Appraisal and validation of a method used for detecting heavy metals in poultry feed in Bangladesh

Md Mosharaf Hossain 1, Abu Sayeed Md Abdul Hannan 1, Md Mostofa Kamal 1, Mohammad Abul Hossain 2,, Shamshad B Quraishi 3
PMCID: PMC9631369  PMID: 36341057

Abstract

Background and Aim:

Low concentrations of heavy metals are toxic and pose a serious threat to human health and the environment. Therefore, regular assessments of the toxic metal content in poultry feed are crucial for evaluating feed quality and customer safety. It is difficult to determine the heavy metals in the poultry feed at the trace amount. Therefore, this study aimed to validate this method through the detection of three heavy metals, chromium (Cr), cadmium (Cd), and lead (Pb), in poultry feed samples.

Materials and Methods:

Graphite furnace atomic absorption spectrometry (GF-AAS) method was used to analyze the heavy metals in poultry feed according to the guidelines given by the Council Directive 333/2007/EC, Commission Decision 657/2002/EC. In this study, various parameters such as linearity check, limit of detection (LOD), limit of quantification (LOQ), recovery percentage, precision checks, repeatability, reproducibility, and uncertainty measurement were considered to validate and assess the method following international guidelines. Heavy metals, such as Pb, Cr, and Cd, were analyzed from the feed samples in the laboratory using the GF-AAS method (Model: AA-7000 Shimadzu, Japan) with high purity argon as the inert gas, and the absorbance was read at wavelengths of 283.0, 357.9, and 228.8 nm, respectively.

Results:

The coefficient of variation (CV%) for system suitability and precision data was <10% for all the metals (Pb, Cr, and Cd) detected in this study. The overall CV% of repeatability and reproducibility ranged from 8.70% to 8.76% and 8.65% to 9.96%, respectively. The linearity of the calibration curves was excellent (r2 > 0.999) at various concentration levels for the three different metals. The recovery (%) was found to be 94.53, 93.97, and 101.63% for Pb, Cr, and Cd, respectively. The LOD values in feed were 0.065, 0.01, and 0.11 mg/kg, and the LOQ values were 0.22, 0.03, and 0.38 mg/kg for Cr, Cd, and Pb, respectively. The values recorded for the measurement uncertainty (%) were 11.48, 4.43, and 12.42% for Cr, Cd, and Pb, respectively.

Conclusion:

The results show that these study criteria or parameters have met the validated or acceptable range. Therefore, it is a reliable technique that can be used undoubtedly for the routine analysis of heavy metals in poultry feed samples across the globe.

Keywords: graphite furnace atomic absorption spectrometry, method validation, poultry feed, toxic metals (Pb, Cr, and Cd)

Introduction

Bangladesh is an agro-based developing country. Bangladesh’s small- and large-scale chicken farms are fast growing and supply meat, eggs, and employment opportunities. More than 50% of this country’s population is involved in agricultural and livestock farming. In a developing nation like Bangladesh, the poultry industry provides both direct and indirect employment opportunities and is a vital part of agricultural farming systems [1]. Both developed and developing countries face an alarming problem of environmental pollution with heavy metals and metalloids [2]. Heavy metals are important and have been extensively studied for their detrimental effects and bioaccumulation in food chains or feeding systems [3]. Feeding poultry requires many different feedstuffs to balance the ration. Most of these feedstuffs are mainly derived from plants grown in the soil. The feedstuffs might be adulterated and contaminated differently from growing to supplying the birds. For example, fish meal collected from contaminated water can transfer heavy metals to poultry feed, as we know that the poultry diet is composed of various raw materials [4]. Feeding feed or food is a very important element for all living beings because it contains all essential nutrients for the proper nourishment of the body. All categories of people use poultry meat and egg as one of the best food items, growing rapidly worldwide. It is obvious that the rapid growth of poultry depends on the feeds and that feeds contain different pollutants, including heavy metals. These metals spread widely and release toxic substances into the environment due to natural and anthropogenic activities [5]. As a result, environmental pollution is increasing alarmingly to heavy metal toxicity, posing a great threat to human health [6]. Some heavy metals, notably cadmium (Cd), chromium (Cr), lead (Pb), and arsenic (As), are reported detrimental at low concentrations. According to the United States Environmental Protection Agency [7], metals (e.g., Cr, As, Cd, and Pb) are considered the most toxic elements for the environment. These are commonly available heavy metals that remain intact, are not degraded easily, and cause environmental and food contamination. The metals, in turn, trigger many acute or chronic disorders in humans due to their gradual accumulation in various animal body organs for a longer period [8].

In view of the above, it is obvious that poultry feeds could contain toxic heavy metals at a sufficient level and their exposure might occur after the birds consume these feed materials. The data on metal contamination or detection in poultry feedstuffs are insufficient in our country. Although there are many methods for quantifying heavy metals, further analytical laboratory results may be necessary for official control and quality assurance. For this reason, global demand is growing to develop and establish new modern laboratory techniques for providing accurate and reliable data or measurements with low concentration levels [9]. According to Food and Agricultural Organization [10], sustainable agricultural development relies on feed quality, which is considered one of Bangladesh’s main barriers to improving livestock.

For this reason, a method of graphite furnace atomic absorption spectrometry (GF-AAS) has been developed to detect toxic elements, mainly Pb, Cd, and Cr, in poultry feed samples. This method appears reliable, precise, accurate, sensitive, selective, and economical [11]. The poultry feed quality can be assessed regularly if the method is implemented and validated globally. Hence, the developed method will help to identify the toxic elements, which has a great application in laboratory examination of poultry feed samples [12]. In addition, this study would greatly benefit the poultry industry, feed miller companies, and poultry integrators.

There are several other analytical techniques available to detect toxic metals reported by previous researchers; these include flame AAS, GF-AAS, and inductively coupled plasma mass spectrometry, which are being used for the quantification of toxic metals in feedstuffs [1315]. This study aimed to analyze the toxic metals (e.g., Pb, Cd, and Cr) in the poultry feed usually manufactured in the different feed mills of Bangladesh.

Materials and Methods

Ethical approval

No ethical approval is needed for conducting this sort of study, as it did not deal with any animals or living beings.

Study period and location

The study was conducted from November 2020 to October 2021 at the Quality Control Laboratory, Department of Livestock Services, Savar, Dhaka, Bangladesh.

Tools, chemicals, and reagents

An atomic absorption spectrophotometer (Model: AA-7000, Shimadzu, Japan) equipped with a GF (GFA 7000) and an autosampler (ASC 7000) was used. A cathode lamp was used and operated according to the manufacturer’s instructions for the analyses of Pb (283.0 nm, 10 mA, and slit 0.7 nm), Cd (228.8 nm, 10 mA, and slit 0.7 nm), and Cr (357.9 nm, 10 mA, and slit 0.7 nm), respectively. Nitric acid (HNO3, 69%), hydrogen peroxide (H2O2, 30%), and standards of Pb, Cd, and Cr were used for sample preparation. A microwave acid digestion system (MADS) (Ethos Easy Milestone) was used for conducting the digestion process. The standard bulk solutions were made daily by diluting respective metals stock standard solutions using 1% (w/w) (HNO3, 69%). The autosampler was used for preparing standard solutions. Deionized water (18 MW/cm) produced using an E-pure system (Thermo Scientific, United States) was used to prepare all the solutions and clean and wash all containers and glassware before use.

Sample preparation and analytical procedure

Poultry feed was prepared with the conventional feed ingredients locally available in Bangladesh following the standard prescribed by the respective breeder company or National Research Council [16], as shown in Table-1. The samples were then dried and ground by mortar and pestle and put in an air sealable bag before laboratory analysis. For the digestion of samples, approximately 0.5 g of feed sample was weighed and digested with 8 mL of HNO3 (69%) and 2 mL of H2O2 (30%) in acid-prewashed Teflon vessels. A MADS (Ethos Easy Milestone) was used for the digestion procedure. After digestion, the sample was diluted with 50 mL final volume with deionized water. The reference material analytical blanks were also prepared with each batch of digestion sets. All samples were prepared in triplicate. Diluted samples and the standard solution were separately put into a set of fresh tubes for atomic absorption spectrophotometer (AAS Shimadzu AA-7000). Final heavy metals, Pb, Cr, and Cd, were measured at 283.0, 357.9, and 228.8 nm, respectively.

Table-1.

Composition of broiler grower feed.

Ingredients Amount (%)
Corn/maize 59.00
Wheat 3.32
Soybean meal (46%) 23.00
Corn gluten (60%) 6.40
Palm oil 4.20
Dicalcium phosphate 1.70
Limestone 1.40
Table salt (NaCl) 0.25
Choline chloride 0.06
Vitamin mineral premix 0.15
L-lysine 0.30
DL-methionine 0.22
Nutrient
 ME 3200 Kcal/kg
 CP 20.0%

DL = Dextrorotation and levorotation, ME = Metabolizable energy, CP = Crude protein.

Preparation of spiking solution for Cd, Cr, and Pb

About 0.5 g (500 mg) sample was obtained from the digestion vessel, and 8 mL HNO3 (69%), 2 mL H2O2 (30%), and 0.5 mL (500 uL) of each 1000 ppb of Cd, Cr, and Pb solutions were added before the solution was mixed and left to digest. Finally, the sample was prepared with water up to 50 mL, and 5 mL of this solution was then diluted to 10 mL with a solution of 1% nitric acid.

Criteria for the validation of the proposed method

The parameters or criteria for the validation of the proposed method in this study followed were linear range, limits for detection and quantification, accuracy percentage, precision checks, and degree of uncertainty measurement [1719]. The validation criteria were assessed following suggested instructions or guidelines of the International Conference on Harmonization [20] and the United States Food and Drug Administration [21].

Linearity

Standard mixtures of Pb, Cr, and Cd were prepared and a linear equation was established for each metal by plotting the absorbance versus the concentrations to measure linearity. Three calibration curves were obtained on 3 consecutive days with a specified standard concentration of each metal. Linearity was calculated by running aqueous standard solution of each metal at final concentrations of 0.25, 0.50, 1.0, 2.0, and 4 μg/mL for Cd; 2, 4, 8, and 16 μg/mL for Cr; and 2, 5, 10, 15, 20, and 40 μg/mL for Pb. The slope, intercept, and regression coefficient (r2) values were retrieved from the linear regression and correlation method.

Recovery percentage

Three sets of spike samples (2.5, 5.0, and 7.5 mg/kg) were prepared, and each set was replicated seven times. Sample reading was taken by measuring two times. For the recovery estimation, an accurate amount of each metal’s three concentration levels (2.5, 5.0, and 7.5 mg/kg for Cr and Pb and 0.25, 0.5, and 0.75 mg/kg for Cd) was added to approximately 1.0 g of blank matrix powder. The powder was then extracted and analyzed for recovery using the formula: Recovery (%) = (amount obtained/amount spiked) × 100.

Limit of detection (LOD), limit of quantification (LOQ), instrument quantification limit (IQL), and instrument detection limit (IDL)

The lowest qualitative and quantitative concentrations for the tested linearity range were calculated for each metal according to the guidelines of ICH2000. LOD and LOQ were calculated using the expression: k × S.D/b, where k = 3.3 for the LOD and 10 for the LOQ, SD is the standard deviation of the intercept, and b is the slope of the calibration curve tested for linearity. IDL and IQL are calculated using following formula: IDL = 3s and IQL = 10s; where, s = standards deviation.

Repeatability and reproducibility precision check

The precision of the method was evaluated based on repeatability and intermediate precision. The repeatability was calculated on the results obtained from the same day for seven independent mixer solutions of the variable concentration, and the intermediate precision was evaluated by calculating the repeatability of the similar concentration by two analysts on different days. The percentage of relative standard deviation was calculated to estimate the precision of this study. For reproducibility precision check, three sets of samples were prepared again after 15 days by the same analyst and after 18 days with different analysts spiking at 0.25, 0.5, and 0.75 and 2.5, 5.0, and 7.5 level and analyzed as before. The mean concentration, standard deviation, and coefficient of variation (CV%) of each level of fortified samples have been calculated.

Statistical analysis

Raw data were inserted into a Microsoft Excel (Microsoft Office, USA) sheet for statistical analysis. The data were statistically analyzed using a Minitab statistical software version 16 [22]. The means and standard deviations of the metal concentrations in samples were calculated. Descriptive analysis was performed using different variables’ percentages, the mean, and standard deviation. Finally, a one-way analysis of variance was used to compare the level of heavy metal residues in poultry feed.

Results

Linearity

Table-2 shows the linear regression data of heavy metals (Pb, Cr, and Cd) found in poultry feed samples. It is obvious from the data that the values of the r2 were 0.999, 0.9996, and 0.9998 for Pb, Cd, and Cr, respectively.

Table-2.

Linear regression data of heavy metals in poultry feed.

Metal Linear range (μg/L) Calibration with an aqueous standard solution

Slope±SD (µg/L) Intercept±SD Regression coefficient (r2)
Pb 2–40 0.0101±0.0006 0.0193±0.0125 0.9996
Cr 2–16 0.0309±0.0018 0.1396±0.0559 0.9998
Cd 0.25–4.0 0.2332±0.0055 0.5036±0.0911 0.9996

SD=Standard deviation, Cr=Chromium, Cd=Cadmium, Pb=Lead

Accuracy or recovery (%) of heavy metals in poultry feed

The overall recovery % for Pb, Cr, and Cd in poultry feed samples was 94.63%, 93.97%, and 101.63%, respectively (Table-3).

Table-3.

The recovery (%) of heavy metals in poultry feed.

Metal Spiked analyte concentration (mg/L) Calculated analyte concentration (n = 7) (mg/kg) Recovery (%) (n = 7) Overall recovery % (n = 21)
Pb 2.5 2.17 86.97 94.63
5 4.6 91.90
7.5 7.87 104.93
Cr 2.5 2.48 102.01 93.97
5 4.50 89.92
7.5 6.97 92.88
Cd 0.25 0.27 109.64 101.63
0.50 0.44 88.84
0.75 0.80 106.42

Cr=Chromium, Cd=Cadmium, Pb=Lead

Determination of IDL, IQL, LOD, and LOQ of toxic metals in poultry feed

The values of IDL for Cr, Cd, and Pb were 0.65, 0.09, and 1.10 mg/L, whereas the values of IQL were 2.16, 0.29, and 3.68 mg/L, respectively. The results of method LOD were 0.065, 0.01, and 0.11 mg/kg, and the values for LOQ were 0.22, 0.03, and 0.38 mg/kg in the metals of Cr, Cd, and Pb, respectively (Table-4).

Table-4.

The results of IDL, IQL, and method LoD and LoQ of the heavy metals in poultry feed.

Parameter Heavy metal Average

Cr Cd Pb
Instrumental IDL (μg/L) 0.65 0.09 1.10 0.61
Instrumental IQL (μg/L) 2.16 0.29 3.68 2.04
Method LoD (μg/kg) 0.065 0.01 0.11 0.061
Method LoQ (μg/kg) 0.22 0.03 0.38 0.21

IDL=Instrument detection limit, IQL=Instrument quantification limit, LoD=Limit of detection, LoQ=Limit of quantification, Cr=Chromium, Cd=Cadmium, Pb=Lead

Repeatability and reproducibility data of heavy metals in poultry feed

The precision studies are performed by measuring the repeatability and reproducibility data, as shown in Tables-5 and 6. The data showed that the values of repeatability of CV% were 8.70%, 8.76%, and 8.75% for the Pb, Cr, and Cd, respectively (Table-5). The reproducibility data of CV% were 8.86%, 9.96%, and 8.65% for Pb, Cr, and Cd, respectively (Table-6).

Table-5.

The repeatability data of heavy metals in poultry feed.

Metal Day Spike concentration (mg/kg) Overall mean concentration (n = 21) (mg/kg) SD CV% Overall CV (%)
Pb 1–3 2.50 2.33 0.21 9.07 8.70
1–3 5.00 4.78 0.46 9.61
1–3 7.50 7.22 0.54 7.42
Cr 1–3 2.50 2.58 0.32 12.39 8.76
1–3 5.00 4.58 0.42 9.27
1–3 7.50 6.91 0.32 4.61
Cd 1–3 0.25 0.26 0.02 8.70 8.75
1–3 0.50 0.46 0.05 10.96
1–3 0.75 0.78 0.05 6.62

SD=Standard deviation, CV%=Coefficient of variation, Cr=Chromium, Cd=Cadmium, Pb=Lead

Table-6.

The reproducibility data of heavy metals in poultry feed.

Metal Day Spike concentration (mg/kg) Overall mean concentration (n = 21) (mg/kg) SD CV% Overall CV%
Pb 1–3 2.50 2.48 0.26 10.45 8.86
1–3 5.00 4.81 0.45 9.27
1–3 7.50 7.57 0.52 6.85
Cr 1–3 2.50 2.54 0.32 9.83 9.96
1–3 5.00 4.63 0.48 9.04
1–3 7.50 6.96 0.84 11.01
Cd 1–3 0.25 0.26 0.02 7.52 8.65
1–3 0.50 0.49 0.06 11.37
1–3 0.75 0.79 0.06 7.07

SD=Standard deviation, CV%=Coefficient of variation, Cr=Chromium, Cd=Cadmium, Pb=Lead

Measurement of uncertainty (%) of heavy metals in the poultry feed

The values of measuring uncertainty (MU) for Pb, Cr, and Cd were 12.42%, 11.48%, and 4.43%, respectively, which were shown through a graph (Figure-1). The highest uncertainty (12.42%) was observed for Pb, whereas the lowest value (4.43%) was found in Cd. Lower uncertainty values characterize higher accuracy.

Figure-1.

Figure-1

Uncertainty percentage (mean ± SE) of measured heavy metals in poultry feed.

Discussion

The linearity data obtained from this study showed that it had met the acceptance level, that is, r2 = 0.995. The data indicate that the r2 value (0.999) found in this study is greater than the acceptable level (0.995), and it implies that the result is very good as the linear relationship found in this study is precise and authentic [23]. Our result is consistent with the report of previous investigators who found similar results for the validation of heavy metals in milk and fish feed samples [9, 24].

The overall recovery % was 92.63–101.63% in the detected metals in this study. The acceptable level for the recovery % ranges from –20% to +10%. Based on this fact, we can say that our analytical value of recovery % has met the standard level or criteria. Recovery (%) refers to how much extraction can be obtained by the analytical process while conducting sample testing in the laboratory [21]. Although no verified data or reference materials for poultry feed samples are available, the recovery % was done to justify the method’s accuracy [23].

LOD and LOQ are measured to detect the performance or efficacy of an instrument or an analytical method. Our results clearly show that the LODs and LOQs for the identified metals in feed samples were determined to be lower than the highest range as given by Regulations (EC) No. 1275/[25], and it has matched well according to Regulation 333/2007/EC [17].

The CV% of repeatability and reproducibility data was <10% in our analytical values of the study. The acceptable value for these parameters is considered 10% [26]. Therefore, it can be stated that the method exhibited good repeatability and reproducibility precision based on the values obtained in this study. The analytical precision was established in accordance with EC commission regulation [27]. CV% normally assesses the precision, which denotes the proximity of compromising data [28]. Measuring uncertainty (MU) calculation is necessary for every measurement according to ISO/IEC 17025 [18] because it checks the errors and omissions that occurred while conducting any assay. The reproducibility data can be used to estimate MU [29, 30]. However, despite the uncertainty or limitations, the method we developed here for detecting toxic metals in the feed samples is very simple, precise, accurate, reliable, and cost-effective as well. Thus, the proposed method can be used for routine analyses or simultaneous detection and quantification of heavy metals, which has a great implication for laboratory examination of poultry feed samples across the globe [12]. Our results are also consistent with the previous investigators. They stated similar types of remarks regarding method validation for the fish feed [9], milk [24], and detection and assessment of heavy metals in poultry foodstuffs [12, 3135].

Conclusion

It can be concluded that all the criteria for method validation (e.g., linearity, precision, recovery percentage, LOD and LOQ, repeatability, reproducibility, and uncertainty measurement) had been satisfied. We think that our developed method to detect heavy metals from poultry feed is very simple, easy to conduct, reliable, and cost-effective compared to other methods.

Authors’ Contributions

MMK and SBQ: Planned and designed the study. MMH and ASMAH: Carried out the sample preparation and laboratory analysis. MMH and MAH: Statistical analyses, tabulation, interpretation, compilation, revision, edition, validation, and manuscript preparation. All authors have read and approved the final manuscript.

Acknowledgments

The study was funded by the Quality Control laboratory for livestock inputs and food products establishment project (No. 224071600), Department of Livestock Services , Savar, Dhaka, Bangladesh.

Competing Interests

The authors declare that they have no competing interests.

Publisher’s Note

Veterinary World remains neutral with regard to jurisdictional claims in published institutional affiliation.

References

  • 1.Hamid M.A, Rahman M.A, Ahmed S, Hossain K.M. Status of poultry industry in Bangladesh and the role of private sector for its development. Asian J. Poult. Sci. 2017;11(1):1–13. [Google Scholar]
  • 2.Ullah A.K.M, Maksud M.A, Khan S.R, Shamshad L.N, Quraishi B. Dietary intake of heavy metals from eight highly consumed species of cultured fish and possible human health risk implications in Bangladesh. Toxicol dRep. 2017a;4:574–579. doi: 10.1016/j.toxrep.2017.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Ali H, Khan E. Trophic transfer, bioaccumulation, and bio-magnification of non-essential hazardous heavy metals and metalloids in food chains/webs concepts and implications for wildlife and human health. Hum. Ecol. Risk Assess. 2019;25(6):1353–1376. [Google Scholar]
  • 4.Sadeghi A, Hashemi M, Jamali-Behnam F, Zohani A, Esmaily H, Dehghan A.A. Determination of chromium, lead and cadmium levels in edible organs of marketed chickens in Mashhad. Iran. J. Food Qual. Hazards Contrl. 2015;2:134–138. [Google Scholar]
  • 5.Dahri S, Pathan A.A, Balouch A, Mallah S. Assessment of Heavy Metals in Poultry Feed of Hyderabad, Sindh. 2nd International Conference on Sustainable Development in Civil Engineering, MUET, Pakistan, December. 2019;05–07:284–286. [Google Scholar]
  • 6.Ali H, Khan F, Ilahi I. Environmental chemistry and ecotoxicology of hazardous heavy metals:Environmental persistence, toxicity, and bioaccumulation. J. Chem. 2019;2019:6730305. [Google Scholar]
  • 7.Lei M, Zhang Y, Khan S, Qin P, Liao B. Pollution, fractionation and mobility of Pb, Cd, Cu, and Zn in garden and paddy soils from a Pb/Zn mining area. Environ. Monit. Assess. 2010;168((1–4)):215–222. doi: 10.1007/s10661-009-1105-4. [DOI] [PubMed] [Google Scholar]
  • 8.Saei-Dehkordi S.S, Fallah A.A. Determination of copper, lead, cadmium and zinc content in commercially valuable fish species from the Persian gulf using derivative potentiometric stripping analysis. Microchem. J. 2011;98(1):156–162. [Google Scholar]
  • 9.Ullah A.K.M, Maksud M.A, Khan S.R, Shamshad L.N, Quraishi B. Development and validation of a GF-AAS method and its application for the trace level determination of Pb Cd, and Cr in fish feed samples commonly used in the hatcheries of Bangladesh. J Anal. Sci. Technol. 2017b;8(4):15. [Google Scholar]
  • 10.Food and Agricultural Organization. Fisheries and Aquaculture Department, Fishery and Aquaculture Country Profiles, The People's Republic of Bangladesh. Rome, Italy: Food and Agricultural Organization; 2014. [Google Scholar]
  • 11.Zhong W.S, Ren T, Zhao L.J. Determination of Pb (lead), Cd (cadmium), Cr (chromium), Cu (copper), and Ni (nickel) in Chinese tea with high-resolution continuum source graphite furnace atomic absorption spectrometry. J. Food Drug Anal. 2016;24(1):46–55. doi: 10.1016/j.jfda.2015.04.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Hosain M.Z, Islam S.M.S, Kamal M.M. Development of a rapid and reliable high-performance liquid chromatography method for determination of water-soluble vitamins in veterinary feed premix. Vet. World. 2021;14(12):3084–3090. doi: 10.14202/vetworld.2021.3084-3090. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Akter M, Sikder T, Ullah A.K.M. Water quality assessment of an industrial zone polluted aquatic body in Dhaka, Bangladesh. Am. J. Environ. Protect. 2014;3(5):232–237. [Google Scholar]
  • 14.Ullah A.K.M, Khalil M.I, Kibria A.K.M. Determination of uranium in stream water, beach sand and rocks by stripping voltammetry following adsorptive accumulation of the U(VI)-chloranilic acid complex. Int. Res. J. Eng. Technol. 2015;2(8):54–63. [Google Scholar]
  • 15.Hwang B.G, Lee S.J, Gil K. In-vivo analysis of the uptake process of heavy metals through maize roots by using synchrotron X-ray fluorescence spectroscopy. J. Korean Phys. Soc. 2016;6(12):1824–1829. [Google Scholar]
  • 16.National Research Council. Nutrient Requirements of Poultry. 9th ed. Washington DC: National Academic Press; 1994. [Google Scholar]
  • 17.European Commission. Regulation (EC) No 333/2007 of 28 March 2007 Laying down the Methods of Sampling and Analysis for the Official Control of the Levels Of Lead, Cadmium, Mercury, Inorganic tin, 3-MCPD and benzo(a) Pyrene in Foodstuffs. Brussels: European Commission; 2007. pL88/29–L88/38. [Google Scholar]
  • 18.ISO/IEC 17025. General Requirements for the Competence of Testing and Calibration Laboratories. Geneva, Switzerland: International Organization for Standardization; 2005. [Google Scholar]
  • 19.Eurachem. The Fitness for the Purpose of Analytical Methods. Teddington: LGC; 2012. 1998. EURACHEM/CITAC. Quantifying Uncertainty in Analytical Measurement. LGC, United Kingdom. [Google Scholar]
  • 20.Ravisankar P, Navya C.N, Pravallika D, Sri N. A review on step-by-step analytical method validation. IOSR J. Pharm. 2015;5(10):7–19. [Google Scholar]
  • 21.Food and Drug Administration. Bioanalytical Methods Validation guidance for Industry, US Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER), Center for Veterinary Medicine (CVM) Maryland: Food and Drug Administration, Silver Spring; 2018. pp. 1–41. [Google Scholar]
  • 22.Minitab. Minitab Statistical Software User Guide 2:Data Analysis and Quality Tools. USA: Minitab Inc., State College, PA; 2000. [Google Scholar]
  • 23.Ertasa O.S, Tezel H. A validated cold vapor AAS method for determining cadmium in human red blood cells. J. Pharm. Biomed. Anal. 2005;36(4):893–897. doi: 10.1016/j.jpba.2004.08.014. [DOI] [PubMed] [Google Scholar]
  • 24.Hossain M.M, Hannan A.S.M, Kamal M.M, Hossain M.A, Zaman S. Development and ratification of a precise method (GF-AAS) used for the determination of poisonous metal lead (Pb) in dairy cow milk sample commonly available in the market of Bangladesh. Austin. J. Anal. Pharm. Chem. 2022;9(1):1142. [Google Scholar]
  • 25.European Commission. Regulation (EC) No 1275/2013 of 6 December 2013 amending Regulation (EC) No. 2002/32/EC on Undesirable Substance in Animal Feed. 2013. Brussels: European Commission; 2013. L328/86–L328/92. [Google Scholar]
  • 26.European Commission. Decision of 12 August 2002 Implementing Council Directive 96/23/EC Concerning the Performance of Analytical Methods and the Interpretation of Results No. 657/EC. Brussels: European Commission; 2002a. pL221/8–L221/36. [Google Scholar]
  • 27.European Commission. Directive 2002/32/EC of the European Parliament and of the Council of 7 May 2002 on Undesirable Substances in Animal Feed. Brussels: European Commission; 2002b. pL140/10–L140-21. [Google Scholar]
  • 28.Jeevanaraj P, Hashim Z, Elias S.M, Aris A.Z. Total mercury (THg), lead (Pb), cadmium (Cd) and arsenic (As) in hair samples:Method validation and quantification among women at reproductive age in Selangor. Int. J. Sci. Basic Appl. Res. 2015;24(4):332–347. [Google Scholar]
  • 29.Nordtest Report TR537. Hand Book for Calculation of Measurement Uncertainty in Environmental Laboratories Nordtest Report TR537-2. (2nd ed) 2004 [Google Scholar]
  • 30.Gonzaler A.G, Herrador M.A. A practical guide to analytical method validation, including measurement uncertainty and accuracy profiles. Trends Anal. Chem. 2007;26(3):227–238. Huber L. (1998). Validation and Quantification in Analytical Laboratories. Interpharm Press, East Englewood. [Google Scholar]
  • 31.Rawar E.A, Rohman A. Validation of atomic absorption spectrophotometry and mercury analyzer for analysis of cadmium, lead and mercury in red chili pepper. Asian J. Agric. Res. 2016;10(1):65–71. [Google Scholar]
  • 32.Shakeel M, Adnan A, Durham J. Heavy metals in poultry products in Bangladesh:A possible death threat to future generations. J. Social Political Sci. 2019;2(1):98–105. [Google Scholar]
  • 33.Mathaiyan M, Natarajan A, Rajarathinam X, Rajeshkumar S. Assessment of Pb, Cd, As and Hg concentration in edible parts of broiler in major metropolitan cities of Tamil Nadu, India. Toxicol. Rep. 2021;8:668–675. doi: 10.1016/j.toxrep.2021.03.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Rashid M.A, Sarker M.S.K, Khatun H, Sarker N.R, Ali M.D.Y, Islam M.N. Detection of heavy metals in poultry feed, meat and eggs. Asian Australas. J. Food Saf. Secur. 2018;2(1):1–5. [Google Scholar]
  • 35.Korish M.M, Attia Y.A. Evaluation of heavy metal content in feed, litter, meat, meat products, liver, and table eggs of chickens. Animals (Basel) 2020;10(4):727. doi: 10.3390/ani10040727. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Veterinary World are provided here courtesy of Veterinary World

RESOURCES