD-free group (Table 3). However, comparisons between BOS and RAS groups revealed significant differences in R5 at time of CLAD onset (Bonferroni adjusted p = 0.0234, Table 3).
TABLE 2.
CLAD-free (n = 225) | BOS (n = 32) | RAS (n = 6) | p-value | |
---|---|---|---|---|
Spectral oscillometry | ||||
R5 a | 3.46 [2.88, 4.40] | 4.60 [3.77, 5.46] | 3.55 [3.40, 3.96] | 0.0002 |
R5-19 a | 0.57 [0.23, 1.05] | 1.30 [0.88, 1.77] | 0.89 [0.75, 1.37] | <0.0001 |
X5 a | −1.58 [−2.21, −1.18] | −2.91 [−4.81, −2.15] | −2.41 [−2.87, −1.55] | <0.0001 |
Ax b | 9.0 [5.0, 17.1] | 22.6 [13.8, 38.5] | 18.1 [10.4, 18.8] | <0.0001 |
Fres (Hz) | 18.3 [13.8, 22.0] | 24.1 [21.0, 28.2] | 21.6 [18.1, 23.4] | <0.0001 |
Intrabreath Oscillometry | ||||
ReE a | 2.84 [2.32, 3.67] | 3.92 [3.25, 4.39] | 2.65 [2.37, 3.18] | 0.0004 |
ReI a | 2.29 [1.91, 2.80] | 2.58 [2.21, 3.22] | 2.25 [1.95, 2.52] | 0.0258 |
XeE a | −0.24 [−0.67, 0.08] | −1.44 [−2.44, −0.61] | −0.62 [−1.78, −0.35] | <0.0001 |
XeI a | −0.30 [−0.61, −0.10] | −0.74 [−1.27, −0.48] | −0.84 [−1.28, −0.62] | <0.0001 |
ARV c | −0.22 [−0.50, 0.01] | −0.24 [−0.53, 0.08] | −0.27 [−0.37, −0.13] | 0.8967 |
ARV′ d | 1.44 [0.93, 2.38] | 2.62 [1.90, 3.32] | 1.76 [1.15, 2.20] | <0.0001 |
AXV c | 0.30 [0.12, 0.74] | 0.69 [0.34, 1.40] | 0.42 [0.29, 0.59] | 0.0006 |
AXV′ d | −0.07 [−0.53, 0.20] | −1.19 [−2.52, −0.50] | −0.08 [−2.33, 0.30] | <0.0001 |
Units of measures: a cmH2O∙s/L; b cmH2O/L; c cmH2O∙s; d cmH2O.
R5, resistance at 5 Hz; R19, resistance at 19 Hz; R5–19, difference in resistance between 5 and 19 Hz; Ax, reactance area between 5 Hz and Fres; Fres, resonance frequency; ReE, resistance at end-expiration; ReI, resistance at end-inspiration; XeE, reactance at end-expiration; XeI, reactance at end-inspiration; ARV, area of resistance volume loop; ARV’, area of resistance flow loop; AXV, area of reactance volume loop; AXV’, area of reactance flow loop. Data are shown as median [IQR]. Statistics were performed with Kruskal–Wallis one-way analysis of variance.
TABLE 3.
BOS vs. CLAD-free | RAS vs. CLAD-free | BOS vs. RAS | ||||||
---|---|---|---|---|---|---|---|---|
Spectral oscillometry | ||||||||
R5 | <0.0001 | 0.7925 | 0.0234 | |||||
R5-19 | <0.0001 | 0.1665 | 0.2285 | |||||
X5 | <0.0001 | 0.2458 | 0.2133 | |||||
Ax | <0.0001 | 0.2262 | 0.1987 | |||||
Fres | <0.0001 | 0.1823 | 0.2446 | |||||
Intrabreath Oscillometry | ||||||||
ReE | 0.0001 | 0.7877 | 0.0410 | |||||
ReI | 0.0092 | 0.6183 | 0.0816 | |||||
XeE | <0.0001 | 0.1255 | 0.2970 | |||||
XeI | <0.0001 | 0.0170 | 0.7108 | |||||
ARV | 0.6694 | 0.8357 | 1.0000 | |||||
ARV′ | <0.0001 | 0.4230 | 0.1591 | |||||
AXV | 0.0001 | 0.5058 | 0.2285 | |||||
AXV′ | <0.0001 | 0.9728 | 0.1987 |
For definitions and units, see legend to Table 2.
Intrabreath oscillometry and CLAD
Intrabreath oscillometry at the time of CLAD onset (or the time-matched date in the CLAD-free group) was also significantly different amongst CLAD-free, BOS, and RAS patients (Table 2). These were observed in the post-hoc analysis of the resistance measurements at end-inspiration (ReI) and end-expiration (ReE) as well as the reactance values at end-inspiration (XeI) and end-expiration (XeE) (Table 3). Similarly, nearly all intrabreath oscillometry parameters, with the exception of the area of resistance-volume loop (ARV), were significantly different between BOS and its time-matched CLAD-free cohort (Table 3). In contrast to the findings of spectral oscillometry, the intrabreath metric of reactance at end-inspiration, XeI, was significantly different between the RAS-CLAD and the CLAD-free group (Bonferroni adjusted p = 0.0170, Table 3). Comparisons between BOS and RAS groups revealed significant differences in ReE at time of CLAD onset (Bonferroni adjusted p = 0.0410, Table 3).
The intrabreath impedance-volume loops (Figure 2, middle panels) and impedance-flow loops (Figure 2, right panels) are markedly different in the patients who remained CLAD-free (A) and those who had BOS (B) or RAS (C). In CLAD-free patients (A), there is minimal change in the resistance and reactance values during normal tidal inspiration and expiration. In contrast, patients with BOS (B) exhibit progressively lower resistance values during inspiration that recovers during expiration while there is a markedly steep drop in the reactance values as the breath reaches end expiration. This club shaped reactance-volume loop is characteristic of expiratory flow limitation observed in obstructive lung disease. In patients with RAS (C, middle panel), the main finding is the thin reactance-volume loop and linear decrease in the reactance values during inspiration that is followed with the same slope of increase during expiration, while the resistance-flow loop is similar to that observed in the CLAD-free state (A, middle panel).
The differences in the resistance and reactance measurements at end-inspiration and end-expiration, at a state of zero flow, is illustrated in the impedance-flow loops (Figure 2, far right panels). The reactance-flow loops are markedly different amongst the CLAD-free (A), BOS (B) and RAS (C) states, where the strong negative volume dependence of reactance resulted in low end-inspiratory reactance (a restrictive feature) and clockwise looping in the reactance-flow diagram in RAS as opposed to the counter-clockwise pattern in BOS. In contrast, the differences in the resistance-flow loops were less evident.
Discussion
This is the first study to characterize lung transplant patients at the time of CLAD onset by both standard (spectral) and intrabreath oscillometry. We observed distinct spectral oscillogram patterns among CLAD-free, BOS and RAS as well as significant differences in both spectral and intrabreath oscillometry parameters that can distinguish CLAD and its phenotypes from CLAD-free states at the time of CLAD onset. By matching the date of the oscillometry/pulmonary function test of the CLAD-free patients to within 2 weeks of the CLAD-onset date for the CLAD patients, we were able to account for the time-dependence of changing lung function post-transplant, allowing us to compare CLAD and CLAD-free patients with similar periods of follow-up time post-lung transplant.
We observed significant differences in both spectral and intrabreath oscillometry parameters depending on CLAD status. Larger differences were found in the BOS than the RAS group compared to CLAD-free. BOS and RAS were different with respect to the spectral oscillometry parameter, R5, and the intrabreath oscillometry metric, ReE. BOS is an obstructive disease that develops from progressive obliteration of the small airways and presents with a concurrent drop in the FEV1/FVC (forced vital capacity) ratio at time of CLAD-onset (Glanville et al., 2019; Meyer et al., 2014) (Table 1). This is reflected in the pattern of the resistance and reactance curves that resemble those of patients with COPD, with high R5-19 and Ax, and low X5 values (Eddy et al., 2019; Hantos, 2021). The intrabreath oscillometry in BOS-patients reveals dynamic changes in airway closure and gas-trapping during tidal breathing with significantly higher ARV’ (area of resistance-flow volume loop) and AXV’ (area of reactance-flow loop), respectively. In contrast, RAS develops as a consequence of fibroelastosis of the lung allograft (Glanville et al., 2019; Verleden et al., 2019) and presents with restrictive physiology (Table 1) and an oscillometry pattern that is similar to patients with pulmonary fibrosis (Sugiyama et al., 2013; Wu et al., 2022) that include normal resistance measurements and a right-ward shift in the reactance curve, with increased Ax and low X5. The intrabreath oscillometry measurements in RAS are notable for the markedly abnormal XeI, that reflects increased distension at end-inspiration in the context of a stiff fibrotic lung. In the pulmonary fibrosis study, XeI was found to be the single oscillometry parameter most correlated with FVC and total lung capacity (TLC) (Wu et al., 2022).
The demographic and clinical characteristics of the patients at time of transplant were similar for the CLAD-free, BOS and RAS patients. The most common manifestation of CLAD is BOS and only up to 30% of CLAD patients develop RAS (Sato et al., 2011). Our study cohort is similar to those previously reported (Sato et al., 2011), with 20% of our patients being classified with the RAS phenotype. The median follow-up time in the current study is less than 1 year.
The short follow-up is a major study limitation and the main reason for the relatively small number of CLAD cases, particularly those with RAS. Since CLAD occurs in a time-dependent manner, the number of patients with RAS and/or BOS will increase with longer follow-up and ongoing recruitment of patients. A larger RAS patient cohort will improve the generalizability of the results and the power of the study. With longer follow-up and larger cohort, the analysis of the oscillometry data may show differences that are currently undetected between RAS and CLAD-free patients.
The current methods for diagnosis of CLAD are not without limitations. Despite the published consensus guidelines used to define CLAD, the threshold drop in spirometry (currently defined as persistent ≥20% decline of at least 12 weeks in FEV1 compared to the baseline value achieved post-transplant) is subject to controversy due to the differing frequency of pulmonary function testing in various centres, although they clearly have an impact on the diagnosis and date of CLAD-onset (Glanville et al., 2019). Moreover, the extent of investigations to exclude other causes for the drop in lung function is not defined and differs amongst transplant centers (Glanville et al., 2019). Many lung transplant centers also follow patients with spirometry only, complicating the diagnosis of RAS, as restrictive defects can only be made with the additional measurements of lung volumes. This can be further exacerbated when the BOS patients transition into the mixed phenotype, where spirometry, lung volumes and chest imaging are needed to detect this development.
The diagnosis of CLAD in the absence of pulmonary function testing is not recommended, although patients can present with imaging or pathologic findings that are characteristic for CLAD but are too sick to perform spirometry, as accurate measurements require forced expiratory maneuvers that must be repeated in order to meet technical quality control standards. Oscillometry is particularly attractive in this clinical context as it is performed during normal quiet breathing and can be tolerated by anyone who can breathe while wearing a nose clip. Furthermore, the commercial oscillometry devices are small, portable, and easily deployed in a clinic setting or at the bedside. While strict quality control and assurance standards are needed for accurate measurements of oscillometry, no special expertise or infrastructure is required. As such, implementation of oscillometry for lung function assessment post-lung transplant improves accessibility for patients who cannot visit diagnostic pulmonary function laboratories due to mobility issues or who live in remote, underserviced regions of the country.
The need for non-invasive, sensitive methods for early detection CLAD has stimulated much research. Many of these offer promise but are hampered by need for technical expertise and infrastructure support and/or are invasive. They include multi-breath nitrogen wash-out to evaluate small airway function, different imaging modalities and analytic strategies (e.g. high-resolution computed tomography), novel assessments of immune cell counts, chemokines, cytokines, and circulating cell free DNA (Belloli et al., 2021; Shtraichman and Diamond, 2020; Tissot et al., 2019; Veraar et al., 2021). Some groups are also investigating computational models to estimate a recipient’s risk of CLAD using clinical and multi-omic databases (Koutsokera et al., 2017; Pison et al., 2014).
These new techniques and tools, including oscillometry, need further validation and standardization before an individual patient approach can be developed. Timely diagnosis and accurate distinction between CLAD phenotypes are important as they would allow for earlier intervention and offers the possibility of evaluating specific therapies for the early treatment of CLAD. Early diagnosis of RAS is particularly beneficial as these patients have a lower survival rate than BOS (Kulkarni et al., 2019; Todd et al., 2014).
In conclusion, both spectral and intrabreath oscillometry can differentiate between patients with CLAD and those who are remaining CLAD-free and can clearly distinguish between different CLAD phenotypes at the time of CLAD onset. Significant changes in multiple spectral and intrabreath oscillometry parameters may indicate BOS development whereas significant changes observed only in XeI can be an early biomarker to identify RAS. The frequency of lung function monitoring decreases with longer durations post-transplant. More frequent follow-up with oscillometry, regardless of duration post-transplant and evaluation of a larger patient population are needed to demonstrate the potential of oscillometry as a sensitive measure of early diagnosis of CLAD, i.e. at the time of or before the first drop in the FEV1.
Data availability statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics statement
The studies involving human participants were reviewed and approved by The Toronto General Hospital Research Institute Research Ethics Board. The patients/participants provided their written informed consent to participate in this study.
Author contributions
AF contributed to the statistical analysis plan, performed the data analysis and drafted the manuscript. AV conducted the research, compiled and verified the clinical data, helped with the data analysis and manuscript draft. NH developed the statistical plan, oversaw the data analysis, and help draft the manuscript. NB verified clinical data, adjudicated the CLAD status and edited the manuscript. JW maintained the research ethics protocol, developed the standard operating procedures and quality control of the oscillometry tests, and quality assurance of the data. SR conducted the intrabreath analysis. CR developed the standard operating procedures and quality control of pulmonary function testing in the Toronto General Pulmonary Function Laboratory and maintained the quality assurance of the data. ZH developed the intrabreath technique, oversaw the intrabreath analysis and edited the manuscript. C-WC developed the concept, study protocol and oversaw all aspects of the project.
Funding
The study is supported by a grant-in-aid from the Lung Health Foundation, the Pettit Block Term Grants, the CIHR/NSERC Collaborative Health Research Program (grant #415013) and the Ajmera Foundation Multi-Organ Transplant Innovation Fund. AF is supported by the Peterborough K. M. Hunter Charitable Foundation Graduate Award. AV was supported by a University of Toronto Fellowship and Queen Elizabeth II/Institute of Medical Science Graduate Scholarships In Science and Technology (QEII-GSST). JW is supported by Canadian Lung Association–Breathing as One: Allied Health Grant. ZH is supported by Hungarian Scientific Research Fund Grant K128701. ZH is recipient of the ERS Clinical Research Collaboration Award as a host of the intrabreath oscillometry development. Thorasys Thoracic Medical Systems Inc. (Montreal, Canada) provided the software for the 10 Hz sinusoidal frequency recordings. We thank the Registered Pulmonary Technologists at Toronto General Hospital for helping to conduct the study.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
- Bates J. H. T., Irvin C. G., Farré R., Hantos Z. (2011). Oscillation mechanics of the respiratory system. Compr. Physiol. 1, 1233–1272. 10.1002/cphy.c100058 [DOI] [PubMed] [Google Scholar]
- Belloli E. A., Gu T., Wang Y., Vummidi D., Lyu D. M., Combs M. P., et al. (2021). Radiographic graft surveillance in lung transplantation: Prognostic role of parametric response mapping. Am. J. Respir. Crit. Care Med. 204, 967–976. 10.1164/rccm.202012-4528OC [DOI] [PMC free article] [PubMed] [Google Scholar]
- Burgel P.-R., Bergeron A., de Blic J., Bonniaud P., Bourdin A., Chanez P., et al. (2013). Small airways diseases, excluding asthma and COPD: An overview. Eur. Respir. Rev. 22, 131–147. 10.1183/09059180.00001313 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chang E., Vasileva A., Nohra C., Ryan C. M., Chow C-W., Yan Wu J. (2022). Conducting respiratory oscillometry in an outpatient setting. J. Vis. Exp.. 10.3791/63243 [DOI] [PubMed] [Google Scholar]
- Chiabai J., Friedrich F. O., Fernandes M. T. C., Serpa F. S., Antunes M. O. B., Neto F. B., et al. (2021). Intrabreath oscillometry is a sensitive test for assessing disease control in adults with severe asthma. Ann. Allergy Asthma Immunol. 127, 372–377. 10.1016/j.anai.2021.06.005 [DOI] [PubMed] [Google Scholar]
- Cho E., Wu J. K. Y., Birriel D. C., Matelski J., Nadj R., DeHaas E., et al. (2020). Airway oscillometry detects spirometric-silent episodes of acute cellular rejection. Am. J. Respir. Crit. Care Med. 201, 1536–1544. 10.1164/rccm.201908-1539OC [DOI] [PubMed] [Google Scholar]
- Cosio M., Ghezzo H., Hogg J. C., Corbin R., Loveland M., Dosman J., et al. (1978). The relations between structural changes in small airways and pulmonary-function tests. N. Engl. J. Med. 298, 1277–1281. 10.1056/NEJM197806082982303 [DOI] [PubMed] [Google Scholar]
- Dellacà R. L., Santus P., Aliverti A., Stevenson N., Centanni S., Macklem P. T., et al. (2004). Detection of expiratory flow limitation in COPD using the forced oscillation technique. Eur. Respir. J. 23, 232–240. 10.1183/09031936.04.00046804 [DOI] [PubMed] [Google Scholar]
- Eddy R. L., Westcott A., Maksym G. N., Parraga G., Dandurand R. J. (2019). Oscillometry and pulmonary magnetic resonance imaging in asthma and COPD. Physiol. Rep. 7, e13955. 10.14814/phy2.13955 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fujii M., Shirai T., Mori K., Mikamo M., Shishido Y., Akita T., et al. (2015). Inspiratory resonant frequency of forced oscillation technique as a predictor of the composite physiologic index in interstitial lung disease. Respir. Physiol. Neurobiol. 207, 22–27. 10.1016/j.resp.2014.12.009 [DOI] [PubMed] [Google Scholar]
- Glanville A. R., Verleden G. M., Todd J. L., Benden C., Calabrese F., Gottlieb J., et al. (2019). Chronic lung allograft dysfunction: Definition and update of restrictive allograft syndrome―A consensus report from the Pulmonary Council of the ISHLT. J. Heart Lung Transpl. 38, 483–492. 10.1016/j.healun.2019.03.008 [DOI] [PubMed] [Google Scholar]
- Gray D. M., Czovek D., McMillan L., Turkovic L., Stadler J. A. M., Vanker A., et al. (2019). Intra-breath measures of respiratory mechanics in healthy African infants detect risk of respiratory illness in early life. Eur. Respir. J. 53, 1800998. 10.1183/13993003.00998-2018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hantos Z. (2021). Intra-breath oscillometry for assessing respiratory outcomes. Curr. Opin. Physiol. 22, 100441. 10.1016/j.cophys.2021.05.004 [DOI] [Google Scholar]
- King G. G., Bates J., Berger K. I., Calverley P., de Melo P. L., Dellaca R. L., et al. (2020). Technical standards for respiratory oscillometry. Eur. Respir. J. 55, 1900753. 10.1183/13993003.00753-2019 [DOI] [PubMed] [Google Scholar]
- Kotecha S., Paraskeva M. A., Levin K., Snell G. I. (2020). An update on chronic lung allograft dysfunction. Ann. Transl. Med. 8, 417. 10.21037/atm.2020.01.05 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kouri A., Dandurand R. J., Usmani O. S., Chow C.-W. (2021). Exploring the 175-year history of spirometry and the vital lessons it can teach us today. Eur. Respir. Rev. 30, 210081. 10.1183/16000617.0081-2021 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Koutsokera A., Royer P. J., Antonietti J. P., Fritz A., Benden C., Aubert J. D., et al. (2017). Development of a multivariate prediction model for early-onset bronchiolitis obliterans syndrome and restrictive allograft syndrome in lung transplantation. Front. Med. 4, 109. 10.3389/fmed.2017.00109 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kulkarni H. S., Cherikh W. S., Chambers D. C., Garcia V. C., Hachem R. R., Kreisel D., et al. (2019). Bronchiolitis obliterans syndrome–free survival after lung transplantation: An international society for Heart and lung transplantation thoracic transplant registry analysis. J. Heart Lung Transpl. 38, 5–16. 10.1016/j.healun.2018.09.016 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Makan G., Dandurand R. J., Gingl Z., Hantos Z. (2022). Intra-breath changes in respiratory mechanics assessed from multi-frequency oscillometry measurements. Physiol. Meas. 43, 045004. 10.1088/1361-6579/ac5bef [DOI] [PubMed] [Google Scholar]
- Meyer K. C., Raghu G., Verleden G. M., Corris P. A., Aurora P., Wilson K. C., et al. (2014). An international ISHLT/ATS/ERS clinical practice guideline: Diagnosis and management of bronchiolitis obliterans syndrome. Eur. Respir. J. 44, 1479–1503. 10.1183/09031936.00107514 [DOI] [PubMed] [Google Scholar]
- Paredi P., Goldman M., Alamen A., Ausin P., Usmani O. S., Pride N. B., et al. (2010). Comparison of inspiratory and expiratory resistance and reactance in patients with asthma and chronic obstructive pulmonary disease. Thorax 65, 263–267. 10.1136/thx.2009.120790 [DOI] [PubMed] [Google Scholar]
- Parulekar A. D., Kao C. C. (2019). Detection, classification, and management of rejection after lung transplantation. J. Thorac. Dis. 11, S1732–S1739. 10.21037/jtd.2019.03.83 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pison C., Magnan A., Botturi K., Seve M., Brouard S., Marsland B. J., et al. (2014). Prediction of chronic lung allograft dysfunction: A systems medicine challenge. Eur. Respir. J. 43, 689–693. 10.1183/09031936.00161313 [DOI] [PubMed] [Google Scholar]
- Sato M., Waddell T. K., Wagnetz U., Roberts H. C., Hwang D. M., Haroon A., et al. (2011). Restrictive allograft syndrome (RAS): A novel form of chronic lung allograft dysfunction. J. Heart Lung Transpl. 30, 735–742. 10.1016/j.healun.2011.01.712 [DOI] [PubMed] [Google Scholar]
- Shtraichman O., Diamond J. M. (2020). Emerging biomarkers in chronic lung allograft dysfunction. Expert Rev. Mol. diagn. 20, 467–475. 10.1080/14737159.2020.1738929 [DOI] [PubMed] [Google Scholar]
- Stockley J., Cooper B., Stockley R., Sapey E. (2017). Small airways disease: Time for a revisit? Int. J. Chron. Obstruct. Pulmon. Dis. 12, 2343–2353. 10.2147/COPD.S138540 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sugiyama A., Haruta Y., Nakamura I., NakagawaM., Miyamoto S., et al. (2013). Characteristics of inspiratory and expiratory reactance in interstitial lung disease. Respir. Med. 107, 875–882. 10.1016/j.rmed.2013.03.005 [DOI] [PubMed] [Google Scholar]
- Thabut G., Mal H. (2017). Outcomes after lung transplantation. J. Thorac. Dis. 9, 2684–2691. 10.21037/jtd.2017.07.85 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tissot A., Danger R., Claustre J., Magnan A., Brouard S. (2019). Early identification of chronic lung allograft dysfunction: The need of biomarkers. Front. Immunol. 10, 1681. 10.3389/fimmu.2019.01681 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Todd J. L., Jain R., Pavlisko E. N., Finlen Copeland C. A., Reynolds J. M., Snyder L. D., et al. (2014). Impact of forced vital capacity loss on survival after the onset of chronic lung allograft dysfunction. Am. J. Respir. Crit. Care Med. 189, 159–166. 10.1164/rccm.201306-1155OC [DOI] [PMC free article] [PubMed] [Google Scholar]
- Veraar C., Kliman J., Benazzo A., Oberndorfer F., Laggner M., Hacker P., et al. (2021). Potential novel biomarkers for chronic lung allograft dysfunction and azithromycin responsive allograft dysfunction. Sci. Rep. 11, 6799. 10.1038/s41598-021-85949-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Verleden G. M., Glanville A. R., Lease E. D., Fisher A. J., Calabrese F., Corris P. A., et al. (2019). Chronic lung allograft dysfunction: Definition, diagnostic criteria, and approaches to treatment―A consensus report from the Pulmonary Council of the ISHLT. J. Heart Lung Transpl. 38, 493–503. 10.1016/j.healun.2019.03.009 [DOI] [PubMed] [Google Scholar]
- Verleden G. M., Vos R., Vanaudenaerde B., Dupont L., Yserbyt J., Van Raemdonck D., et al. (2015). Current views on chronic rejection after lung transplantation. Transpl. Int. 28, 1131–1139. 10.1111/tri.12579 [DOI] [PubMed] [Google Scholar]
- Verleden S. E., de Jong P. A., Ruttens D., Vandermeulen E., van Raemdonck D. E., Verschakelen J., et al. (2014). Functional and computed tomographic evolution and survival of restrictive allograft syndrome after lung transplantation. J. Heart Lung Transpl. 33, 270–277. 10.1016/j.healun.2013.12.011 [DOI] [PubMed] [Google Scholar]
- Wu J. K., DeHaas E., Nadj R., Cheung A. B., Dandurand R. J., Hantos Z., et al. (2020). Development of quality assurance and quality control guidelines for respiratory oscillometry in clinic studies. Respir. Care 65, 1687–1693. 10.4187/respcare.07412 [DOI] [PubMed] [Google Scholar]
- Wu J. K. Y., Ma J., Nguyen L., Dehaas E. L., Vasileva A., Chang E., et al. (2022). Correlation of respiratory oscillometry with CT image analysis in a prospective cohort of idiopathic pulmonary fibrosis. BMJ Open Respir. Res. 9, e001163. 10.1136/bmjresp-2021-001163 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.