. 2022 Nov 18;18(10):836–839. doi: 10.4244/EIJ-D-22-00363

Extracardiac basal annuloplasty for the treatment of secondary mitral regurgitation

Jérémy Bernard 1, Erwan Salaun 2, Chandrasekar Padmanabhan 3, Marek Deja 4, Milind Hote 5, Shiv Kumar Choudhary 6, Jan Hlavička 7, Richard Saldanha 8, Radim Brát 9, Anil Jain 10, Naman Shastri 11, Seetharam Bhat 12, Chandana NC 13, Manoj Durairaj 14, Bikash Rai Das 15, Anil Kumar Agarwal 16, Vivek Rao 17, Krishna Talluri 18, Jonathan Beaudoin 19, Mathieu Bernier 20, Nancy Côté 21, Jaishankar Raman 22, Philippe Pibarot 23,*
PMCID: PMC9725004  PMID: 35950967

Introduction

Among the 24 million people affected worldwide by mitral regurgitation1, almost two-thirds suffer from secondary mitral regurgitation (SMR)2. Moreover, the presence of symptomatic SMR in patients with heart failure (HF) remains a marker of increased mortality and rehospitalisation risk whatever its severity2. Thus, recent guidelines recommend downsized mitral annuloplasty as a standalone procedure (Class IIb) or as a concomitant procedure combined with coronary artery bypass grafting (CABG; Class I in European and IIa in American guidelines) for the management of SMR3,4. However, standard annuloplasty remains an intracardiac procedure requiring aortic cross-clamping and cardiopulmonary bypass and is thus associated with an increased risk of perioperative complications.

We aimed to evaluate the safety and efficacy of a new surgical procedure of extracardiac annuloplasty using the BACE (Basal Annuloplasty of the Cardia Externally) device (Phoenix Cardiac Devices) for the management of SMR in patients with systolic HF.

Methods

Forty-seven symptomatic patients with significant SMR (i.e., at least moderate or grade 2+) and systolic HF (i.e., left ventricular ejection fraction [LVEF] between 25% and 50%) referred for a surgical mitral valve (MV) intervention were prospectively recruited in 12 multinational centres between November 2012 and July 2019. Specific inclusion and exclusion criteria are described at ClinicalTrials.gov: NCT02701972. Patients had baseline, preoperative, 1-, 3-, 6-, 12-, and 24-month assessments including medical history, functional status (i.e., New York Heart Association [NYHA]), quality of life (i.e., Minnesota Living with Heart Failure Questionnaire [MLHFQ]) and echocardiography analysed by the independent echocardiography core laboratory based at the Quebec Heart & Lung Institute and following the American Society of Echocardiography standards and recommendations5. This study was conducted in conformity with the Declaration of Helsinki and Good Clinical Practice principles and approved by local ethics committees and respective health authorities. All patients provided informed written consent.

The primary safety endpoint was the freedom from major device- or surgery-related adverse events at 6 months following the procedure and was adjudicated by an independent Data Safety & Monitoring Board. The primary efficacy endpoint was the reduction of the SMR grade to ≤mild at 6 months. It was then evaluated in the subset (n=35) which had the 6-month follow-up (FU). Secondary endpoints were freedom from major device-related adverse events, and changes in the SMR grade, NYHA class and MLHFQ score between baseline and FU.

Results

Among the 47 patients recruited, implantation of the device was attempted but was not completed in 3 patients. Thus, procedural success was 94% (Figure 1). In the 44 patients (mean age±standard deviation [SD]: 62±12 years, 73% male, median [interquartile range] Charlson Comorbidity Index: 3 [2-4]) who were treated with the device, the majority (35 patients; 80%) underwent concomitant CABG, whereas 9 (20%) patients had a standalone procedure. The procedure was performed off-pump in 20 (57%) patients, and the median number of grafts was 3 (2–4).

Figure 1. Study endpoints and changes in mitral regurgitation severity, LV remodelling, quality of life and functional status at 6 months following the implantation of the BACE device.

Figure 1

Illustration of the BACE device, which provides external annuloplasty of the cardiac base in order to reduce secondary mitral regurgitation severity and left ventricular dilation, and the main characteristics of the device (top); results of feasibility, safety, and efficacy endpoints (middle); and graphs for LV remodelling, NYHA Functional Class, and quality of life improvements (bottom) at 6 months after the implantation of the device in patients with at least moderate SMR and reduced ejection fraction. AE: adverse events; BACE: basal annuloplasty of the cardia externally; CABG: coronary artery bypass grafting; DSMB: Data Safety & Monitoring Board; LV: left ventricular; LVEDV: LV end-diastolic volume; MLHFQ: Minnesota Living with Heart Failure Questionnaire; NYHA: New York Heart Association; SMR: secondary mitral regurgitation

The primary safety endpoint was met in 70% of patients (31 of 44 patients implanted). Freedom from major device-related events was 100% and 98% at 6 and 12 months, respectively (i.e., compression of right ventricular inflow and outflow causing underfilling and reduced function in 1 patient who underwent surgical device removal). Two patients (5%) died before 30 days, and 6 patients died between 30 days and 6 months. None of these deaths were related to the device.

The primary efficacy endpoint was achieved in 82% of patients (SMR severity: 8 [24%] none or trace, 19 [58%] mild, 4 [12%] moderate and 2 [6%] severe; p-value vs baseline by Friedman test: p<0.001). In terms of secondary efficacy endpoints, 87% of patients had a reduction of at least one grade at 30 days, and 88% at 6 months. Echocardiographic parameters of left ventricular (LV) geometry/function and MV morphology (i.e., MV annulus diameter, tenting area, leaflet coaptation distance, indexed left atrial volume, LV end-diastolic volume, and LV end-systolic and end-diastolic diameters) significantly improved from baseline to 6 months (Figure 1) (all p-values by paired t-tests ≤0.003).

At 6 months, patients experienced significant improvement in NYHA class (p<0.001 by Friedman test), with percentages decreasing from 34% in Class III-IV to 12%, and from 63% in Class II to 47% (Figure 1). The MLHFQ score showed a marked improvement (i.e., reduction; paired t-tests: p<0.001) at 6 months (Figure 1). MR reduction (i.e., ≤mild) persists at 12 and 24 months, respectively, in 83% (20/24) and 91% (21/23) of patients who underwent these subsequent FU.

Discussion

The purpose of the BACE device is to achieve an extracardiac, indirect, and targeted restrictive mitral annuloplasty with the advantages of being less invasive than intracardiac downsized/restrictive annuloplasty and of offering the possibility of correcting persistent/recurrent MR by adjustment of the degree of annular restriction via the subcutaneous ports. Recurrent MR is believed to be the main reason for the lack of superiority of downsized annuloplasty versus mitral valve replacement for the treatment of ischaemic MR. The possibility of adjusting the degree of annular restriction achieved by the BACE device at any time during post-procedural follow-up may overcome this important limitation of standard intracardiac annuloplasty.

Limitations

The main limitations of this study are the single-arm design and the high prevalence of concomitant CABG procedures, which implicate that we cannot ascertain the respective effects of extracardiac annuloplasty versus myocardial revascularisation on study endpoints.

Conclusions

In summary, extracardiac basal annuloplasty with the BACE device appears to be safe and feasible and is associated with a significant reduction of SMR severity, positive LV remodelling, and improvement in the patient’s quality of life and functional status. Future randomised controlled trials comparing BACE±CABG versus intracardiac annuloplasty±CABG are needed.

Acknowledgments

Acknowledgements

We thank the participants of this study and the sponsor who was involved in the study design, the determination of performance goals and assisted with data interpretation, but who did not participate in data collection and data analysis.

Funding

This study was funded by Phoenix Cardiac Devices, Inc., Cary, NC, USA. J. Bernard is supported by a PhD research grant from the Canadian Institutes of Health Research (CIHR), Ottawa, Ontario, Canada. J. Beaudoin is supported by the Fonds de Recherche Québec - Santé, Montréal, Québec, Canada. P. Pibarot holds the Canada Research Chair in Valvular Heart Diseases and is supported by a Foundation grant (FDN-143225) from CIHR.

Conflict of interest statement

K. Talluri was the vice-president of clinical development at Phoenix Cardiac Devices. P. Pibarot received funding from Edwards Lifesciences, Medtronic and Phoenix Cardiac Devices for echocardiography core laboratory analyses, with no personal compensation. The other authors have no conflicts of interest to declare relevant to this study.

Contributor Information

Jérémy Bernard, Institut universitaire de cardiologie et de pneumologie de Québec – Université Laval/Québec Heart & Lung Institute, Québec City, Québec, Canada.

Erwan Salaun, Institut universitaire de cardiologie et de pneumologie de Québec – Université Laval/Québec Heart & Lung Institute, Québec City, Québec, Canada.

Chandrasekar Padmanabhan, G. Kuppuswamy Naidu Memorial Hospital, Coimabatore, India.

Marek Deja, Medical University of Silesia, Katowice, Poland.

Milind Hote, All India Institute of Medical Sciences, New Delhi, India.

Shiv Kumar Choudhary, All India Institute of Medical Sciences, New Delhi, India.

Jan Hlavička, University Hospital Kralovske Vinohrady, Prague, Czech Republic.

Richard Saldanha, K.L.E.S. Dr Prabhakar Kore Hospital & MRC, Belagavi, India.

Radim Brát, Department of Cardiac Surgery, University Hospital Ostrava, Ostrava, Czech Republic.

Anil Jain, EPIC Hospital (Unit of Vatsalya Healthcare LLP), Ahmedabad, India.

Naman Shastri, SAL Hospital, Ahmedabad, India.

Seetharam Bhat, Sri Jayadeva Institute of Cardiovascular Sciences and Research, Bengaluru, India.

Chandana NC, Sri Jayadeva Institute of Cardiovascular Sciences and Research, Bengaluru, India.

Manoj Durairaj, Ruby Hall Clinic, Pune, India.

Bikash Rai Das, GNRC Hospital, Guwahati, India.

Anil Kumar Agarwal, Maharaja Agrasen Hospital, New Delhi, India.

Vivek Rao, Toronto General Hospital, Toronto, ON, Canada.

Krishna Talluri, Phoenix Cardiac Devices, Inc., Cary, NC, USA.

Jonathan Beaudoin, Institut universitaire de cardiologie et de pneumologie de Québec – Université Laval/Québec Heart & Lung Institute, Québec City, Québec, Canada.

Mathieu Bernier, Institut universitaire de cardiologie et de pneumologie de Québec – Université Laval/Québec Heart & Lung Institute, Québec City, Québec, Canada.

Nancy Côté, Institut universitaire de cardiologie et de pneumologie de Québec – Université Laval/Québec Heart & Lung Institute, Québec City, Québec, Canada.

Jaishankar Raman, Austin Hospital and St Vincent’s Hospital, University of Melbourne, Melbourne, Australia.

Philippe Pibarot, Institut universitaire de cardiologie et de pneumologie de Québec – Université Laval/Québec Heart & Lung Institute, Québec City, Québec, Canada.

References

  1. Coffey S, Roberts-Thomson R, Brown A, Carapetis J, Chen M, Enriquez-Sarano M, Zühlke L, Prendergast BD. Global epidemiology of valvular heart disease. Nat Rev Cardiol. 2021;18:853–64. doi: 10.1038/s41569-021-00570-z. [DOI] [PubMed] [Google Scholar]
  2. Rossi A, Dini FL, Faggiano P, Agricola E, Cicoira M, Frattini S, Simioniuc A, Gullace M, Ghio S, Enriquez-Sarano M, Temporelli PL. Independent prognostic value of functional mitral regurgitation in patients with heart failure. A quantitative analysis of 1256 patients with ischaemic and non-ischaemic dilated cardiomyopathy. Heart. 2011;97:1675–80. doi: 10.1136/hrt.2011.225789. [DOI] [PubMed] [Google Scholar]
  3. Writing Committee Members, Otto CM, Nishimura RA, Bonow RO, Carabello BA, Erwin JP, 3rd, Gentile F, Jneid H, Krieger EV, Mack M, McLeod C, O'Gara PT, Rigolin VH, Sundt TM, 3rd, Thompson A, Toly C. 2020 ACC/AHA Guideline for the Management of Patients With Valvular Heart Disease:Executive Summary: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2021;77:450–500. doi: 10.1016/j.jacc.2020.11.035. [DOI] [PubMed] [Google Scholar]
  4. Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J, Capodanno D, Conradi L, De Bonis, De Paulis, Delgado V, Freemantle N, Haugaa KH, Jeppsson A, Jüni P, Pierard L, Prendergast BD, Sádaba JR, Tribouilloy C, Wojakowski W. 2021 ESC/EACTS Guidelines for the management of valvular heart disease. EuroIntervention. 2022;17:e1126–96. doi: 10.4244/EIJ-E-21-00009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Zoghbi WA, Adams D, Bonow RO, Enriquez-Sarano M, Foster E, Grayburn PA, Hahn RT, Han Y, Hung J, Lang RM, Little SH, Shah DJ, Shernan S, Thavendiranathan P, Thomas JD, Weissman NJ. Recommendations for Noninvasive Evaluation of Native Valvular Regurgitation: A Report from the American Society of Echocardiography Developed in Collaboration with the Society for Cardiovascular Magnetic Resonance. J Am Soc Echocardiogr. 2017;30:303–71. doi: 10.1016/j.echo.2017.01.007. [DOI] [PubMed] [Google Scholar]

Articles from EuroIntervention are provided here courtesy of Europa Group

RESOURCES