Bioprosthetic valve dysfunction and failure after TAVI in bicuspid aortic valve stenosis during one-year follow-up according to VARC-3

Verena Veulemans 1,3,, Philippe Nuyens 2, Shouheng Goh 1, Oliver Maier 1, Stephan Binnebößel 1, Jacqueline Heermann 1, Christian Jung 1, Ralf Westenfeld 1, Malte Kelm 1,3, Ole de Backer 2,#, Tobias Zeus 1,#
PMCID: PMC9681687  PMID: 35767098

Abstract

Background

Transcatheter aortic valve implantation (TAVI) in bicuspid aortic valve (BAV) stenosis has become more frequent in the last years. This may pose challenges for long-time valve durability. Therefore, we aimed to evaluate the prevalence of bioprosthetic valve dysfunction (BVD) with the newest-generation devices in BAV stenosis up to one-year follow-up (FU).

Methods

The primary endpoint was defined as the prevalence of BVD during the first procedural year according to Valve Academic Research Consortium (VARC)-3 criteria. Secondary endpoints were defined as failure in device success and clinical endpoints according to VARC-3.

Results

A total of 107 patients were included. Of these, 34 subjects (31.8%) met the criteria for BVD during a mean FU of 263 ± 180 days, of which 20.2% were already documented after thirty days. Device success after one year was lower in the + BVD cohort (57.6% vs. 98.7%, p < 0.0001*). The rates of structural valve deterioration were 6.5%, non-structural valve deterioration (NSVD) 17.8%, subclinical leaflet thickening 10.3%, and endocarditis 0.9%. NSVD was foremost triggered by patient prosthesis mismatch in balloon-expandable valves. Hemodynamic valve deterioration stage 1 and 2 was confirmed in 16.8% of + BVD patients, while stage 1 and 3 bioprosthetic valve failure occurred in 1.9%. There was no impact of BVD on mortality.

Conclusion

There is critical evidence of early BVD after TAVI in BAV during one-year FU in one-third of patients, also lowering device success. The most frequently observed bioprosthetic valve dysfunction was NSVD due to patient prosthesis mismatch following TAVI with a balloon-expandable valve.

Supplementary Information

The online version contains supplementary material available at 10.1007/s00392-022-02052-9.

Keywords: TAVI, Bioprosthetic valve dysfunction, Bicuspid valves, Elderly, Complication, Device success

Introduction

Transcatheter aortic valve implantation (TAVI) as treatment of aortic stenosis in bicuspid aortic valve (BAV) has become more frequent in the last years. The specific anatomic configurations of BAV pose challenges toward valve performance, outcome, and long-time durability, especially when treating patients with longer life-expectancy [1]. Recent data have shown that newer-generation devices offer similar device success in selected bicuspid compared to tricuspid aortic valves [2, 3], although little is known about prevalence and possible predictors of bioprosthetic valve dysfunction (BVD) during follow-up. Recently, the Valve Academic Research Consortium 3 endpoint definitions for aortic valve clinical research [4] were released with critical updates concerning an adequate interpretation of BVD. Structured data on the prevalence and an in-depth analysis of clinical, anatomical, and technical risk factors for BVD are lacking. Therefore, we aimed to evaluate the prevalence of BVD during one-year follow-up in patients who underwent TAVI in BAV stenosis.

Methods

Study population

Consecutive patients with symptomatic bicuspid severe aortic stenosis (AS) who underwent either transfemoral (TF), transapical (TA) or alternative access TAVI with newer-generation devices (SAPIEN 3, Edwards Lifesciences; Evolut R/Pro platform, Medtronic) were included in this retrospective analysis. Valve-in-valve procedures, pure aortic regurgitation, TAVI with other bioprosthetic valves, and procedures without appropriate MSCT data were excluded. All procedures were performed according to current guideline recommendations and under local or general anesthesia.

All patients provided written informed consent for TAVI and for the use of their clinical, procedural, and follow-up data in research. The study procedures were conducted in accordance with the Declaration of Helsinki, and the institutional Ethics Committee of Heinrich-Heine University approved the study protocol (4080).

Study endpoints

The primary endpoint was defined as the prevalence of BVD during one-year follow-up (FU) according to Valve Academic Research Consortium (VARC)-3 criteria [4]. Secondary endpoints were defined as failure in device success and clinical endpoints according to VARC-3.

Procedural details and 3D image analysis of MSCT

MSCT images were transferred to a dedicated workstation for 3-dimensional (3D) volume-rendered reconstruction (3 mensio Structural Heart™, Pie Medical Imaging BV, Maastricht, The Netherlands) after the procedure according to TAVI-related standardized recommendations for CT image acquisition [5]. Dimensions were determined with the use of workstation tools. Tubular (tube), flared or tapered configurations of the aortic root were identified from the relation of the annulus to the inter-commissural distance (ICD) according to the BAVARD-registry [6]. The total aortic valve calcification (AVC) and calcium amount of the upper LVOT expressed as recalculated Agatston Units (AU) adapted from the calcium volume was calculated. All MSCT-reconstructions, echocardiographic interpretations, and analyses were done by experienced level 3 readers. In a subset of patients (23%), post-procedural MSCT was performed. The implantation strategy was retrospectively analyzed according to the CASPAR algorithm [7] as described before.

VARC-3 definitions of bioprosthetic valve dysfunction (BVD) (4)

For the details, please see Supplementary material – Methods.

Statistical analysis

The collected data included patient characteristics, imaging findings, periprocedural in-hospital data, laboratory results and follow-up data. Continuous data are described by the mean and standard deviation, and categorical variables by frequencies and percentages. Continuous variables were compared using Student’s t-test, categorical variables were compared using Fisher’s exact test. The data analysis was performed using the statistical software SPSS (version 27.0.1, SPSS Inc., Chicago, IL, USA), GraphPad Prism (version 6.0, GraphPad Software, San Diego, CA, USA), and Wizard 2- Statistics & Analysis (Evan Miller). All statistical tests were 2-tailed, and a value of p < 0.05 was considered statistically significant.

Results

From 2016 to December 2020, in total 109 patients were identified at the Heart Centre Düsseldorf in Germany or at Rigshospitalet in Copenhagen, Denmark. Since one patient had died (early mortality) and one patient was immediately converted to surgery, we finally included 107 patients in the analyses. Follow-up (FU) was completed 3 months to 1-year FU in this. The study cohort was further separated into patients without and with detection of BVD during a mean FU of 263 ± 180 days (-BVD: n = 73; 68.2%; + BVD: n = 34; 31.8%).

Baseline characteristics

Baseline characteristics did differ according to the particular risk profile of patients without and with BVD. For example, patients with BVD had a lower left ventricular function (LVEF; -BVD vs. + BVD: 54.0 ± 10.9 vs. 48.4 ± 13.5; p = 0.023*) and smaller aortic valve area (AVA; -BVD vs. + BVD: 0.75 ± 0.2 vs. 0.64 ± 0.2; p = 0.007*). All other dimensions of the aortic root as well as calcification distribution were similar (Supplementary material – Table S1).

General procedural characteristics

Procedural details and clinical outcomes are displayed in Table 1. Implantation based on annular sizing was realized in more than two-thirds of cases without differences between the cohorts. Implantation strategy according to the CASPAR algorithm was lower in the + BVD cohort (48.9%) but statistical similar compared to the non-BVD cohort (56.9%). Post-dilatation was more frequently observed in BVD patients (-BVD vs. + BVD: 17.3% vs. 38.2%; p = 0.018*).

Table 1.

Procedural Characteristics

Procedural data Overall
(n = 109; 100%)
No BVD
(n = 75; 68.8%)
BVD
(n = 34; 31.2%)
p value
TF access 106 (97.3) 73 (97.3) 33 (97.1) 0.935
TA access 2 (1.8) 2 (2.7) 0 (0.0) 0.337
Other access 1 (0.9) 0 (0.0) 1 (2.9) 0.136
EDWARDS Sapien 3™ 35 (32.1) 23 (30.7) 12 (35.3) 0.632
 23 mm 2 (5.71) 2 (8.7) 0 (0.0) 0.293
 26 mm 10 (28.6) 6 (26.2) 4 (33.3) 0.652
 29 mm 23 (65.7) 15 (65.2) 8 (66.7) 0.932
Evolut R™ 54 (49.5) 40 (53.3) 14 (41.2) 0.240
 23 mm 1 (1.9) 1 (2.5) 0 (0.0) 0.550
 26 mm 13 (24.1) 11 (27.5) 2 (14.3) 0.320
 29 mm 16 (29.6) 12 (30.0) 4 (28.6) 0.920
 34 mm 24 (44.4) 16 (40.0) 8 (57.1) 0.267
Evolut Pro™ 20 (18.4) 12 (16.0) 8 (23.5) 0.347
 26 mm 5 (25.0) 4 (33.3) 1 (12.5) 0.292
 29 mm 15 (75.0) 8 (66.7) 7 (87.5) 0.292
Implantation based on annular sizing 92 (84.4) 62 (82.7) 30 (88.2) 0.458
Implantation according to CASPAR 57 (54.3) 41 (56.9) 16 (48.9) 0.419
Contrast, ml 104.2 ± 35.2 100.3 ± 35.9 112.7 ± 39.0 0.089
Fluoroscopy time, min 23.0 ± 8.3 22.2 ± 8.1 24.8 ± 8.5 0.139
Dose Area Product, Gy x cm2

1.960

[950–3.772]

2.025

[1.070–3.704]

1.516

[918–4.180]

0.678
Pre-dilatation 88 (80.7) 59 (78.7) 29 (85.3) 0.416
Post-dilatation 26 (23.9) 13 (17.3) 13 (38.2) 0.018*
AR ≥ II° 4 (3.8) 1 (1.3) 3 (8.8) 0.054
Intraprocedural complications
 Immediate stroke 1 (0.9) 0 (0) 1 (2.9) 0.136
 Aortic dissection 0 (0) 0 (0) 0 (0) 1.000
 Annulus rupture 0 (0) 0 (0) 0 (0) 1.000
 Coronary obstruction 2 (1.83) 0 (0) 2 (5.88) 0.034*
 Vascular complications 10 (9.2) 7 (9.3) 3 (8.8) 0.932
 Valve dislocation 0 (0) 0 (0) 0 (0) 1.000
 Conversion to surgery 1 (0.9) 1 (1.3) 0 (0) 0.499
 Need of 2nd valve 0 (0) 0 (0) 0 (0) 1.000
Tamponade 1 (0.9) 1 (1.3) 0 (0) 0.499
 CPR 1 (0.9) 0 (0) 1 (2.9) 0.136
 Immediate Procedural death 0 (0.0) 0 (0.0) 0 (0.0) 1.000

Values are mean ± SD, median ± interquartile range or n (%)

AR aortic regurgitation, CPR cardiopulmonary resuscitation, TA transapical, TF transfemoral

*p value < 0.05

Thirty-day outcome and functional status

Table 2 summarizes the main clinical and echocardiographic outcomes observed at 30 days according to the recently released VARC-3 criteria. In general, the clinical outcomes were similar between both groups except from a higher number of cerebrovascular events in + BVD patients (CVE; -BVD vs. + BVD: 0.0% vs. 5.9%; p = 0.034*) and to BVD linked adverse events that are listed in Table 2 and of which two-third already were documented during 30-days FU. The primary device success was statistically comparable in both groups but tended to be lower in the + BVD collective (85.3% vs. 96.0%).

Table 2.

30-day outcome according to VARC-3

Post-procedural outcome Overall
(n = 109; 100%)
No BVD
(n = 75; 68.8%)
BVD
(n = 34; 31.2%)
p value
30-day mortality 1 (0.9) 1 (1.3) 0 (0) 0.499
Peri- and post-procedural MI 0 (0) 0 (0) 0 (0) 1.000
CVE (NeuroARC 1) 2 (1.83) 0 (0) 2 (5.9) 0.034*
 Hemorrhagic stroke 2 (1.83) 0 (0) 2 (5.9) 0.034*
 Severity (mRS < 2) 1 (0.9) 0 (0) 1 (2.9) 0.136
 Severity (mRS > 2) 1 (0.9) 0 (0) 1 (2.9) 0.136
 TIA 0 (0) 0 (0) 0 (0) 1.000
Bleeding 16 (14.7) 13 (17.3) 3 (8.8) 0.245
 Type I 7 (6.4) 6 (8.0) 1 (2.9) 0.318
 Type 2 6 (5.5) 5 (6.7) 1 (2.9) 0.429
 Type 3 3 (2.8) 2 (2.7) 1 (2.9) 0.935
 Type 4 0 (0) 0 (0) 0 (0) 1.000
VASC 21 (19.3) 16 (21.3) 5 (14.7) 0.416
 Major VASC 6 (5.5) 5 (6.7) 1 (2.9) 0.429
 Minor VASC 14 (12.8) 10 (13.3) 4 (11.8) 0.821
 Closure Device Failure 6 (5.5) 3 (4.0) 3 (8.8) 0.306
 Access-related non-VASC 0 (0) 0 (0) 0 (0) 1.000
AKI I-IV 5 (4.6) 5 (6.7) 0 (0) 0.123
 AKI I 2 (1.8) 2 (2.7) 0 (0) 0.337
 AKI II 1 (0.9) 1 (1.3) 0 (0) 0.499
 AKI III 2 (1.8) 2 (2.7) 0 (0) 0.337
 AKI IV (new RRT) 0 (0) 0 (0) 0 (0) 1.000
New AVB (I-III°) 20 (18.4) 16 (21.3) 4 (11.8) 0.232
New LBBB/RBBB 22 (20.2) 12 (16.0) 10 (29.4) 0.213
New PPI 17 (15.6) 13 (17.3) 4 (11.8) 0.458
TAVI-related Reintervention 4 (3.7) 3 (4.0) 1 (2.9) 0.785
Cardiac Structutal Complications 4 (3.7) 2 (2.7) 2 (5.9) 0.263
 Minor 3 (2.8) 1 (1.3) 2 (5.9) 0.179
 Major 2 (1.8) 1 (1.3) 1 (2.9) 0.337
Functional data
 Hint on BVD 22 (20.2) 0 (0) 22 (64.7)  < 0.001*
 SVD 5 (4.6) 0 (0) 5 (14.7)  < 0.001*
 NSVD 11 (10.1) 0 (0) 11 (32.4)  < 0.001*
 Endocarditis TAVI 0 (0) 0 (0) 0 (0) 1.000
 Leafletthrombosis/HALT 7 (6.4) 0 (0) 7 (20.6)  < 0.001*
 Vmax (m/s) 2.1 ± 0.5 2.0 ± 0.4 2.2 ± 0.5 0.037*
 dPmean (mmHg) 10.0 ± 4.4 9.3 ± 3.9 11.5 ± 5.2 0.013*
 EOAi (cm2/m2) 1.0 ± 0.2 1.1 ± 0.2 0.9 ± 9.2 0.001*
 Moderate-to-severe PPM 8 (7.4) 0 (0) 8 (23.5)  < 0.001*
 EOAi < 0.85 cm2/m2 (BMI<30) 8 (11.0) 0 (0) 8 (23.5)  < 0.001*
 EOAi < 0.70 cm2/m2 (BMI≥30) 0 (0) 0 (0) 0 (0) 1.000
 PVL ≥ II° 3 (2.8) 0 (0) 3 (8.8) 0.009*
 Device success 101 (92.7) 72 (96.0) 29 (85.3) 0.105
Medication at discharge/30-days
 APT mono 14 (13.1) 10 (13.7) 4 (11.8) 0.782
 DPT 43 (40.2) 46 (63.0) 18 (52.9) 0.322
 OAC mono 19 (17.8) 11 (15.1) 8 (23.5) 0.286
 OAC + APT 9 (8.4) 5 (6.9) 4 (11.8) 0.394

Values are mean ± SD, median ± interquartile range or n (%)

AKI acute kidney injury, APT antiplatelet therapy, AVB atrioventricular block, BMI body mass index, CVE cerebrovascular events, DPT dual antiplatelet therapy, EOAi effective orifice area (index), HVD Hemodynamic valve deterioration, LBBB left-bundle branch block, NSVD Non-structural valve deterioration, OAC oral anticoagulation, PPI permanent pacemaker therapy, PVL paravalvular leakage, RBBB right-bundle branch block, RRT renal replacement therapy,

*p value < 0.05

One-year outcome and functional status

After one year, two deaths were documented in the -BVD cohort of unknown causes (i.e. cardiovascular death) and one death in the + BVD cohort due to endocarditis/BVF stage 3 (-BVD vs. + BVD: 2.7% vs. 2.9%; p = 0.814).

SVD was documented in 7 patients (6.5%), NSVD in 19 patients (17.8%), HALT/RELM in 11 patients (10.3%), and endocarditis in one patient (0.9%). In-depth differentiation of BVD is illustrated in the Central Illustration. Moderate PVL was the reason in 6.5% of NSVD, while PPM was documented in 12.2% of NSVD cases. HVD was confirmed in 16.8% of + BVD patients, while BVF occurred in 2 patients (stage 1 and 3; 1.9%; Fig. 1).

Fig. 1.

Fig. 1

Central Illustration: Prevalence of bioprosthetic Valve Dysfunction (BVD) during 30-day and One-year FU. Prevalence of Hemodynamic Valve Deterioration (HVD) and Bioprosthetic Valve Failure (BVF) during 30-day and One-year FU

When focusing on a potential impact of the kind of transcatheter heart valve (THV) on BVD event rate, no difference in overall BVD between self-expandable (SEV) and balloon-expandable (BEV) valves, but a higher incidence of overall NSVD (12.2% vs. 30.2%; p = 0.03), foremost triggered by PPM (5.4% vs. 27.3%; p = 0.003) in BEV, was documented (Fig. 2A).

Fig. 2.

Fig. 2

Event rate according to the type of Transcatheter Heart Valve (THV) and Device success. A Frequency distribution of BVD in self- (SEV) and balloon-expandable valves (BEV). B Device success during time course in -BVD- and + BVD patients

Compared to the 30-day FU, device success after one year was significant lower in the + BVD collective (57.6% vs. 98.7%, p < 0.0001*; Fig. 2B), comparably triggered by SEV and BEV. Concerning functional parameters, the mean valvular gradients were higher by average in the + BVD cohort as compared to -BVD patients (dPmean; -BVD vs. + BVD: 7.9 ± 3.7 vs. 11.6 ± 6.7; p = 0.037*, data not shown).

Discussion

As staging and interpretation of BVD were handled heterogenous in the last years, the recently released VARC-3 definitions now enable a streamlined interpretation algorithm for more homogenous results. To our knowledge, this is the first study addressing VARC-3 definitions for BVD in bicuspid TAVI. The main read-outs of our dual-center retrospective study revealed that:

  1. There is evidence of frequently observed BVD after TAVI in bicuspid aortic valves during 30-day (20.2%) and 1-year FU (31.8%), also lowering device success.

  2. The most frequently observed component of BVD was NSVD (17.8%), foremost triggered by patient prosthesis mismatch (PPM) in balloon-expandable valves (BEV).

  3. There was no impact of BVD on thirty-day or one-year mortality.

Bioprosthetic dysfunction is known to consist of both structural and non-structural alterations that have now been systemized according to etiology and severity, in line with current literature [8, 9]. Thus, four columns of BVD have been classified: SVD, NSVD, valve thrombosis with subclinically or clinically relevant HALT/RELM, and endocarditis. The new VARC-3 criteria now streamlined the current and slightly differing definitions of SVD. They allow for differentiation of the SVD-related new term of hemodynamic valve deterioration (HVD) from PPM-related residually high transprosthetic gradients in the meaning of NSVD to avoid an overestimation of SVD definition based on elevated gradients. Furthermore, in-depth subclassifications of SVD have also been provided, including morphological changes (stage 1) despite elevated gradients. In terms of PPM, the EOAi was thoroughly distinguished, differentiating between an EOAi < 0.85 cm2/cm2 and EOAi < 0.70 cm2/cm2 in obese patients to avoid an overestimation of NSVD. In tricuspid valves, the differing interpretation of BVD became already evident in the last years using different criteria for both surgical and transcatheter devices. Par example, the definition of SVD in the VIVID consensus document [9] introduced the concept of progressive SVD stages into the main criteria of the EAPCI/ESC/EACTS definitions [8], also including specific management recommendations according to the degree and kind of dysfunction. Thus, different stages of bioprosthetic valve failure (BVF) were incorporated into VARC-3, allowing to assess clinically relevant consequences of BVD and including the need for valve-related death and re-intervention. These changes might help to better differentiate the underlying conditions and resulting consequences in the management of BVD, also giving in-detail recommendations for an appropriate imaging following bioprosthetic valve implantation.

Considering all possible alterations, BVD showed a high prevalence (31.8%) in bicuspid valves during one-year FU in this study, of which two-thirds of all events were already documented during 30-day FU (20.2%). However, BVD in this study was primarily due to NSVD, foremost triggered by PPM in balloon-expandable valves that are known to lead to PPM in more than fifty percent of tricuspid TAVI [10]. According to VARC-3, PPM was thoroughly adapted to the recommended EOAi differentiating between obese and non-obese patients. In general, TAVI is associated with a lower prevalence of PPM than surgical aortic valve replacement, showing the lowest prevalence using self-expanding valves with a supra-annular design [11]. Regarding the association to longer-term outcomes, a valuable hemodynamic performance across bicuspid patients undergoing TAVI may be crucial for long-term durability [12].

In contrast, SVD was recognized with 6.5% within our study, nearly 5% already during 30-day FU. Until now, there were no data on early SVD stages in bicuspid TAVI. According to current literature, the medium long-term risk of severe SVD seems to be low (up to 2.5%) for 4–8 years FU [13] in tricuspid valves. Thus, the amount of short- and mid-term SVD in this study seems to be comparatively high. Recently, Forrest et al. [14] showed favorable 30-day results in bicuspid low-risk patients undergoing TAVI with self-expandable valves, counting for significantly elevated gradients in only 1.4% of cases. However, the analysis was almost based on VARC-2 criteria and realized without in-detail BVD analysis. The overall device success with 95% was comparable to this study with 93% after thirty days, which sounds valuable for both studies at first glance. Nevertheless, when differentiating into BVD and non-BVD cases, device success after thirty days and one year was dramatically lower in the + BVD cohort (85% vs. 57% respectively). To reiterate: More than one-third of patients showing any BVD event, formally performed with device failure during the one-year course. A phenomenon that must be considered when widely expanding TAVI in bicuspid low-risk patients.

Subclinical HALT with reduced leaflet motion (RELM, i.e. leaflet restriction) was diagnosed in 10.3% of our cases by imaging findings (MSCT and TEE), approximately 7% already documented during 30-day FU and the majority in self-expandable valves. At this point, it must be pronounced that the evaluation of HALT should be performed by cardiac CT with high spatial resolution and multiplanar reformats. Post-procedural MSCT was only performed in approximately 23% of cases in this study but predominantly in self-expandable devices (70% vs. 30% of all performed post-treatment CTs), supposing a possible under-estimation and selection bias as well, as TAVI in BAV was predominantly performed with self-expandable valves (~ 68% of cases). However, leaflet thrombosis can alternatively be diagnosed by TEE to evaluate leaflet thickness and motion [4], which was the case in 4% of patients due to already visible alterations and abnormal flow conditions recorded by transthoracic echocardiography. Furthermore, the frequent but commonly subclinical phenomenon of HALT, also synonymous described as subclinical leaflet thrombosis, must be distinguished from clinically relevant valve thrombosis. However, the prevalence of HALT in this study is in line with the current reported evidence in bicuspid valves [15]. In general, HALT has been described in 5–40% of all patients who undergo post-procedural MSCT [16] and is known to be probably associated with composites of strokes or TIA when becoming clinically relevant. As a result, one case of our study was probably associated with the history of stroke but some time delay between the events, so that it was not classified as clinically relevant leaflet thrombosis (also taken a RELM < 25% as diagnosed by MSCT into consideration). To conclude, there might be a need for a structured post TAVI imaging attempt, especially after bicuspid procedures. Mal-apposition, platelet activation, and high annular calcium amounts in BAV might impact sinus blood flow conditions and, therefore, the priming of HALT. Endocarditis is another reversible mechanism of non-SVD in the development of valve failure, described with an average of 5% in nearly all studies and a climax in the first two years [17], unfortunately still combined with high mortality. Endocarditis in this study was a rare event with only 0.9% reflecting one case that likewise led to one-year death but might be underestimated compared to current literature.

Hemodynamic valve deterioration and bioprosthetic valve failure

HVD may be caused by SVD, valve thrombosis, or endocarditis. Therefore, an in-depth assessment of valve leaflet morphology and structure is crucial to differentiate between the given entity. We could identify early HVD in 16.8% of patients, distributed on 11.2% stage 1 and 5.6% stage 2 HVD according to the currently recommended classification [4]. Again, approximately 11% of cases were already documented during 30-day FU, which must be considered as high compared to current knowledge but seem to be an effect of the new streamlined algorithm that also includes morphological changes (HVD stage 1). However, nearly 4% of stage 2 HVD (i.e., non-PPM caused relevant gradient elevation) during the 30-day course is an issue to address and to further investigate in TAVI-treated bicuspid patients. As bioprosthetic valve failure (BVF) should be reported separately from subclinical BVD to adequately address the clinical relevance, we documented two events of BVF (1.9%), one stage 1 BVF with worsening of symptoms due to SVD-triggered HVD stage 2 and one stage 3 BVF (valve-related death) due to endocarditis of the bio-prosthesis. Compared to the current literature and the remarks above, the amount of BVF seems to be again higher in short-to-mid-term FU. However, the new classification according to VARC-3 was never addressed so far.

Conclusion

There is critical evidence of early BVD after TAVI in BAV during one-year FU in one-third of patients, also lowering device success. The most frequently observed bioprosthetic valve dysfunction was NSVD due to patient prosthesis mismatch following TAVI with a balloon-expandable valve. Identifying predictors of BVD may help to optimize long-term valve performance in the future.

Limitations

This study is a dual-center, retrospective analysis with associated unavoidable limitations due to its design. The frequencies of HALT/RELM must be handled with caution as cardiac CT wasn’t performed regularly. However, taking the low percentage of bicuspid valves in western countries into account, we provided first insights into the prevalence of BVD according to the new VARC-3 criteria in a bicuspid cohort covering more than a hundred patients.

Supplementary Information

Below is the link to the electronic supplementary material.

Abbreviations

BVD

Bioprosthetic dysfunction

BVF

Bioprosthetic valve failure

HALT

Hypo-attenuating leaflet thickening

HVD

Hemodynamic valve deterioration

LVOT

Left ventricular outflow tract

NSVD

Non-structural valve dysfunction

MSCT

Multislice computer tomography

PPI

Permanent pacemaker implantation

PPM

Permanent pacemaker implantation

PPI

Patient prosthesis mismatch

SVD

Structural valve deterioration

Funding

Open Access funding enabled and organized by Projekt DEAL. None.

Declarations

Conflict of interest

VV, TZ, CJ, and RW have received consulting fees, travel expenses, or study honoraria from Medtronic, Edwards Lifesciences, and/or Boston Scientific. ODB has received institutional research grants and consulting fees from Abbott, Medtronic, and/or Boston Scientific. All other authors have nothing to disclose regarding this project.

Footnotes

Ole de Backer and Tobias Zeus have contributed equally to the manuscript.

References

  • 1.Yoon SH, Kim WK, Dhoble A, et al. Bicuspid aortic valve morphology and outcomes after transcatheter aortic valve replacement. J Am Coll Cardiol. 2020;76:1018–1030. doi: 10.1016/j.jacc.2020.07.005. [DOI] [PubMed] [Google Scholar]
  • 2.Makkar RR, Yoon SH, Leon MB, Chakravarty T, Rinaldi M, Shah PB, Skipper ER, Thourani VH, Babaliaros V, Cheng W, Trento A, Vemulapalli S, Kapadia SR, Kodali S, Mack MJ, Tang GHL, Kaneko T. Association between transcatheter aortic valve replacement for bicuspid vs tricuspid aortic stenosis and mortality or stroke. JAMA. 2019;321:2193–2202. doi: 10.1001/jama.2019.7108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Makkar RR, Yoon SH, Chakravarty T, et al. Association between transcatheter aortic valve replacement for bicuspid vs tricuspid aortic stenosis and mortality or stroke among patients at low surgical risk. JAMA. 2021;326:1034–1044. doi: 10.1001/jama.2021.13346. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.VARC-3 WRITING COMMITTEE. Généreux P, Piazza N, et al. Valve academic research consortium 3: updated endpoint definitions for aortic valve clinical research. J Am Coll Cardiol. 2021;77:2717–2746. doi: 10.1016/j.jacc.2021.02.038. [DOI] [PubMed] [Google Scholar]
  • 5.Achenbach S, Delgado V, Hausleiter J, Schoenhagen P, Min JK, Leipsic JA. SCCT expert consensus document on computed tomography imaging before transcatheter aortic valve implantation (TAVI)/transcatheter aortic valve replacement (TAVR) J Cardiovasc Comput Tomogr. 2012;6:366–380. doi: 10.1016/j.jcct.2012.11.002. [DOI] [PubMed] [Google Scholar]
  • 6.Tchetche D, de Biase C, van Gils L, et al. Bicuspid aortic valve anatomy and relationship with devices: the BAVARD multicenter registry. Circ Cardiovasc Interv. 2019;12:e007107. doi: 10.1161/CIRCINTERVENTIONS.118.007107. [DOI] [PubMed] [Google Scholar]
  • 7.Petronio AS, Angelillis M, De Backer O, et al. Bicuspid aortic valve sizing for transcatheter aortic valve implantation: development and validation of an algorithm based on multi-slice computed tomography. J Cardiovasc Comput Tomogr. 2020;14:452–461. doi: 10.1016/j.jcct.2020.01.007. [DOI] [PubMed] [Google Scholar]
  • 8.Capodanno D, Petronio AS, Prendergast B, et al. Standardized definitions of structural deterioration and valve failure in assessing long-term durability of transcatheter and surgical aortic bioprosthetic valves: a consensus statement from the European association of percutaneous cardiovascular interventions (EAPCI) endorsed by the European society of cardiology (ESC) and the European association for cardio-thoracic surgery (EACTS) Eur Heart J. 2017;38:3382–3390. doi: 10.1093/eurheartj/ehx303. [DOI] [PubMed] [Google Scholar]
  • 9.Dvir D, Bourguignon T, Otto CM, et al. Standardized definition of structural valve degeneration for surgical and transcatheter bioprosthetic aortic valves. Circulation. 2018;137:388–399. doi: 10.1161/CIRCULATIONAHA.117.030729. [DOI] [PubMed] [Google Scholar]
  • 10.Mack MJ, Leon MB, Thourani VH, et al. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med. 2019;380:1695–1705. doi: 10.1056/NEJMoa1814052. [DOI] [PubMed] [Google Scholar]
  • 11.Mangieri A, Chieffo A, Kim WK, et al. Transcatheter aortic valve implantation using the ACURATE neo in bicuspid and tricuspid aortic valve stenosis: a propensity-matched analysis of a European experience. EuroIntervention. 2018;14:e1269–e1275. doi: 10.4244/EIJ-D-18-00281. [DOI] [PubMed] [Google Scholar]
  • 12.Herrmann HC, Daneshvar SA, Fonarow GC, et al. Prosthesis-patient mismatch in patients undergoing transcatheter aortic valve replacement: from the STS/ACC TVT registry. J Am Coll Cardiol. 2018;72:2701–2711. doi: 10.1016/j.jacc.2018.09.001. [DOI] [PubMed] [Google Scholar]
  • 13.Barbanti M, Costa G, Zappulla P, et al. Incidence of long-term structural valve dysfunction and bioprosthetic valve failure after transcatheter aortic valve replacement. J Am Heart Assoc. 2018;7:e008440. doi: 10.1161/JAHA.117.008440. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Forrest JK, Ramlawi B, Deeb GM, et al. Transcatheter aortic valve replacement in low-risk patients with bicuspid aortic valve stenosis. JAMA Cardiol. 2021;6:50–57. doi: 10.1001/jamacardio.2020.4738. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Waksman R, Craig PE, Torguson R, et al. Transcatheter aortic valve replacement in low-risk patients with symptomatic severe bicuspid aortic valve stenosis. JACC Cardiovasc Interv. 2020;13:1019–1027. doi: 10.1016/j.jcin.2020.02.008. [DOI] [PubMed] [Google Scholar]
  • 16.Chakravarty T, Søndergaard L, Friedman J, et al. Subclinical leaflet thrombosis in surgical and transcatheter bioprosthetic aortic valves: an observational study. Lancet. 2017;389:2383–2392. doi: 10.1016/S0140-6736(17)30757-2. [DOI] [PubMed] [Google Scholar]
  • 17.Prasitlumkum N, Thangjui S, Leesutipornchai T, Kewcharoen J, Limpruttidham N, Pai RG. Comparison of infective endocarditis risk between balloon and self-expandable valves following transcatheter aortic valve replacement: systematic review and meta-analysis. Cardiovasc Interv Ther. 2021;36:363–374. doi: 10.1007/s12928-020-00675-1. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials


Articles from Clinical Research in Cardiology are provided here courtesy of Springer

RESOURCES