Abstract
Selective internal radiation therapy (SIRT) is one of the treatment options for liver tumors. Microspheres labelled with a therapeutic radionuclide (90Y or 166Ho) are injected into the liver artery feeding the tumor(s), usually achieving a high tumor absorbed dose and a high tumor control rate. This treatment adopts a theranostic approach with a mandatory simulation phase, using a surrogate to radioactive microspheres (99mTc-macroaggregated albumin, MAA) or a scout dose of 166Ho microspheres, imaged by SPECT/CT. This pre-therapy imaging aims to evaluate the tumor targeting and detect potential contraindications to SIRT, i.e., digestive extrahepatic uptake or excessive lung shunt. Moreover, the absorbed doses to the tumor(s) and the healthy liver can be estimated and used for planning the therapeutic activity for SIRT optimization. The aim of this review is to evaluate the accuracy of this theranostic approach using pre-therapy imaging for simulating the biodistribution of the microspheres. This review synthesizes the recent publications demonstrating the advantages and limitations of pre-therapy imaging in SIRT, particularly for activity planning.
Keywords: theranostic, SIRT, radioembolization, pre-therapy imaging, MAA SPECT/CT
1. Introduction
Selective internal radiation therapy (SIRT) is a treatment option for liver tumors by delivering radioactive microspheres in the tumors’ feeding arteries, leading to very high absorbed doses to tumors and usually achieving a high tumor control rate [1]. Hepatic tumors are mainly vascularized by the liver artery as opposed to the healthy liver, whose blood supply is preferentially ensured by the portal vein [2]. This unique characteristic is the milestone of SIRT efficacy. By injecting radioactive microspheres in the liver artery or any of its branches, tumor(s) receive very high absorbed doses while sparing the healthy liver.
Three types of radioactive microspheres are commercially available, differing by their physical and radioactive characteristics: yttrium-90 (90Y)-resin microspheres (Sir-Spheres®, Sirtex Medical GmbH, Sydney, Australia), 90Y-glass microspheres (Therasphere®, Boston Scientific, Boston, MA, USA), and holmium-166 (166Ho)-poly-L-lactic acid (PLLA) microspheres (QuiremSpheres®, Quirem Medical B.V., Deventer, The Netherlands) [3]. Only Sir-Spheres® and Therasphere® have FDA approval for the U.S. market.
The SIRT procedure applies a theranostic approach by using similar radiopharmaceuticals agents for both therapy planning and treatment [4]. During the first step, the treatment is simulated using a surrogate: 99mTc-macroaggregated albumin (MAA) or a scout dose of the radioactive microspheres (166Ho-scout dose), enabling a selection of patients who have good tumor targeting and no absolute contraindications for treatment such as excessive lung shunt and digestive uptake. These radionuclides emit low energy gamma rays (Table 1), allowing for detection by nuclear imaging systems to precisely localize the distribution in the liver or in non-targeted tissues. In particular, 3D imaging using single photon emission computed tomography combined with CT scan (SPECT/CT) allows an accurate evaluation of the MAA distribution (MAA SPECT/CT) for evaluating extra hepatic arterial shunting to limit the risks of lung and gastrointestinal complications during treatment [5,6].
Table 1.
Labelled Particle | Gamma | Positron | |
---|---|---|---|
Radionuclide | Emission Detected by SPECT/CT | Emission Detected by PET/CT | |
(Radioactive Half Life) | (Energy and Abundance) | (Abundance) | |
99mTc | MAA | Direct g radiation (140 keV, 89%) |
NA |
(6 h) | |||
166Ho | PLLA microspheres | Direct g radiation (82 KeV, 6,7%) |
NA |
(27 h) | |||
90Y | resin and glass microspheres | Indirect radiation (bremsstrahlung, continuous spectrum radiation from 0 to 2.3 MeV) |
0.0036% |
(64 h) |
NA: not available.
The MAA distribution in the tumor and healthy liver compartments can also be analyzed to estimate the absorbed doses (multicompartmental model) and establish the amount of radioactive microspheres to be scheduled for the treatment (activity planning) [7].
Using 166Ho microspheres, a minimal dose (scout dose) can be used as an alternative to MAA particles. This scout dose can simulate the treatment without risks of radiation toxicity [8].
After treatment, the correct distribution of the radioactive microspheres is confirmed by SPECT/CT for 90Y (bremsstrahlung) and for 166Ho by 90Y PET/CT (radiation characteristics in Table 1). In addition, the paramagnetic nature of holmium metal enables MRI imaging, evaluating as well the biodistribution of Ho microspheres [9].
Post-therapy 90Y PET/CT imaging evaluates with accuracy the actual absorbed doses to the tumor(s) and to the healthy liver [3].
Pre-therapy MAA SPECT/CT/CT and post-therapy 90Y PET/CT absorbed doses are strongly correlated with tumor response and patient outcome (high tumor absorbed dose) or with liver toxicity (high healthy liver absorbed dose) [10,11,12,13].
Figure 1 illustrates pre-therapy and post-therapy imaging in a patient treated by resin microspheres for colorectal metastases, showing a nice match between 99mTc-MAA and 90Y-resin spheres distribution.
With regards to activity planning, different simple and safe methods are recommended by manufacturers. Activity planning is performed using a monocompartmental model, targeting a liver absorbed dose of 80–150 Gy with glass microspheres and 60 Gy for 166Ho microspheres [3]. For resin microspheres, activity planning is often performed with a semi-empirical method, using a calculation based on the patient’s body surface area and the tumor burden. As an alternative, Sirtex Medical also recommends the use of a more complex method named the partition model or multicompartmental model, which takes into account separately the absorbed doses to the non-tumoral liver, to the tumor(s), and to the lungs using MAA pre-therapy imaging [14]. Article 56 of the EU Council Directive 2013/59 and recent guidelines recommend the use of this personalized method of activity planning, differentiating tumor and healthy liver compartments [15,16,17]. As demonstrated by previous data, this multicompartmental model optimizes treatment planning and improves the effectiveness of SIRT [13]. With this model, the DOSISPHERE-01 randomized controlled trial demonstrated a significant increase in tumor response and better patient outcome in hepatocellular carcinoma (HCC), without increasing the toxicity [10].
This review aims to evaluate the accuracy of pre-therapy imaging using MAA or a scout dose for predicting the treatment efficacy. This review will synthesize the recent publications demonstrating the advantages and limitations of pre-therapy imaging in SIRT, particularly for activity planning.
2. MAA Particles as a Surrogate to Radioactive Microspheres: Physical and Technical Limitations
Prior to SIRT, a mandatory simulation phase is performed with diagnostic liver arteriography for mapping the tumor vascularization coil embolization if needed (e.g., in case of proximity to arterial digestive branches) and ends with the injection of MAA particles through the angiographic catheter, well positioned in the vascular territory of the tumor(s). This angiographic procedure is followed by planar and SPECT/CT imaging of the MAA distribution. Like 90Y or 166Ho microspheres, MAA is trapped in the arterial microvasculature of the liver and can therefore simulate the distribution of the microspheres. Nevertheless, MAA is not the perfect surrogate because of their physical characteristics, which differ from those of the therapeutic microspheres.
Compared to microspheres, MAA particles have a variable shape, a different size distribution (90% within 10 to 90 μm) with a lower mean size (15 μm versus 25–32 μm), and are injected in much smaller numbers (Table 2) [3,18,19,20]. These differences can explain some discrepancy between the MAA and the microsphere biodistribution, especially the risk of increased shunt to the lungs with MAA.
Table 2.
Characteristics | 99mTc-MAA Particles |
90Y-Resin Microspheres |
90Y-Glass Microspheres |
166Ho-PLLA Microspheres |
---|---|---|---|---|
Diameter (mean) | 15 μm | 32 μm | 25 μm | 30 μm |
Usual number of injected particles (in millions) | 0.3–0.7 | 50 * | 4 * | 30 * |
* For a 3 GBq activity at usual calibration time.
The smallest MAA particles can pass through the liver capillaries and be responsible for over-estimation of the lung shunt [21]. Nevertheless, an excessive MAA uptake in the lungs indicates high hepato-pulmonary shunting and hence a relative contraindication for SIRT (risks of radiation-induced pneumonitis and fibrosis) [22]. A lung shunt superior to 20% or an estimated lung absorbed dose in excess of 30 Gy are contraindications for SIRT [23]. Moreover, these physical differences can be responsible for distribution variations in the healthy and tumor(s) compartments. More precisely, the variable size of MAA particles seems to be a limiting factor in tumor absorbed dose prediction. Indeed, by simulating the treatment with a scout dose of microspheres instead of MAA particles, tumor absorbed doses were more precisely predicted [24,25].
In addition, the lack of precise replication of positioning between the liver arteriographies at the time of simulation and treatment can result in a mismatch between liver distribution of MAA and microspheres. Both procedures require the use of similar angiographic catheters and positioning as well as a well-matched orientation in the artery lumen, and end with an injection of particles at a distance from arterial bifurcations, after controlling for the absence of any vasospasm [26,27]. In addition, MAA particles must also be injected slowly (over 20 s) to better match with the treatment procedure [28].
3. Accuracy of MAA SPECT/CT for Predicting Tumor and Non-Tumoral Whole Liver Absorbed Doses
Some previous studies compared the simulated MAA tumor and non-tumoral liver absorbed doses to the real absorbed doses calculated with post-therapy 90Y imaging. The non-tumoral whole liver absorbed dose (NTWLD) refers to the whole healthy liver (targeted or not by SIRT) as defined recently by the EANM dosimetry committee [18]. In these studies, the strength of the correlation between the MAA absorbed doses and the 90Y absorbed doses was assessed and quantified using the Pearson coefficient R, ranging from 0 (no linear correlation) to 1 (perfect linear correlation) [29]. Figure 2 represents the results of previous studies where the Pearson test was used for comparing the biodistribution of MAA and the radioactive microspheres (more details in Supplementary Materials Table S1). Regarding the estimate of tumor absorbed doses, they demonstrated a variable correlation, with a Pearson coefficient ranging from 0.56 to 0.91. Nevertheless, a large majority of these studies demonstrated a moderate correlation (R < 0.7), and only two studies with a small number of patients showed a higher correlation (R close to 0.9). Another parameter was also compared, i.e., tumor to liver ratio (TNR), and will be discussed further.
Some studies also measured the agreement between these absorbed doses using Bland–Altman analyses [41]. By measuring the differences between these pairs, this statistical test evaluated the risk of errors in MAA imaging for calculating the absorbed doses, after defining the limits of agreement (95% confidence interval). Table 3 summarizes studies performing Bland–Altman analyses and where the relative confidence intervals were available or estimated from absolute values. All these studies demonstrated a similar risk of under- and over-estimation of the tumor absorbed dose (similar plotting over and under the reference X axis). Moreover, the limits of agreement were large in a majority of these studies, with a maximum error over 100% of the tumor absorbed dose estimation in 3 studies.
Table 3.
Studies | Type of Microspheres | Post Therapy Imaging | Type of Tumors | Nb of Patients | Nb of Tumors | TD Estimation (Relative 95% CI) + |
NTWLD Estimation (Relative 95% CI) + |
---|---|---|---|---|---|---|---|
Kafrouni et al. 2019 [36] | glass | 90Y PET/CT | HCC | 23 | 24 | −29%; +29% | −34%; +34% |
Richetta et al. 2019 [32] | resin | 90Y PET/CT | HCC | 10 | 10 | −39%; +33% | −24%; +31% |
Jafargholi Rangraz et al. 2020 [33] | resin | 90Y PET/CT | HCC and mets | 31 | 67 | −169%; +146% | −30%, +23% |
Jadoul et al. 2020 [42] |
resin and glass | 90Y PET/CT | HCC and mets | 57 | 137 | −100%; +100% | −36%, +36% |
d’Abadie et al. 2021 [34] | resin | 90Y PET/CT | HCC and mets | 66 | 171 | −76%; +320% | −19%, +24% |
HCC: hepatocellular carcinoma; mets: metastases; TD: Tumor absorbed dose; Nb: number; NTWLD: non-tumoral whole liver absorbed dose; + In some studies, absolute 95% CI were converted in relative 95% CI (comparison to the mean).
Regarding the prediction of the normal liver absorbed dose, all studies were congruent and demonstrated the accuracy of MAA SPECT/CT. The NTWLD calculated with pre-therapy MAA SPECT/CT was strongly correlated with the NTWLD calculated with post-therapy 90Y PET/CT, with a Pearson coefficient over 0.9 in all studies (Figure 2, Supplementary Materials: Table S1). Moreover, Bland–Altman analyses confirmed this good agreement with a maximal deviation ranging from −36% to +36% (Table 3).
Using alternative methods of comparison in 16 patients, Thomas et al. showed similar results, demonstrating a lack of prediction of the tumor absorbed dose but a highly reliable prediction of the normal liver absorbed dose with MAA imaging [43].
4. Value of MAA SPECT/CT to Predict the Tumor to Normal Liver Uptake Ratio
Some recent studies compared pre-therapy MAA imaging to post-therapy 90Y imaging using the tumor to normal liver uptake ratio (TNR). As previously mentioned, tumors are preferentially vascularized by the liver artery while the healthy liver is vascularized by the portal vein; therefore, TNR is usually greater than 1 [44]. This ratio is also directly correlated to the tumor absorbed dose [45]. The results of these studies are summarized in Figure 2B and detailed in Supplementary Materials (Table S2). They demonstrated results similar to the prediction of tumor absorbed doses. Indeed, TNR was moderately predicted, with a Pearson coefficient ranging from 0.53 to 0.9.
Other studies were performed using an alternative Spearman’s rank test, also demonstrating a moderate correlation in a study analyzing colorectal metastases (ρ = 0.51) and in a study pooling a mix of tumors (ρ = 0.65) [45,46].
Nevertheless, a high TNR uptake seems very well predicted by MAA imaging. In a study evaluating 171 tumors, MAA SPECT/CT predicted a high TNR ratio (≥1.5) with an accuracy of 85% [34]. Moreover, in liver metastases from colorectal cancer (CRC), a TNR ratio superior to 1.7 was a predictor of SIRT efficacy [45]. Therefore, the TNR ratio estimated from MAA imaging can generally select patients suitable for SIRT, avoiding ineffective procedures for patients with TNR ratios close to 1 [13,17].
5. MAA Tumor Absorbed Doses Correlate with Clinical Outcome after SIRT
Despite the low accuracy of MAA SPECT/CT for predicting tumor absorbed doses, numerous data are available for demonstrating a strong relationship between the MAA tumor absorbed dose and SIRT efficacy. In particular, when a certain MAA tumor absorbed dose threshold was achieved, the radiological response and the patient outcome were significantly improved. Table 4 summarizes studies showing this correlation between MAA tumor absorbed doses and the clinical outcome, demonstrating the high significance of this pre-therapy dosimetry. Nevertheless, a distinction must be made between determining an exact tumor absorbed dose and predicting a high level of tumor absorbed dose. Based on previous data, an exact tumor absorbed dose cannot be predicted from MAA imaging, but its level (low versus high) can be determined with high accuracy. When a high tumor absorbed dose is simulated with MAA SPECT/CT, the tumor will receive a high tumor absorbed dose in a large majority of the cases, achieving a threshold of high tumor absorbed dose. In a series of patients treated for HCC by resin microspheres, a threshold dose of 100 Gy simulated with MAA imaging was achieved in 90% of them on 90Y post-therapy imaging [34].
Table 4.
Studies + | Microspheres | Tumors | Number of Patients | Tumor Absorbed Dose Threshold | Predictor of Tumor Response | Predictor of Better Survival (Median) |
---|---|---|---|---|---|---|
Chiesa et al. 2011 [47] | glass | HCC | 46 | ≥257 Gy | 85% | NA |
Garin et al. 2012 [48] | glass | HCC | 36 | ≥205 Gy | 100% | 18 mo ° |
(vs. 9 mo) | ||||||
Mazzaferro et al. 2013 [49] | glass | HCC | 52 | ≥500 Gy | 80% | NA |
Chiesa et al. 2015 [50] | glass | HCC | 52 | ≥217 Gy | 100% | NA |
Garin et al. 2017 [51] | glass | HCC | 85 | ≥205 Gy | 98% | 21 mo ° |
(vs. 6.5 mo) | ||||||
Bourien et al. 2019 [52] | glass | CGC | 64 | ≥260 Gy | 88% * | 28.2 mo ° |
(vs. 11.4 mo) | ||||||
Lam et al. 2013 [53] | resin | CRC | 25 | ≥55 Gy | NA | 32.8 mo ° |
(vs. 7.2 mo) | ||||||
Chansanti et al. 2017 [54] | resin | NET | 15 | ≥191 Gy | 83% | NA |
Piasecki et al. 2018 [45] | resin | CRC | 21 | ≥70 Gy | 99% | NA |
Hermann et al. 2020 [55] | resin | HCC | 121 | ≥100 Gy | 72% | 14.1 mo ° |
(vs. 6.8 mo) | ||||||
Son et al. 2021 [38] | resin | HCC | 34 | ≥125 Gy | 86% | NA |
HCC: hepatocellular carcinoma; CGC: cholangiocarcinoma; NET: neuroendocrine tumors; CRC: colorectal cancer; NA: not available; * subgroup of 29 patients; + Studies with mixed tumor types and without distinction of types of treatment (resin or glass) were excluded. ° differences statistically significant (p < 0.05).
6. Dosimetry Considerations Using MAA SPECT/CT
Using the multicompartmental model with MAA SPECT/CT, activity planning can be determined to reach a tumor absorbed dose threshold. Regarding HCC, it is recommended to target a tumor absorbed dose equal to or above 100–120 Gy with resin microspheres and equal to or above 205–300 Gy with glass microspheres [16,17].
Two dosimetric factors predict SIRT efficacy: a high tumor absorbed dose and homogeneous distribution of the microspheres in the tumor [3,11,56]. The tumor response rate will proportionally increase with the absorbed dose, reaching a plateau of complete response for the highest doses [50,55,56,57,58]. From a dosimetric point of view, the main limitation of SIRT effectiveness is the heterogeneity of the microsphere distribution in the tumor [3]. Tumor absorbed doses can be corrected for heterogeneity using Equivalent Uniform Dose (EUD) [59]. In a series of HCC tumors, the tumor control probability reached 95% for an EUD over 100 Gy, estimated with post therapy 90Y PET/CT [56]. This EUD was achieved for a large range of tumor absorbed doses, varying from 190 (homogeneous distribution) to 1800 Gy (heterogeneous distribution). Therefore, targeting a precise tumor absorbed dose with MAA SPECT/CT seems inappropriate, and the tumor absorbed dose must be the highest possible for increasing the probability of complete response. In addition, as demonstrated in previous data (Table 3), the tumor absorbed dose is overestimated in approximately half of cases with MAA imaging, resulting sometimes in very low and ineffective tumor absorbed doses, as demonstrated in Table 5. Table 5 estimates the confidence intervals of the real (post-therapy) tumor absorbed dose by targeting a precise MAA tumor absorbed dose (derived from data in Table 3). Therefore, the recommended thresholds of tumor absorbed doses targeted with MAA imaging must be considered with caution, and one should always aim for the highest tumor absorbed dose taking into account the liver tolerance. The DOSISPHERE-01 trial illustrated well this concept by adopting an interesting method of activity planning in the arm of patients treated with the multicompartmental model, trying to reach very high levels of tumor absorbed doses, beyond the recommended thresholds [10]. Indeed, the activity was planned to target a tumor absorbed dose over 205 Gy but, if possible, over 250 Gy. In this study arm, the mean tumor absorbed dose planned with MAA SPECT/CT ended up at 332 Gy, largely above the recommended thresholds. Thereafter, the clinical efficacy of this method was strongly confirmed during the follow up.
Table 5.
TD | NTWLD | Studies | |||
---|---|---|---|---|---|
Pre-Therapy (MAA) Target Dose | 100 Gy (Resin) | 205 Gy (Glass) | 40 Gy (Resin) | 70 Gy (Glass) | Nb Patients- Nb Tumors [Reference] |
Confidence intervals of the real absorbed dose
[95% CI] |
[71–129 Gy] | [146–264 Gy] | [26–54 Gy] | [46–94 Gy] | 23–24 |
[36] | |||||
[61–133 Gy] | [125–273 Gy] | [30–52 Gy] | [53–92 Gy] | 10–10 | |
[32] | |||||
[0–246 Gy] | [0–504 Gy] | [28–49 Gy] | [49–86 Gy] | 31–67 | |
[33] | |||||
[0–200 Gy] | [0–410 Gy] | [26–54 Gy] | [45–95 Gy] | 57–137 | |
[42] | |||||
[24–420 Gy] | [49–861 Gy] | [32–50 Gy] | [57–87 Gy] | 66–171 | |
[34] |
TD: Tumor absorbed dose; NTWLD: non-tumoral whole liver absorbed dose; CI: confidence intervals.
Moreover, due to the low predictive value of MAA imaging for tumor absorbed doses, the thresholds of MAA tumor absorbed doses cannot be used for patient selection during the workup. Indeed, a tumor absorbed dose threshold not reached with pre-therapy MAA dosimetry can be finally reached with post-therapy dosimetry. This was demonstrated by previous data analyzing post-therapy 90Y tumor dosimetry, showing a clinical efficacy under these MAA recommended tumor absorbed dose thresholds [13].
Using the multicompartmental model, an alternative strategy can be used: targeting the maximum tolerable absorbed dose to the non-tumoral liver [12,34]. This method can be used in patients with a TNR significantly higher than 1. Therefore, the activity can be significantly increased, as can the tumor absorbed dose and then the tumor control probability [34].
Activity planning is limited by the risk of toxicity related to the radiation of the healthy liver. Radioembolization-induced liver disease (REILD) is the more serious complication of SIRT, occurring in less than 5% of patients. REILD is defined by liver damage occurring within six months after SIRT in absence of tumor progression [16]. Patients with underlying liver disease and especially advanced cirrhosis (elevated baseline bilirubinemia or Child score B) and a low liver reserve (<30% of liver not targeted by SIRT) are at higher risk [60,61]. Additionally, REILD is strongly correlated with the NTWLD, with a significant risk for absorbed doses above 40–50 Gy with resin microspheres and above 70–90 Gy with glass microspheres [12,57,62,63].
In addition, using glass microspheres, Garin et al. demonstrated permanent liver toxicity (≥grade 3) for patients with a healthy liver dose in the targeting liver above 120 Gy when the liver reserve was inferior to 30% [51,61]. Interestingly, 120 Gy in 70% of the healthy liver corresponds to a NTWLD of 80 Gy, equal to the well-known threshold of toxicity. Radiobiological models demonstrated that liver tolerance to radiation is dependent on the liver volume and is very strong for an irradiated healthy liver volume inferior to 40%. In addition, the absorbed dose averaged from the whole normal liver was the strongest parameter correlated with liver toxicity [18]. Therefore, by injecting an activity reaching or staying behind a limit of NTWLD of 40 Gy for resin microspheres and 70 Gy for glass microspheres, the treatment planning can be significantly optimized for clinical efficacy while controlling for the risk of liver toxicity. Indeed, the accurate prediction of the healthy liver dose with MAA imaging allows for performing this safe planning. In the worst scenario, the NTWLD will stay around the safe thresholds (Table 5).
The formula for calculating the activity based on this non-tumoral liver absorbed dose reads as follows [34]:
With A: the activity planned (GBq), CWL: the counts in the whole liver defined on MAA SPECT, CNTL: the counts in the non-tumoral liver defined on MAA SPECT, NTLD: the established absorbed dose to the non-tumoral liver (Gy), MNTL: the mass of the non-tumoral liver (kg), LSF: the lung shunt fraction.
7. Pre-Therapy Imaging with 166Ho SPECT/CT
166Ho (PLLA) microspheres can be used as an alternative to 90Y (resin or glass) microspheres. The 166Ho radionuclide emits a beta radiation responsible for its therapeutic effect but also a small amount of gamma radiation (Table 1) that allows imaging by a SPECT system (166Ho SPECT/CT) [3]. 166Ho scout dose microspheres (maximal activity of 250 MBq) is a safe alternative to MAA particles for evaluating the feasibility of the treatment [8].
166Ho scout microspheres have the same shape and size as 166Ho microspheres used for therapy. Therefore, pre-therapy 166Ho SPECT/CT is expected to be more predictive than MAA SPECT/CT. A previous study comparing these imaging methods demonstrated that the lung absorbed dose was more precisely evaluated with this scout dose compared to MAA imaging [64]. Indeed, MAA particles can overestimate the lung shunt due to a fraction of some small particles (i.e., 10 μm) which can pass through the liver capillaries and reach the lungs [65]. Regarding the intrahepatic biodistribution, pre-therapy 166Ho SPECT/CT can also estimate the healthy liver and tumor absorbed doses. Smits et al. demonstrated similar values of pre-therapy 166Ho SPECT/CT and pre-therapy MAA SPECT/CT for predicting the healthy liver absorbed dose. Regarding the tumor absorbed dose, pre-therapy 166Ho SPECT/CT was more predictive than MAA SPECT/CT, but the absolute confidence intervals of Bland–Altman analyses were still significant, demonstrating only a moderate prediction (CI 95%: −90.3 Gy; + 105.3 Gy) [24].
The predictive value of the 166Ho scout dose is also limited by the spatial resolution of 166Ho SPECT/CT. Indeed, the 166Ho radionuclide emits a multitude of gamma radiations of high energy (MeV range), which interact with the patient’s body and mainly with the lead collimator of the SPECT/CT [66]. Therefore, secondary lead X-rays are produced that fall in the energy window set around the 81 keV photopeak of 166Ho, limiting significantly the spatial resolution and the quantitative assessment of the microsphere distribution. Last generation SPECT/CT systems using tungsten collimators would not be affected by this problem, improving in theory the spatial resolution and the quantification. Additionally, a loss of the spatial resolution is also due to the use of a medium energy collimator used during the SPECT acquisition [67]. Therefore, previous dosimetry studies used complex reconstruction algorithms, with Monte Carlo simulation modelling, for improving the resolution of the SPECT/CT and the accuracy of dosimetry [8,21,24,66,68,69]. These Monte Carlo simulations are for now only used in research and development, involving important computer processing resources and time, with software that is not commercialized nor available for routine clinical use. Therefore, the monocompartmental model is recommended using 166Ho microspheres, targeting a maximum of 60 Gy to the whole liver volume [70]. Nevertheless, a recent study evaluating post-therapy 166Ho SPECT/CT dosimetry demonstrated a dose response relationship and a better outcome for patients treated with a tumor absorbed dose of at least 90 Gy (for colorectal metastases). In addition, no liver toxicity was shown in this population of patients with a NTWLD reaching 55 Gy [69]. Similar studies evaluating dose effect relationships using the pre-therapy 166Ho scout dose would support in the future the use of the partition model for activity planning.
8. Conclusions
Pre-therapy imaging in SIRT is strongly involved in a theranostic approach for conducting treatment planning. MAA SPECT/CT plays a crucial role in selecting potential good responders to SIRT. Patients with high MAA tumor uptake have a high probability of receiving high and efficient tumor absorbed doses after SIRT. Moreover, pre-therapy imaging accurately predicts the healthy liver absorbed dose and, therefore, the planned therapeutic activity can be optimized while still staying behind the threshold dose of liver toxicity. Using this model, treatment planning really becomes personalized, improving the safety and the clinical benefits of SIRT.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/jcm11237245/s1, Table S1: Studies comparing the tumor and normal liver absorbed doses using MAA SPECT/CT to 90Y post treatment imaging using the Pearson coefficient for measuring the strength of the correlation; Table S2: Studies evaluating the accuracy of the prediction of tumor to normal liver uptake (TNR) with MAA SPECT/CT.
Author Contributions
Conceptualization, P.d. and F.J.; methodology, P.d., S.W. and F.J.; software, P.d. and S.W.; validation, all authors; formal analysis, P.d., S.W. and F.J.; investigation, P.d., S.W., M.H. and F.J.; resources, R.L. and F.J.; data curation, P.d., F.J., S.W. and M.H.; writing—original draft preparation, P.d., S.W., M.H. and F.J.; writing—review and editing, all authors; visualization, all authors; supervision, P.d. and F.J., project administration, P.d. and F.J., funding acquisition, F.J. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
No new data were created in this study. Data sharing is not applicable in this article.
Conflicts of Interest
The authors declare no conflict of interest.
Funding Statement
This research received no external funding.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Bastiaannet R., Kappadath S.C., Kunnen B., Braat A., Lam M., de Jong H. The physics of radioembolization. EJNMMI Phys. 2018;5:22. doi: 10.1186/s40658-018-0221-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Filippi L., Braat A.J. Theragnostics in primary and secondary liver tumors: The need for a personalized approach. Q. J. Nucl. Med. Mol. Imaging. 2021;65:353–370. doi: 10.23736/S1824-4785.21.03407-5. [DOI] [PubMed] [Google Scholar]
- 3.D’Abadie P., Hesse M., Louppe A., Lhommel R., Walrand S., Jamar F. Microspheres Used in Liver Radioembolization: From Conception to Clinical Effects. Molecules. 2021;26:3966. doi: 10.3390/molecules26133966. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Gomes Marin J.F., Nunes R.F., Coutinho A.M., Zaniboni E.C., Costa L.B., Barbosa F.G., Queiroz M.A., Cerri G.G., Buchpiguel C.A. Theranostics in Nuclear Medicine: Emerging and Re-emerging Integrated Imaging and Therapies in the Era of Precision Oncology. Radiographics. 2020;40:1715–1740. doi: 10.1148/rg.2020200021. [DOI] [PubMed] [Google Scholar]
- 5.Georgiou M.F., Kuker R.A., Studenski M.T., Ahlman P.P., Witte M., Portelance L. Lung shunt fraction calculation using (99m)Tc-MAA SPECT/CT imaging for (90)Y microsphere selective internal radiation therapy of liver tumors. EJNMMI Res. 2021;11:96. doi: 10.1186/s13550-021-00837-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Theysohn J.M., Ruhlmann M., Muller S., Dechene A., Best J., Haubold J., Umutlu L., Gerken G., Bockisch A., Lauenstein T.C. Radioembolization with Y-90 Glass Microspheres: Do We Really Need SPECT-CT to Identify Extrahepatic Shunts? PLoS ONE. 2015;10:e0137587. doi: 10.1371/journal.pone.0137587. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Sangro B., Inarrairaegui M., Bilbao J.I. Radioembolization for hepatocellular carcinoma. J. Hepatol. 2012;56:464–473. doi: 10.1016/j.jhep.2011.07.012. [DOI] [PubMed] [Google Scholar]
- 8.Braat A., Prince J.F., van Rooij R., Bruijnen R.C.G., van den Bosch M., Lam M. Safety analysis of holmium-166 microsphere scout dose imaging during radioembolisation work-up: A cohort study. Eur. Radiol. 2018;28:920–928. doi: 10.1007/s00330-017-4998-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Van de Maat G.H., Seevinck P.R., Elschot M., Smits M.L., de Leeuw H., van Het Schip A.D., Vente M.A., Zonnenberg B.A., de Jong H.W., Lam M.G., et al. MRI-based biodistribution assessment of holmium-166 poly(L-lactic acid) microspheres after radioembolisation. Eur. Radiol. 2013;23:827–835. doi: 10.1007/s00330-012-2648-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Garin E., Tselikas L., Guiu B., Chalaye J., Edeline J., de Baere T., Assenat E., Tacher V., Robert C., Terroir-Cassou-Mounat M., et al. Personalised versus standard dosimetry approach of selective internal radiation therapy in patients with locally advanced hepatocellular carcinoma (DOSISPHERE-01): A randomised, multicentre, open-label phase 2 trial. Lancet Gastroenterol. Hepatol. 2021;6:17–29. doi: 10.1016/S2468-1253(20)30290-9. [DOI] [PubMed] [Google Scholar]
- 11.D’Abadie P., Walrand S., Hesse M., Annet L., Borbath I., Van den Eynde M., Lhommel R., Jamar F. Prediction of tumor response and patient outcome after radioembolization of hepatocellular carcinoma using 90Y-PET-computed tomography dosimetry. Nucl. Med. Commun. 2021;42:747–754. doi: 10.1097/MNM.0000000000001395. [DOI] [PubMed] [Google Scholar]
- 12.Chiesa C., Mira M., Bhoori S., Bormolini G., Maccauro M., Spreafico C., Cascella T., Cavallo A., De Nile M.C., Mazzaglia S., et al. Radioembolization of hepatocarcinoma with (90)Y glass microspheres: Treatment optimization using the dose-toxicity relationship. Eur. J. Nucl. Med. Mol. Imaging. 2020;47:3018–3032. doi: 10.1007/s00259-020-04845-4. [DOI] [PubMed] [Google Scholar]
- 13.D’Abadie P., Walrand S., Lhommel R., Hesse M., Borbath I., Jamar F. Optimization of the Clinical Effectiveness of Radioembolization in Hepatocellular Carcinoma with Dosimetry and Patient-Selection Criteria. Curr. Oncol. 2022;29:2422–2434. doi: 10.3390/curroncol29040196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Lau W.Y., Kennedy A.S., Kim Y.H., Lai H.K., Lee R.C., Leung T.W., Liu C.S., Salem R., Sangro B., Shuter B., et al. Patient selection and activity planning guide for selective internal radiotherapy with yttrium-90 resin microspheres. Int. J. Radiat. Oncol. Biol. Phys. 2012;82:401–407. doi: 10.1016/j.ijrobp.2010.08.015. [DOI] [PubMed] [Google Scholar]
- 15.Cicone F., Gnesin S., Cremonesi M. Dosimetry of nuclear medicine therapies: Current controversies and impact on treatment optimization. Q. J. Nucl. Med. Mol. Imaging. 2021;65:327–332. doi: 10.23736/S1824-4785.21.03418-X. [DOI] [PubMed] [Google Scholar]
- 16.Weber M., Lam M., Chiesa C., Konijnenberg M., Cremonesi M., Flamen P., Gnesin S., Bodei L., Kracmerova T., Luster M., et al. EANM procedure guideline for the treatment of liver cancer and liver metastases with intra-arterial radioactive compounds. Eur. J. Nucl. Med. Mol. Imaging. 2022;49:1682–1699. doi: 10.1007/s00259-021-05600-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Levillain H., Bagni O., Deroose C.M., Dieudonne A., Gnesin S., Grosser O.S., Kappadath S.C., Kennedy A., Kokabi N., Liu D.M., et al. International recommendations for personalised selective internal radiation therapy of primary and metastatic liver diseases with yttrium-90 resin microspheres. Eur. J. Nucl. Med. Mol. Imaging. 2021;48:1570–1584. doi: 10.1007/s00259-020-05163-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Chiesa C., Sjogreen-Gleisner K., Walrand S., Strigari L., Flux G., Gear J., Stokke C., Gabina P.M., Bernhardt P., Konijnenberg M. EANM dosimetry committee series on standard operational procedures: A unified methodology for (99m)Tc-MAA pre- and (90)Y peri-therapy dosimetry in liver radioembolization with (90)Y microspheres. EJNMMI Phys. 2021;8:77. doi: 10.1186/s40658-021-00394-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Reinders M.T.M., Smits M.L.J., van Roekel C., Braat A. Holmium-166 Microsphere Radioembolization of Hepatic Malignancies. Semin. Nucl. Med. 2019;49:237–243. doi: 10.1053/j.semnuclmed.2019.01.008. [DOI] [PubMed] [Google Scholar]
- 20.Al-Janabi M.A., Yousif Z.M., Kadim A.H., Al-Salem A.M. A new technique for the preparation of ready-to-use macroaggregated albumin (MAA) kits to be labelled with 99mTc for lung scanning. Int. J. Appl. Radiat. Isot. 1983;34:1473–1478. doi: 10.1016/0020-708X(83)90050-9. [DOI] [PubMed] [Google Scholar]
- 21.Elschot M., Nijsen J.F., Lam M.G., Smits M.L., Prince J.F., Viergever M.A., van den Bosch M.A., Zonnenberg B.A., de Jong H.W. ((9)(9)m)Tc-MAA overestimates the absorbed dose to the lungs in radioembolization: A quantitative evaluation in patients treated with (1)(6)(6)Ho-microspheres. Eur. J. Nucl. Med. Mol. Imaging. 2014;41:1965–1975. doi: 10.1007/s00259-014-2784-9. [DOI] [PubMed] [Google Scholar]
- 22.Leung T.W., Lau W.Y., Ho S.K., Ward S.C., Chow J.H., Chan M.S., Metreweli C., Johnson P.J., Li A.K. Radiation pneumonitis after selective internal radiation treatment with intraarterial 90yttrium-microspheres for inoperable hepatic tumors. Int. J. Radiat. Oncol. Biol. Phys. 1995;33:919–924. doi: 10.1016/0360-3016(95)00039-3. [DOI] [PubMed] [Google Scholar]
- 23.Ward T.J., Tamrazi A., Lam M.G., Louie J.D., Kao P.N., Shah R.P., Kadoch M.A., Sze D.Y. Management of High Hepatopulmonary Shunting in Patients Undergoing Hepatic Radioembolization. J. Vasc. Interv. Radiol. 2015;26:1751–1760. doi: 10.1016/j.jvir.2015.08.027. [DOI] [PubMed] [Google Scholar]
- 24.Smits M.L.J., Dassen M.G., Prince J.F., Braat A., Beijst C., Bruijnen R.C.G., de Jong H., Lam M. The superior predictive value of (166)Ho-scout compared with (99m)Tc-macroaggregated albumin prior to (166)Ho-microspheres radioembolization in patients with liver metastases. Eur. J. Nucl. Med. Mol. Imaging. 2020;47:798–806. doi: 10.1007/s00259-019-04460-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Kokabi N., Webster L.A., Elsayed M., Switchenko J.M., Chen B., Brandon D., Galt J., Sethi I., Cristescu M., Kappadath C., et al. Accuracy and Safety of Scout Dose Resin Yttrium-90 Microspheres for Radioembolization Therapy Treatment Planning: A prospective single-arm clinical trial. J. Vasc. Interv. Radiol. 2022;33:1578–1587. doi: 10.1016/j.jvir.2022.08.027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Garin E., Rolland Y., Laffont S., Edeline J. Clinical impact of (99m)Tc-MAA SPECT/CT-based dosimetry in the radioembolization of liver malignancies with (90)Y-loaded microspheres. Eur. J. Nucl. Med. Mol. Imaging. 2016;43:559–575. doi: 10.1007/s00259-015-3157-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.D’Abadie P., Walrand S., Goffette P., Amini N., Maanen A.V., Lhommel R., Jamar F. Antireflux catheter improves tumor targeting in liver radioembolization with resin microspheres. Diagn. Interv. Radiol. 2021;27:768–773. doi: 10.5152/dir.2021.20785. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Garin E., Guiu B., Edeline J., Rolland Y., Palard X. Trans-arterial Radioembolization Dosimetry in 2022. Cardiovasc. Interv. Radiol. 2022;45:1608–1621. doi: 10.1007/s00270-022-03215-x. [DOI] [PubMed] [Google Scholar]
- 29.Schober P., Boer C., Schwarte L.A. Correlation Coefficients: Appropriate Use and Interpretation. Anesth. Analg. 2018;126:1763–1768. doi: 10.1213/ANE.0000000000002864. [DOI] [PubMed] [Google Scholar]
- 30.Song Y.S., Paeng J.C., Kim H.C., Chung J.W., Cheon G.J., Chung J.K., Lee D.S., Kang K.W. PET/CT-Based Dosimetry in 90Y-Microsphere Selective Internal Radiation Therapy: Single Cohort Comparison With Pretreatment Planning on (99m)Tc-MAA Imaging and Correlation with Treatment Efficacy. Medicine. 2015;94:e945. doi: 10.1097/MD.0000000000000945. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Gnesin S., Canetti L., Adib S., Cherbuin N., Silva Monteiro M., Bize P., Denys A., Prior J.O., Baechler S., Boubaker A. Partition Model-Based 99mTc-MAA SPECT/CT Predictive Dosimetry Compared with 90Y TOF PET/CT Posttreatment Dosimetry in Radioembolization of Hepatocellular Carcinoma: A Quantitative Agreement Comparison. J. Nucl. Med. 2016;57:1672–1678. doi: 10.2967/jnumed.116.173104. [DOI] [PubMed] [Google Scholar]
- 32.Richetta E., Pasquino M., Poli M., Cutaia C., Valero C., Tabone M., Paradisi B.P., Pacilio M., Pellerito R.E., Stasi M. PET-CT post therapy dosimetry in radioembolization with resin (90)Y microspheres: Comparison with pre-treatment SPECT-CT (99m)Tc-MAA results. Phys. Med. 2019;64:16–23. doi: 10.1016/j.ejmp.2019.05.025. [DOI] [PubMed] [Google Scholar]
- 33.Jafargholi Rangraz E., Tang X., Van Laeken C., Maleux G., Dekervel J., Van Cutsem E., Verslype C., Baete K., Nuyts J., Deroose C.M. Quantitative comparison of pre-treatment predictive and post-treatment measured dosimetry for selective internal radiation therapy using cone-beam CT for tumor and liver perfusion territory definition. EJNMMI Res. 2020;10:94. doi: 10.1186/s13550-020-00675-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.D’Abadie P., Walrand S., Hesse M., Amini N., Lhommel R., Sawadogo K., Jamar F. Accurate non-tumoral 99mTc-MAA absorbed dose prediction to plan optimized activities in liver radioembolization using resin microspheres. Phys. Med. 2021;89:250–257. doi: 10.1016/j.ejmp.2021.07.032. [DOI] [PubMed] [Google Scholar]
- 35.Haste P., Tann M., Persohn S., LaRoche T., Aaron V., Mauxion T., Chauhan N., Dreher M.R., Johnson M.S. Correlation of Technetium-99m Macroaggregated Albumin and Yttrium-90 Glass Microsphere Biodistribution in Hepatocellular Carcinoma: A Retrospective Review of Pretreatment Single Photon Emission CT and Posttreatment Positron Emission Tomography/CT. J. Vasc. Interv. Radiol. 2017;28:722–730. doi: 10.1016/j.jvir.2016.12.1221. [DOI] [PubMed] [Google Scholar]
- 36.Kafrouni M., Allimant C., Fourcade M., Vauclin S., Guiu B., Mariano-Goulart D., Ben Bouallegue F. Analysis of differences between (99m)Tc-MAA SPECT- and (90)Y-microsphere PET-based dosimetry for hepatocellular carcinoma selective internal radiation therapy. EJNMMI Res. 2019;9:62. doi: 10.1186/s13550-019-0533-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Debebe S.A., Adjouadi M., Gulec S.A., Franquiz J., McGoron A.J. (90) Y SPECT/CT quantitative study and comparison of uptake with pretreatment (99 m) Tc-MAA SPECT/CT in radiomicrosphere therapy. J. Appl. Clin. Med. Phys. 2019;20:30–42. doi: 10.1002/acm2.12512. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Son M.H., Ha L.N., Bang M.H., Bae S., Giang D.T., Thinh N.T., Paeng J.C. Diagnostic and prognostic value of (99m)Tc-MAA SPECT/CT for treatment planning of (90)Y-resin microsphere radioembolization for hepatocellular carcinoma: Comparison with planar image. Sci. Rep. 2021;11:3207. doi: 10.1038/s41598-021-82887-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Villalobos A., Cheng B., Wagstaff W., Sethi I., Bercu Z., Schuster D.M., Brandon D.C., Galt J., Kokabi N. Tumor-to-Normal Ratio Relationship between Planning Technetium-99 Macroaggregated Albumin and Posttherapy Yttrium-90 Bremsstrahlung SPECT/CT. J. Vasc. Interv. Radiol. 2021;32:752–760. doi: 10.1016/j.jvir.2020.12.023. [DOI] [PubMed] [Google Scholar]
- 40.Meine T.C., Brunkhorst T., Werncke T., Schutze C., Vogel A., Kirstein M.M., Dewald C.L.A., Becker L.S., Maschke S.K., Kretschmann N., et al. Comparison of the Uptake of Hepatocellular Carcinoma on Pre-Therapeutic MDCT, CACT, and SPECT/CT, and the Correlation with Post-Therapeutic PET/CT in Patients Undergoing Selective Internal Radiation Therapy. J. Clin. Med. 2021;10:3837. doi: 10.3390/jcm10173837. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Giavarina D. Understanding Bland Altman analysis. Biochem. Med. 2015;25:141–151. doi: 10.11613/BM.2015.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Jadoul A., Bernard C., Lovinfosse P., Gerard L., Lilet H., Cornet O., Hustinx R. Comparative dosimetry between (99m)Tc-MAA SPECT/CT and (90)Y PET/CT in primary and metastatic liver tumors. Eur. J. Nucl. Med. Mol. Imaging. 2020;47:828–837. doi: 10.1007/s00259-019-04465-7. [DOI] [PubMed] [Google Scholar]
- 43.Thomas M.A., Mahvash A., Abdelsalam M., Kaseb A.O., Kappadath S.C. Planning dosimetry for (90) Y radioembolization with glass microspheres: Evaluating the fidelity of (99m) Tc-MAA and partition model predictions. Med. Phys. 2020;47:5333–5342. doi: 10.1002/mp.14452. [DOI] [PubMed] [Google Scholar]
- 44.Ilhan H., Goritschan A., Paprottka P., Jakobs T.F., Fendler W.P., Todica A., Bartenstein P., Hacker M., Haug A.R. Predictive Value of 99mTc-MAA SPECT for 90Y-Labeled Resin Microsphere Distribution in Radioembolization of Primary and Secondary Hepatic Tumors. J. Nucl. Med. 2015;56:1654–1660. doi: 10.2967/jnumed.115.162685. [DOI] [PubMed] [Google Scholar]
- 45.Piasecki P., Narloch J., Brzozowski K., Ziecina P., Mazurek A., Budzynska A., Korniluk J., Dziuk M. The Predictive Value of SPECT/CT imaging in colorectal liver metastases response after 90Y-radioembolization. PLoS ONE. 2018;13:e0200488. doi: 10.1371/journal.pone.0200488. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Rhee S., Kim S., Cho J., Park J., Eo J.S., Park S., Lee E., Kim Y.H., Choe J.G. Semi-Quantitative Analysis of Post-Transarterial Radioembolization (90)Y Microsphere Positron Emission Tomography Combined with Computed Tomography (PET/CT) Images in Advanced Liver Malignancy: Comparison With (99m)Tc Macroaggregated Albumin (MAA) Single Photon Emission Computed Tomography (SPECT) Nucl. Med. Mol. Imaging. 2016;50:63–69. doi: 10.1007/s13139-015-0366-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Chiesa C., Maccauro M., Romito R., Spreafico C., Pellizzari S., Negri A., Sposito C., Morosi C., Civelli E., Lanocita R., et al. Need, feasibility and convenience of dosimetric treatment planning in liver selective internal radiation therapy with (90)Y microspheres: The experience of the National Tumor Institute of Milan. Q. J. Nucl. Med. Mol. Imaging. 2011;55:168–197. [PubMed] [Google Scholar]
- 48.Garin E., Lenoir L., Rolland Y., Edeline J., Mesbah H., Laffont S., Poree P., Clement B., Raoul J.L., Boucher E. Dosimetry based on 99mTc-macroaggregated albumin SPECT/CT accurately predicts tumor response and survival in hepatocellular carcinoma patients treated with 90Y-loaded glass microspheres: Preliminary results. J. Nucl. Med. 2012;53:255–263. doi: 10.2967/jnumed.111.094235. [DOI] [PubMed] [Google Scholar]
- 49.Mazzaferro V., Sposito C., Bhoori S., Romito R., Chiesa C., Morosi C., Maccauro M., Marchiano A., Bongini M., Lanocita R., et al. Yttrium-90 radioembolization for intermediate-advanced hepatocellular carcinoma: A phase 2 study. Hepatology. 2013;57:1826–1837. doi: 10.1002/hep.26014. [DOI] [PubMed] [Google Scholar]
- 50.Chiesa C., Mira M., Maccauro M., Spreafico C., Romito R., Morosi C., Camerini T., Carrara M., Pellizzari S., Negri A., et al. Radioembolization of hepatocarcinoma with (90)Y glass microspheres: Development of an individualized treatment planning strategy based on dosimetry and radiobiology. Eur. J. Nucl. Med. Mol. Imaging. 2015;42:1718–1738. doi: 10.1007/s00259-015-3068-8. [DOI] [PubMed] [Google Scholar]
- 51.Garin E., Rolland Y., Pracht M., Le Sourd S., Laffont S., Mesbah H., Haumont L.A., Lenoir L., Rohou T., Brun V., et al. High impact of macroaggregated albumin-based tumour dose on response and overall survival in hepatocellular carcinoma patients treated with (90) Y-loaded glass microsphere radioembolization. Liver Int. 2017;37:101–110. doi: 10.1111/liv.13220. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Bourien H., Palard X., Rolland Y., Le Du F., Beuzit L., Uguen T., Le Sourd S., Pracht M., Manceau V., Lievre A., et al. Yttrium-90 glass microspheres radioembolization (RE) for biliary tract cancer: A large single-center experience. Eur. J. Nucl. Med. Mol. Imaging. 2019;46:669–676. doi: 10.1007/s00259-018-4199-5. [DOI] [PubMed] [Google Scholar]
- 53.Lam M.G., Goris M.L., Iagaru A.H., Mittra E.S., Louie J.D., Sze D.Y. Prognostic utility of 90Y radioembolization dosimetry based on fusion 99mTc-macroaggregated albumin-99mTc-sulfur colloid SPECT. J. Nucl. Med. 2013;54:2055–2061. doi: 10.2967/jnumed.113.123257. [DOI] [PubMed] [Google Scholar]
- 54.Chansanti O., Jahangiri Y., Matsui Y., Adachi A., Geeratikun Y., Kaufman J.A., Kolbeck K.J., Stevens J.S., Farsad K. Tumor Dose Response in Yttrium-90 Resin Microsphere Embolization for Neuroendocrine Liver Metastases: A Tumor-Specific Analysis with Dose Estimation Using SPECT-CT. J. Vasc. Interv. Radiol. 2017;28:1528–1535. doi: 10.1016/j.jvir.2017.07.008. [DOI] [PubMed] [Google Scholar]
- 55.Hermann A.L., Dieudonne A., Ronot M., Sanchez M., Pereira H., Chatellier G., Garin E., Castera L., Lebtahi R., Vilgrain V., et al. Relationship of Tumor Radiation-absorbed Dose to Survival and Response in Hepatocellular Carcinoma Treated with Transarterial Radioembolization with (90)Y in the SARAH Study. Radiology. 2020;296:673–684. doi: 10.1148/radiol.2020191606. [DOI] [PubMed] [Google Scholar]
- 56.D’Abadie P., Walrand S., Hesse M., Borbath I., Lhommel R., Jamar F. TCP post-radioembolization and TCP post-EBRT in HCC are similar and can be predicted using the in vitro radiosensitivity. EJNMMI Res. 2022;12:40. doi: 10.1186/s13550-022-00911-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Strigari L., Sciuto R., Rea S., Carpanese L., Pizzi G., Soriani A., Iaccarino G., Benassi M., Ettorre G.M., Maini C.L. Efficacy and toxicity related to treatment of hepatocellular carcinoma with 90Y-SIR spheres: Radiobiologic considerations. J. Nucl. Med. 2010;51:1377–1385. doi: 10.2967/jnumed.110.075861. [DOI] [PubMed] [Google Scholar]
- 58.Romano C., Mazzaglia S., Maccauro M., Spreafico C., Gabutti A., Maffi G., Morosi C., Cascella T., Mira M., De Nile M.C., et al. Radioembolization of Hepatocellular Carcinoma with (90)Y Glass Microspheres: No Advantage of Voxel Dosimetry with Respect to Mean Dose in Dose-Response Analysis with Two Radiological Methods. Cancers. 2022;14:959. doi: 10.3390/cancers14040959. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.D’Abadie P., Hesse M., Jamar F., Lhommel R., Walrand S. (90)Y TOF-PET based EUD reunifies patient survival prediction in resin and glass microspheres radioembolization of HCC tumours. Phys. Med. Biol. 2018;63:245010. doi: 10.1088/1361-6560/aaf205. [DOI] [PubMed] [Google Scholar]
- 60.Gil-Alzugaray B., Chopitea A., Inarrairaegui M., Bilbao J.I., Rodriguez-Fraile M., Rodriguez J., Benito A., Dominguez I., D’Avola D., Herrero J.I., et al. Prognostic factors and prevention of radioembolization-induced liver disease. Hepatology. 2013;57:1078–1087. doi: 10.1002/hep.26191. [DOI] [PubMed] [Google Scholar]
- 61.Garin E., Lenoir L., Edeline J., Laffont S., Mesbah H., Poree P., Sulpice L., Boudjema K., Mesbah M., Guillygomarc’h A., et al. Boosted selective internal radiation therapy with 90Y-loaded glass microspheres (B-SIRT) for hepatocellular carcinoma patients: A new personalized promising concept. Eur. J. Nucl. Med. Mol. Imaging. 2013;40:1057–1068. doi: 10.1007/s00259-013-2395-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Cremonesi M., Ferrari M., Bartolomei M., Orsi F., Bonomo G., Arico D., Mallia A., De Cicco C., Pedroli G., Paganelli G. Radioembolisation with 90Y-microspheres: Dosimetric and radiobiological investigation for multi-cycle treatment. Eur. J. Nucl. Med. Mol. Imaging. 2008;35:2088–2096. doi: 10.1007/s00259-008-0857-3. [DOI] [PubMed] [Google Scholar]
- 63.Alsultan A.A., van Roekel C., Barentsz M.W., Smits M.L.J., Kunnen B., Koopman M., Bruijnen R.C.G., de Keizer B., Lam M. Dose-response and dose-toxicity relationships for yttrium-90 glass radioembolization in patients with colorectal cancer liver metastases. J. Nucl. Med. 2021;120:255745. doi: 10.2967/jnumed.120.255745. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Prince J.F., van Rooij R., Bol G.H., de Jong H.W., van den Bosch M.A., Lam M.G. Safety of a Scout Dose Preceding Hepatic Radioembolization with 166Ho Microspheres. J. Nucl. Med. 2015;56:817–823. doi: 10.2967/jnumed.115.155564. [DOI] [PubMed] [Google Scholar]
- 65.Chiesa C., Maccauro M. (166)Ho microsphere scout dose for more accurate radioembolization treatment planning. Eur. J. Nucl. Med. Mol. Imaging. 2020;47:744–747. doi: 10.1007/s00259-019-04617-9. [DOI] [PubMed] [Google Scholar]
- 66.Elschot M., Smits M.L., Nijsen J.F., Lam M.G., Zonnenberg B.A., van den Bosch M.A., Viergever M.A., de Jong H.W. Quantitative Monte Carlo-based holmium-166 SPECT reconstruction. Med. Phys. 2013;40:112502. doi: 10.1118/1.4823788. [DOI] [PubMed] [Google Scholar]
- 67.Elschot M., Nijsen J.F., Dam A.J., de Jong H.W. Quantitative evaluation of scintillation camera imaging characteristics of isotopes used in liver radioembolization. PLoS ONE. 2011;6:e26174. doi: 10.1371/journal.pone.0026174. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Van Roekel C., van den Hoven A.F., Bastiaannet R., Bruijnen R.C.G., Braat A., de Keizer B., Lam M., Smits M.L.J. Use of an anti-reflux catheter to improve tumor targeting for holmium-166 radioembolization-a prospective, within-patient randomized study. Eur. J. Nucl. Med. Mol. Imaging. 2021;48:1658–1668. doi: 10.1007/s00259-020-05079-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Van Roekel C., Bastiaannet R., Smits M.L.J., Bruijnen R.C., Braat A., de Jong H., Elias S.G., Lam M. Dose-Effect Relationships of (166)Ho Radioembolization in Colorectal Cancer. J. Nucl. Med. 2021;62:272–279. doi: 10.2967/jnumed.120.243832. [DOI] [PubMed] [Google Scholar]
- 70.Stella M., Braat A., van Rooij R., de Jong H., Lam M. Holmium-166 Radioembolization: Current Status and Future Prospective. Cardiovasc. Interv. Radiol. 2022;45:1634–1645. doi: 10.1007/s00270-022-03187-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
No new data were created in this study. Data sharing is not applicable in this article.