Abstract
Over the past 30 years, the significance of nitric oxide (NO) has become increasingly apparent in mammalian physiology. It is biosynthesized by three isoforms of nitric oxide synthases (NOS): neuronal (nNOS), endothelial (eNOS), and inducible (iNOS). Neuronal and eNOS both produce low levels of NO (nM) as a signaling agent and vasodilator, respectively. Inducible (iNOS) is present in activated macrophages at sites of infection to generate acutely toxic (μM) levels of NO as part of the mammalian immune defense mechanism. These discoveries have led to numerous animal and clinical studies to evaluate the potential therapeutic utility of NO in various medical operations/treatments, primarily using NO gas (via gas-cylinders) as the NO source. In this review, we focus specifically on recent advances in the electrochemical generation of NO (E-NOgen) as an alternative means to generate NO from cheap and inert sources, and the fabrication and testing of biomedical devices that utilize E-NOgen to controllably generate NO for medical applications.
Keywords: nitric oxide, copper catalyst, electrochemical nitrite reduction, electrocatalysis, IV-catheters, inhaled nitric oxide, silicone-rubber patches
1. INTRODUCTION
Nitric oxide (NO) is an endogenous gas that plays several key physiological roles in humans, including prevention of platelet activation/adhesion, inhibiting bacterial adhesion and proliferation, enhancing vasodilation, promoting angiogenesis, mediating nerve-signal transduction, and aiding in wound healing.1-7 At low NO fluxes of 0.5 – 4.0 x 10−10 mol min−1 cm−2, as emitted from a normal-functioning endothelium, NO acts as an anticoagulant, both controlling arteriolar tone and preventing activation of platelets that approach this surface.8 NO is also released by macrophages and nasal epithelial cells at higher fluxes, where it acts as a potent antimicrobial agent to combat bacterial infection.9, 10 The antithrombotic and antibacterial properties of NO have resulted in its use to combat clotting and bacterial biofilm formation in several medical applications.
Inhaled nitric oxide (INO) therapy has become a mainstay of intensive care units for lung failure, resulting in preferential pulmonary vasodilation.11 In cases where neonates have hypoxic respiratory failure (HRF), INO has been demonstrated to improve oxygenation and reduce the need for extracorporeal membrane oxygenation (ECMO) therapy.12, 13 Though INO is only approved for HRF at present, it has been demonstrated to be beneficial in a numerous other treatments, including for pneumonia,14 acute respiratory distress syndrome (ARDS),15 stroke, ,16 cystic fibrosis,17 tuberculosis,18 pulmonary hypertension,19, 20 pulmonary fibrosis,21 COPD,22 and most recently for potential use in treating COVID-19.23, 24 During cardiopulmonary bypass (CPB), NO has also been shown to prevent blood activation (due to the exposure of blood to air),25 inflammation, and organ failure during heart surgery.26, 27 When NO is added to the sweep gas of an oxygenator during simulated CPB, it has been shown to attenuate blood cell activation.28-31
The major limitation to the routine use of NO in emergency rooms and in the field is the prohibitively high cost and low portability of current commercial systems for gaseous NO delivery. These require delivering NO via gas cylinders, which are limited to critical care facilities, contain NO at concentrations that can lead to decomposition (due to disproportionation over time) to generate NO2, a highly toxic gas, and are expensive (~$3000/patient/day).32 Alternative approaches have been tested to develop more compact/portable NO-delivery systems. For example, a liquid N2O4 device as a source of NO is under development (GeNO, LLC). N2O4 exists in equilibrium with NO2(g) that gets passed over a reducing cartridge, ultimately generating NO.33 The focus of this review is on recent research into alternative, portable, tunable, and low cost means of generating NO electrochemically from inorganic nitrite ions for medical applications.
2 ∣. BIOMEDICAL APPLICATIONS OF ELECTROCHEMICAL NITRIC OXIDE GENERATION
Taking inspiration from biological copper-nitrite reductases (CuNIR),34, 35 our team has developed a new, low-cost electrochemical method to generate NO from benign precursors for potential use in clinics, ambulances, and in the field. Using CuNIR inorganic model complexes as a starting point, the electrochemical reduction of nitrite (NO2−) ions to NO gas can be achieved using specially developed Cu(II)-ligand (Cu(II)-L) complexes as catalysts. Here, the Cu(II)-L complex is initially reduced to the Cu(I)-L form via either an applied current or potential, then binds NO2− and rapidly reduces it to NO and water, thereby regenerating the Cu(II)-L form of the catalyst (Fig. 1, eq. 1). Importantly, NO gas levels are tuned at rates dependent on the applied voltage or current. The ability to generate NO on demand from low-cost and stable sources
| (eq. 1) |
Fig. 1.

Mechanism of electrochemical reduction of NO2− to generate NO via a Cu(II)-ligand catalyst.
(NaNO2), and the ability to tune the concentration of NO generated (applied current/voltage) are both highly desirable in applying this method of NO generation for a variety of biomedical applications. Below we discuss advances in the controlled electrochemical generation of NO via this method in three different configurations for three different biomedical applications: intravenous (IV)-catheters, gas phase NO for INO therapy, and silicone rubber patches for wound healing (Fig. 2).
Fig. 2.
Overview of three E-NOgen devices utilized for biomedical applications. (A) Two- and three-electrode single-lumen IV-catheters and dual lumen catheter containing a second PO2 sensing lumen, (B) Bulk NO generator for INO, and (C) NO-releasing silicone rubber patch for wound healing.
2.1 ∣. IV-CATHETERS
The use of IV-catheters comes with two primary risk factors – thrombus formation and bacterial infection. These risks can range from simple interference with medical care to fatality, with an estimated 28,000 deaths and ~$2.3 billion in additional healthcare costs per year associated with these two complications.36 Nevertheless, this risk is tolerated because vascular access is essential for diagnoses and treatments. Due to the antithrombotic and antibacterial effects of NO, the generation of NO within catheter devices has been studied extensively in the literature as a potentially novel and convenient solution to mitigate IV-catheter risk factors.37, 38
A large body of literature details the use of the non-electrochemical generation of NO via NO donor molecules within polymeric materials, or as coatings on implantable devices, including IV-catheters.10, 39-53 For example, silicone foley catheters impregnated with S-nitroso-N-acetylpenicillamine (SNAP) generate surface NO fluxes > 0.7 × 10−10 mol min−1 cm−2 for over 31 days and result in up to 2.5 log unit decreases in viable S. epidermis biofilms at the catheter surface compared to controls.51 Diazeniumdiolate derivatives (also frequently referred to as NONOates in the literature) have also been studied in catheter/sensor applications. For example, glucose sensors containing a diazeniumdiolated dibutylhexyldiamine coating tested in bovine serum exhibited continuous NO fluxes > 1.0 × 10−10 mol min−1 cm−2 for 6 days leading to more accurate venous blood glucose sensing in rabbit jugular veins compared to controls without the coating.52 While long-term NO release and anti-thrombotic activity have been demonstrated for several NO donors/coatings, the use of common NO donors such as diazeniumdiolates or S-nitrosothiols in NO-releasing catheter devices/coatings is limited by their instability to temperature, photolysis, leaching, etc., negatively impacting their storability over time. Furthermore, these NO donor-based methods suffer from low tunability of their NO release profiles – dependent on the nature of the molecule (half-life/kinetics of NO release is molecule/material dependent), can potentially leach out of the material, and may require rigorous storage and handling conditions to prevent decomposition prior to their use. Consequently, the commercialization of these technologies has been challenging.
Over the past ~10 years our laboratory has devised and tested a series of copper-based E-NOgen catalysts for use in IV-catheter devices. In our initial report of an E-NOgen-based catheter device, we demonstrated that a single-lumen silicone-rubber catheter containing a Cu0 working electrode and Ag/AgCl working/counter electrode with a buffer/nitrite/EDTA solution can generate low concentrations of NO from trace Cu(I) generated at the surface of the Cu0 electrode under specific conditions.54 Due to the initial presence of a fouling copper-oxide layer at the surface of the Cu0 electrode, holding solely at an anodic potential for long periods results in negligible surface oxidation of Cu0 to Cu(I), and correspondingly, negligible NO generation. Instead, a two-step pulse sequence was required, whereby an initial 3-minute cathodic potential at −0.7 V vs. NHE was applied to strip the copper-oxide layer, followed by a subsequent 3-minute anodic potential at 0.2 V vs. NHE to generate Cu(I) at the exposed Cu0 wire for nitrite reduction. Due to the rapid formation of new copper-oxide layers during the anodic phase, the continued generation of NO from this system required a constant three-minute alternating cathodic/anodic voltage cycle. Low NO fluxes of 0.5-1.0 × 10−10 mol min−1 cm−2 were emitted from the outside surface of the catheter, measured via the gold-standard chemiluminescence NO detection method.55 Further, in a 7-day antimicrobial study, viable bacterial counts of E. coli and A. baumanni strains on the outside surface of these NO emitting catheters were reduced by 97% and 98%, respectively. In a subsequent study, the pulse sequence was refined to produce more consistent and increased NO fluxes by reducing the pulse cycle from 3 min to 30 s and by using a −1.3 V applied cathodic potential, respectively.56 Overall, the limitations of using a Cu0 working electrode as source of Cu(I) ions were somewhat prohibitive (requiring sequentially alternating the applied voltage, etc.), but nevertheless, these studies provided preliminary evidence that the preparation of alternative Cu(I)-ligand catalysts could potentially be employed to electrocatalytically generate pure NO from a NO2− source.
The next generation of E-NOgen catalysts for IV-catheters took an alternate approach: the design and synthesis of Cu(II)-L catalysts that could be electrocatalytically reduced to the Cu(I)-L state for nitrite reduction (see Fig. 1A). This work took inspiration from the type II copper center (T2-Cu) active site of bacterial CuNIRs,57, 58 as well as other reports of Cu(II)/Cu(I) model complexes that can catalyze nitrite reduction chemically/electrochemically.59-61 In our initial report using this approach, tris(2-pyridylmethyl)amine (abbreviated as TMPA here, but also commonly abbreviated as TPA or TPMA) was used to generate a Cu(II)-TMPA catalyst (Fig. 3A), that exhibits a reversible Cu(II)/Cu(I) couple with ER = −380 mV vs. Ag/AgCl in the absence of nitrite.62 Single-lumen silicone rubber catheters containing a Pt wire working electrode and an Ag/AgCl reference/counter electrode with the [Cu(II)(TMPA)]2+ complex (4 mM in a 0.5 M, pH 7.2 MOPS buffer with 0.2 M NaCl electrolyte and 0.4 M NaNO2) in the bulk solution/reservoir were fabricated. These catheters generated NO fluxes ranging from ~0.5 - 3.0 × 10−10 mol min−1 cm−2 and were tunable based on cathodic hold potentials applied to the working Pt electrode (ranging from −0.20 V to −0.40 V). Furthermore, continuous NO release at physiologically relevant levels was demonstrated for up to 7 days. In clinical settings, the application of this technology requires the use of dual-lumen catheters (see Fig. 2A), whereby one lumen acts as a dedicated E-NOgen compartment while the second lumen is used for clinical purposes (i.e., drug infusion or electrochemical sensing of blood gases, electrolytes, glucose, etc.). As such, both single- and dual-lumen catheters were tested in both in vivo anti-thrombosis trials within a rabbit model, as well as for antimicrobial/biofilm properties using a drip-flow bioreactor (as a model for in vivo conditions). Significant reductions in thrombosis were achieved with both single and dual lumen catheters after 7 h of NO release compared to controls (83 ± 12% reduction; p < 0.05, n = 3). Likewise, upon an initial inoculation of E. coli on the surface of single- and dual-lumen catheters, followed by a continuous flow of medium containing E. coli for 3 days in the bioreactor, a > 102 decrease of viable bacteria were found on the surfaces of the catheters when E-NOgen was turned on for only 3 h per day at a NO flux of 0.6 × 10−10 mol min−1 cm−2.
Fig. 3.
Overview of the Cu(II)-L catalysts tested in E-NOgen devices.
One potential concern with using non-centrosymmetric dual-lumen catheters was whether E-NOgen within one lumen of the catheter could generate adequate NO fluxes at the surface of the other lumen. To address this, a detailed finite element analysis study was performed on single- and dual-lumen catheters, as well as a hypothetical 3-lumen catheter for more centrosymmetric NO release profiles.63 The transport of NO from the internal E-NOgen lumen to the external surface of a catheter is dictated by NO transport in the bulk E-NOgen medium, as well as through the polymeric catheter material – i.e. its partition and diffusion coefficients. Importantly, this study demonstrated that dual-lumen catheters composed of silicone rubber and Elast-Eon 5-325 (silicone polyurethane), both polymers with high NO diffusion and partition coefficients, exhibited asymmetric NO release profiles at 10 min response times, with increasing symmetry in NO release after 20 min. Across the board, polymer materials with higher diffusion coefficients gave faster response times to reach a steady-state and more symmetric NO outer surface fluxes, while the effect of the partition coefficient on response times and symmetry of NO release was more complex. Poor distribution/symmetry of NO was predicted at short response-times (10-20 min) with polymers containing high partition coefficients for NO (likely due to needing higher NO concentrations in the polymer before it is released at the surface of the material), though these ultimately gave the highest NO fluxes at longer response-times (30-40 min). In contrast, rapid and more symmetric distributions of NO at 10-20 min time points were predicted in polymers with lower partition coefficients, ultimately generating lower maximum NO fluxes with longer response-times. The results from this study indicated that both silicone rubber and Elast-Eon 5-325 are the best polymer materials to use as IV-catheters for rapid and more symmetric NO release, and that within 20-40 minutes NO is produced at physiologically relevant fluxes at all sides of the catheter.
Our laboratory has also demonstrated that E-NOgen from dual-lumen catheters can in fact enhance the accuracy of electrochemical intravascular sensing of the partial pressure of O2 (PO2).64 Typically, blood analytes are tested periodically in vitro by drawing blood samples, leaving gaps between analyte monitoring that could potentially be detrimental for patient care/diagnostics, especially with critically ill patients. To address this challenge, Ren et al. devised a dual-lumen catheter containing one E-NOgen lumen with a Pt wire working electrode and Ag/AgCl reference/counter electrode containing the Cu(II)-TMPA catalyst in a nitrite/buffer/electrolyte solution (see Fig. 2A). The PO2 sensing lumen consisted of a PFA-coated Pt wire working electrode with an Ag/AgCl wire counter/reference electrode with a 0.1 M bicarbonate buffer at pH 10. The leads from both lumens were connected to separate channels of a multi-channel potentiostat to separately induce E-NOgen by holding the applied potential at −400 mV, and PO2 sensing by holding the applied potential at −700 mV (both vs. Ag/AgCl) to induce O2 reduction. In benchtop experiments, the O2 sensing was fully compatible with E-NOgen - the low NO fluxes produced from E-NOgen do not affect O2 sensing results relative to controls where the E-NOgen process was not turned on (i.e., the second order reaction of NO with O2 is negligible at the concentrations of NO being produced). When tested in vivo within rabbit veins for 7 h and porcine arteries for 21 h, IV-catheters that had E-NOgen turned on were able to continuously measure accurate PO2 levels over the entire 7 h and 21 h duration of the rabbit and porcine experiments, respectively. In contrast, in the controls where E-NOgen was turned off, PO2 measurements were accurate for only 4 h in rabbit veins and 6 h in porcine arteries before large negative deviations in the sensed PO2 levels were measured due to the formation of blood clots on the catheter surface (with active platelets in these clots consuming oxygen, lowering the PO2 levels locally at the surface of the catheters in vivo). This technology could potentially be expanded towards the development of other in vivo sensors for continuous and accurate real-time measurements of different blood analytes to greatly enhance real-time medical diagnostic capabilities.
More recently, we have tested several other Cu(II)-L catalysts to improve catalyst lifetime and NO fluxes (see Fig. 3B, 3C).65, 66 The Cu(II)-Me3TACN (Me3TACN = 1,4,7-trimethyl-1,4,7-triazacyclononane) catalyst (Fig. 3B) exhibits a reduction potential of ~ −340 mV vs. Ag/AgCl. Single- and dual-lumen catheters containing a Pt-wire working electrode and Ag/AgCl wire reference/counter electrode and a solution of 2 mM Cu(II)-Me3TACN/0.4 M NaNO2/HEPES buffer, pH 7.3, within the E-NOgen lumen, were fabricated. These catheters were tested for tunability of NO flux, long-term stability/lifetime, antibacterial properties, and O2 sensitivity. Notably, NO fluxes were generated with a high degree of tunability – ranging from 0 to 16 × 10−10 mol min−1 cm−2 (single-lumen) and 0 to 14 × 10−10 mol min−1 cm−2 (dual-lumen), corresponding to applied currents of 0 to 20 μA and from 0 to 17 × 10−10 mol min−1 cm−2 (single-lumen) and 0 to 15 × 10−10 mol min−1 cm−2 (dual-lumen) corresponding to applied potentials of 0 to −400 mV vs. Ag/AgCl. The Cu(II)-Me3TACN catalyst produced NO with average Faradaic efficiencies, ranging from 84% at 1 μA applied current to 57% at 11 μA applied current. Furthermore, the continuous release of NO at physiologically useful NO fluxes (at ~ 5 × 10−10 mol min−1 cm−2) was demonstrated for up to 8 days. In antimicrobial studies, dual-lumen catheters with one lumen containing the catalyst (and electrodes) exhibited a greater than two-log unit decrease in viable P. aeruginosa and slightly under a 3-log unit decrease in viable S. aureus cell counts when E-NOgen was only turned on for 6 h per day at a −300 mV hold potential vs. Ag/AgCl for 5 days. Despite the significant advances in catalyst lifetime/NO flux/antimicrobial properties compared to previous catalysts, the Cu(II)-Me3TACN catalyst suffered from two major drawbacks. Owing to its negative reduction potential, this catalyst is O2 sensitive, and NO fluxes decreased by ~75% when the vessel housing of the catheter was switched from a 100% N2 sparge gas to a 3% O2 sparge gas. Additionally, in the presence of low levels of nitrite, CV experiments suggested that disproportionation of the Cu(I) form of the complex to Cu(II) and Cu0 occurred, the latter of which plated out on the surface of the working electrode over time.
In order to draw detailed comparisons between ligand structure and catalyst properties (i.e. reduction potential, Faradaic efficiencies, turnover frequencies), a series six of Cu(II)-BMPA-/BEPA-/Et/Pr/Bu (BMPA = bis(2-methylpyridyl)amine; BEPA = bis(2-ethylpyridyl)amine) catalysts containing appended carboxylate arms (Et = ethylate, Pr = propionate, Bu = butylate) were recently studied (Fig. 3C).66 Remarkably, these catalysts span Cu(II)/Cu(I) reduction potentials ranging from −380 mV in Cu(II)-BMPA-Pr to +10 mV for Cu(II)-BEPA-Bu (both vs. Ag/AgCl), whereby the BMPA-carboxylates all exhibit more negative reduction potentials compared to their BEPA-carboxylate analogs. Furthermore, within the series of BMPA-/BEPA-carboxylates, the shorter carboxylate chain lengths correlate with more negative reduction potentials (e.g., BEPA-Et = −250 mV, BEPA-Pr = −80 mV, and BEPA-Bu =10 mV). DFT modelling of this series suggests that the ligand steric constraints affect optimal metal-ligand binding geometries, leading to contracted Cu(II)-pyridine bonds across the BMPA-carboxylate series, causing more negative potentials for the BMPA-carboxylates. Additionally, the optimized BMPA- and BEPA-Et structures had elongated Cu-OOC bond-lengths, leading to further contraction of Cu(II)-pyridine and tertiary amine bonds to compensate for poor donicity from the ethylate arm, which is not long enough to optimally coordinate to the Cu(II) center. With the large range of reduction potentials across the series, several trends were identified in the catalyst properties. For example, the turnover frequencies were directly proportional to the Cu(II)/Cu(I) reduction potential, whereby more reducing catalysts had faster turnover frequencies (BMPA-Et > BMPA-Pr > BMPA-Bu > BEPA-Et > BEPA-Pr > BEPA-Bu), ranging from 11.82 s−1 with Cu(II)-BMPA-Et to 1.18 s−1 with Cu(II)-BEPA-Bu. On the other hand, the Faradaic efficiencies were quite low for BMPA-Et, BMPA-Pr, and BEPA-Et at 15%, 47%, and 23% respectively. The BMPA-Bu, BEPA-Pr, and BEPA-Bu Cu(II) complexes had Faradaic efficiencies of 84%, 94%, and 91%, respectively. These values were ascribed to two competing factors: (1) The more reducing complexes could further reduce/disproportionate generated NO to yield N2O, and (2) The steric strain in BMPA-/BEPA-Et complexes enhances this disproportionation reaction, leading to deviations in the general trend correlating Er to Faradaic efficiency. The combined data for this series demonstrates that a balance must be struck between reduction potential and catalyst Faradaic efficiency: more positive Cu(II)/Cu(I) couples inhibit the over-reduction of NO to N2O but such complexes are also slower catalysts.
In catheter studies, Cu(II)-BMPA-Pr was demonstrated to release NO fluxes between 0 – 8 × 10−10 mol min−1 cm−2 using cathodic hold potentials ranging from 0 to −0.45 V vs Ag/AgCl, and similarly generated NO fluxes between 0 – 7 × 10−10 mol min−1 cm−2 using applied currents between 0 – 15 μA within single-lumen catheters.65 Cu(II)-BEPA-Pr is the best performing catalyst reported to date, with NO fluxes ranging from 0 – 24 × 10−10 mol min−1 cm−2 and 0 – 17 × 10−10 mol min−1 cm−2 in single- and dual-lumen catheter configurations, respectively, when applied potentials ranged from 0 to −0.35 V vs. Ag/AgCl. In antimicrobial studies with dual-lumen catheters containing the Cu(II)-BEPA-Pr complex, > 2 log unit reductions in viable strains of both P. aeruginosa and S. aureus were reported. Nevertheless, the Cu(II)-BEPA-Pr catalyst suffers from similar O2 sensitivity problems as Cu(II)-Me3TACN, with a ~70 % decrease in NO flux when switching from N2 to 3% O2 sparge gasses. Even so, the performance and NO fluxes generated from the catalyst were still adequate to achieve significant anti-thrombotic and antimicrobial activities under ambient conditions.
2.2 ∣. INHALED NITRIC OXIDE AND CARDIOPULMONARY BYPASS SURGURY
The application of our E-NOgen technology using nitrite ions and Cu(II)-ligand complexes has also been studied and optimized for other potential medical applications including inhaled nitric oxide (INO) to treat persistent pulmonary hypertension (PPHN) and during cardiopulmonary bypass (CPB) surgery.67 The major risk factor of CPB is white blood cell activation due to the contact of extracorporeal blood with the ambient atmosphere during cardiotomy suction, and a corresponding systemic inflammatory response (SIR) during the procedure.68 Recently, our laboratories have sought to test E-NOgen as an alternate and cheaper source of NO(g) for medical procedures like PPHN and CBP. Using the E-NOgen principles established for the IV-catheter applications above, a portable gas-phase NO generator was fabricated and tested during CPB with extracorporeal circulation in a porcine model.
The first generation portable E-NOgen device consisted of an 80 mL bulk E-NOgen cell containing 7 mM Cu(II)-Me3TACN catalyst, 1.0 M NaNO2, and 0.5 M HEPES buffer (pH 7.3), as well as a high-surface area 50 cm2 Au-mesh working electrode and 25 cm2 Pt-mesh counter/reference electrode. In the optimal configuration for NO generation (Fig. 2B), the E-NOgen cell solution is circulated rapidly into a gas-extraction cell at flow rates up to 700 mL/min via a micro-liquid pump. A silicone-fiber-based gas exchanger rapidly extracts the highly permeable NO gas from the circulating bulk solution while the remaining solution is recycled back to the pump reservoir and ultimately to the E-NOgen cell to continue generating more NO from the nitrite reservoir. We found that this method of NO(g) extraction was superior to a previously reported configuration using an iron meso-tetrakis(4-N-methylpyridiniumyl)porphyrin as the nitrite reduction catalyst,69 whereby the NO extraction cell consisted of a bubbler containing the sweep gas with an outlet gas port for the extracted NO plus carrier gas. With the silicone-fiber based gas exchanger method, upon extraction/transport of NO through the walls of the hollow silicone fibers, a recipient carrier gas (N2 or air) delivers the NO for potential medical applications. To test the levels of NO being generated in the gas flow, the stream can be monitored continuously with NO/NO2 selective electrochemical sensors. At a set 0.1 L/min carrier-gas flow rate, output NO levels could be tuned between ~200 – 400 PPM (with N2 as the carrier gas) by varying the applied current to the E-NOgen cell from 5 – 30 μA, with only a 5-min response time between the adjustment of current and generation of steady-state levels of NO. The continuous generation of 400 PPM NO was sustained for up to 24 h. Upon using ambient air as the carrier gas, NO levels decreased by 50%, though NO can still be generated at suitable levels (up to ~180 PPM) for most desired medical applications. The use of ambient air as the carrier gas is ideal, as the device would eliminate the need for any NO and N2 gas cylinders, greatly enhancing its portability, cost, and utility. However, the reaction of NO with O2 presents a concern at these higher NO concentrations where the reactivity of NO with O2 could potentially generate hazardous levels of NO2. To test NO2 output levels, an experiment where NO was continuously generated at 60 PPM with air as the carrier gas was conducted. This experiment yielded relatively low concentrations of NO2 (< 1.5 PPM), consistent with the reaction rate of NO in air.70 The disproportionate decrease in NO output relative to NO2− detection suggests that the diminished NO release under increased concentrations of O2 in the sweep gas is primarily due to the O2 sensitivity of the Cu(I)-L catalyst and competition between NO2− and O2 binding to the reduced Cu(I)-L center.
The application of this E-NOgen device in CBP was tested in a porcine model where 500 PPM of NO was generated and delivered in a 1:1 N2:O2 carrier gas to the blood-oxygenator of a cardiotomy suction unit. CD11B expression in granulocytes and monocytes was monitored as a biomarker for the degree of white blood cell activation. In the experimental group that received 500 PPM of NO, normal values of CD11B in both granulocytes and monocytes were sustained during the 2 h operation, as well as during the subsequent 72 h of post-op monitoring. In contrast, a large increase in CD11B expression was measured in the control group during the procedure (up to 600% relative to the baseline in granulocytes and 250% in monocytes), suggesting an inflammatory response occurred during the operation (termed “systemic inflammatory response syndrome (SIRS)). These results provide promising evidence that the E-NOgen technology not only can be utilized specifically for CPB, but due to its robust tunability and low cost, could also be further adapted for other medical applications requiring airway delivery of gas phase NO (i.e., PPHH). The device does not require NO gas cylinders (which are expensive and unstable over long time periods) and can generate adequate concentrations of NO using ambient air as a carrier gas source.
2.3 ∣. WOUND-HEALING PATCHES
The antimicrobial properties of NO combined with many of its reported secondary benefits, e.g. inhibition of inflammatory responses, increasing angiogenesis, etc. have led to its testing in NO-based therapeutics for wound-healing applications.71 This approach is especially useful in cases of chronic wounds and infections, where antimicrobials or antibiotics would typically be utilized over long time periods. With increasing drug-resistant bacterial strains, alternative methods for chronic wound treatment, including NO, have been evaluated. Indeed, several reports utilizing the chemical release of NO have demonstrated that NO-releasing materials such as nanoparticles and hydrogels can be quite useful in wound-healing applications.72-75
We have reported the adaptation of our E-NOgen technology within silicone-rubber patches that can controllably release NO directionally from one surface of the patch (see Fig. 2C).76 The device consists of a 10-40 mm x 10-40 mm x 5-10 mm patch (several dimensions were fabricated/tested) made from polydimethylsiloxane (PDMS) housing, a cheap and flexible material. Within the housing lies: (1) a physical aluminum barrier to force directional NO release at the surface of the patch that adheres to the skin; (2) a Ag/AgCl mesh as the reference/counter electrode; (3) a porous PDMS spacer with an inner chamber housing a 4 mM Cu(II)-TMPA catalyst solution with 0.4 M NaNO2, 0.2 M NaCl, and 0.5 M HEPES (pH 7.2); (4) a gold-sputtered stainless steel mesh working electrode; and lastly (5) an outer surface that would contact the wound site that is sealed with a silicone rubber membrane and adhesive. Since portability is essential for such a device (where the E-NOgen application is not directly involved with ICU medical care), the electrode leads were tested with a portable, battery-powered, voltage supply that can provide potentials within a ± 1V range. Across a range of applied potentials between −0.1 V to −0.5 V, NO fluxes from 1 – 8 × 10−10 mol min−1 cm−2 were achieved (patch dimension: 40 x 40 x 6 mm). Additionally, continuous E-NOgen of fluxes between 8 – 5 × 10−10 mol min−1 cm−2 were sustained over 4 days, and recovery of 8 × 10−10 mol min−1 cm−2 fluxes were demonstrated upon replacement of the internal catalyst solution with a fresh nitrite/Cu(II)-TMPA stock solution after 4 days. In antimicrobial studies with the E-NOgen patches placed on blood agar plates, viable E. coli counts were reduced by 35% after 6 h of E-NOgen, 2 log units after 24 h of E-NOgen, and 3 log units after 48 h of E-NOgen (all with respect to control patches where E-NOgen was not powered on to provide NO). Similarly, viable S. aureus decreased by 20% after 6 h and 3 log units after 24 h of continuous E-NOgen. These results demonstrate that low-cost PDMS patches with mesh electrodes in contact with a solution of nitrite and Cu(II)-ligand complex can serve as an inexpensive and effective means of controllably inhibiting the proliferation of infectious bacteria at sites of chronic infection.
3 ∣. CONCLUSIONS
The beneficial biomedical properties of NO have been investigated for some time. A large body of the research to date has focused on the use of gaseous NO or NO-donor molecules (i.e. S-nitrosothiols or diazeniumdiolates) to deliver NO for vasodilation, antimicrobial and anti-thrombotic purposes. Indeed, inhaled NO gas is already approved for treating pulmonary hypertension in newborn babies and adults. Gas phase NO has also been used off-label in certain medical applications (CBP, see above) and shown to be beneficial in numerous others, but present technologies for NO administration are limited to emergency rooms/hospital settings due to portability and cost constraints. The recent advances in the design and testing of new biomedical devices that carry out E-NOgen highlight a potential means of making NO treatments much more cost efficient by using cheap/inert NaNO2 as an NO reservoir, and more portable by the design of compact electrochemical devices and portable voltage sources. For the devices discussed herein, one major limitation to the overall catalyst lifetime and concentration of NO generated is the O2 sensitivity of the Cu(II)-L catalysts in their reduced form. The continued tuning of the Cu(II)/Cu(I) reduction potentials via synthesis of new Cu(II)-ligand catalysts may provide a means to reduce or eliminate O2 sensitivity, allowing for significant elongation of catalyst lifetimes. At even more positive Cu(II)/Cu(I) potentials than those reported for our BMPA-/BEPA-carboxylate ligand series,65, 66 over-reduction of NO to N2O would certainly be mitigated, but a new balance between minimizing/eliminating catalyst O2 sensitivities and the decreasing turnover frequencies of such catalysts will need to be struck. For IV-catheter applications, catalysts that can tolerate up to 11-15% O2 (for use in more O2-rich arteries) while still generating NO fluxes of 1 – 4 × 10−10 mol min−1 cm−2 are desirable, while for CBP, INO, and wound-healing patches, O2 tolerance of up to 20% or greater would be ideal. For CBP/INO, this would allow for ambient air to be used as the carrier gas. Since the wound-healing patch design has no way to sparge or eliminate atmospheric O2 levels in the Cu(II)-L/NO2− reservoir, tolerance to ambient O2 levels would be even more desirable. Further catalyst development in this direction is in progress. Additionally, for NO-generating technologies to be used for medical purposes, the tight control of output NO levels is essential for their use in the field. While the Cu(II)-L catalysts discussed herein can generate a wide range of NO fluxes by tuning either the applied potential or current, the bulk catalyst solution slowly becomes more basic. This leads to a slow decrease in catalyst activity, and correspondingly, lower NO fluxes over time when holding at either a constant current or potential. The incorporation and optimization of feedback controls to allow for precise NO monitoring and automated adjustment of the applied currents/voltages to achieve consistent levels of NO generation would greatly benefit the development of these technologies for ultimate use in the medical field.
The physiological functions of NO and the potential for corresponding biomedical applications are numerous and extend beyond the anti-thrombotic, antimicrobial and anti-inflammatory emphasis of this review. For example, one recent publication reports the design of an implantable E-NOgen device that utilizes Fe3S4 nanoparticles containing a surface Pt-dopant to reduce a NaNO2 source to NO.77 Here, the device was implanted into a mouse brain and the effect of controlled electrochemical-mediated NO release on NO-dependent neuronal signaling was visualized. We expect that the continued optimization of cost, portability, tunability, and stability/lifetime of the catalysts will be essential in the advancement of existing and newer E-NOgen-based biomedical devices for applications in clinics at scale.
ACKNOWLEDGEMENTS
The authors acknowledge funding from the National Institute of Health (HL132037).
REFERENCES
- (1).Mellion BT, Ignarro LJ, Ohlstein EH, Pontecorvo EG, Hyman AL, Kadowitz PJ, Evidence for the inhibitory role of guanosine 3′, 5′ -monophosphate in ADP-induced human platelet aggregation in the presence of nitric oxide and related vasodilators. Blood. 1981, 57, 946–955. [PubMed] [Google Scholar]
- (2).Radomski MW, Palmer RM, Moncada S, Comparative pharmacology of endothelium-derived relaxing factor, nitric oxide and prostacyclin in platelets. Br. J. Pharmacol 1987, 92, 181–187. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (3).Moncada S, Palmer RM, Higgs EA, Nitric oxide: physiology, pathophysiology, and pharmacology. Pharmacol. Rev 1991, 43, 109–142 [PubMed] [Google Scholar]
- (4).Fang FC, Perspectives series: host/pathogen interactions. Mechanisms of nitric oxide-related antimicrobial activity. J. Clin. Invest 1997, 99, 2818–2825. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (5).Chaux A, Ruan XM, Fishbein MC, Ouyang Y, Kaul S, Pass JA, Matloff JM, Perivascular delivery of a nitric oxide donor inhibits neointimal hyperplasia in vein grafts implanted in the arterial circulation. J. Thorac. Cardiovasc. Surg 1998, 115, 604–614. [DOI] [PubMed] [Google Scholar]
- (6).Luo JD, Chen AF, Nitric oxide: a newly discovered function on wound healing. Acta Pharmacol. Sin 2005, 26, 259–264. [DOI] [PubMed] [Google Scholar]
- (7).Wallis JP, Nitric oxide and blood: a review. Transfus. Med 2005, 15, 1–11. [DOI] [PubMed] [Google Scholar]
- (8).Vaughn MW, Kuo L, Liao JC, Estimation of nitric oxide production and reaction rates in tissue by use of a mathematical model. Am. J. Physiol 1998, 274, H2163–2176. [DOI] [PubMed] [Google Scholar]
- (9).Rouby JJ, The nose, nitric oxide, and paranasal sinuses: the outpost of pulmonary antiinfectious defenses? Am. J. Respir. Crit Care Med 2003, 168, 265–266. [DOI] [PubMed] [Google Scholar]
- (10).Genevieve MH, Pradip KM, Emerging antimicrobial applications of nitric oxide (NO) and NO-releasing materials. Anti-Infect. Agents Med. Chem 2010, 9, 187–197. [Google Scholar]
- (11).Creagh-Brown BC, Griffiths MJD, Evans TW, Bench-to-bedside review: Inhaled nitric oxide therapy in adults. Crit. Care 2009, 13, 221–221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (12).Neonatal Inhaled Nitric Oxide Study Group. Inhaled nitric oxide in full-term and nearly full-term infants with hypoxic respiratory failure. N. Engl. J. Med 1997, 336, 597–604. [DOI] [PubMed] [Google Scholar]
- (13).Roberts JD, Fineman JR, Morin FC, Shaul PW, Rimar S, Schreiber MD, Polin RA, Zwass MS, Zayek MM, Gross I, Heymann MA, Zapol WM, Thusu KG, Zellers TM, Wylam ME, Zaslavsky A, Inhaled nitric oxide and persistent pulmonary hypertension of the newborn. N. Engl. J. Med 1997, 336, 605–610. [DOI] [PubMed] [Google Scholar]
- (14).Blomqvist H, Wickerts CJ, Andreen M, Ullberg U, Ortqvist A, Frostell C, Enhanced pneumonia resolution by inhalation of nitric oxide? Acta Anaesthesiol. Scand 1993, 37, 110–114. [DOI] [PubMed] [Google Scholar]
- (15).Dellinger RP, Trzeciak SW, Criner GJ, Zimmerman JL, Taylor RW, Usansky H, Young J, Goldstein B, Association between inhaled nitric oxide treatment and long-term pulmonary function in survivors of acute respiratory distress syndrome. Crit. Care 2012, 16, R36. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (16).Charriaut-Marlangue C, Bonnin P, Gharib A, Leger PL, Villapol S, Pocard M, Gressens P, Renolleau S, Baud O, Inhaled nitric oxide reduces brain damage by collateral recruitment in a neonatal stroke model. Stroke. 2012, 43, 3078–3084. [DOI] [PubMed] [Google Scholar]
- (17).Ratjen F, Gärtig S, Wiesemann HG, Grasemann H, Effect of inhaled nitric oxide on pulmonary function in cystic fibrosis. Respir. Med 1999, 93, 579–583. [DOI] [PubMed] [Google Scholar]
- (18).Long R, Jones R, Talbot J, Mayers I, Barrie J, Hoskinson M, Light B, Inhaled nitric oxide treatment of patients with pulmonary tuberculosis evidenced by positive sputum smears. Antimicrob. Agents Chemother 2005, 49, 1209–1212. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (19).Channick RN, Newhart JW, Johnson FW, Williams PJ, Auger WR, Fedullo PF, Moser KM, Pulsed delivery of inhaled nitric oxide to patients with primary pulmonary hypertension: an ambulatory delivery system and initial clinical tests. Chest. 1996, 109, 1545–1549. [DOI] [PubMed] [Google Scholar]
- (20).Pérez-Peñate GM, Juliá-Serdà G, Ojeda-Betancort N, García-Quintana A, Pulido-Duque J, Rodríguez-Pérez A, Cabrera-Navarro P, Gómez-Sánchez MA, Long-term inhaled nitric oxide plus phosphodiesterase 5 inhibitors for severe pulmonary hypertension. J. Heart Lung Transplant 2008, 27, 1326–1332. [DOI] [PubMed] [Google Scholar]
- (21).Yung GL, Kriett JM, Jamieson SW, Johnson FW, Newhart J, Kinninger K, Channick RN, Outpatient inhaled nitric oxide in a patient with idiopathic pulmonary fibrosis: a bridge to lung transplantation. J. Heart Lung Transplant 2001, 20, 1224–1227. [DOI] [PubMed] [Google Scholar]
- (22).Vonbank K, Ziesche R, Higenbottam TW, Stiebellehner L, Petkov V, Schenk P, Germann P, Block LH, Controlled prospective randomised trial on the effects on pulmonary haemodynamics of the ambulatory long term use of nitric oxide and oxygen in patients with severe COPD. Thorax. 2003, 58, 289–293. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (23).Lotz C, Muellenbach RM, Meybohm P, Mutlak H, Lepper PM, Rolfes CB, Peivandi A, Stumpner J, Kredel M, Kranke P, Torje I, Reyher C, Effects of inhaled nitric oxide in COVID-19-induced ARDS - Is it worthwhile? Acta. Anaesthesiol. Scand 2021, 65, 629–632. [DOI] [PubMed] [Google Scholar]
- (24).Parikh R, Wilson C, Weinberg J, Gavin D, Murphy J, Reardon CC, Inhaled nitric oxide treatment in spontaneously breathing COVID-19 patients. Ther. Adv. Respir. Dis 2020, 14, 1753466620933510. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (25).El-Sabbagh AM, Toomasian CJ, Toomasian JM, Ulysse G, Major T, Bartlett RH, Effect of air exposure and suction on blood cell activation and hemolysis in an in vitro cardiotomy suction model. ASAIO J. 2013, 59, 474–479. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (26).Laffey JG, Boylan JF, Cheng DC, The systemic inflammatory response to cardiac surgery: implications for the anesthesiologist. Anesthesiology 2002, 97, 215–252. [DOI] [PubMed] [Google Scholar]
- (27).Salis S, Mazzanti VV, Merli G, Salvi L, Tedesco CC, Veglia F, Sisillo E, Cardiopulmonary bypass duration is an independent predictor of morbidity and mortality after cardiac surgery. J. Cardiothorac. Vasc. Anesth 2008, 22, 814–822. [DOI] [PubMed] [Google Scholar]
- (28).Chello M, Mastroroberto P, Marchese AR, Maltese G, Santangelo E, Amantea B, Nitric oxide inhibits neutrophil adhesion during experimental extracorporeal circulation. Anesthesiology. 1998, 89, 443–448. [DOI] [PubMed] [Google Scholar]
- (29).Mellgren K, Friberg LG, Mellgren G, Hedner T, Wennmalm A, Wadenvik H, Nitric oxide in the oxygenator sweep gas reduces platelet activation during experimental perfusion. Ann. Thorac. Surg 1996, 61, 1194–1198. [DOI] [PubMed] [Google Scholar]
- (30).Checchia PA, Bronicki RA, Muenzer JT, Dixon D, Raithel S, Gandhi SK, Huddleston CB, Nitric oxide delivery during cardiopulmonary bypass reduces postoperative morbidity in children--a randomized trial. J. Thorac. Cardiovasc. Surg 2013, 146, 530–536. [DOI] [PubMed] [Google Scholar]
- (31).James C, Millar J, Horton S, Brizard C, Molesworth C, Butt W, Nitric oxide administration during paediatric cardiopulmonary bypass: A randomised controlled trial. Intensive Care Med. 2016, 42, 1744–1752. [DOI] [PubMed] [Google Scholar]
- (32).Cole FS, Alleyne C, Barks JD, Boyle RJ, Carroll JL, Dokken D, Edwards WH, Georgieff M, Gregory K, Johnston MV, Kramer M, Mitchell C, Neu J, Pursley DM, Robinson WM, Rowitch DH, NIH Consensus Development Conference statement: inhaled nitric-oxide therapy for premature infants. Pediatrics. 2011, 127, 363–369. [DOI] [PubMed] [Google Scholar]
- (33).Shen Q, Hedberg K, Investigation of the equilibrium N2O4 ⇌ 2NO2 by electron diffraction: Molecular structures and effective temperature and pressure of the expanding gas with implications for studies of other dimer–monomer equilibria. J. Phys. Chem. A 1998, 102, 6470–6476. [Google Scholar]
- (34).Merkle AC, Lehnert N, Binding and activation of nitrite and nitric oxide by copper nitrite reductase and corresponding model complexes. Dalton Trans. 2012, 41, 3355–3368. [DOI] [PubMed] [Google Scholar]
- (35).Horrell S, Kekilli D, Strange RW, Hough MA, Recent structural insights into the function of copper nitrite reductases. Metallomics. 2017, 9, 1470–1482. [DOI] [PubMed] [Google Scholar]
- (36).O'Grady NP, Alexander M, Dellinger EP, Gerberding JL, Heard SO, Maki DG, Masur H, McCormick RD, Mermel LA, Pearson ML, Raad II, Randolph A, Weinstein RA, guidelines for the prevention of intravascular catheter-related infections. Centers for Disease Control and Prevention. Morbidity Mortality Wkly. Rep 2002, 51, 1–26. [PubMed] [Google Scholar]
- (37).Wo Y, Brisbois EJ, Bartlett RH, Meyerhoff ME, Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO). Biomater. Sci 2016, 4, 1161–1183. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (38).Varu VN, Tsihlis ND, Kibbe MR, Basic science review: Nitric oxide releasing prosthetic materials. Vasc. Endovasc. Surg 2009, 43, 121–131. [DOI] [PubMed] [Google Scholar]
- (39).Schoenfisch MH, Mowery KA, Rader MV, Baliga N, Wahr JA, Meyerhoff ME, Improving the thromboresistivity of chemical sensors via nitric oxide release: Fabrication and in vivo evaluation of NO-releasing oxygen-sensing catheters. Anal. Chem 2000, 72, 1119–1126. [DOI] [PubMed] [Google Scholar]
- (40).Zhang H, Annich GM, Miskulin J, Osterholzer K, Merz SI, Bartlett RH, Meyerhoff ME, Nitric oxide releasing silicone rubbers with improved blood compatibility: preparation, characterization, and in vivo evaluation. Biomaterials. 2002, 23, 1485–1494. [DOI] [PubMed] [Google Scholar]
- (41).Coneski PN, Schoenfisch MH, Synthesis of nitric oxide-releasing polyurethanes with S-nitrosothiol-containing hard and soft segments. Polym. Chem 2011, 2, 906–913. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (42).Parzuchowski PG, Frost MC, Meyerhoff ME, Synthesis and characterization of polymethacrylate-based nitric oxide donors. J. Am. Chem. Soc 2002, 124, 12182–12191. [DOI] [PubMed] [Google Scholar]
- (43).Buergler JM, Tio FO, Schulz DG, Khan MM, Mazur W, French BA, Raizner AE, Ali NM, Use of nitric-oxide-eluting polymer-coated coronary stents for prevention of restenosis in pigs. Coron. Artery Dis 2000, 11, 351–357. [DOI] [PubMed] [Google Scholar]
- (44).Zhang H, Annich GM, Miskulin J, Stankiewicz K, Osterholzer K, Merz SI, Bartlett RH, Meyerhoff ME, Nitric oxide-releasing fumed silica particles: synthesis, characterization, and biomedical application. J. Am. Chem. Soc 2003, 125, 5015–5024. [DOI] [PubMed] [Google Scholar]
- (45).Reynolds MM, Hrabie JA, Oh BK, Politis JK, Citro ML, Keefer LK, Meyerhoff ME, Nitric oxide releasing polyurethanes with covalently linked diazeniumdiolated secondary amines. Biomacromolecules. 2006, 7, 987–994. [DOI] [PubMed] [Google Scholar]
- (46).Wu Y, Rojas AP, Griffith GW, Skrzypchak AM, Lafayette N, Bartlett RH, Meyerhoff ME, Improving blood compatibility of intravascular oxygen sensors via catalytic decomposition of S-nitrosothiols to generate nitric oxide in situ. Sens. Actuators B Chem 2007, 121, 36–46. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (47).Major TC, Brant DO, Burney CP, Amoako KA, Annich GM, Meyerhoff ME, Handa H, Bartlett RH, The hemocompatibility of a nitric oxide generating polymer that catalyzes S-nitrosothiol decomposition in an extracorporeal circulation model. Biomaterials. 2011, 32, 5957–5969. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (48).Jen MC, Serrano MC, van Lith R, Ameer GA, Polymer-based nitric oxide therapies: recent insights for biomedical applications. Adv. Funct. Mater 2012, 22, 239–260. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (49).Handa H, Brisbois EJ, Major TC, Refahiyat L, Amoako KA, Annich GM, Bartlett RH, Meyerhoff ME, In vitro and in vivo study of sustained nitric oxide release coating using diazeniumdiolate-doped poly(vinyl chloride) matrix with poly(lactide-co-glycolide) additive. J. Mater. Chem. B 2013, 1, 3578–3587. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (50).Handa H, Major TC, Brisbois EJ, Amoako KA, Meyerhoff ME, Bartlett RH, Hemocompatibility comparison of biomedical grade polymers using rabbit thrombogenicity model for preparing nonthrombogenic nitric oxide releasing surfaces. J. Mater. Chem. B 2014, 2, 1059–1067. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (51).Colletta A, Wu J, Wo Y, Kappler M, Chen H, Xi C, Meyerhoff ME, S-Nitroso-N-acetylpenicillamine (SNAP) impregnated silicone foley catheters: A potential biomaterial/device to prevent catheter-associated urinary tract infections. ACS Biomater. Sci. Eng 2015, 1, 416–424. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (52).Wolf AK, Qin Y, Major TC, Meyerhoff ME, Improved thromboresistance and analytical performance of intravascular amperometric glucose sensors using optimized nitric oxide release coatings. Chin. Chem. Lett 2015, 26, 464–468. [Google Scholar]
- (53).Ketchum AR, Kappler MP, Wu J, Xi C, Meyerhoff ME, The preparation and characterization of nitric oxide releasing silicone rubber materials impregnated with S-nitroso-tert-dodecyl mercaptan. J. Mater. Chem. B 2016, 4, 422–430. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (54).Höfler L, Koley D, Wu J, Xi C, Meyerhoff ME, Electromodulated release of nitric oxide through polymer material from reservoir of inorganic nitrite salt. RSC Adv. 2012, 2, 6765–6767. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (55).Coneski PN, Schoenfisch MH, Nitric oxide release: Part III. Measurement and reporting. Chem. Soc. Rev 2012, 41, 3753–3758. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (56).Ren H, Colletta A, Koley D, Wu J, Xi C, Major TC, Bartlett RH, Meyerhoff ME, Thromboresistant/anti-biofilm catheters via electrochemically modulated nitric oxide release. Bioelectrochemistry. 2015, 104, 10–16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (57).Kukimoto M, Nishiyama M, Murphy MEP, Turley S, Adman ET, Horinouchi S, Beppu T, X-ray structure and site-directed mutagenesis of a nitrite reductase from alcaligenes faecalis S-6: Roles of two copper atoms in nitrite reduction. Biochemistry. 1994, 33, 5246–5252. [DOI] [PubMed] [Google Scholar]
- (58).Adman ET, Godden JW, Turley S, The structure of copper-nitrite reductase from Achromobacter cycloclastes at five pH values, with NO2- bound and with type II copper depleted. J. Biol. Chem 1995, 270, 27458–27474. [DOI] [PubMed] [Google Scholar]
- (59).Hsu SCN, Chang Y-L, Chuang W-J, Chen H-Y, Lin IJ, Chiang MY, Kao C-L, Chen H-Y, Copper(I) nitro complex with an anionic [HB(3,5-Me2Pz)3]– ligand: A synthetic model for the copper nitrite reductase active site. Inorg. Chem 2012, 51, 9297–9308. [DOI] [PubMed] [Google Scholar]
- (60).Kumar M, Dixon NA, Merkle AC, Zeller M, Lehnert N, Papish ET, Hydrotris(triazolyl)borate complexes as functional models for Cu nitrite reductase: The electronic influence of distal nitrogens. Inorg. Chem 2012, 51, 7004–7006. [DOI] [PubMed] [Google Scholar]
- (61).Chang Y-L, Lin Y-F, Chuang W-J, Kao C-L, Narwane M, Chen H-Y, Chiang MY, Hsu SCN, Structure and nitrite reduction reactivity study of bio-inspired copper(i)–nitro complexes in steric and electronic considerations of tridentate nitrogen ligands. Dalton Trans. 2018, 47, 5335–5341. [DOI] [PubMed] [Google Scholar]
- (62).Ren H, Wu J, Xi C, Lehnert N, Major T, Bartlett RH, Meyerhoff ME, Electrochemically modulated nitric oxide (NO) releasing biomedical devices via copper(II)-tri(2-pyridylmethyl)amine mediated reduction of nitrite. ACS Appl. Mater. Interfaces 2014, 6, 3779–3783. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (63).Ren H, Bull JL, Meyerhoff ME, Transport of nitric oxide (NO) in various biomedical grade polyurethanes: measurements and modeling impact on NO release properties of medical devices. Acs Biomater. Sci. Eng 2016, 2, 1483–1492. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (64).Ren H, Coughlin MA, Major TC, Aiello S, Rojas Pena A, Bartlett RH, Meyerhoff ME, Improved in vivo performance of amperometric oxygen (PO2) sensing catheters via electrochemical nitric oxide generation/release. Anal. Chem 2015, 87, 8067–8072. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (65).Konopińska KK, Schmidt NJ, Hunt AP, Lehnert N, Wu J, Xi C, Meyerhoff ME, Comparison of copper(II)–ligand complexes as mediators for preparing electrochemically modulated nitric oxide-releasing catheters. ACS Appl. Mater. Interfaces 2018, 10, 25047–25055. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (66).Hunt AP, Batka AE, Hosseinzadeh M, Gregory JD, Haque HK, Ren H, Meyerhoff ME, Lehnert N, Nitric oxide generation on demand for biomedical applications via electrocatalytic nitrite reduction by copper BMPA- and BEPA-carboxylate complexes. ACS Catal. 2019, 9, 7746–7758. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (67).Qin Y, Zajda J, Brisbois EJ, Ren H, Toomasian JM, Major TC, Rojas-Pena A, Carr B, Johnson T, Haft JW, Bartlett RH, Hunt AP, Lehnert N, Meyerhoff ME, Portable nitric oxide (NO) generator based on electrochemical reduction of nitrite for potential applications in inhaled NO therapy and cardiopulmonary bypass surgery. Mol. Pharmaceutics 2017, 14, 3762–3771. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (68).Lau K, Shah H, Kelleher A, Moat N, Coronary artery surgery: cardiotomy suction or cell salvage? J. Cardiothorac. Surg 2007, 2, 46. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (69).Chi Y, Chen J, Aoki K, Electrochemical generation of free nitric oxide from nitrite catalyzed by iron meso-tetrakis(4-N-methylpyridiniumyl)porphyrin. Inorg. Chem 2004, 43, 8437–8446. [DOI] [PubMed] [Google Scholar]
- (70).Schedin U, Frostell CG, Gustafsson LE, Formation of nitrogen dioxide from nitric oxide and their measurement in clinically relevant circumstances. Br. J. Anaesth 1999, 82, 182–192. [DOI] [PubMed] [Google Scholar]
- (71).Carpenter AW, Schoenfisch MH, Nitric oxide release: part II. Therapeutic applications. Chem. Soc. Rev 2012, 41, 3742–3752. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (72).Bokhari AR, Murrell GA, The role of nitric oxide in tendon healing. J. Shoulder Elbow Surg 2012, 21, 238–244. [DOI] [PubMed] [Google Scholar]
- (73).Champeau M, Póvoa V, Militão L, Cabrini FM, Picheth GF, Meneau F, Jara CP, de Araujo EP, de Oliveira MG, Supramolecular poly(acrylic acid)/F127 hydrogel with hydration-controlled nitric oxide release for enhancing wound healing. Acta Biomater. 2018, 74, 312–325. [DOI] [PubMed] [Google Scholar]
- (74).Santos MI, da Silva LCE, Bomediano MP, Catori DM, Gonçalves MC, de Oliveira MG, 3D printed nitric oxide-releasing poly(acrylic acid)/F127/cellulose nanocrystal hydrogels. Soft Matter. 2021, 17, 6352–6361. [DOI] [PubMed] [Google Scholar]
- (75).Yang C, Hwang HH, Jeong S, Seo D, Jeong Y, Lee DY, Lee K, Inducing angiogenesis with the controlled release of nitric oxide from biodegradable and biocompatible copolymeric nanoparticles. Int. J. Nanomed 2018, 13, 6517–6530. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (76).Lee WH, Ren H, Wu J, Novak O, Brown RB, Xi C, Meyerhoff ME, Electrochemically modulated nitric oxide release from flexible silicone rubber patch: Antimicrobial activity for potential wound healing applications. ACS Biomater. Sci. Eng 2016, 2, 1432–1435. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (77).Park J, Jin K, Sahasrabudhe A, Chiang P-H, Maalouf JH, Koehler F, Rosenfeld D, Rao S, Tanaka T, Khudiyev T, Schiffer ZJ, Fink Y, Yizhar O, Manthiram K, Anikeeva P, In situ electrochemical generation of nitric oxide for neuronal modulation. Nat. Nanotechnol 2020, 15, 690–697. [DOI] [PMC free article] [PubMed] [Google Scholar]