Monobloc press-fit cups with large-diameter bearings are safe in revision total hip arthroplasty

Paul-André Synnott 1, Sivan Sivaloganathan 2, Marc-Olivier Kiss 3, Benoit Binette 2, Mina W Morcos 2, Pascal-André Vendittoli 4,
PMCID: PMC9635988  PMID: 36349354

Abstract

BACKGROUND

Instability is a major cause of failure of revision total hip arthroplasty (THA) which can be avoided with the use of monobloc press-fit cups with large diameter heads (LDH).

OBJECTIVE

This consecutive case series analyses whether LDH monobloc components are a safe and clinically beneficial option for revision THA.

METHODS

This consecutive case series includes 47 revision THA with LDH monobloc acetabular cup. Acetabular bone defects were Paprosky type I (42), type IIA (2) and type IIC (3). The Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Forgotten Joint Score (FJS) and the Patient’s Joint Perception (PJP) scores were analysed and a radiographic evaluation for signs of implant dysfunction was performed.

RESULTS

After a mean follow-up of 4.5 years, there were 5 (10.6%) acetabular cup re-revisions: 2 loss of primary fixation and 3 instabilities. Implant survivorship at 4-years was 89.4% (95% CI: 89.3 to 89.5). Recurrent hip dislocation was reported in 1 patient (2.1%) and remains under conservative treatment. The mean WOMAC and FJS were 19.5 (16.8; 0.0 to 58.3) and 57.3 (28.9; 6.3 to 100.0), respectively. Regarding the PJP, 3 (8.8%) patients perceived their hip as natural, 8 (23.5%) as an artificial joint with no restriction, 14 (41.2%) with minor restriction and 9 (26.5%) with major restriction.

CONCLUSION

In cases of revision THA with limited bone loss, press-fit primary fixation with LDH monobloc acetabular components are valuable options that offer excellent mid-term out-comes with low re-revision and joint instability rates.

Keywords: Total Hip Replacement, Revision Surgery, Hip Prosthesis, Acetabulum

INTRODUCTION

Revision total hip arthroplasty (THA) is a challenging procedure associated with inferior outcomes when compared with primary THA.1,2 Following revision THA, the most common causes of failure are aseptic loosening and instability.3 These complications are likely to be more frequent as the number of revision THA increases over the next decade.4 Large diameter head (LDH) prostheses offer a valuable option for tackling the risk of instability as they have been demonstrated to decrease the incidence of instability after revision THA.5

LDH THA can be defined as femoral head diameter ≥36 mm or wall thickness of the acetabular cup ≤12mm. Different designs of LDH THA exist, such as LDH ceramic-on-ceramic (CoC), LDH metal-on-metal (MoM) and dual mobility (DM) articulation. To maximise the bearing diameter, the wall thickness of the acetabular cup is kept minimal. Therefore, acetabular components of LDH THA are offered as a monobloc design or factory preassembled for CoC LDH. The primary stability of these components relies only on press-fit, as no supplementary screw fixation is possible.

The primary objective of this study is to assess the mid-term aseptic loosening rate of LDH monobloc press-fit components following revision THA. Secondary objectives are to evaluate complications by cause, assess for radiological evidence of implant failure, and measure patient-reported outcomes (PROMs). The hypothesis of this study is that LDH monobloc components implanted during revision THA offer satisfactory survival rate and clinical outcomes.

MATERIALS AND METHODS

For this consecutive single-centre case series, the local arthroplasty database was used to retrospectively identify revision THA using press-fit only LDH monobloc acetabular components between 2007 and 2019. Inclusion criteria were a minimum follow-up of 2 years after revision THA and the availability of preoperative, postoperative, and last follow-up hip radiographs (Figure 1). Patients’ characteristics are summarized in Table 1.

Figure 1. Study flow diagram.

Figure 1.

Table 1. Patient characteristics for revision THA using monobloc press-fit cups with large diameter bearings.

Criteria
Hips (patients), n 47 (44)
Mean age at surgery, years (SD; range) 58.8 (10.8; 30.0–79.0)
Gender, n
Female 25 (53.2%)
Male 22 (46.8%)
Side, n
Right 27 (57.4%)
Left 20 (42.6%)
Mean body mass index, kg/m2 (SD; range) 27.8 (4.9; 20.2-44.9)
Cause of revision, n
ARMD 29 (61.7%)
Infection 4 (8.5%)
Infection and ARMD 1 (2.1%)
Instability 4 (8.5%)
Aseptic loosening 4 (8.5%)
Residual pain 3 (6.4%)
Liner fracture 2 (4.3%)
Acetabular defect – Paprosky classification, n
Type 1 42 (89.4%)
Type 2A 2 (4.2%)
Type 2B 0 (0.0%)
Type 2C 3 (6.4%)
Type 3 0 (0.0%)
Pre-revision mean acetabular cup abduction angle, ° (SD; range) 45.4 (6.6; 29.4-60.0)
Pre-revision acetabular protrusion, n 5 (10.6%)
Pre-revision heterotopic ossification, n
Brooker Class I 5 (10.6%)
Brooker Class II 1 (2.1%)
Brooker Class III 1 (2.1%)
Pre-revision mean acetabular cup diameter, mm (SD; range) 53.2 (4.8; 41.0-64.0)
Prosthesis, n
CoC LDH
Maxera (Zimmer Biomet, Warsaw, Indiana) 38 (80.8%)
MoM LDH
Durom Metasul (Zimmer Biomet, Warsaw, Indiana) 3 (6.4%)
DM
Polarcup (Smith & Nephew, Memphis, Tennessee) 6 (12.8%)
Mean acetabular cup diameter, mm (SD; range) 56.6 (4.4; 46.0-66.0)
Mean femoral head diameter, mm (SD; range) 44.9 (3.5; 36.0-54.0)
Mean follow-up, years (SD; range) 4.5 (2.7; 2.0-13.7)

SD, standard deviation; ARMD, adverse reaction to metal debris; CoC, ceramic-on-ceramic; MoM, metal-on-metal; LDH, large diameter head; DM, dual mobility

Medical files were reviewed to gather perioperative data and identify complications during follow-up. Acetabular bone defects during revision THA were described according to the Paprosky classification.6 Revision failures for any cause were recorded. A radiographic evaluation on postoperative and last follow-up radiographs was conducted to identify signs of aseptic loosening of the acetabular component. Such signs included the presence of periacetabular radiolucent lines described with de DeLee and Charnley classification7 of more than 2 mm or acetabular component migration assessed using the technique described by Massin et al.8 Significant horizontal or vertical migration was defined as >3 mm,9 and significant variation of the acetabular inclination was defined as >5°.8 Post-revision radiographs were also assessed for heterotopic ossification described according to the Brooker classification,10 the presence of acetabular protrusion11 and superolateral acetabular uncoverage estimated by the technique described by Li et al.12 Functional outcomes were assessed at the last follow-up with the Western Ontario McMaster Universities Osteoarthritis Index (WOMAC, the best score is 0),13 the Forgotten Joint Score (FJS, the best score is 100),14 and the Patient’s Joint Perception question (PJP).15

Ethical Considerations

The study was conducted in accordance with the Declaration of Helsinki. The patient’s written consent was obtained, and ethical approval was received by the Institutional Ethics Committee of CIUSSS de l’Est-de-l’Île-de-Montréal (2021-2469; 01-2021).

Statistics

Continuous variables were described using means, standard deviations, and ranges. Categorical variables were defined using absolute and relative frequencies. Implant survivorship analyses were conducted using the Kaplan-Meier method and reported with a 95% confidence interval (CI), with acetabular component re-revision as the endpoint.16 Statistical analysis was performed using SPSS version 27 (IBM Corporation, Armonk, NY, USA).

Surgical technique

The procedures were performed by five experienced arthroplasty surgeons in our academic institutions. All revision THAs were performed using a posterior approach. The revision was limited to the acetabular component in 26 hips (55%). To retrieve the primary femoral stem, a femoral osteotomy was required in 3 cases (6%). To limit acetabular host bone deficits, the Explant (Zimmer Biomet, Warsaw, Indiana) was used to retrieve the acetabular cup. Allografts were used in 16 cases (34%) to augment acetabular deficits. Press-fit fixation of the acetabular component was obtained by 1 to 2 mm under reaming versus the implant’s outside diameter. Press-fit between the anterior and posterior acetabular rims was achieved in cases where a superior rim deficit was present. One such technique to achieve this was to medialize the implant to increase anterior and posterior rim contact. If superior contact was the preferred option where smaller defects were present; the center of rotation for the hip was elevated by 10 mm. Examples of these principles are demonstrated in Figure 2 and Figure 3.

Figure 2. Anteroposterior (AP) pelvis radiograph of a woman of 39 years of age who had a THA with 22mm head and structural bone graft for hip dysplasia (left). She sustained recurrent anterior THA dislocation and required surgical revision. AP pelvis radiograph after revision with CoC LDH THA with 36 mm bearing diameter (right). After removing the acetabular cup and accessible screws, the acetabular cavity was deepened and enlarged by 5 mm (original cup diameter 41 mm, new monobloc cup 46 mm).

Figure 2.

Figure 3. AP pelvis radiograph of a 70 years-old male who had a MoM LDH THA for avascular necrosis following hip fracture fixation with cannulated screws (left). He required surgical revision for Adverse reaction to metal debris (ARMD). AP pelvis radiograph after revision with CoC LDH THA 44 mm bearing diameter (right). After removing the acetabular cup, the acetabular cavity was deepened and enlarged by 4 mm (original cup diameter 52 mm, new monobloc cup 56 mm). Cup medialization with slight high hip center was performed to obtain adequate primary press-fit fixation.

Figure 3.

RESULTS

After an average follow-up of 4.5 years (2.7; 2.0 to 13.7), no aseptic loosening of the acetabular component was observed. Implant survivorship for any revision, at 4-years was 89.4% (95% CI: 89.3 to 89.5). There were 5 re-revisions (10.6%): 2 (4.3%) for loss of primary fixation (≤ 6 weeks) and 3 (6.4%) for instability, two of whom were primarily revised for instability. There was 1 (2.1%) non-revision reoperation for plate removal and femoral allograft in a patient with residual pain in whose revision THA had required femoral osteotomy which had been managed with a hook plate fixated with cerclage cables.

Regarding complications that did not require reoperation, there was 1 (2.1%) excessive postoperative blood loss that necessitated hospitalisation and transfusion, 1 (2.1%) peroperative acetabular fracture that was managed with six-weeks protective weight-bearing and remains stable six years post-revision, 1 (2.1%) pseudotumor secondary to metal debris of a MoM LDH prosthesis that was confirmed on magnetic resonance imaging (MRI) and managed conservatively, and 1 (2.1%) recurrent hip dislocation that remains under conservative treatment 6 years after the last event.

Last follow-up radiographic analyses revealed no additional signs of acetabular component aseptic loosening. The mean acetabular component abduction angle post-revision surgery was 43.1° (5.8°; 30.2° to 59.0°). Acetabular protrusion was present in 20 hips (42.6%), including 5 cases with pre-revision acetabular protrusion. The uncovered proportion of the acetabular component was 0 to 5% in 25 hips (53.2%), 6 to 10% in 8 hips (17.0%), 11 to 15% in 10 hips (21.3%), 16 to 20% in 3 hips (6.4%) and 21 to 25% in 1 hip (2.1%). At last follow-up, heterotopic ossification was categorized as Brooker class I for 8 hips (17.0%), class II for 1 hip (2.1%) and class III for 1 hip (2.1%). PROMs results are reported in Table 2.

Table 2. Functional outcomes at last follow-up following revision THA using monobloc press-fit cups with large diameter bearings.

Criteria
Mean functional score, (SD; range)
WOMAC 19.5 (16.8; 0.0 - 58.3)
FJS 57.3 (28.9; 6.3 - 100.0)
PJP, n
Natural hip 3 (8.8%)
Artificial hip with no restriction 8 (23.5%)
Artificial hip with minimal restriction 14 (41.2%)
Artificial hip with major restriction 9 (26.5%)
Non-functional hip 0 (0.0%)

WOMAC, Western Ontario and McMaster Universities Osteoarthritis Index; FJS, forgotten joint score; PJP, patient’s joint perception

DISCUSSION

There has been an increase in revision hip surgery over the last decade as the number of primary THA procedures has increased. Based on the UK joint registry alone, there has been an increase from 2730 in 2004 to 8087 in 2019.17 Revision surgery results in a less favourable quality of life; a paper by Lubbeke et al. indicates both Harris Hip Score (HHS) and WOMAC scores are significantly less for revision THA when compared with primary THA.18 The most common causes for revision of primary THA include instability, infection, and aseptic loosening.19,20 There is strong evidence in the literature that patient factors (such as age, gender, co-morbidities), surgeon experience and implant factors all impact the risk of revision.21–26

At our institution, the preferred option is an LDH using either monobloc or DM configuration for the acetabular component in primary THA and revision THA (Paprosky 1 and 2). There are numerous advantages to LDH implants. LDH THA offers an impingement-free range of motion and a reduced risk of dislocation.27–31 The larger head-neck offset allows for a greater supraphysiologic range of motion. This can also compensate for a patient’s abnormal spinopelvic mobility and surgical imprecision.32–34 Additionally, LDH bearing with a small clearance exerts a high suction force, which provides greater hip micro-stability.35,36 With appropriate biomechanical reconstruction, LDH THA can restore normal gait parameters.15,37,38 This results in unrestricted activities and higher patient satisfaction scores.15,38 We use CoC LDH for our patients with a life expectancy of more than 20 years and use DM bearings for all others.

Monobloc acetabular components do not allow for supplemental screw fixation, thus relying on press-fit for primary fixation. Compared to modular cups, monobloc components have not been associated with an increased rate of acetabular cup migration, aseptic loosening, or revision.39,40 Monobloc components offer optimal liner-shell conformity and eliminates liner-shell micromotion which has been proposed to reduce the risk of backside wear.41 Additionally, the absence of screw holes eliminates the access channels of wear debris to the subchondral bone, reducing the risk of acetabular osteolysis.42 Monobloc acetabular components in revision THA have been associated with favourable outcomes i.e., low aseptic loosening rates, significant improvement in functional outcomes and low re-revision rates.43,44

In our case series, using LDH cups for revision surgery, we found that implant survivorship at 4-years was 89.4% (95% CI: 89.3 to 89.5), wish is slightly superior to the 5-year survivorship of revision THA of 86.8% reported in the literature.21 After an average follow-up of 4.5 years (2.7; 2.0 to 13.7), no aseptic loosening of the acetabular component was observed. The reason for the 5 (10.6%) re-revisions varied and included: 2 loss of primary fixation (≤ 6 weeks) and 3 instabilities, 2 of whom were primarily revised for instability which have been associated with a higher risk of revision failure for instability.28 The third case of re-revision for instability was a revision of a MoM prosthesis which have been associated with higher complication rates, namely instability.45 Additionally, 2 of the failed revision for instability had a LDH prosthesis implanted during primary THA, which mitigates the advantages of using a LDH prosthesis for revision THA. Instability rates of up to 28% following revision THA have been reported in the literature, which are much higher than the instability rate of 8.5% reported in this series.46 Secondary outcome measures based on radiological imaging found no evidence of acetabular cup aseptic loosening.

The available literature supports these findings. A study by Klemt et al. using DM in revision THA with associated abductor deficiency demonstrated significantly lower dislocation rates at 4 years follow-up with 0% dislocations compared with 15.5% and 4.1% in conventional liners and constrained liners, respectively.47 Chalmers et al. found that LDH implants were a viable salvage option for high-risk cases with recurrent dislocation despite using constrained liners at a median follow up of 37 months.48 A study by Eckar et al. followed 216 patients for a median period of 69 months to assess the revision rate and dislocation rate of DM cups used in complex revision THA. The dislocation-free survival rate was 96% at 5 years and 82% at 9 years.49 A systematic review and meta-analysis performed by Pituckanotai et al. that reviewed 11 articles concluded that LDH is the implant of choice to reduce the risk of revision and dislocation when reviewing short term (0-5 year) outcome studies.50 LDH cups offer numerous advantages; the key advantage for revision is that it allows a larger area of contact between the cementless cup and the host’s bone.

The mode of fixation of the acetabular implant is widely discussed in the literature. For revision THA, the evidence in the literature suggests cemented fixation has been associated with poor outcomes and a high rate of loosening, whereas excellent outcomes have been documented with press-fit fixation in the revision THA.51–53 Our results also support that in revision THA with limited to moderate bone deficits that do not require structural allograft, adequate press-fit fixation of monobloc acetabular components can be obtained. An adequate press-fit on the anterior and posterior acetabular rims should be sought to optimize fixation. Therefore, if needed, the acetabular component should be medialized before attempting to increase the cup size. In cases of superior bone deficits, a high hip center of up to 10 mm can ensure adequate host bone contact on the superior dome, which is essential for the press-fit stability.54,55 While reaming, surgeons should use increments of 1 mm when approaching the desired diameter to ensure the sphericity of the reamed cavity. If desired, to gain confidence with press-fit only fixation, surgeons could seek adequate fixation with modular components without using screws before switching to press-fit monobloc acetabular components.

Khatod et al. investigated factors associated with re-revision and found the leading cause to be instability (42.9%), followed by infection (28.6%), and aseptic loosening (9.5%).21 Likewise, Springer et al. found instability to be the leading cause of re-revision (35%), followed by aseptic loosening (30%), and infection (12%).19 Interestingly, when reviewing the risk of re-revision, the evidence from the UK joint registry suggests a strong relationship between the time from primary THA to revision THA is a higher risk of re-revision. Causes for shorter time periods to revision would include instability as the leading cause, whereas more extended time periods to a revision include aseptic loosening.17 This is supported by Springer et al., whose research indicated that the survivorship of revisions at 10 years was 8.2%, with instability and aseptic loosening being the leading causes for the re-revision.19 It is stated that 20% of hips revised within a year of primary surgery are re-revised within 10 years; if primary hips lasted at least 5 years then the revision rate is 10.8% at 10 years.17

Implant survivorship is only one measure of success; it cannot be used to indicate satisfaction or improvement in function. While there is strong evidence in the literature supporting the use of LDH to improve functional range of motion in primary hips, there is limited data available for functional outcomes for the use of LDH in the revision setting. The benefits of using LDH in the primary setting includes a reduction in the risk of impingement (allowing for a greater range of movement and a decreased risk of dislocation); improved abductor function and hip kinematics; and a reduced long-term risk of implant loosening.56 Our study analysis functional outcomes for the use of LDH in the revision setting. This study demonstrates an improvement in functional outcomes when using LDH for revision surgery. The functional outcome scores highlighted include a WOMAC score of 19.5 (16.8; 0.0 to 58.3) and an FJS of 57.3 (28.9; 6.3 to 100.0); LDH for revision surgery is a good option. One of the few other studies that analyzes functional outcomes of LDH in the revision setting includes the study by Civinini et al. This study prospectively followed 33 patients for a mean of 3 years. The HHS improved from 48 preoperatively to 86 postoperatively, and the average total WOMAC score improved from 79 to 19 postoperatively.57

Limitations of the study

This study was performed in a tertiary centre with arthroplasty surgeons performing high volume complex revision THAs using monobloc LDH. This is not a true reflection of what is commonly practised. The results illustrated may not be replicated in all hospitals using the same implants if they were to adopt this technique. In addition, most acetabular defects were the type 1 Paprosky classification. The true extent of the application with other types of defects remains unclear. Finally, due to the retrospective nature of the study, cases planned for revision THA with a monobloc cup that were converted to a modular component intraoperatively were not included. The choice of implant type was based on the intraoperative judgment and technical abilities of the surgeon.

CONCLUSION

Based on our findings, the press-fit primary fixation of LDH monobloc components implanted in revision THAs is a safe and successful option. These acetabular components have demonstrated good mid-term results with low rates of relative complications while maintaining good functional outcomes.

AUTHOR CONTRIBUTIONS

Conceptualization, PAV, MWM and MOK.; methodology, PAS.; data acquisition and analysis, PAS and BB; writing - original draft preparation, PAS and SS; writing - review and editing, PAS, SS, BB, MWM, MOK and PAV; supervision, PAV and MOK. All authors have read and agreed to the published version of the manuscript.

DECLARATION OF CONFLICTING INTERESTS

PAV declares being a consultant for Stryker and Johnson & Johnson; receives royalties from Microport; and received research funding from Zimmer, Smith & Nephew, Johnson & Johnson, and Stryker. MOK declares he received research funding from Zimmer, Smith & Nephew, Johnson & Jonhson and Styker. PAS, SS and BB declare that they have no conflict of interest.

Acknowledgments

ACKNOWLEDGEMENTS

We thank Dr. Alain Roy, Dr. Martin Lavigne, and Dr. Vincent Massé who contributed to the surgical procedures

Funding Statement

Unrestricted grant was obtained from the Maisonneuve-Rosemont hospital Foundation to support PAS Master’s degree University study tuition fees. The authors did not receive financial support for the research, authorship, and/or publication of this article.

References

  1. 1. Patil S, Garbuz DS, Greidanus NV, Masri BA, Duncan CP. Quality of life outcomes in revision vs primary total hip arthroplasty: a prospective cohort study. J Arthroplasty. 2008;23(4):550-553. doi:10.1016/j.arth.2007.04.035 [DOI] [PubMed]
  2. 2. Zhang S, Chen JY, Pang HN, Lo NN, Yeo SJ, Liow MHL. Revision total hip arthroplasty is associated with poorer clinically meaningful improvements and patient satisfaction compared to primary total hip arthroplasty. J Orthop. 2021;28:96-100. doi:10.1016/j.jor.2021.11.008 [DOI] [PMC free article] [PubMed]
  3. 3. Kenney C, Dick S, Lea J, Liu J, Ebraheim NA. A systematic review of the causes of failure of Revision Total Hip Arthroplasty. Journal of Orthopaedics. 2019;16(5):393-395. doi:10.1016/j.jor.2019.04.011 [DOI] [PMC free article] [PubMed]
  4. 4. Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89(4):780-785. doi:10.2106/jbjs.f.00222 [DOI] [PubMed]
  5. 5. Howie DW, Holubowycz OT, Middleton R. Large femoral heads decrease the incidence of dislocation after total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2012;94(12):1095-1102. doi:10.2106/jbjs.k.00570 [DOI] [PubMed]
  6. 6. Paprosky WG, Perona PG, Lawrence JM. Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty. 1994;9(1):33-44. doi:10.1016/0883-5403(94)90135-x [DOI] [PubMed]
  7. 7. DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;(121):20-32. doi:10.1097/00003086-197611000-00003 [PubMed]
  8. 8. Massin P, Schmidt L, Engh CA. Evaluation of cementless acetabular component migration. An experimental study. J Arthroplasty. 1989;4(3):245-251. doi:10.1016/s0883-5403(89)80020-8 [DOI] [PubMed]
  9. 9. Malchau H, Kärrholm J, Wang YX, Herberts P. Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients. Acta Orthop Scand. 1995;66(5):418-424. doi:10.3109/17453679508995578 [DOI] [PubMed]
  10. 10. Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr. Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am. 1973;55(8):1629-1632. doi:10.2106/00004623-197355080-00006 [PubMed]
  11. 11. McBride MT, Muldoon MP, Santore RF, Trousdale RT, Wenger DR. Protrusio acetabuli: diagnosis and treatment. J Am Acad Orthop Surg. 2001;9(2):79-88. doi:10.5435/00124635-200103000-00002 [DOI] [PubMed]
  12. 12. Li H, Mao Y, Oni JK, Dai K, Zhu Z. Total hip replacement for developmental dysplasia of the hip with more than 30% lateral uncoverage of uncemented acetabular components. Bone Joint J. 2013;95-B(9):1178-1183. doi:10.1302/0301-620x.95b9.31398 [DOI] [PubMed]
  13. 13. Bellamy N, Buchanan WW, Goldsmith CH, Campbell J, Stitt LW. Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol. 1988;15(12):1833-1840. [PubMed]
  14. 14. Behrend H, Giesinger K, Giesinger JM, Kuster MS. The “forgotten joint” as the ultimate goal in joint arthroplasty: validation of a new patient-reported outcome measure. J Arthroplasty. 2012;27(3):430-436.e1. doi:10.1016/j.arth.2011.06.035 [DOI] [PubMed]
  15. 15. Collins M, Lavigne M, Girard J, Vendittoli PA. Joint perception after hip or knee replacement surgery. Orthopaedics & Traumatology: Surgery & Research. 2012;98(3):275-280. doi:10.1016/j.otsr.2011.08.021 [DOI] [PubMed]
  16. 16. Goel MK, Khanna P, Kishore J. Understanding survival analysis: Kaplan-Meier estimate. Int J Ayurveda Res. 2010;1(4):274-278. doi:10.4103/0974-7788.76794 [DOI] [PMC free article] [PubMed]
  17. 17. NJR U. Number and percentage of hip revisions by procedure type and year. Accessed April 28, 2022. https://reports.njrcentre.org.uk/Portals/0/PDFdownloads/NJR%2018th%20Annual%20Report%202021.pdf
  18. 18. Lübbeke A, Katz JN, Perneger TV, Hoffmeyer P. Primary and revision hip arthroplasty: 5-year outcomes and influence of age and comorbidity. J Rheumatol. 2007;34(2):394-400. [PubMed]
  19. 19. Springer BD, Fehring TK, Griffin WL, Odum SM, Masonis JL. Why revision total hip arthroplasty fails. Clin Orthop Relat Res. 2009;467(1):166-173. doi:10.1007/s11999-008-0566-z [DOI] [PMC free article] [PubMed]
  20. 20. Clohisy JC, Calvert G, Tull F, McDonald D, Maloney WJ. Reasons for revision hip surgery: a retrospective review. Clin Orthop Relat Res. 2004;(429):188-192. doi:10.1097/01.blo.0000150126.73024.42 [DOI] [PubMed]
  21. 21. Khatod M, Cafri G, Inacio MCS, Schepps AL, Paxton EW, Bini SA. Revision total hip arthoplasty: factors associated with re-revision surgery. J Bone Joint Surg Am. 2015;97(5):359-366. doi:10.2106/jbjs.n.00073 [DOI] [PubMed]
  22. 22. Mahomed NN, Barrett JA, Katz JN, et al. Rates and outcomes of primary and revision total hip replacement in the United States medicare population. J Bone Joint Surg Am. 2003;85(1):27-32. doi:10.2106/00004623-200301000-00005 [DOI] [PubMed]
  23. 23. Lee PTH, Lakstein DL, Lozano B, Safir O, Backstein J, Gross AE. Mid-to long-term results of revision total hip replacement in patients aged 50 years or younger. Bone Joint J. 2014;96-B(8):1047-1051. doi:10.1302/0301-620x.96b8.31587 [DOI] [PubMed]
  24. 24. Radcliffe GS, Tomichan MC, Andrews M, Stone MH. Revision hip surgery in the elderly: is it worthwhile? J Arthroplasty. 1999;14(1):38-44. doi:10.1016/s0883-5403(99)90200-0 [DOI] [PubMed]
  25. 25. Zhan C, Kaczmarek R, Loyo-Berrios N, Sangl J, Bright RA. Incidence and short-term outcomes of primary and revision hip replacement in the United States. J Bone Joint Surg Am. 2007;89(3):526-533. doi:10.2106/jbjs.f.00952 [DOI] [PubMed]
  26. 26. Dobzyniak M, Fehring TK, Odum S. Early failure in total hip arthroplasty. Clin Orthop Relat Res. 2006;447:76-78. doi:10.1097/01.blo.0000203484.90711.52 [DOI] [PubMed]
  27. 27. Garbuz DS, Masri BA, Duncan CP, et al. The Frank Stinchfield Award: Dislocation in Revision THA: Do Large Heads (36 and 40 mm) Result in Reduced Dislocation Rates in a Randomized Clinical Trial? Clinical Orthopaedics Related Research. 2012;470(2):351-356. doi:10.1007/s11999-011-2146-x [DOI] [PMC free article] [PubMed]
  28. 28. Hoskins W, Bingham R, Hatton A, de Steiger RN. Standard, Large-Head, Dual-Mobility, or Constrained-Liner Revision Total Hip Arthroplasty for a Diagnosis of Dislocation: An Analysis of 1,275 Revision Total Hip Replacements. J Bone Joint Surg Am. 2020;102(23):2060-2067. doi:10.2106/jbjs.20.00479 [DOI] [PubMed]
  29. 29. Tsikandylakis G, Mohaddes M, Cnudde P, Eskelinen A, Kärrholm J, Rolfson O. Head size in primary total hip arthroplasty. EFORT Open Rev. 2018;3(5):225-231. doi:10.1302/2058-5241.3.170061 [DOI] [PMC free article] [PubMed]
  30. 30. Cinotti G, Lucioli N, Malagoli A, Calderoli C, Cassese F. Do large femoral heads reduce the risks of impingement in total hip arthroplasty with optimal and non-optimal cup positioning? International Orthopaedics. 2011;35(3):317-323. doi:10.1007/s00264-010-0954-3 [DOI] [PMC free article] [PubMed]
  31. 31. Stulberg SD. Dual mobility for chronic hip instability: a solution option. Orthopedics. 2010;33(9):637. doi:10.3928/01477447-20100722-51 [DOI] [PubMed]
  32. 32. McKnight BM, Trasolini NA, Dorr LD. Spinopelvic Motion and Impingement in Total Hip Arthroplasty. The Journal of Arthroplasty. 2019;34(7):S53-S56. doi:10.1016/j.arth.2019.01.033 [DOI] [PubMed]
  33. 33. Ike H, Dorr LD, Trasolini N, Stefl M, McKnight B, Heckmann N. Spine-Pelvis-Hip Relationship in the Functioning of a Total Hip Replacement. The Journal of Bone and Joint Surgery. 2018;100(18):1606-1615. doi:10.2106/jbjs.17.00403 [DOI] [PubMed]
  34. 34. Blumenfeld TJ. Pearls: Clinical Application of Ranawat’s Sign. Clin Orthop Relat Res. 2017;475(7):1789-1790. doi:10.1007/s11999-017-5376-8 [DOI] [PMC free article] [PubMed]
  35. 35. Blakeney WG, Epinette JA, Vendittoli PA. Reproducing the Proximal Femoral Anatomy: Large-Diameter Head THA. In: Rivière C, Vendittoli PA, eds. Personalized Hip and Knee Joint Replacement. Springer; 2020:65-73. doi:10.1007/978-3-030-24243-5_7 [PubMed]
  36. 36. Clarke MT, Lee PTH, Villar RN. Dislocation after total hip replacement in relation to metal-on-metal bearing surfaces. J Bone Joint Surg Br. 2003;85-B(5):650-654. doi:10.1302/0301-620x.85b5.13993 [PubMed]
  37. 37. Lavigne M, Therrien M, Nantel J, Roy A, Prince F, Vendittoli PA. The John Charnley Award: The functional outcome of hip resurfacing and large-head THA is the same: a randomized, double-blind study. Clin Orthop Relat Res. 2010;468(2):326-336. doi:10.1007/s11999-009-0938-z [DOI] [PMC free article] [PubMed]
  38. 38. Puliero B, Blakeney WG, Beaulieu Y, Vendittoli PA. Joint Perception After Total Hip Arthroplasty and the Forgotten Joint. J Arthroplasty. 2019;34(1):65-70. doi:10.1016/j.arth.2018.09.086 [DOI] [PubMed]
  39. 39. Halma JJ, Vogely CH, Dhert WJ, Van Gaalen SM, De Gast A. Do Monoblock Cups Improve Survivorship, Decrease Wear, or Reduce Osteolysis in Uncemented Total Hip Arthroplasty? Clinical Orthopaedics Related Research. 2013;471(11):3572-3580. doi:10.1007/s11999-013-3144-y [DOI] [PMC free article] [PubMed]
  40. 40. Weiss RJ, Hailer NP, Stark A, Kärrholm J. Survival of uncemented acetabular monoblock cups: evaluation of 210 hips in the Swedish Hip Arthroplasty Register. Acta Orthop. 2012;83(3):214-219. doi:10.3109/17453674.2012.688726 [DOI] [PMC free article] [PubMed]
  41. 41. Young AM, Sychterz CJ, Hopper RH Jr, Engh CA. Effect of acetabular modularity on polyethylene wear and osteolysis in total hip arthroplasty. J Bone Joint Surg Am. 2002;84(1):58-63. doi:10.2106/00004623-200201000-00009 [DOI] [PubMed]
  42. 42. Huk OL, Bansal M, Betts F, et al. Polyethylene and metal debris generated by non-articulating surfaces of modular acetabular components. J Bone Joint Surg Br. 1994;76-B(4):568-574. doi:10.1302/0301-620x.76b4.8027142 [PubMed]
  43. 43. Unger AS, Lewis RJ, Gruen T. Evaluation of a porous tantalum uncemented acetabular cup in revision total hip arthroplasty: clinical and radiological results of 60 hips. J Arthroplasty. 2005;20(8):1002-1009. doi:10.1016/j.arth.2005.01.023 [DOI] [PubMed]
  44. 44. Fernández-Fairen M, Murcia A, Blanco A, Meroño A, Murcia A Jr, Ballester J. Revision of failed total hip arthroplasty acetabular cups to porous tantalum components: a 5-year follow-up study. J Arthroplasty. 2010;25(6):865-872. doi:10.1016/j.arth.2009.07.027 [DOI] [PubMed]
  45. 45. Stryker LS, Odum SM, Fehring TK, Springer BD. Revisions of monoblock metal-on-metal THAs have high early complication rates. Clin Orthop Relat Res. 2015;473(2):469-474. doi:10.1007/s11999-014-3791-7 [DOI] [PMC free article] [PubMed]
  46. 46. Parvizi J, Picinic E, Sharkey PF. Revision total hip arthroplasty for instability: surgical techniques and principles. Instr Course Lect. 2009;58:183-191. [PubMed]
  47. 47. Klemt C, Smith EJ, Oganesyan R, Limmahakhun S, Fitz D, Kwon YM. Outcome of Dual Mobility Constructs for Adverse Local Tissue Reaction Associated Abductor Deficiency in Revision Total Hip Arthroplasty. J Arthroplasty. 2020;35(12):3686-3691. doi:10.1016/j.arth.2020.06.043 [DOI] [PubMed]
  48. 48. Chalmers BP, Pallante GD, Taunton MJ, Sierra RJ, Trousdale RT. Can Dislocation of a Constrained Liner Be Salvaged With Dual-mobility Constructs in Revision THA? Clin Orthop Relat Res. 2018;476(2):305-312. doi:10.1007/s11999.0000000000000026 [DOI] [PMC free article] [PubMed]
  49. 49. Unter Ecker N, Kocaoğlu H, Zahar A, Haasper C, Gehrke T, Citak M. What Is the Dislocation and Revision Rate of Dual-mobility Cups Used in Complex Revision THAs? Clin Orthop Relat Res. 2021;479(2):280-285. doi:10.1097/corr.0000000000001467 [DOI] [PMC free article] [PubMed]
  50. 50. Pituckanotai K, Arirachakaran A, Tuchinda H, et al. Risk of revision and dislocation in single, dual mobility and large femoral head total hip arthroplasty: systematic review and network meta-analysis. Eur J Orthop Surg Traumatol. 2018;28(3):445-455. doi:10.1007/s00590-017-2073-y [DOI] [PubMed]
  51. 51. Pulido L, Rachala SR, Cabanela ME. Cementless acetabular revision: past, present, and future. Revision total hip arthroplasty: the acetabular side using cementless implants. Int Orthop. 2011;35(2):289-298. doi:10.1007/s00264-010-1198-y [DOI] [PMC free article] [PubMed]
  52. 52. Elke R, Berli B, Wagner A, Morscher EW. Acetabular revision in total hip replacement with a press-fit cup. J Bone Joint Surg Br. 2003;85-B(8):1114-1119. doi:10.1302/0301-620x.85b8.13590 [DOI] [PubMed]
  53. 53. Massin P, Besnier L. Acetabular revision using a press-fit dual mobility cup. Orthop Traumatol Surg Res. 2010;96(1):9-13. doi:10.1016/j.rcot.2009.11.006 [DOI] [PubMed]
  54. 54. Bozic KJ, Freiberg AA, Harris WH. The high hip center. Clin Orthop Relat Res. 2004;(420):101-105. doi:10.1097/00003086-200403000-00014 [DOI] [PubMed]
  55. 55. Spears IR, Pfleiderer M, Schneider E, Hille E, Morlock MM. The effect of interfacial parameters on cup-bone relative micromotions. A finite element investigation. J Biomech. 2001;34(1):113-120. doi:10.1016/s0021-9290(00)00112-3 [DOI] [PubMed]
  56. 56. Castagnini F, Cosentino M, Bracci G, Masetti C, Faldini C, Traina F. Ceramic-on-Ceramic Total Hip Arthroplasty with Large Diameter Heads: A Systematic Review. Med Princ Pract. 2021;30(1):29-36. doi:10.1159/000508982 [DOI] [PMC free article] [PubMed]
  57. 57. Civinini R, Carulli C, Matassi F, Nistri L, Innocenti M. A dual-mobility cup reduces risk of dislocation in isolated acetabular revisions. Clin Orthop Relat Res. 2012;470(12):3542-3548. doi:10.1007/s11999-012-2428-y [DOI] [PMC free article] [PubMed]

Articles from Orthopedic Reviews are provided here courtesy of Open Medical Publishing

RESOURCES