Correction to: Quantitative analysis of phantom studies of 111In and 68Ga imaging of neuroendocrine tumours

L Jönsson 1,, A Stenvall 1, E Mattsson 2, E Larsson 3, A Sundlöv 4, T Ohlsson 3, C Hindorf 3
PMCID: PMC9622970  PMID: 36315296

Correction to: EJNMMI Phys 5, 5 (2018). https://doi.org/10.1186/s40658-018-0204-0

Following publication of the original article [1], it was reported that Eq. 1 was erroneously published as:

Tadapt=T×ACmax-BG+BG 1

This equation should read as follows:

Tadapt=T×ACmax-BG+BGACmax 1

The numerical values presented in the paper are calculated according to this corrected equation.