. 2021;16(2):103–108. doi: 10.22037/iej.v16i2.30586

Shaping Ability of ProTaper Next, Hyflex CM, and V-Taper 2H Nickel-Titanium Files in Mandibular Molars: A Micro-computed Tomographic Study

Mateusz Radwański a,*, Michał Łęski a, Adam K Puszkarz b, Izabella Krucińska b
PMCID: PMC9709896  PMID: 36704222

Abstract

Introduction:

Although micro-computed tomography (MCT) evaluation of the shaping ability of ProTaper Next (PTN) and Hyflex CM (HCM) files has been reported, to our knowledge, no study has assessed the performance of V-Taper 2H (VT) files. The aim of this study was to evaluate and compare the shaping ability of PTN, HCM, and VT systems in the mesial canals of mandibular molars using MCT.

Methods and Materials:

Thirty extracted first and second mandibular molars were scanned using MCT and randomly assigned to HCM, PTN, and VT groups. Images obtained before and after preparation were evaluated for the increase in the root canal volume, untouched surface area, and amount of accumulated hard tissue debris. One-way analysis of variance (ANOVA) and Kruskal-Wallis test were used to compare the variables in the groups (α=5%).

Results:

There were no statistically significant between-group differences in the postoperative measurements (P>0.05). The canal volume increased in all three groups: PTN (73.84%), VT (73.48%), and HCM (49.29%). The largest and smallest untouched areas were observed in the PTN (41.37%) and VT (30.85%) groups, respectively (P>0.05). The debris formed during canal preparation was 1.84%, 2.16%, and 2.42% in the VT, PTN, and HCM groups, respectively (P>0.05).

Conclusions:

Based on our in vitro study, the PTN, HCM, and VT systems showed similar shaping abilities. None of the tested canals were completely free from debris, while the untouched surface area was considerably large. The VT system had the most favorable results with the smallest untouched surface area and least debris were. We would recommend further trials to endorse these findings.

Key Words: Micro-computed Tomography, Nickel-Titanium Instruments, Root Canal Preparation, Rotary Instrumentation

Introduction

Cleaning and shaping of the root canal system are crucial steps during endodontic treatment. Chemo-mechanical preparation with preservation of the original canal anatomy, without iatrogenic events, allows adequate irrigation and canal obturation [1-3]. The introduction of rotary nickel–titanium (NiTi) files has reduced the working time and minimized the risk of procedural errors. Over the years, many rotary systems have been introduced, with files showing variations in the taper, cross-section of the working part, shape of the tip and cutting edges, composition of the NiTi alloy, and manufacturing process [4].

However, most of the available rotary systems, both continuous and reciprocating, have failed to achieve complete debridement of the canal space. They tend to leave areas of untouched walls and accumulated hard tissue debris in irregularities, which lead to residual bacteria and further infection [5]. Thus, the optimal preparation of root canals remains a challenge.

The ProTaper Next (PTN) (PTN, Dentsply Sirona, Ballaigues, Switzerland) system comprises five files (X1-X5) made from M-Wire, a metallurgically improved version of NiTi. The instruments are characterized by a variable taper on a single file and an offset mass of rotation [6]. These features serve to decrease engagement and increase flexibility and the capacity to extract debris from canals without changing the shaping ability [7].

Hyflex CM (HCM) files (Coltene-Whaledent, Allstätten, Switzerland) are manufactured from a NiTi alloy with a lower percentage weight of nickel (52%) [8]. The files can be precurved before use and are more flexible. HCM files respond to pressure by lengthening their spirals, which prevents binding of the files to the walls. Heat treatment (e.g. during autoclaving) enables recovery of the original shape and reuse of the instruments. The basic set includes six NiTi instruments that are available in two lengths: 21 and 25 mm. Only the orifice opener (25/0.08) is 19 mm long [9].

V-Taper 2H (VT) files (SS White Dental Inc., Lakewood, NJ, USA) are made from CM-wire and exhibit a noncutting tip, variable taper, and variable pitch design. The parabolic core design enables high flexibility for the successful negotiation of curved canals and strengthens the files against separation [10]. The files can selectively remove dentin, which leads to the preservation of more tissue during the procedure [16]. The “VT Standard Kit” comprises six files.

Several methodologies have been used for the evaluation of the shaping ability of endodontic instruments. Among these, micro-computed tomography (MCT) has become the gold standard and shown to be an accurate method for the evaluation of the three-dimensional (3D) aspects of root canal shaping, such as the increase in the root canal volume, untouched surface area, and amount of accumulated hard tissue debris [5, 11]. Although MCT evaluation of the shaping ability of PTN [7, 12-14] and HCM [9] files has been reported, to our knowledge, no study has assessed the performance and effectiveness of VT files. The aim of this study was to evaluate and compare the shaping ability of PTN, HCM, and VT systems in the mesial canals of mandibular molars using MCT. The null hypothesis was that there is no difference between the three NiTi systems in terms of the 3D shaping parameters.

Materials and Methods

Sample selection

After local research ethics committee approval (RNN/35/20/KE), we selected 30 freshly extracted first and second mandibular molars and stored them in a 0.5% thymol solution before the experiment. A sample size calculation was conducted in each group and was considered an alpha of 5% and power of 80% or upper, which resulted in five samples. To make our results stronger, we have chosen 10 samples for each group.

The inclusion criteria were fully formed roots, curvatures between 10° and 30° [15], and an initial foramen diameter equivalent to that of a size 10 K-file (Dentsply Sirona Endodontics, Ballaigues, Switzerland).

After endodontic access, we inserted a size 10 K-file (Dentsply Sirona Endodontics, Ballaigues, Switzerland) into the mesial canals until its tip could be seen to exit the apical foramen. The working length (WL) was established 1 mm short of the apical foramen. After sectioning the distal root of the specimen, we scaled the surface of the mesial root with periodontal curettes.

Pre-instrumentation MCT images

Before instrumentation, we scanned the mesial roots of all specimens using MCT (SkyScan 1272; Bruker, Kontich, Belgium) under the following scanning conditions: X-ray source voltage, 90 kV; X-ray source current, 111 µA; and pixel size, 21 µm. Rotation of 180° was performed with a rotation step of 0.5°, and a copper (0.038 mm)+aluminum (0.5 mm) filter was selected. Before instrumentation, the initial volume (mm3), volume of dentine (mm3), and surface area of dentine (mm2) were calculated, and the samples were randomly divided into three groups including PTN, HCM, and VT groups (n=10).

Root canal preparation

The same experienced operator instrumented the canals by using the X-smart Endodontic Motor (Dentsply Sirona Endodontics, Ballaigues, Switzerland) according to the manufacturer’s instructions. In all groups, a glide path was created by scouting the canals to an apical size of 15 using C-pilot (VDW, Munich, Germany). Apical patency was controlled with a size 10 K-file (Dentsply Sirona Endodontics, Ballaigues, Switzerland).

PTN group

The PTN files were used with a constant speed of 300 revolutions per min (rpm) and torque set up to 2 N/cm. The sequence was as follows: X1 (size 17/0.04 taper), X2 (size 25/0.06 taper), and X3 (size 30/0.07 taper) up to the full WL.

HCM group

The HCM files were used with a constant speed of 500 rpm and torque set up to 2.5 N/cm. The operating sequence was as follows: 25/0.08 up to one third of WL, followed by 20/0.04, 25/0.04, 20/0.06, and 30/0.04 up to the full WL.

VT group

VT files were used with a constant speed of 300 rpm and torque set up to 3.0 N/cm. The following sequence was used: 17/0.04, 20/0.06, 25/0.06, and 30/0.06 up to the full WL.

Figure 1.

Figure 1

Three-dimensional reconstruction of canals before (red) and after (green) preparation with the different systems. The red color on the superimposed images represents the untouched surface area. I: ProTaper Next, II: Hyflex CM, III: V-Taper 2H

In each group, we prepared the canals to an apical size of 30. Each set of instruments was used to enlarge two root canals in one root and discarded. After the use of each instrument, we irrigated the canals with 1 mL of 5.25% NaOCl. The final irrigation protocol was standardized for every canal in each group: 10 mL of 5.25% of NaOCl, followed by 5 mL of 17% EDTA for 3 min and, finally, 5 mL of 0.9% saline. We irrigated the canals using a 5-mL disposable plastic syringe with a 30-gauge Endo-Eze Tip (Ultradent, South Jordan, UT, USA) that was passively placed without binding in the canal, up to 2 mm from the apical foramen. Finally, we dried the canals with absorbent paper points. After root canal preparation, postoperative MCT scans of each specimen were acquired using the previously described parameters.

MCT evaluation

The images were reconstructed using NRecon v.1.6.0 software (Bruker, Kontich, Belgium), and the parameters were calculated using CTAn v1.14.4 software (Bruker, Kontich, Belgium). We established a region of interest (ROI) from the furcation level to the apex of the root. 3D visualization was achieved using CTvol v2.3.2.0 software (Bruker, Kontich, Belgium).

Statistical analysis

All data are presented as mean values with standard deviations. Normality testing was performed using the Shapiro–Wilk test. One-way analysis of variance (one-way ANOVA) was used for comparison of the percentage of untouched surface area among the three groups, while the percentage increase in the canal volume and the percentage of accumulated debris were compared using the Kruskal–Wallis test. A P-value less than 0.05 was considered statistically significant. All calculations and statistical analyses were performed using SPSS Statistics version 25 (IBM, IL, USA).

Results

Statistical analysis [Shapiro-Wilk, and one-way ANOVA] revealed that there were no significant differences in the evaluated parameters among the three groups before preparation (P>0.05; Table 1). The results of 3D analysis are presented in Table 2. There were no significant differences among the three NiTi systems in terms of the postoperative 3D measurements including increase in the root canal volume, untouched canal surface area, and accumulated hard tissue debris (P>0.05). The 3D reconstruction of representative specimens in each group is shown in Figure 1.

Discussion

The results of this study indicated no statistically differences among the PTN, VT, and HCM systems in terms of all 3D parameters related to their shaping ability. Thus, the null hypothesis was accepted. Two experimental models are most frequently used for the evaluation of root canal preparations: simulated root canals in resin blocks or root canals in extracted human teeth [9, 16]. Even though resin blocks ensure the standardization of evaluated parameters, we chosed human teeth for the present study because they reflect more realistic clinical conditions due to the properties of dentine [17].

In the present study, the following 3D parameters were assessed using MCT: increase in the canal volume, untouched surface area, and the amount of hard tissue debris formed during instrumentation.

Regarding the canal volume, there was an increase in all three systems: PTN, 73.84%; VT, 73.48%; and HCM, 49.29%. The size of the file tip was the same in all three groups (size 30); thus, the increase in volume was most likely affected by the file taper. We surmised that a smaller diameter increases for every millimeter of the working part in the specimens prepared with HCM files (0.04) resulted in a smaller change in the canal volume.

Assessment of the untouched surface area and the amount of hard tissue debris accumulated during root canal shaping is extremely important in the evaluation of the long-term results of endodontic treatment, because debris and untreated surfaces could serve as a scaffold for residual bacteria and their growth, thus contributing to the complications after treatment [7, 12, 18, 19].

Despite continuous technological progress, it has not been possible to introduce a NiTi system that prepares most of the canal surfaces. In previous studies, the percentage area of untreated canal walls ranged from 9.6% to 79% [9, 20-22]. These differences are primarily attributed to the anatomy of the tested roots. The mandibular molars are most used tooth. The complexity of the root canal anatomy and the presence of structural irregularities (i.e., isthmuses, anastomoses, and additional canals) could contribute to an increase in the untouched surface area and may be the reason for more accumulated debris during shaping, which cannot be removed from such areas. It should be emphasized that the cross-section of root canals is rarely round, and their shape enables preparation with rotary files. In the present study, none of the systems exhibited an effective shaping ability. The largest and smallest untouched areas were observed in the PTN (41.37%)

and VT (30.85%) groups, respectively. In general, files with a greater taper should prepare a larger surface area [4]; this was not observed in the present study. The PTN files, which have the maximum taper (0.07), left the largest area unprepared. This result could be explained by the largest pre-instrumentation surface area of the canals in the PTN group (83.97 mm2), which probably resulted in more canal irregularities and a larger untouched area during shaping. Our observations are consistent with those of Paqué et al. [23] and Drukteinis et al. [7], who concluded that the degree of taper was not the most important factor for determining the ability to prepare a larger surface area. In addition, Gagliardi et al. [24] observed a significantly larger percentage of untouched area with the PTN system than with the ProTaper Gold and ProTaper Universal systems; this was because of the reduced contact area in the canal wall.

Table 1.

Evaluation of preoperative (A) and postoperative (B) parameters between the tested systems; the data are presented as mean (SD)

System P -value
ProTaper Next (I) Hyflex CM (II) V-Taper 2H (III)
Dentine (mm 3 ) A 206.42 (24.68) 212.13 (28.50) 208.35 (43.06) 0.925
B 197.14 (23.16) 207.52 (30.86) 198.23 (40.93)
Volume (mm 3 ) A 7.92 (2.43) 6.17 (2.43) 5.54 (2.76) 0.115
B 13.76 (2.74) 9.22 (1.93) 9.61 (2.93)
Surface area (mm 2 ) A 83.97 (24.09) 75.03 (25.77) 63.82 (11.90) 0.130
B 94.05 (14.56) 81.03 (14.45) 70.84 (14.34)

Table 2.

Mean values and standard deviations for the percentage increase in the canal volume, percentage of untouched surface area, and percentage of debris formed during canal preparation with the tested systems

System P -value
ProTaper Next (I) Hyflex CM (II) V-Taper 2H (III)
% Increase in the canal volume 73.84 (50.57) 49.29 (31.71) 73.48 (81.27) 0.576
% Untouched surface area 41.37 (12.84) 31.67 (12.16) 30.85 (19.83) 0.251
% Of debris 2.16 (0.44) 2.42 (0.79) 1.84 (0.33) 0.116

The debris formed during canal preparation was reported as 1.84% for VT files, 2.16% for PTN files, and 2.42% for HCM files. Lopes et al. [21] noticed a smaller amount of debris after preparation with PTN files (0.10%). According to the authors, the difference between systems may result from the irrigation protocol used during shaping, such as the total volume of rinsing solution per canal, time of irrigation, and positioning of the irrigation needle. In the literature, the taper, final tip size, and cross-section of the files used for instrumentation are considered factors influencing the amount of debris. Shaheen [25] concluded that greater accumulation of debris in the canal, particularly in the apical part, may be affected by greater taper of the files. The author observed a larger amount of debris after shaping with WaveOne files (0.08) compared with One Shape files (0.06). In addition, he noticed that the offset design of the file, which provides a larger space for chips, contributes to lesser accumulation of residue and its transport in the coronal direction. These findings are not consistent with those in our study, where HCM files with a 0.04 taper resulted in greater debris accumulation (2.42%) than did PTN and HCM files with 0.07 and 0.06 tapers, respectively. The offset design of PTN files does not maximize debris elimination from canals [21]. Xu et al. [19] compared the efficacy of hard-tissue debris removal from the mandibular first molars prepared to various apical sizes and summarized that a greater final apical size resulted in significantly lower debris accumulation. It should be considered that root canal preparation using NiTi files with a large taper (>.06) might reduce the resistance of roots and increase the risk of fracture [26].

Conclusions

Our in vitro study showed similar shaping abilities for the PTN, HCM, and VT rotary systems. None of the tested canals were completely free from debris, while the untouched surface area was considerably large. It seems that other measures such as irrigation and use of intracanal dressings are necessary to achieve the optimal chemo-mechanical canal preparation and disinfection.

Conflict of Interest:

‘None declared’.

Acknowledgments

The study was funded by Medical University of Lodz, Poland (502-03/2-044-02/502-24-069-18). This study was partially financed from funds: I42/501/14-148-1-19 statutory activity by the Lodz University of Technology, Institute of Material Science of Textiles and Polymer Composites, and the “Innovative Textiles 2020+” investment project within the Regional Operational Program for Lodzkie 2014-2020.

References

  • 1.Hülsmann M, Peters OA, Dummer PMH. Mechanical preparation of root canals: shaping goals, techniques and means. Endodontic Topics. 2005;10(1):30–76. [Google Scholar]
  • 2.Shen Y, Zhou HM, Zheng YF, Peng B, Haapasalo M. Current challenges and concepts of the thermomechanical treatment of nickel-titanium instruments. J Endod. 2013;39(2):163–72. doi: 10.1016/j.joen.2012.11.005. [DOI] [PubMed] [Google Scholar]
  • 3.Santa-Rosa J, de Sousa-Neto MD, Versiani MA, Nevares G, Xavier F, Romeiro K, Cassimiro M, Leoni GB, de Menezes RF, Albuquerque D. Shaping Ability of Single-file Systems with Different Movements: A Micro-computed Tomographic Study. Iran Endod J. 2016;11(3):228–33. doi: 10.7508/iej.2016.03.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Haapasalo M, Shen Y. Evolution of nickel–titanium instruments: from past to future. Endodontic Topics. 2013;29(1):3–17. [Google Scholar]
  • 5.Versiani MA, Carvalho KKT, Mazzi-Chaves JF, Sousa-Neto MD. Micro-computed Tomographic Evaluation of the Shaping Ability of XP-endo Shaper, iRaCe, and EdgeFile Systems in Long Oval-shaped Canals. Journal of Endodontics. 2018;44(3):489–95. doi: 10.1016/j.joen.2017.09.008. [DOI] [PubMed] [Google Scholar]
  • 6.Neto F, Ginjeira A. Comparative analysis of simulated root canals shaping, using ProTaper Universal, Next and Gold. Revista Portuguesa de Estomatologia, Medicina Dentária e Cirurgia Maxilofacial. 2016;57(2):82–6. [Google Scholar]
  • 7.Drukteinis S, Peciuliene V, Dummer PMH, Hupp J. Shaping ability of BioRace, ProTaper NEXT and Genius nickel-titanium instruments in curved canals of mandibular molars: a MicroCT study. Int Endod J. 2019;52(1):86–93. doi: 10.1111/iej.12961. [DOI] [PubMed] [Google Scholar]
  • 8.Gambarini G, Gerosa R, De Luca M, Garala M, Testarelli L. Mechanical properties of a new and improved nickel-titanium alloy for endodontic use: an evaluation of file flexibility. Oral Sur, Oral Med, Oral Pathol, Oral Radiol Endod. 2008;105(6):798–800. doi: 10.1016/j.tripleo.2008.02.017. [DOI] [PubMed] [Google Scholar]
  • 9.Zhao D, Shen Y, Peng B, Haapasalo M. Micro-computed tomography evaluation of the preparation of mesiobuccal root canals in maxillary first molars with Hyflex CM, Twisted Files, and K3 instruments. J Endod. 2013;39(3):385–8. doi: 10.1016/j.joen.2012.11.030. [DOI] [PubMed] [Google Scholar]
  • 10.Shenoi PR, Luniya DA, Badole GP, Makade CS, Kubde R, Khode RT. Comparative evaluation of shaping ability of V-Taper 2H, ProTaper Next, and HyFlex CM in curved canals using cone-beam computed tomography: An. Indian J Dent Res. 2017;28(2):181–6. doi: 10.4103/ijdr.IJDR_34_16. [DOI] [PubMed] [Google Scholar]
  • 11.Marques da Silva B, Scaini F, Tomazinho FSF, Gonzaga CC, Leão Gabardo MC, Baratto-Filho F. Root Preparation of Deciduous Teeth: Efficacy of WaveOne and ProTaper Systems with and without Passive Ultrasonic Irrigation. Iran Endod J. 2018;13(3):362–6. doi: 10.22037/iej.v13i3.17094. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Stringheta CP, Bueno CES, Kato AS, Freire LG, Iglecias EF, Santos M, Pelegrine RA. Micro-computed tomographic evaluation of the shaping ability of four instrumentation systems in curved root canals. Int Endod J. 2019;52(6):908–16. doi: 10.1111/iej.13084. [DOI] [PubMed] [Google Scholar]
  • 13.Turkistani AK, Gomaa MM, Shafei LA, Alsofi L, Majeed A, AlShwaimi E. Shaping Ability of HyFlex EDM and ProTaper Next Rotary Instruments in Curved Root Canals: A Micro-CT Study. J Contemp Dent Pract. 2019;20(6):680–5. [PubMed] [Google Scholar]
  • 14.Cassimiro M, Romeiro K, Gominho L, de Almeida A, Silva L, Albuquerque D. Effects of Reciproc, ProTaper Next and WaveOne Gold on Root Canal Walls: A Stereomicroscope Analysis. Iran Endod J. 2018;13(2):228–33. doi: 10.22037/iej.v13i2.16327. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971;32(2):271–5. doi: 10.1016/0030-4220(71)90230-1. [DOI] [PubMed] [Google Scholar]
  • 16.Gu Y, Kum K-Y, Perinpanayagam H, Kim C, Kum DJ, Lim S-M, Chang S-W, Baek S-H, Zhu Q, Yoo Y-J. Various heat-treated nickel–titanium rotary instruments evaluated in S-shaped simulated resin canals. J Dent Sci. 2017;12(1):14–20. doi: 10.1016/j.jds.2016.04.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Setzer FC, Kwon TK, Karabucak B. Comparison of apical transportation between two rotary file systems and two hybrid rotary instrumentation sequences. J Endod. 2010;36(7):1226–9. doi: 10.1016/j.joen.2010.03.011. [DOI] [PubMed] [Google Scholar]
  • 18.Kyaw Moe MM, Ha JH, Jin MU, Kim YK, Kim SK. Root Canal Shaping Effect of Instruments with Offset Mass of Rotation in the Mandibular First Molar: A Micro-computed Tomographic Study. Journal of Endodontics. 2018;44(5):822–7. doi: 10.1016/j.joen.2017.11.012. [DOI] [PubMed] [Google Scholar]
  • 19.Xu K, Wang J, Wang K, Gen N, Li J. Micro-computed tomographic evaluation of the effect of the final apical size prepared by rotary nickel-titanium files on the removal efficacy of hard-tissue debris. J Int Med Res. 2018;46(6):2219–29. doi: 10.1177/0300060518757607. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Alves FR, Almeida BM, Neves MA, Moreno JO, Rôças IN, Siqueira JF. Disinfecting oval-shaped root canals: effectiveness of different supplementary approaches. J Endod. 2011;37(4):496–501. doi: 10.1016/j.joen.2010.12.008. [DOI] [PubMed] [Google Scholar]
  • 21.Lopes RMV, Marins FC, Belladonna FG, Souza EM, De-Deus G, Lopes RT, Silva EJNL. Untouched canal areas and debris accumulation after root canal preparation with rotary and adaptive systems. Aust Endod J. 2018;44(3):260–6. doi: 10.1111/aej.12237. [DOI] [PubMed] [Google Scholar]
  • 22.Siqueira JF, Alves FR, Versiani MA, Rôças IN, Almeida BM, Neves MA, Sousa-Neto MD. Correlative bacteriologic and micro-computed tomographic analysis of mandibular molar mesial canals prepared by self-adjusting file, reciproc, and twisted file systems. J Endod. 2013;39(8):1044–50. doi: 10.1016/j.joen.2013.04.034. [DOI] [PubMed] [Google Scholar]
  • 23.Paqué F, Peters OA. Micro-computed tomography evaluation of the preparation of long oval root canals in mandibular molars with the self-adjusting file. J Endod. 2011;37(4):517–21. doi: 10.1016/j.joen.2010.12.011. [DOI] [PubMed] [Google Scholar]
  • 24.Gagliardi J, Versiani MA, de Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the Shaping Characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in Curved Canals. J Endod. 2015;41(10):1718–24. doi: 10.1016/j.joen.2015.07.009. [DOI] [PubMed] [Google Scholar]
  • 25.Shaheen N. The efficacy of different single file systems in cleaning oval shaped root canal. Tanta Dental Journal. 2019;16(2):73–9. [Google Scholar]
  • 26.Bier CA, Shemesh H, Tanomaru-Filho M, Wesselink PR, Wu MK. The ability of different nickel-titanium rotary instruments to induce dentinal damage during canal preparation. J Endod. 2009;35(2):236–8. doi: 10.1016/j.joen.2008.10.021. [DOI] [PubMed] [Google Scholar]

Articles from Iranian Endodontic Journal are provided here courtesy of Iranian Center for Endodontic Research

RESOURCES