The SNAP Trial: 2-Year Results of a Double-Blind Multicenter Randomized Controlled Trial of a Silicon Nitride Versus a PEEK Cage in Patients After Lumbar Fusion Surgery

R F M R Kersten 1,2,, F C Öner 2, M P Arts 3, M Mitroiu 4, K C B Roes 5, A de Gast 1, S M van Gaalen 1,6
PMCID: PMC9609539  PMID: 33406905

Abstract

Study Design:

Randomized controlled trial.

Objectives:

Lumbar interbody fusion with cages is performed to provide vertebral stability, restore alignment, and maintain disc and foraminal height. Polyetheretherketone (PEEK) is commonly used. Silicon nitride (Si3N4) is an alternative material with good osteointegrative properties. This study was designed to assess if Si3N4 cages perform similar to PEEK.

Methods:

A non-inferiority double-blind multicenter RCT was designed. Patients presenting with chronic low-back pain with or without leg pain were included. Single- or double-level instrumented transforaminal lumbar interbody fusion (TLIF) using an oblique PEEK or Si3N4 cage was performed. The primary outcome was the Roland-Morris Disability Questionnaire (RMDQ). The non-inferiority margin for the RMDQ was 2.6 points on a scale of 24. Secondary outcomes included the Oswestry Disability Questionnaire (ODI), Visual Analogue Scales (VAS), SF-36 Physical Function, patient and surgeon Likert scores, radiographic evaluations for subsidence, segmental motion, and fusion. Follow-up was planned at 3, 6, 12, and 24-months.

Results:

Ninety-two patients were randomized (i.e. 48 to PEEK and 44 to Si3N4). Both groups showed good clinical improvements on the RMDQ scores of up to 5-8 points during follow-up. No statistically significant differences were observed in clinical and radiographic outcomes. Mean operative time and blood loss were statistically significantly higher for the Si3N4 cohort. Although not statistically significant, there was a higher incidence of complications and revisions associated with the Si3N4 cage.

Conclusions:

There was insufficient evidence to conclude that Si3N4 was non-inferior to PEEK.

Keywords: polyetheretherketone (PEEK), silicon nitride (Si3N4), lumbar spinal fusion, degenerative disc disease, randomized controlled trial

Introduction

Intervertebral fusion is one of the methods to treat chronic low back pain. Mechanical and biological factors play an important role in creating an optimal environment for bony fusion. Originally, stand-alone bone grafts were used, but they are associated with nonunion, collapse and donor side morbidity. Therefore they were succeeded by the use of interbody cages. 1 Interbody cages can be used to restore alignment and maintain disc- and foraminal height while facilitating bony fusion. PEEK has become one of the most frequently used materials with high fusion rates and good clinical results. 2 However, there are also disadvantages. PEEK's hydrophobic surface discourages direct appositional bone growth, which may lead to the formation of a fibrous layer around the implant. 3 Ti surfaces can be more osteoinductive than PEEK, but they produce artifacts on CT and MRI and are associated with an increased risk of subsidence compared to PEEK. 4 Latest developments focus on combining the 2 materials to optimize intervertebral fusion. For example, the enhancement of PEEK cages with Ti-coated endplates 5 and hydroxyapatite coated PEEK cages can improve osteointegration. 6 Still, no differences in clinical outcomes and fusion rates are reported between these materials. 7

New materials like ceramics have been introduced. Silicon nitride (Si3N4) is such a (non-oxide) ceramic with high strength and toughness. Si3N4 minimizes scatter and artifacts on CT and MRI imaging. 8 Due to its surface chemistry it allows a decreased bacterial activity compared to PEEK and Ti. 9 Si3N4 received the CE Mark and FDA market clearance for its use as an interbody cage in 2008. It's mechanical, chemical and osteoconductive qualities were extensively described in literature. 10 A recent animal study showed similar results in mechanical stability and bone formation of Si3N4 cages compared to the PEEK. 11 A RCT comparing PEEK and Si3N4 cages after anterior cervical discectomy with fusion (ACDF) reported no statistically significant differences in clinical outcome and fusion rates. 12 At the time of our study design, no clinical trial data of Si3N4 in the lumbar spine were published yet. Therefore, the Silicon Nitride And PEEK (SNAP) trial was designed to compare a PEEK cage with a Si3N4 cage in patients after lumbar fusion surgery. 13 Primary objective was to show that lumbar spinal fusion with a Si3N4 cage produces similar improvement in clinical outcome compared to a PEEK cage. This article reports the 2-year outcomes.

Materials and Methods

Study Design

One hundred patients presenting with chronic low back pain with or without leg pain were treated with either a PEEK or Si3N4 cage. The study protocol was published in detail previously. 13 In short, the study was designed as a non-inferiority multicenter clinical observer and patient blinded RCT. Inclusion criteria are listed in Table 1. Patients were randomly allocated by use of a centralized 24-hour online computerized randomization system (Sealed Envelope Ltd. London). Measurements were performed pre-operative and at 3, 6, 12 and 24 months.

Table 1.

Inclusion and Exclusion Criteria.

Inclusion criteria
  • – Male and female patients age 18-75 years

  • – Chronic low back pain unresponsive to at least 6 months of conservative care

  • – MRI and standing x-ray evidence of Pfirrmann Grade III or greater disc degeneration

  • – Degeneration and/or degenerative or isthmic spondylolisthesis of Grade I or II

  • – Signed informed consent

Exclusion criteria
  • – Osteoporosis

  • – Patients with prior failed fusion at the same level

  • – Degenerative scoliosis

  • – Degenerative spondylolisthesis greater than Grade II

  • – Pregnancy

  • – Psychiatric or mental disease

  • – Alcoholism (drinking more than 5 units per day)

  • – Active infection or prior infection at the surgical site

  • – Active cancer

  • – Insufficient language skills to complete questionnaires

  • – Participation in another study

  • – More than 2 symptomatic levels that need fusion

  • – Planned emigration abroad in the year after inclusion

Ethical Considerations

This study was performed in line with the principles of the Declaration of Helsinki. The protocol has been reviewed and approved by the local medical ethics committee (Verenigde Commissies Mensgebonden Onderzoek, as of Jan 1th 2015 known as Medical Research Ethics Committee United, Nieuwegein, the Netherlands. Approval number NL34808.100.10). Written informed consent was obtained from all individual participants included in the study. Authors were not exempt from requirement.

Surgical Procedure

Single- or double-level transforaminal lumbar interbody fusion (TLIF) with pedicle screw fixation was performed with either an oblique PEEK or Si3N4 cage (Phantom™PLIF and Valeo®OL, respectively, CTL Medical, Dallas, TX, USA) (Figure 1A ∼ B). The Si3N4 cage had a lordosis of 0° whereas the PEEK implant had 6° of lordosis. After adequate exposure and placement of pedicle screws, a facetectomy was performed followed by an appropriate decompression on the symptomatic site. The disc space was cleared from disc material and endplates were prepared. Cages were packed with autograft derived from locally harvested bone. A single oblique cage was placed in the prepared disc space. Final fixation of screws and rods was performed under compression. Patients were mobilized on the first day after surgery without bracing.

Figure 1.

Figure 1.

Lumbar intervertebral cages used in this study: (A) Valeotm OL Si3N4 cage and (B) Phantomtm PLIF PEEK cage.

Outcome Measures

Clinical assessment

The primary outcome measure was the Roland Morris Disability Questionnaire (RMDQ) (0-24 scale). 14 Secondary outcome measures included scores from the generic quality of life questionnaire SF-36, 15 Oswestry Disability Index (ODI, 0-50 scale), 16 Visual Analog Scales for leg and back pain (VAS, 0 to 100 mm) 17 and the 7-point Likert score for patient and surgeon perceived recovery in which “complete recovery” and “almost complete recovery” were considered good outcomes. 18 In addition, a neurological examination was conducted at each follow-up.

Radiological assessment

Fusion status was evaluated according to the criteria described by Burkus et al. 19 which included: (i) the presence of bridging bone on a computed tomography (CT) scan (Siemens Sensation 16, Malvern, PA, USA, 3.0 mm slice) at 12-months follow-up; (ii) disc height and angular changes in segmental alignment on lateral conventional radiographs (CR) during follow-up; and (iii) an assessment of device-host interface on a CT scan at 12-months follow-up. 19 Standing anterior-posterior (AP) and lateral CR’s were collected at 3, 6, 12, and 24-months of follow-up (Figure 2). Average disc heights were determined as the mean of the anterior and posterior measurements. Subsidence was defined as a loss of >1 mm in average disc height. At 24-months, additional flexion/extension standing lateral radiographs were obtained to monitor angular motion. Fusion was defined as an angular motion of <2° and translational motion of < 0.5 mm. Each level was analyzed separately in cases with 2-level fusion. All radiological analyses were performed by radiologists from an independent organization (Medical Metrics, Houston, TX, USA).

Figure 2.

Figure 2.

Lateral X-rays of L4-L5 fusion at 24-months for: (A) Si3N4 cage and (B) PEEK cage. Note that fusion was achieved with both cages as indicated by bone bridging between the endplates.

Statistical Analyses

Primary efficacy analysis

The primary outcome was the RMDQ score. Primary objective was to demonstrate that the Si3N4 cage was non-inferior to the PEEK cage based on the primary comparison at 12 months. The considered non-inferiority margin was 2.6 points for the difference in RMDQ between the treatment arms.13,20 The analysis was based on a mixed-effects model for repeated measurements (MMRM). No imputation of missing data was performed. The MMRM model included treatment (type of cage) and center as factors, baseline RMDQ as covariate (fixed effects), and patient as random effect. An unstructured covariance matrix was assumed to model the within-patient variance and estimation was performed by restricted maximum likelihood method. Based on the model, the result of the contrast at 12 months is expressed with point estimate for difference in mean RMDQ between the 2 cages (Si3N4-PEEK) and one-sided confidence interval with significance level of 2.5%. Non-inferiority was to be demonstrated if the upper boundary of this confidence interval does not exceed the non-inferiority margin of 2.6 points. Assuming a standard deviation of 4 points, 50 patients per arm provide 90% power to demonstrate non-inferiority. 13

Sensitivity analysis

To assess impact of drop outs, sensitivity analysis was performed. This analysis was conducted following Last Observation Carried Forward (LOCF) imputation. The analytical and estimation method for the sensitivity analysis was based on the same mixed-effects model for repeated measurements with the same terms as employed for the primary efficacy analysis (MMRM on the completed dataset).

Secondary efficacy analyses

The secondary efficacy outcomes assessed at each visit (ODI, Vas leg, VAS back, SF36 and radiological measurements) were analyzed using the same mixed-effects model for repeated measurements with the same terms as employed for the primary efficacy analysis. Dichotomous outcomes (dichotomized Likert scales for patient and surgeon perception) were compared between treatment groups based on Z-tests for comparing proportions. Statistical analyses were performed using RStudio and nlme. Plots were created using R base plotting functions and ggplot2.

Results

Baseline Characteristics

Between 2012 and 2015, 100 patients were included in 2 centers (49 and 51). Eight patients were subsequently excluded due to protocol violations (no randomization pre-operative, proof of osteoporosis after inclusion, age during surgery) or cancellation of surgery by the patient after inclusion. Of the remaining 92 patients (46 per each center), 48 were randomized for PEEK and 44 for Si3N4. Eight patients in the Si3N4 group received a 2-level fusion compared to 5 patients in the PEEK group. Baseline characteristics are shown in Table 2. At 24 months, 7 patients were lost to follow-up (7.6% drop-out rate).

Table 2.

Baseline Characteristics.

PEEK Si3N4
n 48 44
Age (mean (sd)) 53.3 (9.2) 55.4 (11.5)
Gender = Female (%) 33 (68.8) 28 (63.6)
BMI (mean(sd)) 27.1 (4.3) 27.1 (5.1)
Smoking = Yes (%) 31 (64.6) 32 (72.7)
Duration of complaints (mean (sd)) 10.6 (9.3) 8.9 (6.1)
Type of complaints, n (%)
 Radicular pain 8 (16.7) 9 (20.5)
 Combination back/radiculair 39 (81.2) 30 (68.2)
 Back pain 1 (2.1) 5 (11.4)
Clinical diagnosis, n (%)
 Degenerative disc disease 10 (20.8) 13 (29.5)
 Isthmic spondylolisthesis grade 1 12 (25.0) 11 (25.0)
 Isthmic spondylolisthesis grade 2 5 (10.4) 6 (13.6)
 Degenerative spondylolisthesis grade 1 20 (41.7) 14 (31.8)
 Degenerative spondylolisthesis grade 2 1 (2.1) 0 (0.0)
Operated levels, n (%)
 1-level: L3-L4 4 (8.5) 5 (11.4)
  L4-L5 15 (31.9) 11 (25.0)
  L5-S1 21 (44.7) 19 (43.2)
  L5-L6 1 (2.1) 0 (0.0)
  L6-S1 1 (2.1) 1 (2.3)
 2-level: L3-L5 2 (4.3) 1 (2.3)
  L4-L6 0 (0.0) 1 (2.3)
  L4-S1 3 (6.3) 5 (11.4)
  L5-L6-S1 0 (0.0) 1 (2.3)
RMDQ (mean (sd)) 14.2 (4.3) 14.8 (4.3)
ODI (mean (sd)) 23.1 (7.4) 22.5 (7.0)
VAS leg (mean (sd)) 60.9 (20.7) 58.9 (27.8)
VAS back (mean sd)) 62.3 (22.3) 61.7 (21.9)
SF-36 Physical Functioning (mean (sd)) 37.0 (19.5) 39.9 (19.4)

Perioperative Results

Peri-operative data are shown in Table 3. There were no differences in length of hospital stay between both groups. Average operative time (Si3N4 72-290 min vs PEEK 75-240 min) and blood loss (Si3N4 120-1700 ml vs PEEK 100-700 ml) was significantly higher in the Si3N4 group. There was also a slightly higher peri-operative complication rate in the Si3N4 group, although these differences were not statistically significant (Table 3).

Table 3.

Peri-Operative Characteristics.

PEEK (n = 48) Si3N4 (n = 44) P value
Operative time min (mean (sd)) 127 (46) 150 (51) 0.03*
Blood loss ml (mean (sd)) 317 (150) 473 (332) 0.01*
Hospital stay days (mean (sd)) 3.8 (2.2) 3.8 (1.6) 0.90*
Complications, n (%):
 Dural tear 1 (2.1) 4 (9.1) 0.14**
 Implant malposition 0 (0.0) 3 (6.8) 0.07**
 Sensory deficit 1 (2.1) 3 (6.8) 0.27**
 Motor deficit (MRC grade 4/5) 2 (4.2) 2 (4.2) 0.93**

*Two-sample t-test, p-value for difference 2-sided.

**Two-sample Z-test for equality of proportions, p-value for difference 2-sided.

Clinical Outcome

Both treatment arms showed good improvements in RMDQ scores during the 24-months follow-up (Figure 3). Although patients treated with PEEK had better outcomes at 3, 6, 12 and 24 months compared to Si3N4, these differences were not significant.

Figure 3.

Figure 3.

RMDQ scores during follow-up.

Using the a priori selected non-inferiority margin of 2.6, the null hypothesis that Si3N4 is non-inferior to PEEK could not be rejected. This is graphically shown in Figure 4. The upper boundary of the confidence interval exceeds the non-inferiority margin of 2.6 at each follow-up period.

Figure 4.

Figure 4.

95% CI of RMDQ difference between PEEK and Si3N4 by visit interval.

Secondary outcomes are shown in Table 4. All patients showed good improvements during follow-up. There were no significant differences in VAS leg, VAS back, SF36 and ODI scores between the 2 groups. Although both surgeons and patients reported generally better recovery rates for the PEEK group at each follow-up time point, these differences did not reach statistical significance.

Table 4.

Outcome During Follow-Up.

PEEK Si3N4 p value
Roland Morris Disability Questionnaire (sd) 0-24 scale
 3 months 9.5 (5.5) 11.0 (5.2) 0.19*
 6 months 7.0 (6.3) 9.2 (6.7) 0.16*
 12 months 5.7 (5.8) 7.9 (6.4) 0.11*
 24 months 6.1 (6.5) 8.5 (7.0) 0.20*
Oswestry Disability Questionnaire (sd) 0-50 scale
 3 months 14.8 (9.4) 15.7 (8.0) 0.47*
 6 months 10.7 (10.3) 12.3 (8.7) 0.49*
 12 months 10.2 (10.2) 9.7 (9.1) 0.17*
 24 months 11.9 (10.3) 11.2 (10.9) 0.40*
VAS leg (sd) 0-100 scale
 3 months 26.5 (27.7) 26.4 (26.2) 0.49*
 6 months 23.6 (29.2) 26.2 (28.8) 0.47*
 12 months 24.6 (28.4) 26.5 (22.9) 0.28*
 24 months 26.3 (25.3) 30.0 (31.3) 0.31*
VAS back (sd) 0-100 scale
 3 months 34.9 (18.1) 37.9 (22.3) 0.49*
 6 months 28.9 (23.3) 26.4 (25.4) 0.27*
 12 months 30.2 (21.9) 31.0 (22.9) 0.39*
 24 months 34.8 (24.7) 38.2 (25.7) 0.45*
SF36 physical functioning (sd)
 3 months 59.2 (20.6) 58.0 (17.2) 0.44*
 6 months 66.0 (23.9) 61.2 (19.7) 0.25*
 12 months 73.1 (23.1) 68.5 (21.2) 0.24*
 24 months 71.5 (24.6) 64.9 (23.0) 0.20*
Surgeon perceived Likert (%)
 3 months 73.3 58.1 0.20**
 6 months 76.2 61.0 0.20**
 12 months 78.6 61.9 0.15**
 24 months 78.0 56.3 0.08**
Patient perceived Likert (%)
 3 months 64.3 58.1 0.72**
 6 months 76.2 56.1 0.09**
 12 months 78.6 64.3 0.23**
 24 months 75.0 50.0 0.05**
Disc height (mm)
 postoperative 8.3 8.1 0.63*
 3 months 7.3 7.3 0.91*
 6 months 7.2 7.0 0.67*
 12 months 7.1 6.9 0.67*
 24 months 7.1 6.9 0.68*
Translational motion (mm) 24 months 0.12 0.14 0.70**
Angular motion (°) 24 months 0.94 1.18 0.24**

* Estimated from an MMRM model with the same specification as for the primary outcome analysis, p-value for difference one-sided.

**Two-sample t-test, p-value for difference 2-sided.

Radiological Outcomes

The radiographic data are also provided in Table 4. There were no significant differences in average disc heights between groups. Also, no significant differences in fusion rates were seen between the PEEK and Si3N4 based on the flexion/extension analysis of angular or translational motion (88% vs 82% respectively, p = 0.40). Bony bridging, measured on CT at 12 months, was seen in 42% vs 57% of patients in the PEEK and Si3N4 group respectively (p = 0.13). Sagittal and coronal views for a Si3N4 implant are shown in Figure 5 A∼B. Due to the fact that PEEK cages are radiolucent, the interface between the endplates and these cages could not be adequately ascertained. An assessment of the device-bone interface (i.e., radiolucency or osseous integration) was therefore deemed to be unreliable and could not be incorporated into the analyses.

Figure 5.

Figure 5.

CT imaging of a Si3N4 cage at 12 months, showing bone bridging in the (A) sagittal and (B) coronal views. No signs of lucency were seen at the device-bone interface.

Complications and Revisions

During 24 months follow-up there were 14 revisions (15.2% revision rate). Specifications are provided in Table 5. In the PEEK group 4 out of 48 patients (8.3%) were revised, compared to 10 out of 44 patients (22.7%) in the Si3N4 group (p = 0.10). Almost one-third of revisions were performed due to adjacent level problems (5 out of 14).

Table 5.

Revision Surgery.

Cage Index level Time Revision
PEEK L4-S1 5 months redecompression L5-S1
PEEK L5-S1 7 months redecompression L5-S1 + screw removal S1
PEEK L4-L5 10 months adjacent level L5-S1
PEEK L3-L4 14 months adjacent level L4-S1
Silicon Nitride L5-S1 1 day revision cage due to implant malposition
Silicon Nitride L5-S1 2 days revision screw L6 due to neurological disorder
Silicon Nitride L5-S1 6 months revision screw due to lose endcap
Silicon Nitride L5-S1 7 months redecompression L5-S1
Silicon Nitride L4-S1 8 months adjacent level L3-L4
Silicon Nitride L3-L4 10 months adjacent level L4-S1
Silicon Nitride L5-S1 18 months revision cage due to non-union/loosening screws
Silicon Nitride L5-S1 18 months revision cage due to loosening cage
Silicon Nitride L3-L4 19 months adjacent level L4-L5
Silicon Nitride L5-S1 20 months revision cage due to non-union.

Discussion

The SNAP trial was designed to compare the clinical and radiological outcomes for Si3N4 cages versus PEEK cages in patients undergoing lumbar fusion surgery. The overall results indicate that patients treated with either cage material had comparable outcomes with respect to disability, pain, and fusion. In particular, the RMDQ improvements observed in this trial were in line with the results from other spinal fusion studies,21,22 thereby reflecting good 2-year clinical outcomes for both groups. The secondary outcome scores were also consistent with reported literature using PEEK cages, ranging from 24 to 36 for VAS back pain, 26 to 42 for VAS leg pain 2 and 9 to 20 for ODI. 23 Lastly, the fusion results observed were also found to be similar to values reported in literature. 2

Primary Outcome

In this study, it was hypothesized that Si3N4 would be non-inferior to PEEK as measured by a non-inferiority margin of 2.6 points on RMDQ scores at 12-months follow-up. Although both implant groups had improvement scores of up to 5-8 points, there was insufficient evidence to conclude that Si3N4 was non-inferior to PEEK. As with any non-inferiority study, this does depend directly on the non-inferiority margin of 2.6 points improvement on RMDQ that was pre-determined. Our considerations are part of the protocol, 13 but other perspectives could have been taken. For example, Stratford et al 24 reported that the minimum detectable difference between pre- and post-treatments in patient with low back pain varied based on the patient’s initial RMDQ score. They concluded that clinically important changes in the RMDQ were 2 (for an initial score of 0 to 8), 4 (for an initial score of 5 to 12), 5 (for an initial score of 9 to 16), 8 (for an initial score of 13 to 20), and 8 (for an initial score of 17 to 24). Since in our study the initial RMDQ score was 14, a higher non-inferiority margin might have been chosen, although such a margin does not only depend on the minimal detectable difference at individual patient level. It does stress the importance of stratification of the patient population in assessing relevant pain scores, and should be taken into consideration for future studies.

Perioperative Outcomes

A significant difference was found in operative time and blood loss in favor of the PEEK cohort (i.e.127 min vs 150 min and 317 ml vs 473 m respectively). The greater amount of blood loss was directly linked to a longer operative time for the Si3N4 cohort. However, this result is skewed due to an outlier value of one patient in the Si3N4 group whose blood loss was 1700 ml. The difference in operative time can also be partially explained by a higher number of 2-level procedures in the Si3N4 cohort compared to PEEK (i.e., 8 versus 5). Additionally, upon rotating the Si3N4 cage during insertion, in 2 patients a fracture occurred at the insertor-cage interface. These cages needed to be replaced, extending the operative time. After thorough ana-lysis, the cause of these 2 incidents was found to be a lack of stability in the insertor-cage interface. After adjusting the tip of the insertor, which created a more stable grip while inserting the cage, no additional fractures occurred. Other perioperative complications were evenly distributed over the length of the study.

Radiological Outcomes

There is considerable controversy in the scientific literature as to when a lumbar segment is radiologically fused.19,25 Various criteria of angular and translational motions have been proposed, coupled with the presence of anterior bridging bone (i.e., the “sentinel sign”) without radiolucencies at the superior or inferior surfaces of the implant. In this study, as the PEEK cages were radiolucent, an assessment of either radiolucencies around these cages and their osseous integration was deemed unreliable and therefore unusable for this analyses. However, several other criteria were usable. First, bony bridging was measured on CT at 12 months and defined as the presence of a bony bridge from one endplate to the next. Secondly, disc height measurements were used for analyses of potential subsidence. In both groups no statistically significant differences were seen in average amount of subsidence or bony bridging. Thirdly, segmental motion measured on flexion/extension radiograms was used to analyze fusion, defining angular motion < 2° and translational motion < 0.5 mm as fusion. This study showed fusion rates consistent with results found in literature. 2 However, a technically and/or radiographically insufficient fusion does not necessarily equates an unsuccessful clinical outcome because vertebral stability may occur before it is radiographically evident. 26 This could explain there is no clear evidence that a bony fusion correlates with a good clinical outcome.

Complications

There were more revisions within the Si3N4 group compared to the PEEK group (10 vs 4), however this difference was not statistically significant (p = 0.10) (Table 5). Most revisions were performed due to adjacent level problems. Also, as described earlier, two Si3N4 cages fractured during surgery at the insertor-cage interface due to a technical problem with the insertor.

Limitations

The design of the SNAP trial had several limitations. Firstly, the use of a single oblique cage was chosen to allow for more accurate fusion measurements on CT. However, a single cage is mechanically less stable compared to 2 parallel placed cages. 27 This could have biased the results and can also explain the high revision rate of 15.2%. Secondly, as discussed earlier we can reiterate the way our non-inferiority margin of 2.6 points improvement on RMDQ was determined. Thirdly, this study was funded by Amedica Corporation (Salt Lake City, UT, USA), the manufacturer of the Si3N4 cage. Every effort was made to eliminate bias in the study design, protocol, and management of the study. Independent Clinical Research Organization (CRO) managed the study together with the principal investigator’s institution, the statistical analyses were performed by an independent organization employing their own statisticians (Julius Centre for Health Sciences and Primary Care, University Medical Center Utrecht, The Netherlands) and yet another independent unit performed the radiographic measurements (Medical Metrix). With those precautions, the authors have implemented the most reasonable procedure to minimize bias.

Conclusions

Despite the fact that both groups in our trial had good clinical improvements on the RMDQ scores during follow-up of up to 5-8 points after 24 months, there is insufficient evidence to conclude that the Si3N4 cage is non-inferior to the PEEK cage. Perioperative blood loss and surgery time were significantly higher in the Si3N4 group. Additionally, a higher incidence of complications and a higher incidence of revisions seemed to be associated with the Si3N4 cage, although not statistically significant.

Acknowledgments

Robin van Zijl, Ditte Varkevisser, Lidewij Ochtman and Anneli Duits are acknowledged for their assistance in organizing and data gathering.

Authors’ Note: The research departments of both participating centers were compensated for their time. The device(s)/drug(s) is/are FDA-approved or approved by corresponding national agency for this indication. Trial registration number: clinicaltrials.gov, Identifier: NCT01557829.

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This study was funded by Amedica Corporation, Salt Lake City, UT, USA.

ORCID iD: R. F. M. R. Kersten, MD Inline graphic https://orcid.org/0000-0002-8096-818X

References

  • 1.McAfee PC. Interbody fusion cages in reconstructive operations on the spine. J Bone Joint Surg Am. 1999;81(6):859–880. [DOI] [PubMed] [Google Scholar]
  • 2.Jiya TU, Smit T, Van Royen BJ, Mullender M. Posterior lumbar interbody fusion using non resorbable poly-ether-ether-ketone versus resorbable poly-L-lactide-co-D, L-lactide fusion devices. Clinical outcome at a minimum of 2-year follow-up. Eur Spine J. 2011;20(4):618–622. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Phan K, Hogan JA, Assem Y, Mobbs RJ. PEEK-Halo effect in interbody fusion. J Clin Neurosci. 2016;24:138–140. [DOI] [PubMed] [Google Scholar]
  • 4.Seaman S, Kerezoudis P, Bydon M, Torner JC, Hitchon PW. Titanium vs. polyetheretherketone (PEEK) interbody fusion: meta-analysis and review of the literature. J Clin Neurosci. 2017;44:23–29. [DOI] [PubMed] [Google Scholar]
  • 5.Chong E, Mobbs RJ, Pelletier MH, Walsh WR. Titanium/polyetheretherketone cages for cervical arthrodesis with degenerative and traumatic pathologies: early clinical outcomes and fusion rates. Orthop Surg. 2016;8(1):19–26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Johansson P, Barkarmo S, Hawthan M, Peruzzi N, Kjellin P, Wennerberg A. Biomechanical, histological, and computed X-ray tomographic analyses of hydroxyapatite coated PEEK implants in an extended healing model in rabbit. J Biomed Mater Res A. 2018;106(5):1440–1447. [DOI] [PubMed] [Google Scholar]
  • 7.Kersten RF, van Gaalen SM, de Gast A, Oner FC. Polyetheretherketone (PEEK) cages in cervical applications: a systematic review. Spine J. 2013;15(6):1446–1460. [DOI] [PubMed] [Google Scholar]
  • 8.Bock RM, McEntire BJ, Bal BS, Rahaman MN, Boffelli M, Pezzotti G. Surface modulation of silicon nitride ceramics for orthopaedic applications. Acta Biomater. 2015;26:318–330. [DOI] [PubMed] [Google Scholar]
  • 9.Gorth DJ, Puckett S, Ercan B, Webster TJ, Rahaman M, Bal BS. Decreased bacteria activity on Si(3)N(4) surfaces compared with PEEK or titanium. Int J Nanomedicine. 2012;7:4829–4840. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Pezzotti G, McEntire BJ, Bock R, et al. Silicon nitride: a synthetic mineral for vertebrate biology. Sci Rep. 2016;6:31717. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Kersten RFMR, Wu G, Pouran B, et al. Comparison of polyetheretherketone versus silicon nitride intervertebral spinal spacers in a caprine model. J Biomed Mater Res B Appl Biomater. 2018;107(3):680–699. [DOI] [PubMed] [Google Scholar]
  • 12.Arts MP, Wolfs JF, Corbin TP. Porous silicon nitride spacers versus PEEK cages for anterior cervical discectomy and fusion: clinical and radiological results of a single-blinded randomized controlled trial. Eur Spine J. 2017;26(9):2372–2379. [DOI] [PubMed] [Google Scholar]
  • 13.Kersten RF, van Gaalen SM, Arts MP, et al. The SNAP trial: a double blind multi-center randomized controlled trial of a silicon nitride versus a PEEK cage in transforaminal lumbar interbody fusion in patients with symptomatic degenerative lumbar disc disorders: study protocol. BMC Musculoskelet Disord. 2014;15:57–2474–15–57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Brouwer S, Kuijer W, Dijkstra PU, Goeken LN, Groothoff JW, Geertzen JH. Reliability and stability of the Roland Morris disability questionnaire: intra class correlation and limits of agreement. Disabil Rehabil. 2004;26(3):162–165. [DOI] [PubMed] [Google Scholar]
  • 15.Ware JE, Jr, Sherbourne CD. The MOS 36-item short-form health survey (SF-36). I. Conceptual framework and item selection. Med Care. 1992;30(6):473–483. [PubMed] [Google Scholar]
  • 16.van Hooff ML, Spruit M, Fairbank JC, van Limbeek J, Jacobs WC. The Oswestry Disability Index (version 2.1a): validation of a Dutch language version. Spine (Phila Pa 1976). 2015;40(2):E83–E90. [DOI] [PubMed] [Google Scholar]
  • 17.Collins SL, Moore RA, McQuay HJ. The visual analogue pain intensity scale: what is moderate pain in millimetres? Pain. 1997;72(1-2):95–97. [DOI] [PubMed] [Google Scholar]
  • 18.Bombardier C. Outcome assessments in the evaluation of treatment of spinal disorders: summary and general recommendations. Spine (Phila Pa 1976). 2000;25(24):3100–3103. [DOI] [PubMed] [Google Scholar]
  • 19.Burkus JK, Foley K, Haid RW, LeHuec JC. Surgical interbody research group—radiographic assessment of interbody fusion devices: fusion criteria for anterior lumbar interbody surgery. Neurosurg Focus. 2001;10(4):E11. [DOI] [PubMed] [Google Scholar]
  • 20.Ostelo RW, de Vet HC, Knol DL, van den Brandt PA. 24-item Roland-Morris disability questionnaire was preferred out of six functional status questionnaires for post-lumbar disc surgery. J Clin Epidemiol. 2004;57(3):268–276. [DOI] [PubMed] [Google Scholar]
  • 21.Robertson PA, Plank LD. Prospective cohort analysis of disability reduction with lumbar spinal fusion surgery in community practice. J Spinal Disord Tech. 2008;21(4):235–240. [DOI] [PubMed] [Google Scholar]
  • 22.Ohba T, Ebata S, Haro H. Comparison of serum markers for muscle damage, surgical blood loss, postoperative recovery, and surgical site pain after extreme lateral interbody fusion with percutaneous pedicle screws or traditional open posterior lumbar interbody fusion. BMC Musculoskelet Disord. 2017;18(1):415–017–1775–y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Yu CH, Wang CT, Chen PQ. Instrumented posterior lumbar interbody fusion in adult spondylolisthesis. Clin Orthop Relat Res. 2008;466(2):3034–3043. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Stratford PW, Binkley JM, Riddle DL, Guyatt GH. Sensitivity to change of the Roland-Morris back pain questionnaire: part 1. Phys Ther. 1998;78(1):1186–1196. [DOI] [PubMed] [Google Scholar]
  • 25.Goldstein C, Drew B. When is a spine fused? Injury. 2011;42(3):306–313. [DOI] [PubMed] [Google Scholar]
  • 26.Kroeze RJ, van der Veen AJ, van Royen BJ, Bank RA, Helder MN, Smit TH. Relation between radiological assessment and biomechanical stability of lumbar interbody fusion in a large animal model. Eur Spine J. 2013;22(12):2731–2739. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Landham PR, Don AS, Robertson PA. Do position and size matter? An analysis of cage and placement variables for optimum lordosis in PLIF reconstruction. Eur Spine J. 2017;26(11):2843–2850. [DOI] [PubMed] [Google Scholar]