Skip to main content
Micromachines logoLink to Micromachines
. 2022 Sep 28;13(10):1622. doi: 10.3390/mi13101622

Characterization of a Droplet Containing the Clustered Magnetic Beads Manipulation by Magnetically Actuated Chips

Sheng-Huang Yen 1, Pei-Chieh Chin 1, Jun-Yu Hsu 1, Jr-Lung Lin 1,*
Editor: Eui-Hyeok Yang1
PMCID: PMC9610654  PMID: 36295975

Abstract

A magnetically actuated chip was successfully developed in this study to perform the purpose of transportation for a droplet containing clustered magnetic beads. The magnetic field gradient is generated by the chip of the two-layer 4 × 4 array micro-coils, which was commercially fabricated by printing circuit board (PCB) technology. A numerical model was first established to investigate the magnetic field and thermal field for such a micro-coil. Consequently, the numerical simulations were in reasonable agreement with the experimental results. Moreover, a theoretical analysis was derived to predict the dynamic behaviors of the droplets. This analysis will offer the optimal operation for such a magnetically actuated chip. This study aims to successfully implement the concept of “digital microfluidics” in “point-of-care testing” (POCT). In the future, the micro-coil chip will be of substantial benefit to genetic analysis and infectious disease detection.

Keywords: magnetically actuated, droplet, point-of-care testing, digital microfluidics

1. Introduction

In past decades, micro-electro-mechanical systems technologies have contributed to the wide and rapid development of microfluidic devices with many different functions. The functions of transporting, mixing, separating, detecting, and sensing are exhibited in micro-total analysis systems (µTAS). In implementing the aforementioned functions, transporting the sample and reagent plays an imperative role, which is required in the applications of medicine, biology, and the environment. Two microfluidic technologies, namely continuous flow or droplet-driven schemes, have successfully demonstrated miniaturized μTAS for transportation [1]. The aforementioned approaches are promising, but major challenges still exist in real-world biosamples with microfluidic devices and their reliance on equipment for actuation [2]. In such tasks, droplet-based platforms were used for the storage and processing of samples and reagents demonstrating significant potential without the aid of micropumps [1]. Droplet-actuated microfluidic devices have made substantial progress toward integrating essential manipulation for bio-analysis.

In a continuous flow scheme, transport is facilitated by the devices of micropumps. Briefly, micropumps can be divided into either non-mechanical or mechanical depending on whether the elements are fixed or movable [3]. Non-mechanical pumps, which have fixed elements, must import external energies to convert momentum to facilitate fluid flow. Converted methods include electro-hydrodynamic, magneto-hydrodynamic, electro-osmotic, electrowetting, bubble type, and electrochemical approaches [3]. Conversely, mechanical pumps must integrate movable elements (i.e., actuators) into a body chamber. To drive the fluid flow, external energies must also be imported into the actuators. The driven modes of actuators include electrical, electrostatic, pneumatic, piezoelectric, and magnetic types [3]. However, valves must be used in mechanical micropumps to restrict the flow direction [4]. These valves can generally be categorized as check valves [5], flapping valves [6,7], or peristaltic valves [8,9]. In achieving valveless pumping, mechanical micropumps have been successfully developed with simple geometries [10]. However, these valves with mechanical micropumps can damage and clog the cells or particles in the process of fluid flow [11]. Additionally, these valves continuously consume reagents and biosamples during transportation. Furthermore, these valves are constantly applied under such a high pressure; thus, the valve materials should overcome fatigue damage for long-term operation [12].

Droplet-based microfluidics, also named digital microfluidics, have recently gradually gained a considerable amount of attention. One major advantage of droplet-based microfluidic devices lies in their manipulation of droplets on an open surface. In addition, droplets are not only self-contained systems but also function as reaction chambers and transportation units. The droplet-actuated platforms are particularly useful for point-of-care testing because of their simplicity, portability, and ability to store samples and reagents on the chip. Many actuated methods have been successfully proposed to manipulate the movement of the droplets, including electrowetting on dielectric [13], thermocapillary force [14], surface acoustic wave [15], electrophoresis [16], optical force [17], and magnetic force [18,19,20,21,22]. Among all the actuation methods, magnetic-actuated droplet manipulation [18,19,20,21,22] demonstrates some advantages because of its flexibility, long distance, large driving force, and easy operation.

Magnetic beads not only serve as an actuated medium but are also carriers for biomolecules; for example, super-paramagnetic particles are suitable for nucleic acid binding and transfer [2]. Magnetic beads are also attractive from a practical perspective because they can be used to capture proteins or nucleic acids [23]. The extraction from crude biosamples uses various commercially available reagents and surface-coated particles to perform on-chip DNA extraction and purification [18,19,20,22]. Permanent magnets [19,20] or micro-coils [18,21,22] have been employed to transport [19,20,22], split [18,20,22], and merge droplets [19] for genetic assay. Nonetheless, the numerical simulation remains inadequate for analyzing the magnetic and thermal fields of such planar coils. If numerical simulation can be utilized to analyze the distribution of magnetic and temperature fields for the planar micro-coils, then this simulation can provide an optimal design for these micro-coils. Therefore, the time and cost of trial-and-error methods can be significantly reduced.

Therefore, a magnetic-actuated chip for transporting droplets that utilizes two-layer 4 × 4 array micro-coils was designed in this study by fabricating the PCB technology. Numerical simulation was used to investigate the magnetic and thermal fields for a micro-coil. Moreover, theoretical analysis was derived to predict the moving speed of a droplet considering the parameters, including the loading mass of magnetic beads, droplet volumes, and applied DC currents. Finally, fully automated manipulation of a droplet on an open surface was demonstrated for conducting sample transportation.

2. Theoretical Analysis

In a previous study, Long et al. [19] derived a theoretical analysis to categorize an “operating diagram” that describes the following three droplet regions: stable motion, breakage, and magnet release. They have derived various formulas to analyze the three different regions. A theoretical model is developed in this study to characterize the dynamic behaviors of the droplet containing clustered magnetic beads. The kinematics of the droplet is determined by the Newtonian law acting on the magnetic force (FB), Stoke force (FSt), and the surface tension force (Fγ). Magnetic force is expressed as follows [24]:

FB=Vbχ(B)Bμo (1)

where Vb is the volume of magnetic beads, χ is the magnetic susceptibility of the beads, μo is the permeability of free space (4π × 10−7 H/m), and B is the magnetic flux density (i.e., magnetic field).

In principle, the droplet is driven to move by the x-axis magnetic force (FB,x), which is simplified by [24]:

FB,x=mb χ BmBz,Maxμoρblx (2)

where mb and ρb are the mass and density of the clustered magnetic beads, respectively; lx is the axial distance from the center to the rim of a micro-coil; and Bm is the z-axial magnetic field generated by the Helmholtz coil. Notably, the maximum z-direction magnetic field, that is, Bz,Max, is induced by the micro-coil applied in a given DC current.

The motion of a droplet surrounded by silicon oil follows Stoke’s rule. The Stoke’s force (FSt) is obtained as follows:

FSt=3πηOdDuD (3)

where ηO is medium (oil-phase) viscosity, dD is the diameter of the droplet, and uD is the moving speed of a droplet. The frictional force (Fμ) between a liquid droplet and a substrate surface can be expressed as the following:

Fμ=μkΔρVDg (4)

where μk is the friction coefficient of the substrate surface. Δρ is the difference density between liquid droplets and the surrounding medium. g is the gravitational acceleration. Upon moving onto a substrate, a droplet’s surface tension is expressed in the following manner:

Fγ=πdDγOWcosθ (5)

where θ is the contact angle. γO-W is the relative surface tension between liquid water and oil medium. In order to present the ratio between frictional force and surface tension force, the coefficient of ζ is introduced to define as follows:

ζ=FμFγ=μkΔρVDgπdDγOWcosθ (6)

By simplifying Equation (6), the coefficient of ζ is obtained as follows (see Supplementary Document S1):

ζ=(2(1cosθ)3π2sin3(2θ))1/3μkΔρ(VD)2/3gγOW (7)

A droplet moving speed (uD) can be determined by the force balance between the magnetic force, frictional force, and Stoke’s drag force, as shown below:

ρDVDduDdt=3πηOdDuD+mbχBmBz,MaxμolxμkΔρVDg (8)

When a droplet moves, its shape is assumed to remain unchanged. The surface tension of a droplet will not affect the droplet’s motion since it is an internal force. Additionally, when the droplet moves, determining the frictional force can be a challenge. Therefore, the surface tension force is used instead. Equation (8) can be rewritten as follows:

ρDVDduDdt=3πηOdDuD+mbχBmBz,MaxμolxζπdDγOWcosθ (9)

The moving speed of the droplet (uD) can be solved as follows:

uD(t)=(mbχBmBz,MaxρDVDρbμOlx6ζγOWcosθρDdD2)(1et/τs) (10)

where τS is Stoke’s time, which is defined as τS=ρdD3/18ηO. As t approaches infinity or is substantially larger than Stoke’s time and UD is called the terminal velocity. UD can be obtained by:

UD=(mbρDVDρbχBmBz,MaxμOlx(36cosθπVD2)1/3ζγOWρD) (11)

2.1. Design and Fabrication

A droplet containing the clustered magnetic beads manipulated by the two-layer micro-coils is schematically shown in Figure 1. The Helmholtz coil is used to produce a uniform transversal magnetic field (Bm). The two-layer micro-coils are utilized to generate the magnetic field gradient (▽B). The magnetic force is linearly proportional to Bm (▽B), which actuates the movement of the droplet. Initially, the droplet containing the clustered magnetic beads is fixed at the center of micro-coil #1 while coil #1 is turned on. Afterward, coil #1 is turned off while coil #2 is turned on. The droplet is then driven to move at the center of coil #2. The magnetic field is produced by the sequential turning on/off of the micro-coil at a constantly driving frequency. Thus, the droplet is driven to achieve transportation performance.

Figure 1.

Figure 1

Schematic of the working principle of the two-layer micro-coil chip magnetically actuated droplet movement. (a) Fixed at the center of coil #1 while coil #1 is turned on. (b) Moved to the center of coil #2 while coil #2 is turned on and coil #1 is turned off.

The photograph of the magnetic-actuated chip is displayed in Figure 2a. The proposed chip of the two-layer 4 × 4 array micro-coils was printed by PCB technology as shown in Figure 2b. Each layer is separated into 100 μm thickness. Micro-coils of the upper and lower layers partially overlap with a distance of 3.0 mm between them. Micro-coils consist of eight windings in a square shape. These micro-coils are designed to be 150 μm in width and gap. In order to generate hydrophobic properties, Teflon coatings are applied. The magnetic-actuated chip was subsequently dip-coated in a solution of 1% w/w Teflon AF 1600 (DuPont Corp., Wilmington, DE, USA) mixed with FC-40 solvent (3M Company, St. Paul City, MN, USA) and was then baked overnight at 80 °C.

Figure 2.

Figure 2

(a) Photograph of the magnetic-actuated chip. (b) Two-layer 4 × 4 array micro-coils of the PCB layout on the upper and bottom layers.

2.2. Experiments

The experimental setup can be seen in Figure 3. The Helmholtz coil is placed on the top and bottom of the magnetically actuated chip with a given gap. This generates a uniform transversal magnetic field measuring approximately 50.0 mT at a 50 × 50 mm2 area. The magnetically actuated chip was placed on a platform for observation of the droplet’s kinetics. A power supply was employed to apply the DC currents on the two-layer 4 × 4 array micro-coil chip. A droplet was then dropped on the top surface of the chip. Afterward, 2.88 μm magnetic particles (MF-DEX-3000, MagQu LLC, Surprise City, AZ, USA) were dispensed into the water droplet. The magnetic beads are formed by the core of a single crystal Fe3O4 metallic sphere coated with dextran. The induced magnitude and direction of the magnetic fields were determined by adjusting the driving DC currents, while the direction of the magnetic fields produced by each coil was controlled by a custom-designed analog circuit. Images of the droplet movement were captured using a digital camera attached to a charge-coupled device (CCD, Cool SNAP HQ2, Photometrics, Huntington Beach, CA, USA). Customized software edited in LabVIEW (National Instruments, Austin, TX, USA) was utilized to control the magnitude, duration, and direction of DC currents sequentially. Notably, a thermoelectric cooler integrated with a K-type thermocouple was equipped underneath the PCB as a cooling system. The cooling temperature was controlled by utilizing a commercial PID controller to dissipate the effects of Joule heating.

Figure 3.

Figure 3

Experimental setup of the magnetically actuated platform. a: digital camera attached with a CCD; b: Helmholtz coil; c: magnetically actuated chip.

3. Results and Discussion

3.1. Magnetic Characterizations

A numerical simulation was performed in this study to investigate a planar micro-coil considering magnetic and thermal fields. These fields were numerically simulated using commercial software (CFD-ACE+, CFD-RC, Huntsville, AL, USA). The modules of thermal, electric, and magnetic fields were selected to study the magnetic and thermal phenomena [25]. Heat generation and DC currents were employed on an 8-winding coil. The heat transfer of the proposed system is mainly governed by the mechanisms of heat conduction coupling with free convection. In this simulation, a 3D geometric domain was discretely created with structured hexahedral meshes. A residual criterion of 10−8 was employed to guarantee the convergence of each iterative solution step in the simulated process. The physical properties of the materials, i.e., copper, Teflon, and air, are listed in Table 1.

Table 1.

Properties of materials.

Property Copper Teflon Air
density (kgm−3) 8960 940 1.23
electric conductivity (Sm−1) 5.996 × 107 1.0 × 10−4 10−9
thermal conductivity (Wm−1K−1) 386 0.3 0.025
specific heat capacity (Jkg−1K−1) 385 1500 1006
convection coefficient (Wm−2) none none 7.5
relative permeability 1 1 1

Figure 4a shows the simulated geometries of two micro-coils. The dimensions of an 8-winding micro-coil are listed in Table 2. Two different directions of DCs were separately applied to each of the coil centers. The 2D and 3D contours of the magnetic field were numerically demonstrated as shown in Figure 4b,c, respectively. Considering the application of a DC of 1.0 A in different directions, the 2D or 3D profiles of the z-axis magnetic field were symmetrical to the wave-shaped surfaces. The maximum value was numerically obtained as 2.84 mT at a DC of 1.0 A.

Figure 4.

Figure 4

Figure 4

(a) Numerical geometry of two micro-coils. (b) 2D and (c) 3D numerical contours of the z-axial magnetic field at an applied electric current of 1.0 A.

Table 2.

Micro-coil geometry.

Geometry Value
width (μm) 150
height (μm) 35
gap (μm) 150
winding 8
length (mm) 17.4

The numerical simulation results of different DC currents revealed that the 2D and 3D contours of the magnetic field were remarkably similar but demonstrated differences in the maximum values. Therefore, dimensionless parameters were introduced and analyzed in this study. Herein, the dimensionless parameters are defined as follows: B*Z = BZ/BZ,Max for the z-axial magnetic field, F*B,X = FB,X/FB,X,Max for the x-axial magnetic force, and x* = 2x/W for the x-axial distance. BZ,Max is the z-axial maximum magnetic field, W is the width of a micro-coil, and FB,X,Max is the x-axial maximum magnetic force. The profile of B*z varied along with x* as shown in Figure 5a. The maximum value is located near the center but not at the initial point of a coil. The value of B*Z gradually decreased with increasing x*. The zigzag-shaped profile was observed at the upper layer of the coil. However, the profile demonstrated smoothness at the lower layer of the coil. The difference in B*z at the lower layer is approximately 0.834 times that of the upper layer. The profiles of F*B,X were numerically varied along with the x* for the two different layers, as shown in Figure 6b. However, the profiles for different layers are substantially different. Notably, F*B,X,Max at the lower layer is approximately 0.3 times that of the upper layer.

Figure 5.

Figure 5

Numerical profiles of the dimensionless (a) B*Z and (b) F*B,X varied along with x*.

Figure 6.

Figure 6

(a) Numerical and experimental results of BZ,Max and (b) calculations of FB,X,Max versus different DC currents.

The numerical simulations of the Bz,Max were then compared with the experimental measurements using a Tesla meter (TM-401, KANETEC, Tokyo, Japan) versus different applied currents, as shown in Figure 6a. The simulation and experiment results of Bz,Max quasi-linearly varied with the applied DC currents for the planar coil. Additionally, the numerical simulations are in reasonable agreement with the experimental measurements. The deviation between the numerical calculations and experimental measurements was less than 18.5%. Similarly, the FB,x,Max was then calculated as displayed in Figure 6b. The FB,x,Max results also presented linear proportions with the different DC currents.

3.2. Thermal Characterizations

The temperature effect is a crucial issue for magnetic actuation. Elevated temperature cannot be immediately reduced, thus decreasing the efficiency of the magnetic manipulation or damaging biosamples. The 3D steady-state heat conduction equation can be expressed as:

(kT)+q˙=0 (12)

where k is the thermal conductivity and q˙ is the heat generation density. The heat generation density (q)˙ due to Joule heating is obtained as follows:

q˙=Q˙VC=I2RVC=LC×I2σVC(AC) (13)

where I, σ, AC and VC are electric current, electric conductivity, cross-sectional area, and occupied volume of a micro-coil, respectively. The surrounding surfaces were set under an adiabatic condition. The top and bottom surfaces were dissipated by heat convection are expressed as follows:

kT=h(TST) (14)

where h is the heat convection coefficient. TS and T represent the surface and circumstance temperature, respectively. In particular, the free convection coefficient (h) is a function of surface geometry, fluid motion, and fluid properties. Therefore, the value of h was set to 7.5 Wm−2K−1 [26] to evaluate heat generation coupled with free convection [26]. The 2D and 3D numerical contours of temperature distribution are shown in Figure 7a,b, respectively. The maximum temperature reached is 68.9 °C. Thus, the heat source generated by the 8-winding micro-coil dissipated and uniformly distributed on the chip. A dimensionless temperature is defined as T* = T/TMax. The profile of the dimensionless temperature is also shown in Figure 8a. The temperature displayed a zigzag shape due to the windings of a micro-coil. However, these differences in the temperature between the 8-winding of a coil and the gaps are substantially small.

Figure 7.

Figure 7

(a) 2D and (b) 3D numerical contours of the temperature field at an applied electric current of 1.0 A.

Figure 8.

Figure 8

(a) Numerical profiles of T* varied along with the dimensionless distance of x*. (b) Numerical and experimental comparison of maximum temperature (TMax) versus different applied currents.

The temperature of the planar coils was experimentally measured by a top-attached K-type thermocouple. The numerical simulation and measurement comparison of the maximum temperature against the applied current are shown in Figure 8b. The figure reveals that the maximum temperatures are quadratically increased with the rise in applied currents. The numerical simulation could achieve consistency with the experimental measurements. Furthermore, the deviation between the numerical simulations and experimental measurements was less than 5.9%. This finding proved that the numerical model of the temperature field can be applied to optimal designs (such as micro-coils) in the future.

3.3. Transportation Characterizations

A droplet containing the clustered magnetic beads, which is located at the center of an upper-layer micro-coil as indicated in Figure 9a. Afterward, a constant electric current was applied at the lower layer of a micro-coil to generate the magnetic field. The clustered magnetic beads were then induced to move the front rim of a droplet as shown in Figure 9b, thus facilitating the movement of the droplet to the center of a micro-coil (Figure 9c). The moving time is experimentally measured in the movement process. The motion distance, that is, half a width of the micro-coil (lx = 3.0 mm), was then divided by the moving time to convert the moving speed of a droplet.

Figure 9.

Figure 9

A series of photographs depicting a droplet actuating from (a) intial postion, (b) movement, to (c) final destination.

Equation (11) is used to predict the moving speed of a droplet containing clustered magnetic beads. The parameters included the loading mass of the clustered magnetic beads, droplet volumes, and applied DC currents. The electric properties of materials are listed in Table 3. Figure 10 shows the moving speed of a droplet considering the mass of magnetic beads at a constant droplet volume of 12 μL and an applied current of 0.4 A. This figure indicates that the loading mass of magnetic beads must exceed “the threshold mass” to actuate the droplets for the given droplet volume. In principle, the “loading mass” is too small (i.e., mb < 60 μg) to actuate the droplet movement. Moreover, the droplet moving speed linearly increased with the rising loading mass of magnetic beads for the theoretical analysis and experimental data.

Table 3.

Properties of materials.

Property Magnetic Bead [27] Silicon Oil [28] Water
density (kgm−3) 1500 970 1000
viscosity (Pa-s) none 0.0149 0.001
magnetic susceptibility 0.235 none none
surface tension (Nm−1) none 0.021 0.072

Figure 10.

Figure 10

Moving speed of a droplet versus the loading mass of the magnetic beads, where ζ = 0.001 was chosen in this case.

The mass of the clustered magnetic beads is maintained at mb = 300 μg with the application of two different DC currents and the current magnitudes are 0.4 and 1.0 A. The volumes of a droplet were pipetted to be 3.0, 6.0, 9.0, 12.0, and 15.0 μL. The experimental results revealed that the movement speeds are inversely proportional to the droplet volumes for a constant DC. As expected, a small droplet volume and high DC can achieve a fast moving speed. On the contrary, a large droplet volume and low DC could lead to a slow movement speed. The theoretical analysis of moving speed, that is, the formula of Equation (11), was also compared with the experimental measurements as shown in Figure 11. The prediction results are in reasonable agreement with the measurements at a low DC of 0.4 A. However, the prediction results are substantially higher than the experiments at a high DC of 1.0 A. This phenomenon is due to the thermal effect, which was dissipated at a high electric current. Therefore, the theoretical analysis of a droplet moving speed can be used for predictions at a DC of 0.4 A while the thermal dissipation is neglected.

Figure 11.

Figure 11

Theoretical and experimental comparisons of a droplet at different volumes versus moving speeds.

Finally, a liquid droplet was tested to facilitate transportation on an open surface by the proposed micro-coil chip. To reduce movement resistance, silicon oil was uniformly distributed on the PCB surface. The operating condition is as follows: mb = 300 ug, VD = 12.0 uL, and I = 1.0 A. Every adjacent micro-coil turns on/off by switching a frequency of 0.5 s. Figure 12 shows a series of photographs to perform the transportation of a droplet. A droplet containing the clustered magnetic beads was smoothly and continuously transported by a two-layer 4 × 4 array micro-coil chip. These results verified that the digital microfluidic chip could be used for transportation applications. In the current prototype, magnetically actuated chips are capable of transporting droplets either vertically or horizontally. The proposed system, however, does not implement curved trajectory functions. The sequence of a full cycle is displayed in Supplementary Video S1.

Figure 12.

Figure 12

Figure 12

Water droplet moves along the two-layer 4 × 4 array micro-coils chip at different times. (a) t = 0; (b) t = 3.0 s; (c) t = 3.5 s; (d) t = 6.5 s; (e) t = 7.0 s; (f) t = 10.0 s; (g) t = 10.5 s; (h) t = 13.5 s; (i) t = 14.0 s; (j) t = 17.0 s; (k) t = 17.5 s; and (l) t = 20.5 s. The white arrows indicate the motion direction.

4. Conclusions

The major contribution of the droplet-based manipulation lies in its flexibility because this magnetic actuation can simultaneously handle a diverse range of biosamples and reagents. It is also easily integrated with thermal control and optical detection. Overall, the z-axial maximum magnetic fields linearly vary with the DC currents with the function of 2.86 × I. The temperature differences also vary quadratically proportional to 45.5 × I2. The numerical simulations were compared with the experimental measurements of magnetic and temperature fields, with a deviation of less than 18.5% and 5.9%, respectively. Thus, the numerical model could be a powerful tool for the optimal design of the micro-coils. Meanwhile, the proposed theoretical analysis for the dynamics of the droplet could be used to predict the motion behavior. The developed platform is expected to deliver nucleic-acid-based diagnostic assays to point-of-care testing (POCT).

Acknowledgments

The authors would like to thank Chiou, Chi-Han for his assistance in completing the experiments.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/mi13101622/s1, Document S1: Deriving procedure of coefficient ζ; Video S1: Magnetically actuated manipulation of a droplet.

Author Contributions

S.-H.Y. and J.-L.L. conceived and designed the experiments; J.-Y.H. performed the experiments; P.-C.C. and J.-L.L. conducted a numerical simulation; P.-C.C. and J.-L.L. analyzed the numerical and experimental data; S.-H.Y. edited actuation software; J.-L.L. wrote the paper. All authors have read and agreed to the published version of the manuscript.

Data Availability Statement

The data presented in this study are available within the article. The other data that support the findings of this study are available upon request from the corresponding author.

Conflicts of Interest

The authors declare no conflict of interest.

Funding Statement

The authors would like to thank financial support from I-Shou University under the grant number ISU 110-MCRP-06.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Park S., Zhang Y., Lin S., Wang T.-H., Yang S. Advances in microfluidic PCR for point-of-care infectious disease diagnostics. Biotechnol. Adv. 2011;29:830–839. doi: 10.1016/j.biotechadv.2011.06.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Zhang Y., Park S., Yang S., Wang T.-H. An all-in-one microfluidic device for parallel DNA extraction and gene analysis. Biomed. Microdevices. 2010;12:1043–1049. doi: 10.1007/s10544-010-9458-6. [DOI] [PubMed] [Google Scholar]
  • 3.Wang Y.-N., Fu L.-M. Micropumps and biomedical applications—A review. Microelectron. Eng. 2018;195:121–138. doi: 10.1016/j.mee.2018.04.008. [DOI] [Google Scholar]
  • 4.Qian J.-Y., Hou C.-W., Li X.-J., Jin Z.-J. Actuation mechanism of microvalves: A review. Micromachines. 2020;11:172. doi: 10.3390/mi11020172. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Zengerle R., Ulrich J., Kluge S., Richter M., Richter A. A bi-directional silicon micropump. Sens. Actuators A. 1995;50:81–86. doi: 10.1016/0924-4247(96)80088-4. [DOI] [Google Scholar]
  • 6.Nguyen T.T., Goo N.S., Nguye V.K., Yoo Y., Park S. Design, fabrication, and experimental characterization of a flap valve IPMC micropump with a flexibly supported diaphragm. Sens. Actuators A. 2008;141:640–648. doi: 10.1016/j.sna.2007.09.017. [DOI] [Google Scholar]
  • 7.Bui G.T., Wang J.-H., Lin J.L. Optimization of micropump performance utilizing a single membrane with an active check valve. Micromachines. 2018;9:1. doi: 10.3390/mi9010001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Huang C.-W., Huang S.-B., Lee G.-B. Pneumatic micropumps with serially connected actuation chambers. J. Micromech. Microeng. 2006;16:2265–2272. doi: 10.1088/0960-1317/16/11/003. [DOI] [Google Scholar]
  • 9.Chiou C.-H., Yeh T.-Y., Lin J.-L. Deformation analysis of a pneumatically-activated polydimethylsiloxane (PDMS) membrane and potential micro-pump applications. Micromachines. 2015;6:216–229. doi: 10.3390/mi6020216. [DOI] [Google Scholar]
  • 10.Lin J.-L. Design and fabrication of pneumatically actuated valveless pumps. Micromachines. 2022;13:16. doi: 10.3390/mi13010016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.van den Berg A., Lammerink T.S.J. Micro total analysis systems: Microfluidic aspects, integration concept and applications. Microsyst. Technol. Chem. Life Sci. 1998;194:21–49. [Google Scholar]
  • 12.Gietzelt T., Jacobi O., Piotter V., Ruprecht R., Hausselt J. Development of a micro annular gear pump by micro powder injection molding. J. Mater. Sci. 2004;39:2113–2119. doi: 10.1023/B:JMSC.0000017774.64153.d9. [DOI] [Google Scholar]
  • 13.Cho S.K., Moon H., Kim C.J. Creating, transporting, cutting and merging liquid droplets by electrowetting-based actuation for digital microfluidic circuits. J. Microelectromech. Syst. 2003;12:70–80. [Google Scholar]
  • 14.Darhuber A.A., Valentino J.P. Thermocapillary actuation of droplets on chemically patterned surfaces by programmable microheater arrays. J. Microelectromech. Syst. 2003;12:873–879. doi: 10.1109/JMEMS.2003.820267. [DOI] [Google Scholar]
  • 15.Wixforth A. Flat fluidics: Acoustically driven planar microfluidic devices for biological and chemical applications. Transducers. 2005;1:143–146. [Google Scholar]
  • 16.Schwartz J.A., Vykoukal J.V., Gascoyne R.C. Droplet-based chemistry on a programmable micro-chip. Lab Chip. 2004;4:11–17. doi: 10.1039/b310285h. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Ichimura K., Oh S.K., Nakagawa M. Light-driven motion of liquids on a photo responsive surface. Science. 2000;288:1624–1626. doi: 10.1126/science.288.5471.1624. [DOI] [PubMed] [Google Scholar]
  • 18.Lehmann U., Vandevyver C., Parashar V.K., Gijs M.A.M. Droplet-based DNA purification in a magnetic lab-on-a-chip. Angew. Chem. Int. Ed. 2006;45:3062–3067. doi: 10.1002/anie.200503624. [DOI] [PubMed] [Google Scholar]
  • 19.Long Z., Shetty A.M., Solomon M.J., Larson R.G. Fundamentals of magnet-actuated droplet manipulation on an open hydrophobic surface. Lab Chip. 2009;9:1567–1575. doi: 10.1039/b819818g. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Zhang Y., Park S., Liu K., Tsuan J., Yang S., Wang T.-H. A surface topography assisted droplet manipulation platform for biomarker detection and pathogen identification. Lab Chip. 2011;11:398–406. doi: 10.1039/C0LC00296H. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Nguyen N.-T., Ng K.M., Huang X. Manipulation of ferrofluid droplets using planar coils. Appl. Phys. Lett. 2006;89:052509. doi: 10.1063/1.2335403. [DOI] [Google Scholar]
  • 22.Chiou C.-H., Jin Shin D., Zhang Y., Wang T.-H. Topography-assisted electromagnetic platform for blood-to-PCR in a droplet. Biosens. Bioelectron. 2013;50:91–99. doi: 10.1016/j.bios.2013.06.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Li Z., Zu X., Du Z., Hu Z. Research on magnetic bead motion characteristics based on magnetic beads present technology. Sci. Rep. 2021;11:19995. doi: 10.1038/s41598-021-99331-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Beyzavi X.A., Nguyen N.-T. Modeling and optimization of planar microcoils. J. Micromech. Microeng. 2008;18:095018. doi: 10.1088/0960-1317/18/9/095018. [DOI] [Google Scholar]
  • 25.Lien K.-Y., Lin J.-L., Liu C.-Y., Lei H.-Y., Lee G.-B. Purification and enrichment of virus samples utilizing magnetic beads on a microfluidic system. Lab Chip. 2007;7:868–875. doi: 10.1039/b700516d. [DOI] [PubMed] [Google Scholar]
  • 26.Lin J.-L., Wang S.-S., Wu M.-H., Oh-Yang C.-C. Development of an integrated microfluidic perfusion cell culture system for real-time microscopic observation of biological cells. Sensors. 2011;11:8395–8411. doi: 10.3390/s110908395. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Grob D.T., Wise N., Oduwole O., Sheard S. Magnetic susceptibility characterization of superparamagnetic microspheres. J. Magn. Magn. Mater. 2018;452:134–140. doi: 10.1016/j.jmmm.2017.12.007. [DOI] [Google Scholar]
  • 28.Jiao Z., Huang X., Nguyen N.-T., Abgrall P. Thermocapillary actuation of droplet in a planar microchannel. Microfluid. Nanofluid. 2007;5:205–214. doi: 10.1007/s10404-007-0235-7. [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

The data presented in this study are available within the article. The other data that support the findings of this study are available upon request from the corresponding author.


Articles from Micromachines are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES