Abstract
Pump thrombosis is a devastating complication after left ventricular assist device implantation. This study aims to elucidate the relation between left ventricular assist device implantation angle and risk of pump thrombosis. Between November 2010 and March 2020, 53 left ventricular assist device-patients underwent a computed tomography scan. Using a 3-dimensional multiplanar reformation the left ventricular axis was reconstructed to measure the implantation angle of the inflow cannula. All patients were retrospectively analyzed for the occurrence of pump thrombosis. In 10 (91%) patients with a pump thrombosis, the implantation angle was towards the lateral wall of the left ventricle. In only 20 patients (49%) of the patients without a pump thrombosis the inflow cannula pointed towards the lateral wall of the left ventricle. The mean angle in patients with a pump thrombosis was 10.1 ± 11.9 degrees towards the lateral wall of the left ventricle compared to 4.1 ± 19.9 degrees towards the septum in non-pump thrombosis patients (P = 0.005). There was a trend towards a significant difference in time to first pump thrombosis between patients with a lateral or septal deviated left ventricular assist device (hazard ratio of 0.15, P = 0.07). This study demonstrates that left ventricular assist device implantation angle is associated with pump thrombosis. Almost all patients in whom a pump thrombosis occurred during follow-up had a left ventricular assist device implanted with the inflow-cannula pointing towards the lateral wall of the left ventricle.
Keywords: CT based angle measurement, Left ventricular assist device, Pump thrombosis
Introduction
Developments in the field of mechanical circulatory support for end-stage heart failure have resulted in an increased use of left ventricular assist devices (LVAD), especially as destination therapy [1]. The excitement derived from improved survival after LVAD implantation, however, has been tempered by the risk of devastating complications including major bleedings and pump thrombosis (PT). Accordingly, reducing thrombogenic potential and thereby improving long-term outcomes is a major research topic.
Recently, the relation between PT and the angulation of the left ventricular (LV) inflow cannula has gained interest since the angulation is thought to affect intraventricular flow dynamics, in specific flow obstruction, which can predispose thrombus formation. Various hemodynamic simulations have been published to model the flow through the LV into the inflow cannula suggesting a relation between axis deviation and increased thrombogenicity [2–5].
Clinical investigations of inflow cannula position are scarce. Two previous studies showed that the cannula in patients with low incidence of PT was aligned parallel to the intraventricular septum [6] and directed at the mitral valve [7]. Most probably, this alignment promotes better flow patterns and efficient LV unloading. In a third study, Sorensen et al. showed that malpositioning towards the intraventricular septum might contribute to PT [8]. However, in this study only 2 of the 68 (2.9%) patients developed PT.
The present study aims to explore the relation between inflow cannula angulation and PT in the clinical setting. We hypothesize that deviations of the LVAD inflow cannula alignment lead to an increased risk of PT by means of unfavorable hemodynamics.
Materials and methods
Patient population
All 79 patients in whom a continuous flow LVAD (HeartWare Inc., Framingham, MA) was implanted as destination therapy for heart failure in the Leiden University Medical Center, between November 2010 and March 2020 were eligible for inclusion. Our center has a registration for destination therapy only. For the current study, patients in whom no computed tomography (CT) scan was performed after LVAD implantation were excluded (n = 26). Of these 26 patients only 1 patient had a PT. For the present analysis patients with a congenital heart disease requiring a ventricular assist device were not included. Informed consent was obtained from all included patients that were alive when the study was conducted. The study protocol was reviewed by the local medical ethics committee (G20.182) who waived the need for official approval according to the Medical Research Involving Human Subjects Act.
Data analysis
Clinical, laboratory and survival data were collected and analyzed retrospectively from the patient information systems (EPD-Vision; Leiden University Medical Center, Leiden, the Netherlands; HiX 6.1, Chipsoft, Amsterdam, the Netherlands). Baseline variables, including etiology of heart failure and Standard Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) classification, were collected at date of LVAD implantation. Medical history was screened for the occurrence of PT. PT was defined as ≥ 2 signs or symptoms of PT in combination with an accompanying intervention such as intensified treatment with anti-coagulation (standard regimen consisted of vitamin K antagonist in combination with clopidogrel), intravenous thrombolytics or pump replacement. The following signs and symptoms were considered suggestive of PT in line with the consensus document by Kormos et al.: (1) presence of hemolysis, (2) worsening of heart failure and (3) abnormal pump parameters [9, 10]. Any available CT visualizing the thorax that was performed for any clinical indication during follow up, was used for measurement of LVAD implantation angle.
CT measurements
Non-ECG gated CT data sets were used to measure the implantation angle of the inflow cannula of the LVAD. The LV long axis was reconstructed using a 3-dimensional multiplanar reformation (MPR) (Fig. 1). First, on a transversal plane the MPR crosshair was aligned with the mitral valve annulus and the perpendicular part of the crosshair parallel to the LV long-axis (Panel A). This was repeated in the sagittal view to create a double oblique short-axis view of the LV (Panel B). Then, the crosshair was further aligned to the septum (Panel C) to create a double oblique four-chamber and two-chamber view, with the crosshair parallel to the LV long-axis. By adjusting the window settings (width 3000 Hounsfield units, level 800 Hounsfield units) the inflow cannula of the LVAD was clearly visualized (Panel D and E). The angle between the inflow cannula and the LV-axis was then measured in the four-chamber (Panel E) (septal or lateral) and two-chamber view (Panel D) (anterior or inferior). To improve reproducibility the septum was mainly used as a reference instead of the neo-apex which might be influenced as a result of the surgical procedure. Measurements were performed by two experienced observers (MG and KK), in two different readings. The results were compared to assess the inter-observer variability. For the final analysis the average between the two measurements was used. After a period of several months one observer (MG) re-measured all cases to assess intra-observer variability blinded to initial results.
Follow-Up/Statistical analysis
Continuous variables are expressed as mean ± SD when normally distributed or otherwise as the median and interquartile range (IQR). Categoric variables are presented as numbers and percentages. The implantation angle of patients with and without PT was compared. Comparison of continuous data was performed using two-tailed unpaired Student t test for normally distributed variables.
Patients were included in the current study on the date of LVAD implantation. The first CT scan visualizing the thorax after LVAD implantation was used for the measurements of the implantation angle. All included patients had a CT scan before occurrence of a PT. For the present study we assumed that the angle of the LVAD did not change over time. Therefore, the measurement on the date of CT was used as a proxy for the angle on the date of implant. Time to first PT from study entry at date of LVAD implantation was compared between patients with a septal versus a lateral deviated LVAD. To account for competing risks, we carried out a cause specific hazard analysis for the time from implant to first PT based on a Cox regression analysis. The extent of agreement between the two measurements of the implantation angle of the inflow cannula of the LVAD was assessed with the Bland–Altman test and intraclass correlation coefficient for both inter and intra-observer variability. P values less than 0.05 were considered statistically significant and are marked bold in the tables. Statistical analysis was performed using SPSS 25.0 software (IBM, Armonk, NY).
Results
Patient population
The baseline characteristics of the 53 included patients are depicted in Table 1 and 2. Mean age was 62.4 ± 9.2 years and 40 patients (75%) were male. The majority of patients had an ischemic etiology of their heart failure. There were no significant differences in baseline characteristics between patients with and without pump thrombosis except that patients with pump thrombosis had a higher Log EuroScore and more frequently diabetes mellitus. In most patients (94%), a median sternotomy was used for LVAD implantation. The median time between the date of LVAD implantation and the scan used for assessment of the LVAD inflow cannula position was 83 days (IQR 15–656). Of the 53 CT scans used, 32 were contrast-enhanced (60%).
Table 1.
Baseline characteristics | Total (N = 53) | PT (N = 11) | No PT (N = 42) | P-value |
---|---|---|---|---|
Age (years) | 62.4 ± 9.2 | 62.3 ± 8.5 | 62.5 ± 9.5 | 0.955 |
Gender (male) | 40 (75%) | |||
BMI (kg/m2) | 26.0 ± 3.6 | 27.0 ± 2.9 | 25.6 ± 3.8 | 0.334 |
Log EuroScore (%) | 20.0 (IQR 10.0–33.5) | 29.0 (IQR 23.0–47.0) | 18.5 (IQR 9.75–28.0) | 0.035 |
INTERMACS profile | NA | |||
1.Critical cardiogenic shock | 2 (4%) | 0 | 2 (5%) | |
2.Progressive decline | 3 (6%) | 1 (9%) | 2 (5%) | |
3.Stable but inotrope dependent | 27 (51%) | 6 (55%) | 21 (50%) | |
4.Resting symptoms | 9 (17%) | 1 (9%) | 8 (19%) | |
5.Exertion intolerant | 11 (21%) | 3 (27%) | 8 (19%) | |
6.Exertion limited | 0 | 0 | 0 | |
7.Advanced NYHA class 3 | 1 (2%) | 0 | 1 (2%) | |
Etiology of heart failure | 0.804 | |||
Ischemic cardiomyopathy | 32 (60%) | 7 (64%) | 25 (60%) | |
Non-ischemic cardiomyopathy | 21 (40%) | 4 (36% | 17 (41%) | |
Left ventricular ejection fraction before implant (%) | 23 (18–29) | 27 (23–30) | 21 (17–29) | 0.062 |
Left ventricular end-diastolic diameter (mm) | 67 (64–73) | 70 (67–78) | 67 (63–73) | 0.079 |
Previous thoracotomy | 25 (47%) | 7 (64%) | 18 (43%) | 0.219 |
Previous atrial fibrillation | 23 (43%) | 6 (55%) | 17 (41%) | 0.402 |
Previous ischemic stroke | 4 (8%) | 0 (0%) | 4 (10%) | 0.287 |
Previous venous thromboembolism | 4 (8%) | 0 (0%) | 4 (10%) | 0.287 |
Previous diabetes mellitus | 11 (21%) | 5 (46%) | 6 (14%) | 0.023 |
Cardiac resynchronization therapy | 38 (72%) | 9 (82%) | 29 (69%) | 0.403 |
INTERMACS Interagency Registry for Mechanically Assisted Circulatory Support, NYHA New York Heart Association, PT pump thrombosis
Patient data at baseline. Values are shown as mean ± SD, n (%) or median (interquartile range)
Table 2.
Surgical characteristics | Total (N = 53) | PT (N = 11) | No PT (N = 42) | P-value |
---|---|---|---|---|
Median sternotomy | 50 (94) | 9 (82) | 41 (98) | 0.044 |
Concomitant surgery | 47 (89) | 9 (82) | 38 (90) | 0.420 |
Tricuspid valve annuloplasty | 38 (72) | 7 (64) | 31 (74) | 0.505 |
LAA excision | 31 (59) | 4 (36) | 27 (64) | 0.094 |
Left ventricular reconstruction | 9 (17) | 2 (19) | 7 (17) | 0.905 |
Aortic valve replacement | 8 (15) | 2 (19) | 6 (14) | 0.748 |
PFO/ASD closure | 2 (4) | 0 (0) | 2 (5_ | 0.461 |
ASD Atrial Septal Defect, LAA left atrial appendage, PFO patent foramen ovale, PT pump thrombosis.
Surgical data at baseline. Values are shown as n (%)
Clinical follow-up
During a median follow-up of 582 days (IQR 208–1561), 38 patients (72%) died with a median time from implant to death of 478 days (IQR 59–1378). In total, eleven of the 53 patients experienced one or more LVAD PT. Median time from LVAD implantation to first PT was 332 days (IQR 74–651). INR values of patients who experienced a PT were all above 2.0 (2.2–4.3) except for two patients. One patient used therapeutic low molecular weight heparin because of recent hemorrhagic stroke (oral anticoagulation was discontinued). Another patient used unfractionated heparin because he was still admitted to the intensive care unit after LVAD implantation (APTT 63–72).
LVAD implantation angle measurement: inter and intra-observer variability
The CT scans of all 53 patients were reviewed by two independent observers (Fig. 2a). In one patient the angle could not be assessed on the four-chamber view by both reviewers due to insufficient scan range; this patient did not have a PT. Correlation for the implantation angle between the observers on the four-chamber view was 0.962 (P < 0.001) and on the two-chamber view 0.923 (P < 0.001). Bland–Altman analysis for the four-chamber view demonstrated a bias of − 1,9 degrees with limits of agreement ranging from − 9.8–6.0 degrees. Similarly, for the two-chamber view the bias was 0.9 degrees with limits of agreement ranging from − 9.2–11.1 degrees. Similar results were observed for intra-observer variability. (Fig. 2b). No clear differences were observed in reproducibility between contrast or non-contrast enhanced CT-scans (data not shown).
LVAD implantation angle
The LVAD implantation angle was assessed on the four-chamber view. Of the 52 patients, 22 (42%) had an LVAD implanted towards the LV septum and 30 (58%) towards the LV lateral free wall. Overall, the mean implantation angle on the four-chamber view was 1.0 ± 1.9 degrees towards the septal wall. On the two-chamber view in 50 patients (94%) the LVAD implantation angle on the two-chamber view was towards inferior, whereas in only 3 (6%) patients the LVAD was facing the anterior wall of the LV. Overall, on the two-chamber view the angle was 20.1 ± 12.8 degrees towards the inferior wall.
Relation between PT and implantation angle
To assess the association between implantation angle and PT, a comparison was made between the 11 patients with and the 42 patients without PT. On the four-chamber view, in the 11 patients with PT, 10 patients (91%) had an implantation angle towards the lateral wall of the left ventricle (LV). Interestingly, of the patients without PT, only 20 patients (49%) had an implantation angle towards the lateral wall of the LV (P = 0.012). As depicted in Fig. 3, the mean angle in patients with an LVAD PT was 10.1 ± 11.9 towards the lateral wall of the LV. In the non-PT group the mean angle was 4.1 ± 19.9 towards the septum (P = 0.005). On the two-chamber view, none of the patients with a PT had an LVAD angulated towards the anterior wall. In non-PT patients, only 3 of the 42 patients had an LVAD facing the anterior wall (7%). The mean angle was 19.4 ± 14.1 in patients with a PT compared to 20.3 ± 12.6 for non-PT patients (P = 0.844) (Fig. 3). Figure 4 demonstrates the relation between the implantation angle on the four- and two-chamber view in relation to the occurrence of LVAD PT. Of note, the majority of patients with PT had an implantation angle towards the inferolateral wall of the LV.
Using Cox regression survival analyses, the difference in time between implant to first PT between patients with a lateral or septal deviated LVAD was assessed. With a hazard ratio of 0.15 (95% confidence interval 0.02–1.16) (P = 0.068), there was a trend towards a difference. Septal deviation is associated with a lower hazard ratio.
Discussion
The main finding of the current study is that LVAD implantation angle, evaluated by 3-dimensional assessment on computed tomography, is associated with PT. Almost all patients in whom a PT occurred during follow-up had an LVAD implanted with the inflow-cannula pointing towards the lateral wall of the LV.
LVAD PT is one of the most feared complications occurring in 2–6% of the patients within the first six months after implant [9, 11]. Treatment remains challenging. Adherence to a structured surgical implant to create an unobstructed blood flow path and clinical management is associated with lower incidence of (early) PT. Survival after LVAD PT is around 70–92% at 30 days [12, 13]. The mortality in our study is higher because our patient population only consisted of destination therapy patients and our follow-up period was longer than 30 days.
In the current study, patients with PT had an average angle between the inflow cannula and the LV axis of 10 degrees towards the lateral wall of the LV. This is in line with the findings by Chivukula et al. who used computational fluid dynamics to model the inflow for various inflow cannula angulations in vitro [14]. They concluded that angulation of the inflow cannula > 7 degrees from the apical axis leads to unfavorable hemodynamics and thereby potentially increased thrombogenicity. It should be noted that this was a computational study, not including real patient data. Also, the left ventricular apex position is associated with the occurrence of PT [15]. There is some data on relation between LVAD angle and PT derived from pathology studies suggesting that minor thrombosis of the inflow cannula is frequently observed and unrelated to clinically relevant PT. In a pathologic case serie of eight patients with an LVAD who were successfully bridged to transplant, the position of the cannula was assessed by chest X-rays, CT scans and echocardiography [16]. All patients were found to have thrombus associated with the outer aspect of the LVAD inflow cannula but none had signs of clinical PT. Only one patient showed a sub-optimal positioning of the inflow cannula (towards posterior). It is important that these were patients successfully bridged to transplant, so no cases with clinically manifest PT were included. Limited studies on PT and LVAD angle in-vivo are available. In a recent study, 63 patients with an LVAD were followed for one year from index discharge [17]. Cannula coronal angle was measured on chest X-rays and linked to a composite endpoint of stroke and PT. There was no relation with cannula angle and this endpoint. However, only two-dimensional chest X-ray was used. The only study on the relation between PT and dedicated 3-dimensional assessment of LVAD inflow angle was recently published by Sorensen et al. [8] In 68 HeartWare LVAD and 54 HeartMate LVADs they showed that a deviation towards the interventricular septum was associated with an increased risk of thrombosis. However, it should be noted that in their study only 2 PTs occurred in patients with a HeartMate.
Potentially is it not the direction of which the inflow cannula is deviating but merely the deviation itself is related to PT. Noteworthy, during implantation of an LVAD, the surgeon is able to determine the location of the inflow cannula. However influencing the angle between the LV septum and LV inflow cannula in an empty LV during cardiopulmonary bypass is challenging. A solution for this problem might be a gimbaled sewing ring. With this ring, the angle can be adapted for several degrees post-implant [18].
Strengths and limitations
The main strength of the present analysis is the dedicated 3-dimensional analysis of LVAD inflow cannula implantation angle. Previous studies assessed the LVAD implantation angle on a 2-dimensional anterior posterior chest X-ray [2, 7]. However, the LV is a conical shaped structure, and especially in patients with (dilated) cardiomyopathies, the LV can be extremely deformed with an aberrant position within the thorax. For a dedicated, precise analysis, a 3-dimensional approach is needed. In the present analysis, we assumed that LVAD implantation angle is stable over time. This assumption is supported by two studies by Kazui et al. and Adamson et al., showing a stable position over time using routine chest radiographs in patients with an LVAD [19, 20]. There is no current proof of LVAD position stability on 3-dimensional imaging. Another strength of the present cohort is the fact that it consists of patients in whom an LVAD was implanted as destination therapy which allows for long(er) follow-up.
There are some limitations which need consideration when interpreting the results of the present study. First, this is a single-center study and the results need duplication in a larger cohort. Secondly, for this analysis only patients with a HeartWare LVAD were included. Probably the results can be extrapolated to other types of inflow cannulas but this remains to be investigated. The global sale of the HeartWare system has recently been stopped due to observational evidence associated with increased neurological adverse events and mortality. However, ongoing support is necessary for patients who currently have this type of LVAD, approximately 4000 people worldwide. Despite the fact that the present investigation is a retrospective study, there is no loss to follow-up. All patients are followed at the out-patient clinic of the Leiden University Medical Center and informed consent was obtained in all patients. Potentially, bias could have been introduced by the fact that the study only included patients in whom, for clinical purposes, a CT-scan was performed. Only one patient with a PT had no scan in our cohort. Hypothetically in patients with more co-morbidities or worse cardiac functional status more often CT-scans are performed and these co-morbidities might correlate with PT. Moreover, for the present study only static CT scan was available. In none of our patients a full cardiac cycle ECG-gated CTA was available. Therefore, the influence of the cardiac cycle itself on the orientation of the inflow canula in relation to the LV axis can not be evaluated.
Conclusion
This study, evaluating 3-dimensional LVAD inflow cannula position in the LV by means of computed tomography, demonstrates that angulation deviations are associated with PT. Especially, an inflow cannula position towards the lateral wall of the LV is associated with increased thrombotic risk. This underlines the need for careful positioning of the inflow cannula during LVAD implantation.
Author contributions
Conception and design: KK, MdG. Provision of study materials or patients: MP, JB, LT, SB. Collection and assembly of data: KK, MdG. Data analysis and interpretation: KK, MdG, BM. Manuscript writing: All authors. Final approval of manuscript: All authors.
Funding
The department of Cardiology received research grants from Biotronik, Medtronic and Boston Scientific. This funding was not applied to the current research.
Data availability
Individual patients data are not available for data sharing. The data that support the findings of this study might be available from the corresponding author upon reasonable request.
Declarations
Conflict of interest
There is no conflict of interest for the present manuscript.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Kirsten A. Kortekaas and Michiel A. de Graaf have contributed equally.
References
- 1.Michaels A, Cowger J. Patient selection for destination LVAD therapy: predicting success in the short and long term. Curr Heart Fail Rep. 2019;16(5):140–149. doi: 10.1007/s11897-019-00434-1. [DOI] [PubMed] [Google Scholar]
- 2.Bhama JK, Bansal A. Left ventricular assist device inflow cannula position may contribute to the development of HeartMate II left ventricular assist device pump thrombosis. Ochsner J. 2018;18(2):131–135. doi: 10.31486/toj.17.0070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Fraser KH, Taskin ME, Griffith BP, Wu ZJ. The use of computational fluid dynamics in the development of ventricular assist devices. Med Eng Phys. 2011;33(3):263–280. doi: 10.1016/j.medengphy.2010.10.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Laumen M, Kaufmann T, Timms D, Schlanstein P, Jansen S, Gregory S, et al. Flow analysis of ventricular assist device inflow and outflow cannula positioning using a naturally shaped ventricle and aortic branch. Artif Organs. 2010;34(10):798–806. doi: 10.1111/j.1525-1594.2010.01098.x. [DOI] [PubMed] [Google Scholar]
- 5.Ong C, Dokos S, Chan B, Lim E, Al Abed A, Bin Abu Osman NA, et al. Numerical investigation of the effect of cannula placement on thrombosis. Theor Biol Med Model. 2013;10:35. doi: 10.1186/1742-4682-10-35. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Adamson RM, Mangi AA, Kormos RL, Farrar DJ, Dembitsky WP. Principles of HeartMate II implantation to avoid pump malposition and migration. J Card Surg. 2015;30(3):296–299. doi: 10.1111/jocs.12478. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Imamura T, Nguyen A, Chung B, Rodgers D, Sarswat N, Kim G, et al. Association of inflow cannula position with left ventricular unloading and clinical outcomes in patients with heartmate II left ventricular assist device. Asaio J. 2019;65(4):331–335. doi: 10.1097/MAT.0000000000000823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Sorensen EN, Kon ZN, Feller ED, Pham SM, Griffith BP. Quantitative assessment of inflow malposition in two continuous-flow left ventricular assist devices. Ann Thorac Surg. 2018;105(5):1377–1383. doi: 10.1016/j.athoracsur.2017.12.004. [DOI] [PubMed] [Google Scholar]
- 9.Maltais S, Kilic A, Nathan S, Keebler M, Emani S, Ransom J, et al. PREVENtion of HeartMate II pump thrombosis through clinical management: the PREVENT multi-center study. J Heart Lung Transpl. 2017;36(1):1–12. doi: 10.1016/j.healun.2016.10.001. [DOI] [PubMed] [Google Scholar]
- 10.Kormos RL, Antonides CFJ, Goldstein DJ, Cowger JA, Starling RC, Kirklin JK, et al. Updated definitions of adverse events for trials and registries of mechanical circulatory support: a consensus statement of the mechanical circulatory support academic research consortium. J Heart Lung Transpl. 2020;39(8):735–750. doi: 10.1016/j.healun.2020.03.010. [DOI] [PubMed] [Google Scholar]
- 11.Kirklin JK, Naftel DC, Pagani FD, Kormos RL, Myers S, Acker MA, et al. Pump thrombosis in the thoratec HeartMate II device: an update analysis of the INTERMACS registry. J Heart Lung Transpl. 2015;34(12):1515–1526. doi: 10.1016/j.healun.2015.10.024. [DOI] [PubMed] [Google Scholar]
- 12.Luc JGY, Tchantchaleishvili V, Phan K, Dunlay SM, Maltais S, Stulak JM. Medical therapy as compared to surgical device exchange for left ventricular assist device thrombosis: a systematic review and meta-analysis. Asaio J. 2019;65(4):307–317. doi: 10.1097/MAT.0000000000000833. [DOI] [PubMed] [Google Scholar]
- 13.Scandroglio AM, Kaufmann F, Pieri M, Kretzschmar A, Müller M, Pergantis P, et al. Diagnosis and treatment algorithm for blood flow obstructions in patients with left ventricular assist device. J Am Coll Cardiol. 2016;67(23):2758–2768. doi: 10.1016/j.jacc.2016.03.573. [DOI] [PubMed] [Google Scholar]
- 14.Chivukula VK, Beckman JA, Prisco AR, Dardas T, Lin S, Smith JW, et al. Left ventricular assist device inflow cannula angle and thrombosis risk. Circ Heart Fail. 2018;11(4):e004325. doi: 10.1161/CIRCHEARTFAILURE.117.004325. [DOI] [PubMed] [Google Scholar]
- 15.Yarboro LT, Mehaffey JH, Hawkins RB, Kron IL, Ailawadi G, Kern JA, et al. Pre-implant left ventricular apex position predicts risk of HeartMate II pump thrombosis. J Card Surg. 2017;32(12):837–842. doi: 10.1111/jocs.13382. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Strickland KC, Watkins JC, Couper GS, Givertz MM, Padera RF. Thrombus around the redesigned HeartWare HVAD inflow cannula: a pathologic case series. J Heart Lung Transpl. 2016;35(7):926–930. doi: 10.1016/j.healun.2016.01.1230. [DOI] [PubMed] [Google Scholar]
- 17.Imamura T, Narang N, Nitta D, Fujino T, Nguyen A, Chung B, et al. HeartWare ventricular assist device cannula position and hemocompatibility-related adverse events. Ann Thorac Surg. 2020;110(3):911–917. doi: 10.1016/j.athoracsur.2019.12.049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Tamez D, LaRose JA, Shambaugh C, Chorpenning K, Soucy KG, Sobieski MA, et al. Early feasibility testing and engineering development of the transapical approach for the HeartWare MVAD ventricular assist system. Asaio J. 2014;60(2):170–177. doi: 10.1097/MAT.0000000000000038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Kazui T, Zhang A, Greenberg J, Itoh A, Tran PL, Keith AD, et al. Left ventricular assist device inflow angle and pump positional change over time adverse impact on left ventricular assist device function. Ann Thorac Surg. 2016;102(6):1933–1940. doi: 10.1016/j.athoracsur.2016.05.025. [DOI] [PubMed] [Google Scholar]
- 20.Adamson RM, Bower BL, Sundareswaran KS, Farrar DJ, Dembitsky WP. Radiologic assessment of HeartMate II position: minimal pump migration after long-term support. J Heart Lung Transpl. 2015;34(12):1617–1623. doi: 10.1016/j.healun.2015.07.006. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Individual patients data are not available for data sharing. The data that support the findings of this study might be available from the corresponding author upon reasonable request.