Abstract
Stimulated by the COVID-19 outbreak, the global healthcare industry better acknowledges the necessity of innovating novel methods for remote healthcare monitoring and treating patients outside clinics. Here we report the development of two different types of graphene textile electrodes differentiated by the employed fabrication techniques (i.e., dip-coating and spray printing) and successful demonstration of ergonomic and truly wearable, single-arm diagnostic electrocardiography (SADE) using only 3 electrodes positioned on only 1 arm. The performance of the printed graphene e-textile wearable systems were benchmarked against the “gold standard” silver/silver chloride (Ag/AgCl) “wet” electrodes; achieving excellent correlation up to ∼ 96% and ∼ 98% in ECG recordings (15 s duration) acquired with graphene textiles fabricated by dip-coating and spray printing techniques, respectively. In addition, we successfully implemented automatic detection of heartrate of 8 volunteers (mean value: 74.4 bpm) during 5 min of static and dynamic daily activities and benchmarked their recordings with a standard fingertip photoplethysmography (PPG) device. Heart rate variability (HRV) was calculated, and the root means successive square difference (rMMSD) metric was 30 ms during 5 min of recording. Other cardiac parameters such as R-R interval, QRS complex duration, S-T segment duration, and T-wave duration were also detected and compared to typical chest ECG values.
Keywords: ECG, Single-arm electrocardiography, Heart-rate variability, Graphene textile, Electrocardiogram, Wearable electronics, Biopotential monitoring, Conductive nanomaterials, EHealth, E-textiles, Internet of things (IoT), Long-term monitoring, Personalized healthcare, Reduced graphene oxide (rGO), Smart garments, Vital signs
Graphical Abstract
1. Introduction
Wearable technologies (wearables) can facilitate effortless remote health monitoring in the comfort of the bedside and steer telemedicine to continuous mobile health tracking and medical assessment. Wearables in remote healthcare (mobile health, mHealth) expand rapidly in line with the expansion of “Healthcare 4.0″ and the growing adoption of the internet of medical things (IoMT). Also, endorsement of machine learning (mL) and artificial intelligence (AI) algorithms with the merger of big data analysis in pathology, and application of 5 G for securely and faster than ever medical data transmission to healthcare staff further facilitates this adoption [1], [2]. On the other hand, given both the aging of the population and the growing awareness, which sparked during the COVID-19 global pandemic, remote health trackers are not only critical in monitoring vitals of senior citizens, they also could serve as a valuable and accessible alternative to the conventional patient monitoring systems when the hospitals and medical clinics are under a heavy burden [3], [4]. This, together with an increased interest in fitness tracking, calls for affordable and wearable health monitoring devices [5].
World Health Organization reports cardiovascular diseases (CVDs) to be among the leading causes of mortality, and patients with CVD have a much higher probability of untimely death [6]. It goes without saying that wearable cardiac health tracker devices can provide an excellent convenience in the routine remote monitoring of people with cardiovascular complaints. Additionally, they can enable early diagnostics or set lifesaving alert mechanisms and inform first responders in case of cardiovascular complication events such as cardiac arrest [7].
Electrocardiography (ECG) is the conventional examination method to access vital information about the heart including rhythm abnormalities and various CVDs, through electrocardiograms (waveforms representing electrical activity of the heart) which are picked up while the heart muscle fibers contract in response to electrical depolarization. These electrical activities (cardiac biopotentials) are commonly recorded by placing electrodes on the chest and limbs [8]. Another competing technology for recording cardiac information is photoplethysmography (PPG). Although PPG devices are inherently optimal for mobile health monitoring and can be integrated even to an earbud or a ring [9], [10], the amount of extracted cardiac features from PPG is not comparable to the rich ECG data.
However, one of the inherent challenges of mobile ECG recording is optimizing the number of channels and corresponding electrode placement since conventional clinical ECG involves a complex 12-lead monitoring mechanism and hinders its mobile function. Newly introduced, smartwatch ECG approaches use a derivative of typical lead-I ECG and are adequate for the mobility specifications, but they best perform in rest conditions and lack continuous monitoring capability due to their electrode arrangement [11]. On the other hand, wrist ECG addresses ongoing monitoring issues of smartwatch ECG, but they are still not optimal for easy integration to everyday wearable clothes like smartwatch ECG devices [12]. Ear-ECG and neck ECG are other alternatives, which seems to address both continuous monitoring and easy integration to daily clothing challenges, but either they have weaker signal due to distance from the heart [13], [14], or are more affected by motion artifacts due to their arrangement [12].
Another possibility is exploiting single-arm ECG. Not long ago, a study looked into finite element method simulation of ECG over body surface by constructing the cardiac current dipole by superimposing three orthogonal current dipoles with the amplitude variations in each dimension and simulated the actual electrical behavior of the heart [15]. It illustrated by simulation that the intensity of the R-wave peak is more intense in the left arm compared to any other non-standard ECG electrode configuration on the upper torso and head surface. Following this insight, we present measurement results and a biopotential map confirming that upper left arm region below the shoulder deltoid yields the highest R-wave amplitudes ( Fig. 1). Thus, left arm ECG’s arrangement advances are two-fold; first, ECG is more intense in this spot than ear-ECG and others; second, motion artifacts less affect the signal than neck-ECG. Both advances promise better signal-to-noise ratio (SNR) measurement and render extraction of more cardiac parameters such as event durations of R-R interval, QRS complex, S-T segment, and T-wave from ECG data, which was not recorded in high accuracy with other non-standard ECG arrangements.
Fig. 1.
Measured R-peak amplitudes recorded from different locations across the left arm displayed as a biopotential heat map.
On the other hand, upon choosing electrodes’ arrangement, selecting sensing electrodes’ type is the next design challenge for a mobile ECG monitoring device. Due to their stable characteristics, silver/silver chloride (Ag/AgCl) “wet” electrodes supported by adhesive backrest and gel layer, to improve skin-electrode contact, are widely used in clinical ECG [16]. However, the conductive gel dehydrates in time and degrades the electrode’s SNR performance. Therefore, electrodes need to be replaced in a few hours or gel needs to be re-applied, which are neither practicable in everyday use nor easy to integrate into any mobile ECG solution. Moreover, the gel can cause an itching sensation and develop red and swollen skin immediately upon peeling the electrode. Such irritations and allergic reactions may last for several hours [17] or, in certain conditions, are reported to lead to dermatitis [18]. Therefore, it is preferable to use gel-free “dry” electrodes in continuous monitoring applications due to their superior long-term and multiple-day measurement performance [19]. Compared to other variants of dry electrodes, electronic textiles (e-textiles) appear to be the key technology in wearables non-invasive monitoring of biological indicators [20]—other competing technologies include tattoo-based and bio fluidic-based wearables [21], [22].
E-textiles promise flexibility, permeability to air and moisture, and accessible integration to daily clothing. Flexibility is essential to enable skin-compatible devices by matching the body’s natural contours, providing wearability, and achieving better skin-electrode coupling, whereas permeability to air and moisture alleviates the possibility of skin irritations [17]. Moreover, intelligent clothing—clothes enhanced with e-textiles to add functionality beyond traditional use, can embrace a critical role in a more extensive wireless body area network (WBAN) by including embedded flexible and stretchable antennas [23] and wireless transmission modules to enable continuous transfer of physiological information, which had sensed by the fabric itself, to a remote medical unit or the cloud; to be examined on the spot or later by professional healthcare staff. On the other hand, the flexible interconnection of the sensing sections of the intelligent clothes with electronic circuits is possible [24]. Additionally, in order to have truly “all-textile” intelligent clothes, research is actively progressing on the development of textile-based lithium-ion batteries [25] and textile-based energy harvesting mechanisms [26].
Fabrication techniques in textile electrodes also play a critical role in expanding intelligent clothing. Textile electrodes can be attached to daily clothes and accessories thanks to instream fabric manufacturing/decoration approaches such as knitting, weaving, and embroidery [27]. On the other hand, the sensing electrode section can also be applied onto finished textiles directly using techniques such as electroplating [28], physical vapor deposition (PVD) [29], chemical polymerization [30], dip-coating [31], or printing methods [32]. Especially, dip-coating and spray printing are promising since low manufacturing cost due to the adaptability of the existing commercial roll-to-roll (R2R) textile fabrication processes presents a serious game changer over the other e-textile fabrication methods, which usually require special equipment or processes that hinder their mass production (e.g., deposition functional thin films or patterning with lithography-based techniques in a cleanroom environment) [33], [34].
Graphene-based nanocomposites are emerging as attractive materials for biomedical applications, and various highly flexible and stretchable sensors have been fabricated using graphene for wearable technologies, including flexible strain gauges and electrochemical biosensors [35], [36]. Our group has been pioneering the synthesis of graphene e-textiles for biopotential signal monitoring, and we have introduced a cost- and time-effective scalable process based on dip-coating whereby conductive textiles are formed using graphene as a cladding layer [31]. Furthermore, previously, we benchmarked the performance of the developed graphene textiles with “gold standard” electrodes for acquisition of cardiac (electrocardiography, ECG) [31], neural (electroencephalography, EEG) [37], muscular (electromyography, EMG) [38], and ocular biopotentials (electrooculography, EOG) [39]. Later, we reported textile electrode integrated headband for human-computer interaction (HCI) and eye tracking [40], [41] and wearable band for step counting and muscle fatigue detection [42].
In this study, textile electrodes with different coating methods were introduced—one fabricated by dip-coating, and in the other one, the sensing area directly painted into the intelligent cloth by spray printing. We have custom-designed a front-end read-out circuitry for real-time acquisition, signal conditioning, and transmission of the single-arm ECG signals to a nearby laptop. Both devices were benchmarked in 5-minute long ECG recordings from multiple participants in static and dynamic settings in which they reached correlation values up to 98% with respect to “gold standard” electrodes. The embedded system detects heartrate (beat per minute, BPM) by counting R-wave peaks, and we simultaneously compared its readings with a standard fingertip photoplethysmography (PPG) device and observed excellent accuracy in heartrate detection. The heart-rate variability (HRV) was also extracted from the elapsed time between R-wave peaks where the root means successive square difference (rMMSD) metric was calculated as 30 ms during a 5 min long recording. Additionally, we studied other important cardiac parameters such as QRS complex duration, S-T segment duration, T-wave duration and compared the recorded values with the typical values of each participant. Results of our study demonstrate the feasibility of the developed wearable graphene textile-based single-arm diagnostic ECG prototype which we term as “SADE”.
2. Materials and methods
2.1. Fabrication of graphene textiles by dip-coating and spray printing
Graphene, a 2D single-layer structure of sp2 bonded carbon atoms formed in a honeycomb pattern, has recently attracted significant attention due to its unique electrical and structural properties [43]. Graphene-coated conductive textiles were prepared following a scalable and inexpensive, three-step “dip-dry-reduce” process to obtain conformal claddings of graphene over textile fibers. The fabrication process started with synthesis of graphene oxide (GO) following Hummer’s method and applying the GO solution (4 mg GO dispersed in mL of distilled water) onto textile surface using dip-coating and spray printing techniques. The dip-coating was applied by immersing the previously cut plain textile in GO solution with 3 × 3 cm2 dimension ( Fig. 2).
Fig. 2.
Fabrication of graphene e-textiles with (a) “dip-dry-reduce” cycle (∼ 3 ×3 cm2) and (b) “print-dry-reduce” (each spot ∼ 3 ×3 cm2).
Then, each coated piece was dried to achieve homogeneous GO-coated textiles. The surface texture, tightness, and thickness are essential factors in textile coating; these parameters were controlled during the process. Various textiles such as nylon, polyester, and cotton have been tested to obtain optimal deposition. Due to the low surface roughness, which resulted in homogeneous GO coating, nylon is selected as a plain textile to provide the best coating uniformity.
In parallel, the spray printing technique was engineered to patter the graphene textile coatings, which requires predetermined electrode positioning. These positions were selected through trial-and-error ECG experiments with the dip-coated electrodes on single-arm recordings. Then, stencil and textile were placed.
on a hot plate to accelerate the drying of the textile. The spray was held at ∼ 30 cm distance from the textile during the printing of GO and was followed by the “print-dry-reduce” fabrication cycle (Fig. 2). Next, chemical reduction was performed to create conformal electroconductive graphene structures using 4.33 mL of a 10 vol% saturated sodium hydroxide (NaOH) buffer solution and 227 mg of sodium borohydride (NaBH4) as a reducing agent following the process described by Guex [44]. The chemical reduction process disrupts the sp2 bond of graphite, making the epoxide and hydroxyl groups hydrophilic [45], [46]. The quality of the properties and potential application areas of the fabricated graphene may vary according to the selected reduction process [47].
After reduction, textiles were washed by immersing them in deionized water to remove unbounded chemical residues from the fiber surface, and only stable and well-adhered reduced graphene oxide (rGO) flakes remained on the textiles. Finally, they were placed in a vacuum oven and left to dry for a couple of hours. During the reduction of GO, there was a noticeable change in the textile color from brown to black due to the removal of oxygen-containing groups, which accompanies the increase of the electrical conductivity. Thus, the color change of the textile is a qualitative indicator of the effectiveness of the reduction process.
For prototyping, the prepared e-textile piece from the dip-coating procedure was fixed on a flexible sticky foam and sandwiched on an ordinary sports headband between metallic snap fasteners to establish electrical connection with the front-end circuitry. To acquire higher SNR and guarantee skin-electrode interface stability, we have used elastic bands with Velcro straps and polyethylene-based foam paddings to provide pressures in the range of a few mmHg (up to 5 mmHg) to support the contact of graphene-based e-textile sensors and form a stable high contact area on the skin surface. The procedure for printed textiles was more straightforward, and they required no cutting since graphene was homogeneously distributed on all its surfaces, including the edges.
In the dip-coating, the color of the edges was slightly darker compared to the central spots, indicating looser homogeneity distribution of the coating. In contrast, spray printing enables denser coating of the textile surface with the transfer of liquid droplets using pressurized air as opposed to bulk liquid transfer in the dip coating. Moreover, spray coating also allows for precise adjustment of the electrode locations in the final cloth as this approach enables patterning of the functional graphene layer onto a larger fabric. To expand the spray printing technique to accommodate additional electrodes and recording channels, the stencil (template) can be redesigned and produced using 3D printing or injection molding, making this approach adaptable for rapidly configuring and fabricating smart textiles on demand.
2.2. Scanning electron microscopy imaging and Raman spectroscopy on graphene textiles
Graphene textiles were imaged with a scanning electron microscope (SEM) and Raman spectroscopy was performed as a means to quickly confirm the coating and GO reduction steps. As illustrated in the scanning electron micrographs ( Fig. 3a), surfaces of the textile fibers show high contrast due to fairly homogenous layering of rGO while the electrically insulating cores of the fibers are bright and display charging. This is one visual indication of the existence of a conductive rGO coating on textile surfaces. Along with a visual confirmation, the changes in chemical structure during the reduction process of GO can be observed with Raman spectroscopy. Typically, crystalline graphite shows only a sharp Raman shift called the G-peak in the 1575–1581 cm−1 region, while GO and rGO display two broader Raman shifts named as the D- and G-band (ID, IG) occurring at slightly different locations and with different intensities [48], [49]. Reports in the literature have extensively studied the Raman spectrum of GO and rGO obtained with different synthesis and reduction chemistries, and while there may be slight differences in the D and G vibration bands, typically for GO the D-band can peak in the range of 1353–1363 cm−1 and G-band in the 1588–1598 cm−1 range; while for rGO, the D-band can peak in the range of 1349–1352 cm−1 and G-band as low as 1574 cm−1 and up to 1603 cm−1 range have been observed [49], [50], [51]. To verify the existence of the typical D and G vibration bands in the synthesized GO and rGO-textiles, we have performed Raman spectroscopy (Renishaw, England, UK) using 532 nm laser excitation. The recorded Raman spectrum of GO and rGO-textiles (Fig. 3b) display D and G vibration bands well aligned with the literature. Moreover, the ratio of the D and G peak intensities (ID/IG ratio) is used as a typical signature of GO to rGO conversion, whereby, an increase in the ID/IG ratio is expected upon reduction [52], [53]. While the ID/IG ratio was 0.89 for GO coating on textile, it increased to 1.38 upon chemical treatment of the GO-coated textile with NaBH4, which validates the reduction process and formation of reduced graphene oxide (rGO) on textile surfaces.
Fig. 3.
(a) Field emission scanning electron microscopy (FE-SEM) images of graphene textiles (insets show fiber cores and surfaces), and (b) Raman spectra of graphene oxide-coated textile (GO-textile, red), and the same textile after reduction (rGO-textile, black).
2.3. Skin-electrode impedance, sheet resistance, conductivity and washability
The skin-electrode impedance is a prominent performance metric for biopotential electrodes, where small and stable impedance values indicate better performance [54], [55]. The impedance of the graphene textile electrodes and the Ag/AgCl electrodes were compared. The three electrodes were placed on the forearm, ∼ 5 cm away from the wrist. The produced current was applied to the skin by the placed electrodes, and the reference electrode was held at ground potential. The measuring electrode was either an Ag/AgCl or textile electrode. The impedance of the Ag/AgCl electrode in the frequency range of 1 Hz to 1 kHz was measured between 74 kΩ and 16 kΩ, while the impedance value of the graphene-coated textile electrode ranged from 67 kΩ to 16 kΩ ( Fig. 4).
Fig. 4.
(a) Plot of the skin-electrode impedance of the Ag/AgCl electrodes and graphene textiles, and (b) conductivity map of a 10×8 cm sample of synthesized graphene textile extracted from 4-point probe sheet resistance measurements over 20 pieces of 2×2 cm textile pieces cut-out from the larger piece.
Lower impedance values for the graphene textile electrodes were attributed to having a larger area than the Ag/AgCl electrodes, 9 cm2 vs. 2.3 cm2, respectively, due to the inverse linear relationship between electrode area and skin-electrode impedance. Impedance values are affected by the electrode surfaces that are in contact with the skin, and since textile electrodes do not incorporate adhesives, their surface contact is not excellent, and the physical electrode area does not directly translate to the active electrode area.
In addition, the sheet resistance of graphene textiles was measured with a four-point probe unit. The sheet resistance and conductivity were found as ∼14 kΩ/square and ∼ 0.33 S/m, respectively, for a textile thickness of ∼ 0.2 mm (Fig. 4b). The developed graphene textile electrodes are highly flexible due to their soft, fabric nature and only weighing ∼ 50 mg.
One of the characterizations required for the long-term usability of textile electrodes is to assess their level of degradation to washing. Indeed, loss in conductivity is expected after a certain number of washing cycles due to absorption of water and mechanical stress load on textile fibers which may decrease or damage the conductive network. Starting with two identical electrodes, one was cleaned in a washing machine (40 °C at 400 rpm for 30 min), and the other was gently dipped to detergent water and kept without movement for 30 min (hand washed). It was found that after five washing cycles, hand washed electrode’s resistivity increased 50%. In contrast, the resistivity of the other electrode increased by 150% due to the high mechanical stress induced by the washing machine, indicating hand wash should be preferred for graphene textiles to preserve their long-term usage. Additionally, surface resistance increased from ∼ 45 kΩ to ∼ 65 kΩ in five handwashing cycles. However, considering the overall circuit model of the skin-electrode (with impedance ZE) and analog front-end (input impedance ZIN) as two impedances in series, the washing cycles did not prevent reusage of the electrodes for ECG measurements, due to much higher input impedance of the analog front-end amplifier compared to the skin-electrode impedance, and the gain provided by the amplifier [17].
2.4. Design of the front-end readout circuitry
It is imperative to condition the signal and filter the motion artifacts and power-line noise to record ECG signals with an excellent SNR. Overview of the complete, wearable single arm ECG monitoring and acquisition system with printed graphene electrodes and embedded electronics is shown in Fig. 5, along with the schematic of the custom-designed readout unit. The values of circuit components are summarized in Table 1. At the analog front-end, the circuit obtains its surface potentials via graphene textile electrodes, and after onboard conditioning, they are sent to the microcontroller (MCU) for digitization and further processing. Then, signals are transmitted via a Bluetooth connection (HC06) to the PC with a custom-developed graphical user interface in LabVIEW (National Instruments, USA), where real-time data is stored and tracked.
Fig. 5.
Overview of the single arm, real-time ECG monitoring and acquisition system with printed graphene electrodes and readout circuit all built onto a wearable elastic armband, along with the hardware-level schematic of the embedded system and the custom-designed graphical user interface (GUI).
Table 1.
List of components and specifications for the signal acquisition system.
| Circuit Blocks | Specifications | |
|---|---|---|
| Pre-amplifier | R1,R2: 1 kΩ C1,C2: 1 nF C3: 100nF | |
| Protection circuitry | R3,R4: 20 kΩ | |
| DRL circuitry | R5: 10 kΩ R6,R7: 390 kΩ C4: 1µF C5: 270nF | |
| High-Pass Filter | R8,R9: 36 kΩ R10: 4.7 kΩ R11,R12: 1 kΩ C6,C7: 1µF C8: 1nF | |
| Low-Pass Filter | R13,R14: 10 kΩ R15: 160 Ω R16: 330 kΩ C9,C11: 1µF C10,C12: 270nF |
The onboard analog circuitry has second- and fourth-order Sallen-Key Butterworth high-pass and low-pass filters with 2 Hz and 39 Hz cut-off frequencies, respectively. The values for the filters were optimized and simulated using Texas Instrument’s online filter design tool. The instrumentation amplifiers (INA122, Texas Instruments, USA) and op-amps (OPA2365, Texas Instruments, USA) suit the battery operation with a single lithium-ion polymer battery (3.7 V and 500 mAh). A voltage divider (MAX5421, Maxim, USA) was used to adjust the gain in the post-amplification level. MCP73831 (Microchip, USA) and TPS61090 (Texas Instruments, USA) were used in the circuitry to add battery charging capability and DC-DC boosting. A rail-splitter (TLE2426, Texas Instruments, USA) with a voltage divider with a buffer circuit was applied to avoid unbalanced occurrences when dividing the regulated 5 V. The entire system was placed in a 6.5×5 x 2.75 cm3 transparent case and it weights ∼ 53 g.
3. Results and discussion
3.1. Single-arm ECG with dip-coated graphene textiles
In order to verify the feasibility of the developed graphene textile electrodes with dip-coating in sensing single-arm electrocardiograms, they were benchmarked against the standard Ag/AgCl electrodes (Foam Monitoring Electrodes Ref 2228 3M Red Dot, USA). ECG recording experiments were performed in static and dynamic conditions for 5 min each, where the subjects were instructed to sit in a relaxed position and walk, respectively. Due to the importance of the electrode position on the signal quality, signals were asynchronously recorded so that the comparisons could be from the exact electrode locations (first with armbands, then wet electrodes). There were 8 participants (3 female, 5 male), and none reported a CVD history. Since the measurement was done asynchronously, the heartbeats and, therefore, the P-QRST-T complex were naturally not synchronous in recordings of different electrode types. Thus, we selected a 15 s spot with matching heartrates from each recording and synchronized them in data processing for obtaining correlation values. To quantify the overlap between the optioned signals, the built-in linear correlation function of MATLAB® (Mathworks, USA) was used. The correlation coefficients summarized in Table 2 reveal a maximum correlation of 96% obtained ECG recordings from participant 1 (P1) and minimum of 84% observed in participant 8 (P8) with an overall average of 88.6% in the static trials, and a maximum of 93% for P1 and minimum of 70% for P3 with an overall average of 84% in the dynamic trials.
Table 2.
Correlation and SNR comparison of single-arm ECG recordings obtained with dip-coated and spray-printed graphene textile electrodes against commercial Ag/AgCl “wet” electrodes.
| Participant (P) No / Age |
Correlation [%] |
SNR Sitting [dB] |
SNR Walking [dB] |
|||
|---|---|---|---|---|---|---|
| Sitting | Walking | Ag/AgCl | Dip-coated Graphene Textile | Ag/AgCl | Dip-coated Graphene Textile | |
| P1 / 26 | 96% | 93% | 13.7 | 13.3 | 14.1 | 12.3 |
| P2 / 22 | 93% | 85% | 25.9 | 24.1 , |
16.0 | 20.2 - |
| P3 / 45 | 85% | 70% | 22.0 | 24.1 | 21.4 | 20.2 |
| P4 / 31 | 90% | 89% | 23.0 | 21.9 | 18.9 | 19.4 |
| P5 / 24 | 87% | 86% | 24.1 | 21.4 | 15.7 | 18.6 |
| P6 / 52 | 87% | 74% | 20.7 | 22.6 . |
18.0 , |
23.7 |
| P7 / 27 | 87% | 89% | 15.1 | 13.8 | 15.0 | 15.9 |
| P8 / 26 | 84% | 86% | 18.3 | 23.3 | 18.1 | 23.1 |
| Mean: 31.6 | 88.6% | 84% | 20.4 | 20.6 | 17.2 | 19.2 |
| Sitting | Walking | Ag/AgCl | Spray-coated Graphene Textile | Ag/AgCl | Spray-coated Graphene Textile | |
| P7’ / 27 | 93% | 84% | 25.8 | 27.5 | 22.0 | 22.2 |
| P8’ / 26 | 95% | 92% | 26.9 | 30.6 | 22.1 | 25.0 |
| Mean: 26.5 | 94% | 88% | 26.3 | 29.1 | 22.1 | 23.6 |
Fig. 6 illustrates the results from participant #1 who had the highest correlation coefficient on both trials, and each segment of P-QRS-T morphology is observed both in static and dynamic trials in single-arm ECG recordings obtained with Ag/AgCl electrodes and dip-coated graphene textiles. During the walking trial, P-wave was the most challenging feature to observe from all the participants. This is attributed to the nature of single-arm ECG measurement approach, which results in lower P-wave intensities. During the dynamic trial, it was observed that dip-coated graphene textiles were more prone to movement artifacts than standard electrodes (especially in middle-aged participants); however, during the static trial, no such differences were observed. As seen in Fig. 7, dip-coated graphene textiles record more fluctuation in the ECG signals’ baseline than standard electrodes, emphasizing the importance of developing robust algorithms to ensure stable recordings in wearable technologies.
Fig. 6.
Single-arm ECG of a participant with the highest correlation coefficient in the static and dynamic condition, recorded by the dip-coated graphene textile electrodes.
Fig. 7.
Single-arm ECG with spray printed-based prototype in the static and dynamic experiments. Compared to the dip-coated-based prototype, correlation coefficients are high even in longer durations, thanks to the superior coating.
The correlation coefficient assesses the similarity of the picked-up biopotentials by electrodes, and SNR values determine the quality of the ECG measurement itself. SNR values were calculated by applying wavelet transform in Daubechies-4 (db-4) basis to the signals after analog conditioning, and Donoho’s threshold was used to separate signal information (P-QRST-T morphology) from the noise (baseline and its fluctuation) [56]. It was observed that signals taken from graphene textiles were slightly better than Ag/AgCl electrodes (Table 2) since graphene electrodes had slightly better skin-electrode impedance values in low frequencies. As SNR values inherently differ from person to person due to physiological variations, a fair SNR assessment would be to compare the different electrodes and trials for the same participant as opposed to comparing the averaged values of all the participants. This is fundamentally due to the ECG intensity difference per person (especially R-wave peak intensity); that is, if the ECG signal’s overall intensity is more than the baseline, it will have high SNR.
Factors such as fitness and age have a role in achieving high SNR from ECG signals. For instance, if recording happens to be picked up from individuals with more fat tissue on their arm, due to the low-pass filtering of these tissues, surface electrodes will pick lower ECG signal intensity than more slim participants, and their SNR values will be lower. However, this does not mean their correlation coefficient is low [57]. Likewise, individuals with body hair on arms also record less SNR intensity with textile and clinical electrodes. For instance, participant 1 had the highest correlation value, but his SNR values were below average. As discussed, this is attributed to the quality of the ECG signal picked up irrespective of the type of electrodes.
3.2. Single-arm ECG with spray-printed graphene textiles
Right after experimentation with the first prototype based on dip-coated graphene textiles, participants 7 and 8 who were among the “poor” performers in terms of both correlation and SNR values, were selected to test the spray-printed graphene textiles and ECG signals were recorded using the same experimental protocols. As previously mentioned, three graphene textile electrodes separated roughly ∼ 10 cm apart were directly printed onto a bare nylon textile surface, unlike the dip-coating approach where the textile electrodes were sewn onto a piece of nylon textile which acted as a “carrier”. Fig. 7 shows sample of a recorded ECG signal from participant 8 under static and dynamic conditions for a duration of 15 s with a clear display of the P-QRS-T complex, as well as, a longer recorded signal of more than 1 min where R-peaks are visible.
The correlation coefficients among signals recorded with two different electrodes (spray-printed graphene textiles and “wet” Ag/AgCl electrodes) along with SNR values for each participant under static and dynamic conditions are listed in Table 2 as P7’ and P8’. Results show that, on average, the second prototype based on spray-printed graphene textiles achieves better correlation rates in single-arm ECG. For participant 7, the correlation value improved from 87% to 93% for the static case (sitting), while for the dynamic case (walking) there was a decrease from 89% to 84%. On the other hand, for participant 8, the correlation values improved both for the static and dynamic scenarios, from 84% to 95% and 86–92%, respectively. Also, stronger signals were recorded for spray-printed graphene textiles despite using the same acquisition circuit. For example, for participant 8, the signal intensity was around ∼ 500 mV peak to peak in sitting position with the dip-coated graphene textiles. In contrast, spray-printed graphene textiles displayed almost double the voltage value (∼ 1 V) under the same testing conditions. A similar improvement in the recorded signal intensity was also observed in the ECG data in the walking trial and with the other participant.
Likewise, the average of the SNRs for participants seven and eight increased from 18.55 dB (dip-coated textile) to 29.1 dB (spray-printed textile); and 19.5 dB (dip-coated textile) to 23.6 dB (spray-printed textile); for sitting and walking cases, respectively. The typical ECG characteristics, P-wave, QRS complex, and T wave, were clearly visible in the recorded electrocardiograms. This result motivated us to pursue automatic detection of specific ECG parameters with the embedded system, which will be described in detail in the next section. The improvement observed in the ECG acquisition performance of spray-printed graphene textile electrodes compared to dip-coated ones is attributed to improved coating quality and better optimization of electrode positions on the armband. Coupling these with the inherent advantages of spray printing in fabrication (i.e., fast patterning/placement and scalability to mass production), we can argue that among the two approaches presented here, spray printing is the superior alternative for fabricating graphene textile electrodes.
3.3. Real-time detection of heartrate (BPM) and heart-rate variability (HRV)
In parallel with the real-time data acquisition, the system detects critical ECG parameters such as R-R interval, heartrate (BPM), and heart-rate variability (HRV) and reports them to the operator in real-time. HRV indicates the change in the intervals of the consecutive heartbeats and is a critical index for monitoring the cardiovascular system and routinely used for detecting different types of arrhythmia [58]. HRV is also associated with physical fitness and psychological well-being and is used in recovery and stress analysis [59]. During the COVID-19 global pandemic, the HRV data is heavily analyzed for early disease diagnosis [60], and also to study the effect of isolation on mental health and the stress levels of individuals [61].
R-wave peaks were detected with a thresholding-based algorithm in which a single-arm ECG signal was smoothed twice with a moving root mean square (RMS) filter with a 150 ms window. The algorithm calculates the threshold parameter with an automated calibration upon starting the recording. In every 2-minutes of R-wave peak detection, approximately two T-waves were mistakenly overcounted as R-peaks, which is still an acceptable error rate for a simple thresholding-based heartrate detection algorithm. After calculating the R-R interval, root mean square successive difference (rMSSD) was calculated, which is one of the popular time-domain detection metrics for HRV [62]. For the previously mentioned 5-minute trial, the rMSSD-HRV value was calculated as 30 ms, which lies inside the typical expected range (i.e., 18–74 ms) for a 30-year old participant.
Since the developed system can capture the intricate features of the P-QRS-T complex with an acceptable SNR and sufficient intensity ( Fig. 8a); information on other critical cardiac parameters can be extracted from the P-QRS-T complex ( Table 3). Accordingly, the mean heartrate for each of the previously introduced eight participants was recorded in different conditions such as sitting, standing, and walking; and average of the means and standard deviations (STD) for all participants were listed, along with the R-R interval (essential for HRV calculation) and durations of the QRS complex, ST segment, and T-wave (Table 3).
Fig. 8.
(a) ECG waveform recorded from a single-arm using spray printed graphene textiles showing the P-QRS-T complex, P-wave duration (red), QRS complex duration (orange), S-T segment’s duration (green), T-wave duration (blue). The total number of R-waves correspond to the heartrate and the time elapsed between two successive R-waves represent the heart-rate variability (HRV), and a custom-written offline thresholding-based algorithm automatically detects both parameters. (b) Snapshot from benchmarking BPM values between commercially available photoplethysmography (PPG) based pulse oximeter (BPM = 70) and by the developed single-arm ECG (BPM = 70).
Table 3.
Tabular summary of heartrate (BPM) values for eight participants in sitting, standing, and walking conditions, and other critical cardiac parameters (R-R interval, durations of QRS, ST segment, and T-wave) obtained through single-arm ECG measurements with the custom-made wearable graphene textile armband system and a commercial system; along with typical chest ECG parameters.
| Single-arm ECG * Custom system |
Single-arm ECG § Commercial system |
Typical chest ECG ┬ Clinical system |
|
|---|---|---|---|
| Parameter | Mean + STD | Mean + STD | Reference range |
| BPM sitting [count] | 74.4 ± 8.6 | 72.4 ± 6.2 | 60 –100 [63] |
| BPM standing [count] | 87.3 ± 8.7 | 76.2 ± 6.6 | 70 – 110 [63] |
| BPM walking [count] | 83.6 ± 9.9 | 76.4 ± 6.3 | 80 – 140 [63] |
| R-R interval [ms] | 834.1 ± 81.4 | 760 ± 92.2 | 600 – 1200 [64] |
| QRS duration [ms] | 144.3 ± 11.9 | 90.2 ± 14.2 | ≤ 120 [65] |
| ST duration [ms] | 129.8 ± 10.7 | 120.4 ± 10.1 | 80 – 120 [66] |
| T duration [ms] | 220.9 ± 16.7 | 204.6 ± 19.2 | 100 – 250 [67] |
STD: standard deviation, BPM: beat per minute
Custom system: wearable graphene textile electrodes and custom-designed ECG acquisition board
Commercial system: pre-gelled Ag/AgCl “wet”electrodes and OpenBCI Cyton Board
Clinical system: pre-gelled Ag/AgCl “wet” electrodes and clinical ECG monitoring systems
In addition, to benchmark the extracted values from the developed system against a reference system, gold-standard pre-gelled Ag/AgCl wet electrodes were also used in conjunction with a commercial general purpose biopotential signal acquisition unit (OpenBCI Cyton Board, USA) and the single-arm ECG recording experiment was repeated. Although the single-arm ECG data recorded using the SADE system (wearable graphene textiles and custom-made readout circuit in this work) and the reference system (Ag/AgCl electrodes and OpenBCI Cyton Board) were obtained from different participants, the standard deviation of all extracted parameters turn out to be similar, and also in alignment with clinically standard chest ECGs from literature [63], [64], [65], [66], [67], verifying the high accuracy of the developed system.
An off-the-shelf photoplethysmography (PPG) heartbeat tracker was used to benchmark the recordings for BPM measurements (Fig. 8b). During the measurements, no significant misreading (i.e., error of more than two counts) was observed between the recorded heartrate from the PPG and single-arm ECG. All the participants’ heartbeats in standing conditions are more than sitting and walking cases. Due to the experimental method that we followed, the participants were not briefed about the experiment procedure, and when they were asked to stand up after ∼ 5 min of stationary sitting, this might induce anxiety and increase their average heartbeat.
Medical ECG interpretation includes assessment of the wave morphology, their intensity, and duration of the intervals of the ECG curve. The P-wave typically reflects atrial depolarization, and its evaluation is primarily based on validating whether the impulse conduction from the atria to the ventricles is normal or not [68]. Although P-wave was observed during the single-arm ECG, its intensity was poor. The QRS complex dominantly represents the depolarization of the left ventricle. Short or broad depolarization of the QRS complex is among the features that professional healthcare staff evaluate during medical diagnosis [69]. In our single-arm ECG measurements, we observed a slightly larger QRS complex duration than typical chest electrocardiograms. The S-T segment which corresponds to the plateau phase of the action potential is yet another critical feature, whereby both its magnitude and duration are essential in diagnosing different cardiac abnormalities, and in particular to identify the onset of acute myocardial ischemia [70]. The single-arm ECG slightly exceeds the typical S-T segment duration of chest ECG. On the other hand, the T-wave reflects the rapid repolarization of contractile cells and the change in this wave occurs in an extensive range of conditions [71]. In our study, the recorded T-waves from single-arm ECG matched with typical chest ECG waveforms.
3.4. Electrocardiogram recording as a screening tool for at-home sleep monitoring
A crucial consideration for the successful development of wearable electronics is their robust long-term performance which is fundamentally related to the sensor’s fabrication, front-end readout hardware design, and system packaging. To demonstrate the superior performance and comfortable use of the developed single-arm ECG monitoring system (SADE), we performed a 45-minute-long sleep study ( Fig. 9). The system was worn by a 32-year-old male participant without any diagnosed cardiovascular diseases and a long-term ECG measurement was successfully performed while the participant was taking an afternoon nap. The noise of the single-arm ECG signal can be evaluated using RMS analysis, which indicates low fluctuations of the recorded ECG signal over time. Here it changed from 3.69 μV to 3.48, to 3.67, to 3.82, and finally to 3.57 μV recorded in the 8th, 16th, 24th, 32th, and 40th minutes, respectively. The coefficient of variation (CV) is 3.54%, showing a minimal change of RMS noise during 45-minute of continuous monitoring, concluding that SADE can be used for long-term healthcare monitoring, as supported by high-quality ECG recordings.
Fig. 9.
Long-term monitoring of single-arm ECG using smart graphene textiles for 45 min during a sleep study.
3.5. Figure of merit
After presenting the system’s feasibility, we compared our developed single-arm ECG diagnostic unit with commercially available gel-less wearable ECG units ( Table 4). “Placement” refers to where the biopotential sensing electrodes are attached. In this work, they are in contact solely with the left arm. In commercial units, they are attached to the chest area. For “usability,” the classification was based on how wearable the system is. We categorized socially-discreet systems integrated into daily clothing as “highly wearable.” If they have authentic and stylish packaging but are not embedded into everyday clothing, we categorized them as “moderately wearable”, and the others as “slightly wearable”. The developed system here has the lowest cost while still offering acceptable size and weight matching commercially available units. It stands out as a highly wearable and compact system, thanks to its unique single-arm electrode placement location. On the contrary, the other highly wearable units are in fact much larger (e.g., the size of a T-shirt) and rely on ECG acquisition from the chest which hinders their use a separate add-on accessory to ones’ existing clothing preference or style. We envision that with better packaging, the system presented here can be even lighter and more compact since each electrode only takes 30 × 30 mm with negligible height and weight of only ∼ 50 mg.
Table 4.
Comparison of the current study with commercially available wearable ECG units.
| Price | Weight (g) | Placement | Size (mm) | Usability | |
|---|---|---|---|---|---|
| Thıs Work → | < $100 | 155 | Left Arm | 70 × 54 × 28 | Highly Wearable |
| Garmin's HRM-Pro | $130 | 59 | Chest | 10 × 0.3 × 0.12 | Moderately Wearable |
| Qardio | $500 | 130 | Chest | 193 × 102 × 10 | Moderately Wearable |
| Vivalink | − | 7.5 | Chest | 90 × 20 × 7.9 | Moderately Wearable |
| LevMed ECG Strap | $300 | NA | Chest | − | Slightly Wearable |
| HeartIn Fit ECG Shirt | $300 | NA | Torso | 61.3 × 45.6 × 5.5 | Highly Wearability |
| Hexoskin Pro Kit | $650 | NA | Torso | − | Highly Wearability |
4. Conclusion
Health monitoring of daily patient activities provides valuable information to medical staff, and portable systems are ideal candidates for remote health surveillance. In this study, we have developed two wearable single-arm electrocardiography (ECG) prototype systems with embedded graphene textile electrodes made by different coating methods: dip-coating and spray printing. Comparing the skin-electrode impedance values to commercial silver/silver chloride (Ag/AgCl) electrodes presented a very similar level of performance for the graphene textile electrodes. Furthermore, wearable systems were tested on 8 different participants to benchmark their performance. Signals taken asynchronously from Ag/AgCl and graphene textile electrodes were compared in terms of correlation percentage and signal-to-noise ratio (SNR) values; on both of these, textile electrodes performed toe-to-toe with Ag/AgCl electrodes. Average correlation values obtained from 8 participants reach ∼ 90% in static conditions (sitting) and ∼ 85% in dynamic conditions (walking), while SNR values were nearly equal. Spray printed electrodes performed better than dip-coated ones, indicating that spray printing is the preferred textile finishing technique for rGO textile fabrication. Among the non-conventional ECG electrode arrangements, single-arm ECG achieved continuous monitoring with less movement and high signal intensity. The integration of the developed technology will make the devices significantly cheaper and attractive for everyday applications ranging from cardiac health trackers to psychological stress monitoring and human-computer/human-machine interactions (HCI/HMI).
Funding
This work was supported in part by Sabanci University and The Scientific and Technological Research Council of Turkey (TUBITAK) grant number 20AG028 and 119C091. Professor Murat Kaya Yapici appreciates the support of the Turkish Academy of Sciences (TUBA) within the framework of the TUBA Outstanding Young Scientist Award Program (GEBIP).
Author contributions
OO performed experiments, modified the acquisition circuit, wrote scripts for signal processing and graphical user interface, constructed the board enclosure, performed conductivity and skin-electrode impedance measurements. AG designed the initial acquisition unit. GA synthesized the electroconductive textiles and performed experiments. SG assisted in performing the measurements. AG, GA, and MKY analyzed the data. OO, AG, GA and MKY wrote the manuscript. OO, AG, SG, and MKY revised the manuscript. MKY conceived the research plan and designed the experiments, reviewed the manuscript, supervised and led the project.
CRediT authorship contribution statement
Ozberk Ozturk: Methodology, Measurements, Validation, Formal analysis, Investigation, Data curation, Software, Visualization, Writing – original draft, Writing – review & editing. Ata Golparvar: Methodology, Measurements, Validation, Formal analysis, Investigation, Data curation, Software, Visualization, Writing – original draft, Writing – review & editing. Gizem Acar: Measurements, Investigation, Data curation, Writing – original draft. Saygun Guler: Measurements, Investigation, Data curation, Writing – review & editing. Murat Kaya Yapici: Conceptualization, Methodology, Validation, Formal analysis, Investigation, Visualization, Resources, Funding acquisition, Supervision, Project Administration, Writing – original draft, Writing – review & editing.
Declaration of Competing Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
This article is partially based on the M.Sc. thesis of G. Acar. The authors gratefully thank the participants involved in this study. The authors thank Osman Sahin for his help in acquiring the SEM image of graphene textile fibers.
Institutional review board statement
The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the Ethics Committee of Sabanci University (FENS-2020–48).
Informed consent statement
The authors declare no conflict of interest.
Biographies

Ozberk Ozturk received his B.S. from Middle East Technical University of Ankara, Turkey in 2017 and his M.Sc. from Sabanci University of Istanbul, Turkey in 2020. Currently, he is a PhD candidate in the electronics engineering program of Sabanci University. His research interests include wearable electronics, textile electrodes and MEMS system design and fabrication.

Ata Golparvar received the B.S. degree in electrical and electronics engineering from the Azad University of Tabriz, Iran in 2016, and M.Sc. degree in electronic engineering from Sabanci University, Istanbul, Turkey in 2019. Currently he is a Ph.D. candidate in the Microsystems and Microelectronics program of Swiss Federal Institute of Technology Lausanne (EPFL), Switzerland. His research interests include wearable electronics and their applications, textile electrodes, biomedical instrumentation and embedded systems.

Gizem Acar received her B.S. degree in Physics from the Middle East Technical University-METU, Ankara, Turkey in 2017, and her M.Sc. degree from Sabanci University, Turkey in 2020. During her studies, she worked part-time at METU Technopark and METU Physics department primarily on subjects related to digital design. Later in her Masters, she focused on wearable electronics and textile electrodes. She is currently a PhD student in Lancaster University, UK.

Saygun Guler completed his MSc-Biomedical Engineering at the University of Dundee. Later he worked in digital healthcare industry for two years in Southampton Science Park, UK, where he served as a biosignal engineer. He is currently a graduate researcher pursuing his PhD in SU-MEMS research group in Sabanci University, Istanbul, Turkey.

Murat Kaya Yapici received the B.S. and Ph.D. degrees in electrical engineering from Texas A&M University, College Station, TX, USA, in 2004 and 2009, respectively. From 2009–2010, he was a Postdoctoral Research Associate with the Solid State Electronics, Photonics and Nano-Engineering Laboratory at Texas A&M University. He is the director and principal investigator of Sabanci University Micro/Nano Devices and Systems Lab (SU-MEMS), a faculty member in the Electronics Engineering Program of Sabanci University, Istanbul, Turkey, and an affiliate faculty member in the Department of Electrical Engineering, University of Washington, Seattle, WA, USA. His research interests include MEMS/NEMS, nanotechnology, sensors/actuators, wearable and flexible electronics, point-of-care devices, microfluidics, acousto-optic devices for biological and medical applications, as well as semiconductor process technology, novel nanofabrication based on scanning probes and system-level integration of nanomaterials, MEMS, and CMOS. Dr. Yapici is a member of SPIE and URSI. He has been on the technical program committee of IEEE Sensors, IEEE BSN-BHI conferences.
Data availability
No data was used for the research described in the article.
References
- 1.Franchi A., Franchi L., Franchi T. Digital health, big data and connectivity: 5G and beyond for patient-centred care. Int. J. Digit. Health. 2021;1 [Google Scholar]
- 2.Jayaraman P.P., Forkan A.R.M., Morshed A., Haghighi P.D., Kang Y.B. Healthcare 4.0: a review of frontiers in digital health. Wiley Interdiscip. Rev.: Data Min. Knowl. Discov. 2020;10:234–238. e1350. [Google Scholar]
- 3.Al-Khafajiy M., Baker T., Chalmers C., Asim M., Kolivand H., Fahim M., Waraich A. Remote health monitoring of elderly through wearable sensors. Multimed. Tools Appl. 2019;78:24681–24706. [Google Scholar]
- 4.Tavakoli M., Carriere J., Torabi A. Robotics, smart wearable technologies, and autonomous intelligent systems for healthcare during the COVID‐19 pandemic: an analysis of the state of the art and future vision. Adv. Intell. Syst. 2020;2 [Google Scholar]
- 5.Caizzone A. An ultra low-noise micropower PPG sensor. EPFL. 2020 [Google Scholar]
- 6.Kendir C., van den Akker M., Vos R., Metsemakers J. Cardiovascular disease patients have increased risk for comorbidity: a cross-sectional study in the Netherlands. Eur. J. Gen. Pract. 2018;24:45–50. doi: 10.1080/13814788.2017.1398318. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Bayoumy K., Gaber M., Elshafeey A., Mhaimeed O., Dineen E.H., Marvel F.A., et al. Smart wearable devices in cardiovascular care: where we are and how to move forward, Nature Reviews. Cardiology. 2021:1–19. doi: 10.1038/s41569-021-00522-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.AlGhatrif M., Lindsay J. A brief review: history to understand fundamentals of electrocardiography. J. Community Hosp. Intern. Med. Perspect. 2012;2:14383. doi: 10.3402/jchimp.v2i1.14383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Boukhayma A., Barison A., Haddad S., Caizzone A. Ring-embedded micro-power mm-sized optical sensor for accurate heart beat monitoring. IEEE Access. 2021;9:127217–127225. [Google Scholar]
- 10.Boukhayma A., Barison A., Haddad S., Caizzone A. Earbud-embedded micro-power mm-sized optical sensor for accurate heart beat monitoring. IEEE Sens. J. 2021;21:19967–19977. [Google Scholar]
- 11.Isakadze N., Martin S.S. How useful is the smartwatch ECG? Trends Cardiovasc. Med. 2020;30:442–448. doi: 10.1016/j.tcm.2019.10.010. [DOI] [PubMed] [Google Scholar]
- 12.Yapici M.K., Alkhidir T.E. Intelligent medical garments with graphene-functionalized smart-cloth ECG sensors. Sensors. 2017;17:875. doi: 10.3390/s17040875. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.N.K. Jacob, E. Balaban, R. Saunders, J.C. Batchelor, S.G. Yeates, A.J. Casson, An exploration of behind-the-ear ECG signals from a single ear using inkjet printed conformal tattoo electrodes, 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), IEEE 2018, pp. 1283–6. [DOI] [PubMed]
- 14.Guler S., Golparvar A., Ozturk O., Yapici M.K. Ear electrocardiography (ECG) with soft graphene textiles for hearable applications. IEEE Sens. Lett. 2022;6:1–4. [Google Scholar]
- 15.von Rosenberg W., Chanwimalueang T., Goverdovsky V., Peters N.S., Papavassiliou C., Mandic D.P. Hearables: feasibility of recording cardiac rhythms from head and in-ear locations. R. Soc. Open Sci. 2017;4 doi: 10.1098/rsos.171214. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Webster J.G. John Wiley & Sons; 2009. Medical instrumentation: application and design. [Google Scholar]
- 17.Acar G., Ozturk O., Golparvar A.J., Elboshra T.A., Böhringer K., Yapici M.K. Wearable and flexible textile electrodes for biopotential signal monitoring: a review. Electronics. 2019;8:479. [Google Scholar]
- 18.Uter W., Schwanitz H. Contact dermatitis from propylene glycol in ECG electrode gel. Contact Dermat. 1996;34:230–231. doi: 10.1111/j.1600-0536.1996.tb02190.x. [DOI] [PubMed] [Google Scholar]
- 19.Niu X., Gao X., Liu Y., Liu H. Surface bioelectric dry electrodes: a review. Measurement. 2021;183 [Google Scholar]
- 20.Pani D., Achilli A., Bonfiglio A. Survey on textile electrode technologies for electrocardiographic (ECG) monitoring, from metal wires to polymers. Adv. Mater. Technol. 2018;3 [Google Scholar]
- 21.Ha T., Tran J., Liu S., Jang H., Jeong H., Mitbander R., et al. A chest‐laminated ultrathin and stretchable E‐tattoo for the measurement of electrocardiogram, seismocardiogram, and cardiac time intervals. Adv. Sci. 2019;6 doi: 10.1002/advs.201900290. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Iqbal S.M., Mahgoub I., Du E., Leavitt M.A., Asghar W. Advances in healthcare wearable devices, npj Flexible. Electronics. 2021;5:1–14. [Google Scholar]
- 23.Xie Z., Avila R., Huang Y., Rogers J.A. Flexible and stretchable antennas for biointegrated electronics. Adv. Mater. 2020;32 doi: 10.1002/adma.201902767. [DOI] [PubMed] [Google Scholar]
- 24.Agcayazi T., Chatterjee K., Bozkurt A., Ghosh T.K. Flexible interconnects for electronic textiles. Adv. Mater. Technol. 2018;3 [Google Scholar]
- 25.Chen X., Ma Y. Wearable lithium ion batteries based on carbon nanotubes and graphene. Adv. Mater. Technol. 2018;3 [Google Scholar]
- 26.Yang B., Xiong Y., Ma K., Liu S., Tao X. Recent advances in wearable textile‐based triboelectric generator systems for energy harvesting from human motion. EcoMat. 2020;2 e12054. [Google Scholar]
- 27.Stoppa M., Chiolerio A. Wearable electronics and smart textiles: a critical review. Sensors. 2014;14:11957–11992. doi: 10.3390/s140711957. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Cho G., Jeong K., Paik M.J., Kwun Y., Sung M. Performance evaluation of textile-based electrodes and motion sensors for smart clothing. IEEE Sens. J. 2011;11:3183–3193. [Google Scholar]
- 29.Wei Q., Xu Y., Wang Y. Surface modification of textiles. Elsevier,; 2009. Textile surface functionalization by physical vapor deposition (PVD) pp. 58–90. [Google Scholar]
- 30.Lv J., Zhou P., Zhang L., Zhong Y., Sui X., Wang B., Chen Z., Xu H., Mao Z. High-performance textile electrodes for wearable electronics obtained by an improved in situ polymerization method. Chem. Eng. J. 2019;361:897–907. [Google Scholar]
- 31.Yapici M.K., Alkhidir T., Samad Y.A., Liao K. Graphene-clad textile electrodes for electrocardiogram monitoring. Sens. Actuators B: Chem. 2015;221:1469–1474. [Google Scholar]
- 32.Takamatsu S., Lonjaret T., Crisp D., Badier J.-M., Malliaras G.G., Ismailova E. Direct patterning of organic conductors on knitted textiles for long-term electrocardiography. Sci. Rep. 2015;5:1–7. doi: 10.1038/srep15003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Paul G., Torah R., Beeby S., Tudor J. Novel active electrodes for ECG monitoring on woven textiles fabricated by screen and stencil printing. Sens. Actuators A: Phys. 2015;221:60–66. [Google Scholar]
- 34.Karaguzel B., Merritt C.R., Kang T., Wilson J.M., Nagle H.T., Grant E., Pourdeyhimi B. Utility of nonwovens in the production of integrated electrical circuits via printing conductive inks. J. Text. Inst. 2008;99:37–45. [Google Scholar]
- 35.Krishnan S.K., Singh E., Singh P., Meyyappan M., Nalwa H.S. A review on graphene-based nanocomposites for electrochemical and fluorescent biosensors. RSC Adv. 2019;9:8778–8881. doi: 10.1039/c8ra09577a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Irani F.S., Shafaghi A.H., Tasdelen M.C., Delipinar T., Kaya C.E., Yapici G.G., Yapici M.K. Graphene as a piezoresistive material in strain sensing applications. Micromachines. 2022;13:119. doi: 10.3390/mi13010119. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Golparvar A., Ozturk O., Yapici M.K. Gel-free wearable electroencephalography (EEG) with soft graphene textiles, 2021 Ieee. Sens., IEEE. 2021:1–4. [Google Scholar]
- 38.Ozturk O., Yapici M.K. Muscular activity monitoring and surface electromyography (sEMG) with graphene textiles, 2019 IEEE. Sens, IEEE. 2019:1–4. [Google Scholar]
- 39.Golparvar A.J., Yapici M.K. Electrooculography by wearable graphene textiles. IEEE Sens. J. 2018;18:8971–8978. [Google Scholar]
- 40.Golparvar A.J., Yapici M.K. Graphene smart textile-based wearable eye movement sensor for electro-ocular control and interaction with objects. J. Electrochem. Soc. 2019;166:B3184–B3193. [Google Scholar]
- 41.Golparvar A.J., Yapici M.K. Toward graphene textiles in wearable eye tracking systems for human–machine interaction. Beilstein J. Nanotechnol. 2021;12:180–189. doi: 10.3762/bjnano.12.14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Ozturk O., Yapici M.K. Surface electromyography with wearable graphene textiles. IEEE Sens. J. 2021;21:14397–14406. [Google Scholar]
- 43.Brisebois P., Siaj M. Harvesting graphene oxide–years 1859 to 2019: a review of its structure, synthesis, properties and exfoliation. J. Mater. Chem. C. 2020;8:1517–1547. [Google Scholar]
- 44.Guex L.G., Sacchi B., Peuvot K.F., Andersson R.L., Pourrahimi A.M., Ström V., Farris S., Olsson R.T. Experimental review: chemical reduction of graphene oxide (GO) to reduced graphene oxide (rGO) by aqueous chemistry. Nanoscale. 2017;9:9562–9571. doi: 10.1039/c7nr02943h. [DOI] [PubMed] [Google Scholar]
- 45.Moon I.K., Lee J., Ruoff R.S., Lee H. Reduced graphene oxide by chemical graphitization. Nat. Commun. 2010;1:1–6. doi: 10.1038/ncomms1067. [DOI] [PubMed] [Google Scholar]
- 46.Phiri J., Gane P., Maloney T.C. General overview of graphene: production, properties and application in polymer composites. Mater. Sci. Eng.: B. 2017;215:9–28. [Google Scholar]
- 47.Bonaccorso F., Lombardo A., Hasan T., Sun Z., Colombo L., Ferrari A.C. Production and processing of graphene and 2d crystals. Mater. Today. 2012;15:564–589. [Google Scholar]
- 48.Tuinstra F., Koenig J.L. Raman spectrum of graphite. J. Chem. Phys. 1970;53(3):1126–1130. [Google Scholar]
- 49.Stankovich S., Dikin D.A., Piner R.D., Kohlhaas K.A., Kleinhammes A., Jia Y., Wu Y., Nguyen S.T., Ruoff R.S. Synthesis of graphene-based nanosheets via chemical reduction of exfoliated graphite oxide. Carbon. 2007;45:1558–1565. [Google Scholar]
- 50.Bose S., Kuila T., Mishra A.K., Kim N.H., Lee J.H. Dual role of glycine as a chemical functionalizer and a reducing agent in the preparation of graphene: an environmentally friendly method. J. Mater. Chem. 2012;22:9696–9703. [Google Scholar]
- 51.Low F.W., Lai C.W., Abd Hamid S.B. Easy preparation of ultrathin reduced graphene oxide sheets at a high stirring speed. Ceram. Int. 2015;41:5798–5806. [Google Scholar]
- 52.Cui P., Lee J., Hwang E., Lee H. One-pot reduction of graphene oxide at subzero temperatures. Chem. Commun. 2011;47:12370–12372. doi: 10.1039/c1cc15569e. [DOI] [PubMed] [Google Scholar]
- 53.Wang J., Salihi E.C., Šiller L. Green reduction of graphene oxide using alanine. Mater. Sci. Eng.: C. 2017;72:1–6. doi: 10.1016/j.msec.2016.11.017. [DOI] [PubMed] [Google Scholar]
- 54.Grimnes S. Impedance measurement of individual skin surface electrodes. Med. Biol. Eng. Comput. 1983;21:750–755. doi: 10.1007/BF02464038. [DOI] [PubMed] [Google Scholar]
- 55.Bergey G.E., Squires R.D., Sipple W.C. Electrocardiogram recording with pasteless electrodes. IEEE Trans. Biomed. Eng. 1971;BME-18:206–211. [Google Scholar]
- 56.Yadav T., Mehra R. Denoising and SNR improvement of ECG signals using wavelet based techniques. 2016 2nd Int. Conf. Gener. Comput. Technol. (NGCT), IEEE. 2016:678–682. [Google Scholar]
- 57.Doheny E.P., Goulding C., Flood M.W., Mcmanus L., Lowery M.M. Feature-based evaluation of a wearable surface EMG sensor against laboratory standard EMG during force-varying and fatiguing contractions. IEEE Sens. J. 2019;20:2757–2765. [Google Scholar]
- 58.Acharya U.R., Joseph K.P., Kannathal N., Lim C.M., Suri J.S. Heart rate variability: a review. Med. Biol. Eng. Comput. 2006;44:1031–1051. doi: 10.1007/s11517-006-0119-0. [DOI] [PubMed] [Google Scholar]
- 59.Sztajzel J. Heart rate variability: a noninvasive electrocardiographic method to measure the autonomic nervous system. Swiss Med. Wkly. 2004;134:514–522. doi: 10.4414/smw.2004.10321. [DOI] [PubMed] [Google Scholar]
- 60.Hasty F., García G., Dávila H., Wittels S.H., Hendricks S., Chong S. Heart rate variability as a possible predictive marker for acute inflammatory response in COVID-19 patients. Mil. Med. 2021;186:e34–e38. doi: 10.1093/milmed/usaa405. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Miller J.G., Chahal R., Kirshenbaum J.S., Ho T.C., Gifuni A.J., Gotlib I.H. Heart rate variability moderates the effects of COVID-19-related stress and family adversity on emotional problems in adolescents: Testing models of differential susceptibility and diathesis stress. Dev. Psychopathol. 2020:1–12. doi: 10.1017/S095457942100033X. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Umetani K., Singer D.H., McCraty R., Atkinson M. Twenty-four hour time domain heart rate variability and heart rate: relations to age and gender over nine decades. J. Am. Coll. Cardiol. 1998;31:593–601. doi: 10.1016/s0735-1097(97)00554-8. [DOI] [PubMed] [Google Scholar]
- 63.Kang J.J., Luan T.H., Larkin H. Data processing of physiological sensor data and alarm determination utilising activity recognition. Int. J. Inf., Commun. Technol. Appl. 2016;2:108–131. [Google Scholar]
- 64.Nasario-Junior O., Benchimol-Barbosa P.R., Nadal J. Beat-to-beat T-peak T-end interval duration variability assessed by RR-interval histogram analysis in health sedentary and athlete. 2017 Comput. Cardiol. (CinC), IEEE. 2017:1–4. [Google Scholar]
- 65.Wang N.C., Maggioni A.P., Konstam M.A., Zannad F., Krasa H.B., Burnett JC Jr, Grinfeld L., Swedberg K., Udelson J.E., Cook T., Traver B., Zimmer C., Orlandi C., Gheorghiade M. I.Efficacy of Vasopressin Antagonism in Heart Failure Outcome Study With Tolvaptan (EVEREST), Clinical implications of QRS duration in patients hospitalized with worsening heart failure and reduced left ventricular ejection fraction. Jama. 2008;299:2656–2666. doi: 10.1001/jama.299.22.2656. [DOI] [PubMed] [Google Scholar]
- 66.G.D. Clifford, F. Azuaje, P. McSharry, Advanced methods and tools for ECG data analysis: Artech house Boston; 2006.
- 67.Zhang L., Timothy K.W., Vincent G.M., Lehmann M.H., Fox J., Giuli L.C., Shen J., Splawski I., Priori S.G., Compton S.J., Yanowitz F., Benhorin J., Moss A.J., Schwartz P.J., Robinson J.L., Wang Q., Zareba W., Keating M.T., Towbin J.A., Napolitano C., Medina A. Spectrum of ST-T–wave patterns and repolarization parameters in congenital long-QT syndrome: ECG findings identify genotypes. Circulation. 2000;102:2849–2855. doi: 10.1161/01.cir.102.23.2849. [DOI] [PubMed] [Google Scholar]
- 68.Baranchuk A., de Luna A.B. The P-wave morphology: what does it tell us? Herzschrittmacher + Elektro. 2015;26:192–199. doi: 10.1007/s00399-015-0385-3. [DOI] [PubMed] [Google Scholar]
- 69.Mousavi S., Afghah F., Khadem F., Acharya U.R. ECG Language processing (ELP): A new technique to analyze ECG signals. Comput. Methods Prog. Biomed. 2021;202 doi: 10.1016/j.cmpb.2021.105959. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Kleber A.G. ST-segment elevation in the electrocardiogram: a sign of myocardial ischemia. Cardiovasc. Res. 2000;45:111–118. doi: 10.1016/s0008-6363(99)00301-6. [DOI] [PubMed] [Google Scholar]
- 71.L. Burattini, Electrocardiographic T-wave Alternans detection and significance: University of Rochester; 1998.