Abstract
Background:
The Tandem Control-IQ (CIQ) system has demonstrated significant glycemic improvements in large randomized controlled and real-world trials. Use of this system is lower in people with type 1 diabetes (T1D) government-sponsored insurance and those with type 2 diabetes (T2D). This analysis aimed to evaluate the performance of CIQ in these groups.
Methods and Materials:
A retrospective analysis of CIQ users was performed. Users age ≥6 years with a t:slim X2 Pump and >30 days of continuous glucose monitoring (CGM) data pre-CIQ and >30 days post-CIQ technology initiation were included.
Results:
A total of 4243 Medicare and 1332 Medicaid CIQ users were analyzed among whom 5075 had T1D and 500 had T2D. After starting CIQ, the Medicare beneficiaries group saw significant improvement in time in target range 70–180 mg/dL (TIR; 64% vs. 74%; P < 0.0001), glucose management index (GMI; 7.3% vs. 7.0%; P < 0.0001), and the percentage of users meeting American Diabetes Association (ADA) CGM Glucometrics Guidelines (12.8% vs. 26.3%; P < 0.0001). The Medicaid group also saw significant improvement in TIR (46% vs. 60%; P < 0.0001), GMI (7.9% vs. 7.5%; P < 0.0001), and percentage meeting ADA guidelines (5.7% vs. 13.4%; P < 0.0001). Patients with T2D and either insurance saw significant glycemic improvements.
Conclusions:
The CIQ system was effective in the Medicare and Medicaid groups in improving glycemic control. The T2D subgroup also demonstrated improved glycemic control with CIQ use. Glucometrics achieved in this analysis are comparable with those seen in previous randomized controlled clinical trials with the CIQ system.
Keywords: Type 1 diabetes, Type 2 diabetes, Hybrid closed loop, Special populations, Disparities
Introduction
Hybrid closed-loop (HCL) systems combining an insulin pump, continuous glucose monitor (CGM), and automated control algorithm have demonstrated significant improvements in glycemic control in both randomized controlled and real-world observational trials.1–5 The Tandem t:slim X2 Control-IQ (CIQ) system (Tandem Diabetes, San Diego, CA) combines the t:slim X2 insulin pump, the Dexcom G6 CGM (Dexcom, San Diego, CA), and the CIQ HCL control algorithm to reduce hypo- and hyperglycemia and maximize CGM time in target range (TIR; 70–180 mg/dL).6,7 A National Institutes of Health (NIH)-sponsored randomized controlled trial of CIQ demonstrated significant improvement of +11% TIR compared to a sensor augmented pump (SAP) control group over 6 months among adults and adolescents with type 1 diabetes (T1D).1
Similar TIR improvement of +11% was demonstrated among children 6–13 years of age in a similar NIH-sponsored randomized controlled trial.2 Real-world analysis of 9451 CIQ users demonstrated attainment of 74%–75% TIR over 12-months of CIQ.3 Subanalysis of this cohort focused on those with the highest baseline glucose values demonstrated a +27% improvement in TIR among those with baseline glucose management indicator (GMI) ≥9% and a +28% improvement in TIR among those with baseline GMI ≥10%.8
Despite these noted benefits of HCL technology, there have been many barriers to technology adoption, particularly among people with insulin-requiring type 2 diabetes (T2D), Medicare beneficiaries and those covered by Medicaid insurance.9–11 Lower socioeconomic status, presence of government-sponsored insurance, and minority racial-ethnic status have been associated with lower rates of technology use (both insulin pump and CGM) and higher hemoglobin A1c (HbA1c) values in numerous studies.9,10,12–15 Evidence of the benefits of insulin pump therapy in T2D has been growing, with consistent improvements in HbA1c, reduction of hypoglycemia as well as total daily dose of insulin, both short and long term.16–23 In the Medicare and Medicaid beneficiaries in particular, coverage criteria determined by the Centers for Medicare and Medicaid Services currently limit access to insulin pumps, which contribute to these disparities.24,25
Furthermore, within Medicare beneficiaries, significant race-ethnicity-associated differences exist, and greater gaps in diabetes technology adoption have been noted.11 In this real-world retrospective analysis, we aimed to assess glycemic control outcomes with CIQ use among Medicare and Medicaid-beneficiaries with any type of diabetes and those with T2D with either type of insurance.
Methods and Materials
Study design/data sources
We performed a retrospective analysis of users of CIQ technology in the United States who had descriptive data available in Tandem's Customer Relations Management database and who had uploaded their glycemic data—either through Tandem's t:connect uploader or through the mobile app—to Tandem's t:connect web application from January 11, 2020 to January 11, 2022. The data extracted for analysis were deidentified. Participants consented to the use of their data for research purposes as part of their onboarding to Tandem when initiating their t:connect account. Glycemic outcomes were calculated for all participants who had at least 30 days of CGM data with ≥75% CGM availability before and after CIQ initiation.
The outcomes were calculated by participant and were limited to data collected while the participants were using Software versions 6 (t:slim X2 with Basal-IQ) and Software version 7 (t:slim X2 with CIQ). CGM values outside the valid ranges of 40–400 mg/dL were filtered out, with glucose value <40 and >400 mg/dL being saturated at 40 and 400 mg/dL, respectively. No CGM interpolation was performed. Glycemic outcomes are reported regardless of closed-loop status. No Institutional Review Board approval was sought for this retrospective analysis.
Study population/participants
Data from users who were aged 6 years and older, had at least 12 consecutive months of data available on CIQ, and had at least 30 days of ≥75% CGM data availability before and after CIQ initiation were included in the analysis. Any CIQ users with data uploaded in t:connect that did not meet these criteria were excluded.
Outcome metrics
Our primary outcome of interest was difference in TIR after starting CIQ use for at least 30 days after CIQ initiation, compared to baseline. We also analyzed change in GMI, time below range (TBR) 54–<70 mg/dL, TBR <54 mg/dL, time above range (TAR) >180 mg/dL, and TAR >250 mg/dL.
Definitions
CGM metrics are calculated and presented as recommended by the international consensus on time in range and using the American Diabetes Association (ADA) targets for TIR metrics.26 Percent time in closed-loop automation was calculated as the percentage of the total basal rates delivered by the pump, in 5-min increments, which were decided by the CIQ algorithm.
Metrics for analysis were defined as follows:
Time in range (TIR): Median of the % time spent in the clinically defined range of 70–180 mg/dL.
TBR: Median of the % time spent in the clinically defined range of <70 mg/dL.
Time in automation: Median percent of the total basal rates delivered by the pump, in 5-min increments, which were decided by the CIQ algorithm.
GMI27: GMI = 3.31 + 0.02392 × [mean glucose in mg/dL]. The average glucose is calculated over the entire time a customer used a Tandem pump in accordance with the guidelines above.
User Meeting ADA guidelines for CGM glucometrics28: GMI Below 7% with a Time in Range of >70% and either TBR of <4% for those younger than the age of 65 years or a TBR of <1% for those of age 65 years and older.
Exposure variables
Descriptive data were organized according to diabetes type, age, gender, time since diabetes diagnosis, GMI, previous Tandem pump software versions, and time in CIQ technology automation.
Definitions
Prior HbA1c (Reported HbA1c): The last recorded HbA1c reported by a customer as part of the sales process. Values were included if they were taken within 6 months of initial t:connect data.
Diabetes type: Self-reported clinical diabetes type 1 or type 2.
Age: Age of customer at the report run-time.
Sex: Self-reported sex
Payer: The last payer the customer used for a Tandem pump or supply purchase available in Tandem's Customer Relations Management database.
Statistical analysis
Outcomes were analyzed using Wilcoxon signed rank test and are reported as median (quartiles). Differences in meeting ADA guidelines for CGM glucometrics28 were analyzed using a test of proportions. Significance of differences in Time in Range and GMI were tested using Wilcoxon tests. Differences in meeting ADA guidelines were analyzed using a test of proportions.
Results
Study population
There were 5575 CIQ users with sufficient pre- and post-CIQ data for analysis (Table 1). The Medicare beneficiaries group consisted of 4243 CIQ users, with a mean ± standard deviation (SD) age of 67.4 ± 10.9 years, 48.3% male, of which 89% had T1D, 11% had T2D, and median (interquartile range [IQR]) baseline GMI was 7.3% (6.9%–7.7%). The Medicaid group consisted of 1332 CIQ users with a mean ± SD age of 22.3 ± 14.2 years, 44.1% male, of which 98% had T1D, 2% had T2D, and median (IQR) baseline GMI was 7.9% (7.4%–8.6%). The combined T2D subgroup with Medicare and/or Medicaid insurance consisted of 500 CIQ users with a mean ± SD age of 69.2 ± 10.5 years, 55.2% male, and with a median (IQR) baseline GMI of 7.3% (6.9%–7.7%).
Table 1.
Cohort Baseline Characteristics
All | Insurance cohorts |
Type 2 diabetes (Medicare + Medicaid) | ||
---|---|---|---|---|
Medicare | Medicaid | |||
n | 5575 | 4243 | 1332 | 500 |
Mean age (years) | 56.7 ± 22.5 | 67.4 ± 10.9 | 22.3 ± 14.2 | 69.2 ± 10.5 |
Male (%) | 47.3 | 48.3 | 44.1 | 55.2 |
Diabetes type (%) | ||||
Type 1 | 91 | 89 | 98 | 0 |
Type 2 | 9 | 11 | 2 | 100 |
GMI (%) | 7.4 (7.0–7.9) | 7.3 (6.9–7.7) | 7.9 (7.4–8.6) | 7.3 (6.9–7.7) |
Mean SG (mg/dL) | 171.0 | 166.8 | 191.9 | 166.8 |
TIR 70–180 mg/dL (%) | 60 | 64 | 46 | 64 |
TBR 54–69 mg/dL (%) | 0.75 | 0.74 | 0.74 | 0.26 |
TBR <54 mg/dL (%) | 0.12 | 0.13 | 0.15 | 0.04 |
TAR 181–250 mg/dL (%) | 28 | 26 | 27 | 27 |
TAR >250 mg/dL (%) | 9.9 | 8 | 21 | 7 |
GMI, glucose management indicator; SG, sensor glucose; TAR, time above range; TBR, time below range; TIR, time in target range.
The cohort with Medicaid insurance was then subdivided by age group using standard age divisions: 6–13, 14–18, 19–64, and 65+ years. The 6–13 years age group (n = 394) was 10.4 ± 2.2 years of age and 49.5% male. The 14–18 years age group (n = 367) was 15.8 ± 1.4 years of age and 50.1% male. The 19–64 years age group (n = 554) was 33.7 ± 12.4 years of age and 35.6% male. The 65+ group (n = 17) was 69.0 ± 4.4 years of age and 64.7% male.
Glycemic control in the Medicare cohort
Sufficient data were available for 4243 CIQ users who listed Medicare as their insurance. This cohort was 67.4 ± 10.9 years of age. After starting CIQ, the Medicare group saw a significant decrease in mean sensor glucose (SG) from 166.8 to 154.3 mg/dL (P < 0.0001), a significant decrease in GMI from 7.3% to 7.0% (P < 0.0001), and a significant increase in TIR from 64% to 74% (P < 0.0001; Table 2). This was seen with no change in level 1 hypoglycemia (0.74% vs. 0.74%; P = 0.327) and with a statistically significant, although overall minimal change in level 2 hypoglycemia (0.11% vs. 0.13%; P < 0.0001; Fig. 1).
Table 2.
Glucose Management Indicator and Time in Range Using Tandem Control-IQ by Cohorts
Medicare |
Medicaid |
Type 2 diabetes |
|||||||
---|---|---|---|---|---|---|---|---|---|
Pre-CIQ | Post-CIQ | P | Pre-CIQ | Post-CIQ | P | Pre-CIQ | Post-CIQ | P | |
GMI (%) | 7.3 | 7.0 | <0.0001 | 7.9 | 7.5 | <0.0001 | 7.3 | 7.1 | <0.0001 |
Mean SG (mg/dL) | 166.8 | 154.3 | <0.0001 | 191.9 | 175.2 | <0.0001 | 166.8 | 158.4 | <0.0001 |
TIR 70–180 mg/dL (%) | 64 | 74 | <0.0001 | 46 | 60 | <0.0001 | 64 | 72 | <0.0001 |
TBR 54–69 mg/dL (%) | 0.74 | 0.74 | 0.327 | 0.74 | 0.75 | 0.518 | 0.26 | 0.28 | 0.719 |
TRR <54 mg/dL (%) | 0.11 | 0.13 | <0.0001 | 0.15 | 0.18 | <0.0001 | 0.04 | 0.06 | <0.0001 |
TAR 181–250 mg/dL (%) | 26 | 20 | <0.0001 | 27 | 24 | <0.0001 | 27 | 22 | <0.0001 |
TAR >250 mg/dL (%) | 8 | 5 | <0.0001 | 21 | 13 | <0.0001 | 7 | 5 | <0.0001 |
Users meeting ADA goals (%) | |||||||||
All guidelines | 12.8 | 26.3 | <0.0001 | 5.7 | 13.4 | <0.0001 | 19.4 | 30.2 | <0.0001 |
GMI <7% | 30.6 | 50.9 | <0.0001 | 10.3 | 18.1 | <0.0001 | 28.8 | 40.8 | <0.0001 |
TIR >70% | 35.1 | 62.1 | <0.0001 | 11.0 | 22.3 | <0.0001 | 37.4 | 54.8 | <0.0001 |
TBR <4%* | 91 | 93 | 0.551 | 92 | 93 | 0.306 | 98 | 97 | 0.936 |
TBR <1%** | 57 | 56 | 0.694 | 53 | 41 | 0.616 | 81 | 81 | 0.999 |
ADA, American Diabetes Association; CIQ, Control-IQ.
TBR of < 4% for those younger than the age of 65 years.
TBR of <1% for those of age 65 years and older.
FIG. 1.
Medicare CIQ users pre- and post-CIQ glycemic metrics. (A) Change in time in ranges pre- and post-CIQ, P-value shown for pre-/postchange in TIR 70–180 mg/dL. (B) Change in GMI pre- and post-CIQ. (C) Percent of people meeting ADA TIR target, level 2 hypoglycemia target, level 1 hypoglycemia target, GMI <7%, 8%, and 9% targets. ADA, American Diabetes Association; CIQ, Control-IQ; GMI, glucose management indicator; TIR, time in target range.
At baseline, 12.8% of the Medicare cohort was meeting all ADA CGM guidelines for GMI, TIR, and TBR while after CIQ use, this percentage significantly increased to 26.3% (P < 0.0001). The percentage of users meeting the individual guideline for GMI <7% significantly increased from 30.6% to 50.9% (P < 0.0001). The percentage meeting the TIR goal of >70% also significantly increased from 35.1% at baseline to 62.1% with CIQ use (P < 0.0001). The percentage meeting TBR goals of <4% for age <65 years and <1% for age ≥65 years remained unchanged (Table 2).
Within the Medicare cohort, there were 806 users who transitioned from multiple daily injection (MDI) therapy to CIQ therapy. This subgroup had a higher baseline GMI at 7.9% and saw a significant decline in GMI to 7.1% (difference of −0.8%; P < 0.0001).
Glycemic control in the Medicaid cohort
Sufficient data were available for 1332 users who listed Medicaid as their insurance. After starting CIQ, the Medicaid population saw a significant decrease in mean SG from 191.9 to 175.2 mg/dL (P < 0.0001), a significant decrease in GMI from 7.9% to 7.5% (P < 0.0001), and a significant increase in TIR from 46% to 60% (P < 0.0001) (Table 2). This was seen with no change in level 1 hypoglycemia (0.76% vs. 0.72%; P = 0.518) and with a statistically significant, although overall minimal change in level 2 hypoglycemia (0.15% vs. 0.18%; P < 0.0001; Fig. 2).
FIG. 2.
Medicaid CIQ users pre- and post-CIQ glycemic metrics. (A) Change in time in ranges pre- and post-CIQ, P-value shown for pre-/postchange in TIR 70–180 mg/dL. (B) Change in GMI pre- and post-CIQ. (C) Percent of people meeting ADA TIR target, level 2 hypoglycemia target, level 1 hypoglycemia target, GMI <7%, 8%, and 9% targets.
At baseline, only 5.7% of the Medicaid cohort was meeting all the ADA CGM guidelines for GMI, TIR, and TBR, while after CIQ use this number more than doubled to 13.4% (P < 0.0001). The percentage of users meeting the individual guideline for GMI <7% significantly increased from 10.3% to 18.1% (P < 0.0001) while the percentage meeting the TIR goal of >70% more than doubled from 11.0% to 22.3% (P < 0.0001). The percentage meeting TBR goals (levels 1 and 2) remained unchanged.
The Medicaid cohort was further subanalyzed by age group (Table 3). The age-group based analysis demonstrated that significant improvement for GMI was seen for children 6–13 years of age (8.2% vs. 7.6%; P < 0.0001), adolescents 14–18 years of age (8.2% vs. 7.6%; P < 0.0001), adults 19–64 years of age (7.7% vs. 7.3%; P < 0.0001), and seniors 65+ years of age (7.4% vs. 7.2%; P = 0.0089). Across all age groups TIR was also significantly increased without significant change in level 1 or level 2 hypoglycemia (Fig. 3).
Table 3.
Glucose Management Indicator and Time in Range Using Tandem Control-IQ among Medicaid Patients by Age Group
6–13 years |
14–18 years |
19–64 years |
65+ years |
|||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Pre-CIQ | Post-CIQ | P | Pre-CIQ | Post-CIQ | P | Pre-CIQ | Post-CIQ | P | Pre-CIQ | Post-CIQ | P | |
GMI (%) | 8.2 | 7.6 | <0.0001 | 8.2 | 7.6 | <0.0001 | 7.7 | 7.3 | <0.0001 | 7.4 | 7.2 | 0.0089 |
Mean SG (mg/dL) | 204.4 | 179.4 | <0.0001 | 204.4 | 179.4 | <0.0001 | 183.5 | 166.8 | <0.0001 | 171.0 | 162.6 | 0.0089 |
TIR 70–180 mg/dL (%) | 43 | 57 | <0.0001 | 42 | 56 | <0.0001 | 53 | 65 | <0.0001 | 58 | 63 | 0.0011 |
TBR 54–69 mg/dL (%) | 0.72 | 0.81 | 0.0281 | 0.65 | 0.69 | 0.4642 | 0.84 | 0.80 | 0.1655 | 0.83 | 1.06 | 0.7119 |
TRR <54 mg/dL (%) | 0.14 | 0.19 | 0.0008 | 0.13 | 0.18 | <0.0001 | 0.16 | 0.18 | 0.0174 | 0.21 | 0.17 | 0.3627 |
TAR 181–250 mg/dL (%) | 28 | 24 | <0.0001 | 27 | 24 | <0.0001 | 27 | 23 | <0.0001 | 26 | 22 | 0.0008 |
TAR >250 mg/dL (%) | 26 | 17 | <0.0001 | 26 | 17 | <0.0001 | 16 | 10 | <0.0001 | 11 | 11 | 0.0056 |
Users meeting ADA goals (%) | ||||||||||||
All guidelines | 2.8 | 7.6 | 0.0023 | 3.5 | 9.3 | 0.0015 | 9.2 | 20.6 | <0.0001 | 5.9 | 5.9 | 0.9999 |
GMI <7% | 5.9 | 10.4 | 0.0189 | 6 | 10.9 | 0.0169 | 15.9 | 27.8 | <0.0001 | 23.5 | 35.3 | 0.4516 |
TIR >70% | 5.6 | 11.4 | 0.0033 | 6.3 | 15 | <0.0001 | 15.9 | 27.8 | <0.0001 | 23.5 | 35.3 | 0.4516 |
TBR <4%* | 92 | 93 | 0.5792 | 94 | 96 | 0.1600 | 90 | 90 | 0.7640 | n/a | n/a | n/a |
TBR <1%** | n/a | n/a | n/a | n/a | n/a | n/a | n/a | n/a | n/a | 53 | 41 | 0.4920 |
TBR of < 4% for those younger than the age of 65 years.
TBR of <1% for those of age 65 years and older.
FIG. 3.
Medicaid CIQ users pre- and post-CIQ analysis by age group. (A) Change in time in ranges pre- and post-CIQ by age group, P-values shown for pre-/postchange in TIR 70–180 mg/dL by age group. (B) Change in GMI pre- and post-CIQ.
The percentage of users meeting all ADA guidelines for CGM glucometrics was significantly improved in the 6–13-, 14–18-, and 19–64-year age groups and unchanged in the 65+-year age group. The percentage meeting the individual guideline for GMI <7% and TIR >70% was significantly improved in the 6–13-, 14–18-, and 19–64-year age groups, and unchanged in the 65+-year age group (Table 3).
Within the Medicaid group, there were 659 users who transitioned from MDI therapy to CIQ therapy. This group had a baseline GMI of 8.6% and saw a significant decline in GMI to 7.4% (difference of −1.2%; P < 0.0001).
Glycemic control among Medicaid and Medicare CIQ users with T2D
There were 500 individuals with Medicare or Medicaid insurance who listed their diabetes category as T2D. The population with T2D saw a significant decline in mean SG from 166.8 to 158.4 mg/dL (P < 0.0001), a significant decrease in GMI from 7.3% to 7.1% (P < 0.0001), and a significant increase in TIR from 64% to 72% (P < 0.0001) (Table 2). This was seen without a change in level 1 hypoglycemia and with a small, but statistically significant change in level 2 hypoglycemia (0.04%–0.06%; P < 0.0001) (Figure 4).
FIG. 4.
Medicare + Medicaid CIQ users with type 2 diabetes pre- and post-CIQ glycemic metrics. (A) Change in time in ranges pre- and post-CIQ, P-value shown for pre-/postchange in TIR 70–180 mg/dL. (B) Change in GMI pre- and post-CIQ. (C) Percent of people meeting ADA TIR target, level 2 hypoglycemia target, level 1 hypoglycemia target, GMI <7%, 8%, and 9% targets.
The percentage of people with T2D meeting all ADA CGM guidelines for GMI, TIR, and TBR significantly increased with CIQ use from 19.4% to 30.2% (P < 0.0001). For the individual guidelines, there was significant improvement in the percentage meeting ADA CGM guidelines for GMI and TIR with CIQ use. The percentage meeting guidelines for TBR (levels 1 and 2) did not change among the T2D cohort.
Among the T2D subgroup, there were 238 users who transitioned from MDI therapy to CIQ therapy. They had a baseline GMI of 8.1% and saw a significant improvement in GMI to 7.2% (difference of −0.9%; P < 0.0001).
Discussion
The CIQ system was effective in the Medicare and Medicaid beneficiaries' groups, including a subset of patients with T2D, at improving glycemic control as assessed by a variety of glycemic metrics and target goals. There was significant reduction in GMI in the Medicare group by 0.3%, in the Medicaid group by 0.4%, and among the T2D subset by 0.2%. There was also significant improvement in TIR in the Medicare group by 10%, in the Medicaid group by 14%, and in the T2D subset by 8%. These improvements were seen without significant change in level 1 hypoglycemia in any of the analyses, although with slight increases in level 2 hypoglycemic exposure. There were also significant improvements in the percentage of people meeting ADA CGM guidelines for all targets, GMI, and TIR for all the groups analyzed.
The CIQ system has previously been tested in a large NIH-sponsored series of randomized controlled trials.1,2 The data published by Brown et al., demonstrated that compared to SAP, CIQ use resulted in an 11% improvement in TIR, 0.33% reduction in HbA1c, 13 mg/dL reduction in average sensor glucose, and 0.88% reduction in level 1 hypoglycemia.1 In that study, CIQ users reached a TIR of 71%, HbA1c of 7.06%, and TBR % <70 mg/dL of 1.58%. In the present analysis, users with Medicare saw a similar decline in HbA1c reaching a similar value of about 7.0%.
They also saw a similar improvement in TIR reaching a slightly higher TIR of 74%. This was done while maintaining a TBR % <70 mg/dL of under 1%. There was a statistically significant, although clinically insignificant, increase in level 2 hypoglycemia from 0.11% to 0.13%. These findings demonstrate that the Medicare beneficiaries group is seeing similar benefits with CIQ use to those seen in the Tandem t:slim with CIQ pivotal study by Brown et al.1
The Medicaid cohort contained pediatric, adolescent, and adult participants. Among the 14–18-, 19–64-, and 65+-year old Medicaid CIQ users, GMI and TIR were significantly improved across all age groups. This was seen without a change in level 1 hypoglycemia. There was a statistically significant, although likely clinically insignificant, increase in level 2 hypoglycemia from 0.15% to 0.18%. The GMI reductions were −0.6%, −0.4%, and −0.2%, respectively. The TIR improvements were +14%, +13%, and +5%, respectively. The changes for the age cohorts of 14–18 and 19–64 year also compare favorably with the improvements seen in the study by Brown et al.1 The improvements seen in the 65+-year cohort were somewhat smaller, although still statistically significant within this analysis.
The data published by Breton et al., compared SAP use to CIQ use in an NIH-sponsored randomized controlled trial.2 In this pediatric study, CIQ use resulted in an 11% improvement in TIR, 0.4% reduction in HbA1c, 13 mg/dL reduction in average glucose, and no change in % <70 mg/dL. In that study, CIQ users reached a TIR of 67%, HbA1c of 7.0%, and TBR % <70 mg/dL of 1.6%. The 6–13-year-old CIQ Medicaid users saw a GMI improvement of 0.6% reaching a GMI of 7.6% with a TIR improvement of 14% reaching a TIR of 57%, and with a TBR % <70 mg/dL below 1%. These results demonstrate that the real-world pediatric Medicaid population is showing a similar or even greater benefit from CIQ use compared to that seen in the NIH-trial by Breton et al.2
The evidence on insulin pump therapy benefits in T2Ds and older adults with T2Ds, such Medicare age beneficiaries, continues to increase. A multicenter open-label, clinical trial with 56 T2D participants (mean age 57 ± 10 years) studied the effects of insulin pump initiation after discontinuation of all other oral medications except for metformin. After 16 weeks, the mean HbA1c decreased by 1.2 ± 1.2% (P < 0.001) with no episodes of severe hypoglycemia.16
The Opt2mise Glucose Control in Type 2 Diabetes Mellitus With Insulin Pump Therapy (OpT2mise) was the first large scale randomized trial that studied the effects of insulin pump therapy in 331 adults with T2D, including 69 subjects aged 65 years and older. In this trial, the decrease in HbA1c was greater in the insulin pump group compared to the MDI group with a between-group treatment difference of −0.7% (95% confidence interval −0.9 to −0.4; P < 0.0001) and a 21% reduction of mean total daily insulin dose in the pump treatment group (P < 0.0001).17
In addition, durable effects of insulin pump therapy in T2D have been reported as well. A cohort of 161 T2D (mean age at insulin pump start 58.3 ± 9.8 years) was followed for up to 9 years (mean follow-up 5.1 ± 3.2 years). After 1 year, the HbA1c had decreased by 1.3% from baseline (P < 0.001), and after 9 years of follow-up, the HbA1c decrease was maintained (P < 0.05).29
Very recently, a retrospective observational study of 3952 T2D participants using a tubeless insulin pump showed a reported reduction in mean HbA1c by −1.3 ± 1.7% (P < 0.0001) after 90 days of use. Within this cohort, 914 participants (25%) were aged 65 years or older, and they also experienced a decrease in HbA1c of −0.9% ± 1.3 (P < 0.0001).23
Finally, an open label, randomized, crossover pilot study compared glucose control under single hormone HCL versus MDI in adults with T2D older than the age of 55 (mean age 63.6 ± 6.7 years old). In this short study, TIR improved by 21.6% overnight and 7.3% over 24 h without increasing hypoglycemia risk when using HCL.4
Despite the growing evidence of clinical benefits of insulin pump therapy in T2D, the value of insulin pump therapy with HCL has been underestimated, potentially due to perceived overall less risk for hypoglycemia in this population if early in the disease and using newer classes of pharmacologic agents.28 However, older people with T2D such as those on Medicare, and with advanced disease requiring insulin, are a greater risk of hypoglycemia, not only due to insulin therapy but also due to the likely presence of diabetic complications such as renal disease.30,31 The analysis of a retrospective cohort study of 50,439 T2D subjects from 2006 to 2015 revealed that the incidence of severe hypoglycemia increased from 0.12% in 2006 to 0.31% in 2015 (P = 0.01) and the presence of severe hypoglycemia was higher in those with previous diagnosis of nonsevere hypoglycemia (9% vs. 2%, P < 0.001).32
Moreover, fear of hypoglycemia in the patient with T2D is a very real entity. A survey of 424 T2D participants, of which 53.3% on insulin, studied the hypoglycemic attitudes and behavioral scale. The insulin using group had longer duration of diabetes, higher HbA1c, history of severe hypoglycemia (all P < 0.001), and greater fear of hypoglycemia than the noninsulin using group (P < 0.001).33 Therefore, automated insulin delivery systems use in T2D present an excellent regimen modality to not only improve glucometrics, but also to reduce hypoglycemia regardless of the HbA1c levels.
As demonstrated by this real-world analysis of CIQ use, in the 500 individuals with Medicare or Medicaid who listed their diabetes category as T2D, the HbA1c significantly decreased from 7.3% to 7.1% (P < 0.001) with significant increase in TIR to 72% (P < 0.001), without changes in level 1 hypoglycemia and a small but statistically significant reduction of level 2 hypoglycemia (P < 0.001). In addition, in the 238 users who transitioned from MDI therapy to CIQ therapy, the GMI decreased from baseline by −0.9% (8.1%–7.2%, P < 0.001). CIQ users with T2D who met all the ADA CGM guidelines for GMI, TIR, and TBR increased to 30.2% (P < 0.001).
Strengths of the present study include the large sample sizes for all of the groups analyzed, which are significantly larger than those available for device prospective trials. The analysis also benefits from a real-world setting, which makes the data more generalizable to clinical practice. Weaknesses of the present study include the lack of follow-up biological HbA1c data and thus the reliance on GMI as a surrogate. The need to have uploaded device data may decrease the generalizability of results as those device users who did not upload their data would not be represented. The analyses were performed using a reporting dashboard of real-world data and are limited to predetermined analyses existing within the dashboard tools.
Other metrics of interest such as daytime versus nighttime analysis, mealtime analysis, analysis of coefficient of variation of sensor glucose, and analysis of newer metrics such as Glycemia Risk Index are not currently available within these tools.
In conclusion, the present analysis demonstrates that the use of CIQ HCL technology is highly successful at improving glycemic control in Medicare and Medicaid beneficiaries, including those with T2D when compared to their previous therapy. The glycemic targets achieved compares favorably in terms of GMI, mean SG, and TIR to the levels achieved in the HCL group within prior randomized controlled clinical trials of this system. These findings support our hypothesis that Medicare and Medicaid beneficiaries with both T1D and T2D benefit from the CIQ HCL system, and prescribers should utilize this technology to achieve glycemic benefits within these individuals.
Acknowledgments
We thank our partners at Tandem Diabetes Care who contributed to data acquisition, cleaning, and analysis, including Steph Habif, Brian Hansen, Chris Dawson, Joey Dagostino, Lars Mullen, and Alexandra Constantin. We thank the people living with diabetes and using the CIQ system, who contributed data to this analysis.
Authors' Contributions
G.P.F. wrote sections of the article and contributed to critical review and revision of the work. A.L.C., R.J.G., D.F.K., C.J.L., J.B.M., and G.U. all contributed to critical review and revision of the work. G.A. wrote sections of the article and contributed to critical review and revisions of the work.
Author Disclosure Statement
G.P.F. conducts research sponsored by Medtronic, Dexcom, Abbott, Tandem, Insulet, Beta Bionics, and Lilly and has served as a consultant/speaker/ad board member for Medtronic, Dexcom, Abbott, Tandem, Insulet, Beta Bionics, and Lilly. A.L.C. conducts research sponsored by Medtronic, Abbott, Insulet, Dexcom, Sanofi, UnitedHealth, and Eli Lilly and Novo Nordisk, and has served as a consultant/speaker/advisory board member for Mannkind, Medtronic, and Sanofi. R.J.G. received research support to Emory University for investigator-initiated studies from Novo Nordisk, Dexcom, and Eli Lilly and consulting fees from Sanofi, Eli Lilly, and Weight Watchers, outside of this work. D.F.K. has served on advisory boards and/or speaker bureaus for Dexcom and Abbott Diabetes Care, and her institution received research funding from Dexcom, Abbott Diabetes and Beta bionics.
C.J.L. reports research support from Abbott Diabetes, Dexcom, Insulet, Tandem Diabetes Care, and Senseonics paid to her institution, nonfinancial device support from Dexcom, Tandem Diabetes Care, and Abbott Diabetes, service as a consultant for Sanofi, Eli Lilly, and Dexcom, and Novo Nordisk outside the submitted work. J.B.M. conducts research sponsored by Beta Bionics and Novo Nordisk; and has served as a consultant/advisor for Bayer, Boehringer Ingelheim, Provention Bio, and Mannkind Novo Nordisk. G.A. has received research support to Northwestern University from Dexcom, Eli-Lilly, Emmes, Fractyl Health, and Insulet; and has served as consultant for Bayer, Dexcom, Eli Lilly and Insulet.
Funding Information
R.J.G. is partially supported by the National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK) of the National Institutes of Health (NIH) under Award Numbers P30DK111024 (R.J.G.) and K23DK123384 (R.J.G.). G.U. is partly supported by research grants from the NIH/NATS UL1 TR002378, 1P30DK111024-05, and P30DK111024-05S and has received unrestricted research support (to Emory University) from AstraZeneca, Novo Nordisk, and Dexcom. C.J.L. is or recently has been partially supported by grants from NIH/NIDDK (R01DK125856, P30DK020541, 1R01DK120358-0 ) and the Helmsley foundation.
References
- 1. Brown SA, Kovatchev BP, Raghinaru D, et al. : Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes. N Engl J Med 2019;381:1707–1717. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Breton MD, Kanapka LG, Beck RW, et al. : A randomized trial of closed-loop control in children with type 1 diabetes. N Engl J Med 2020;383:836–845. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Breton MD, Kovatchev BP: One year real-world use of the control-IQ advanced hybrid closed-loop technology. Diabetes Technol Ther 2021;23:601–608. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Taleb N, Carpentier AC, Messier V, et al. : Efficacy of artificial pancreas use in patients with type 2 diabetes using intensive insulin therapy: a randomized crossover pilot trial. Diabetes Care 2019;42:e107–e109. [DOI] [PubMed] [Google Scholar]
- 5. Da Silva J, Lepore G, Battelino T, et al. : Real-world Performance of the MiniMed™ 780G system: first report of outcomes from 4'120 users. Diabetes Technol Ther 2022;24:113–119. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Anderson SM, Dassau E, Raghinaru D, et al. : The international diabetes closed-loop study: testing artificial pancreas component interoperability. Diabetes Technol Ther 2019;21:73–80. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Patek SD, Magni L, Dassau E, et al. : Modular closed-loop control of diabetes. IEEE Trans Biomed Eng 2012;59:2986–2999. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Forlenza GP, Breton MD, Kovatchev B: Candidate selection for hybrid closed loop systems. Diabetes Technol Ther 2021;23:760–762. [DOI] [PubMed] [Google Scholar]
- 9. Addala A, Auzanneau M, Miller K, et al. : A decade of disparities in diabetes technology use and HbA(1c) in pediatric type 1 diabetes: a transatlantic comparison. Diabetes Care 2021;44:133–140. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Agarwal S, Crespo-Ramos G, Long JA, Miller VA: “I didn't really have a choice”: qualitative analysis of racial-ethnic disparities in diabetes technology use among young adults with type 1 diabetes. Diabetes Technol Ther 2021;23:616–622. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Wherry K, Zhu C, Vigersky RA: Inequity in adoption of advanced diabetes technologies among Medicare fee-for-service beneficiaries. J Clin Endocrinol Metab 2021;107:e2177–e2185. [DOI] [PubMed] [Google Scholar]
- 12. O'Connor MR, Carlin K, Coker T, et al. : Disparities in insulin pump therapy persist in youth with type 1 diabetes despite rising overall pump use rates. J Pediatr Nurs 2019;44:16–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Monkemoller K, Muller-Godeffroy E, Lilienthal E, et al. : The association between socio-economic status and diabetes care and outcome in children with diabetes type 1 in Germany: The DIAS study (diabetes and social disparities). Pediatr Diabetes 2019;20:637–644. [DOI] [PubMed] [Google Scholar]
- 14. Miller KM, Beck RW, Foster NC, Maahs DM: HbA1c levels in type 1 diabetes from early childhood to older adults: a deeper dive into the influence of technology and socioeconomic status on HbA1c in the T1D exchange clinic registry findings. Diabetes Technol Ther 2020;22:645–650. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Fredette ME, Zonfrillo MR, Park S, et al. : Self-reported insulin pump prescribing practices in pediatric type 1 diabetes. Pediatr Diabetes 2021;22:758–765. [DOI] [PubMed] [Google Scholar]
- 16. Edelman SV, Bode BW, Bailey TS, et al. : Insulin pump therapy in patients with type 2 diabetes safely improved glycemic control using a simple insulin dosing regimen. Diabetes Technol Ther 2010;12:627–633. [DOI] [PubMed] [Google Scholar]
- 17. Reznik Y, Cohen O, Aronson R, et al. : Insulin pump treatment compared with multiple daily injections for treatment of type 2 diabetes (OpT2mise): a randomised open-label controlled trial. Lancet 2014;384:1265–1272. [DOI] [PubMed] [Google Scholar]
- 18. Aronson R, Reznik Y, Conget I, et al. : Sustained efficacy of insulin pump therapy compared with multiple daily injections in type 2 diabetes: 12-month data from the OpT2mise randomized trial. Diabetes Obes Metab 2016;18:500–507. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Layne JE, Parkin CG, Zisser H: Efficacy of a tubeless patch pump in patients with type 2 diabetes previously treated with multiple daily injections. J Diabetes Sci Technol 2017;11:178–179. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Singh P, Pandey D, Trivedi N: Insulin pump in difficult to control type 2 diabetes: a single center, five years' experience. Cureus 2018;10:e3240. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Chlup R, Runzis S, Castaneda J, et al. : Complex assessment of metabolic effectiveness of insulin pump therapy in patients with type 2 diabetes beyond HbA1c reduction. Diabetes Technol Ther 2018;20:153–159. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Grunberger G, Bhargava A, Ly T, et al. : Human regular U-500 insulin via continuous subcutaneous insulin infusion versus multiple daily injections in adults with type 2 diabetes: The VIVID study. Diabetes Obes Metab 2020;22:434–441. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Carlson AL, Huyett LM, Jantz J, et al. : Improved glycemic control in 3,592 adults with type 2 diabetes mellitus initiating a tubeless insulin management system. Diabetes Res Clin Pract 2021;174:108735. [DOI] [PubMed] [Google Scholar]
- 24. Aleppo G, Parkin CG, Carlson A, et al. : Lost in translation. A disconnect between the science and medicare coverage criteria for continuous subcutaneous insulin infusion (CSII). Diabetes Technol Ther 2021;23:715–725. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Galindo RJ, Parkin CG, Aleppo G, et al. : What's wrong with this picture? A critical review of current Centers for Medicare & Medicaid Services Coverage Criteria for Continuous Glucose Monitoring. Diabetes Technol Ther 2021;23:652–660. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Battelino T, Danne T, Bergenstal RM, et al. : Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range. Diabetes Care 2019;42:1593–1603. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Bergenstal RM, Beck RW, Close KL, et al. : Glucose management indicator (GMI): a new term for estimating A1C from continuous glucose monitoring. Diabetes Care 2018;41:2275–2280. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. American Diabetes Association Professional Practice Committee: 6. Glycemic Targets: Standards of Medical Care in Diabetes—2022. Diabetes Care 2021;45(Suppl. 1):S83–S96. [DOI] [PubMed] [Google Scholar]
- 29. Morera J, Joubert M, Morello R, et al. : Sustained efficacy of insulin pump therapy in type 2 diabetes: 9-year follow-up in a cohort of 161 patients. Diabetes Care 2016;39:e74–e75. [DOI] [PubMed] [Google Scholar]
- 30. American Diabetes Association Professional Practice Committee: 13. Older Adults: Standards of Medical Care in Diabetes—2022. Diabetes Care 2021;45(Suppl. 1):S195–S207. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Ibrahim M, Baker J, Cahn A, et al. : Hypoglycaemia and its management in primary care setting. Diabetes Metab Res Rev 2020;36:e3332. [DOI] [PubMed] [Google Scholar]
- 32. Misra-Hebert AD, Pantalone KM, Ji X, et al. : Patient characteristics associated with severe hypoglycemia in a type 2 diabetes cohort in a large, integrated health care system from 2006 to 2015. Diabetes Care 2018;41:1164–1171. [DOI] [PubMed] [Google Scholar]
- 33. Polonsky WH, Fisher L, Hessler D, Edelman SV: Identifying the worries and concerns about hypoglycemia in adults with type 2 diabetes. J Diabetes Complicat 2015;29:1171–1176. [DOI] [PubMed] [Google Scholar]