Journal of Pain Research logoLink to Journal of Pain Research
. 2022 Nov 22;15:3625–3638. doi: 10.2147/JPR.S357416

Off-Label Magnetic Resonance Imaging (MRI) in Patients with Persistent Pain with Spinal Cord Stimulators: A Case Series

Thomas Ragukonis 1,
PMCID: PMC9700432  PMID: 36444170

Abstract

Objective

Advances in spinal cord stimulator (SCS) technology and increasing prevalence of magnetic resonance imaging (MRI) diagnostic testing require empirical evidence describing the presence of MRI-related SCS adverse events related to off-label use of imaging. MRI safety recommendations vary based on the type of stimulator used with scant availability regarding adverse events associated with off-label MRI use. The aim of this case series is to describe the type and frequency of adverse events associated with off-label MRI use in patients with implanted SCSs.

Materials and Methods

Convenient samples of records of patients who had SCS and subsequently underwent MRI were included. Descriptive data including patient demographics, stimulator information, and frequency of adverse events were described.

Results

Sixty-nine individuals with implanted SCSs were included. The total number of scans was 78. Sixty-two percent of the sample was female. Over 92% of the MRI scans were considered off-label and the overall adverse event rate for off-label use was 9.72%. No serious adverse events were reported. Seven clinical adverse events were reported, all of which were related to the spinal cord stimulation and resolved.

Conclusion

This case series demonstrates that individuals implanted with SCSs experienced no serious adverse events associated with off-label MRI use. While these results represent a convenient sample, they provide important preliminary information about using MRI when medically necessary for patients with older spinal cord stimulator models. Specifically, these data demonstrate that the rate of observed adverse events related to MRI was low and suggest that the benefits of acquiring these images for pain management may outweigh the risks of not acquiring MRI for appropriate pain management.

Keywords: spinal cord stimulation, chronic pain, magnetic resonance imaging

Introduction

Spinal cord stimulators (SCS) are implantable devices used to manage intractable and persistent pain and involve low-voltage current intended to block nociceptive input.1 Spinal cord stimulators have been shown to be safe and effective for persistent pain conditions, including spinal pain and complex regional pain syndrome.2–6

Safety concerns exist for patients with implanted SCSs and the use of magnetic resonance imaging (MRI) due to the magnetic field interfering with the device and its functionality. For example, one concern is that the magnetic field may alter the position of the device and/or accelerate the device into the bore of the magnet, causing severe damage to patient tissue. Also, the radiofrequency current may lead to heating of SCS, resulting in thermal and/or electrical burns to the patient and potential device malfunction. Another concern is that the presence of the device may affect the quality of the image(s) depending on the location of the image relative to the device. If the goal of the MRI is to evaluate areas in which to potentially intervene, image quality is of obvious importance.7,8 Although individual studies have reported safety outcomes, no consensus guidelines exist on how to safely perform imaging of patients with various implanted SCS. Rather, manufacturers provide device-specific guidelines, but these guidelines vary on whether or not MRI is contraindicated, and which parts of the body are considered safe to image. For example, some devices have been developed to allow for full body scans while others allow only for head scans. In some devices, MRI may be an absolute contraindication and in others, safety has yet to be determined. De Andres et al9 described the frequency and type of adverse events associated with a single manufacturer and reported that in 31 patients with SCSs who underwent MRI, seven reported events ranging from feeling stimulation during the MRI and increased thermal sensations at the lead sites. Others have reported on SCS-related adverse events but not necessarily related to MRI.10

On-label versus off-label MRI in patients with SCSs is a related concern and one that has not been widely explored. Patients post-laminectomy syndrome or with failed back surgery syndrome (FBSS) are the most common historical indication for SCS therapy and are among those that may require repeat imaging.11,12 Prior to 2012, all spinal cord stimulation systems were not considered “on label” for MRI use. Spinal cord stimulator models, including Precision, Protégé, and Eon Mini systems, were excluded for MRI. In 2012, the Medtronic system gained MRI conditionality. As newer generation implantable pulse generators (IPGs) have gained MRI conditionality, some of these are specific only for head or extremities. Others require specifications inherent to the leads, such as normalized impedance values or placement within a particular anatomical region, T7-T12.

Given the limited information available regarding MRI conditionality in patients with SCS, the purpose of this case series is to describe the frequency and type of long-term adverse events associated with off-label MRI use in patients with SCSs.

Methods

This case series included a convenience sample of patient records from an outpatient interventional pain clinic. The records represent with a permanent SCS who subsequently underwent MRI. The frequency and type of long-term adverse events reported by patients were evaluated following MRI. This study involved a retrospective review of patient records and thus a waiver of informed consent was obtained and the study was approved by the Advarra Institutional Review Board; the data accessed complied with relevant data protection and privacy regulations. The primary outcomes of interest were the frequency and type of adverse events following MRI. Adverse events were coded as serious or clinical, SCS or MRI-related, and as mild, moderate, or severe. Demographic information, duration of clinical symptoms, manufacturer and model type of SCS, on- versus off- label MRI use, and adverse event frequency were described. All data were analyzed in Microsoft Excel.

Results

Sixty-nine patient records were included in this study and Table 1 includes demographic information about the sample. Of these 69 records, 7 included more than one MRI, thus the total scan count for this sample was 78. Forty-two (62%) were female and the average age of the sample was 60.5 years. The duration of pain symptoms ranged from 5 to 47 years, with an average of 12 years. The primary medical diagnoses included failed back surgery syndrome (FBSS), failed neck surgery syndrome (FNSS), and chronic regional pain syndrome. The primary rationale to opt for off-label MRI versus other imaging methods is that MRI in this sample was the best available option to visualize the spinal cord and nerves; MRI may show spinal abnormalities, injuries, and disease that may not be seen with other methods.

Table 1.

Demographic Information

Patient # Age Sex Diagnosis Pain Duration (Years)
1 54 F Failed back surgery syndrome 11
2 62 M Failed back surgery syndrome 5
3 62 F Low Back pain; neck pain 19
4 49 F Failed back surgery syndrome 20
5 66 M Failed back surgery syndrome 10
6 53 M Failed back surgery syndrome 8
7* 44 F Failed back surgery syndrome; cervicalgia 20
7A* 45 F Post laminectomy syndrome 20
7B* 46 F Post laminectomy syndrome 21
7C* 47 F Failed back surgery syndrome; cervicalgia 22
8 65 M Post laminectomy syndrome 2
9* 59 M Failed back surgery syndrome 10
9A* 60 M Failed back surgery syndrome 11
10 50 F Chronic regional pain syndrome, bilateral 13
11 50 F Low Back pain 13
12 71 F Lumbalgia 7
13 57 F Failed back surgery syndrome 11
14 51 M Failed back surgery syndrome 15
15 71 F Low back pain 6
16 51 M Failed back surgery syndrome 7
17 59 F Multilevel disc disease 8
18 71 F Low back pain 10
19 83 F Multilevel disc disease 4
20 54 F Low back pain 13
21 79 M Low back pain 14
22 69 M Failed back surgery syndrome 15
23 51 M Chronic regional pain syndrome 5
24 42 F Failed back surgery syndrome 4
25 53 F Failed back surgery syndrome 36
26 77 F Failed back surgery syndrome 19
27 44 M Failed back surgery syndrome 11
28 68 F Multilevel disc disease 12
29 63 M Low back pain 5
30 65 F Failed back surgery syndrome 8
31 88 M Degenerative disc disease 5
32 52 M Failed back surgery syndrome 4.5
33 69 F Multilevel disc disease 9
34 63 F Failed back surgery syndrome 10
35 57 F Failed back surgery syndrome 5
36 60 F Failed back surgery syndrome 12
37* 80 M Facet arthroplasty 13
37A* 80 M Facet arthroplasty 13
38 66 F Multilevel disc herniation 10
39 55 F Lumbalgia 7
40* 79 F Multilevel disc herniation 11
40A* 80 F Multilevel disc herniation 12
41* 57 M Failed back surgery syndrome 6
41A* 59 M Failed back surgery syndrome 8
42* 38 M Failed back surgery syndrome 18
42A* 40 M Failed back surgery syndrome 20
43 47 F Failed back surgery syndrome 9
44 61 F Failed back surgery syndrome 28
45 77 F Multilevel disc disease 13
46 75 F Failed back surgery syndrome 8
47* 70 F Multilevel disc herniation 12
47A* 70 H Multilevel disc herniation 12
48 58 F Failed back surgery syndrome 8.75
49 65 M Failed back surgery syndrome 3
50 63 M Failed back surgery syndrome 3.25
51 67 M Failed back surgery syndrome; peripheral neuropathy 7
52 69 F Failed back surgery syndrome 9
53 39 M Failed back surgery syndrome 10
54 72 F Failed back surgery syndrome 10
55 51 F Failed back surgery syndrome 6
56 51 F Failed back surgery syndrome 13
57 59 F Failed back surgery syndrome 28
58 68 F Other intervertebral disc displacement, lumbar 47
59 66 F Other intervertebral disc displacement, lumbar 12
60 56 M Failed back surgery syndrome 20
61 53 M Failed back surgery syndrome 14
62 49 F Failed neck surgery syndrome 4.5
63 63 M Failed neck surgery syndrome 17
64 48 F Failed neck surgery syndrome 24
65 45 F Failed back surgery syndrome, chronic regional pain syndrome 18
66 54 M Failed back surgery syndrome 13
67 72 M Failed back surgery syndrome 7
68 49 F Failed neck surgery syndrome, failed back surgery syndrome 15
69 67 M Radiculopathy 8

Note: *Indicates the same individual who underwent more than one MRI.

The duration from implant to MRI and duration from MRI to long-term follow-up are in Table 2. The time from permanent generator implant to MRI ranged from less than 1 month to 87 months (mean 19.4 months, standard deviation 17.7 months). The time from MRI to long-term follow-up ranged from 2 months to 17 months (mean 10.4 months, standard deviation 3.2 months). All patients underwent MRI in a 1.5 Tesla scanner (Siemens Medical Solutions USA, Inc, Malvern, PA).

Table 2.

MRI and Follow-Up Information

Patient # Duration from IPG Implant to MRI (Months) Duration from MRI to Follow-up (Months) Clinical Event Severity of Clinical Event Related to SCS Related to MRI? Event Resolved?
1 9 13
2 7 11
3 4 11
4 25 9 Radicular chest wall or abdominal stimulation Mild No No Yes
5 35 15
6 4 8
7* 12 11 Undesirable changes in stimulation Mild Yes No Yes
7A* 1 1
7B* 11 13
7C* 19 13
8 19 13
9* 25 14
9A* 5 11
10 4 11
11 26 9 Undesirable changes in stimulation Mild Yes No Yes
12 22 12
13 5 7
14 87 11
15 13 8
16 39 4
17 28 12
18 4 13
19 3 16
20 3 13
21 14 12
22 49 14
23 5 9
24 5 11
25 0 2
26 26 13
27 8 11
28 3 11
29 7 4
30 1 11
31 9 11
32 1 10
33 7 11 Persistent pain or numbness at the electrode or IPG site Moderate Yes No Yes
34 17 8
35 9 10 Persistent pain or numbness at the electrode or IPG site Moderate Yes No Yes
36 12 11
37* 37 12
37A* 33 12
38 5 11
39 9 12
40* 34 21
40A* 55 6
41* 28 8
41A* 4 1
42* 33 12
42A* 60 5
43 17 10
44 30 13
45 47 13
46 10 11
47* 2 6
47A* 7 9
48 12 17
49 5 0
50 20 10
51 5 8 Undesirable changes in stimulation Mild Yes No Yes
52 46 9
53 17 10
54 34 11
55 30 11
56 30 12
57 14 12
58 34 7 Undesirable changes in stimulation Mild Yes No Yes
59 0 11
60 27 12
61 16 11
62 4 12
63 10 9
64 3 13
65 49 13
66 69 11
67 42 2
68 33 15
69 26 24

Note: *Indicates the same individual who underwent more than one MRI.

The manufacturers and IPG models are listed in Table 3. Based on the IPG model, lead types, locations, and configurations, the overwhelming majority of patients in this sample (94.1%) underwent off-label MRI use. The frequency of adverse events for patients who underwent off-label MRI use was 10.9%. The most common reported as undesirable changes in stimulation (4/7), followed by persistent pain or numbness (2/7), and radicular chest wall or abdominal stimulation (1/7). All seven events were considered non-serious and all were resolved. For all adverse events, a representative from the company was called for troubleshooting and resolved with re-programming. If adequate stimulation could not be achieved, patients were sent for additional imaging to examine the leads and IPG under fluoroscopy. After testing the functionality of equipment, devices were re-programmed until optimal results were achieved.

Table 3.

Spinal Cord Stimulator Information

Patient IPG Manufacturer Model IPG Placement Lead 1 Type Lead 1 Location Lead 2 Type Lead 1 Location Lead 1 Configuration Lead 2 Configuration MRI Location Off Label Use? Clinical Event?
1 Boston Scientific Precision Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar+ Cervical Yes
2 St. Jude Medical Protégé Upper buttock Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar Yes
3 St. Jude Medical Protégé Upper buttock Percutaneous C2-C3 Percutaneous C2-C3 1x8 1x8 Lumbar+ Cervical Yes
4 Boston Scientific Precision Upper buttock Percutaneous L2-L3 Percutaneous L2-L3 1x8 1x8 Lumbar+ Cervical Yes Yes
5 Boston Scientific Precision Upper buttock Percutaneous T7-T8 Percutaneous T7-T8 1x8 1x8 Lumbar Yes
6 St. Jude Medical Eon Mini Upper buttock Percutaneous T12-L1 Percutaneous T12-L1 1x8 1x8 Lumbar Yes
7* St. Jude Medical Protégé Upper buttock Percutaneous C2-C3 Perc paddle T7-T8 1x8 1x8 Lumbar+ Cervical Yes Yes
7A* St. Jude Medical Protégé Flank Percutaneous C2-C3 Percutaneous T7-T8 1x8 1x8 Lumbar Yes
7B* St. Jude Medical Protégé Flank Percutaneous C2-C3 Percutaneous T7-T8 1x8 1x8 Cervical Yes
7C* St. Jude Medical Protégé Upper buttock Percutaneous C2-C3 Perc paddle T7-T8 1x8 1x8 Lumbar+ cervical Yes
8 Boston Scientific Precision Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes
9* St. Jude Medical Eon Mini Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar+ Cervical Yes
9A* St. Jude Medical Proclaim Flank Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes
10 St. Jude Medical Protégé Upper buttock Percutaneous T12-L1 Percutaneous T12-L1 1x8 1x8 Lumbar Yes
11 St. Jude Medical Eon Mini Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes Yes
12 Boston Scientific Precision Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar+ Cervical Yes
13 St. Jude Medical Protégé Upper buttock Percutaneous T7-T9 Perc paddle T7-T9 1x8 1x8 Lumbar+ Cervical Yes
14 Boston Scientific Precision Upper buttock Percutaneous T9-T10 Percutaneous T9-T10 1x8 1x8 Lumbar Yes
15 St. Jude Medical Protégé Upper buttock Percutaneous T7-T9 Perc paddle T7-T9 1x8 1x8 Lumbar+ Cervical Yes
16 Boston Scientific Precision Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar+ Thoracic Yes
17 Boston Scientific Precision Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes
18 St. Jude Medical Protégé Upper buttock Percutaneous T12-L1 Perc paddle T12-L1 1x8 1x8 Lumbar Yes
19 St. Jude Medical Protégé Upper buttock Percutaneous T8/T9 Perc paddle T9/T10 1x8 1x8 Lumbar Yes
20 St. Jude Medical Protégé Upper buttock Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar+ Cervical Yes
21 St. Jude Medical Protégé Upper buttock Percutaneous T12-L1 Perc paddle T12-L1 1x8 1x8 Lumbar Yes
22 St. Jude Medical Protégé Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes
23 St. Jude Medical Eon Mini Upper buttock Percutaneous C1-C3 Percutaneous C1-C3 1x8 1x8 Cervical Yes
24 St. Jude Medical Protégé Upper buttock Percutaneous T7-T9 Perc paddle T7-T9 1x8 1x8 Lumbar Yes
25 St. Jude Medical Eon Mini Upper buttock Percutaneous T8-T10 Percutaneous T8-T10 1x8 1x8 Lumbar Yes
26 St. Jude Medical Protégé Upper buttock Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar Yes
27 St. Jude Medical Protégé Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes
28 St. Jude Medical Protégé Upper buttock Percutaneous T7-T8 Perc paddle T7-T8 1x8 1x8 Lumbar Yes
29 St. Jude Medical Protégé Upper buttock Percutaneous T8-T9 Perc paddle T8-T9 1x8 1x8 Lumbar Yes
30 St. Jude Medical Protégé Upper buttock Percutaneous T7-T8 Perc paddle T7-T8 1x8 1x8 Lumbar Yes
31 Boston Scientific Precision Upper buttock Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar Yes
32 St. Jude Medical Protégé Upper buttock Percutaneous T7-T9 Perc paddle T7-T9 Lumbar Yes
33 Boston Scientific Precision Upper buttock Percutaneous T7-T8 Percutaneous T7-T8 1x8 1x8 Lumbar Yes Yes
34 St. Jude Medical Eon Mini Upper buttock Percutaneous T9-T10 Perc paddle T9-T10 1x8 1x8 Lumbar Yes
35 St. Jude Medical Protégé MRI Upper buttock Percutaneous T7-T8 Percutaneous T7-T8 1x8 1x8 Lumbar Yes Yes
36 St. Jude Medical Proclaim 7 Right flank Percutaneous T7-T9 Perc paddle T7-T9 1x8 1x8 Lumbar Yes
37* St. Jude Medical Protégé MRI Right flank Percutaneous T7-T8 Perc paddle T7-T8 1x8 1x8 Cervical Yes
37A* St. Jude Medical Protégé MRI Right flank Percutaneous T7-T8 Perc paddle T7-T8 1x8 1x8 Cervical Yes
38 St. Jude Medical Protégé Lower flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar Yes
39 St. Jude Medical Protégé MRI Right flank Percutaneous T7-T9 Perc paddle T7-T9 1x8 1x8 Lumbar Yes
40* St. Jude Medical Eon Mini Right flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar Yes
40A* St. Jude Medical Eon Mini Right flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar+ cervical Yes
41* St. Jude Medical Proclaim 5 Right flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar  No
41A* St. Jude Medical Proclaim 5 Right flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar+ thoracic No
42* St. Jude Medical Eon Mini Right flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar Yes
42A* St. Jude Medical Eon Mini Right flank Percutaneous T7-T10 Percutaneous T7-T10 1x8 1x8 Lumbar+ thoracic Yes
43 St. Jude Medical Eon Mini Right flank Percutaneous T8-T9 Perc paddle T7-T8 1x8 1x8 Lumbar Yes
44 St. Jude Medical Protégé IPG Right flank Perc paddle T7-T9 1x8 1x8 Lumbar Yes
45 St. Jude Medical Eon Mini Right flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar Yes
46 Boston Scientific Precision Right flank Laminectomy T8-T9 Laminectomy T8-T9 1x8 1x8 Lumbar Yes
47* St. Jude Medical Proclaim 7 Right flank Percutaneous T7-T8 Percutaneous T7-T8 1x8 1x8 Cervical  No
47A* St. Jude Medical Proclaim 8 Right flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Lumbar+ thoracic No
48 St. Jude Medical Protégé MRI Right flank Percutaneous T7-T8 Perc paddle T7-T8 1x8 1x8 Lumbar Yes
49 St. Jude Medical Proclaim 7 Right flank Percutaneous C7-T1 Percutaneous C7-T1 1x8 1x8 Lumbar Yes
50 St. Jude Medical Proclaim 7 Right flank Percutaneous T7-T9 Percutaneous T7-T9 1x8 1x8 Right shoulder  No
51 St. Jude Medical Proclaim 7 Right flank Percutaneous T12-L1 Percutaneous T12-L1 1x8 1x8 Lumbar Yes Yes
52 St. Jude Medical Protégé IPG Right flank Percutaneous T7-T8 Perc paddle T7-T8 1x8 1x8 Cervical Yes
53 St. Jude Medical Proclaim Right flank Percutaneous T7-T9 Perc paddle T7-T9 1x8 1x8 Lumbar Yes
54 St Jude Medical Eon Mini Right flank Percutaneous T7-T8 Perc paddle T7-T8 1x8 1x8 Lumbar Yes
55 St. Jude Medical Protégé IPG Right flank Percutaneous T8-T9 Perc paddle T8-T9 1X8 1X8 Lumbar Yes
56 Boston Scientific Precision Right flank Percutaneous T8-T10 Percutaneous T8-T10 1x8 1x8 Lumbar Yes
57 St. Jude Medical Protégé IPG Right flank Percutaneous T7-T9 Perc paddle T7-T9 1x8 1x8 Lumbar Yes
58 St. Jude Medical Protégé IPG Right flank Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes Yes
59 St. Jude Medical Prodigy Right flank Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar, Thoracic, Cervical  No
60 St. Jude Medical Eon Mini Right flank Percutaneous T12-L1 Percutaneous T12-L1 1x8 1x8 Lumbar Yes
61 St. Jude Medical Protégé IPG Right flank Percutaneous T6-T7 Percutaneous T8-T9 1x8 1x8 Lumbar Yes
62 St. Jude Medical Protégé Right flank Percutaneous C2-C3 Percutaneous C2-C3 1x8 1x8 Right hip+ Lumbar Yes
63 St. Jude Medical Proclaim 7 Right flank Percutaneous C2-C3 Percutaneous C2-C3 1x8 1x8 Right knee Yes
64 St. Jude Medical Proclaim Right flank Percutaneous C2-C3 Percutaneous C2-C3 1x8 1x8 Lumbar Yes
65 St. Jude Medical Eon Mini Right flank Percutaneous T12-L1 Percutaneous T12-L1 1x8 1x8 Lumbar Yes
66 Boston Scientific Precision Upper buttock Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes
67 St. Jude Medical Protégé Upper buttock Percutaneous T8-T10 Percutaneous T8-T10 1x8 1x8 Lumbar Yes
68 Boston Scientific Precision Upper buttock Percutaneous L2-L3 Percutaneous L2-L3 1x8 1x8 Cervical Yes
69 St. Jude Medical Eon mini Right flank Percutaneous T8-T9 Percutaneous T8-T9 1x8 1x8 Lumbar Yes

Note: *Indicates the same individual who underwent more than one MRI.

Discussion

This case series indicates that off-label MRI use did not result in long-term serious adverse events. While these represent pilot data, this work is notable and relevant to clinical management of chronic pain for several reasons. These data add to existing work in this area regarding MRI safety in patients with SCS and have potentially important implications for both diagnostic radiology and interventional pain medicine. A recent case report from Dr Andres et al7 reported on the within-session adverse events associated with full-body MRI immediately following SCS implant. The patient in that case reported no serious or clinical events during the MRI. Second, the current case series expands on previous publications by examining off-label imaging and the frequency of long-term serious and clinical adverse events associated with this imaging. Previous recommendations for MRI in patients with SCS suggested that physicians should carefully select patients for MRI based on their potential to experience adverse events, including movement of the device, risk of thermal injury or burning, and alteration of the neurostimulation program.13 A small percentage of individuals from the current study experienced mild to moderate clinical events; however, all of the clinical events were related to the device rather than the MRI and the frequency rate of adverse events was similar to published trials of SCS long-term safety.14

Given the low rate of adverse events when MRI was used in an off-label manner, these results suggest that the benefits of obtaining MRI for the purposes of identifying structure(s) that may be contributing to a patient’s pain may outweigh the risks of not obtaining imaging in patients who have older SCS units. Future work may expand on the current study by exploring differences in adverse events in patients with on-label versus off-label MRI use. Although the vast majority of patients in the current study underwent off-label MRI, it is unclear if the frequency of adverse events may be different compared to individuals undergoing on-label MRI. These data would provide further information to aid clinical decision-making for diagnostic imaging in patients with persistent pain. Last, and as others have recommended,15,16 consensus guidelines for the use of MRI for patients with SCS should be developed and evaluated beyond individual manufacturer’s recommendations to provide in-depth information related to the safety parameters and utility of MRI in patients with chronic pain conditions.

Limitations

The current study included a convenience sample of patient records. Records were not included if follow-up data was not available (eg, the patient did not return to the clinic) or if demographic or other clinical information was not available in the patient record. The clinic from which these data were collected did not systematically assess (ie, at a specific time frame or manner) the presence of adverse events; thus, it is possible that the frequency of adverse events could be higher or lower based on the systematic evaluation of adverse events. Last, our follow-up time was relatively short at an average of 10.4 months. While we expect that adverse events associated with MRI would occur in the short-term (ie during the scan or shortly thereafter), it is possible that the MRI was associated with longer term effects that our follow-up period did not adequately capture.

Disclosure

The author reports no conflicts of interest in this work.

References

  • 1.Melzack R, Wall PD. Pain mechanisms: a new theory. Science. 1965;150(3699):971–979. doi: 10.1126/science.150.3699.971 [DOI] [PubMed] [Google Scholar]
  • 2.Deer TR, Krames E, Mekhail N, et al. The appropriate use of neurostimulation: new and evolving neurostimulation therapies and applicable treatment for chronic pain and selected disease states. Neuromodulation appropriateness consensus committee. Neuromodulation. 2014;17(6):599–615; discussion 615. doi: 10.1111/ner.12204 [DOI] [PubMed] [Google Scholar]
  • 3.Deer TR, Grider JS, Lamer TJ, et al. A systematic literature review of spine neurostimulation therapies for the treatment of pain. Pain Med. 2020;21(7):1421–1432. [DOI] [PubMed] [Google Scholar]
  • 4.Maheshwari A, Pope JE, Deer TR, Falowski S. Advanced methods of spinal stimulation in the treatment of chronic pain: pulse trains, waveforms, frequencies, targets, and feedback loops. Expert Rev Med Devices. 2019;16(2):95–106. doi: 10.1080/17434440.2019.1567325 [DOI] [PubMed] [Google Scholar]
  • 5.Hunter CW, Carlson J, Yang A, Deer T. Spinal cord stimulation for the treatment of failed neck surgery syndrome: outcome of a prospective case series. Neuromodulation. 2018;21(5):495–503. doi: 10.1111/ner.12769 [DOI] [PubMed] [Google Scholar]
  • 6.Lad SP, Babu R, Bagley JH, et al. Utilization of spinal cord stimulation in patients with failed back surgery syndrome. Spine. 2014;39(12):E719–727. doi: 10.1097/BRS.0000000000000320 [DOI] [PubMed] [Google Scholar]
  • 7.De Andres J, Martinez-Sanjuan V, Fabregat-Cid G, Asensio-Samper JM, Sanchis-Lopez N, Villanueva-Perez V. MRI-compatible spinal cord stimulator device and related changes in patient safety and imaging artifacts. Pain Med. 2014;15(10):1815–1819. doi: 10.1111/pme.12522 [DOI] [PubMed] [Google Scholar]
  • 8.Manfield J, Bartlett R, Safety PN. Utility of spinal magnetic resonance imaging in patients with high-frequency spinal cord stimulators: a prospective single-centre study. Stereotact Funct Neurosurg. 2019;97(4):272–277. doi: 10.1159/000503730 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.De Andres J, Valia JC, Cerda-Olmedo G, et al. Magnetic resonance imaging in patients with spinal neurostimulation systems. Anesthesiology. 2007;106(4):779–786. doi: 10.1097/01.anes.0000264776.17533.06 [DOI] [PubMed] [Google Scholar]
  • 10.Kapural L, Gupta M, Paicius R, et al. Treatment of chronic abdominal pain with 10-kHz spinal cord stimulation: safety and efficacy results from a 12-month prospective, multicenter, feasibility study. Clin Transl Gastroenterol. 2020;11(2):e00133. doi: 10.14309/ctg.0000000000000133 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Amirdelfan K, Webster L, Poree L, Sukul V, McRoberts P. Treatment options for failed back surgery syndrome patients with refractory chronic pain: an evidence based approach. Spine. 2017;42(Suppl 14):S41–S52. doi: 10.1097/BRS.0000000000002217 [DOI] [PubMed] [Google Scholar]
  • 12.Sitzman BT, Provenzano DA. Best practices in spinal cord stimulation. Spine. 2017;42(Suppl 14):S67–S71. doi: 10.1097/BRS.0000000000002220 [DOI] [PubMed] [Google Scholar]
  • 13.Liem LA, van Dongen VC. Magnetic resonance imaging and spinal cord stimulation systems. Pain. 1997;70(1):95–97. doi: 10.1016/S0304-3959(96)03298-8 [DOI] [PubMed] [Google Scholar]
  • 14.Mekhail N, Levy RM, Deer TR, et al. Long-term safety and efficacy of closed-loop spinal cord stimulation to treat chronic back and leg pain (Evoke): a double-blind, randomised, controlled trial. Lancet Neurol. 2020;19(2):123–134. doi: 10.1016/S1474-4422(19)30414-4 [DOI] [PubMed] [Google Scholar]
  • 15.Ghaly RF, Tverdohleb T, Candido KD, Knezevic NN. Do we need to establish guidelines for patients with neuromodulation implantable devices, including spinal cord stimulators undergoing nonspinal surgeries? Surg Neurol Int. 2016;7:18. doi: 10.4103/2152-7806.176373 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Harned ME, Gish B, Zuelzer A, Grider JS. Anesthetic considerations and perioperative management of spinal cord stimulators: literature review and initial recommendations. Pain Physician. 2017;20(4):319–329. doi: 10.36076/ppj.2017.329 [DOI] [PubMed] [Google Scholar]

Articles from Journal of Pain Research are provided here courtesy of Dove Press