Abstract
Metals, acrylics, zirconia, and other such materials have been conventionally used in dentistry. The development of polymers has facilitated significant changes in clinical dentistry. High-performance polymer materials are at the forefront of dentistry. Polyaryletherketone (PAEK) is a biocompatible polycyclic, aromatic, thermoplastic polymer having good mechanical and thermal properties. It has two members: polyetheretherketone (PEEK) and polyetherketoneketone (PEKK). The difference in the ratio and structure of ether to ketone group affects the melting point and glass transition temperature. PEEK and PEKK have a high impact with their physical and mechanical characteristics similar to that of the bone and mimic the natural tooth structure. It can be used as a substitute for metals and other materials owing to its non-allergic properties and acceptable aesthetics. Currently, to modify the properties of both materials, additives were used. This semicrystalline structure does not provide any kind of mutagenicity and cytotoxicity. This review provides insight into the properties and applications of polymer in dentistry and the medical field as well. There is room for metal-free restorations in modern dental practice due to the rising demand for aesthetics, a few disadvantages with existing materials, and clinicians changing their paradigms toward metal-free restorations. The objective of this review is to provide a thorough understanding of PEEK and PEKK and their multiple uses in prosthetic, implant frameworks, abutments, crowns, and ortho wires, as well as in restorative dentistry, while demonstrating their potential for clinical applications.
Keywords: bioactivity, implant material, polyetherketoneketone polymer, polyetheretherketone polymer, polyaryletherketone polymer
Introduction and background
In order to accomplish the ideal, dentistry has been actively evolving its methods and utilizing advanced materials. Metals, acrylics, zirconia, and other such materials have been conventionally used in dentistry. The development of polymers has facilitated significant changes in clinical dentistry [1]. It has great mechanical and physical properties, and it is biocompatible in nature [2]. In search of newer materials and to overcome the limitations of materials used in day-to-day life, more advanced materials are introduced; one of them is polyaryletherketone (PAEK). Since the 1980s, PAEK is used in the engineering field showing great machinability [3]. It is a thermoplastic polymer and performs extraordinarily well with its effective mechanical and chemical resistance [4]. High-performance polymer materials are at the forefront of dentistry research because they might enhance framework properties and possibly lower rehabilitation costs.
PAEK is a semicrystalline thermoplastic polymer; it has a glass transition temperature of about 157°C and a melting temperature of 370°C [5]. PAEK polymer contains ether and ketone groups. The difference in the ratio and structure of ether to ketone group affects the melting point and glass transition temperature. The more the ketone group, the more the polarity and rigidity and thus the more the glass transition temperature and melting point. PAEK has two family members; the first is polyetheretherketone (PEEK), the monomer unit of ether ether ketone, and the other one is polyetherketoneketone (PEKK), the monomer unit of ether ketone ketone. Polymers with semicrystalline structure do not provide any kind of mutagenicity and cytotoxicity [2-5]. PEEK and PEKK have a high impact with their physical and mechanical characteristics similar to that of the bone.
Additionally, it was discovered that PEEK has biocompatible qualities [2]. The combination of positive in vivo and in vitro results made polymer very popular for medical uses such as orthopedic and social implants, though its potential for dental applications has also been studied for over a decade. Its use has been recommended for a variety of intraoral fixed and removable prosthesis and restoration. It can also be used as ortho wires because it might exert more favorable forces than standard wires [6]. Recently, PEKK, which is also a biocompatible high-performance polymer (Bio-HPP), was shown as an innovative dental material. It is a potential replacement to metal and glass ceramics in dental application because of its fracture resistance, improved stress distribution, and shock-absorbing properties. Its aforementioned characteristics make it highly compatible in the medical industry. Though one of the organic polymers of PAEK family is commonly used as implant material, it has been acknowledged as a suitable substitute for long-established titanium in ortho application [3]. The objective of this review is to provide a thorough understanding of PEEK and PEKK and their multiple uses in prosthetic, implant frameworks, abutments, crowns, and ortho wires, as well as in restorative dentistry, while demonstrating their potential for clinical applications [2].
Review
PAEK, a semicrystalline polycyclic, aromatic, thermoplastic polymer, has five family members. This polymer has monomers and differs in ratio and structure of ether to ketone group: polyetherketone (PEK), polyetheretherketone (PEEK), polyetherketoneketone (PEKK), polyetheretherketoneketone (PEEKK), and polyetherketoneetherketoneketone (PEKEKK). PEEK and PEKK are polymers of PAEK family that can be used widely in dentistry (Figure 1). Further article shows a detailed review of PEEK and PEKK including structure, synthesis, properties, and applications.
Figure 1. Showing two family members of PAEK.
Structure and synthesis of PEEK and PEKK
PEEK is a semicrystalline thermoplastic, polycyclic, aromatic polymer. When 4,4'-difluorobenzophenone is added with disodium hydroquinone at temperature of 300°C and melting point of 335°C, PEEK is formed (Figure 2 and Figure 3) [7]. The manufacturing of PEEK polymer is done by injection molding, compression molding, computer-aided design/computer-assisted manufacture (CAD/CAM), or rapid prototyping.
Figure 2. Synthesis of PEEK.
PEEK: polyetheretherketone
Source: [8]
Figure 3. Structure of PEEK polymer.
PEEK: polyetheretherketone
PEKK is an ultra-high-molecular-weight polymer with linear aromatic polyether ketone group introduced by Bonner in the year 1962 [9]. PEKK is a product of diphenyl ether and terephthaloyl chloride in which aluminum chloride and nitrobenzene are added (Figure 4 and Figure 5). Compression molding and particle leaching are such process used by Converse et al. for developing a PEKK [10]. It differs from PEEK in the presence of extra ketone group in its aromatic ring, which increases polarity and backbone rigidity of the material. PEKK has a melting point of 305°C [2].
Figure 4. Synthesis of PEKK.
PEKK: polyetherketoneketone
Figure 5. Structure of PEKK polymer.
PEKK: polyetherketoneketone
Properties of PEEK and PEKK
PAEK polymer shows various enhanced physical, mechanical, and biological properties (Table 1) [2,11-15].
Table 1. Properties of PEEK and PEKK polymer.
PEEK: polyetheretherketone; PEKK: polyetherketoneketone; VHN: Vickers hardness number; FEFF1.3: free energy force field 1.3
| Properties | PEEK | PEKK |
| Physical properties | Biocompatible, minimal plaque affinity, aesthetics, dimensional stability, good polishability, and wear resistance | Biocompatible, shock absorbance, aesthetics, flexibility, microporosity, and hydrophilic nature |
| Mechanical properties | ||
| Tensile strength (MPa) | 100.69 | 115 |
| Elastic modulus (GPa) | 3.5 | 5.1 |
| Flexural strength (MPa) | 163.88 | 200 |
| Compressive strength (MPa) | 118-169 | 246 |
| Melting temperature (°C) | 334-350 | 363-386 |
| Hardness | 26-29 VHN | 252 MPa |
| Water absorption (mg/mm3) | 0.1-0.5 | 8.7 |
| Density (g/cm3) | 1.3 | FEFF1.3 |
| Biological properties | Osteointegration property | Osteointegration property and antibacterial |
Applications of PEEK and PEKK
There are several applications of PEEK and PEKK due to its various benefits in dentistry. It can be used widely as implants, prosthetic material, ceramic filler, and so on. Detailed explanatory benefits of PEEK and PEKK are shown further.
Implant Material
PEEK is a material that can be utilized as implants. Titanium is relentlessly used as an implant material, but due to some of its limitations, PEEK has been introduced as such; titanium is metal, whereas PEEK is metal-free material. Titanium is heavier, and PEEK is lighter in weight; also, titanium causes hypersensitivity as it is a metal. The osseointegration property of PEEK is good, but PEEK material alone does not have certain enhancing properties. It requires certain surface modifications to increase its osseoconductive property significantly [16,17]. The best surface treatment for polymer is carbon fiber reinforced (CFR), which increases its modulus of elasticity nearer to cortical bone [18]. According to documentation, the elastic modulus of PEEK is 3.6 GPa, but after reinforcing it with carbon fibers, elastic modulus increases to 18 GPa. The elastic modulus of cortical bone is 15 GPa. In comparison with elastic modulus of titanium that is 110 GPa, the elastic modulus of PEEK is close to the elastic modulus of the bone [17,19,20]. PEEK also has flexural strength of 140-170 MPa [17]. Many modifications are done with PEEK polymer such as spin coating, that is, covering a PEEK surface with nanoparticles of calcium hydroxyapatite, and get incorporated into the implant surface creating a better bone implant contact [21]. Another method for CFR-PEEK is by plasma spraying with titanium oxide and hydroxyapatite crystals; then, immersing it in alpha tricalcium phosphate shows better properties, and this was currently given by Nakahara et al. [22]. It is also modified with tantalum ions to form Ta2O5 nanoparticles, which provides better osteoconductivity, elasticity, and attachment with cortical bone, in which PEEK limits as it is bioinert in nature [23]. PEEK can also be modified by processing it with chemicals by amination, sulfonation, and nitration in pre-polymerization or post-polymerization [7]. Sulfonation is one of the methods incorporated with PEEK polymer for preventing the formation of biofilm, in return making it Staphylococcus aureus and Escherichia coli resistant [24]. Due to mastication or any parafunctional habits such as bruxism, titanium abutment or screws are replaced by PEEK as it does not cause hypersensitivity unlike titanium and can bear forces such as titanium [25].
PEKK is another member of PAEK family that also shows better physical, mechanical, and biological characteristics. It is also used as an alternative to titanium in implants because it shows better compatibility with other materials. PEKK already has better elastic modulus than PEEK; even surface treatment is not much required. But to make it more effective, additives with PEKK were introduced. Sulfonation was done with PEKK; also, it was observed that PEKK was having early healing of the bone and good osteointegration property compared to sulfonated PEEK. The combination of PEKK and titanium is used to provide retention for a very long period of time [26]. The combination of PT40, that is, PEKK and tantalum with 40 v%, is also used, as the ability of PT40 to bear load and serve as bony substitute is excellent [27]. After temporalis myofascial flap surgery, hollowing of the bone was replaced by temporal implants achieving smooth healing process, and patients found it to be aesthetically pleasing [28]. PEKK offers better retention when inserted in overdentures. Other than dentistry, PEKK is also used as orthopedic implants [29].
Prosthetic Material
Metals such as cobalt-chromium and titanium were previously utilized as prosthesis; however, they are aesthetically unacceptable, heavy, and metallic in taste, and some people are allergic to it. PEEK polymer was introduced to use it as a prosthetic material in removable partial dentures (RPDs) and fixed partial dentures (FPDs), crowns, and so on. In removable prosthesis, it can be used as clasps and can replace cobalt-chromium alloys [30]. Obturators are also prosthetic devices that are generally made up of acrylic, but the superior properties of PEEK can overcome these [31]. For people who are partially edentulous, RPDs involve replacing teeth and other structures with artificial ones. When cobalt-chromium, titanium alloy, and PEEK were compared, PEEK was more useful for the protection of periodontal fibers, as well as distributed masticatory force evenly all over the denture base [32]. However, some studies are still going on to prove the ability of the polymer. CAD/CAM is the smart advance for making digital impression and to reduce chairside time. Enhanced biocompatibility, the elimination of allergic reactions, durability, and elasticity, as well as being metal-free, make it more aesthetically appealing, and economical advantages would be made possible by combining polymer-based materials and digital fabrication techniques [33]. Polymethylmethacrylate (PMMA) is used to make FPDs machined with CAD/CAM, but PEEK is the different material that can be used as an alternative to PMMA. Fracture resistance of PEEK is higher than those of lithium disilicate glass-ceramic (950N), alumina (851N), and zirconia (981-1331N); also, the abrasive property of PEEK is fantastic [34-37]. PEEK shows resistance to chemical and radiation as well, to prevent damage from glass and carbon fibers [38]. PEEK with nanoparticles of zirconia is another modified form that shows low wear and friction resistance. Nanoparticles such as titanium dioxide, fluorapatite, and hydroxyapatite were added to PEEK in order to increase their elastic recovery, antibacterial property, stability, and biocompatibility [39,40].
PEEK is a better material than metals, but PEKK shows superior shock absorbance and less stress concentration of 49 MPa, whereas PEEK shows higher stress on the base of prosthesis of 58 MPa [41]. The lesser the stress concentration, the lesser the fracture risk on the denture base. PEKK is the topmost member of PAEK family and the resultant of amorphous and semicrystalline structure. PEKK has lower crystallization rate, and it is less affected by cooling when heated in lower temperature, so irrespective of CAD/CAM and conventional denture-making procedures, it can be machined by printers with a low build chamber temperature of less than 200°C [42-44]. Retention is the most important factor for dentures. In removable prosthesis, clasps made up of PEKK can be used for longer duration and also can be used as an alternative to nylon inserts as it is more retentive and abrasive [45]. Acrylic resin shows poor marginal fit after a decade, which causes plaque retention, recurrent caries, periodontal problems, and prosthetic failure. CAD/CAM-based digital technique with PEKK polymer along with surface modifications makes higher retentive and perfectly fitted denture. Thus, PEKK coping was proved better than zirconia coping [46,47]. PEKK is also used as obturator such as in removable speech bulb prosthesis [48]. PEEK and PEKK polymers can be used in making crowns and veneers. Another enhanced form of PEEK is biocompatible high-performance polymer (Bio-HPP) with the addition of ceramic filler of 20% with size of 0.3-0.5 μm in resin matrix, given by Bredent in Germany [49]. Bio-HPP PEEK has excellent polishing capabilities, low plaque affinity, nonmetallic taste, and strong wear resistance, which are all characteristics that make it anti-allergic [50]. PEEK and PEKK can be used with jiffy dentures to produce interim dentures or intermediate dentures.
Endodontic Material
Whenever more than 50% of the tooth structure is lost and the tooth cannot resist torsional stress, post and core is used. There are two types of posts: metallic and nonmetallic. Titanium, stainless steel, and gold alloys are metals used to fabricate post; due to increasing demands of aesthetics and allergic reactions of patients to metal, metallic posts have been disregarded. Other types of post are nonmetallic such as zirconia, ceramic, carbon fiber, and glass fiber. These materials have their certain limitations such as high modulus of elasticity, brittleness, and leaving voids on the surface; exposure to moisture will lead to failure and can cause root fracture. To overcome all limitations of such materials, a low-elastic-modulus, higher-strength, tissue-compatible material of PAEK family was smartly made. In endodontically treated tooth, PEEK polymer can be used as a post material as it will provide less stress on the tooth structure, as well as in the core material [51].
For restoration purpose, as PEEK is bioinert, the addition of hydroxyapatite makes it bioactive and hydrophilic in nature [52]. In order to prove the ability of fracture resistance of PEEK polymer, a comparative study was done between polymer-infiltrated ceramic, fiber-reinforced composite post, and PEEK. Higher fracture resistance was shown by PEEK polymer among all aesthetic post material [53]. Since the PEEK material's elastic modulus is so near to that of dentin tissue, less stress development in the tooth and after core restoration than in conventional post systems was observed [54]. For irrigation during root canal treatment, tips used to irrigate the canals can be made from PEEK polymer as it is chemical resistant and does not react with irrigants such as sodium hypochlorite, chlorhexidine, and ethylenediaminetetraacetic acid [14,55].
PEKK and polymer other than PEEK can also be used as endodontic posts and endo crowns [56,57]. It also possess similar properties of that of PEEK in relation with the fabrication of endodontic post. PEKK has modulus of elasticity similar to that of dentin and mimics natural tooth structure. PEKK has low modulus of elasticity and flexural strength compared with other metallic and nonmetallic conventional post materials, which helps in less stress distribution and dispersion of stress in intraradicular space in order to prevent the fracture of root surface [56]. These also require ferrule of at least 1.5 mm to be present for fracture resistance. There are two methods of fabrication of post: prefabrication and custom-made. When compared to prefabricated PEKK posts, custom-made PEKK posts showed stronger bonds [57].
Orthodontic Material
Materials usually used for ortho wires and brackets are acrylic, metals, or ceramics. PEEK and PEKK are the materials that can replace metal ortho wires and provide proper force normally required. Other than metal, ceramic brackets are also used as brackets for aesthetic purpose. Ceramic brackets have some shortcomings in relation with properties such as strength, and the force applied by brackets made up of ceramic is comparatively less, which in turn increases the treatment span. To overcome this limitation, PEEK is a more feasible material that can be used. CAD/CAM, which is a new digital technology, when combined with PEEK polymer can be used as an alternative to conventional self-cure and heat cure space maintainers, making it favorable and aesthetically pleasing for the patient [58]. Besides aesthetics, metal ortho wires, brackets, and appliances have certain limitations. For example, metal releases nickel and chromium ions. When metal wire and brackets are coupled, they undergo galvanic reaction and corrosion. PEEK wire is a material that shows promise in terms of stain resistance. 16-22 PEEK and 19-25 PEEK had greater load reductions than nickel-titanium (Ni-Ti). Intentionally, 70% and 80% of the initial load was kept, and the preserved amount was adequate for orthodontic treatment. Thus, PEEK can be used as an alternative to metal for better orthodontic treatment [59]. The novel PEEK tube also demonstrated a good balance of aesthetic and practical qualities along with other orthodontic appliances. Wire is easily incorporated in the hole of the tube because of its elasticity and flexibility [60].
Difference between PEEK and PEKK
PAEK polymer family members PEEK and PEKK have similar properties except for some that are mentioned in Table 2.
Table 2. Difference between PEEK and PEKK.
PEEK: polyetheretherketone; PEKK: polyetherketoneketone
| PEEK | PEKK |
| Contains two ether group and one keto group | Contains two keto group and one ether group |
| Higher extrusion temperature (almost 400°C) | Lower extrusion temperature (340°C-360°C) |
| Lower glass transition temperature (143°C) | Higher glass transition temperature than PEEK (159°C) |
| Low modulus of elasticity than PEKK | Greater modulus of elasticity than PEEK |
| As PEEK has low modulus of elasticity, it requires surface treatment | It works if surface treatment is not done with PEKK polymer |
| 80% straight and 20% kinked segments melt at 360°C | 60% straight and 40% kinked segments melt at 305°C |
Conclusions
The polymer-based material PAEK is a polycyclic, thermoplastic, semicrystalline structure with superior properties. PEEK and PEKK are polymers that can be used widely in dentistry as implants, RPDs, FPDs, crowns, ortho wires, and so on. Other than dentistry, it can be used in the medical field also. These materials benefit not only clinicians but also patients. As it is metal-free and has an elastic structure, it differs from normal metallic substances used in day-to-day life. Currently, to modify the properties of both materials, additives are used. In light of this review, practicing this material in various procedures can provide improvised and beneficiary treatment options. Various studies are still going to prove the efficacy of this polymer.
The content published in Cureus is the result of clinical experience and/or research by independent individuals or organizations. Cureus is not responsible for the scientific accuracy or reliability of data or conclusions published herein. All content published within Cureus is intended only for educational, research and reference purposes. Additionally, articles published within Cureus should not be deemed a suitable substitute for the advice of a qualified health care professional. Do not disregard or avoid professional medical advice due to content published within Cureus.
Footnotes
The authors have declared that no competing interests exist.
References
- 1.The dentin organic matrix - limitations of restorative dentistry hidden on the nanometer scale. Bertassoni LE, Orgel JP, Antipova O, Swain MV. Acta Biomater. 2012;8:2419–2433. doi: 10.1016/j.actbio.2012.02.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Polyetherketoneketone (PEKK): an emerging biomaterial for oral implants and dental prostheses. Alqurashi H, Khurshid Z, Syed AU, Rashid Habib S, Rokaya D, Zafar MS. J Adv Res. 2021;28:87–95. doi: 10.1016/j.jare.2020.09.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.PEEK biomaterials in trauma, orthopedic, and spinal implants. Kurtz SM, Devine JN. Biomaterials. 2007;28:4845–4869. doi: 10.1016/j.biomaterials.2007.07.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Choupin T. Paris, France: ENSAM; 2017. Mechanical performances of PEKK thermoplastic composites linked to their processing parameters. [Google Scholar]
- 5.Polyaryletherketone polymer nanocomposite engineered with nanolaminated Ti3SiC2 ceramic fillers. Mahesh KV, Balanand S, Raimond R, Peer Mohamed A, Ananthakumar S. Mater Des. 2014;63:360–367. [Google Scholar]
- 6.Peek high performance polymers: a review of properties and clinical applications in prosthodontics and restorative dentistry. Alexakou E, Damanaki M, Zoidis P, Bakiri E, Mouzis N, Smidt G, Kourtis S. Eur J Prosthodont Restor Dent. 2019;27:113–121. doi: 10.1922/EJPRD_01892Zoidis09. [DOI] [PubMed] [Google Scholar]
- 7.Synthesis, characterization and study of the thermal properties of new polyarylene ethers. Staniland PA, Wilde CJ, Bottino FA, Di Pasquale G, Pollicino A, Recca A. Polymer. 1992;33:1976–1981. [Google Scholar]
- 8.Influence of interphase layer on the overall elasto-plastic behaviors of HA/PEEK biocomposite. Fan JP, Tsui CP, Tang CY, Chow CL. Biomaterials. 2004;25:5363–5373. doi: 10.1016/j.biomaterials.2003.12.050. [DOI] [PubMed] [Google Scholar]
- 9.Synthesis and properties of novel copolymers of poly (ether ketone diphenyl ketone ether ketone ketone) and poly (ether amide ether amide ether ketone ketone) Huang B, Qian J, Wang G, Cai M. Polym Eng Sci. 2014;54:1757–1764. [Google Scholar]
- 10.Hydroxyapatite whisker-reinforced polyetherketoneketone bone ingrowth scaffolds. Converse GL, Conrad TL, Merrill CH, Roeder RK. Acta Biomater. 2010;6:856–863. doi: 10.1016/j.actbio.2009.08.004. [DOI] [PubMed] [Google Scholar]
- 11.Wypych G. Ontario, Canada: ChemTec Publishing; 2022. Handbook of polymers. [Google Scholar]
- 12.Relationship between molecular structure and thermal properties of poly (aryl ether ketone) s. Shibata M, Yosomiya R, Wang J, Zheng Y, Zhang W, Wu Z. Macromol Rapid Commun. 1997;18:99–105. [Google Scholar]
- 13.Superficial modifications in TiO2 and Al2O3 ceramics. Santos FP, Campos ED, Costa M, Melo FC, Honda RY, Mota RP. Mater Res. 2003;6:353–357. [Google Scholar]
- 14.Peek: a futuristic dental material in pediatric dentistry. Mushtaq A, Chawla S, Perween N, Gupta R. J Pediatr Dent. 2021;7:72–74. [Google Scholar]
- 15.Erratum: antibacterial properties of PEKK for orthopedic applications [corrigendum] Int J Nanomedicine. 2021;16:4189. doi: 10.2147/IJN.S322675. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Osseointegration of one-piece zirconia implants compared with a titanium implant of identical design: a histomorphometric study in the dog. Koch FP, Weng D, Krämer S, Biesterfeld S, Jahn-Eimermacher A, Wagner W. Clin Oral Implants Res. 2010;21:350–356. doi: 10.1111/j.1600-0501.2009.01832.x. [DOI] [PubMed] [Google Scholar]
- 17.PEEK dental implants: a review of the literature. Schwitalla A, Müller WD. J Oral Implantol. 2013;39:743–749. doi: 10.1563/AAID-JOI-D-11-00002. [DOI] [PubMed] [Google Scholar]
- 18.In vitro biocompatibility testing of polymers for orthopaedic implants using cultured fibroblasts and osteoblasts. Morrison C, Macnair R, MacDonald C, Wykman A, Goldie I, Grant MH. Biomater. 1995;16:987–992. doi: 10.1016/0142-9612(95)94906-2. [DOI] [PubMed] [Google Scholar]
- 19.Biomechanical rigidity of an all-polyetheretherketone anterior thoracolumbar spinal reconstruction construct: an in vitro corpectomy model. Moon SM, Ingalhalikar A, Highsmith JM, Vaccaro AR. Spine J. 2009;9:330–335. doi: 10.1016/j.spinee.2008.11.012. [DOI] [PubMed] [Google Scholar]
- 20.Strength of poly-etherether-ketone: effects of sterilization and thermal ageing. Xin H, Shepherd D, Dearn K. Polym Test. 2013;32:1001–1005. [Google Scholar]
- 21.Nano-hydroxyapatite-coated PEEK implants: a pilot study in rabbit bone. Barkarmo S, Wennerberg A, Hoffman M, Kjellin P, Breding K, Handa P, Stenport V. J Biomed Mater Res A. 2013;101:465–471. doi: 10.1002/jbm.a.34358. [DOI] [PubMed] [Google Scholar]
- 22.Interfacial shear strength of bioactive-coated carbon fiber reinforced polyetheretherketone after in vivo implantation. Nakahara I, Takao M, Goto T, Ohtsuki C, Hibino S, Sugano N. J Orthop Res. 2012;30:1618–1625. doi: 10.1002/jor.22115. [DOI] [PubMed] [Google Scholar]
- 23.Enhanced osteointegration on tantalum-implanted polyetheretherketone surface with bone-like elastic modulus. Lu T, Wen J, Qian S, et al. Biomaterials. 2015;51:173–183. doi: 10.1016/j.biomaterials.2015.02.018. [DOI] [PubMed] [Google Scholar]
- 24.Biofilm behavior on sulfonated poly(ether-ether-ketone) (sPEEK) Montero JF, Tajiri HA, Barra GM, et al. Mater Sci Eng C Mater Biol Appl. 2017;70:456–460. doi: 10.1016/j.msec.2016.09.017. [DOI] [PubMed] [Google Scholar]
- 25.PEEK materials as an alternative to titanium in dental implants: a systematic review. Mishra S, Chowdhary R. Clin Implant Dent Relat Res. 2019;21:208–222. doi: 10.1111/cid.12706. [DOI] [PubMed] [Google Scholar]
- 26.Long-term retention behaviour of resin matrix attachment systems for overdentures. Passia N, Ghazal M, Kern M. J Mech Behav Biomed Mater. 2016;57:88–94. doi: 10.1016/j.jmbbm.2015.11.038. [DOI] [PubMed] [Google Scholar]
- 27.Implantable PEKK/tantalum microparticles composite with improved surface performances for regulating cell behaviors, promoting bone formation and osseointegration. Hu X, Mei S, Wang F, et al. Bioact Mater. 2021;6:928–940. doi: 10.1016/j.bioactmat.2020.09.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Accuracy of patient-specific temporal implants using PEKK. Khashaba MM, Shaheen HA, Ibrahim WH, AlDainy DG. J Craniomaxillofac Surg. 2021;49:943–949. doi: 10.1016/j.jcms.2021.05.001. [DOI] [PubMed] [Google Scholar]
- 29.Antibacterial properties of PEKK for orthopedic applications. Wang M, Bhardwaj G, Webster TJ. Int J Nanomedicine. 2017;12:6471–6476. doi: 10.2147/IJN.S134983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Retentive forces and fatigue resistance of thermoplastic resin clasps. Tannous F, Steiner M, Shahin R, Kern M. Dent Mater. 2012;28:273–278. doi: 10.1016/j.dental.2011.10.016. [DOI] [PubMed] [Google Scholar]
- 31.Use of polyetheretherketone in the fabrication of a maxillary obturator prosthesis: a clinical report. Costa-Palau S, Torrents-Nicolas J, Brufau-de Barberà M, Cabratosa-Termes J. J Prosthet Dent. 2014;112:680–682. doi: 10.1016/j.prosdent.2013.10.026. [DOI] [PubMed] [Google Scholar]
- 32.A three-dimensional finite element analysis of mechanical function for 4 removable partial denture designs with 3 framework materials: CoCr, Ti-6Al-4V alloy and PEEK. Chen X, Mao B, Zhu Z, et al. Sci Rep. 2019;9:13975. doi: 10.1038/s41598-019-50363-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.CAD/CAM constructed poly(etheretherketone) (PEEK) framework of Kennedy class I removable partial denture: a clinical report. Harb IE, Abdel-Khalek EA, Hegazy SA. https://doi.org/10.1111/jopr.12968. J Prosthodont. 2019;28:0–8. doi: 10.1111/jopr.12968. [DOI] [PubMed] [Google Scholar]
- 34.Fracture strength of temporary fixed partial dentures: CAD/CAM versus directly fabricated restorations. Alt V, Hannig M, Wöstmann B, Balkenhol M. Dent Mater. 2011;27:339–347. doi: 10.1016/j.dental.2010.11.012. [DOI] [PubMed] [Google Scholar]
- 35.Load-bearing capacity of CAD/CAM milled polymeric three-unit fixed dental prostheses: effect of aging regimens. Stawarczyk B, Ender A, Trottmann A, Özcan M, Fischer J, Hämmerle CH. Clin Oral Investig. 2012;16:1669–1677. doi: 10.1007/s00784-011-0670-4. [DOI] [PubMed] [Google Scholar]
- 36.Load-bearing capacity of all-ceramic three-unit fixed partial dentures with different computer-aided design (CAD)/computer-aided manufacturing (CAM) fabricated framework materials. Beuer F, Steff B, Naumann M, Sorensen JA. Eur J Oral Sci. 2008;116:381–386. doi: 10.1111/j.1600-0722.2008.00551.x. [DOI] [PubMed] [Google Scholar]
- 37.Fracture force of tooth-tooth- and implant-tooth-supported all-ceramic fixed partial dentures using titanium vs. customised zirconia implant abutments. Kolbeck C, Behr M, Rosentritt M, Handel G. Clin Oral Implants Res. 2008;19:1049–1053. doi: 10.1111/j.1600-0501.2008.01551.x. [DOI] [PubMed] [Google Scholar]
- 38.An insight into a novel material: PEEK. Razzaque A, Dhaded S. Res Rev Prost Dent. 2017;1:1–6. [Google Scholar]
- 39.Polyetheretherketone/nano-fluorohydroxyapatite composite with antimicrobial activity and osseointegration properties. Wang L, He S, Wu X, et al. Biomaterials. 2014;35:6758–6775. doi: 10.1016/j.biomaterials.2014.04.085. [DOI] [PubMed] [Google Scholar]
- 40.Bioactivity and osseointegration of PEEK are inferior to those of titanium: a systematic review. Najeeb S, Khurshid Z, Zohaib S, Zafar MS. J Oral Implantol. 2016;42:512–516. doi: 10.1563/aaid-joi-D-16-00072. [DOI] [PubMed] [Google Scholar]
- 41.Mechanical response of PEKK and peek as frameworks for implant-supported full-arch fixed dental prosthesis: 3D finite element analysis. Villefort RF, Diamantino PJ, Zeidler SL, Borges AL, Silva-Concílio LR, Saavedra GD, Tribst JP. Eur J Dent. 2022;16:115–121. doi: 10.1055/s-0041-1731833. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Evaluation of decomposition kinetics of poly (ether-ether-ketone) by thermogravimetric analysis. da Cunha Vasconcelos G, Mazur RL, Botelho EC, Rezende MC, Costa ML. J Aerosp Technol Manag. 2010;2:155–162. [Google Scholar]
- 43.Shrinkage behaviour of semi-crystalline polymers in laser sintering: PEKK and PA12. Benedetti L, Brulé B, Decreamer N, Evans KE, Ghita O. Mater Des. 2019;12:181. [Google Scholar]
- 44.Three-dimensional analysis of marginal and internal fit of copings fabricated with polyetherketoneketone (PEKK) and zirconia. Bae SY, Park JY, Jeong ID, Kim HY, Kim JH, Kim WC. J Prosthodont Res. 2017;61:106–112. doi: 10.1016/j.jpor.2016.07.005. [DOI] [PubMed] [Google Scholar]
- 45.Retentive properties of two stud attachments with polyetherketoneketone or nylon insert in mandibular implant overdentures. Choi JW, Yun BH, Jeong CM, Huh JB. Int J Oral Maxillofac Implants. 2018;33:1079–1088. doi: 10.11607/jomi.6023. [DOI] [PubMed] [Google Scholar]
- 46.Effect of margin designs on the marginal adaptation of zirconia copings. Habib SR, Al Ajmi MG, Al Dhafyan M, Jomah A, Abualsaud H, Almashali M. Acta Stomatol Croat. 2017;51:179–187. doi: 10.15644/asc51/3/1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Marginal adaptation and CAD-CAM technology: a systematic review of restorative material and fabrication techniques. Papadiochou S, Pissiotis AL. J Prosthet Dent. 2018;119:545–551. doi: 10.1016/j.prosdent.2017.07.001. [DOI] [PubMed] [Google Scholar]
- 48.A speech bulb prosthesis for a soft palate defect with a polyetherketoneketone (PEKK) framework fabricated by multiple digital techniques: a clinical report. Sun F, Shen X, Zhou N, Gao Y, Guo Y, Yang X, Wu G. J Prosthet Dent. 2020;124:495–499. doi: 10.1016/j.prosdent.2019.10.020. [DOI] [PubMed] [Google Scholar]
- 49.Clinical evaluation of performance of single unit polyetheretherketone crown restoration-a pilot study. Sulaya K, Guttal SS. J Indian Prosthodont Soc. 2020;20:38–44. doi: 10.4103/jips.jips_228_19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Comparative prospective clinical evaluation of computer aided design/ computer aided manufacturing milled BioHPP PEEK inlays and zirconia inlays. Rajamani VK, Reyal SS, Gowda EM, Shashidhar MP. J Indian Prosthodont Soc. 2021;21:240–248. doi: 10.4103/jips.jips_57_21. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Mechanical and thermal stress evaluation of PEEK prefabricated post with different head design in endodontically treated tooth: 3D-finite element analysis. Ibrahim RO, Al-Zahawi AR, Sabri LA. Dent Mater J. 2021;40:508–518. doi: 10.4012/dmj.2020-053. [DOI] [PubMed] [Google Scholar]
- 52.Fabrication of a novel hydroxyapatite/polyether ether ketone surface nanocomposite via friction stir processing for orthopedic and dental applications. Almasi D, Lau WJ, Rasaee S, Sharifi R, Mozaffari HR. Prog Biomater. 2020;9:35–44. doi: 10.1007/s40204-020-00130-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Fracture resistance of endodontically treated teeth restored with various esthetic posts. Haralur SB. Technol Health Care. 2021;29:243–252. doi: 10.3233/THC-202228. [DOI] [PubMed] [Google Scholar]
- 54.Evaluation of the use of PEEK material in post-core and crown restorations using finite element analysis. Tekin S, Adiguzel O, Cangul S, Atas O, Erpacal B. https://pubmed.ncbi.nlm.nih.gov/33017528/ Am J Dent. 2020;33:251–257. [PubMed] [Google Scholar]
- 55.Effects of endodontic irrigants on material and surface properties of biocompatible thermoplastics. Kucher M, Dannemann M, Modler N, Hannig C, Weber MT. Dent J (Basel) 2019;7:26. doi: 10.3390/dj7010026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Biomechanical evaluation of a tooth restored with high performance polymer PEKK post-core system: a 3D finite element analysis. Lee KS, Shin JH, Kim JE, Kim JH, Lee WC, Shin SW, Lee JY. Biomed Res Int. 2017;2017:1373127. doi: 10.1155/2017/1373127. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Custom and prefabricated polyetherketoneketone (PEKK) post-core systems bond strength: scanning electron microscopy evaluation. Güven MÇ, Dayan SÇ, Yıldırım G, Mumcu E. Microsc Res Tech. 2020;83:804–810. doi: 10.1002/jemt.23471. [DOI] [PubMed] [Google Scholar]
- 58.Computer-aided design of polyetheretherketone for application to removable pediatric space maintainers. Guo H, Wang Y, Zhao Y, Liu H. BMC Oral Health. 2020;20:201. doi: 10.1186/s12903-020-01184-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Load-deflection and friction properties of peek wires as alternative orthodontic wires. Tada Y, Hayakawa T, Nakamura Y. Materials (Basel) 2017;10:914. doi: 10.3390/ma10080914. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Mechanical properties of orthodontic wires covered with a polyether ether ketone tube. Shirakawa N, Iwata T, Miyake S, Otuka T, Koizumi S, Kawata T. Angle Orthod. 2018;88:442–449. doi: 10.2319/082417-572.1. [DOI] [PMC free article] [PubMed] [Google Scholar]