Abstract
The irregular shape of the tumor and the viscoelastic mechanical behavior of the brain tissue cause a position- and time-dependent pressure profile at the interface of the tumor and brain during tumor growth, respectively. Calculating the resultant pressure profile through numerical approaches, such as finite element method (FEM), may contribute to the surgical and radiotherapeutic treatments designing for the brain tumor patients via precisely build statistical atlases of brain geometry. However, to date, FE modeling of the tumor growth in the brain has been limited to a) application of a constant or a time-independent pressure boundary, b) application of time-independent material properties for the brain tissue, e.g., elastic and hyperelastic, and c) application of a quasi-spherical shape for the tumor FE model. This study proposed a computational framework to calculate the resultant position- and time-dependent pressure profile on the brain tissue due to tumor growth. A FE patch of the brain tissue was constructed and an inverse dynamic FE-optimization algorithm was used to calculate its viscoelastic mechanical properties under compressive uniaxial loading. To calculate the simulation time that results in a stable (independent to shear relaxation modulus) mechanical response, several FE simulations at various displacement rates were performed. Finally, two patient-specific post-tumor resection FE models were input to the FE-optimization algorithm to calculate the optimized 3rd-order position-dependent and normal distribution time-dependent pressure profile parameters. The optimized viscoelastic material properties, the most suitable simulation time, and the optimized 3rd-order position- and -time-dependent pressure profiles were calculated.
Keywords: Brain tumor, Viscoelastic, Inverse Finite Element, Optimization, Pressure Boundary
1. Introduction
Primary brain tumors is a major public health concern (Eichberg et al., 2019). Annually, more than ~20,000 patients are diagnosed with a high-grade glioma, with ~100,000 to 200,000 new cases of brain metastases each year (Jemal et al., 2008). Construction of accurate brain tumor atlases requires a correct 3D registration of the brain tumor images to a stereotactic space (Clatz et al., 2005; Mohamed and Davatzikos, 2005). Although brain tumor atlases would have practical applications for a more comprehensive tumor resection planning and therapeutic approaches via the variable parameters in the tumor, such as grade, size, and progression (Cuadra et al., 2002; Dawant et al., 2002; Kyriacou et al., 1999; Mohamed et al., 2001), to date, there are still some challenges in the tumor vicinity due to the tumor irregular shape and progression complications (Cuadra et al., 2004; Dawant et al., 2002). The contribution of biological factors as well as the biomechanical forces may alter the tumor microenvironment, and in turn the progression and response to the treatment in the tumor may get affected (Jain et al., 2014).
Growth of a tumor inside a stiffer fixed-volume solid matrix of the skull causes a generation and accumulation of forces due to mechanical interactions at the interface of the tumor and the brain following by the mass-effect (Gerard et al., 2017; Mokri, 2001; Orringer et al., 2012) that is a main reason of neurologic injury (Ropper, 1986). These mechanical forces could hinder inducing apoptosis, cancer cell proliferation, and augment the invasive and metastatic potential of cancer cells influencing tumor progression (Cheng et al., 2009; Demou, 2010; Helmlinger et al., 1997; Janet et al., 2012; Kaufman et al., 2005; Stylianopoulos et al., 2012). Moreover, they can invoke up to ~24 mm brain shift during tumor resection (Bayer et al., 2017; Chen et al., 2011; Gerard et al., 2017; Miga et al., 2016; Nimsky et al., 2000).
Finite element method (FEM) is a numerical approach that allows calculating the resultant forces and mass-effects due to tumor growth under various loading and boundary conditions as well as material properties. Various 2D (Kyriacou et al., 1999; Mohamed et al., 2001; Wasserman and Acharya, 1996) and 3D tumor growth (Angeli and Stylianopoulos, 2016; Dumpuri et al., 2007; Hawkins-Daarud et al., 2012; Hogea et al., 2007; Hogea et al., 2006a; Hogea et al., 2006b; Lloyd et al., 2007; Mohamed and Davatzikos, 2005; Nimsky et al., 2004) as well as mass-effect (Dumpuri et al., 2010; Fan et al., 2011; Sun et al., 2014) FE models have been developed to calculate the mechanical forces and mass-effects due to tumor growth. These models have all been limited to whether using a constant outward pressure profile in the tumor boundary to control the growth or using a simplified quasi-spherical tumor shape. However, very recently our group showed that a constant pressure boundary unable to capture the realistic irregular shape of a tumor so the application of a variable pressure boundary as a function of tumor shape is highly preferred (Abdolkarimzadeh et al., 2021). Yu et al., (Yu et al., 2022) employed hyperelastic material model to estimate the brain deformations in patients after tumor removal. However, it is known that soft biological tissues in general (Dehoff, 1978; Estermann et al., 2022; Holzapfel, 2017; Humphrey, 2003; Karimi et al., 2017a; Maccabi et al., 2018) and the brain tissue specifically are highly viscoelastic (Bilston et al., 1997; Budday et al., 2017; Finan et al., 2017; MacManus et al., 2017; Qiu et al., 2020; Sundaresh et al., 2021), so their mechanical response depend on the rate of the applied load. In our prior study (Abdolkarimzadeh et al., 2021), we used an elastic material model to address the mechanical properties of the brain tissue. However, the results revealed a considerable difference when it comes to the nodal comparison between the pre- and post-surgery tumor shapes. The error may relate to the simple elastic material model of the brain tissue, while it has been well documented that the brain tissue is highly viscoelastic (Bilston et al., 1997; Budday et al., 2017; Finan et al., 2017; MacManus et al., 2017; Qiu et al., 2020; Sundaresh et al., 2021). Although the application of a time-independent material model could be able to result in local accuracy when it comes to local comparison, it is difficult to capture the mechanical response of the entire brain throughout the tumor growth cycle. To date, there is a gap in our knowledge about the role of the brain tissue viscoelasticity on the resultant pressure profile in the brain tissue due to the tumor growth. In this study, we hypothesized that the applied pressure boundary on the brain tissue due to the tumor growth is highly position- and time-dependent that stems from the viscoelastic mechanical behavior of the brain tissue (Bilston et al., 1997; Budday et al., 2017; Finan et al., 2017; MacManus et al., 2017; Qiu et al., 2020; Sundaresh et al., 2021) as well as the irregular shape of the tumor (Aghalari et al., 2021; Hussain et al., 2017; Hussain et al., 2018), respectively. Since the strength of the bulk tumor mass-effect and the final tumor size (Zacharaki et al., 2008) in addition to the tumor progression and response to treatment have been reported to be regulated by the resultant pressure parameters during the tumor growth (Angeli and Stylianopoulos, 2016).
This study is therefore aimed to calculate the position- and time-dependent pressure boundary on a viscoelastic brain tissue due to the tumor growth that may contribute to a precise estimation of mass-effect due to tumor resection surgery using a fully coupled inverse dynamic FE-optimization algorithm. Since we have a time-dependent pressure boundary that is being updated at each load increment (time-step), a simple elastic material model unable to consider that while viscoelastic material model understands a time-dependent pressure. In addition, a simple elastic material model results in the same stresses and strains in the tissue regardless of the rate of the applied load but only is sensitive to the magnitude of the load. However, in the case of tumor growth, the brain tissue is constantly being deformed so the applied pressure on the tissue is constantly being updated. Therefore, we required to employ a time- and position-dependent pressure profile to address the tumor growth. To do that, a FE patch of the brain tissue was constructed and subjected to a compressive load to calculate the viscoelastic parameters of the brain tissue using a FE-optimization algorithm (Abdolkarimzadeh et al., 2021; Rahmati et al., 2021) based on the experimental human brain data (Karimi et al., 2017b). Thereafter, a set of FE simulations were performed using the same patch model to calculate the most suitable simulation time that results in stability in the viscoelastic parameters of the brain tissue (independency to the shear relaxation modulus). The calculated viscoelastic parameters and the simulation time were then assigned to volume-meshed post-surgery FE models of the human brain (Abdolkarimzadeh et al., 2021) to calculate the optimized 3rd-order position-dependent and normal distribution time-dependent pressure function based on the minimum distance between the surface-nodes of the post-surgery brain cavity and pre-surgery tumor. To model the tumor growth, the ideal method is to apply load on tumor in the first stages of its growth aiming to reach the final shape of the tumor before tumor resection. However, the data corresponding to the first stages of tumor are rare to access since tumor symptoms appear in the final stages of tumor growth and patients usually make an appointment to visit a doctor. Therefore, to an inverse FE approach was used in which the post-resection model was loaded to reach pre-resection tumor model.
2. Materials and Methods
2.1. Brain Patch Finite Element Model – Viscoelastic Material Properties
It has been well documented that the brain tissue is viscoelastic (Donnelly and Medige, 1997; Fallenstein et al., 1969; Forte et al., 2017) and its mechanical behavior can be addressed through Prony series (Forte et al., 2017; Miller and Chinzei, 1997; Prange and Margulies, 2002; Rashid et al., 2012). A viscoelastic material model has both viscous and elastic properties (Herrmann, 1968). The behavior of a viscoelastic material can be outlined via short- and long-time shear moduli as follows:
| (1) |
where G0 and G∞ are the short-time and long-time shear modulus, respectively, and β is the decay constant. A Jaumann rate formulation is also used as follows:
| (2) |
where the prime denotes the deviatoric part of the stress rate, , and the strain rate, Dij.
To calculate the viscoelastic mechanical properties of the brain tissue, a brain patch FE model was constructed, subjected to a compressive strain, and the resultant stress was compared to the prior experimental brain compressive data by Karimi et al., (Karimi et al., 2017b). The FE model of the brain patch has the length, width, and thickness of 30 mm × 30 mm × 20 mm that match the experimental study (Karimi et al., 2017b) (Figure 5 of that paper) as shown in Fig. 1a. Thereafter, the patch model was subjected to a uniaxial compression in the X direction, where the displacement boundary condition (~40% strain) was applied to the FE model to mimic the uniaxial mechanical testing of human brain (Karimi et al., 2017b). The optimization algorithm of Fminsearch-Unconstrained nonlinear minimization was coupled with the solver of LS-DYNA (Livermore Software Technology Corporation, CA, US) to calculate the viscoelastic parameters of the patch FE model (Abdolkarimzadeh et al., 2021; Rahmati et al., 2021). Fminsearch finds the minimum of a scalar function of several parameters, starting at an initial estimation. An initial guess, herein G0=528 kPa, G∞=168 kPa, and β=700 1/s (Galford and McElhaney, 1970), with the upper and lower boundaries of 34< G0 <528 kPa, 5.5< G∞ <168 kPa, and 30< β <1000 1/s (Turquier et al., 1996; Zhang et al., 2001), were set for the patch FE model and was started to run with the cost function of mean squared error (Schmidt et al., 2009). The mean squared error was the sum of the squared differences between the experimental data (Karimi et al., 2017b) and optimization value. The resultant stress-strain in the gauge at the center of the FE patch model (Fig. 1a) was calculated and plotted versus the experimental data (Karimi et al., 2017b) (Figure 5 of that paper) as displayed in Fig. 1b. The final optimized viscoelastic material properties were assigned to the brain tissue FE models as listed in Table 1.
Fig. 5.

Comparative contour maps of deformation in the post-resection brain cavity versus the pre-resection tumor surface nodes in (a) patient #11 and (b) patient #12.
Fig. 1.

(a) FE model of the brain patch, including the strain gauge and boundary conditions. (b) Stress-strain curves representing the experimental uniaxial compressive testing of human brain patches, compared to identical FE model.
Table 1.
The average von Mises stress at different simulation times.
| Time (μs) | Average von Mises Stress (kPa) |
|---|---|
|
| |
| 720 | 44.3 |
| 7200 | 21.4 |
| 10000 | 19.2 |
| 20000 | 16.3 |
| 30000 | 15.3 |
| 40000 | 14.8 |
| 50000 | 14.4 |
| 60000 | 14.4 |
| 70000 | 14.2 |
| 72e3 | 14.2 |
| 80000 | 14.0 |
| 90000 | 14.0 |
| 100000 | 14.0 |
| 720000 | 13.5 |
| 7200000 | 13.4 |
| 72000000 | 13.3 |
| 720000000 | 13.4 |
FE simulations based on the modeling needs can be done at various time scales (end time) and step times. Although timing would not be a primary concern for a time-independent material model, such as elastic and hyperelastic, it would have a considerable influence in time-dependent material model, such as viscoelastic. In this step, a set of FE simulations were performed using the same patch FE model with optimized viscoelastic parameters extracted from the previously described FE-optimization runs. The patch FE model was subjected to the same loading and boundary conditions as explained before but different simulation times to calculate the stable time that results in the independency of the mechanical response to the shear relaxation modulus of the brain tissue. The FE simulations as well as the corresponding stresses were also reported in Table 2.
Table 2.
The identified material parameters of the human brain tissue and skull (Razaghi et al., 2019; Razaghi et al., 2021).
| Parameters | Short-time shear modulus G0 (kPa) | Long-time shear modulus G∞ (kPa) | Decay constant β (1/s) | Bulk modulus κ (kPa) | Density ρ (kg/m3) |
| Brain tissue | 505 | 110 | 828 | 5625 | 1140 |
| Parameters | Shear Modulus μ (MPa) | Yield stress σy (MPa) | Plastic hardening modulus | Bulk modulus κ (MPa) | Density ρ (kg/m3) |
| Skull | 3470 | 41.80 | 0.0462 | 7120 | 1456 |
2.2. Image acquisition, registration, and segmentation
Pre- and post-resection MRI images of two brain tumor patients (patients #11 & #12) were employed from an online resources at the Montreal Neurological Institute in 2010 (group #3) (Mercier et al., 2012). The segmentation and FE model reconstruction were fully explained in our recent paper (Abdolkarimzadeh et al., 2021). Briefly, the MRI images were first converted into a stack of *JPG images, following by a relative and absolute rotation and/or transformation to fully match the pre- and post-resection images together through affine transformation custom Matlab script. This was a very important stage of the work since the patient may not be able to hold the head at the same spot before and after the tumor resection, so an accurate rotation and/or transformation would be a very critical step for the following FE simulations. An expert neuroradiologist also helped us in addition to the semi-automatic Matlab code to correct the possible mismatches that our code cannot fix since even an insignificant mismatch herein would influence the mass-effect calculation and interpretation of the numerical outcomes.
The brain and the skull were segmented using Mimics (Materialise Inc., Belgium) and the surface meshes were thereafter volume meshed with 10-noded tetrahedral elements (Karimi et al., 2021a; Karimi et al., 2021b; Karimi et al., 2021c), as the 10-noded tetrahedral elements have been resulted the same precision as 20-noded hexahedral elements at considerably a lower computational cost (Karimi et al., 2021b). The FE models of brains after tumor removal from the isometric and top views are presented in Fig. 2. The FE models of the brain #11 and #12 were made of 395,274/550,554 and 144,138/216,166 elements/nodes, respectively. Mesh quality assessment (Ansys, Canonsburg, PA, US) and mesh density analysis (Ansys/LST, Canonsburg, PA, US) were performed to first check the quality of the mesh and then estimate the errors because of element numbers (data are not reported here) (Karimi et al., 2021b). The skull is not represented herein since it would hinder us showing the resultant cavity in the brain due to tumor removal as well as the anatomical location and size of the tumor.
Fig. 2.

The post-resection FE models of (a) patient #11 and (b) patient #12 from the isometric and top views.
A custom Matlab script helped to define the load surface, equivalence the nodes, and write the final LS-Dyna *k file. A 10-core Intel® Xeon® CPU W-2155@3.30 GHz computer with 256GB RAM was used to run the simulations in explicit-dynamic LS-DYNA (Ansys/LST, Canonsburg, PA, US). The simulations were performed in one-step for 80 ms with time steps of 1 ms (80 time steps). Cerebrospinal fluid pressure (CSFP) was applied on the outer surface of the brain from 0–5 ms with the magnitude of −10 mmHg (−1.33 kPa) (Eklund et al., 2016; Turner et al., 1996) that is determined based on the human intracranial pressure in the supine body position. Thereafter, from 5–80 ms the pressure was set to +10 mmHg. The MRI images show the brain under applied CSFP, therefore, a negative CSFP would help to precondition the tissue and thereafter pressure it back to the same load boundary.
2.3. Inverse Dynamic FE-optimization algorithm
A fully coupled FE-optimization algorithm (Abdolkarimzadeh et al., 2021; Rahmati et al., 2021) was employed to optimize the pressure profile at the interface of the brain tissue and tumor. Flow-chart of the applied algorithm is presented in Fig. 3. The position- and time-dependent pressure profiles were the aims of the optimization algorithm while the distance between the brain cavity surface-nodes and the pre-resection tumor surface-nodes was the cost function. When it comes to the FEM, two approaches can be used in terms of the stress calculation, including forward and inverse (Moulton et al., 1995). In the forward approach, the loading and boundary conditions, and the material properties are the known variables while the strain in an element or displacement at the nodes are the unknown variables and need to be calculated using FE solver. However, in the inverse FEM, the loading and boundary conditions as well as the material properties are the unknown parameters while the strain in the elements or nodal displacements are the known parameters. Herein, in the brain FE model, the pressure profile at the interface of the brain tissue and tumor is unknown while we know the nodal coordinates before and after tumor removal that provide us with the nodal displacement. An inverse approach would need more computational hours due to various initial guesses while a forward approach with known parameters would need less hours for the same problem. To solve such an inverse FE model, herein we benefitted from a coupled FE-optimization algorithm (Abdolkarimzadeh et al., 2021; Rahmati et al., 2021). Fminsearch with different initial guesses of P0, P1, P2, P3 and σ for patients #11 and #12 with the upper and lower boundaries of 0 < P0 < 60,0 < P1 < 60,0 < P2 < 60,0 < P3 < 60, and 1 < σ < 10 were set for the pressure profile and was started to run with the cost function of mean squared error. To do that, an initial guess was used to define a 3rd order polynomial function for position-dependent pressure profile and a two-variable (mean and standard deviation) normal distribution function for time-dependent pressure profile. The pressure herein in the optimized one to result in a minimized cost function that is the difference between the nodal coordinates of the reconstructed MRI data and computational data (Fig. 3). The optimized position- and time-dependent pressure profiles include an optimized 3rd-order pressure function and probability density function of normal distribution function that were defined as follows:
| (3) |
| (4) |
where P0, P1, P2, and P3 are the pressure parameters, is the vector of the distance that denotes the normal distance between the center of each element and a surface node on the tumor in Eq. (3), and μ and σ are mean and standard deviation in Eq. (4), respectively. Since the value of pressure increases during the tumor growth, μ should be determined as the end time of the simulation. The other parameters, i.e., σ is computed as an optimization parameter. The final optimized position-time-dependent pressure profile is defined as follows:
| (5) |
Fig. 3.

Flow-chart of the fully coupled inverse FE-optimization algorithm.
The change in time-dependent pressure pattern as a function of σ for a desired time is shown in Fig. 4.
Fig. 4.

The normalized pressure as a function of time for various σ values.
The optimization algorithm of Fminsearch-Unconstrained nonlinear minimization was coupled with the solver of LS-DYNA to calculate the pressure profile parameters (Abdolkarimzadeh et al., 2021). Fminsearch finds the minimum of a scalar function of several parameters, starting at an initial estimate. The coupled FE-optimization algorithm run in Matlab environment using LS-DYNA solver for more than ~90 iterations and the finally converged pressure parameters are listed in Table 2. The cost function was used as follows:
| (6) |
where is the position vector corresponding to the ith node of the brain cavity, pT is the position matrix of all tumor surface-nodes and n is the number of brain cavity surface-nodes. It is worth mentioning that we cannot certify that the optimized pressure profile parameters result in “absolute” minimum for a non-convex cost function. However, herein, the best “possible” minimum was calculated since the solver run more than ~90 iterations for different initial conditions to result in the least possible error so we can say the solution is “consistent” and not a “relative” minimum.
3. Results
The effective stresses at various simulation times in the FE model of the brain patch are listed in Table 1. The optimized pressure profile parameters for both FE cases are also calculated and summarized in Table 2. The contour maps of the tumor surface nodes in the pre- and post-resection before and after loading for both FE cases of #11 and #12 are presented in Fig. 5. The resultant pressure contours in the post-resection tumor surface (brain cavity) after loading for both cases of #11 and #12 are shown in Fig. 5.
The contour maps of displacement in the brain tissues of patients #11 and #12 at various simulation times are shown in Fig. 6. The contour maps of pressure and von Mises stress in the brain tissues of patients #11 and #12 are shown in Figs. 7 & 8.
Fig. 6.

Contour maps of pressure in the post-resection brain cavity in (a) patient #11 and (b) patient #12.
Fig. 7.

The contour maps of displacement in the brain tissue of (a) patient #11 and (b) patient #12 at various simulation times.
Fig. 8.

The contour maps of von Mises stress in the brain tissue of (a) patient #11 and (b) patient #12 at various simulation times.
4. Discussions
Accurate planning and navigation is required to maximize tumor resection is going to be as planned while make sure the nearby healthy tissues will not be damaged. That would require a pressure profile that not only could capture the irregular shape of the tumor during the growth in the brain but also the time-dependent viscoelastic behavior of the brain tissue. However, the currently available approaches whether have failed near the brain due to large deformations and irregular growth pattern of the tumor or have failed to address the time-dependency of the brain biomechanics. In this study, a fully coupled dynamic FE-optimization algorithm (Abdolkarimzadeh et al., 2021; Rahmati et al., 2021) to calculate the position-and time-dependent pressure profile in the brain tumor in post-tumor resection patients.
To calculate the viscoelastic mechanical properties of the brain tissue, a brain patch FE model was constructed and subjected to a compressive uniaxial loading (Fig. 1a). A fully coupled FE-optimization algorithm was employed to calculate the stress-strain in the gauge at the center of the patch FE model and compare it to the experimental data (Karimi et al., 2017b) (Fig. 1b). The calculated viscoelastic mechanical properties were then assigned to the post-resection FE models of the human brain (Fig. 2). A positive pressure was applied to the post-resection brain cavity through a fully coupled inverse FE-optimization algorithm (Fig. 3) to calculate the position- and time-dependent pressure profile at the interface of the brain tissue and the tumor that mimics the tumor growth. The pressure profile proposed in this study benefitted from a time-dependent pressure parameter, σ, which defines the action time that pressure comes into effect on the surface elements of the brain cavity (Fig. 4).
The results revealed a very good fit between the pre-resection surface nodes and that of the post-resection tumor surface after loading (Fig. 5) that implies the accuracy of the proposed position- and time-dependent pressure profile in simulating the tumor growth. The contour maps of the pressure in the brain cavity due to the tumor growth are shown higher magnitudes at the free surfaces where there is a contact with the skull (Fig. 6).
The deformation in the brain tissue was 0.8–1.1 cm (Fig. 7) that is in a good agreement with that of 0.23–1.45 cm (Angeli and Stylianopoulos, 2016). The areas close to the tumor showed the highest deformation while distal regions showed almost no deformation that implies that the biomechanical effects of the tumor growth are mostly local. The mass-effect of 16–44 mm was observed in the regions around the brain cavity (Fig. 7) that is within the range of brain shift due to tumor resection (~ 24 mm) (Bayer et al., 2017; Chen et al., 2011; Gerard et al., 2017; Miga et al., 2016; Nimsky et al., 2000). The accurate prediction of mass-effect depends on the applied pressure boundary since the mass-effect has a pivotal role in defining the premise for justifying surgery of the cerebrum (Bullock et al., 2006; Gonda et al., 2013; McKenna et al., 2012) as well as neurologic diseases, i.e., tumor (Gamburg et al., 2000), stroke (McKenna et al., 2012; Zazulia et al., 1999), and trauma (Bullock et al., 2006; Kim and Gean, 2011).
The brain tissues showed the maximum von Mises stress of 140 kPa (Fig. 8) that is in good agreement with 120 kPa reported by Angeli et al., (Angeli and Stylianopoulos, 2016). In our prior study, the brain tissue was modeled as a homogenous isotropic hyperelastic material (Abdolkarimzadeh et al., 2021) that the tissue mechanical response was entirely time-independent. In this study, the brain tissue was simulated as a time-dependent viscoelastic material model that allows us to have a better understanding of the resultant pressure profile on the brain tissue due to tumor growth. In our prior study, the application of a time-independent hyperelastic model caused an irregular deformation in the brain when the CSFP was applied on the outer surface of the brain FE model (Abdolkarimzadeh et al., 2021). In this study, however, because of the viscoelastic material model, the brain showed stable response and reached a realistic physiological equilibrium before and during the tumor growth. Brain tumors can take anywhere from few months to years to form (Islim et al., 2019; Sallemi et al., 2015; Yamada et al., 2021). Although, to date, based on the current computational resources it is still not plausible to conduct such long FE simulations to predict the tumor growth, it is possible to use time-dependent material models, such as viscoelastic, to “partially” capture the time-dependent mechanical response of the brain tissue due to the tumor growth over a course of reasonable time-scale. The application of time-independent elastic and hyperelastic material models not only cannot address the time-dependent mechanical behavior of the brain tissue but also would result in considerable error in the pressure profile and mass-effect calculations.
A position- and time-dependent pressure boundary would have implications for modeling of structures with unidentified volumetric pressure, i.e., Schlemm’s canal (Maepea and Bill, 1992; Stamer et al., 2015) that the intraocular pressure may vary both as a function of the shape of the canal as well as the aqueous outflow pulse (Johnstone et al., 2021). Understanding of the position- and time-dependent pressure profile in the Schlemm’s canal may contribute in calculating an accurate magnitude of the shear stress applies on the endothelial cells of the Schlemm’s canal (Madekurozwa et al., 2021; McDonnell et al., 2020; Reina-Torres et al., 2021).
The mass-effect plays a key role in defining the premise for justifying surgery of the cerebrum (Bullock et al., 2006; Gonda et al., 2013; McKenna et al., 2012). In neurologic diseases, ranging from trauma (Bullock et al., 2006; Kim and Gean, 2011), stroke (McKenna et al., 2012; Zazulia et al., 1999), to tumor (Gamburg et al., 2000), increased mass-effect is consistently associated with poor prognosis. Our results revealed that considering a variable pressure boundary causes a more realistic pressure contour around the tumor region, so it would result in a more accurate prediction of mass-effect in ventricles. Our results herein showed the mass-effect of ~5–11 mm in the regions around the brain cavity (Fig. 7), which is lower than that of brain shift reported during tumor resection ~24 mm (Bayer et al., 2017; Chen et al., 2011; Gerard et al., 2017; Miga et al., 2016; Nimsky et al., 2000). This difference could be related to the stiffness of the tissues at the vicinity of the tumor that affect the mass-effect. In addition, herein we only modeled two cases while more cases would be required if a wider range of mass-effect estimation is the goal.
This study is limited by the following considerations. First, the quality of the MRI data plays a key role in FE models reconstruction and the following FE simulations. Therefore, even an insignificant mismatch between the pre- and post-resection images would result in error in mass-effect and pressure profile calculation. Herein, the affine transformation was used to register and correct the images in addition to an expert neurologist to minimize any mistakes. Second, the brain tissue in this study was considered to behave as homogenous isotropic viscoelastic, however, it has been well documented that the brain tissue is highly anisotropic having dissimilar material properties for the gray and white matters (Antonovaite et al., 2021; Giordano and Kleiven, 2014; Mijailovic et al., 2021; Velardi et al., 2006). Third, mechanical properties of the human brain tissue in vivo have been poorly understood (Howells et al., 2012; Raboel et al., 2012; Tain and Alperin, 2008); hence, the mechanical properties of a cadaveric human brain tissue were adopted for this study (Karimi et al., 2017b). Fourth, to date, there is no experimentally validated gold standard to compare the computed mass-effect of the tumor versus the experimental data for the pressures and deformations. In this study, since the pre- and post-resection images were provided for the same patient so we could compare the images, calculate the nodal displacement between the two sets of images, and use that as a reference. Finally, herein we used the post-resection data to predict the tumor growth using a position- and time-dependent pressure profile. However, the tumor growth is a forward process as it initiates in the brain and starts to grow by the passage of time. Although this is the real case, for such simulations, we would require several images during the tumor growth and accessing to those data would not be possible unless an accurate follow up. Therefore, in this study, we used an inverse approach and hypothesized that the growth of the tumor in the brain can be inversely simulated.
In summary, herein we proposed a position- and time-dependent pressure profile to address the viscoelastic mechanical behavior of the brain tissue due to tumor growth. The brain tissue has been well shown to have a time-dependent mechanical response under the applied loadings. Therefore, a time-independent pressure profile cannot address the mechanical behavior of the brain tissue and would result in an inaccurate prediction of the brain tumor mass-effect as well as tissue deformation. In this study, the viscoelastic mechanical parameters of the brain tissue were calculated using an inverse FE-optimization algorithm based on the experimental compressive data of the human brain tissue. Several FE simulations were then performed using a patch FE model to calculate the simulation time that results in independency of the tissue mechanical response (shear relaxation modulus) to the time scale of the entire simulation. Finally, the viscoelastic mechanical properties as well as the simulation time were incorporated into the FE model of the post resection human brain with the tumor cavity to inversely calculate the 3rd order position-dependent and normal distribution time-dependent pressure profile at the interface of the brain-tumor. The results revealed that the new proposed pressure profile can well capture the position- and time-dependent mechanical behavior of the brain tissue due to irregular growth of the tumor size and can be considered as a new computational framework for tumor resection surgical planning. The proposed method is computationally inexpensive, robust, accurate, and allows the surgeons to predict the resultant mass-effect before tumor resection, so eventually contributes to better surgical outcomes.
Table 3.
The pressure parameters for patients #11 and #12.
| P0 (kPa) | P1 (kPa) | P2 (kPa) | P3 (kPa) | σ | |
|---|---|---|---|---|---|
|
| |||||
| Patient #11 | |||||
| Pressure parameters | 14.1 | 23 | 10 | 0 | 2 |
|
| |||||
| Patient #12 | |||||
| Pressure parameters | 13.1 | 44 | 1 | 0 | 2.5 |
Funding
This work was supported in part by the National Institutes of Health Grants P30-EY003039 (Bethesda, Maryland); EyeSight Foundation of Alabama (Birmingham, Alabama); and Research to Prevent Blindness (New York, New York).
Footnotes
Declaration of Competing Interest
None declared.
References
- Abdolkarimzadeh F, Ashory MR, Ghasemi-Ghalebahman A, Karimi A, 2021. Inverse dynamic finite element-optimization modeling of the brain tumor mass-effect using a variable pressure boundary. Computer Methods and Programs in Biomedicine 212, 106476, doi: 10.1016/j.cmpb.2021.106476. [DOI] [PubMed] [Google Scholar]
- Aghalari M, Aghagolzadeh A, Ezoji M, 2021. Brain tumor image segmentation via asymmetric/symmetric UNet based on two-pathway-residual blocks. Biomedical Signal Processing and Control 69, 102841, doi:ARTN 102841 10.1016/j.bspc.2021.102841. [DOI] [Google Scholar]
- Angeli S, Stylianopoulos T, 2016. Biphasic modeling of brain tumor biomechanics and response to radiation treatment. J Biomech 49, 1524–1531, doi: 10.1016/j.jbiomech.2016.03.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Antonovaite N, Hulshof LA, Hol EM, Wadman WJ, Iannuzzi D, 2021. Viscoelastic mapping of mouse brain tissue: relation to structure and age. journal of the mechanical behavior of biomedical materials 113, 104159. [DOI] [PubMed] [Google Scholar]
- Bayer S, Maier A, Ostermeier M, Fahrig R, 2017. Intraoperative Imaging Modalities and Compensation for Brain Shift in Tumor Resection Surgery. Int J Biomed Imaging 2017, 6028645, doi: 10.1155/2017/6028645. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bilston LE, Liu Z, Phan-Thien N, 1997. Linear viscoelastic properties of bovine brain tissue in shear. Biorheology 34, 377–85, doi: 10.1016/s0006-355x(98)00022-5. [DOI] [PubMed] [Google Scholar]
- Budday S, Sommer G, Birkl C, Langkammer C, Haybaeck J, Kohnert J, Bauer M, Paulsen F, Steinmann P, Kuhl E, Holzapfel GA, 2017. Mechanical characterization of human brain tissue. Acta Biomater 48, 319–340, doi: 10.1016/j.actbio.2016.10.036. [DOI] [PubMed] [Google Scholar]
- Bullock MR, Chesnut R, Ghajar J, Gordon D, Hartl R, Newell DW, Servadei F, Walters BC, Wilberger J, Surgical Management of Traumatic Brain Injury Author G, 2006. Surgical management of traumatic parenchymal lesions. Neurosurgery 58, S25–46; discussion Si-iv, doi: 10.1227/01.NEU.0000210365.36914.E3. [DOI] [PubMed] [Google Scholar]
- Chen I, Coffey AM, Ding S, Dumpuri P, Dawant BM, Thompson RC, Miga MI, 2011. Intraoperative brain shift compensation: accounting for dural septa. IEEE Trans Biomed Eng 58, 499–508, doi: 10.1109/TBME.2010.2093896. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cheng G, Tse J, Jain RK, Munn LL, 2009. Micro-environmental mechanical stress controls tumor spheroid size and morphology by suppressing proliferation and inducing apoptosis in cancer cells. PLoS One 4, e4632, doi: 10.1371/journal.pone.0004632. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Clatz O, Sermesant M, Bondiau PY, Delingette H, Warfield SK, Malandain G, Ayache N, 2005. Realistic simulation of the 3-D growth of brain tumors in MR images coupling diffusion with biomechanical deformation. IEEE Trans Med Imaging 24, 1334–46, doi: 10.1109/TMI.2005.857217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cuadra MB, Pollo C, Bardera A, Cuisenaire O, Villemure J-G, Thiran J-P, 2004. Atlas-based segmentation of pathological MR brain images using a model of lesion growth. IEEE transactions on medical imaging 23, 1301–1314. [DOI] [PubMed] [Google Scholar]
- Cuadra MB, Gomez J, Hagmann P, Pollo C, Villemure J-G, Dawant BM, Thiran J-P, 2002. Atlas-based segmentation of pathological brains using a model of tumor growth. International Conference on Medical Image Computing and Computer-Assisted Intervention. Springer, pp. 380–387. [Google Scholar]
- Dawant B, Hartmann S, Pan S, Gadamsetty S, 2002. Brain atlas deformation in the presence of small and large space-occupying tumors. Computer Aided Surgery 7, 1–10. [DOI] [PubMed] [Google Scholar]
- Dehoff P, 1978. On the nonlinear viscoelastic behavior of soft biological tissues. Journal of Biomechanics 11, 35–40. [DOI] [PubMed] [Google Scholar]
- Demou ZN, 2010. Gene expression profiles in 3D tumor analogs indicate compressive strain differentially enhances metastatic potential. Ann Biomed Eng 38, 3509–20, doi: 10.1007/s10439-010-0097-0. [DOI] [PubMed] [Google Scholar]
- Donnelly BR, Medige J, 1997. Shear properties of human brain tissue. J Biomech Eng 119, 423–32, doi: 10.1115/1.2798289. [DOI] [PubMed] [Google Scholar]
- Dumpuri P, Thompson RC, Dawant BM, Cao A, Miga MI, 2007. An atlas-based method to compensate for brain shift: preliminary results. Med Image Anal 11, 128–45, doi: 10.1016/j.media.2006.11.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dumpuri P, Thompson RC, Cao AZ, Ding SY, Garg I, Dawant BM, Miga MI, 2010. A Fast and Efficient Method to Compensate for Brain Shift for Tumor Resection Therapies Measured Between Preoperative and Postoperative Tomograms. Ieee Transactions on Biomedical Engineering 57, 1285–1296, doi: 10.1109/Tbme.2009.2039643. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Eichberg DG, Di L, Shah AH, Luther E, Richardson AM, Sarkiss CA, Ivan ME, Komotar RJ, 2019. Brain Tumor Surgery is Safe in Octogenarians and Nonagenarians: A Single-Surgeon 741 Patient Series. World Neurosurgery 132, e185–e192, doi: 10.1016/j.wneu.2019.08.219. [DOI] [PubMed] [Google Scholar]
- Eklund A, Jóhannesson G, Johansson E, Holmlund P, Qvarlander S, Ambarki K, Wåhlin A, Koskinen LOD, Malm J, 2016. The pressure difference between eye and brain changes with posture. Annals of neurology 80, 269–276. [DOI] [PubMed] [Google Scholar]
- Estermann SJ, Pahr DH, Reisinger A, 2022. Material design of soft biological tissue replicas using viscoelastic micromechanical modelling. J Mech Behav Biomed Mater 125, 104875, doi: 10.1016/j.jmbbm.2021.104875. [DOI] [PubMed] [Google Scholar]
- Fallenstein GT, Hulce VD, Melvin JW, 1969. Dynamic mechanical properties of human brain tissue. J Biomech 2, 217–26, doi: 10.1016/0021-9290(69)90079-7. [DOI] [PubMed] [Google Scholar]
- Fan X, Ji S, Fontaine K, Hartov A, Roberts D, Paulsen K, 2011. Simulation of brain tumor resection in image-guided neurosurgery. Medical Imaging 2011: Visualization, Image-Guided Procedures, and Modeling, Vol. 7964. International Society for Optics and Photonics, pp. 79640U. [Google Scholar]
- Finan JD, Sundaresh SN, Elkin BS, McKhann GM, 2nd, Morrison, B., 3rd, 2017. Regional mechanical properties of human brain tissue for computational models of traumatic brain injury. Acta Biomater 55, 333–339, doi: 10.1016/j.actbio.2017.03.037. [DOI] [PubMed] [Google Scholar]
- Forte AE, Gentleman SM, Dini D, 2017. On the characterization of the heterogeneous mechanical response of human brain tissue. Biomech Model Mechanobiol 16, 907–920, doi: 10.1007/s10237-016-0860-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Galford JE, McElhaney JH, 1970. A viscoelastic study of scalp, brain, and dura. J Biomech 3, 211–21, doi: 10.1016/0021-9290(70)90007-2. [DOI] [PubMed] [Google Scholar]
- Gamburg ES, Regine WF, Patchell RA, Strottmann JM, Mohiuddin M, Young AB, 2000. The prognostic significance of midline shift at presentation on survival in patients with glioblastoma multiforme. International Journal of Radiation Oncology Biology Physics 48, 1359–1362, doi:Doi 10.1016/S0360-3016(00)01410-3. [DOI] [PubMed] [Google Scholar]
- Gerard IJ, Kersten-Oertel M, Petrecca K, Sirhan D, Hall JA, Collins DL, 2017. Brain shift in neuronavigation of brain tumors: A review. Med Image Anal 35, 403–420, doi: 10.1016/j.media.2016.08.007. [DOI] [PubMed] [Google Scholar]
- Giordano C, Kleiven S, 2014. Connecting fractional anisotropy from medical images with mechanical anisotropy of a hyperviscoelastic fibre-reinforced constitutive model for brain tissue. J R Soc Interface 11, 20130914, doi: 10.1098/rsif.2013.0914. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gonda DD, Warnke P, Sanai N, Taich Z, Kasper EM, Chen CC, 2013. The value of extended glioblastoma resection: Insights from randomized controlled trials. Surg Neurol Int 4, 110, doi: 10.4103/2152-7806.117173. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hawkins-Daarud A, van der Zee KG, Oden JT, 2012. Numerical simulation of a thermodynamically consistent four-species tumor growth model. Int J Numer Method Biomed Eng 28, 3–24, doi: 10.1002/cnm.1467. [DOI] [PubMed] [Google Scholar]
- Helmlinger G, Netti PA, Lichtenbeld HC, Melder RJ, Jain RK, 1997. Solid stress inhibits the growth of multicellular tumor spheroids. Nat Biotechnol 15, 778–83, doi: 10.1038/nbt0897-778. [DOI] [PubMed] [Google Scholar]
- Herrmann L, 1968. A numerical procedure for viscoelastic stress analysis. Seventh Meeting of ICRPG Mechanical Behavior Working Group, Orlando, FL, 1968. [Google Scholar]
- Hogea C, Davatzikos C, Biros G, 2007. Modeling glioma growth and mass effect in 3D MR images of the brain. International Conference on Medical Image Computing and Computer-Assisted Intervention. Springer, pp. 642–650. [DOI] [PubMed] [Google Scholar]
- Hogea C, Abraham F, Biros G, Davatzikos C, 2006a. Fast solvers for soft tissue simulation with application to construction of brain tumor atlases. submitted to IEEE Transactions on Medical Imaging. [Google Scholar]
- Hogea C, Abraham F, Biros G, Davatzikos C, 2006b. A framework for soft tissue simulations with applications to modeling brain tumor mass-effect in 3-d images. COMPUTATIONAL BIOMECHANICS FOR MEDICINE, 24. [DOI] [PubMed] [Google Scholar]
- Holzapfel GA, 2017. Similarities between soft biological tissues and rubberlike materials. Constitutive models for rubber IV. Routledge, pp. 607–617. [Google Scholar]
- Howells T, Lewen A, Skold MK, Ronne-Engstrom E, Enblad P, 2012. An evaluation of three measures of intracranial compliance in traumatic brain injury patients. Intensive Care Med 38, 1061–8, doi: 10.1007/s00134-012-2571-7. [DOI] [PubMed] [Google Scholar]
- Humphrey JD, 2003. Continuum biomechanics of soft biological tissues. Proceedings of the Royal Society a-Mathematical Physical and Engineering Sciences 459, 3–46, doi: 10.1098/rspa.2002.1060. [DOI] [Google Scholar]
- Hussain S, Anwar SM, Majid M, 2017. Brain tumor segmentation using cascaded deep convolutional neural network. 2017 39th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), pp. 1998–2001. [DOI] [PubMed] [Google Scholar]
- Hussain S, Anwar SM, Majid M, 2018. Segmentation of glioma tumors in brain using deep convolutional neural network. Neurocomputing 282, 248–261, doi: 10.1016/j.neucom.2017.12.032. [DOI] [Google Scholar]
- Islim AI, Mohan M, Moon RDC, Srikandarajah N, Mills SJ, Brodbelt AR, Jenkinson MD, 2019. Incidental intracranial meningiomas: a systematic review and meta-analysis of prognostic factors and outcomes. Journal of Neuro-Oncology 142, 211–221, doi: 10.1007/s11060-019-03104-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jain RK, Martin JD, Stylianopoulos T, 2014. The role of mechanical forces in tumor growth and therapy. Annu Rev Biomed Eng 16, 321–46, doi: 10.1146/annurev-bioeng-071813-105259. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Janet MT, Cheng G, Tyrrell JA, Wilcox-Adelman SA, Boucher Y, Jain RK, Munn LL, 2012. Mechanical compression drives cancer cells toward invasive phenotype. Proceedings of the National Academy of Sciences 109, 911–916. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jemal A, Siegel R, Ward E, Hao Y, Xu J, Murray T, Thun MJ, 2008. Cancer statistics, 2008. CA: a cancer journal for clinicians 58, 71–96. [DOI] [PubMed] [Google Scholar]
- Johnstone M, Xin C, Tan J, Martin E, Wen J, Wang RK, 2021. Aqueous outflow regulation – 21st century concepts. Progress in Retinal and Eye Research 83, 100917, doi: 10.1016/j.preteyeres.2020.100917. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Karimi A, Shojaei A, Razaghi R, 2017a. Viscoelastic mechanical measurement of the healthy and atherosclerotic human coronary arteries using DIC technique. Artery Research 18, 14–21, doi: 10.1016/j.artres.2017.02.004. [DOI] [Google Scholar]
- Karimi A, Rahmati SM, Razaghi R, 2017b. A combination of experimental measurement, constitutive damage model, and diffusion tensor imaging to characterize the mechanical properties of the human brain. Comput Methods Biomech Biomed Engin 20, 1350–1363, doi: 10.1080/10255842.2017.1362694. [DOI] [PubMed] [Google Scholar]
- Karimi A, Razaghi R, Girkin CA, Downs JC, 2021a. Ocular biomechanics due to ground blast reinforcement. Comput Methods Programs Biomed 211, 106425, doi: 10.1016/j.cmpb.2021.106425. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Karimi A, Grytz R, Rahmati SM, Girkin CA, Downs JC, 2021b. Analysis of the effects of finite element type within a 3D biomechanical model of a human optic nerve head and posterior pole. Computer Methods and Programs in Biomedicine 198, 105794, doi: 10.1016/j.cmpb.2020.105794. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Karimi A, Rahmati SM, Grytz RG, Girkin CA, Downs JC, 2021c. Modeling the biomechanics of the lamina cribrosa microstructure in the human eye. Acta Biomater 134, 357–378, doi: 10.1016/j.actbio.2021.07.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kaufman LJ, Brangwynne CP, Kasza KE, Filippidi E, Gordon VD, Deisboeck TS, Weitz DA, 2005. Glioma expansion in collagen I matrices: analyzing collagen concentration-dependent growth and motility patterns. Biophys J 89, 635–50, doi: 10.1529/biophysj.105.061994. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kim JJ, Gean AD, 2011. Imaging for the diagnosis and management of traumatic brain injury. Neurotherapeutics 8, 39–53, doi: 10.1007/s13311-010-0003-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kyriacou SK, Davatzikos C, Zinreich SJ, Bryan RN, 1999. Nonlinear elastic registration of brain images with tumor pathology using a biomechanical model [MRI]. IEEE transactions on medical imaging 18, 580–592. [DOI] [PubMed] [Google Scholar]
- Lloyd BA, Szczerba D, Székely G, 2007. A coupled finite element model of tumor growth and vascularization. International Conference on Medical Image Computing and Computer-Assisted Intervention. Springer, pp. 874–881. [DOI] [PubMed] [Google Scholar]
- Maccabi A, Shin A, Namiri NK, Bajwa N, St John M, Taylor ZD, Grundfest W, Saddik GN, 2018. Quantitative characterization of viscoelastic behavior in tissue-mimicking phantoms and ex vivo animal tissues. PLoS One 13, e0191919, doi: 10.1371/journal.pone.0191919. [DOI] [PMC free article] [PubMed] [Google Scholar]
- MacManus DB, Pierrat B, Murphy JG, Gilchrist MD, 2017. Region and species dependent mechanical properties of adolescent and young adult brain tissue. Sci Rep 7, 13729, doi: 10.1038/s41598-017-13727-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Madekurozwa M, Stamer WD, Reina-Torres E, Sherwood JM, Overby DR, 2021. The ocular pulse decreases aqueous humor outflow resistance by stimulating nitric oxide production. Am J Physiol Cell Physiol 320, C652–C665, doi: 10.1152/ajpcell.00473.2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Maepea O, Bill A, 1992. Pressures in the juxtacanalicular tissue and Schlemm’s canal in monkeys. Exp Eye Res 54, 879–83, doi: 10.1016/0014-4835(92)90151-h. [DOI] [PubMed] [Google Scholar]
- McDonnell F, Perkumas KM, Ashpole NE, Kalnitsky J, Sherwood JM, Overby DR, Stamer WD, 2020. Shear Stress in Schlemm’s Canal as a Sensor of Intraocular Pressure. Sci Rep 10, 5804, doi: 10.1038/s41598-020-62730-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McKenna A, Wilson CF, Caldwell SB, Curran D, 2012. Functional outcomes of decompressive hemicraniectomy following malignant middle cerebral artery infarctions: a systematic review. Br J Neurosurg 26, 310–5, doi: 10.3109/02688697.2012.654835. [DOI] [PubMed] [Google Scholar]
- Mercier L, Del Maestro RF, Petrecca K, Araujo D, Haegelen C, Collins DL, 2012. Online database of clinical MR and ultrasound images of brain tumors. Med Phys 39, 3253–61, doi: 10.1118/1.4709600. [DOI] [PubMed] [Google Scholar]
- Miga MI, Sun K, Chen I, Clements LW, Pheiffer TS, Simpson AL, Thompson RC, 2016. Clinical evaluation of a model-updated image-guidance approach to brain shift compensation: experience in 16 cases. Int J Comput Assist Radiol Surg 11, 1467–74, doi: 10.1007/s11548-015-1295-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mijailovic AS, Galarza S, Raayai-Ardakani S, Birch NP, Schiffman JD, Crosby AJ, Cohen T, Peyton SR, Van Vliet KJ, 2021. Localized characterization of brain tissue mechanical properties by needle induced cavitation rheology and volume controlled cavity expansion. Journal of the Mechanical Behavior of Biomedical Materials 114, 104168. [DOI] [PubMed] [Google Scholar]
- Miller K, Chinzei K, 1997. Constitutive modelling of brain tissue: experiment and theory. J Biomech 30, 1115–21, doi: 10.1016/s0021-9290(97)00092-4. [DOI] [PubMed] [Google Scholar]
- Mohamed A, Davatzikos C, 2005. Finite element modeling of brain tumor mass-effect from 3D medical images. International Conference on Medical Image Computing and Computer-Assisted Intervention. Springer, pp. 400–408. [DOI] [PubMed] [Google Scholar]
- Mohamed A, Kyriacou SK, Davatzikos C, 2001. A statistical approach for estimating brain tumor induced deformation. Proceedings IEEE Workshop on Mathematical Methods in Biomedical Image Analysis (MMBIA 2001). IEEE, pp. 52–59. [Google Scholar]
- Mokri B, 2001. The Monro-Kellie hypothesis: applications in CSF volume depletion. Neurology 56, 1746–8, doi: 10.1212/wnl.56.12.1746. [DOI] [PubMed] [Google Scholar]
- Moulton MJ, Creswell LL, Actis RL, Myers KW, Vannier MW, Szabo BA, Pasque MK, 1995. An inverse approach to determining myocardial material properties. Journal of Biomechanics 28, 935–948, doi: 10.1016/0021-9290(94)00144-S. [DOI] [PubMed] [Google Scholar]
- Nimsky C, Ganslandt O, Von Keller B, Romstock J, Fahlbusch R, 2004. Intraoperative high-field-strength MR imaging: implementation and experience in 200 patients. Radiology 233, 67–78, doi: 10.1148/radiol.2331031352. [DOI] [PubMed] [Google Scholar]
- Nimsky C, Ganslandt O, Cerny S, Hastreiter P, Greiner G, Fahlbusch R, 2000. Quantification of, visualization of, and compensation for brain shift using intraoperative magnetic resonance imaging. Neurosurgery 47, 1070–1079, doi:Doi 10.1097/00006123-200011000-00008. [DOI] [PubMed] [Google Scholar]
- Orringer DA, Golby A, Jolesz F, 2012. Neuronavigation in the surgical management of brain tumors: current and future trends. Expert Rev Med Devices 9, 491–500, doi: 10.1586/erd.12.42. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Prange MT, Margulies SS, 2002. Regional, directional, and age-dependent properties of the brain undergoing large deformation. J Biomech Eng 124, 244–52, doi: 10.1115/1.1449907. [DOI] [PubMed] [Google Scholar]
- Qiu S, Jiang W, Alam MS, Chen S, Lai C, Wang T, Li X, Liu J, Gao M, Tang Y, Li X, Zeng J, Feng Y, 2020. Viscoelastic characterization of injured brain tissue after controlled cortical impact (CCI) using a mouse model. J Neurosci Methods 330, 108463, doi: 10.1016/j.jneumeth.2019.108463. [DOI] [PubMed] [Google Scholar]
- Raboel PH, Bartek J Jr., Andresen M, Bellander BM, Romner B, 2012. Intracranial Pressure Monitoring: Invasive versus Non-Invasive Methods-A Review. Crit Care Res Pract 2012, 950393, doi: 10.1155/2012/950393. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rahmati SM, Razaghi R, Karimi A, 2021. Biomechanics of the keratoconic cornea: Theory, segmentation, pressure distribution, and coupled FE-optimization algorithm. J Mech Behav Biomed Mater 113, 104155, doi: 10.1016/j.jmbbm.2020.104155. [DOI] [PubMed] [Google Scholar]
- Rashid B, Destrade M, Gilchrist MD, 2012. Mechanical characterization of brain tissue in compression at dynamic strain rates. Journal of the Mechanical Behavior of Biomedical Materials 10, 23–38, doi: 10.1016/j.jmbbm.2012.01.022. [DOI] [PubMed] [Google Scholar]
- Razaghi R, Biglari H, Karimi A, 2019. Risk of rupture of the cerebral aneurysm in relation to traumatic brain injury using a patient-specific fluid-structure interaction model. Computer Methods and Programs in Biomedicine 176, 9–16, doi: 10.1016/j.cmpb.2019.04.015. [DOI] [PubMed] [Google Scholar]
- Razaghi R, Biglari H, Karimi A, 2021. A patient-specific fluid-structure interaction model of the cerebrovascular damage in relation to traumatic brain injury. Trauma-England 23, 33–43, doi:Artn 1460408620921729 10.1177/1460408620921729. [DOI] [PubMed] [Google Scholar]
- Reina-Torres E, De Ieso ML, Pasquale LR, Madekurozwa M, van Batenburg-Sherwood J, Overby DR, Stamer WD, 2021. The vital role for nitric oxide in intraocular pressure homeostasis. Prog Retin Eye Res 83, 100922, doi: 10.1016/j.preteyeres.2020.100922. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ropper AH, 1986. Lateral displacement of the brain and level of consciousness in patients with an acute hemispheral mass. N Engl J Med 314, 953–8, doi: 10.1056/NEJM198604103141504. [DOI] [PubMed] [Google Scholar]
- Sallemi L, Njeh I, Lehericy S, 2015. Towards a computer aided prognosis for brain glioblastomas tumor growth estimation. IEEE transactions on nanobioscience 14, 727–733. [DOI] [PubMed] [Google Scholar]
- Schmidt DA, Shi C, Berry RA, Honig ML, Utschick W, 2009. Minimum mean squared error interference alignment. 2009 Conference Record of the Forty-Third Asilomar Conference on Signals, Systems and Computers. IEEE, pp. 1106–1110. [Google Scholar]
- Stamer WD, Braakman ST, Zhou EH, Ethier CR, Fredberg JJ, Overby DR, Johnson M, 2015. Biomechanics of Schlemm’s canal endothelium and intraocular pressure reduction. Prog Retin Eye Res 44, 86–98, doi: 10.1016/j.preteyeres.2014.08.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stylianopoulos T, Martin JD, Chauhan VP, Jain SR, Diop-Frimpong B, Bardeesy N, Smith BL, Ferrone CR, Hornicek FJ, Boucher Y, Munn LL, Jain RK, 2012. Causes, consequences, and remedies for growth-induced solid stress in murine and human tumors. Proc Natl Acad Sci U S A 109, 15101–8, doi: 10.1073/pnas.1213353109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sun K, Pheiffer TS, Simpson AL, Weis JA, Thompson RC, Miga MI, 2014. Near real-time computer assisted surgery for brain shift correction using biomechanical models. IEEE journal of translational engineering in health and medicine 2, 1–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sundaresh SN, Finan JD, Elkin BS, Lee C, Xiao J, Morrison III B, 2021. Viscoelastic characterization of porcine brain tissue mechanical properties under indentation loading. Brain Multiphysics, 100041. [Google Scholar]
- Tain R-W, Alperin N, 2008. Noninvasive intracranial compliance from MRI-based measurements of transcranial blood and CSF flows: indirect versus direct approach. IEEE transactions on biomedical engineering 56, 544–551. [DOI] [PubMed] [Google Scholar]
- Turner CR, Losasso TJ, Muzzi DA, Weglinski MR, 1996. Brain relaxation and cerebrospinal fluid pressure during craniotomy for resection of supratentorial mass lesions. J Neurosurg Anesthesiol 8, 126–32, doi: 10.1097/00008506-199604000-00005. [DOI] [PubMed] [Google Scholar]
- Turquier F, Kang HS, Trosseille X, Willinger R, Lavaste F, Tarriere C, Dömont A, 1996. Validation study of a 3D finite element head model against experimental data. SAE transactions, 1912–1923. [Google Scholar]
- Velardi F, Fraternali F, Angelillo M, 2006. Anisotropic constitutive equations and experimental tensile behavior of brain tissue. Biomech Model Mechanobiol 5, 53–61, doi: 10.1007/s10237-005-0007-9. [DOI] [PubMed] [Google Scholar]
- Wasserman R, Acharya R, 1996. A patient-specific in vivo tumor model. Math Biosci 136, 111–40, doi: 10.1016/0025-5564(96)00045-4. [DOI] [PubMed] [Google Scholar]
- Yamada S, Kinoshita M, Nakagawa T, Hirayama R, Kijima N, Kagawa N, Kishima H, 2021. The Impact of 5-Year Tumor Doubling Time to Predict the Subsequent Long-Term Natural History of Asymptomatic Meningiomas. World Neurosurgery 151, e943–e949, doi: 10.1016/j.wneu.2021.05.023. [DOI] [PubMed] [Google Scholar]
- Yu Y, Bourantas G, Zwick B, Joldes G, Kapur T, Frisken S, Kikinis R, Nabavi A, Golby A, Wittek A, Miller K, 2022. Computer simulation of tumour resection-induced brain deformation by a meshless approach. Int J Numer Method Biomed Eng 38, e3539, doi: 10.1002/cnm.3539. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zacharaki EI, Hogea CS, Biros G, Davatzikos C, 2008. A comparative study of biomechanical simulators in deformable registration of brain tumor images. IEEE Trans Biomed Eng 55, 1233–6, doi: 10.1109/TBME.2007.905484. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zazulia AR, Diringer MN, Derdeyn CP, Powers WJ, 1999. Progression of mass effect after intracerebral hemorrhage. Stroke 30, 1167–73, doi: 10.1161/01.str.30.6.1167. [DOI] [PubMed] [Google Scholar]
- Zhang LY, Yang KH, King AI, 2001. Comparison of brain responses between frontal and lateral impacts by finite element modeling. Journal of Neurotrauma 18, 21–30, doi:Doi 10.1089/089771501750055749. [DOI] [PubMed] [Google Scholar]
