Abstract
Background: Intraocular pressure (IOP) is influenced by body position. The purpose of this study is to compare the IOP measurements obtained with two different devices, to investigate IOP changes in standing, sitting, and supine positions. Methods: In this comparative prospective case series, IOP was measured in sitting, supine, prone, and standing (standing 1) positions and again five minutes after standing (standing 2), utilizing an Icare Pro (ICP) and a Tono-Pen Avia (TPA) in the 64 eyes of 32 healthy subjects. Results: Compared to the sitting position, both devices showed an increase in the IOP both in supine and standing 2 positions (p < 0.05). The mean IOP difference between the two devices was: in the sitting position, 0.57 ± 2.10 mmHg (range: −3.80 to 6.60 mmHg) (p < 0.05), in the supine position, 0.93 ± 2.49 mmHg (range: −4.50 to 7.10 mmHg) (p < 0.05), in the standing 1 position, 0.37 ± 1.96 mmHg (range: −5.20 to 5.00 mmHg) (p = 0.102), and in the standing 2 position 0.73 ± 2.03 mmHg (range: −4.5 to 6.4 mmHg) (p < 0.001). Conclusions: The results highlight an agreement between the TPA and ICP, both confirming not only the increase in IOP in the supine position, but also showing an increase in the standing 2 position. Therefore, it is suggested to perform such measurements in patients with glaucoma, to explain its progression in an apparently normal tension or in compensated patients.
Keywords: intraocular pressure, Icare Pro, standing position, supine position, Tono-Pen Avia
1. Introduction
During an ophthalmic examination, intraocular pressure (IOP) is routinely measured as it is a well-known risk factor for glaucoma onset and progression. [1,2,3]. A single measurement is not sufficient to rule out the development of this condition, as the IOP has circadian variations that can influence the progression of glaucoma [4,5,6]. The body position was suggested as one of the causes of this fluctuation; in fact, several studies have shown that the IOP could significantly rise when a person is lying down compared to when they are in the sitting position, which could explain some nocturnal elevations [7,8]. During the day, several people spend hours in the standing position, so it might be useful to know if a change in IOP occurs in this situation. In a previous paper, utilizing an I Care Pro (ICP) Tonometer, we were able to prove that both in the supine and in the standing position, there was an increase in IOP [9], but we were not able to find other studies that reported a comparison with other instruments, that confirmed such a finding in standing position. The IOP measurements are mainly performed with Goldmann applanation tonometry (GAT), which is nowadays considered the gold standard for these measurements, but this device can be used only with the patient seated at a slit lamp [10]. Some prototypes of GAT have been designed to be used in the reclining position [11], but these are not available on the market. On the contrary, there are several hand-held IOP measuring devices that can be used in a non-seated position.
The purpose of this study was to compare IOP measurements obtained with different devices, to check if they both confirmed these findings and their comparability. For such a purpose, we decided to compare two devices: Icare PRO (ICP) and Tono-Pen AVIA (TPA).
2. Materials and Methods
2.1. Selection of Patients
The study was carried out in adherence to tenets of the World Medical Association’s Declaration of Helsinki, Institutional Review Board (IRB) approval was obtained (Cometico Campania Sud, Italy, prot. n° 16544), and an informed consent was achieved from all participants included in the study. Each subject underwent a general physical checkup, demonstrating the absence of any systemic disease. Patients with systemic and ocular diseases which could potentially interfere with the purpose of the study, such as diabetes, collagenopathies, dry eyes, keratoconus, uveitis, corneal and lens opacities, glaucoma, previous intraocular or corneal surgery, pregnancy, and an inability to fixate on the target, were excluded from the study. All participants underwent a complete ophthalmological examination, including postoperative corrected distance visual acuity, a refractive error, and a fundus examination. Moreover, all the patients did not have a smoking history or an alcohol addiction.
2.2. IOP Measurement
The IOP measurement was performed in the time slot between 10 am and 12 pm by two different observers, who were not aware of the results obtained from the other. The measurements were obtained in the following order: the sitting position, supine after five minutes of lying down, standing (standing 1), and five minutes after the patient was staying in the standing position (standing 2).
IOP was measured with the ICP (Icare Finland Oy, Finland version 1.1) and then with the TPA (Reichert Inc., Depew, NY, USA). Although the ICP does not require anesthesia, before the first measurement, both eyes were anesthetized with oxybuprocaine eye drops, to avoid differences in performing the two measurements, even if it has been shown that oxybuprocaine does not change corneal thickness [12].
The ICP measures the IOP, converting the acceleration of a small probe pushed against the cornea [13]. This procedure does not require local anesthesia or calibration, due to a very brief contact with the corneal surface.
The ICP provides the mean IOP from six consecutive tests, showing the accuracy by a color code shown below the value of the IOP. In case of being within the normal limits of variance, the indicator is green; if the variance is greater than normal, it is yellow; and if the variation is unacceptably high, it is red. In these two last cases, the measurements should be carefully considered [11].
Measurements were accepted only when the indicator was green, otherwise they were repeated.
The TPA uses the same physical principle as the GAT to measure the IOP, but the applanated area is much smaller, since the transducer tip, covered with a new latex tip cover before each measurement, has a diameter of 1.0 mm.
The mean IOP readings were automatically averaged by the instrument, when ten valid readings were obtained, by lightly touching the central cornea. The measurement was shown on the liquid crystal display, which is situated on the side of the device, and together it displayed the “statistical confidence indicator”, indicating that the standard deviation of the valid measurements is 5% or less of the number shown. The higher the value, the more reliable the measurement is. Only values higher than 90 were accepted.
2.3. Central Corneal Thickness and Axial Eye Length
To check if potential differences were related to some ocular parameters, such as the central corneal thickness (CCT) or axial eye length (AL), the patients were examined with a Pentacam HR (Oculus, Wetzlar, Germany, version 1.19r11) and an IOLMaster (Zeiss, Jena, Germany, version 5.4.4.00006) [14,15].
2.4. Statistical Analysis
The exact Kolmogorov–Smirnov test showed a normal distribution (p > 0.05) for all data. For this reason, a paired t-test was used for all statistical evaluations. Moreover, Bland–Altman plots were also performed. A p-value less than 0.05 was considered statistically significant.
A correlation between the TCP and TPA measurements in all position were evaluated by Pearsons’s correlation coefficient (r). In addition, a correlation between the CCT and the difference between the ICP and TPA in all positions, and between the AL and the difference between the ICP and TPA in all positions, were evaluated by the same test. The required sample size was calculated with G*Power software (Version 3.1.9.7, Faul, Erdfelder, Lang, and Buchner, 2020, Available at https://www.gpower.hhu.de (accessed on 9 July 2022)). It was estimated that with a significance level of 5% and a test power of 80%, a sample size of 64 eyes would be necessary to detect a difference in the mean error of 0.50 mmHg, given a within-subject standard deviation (SD) for IOP equal to 1.50 mmHg.
3. Results
3.1. Patient Characteristics
Sixty-four eyes of 32 patients (16 males) aged between 21 and 55 years (mean 29 ± 9 years) and with a refractive error, in terms of being spherically equivalent, between −6.63 and +1 D (mean: −2.32 ± 2.19 D), were included in this comparative prospective observational study.
3.2. Comparison between TPA and ICP IOP Measurements
Table 1 summarizes the IOP measurements obtained by the two instruments in the different positions, showing that there were statistically significant differences between the IOP measurements obtained by the same device (all p < 0.05), except for the difference in IOP obtained by the TPA between the sitting and standing 1 position (p = 0.187).
Table 1.
Sitting | Supine | Standing 1 | Standing 2 | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
ICP | TPA | Δ | ICP | TPA | Δ | ICP | TPA | Δ | ICP | TPA | Δ | |
Mean | 14.97 | 14.34 | 0.57 | 16.58 | 15.61 | 0.93 | 14.45 | 14.03 | 0.37 | 15.70 | 14.88 | 0.73 |
CI 95% | [14.30; 15.65] |
[13.57; 15.12] |
[0.03; 1.12] |
[15.88; 17.30] |
[14.75; 16.47] |
[0.29; 1.58] |
[13.83; 15.07] |
[13.31; 14.75] |
[−0.13; 0.88] |
[14.94; 16.45] |
[14.04; 15.71] |
[0.20; 1.24] |
SD | 2.71 | 3.10 | 2.10 | 2.83 | 3.44 | 2.49 | 2.49 | 2.89 | 1.96 | 3.04 | 3.36 | 2.03 |
Min | 8.60 | 8.00 | −3.80 | 9.10 | 10.00 | −4.50 | 8.30 | 9.00 | −5.20 | 8.50 | 9.00 | −4.50 |
Max | 22.80 | 22.00 | 6.60 | 23.90 | 22.00 | 7.10 | 20.40 | 22.00 | 5.00 | 23.00 | 23.00 | 6.40 |
p value | - | - | 0.019 | - | - | 0.003 | - | - | 0.102 | - | - | 0.002 |
The IOP is higher in the standing 2 and supine position than in the sitting one, both with the ICP and TPA (all p < 0.05). It is interesting to note that, switching from supine to standing 1, there is a significant decrease in IOP (p < 0.001), although such a decrease is not significant when compared to the sitting one with the TPA (p = 0.187). Lastly, after 5 min standing, the IOP becomes higher than in the sitting position (p < 0.05) with both instruments. The comparison between the two devices indicated in Figure 1a–d and in Table 1 shows that, except in the standing 1 position, the measurements are significantly different, with the ICP values being higher than the TPA ones.
3.3. Correlation between TCP and TPA, Correlation between CCT, AL and IOP Measurements
There was a strong positive correlation between the measurements obtained by the ICP and TPA in all positions (all r ≥ 0.699, all p < 0.05). The CCT ranged from 470 to 598 µ (mean: 539.95 ± 30.36 µ) and the AL ranged from 21.70 to 27.16 mm (mean: 24.45 ± 1.29 mm). The correlations between the CCT and the difference between the ICP and TPA in all positions are shown in Table 2.
Table 2.
Difference between ICP and TPA | ||||
---|---|---|---|---|
Sitting | Supine | Standing 1 | Standing 2 | |
CCT | −0.084 | −0.084 | −0.166 | −0.083 |
AL | −0.182 | −0.184 | 0.024 | −0.018 |
There was a weak negative, but not a statistically significant, correlation between the CCT and the difference between the two devices in all positions (all p > 0.05). The correlations between the AL and the difference between the ICP and TPA in all positions are shown in Table 2. In the same way, there was no correlation between the AL and the difference between the two devices in all positions (all p > 0.05).
4. Discussion
In a literary review on PubMed, we found few papers where the IOP was measured with different devices in the sitting and supine position, such as the TPA [11], ICP [16], and Tono-Pen [17,18,19,20,21], and even less comparing different the tonometers [11,22,23], but we were not able to find IOP measurements in the standing position in the references we studied.
Schweier et al. [11] measured IOP in the sitting position and then 10 min after in the reclining position, in the 36 eyes of 36 healthy people with the ICP, TPA, and GAT. They noticed that IOP in the sitting position was lower than the reclining one with both the handheld tonometers and the mean difference with the TPA (1.8 mmHg) being greater than with the ICP (0.8 mmHg).
Barkana et al. [22] tested, with the Tono-Pen XL and ICP, the IOP in the sitting position, and then after 10 min, tested it in the supine position, in 21 eyes of 21 healthy subjects. They observed a mean rise in the IOP of 0.9 mmHg from the sitting to the supine position with the Tono-Pen, but they found a mean decrease in 0.9 mmHg from the sitting to the supine position with the ICP, as compared to Schweier.
Lee et al. [24] compared the IOP measurements obtained using the Icare Pro rebound tonometer and Tono-Pen XL tonometer in the supine and right and left lateral decubitus body positions in the one hundred eyes of 50 healthy volunteers or glaucoma suspects. The IOP readings obtained using the Icare Pro and Tono-Pen showed good correlations in these two positions, although the Icare Pro readings were higher than the Tono-Pen readings in all of them.
Concerning the differences between the sitting and the supine position, our results confirm those obtained by Schweier [11] and Lee [24] but are in contrast with those obtained by Barkana [22]. Indeed, we found an increase in IOP with both the ICP and TPA, when changing the position from sitting to supine, even though we waited 5 min instead of 10.
Our results confirm also those ones of Nakakura et al. [23], who recently found an increase in IOP in the supine compared to in the sitting position using the ICP, Tono-Pen XL, and Kowa.
The present study confirms the results obtained with the ICP in a recently published paper by De Bernardo et al. Indeed, they observed a non-significant decrease in IOP from the sitting to the standing 1 position and a small but significant increase in IOP in the standing 2 position relative to the sitting position [9].
Several reasons could explain the increase in IOP in the supine position. Some authors suggest that there is a choroidal vascular engorgement caused by the redistribution of body fluids [25,26], while others suggest that there is an increase in the episcleral venous pressure [27,28,29].
Regarding the differences between the two devices (TPA e ICP) in the standing position, no comparison can be made within the international literature because we were not able to find similar papers.
Interestingly, both devices detected an IOP decrease immediately after standing, and an IOP rise in the standing position after five minutes.
The IOP rise in the standing position is slightly difficult to explain, as the IOP regulation’s exact mechanism is uncertain. It should be subject to systemic vascular monitoring, that could also explain the slight decrease when going from the supine to the standing position, that conversely becomes an increase 5 min after maintaining the standing position. [30,31,32,33].
The measurements obtained by the two devices (Tono-Pen Avia and ICare Pro) were different, but the detection of an increase and decrease was similar. The dissimilarity of the measures was not related to differences in the CCT or AL. However, our results confirm the importance of ensuring that measurements should be taken with the same instrument when monitoring ocular hypertension or glaucoma.
One criticism that could be made of this study is that we used the ICP and TPA, but not the GAT, which represents the gold standard in measuring IOP.
This criticism could be overcome because, even if the GAT is considered to be the gold standard, it is well known that several corneal parameters such as biomechanical properties [33,34], CCT [35,36,37,38], corneal irregularities [10], and refractive corneal surgery [38] can influence the GAT measurements. These limits could also be present with other tonometers, but the real problem is that to perform such measurements with the GAT in different positions is quite difficult, and as far as we know, no papers have considered it as a gold standard in the standing and supine positions too.
In addition, several papers compared the GAT with TPA or ICP in the sitting position [11,17,22]. Schweier et al. [11] founded a mean IOP measured by the GAT of 14.9 ± 3.5 mmHg, compared to a mean IOP measured by the ICP of 15.6 ± 3.1 mmHg, and with the TPA, of 14.8 ± 2.7 mmHg in the sitting position [11]. Barkana’s studies [17,22] showed, in the sitting position, a mean IOP for the GAT measurements lower than the Icare Pro and TonoPenXL measurements.
The findings show a slight agreement between the ICare Pro and Tono-Pen Avia.
Further studies in a larger group of patients comparing the IOP changes with parameters such as one’s age, gender, race (African ancestry), changes in blood pressure in different positions [27,28,29], nuclear sclerosis, iris color, myopia, EtOH use, and smoking could help in understanding the reasons of such an increase.
In order to better understand the development of glaucoma in apparently normal tension or in compensated patients, the IOP scale in the standing position along with the one in the supine position should be incorporated into clinical practice. In fact, certain occupations (e.g., waiters, policemen, and chefs) require a considerable amount of standing time, and our research not only confirms the increase in IOP in the supine position but, more importantly, indicates an increase in the standing position.
Future studies including glaucomatous patients could be helpful in terms of knowing whether the IOP fluctuations are greater with changes in the body’s position.
Author Contributions
Conceptualization, N.R.; data curation, M.D.B.; formal analysis, M.B. and F.C.; funding acquisition, M.D.B.; investigation, F.C.; methodology, N.R.; project administration, M.D.B.; resources, G.A.; software, G.A.; supervision, N.R.; validation, M.D.B. and N.R.; visualization, N.R.; writing—original draft, G.A.; writing—review and editing, M.B. and M.D.S. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
This study was conducted according to the guidelines of the Declaration of Helsinki and approved by the Ethics Committee “Cometico Campania Sud” (CECS); Prot. Number 16544.
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study.
Data Availability Statement
The data presented in this study are available on request from the corresponding author.
Conflicts of Interest
The authors declare no conflict of interest.
Funding Statement
FARB grant—University of Salerno.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Heijl A., Leske M.C., Bengtsson B., Hyman L., Bengtsson B., Hussein M. Early Manifest Glaucoma Trial Group. Reduction of intraocular pressure and glaucoma progression: Results from the Early Manifest Glaucoma Trial. Acta Ophthalmol. 2002;120:1268–1279. doi: 10.1001/archopht.120.10.1268. [DOI] [PubMed] [Google Scholar]
- 2.Kass M.A., Heuer D.K., Higginbotham E.J., Johnson C.A., Keltner J.L., Miller J.P. The Ocular Hypertension Treatment Study: A randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Acta Ophthalmol. 2002;120:701–713. doi: 10.1001/archopht.120.6.701. [DOI] [PubMed] [Google Scholar]
- 3.Leske M.C., Heijl A., Hussein M., Bengtsson B., Hyman L., Komaroff E. Early Manifest Glaucoma Trial Group. Factors for glaucoma progression and the effect of treatment: The Early Manifest Glaucoma Trial. Acta Ophthalmol. Scand. 2003;121:48–56. doi: 10.1001/archopht.121.1.48. [DOI] [PubMed] [Google Scholar]
- 4.Asrani S., Zeimer R., Wilensky J., Gieser D., Vitale S., Lindenmuth K. Large diurnal fluctuations in intraocular pressure are an independent risk factor in patients with glaucoma. J. Glaucoma. 2000;9:134–142. doi: 10.1097/00061198-200004000-00002. [DOI] [PubMed] [Google Scholar]
- 5.Caprioli J. Intraocular pressure fluctuation: An independent risk factor for glaucoma? Acta Ophthalmol. 2007;125:1124–1125. doi: 10.1001/archopht.125.8.1124. [DOI] [PubMed] [Google Scholar]
- 6.Caprioli J., Coleman A.L. Intraocular pressure fluctuation: A risk factor for visual field progression at low intraocular pressures in the Advanced Glaucoma Intervention Study. Ophthalmology. 2008;115:1123–1129. doi: 10.1016/j.ophtha.2007.10.031. [DOI] [PubMed] [Google Scholar]
- 7.Tarkkanen A., Leikola J. Postural variations of the intraocular pressure as measured with the Mackay-Marg tonometer. Acta Ophthalmol. 1967;45:569–575. doi: 10.1111/j.1755-3768.1967.tb06523.x. [DOI] [PubMed] [Google Scholar]
- 8.Jain M.R., Marmion V.J. Rapid pneumatic and Mackey-Marg applanation tonometry to evaluate the postural effect on intraocular pressure. Br. J. Ophthalmol. 1976;60:687–693. doi: 10.1136/bjo.60.10.687. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.De Bernardo M., Borrelli M., Cembalo G., Rosa N. Intraocular Pressure Measurements in Standing Position with a Rebound Tonometer. Med. 2019;55:701. doi: 10.3390/medicina55100701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Whitacre M.M., Stein R. Sources of error with use of Goldmann-type tonometers. Surv. Ophthalmol. 1993;38:1–30. doi: 10.1016/0039-6257(93)90053-A. [DOI] [PubMed] [Google Scholar]
- 11.Schweier C., Hanson J.V.M., Funk J., Töteberg-Harms M. Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions. BMC Ophthalmol. 2013;13:44. doi: 10.1186/1471-2415-13-44. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Rosa N., De Bernardo M., Borrelli M., Filosa M.L., Lanza M. Effect of oxybuprocaine eye drops on corneal volume and thickness measurements. Optom. Vis. Sci. 2011;88:640–644. doi: 10.1097/OPX.0b013e3182114303. [DOI] [PubMed] [Google Scholar]
- 13.Kontiola A.I. A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol. 2000;78:142–145. doi: 10.1034/j.1600-0420.2000.078002142.x. [DOI] [PubMed] [Google Scholar]
- 14.De Bernardo M., Salerno G., Conretta P., Rosa N. Axial Lenght Shortening After Cataract Surgery: New Approach to Solve the Question. Transl. Vis. Sci. Technol. 2018;7:34. doi: 10.1167/tvst.7.6.34. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.De Bernardo M., Cornetta P., Marotta G., Salerno G., De Pascale I., Rosa N. Measurement of corneal thickness using Pentacam HR versus Nidek CEM-530 specular microscopy. J. Int. Med. Res. 2020;48:300060519892385. doi: 10.1177/0300060519892385. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Lee T.E., Yoo C., Kim Y.Y. Effects of Different Sleeping Postures on Intraocular Pressure and Ocular Perfusion Pressure in Healthy Young Subjects. Ophthalmology. 2013;120:1565–1570. doi: 10.1016/j.ophtha.2013.01.011. [DOI] [PubMed] [Google Scholar]
- 17.Barkana Y. Postural Change in Intraocular Pressure: A Comparison of Measurement with a Goldmann Tonometer, Tonopen XL, Pneumatonometer, and HA-2. J. Glaucoma. 2014;23:e23–e28. doi: 10.1097/IJG.0b013e3182a0762f. [DOI] [PubMed] [Google Scholar]
- 18.Huang R., Ge J., Chen G., Gao X., Laties A.M., Zhang X. Four Measures of Intraocular Pressure Fluctuation: Which Correlates Most Optimally with Actual Office-hour Readings? J. Glaucoma. 2015;24:550–555. doi: 10.1097/IJG.0000000000000059. [DOI] [PubMed] [Google Scholar]
- 19.Lee J.Y., Yoo C., Jung J.H., Hwang Y.H., Kim Y.Y. The effect of lateral decubitus position on intraocular pressure in healthy young subjects. Acta. Ophthalmol. 2012;90:e68–e72. doi: 10.1111/j.1755-3768.2011.02208.x. [DOI] [PubMed] [Google Scholar]
- 20.Fogagnolo P., Orzalesi N., Ferreras A., Rossetti L. The circadian curve of intraocular pressure: Can we estimate its characteristics during office hours? Investig. Ophthalmol. Vis. Sci. 2009;50:2209–2215. doi: 10.1167/iovs.08-2889. [DOI] [PubMed] [Google Scholar]
- 21.Moster S.J., Fakhraie G., Venketesh R., Moster M.L., Zhao Y., Moster M.R. Relationship of central corneal thickness to postural IOP changes in patients with and without glaucoma in southern India. Int. Ophthalmol. 2012;32:307–311. doi: 10.1007/s10792-012-9522-8. [DOI] [PubMed] [Google Scholar]
- 22.Barkana Y., Gutfreund S. Measurement of the difference in intraocular pressure between the sitting and lying body positions in healthy subjects: Direct comparison of the Icare Pro with the Goldmann applanation tonometer, Pneumatonometer and Tonopen XL. Clin. Exp. Ophthalmol. 2014;42:608–614. doi: 10.1111/ceo.12272. [DOI] [PubMed] [Google Scholar]
- 23.Nakakura S., Mori E., Yamamoto M., Tsushima Y., Tabuchi H., Kiuchi Y. Intradevice and Interdevice Agreement Between a Rebound Tonometer, Icare PRO, and the Tonopen XL and Kowa Hand-held Applanation Tonometer When Used in the Sitting and Supine Position. J. Glaucoma. 2015;24:515–521. doi: 10.1097/IJG.0000000000000016. [DOI] [PubMed] [Google Scholar]
- 24.Lee T.E., Yoo C., Hwang J.Y., Lin S., Kim Y.Y. Comparison of Intraocular Pressure Measurements between Icare Pro Rebound Tonometer and Tono-Pen XL Tonometer in Supine and Lateral Decubitus Body Positions. Curr. Eye. Res. 2015;40:923–929. doi: 10.3109/02713683.2014.964416. [DOI] [PubMed] [Google Scholar]
- 25.Smith T.J., Lewis J. Effect of inverted body position intraocular pressure. Am. J. Ophthalmol. 1985;99:617–618. doi: 10.1016/S0002-9394(14)77989-9. [DOI] [PubMed] [Google Scholar]
- 26.Malihi M., Sit A.J. Effect of head and body position on intraocular pressure. Ophthalmology. 2012;119:987–991. doi: 10.1016/j.ophtha.2011.11.024. [DOI] [PubMed] [Google Scholar]
- 27.Sultan M., Blondeau P. Episcleral venous pressure in younger and older subjects in the sitting and supine positions. J. Glaucoma. 2003;12:370–373. doi: 10.1097/00061198-200308000-00013. [DOI] [PubMed] [Google Scholar]
- 28.Blondeau P., Tetrault J.P., Papamarkakis C. Diurnal variation of episcleral venous pressure in healthy patients: A pilot study. J. Glaucoma. 2001;10:18–24. doi: 10.1097/00061198-200102000-00005. [DOI] [PubMed] [Google Scholar]
- 29.Friberg T.R., Sanborn G., Weinreb R.N. Intraocular and episcleral venous pressure increase during inverted posture. Am. J. Ophthalmol. 1987;103:523–526. doi: 10.1016/S0002-9394(14)74275-8. [DOI] [PubMed] [Google Scholar]
- 30.Cowley A.W., Franchini K.G. Neurogenic control of blood vessels. In: Robertson D., Low P.A., Polinsky R.J., editors. Primer on the Autonomic Nervous System. Academic Press; New York, NY, USA: 1996. pp. 49–55. [Google Scholar]
- 31.Hayreh S.S. Factors influencing blood flow in the optic nerve head. J. Glaucoma. 1997;6:412–425. doi: 10.1097/00061198-199712000-00012. [DOI] [PubMed] [Google Scholar]
- 32.De Bernardo M., Salerno G., Gioia M., Capasso L., Russillo M.C., Picillo M., Erro R., Amboni M., Barone P., Rosa N., et al. Intraocular pressure and choroidal thickness postural changes in multiple system atrophy and Parkinson’s disease. Sci. Rep. 2021;11:8936. doi: 10.1038/s41598-021-88250-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.De Bernardo M., Cornetta P., Marotta G., Rosa N. Corneal biomechanical parameters in pellucid marginal degeneration. J. Curr. Ophthalmol. 2018;30:280. doi: 10.1016/j.joco.2018.06.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Rosa N., Lanza M., De Bernardo M., Signoriello G., Chiodini P. Relationship Between Corneal Hysteresis and Corneal Resistance Factor with Other Ocular Parameters. Semin. Ophthalmol. 2015;30:335–339. doi: 10.3109/08820538.2013.874479. [DOI] [PubMed] [Google Scholar]
- 35.De Bernardo M., Rosa N. Intraocular pressure after LASEK. Graefes. Arch. Clin. Exp. Ophthalmol. 2018;256:2009–2010. doi: 10.1007/s00417-018-4047-0. [DOI] [PubMed] [Google Scholar]
- 36.De Bernardo M., Vitiello L., Abbinante G., De Pascale I., Capasso L., Marotta G., Rosa N. Comparison between two devices in the detection of corneal thickness changes after cataract surgery. Sci. Rep. 2021;11:6709. doi: 10.1038/s41598-021-86158-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Chatterjee A., Shah S., Bessant D.A., Naroo S.A., Doyle S.J. Reduction in intraocular pressure after excimer laser photorefractive keratectomy. Correlation with pretreatment myopia. Ophthalmology. 1997;104:355–359. doi: 10.1016/S0161-6420(97)30308-X. [DOI] [PubMed] [Google Scholar]
- 38.De Bernardo M., Capasso L., Caliendo L., Vosa Y., Rosa N. Intraocular Pressure Evaluation after Myopic Refractive Surgery: A Comparison of Methods in 121 Eyes. Semin. Ophthalmol. 2016;31:233–242. doi: 10.3109/08820538.2014.962156. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data presented in this study are available on request from the corresponding author.