Introduction
Bioresorbable vascular scaffolds (BVS)1,2,3 were designed to be an effective substitute to conventional metallic stents and overcome limitations including positive vessel remodelling, stent fracture, neoatherosclerosis, being a perpetual nidus for stent thrombosis and inhibition of physiological vasodilation4,5. The most widely studied polymeric devices comprise poly-L-lactide (PLLA) struts coated with poly-D, L-lactide (PDLLA) and an antiproliferative agent4,5. Although late outcomes with the Absorb™️ bioresorbable vascular scaffold (Abbott Vascular, Santa Clara, CA, USA) are disappointing6, this only confirms the need to gain a better understanding of the late behaviour of these scaffolds. The literature so far suggests that there is some evidence supporting their use in venous grafts7,8,9,10. This case description illustrates the unique findings of optical coherence tomography (OCT) imaging of a BVS in a vein graft at 40 months post implantation.
Methods
A 77-year-old lady had previously undergone two bypass operations, the latter following recurrent restenoses of conventional metallic stents. Approximately seven years post second bypass (April 2013), she had symptoms (angina) with inducible ischaemia on a stress echocardiogram. Angiography found high-grade proximal disease in a saphenous venous graft to the left anterior descending artery. The lesion was adequately prepared using a 3.0 mm predilatation balloon (Tazuna™ [Terumo Corp., Tokyo, Japan] –a semi-compliant percutaneous transluminal coronary angioplasty [PTCA] balloon catheter). A 3.5×28 mm Absorb BVS was deployed at 14 atmospheres and post-dilation was performed with a 3.75 mm non-compliant balloon (Hiryu™ [Terumo Corp.] –a non-compliant PTCA balloon catheter) to 21 atmospheres. A distal protection device was utilised (3.0 mm SpiderFX™️ [Medtronic, Minneapolis, MN, USA] embolic protection device) and the patient was placed on dual antiplatelet therapy for six months (aspirin 100 mg daily and clopidogrel 75 mg daily).
Results
Approximately 40 months post implant, she again developed symptoms (angina); an exercise test was abnormal and repeat coronary angiography was performed. This demonstrated new flow-limiting disease in a small to medium-sized native marginal circumflex artery, successfully treated with plain balloon angioplasty using a 2.0 mm predilatation balloon (Tazuna). Given concerns as to whether the diseased circumflex artery explained her presentation and the fact that the proximal segment of the vein graft to the left anterior descending artery appeared to have reduced volume (Figure 1), an OCT study (Moving image 1, Figure 2) was performed to assess the BVS in the vein graft to the left anterior descending artery. This demonstrated good apposition of the scaffold, tissue coverage with a neointimal depth of 0.48 mm and five uncovered struts. There were several struts manifesting various stages of PLLA degradation ranging from early cellular infiltration of dissolving struts (very high peak intensity at the core with low median intensity) to late degradation of the strut (moderately high peak intensity with a high median intensity). No intervention was performed on the vein graft. Post procedure, she was placed on dual antiplatelet therapy for six months (aspirin 100 mg daily and clopidogrel 75 mg daily) and has remained well since.
Figure 1. Saphenous vein graft to the left anterior descending artery pre implantation, immediately post implantation of the BVS and at 40 months post implantation of the BVS.

BVS: bioresorbable vascular scaffold
Figure 2. Optical coherence tomography image of the BVS.

Discussion
In September 2017, Abbott Vascular ceased supplying BVS, citing low demand but more likely due to disappointing clinical results. They also signalled their intention to release a newer improved device for future clinical trials. Accepting that initial poor implantation technique and inappropriate vessel selection played a role, the four-year ABSORB II data6 showed a statistically higher rate of target lesion failure (TLF) (11.1% vs. 5.6%). There were no cases of stent/scaffold thrombosis between three and four years and some see this as the potential start of scaffold benefit. Data do support a “learning curve”: the three-year data of ABSORB III11 showed only a trend towards a higher rate of TLF in the ABSORB arm (13.4% vs. 10.4%) but with a persistently higher rate of device thrombosis (2.3% vs. 0.7%), whilst the 30-day ABSORB IV (using an optimal implantation technique) reported reduced early events (30-day scaffold thrombosis reduced from 1.1% in ABSORB III to 0.4% in ABSORB IV at 30 days) to the extent that ABSORB IV shows non-inferiority at 30 days. Given that a “learning curve” probably exists, cases such as this and any knowledge gained remain clinically relevant.
The development of a “vascular restoration” strategy in the form of BVS has been considered by many as the “fourth revolution” in the percutaneous management of coronary artery disease12,13. So far, seven trials have been published comparing everolimus-eluting bioresorbable scaffolds and everolimus-eluting metallic stents14,15. The benefit of bioresorbable stents is thought to pay off in the long term: at around twelve months there is a loss of mechanical support that allows conditioning of tissue to enhance healing16, and at three years1 full reabsorption is noted, that assists in physiological function and neo-media formation in coronary arteries. Based on trial evidence at 60 months17, the BVS struts disappear completely.
Recently, Nakatani and colleagues18 performed an intravascular ultrasound echogenicity analysis and light intensity analysis on OCT in a porcine model with a follow-up duration of 48 months. There was good correlation observed for the strut depolymerisation process and integration process post comprehensive bioresorption. In addition, late luminal enlargement was observed at three to four years and was associated with strut integration.
Some bioresorbable stents are still commercially available; however, we are still awaiting the results of ABSORB IV, due for release in the next few years. Newer-generation stents such as the Magmaris™️ (Biotronik, Bülach, Switzerland) are showing promising results after publication of BIOSOLVE II and III19, in which improved stent and study design adds strength to the argument that BVS may still play a role in future percutaneous interventions.
Limitations
Little is known about the role of bioresorbable vascular scaffolds in venous grafts. The publications7,8,9,10,20 so far in this genre are summarised in Table 1. Indications for use in venous grafts are beyond the scope of this report, and the stent inserted was done so prior to release of ABSORB II and III. Our rationale for implantation in this high-risk patient was that she had incessant restenosis of metallic stents and was willing to allow sequential non-invasive imaging with computed tomography.
Table 1. Summary of bioresorbable vascular scaffolds in saphenous venous grafts.
| Author | Country | Cohort | Gender | Age | Size | Dilatation | Imaging | Comments |
|---|---|---|---|---|---|---|---|---|
| Ong7 | Singapore | 1 | M | 52 years | 3.0×18 mm | Pre | CTCA 3 months post insertion | Presentation with NSTEMI and patent stent at 3 months on CTCA |
| Roleder>9 | Poland | 1 | M | 83 years | 2.5×15 mm | N/A | Repeat angiography 2 weeks post insertion | Re-presentation with unstable angina and patent stent at 2 weeks on conventional angiography |
| Roleder10 | Poland | 6 |
M ×4 F ×2 |
Median 73 years (57-83) |
Median RLD 3.13 mm (2.25-3.45). Median MLD 1.12 (0.52-1.86 mm) |
Pre (100%) Post (100%) | Post-procedure OCT, surveillance OCT at 6 months and 18 months | Only one with 18-month and the rest 6-month follow-up with optical coherence imaging, with the majority showing neointimal formation. |
| Yew8 | Malaysia | 1 | M | 53 years | 3.0×28 mm | Pre + Post | Repeat PCI 4 weeks post angiogram | Presentation with ACS followed up with conventional angiography 4 weeks later |
| Wojakowski* | Poland | 1 | M | 54 years | 3.5×12 mm | Pre |
Post-procedure OCT 3 months following insertion |
54-year-old man with stable angina. Optical coherence imaging at 3 months showed complete apposition |
| Picard20 | Canada | 10 |
M ×9 F ×1 |
Median 75 years (64-81) |
Median diameter 3.5 mm (3-3.5). Median length 18 mm (18-28) |
Pre (84%) Post (77%) | N/A | Target vessel revascularisation similar to conventional stents (13 lesions) |
| * TCT 2014 C-167; ACS: acute coronary syndrome; CTCA: computed tomography coronary angiography; F: female; M: male; MLD: minimal lumen diameter; mm: millimetres; N/A: not available; NSTEMI: non-ST elevation myocardial infarction; OCT: optical coherence tomography; PCI: percutaneous coronary intervention; Post: post-dilation; Pre: predilation; RLD: reference lumen diameter | ||||||||
Conclusion
This report is unique given that it is the longest follow-up period to date of such an intervention with an Absorb scaffold in a saphenous vein graft. In an era when the use of BVS has been questioned, in this particular case the scaffold demonstrated excellent late angiographic patency confirmed on invasive OCT imaging. Meticulous attention to Absorb scaffold implantation technique is critical to providing good short-term as well as long-term clinical results21,22. Despite the Absorb BVS being recalled due to an increased incidence of target vessel myocardial infarctions11, there is still potential for benefit after scaffold resorption, and newer-generation BVS devices remain a viable option. Published data are still lacking to debunk its use entirely6.
Impact on daily practice
Our findings suggest that late OCT images of the scaffold treatment site are consistent with the existing chronology of events in intracoronary imaging of Absorb scaffolds in native coronary arteries. Further research and ideally a robust randomised trial comparing Absorb bioresorbable vascular scaffolds to conventional stenting are needed to clarify their role in this cohort.
Supplementary data
Optical coherence tomography study of the saphenous vein graft bioresorbable vascular scaffold.
Acknowledgments
Acknowledgements
The authors would like to acknowledge the input from Professor Peter Barlis and Dr Vikas Thondapu from the University of Melbourne regarding the optical coherence tomography images.
Conflict of interest statement
The authors have no conflicts of interest to declare.
Contributor Information
Vinayak Nagaraja, Department of Cardiology, Prince of Wales Hospital, Sydney, Australia.
Stuart Moss, St George Hospital, Sydney, Australia.
Nigel S. Jepson, Department of Cardiology, Prince of Wales Hospital, Sydney, Australia.
Mark R. Pitney, Sutherland Heart Clinic, Sutherland Hospital, Sydney, Australia.
References
- Onuma Y, Serruys PW, Perkins LE, Okamura T, Gonzalo N, García-García HM, Regar E, Kamberi M, Powers JC, Rapoza R, van Beusekom, van der, Virmani R. Intracoronary optical coherence tomography and histology at 1 month and 2, 3, and 4 years after implantation of everolimus-eluting bioresorbable vascular scaffolds in a porcine coronary artery model: an attempt to decipher the human optical coherence tomography images in the ABSORB trial. Circulation. 2010;122:2288–300. doi: 10.1161/CIRCULATIONAHA.109.921528. [DOI] [PubMed] [Google Scholar]
- Serruys PW, Onuma Y, Ormiston JA, de Bruyne, Regar E, Dudek D, Thuesen L, Smits PC, Chevalier B, McClean D, Koolen J, Windecker S, Whitbourn R, Meredith I, Dorange C, Veldhof S, Miquel-Hebert K, Rapoza R, García-García HM. Evaluation of the second generation of a bioresorbable everolimus drug-eluting vascular scaffold for treatment of de novo coronary artery stenosis: six-month clinical and imaging outcomes. Circulation. 2010;122:2301–12. doi: 10.1161/CIRCULATIONAHA.110.970772. [DOI] [PubMed] [Google Scholar]
- Abizaid A, Ribamar Costa, Bartorelli AL, Whitbourn R, van Geuns, Chevalier B, Patel T, Seth A, Stuteville M, Dorange C, Cheong WF, Sudhir K, Serruys PW ABSORB EXTEND investigators. The ABSORB EXTEND study: preliminary report of the twelve-month clinical outcomes in the first 512 patients enrolled. EuroIntervention. 2015;10:1396–401. doi: 10.4244/EIJV10I12A243. [DOI] [PubMed] [Google Scholar]
- Ormiston JA, Serruys PW, Regar E, Dudek D, Thuesen L, Webster MW, Onuma Y, Garcia-Garcia HM, McGreevy R, Veldhof S. A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial. Lancet. 2008;371:899–907. doi: 10.1016/S0140-6736(08)60415-8. [DOI] [PubMed] [Google Scholar]
- Ormiston JA, Serruys PW. Bioabsorbable coronary stents. Circ Cardiovasc Interv. 2009;2:255–60. doi: 10.1161/CIRCINTERVENTIONS.109.859173. [DOI] [PubMed] [Google Scholar]
- Serruys PW, Chevalier B, Sotomi Y, Cequier A, Carrié D, Piek JJ, Van Boven, Dominici M, Dudek D, McClean D, Helqvist S, Haude M, Reith S, de Sousa, Campo G, Iñiguez A, Sabaté M, Windecker S, Onuma Y. Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trial. Lancet. 2016;388:2479–91. doi: 10.1016/S0140-6736(16)32050-5. [DOI] [PubMed] [Google Scholar]
- Ong PJ, Jafary FH, Ho HH. “First-in-man” use of bioresorbable vascular scaffold in saphenous vein graft. EuroIntervention. 2013;9:165. doi: 10.4244/EIJV9I1A24. [DOI] [PubMed] [Google Scholar]
- Yew KL. Novel use of absorb bioresorbable vascular scaffold and STENTYS self-apposing coronary stent for complex saphenous vein grafts intervention. Int J Cardiol. 2014;177:e184–5. doi: 10.1016/j.ijcard.2014.08.043. [DOI] [PubMed] [Google Scholar]
- Roleder T, Ochala A, Smolka G, Wanha W, Wojakowski W. Implantation of a bioabsorbable vascular scaffold into a coronary vein graft: a two-week angiography follow-up. Kardiol Pol. 2014;72:281. doi: 10.5603/KP.2014.0052. [DOI] [PubMed] [Google Scholar]
- Roleder T, Wanha W, Smolka G, Zimoch J, Ochala A, Wojakowski W. Bioresorbable vascular scaffolds in saphenous vein grafts (data from OCTOPUS registry). Postepy Kardiol Interwencyjnej. 2015;11:323–6. doi: 10.5114/pwki.2015.55604. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kereiakes DJ, Ellis SG, Metzger C, Caputo RP, Rizik DG, Teirstein PS, Litt MR, Kini A, Kabour A, Marx SO, Popma JJ, McGreevy R, Zhang Z, Simonton C, Stone GW ABSORB III Investigators. 3-Year Clinical Outcomes With Everolimus-Eluting Bioresorbable Coronary Scaffolds: The ABSORB III Trial. J Am Coll Cardiol. 2017;70:2852–62. doi: 10.1016/j.jacc.2017.10.010. [DOI] [PubMed] [Google Scholar]
- Onuma Y, Serruys PW. Bioresorbable scaffold: the advent of a new era in percutaneous coronary and peripheral revascularization? Circulation. 2011;123:779–97. doi: 10.1161/CIRCULATIONAHA.110.971606. [DOI] [PubMed] [Google Scholar]
- Wykrzykowska JJ, Onuma Y, Serruys PW. Vascular restoration therapy: the fourth revolution in interventional cardiology and the ultimate “rosy” prophecy. EuroIntervention. 2009;5 Suppl F:F7–8. doi: 10.4244/EIJV5IFA1. [DOI] [PubMed] [Google Scholar]
- Sorrentino S, Giustino G, Mehran R, Kini AS, Sharma SK, Faggioni M, Farhan S, Vogel B, Indolfi C, Dangas GD. Everolimus-Eluting Bioresorbable Scaffolds Versus Everolimus-Eluting Metallic Stents. J Am Coll Cardiol. 2017;69:3055–66. doi: 10.1016/j.jacc.2017.04.011. [DOI] [PubMed] [Google Scholar]
- Zhang XL, Zhu QQ, Kang LN, Li XL, Xu B. Mid- and Long-Term Outcome Comparisons of Everolimus-Eluting Bioresorbable Scaffolds Versus Everolimus-Eluting Metallic Stents: A Systematic Review and Meta-analysis. Ann Intern Med. 2017;167:642–54. doi: 10.7326/M17-1101. [DOI] [PubMed] [Google Scholar]
- Gogas BD, Benham JJ, Hsu S, Sheehy A, Lefer DJ, -Goodchild TT, Polhemus DJ, Bouchi YH, Hung OY, Yoo SY, Joshi U, Giddens DP, Veneziani A, Quyyumi A, Rapoza R, King SB, Samady H. Vasomotor Function Comparative Assessment at 1 and 2 Years Following Implantation of the Absorb Everolimus-Eluting Bioresorbable Vascular Scaffold and the Xience V Everolimus-Eluting Metallic Stent in Porcine Coronary Arteries: Insights From In Vivo Angiography, Ex Vivo Assessment, and Gene Analysis at the Stented/Scaffolded Segments and the Proximal and Distal Edges. JACC Cardiovasc Interv. 2016;9:728–41. doi: 10.1016/j.jcin.2015.12.018. [DOI] [PubMed] [Google Scholar]
- Serruys PW, Ormiston J, van Geuns, de Bruyne, Dudek D, Christiansen E, Chevalier B, Smits P, McClean D, Koolen J, Windecker S, Whitbourn R, Meredith I, Wasungu L, Ediebah D, Veldhof S, Onuma Y. A Polylactide Bioresorbable Scaffold Eluting Everolimus for Treatment of Coronary Stenosis: 5-Year Follow-Up. J Am Coll Cardiol. 2016;67:766–76. doi: 10.1016/j.jacc.2015.11.060. [DOI] [PubMed] [Google Scholar]
- Nakatani S, Ishibashi Y, Sotomi Y, Perkins L, Eggermont J, Grundeken MJ, Dijkstra J, Rapoza R, Virmani R, Serruys PW, Onuma Y. Bioresorption and Vessel Wall Integration of a Fully Bioresorbable Polymeric Everolimus-Eluting Scaffold: Optical Coherence Tomography, Intravascular Ultrasound, and Histological Study in a Porcine Model With 4-Year Follow-Up. JACC Cardiovasc Interv. 2016;9:838–51. doi: 10.1016/j.jcin.2016.01.030. [DOI] [PubMed] [Google Scholar]
- Haude M, Ince H, Kische S, Abizaid A, Tölg R, Alves Lemos, Van Mieghem, Verheye S, von Birgelen, -Christiansen EH, Barbato E, Garcia-Garcia HM, Waksman R -BIOSOLVE-II and III investigators. Safety and clinical performance of a drug eluting absorbable metal scaffold in the treatment of subjects with de novo lesions in native coronary arteries: Pooled 12-month outcomes of BIOSOLVE-II and BIOSOLVE-III. Catheter Cardiovasc Interv. 2018;92:E502–11. doi: 10.1002/ccd.27680. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Picard F, Marquis-Gravel G, Avram R, Ly HQ, Dorval JF, Doucet S, de Hemptinne, L’allier PL, Tanguay JF. Everolimus-eluting bioresorbable vascular scaffold implantation to treat saphenous vein graft disease, single-center initial experience. J Interv Cardiol. 2017;30:433–9. doi: 10.1111/joic.12425. [DOI] [PubMed] [Google Scholar]
- Tanaka A, Latib A, Kawamoto H, Jabbour RJ, Sato K, Miyazaki T, Naganuma T, Mangieri A, Pagnesi M, Montalto C, Chieffo A, Carlino M, Montorfano M, Colombo A. Clinical outcomes of a real-world cohort following bioresorbable vascular scaffold implantation utilising an optimised implantation strategy. EuroIntervention. 2017;12:1730–7. doi: 10.4244/EIJ-D-16-00247. [DOI] [PubMed] [Google Scholar]
- Oberhauser JP, Hossainy S, Rapoza RJ. Design principles and performance of bioresorbable polymeric vascular scaffolds. EuroIntervention. 2009;5 Suppl F:F15–22. doi: 10.4244/EIJV5IFA3. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Optical coherence tomography study of the saphenous vein graft bioresorbable vascular scaffold.