Study Protocol COVER-ALL: Clinical impact of a volumetric image method for confirming tumour coverage with ablation on patients with malignant liver lesions

Yuan-Mao Lin 1, Iwan Paolucci 1, Brian M Anderson 2, Caleb S O’Connor 2, Bastien Rigaud 2, Maria Briones-Dimayuga 1, Kyle A Jones 2, Kristy K Brock 2, Bryan M Fellman 3, Bruno C Odisio 1
PMCID: PMC9712233  NIHMSID: NIHMS1835162  PMID: 36058995

Abstract

Purpose:

This study aims to evaluate the intra-procedural use of a novel ablation confirmation (AC) method, consisting of biomechanical deformable image registration incorporating AI-based auto-segmentation, and its impact on tumor coverage by quantitative three-dimensional MAM (3D-MAM) CT-generated assessment.

Materials and methods:

This single-center, randomized, phase II, intent-to-treat trial is enrolling 100 subjects with primary and secondary liver tumors (≤3 tumors, 1–5 cm diameter) undergoing microwave or radiofrequency ablation with a goal of achieving ≥ 5 mm MAM. For the experimental arm, the proposed novel AC method is utilized for ablation applicator(s) placement verification and 3D-MAM assessment. For the control arm, the same variables are assessed by visual inspection and anatomical landmarks-based quantitative measurements aided by co-registration of pre- and post-ablation contrast-enhanced CT images. The primary objective is to evaluate the impact of the proposed AC method on the 3D-MAM. Secondary objectives are 2-year LTP-free survival, complication rates, quality of life, liver function, other oncological outcomes, and impact of AC method on procedure workflow.

Discussion:

The COVER-ALL trial will provide information on the role of a biomechanical deformable registration-based ablation confirmation method incorporating AI-based auto-segmentation for improving 3D-MAM, which might translate in improvements of liver ablation efficacy.

Conclusion:

The COVER-ALL trial aims to provide information on the role of a novel intra-procedural AC method for improving 3D-MAM, which might translate in improvements of liver ablation efficacy.

Keywords: Biomechanical deformable image registration, Liver cancer, Local tumor progression, Minimal ablation margin, Thermal ablation

INTRODUCTION

Liver cancer is a leading cause of cancer deaths globally [1]. Local therapies that are safe, cost-effective, and repeatable are urgently needed. Percutaneous thermal ablation has become a widely utilized local therapy for patients with both primary and secondary liver cancers not eligible for surgical resection, with recent series demonstrating encouraging overall survival rates in par with surgical resection in selected patients [28]. Nevertheless, its associated higher rates of local recurrence and disease-free survival when compared to surgical resection still remains as a major limitation for its widespread acceptance [4; 9-15].

Complete tumor coverage with minimum ablation margin (MAM) with at least ≥ 5 mm is considered one of the major factors associated with improved local tumor control following thermal ablation [1621]. Moreover, it is the only prognostic factor that can be modified during delivery of ablation therapy. Aside from proper patient selection, achieving sufficient MAM requires proper tumor mapping, targeting, and margins evaluation. Unfortunately, several intrinsic factors related to ablation therapy poses challenges to achieve such required steps. Namely, changes in liver shape, position, and volume due to patient’s breathing and ablation-related tissue dehydration, image quality degradation due to probe placement, hydrodissection, and limitations on contrast-media dose exposure are known limitations for ablation therapy accuracy [10; 22-24].

The retrospective evaluation of ablation confirmation (AC) software packages on patients previously submitted to percutaneous thermal ablation has been reported by several investigators, demonstrating a correlation between local tumor progression (LTP) and MAM thresholds estimated by such AC methods [1620]. Nevertheless, the prospective use of ablation confirmation (AC) methods intra-procedurally and its impact on ablation outcomes, along with its potential consequent clinical benefits, remains unclear, currently limiting the standardization and validation of such methods. We hypothesize that 3D-MAM following percutaneous thermal ablation of liver cancers will be significantly improved with the use of a novel AC consisting of a deformable image registration (DIR) method incorporating ablation-specific artificial intelligence (AI)-based auto-segmentation. This enables accurate mapping of the target tumor, verify proper ablation applicator(s) tumor targeting, and assess MAM intra-procedurally, while taking in account the intrinsic factors that currently limits such required steps. Therefore, potentially increasing percutaneous ablation efficacy and local oncological outcomes.

MATERIALS AND METHODS

Trial Design and Study Setting

The COVER-ALL trial is a prospective, randomized, two-arm, intent-to-treat phase II study, being conducted at The University of Texas MD Anderson Cancer Center. The trial started recruitment in January 2020. This study is funded by the National Institutes of Health.

Study population

Patients referred to percutaneous liver ablation at the Interventional Radiology Department will be screened for trial eligibility and enrolment based on the review of the electronic medical chart. Subjects over 18 years with confirmed primary and secondary liver cancers planned to undergo percutaneous thermal ablation as per standard of care are eligible. Subject can only be enrolled in the study once. Full inclusion and exclusion criteria are provided in Table 1.

Table 1:

Eligibility criteria for COVER-ALL

Inclusion criteria Exclusion criteria
Age > 18 years-old Active bacterial infection or fungal infection on the day of the ablation that would interfere the safety of procedure or the primary outcome assessment
ECOG Performance status 0–2 ASA score of > 4
Patients presenting with ≤ 3 liver tumors (biopsy-proven or documented by imaging) measuring 1 to 5 cm planned to undergo percutaneous thermal ablation with either microwave or radiofrequency ablation. Patients with more than 3 tumors might also be eligible in case other tumors can be treated with another curative-intended loco-regional therapy (i.e., surgical resection, radiation therapy) Any locoregional therapies at the target tumor within 30 days before the ablation procedure
Ability to completely cover the target tumor with at least a 5 mm ablation margin on the intraprocedural contrast-enhanced CT INR > 1.5 and Platelet < 50,000/mm3 or uncorrectable coagulopathy
Adequate estimated glomerular filtration rate Currently breastfeeding or pregnant
Target tumor is visualized on intra-procedural per-ablation contrast-enhanced CT Physical or psychological condition which would impair study participation

Abbreviations: ASA American Society of Anesthesiologists, ECOG Eastern Cooperative Oncology Group, INR International Normalized Ratio, CT Computed tomography.

Interventions

The ablation procedure, peri-procedural care, and follow-up will be performed in accordance with our standard of care institutional practice. All ablations are performed under CT-guidance under general anesthesia support. All CT images are acquired during apnea. Target tumors are treated with the intent to obtain complete tumor ablation with ≥ 5 mm MAM. Only radiofrequency ablation (Cool-tip, Medtronic Inc, Dublin, Ireland) and microwave ablation (Neuwave, Ethicon inc, Raritan, NJ, USA) are allowed in this study. Given our current practice, we expect that over 95% of enrolled patients will be treated with microwave ablation. Hydrodissection or other adjunctive techniques to prevent thermal damage to adjacent critical structures are allowed during any steps of the ablation procedure.

A schematic review of the study workflow is depicted on Figure 1. A dual-phase pre-ablation contrast-enhanced CT (CECT) is acquired to identify the target tumor and confirm trial eligibility using a standardized CT imaging protocol (Appendix 1). After percutaneous placement of ablation applicator(s), a native CT is acquired to verify the position of ablation applicator(s) at the target tumor by visual inspection. Then, both pre-ablation CECT and native CT images with ablation applicator(s) are transferred to RayStation (RaySearch Laboratories, Stockholm, Sweden) for AI-based auto-segmentation and biomechanical DIR for target tumor mapping on native CT image (Appendix 2). The results will not be disclosed to the interventional radiologist before randomization. The subject will not be enrolled if tumor cannot be well visualized or segmented. After ablation applicator placement is deemed appropriate by the interventional radiologist, subjects are randomized. For the experimental arm, the spatial correlation between the target tumor and ablation applicator(s) will be disclosed to the interventional radiologists on 2D and 3D images. Reposition of ablation applicator(s) and ablation will be performed accordingly to interventional radiologist’s discretion. For the control arm, the information generated by the AI-based auto-segmentation and biomechanical DIR will not be disclosed to the interventional radiologists.

Figure 1.

Figure 1.

COVER-ALL study design. CECT, contrast-enhanced CT

After ablation delivery, a post-ablation dual-phase CECT will be performed to confirm tumor coverage by ablation zone and 3D-MAM quantification. Pre- and post-ablation CECT images will be transferred to RayStation for biomechanical DIR and AI-based auto-segmentation. Then, the target tumor will be mapped onto post-ablation CECT image and the 3D-MAM will be automatically quantified (Figure 2). For the experimental arm, 3D-MAM quantification generated by the AC method will be disclosed to the interventional radiologist, along with its spatial localization on 2D and 3D images. For the control arm, the results of 3D-MAM quantification will not be disclosed to the interventional radiologist, who will determine the MAM as per our current standard of practice, consisting of anatomical landmarks-based quantitative measurements aided by co-registration of pre- and post-ablation CECT images [16]. Re-ablation is permitted in both arms. Final 3D-MAM quantification will be performed utilizing the final post-ablation CECT.

Figure 2.

Figure 2.

Artificial intelligence-based auto-segmentation, biomechanical deformable image registration, spatial correlation between the target tumor and ablation applicator(s), and 3D minimum ablation margin quantification in a 57-year-old man treated with microwave ablation. A) pre-ablation contrast-enhanced CT (CECT), auto-segmentation of liver (light blue) and target tumor (green) were performed on axial pre-ablation CECT. B) After placement of ablation applicator(s), a native CT was acquired to verify the position of ablation applicators (arrows). Then, biomechanical deformable image registration was performed with both pre-ablation CECT and native CT and the target tumor (green) was mapped on native CT. C) 3D volume rendering image of native CT for spatial correlation between the target tumor (green) and ablation applicators (dark blue). D) A final CECT was acquired to verify ablation zone (orange). Then, biomechanical deformable image registration was performed with both pre-ablation CECT and final CECT and the target tumor (green) was mapped on final CECT. The 3D minimum ablation margin was computed, which was 5.5 mm, located on a plane between the sagittal and coronal planes (not shown).

Randomization methodology

Subjects will be randomly assigned 1:1 to two treatment arms using the Pocock-Simon dynamic allocation method [25] with a minimization probability parameter of 0.90 to balance the baseline covariates: tumor histology (colorectal liver metastases or other histology), RAS mutation status (for colorectal liver metastasis only, mutated, wild-type, or undetermined), tumor size (<2 cm, 2 to 3 cm, or >3 to ≤5 cm), subcapsular location (defined as tumor within 1 cm from the liver capsule, yes or no), and presence of multiple tumors (yes or no).

Follow-up

Subjects will be followed up to 24 months after intervention. Subjects will undergo chest and abdominal CECT or contrast-enhanced liver magnetic resonance imaging 4 to 12 weeks after the ablation procedure. Then, chest and abdominal CECT, contrast-enhanced liver magnetic resonance imaging, or fluorine 18 fluorodeoxyglucose positron emission tomography will be acquired 3–4 months thereafter until death per institutional practice. Follow-up images will be reviewed by two experienced abdominal radiologists to determine residual unablated tumor, LTP, intrahepatic tumor progression other than target tumor, and extrahepatic disease progression. The abdominal radiologist will be blinded to the allocation of subjects (experimental vs control arm). The interpreting radiologists will be blinded to the results of 3D-MAM and the disagreement will be resolved by consensus. All the ablation outcomes will be assessed according to image-guided tumor ablation standardized terminology and reporting criteria [26; 27].

Objectives and outcomes

Clinical trial schema is depicted on Figure 3. The primary objective is to evaluate if the intra-procedural feedback of the proposed AC method (biomechanical DIR incorporating ablation-specific AI-based auto-segmentation) will increase the 3D-MAM on a three-dimensional computed tomography-generated analysis. Secondary objectives are assessing whether applying the proposed AC method improves 2-year LTP-free survival (LTPFS) rates and other oncological outcomes (i.e., intra-hepatic and overall progression-free survivals and overall survival), and to evaluate the impact of its use on procedure workflow, complication rates, quality of life, and liver function.

Figure 3.

Figure 3.

Clinical Trial Schema. * Interim look for superiority once half of the evaluable patients (n = 50) have been enrolled. ** Final analysis will be performed on 100 unique evaluable subjects.

Sample size

In order to avoid within patient correlation for subjects with multiple tumors, only the largest tumor per subject will be evaluated on this trial. A retrospective analysis at our institution using the proposed AC method demonstrated a mean 3D-MAM of 2 mm (standard deviation of 2 mm) without proposed AC method guiding intra-procedurally [28]. Assuming the pooled standard deviation of 3D-MAM is 2 mm, a sample size of 50 evaluable subjects in each arm will have 80% power to detect a difference of 1.132 mm using an independent 2-sample t-test with a two-sided 0.05 level of significance. In addition, we expect approximately 20 subjects to be screen failures and 20% subjects to be dropout after randomization due to tumor progression, inability to clearly depict target tumors on CECT, technical limitations, and complications that will preclude further ablation. To account these unevaluable subjects, we will screen a total of 140 subjects and enroll 120 subjects to ensure 100 evaluable unique subjects. Assuming an accrual rate of 35 subjects per year, the study accrual duration will be around 3 years, with the total study duration around 5 years to account for a follow-up period of at least 2 years.

Interim analysis

An interim look for superiority with the use of the proposed AC method will be performed once half the evaluable subjects (n = 50) have been enrolled. A Lan-Demets α-spending function using an Obrien-Flemming boundary will be used for superiority stopping boundaries [29]. We will stop enrolment at the control arm at our interim look if the differences on the 3D-MAM between the two arms disclose a p-value less than 0.003. In that case, the next 50 subjects will be enrolled on the experimental arm only to allow further development of the proposed method on clinical practice and allow more interventional radiologists at our institution to participate in this trial. East v6.5 (Cytel, Cambridge, MA, USA) was used for sample size calculation.

Statistical methods

For the primary objective, the average 3D-MAM will be compared between two arms using a 2-sample t-test (or Wilcoxon rank-sum test). The means and corresponding 95% confidence intervals will be reported for both arms. As secondary objectives, the Kaplan-Meier method will be used to estimate LTPFS and 95% confidence intervals for the quantiles of the survival function based on the method of Brookmeyer and Crowley [30] will be calculate. Time point probabilities (e.g., 2-year LTPFS) and the associated log-log transformed pointwise 95% confidence will also be reported. A multivariate Cox-proportional hazards model will be fitted to the data with 3D-MAM as a continuous variable to assess the significance of 3D-MAM on LTPFS while simultaneously adjusting for other known risk factors. LTPFS will be measured from date of ablation to earliest date of progression at the ablated tumor or death. Those progression free and alive will be censored at their date of last clinic visit. Similar analysis will be used for 2-years intra-hepatic progression-free survival and overall survival. Standard summary statistics will be computed for complication rates, quality of life, and liver function and compared between both arms. Statistical significance will be defined as p <0.05.

DISCUSSION

The increased use of thermal ablation as an alternative to surgical resection has been predicated by its minimally invasive nature, rational use of rising healthcare costs, lower complication rates, and faster recovery [9; 11; 13; 31]. Although studies have shown similar oncological outcomes between thermal ablation and surgery for small primary liver cancers, historically worse rates of local recurrence following thermal ablation when compared to surgery have hindered its application as a first local curative-intent modality for patients with primary and secondary liver cancers.

Several investigators retrospectively evaluated the impact of MAM on local tumor control rates, demonstrating an association of larger MAM with improved local tumor control. Currently, providing potential differences in respect tumor histology and subtypes, the optimal MAM is recommended to be at least ≥ 5mm [1620]. Nevertheless, such recommendations are based on retrospective data utilizing cross-sectional imaging that were not obtained intra-procedurally. Therefore, its extrapolation as an immediate surrogate for MAM is not possible given that immediate ablation-related changes depicted on imaging were not factored in. Moreover, the use of cross-sectional images acquired several days/weeks after the ablation to account for MAM also adds significant challenges for accounting tissue contraction into MAM estimation. Finally, manual segmentation and registration of tumor and ablation zones invariably adds operator bias on MAM analysis. We believe that the use of AI-based methods for tumor and ablation zone segmentations as proposed in this present study is poised to reduce operator input and consequent associated biases on MAM quantification.

Currently, the prospective use of AC methods for intra-procedural decision-making on an intent-to-treat approach and its consequent potential translation into clinical benefit remains elusive. It is expected that the COVER-ALL trial will allow to specifically evaluate the impact of the proposed AC method as an intra-procedural tool for decision-making and 3D-MAM quantification. It is also expected that this study will allow us to understand the impact of the use of this AC method on procedure workflow. We speculate that the use of the proposed AC method based on a DIR incorporating ablation-specific AI-based auto-segmentation for ablation applicator placement verification and MAM quantification will allow optimal coverage of the target tumor and surrounding tissue at risk, while reducing the ablation of non-target tissue (i.e., surrounding non-tumor liver parenchyma tissue).

Our study design has limitations. Firstly, this is a histology-agnostic study, which might limit the correlation between the 3D-MAM quantification generated using this novel AC method and LTP outcomes, which is a secondary objective of the study. Such limitation arises from the single-center nature of this study, which would make significantly challenging to accrue the required number of patients with a single tumor histology. Moreover, this study has been designed to translate a novel AC methodology consisting of DIR and artificial intelligence-based tumor and ablation zones segmentation into clinical practice. Therefore, a larger patient population with a wider variety of tumor types would better reflect the current unmet needs in clinical practice. Multi-institutional studies investigating the use AC methods focused on a single tumor histology such as the Prometheus (hepatocellular carcinoma) and ACCLAIM trial (colorectal liver metastasis) are better suited to correlate ablation margins with local tumor outcomes [32; 33]. Secondly, given the current constrains in performing AI-based segmentation on tumors < 1 cm, sub-centimeters tumors are excluded from the trial, adding a selection bias to trial design. Third, acquisition of more than one post-ablation CECT might be required if intra-procedural 3D-MAM ≥5 mm is not achieved in first attempt, which may increase the risk of contrast-associated acute kidney injury. Thus, we only include subjects with preserved renal function in the trial. Finally, due to the unblinded nature of this study, it is conceivable that operator bias might occur in respect performing more extensive ablations and more careful planning in one of the study’s arms. In order to gain further insight on this potential bias, we will perform volumetric analysis on the amount of non-tumorous liver parenchyma ablated tissue between the two arms.

In conclusion, the COVER-ALL trial aims to provide information on the role of a novel intra-procedural ablation confirmation method for improving 3D-MAM, which might translate in improvements of liver ablation efficacy. Trial registration: ClinicalTrials.gov identifier: NCT04083378

Supplementary Material

1835162_Appendix_1-2

Funding:

This study is funded by the National Institutes of Health grants: 1R01CA235564.

Footnotes

Compliance with Ethical Standards: All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Informed consent: Informed consent will be obtained from all individual participants included in the study.

Consent for publication: For this type of study, consent for publication is not required.

Trial registration: ClinicalTrials.gov identifier: NCT04083378

Conflict of interest: The authors declare that they have no conflict of interest.

REFERENCES

  • 1.Sung H, Ferlay J, Siegel RL, et al. (2021) Global Cancer Statistics 2020: GLOBOCAN Estimates of Incidence and Mortality Worldwide for 36 Cancers in 185 Countries. CA Cancer J Clin, 71(3):209–249 [DOI] [PubMed] [Google Scholar]
  • 2.Solbiati L, Ahmed M, Cova L, Ierace T, Brioschi M, Goldberg SN (2012) Small liver colorectal metastases treated with percutaneous radiofrequency ablation: local response rate and long-term survival with up to 10-year follow-up. Radiology, 265(3):958–968 [DOI] [PubMed] [Google Scholar]
  • 3.Shady W, Petre EN, Gonen M, et al. (2016) Percutaneous Radiofrequency Ablation of Colorectal Cancer Liver Metastases: Factors Affecting Outcomes--A 10-year Experience at a Single Center. Radiology, 278(2):601–611 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Shi J, Sun Q, Wang Y, et al. (2014) Comparison of microwave ablation and surgical resection for treatment of hepatocellular carcinomas conforming to Milan criteria. J Gastroenterol Hepatol, 29(7):1500–1507 [DOI] [PubMed] [Google Scholar]
  • 5.Shiina S, Tateishi R, Arano T, et al. (2012) Radiofrequency ablation for hepatocellular carcinoma: 10-year outcome and prognostic factors. Am J Gastroenterol, 107(4):569–577; quiz 578 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Peng ZW, Zhang YJ, Chen MS, Lin XJ, Liang HH, Shi M (2010) Radiofrequency ablation as first-line treatment for small solitary hepatocellular carcinoma: long-term results. Eur J Surg Oncol, 36(11):1054–1060 [DOI] [PubMed] [Google Scholar]
  • 7.Wang Z, Liu M, Zhang DZ, et al. (2022) Microwave ablation versus laparoscopic resection as first-line therapy for solitary 3–5-cm HCC. Hepatology, 76(1):66–77 [DOI] [PubMed] [Google Scholar]
  • 8.Bale R, Widmann G, Schullian P, et al. (2012) Percutaneous stereotactic radiofrequency ablation of colorectal liver metastases. Eur Radiol, 22(4):930–937 [DOI] [PubMed] [Google Scholar]
  • 9.Weis S, Franke A, Mössner J, Jakobsen JC, Schoppmeyer K (2013) Radiofrequency (thermal) ablation versus no intervention or other interventions for hepatocellular carcinoma. Cochrane Database Syst Rev(12):Cd003046. [DOI] [PMC free article] [PubMed]
  • 10.Gillams A, Goldberg N, Ahmed M, et al. (2015) Thermal ablation of colorectal liver metastases: a position paper by an international panel of ablation experts, The Interventional Oncology Sans Frontières meeting 2013. Eur Radiol, 25(12):3438–3454 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Abdalla EK, Vauthey JN, Ellis LM, et al. (2004) Recurrence and outcomes following hepatic resection, radiofrequency ablation, and combined resection/ablation for colorectal liver metastases. Ann Surg, 239(6):818–825; discussion 825–817 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Di Martino M, Rompianesi G, Mora-Guzmán I, Martín-Pérez E, Montalti R, Troisi RI (2020) Systematic review and meta-analysis of local ablative therapies for resectable colorectal liver metastases. Eur J Surg Oncol, 46(5):772–781 [DOI] [PubMed] [Google Scholar]
  • 13.Otto G, Düber C, Hoppe-Lotichius M, König J, Heise M, Pitton MB (2010) Radiofrequency ablation as first-line treatment in patients with early colorectal liver metastases amenable to surgery. Ann Surg, 251(5):796–803 [DOI] [PubMed] [Google Scholar]
  • 14.Nishiwada S, Ko S, Mukogawa T, et al. (2014) Comparison between percutaneous radiofrequency ablation and surgical hepatectomy focusing on local disease control rate for colorectal liver metastases. Hepatogastroenterology, 61(130):436–441 [PubMed] [Google Scholar]
  • 15.Kron P, Linecker M, Jones RP, Toogood GJ, Clavien PA, Lodge JPA (2019) Ablation or Resection for Colorectal Liver Metastases? A Systematic Review of the Literature. Front Oncol, 9:1052. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Wang X, Sofocleous CT, Erinjeri JP, et al. (2013) Margin size is an independent predictor of local tumor progression after ablation of colon cancer liver metastases. Cardiovasc Intervent Radiol, 36(1):166–175 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Shady W, Petre EN, Do KG, et al. (2018) Percutaneous Microwave versus Radiofrequency Ablation of Colorectal Liver Metastases: Ablation with Clear Margins (A0) Provides the Best Local Tumor Control. J Vasc Interv Radiol, 29(2):268–275.e261 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Shady W, Petre EN, Vakiani E, et al. (2017) Kras mutation is a marker of worse oncologic outcomes after percutaneous radiofrequency ablation of colorectal liver metastases. Oncotarget, 8(39) [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Calandri M, Yamashita S, Gazzera C, et al. (2018) Ablation of colorectal liver metastasis: Interaction of ablation margins and RAS mutation profiling on local tumour progression-free survival. Eur Radiol, 28(7):2727–2734 [DOI] [PubMed] [Google Scholar]
  • 20.Nakazawa T, Kokubu S, Shibuya A, et al. (2007) Radiofrequency ablation of hepatocellular carcinoma: correlation between local tumor progression after ablation and ablative margin. AJR Am J Roentgenol, 188(2):480–488 [DOI] [PubMed] [Google Scholar]
  • 21.Laimer G, Jaschke N, Schullian P, et al. (2021) Volumetric assessment of the periablational safety margin after thermal ablation of colorectal liver metastases. Eur Radiol, 31(9):6489–6499 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Brace CL, Diaz TA, Hinshaw JL, Lee FT, Jr. (2010) Tissue contraction caused by radiofrequency and microwave ablation: a laboratory study in liver and lung. J Vasc Interv Radiol, 21(8):1280–1286 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Mulier S, Ni Y, Jamart J, Ruers T, Marchal G, Michel L (2005) Local recurrence after hepatic radiofrequency coagulation: multivariate meta-analysis and review of contributing factors. Ann Surg, 242(2):158–171 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.van Duijnhoven FH, Jansen MC, Junggeburt JM, et al. (2006) Factors influencing the local failure rate of radiofrequency ablation of colorectal liver metastases. Ann Surg Oncol, 13(5):651–658 [DOI] [PubMed] [Google Scholar]
  • 25.Pocock SJ, Simon R (1975) Sequential treatment assignment with balancing for prognostic factors in the controlled clinical trial. Biometrics, 31(1):103–115 [PubMed] [Google Scholar]
  • 26.Ahmed M, Solbiati L, Brace CL, et al. (2014) Image-guided tumor ablation: standardization of terminology and reporting criteria--a 10-year update. Radiology, 273(1):241–260 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Puijk RS, Ahmed M, Adam A, et al. (2021) Consensus Guidelines for the Definition of Time-to-Event End Points in Image-guided Tumor Ablation: Results of the SIO and DATECAN Initiative. Radiology, 301(3):533–540 [DOI] [PubMed] [Google Scholar]
  • 28.Anderson BM, Lin YM, Lin EY, et al. (2021) A novel use of biomechanical model-based deformable image registration (DIR) for assessing colorectal liver metastases ablation outcomes. Med Phys, 48(10):6226–6236 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Kalbfleisch JD, Prentice RL (2011) The statistical analysis of failure time data, John Wiley & Sons [Google Scholar]
  • 30.Brookmeyer R, Crowley J (1982) A Confidence Interval for the Median Survival Time. Biometrics, 38(1):29–41 [Google Scholar]
  • 31.Mulier S, Ni Y, Jamart J, Michel L, Marchal G, Ruers T (2008) Radiofrequency ablation versus resection for resectable colorectal liver metastases: time for a randomized trial? Ann Surg Oncol, 15(1):144–157 [DOI] [PubMed] [Google Scholar]
  • 32.Oosterveer TTM, van Erp GCM, Hendriks P, et al. (2022) Study Protocol PROMETHEUS: Prospective Multicenter Study to Evaluate the Correlation Between Safety Margin and Local Recurrence After Thermal Ablation Using Image Co-registration in Patients with Hepatocellular Carcinoma. Cardiovasc Intervent Radiol, 45(5):606–612 [DOI] [PMC free article] [PubMed] [Google Scholar]