Abstract
Purpose:
To develop and validate a method for B0 mapping for knee imaging using the quantitative Double-Echo in Steady-State (qDESS) exploiting the phase difference (Δθ) between the two echoes acquired. Contrary to a two-gradient-echo (2-GRE) method, Δθ depends only on the first echo time.
Methods:
Bloch simulations were applied to investigate robustness to noise of the proposed methodology and all imaging studies were validated with phantoms and in vivo simultaneous bilateral knee acquisitions. Two phantoms and 5 healthy subjects were scanned using qDESS, water saturation shift referencing (WASSR), and multi-GRE sequences. ΔB0 maps were calculated with the qDESS and the 2-GRE methods and compared against those obtained with WASSR. The comparison was quantitatively assessed exploiting pixel-wise difference maps, Bland-Altman (BA) analysis, and Lin’s concordance coefficient (ρc). For in-vivo subjects, the comparison was assessed in cartilage using average values in 6 sub-regions.
Results:
The proposed method for measuring ΔB0 inhomogeneities from a qDESS acquisition provided ΔB0 maps that were in good agreement with those obtained using WASSR. ΔB0 ρc values were ≥ 0.98 and 0.90 in phantoms and in vivo, respectively. The agreement between qDESS and WASSR was comparable to that of a 2-GRE method.
Conclusion:
The proposed method may allow B0 correction for qDESS T2 mapping using an inherently co-registered ΔB0 map without requiring an additional B0 measurement sequence. More generally, the method may help shorten knee imaging protocols that require an auxiliary ΔB0 map by exploiting a qDESS acquisition that also provides T2 measurements and high-quality morphological imaging.
Keywords: qMRI, B0 mapping, qDESS, Bilateral knee imaging
Introduction
Quantitative Magnetic Resonance Imaging (MRI) techniques are valuable tools for assessing macromolecular changes in collagenous tissues and studying osteoarthritis (OA) progression1,2. These parameters are measured using a variety of 2D and 3D sequences, all of which have varying sensitivities to pulse sequence parameters and hardware imperfections3. Magnetic field inhomogeneities can affect quantitative accuracy in many of these techniques, such as T2 relaxation time mapping4,5, chemical exchange saturation transfer of glycosaminoglycans6 (gagCEST). More generally, obtaining information about static magnetic field (B0) inhomogeneities is beneficial for a variety of musculoskeletal (MSK) MRI methodologies. For example, correction of field-induced distortions has shown to improve localization of bone in diffusion-weighted imaging (DWI)7. Furthermore, B0 mapping can be used to check if the right condition exists for effective fat saturation, chemical shift-based imaging methodologies8,9, or simultaneous fat-water imaging10. B0 inhomogeneity corrections usually require ancillary scans for B0 mapping using the Water Saturation Shift Referencing11 (WASSR) method or the two Gradient-Echo (2-GRE) method12. These additional scans can lead to time-consuming MRI protocols13 that can limit the application of these methods, especially in studies with large cohorts.
The 3D double-echo in steady-state (DESS) pulse sequence14,15,16 has been widely applied in musculoskeletal imaging due to its image contrast and high SNR efficiency. DESS enables high-resolution imaging of cartilage, including assessment of cartilage thickness and defects17,18,19. Separation of the two echoes in the DESS sequence in a method known as quantitative DESS20,21 (qDESS) can further provide accurate and reproducible T2 measurements22,23 by fitting their signals to an extended phase graph signal model24,25,26. Additionally, the SNR efficiency and multiple contrasts provided by this approach have shown promise for bilateral clinical knee assessment in five minutes or less27,28,29,30. However, as for other T2 mapping methods, T2 estimation accuracy is affected by field inhomogeneities.
Since qDESS acquires two echoes it could be possible to exploit the phase difference between them to obtain a ΔB0 map as commonly done for the 2-GRE method. However, phase evolution in qDESS is more complicated than a GRE multi-echo sequence. In the latter, all echoes are gradient-recalled, and the phase accumulation between two echoes is proportional to the echo time difference. In qDESS, neglecting higher order pathways in the steady state22, the first echo is gradient-recalled, whereas the second echo records the rephasing of the gradient-recalled echo created by the radio-frequency (RF) pulse in the preceding cycle. Thus, the phase evolution is very different from a 2-GRE case and, as it will be explained in the Theory section, the phase difference between the two echoes is proportional to twice the first echo time, whereas it is independent on the choice of the second echo time. Recently, an analytical equation for performing B0 mapping exploiting any two unbalanced Steady State Free Precession (ubSSFP) signals was proposed by Leupold31. Since the qDESS sequence belongs to the family of ubSSFP sequences, the work provided a theoretical setting to measure ΔB0 maps exploiting the difference between the two echoes acquired with qDESS. However, the work by Leupold was purely theoretical and did not provide insights into the actual applicability of the method for in-vivo B0 mapping in complex body regions such as the knee. For example, as will be further explained in the Theory section, the wrap-free bandwidth (the off-resonances for which the phase difference between the two echoes does not present wraps) of the qDESS method is much smaller than that of a standard 2-GRE method. Thus, highly wrapped phase images could be obtained, which can be hard to unwrap in complicated regions such as the knee even with current phase unwrapping algorithms. Inability to correctly perform phase unwrapping would hinder the practical applicability of the method despite theoretical feasibility.
This work aims i) to provide an empirical verification and validation of the theoretical framework proposed by Leupold for B0 mapping using qDESS with phantom and in-vivo experiments again standard methods, ii) to set up and describe an optimal processing pipeline, and iii) to analyze strengths and limitations for a practical application to simultaneous bilateral knee imaging. These contributions may allow implementing a B0 correction method for qDESS T2 mapping without acquiring an additional scan with the additional benefit of inherent co-registration of the T2 and ΔB0 maps. More generally, the method may help shorten knee imaging protocols that require an auxiliary ΔB0 map, such as gagCEST imaging, by exploiting a qDESS acquisition that also provides T2 measurements and high-quality morphological imaging. Preliminary results for this work were presented at the 29th ISMRM annual meeting & exhibition32.
Theory
The qDESS is a gradient-spoiled steady-state sequence that samples two echoes per repetition (TR) that are separated by a spoiler gradient. The two signals have different contrast and are functions of the extrinsic scan parameters (TR, TE, flip angle α, gradient amplitude G, and gradient duration τ) as well as the intrinsic tissue parameters (T1, T2, and diffusivity). Figure 1(A) shows the pulse diagram of the qDESS sequence along with a phase graph representation. Multiple echo pathways contribute to the origin of the first and second echoes, S1 and S2, respectively. Neglecting the contribution from high dephasing orders pathways22, S1 can be thought of as a Free Induction Decay (FID) signal, and thus it has weighting. The second echo (S2), conversely, is a combination of a spin-echo pathway, formed by pulse refocusing the first echo S1 of the previous TR, and a gradient echo pathway and it has a higher T2 and diffusion weighting. The pulse diagram of a multi-echo GRE with a phase graph representation is reported in Fig. 1(B) to highlight the difference between qDESS and a multi-echo GRE sequence.
Figure 1:
(A) qDESS pulse sequence diagram with phase graph representation. FGRE pulse sequence diagram with phase graph representation. (C) The schematization of the processing pipeline that was used to compute ΔB0 map in Femoral Cartilage using qDESS. The coil-combined phase difference between the second and first echo is computed using the HiP. Femoral cartilage is segmented using a deep learning model with the DSOMA framework. The 3D wrapped phase difference is unwrapped and then ΔB0 is computed according to eq. 2. The 3D ΔB0 map in FC is projected onto a 2D space for visualization using DOSMA.
ΔB0 field mapping with qDESS
Recently, Leupold proposed a theoretical equation to estimate the off-resonance, ΔB0, based on any two unbalanced Steady State Free Precession (ubSSFP) signals31. This relationship is reported in equation 1.
| [1] |
Here is the phase difference between the second and first ubSSFP signals acquired at echo times TE2 and TE1, respectively, and ki describes the dephasing order of a ubSSFP signal using the EPG framework. The indicator function is such that if k ≥ 0 and if k < 0.
For a qDESS sequence, the first echo (S1) has dephasing order k1 = 0, whereas the second echo (S2) is has dephasing order k2 = −1. Note that since TE2 = 2TR − TE1 in qDESS, eq 1 can then be expressed as eq. 2:
| [2] |
where the numerator is the phase difference, Δθ, between S2 and S1, obtained applying the angle function (∠) to the Hermitian inner Product33 (HiP), , of the two complex qDESS echo signals. It is worth noting that according to eq. 1, a factor of π is added to the actual phase difference so that in eq. 2 one should subtract π from the phase difference. In our qDESS implementation, the RF pulses are given by alternating the phase of the pulses by a factor of π. As such, no π factor is added to the actual phase difference, and thus it does not need to be subtracted from Δθ in eq. 2.
Contrary, with the 2-GRE method, both echoes have dephasing order k1 = 0 and eq. 1 becomes the well known relationship:
| [3] |
Note that in the qDESS sequence, the second echo has a pulse-refocused component, so the phase accumulation due to off-resonances is also refocused. Thus, once the first echo time is set, varying TE2 does not change the phase accumulation due to off-resonances. The ΔB0 region for which Δθ does not present phase wraps, the wrap-free bandwidth, is then given by Eq. 4
| [4] |
Thus, it is possible to observe that the wrap-free bandwidth is much smaller than that of a 2-GRE method run using an echo spacing similar to qDESS TE1. The minimum TE1 values for qDESS are in the order of a few units of ms, implying that off-resonances in the order of a few tens of Hz give rise to phase wraps in qDESS difference maps. In MSK imaging, it is not uncommon to observe ΔB0 in the order of a few hundreds of Hz. In such cases, phase unwrapping algorithms are necessary to restore the correct phase difference maps. However, for complex phase topographies such as those encountered in MSK imaging, phase unwrapping is a difficult task that can potentially introduce severe errors in phase estimation and thus in ΔB0 estimation.
Effect of ΔB0 off-resonances in T2 estimation with qDESS
Although not investigated in the current work, it is worth providing insights on the effect of ΔB0 off-resonances in qDESS T2 mapping estimation accuracy. Sveinsson et al.22 proposed a simple analytic model that can be used for T2 estimation from the ratio between the second and first echoes of qDESS. The model is expressed in eq. 5:
| [5] |
Here, TR and TE represent the repetition and the first echo time, respectively, α is the flip angle, D is the diffusivity, and the dephasing per unit length induced by the unbalanced gradient is denoted by Δk = γGτ, where G and τ are the spoiler amplitude and duration, respectively, and γ is the gyromagnetic ratio. This simple model, which allows estimating T2 from qDESS, also allows one to appreciate the mechanism for which ΔB0 off-resonances can introduce errors in qDESS T2 mapping. The flip angle must be known to correctly estiamte T2 from eq. 5. It is clear how B1 inhomogeneities across the imaged volume directly affect the actual flip angle experienced by the spins within voxels. When selective pulses are used, which is the standard for fat-suppressed qDESS acquisition, ΔB0 inhomogeneities also affect the actual flip angle experienced by the spins within the voxels accordingly to the frequency profile of the selective pulse used.
Figure 2 reports Bloch simulations run to show the dependence of the T2 estimation error as a function of the off-resonance experienced by the spin ensemble when considering the frequency profile of the selective pulse used in the qDESS sequence.
Figure 2: T2 error as a function of the off-resonance ΔB0.
(A) qDESS RF pulse shape with slice-selective gradient (top) and corresponding spectral (bottom left) and slice (bottom right) profiles. (B) T2 error as a function of the off-resonance ΔB0 obtained running Bloch simulations considering the qDESS RF slice profile for a voxel composed of cartilage (T2 = 45 ms).
Methods
Simulation experiment
Bloch simulations of the qDESS pulse sequence were performed using an in-house developed MATLAB (MathWorks) library34. The simulations considered the spectral profile of the selective pulse used for fat suppression in real qDESS acquisitions. The pulse and its corresponding frequency profile are reported in Fig. 2.
The accuracy and robustness to noise of the qDESS B0 mapping method (eq. 2) were investigated by simulating the qDESS signal (#TR = 400) for different combinations of TE1 and TE2 considering a spin ensemble with T1 = 1400 ms and T2 = 45 ms (relaxation parameters of articular cartilage). Three first echo times were selected: TE1 = 5, 10 and 15ms. For each TE1, the simulations were run three times, progressively increasing TE2. For each pair of (TE1, TE2), the signal was simulated for ΔB0 values ranging from −90Hz to 90Hz with a step size of 0.25Hz resulting in an array of dimensions n x m, where n is the number of ΔB0 values (n = 721) and m is the number of pulse repetition (m = 400). For each ΔB0 value, white Gaussian noise was added progressively increasing the variance of the noise such that the signal-to-noise ratio (SNR) for the qDESS signal with TE1 = 5ms and TE2 = 25ms ranged from 250 to 5. The SNR was defined according to eq. 6, where Psignal and Pnoise represent the average power of the qDESS signal and the noise, respectively, and E(signal) indicates the expectation value of the qDESS signal.
| [6] |
The noise addition procedure was repeated 721 times for each ΔB0 and variance value. The first and second noisy echoes of the last time repetition (the steady state) were retained. For each echo, the data were then reshaped into a n x n matrix, and the phase difference between the second and first echo was computed pixel-wise, applying the angle function to the HiP of the two echoes (see eq. 2. Since phase unwrapping was always necessary given the values of TE1 and the range of ΔB0 considered, 2D unwrapping algorithms are more robust to noise than 1D unwrapping algorithms. Thus, a 2D phase unwrapping algorithm based on the transport of intensity equation35 was used to unwrap the phase maps. The need for effective phase unwrapping was the rationale for reshaping the data in a 2D image, which was also considered closer to a real scenario.
Mean Error (ME) and Root Mean Squared Error (RMSE) between estimated ΔB0 values and ground truth values were estimated for each noise variance and averaged across ΔB0 values.
MRI acquisitions
All MRI acquisitions were performed on a GE 3T SIGNA Premier scanner (GE Healthcare, Milwaukee, WI, USA) using 16-channel flexible phased-array receive-only coils (NeoCoil, Pewaukee, WI, USA). Informed consent was obtained from all subjects in accordance with the Institutional Review Board protocol at our institution. Simultaneous bilateral scans used two coils wrapped around the knees.
The MR imaging was performed using 3D qDESS, 3D Magnetization Prepared SPGR (WASSR) and 2D fast multi-echo GRE (FGRE) sequences. 3D qDESS used a spectrally-selective excitation with nominal flip angle α = 20° and a spoiler gradient, which had a spoiler moment of 3.13mT/mms with duration τ = 1.6ms. WASSR magnetization preparation utilized a 200ms saturation pulse train with a root mean squared B1 of 0.3μT, from −1ppm to 1ppm with a step size of 0.1ppm. Other imaging parameters are specified for each acquisition in the following sections.
Phantoms scans
To validate eq. 2 and study accuracy for different combinations of TE1 and TE2, an agar gel phantom was scanned multiple times using qDESS with three different first echo times: TE1 = 5, 10 and 15ms. For each TE1, the sequence was run three times, progressively increasing the TE2. Using multiple first and second echo times also allowed us to evaluate the effect of SNR on qDESS ΔB0 estimation accuracy. The voxel resolution was 0.7 × 0.7 × 4.0mm3, and all other qDESS imaging parameters were kept the same among acquisitions. A reference ΔB0 map was acquired using WASSR with readout parameters of: TR = 3ms, TE = 1.2ms, α = 5°, voxel size 0.7 × 0.7 × 4.0mm3. The experiment was repeated by modifying the shimming gradient along the x-axis, from −12 mT/m (default prescan value) to 28mT/m, to obtain a greater B0 heterogeneity. The ΔB0 maps obtained using qDESS were compared to the WASSR method obtained reference ΔB0 using pixel-wise difference and Bland-Altman (BA) analysis.
A second phantom, made of 6 vials containing doped water, was scanned using qDESS, WASSR and the Fast GRadient-Echo (FGRE) sequences. qDESS scan parameters were TE1 = 5ms, TR = 25ms (i.e. TE2 = 45ms), voxel resolution = 1 × 1 × 2mm3. WASSR scan parameters were TR = 3.5ms, TE = 2.5ms and voxel size = 1 × 1 × 2mm3. The FGRE scan parameters were TR = 30ms, TE = 2.5, 5, 7.5 and 10ms, voxel size = 1 × 1 × 2mm3. The first two echoes were used to compute the ΔB0 map, i.e. 2-GRE method. The duration of the acquisitions (expressed in minutes:seconds) were 1:42, 3:28, and 3:58 for qDESS, WASSR, and FGRE, respectively. The ΔB0 maps obtained with the qDESS and the 2-GRE methods were compared to the reference map obtained using the WASSR method. The 2-GRE method was added to the comparison to observe the agreement between another standard ΔB0 mapping method with the WASSR reference method and also to observe the agreement between the qDESS and 2-GRE method.
The rationale for considering WASSR the reference method for ΔB0 mapping was because this method has shown sub-Hz accuracy11. This has made it the benchmark for CEST asymmetry ΔB0 correction, which is particularly sensitive to these inhomogeneities. The 2-GRE method also provides accurate ΔB0 mapping, but its absolute accuracy has shown dependence on the ΔTE used36. Thus, the WASSR method was considered a more appropriate reference for evaluating the accuracy of the proposed qDESS ΔB0 mapping method.
In vivo simultaneous bilateral knee scans
To validate the qDESS B0 mapping method in vivo, five healthy subjects were scanned in the sagittal plane with the MRI protocol described above using a simultaneous bilateral knee acquisition27. qDESS scan parameters were TE1 = 5ms, TR = 15ms (i.e. TE2 = 25ms), voxel size = 0.3 × 0.3 × 1.4mm3. WASSR scan parameter were TR = 3.5ms, TE = 2.5ms and voxel size = 0.6 × 0.6 × 4mm3. The FGRE scan parameters were TR = 30ms, TE = 1.4, 3.5 and 5.7ms, voxel size 0.6 × 0.6 × 4mm3. As for the phantoms, the first two echoes were used to compute the ΔB0 map with the 2-GRE method. The voxel size was kept the same among acquisitions to minimize differences in the ΔB0 maps that could arise from imaging parameters instead of reflecting a method-specific difference. Acquisition times (expressed in minutes:seconds) were 2:40, 11:27, and 3:29 for qDESS, WASSR, and 2-GRE, respectively.
Data processing and analysis
Multi-channel data: phase difference reconstruction
For qDESS and FGRE, phase difference images between two echoes were obtained to compute ΔB0 maps. The phase difference images were obtained by applying the angle function to the coil-combined complex difference between the two complex echo signals using the HiP according to eq. 737.
| [7] |
Here, is the complex conjugate of the complex signal from the m-th channel at the first echo time TE1. Sm(TE2) is the complex signal from the m-th channel at the second echo time TE2. N is the number of total channels, which was equal to 16. λm is a weight factor that accounts for varying noise in different coil channels38 and was computed according to eq. 8.
| [8] |
In the equation above, σm is the noise amplitude in the corresponding m-th channel. For each channel, σm is calculated by computing the root mean squared error of pixel intensities from a 10×10 pixel area in the background region of the magnitude images38.
Once obtained the Δθ maps, ΔB0 maps were obtained using eq. 2, for qDESS, and using the 2-GRE method (eq. 3) for FGRE data.
ΔB0 map analysis in femoral cartilage
The ΔB0 maps were evaluated in the femoral cartilage (FC). The pipeline used to obtain the ΔB0 maps in the FC with qDESS is summarized in Fig. 1 (C). For both knees, after coil-combination and phase difference computation, the femoral cartilage was segmented using the Deep Open-Source Medical Image Analysis (DOSMA) framework39,40(https://github.com/ad12/DOSMA). When the 3D phase maps presented wraps, 3D phase unwrapping was used to perform 3D phase unwrapping. Since multiple phase unwrapping algorithms were tested, details are given in the following subsection. After phase unwrapping, the qDESS ΔB0 maps in FC were computed using Eq. 2 and projected onto a 2D space for visualization41. The FC was automatically subdivided into 6 sub-regions: anterior/central/posterior compartments for the medial and lateral condyles.
WASSR and FGRE images were registered to qDESS using Elastix42,43. The FC segmentation mask computed from qDESS was then applied to the registered WASSR and FGRE images. The 3D ΔB0 maps of FC were then obtained and processed with DOSMA as described above.
Phase unwrapping
Due to the relatively narrow wrap-free bandwidth of qDESS phase difference (see 4), phase unwrapping is likely to be necessary for restoring the correct phase difference and thus obtaining accurate ΔB0 maps.
To investigate the most reliable phase unwrapping algorithm, we performed phase unwrapping of in vivo knee qDESS phase difference maps with 3 different algorithms: Phase Region Expanding Labeller for Unwrapping Discrete Estimates44 (PRELUDE), Speedy rEgion-Growing algorithm for Unwrapping Estimated phase45 (SEGUE) and the Rapid Opensource Minimum spanning treE algOrithm46 (ROMEO). PRELUDE belongs to the region-growing spatial unwrapping approach class. Although it is considered the gold standard, it is computationally demanding and masking of the image is normally required to speed up computation. SEGUE is based on PRELUDE, but it can reduce computation time by simultaneously unwrapping and merging multiple regions. In contrast, ROMEO is a recently proposed phase unwrapping method belonging to the class of path-following approaches, which usually provide solutions within seconds, even for highly wrapped images with large matrix sizes. Thus, masking is not necessary.
For each algorithm, the phase unwrapping procedure was tested with and without applying a cartilage mask to the 3D volume. However, a mask removing the background and low-intensity regions (such as the bones) was applied to reduce computation time for PRELUDE and SEGUE. The background mask was automatically obtained by applying intensity thresholding to the 3D volume of the first echo of qDESS. The cartilage mask was added to the background mask to ensure that no pixels belonging to cartilage was wrongly removed by the binary operation.
Successful phase unwrapping was assessed by manually comparing the unwrapped phase maps with the reference WASSR B0 maps. Since eq. 2 describes the relationship between Δϕ and ΔB0 for qDESS it is possible to verify if phase unwrapping was successful.
Statistical analysis
The comparison between ΔB0 maps was quantitatively assessed by exploiting pixel-wise difference maps, the mean deviation (MD), root-mean-squared-deviation (RMSD) and Bland-Altman (BA) analysis. The Lin’s concordance coefficient47 (ρc) was also evaluated as a global metric for assessing the agreement between ΔB0 maps. For the in vivo acquisitions, the BA analysis was performed considering the average ΔB0 values measured in the 6 FC sub-regions. All statistical analysis was performed using in-house scripts written in Python 3
Results
Bloch simulations
Figure 3 reports the RMSE and ME between qDESS estimated ΔB0 values and ground truth values as a function of the noise variance for different combinations of TE1 and TE2. The ME was always below 0.5 regardless of the level of noise that was injected into the data. This implies that there’s no bias between qDESS ΔB0 values and ground truth values, and the method was very robust to noise in terms of bias. Regarding the RMSE, it rapidly fell below 5 and 2Hz when the SNR of the first echo was above 30 and 70, respectively. Once fixed the first echo time, increasing the value of the second echo time lowered the SNR, which increased the RMSE. Using a TE1 equal to 15ms yielded the best SNR for the first echo, which resulted in the best RMSE values despite the SNR for the second echo being the worst among the other values of TE1 used.
Figure 3: Robustness to noise investigated with Bloch simulations.
RMSE (blue line) and ME (orange line) between qDESS estimated ΔB0 values and ground truth values as a function of the noise variance for different combinations of TE1 and TE2. The coordinate axis is expressed as the reciprocal of the noise variance. The top coordinate axis reports the corresponding SNRs for the first and second echos.
Validation in phantoms
Agar phantom
The qDESS ΔB0 maps obtained using different combinations of TE1s and TE2s are reported in Fig. 4 along with their difference with the reference ΔB0 (left column). Due to the small range of ΔB0 values (−15 to 15Hz), phase unwrapping was not necessary. Furthermore, isolated dark red spots in the difference maps are due to the presence of air bubbles in the agar gel phantom (See Figure S1 of supplementary material). All the computed ΔB0 maps were similar across different (TE1, TE2) combinations, and had a small positive MD with the reference values (bias) ranging from 2.3 to 4.6Hz and a RMSE ranging from 2.5 yo 4.8Hz. The highest deviation and RMSE were observed for the pair TE1 = 5ms and TE1 = 85ms, which was consistent with what was found in the simulation experiment. Overall, there was good agreement between the qDESS and WASSR ΔB0 maps as supported by the Lin’s coefficients, which were all greater or equal to 0.72, with the only exception of the map computed from TE1 = 5ms and TE1 = 85ms (ρc = 0.65) for which SNR was the smallest.
Figure 4: Results of the experiment with the agar phantom.

(left panel) Reference ΔB0 map obtained with WASSR. (right panel) qDESS ΔB0 maps for different combinations of TE1 and TE2 in the agar phantom. The estimated off-resonance ΔB0 maps have good agreement with the reference map regardless of the combination of TE1 and TE2. The worst agreement is when TE1 = 5ms and TE2 = 85ms, which also shows the lowest SNR. Isolated dark red spots in the difference maps are due to the presence of air bubbles in the agar gel phantom (See Fig. S1 of supplementary material).
The results obtained modifying the shimming gradient along the x-axis are summarized in Fig.5. Figure 5 (A) reports the ΔB0 map obtained using the WASSR method, and the Δθ maps obtained from qDESS for TE1s equal to 5ms and 15ms. The ΔB0 map spanned the range −40Hz to 40Hz, with a linear dependence with the x-position. The qDESS phase difference Δθ obtained using TE1 = 15ms presented phase wraps.
Figure 5: Relationship between Δθ and ΔB0 evaluated in the agar phantom.
(A) ΔB0 map obtained using the WASSR method, and the Δθ maps obtained from qDESS for TE1s equal to 5ms and 15ms. (B) and (C) show the scatter plots between the qDESS phase difference Δθ as a function of the reference ΔB0 evaluated along a horizontal line across the phantom (black dashed line in (A)). For the case in which TE1 was set equal to 15ms, the scatter plot is also presented after phase unwrapping was applied to the Δθ map (bottom of (C)). A linear model was fit to each scatter plot and the experimental slope of the model is reported along with the theoretical slope obtained by computing the denominator of eq. 2. The theoretical ΔB0 wrap-free bandwidth (Δν), for which the qDESS Δθ does not present wraps, is also reported for both the selected TE1s.
Figures 5 (B) and (C) show the scatter plots between the qDESS Δθ as a function of the reference ΔB0 evaluated along a horizontal line across the phantom (black dashed line in Fig. 5 (A)). For the case in which TE1 was set equal to 15ms, the scatter plot is also presented after phase unwrapping was applied to the Δθ map (bottom of Fig. 5 (C)). A linear model was fit to each scatter plot and the experimental slope was compared to the theoretical slope (the denominator of eq. 2). The theoretical ΔB0 wrap-free bandwidth, for which the qDESS Δθ does not present wraps, is also reported for both the selected TE1s. For both TE1 values, the theoretical slope was comparable to the experimental slope: 0.063rads vs 0.060rads, and 0.188rads vs 0.185rads for the case in which TE1 was set to 5ms and 15ms, respectively. When TE1 was set to 5ms, the phase difference did not present phase wraps, and the theoretical wrap-free bandwidth was from −50Hz to 50Hz. When TE1 was set to 15ms, the phase difference presented phase wraps in accordance with the theoretical wrap-free bandwidth, which was from −15Hz to 15Hz. Finally, in both cases, when ΔB0 was equal to 0 Hz, Δθ was not equal to 0rad but had a small positive value.
Doped water phantom
Figure 6 shows the comparison between the ΔB0 maps obtained with WASSR, qDESS and the 2-GRE method for the phantoms made of 6 vials of doped water. The ΔB0 maps are reported along with the difference between the reference WASSR map, and the BA plots. For completeness, the difference between the ΔB0 map obtained with the qDESS and the 2-GRE methods is also reported. The ΔB0 values spanned the range (−60Hz, 60Hz). Overall, the qDESS and 2-GRE methods produced ΔB0 maps that were in good agreement with those obtained with WASSR as visually highlighted by the quantitative maps, the difference maps, and statistically confirmed by the BA plots and corresponding statistics (Fig. 6 (B)). ρc was equal to 0.98 and 0.91 with a bias of −1Hz and 9.4Hz for qDESS and 2-GRE, respectively. The LOAs were in the order of ±8Hz for both the qDESS and 2-GRE methods. The qDESS ΔB0 map was also in good agreement with the 2-GRE map. The Lin’s coefficient was equal to 0.90 with a negative bias of −10.5Hz and LOAs of ±5.2Hz.
Figure 6: Results in Phantom.
[A] B0 maps obtained using the WASSR, qDESS, and 2-GRE methods (first row) and the difference maps with WASSR (second row). The third row shows the difference between B0 maps obtained with qDESS and the 2-GRE methods. [B] Bland-Altman plots for assessing the agreement of qDESS (first row) and 2-GRE (second row) with WASSR. The third row reports the BA plot between qDESS and 2-GRE. The solid line indicates the mean of the differences, while the dashed lines represent the upper and lower limits of agreement (LoA). Lin’s coefficients are displayed in each plot.
Validation in vivo
Phase unwrapping algorithms
Figure 7 summarizes the results and processing times of phase unwrapping with PRELUDE, SEGUE, and ROMEO for cases where at least one was unsuccessful. A sagittal representative slice is reported for the reference ΔB0 map, wrapped qDESS Δθ map and unwrapped qDESS Δθ maps obtained with the tested algorithms. In 3 out of 10 total knees, at least one phase unwrapping algorithm was unsuccessful.
Figure 7: In vivo results: phase unwrapping algorithms.
Representative sagittal slices for the knees in which at least one algorithm did not correctly perform phase unwrapping. The first column reports the reference ΔB0 map obtained with the WASSR method. The second column reports the wrapped qDESS phase difference along with the unwrapped maps obtained with: PRELUDE applying a cartilage mask (third column), SEGUE applying a cartilage mask (fourth column), SEGUE without applying a cartilage mask (firth column), ROMEO applying the cartilage mask (sixth column) and ROMEO applying without applying the cartilage mask. For each algorithm, the time required to process a knee is also reported.
PRELUDE. It was not possible to assess PRELUDE unwrapping accuracy without using the cartilage mask. No results were obtained in 5+ hours when not using the cartilage mask. When the cartilage mask was used, 1 over 10 knees were wrongly unwrapped (third row, third column in Fig. 7).
SEGUE. Using the cartilage mask provided the same result as PRELUDE. However, without applying the cartilage mask, SEGUE successfully unwrapped all the knees.
ROMEO. In 2 over 10 cases, ROMEO wrongly estimated the unwrapped phase when not applying the cartilage mask. Interestingly, the two knees belonged to the same subject (first and second row in Fig. 7), for which all the other methods correctly unwrapped the phase maps. On the other hand, with the cartilage mask, the phase was correctly unwrapped in all knees.
Based on these results, we concluded that SEGUE, run without applying the cartilage mask, and ROMEO, run applying the cartilage mask, were the most reliable methods. SEGUE was utilized to obtain the qDESS ΔB0 maps. Representative ΔB0 maps of sagittal slice for WASSR, qDESS, and 2-GRE methods are reported in Figure 8.
Figure 8: In vivo results: representative sagittal slices for a subject.
Left (A) and right (B) representative ΔB0 maps of a saggital slice obtained using the WASSR, qDESS and 2-GRE methods for a simultaneous bilateral knee acquisition. The WASSR and 2-GRE images were registered to the qDESS images. The maps are shown after the application of a mask that removed the background and regions of low signal intensity.
ΔB0 map comparison in articular cartilage
Figure 9 shows the comparison between the ΔB0 maps obtained with WASSR, qDESS, and the 2-GRE method for the left and right knees of a healthy subject. For each knee, the 2D projections of the ΔB0 maps are reported along with their difference with the reference WASSR map and the BA plots. Overall, the qDESS and 2-GRE methods produced ΔB0 maps in good agreement with those obtained with WASSR as visually highlighted by the 2D projected maps, and statistically confirmed by the BA plots and corresponding statistics (Fig. 9 A2 and B2). For the left knee, ρc was equal to 0.98 and 0.92 with a bias of −2.1Hz and 6Hz for qDESS and 2-GRE, respectively. The LOAs were on the order of ±6 Hz for both the qDESS and 2-GRE methods. For the right knee, ρc was equal to 0.74 and 0.85 with a bias of −8.1Hz and 5.3Hz for qDESS and 2-GRE, respectively. The LOAs were on the order of ±6Hz for both the qDESS and 2-GRE methods. Overall, although comparable in magnitude, the qDESS method had a negative bias with WASSR, whereas the 2-GRE method had a positive bias with WASSR.
Figure 9: In vivo results: example of ΔB0 in FC.
Left (A1) and right (B1) unrolled 2D ΔB0 projection maps in FC obtained using the WASSR, qDESS and 2-GRE methods for a simultaneous bilateral knee acquisition. A pixel-wise unrolled 2D difference map with WASSR is also displayed. BA plots for the left (A2) and right (B2) knees to evaluate the agreement between the WASSR method and the qDESS and 2-GRE methods, respectively. The BA plots were made using the average ΔB0 values measured in the 6 sub-regions of FC.
Figure 10 shows the BA plots for the comparison of the reference average ΔB0 values in the 6 cartilage regions for 5 subjects, obtained with the WASSR method, with the corresponding average ΔB0 values obtained with the qDESS (Fig. 10 (A)) and 2-GRE (Fig. 10 (B)) methods, respectively, produced by pooling together the data for all 5 subjects. Left and right knees were considered separately. For completeness, the comparison between ΔB0 values obtained with the qDESS and 2-GRE methods is also reported (Fig. 10 (A)). Regarding the comparison between qDESS and WASSR, for both the left and right knees, a negative bias of about −10Hz was observed with LOA in the order of ±10Hz. ρc was equal to 0.95 and 0.90 for the left and right knees, respectively. Based on Lin’s concordance coefficients, an overall very good agreement was found between the qDESS and the WASSR methods. The agreement was comparable to that observed between the 2-GRE and the WASSR methods in terms of LOA (in the order of ±10Hz) and ρc (0.90 and 0.98 for the left and right knee, respectively). Conversely to the qDESS method, a positive bias with the WASSR method was observed for the 2-GRE method, which was 11.7Hz and 2.8Hz for the left and right knees, respectively. The agreement between the qDESS and the 2-GRE methods is less strong than when each method is compared to the reference WASSR method. The Lin’s coefficient is 0.77 and 0.87 for the left and right knee, respectively.
Figure 10: In vivo results: pooled Bland-Altman plots.
BA plots for the comparison of the reference average ΔB0 values in the 6 FC regions, obtained with the WASSR method, with the corresponding average ΔB0 values obtained with the qDESS (A) and 2-GRE (B) methods, respectively. (C) BA plot between ΔB0 values obtained with the qDESS and 2-GRE methods. The BA plots are made by pooling together the average ΔB0 values of all the 5 subjects measured in the 6 sub-regions of FC, considering left and right knees separately.
Discussion
The Bloch simulations and the experiments run on the agar phantom validated eq. 2 and investigated the impact of SNR on ΔB0 accuracy. The slope of the linear relationship between ΔB0 and the phase accumulated between the first and second qDESS echos, Δθ, was proportional to 2TE1. Thus, in accordance with the theory, for the qDESS B0 mapping method, the choice of TE2 does not affect the relationship between ΔB0 and Δθ. In the Bloch simulations, the qDESS method showed good noise robustness for different combinations of TE1 and TE2. We run the simulations for a spin ensemble characterized by relaxation parameters similar to those of articular cartilage, which was the target of interest in this study. Thus, since SNR will depend on the actual relaxation parameters, the performance of the method might be different for the same combination of TE1 and TE2. However, the results reported in Fig. 5 showed good robustness to noise for the same combination of TE1 and TE2 tested with Bloch simulations, even for a compound with different relaxation properties than cartilage.
In addition, in the agar gel phantom, when ΔB0 was equal to 0Hz, Δθ was not equal to 0rad, which is the theoretical Δθ value for ΔB0 = 0Hz according to eq. 2. This is because the reference ΔB0 was evaluated with WASSR, for which a bias (order of a few Hz) with the qDESS method was observed both in phantom and in vivo acquisitions. Although the origin of the bias was not directly investigated, biases among B0 mapping methods have been previously observed in the literature48. The actual phase accumulation in qDESS can be altered by hardware imperfections such as eddy currents. Furthermore, in the agar phantom, we observed that using long TE1 and TE2 produced a stronger agreement with WASSR than using shorter echo times. This might be because, with a longer time for phase accumulation, phase biases introduced by hardware imperfections might be mitigated more. At the same time, utilizing longer echo times increases scan time and exacerbates phase unwrapping. Nonetheless, ΔB0 maps evaluated using the proposed qDESS method were in good agreement with those obtained using the WASSR method, both in phantom and in-vivo. The observed bias with the reference WASSR method is considered acceptable with the aim of implementing a B0 correction for qDESS T2 mapping. The Bloch simulations investigating the effect of ΔB0 off-resonances in qDESS T2 estimation, reported in Fig. 2 (B), showed that an error in ΔB0 estimation of about ±10Hz underestimates or overestimates T2 by less than 1ms, which is below the reproducibility of T2 estimation with qDESS. Furthermore, the agreement between qDESS and WASSR was comparable to that of a standard 2-GRE method and WASSR.
The reported results demonstrated that we could obtain a reliable ΔB0 map using qDESS, a widely applied sequence in knee MRI that allows combining high-resolution morphological information with quantitative T2 mapping. The ability to obtain ΔB0 map can further benefit the impact of qDESS for qMRI knee imaging protocols for multiple reasons. First, the Bloch simulations shown in Fig.2(B) showed that when ΔB0 was above the ±60Hz range, the error in T2, measured with qDESS, was higher than 1.5ms. This error may be relevant as studies investigating the longitudinal variation of T2 in cartilage in knee OA have found variations in the order of 1.5 – 2ms49,50. Similarly, longitudinal studies of T2 in knee cartilage after ACL reconstruction for over a period of 24 months have observed average differences of about 5 ms between T2 at 6 months after surgery and at 24 months51. As we observed ΔB0 values above the ±60Hz range in each subject, these preliminary considerations suggest that taking into account ΔB0 in qDESS T2 mapping might be beneficial for accurate quantification of longitudinal changes. The proposed ΔB0 mapping method will allow such a correction without the need for additional scan time, with the benefit of having inherently co-registered images.
Future studies could systematically investigate the impact of a B0 correction on qDESS T2 mapping, which may be particularly relevant for simultaneous bilateral knee imaging, where B0 is normally less homogeneous compared to unilateral acquisitions. Furthermore, our method can be used to shorten qMRI protocols that require an ancillary scan for B0 mapping to correct the quantitative analysis, such as gagCEST imaging, DWI and to check if the right condition exists for effective fat saturation, chemical shift-based imaging methodologies or simultaneous fat-water imaging. Since a qDESS sequence is normally already acquired in a knee MRI protocol, the saved scan time may allow for cheaper MRI exams, or it can be exploited to enrich the MRI protocol with sequences that will generate information that are biologically meaningful instead of merely correct for hardware imperfections.
The fact that the dynamic range of off-resonances for which qDESS Δϕ does not present wraps is considerably smaller than a standard 2-GRE method can potentially be a limitation of the proposed methodology. In our study, since TE1 was set equal to 5 ms, the field dynamic range was ± 50 Hz. For most of the knees, off-resonance values reached values in the order of 60–80 Hz in the trochlea region of FC, while in other regions, the values were within the field dynamic range for which phase wraps did not occur. By comparing the performance of three state-of-the-art phase unwrapping algorithms it was interesting to observe that only ROMEO and SEGUE, run with and without and applying the cartilage mask, respectively, correctly unwrapped all the Δϕ maps. However, since SEGUE is based on PRELUDE, it is reasonable to think that the same result would have been achieved by PRELUDE without applying the cartilage mask to the phase difference map. Nonetheless, even without considering the background and areas of low intensity (such as bones), we were unable to perform phase unwrapping in a reasonable time without applying the cartilage mask with PRELUDE. Thus, our experiments showed that phase wraps could be a potential limitation for effective qDESS ΔB0 mapping in MSK, but with the proper phase unwrapping algorithm, we were able to restore the correct phase difference maps for all the knees under study. However, the application of the proposed method to body regions where a higher B0 heterogeneity is expected, such as the hip, might be further challenged by phase wraps that might not be easily recovered. One solution would be selecting a shorter TE1, which would increase the field dynamic range and mitigate phase wraps. However, there are limitations to the shortest TE1 that can be used, and, depending on the application, such a short TE1 might not be ideal for SNR optimization and T2 mapping accuracy. Future studies should investigate the application of the proposed method in other regions of the body to better characterize the limitations of the current method and the feasibility of strategies to overcome such limitations.
Conclusions
A method for accurately measuring B0 field inhomogeneity with qDESS was proposed and validated in phantom and in-vivo against the gold standard WASSR B0 mapping method. The proposed qDESS method provided ΔB0 maps that were in good agreement with those obtained using the reference WASSR method both in phantom and in-vivo. The proposed methodology may allow for implementing a B0 correction method for qDESS T2 mapping using an inherently co-registered ΔB0 map without additional scan time. More generally, the method may help shorten knee imaging protocols requaring an auxiliary ΔB0 map, such as gagCEST imaging, by exploiting a qDESS acquisition that also provides T2 measurements and high-quality morphological imaging.
Supplementary Material
Figure S1 qDESS first (left) and second (right) echoes of a representative slice of the agar gel phantom. Spots of lower signal intensity within the phantom are due to air bubbles.
Acknowledgments
This work was supported by GE Healthcare, Philips, and NIH grants R01-AR077604, R01-EB002524, R01-AR074492, K24-AR062068, R21-EB030180, and R00-EB022634.
Data Availability Statement
The MRI data that support the findings of this study are openly available in Zeonodo at http://doi.org/10.5281/zenodo.7034725.
References
- 1.Fang Liu, Rajeev Chaudhary, Hurley Samuel A., et al. Rapid multicomponent T2 analysis of the articular cartilage of the human knee joint at 3.0T. J. Magn. Reson. Imaging 2014;39(5):1191–1197. [DOI] [PubMed] [Google Scholar]
- 2.Wen Ling, Regatte Ravinder R., Navon Gil, Jerschow Alexej. Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST). Proc. Natl. Acad. Sci. 2008;105(7):2266–2270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Chaudhari Akshay S, Feliks Kogan, Pedoia Valentina, Majumdar Sharmila, Gold Garry E., Hargreaves Brian A. Rapid Knee MRI Acquisition and Analysis Techniques for Imaging Osteoarthritis. Journal of Magnetic Resonance Imaging. 2020;52(5):1321–1339. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Majumdar S, Orphanoudakis SC, Gmitro A, O’Donnell M, Gore JC Errors in the measurements of T2 using multiple-echo MRI techniques. II. Effects of static field inhomogeneity. Magn. Reson. Med. 1986;3(4):562–574. [DOI] [PubMed] [Google Scholar]
- 5.Sled John G., Pike G. Bruce. Correction for B1 and B0 variations in quantitative T2 measurements using MRI. Magnetic Resonance in Medicine. 2000;43(4):589–593. [DOI] [PubMed] [Google Scholar]
- 6.Sun Phillip Zhe, Farrar Christian T., Sorensen A. Gregory Correction for artifacts induced by B0 and B1 field inhomogeneities in pH-sensitive chemical exchange saturation transfer (CEST) imaging. Magn. Reson. Med. 2007;58(6):1207–1215. [DOI] [PubMed] [Google Scholar]
- 7.Digma Leonardino A, Feng Christine H, Conlin Christopher C, et al. Correcting B0 inhomogeneity-induced distortions in whole-body diffusion MRI of bone. Sci. Rep. 2022;12(1). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Gerdes Clint M, Kijowski Richard, Reeder Scott B. IDEAL Imaging of the Musculoskeletal System: Robust Water–Fat Separation for Uniform Fat Suppression, Marrow Evaluation, and Cartilage Imaging. American Journal of Roentgenology. 2007;189(5):W284–W291. PMID: 17954627. [DOI] [PubMed] [Google Scholar]
- 9.Sebastien Bacher, David Hajdu Steven, Yael Maeder, Vincent Dunet, Tom Hilbert, Patrick Omoumi. Differentiation between benign and malignant vertebral compression fractures using qualitative and quantitative analysis of a single fast spin echo T2-weighted Dixon sequence. Eur. Radiol. 2021;31(12):9418–9427. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Beata Bachrata, Bernhard Strasser, Wolfgang Bogner, et al. Simultaneous Multiple Resonance Frequency imaging (SMURF): Fat-water imaging using multi-band principles. Magn. Reson. Med. 2021;85(3):1379–1396. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Mina Kim, Joseph Gillen, Landman Bennett A., Zhou Jinyuan, Zijl Peter C.M. Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magn. Reson. Med. 2009;61(6):1441–1450. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Peter Webb, Daniel Spielman, Albert Macovski. Inhomogeneity correction for in vivo spectroscopy by high-resolution water referencing. Magn. Reson. Med. 1992;23(1):1–11. [DOI] [PubMed] [Google Scholar]
- 13.Peterfy CG, Schneider E, Nevitt M. The osteoarthritis initiative: report on the design rationale for the magnetic resonance imaging protocol for the knee. Osteoarthr. Cartil. 2008;16(12):1433–1441. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Bruder H, Fischer H, Graumann R, Deimling M. A new steady-state imaging sequence for simultaneous acquisition of two MR images with clearly different contrasts. Magn. Reson. Med. 1988;7(1):35–42. [DOI] [PubMed] [Google Scholar]
- 15.Redpath TW, Jones RA FADE–A new fast imaging sequence. Magnetic Resonance in Medicine. 1988;6(2):224–234. [DOI] [PubMed] [Google Scholar]
- 16.Lee SY, Cho ZH Fast SSFP gradient echo sequence for simultaneous acquisitions of FID and echo signals. Magnetic Resonance in Medicine. 1988;8(2):142–150. [DOI] [PubMed] [Google Scholar]
- 17.Eckstein F, Westhoff J, Sittek H, et al. In vivo reproducibility of three-dimensional cartilage volume and thickness measurements with MR imaging.. 10.2214/ajr.170.3.9490936. 2013;170(3):593–597. [DOI] [PubMed] [Google Scholar]
- 18.Moriya S, Miki Y, Yokobayashi T, Ishikawa M. Three-dimensional double-echo steady-state (3D-DESS) magnetic resonance imaging of the knee: Contrast optimization by adjusting flip angle:. http://dx.doi.org.stanford.idm.oclc.org/10.1080/02841850902849444. 2009;50(5):507–511. [DOI] [PubMed] [Google Scholar]
- 19.Hardy Peter A, Recht Michael P, Piraino David, Thomasson David. Optimization of a dual echo in the steady state (DESS) free-precession sequence for imaging cartilage. J. Magn. Reson. Imaging 1996;6(2):329–335. [DOI] [PubMed] [Google Scholar]
- 20.Welsch Goetz H, Scheffler Klaus, Mamisch Tallal C., et al. Rapid estimation of cartilage T2 based on double echo at steady state (DESS) with 3 Tesla. Magn. Reson. Med. 2009;62(2):544–549. [DOI] [PubMed] [Google Scholar]
- 21.Staroswiecki Ernesto, Granlund Kristin L., Alley Marcus T., Gold Garry E., Hargreaves Brian A. Simultaneous estimation of T2 and apparent diffusion coefficient in human articular cartilage in vivo with a modified three-dimensional double echo steady state (DESS) sequence at 3 T. Magn. Reson. Med. 2012;67(4):1086–1096. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Sveinsson B, Chaudhari AS, Gold GE, Hargreaves BA A simple analytic method for estimating T2 in the knee from DESS. Magnetic Resonance Imaging. 2017;38:63–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Chaudhari Akshay S, Black Marianne S, Eijgenraam Susanne, et al. Five-minute knee MRI for simultaneous morphometry and T2 relaxometry of cartilage and meniscus and for semiquantitative radiological assessment using double-echo in steady-state at 3T. J. Magn. Reson. Imaging 2018;47(5):1328–1341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Hennig Jürgen. Echoes—how to generate, recognize, use or avoid them in MR-imaging sequences. Part I: Fundamental and not so fundamental properties of spin echoes. Concepts in Magnetic Resonance. 1991;3(3):125–143. [Google Scholar]
- 25.Jürgen Hennig. Echoes—how to generate, recognize, use or avoid them in MR-imaging sequences. Part II: Echoes in imaging sequences. Concepts in Magnetic Resonance. 1991;3(4):179–192. [Google Scholar]
- 26.Matthias Weigel. Extended phase graphs: Dephasing, RF pulses, and echoes - pure and simple. Journal of Magnetic Resonance Imaging. 2015;41(2):266–295. [DOI] [PubMed] [Google Scholar]
- 27.Feliks Kogan, Evan Levine, Chaudhari Akshay S., et al. Simultaneous bilateral-knee MR imaging. Magn. Reson. Med. 2018;80(2):529–537. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Marco Barbieri, Desai Arjun D., Kogan Feliks, et al. Rapid Quantitative Bilateral Knee Imaging with Fully Automated Femoral Cartilage Analysis. In: Proc. of ISMRM 27th Annual Meeting and Exibition; 11–16 May 2019; Montreal, Canada. [Google Scholar]
- 29.Chaudhari Akshay S, Stevens Kathryn J, Sveinsson Bragi, et al. Combined 5-minute double-echo in steady-state with separated echoes and 2-minute proton-density-weighted 2D FSE sequence for comprehensive whole-joint knee MRI assessment. Journal of Magnetic Resonance Imaging. 2019;49(7):e183–e194. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Chaudhari Akshay S, Grissom Murray J., Fang Zhongnan, et al. Diagnostic Accuracy of Quantitative Multicontrast 5-Minute Knee MRI Using Prospective Artificial Intelligence Image Quality Enhancement. American Journal of Roentgenology. 2021;216(6):1614–1625. PMID: 32755384. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Jochen Leupold. Steady-state free precession signals of arbitrary dephasing order and their sensitivity to T*2. Concepts Magn Reson. Part A 2017;46A(1):e21435. [Google Scholar]
- 32.Marco Barbieri, Chaudhari Akshay S., Moran Catherine J., Gold Garry E., Hargreaves Brian A., Kogan Feliks. A Method for Measuring B0 Field Inhomogeneity using Quantitative DESS (qDESS). In: Proc. of ISMRM 29th Annual Meeting and Exibition; 15–20 May 2021. [Google Scholar]
- 33.Scharnhorst K. Angles in Complex Vector Spaces. Acta Appl. Math. 2001 691 2001;69(1):95–103. [Google Scholar]
- 34.Marco Barbieri, Leonardo Brizi, Paola Fantazzini, Stan Sykora, Claudia Testa. A flexible software for NMR pulse sequence simulations, first applications to MR-Fingerprinting: Quantification of myelin water fraction in white matter and diffusion coefficient measurement. In: ISMRM workshop on Magnetic Resonance Fingerprinting; October 2017; Cleveland, OH, USA. [Google Scholar]
- 35.Zixin Zhao, Hangying Zhang, Zhaoxian Xiao, et al. Robust 2D phase unwrapping algorithm based on the transport of intensity equation. Meas. Sci. Technol. 2019;30(1):015201. [Google Scholar]
- 36.Wenbo Wei, Guang Jia, David Flanigan, Jinyuan Zhou, Knopp Michael V.. Chemical exchange saturation transfer MR imaging of articular cartilage glycosaminoglycans at 3 T: Accuracy of B0 Field Inhomogeneity corrections with gradient echo method. Magn. Reson. Imaging 2014;32(1):41–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Robinson Simon Daniel Bredies Kristian, Diana Khabipova, Barbara Dymerska, Marques José P., Schweser Ferdinand. An illustrated comparison of processing methods for MR phase imaging and QSM: combining array coil signals and phase unwrapping. NMR Biomed..2017;30(4):e3601. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Jie Luo, Jagadeesan Bharathi D., Cross Anne H., Yablonskiy Dmitriy A. Gradient Echo Plural Contrast Imaging — Signal model and derived contrasts: T2*, T1, Phase, SWI, T1f, FST2*and T2*-SWI. Neuroimage. 2012;60(2):1073–1082. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Desai Arjun D, Barbieri Marco, Mazzoli Valentina, et al. DOSMA: A deep-learning,open-source framework for musculoskeletal MRI analysis. In: Proc. of ISMRM 27th Annual Meeting and Exibition; 11–16 May 2019; Montreal, Canada. [Google Scholar]
- 40.Desai Arjun D, Gold Garry E, Hargreaves Brian A, Chaudhari Akshay S. Technical Considerations for Semantic Segmentation in MRI using Convolutional Neural Networks. arXiv:1902.01977v1 [eess.IV]. 2019;. [Google Scholar]
- 41.Monu UD, Jordan CD, Samuelson BL, Hargreaves BA, Gold GE, McWalter EJ Cluster analysis of quantitative MRI T2 and T1ρ relaxation times of cartilage identifies differences between healthy and ACL-injured individuals at 3T. Osteoarthr. Cartil. 2017;25(4):513–520. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Stefan Klein, Marius Staring, Keelin Murphy, Viergever Max A., Pluim Josien P.W..Elastix: A toolbox for intensity-based medical image registration. IEEE Trans. Med. Imaging 2010;29(1):196–205. [DOI] [PubMed] [Google Scholar]
- 43.Shamonin Denis P, Bron Esther E, Lelieveldt Boudewijn PF., Smits Marion, Klein Stefan, Staring Marius. Fast Parallel Image Registration on CPU and GPU for Diagnostic Classification of Alzheimer's Disease. Front. Neuroinform. 2014;0(JAN):50. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Mark Jenkinson. Fast, automated, N-dimensional phase-unwrapping algorithm. Magn. Reson. Med. 2003;49(1):193–197. [DOI] [PubMed] [Google Scholar]
- 45.Anita Karsa, Karin Shmueli. SEGUE: A Speedy rEgion-Growing Algorithm for Unwrapping Estimated Phase. IEEE Trans. Med. Imaging 2019;38(6):1347–1357. [DOI] [PubMed] [Google Scholar]
- 46.Barbara Dymerska, Korbinian Eckstein, Beata Bachrata, et al. Phase unwrapping with a rapid opensource minimum spanning tree algorithm (ROMEO). Magn. Reson. Med. 2021;85(4):2294–2308. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Lin Lawrence I-Kuei. A Concordance Correlation Coefficient to Evaluate Reproducibility. Biometrics. 1989;45(1):255–268. [PubMed] [Google Scholar]
- 48.Rodriguez Esau Poblador, Moser Philipp, Dymerska Barbara, et al. A comparison of static and dynamic ΔB0 mapping methods for correction of CEST MRI in the presence of temporal B0 field variations. Magn. Reson. Med. 2019;82(2):633–646. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Judong Pan, Pialat Jean Baptiste Joseph Tom, et al. Knee cartilage T2 characteristics and evolution in relation to morphologic abnormalities detected at 3-T MR imaging: A longitudinal study of the normal control cohort from the osteoarthritis initiative. Radiology. 2011;261(2):507–515. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Kretzschmar M, Heilmeier U, Yu A, et al. Longitudinal analysis of cartilage T2 relaxation times and joint degeneration in African American and Caucasian American women over an observation period of 6 years- Data from the Osteoarthritis Initiative. Osteoarthritis Cartilage. 2016;24(8):1384. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Lansdown Drew A, Xiao Weiyuan, Zhang Alan L., et al. Quantitative imaging of anterior cruciate ligament (ACL) graft demonstrates longitudinal compositional changes and relationships with clinical outcomes at 2 years after ACL reconstruction. J. Orthop. Res. 2020;38(6):1289–1295. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Figure S1 qDESS first (left) and second (right) echoes of a representative slice of the agar gel phantom. Spots of lower signal intensity within the phantom are due to air bubbles.
Data Availability Statement
The MRI data that support the findings of this study are openly available in Zeonodo at http://doi.org/10.5281/zenodo.7034725.









