Quantitative flow ratio for functional evaluation of in-stent restenosis

In-stent restenosis functional assessment with QFR

Catherine Liontou 1, Hernán Mejía-Rentería 2, Francesco Lauri 3, Sonoka Goto 4,5, Hyun-Jong Lee 6, Masafumi Nakayama 7,8, Alicia Quiros 9, Fernando Macaya 10, Nieves Gonzalo 11, Iván Núñez-Gil 12, Pablo Salinas 13, Maria Del Trigo 14, Javier Escaned 15,*
PMCID: PMC9724976  PMID: 31355754

Introduction

Coronary angiography is the most common diagnostic tool to assess in-stent restenosis (ISR) severity, both in clinical practice and in trials1. However, given its poor ability to depict the relevance of functional stenosis, fractional flow reserve (FFR) has been proposed as the reference standard to ascertain functional ISR severity2. More recently, quantitative flow ratio (QFR) has been validated in de novo lesions as an angiography-based approach to functional stenosis characterisation that does not require intracoronary instrumentation3. We investigated the diagnostic performance of QFR in ISR lesions, using FFR as the reference standard.

Methods

This was a multicentre, international, retrospective, blinded study, enrolling patients from three hospitals in three countries (Hospital ClÃnico San Carlos, Spain; Toda Chuo General Hospital, Japan; and Sejong General Hospital, Republic of Korea). The study population consisted of a group of ISR patients in whom FFR was used to guide coronary revascularisation. Patients with ISR defined as ≥50% diameter stenosis (DS) within the stent, or within 5 mm from the stent edges, luminal narrowing as judged visually were considered for the study. Details regarding data collection and analysis are available in Supplementary Appendix 1.

Results

QFR analysis was performed in 78 vessels (73 patients) with ISR, all investigated with FFR (Figure 1). Supplementary Appendix 2 and Supplementary Table 1 show details of patient demographics and clinical characteristics. Angiographic and physiological variables are shown in Table 1. Stenosis severity was intermediate both in terms of angiography (mean %DS: 51±9%) and FFR (mean value: 0.79±0.09).

Figure 1.

Figure 1

Case example of QFR analysis of intermediate ISR in left anterior descending artery (LAD). A) Long LAD stented segment analysed. B) Two angiographic projections >25° apart allow three-dimensional vessel reconstruction. C) QFR computed based on 3D-QCA and TIMI frame count, resulting in a QFR value of 0.89 (non-significant). Green lines represent the proximal and distal borders of the segment with the most significant lesion and the red line represents the most severe stenosis level. D) FFR value was 0.89.

Table 1. Vessel characteristics.

QFR analysis in ISR N=78 vessels
Lesion location, n (%)
Left anterior descending 46 (59)
Left circumflex artery 12 (15)
Obtuse marginal branch 4 (5)
Right coronary artery 16 (21)
Segment location, n (%)
Proximal 33 (42.3)
Mid 41 (52.6)
Distal 4 (5)
Reference vessel diameter, mm 2.7 (2.2-3.0)
Minimal lumen diameter, mm 1.3 (1.0-1.5)
%DS (mean) 51±9
Vessels with DS by 3D-QCA ≥50%, n (%) 38 (49)
Area stenosis, % (mean) 67±10
Lesion length, mm 19.2 (12.9-31.4)
FFR (per vessel) 0.81 (0.75-0.87)
Vessels with FFR ≤0.80 (%) 33 (42)
QFR (per vessel) 0.80 (0.72-0.87)
Vessels with QFR ≤0.80 (%) 40 (51)

The mean difference between FFR and QFR was only 0.01±0.09 (Supplementary Figure 1). Classification agreement between FFR and QFR (in terms of dichotomous functional significance) was high, i.e., 83%. Functional assessment of ISR lesions with QFR was comparable to that reported in de novo lesions in previous studies (Supplementary Table 2). Additionally, the area under the receiver operating characteristic (ROC) curve (AUC) demonstrated high diagnostic performance of QFR regarding its ability to establish ISR relevance, taking FFR as reference (AUC: 0.90 [0.83-0.97]) (Figure 2). Although there was a difference in classification agreement between vessels, this was not of statistical significance (Supplementary Table 3). The study also confirmed the low diagnostic yield of angiography in ISR: in terms of functional severity a 50% DS criterion correctly classified only 68% of ISR cases (Table 2). QFR analysis of ISR cases correctly reclassified (as judged by FFR) 45% of ISR lesions as functionally non-significant.

Table 2. Diagnostic performance of QFR and 3D-QCA DS in ISR population using FFR as reference.

QFR DS by 3D-QCA ≥50%
Classification agreement, n (%) 65 (83%) 53 (68%)
Spearman/Pearson correlation (rho/r) 0.731 0.433
AUC 0.90 (0.83-0.97) 0.74 (0.63-0.85)
Sensitivity (%) 91 (74-97) 70 (51-83)
Specificity (%) 78 (62-88) 67 (50-79)
PPV 75 (58-86) 61 (43-75)
NPV 92 (77-97) 75 (58-86)
+ LR 4.1 (2.3-7.1) 2.0 (1.3-3.3)
− LR 0.1 (0.0-0.3) 0.4 (0.2-0.7)
3D-QCA: three-dimensional quantitative coronary angiography; AUC: area under the curve; DS: diameter stenosis; FFR: fractional flow reserve; ISR: in-stent restenosis; LR: likelihood ratio; NPV: negative predictive value; PPV: positive predictive value; QFR: quantitative flow ratio

Discussion

Our findings support the use of QFR to outline the functional relevance of ISR, with similar diagnostic efficiency to that reported for QFR in major studies in de novo lesions. Compared with available series, the classification agreement of QFR and FFR in ISR lesions was similar to that of two major pivotal studies of QFR-FFR in de novo lesions4,5. Importantly, 45% of the ISR cases deemed significant by angiographic criteria were judged as functionally non-significant both by QFR and by FFR, showing that, due to its high negative predictive value, QFR can lead to safe deferral of revascularisation in a significant proportion of ISR lesions. Furthermore, QFR can be useful as a research tool in assessing the long-term results of stenting.

Limitations

The main limitation of our study was its retrospective character, with exclusion of cases with suboptimal angiography or vessel overlap, something which may have caused selection bias.

Conclusion

QFR has a high diagnostic performance in assessing ISR lesions, similar to that in de novo lesions. It may therefore facilitate adoption of functional assessment in these lesions.

Impact on daily practice

By not requiring intracoronary instrumentation or drug administration, QFR may facilitate the adoption of functional assessment in ISR. Given its high negative predictive value, QFR will contribute to avoiding unnecessary interventions in patients with ISR.

Supplementary data

Supplementary Appendix 1

Methods.

Supplementary Appendix 2

Results.

Supplementary Figure 1

Agreement between QFR and FFR.

Supplementary Table 1

Demographic and clinical characteristics of the ISR population.

Supplementary Table 2

Classification agreement between QFR and FFR in previous large studies.

Supplementary Table 3

Classification agreement between QFR and FFR according to vessel analysis.

Figure 2.

Figure 2

Significant difference in diagnostic performance of QFR and %DS in identifying significant lesions in the ISR population. The area under the curve using FFR as reference standard shows high diagnostic accuracy of QFR but low diagnostic accuracy of %DS for ISR lesions.

Acknowledgments

Acknowledgements

We are grateful to Pilar Jiménez-Quevedo, MD, PhD, Luis Nombela-Franco, MD, PhD, Carlos Macaya, MD, PhD, and Antonio Fernández-Ortiz, MD, PhD, for their contribution to data acquisition.

Conflict of interest statement

The authors have no conflicts of interest to declare.

Contributor Information

Catherine Liontou, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

Hernán Mejía-Rentería, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

Francesco Lauri, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

Sonoka Goto, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain; Cardiovascular Center, Toda Chuo General Hospital, Toda, Japan.

Hyun-Jong Lee, Division of Cardiology, Department of Internal Medicine, Sejong General Hospital, Bucheon, Republic of Korea.

Masafumi Nakayama, Cardiovascular Center, Toda Chuo General Hospital, Toda, Japan; Joint Graduate School of Tokyo Women’s Medical University and Waseda University, Tokyo, Japan.

Alicia Quiros, Department of Mathematics, Universidad de León, León, Spain.

Fernando Macaya, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

Nieves Gonzalo, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

Iván Núñez-Gil, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

Pablo Salinas, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

Maria Del Trigo, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

Javier Escaned, Hospital Clínico San Carlos IDISSC and Universidad Complutense de Madrid, Madrid, Spain.

References

  1. Alfonso F, Cequier A, Angel J, Martí V, Zueco J, Bethencourt A, Mantilla R, López-Mínguez JR, Gómez-Recio M, Morís C, Pérez-Vizcayno MJ, Fernández C, Macaya C, Seabra-Gomes R Restenosis Intra-stent Balloon angioplasty versus elective Stenting (RIBS) Investigators. Value of the American College of Cardiology/American Heart Association angiographic classification of coronary lesion morphology in patients with in-stent restenosis. Insights from the Restenosis Intra-stent Balloon angioplasty versus elective Stenting (RIBS) randomized trial. Am Heart J. 2006;151:681.e1–9. doi: 10.1016/j.ahj.2005.10.014. [DOI] [PubMed] [Google Scholar]
  2. Nam CW, Rha SW, Koo BK, Doh JH, Chung WY, Yoon MH, Tahk SJ, Lee BK, Lee JB, Yoo KD, Cho YK, Chung IS, Hur SH, Kim KB, Choi CU, Oh DJ. Usefulness of coronary pressure measurement for functional evaluation of drug-eluting stent restenosis. Am J Cardiol. 2011;107:1783–6. doi: 10.1016/j.amjcard.2011.02.328. [DOI] [PubMed] [Google Scholar]
  3. Tu S, Barbato E, Koszegi Z, Yang J, Sun Z, Holm NR, Tar B, Li Y, Rusinaru D, Wijns W, Reiber JH. Fractional flow reserve calculation from 3-dimensional quantitative coronary angiography and TIMI frame count: a fast computer model to quantify the functional significance of moderately obstructed coronary arteries. JACC Cardiovasc Interv. 2014;7:768–77. doi: 10.1016/j.jcin.2014.03.004. [DOI] [PubMed] [Google Scholar]
  4. Westra J, Andersen BK, Campo G, Matsuo H, Koltowski L, Eftekhari A, Liu T, Di Serafino L, Di Girolamo D, Escaned J, Nef H, Naber C, Barbierato M, Tu S, Neghabat O, Madsen M, Tebaldi M, Tanigaki T, Kochman J, Somi S, Esposito G, Mercone G, Mejia-Renteria H, Ronco F, Bøtker HE, Wijns W, Christiansen EH, Holm NR. Diagnostic Performance of In-Procedure Angiography-Derived Quantitative Flow Reserve Compared to Pressure-Derived Fractional Flow Reserve: The FAVOR II Europe-Japan Study. J Am Heart Assoc. 18;7(14) doi: 10.1161/JAHA.118.009603. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Westra J, Tu S, Winther S, Nissen L, Vestergaard MB, Andersen BK, Holck EN, Fox Maule C, Johansen JK, Andreasen LN, Simonsen JK, Zhang Y, Kristensen SD, Maeng M, Kaltoft A, Terkelsen CJ, Krusell LR, Jakobsen L, Reiber JHC, Lassen JF, Bottcher M, Botker HE, Christiansen EH, Holm NR. Evaluation of Coronary Artery Stenosis by Quantitative Flow Ratio During Invasive Coronary Angiography: The WIFI II Study (Wire-Free Functional Imaging II). Circ Cardiovasc Imaging. 2018;11:e007107. doi: 10.1161/CIRCIMAGING.117.007107. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Appendix 1

Methods.

Supplementary Appendix 2

Results.

Supplementary Figure 1

Agreement between QFR and FFR.

Supplementary Table 1

Demographic and clinical characteristics of the ISR population.

Supplementary Table 2

Classification agreement between QFR and FFR in previous large studies.

Supplementary Table 3

Classification agreement between QFR and FFR according to vessel analysis.