Abstract
Objectives
To evaluate a compressed sensing artificial intelligence framework (CSAI) to accelerate MRI acquisition of the ankle.
Methods
Thirty patients were scanned at 3T. Axial T2-w, coronal T1-w, and coronal/sagittal intermediate-w scans with fat saturation were acquired using compressed sensing only (12:44 min, CS), CSAI with an acceleration factor of 4.6–5.3 (6:45 min, CSAI2x), and CSAI with an acceleration factor of 6.9–7.7 (4:46 min, CSAI3x). Moreover, a high-resolution axial T2-w scan was obtained using CSAI with a similar scan duration compared to CS. Depiction and presence of abnormalities were graded. Signal-to-noise and contrast-to-noise were calculated. Wilcoxon signed-rank test and Cohen’s kappa were used to compare CSAI with CS sequences.
Results
The correlation was perfect between CS and CSAI2x (κ = 1.0) and excellent for CS and CSAI3x (κ = 0.86–1.0). No significant differences were found for the depiction of structures between CS and CSAI2x and the same abnormalities were detected in both protocols. For CSAI3x the depiction was graded lower (p ≤ 0.001), though most abnormalities were also detected. For CSAI2x contrast-to-noise fluid/muscle was higher compared to CS (p ≤ 0.05), while no differences were found for other tissues. Signal-to-noise and contrast-to-noise were higher for CSAI3x compared to CS (p ≤ 0.05). The high - resolution axial T2-w sequence specifically improved the depiction of tendons and the tibial nerve (p ≤ 0.005).
Conclusions
Acquisition times can be reduced by 47% using CSAI compared to CS without decreasing diagnostic image quality. Reducing acquisition times by 63% is feasible but should be reserved for specific patients. The depiction of specific structures is improved using a high-resolution axial T2-w CSAI scan.
Key Points
• Prospective study showed that CSAI enables reduction in acquisition times by 47% without decreasing diagnostic image quality.
• Reducing acquisition times by 63% still produces images with an acceptable diagnostic accuracy but should be reserved for specific patients.
• CSAI may be implemented to scan at a higher resolution compared to standard CS images without increasing acquisition times.
Keywords: Ankle, MRI, Deep learning, Compressed Sense
Introduction
The ankle is one of the most complex joints with multiple oblique oriented ligaments and tendons [1–6]. In the United States, approximately 10,000 people suffer an ankle injury per day, the most frequent type of injury being sprains [7, 8]. Magnetic resonance (MR) imaging is one of the main imaging modalities for the assessment of musculoskeletal disorders on account of its high soft tissue contrast. However, the data acquisition process is inherently slow due to long encoding times [9]. This leads to long scan times and subsequently increased exam costs and reduced patient throughput. Moreover, the image quality is frequently compromised by motion artifacts, since remaining motionless for several minutes is a challenge even for healthy subjects [10].
Developments such as parallel imaging (PI) and, later, compressed sensing (CS) have accelerated MR image acquisition. CS reduces the number of acquired lines in k-space and restores the missing data through an iterative reconstruction algorithm [11, 12]. The combination of CS and PI was shown to reduce MR image acquisition times of the ankle by 20% without compromising diagnostic performance [13, 14]. More recently deep learning–based methods such as convolutional neural networks (CNNs) and generative adversarial networks (GAN) have shown promising results to accelerate the MR imaging data acquisition process [15]. These methods apply deep learning–based reconstruction schemes to create high-quality images from undersampled MR data [9, 16–21].
The aim of this study was to conduct a validation study to evaluate a compressed sensing artificial intelligence framework (CSAI) combining PI, CS, and deep learning–based artificial intelligence, for additional twofold and threefold acceleration of multi-contrast and multi-planar ankle MR imaging compared to conventional CS imaging and to compare the image quality and diagnostic performance between both techniques. In a secondary analysis, we compared the depiction of anatomical structures using high-resolution images reconstructed with CSAI.
Materials and methods
Subject selection
Thirty patients were prospectively enrolled in our study between January and June 2021 (15 female, age 19–84 years). Informed consent was obtained from all participants; the study was approved by the local institutional review board (42/21S). All patients were referred by the orthopedic department with various disorders including trauma, degeneration, and unclear pain. Individuals with conditions excluded by MR safety guidelines such as pacemakers, other implanted electronic devices or pregnancy, were not included.
Data acquisition
In this study, the utility of a novel CNN that integrates and enhances the conventional CS algorithm referred to as Adaptive-CS-Network as presented by Pezzotti et al [15] was investigated (CSAI). The Adaptive-CS-Network mimics the iterative shrinkage–thresholding algorithm (ISTA) approach presented by Zhang et al [22] and integrates multiscale sparsification in a problem-specific learnable manner and combines a CNN-based sparsifying approach with the image reconstruction approach of compressed sense, which ensures data consistency and incorporates domain-specific prior knowledge such as coil sensitivity distribution and location of the image background. In this regard the Adaptive-CS-Network basically replaces the wavelet transform by a CNN as sparsifying transform in the compressed sense algorithm, still keeping domain-specific knowledge and a term ensuring data consistency in place in the reconstruction process. In contrast to the network presented by Pezzotti et al [15], the Adaptive-CS-Network employed in this work was pre-trained on about 740,000 sparsifying MR images using both 1.5T and 3T images of various anatomies and contrasts. Furthermore, the algorithm was optimized to allow execution on standard reconstruction hardware.
All examinations were performed on a 3T MR scanner (Ingenia Elition; Philips Healthcare) using a 16-channel ankle coil. The ankles were fixated within the coil to reduce motion artifacts. The examination protocol included the following 2D sequences: axial T2-weighted turbo spin echo (TSE) sequences, coronal T1-weighted TSE sequences and coronal as well as sagittal intermediate-weighted (IM) TSE sequences with spectral presaturation with inversion recovery (SPIR) for fat saturation. All sequences were obtained three times: (1) with an acceleration factor of 2.5 reconstructed using CS, (2) with an acceleration factor of 4.6–5.3 reconstructed using CSAI (CSAI2x), and (3) with an acceleration factor of 6.9–7.7 reconstructed using CSAI (CSAI3x). In addition, we obtained a high-resolution axial T2-weighted sequence reconstructed using CSAI (CSAIHR), with a similar scan duration but increased in-plane resolution and smaller slice thickness compared to the axial T2-weighted reference sequence reconstructed using CS only. To assess signal-to-noise (SNR) and contrast-to-noise (CNR), coronal and axial MR sequences were performed twice in 8 patients on the same ankle without repositioning between scans. The total scan duration of the CS protocol was 12:44 min. The scan duration of the CSAI2x protocol was 6:45 min (47% shorter compared to CS). The scan duration of the CSAI3x protocol was 4:46 min (63% shorter compared to CS). The reconstruction times of a multi-slice 2D scan were in the order of about 2 min. Further details of the imaging protocols are shown in Table 1.
Table 1.
Pulse sequence | ax T2w TSE | ax T2w TSE HR | cor T1w TSE | cor IMw TSE | sag IMw TSE |
---|---|---|---|---|---|
TR [ms] | 2791 | 4733 | 705 | 2288 | 3631 |
TE [ms] | 80 | 80 | 18 | 50 | 50 |
Echo train length (ETL) | 15 | 15 | 6 | 15 | 8 |
Acquired resolution [mm] | 0.3 × 0.5 | 0.2 × 0.3 | 0.3 × 0.4 | 0.4 × 0.6 | 0.4 × 0.5 |
Slice thickness [mm] | 3 | 2 | 3 | 3 | 3 |
Field of view [mm] | 130 × 130 | 130 × 130 | 140 × 140 | 140 × 140 | 140 × 140 |
Number of slices | 31 | 46 | 23 | 23 | 24 |
Acceleration factor/scan time [min] | |||||
CS | 2.5/2:25 | - | 2.5/3:04 | 2.5/2:54 | 2.5/2:47 |
CSAI2x | 5.1/1:30 | 5/2:50 | 5.2/1:44 | 5.2/1:45 | 4.6/1:34 |
CSAI3x | 7.7/1:04 | - | 7.5/1:15 | 7.6/1:13 | 6.9/1:05 |
Quantitative image analysis
The subtraction method was used to determine SNR and CNR values for CS, CSAI2x, and CSAI3x, as described previously [13, 23, 24]. Sequences acquired twice in the same exam session were subtracted to create noise maps. Regions of interest (ROIs) were placed in the same location on three consecutive slices in each series and the noise maps. SNR was calculated for the following tissues: fluid, muscle, and tendon. CNR was calculated for fluid/muscle, fluid/tendon, and muscle/tendon.
Semi-quantitative image analysis
CS, CSAI2x, CSAI3x datasets, and CSAIHR images were analyzed in a randomized order with an interval of 4 weeks between readings to prevent recall bias. MR images were analyzed separately by two radiologists (25 and 7 years of experience; K.W. and S.C.F.), blinded to all clinical and other information. Readers graded depiction and presence of abnormalities of the tibiofibular syndesmosis, the medial and lateral ligament complex, the sinus tarsi ligaments, the extensor, flexor and peroneal tendons, the articular cartilage of the ankle joint, the bone, and the tibial nerve. Depiction of anatomical structures was graded using an ordinal 5-point Likert scale (1 = poor, 2 = below average, 3 = fair, 4 = good, 5 = excellent) evaluating the following criteria: partial volume effect, blurring, discrimination from adjacent structures, and signal homogeneity [23]. Abnormalities of the ligaments were graded in analogy to the Schweitzer classification system [25] as 0, no abnormality; 1, degenerative changes; 2, partial tear; 3, complete tear. Abnormalities of the tendons were graded as present/absent. Degenerative changes of the articular cartilage were graded as 0, no abnormality; 1, abnormal signal; 2, surface defect; 3, osteochondral defect. Bone abnormalities were graded as 0, no abnormality; 1, bone marrow edema pattern (BMEP); and 3, other abnormalities. Diagnostic confidence was recorded for all detected abnormalities with a 5-point ordinal scale as applied previously (1 = not detectable; 5 = 100% depicted, sharp) [23].
Statistical analysis
The statistical analysis was performed with SPSS, version 25.0 (IBM) using a two-sided 0.05 level of significance (S.C.F.). We also differentiated between p ≤ 0.05, p ≤ 0.005, and p ≤ 0.001, since previous studies have shown that using p values above 0.005 may lead to a lack of reproducibility of scientific findings [26]. CS was used as standard of reference for all statistical comparisons since this is currently the standard imaging technique in our clinical routine. The Wilcoxon signed-rank test was used to assess differences in image quality and detection and classification of abnormalities between CS and CSAI protocols. Interobserver correlation and intersequence correlation of detected abnormalities were determined using Cohen’s kappa.
Results
Quantitative image analysis
No significant differences were found in SNR between the CS and CSAI2x protocols. Compared to CS, SNR in the CSAI3x protocol was higher for fluid (94.2 ± 48.3 vs. 163.6 ± 77.7, p = 0.035) and muscle (127.8 ± 45.9 vs. 442.0 ± 17.3, p = 0.011); Fig. 1. Compared to CS, CNR in the CSAI2x protocol of fluid/muscle was higher (81.8 ± 37.6 vs. 241.1 ± 183.1, p = 0.035), while no significant differences were found for other investigated tissues. CNR was higher for all tissues in the CSAI3x protocol compared to CS only (fluid/muscle: 81.8 ± 37.6 vs. 278.4 ± 94.8, p = 0.011; fluid/tendon: 52.6 ± 37.7 vs. 120.6 ± 47.8, p = 0.011; muscle/tendon 92.4 ± 68.9 vs. 399.0 ± 49.2, p = 0.011; Fig. 2).
Depiction of anatomical structures
No significant differences were found for the depiction of all structures when comparing CS and CSAI2x. In CSAI3x the depiction was graded significantly lower (p ≤ 0.001) compared to CS for all anatomical structures apart from the tibionavicular ligament, which already had a low Likert score in CS and only marginally decreased in CSAI3x. Both readers found the high-resolution axial T2-weighted CSAIHR scan to improve the depiction of the peroneal tendon and the tibial bone structure (p ≤ 0.05, respectively), the anterior fibulotalar ligament and the extensor tendon (p ≤ 0.005, respectively), and the tibial nerve (p ≤ 0.001) (Fig. 3). In addition, reader 1 also graded the posterior tibiofibular ligament to be better depicted, reader 2 graded the flexor tendons, and the bone structure of the talus and fibula to be significantly better depicted (p ≤ 0.05, respectively). Detailed information on all graded anatomical structures is shown in Table 2. Magic angle artifacts were observed in all cases for CS, CSAI2x, CSAI3x, and CSAIHR. Moreover, no difference was noted for the markedness of the magic angle artifacts between different MR protocols.
Table 2.
Anatomical complex | Reader 1 | Reader 2 | ||||||
---|---|---|---|---|---|---|---|---|
CS | CSAI 2x | CSAI 3x | CSAI HR | CS | CSAI 2x | CSAI 3x | CSAI HR | |
Tibiofibular syndesmosis | ||||||||
Anterior tibiofibular ligament | 4.1 ± 0.5 | 4.0 ± 0.6 | 3.5 ± 0.6*** | 4.2 ± 0.5 | 4.3 ± 0.6 | 4.1 ± 0.7 | 3.4 ± 0.6*** | 4.2 ± 0.5 |
Posterior tibiofibular ligament | 4.6 ± 0.5 | 4.5 ± 0.5 | 3.9 ± 0.5*** | 4.8 ± 0.4* | 4.7 ± 0.5 | 4.5 ± 0.7 | 4.0 ± 0.7*** | 4.8 ± 0.4 |
Lateral collateral ligament complex | ||||||||
Anterior fibulotalar ligament | 4.3 ± 0.6 | 4.2 ± 0.7 | 3.5 ± 0.7*** | 4.6 ± 0.6** | 4.3 ± 0.6 | 4.3 ± 0.7 | 3.4 ± 0.7*** | 4.7 ± 0.5** |
Calcaneofibular ligament | 4.4 ± 0.5 | 4.3 ± 0.5 | 3.5 ± 0.7*** | 4.1 ± 0.7 | 4.1 ± 0.6 | 4.1 ± 0.6 | 3.4 ± 0.7*** | 4.2 ± 0.7 |
Posterior fibulotalar ligament | 4.5 ± 0.5 | 4.4 ± 0.6 | 3.7 ± 0.6*** | 4.4 ± 0.5 | 4.5 ± 0.6 | 4.4 ± 0.6 | 3.6 ± 0.6*** | 4.4 ± 0.5 |
Medial collateral ligament complex | ||||||||
Anterior tibiotalar ligament | 3.7 ± 0.9 | 3.6 ± 0.9 | 3.2 ± 1.0*** | 3.8 ± 0.6 | 3.7 ± 1.0 | 3.7 ± 1.0 | 3.3 ± 1.0*** | 4.0 ± 0.7 |
Tibionavicular ligament | 2.2 ± 1.2 | 2.1 ± 1.2 | 2.0 ± 1.2 | n/a | 2.1 ± 1.1 | 2.0 ± 1.1 | 2.0 ± 1.2 | n/a |
Tibiospring ligament | 4.5 ± 0.5 | 4.5 ± 0.5 | 3.7 ± 0.4*** | n/a | 4.3 ± 0.7 | 4.3 ± 0.7 | 3.8 ± 0.6*** | n/a |
Tibiocalcaneal ligament | 4.5 ± 0.6 | 4.6 ± 0.6 | 4.0 ± 0.4*** | n/a | 4.5 ± 0.6 | 4.6 ± 0.6 | 3.9 ± 0.5*** | n/a |
Posterior tibiotalar ligament | 4.3 ± 0.5 | 4.6 ± 0.5 | 3.7 ± 0.5*** | 4.6 ± 0.5 | 4.7 ± 0.5 | 4.5 ± 0.6 | 3.7 ± 0.6*** | 4.7 ± 0.5 |
Ligaments within the sinus tarsi | 4.3 ± 0.5 | 4.3 ± 0.5 | 3.8 ± 0.6*** | n/a | 4.4 ± 0.5 | 4.4 ± 0.6 | 3.8 ± 0.7*** | n/a |
Tendons | ||||||||
Extensor tendons | 4.5 ± 0.6 | 4.4 ± 0.6 | 3.8 ± 0.4*** | 5.0 ± 0.2** | 4.5 ± 0.6 | 4.4 ± 0.6 | 4.1 ± 0.2*** | 4.9 ± 0.3** |
Peroneal tendons | 4.7 ± 0.4 | 4.7 ± 0.4 | 3.9 ± 0.5*** | 4.9 ± 0.3* | 4.7 ± 0.5 | 4.7 ± 0.5 | 4.0 ± 0.5*** | 4.9 ± 0.3* |
Flexor tendons | 4.9 ± 0.3 | 4.9 ± 0.3 | 4.0 ± 0.4*** | 5.0 ± 0.0 | 4.9 ± 0.3 | 4.8 ± 0.4 | 4.1 ± 0.2*** | 5.0 ± 0.0* |
Cartilage | ||||||||
Fibulotalar cartilage | 3.3 ± 0.6 | 3.0 ± 0.8 | 2.4 ± 0.7*** | 3.4 ± 0.8 | 3.3 ± 0.6 | 3.2 ± 0.8 | 2.5 ± 0.6*** | 3.5 ± 0.9 |
Tibiotalar cartilage | 3.3 ± 0.5 | 3.2 ± 0.6 | 2.3 ± 0.5*** | n/a | 3.3 ± 0.6 | 3.3 ± 0.7 | 2.4 ± 0.6*** | n/a |
Bone | ||||||||
Talus | 4.9 ± 0.3 | 4.8 ± 0.4 | 3.5 ± 0.6*** | 5.0 ± 0.2 | 4.9 ± 0.3 | 4.8 ± 0.4 | 3.5 ± 0.6*** | 5.0 ± 0.0* |
Fibula | 4.9 ± 0.3 | 4.9 ± 0.3 | 3.7 ± 0.6*** | 5.0 ± 0.0 | 4.8 ± 0.4 | 4.8 ± 0.4 | 3.8 ± 0.6*** | 5.0 ± 0.0* |
Tibia | 4.9 ± 0.3 | 4.8 ± 0.4 | 3.7 ± 0.6*** | 5.0 ± 0.0* | 4.8 ± 0.4 | 4.7 ± 0.5 | 3.8 ± 0.6*** | 5.0 ± 0.0* |
Tibial Nerve | 4.0 ± 0.5 | 4.0 ± 0.5 | 3.2 ± 0.7*** | 4.8 ± 0.4*** | 4.0 ± 0.5 | 4.0 ± 0.6 | 3.2 ± 0.7*** | 4.9 ± 0.3*** |
Data are presented as means ± standard deviations
5-point Likert scale (5 = best; 1 = worst)
*p ≤ 0.05 using CS as standard of reference
**p ≤ 0.005 using CS as standard of reference
***p ≤ 0.001 using CS as standard of reference
Assessment of abnormalities
Details of the assessment of abnormalities and the diagnostic confidence are given in Table 3. All abnormalities detected on images reconstructed with CS were also detected on CSAI2x images, and there was no significant difference in the diagnostic confidence recorded for all detected abnormalities between both protocols (Fig. 4). Moreover, most abnormalities detected on images obtained with CS were also detected on CSAI3x images (n = 47/49, reader 1; n = 44/45, reader 2). Only one longitudinal split tear of the peroneus brevis tendon and one bone abnormality (bone marrow edema of the talus, < 3 mm diameter) recorded on images acquired using CS were not seen in CSAI3x images, due to increased blurring of the anatomical structures. Both readers recorded a significantly lower diagnostic confidence for abnormalities of the lateral ligament complex and the bone on CSAI3x compared to CS images (p ≤ 0.005). Reader 1 also recorded a lower diagnostic confidence for abnormalities of the cartilage, while reader 2 recorded a lower diagnostic confidence for tendon abnormalities on CSAI3x compared to CS images (p ≤ 0.05). No new, artificial abnormalities were created by CSAI reconstructions.
Table 3.
Anatomical complex | Grade | CS | CSAI 2x | CSAI 3x | |||
---|---|---|---|---|---|---|---|
n | Rating | n | Rating | n | Rating | ||
Reader I | |||||||
Tibiofibular syndesmosis | I | ||||||
II | 1 | 1 | 1 | ||||
III | |||||||
Total | 1 | 4.0 ± 0.0 | 1 | 4.0 ± 0.0 | 1 | 4.0 ± 0.0 | |
Lateral collateral ligament complex | I | 5 | 5 | 5 | |||
II | 5 | 5 | 6 | ||||
III | 5 | 5 | 4 | ||||
Total | 15 | 4.5 ± 0.5 | 15 | 4.4 ± 0.5 | 15 | 3.9 ± 0.5** | |
Medial collateral ligament complex | I | 2 | 2 | 2 | |||
II | |||||||
III | |||||||
Total | 2 | 4.0 ± 0.0 | 2 | 4.0 ± 0.0 | 2 | 3.0 ± 0.0 | |
Tendons | 13 | 4.6 ± 0.5 | 13 | 4.5 ± 0.5 | 12 | 4.1 ± 0.5 | |
Cartilage | I | ||||||
II | |||||||
III | 4 | 4 | 4 | ||||
Total | 4 | 5.0 ± 0.0 | 4 | 4.8 ± 0.5 | 4 | 4.0 ± 0.0* | |
Bone | I | 7 | 7 | 6 | |||
II | 7 | 7 | 7 | ||||
Total | 14 | 5.0 ± 0.0 | 14 | 5.0 ± 0.0 | 13 | 4.3 ± 0.5** | |
Reader II | |||||||
Tibiofibular syndesmosis | I | ||||||
II | 1 | 1 | 1 | ||||
III | |||||||
Total | 1 | 4.0 ± 0.0 | 1 | 4.0 ± 0.0 | 1 | 4.0 ± 0.0 | |
Lateral collateral ligament complex | I | 4 | 4 | 4 | |||
II | 5 | 5 | 6 | ||||
III | 5 | 5 | 4 | ||||
Total | 14 | 4.3 ± 0.5 | 14 | 4.2 ± 0.6 | 14 | 3.8 ± 0.4** | |
Medial collateral ligament complex | I | 1 | 1 | 1 | |||
II | |||||||
III | |||||||
Total | 1 | 4.0 ± 0.0 | 1 | 4.0 ± 0.0 | 1 | 3.0 ± 0.0 | |
Tendons | 11 | 4.5 ± 0.5 | 11 | 4.3 ± 0.5 | 11 | 3.7 ± 0.5* | |
Cartilage | I | ||||||
II | 1 | ||||||
III | 4 | 4 | 3 | ||||
Total | 4 | 4.5 ± 0.6 | 4 | 4.5 ± 0.6 | 4 | 4.0 ± 0.0 | |
Bone | I | 7 | 7 | 6 | |||
II | 7 | 7 | 7 | ||||
Total | 14 | 5.0 ± 0.0 | 14 | 5.0 ± 0.0 | 13 | 4.2 ± 0.4** |
Data are presented as means ± standard deviations
5-point Likert scale (5 = best; 1 = worst)
*p ≤ 0.05 using CS as standard of reference
**p ≤ 0.005 using CS as standard of reference
Interreader correlation was excellent for all criteria with κ = 0.86–1.0. Intersequence correlation was perfect for CSAI2x and CS for both readers with κ = 1.0, respectively. Intersequence correlation was excellent for CSAI3x and CS for both readers with κ = 0.93–1.0 for reader 1 and κ = 0.86–1.0 for reader 2, respectively (Fig. 5).
Discussion
In this study, we implemented an artificial intelligence framework for additional twofold and threefold acceleration of ankle MR imaging compared to conventional CS. The correlation was perfect for CS and CSAI2x and excellent for CS and CSAI3x. No significant differences were found for the depiction of anatomical structures between CS and CSAI2x and the same abnormalities were detected in both sequences. For CSAI3x the depiction was graded significantly lower compared to CS; however, most abnormalities were also detected. The high-resolution axial T2-weighted sequence was found to specifically improve the depiction of tendons, bone and the tibial nerve, with only minimally increased scan time compared to the axial T2-weighted CS reference scan.
Previous studies have used deep learning to accelerate MR imaging. One method is to create high-resolution from low-resolution data [27–29]. Chaudhari et al used deep learning CNNs to generate thin-slice knee MR images from thicker input slices [27]. Moreover, different deep learning methods have been applied to reconstruct MR images from undersampled MR data [17, 30]. Akçakaya et al used a k-space-based deep learning technique to estimate missing k-space lines from acquired k-space data [17]. Liu et al used a generative adversarial network (GAN) to enforce data consistency for a robust reconstruction of accelerated MR images [30]. In a similar approach to our study, Hammernik et al used an image-space-based technique for reconstruction of accelerated MR images with a so-called variational network [16]. The overall image quality of the variational network reconstructions with an acceleration factor of 4 was graded equal or better compared to conventional PI compressed sensing images, though anatomical structures and pathologies were not separately evaluated [16]. In accordance, we found no significant differences for the depiction of anatomical structures between CS and CSAI2x with an acceleration factor of 4.6–5.3 equating to approximately half the acquisition time of CS only. Moreover, the same abnormalities were detected in both sequences with no significant difference in the recorded diagnostic confidence. These findings suggest that CSAI2x protocols could be implemented in clinical practice without decreasing diagnostic image quality, though further validation studies are warranted to confirm these results.
In addition, we also evaluated the diagnostic image quality for CSAI3x with a threefold acceleration of acquisition times compared to the CS protocol. The depiction and diagnostic confidence for the detection of abnormalities was graded lower for most structures compared to CS. The reduction in diagnostic image quality was caused by increased image blurring, resulting in a reduced sensitivity in the detection of abnormalities. No spurious artifacts, which overlaid or mimicked a pathology, have been observed. However, it should be noted that only one peroneal tendon disorder and one bone abnormality of the talus recorded in CS were not seen in CSAI3x, while all other abnormalities were detected. These findings suggest that CSAI3x could be used in specific cases when the patient is in significant pain or only able to remain motionless for a short time.
Measured values of SNR were higher in CSAI3x, but not in CSAI2x compared to CS. For CSAI3x the higher SNR values were offset by image blurring and a related, significantly worse depiction of anatomical structures and abnormalities. This indicates that SNR and CNR values do not represent the total perceived image quality, when comparing iterative image reconstructions and image denoising. While image details and abnormalities were depicted worse in CSAI3x, no artificial details or abnormalities were observed.
Despite promising results, some limitations are pertinent to this study that should be addressed in further studies. No external standard of reference, such as arthroscopy, was available to serve as ground truth for detected pathologies of the ankle. Therefore, it was not possible to verify the diagnosed pathologies using an external standard of reference. Moreover, our series included only 30 subjects with a limited number of different pathologies and, thus, further studies including more patients are warranted to confirm these results. The aim of this study was to conduct a validation study on an internal data set. However, further studies are warranted to validate these results. Next steps include temporal validation (evaluation on a second data set from the same center) and external validation (evaluation on data from one or more other centers) [31]. To implement CSAI in the clinical workflow, it may also be beneficial to initially introduce these imaging sequences in addition to standard routine imaging protocols. This approach would allow for radiologists to develop a better understanding of CSAI sequences and when to best implement them.
In conclusion, we conducted a validation study from a single institution to evaluate a compressed sensing artificial intelligence framework and found that acquisition times can be reduced by 47% using CSAI without decreasing diagnostic image quality. Reducing acquisition times by 63% still produces images with an acceptable diagnostic accuracy but should be reserved for specific cases when patients are in significant pain or only able to remain motionless for a short time. This study is the foundation for further validation studies.
Abbreviations
- BMEP
Bone marrow edema pattern
- CNN
Convolutional neural network
- CNR
Contrast-to-noise ratio
- CS
Compressed sensing
- CSAI
Compressed sensing artificial intelligence
- CSAI2x
CSAI with an acceleration factor of 4.6–5.3
- CSAI3x
CSAI with an acceleration factor of 6.9–7.7
- CSAIHR
CSAI high-resolution axial T2-weighted sequence
- ETL
Echo train length
- GAN
Generative adversarial network
- IM
Intermediate-weighted
- ISTA
Iterative shrinkage–thresholding algorithm
- MR
Magnetic resonance
- PI
Parallel imaging
- ROI
Region of interest
- SNR
Signal-to-noise-ratio
- SPIR
Spectral presaturation with inversion recovery
- TSE
Turbo-spin-echo
Funding
Open Access funding enabled and organized by Projekt DEAL.
Declarations
Guarantor
The scientific guarantor of this publication is Klaus Woertler.
Conflict of interest
Kilian Weiss and Johannes M. Peeters are employees of Philips Healthcare.
Statistics and biometry
One of the authors has significant statistical expertise: Sarah C. Foreman.
Informed consent
Written informed consent was obtained from all subjects (patients) in this study.
Ethical approval
Institutional Review Board approval was obtained.
Methodology
• prospective
• cross sectional study
• performed at one institution
Footnotes
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Rosenberg ZS, Beltran J, Bencardino JT (2000) From the RSNA Refresher Courses. Radiological Society of North America. MR imaging of the ankle and foot. Radiographics 20 Spec No:S153-179 [DOI] [PubMed]
- 2.Gehrmann RM, Rajan S, Patel DV, Bibbo C. Athletes' ankle injuries: diagnosis and management. Am J Orthop (Belle Mead NJ) 2005;34:551–561. [PubMed] [Google Scholar]
- 3.Mengiardi B, Pfirrmann CW, Vienne P, Hodler J, Zanetti M. Medial collateral ligament complex of the ankle: MR appearance in asymptomatic subjects. Radiology. 2007;242:817–824. doi: 10.1148/radiol.2423060055. [DOI] [PubMed] [Google Scholar]
- 4.Alves T, Dong Q, Jacobson J, Yablon C, Gandikota G. Normal and injured ankle ligaments on ultrasonography with magnetic resonance imaging correlation. J Ultrasound Med. 2019;38:513–528. doi: 10.1002/jum.14716. [DOI] [PubMed] [Google Scholar]
- 5.Salat P, Le V, Veljkovic A, Cresswell ME. Imaging in foot and ankle instability. Foot Ankle Clin. 2018;23(499-522):e428. doi: 10.1016/j.fcl.2018.07.011. [DOI] [PubMed] [Google Scholar]
- 6.Vega J, Malagelada F, Manzanares Cespedes MC, Dalmau-Pastor M. The lateral fibulotalocalcaneal ligament complex: an ankle stabilizing isometric structure. Knee Surg Sports Traumatol Arthrosc. 2020;28:8–17. doi: 10.1007/s00167-018-5188-8. [DOI] [PubMed] [Google Scholar]
- 7.Holmer P, Sondergaard L, Konradsen L, Nielsen PT, Jorgensen LN. Epidemiology of sprains in the lateral ankle and foot. Foot Ankle Int. 1994;15:72–74. doi: 10.1177/107110079401500204. [DOI] [PubMed] [Google Scholar]
- 8.Herzog MM, Kerr ZY, Marshall SW, Wikstrom EA. Epidemiology of ankle sprains and chronic ankle instability. J Athl Train. 2019;54:603–610. doi: 10.4085/1062-6050-447-17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Johnson PM, Recht MP, Knoll F. Improving the speed of MRI with artificial intelligence. Semin Musculoskelet Radiol. 2020;24:12–20. doi: 10.1055/s-0039-3400265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Zaitsev M, Maclaren J, Herbst M. Motion artifacts in MRI: a complex problem with many partial solutions. J Magn Reson Imaging. 2015;42:887–901. doi: 10.1002/jmri.24850. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Geethanath S, Reddy R, Konar AS, et al. Compressed sensing MRI: a review. Crit Rev Biomed Eng. 2013;41:183–204. doi: 10.1615/CritRevBiomedEng.2014008058. [DOI] [PubMed] [Google Scholar]
- 12.Feng L, Benkert T, Block KT, Sodickson DK, Otazo R, Chandarana H. Compressed sensing for body MRI. J Magn Reson Imaging. 2017;45:966–987. doi: 10.1002/jmri.25547. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Gersing AS, Bodden J, Neumann J, et al. Accelerating anatomical 2D turbo spin echo imaging of the ankle using compressed sensing. Eur J Radiol. 2019;118:277–284. doi: 10.1016/j.ejrad.2019.06.006. [DOI] [PubMed] [Google Scholar]
- 14.Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952–962. doi: 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S. [DOI] [PubMed] [Google Scholar]
- 15.Pezzotti N, et al. An adaptive intelligence algorithm for undersampled knee MRI reconstruction. IEEE Access. 2020;8:204825–204838. doi: 10.1109/ACCESS.2020.3034287. [DOI] [Google Scholar]
- 16.Hammernik K, Klatzer T, Kobler E, et al. Learning a variational network for reconstruction of accelerated MRI data. Magn Reson Med. 2018;79:3055–3071. doi: 10.1002/mrm.26977. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Akcakaya M, Moeller S, Weingartner S, Ugurbil K. Scan-specific robust artificial-neural-networks for k-space interpolation (RAKI) reconstruction: database-free deep learning for fast imaging. Magn Reson Med. 2019;81:439–453. doi: 10.1002/mrm.27420. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Arshad M, Qureshi M, Inam O, Omer H. Transfer learning in deep neural network based under-sampled MR image reconstruction. Magn Reson Imaging. 2021;76:96–107. doi: 10.1016/j.mri.2020.09.018. [DOI] [PubMed] [Google Scholar]
- 19.Zhao D, Huang Y, Zhao F, Qin B, Zheng J. Reference-driven undersampled MR image reconstruction using wavelet sparsity-constrained deep image prior. Comput Math Methods Med. 2021;2021:8865582. doi: 10.1155/2021/8865582. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Herrmann J, Koerzdoerfer G, Nickel D et al (2021) Feasibility and implementation of a deep learning MR reconstruction for TSE sequences in musculoskeletal imaging. Diagnostics (Basel) 11 [DOI] [PMC free article] [PubMed]
- 21.Hahn S, Yi J, Lee HJ et al (2021) Image quality and diagnostic performance of accelerated shoulder MRI with deep learning-based reconstruction. AJR Am J Roentgenol. 10.2214/AJR.21.26577 [DOI] [PubMed]
- 22.Zhang J, Ghanem B ( 2018) ISTA-Net: Interpretable Optimization-Inspired Deep Network for Image Compressive Sensing,. EEE/CVF Conference on Computer Vision and Pattern Recognition. 10.1109:1828-1837
- 23.Notohamiprodjo M, Kuschel B, Horng A, et al. 3D-MRI of the ankle with optimized 3D-SPACE. Invest Radiol. 2012;47:231–239. doi: 10.1097/RLI.0b013e31823d7946. [DOI] [PubMed] [Google Scholar]
- 24.Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging. 2007;26:375–385. doi: 10.1002/jmri.20969. [DOI] [PubMed] [Google Scholar]
- 25.Schweitzer ME, Tran D, Deely DM, Hume EL. Medial collateral ligament injuries: evaluation of multiple signs, prevalence and location of associated bone bruises, and assessment with MR imaging. Radiology. 1995;194:825–829. doi: 10.1148/radiology.194.3.7862987. [DOI] [PubMed] [Google Scholar]
- 26.Johnson VE. Revised standards for statistical evidence. Proc Natl Acad Sci U S A. 2013;110:19313–19317. doi: 10.1073/pnas.1313476110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Chaudhari AS, Fang Z, Kogan F, et al. Super-resolution musculoskeletal MRI using deep learning. Magn Reson Med. 2018;80:2139–2154. doi: 10.1002/mrm.27178. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Koonjoo N, Zhu B, Bagnall GC, Bhutto D, Rosen MS. Boosting the signal-to-noise of low-field MRI with deep learning image reconstruction. Sci Rep. 2021;11:8248. doi: 10.1038/s41598-021-87482-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Koch KM, Sherafati M, Arpinar VE, et al. Analysis and evaluation of a deep learning reconstruction approach with denoising for orthopedic MRI. Radiol Artif Intell. 2021;3:e200278. doi: 10.1148/ryai.2021200278. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Liu F, Samsonov A, Chen L, Kijowski R, Feng L. SANTIS: Sampling-Augmented Neural neTwork with incoherent structure for MR image reconstruction. Magn Reson Med. 2019;82:1890–1904. doi: 10.1002/mrm.27827. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Altman DG, Royston P. What do we mean by validating a prognostic model? Stat Med. 2000;19:453–473. doi: 10.1002/(SICI)1097-0258(20000229)19:4<453::AID-SIM350>3.0.CO;2-5. [DOI] [PubMed] [Google Scholar]