Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2022 Oct 28.
Published in final edited form as: Phys Med. 2021 Aug 5;89:129–139. doi: 10.1016/j.ejmp.2021.07.028

Voxel-based partial volume correction of PET images via subtle MRI guided non-local means regularization

Yuanyuan Gao a,b,*, Yansong Zhu b,c,d, Murat Bilgel e, Saeed Ashrafinia b,c, Lijun Lu a, Arman Rahmim b,c,d,*
PMCID: PMC9614701  NIHMSID: NIHMS1843457  PMID: 34365117

Abstract

Purpose:

Positron emission tomography (PET) images tend to be significantly degraded by the partial volume effect (PVE) resulting from the limited spatial resolution of the reconstructed images. Our purpose is to propose a partial volume correction (PVC) method to tackle this issue.

Methods:

In the present work, we explore a voxel-based PVC method under the least squares framework (LS) employing anatomical non-local means (NLMA) regularization. The well-known non-local means (NLM) filter utilizes the high degree of information redundancy that typically exists in images, and is typically used to directly reduce image noise by replacing each voxel intensity with a weighted average of its non-local neighbors. Here we explore NLM as a regularization term within iterative-deconvolution model to perform PVC. Further, an anatomical-guided version of NLM was proposed that incorporates MRI information into NLM to improve resolution and suppress image noise. The proposed approach makes subtle usage of the accompanying MRI information to define a more appropriate search space within the prior model. To optimize the regularized LS objective function, we used the Gauss-Seidel (GS) algorithm with the one-step-late (OSL) technique.

Results:

After the import of NLMA, the visual and quality results are all improved. With a visual check, we notice that NLMA reduce the noise compared to other PVC methods. This is also validated in bias-noise curve compared to non-MRI-guided PVC framework. We can see NLMA gives better bias-noise trade-off compared to other PVC methods.

Conclusions:

Our efforts were evaluated in the base of amyloid brain PET imaging using the BrainWeb phantom and in vivo human data. We also compared our method with other PVC methods. Overall, we demonstrated the value of introducing subtle MRI-guidance in the regularization process, the proposed NLMA method resulting in promising visual as well as quantitative performance improvements.

Keywords: PET, Partial Volume Correction, NLMA, Non-Local Means, Anatomical guidance

1. Introduction

Positron emission tomography (PET) imaging is a highly sensitive molecular imaging tool that is able to quantify functional processes in vivo. However, due to a number of resolution degrading factors, the so-called partial volume effect (PVE) occurs in PET images [1,2], resulting in underestimation of radiotracer uptake in oncological lesions or plaque aggregation measures. Moreover, statistical noise in the data further impacts the quality of PET images, and can be amplified in partial volume correction (PVC) methods [3,4].

A range of PVC algorithms have been explored to improve the spatial resolution of PET images. One general category involves reconstruction-based methods, and is referred to as point-spread-function (PSF) modeling (or resolution modeling). This has been used to improve the quality in PET images by modeling resolution-degrading phenomena within the system matrix of the reconstruction algorithms, including object-domain (e.g. positron range) and detection-domain (e.g. inter-crystal blurring) resolution degrading factors [5]. PSF modeling has attracted considerable interest in PET over the past decade, and has been adopted by major PET vendors in their state-of-the-art PET scanners. At the same time, PSF modeling requires access to the original listmode or sinogram data, which is not easily achievable, especially when exploring application of PVC methods to longitudinal data (including possibly from multiple centers and scanners) obtained over many years [6].

The other category of PVC algorithms involves post-reconstruction processing, as applied to reconstructed images (whether or not they have been generated with PSF modeling). This category itself consists of region-based vs. voxel-based methods. The most commonly used region-based method is the geometric transfer matrix (GTM) method, which performs PVC correction on mean regional uptake values [7]. The GTM approach assumes that the activity within a region is uniform and can be accurately described by its mean value. Du et al. [8] introduced a more refined so-called perturbation-based GTM method for more accurate PVC estimates. Sattarivand et al. [3] also investigated a symmetric GTM (sGTM) method (equivalent to Labbe’s method [4] but avoiding handling of excessively large matrices), showing similar accuracy as GTM but improved precision. In these methods, volumes of interest (VOIs) can be propagated from the Hammers atlas [9], or other sources, and their performance depends on precise segmentation results.

Voxel-based post-reconstruction methods include the Muller-Gartner (MG) method [10], region-based voxel-wise correction (RBV) which is an extension of the GTM method to the voxel domain [11], and the iterative Yang (IY) technique [12]. At the same time, these methods all make strong assumptions about uniformity of PET uptake within MRI regions. This is not the case for the Van Cittert (VC) and Richardson–Lucy (RL) iterative deconvolution techniques that do not use MRI information at all [13]. For instance, VC method or its more statistically sound version ‘reblurred’ VC (RVC) only requires system point spread function (PSF) information and has the advantage of easy implementation. However, it does not use any regularization, and simple application of RVC method will amplify noise levels [13]. As such, regularization may have to be considered.

The well-known non-local means (NLM) method, introduced by Buades et al. in 2005 [14], employs the high degree of redundant information that typically contains in images and reduces image noise by replacing each pixel intensity with a weighted average of its nonlocal neighbors. The use of NLM filter for medical imaging has shown promising results, yet the search window is normally defined as a fixed sub-dimension of the image to constrain the search space [1517]. Ideally, the entire image should be explored as searching window to enhance the performance of the NLM filter, which comes at a high computation cost. Motivated by the success of the NLM filter, a number of NLM based regularization models have been used for image restoration and reconstruction; e.g. in computed tomography (CT) [1821]. NLM can also be used as a denoising method [17,22,23]. A number of methods combine deblurring and denoising steps by solving an optimization problem (inverse problem) [24,25]. Commonly used least square (LS) objective function only accounts for white Gaussian noise, which oversimplifies the noise property in reconstructed PET image [26,27]. Minimizing the LS cost function usually leads to higher noise results with increasing iterations, similar to the maximum likelihood (ML) approach. A penalty is desired for a penalized ML or MAP solution. In the present work, we specifically focus on regularization utilizing the nonlocal means (NLM) method [14], in order to enable PVC while suppressing image noise amplification. This NLM framework employs a high degree of information redundancy (similar structures) that typically exists in images and reduces image noise by replacing each pixel intensity with a weighted average of its nonlocal neighbors. Chan et al. [28] proposed a NLM filter method with anatomical prior, their results suggest promise in clinical practice for whole-body PET/CT imaging.

A philosophy we pursue in our work is that the ‘subtle’ usage of MRI information in PET reconstruction, as opposed to intense or no usage of MRI information. As additional motivation, clinical PET/MRI systems have been designed and produced in the past decade. Several correction methods based on anatomic images have been developed in the past, although these are not being used routinely in clinical practice [29,30]. In the present work, we use the anatomy (MRI information) to restrict the search window region without further intervention from MRI. Our method introduces a voxel-based PVC method under the least squares framework (LS) employing anatomical non-local means (NLMA) for PET image processing, and considers the Multivariate Gaussian distribution through the data fidelity term and the choice of the regularization parameter. We evaluated our proposed method with NLM as well as RVC and RBV methods. The paper is organized as follows. Sec. 2 introduces the mathematical model and the optimization algorithm. The experimental design as applied to recovery of amyloid PET images is described in Sec. 3, along with the results, which are discussed in Sec. 4. Discussion and conclusion are presented in Sec. 5. Parts of this work were presented at the Society of Nuclear Medicine and Molecular imaging (SNMMI) 2018 Annual Meeting [31] and the 2018 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS-MIC) [32].

2. Methods

2.1. Mathematical model

2.1.1. LS image processing framework

Our aim is to effectively tackle PVE in PET images as deteriorated by a known point spread function (PSF). A simple model for the measured PET image is [33]:

g=Hf+η (1)

where g is the measured image, f stands for the true image, Hf denotes the result of the convolution operation with the system PSF, and η stands for the noise. The task is to estimate f given g and H. This problem is, in general, ill-posed. In order to effectively reduce noise in the recovered images, a common method is to add a regularization term. Image after PVC can be obtained by minimizing the following regularize function.

f^=argminfgHf22+βR(f) (2)

where R(f) is the regularization term which uses the MRI information, β is a regularization parameter that balances the strength of regularization and deconvolution, gHf22 is the fidelity term which controls the degree of agreement between the estimated and the measured image, ‖X2 = XTX, XT denotes transpose of X. In this work, we set regularization term R(f) as NLM/NLMA functions, which we will describe in detail in the following section.

2.1.2. NLM-based regularization term

The NLM algorithm was originally proposed as a non-iterative edge-preserving filter to denoise natural images corrupted by additive white Gaussian noise [26]. Essentially, it is one of the neighbourhood filters, which denoises each pixel with a weighted average of its neighbouring pixels according to similarity (though the neighbourhoods tend to be large; i.e. non-local). Different from other neighbourhood filters, NLM filter calculates the similarity based on patches instead of pixels.

Motivated by the success of the NLM filter, a number of NLM based regularization models have been used for image restoration and reconstruction in computed tomography (CT) [1821]. It can also be used as a denoising method [17,22,23]. (With specific testing for amyloid PET images), where image recovery (based on knowledge of point-spread function model) is coupled with an NLM-based regularization term to control noise amplification. In order to perform PVC in PET imaging, we bring NLM within our PVC efforts. The NLM prior (regularization) term is given as:

Rf=i(fikeSWiwik(f)fk) (3)

where ∅ is a positive potential function and here we choose ∅(f) = f2/2, f represents the PET image, and SWi denotes a search window with ith voxel as the centre. The weighting coefficient wik(f) is determined by how similar local patches around the two voxels (i and k) being compared in the PET image as:

wik(f)=exp(P(fi)P(fk)2,a2/σ2)kSWiexp(P(fi)P(fk)2,a2/σ2) (4)

where P(fm) is the image patch centered at the mth voxel, and 2,a2 computes the squared Euclidean distance weighted by a Gaussian kernel with a > 0 denoting the standard deviation of the Gaussian kernel and the parameter σ acts as a degree of filtering to control the decay of the exponential function. In conventional NLM applications, NLM is commonly used as denoising filter. In the present work, this term is presented as regularization, coupled with the resolution recovery model incorporating the PSF of the system, transferring the problem from noise reduction, to PVC with noise suppression.

2.1.3. Incorporating MRI anatomical information into NLM regularization term (NLMA)

In the past, MR images have been taken as priors due to the excellent soft-tissue contrast, and anatomy-guided image processing methods have been used to improve PET image quality. However, these methods often suffer from the potential risk of over-usage of MRI information (e. g., assuming PET images are relatively uniform within anatomical regions, which is a very strong assumption [29,3438]). By contrast, in our work we seek subtle usage of MRI information without uniformity assumption on PET activity. Specifically, in order to incorporate MRI information into the NLM regularization term, we propose to utilize anatomical region masks to merely restrict the search window, thus restricting usage of non-local information to voxels that belong to the same tissue. As shown in Fig. 1, SW is the search window, fi, fj, fk are from the PET image, while Ai, Aj, Akare from the MRI segmentation regions (eg. GM, WM, Caudate, Putamen, etc.). When pixels i and j belong to the same tissueTk, that is AiTkandAjTk, we use the tissue area to replace the search window. This method is called ‘NLMA’, where the additional ‘A’ denotes anatomical guidance. This also improves the searching efficiency. The regularization function of NLMA is:

Rf=i(fikεSWiwik(f)fkχTi(Ak)) (5)

where χTi is the indicator function of tissue Ti and is defined as:

χTi(Ak)={1, if AkTi0, if AkTi (6)
Fig. 1.

Fig. 1.

Schematic representation of the NLMA method, (A) is the PET image, (B) is the segmented MRI image.

Ti denotes the segmented region which the ith voxel belongs to, and SWi denotes a search window with ith voxel being the centre

2.2. Optimization

Direct optimization of the resulting objective function can be complicated. In practice, an empirical one-step-late implementation is usually employed in iterative approaches to obtain solutions [20,34,39]. In this work, the weighting coefficients are computed on the current image estimate, and are considered constants when the image is updated. Although this strategy inevitably increases the computational load, it may produce a more accurate solution. In our study, we applied the Gauss-Seidel (GS) algorithm combined with the one-step-late scheme to minimize the objective function (2). Following iterative OSEM or MLEM reconstruction, the variance of the image has been shown approximately equal to the activity distribution value of the PET measured image [4042].

The GS method is a special case of (projected) successive over relaxation (+SOR), which is derived via Taylor expansion and, similar to Newton methods, updates each image parameter individually by minimizing the objective function (2) over that parameter while holding other parameters fixed [1821]. According to the OSL strategy, the NLMA weighting coefficients (wik) are always computed on a current image estimate and then assumed to be constant when updating the image. The updating expression in GS without regularization is given by:

fin+1=hiT(gHf)hiThi+fin (7)

where hi is ith column of H. The updating expression for GS with NLMA regularization is:

fin+1=hiT(gHf)+(hiTH)fin+βkSWiwikfknhiThi+βkSWiwik (8)

where β is hyper parameter to control the regularization strength in the method, we set it according to the practical experiment (ranging from 0.04 to 2) and kSWiwikfkn is calculated from the NLMA method.

3. Experimental design

3.1. Phantom simulations

We performed simulation studies using the BrainWeb phantom [43], which provides T1-weighted MR image as shown in Fig. 2(A). Fig. 2(B) shows the corresponding segmented tissue map containing 10 different tissue regions. The anatomical model used to generate simulated brain MRI data consist of a set of 3-dimensional “fuzzy” tissue membership volumes. In addition, a discrete anatomical model is provided which consists of a class label (integer) at each voxel, representing the tissue which contributes the most to that voxel [20,43]. We can use this phantom to create PET activity and attenuation images. The simulated PET model consists of a set of non-uniform activity tissues including the volumes (GM, WM, CSF and others). The tissue volumes are defined at a 1 mm isotropic voxel grid in Talairach space [44], with dimensions of 181 × 181 × 217. Realistic PET activities were assigned into different regions based on 11C-PIB patient data: 12500 Bq ml−1 in GM, 3125 Bq ml−1 in WM, 0 Bq ml−1 in air, CSF and bone, 1000 Bq ml−1 in all other tissues including skin, muscle, connective tissue and fat [45]. Furthermore, non-uniform activity was assigned in GM region. To do this, we convolved grey matter with a Gaussian function with a full-width at half-maximum (FWHM) of 10 mm to create non-uniform distribution in GM. The resulting PET phantom is shown in Fig. 2(C). We realize this method may still simplify realistic scenarios [46]. However, in this manuscript, we use such Gaussian convolution method as a simple way to generate non-uniform activity distribution.

Fig. 2.

Fig. 2.

(A) MRI image, (B) Segmented MRI image (10 tissues map), (C) Ground truth PET image, (D) Attenuation map, (E) Reconstructed PET image, and (F) Segmented MRI image (18 tissues map).

Different degradation factors, including attenuation, detector blurring, and normalization effect, were considered in the projection domain in the context of our analytic simulations [4749]. To build the attenuation map, the tissues were assigned the same attenuation coefficients as in PET-SORTEO [13], which is 0 cm−1 for air, 0.146 cm−1 for bone and 0.096 cm−1 for other tissues to build the attenuation map, as shown in Fig. 2(D) [50]. The normalization coefficients were obtained using a component-based method [51]. The total number of counts was adjusted to generate a medium noise (109 number of coincidence events) in the projection domain. 20 noise realizations were computed in the simulation to evaluate quantitative performance. Image reconstruction was performed using the maximum likelihood expectation maximum (MLEM) algorithm with 240 iterations. Fig. 2(E) shows the reconstructed PET image.

To simulate more delicate segmentation guidance, we need to get another different segmentation guidance map. We used Freesurfer software to segment the MRI image into 18 tissues [52]. The generated tissue map contains many small regions such as caudate, putamen, frontal lobe, temporal lobe, hippocampus, and amygdala, as shown in Fig. 2(F). With this phantom, we are able to evaluate performance of different PVC methods in more regions.

We evaluated our proposed method (NLM and NLMA) against RVC and RBV methods. For implementation of NLM and NLMA, we set the overall search window size to 21 × 21 × 21 voxels and patch size to 7 × 7 × 7 voxels. Different regularization parameters β were chosen which we will specify later. The step size in RVC method was set as 0.5.

Among the four different methods, NLMA and RBV require additional anatomical guidance. In order to evaluate these two methods with imperfect MR guidance that may occur in practice, we simulated both registration mismatch and segmentation mismatch. Different extents of registration mismatches were simulated by shifting anatomical images. For registration mismatch, we generated misregistration MRI images from 1.0 mm to 4.0 mm offsets (along the × axis). To simulate segmentation mismatch, we also used Freesurfer to generate elaborate tissue map with noisy MRI image (the noise level is 9%) which is shown in Fig. 3(A). The original noise-free MRI image we used in the PET simulation phantom is shown in Fig. 3(B). In Fig. 3(C) and (D), we also showed the segmentation maps of the corresponding two MRI images. Red arrow points out an example where difference exists between the two segmentation maps.

Fig. 3.

Fig. 3.

(A) 9% noise MRI phantom, (B) noise-free MRI phantom, (C) MRI segmentation map of (A) from Freesurfer, (D) MRI segmentation map of (B) from Freesurfer.

The simulation and experiment works were implemented in Matlab 2017b on a computing system with a 3.30 GHz Intel Core i5-4590 processor. It takes about 15 min for one noise realization of NLMA method and about 10 min for one noise realization of NLM method.

3.2. Quantitative measurements

To quantitatively evaluate the images resulted from different algorithms, we computed noise, namely coefficient of variability (Cov), as well as bias in different regions. Cov measures image variability over different noise realizations for the same region, and is defined as:

Cov=1Mi=1M1R1r=1R(virv¯i)2v¯×100% (9)

where M is the number of voxels in the same region, R is the number of noise realization, v¯i=1Rr=1Rvir is the mean over different noise realizations, vir is the value of voxel i in rth noise realization, and v¯ is the regional mean value of the ground truth activity image.

Biases in activity uptake quantitation is defined in terms of deviation from the ground truth, calculated as follows:

Bias=1Rr=1R(v¯rv¯)v¯×100% (10)

where v¯r is the regional mean value of the rth noise realization.

3.3. Clinical patient data

Our clinical data consists of one 85-year-old cognitively normal amyloid negative male and one 85-year-old cognitively normal amyloid positive female collected from Baltimore Longitudinal Study of Aging (BLSA). Two scans were performed for each participant with two different tracers. In the first scan, participants were injected with 370 MBq 15O-H2O and 60 s PET scans were performed once the counts reached threshold levels. In the second scan, 555 MBq 11C-PIB was injected PET scans were performed over 70 min. We use the average of motion-corrected time frames corresponding to 50–70 min post PIB injection in this study. These two scans were performed on a GE Advance scanner [45]. Filtered back projection (FBP) algorithm was used for reconstruction to generate 3D images of 128 × 128 × 35 voxels with voxel size of 2mm × 2mm × 4.25mm. The FWHM of PSF was set as 7.5mm × 7.5mm × 7.5mm for GE Advance scanner. Anatomical information was also provided by performing a magnetization-prepared rapid gradient echo (MPRAGE) scan for each participant on a 3 T Philips Achieva scanner (repetition time = 6.8 ms, echo time = 3.2 ms, flip angle=8, image size=256 × 256 × 170, voxel size = 1 × 1 × 1.2mm3). The MPRAGE scans were inhomogeneity-corrected, skull-stripped, co-registered with PET images, and anatomically-labelled using the MUSE algorithm [53]. We assessed our NLMA method, together with RBV, RVC and NLM methods, using these four scans. For NLM and NLMA methods, search window was set as 9 × 9 × 9 and patch size was 5 × 5 × 5. Different regularization parameters were applied for these two methods.

4. Results

4.1. Results for simulation experiments

We evaluated the performance of four different methods, including NLM, RVC, RBV and NLMA. We first show the visual results of these four different methods when perfect MR guidance is provided. In this case, we set the β parameter for NLM as 1.5, and 0.5 for NLMA. For RVC, we set the number of iterations as 30. The visual results are shown in Fig. 4. We can see that NLMA has the lowest noise and depicts more structural details compared with other methods.

Fig. 4.

Fig. 4.

Results for different PVC methods. (A) Ground truth PET image and (B) Uncorrected PET images, followed by PET images that are PVC-corrected using (C) RVC, (D) RBV, (E) NLM, and (F) NLMA. All images use the same colour scale.

Mean percentage bias and coefficient of variability (Cov) are computed for quantitative comparisons. For NLM and NLMA, bias-noise curves were obtained by changing regularization strength β. Specifically, the scale of the β parameter ranging from 0.04to 2. Fig. 5 plots quantitative performance of these four methods with perfect MR guidance. Among different β values, the best bias-noise balance was obtained whenβ is close to 1.2. For RVC, the curve was plotted by ranging the iteration number from 1 to 30. Overall, we see that our proposed NLMA method outperformed other methods and largely improves quantitative performance, compared to non-MRI-guided NLM framework (e.g. by ~50% reduced bias when matching noise).

Fig. 5.

Fig. 5.

Bias vs. noise trade-off curves using different algorithms, for GM, WM, Caudate, Putamen, Frontal lobe, Temporal lobe, Hippocampus and Amygdala tissues. Each curve is generated by varying the hyper parameter β.

Fig. 6 (for GM, WM, Caudate and Putamen) shows quantitative performance of different PVC methods when registration mismatches occur. It is seen that with increasing mis-coregistrations, the quantitative performances degrade, as expected; Still, with small errors, NLMA method outperforms others, especial in the small areas. For putamen even to 4 mm errors, the performance of NLMA is better than NLM method.

Fig. 6.

Fig. 6.

Bias - noise trade-off curves for (1st row) 0 mm, (2nd row) 1 mm, (3rd row) 2 mm and (4th row) 4 mm MR-PET registration errors in GM (1st column), WM (2nd column), Caudate (3rd column) and Putamen (4th column) regions.

As summary of above, Fig. 4 and Fig. 5 depict visual and quantitative results of different PVC methods with perfect MR guidance (no errors in registration and segmentation). With perfect segmentation and no registration mismatch, the two segmentation-based methods, NLMA and RBV show better performance compared to NLM method. RVC method has less bias compared with NLM method. Fig. 6 shows bias-noise curves for RBV, NLM and NLMA by varying co-registration accuracies. Bias-noise trade-off performance assessment is prevalent and valuable in quantitative image analysis. The closer the curve is to the origin, the better its performance. These results demonstrate that NLMA is less influenced by registration error except in putamen. In this region, when the error reaches 4 mm, the performance of NLMA is worse than that of NLM method. We observed faster degradation for RBV compared to other methods. As registration mismatch increased, all MR-guided PVC methods degraded in performance. Within 4 mm registration error is match, NLMA shows stronger resilience to segmentation and registration mismatches.

Fig. 7 plots bias-noise trade-off curves in GM, WM, caudate, putamen, frontal lobe, temporal lobe, hippocampus and amygdala tissues for RBV and the proposed NLMA method, with perfect and mis-segmented MR guidance. We also put the NLM result and input image (measured image) in the plot as the baseline. It is seen that segmentation mismatch can degrade performance of both methods. At the same time, after PVC method, the quality of the images outperforms original non-PVC image in the mismatch segmentation MR guidance. In some tissues, the NLMA still performs better than NLM in both bias-noise. In some tissues, the NLMA still shows better bias.

Fig. 7.

Fig. 7.

Bias vs. noise trade-off curves of NLMA method, RBV using perfect MRI guidance (NLMA, RBV) and with mis-segmentation MRI guidance (NLMA-M, RBV-M), in the cases of original PET image (Input) and NLM showed results (NLM) as comparation. Results shown are for GM, WM, Caudate, Putamen, Frontal Lobe, Temporal Lobe, Hippocampus and Amygdala tissues.

4.2. Results for clinical dataset

Fig. 8 shows the anatomical clinical images of two MRI scans, which were used as anatomical reference in the anatomy-based methods in Fig. 9. The left image is the Subject0′s and the right is the Subject1′s. Standard uptake value ratio (SUVR) is a very important semi-quantitative indicator commonly in PET tumour diagnosis. It refers to the radioactivity of the imaging agent. In this paper, we obtained it by dividing the PET image by the mean value of the reference region. The reference region of 15O-H2O is the whole brain, and the 11C-PIB is the Cerebellar GM. In Fig. 9, we show transverse slices of SUVR for 15O-H2O and 11C-PIB clinical data for different PVC methods (the reference region for 15O is the whole brain and for 11C is cerebellar GM). Fig. 10 shows the NLM and NLMA results with different regularization strength (β = 0.1,0.2,0.5,0.8,1.2) for 15O-H2O clinical data and Fig. 11 shows the similar results for 11C-PIB clinical data.

Fig. 8.

Fig. 8.

Anatomical clinical reference images for Subject0 (left) and Subject1 (right) MRI scan. (a) is the transverse, (b) is the coronal and (c) is the sagittal.

Fig. 9.

Fig. 9.

PVC results for clinical participants with Subject0 (left) and Subject1 (right) 15O-H2O and 11C-PIB PET scans.

Fig. 10.

Fig. 10.

Five different β parameters including 0.1, 0.2, 0.5, 0.8 and 1.2 from left to right for NLM and NLMA results of clinical participants with amyloid negative (left) and amyloid positive (right) 15O-H2O PET scan.

Fig. 11.

Fig. 11.

Five different β parameters including 0.1, 0.2, 0.5, 0.8 and 1.2 from left to right for NLM and NLMA results of clinical participants with amyloid negative (left) and amyloid positive (right) 11C-PIB PET scan.

We then plotted SUVR in different ROIs. Fig. 12 plots ROI SUVR from different PVC methods for amyloid negative (left) and amyloid positive (middle) participants. We choose β parameters as 0.1 and 0.8 for NLM and NLMA methods, respectively.

Fig. 12.

Fig. 12.

SUVR in different ROIs for different PVC methods for amyloid negative (left), amyloid positive (right) participants, for 11C-PIB (1st row) and 15O-H2O (2nd row).

5. Discussion

In this study, we proposed a PVC approach based on subtle MRI information inducing nonlocal regularization under the LS framework in order to recover PET activity loss due to PVE and suppress image noise. Specifically, the presented NLMA method utilizes the geometrical information from the MRI data to improve the measured PET data. The method was evaluated with the digital PET/MRI phantom and in vivo human data. The results presented in Sec. 4 demonstrated that the gains from the presented NLMA PVC algorithm are notable compared with those from the NLM and RBV PVCs in terms of quantitative measurements. From the RBV results, we can see that the noise is much higher than NLMA results. We did experiments using different tissues like 10, 12, 18 tissues, and found that as the number of tissues increasing the results of NLMA will be better.

We also compared different MR-based PVC methods in the context of varying MR information mismatches including mis-segmentation and misregistration. The results showed that our proposed PVC methods with subtle MR guidance, demonstrate stronger resilience to mismatches between PET and MR images compared to methods with stronger assumptions invoked in MR-guided PVC. As registration mismatch increased, all MR-guided PVC methods degraded in performance. We observed faster degradation for RBV relative to other methods. With 4 mm registration mismatch, NLMA shows stronger resilience to registration mismatches. We also experimented with larger (>4 mm) registration mismatches (not shown here), and differences between different methods with subtle MR guidance were no longer obvious, but they all outperformed RBV. We compared MRI-guided PVC methods with both perfect and mis-segmented tissue maps. As shown in Fig. 7, with imperfect segmentation, performances of NLMA and RBV degraded. While RBV with segmentation error was inferior compared to NLMA, which still outperformed other methods. We can see NLMA with mis-segmentation case has lower noise and lower bias level in these eight regions. Under the same noise level, NLMA has the lowest bias among these methods. This shows that the proposed methods are less affected by mis-segmentation compared with other methods. In all cases, NLMA with perfect segmentation always provided best performance, though we noticed good segmentation was still important to guarantee its performance.

For clinical data, we notice our NLMA method generates lower noise compared to RVC, RBV and NLM methods for these two tracers, as shown in Fig. 9. Similar as we observed in simulation experiments, RVC gives improved contrast but also higher noise. If we compare RBV and NLMA, they both keep structure detail of images but NLMA is able to further reduce image noise. Between these two tracers, we observed our method shows better structure detail for 11C-PIB tracer. The beta value affects the results of NLM and NLMA. From the Fig. 10 and Fig. 11, we can see that with the increase of beta value (from 0.1 to 1.2), the results of NLM method are getting worse, but the results of NLMA are getting better. Best NLMA results are obtained near 0.5, which coincides with the conclusion of phantom experiment. In Fig. 9, we choose β for NLM method as 0.1 in 15O-H2O and 0.2 in 11C-PIB. We choose β as 0.5 for both 15O-H2O and 11C-PIB images of NLMA method. To ensure that both methods have the best visual effect. In Fig. 12, we use 0.1 and 0.8 beta parameters for NLM and NLMA to draw these two graphs. ROI SUVR (including Caudate, Putamen, Frontal GM, WM, Frontal lobe, and Temporal lobe) shown in Fig. 12 are all better than no-PVC method. For Amygdala, ROI SUVRs of NLMA method is quite close to those from the uncorrected image for 11C-PIB, while in other ROIs, NLMA method is better than uncorrected image and most of other methods. For WM, Frontal lobe, and Temporal lobe SUVR values are quite close to the uncorrected image for 15O-H2O. In other ROIs, NLMA method is better than uncorrected image. From Fig. 12, it can be observed that RBV method performs better in some ROIs, e.g. Caudate and Frontal GM, compared to other PVC methods.

A number of existing MRI-based PET reconstruction or PVC methods have employed more MRI information in the processing stage [29]. Such approaches risk significantly biasing PET images. As such, our framework attempts to make ‘subtle’ usage of MRI information, meaning that MRI information is only used to define the NLMA search window range. Our approach shows promise, as demonstrated in above results. At the same time, ongoing efforts include application to real clinical longitudinal studies to assess impact on assessment of disease. The process using anatomical images to guide functional image formation typically requires registration of the images from different image modalities. The recent advent of integrated PET/MRI systems apparently brings more opportunities for this task and other forms of synergistic analysis by obtaining anatomical and functional images simultaneously, allowing easier spatial alignment with minimal error.

Some existing image-based PVC methods make significant simplifying assumptions. The geometric transfer matrix (GTM) method, a gold-reference PVC method, only produces quantitative values at the region-of-interest level and does not produce PVC images. In our proposed work, we do assume activity distribution is non-uniform in PET. In the present work, by contrast, we explore a voxel-based PVC framework which performs subtle usage of MRI information, and moreover generates corrected images, while controlling for noise amplification. GTM and RBV make assumptions of uniformity of PET values within anatomical segments of MRI, whereas our proposed methods have the different assumption, aiming to quantify image details more accurately.

There are ways in which the performance of the presented NLMA algorithm could be improved. One of these ways, as we have discussed previous in the paper, is to optimize the algorithm parameters, namely the search-window, the patch-window, the control parameter and the hyper-parameter. In some other studies [54,55], the size of search-window and patch-window that are set in a reasonable range do not have noticeable effect on the resultant PET images quality, and the control parameter is set according to the scalar parameter that is determined by experiments. In addition, the determination of hyper-parameter is tuned manually to achieve a good trade-off between noise and resolution in all the cases. It should be noted that the question, what a “good trade-off” is, depends strongly on the specific tasks at hand, and the preference of the observer. Thus, optimization of parameters would be an interesting topic in our future research efforts.

6. Conclusion

In this work, we have demonstrated that subtle MR guidance under the LS framework can result in improved PET images. Our method specifically involves anatomically guided NLM (NLMA) regularization. Visually, NLMA is superior compared to other investigated methods, and quantitatively, it provides improved bias-noise trade-off performance. We also demonstrated that NLMA is not significantly affected by mis-segmentation and mis-registration. The proposed method depicts stronger resilience to mismatches between PET and MR images compared to other methods. In patient data, visual observations indicated that NLMA method had lower noise relative to other algorithms. From the quantitative results, we observed that PVC methods resulted in higher SUVR values compared to performing no PVC.

Acknowledgements

This work was supported by the Natural Science Foundation of Guangdong Province (Grant No. 2018A030313366), and the National Institutes of Health (R21 Grant No. AG056142). The authors appreciate helpful discussions with Drs. Dean Wong and Susan Resnick.

References

  • [1].Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48:932–45. 10.2967/jnumed.106.035774. [DOI] [PubMed] [Google Scholar]
  • [2].Carson RE, Daube-Witherspoon ME, Herscovitch P. Quantitative Functional Brain Imaging with Positron Emission Tomography 2013. 10.1007/0-306-48551-6_2. [DOI] [Google Scholar]
  • [3].Sattarivand M, Kusano M, Poon I, Caldwell C. Symmetric geometric transfer matrix partial volume correction for PET imaging: principle, validation and robustness. Phys Med Biol 2012;57:7101. 10.1118/1.4740193. [DOI] [PubMed] [Google Scholar]
  • [4].Labbé C, Koepp M, Ashburner J, Spinks T, Richardson M, Duncan J, et al. Absolute PET quantification with correction for partial volume effects within cerebral structures. Quantitative functional brain imaging with positron emission tomography. Elsevier; 1998. p. 59–66. 10.1016/B978-012161340-2/50011-1. [DOI] [Google Scholar]
  • [5].Rahmim A, Qi J, Sossi V. Resolution modeling in PET imaging: theory, practice, benefits, and pitfalls. Med Phys 2013;40:064301. 10.1118/1.4800806. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Gúerit S, Jacques L, Bt M, Lee JA. Post-reconstruction deconvolution of PET images by total generalized variation regularization. Signal Process Conf 2015. 10.1109/EUSIPCO.2015.7362459. [DOI] [Google Scholar]
  • [7].Rousset O, Rahmim A, Alavi A, Zaidi H. Partial volume correction strategies in PET. Pet Clinics 2007;2:235–49. 10.1016/j.cpet.2007.10.005. [DOI] [PubMed] [Google Scholar]
  • [8].Du Y, Tsui BMW, Frey EC. Partial volume effect compensation for quantitative brain SPECT imaging. IEEE Trans Med Imaging 2005;24:969–76. 10.1109/TMI.2005.850547. [DOI] [PubMed] [Google Scholar]
  • [9].Wang Q, D’Agostino E, Seghers D, Maes F, Vandermeulen D, Suetens P. Construction and validation of statistical brain atlases for atlas-based brain image segmentation. Kul/esat/psi 2005;18:5691–702. 10.1093/emboj/18.20.5691. [DOI] [PubMed] [Google Scholar]
  • [10].Meechai T, Tepmongkol S, Pluempitiwiriyawej C. Partial-volume effect correction in positron emission tomography brain scan image using super-resolution image reconstruction. Br J Radiol 2015;88:20140119. 10.1259/bjr.20140119. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Thomas BA, Erlandsson K, Modat M, Thurfjell L, Vandenberghe R, Ourselin S, et al. The importance of appropriate partial volume correction for PET quantification in Alzheimer’s disease. Eur J Nucl Med Mol Imaging 2011;38:1104–19. 10.1007/s00259-011-1745-9. [DOI] [PubMed] [Google Scholar]
  • [12].Erlandsson K, Buvat I, Pretorius PH, Thomas BA, Hutton BF. A review of partial volume correction techniques for emission tomography and their applications in neurology, cardiology and oncology. Phys Med Biol 2012;57:R119–59. 10.1088/0031-9155/57/21/R119. [DOI] [PubMed] [Google Scholar]
  • [13].Tohka J, Reilhac A. Deconvolution-based partial volume correction in Raclopride-PET and Monte Carlo comparison to MR-based method. Neuroimage 2008;39:1570–84. [DOI] [PubMed] [Google Scholar]
  • [14].Buades A, Coll B, Morel JM. A non-local algorithm for image denoising. In: IEEE Computer Society Conference on Computer Vision & Pattern Recognition; 2005. 10.1109/CVPR.2005.38. [DOI] [Google Scholar]
  • [15].Mejia J, Mederos B, Cabrera SD, Domínguez HO, Villegas OOV. Noise reduction in PET sinograms using non-local total variation regularization. Image Anal Interpretation 2014. 10.1109/SSIAI.2014.6806018. [DOI] [Google Scholar]
  • [16].Arabi H, Zaidi H. Spatially-guided non-local mean approach for denoising of PET images. Med Phys 2020;47:1656–69. 10.1002/mp.v47.410.1002/mp.14024. [DOI] [PubMed] [Google Scholar]
  • [17].Dutta J, Fakhri GE, Leahy RM, Li Q. A nonlocal averaging technique for kinetic parameter estimation from dynamic PET data. In: Nuclear Science Symposium & Medical Imaging Conference; 2012. 10.1109/NSSMIC.2011.6153669. [DOI] [Google Scholar]
  • [18].Zhang H, Ma J, Wang J, Liu Y, Lu H, Liang Z. Statistical image reconstruction for low-dose CT using nonlocal means-based regularization. Comput Med Imaging Graph 2014;38:423–35. 10.1016/j.compmedimag.2014.05.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Zhang H, Huang J, Ma J, Bian Z, Feng Q, Lu H, et al. Iterative reconstruction for x-ray computed tomography using prior-image induced nonlocal regularization. IEEE Trans Biomed Eng 2014;61:2367–78. 10.1109/TBME.2013.2287244. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Tang J, Rahmim A. Anatomy assisted PET image reconstruction incorporating multi-resolution joint entropy. Phys Med Biol 2015;60:31–48. 10.1088/0031-9155/60/1/31. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Wang J, Li T, Lu H, Liang Z. Penalized weighted least-squares approach to sinogram noise reduction and image reconstruction for low-dose X-ray computed tomography. IEEE Trans Med Imaging 2006;25:1272–83. 10.1109/TMI.2006.882141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Zhang X, Burger M, Bresson X, Osher S. Bregmanized nonlocal regularization for deconvolution and sparse reconstruction. SIAM J Imag Sci 2010;3:253–76. 10.1137/090746379. [DOI] [Google Scholar]
  • [23].Yang Z, Jacob M. Nonlocal regularization of inverse problems: a unified variational framework. IEEE Trans Image Process 2013;22:3192–203. 10.1109/TIP.2012.2216278. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Gallivanone F, Stefano A, Canevari C, Messa C, Gilardi MC, Castiglioni I. PVE correction in PET from PVE affected images. In: Nuclear Science Symposium Conference Record; 2010. 10.1109/NSSMIC.2009.5401688. [DOI] [Google Scholar]
  • [25].Xu Z, Gao M, Papadakis GZ, Luna B, Jain S, Mollura DJ, et al. Joint solution for PET image segmentation, denoising, and partial volume correction. Med Image Anal 2018;46:229–43. 10.1016/j.media.2018.03.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [26].Teymurazyan A, Riauka T, Jans H-S, Robinson D. Properties of noise in positron emission tomography images reconstructed; with filtered-backprojection and row-action maximum likelihood algorithm. J Digit Imaging 2013;26:447–56. 10.1007/s10278-012-9511-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27].Golla SSV, Lubberink M, Berckel BNMV, Lammertsma AA, Boellaard R. Partial volume correction of brain PET studies using iterative deconvolution in combination with HYPR denoising. Ejnmmi Res 2017;7:36. 10.1186/s13550-017-0284-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Chan C, Fulton R, Barnett R, Feng DD, Meikle S. Postreconstruction nonlocal means filtering of whole-body PET with an anatomical prior. IEEE Trans Med Imaging 2014;33:636–50. 10.1109/TMI.4210.1109/TMI.2013.2292881. [DOI] [PubMed] [Google Scholar]
  • [29].Bai B, Li Q, Leahy RM. Magnetic resonance-guided positron emission tomography image reconstruction. Semin Nucl Med 2013;43:30–44. 10.1053/j.semnuclmed.2012.08.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Yan J, Lim J-S, Townsend DW. MRI-guided brain PET image filtering and partial volume correction. Phys Med Biol 2015;60:961–76. 10.1088/0031-9155/60/3/961. [DOI] [PubMed] [Google Scholar]
  • [31].Gao Y, Bilgel M, Ashrafinia S, Lu L, Rousset O, Resnick S, et al. Evaluation of non-local methods with and without anatomy information for improved quantitative amyloid PET imaging. In: J Nucl Med, 59; 2018. p. 1779. https://jnm.snmjournals.org/content/59/supplement_1/1779.30097506 [Google Scholar]
  • [32].Gao Y, Zhang H, Zhu Y, Bilgel M, Rousset O, Resnick S, et al. Voxel-based partial volume correction of amyloid PET images incorporating non-local means regularization. In: 2018 IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC): IEEE; 2018. p. 1–4. 10.1109/NSSMIC.2018.8824467. [DOI] [Google Scholar]
  • [33].Bettinardi V, Castiglioni I, De Bernardi E, Gilardi MC. PET quantification: strategies for partial volume correction. Clin Transl Imaging 2014;2:199–218. 10.1007/s40336-014-0066-y. [DOI] [Google Scholar]
  • [34].Baete K, Nuyts J, Van Paesschen W, Suetens P, Dupont P. Anatomical-based FDG-PET reconstruction for the detection of hypo-metabolic regions in epilepsy. IEEE Trans Med Imaging 2004;23:510–9. 10.1109/TMI.2004.825623. [DOI] [PubMed] [Google Scholar]
  • [35].Bowsher JE, Johnson VE, Turkington TG, Jaszczak RJ, Floyd CE, Coleman RE. Bayesian reconstruction and use of anatomical a priori information for emission tomography. Med Imaging IEEE Trans 1996;15:673–86. 10.1109/TMI.4210.1109/42.538945. [DOI] [PubMed] [Google Scholar]
  • [36].Leahy RM, Yan X. Incorporation of anatomical MR data for improved dunctional imaging with PET. In: Biennial International Conference on Information Processing in Medical Imaging; 1991. 10.1007/BFb0033746. [DOI] [Google Scholar]
  • [37].Cheng-Liao J, Qi J. PET image reconstruction with anatomical edge guided level set prior. Phys Med Biol 2011;56:6899–918. 10.1088/0031-9155/56/21/009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [38].Somayajula S, Panagiotou C, Rangarajan A, Li Q, Arridge SR, Leahy RM. PET image reconstruction using information theoretic anatomical priors. IEEE Trans Med Imaging 2011;30:537–49. 10.1109/TMI.2010.2076827. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [39].Nuyts J The use of mutual information and joint entropy for anatomical priors in emission tomography. In: IEEE Nuclear Science Symposium Conference Record; 2007. 10.1109/NSSMIC.2007.4437034. [DOI] [Google Scholar]
  • [40].Rahmim A, Tang J. Noise propagation in resolution modeled PET imaging and its impact on detectability. Med Phys Biol 2013;58:6945–68. 10.1088/0031-9155/58/19/6945. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [41].Qi J A unified noise analysis for iterative image estimation. Phys Med Biol 2003; 48:3505–19. 10.1088/0031-9155/48/21/004. [DOI] [PubMed] [Google Scholar]
  • [42].Barrett HH, Wilson DW, Tsui BMW. Noise properties of the EM algorithm. I Theory Phys Med Biol 1994;39:833–46. 10.1088/0031-9155/39/5/005. [DOI] [PubMed] [Google Scholar]
  • [43].Collins DL, Zijdenbos AP, Kollokian V, Sled JG, Kabani NJ, Holmes CJ, et al. Design and construction of a realistic digital brain phantom. Med Imaging IEEE Trans 1998;17:463–8. 10.1109/42.712135. [DOI] [PubMed] [Google Scholar]
  • [44].Weiss KL, Hai P, Storrs J, Strub W, Weiss JL, Li J, et al. Clinical brain MR imaging prescriptions in Talairach space: technologist- and computer-driven methods. AJNR Am J Neuroradiol 2003;24:922–9. 10.1034/j.1600-0455.2003.00077.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [45].Vunckx K, Atre A, Baete K, Reilhac A, Deroose CM, Van Laere K, et al. Evaluation of three MRI-based anatomical priors for quantitative PET brain imaging. IEEE Trans Med Imaging 2012;31:599–612. 10.1109/TMI.2011.2173766. [DOI] [PubMed] [Google Scholar]
  • [46].Belzunce MA, Reader AJ. Technical Note: Ultra high-resolution radiotracer-specific digital PET brain phantoms based on the BigBrain atlas. Med Phys 2020;47: 3356–62. 10.1002/mp.v47.810.1002/mp.14218. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [47].Rahmim A, Cheng J-C, Blinder S, Camborde M-L, Sossi V. Statistical dynamic image reconstruction in state-of-the-art high-resolution PET. Phys Med Biol 2005;50: 4887–912. 10.1088/0031-9155/50/20/010. [DOI] [PubMed] [Google Scholar]
  • [48].Hong IK, Chung ST, Kim HK, Kim YB, Son YD, Cho ZH. Ultra fast symmetry and SIMD-based projection-backprojection (SSP) algorithm for 3-D PET image reconstruction. IEEE Trans Med Imaging 2007;26:789. 10.1109/TMI.2002.808360. [DOI] [PubMed] [Google Scholar]
  • [49].Ashrafinia S, Mohyuddin H, Karakatsanis N, Jha A, Casey M, Dan K, et al. Generalized PSF-modelling for optimized quantitation in PET imaging. Phys Med Biol 2017;62. 10.1088/1361-6560/aa6911. [DOI] [PubMed] [Google Scholar]
  • [50].Burger C, Goerres G, Schoenes S, Buck A, Lonn AHR, Schulthess GK, et al. PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients. Eur J Nucl Med Mol Imaging 2002;29:922–7. 10.1007/s00259-002-0796-3. [DOI] [PubMed] [Google Scholar]
  • [51].Badawi R, Marsden P. Developments in component-based normalization for 3D PET. Phys Med Biol 1999;44:571. 10.1016/j.visres.2005.12.023. [DOI] [PubMed] [Google Scholar]
  • [52].Desikan RS, Ségonne F, Fischl B, Quinn BT, Dickerson BC, Blacker D, et al. An automated labeling system for subdividing the human cerebral cortex on MRI scans into gyral based regions of interest. NeuroImage 2006;31:968–80. 10.1016/j.neuroimage.2006.01.021. [DOI] [PubMed] [Google Scholar]
  • [53].Doshi J, Erus G, Ou Y, Resnick SM, Gur RC, Gur RE, et al. MUlti-atlas region Segmentation utilizing Ensembles of registration algorithms and parameters, and locally optimal atlas selection. NeuroImage 2016;127:186–95. 10.1016/j.neuroimage.2015.11.073. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [54].Zhang H, Zeng D, Lin J, Zhang H, Bian Z, Huang J, et al. Iterative reconstruction for dual energy CT with an average image-induced nonlocal means regularization. Phys Med Biol 2017;62:5556–74. 10.1088/1361-6560/aa7122. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [55].Ma J, Zhang H, Gao Y, Huang J, Liang Z, Feng Q, et al. Iterative image reconstruction for cerebral perfusion CT using a pre-contrast scan induced edge-preserving prior. Phys Med Biol 2012;57:7519–42. 10.1088/0031-9155/57/22/7519. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES