Diagnostic accuracy of autoverification and guidance system for COVID-19 RT-PCR results

Yingmu Cai 1,2,3,#, Mengyu Liu 1,3,#, Zhiyuan Wu 4,5,#, Cuihong Tian 3,5,6,#, Song Qiu 2, Zhen Li 7, Feng Xu 7, Wei Li 1,3, Yan Zheng 8, Aijuan Xu 7, Longxu Xie 2,7,, Xuerui Tan 3,
PMCID: PMC9755791  PMID: 36540610

Abstract

Background

To date, most countries worldwide have declared that the pandemic of COVID-19 is over, while the WHO has not officially ended the COVID-19 pandemic, and China still insists on the personalized dynamic COVID-free policy. Large-scale nucleic acid testing in Chinese communities and the manual interpretation for SARS-CoV-2 nucleic acid detection results pose a huge challenge for labour, quality and turnaround time (TAT) requirements. To solve this specific issue while increase the efficiency and accuracy of interpretation, we created an autoverification and guidance system (AGS) that can automatically interpret and report the COVID-19 reverse transcriptase-polymerase chain reaction (RT-PCR) results relaying on computer-based autoverification procedure and then validated its performance in real-world environments. This would be conductive to transmission risk prediction, COVID-19 prevention and control and timely medical treatment for positive patients in the context of the predictive, preventive and personalized medicine (PPPM).

Methods

A diagnostic accuracy test was conducted with 380,693 participants from two COVID-19 test sites in China, the Hong Kong Hybribio Medical Laboratory (n = 266,035) and the mobile medical shelter at a Shanghai airport (n = 114,658). These participants underwent SARS-CoV-2 RT-PCR from March 28 to April 10, 2022. All RT-PCR results were interpreted by laboratorians and by using AGS simultaneously. Considering the manual interpretation as gold standard, the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy were applied to evaluate the diagnostic value of the AGS on the interpretation of RT-PCR results.

Results

Among the 266,035 samples in Hong Kong, there were 16,356 (6.15%) positive, 231,073 (86.86%) negative, 18,606 (6.99%) indefinite, 231,073 (86.86%, negative) no retest required and 34,962 (13.14%, positive and indefinite) retest required; the 114,658 samples in Shanghai consisted of 76 (0.07%) positive, 109,956 (95.90%) negative, 4626 (4.03%) indefinite, 109,956 (95.90%, negative) no retest required and 4702 (4.10%, positive and indefinite) retest required. Compared to the fashioned manual interpretation, the AGS is a procedure of high accuracy [99.96% (95%CI, 99.95–99.97%) in Hong Kong and 100% (95%CI, 100–100%) in Shanghai] with perfect sensitivity [99.98% (95%CI, 99.97–99.98%) in Hong Kong and 100% (95%CI, 100–100%) in Shanghai], specificity [99.87% (95%CI, 99.82–99.90%) in Hong Kong and 100% (95%CI, 99.92–100%) in Shanghai], PPV [99.98% (95%CI, 99.97–99.99%) in Hong Kong and 100% (95%CI, 99.99–100%) in Shanghai] and NPV [99.85% (95%CI, 99.80–99.88%) in Hong Kong and 100% (95%CI, 99.90–100%) in Shanghai]. The need for manual interpretation of total samples was dramatically reduced from 100% to 13.1% and the interpretation time fell from 53 h to 26 min in Hong Kong; while the manual interpretation of total samples was decreased from 100% to 4.1% and the interpretation time dropped from 20 h to 16 min at Shanghai.

Conclusions

The AGS is a procedure of high accuracy and significantly relieves both labour and time from the challenge of large-scale screening of SARS-CoV-2 using RT-PCR. It should be recommended as a powerful screening, diagnostic and predictive system for SARS-CoV-2 to contribute timely the ending of the COVID-19 pandemic following the concept of PPPM.

Keywords: Predictive preventive personalized medicine (PPPM/3PM), COVID-19, Autoverification and guidance system (AGS), Reverse transcriptase-polymerase chain reaction (RT-PCR), Manual interpretation, Diagnostic accuracy, Prevention and control, Personalized policy, Dynamic COVID-free policy

Introduction

The dynamic COVID-free policy applied by China

The coronavirus disease 2019 (COVID-19) has become the most widespread global pandemic since it was first reported on December 30, 2019 [1, 2], causing huge health and economic burden [3]. Nowadays, there is a trend to declare the ending of COVID-19 globally, while China is still pursuing the dynamic COVID-free strategy due to high population density and limited medical resources. Based on the susceptibility of COVID-19 [4], wearing a mask and presenting a negative nucleic acid testing result within 48 h are necessary when people enter public places such as airport, railway station and supermarket. Under this strict control and prevention measures, the dynamic COVID-free strategy applied by China contributes to the lowest infection and mortality rates, aligning with the concept of predictive, preventive and personalized medicine (PPPM) [5, 6].

Challenges in preventing COVID-19

The diagnosis of COVID-19 is based on the comprehensive analysis of epidemiological history, clinical manifestation, laboratory and imaging examination [7]. Among them, nucleic acid amplification test (NAAT) of SARS-CoV-2 is critically important [8, 9], and the reverse transcriptase-polymerase chain reaction (RT-PCR) test is the gold standard for the identification of SARS-CoV-2 in the specimens collected from the upper and lower respiratory tracts [10, 11], playing a vital role in the pandemic secondary prevention and control with a PPPM approach [12].

Under the dynamic COVID-free policy in China, the number of daily samples sent to a laboratory at the peak time, a period when a mass nucleic acid screening with a 100% coverage of population for several days are conducted in the residential communities, was over 10 times higher than that of the routine time, a period when only one person entering airport or railway station or person working in the industry with high risk for COVID-19 are required to do the nucleic acid testing. It poses a huge challenge for the quality and efficiency of the interpretation of the RT-PCR results. The gigantic workload is far beyond the capacity of personnel audit and thus commonly causes a delay in data reporting. However, all laboratories globally conducting nucleic acid detection of SARS-CoV-2 rely on manual interpretation of RT-PCR results presently, which is labour intensive and time-consuming and sometime fails to meet the requirements of turnaround times (TATs). Manual interpretation of RT-PCR results also demands accredited laboratorian qualified for the standardized audit procedures after training. Both false negative and false positive reports are the major challenges caused by the inconsistency of individual result’s interpretation. For example, the false negative reports of RT-PCR can lead to a wrong management of the positive cases, thus resulting in the processive spread of the pandemic [13, 14]. Moreover, the lack of standardized guidance for indefinite or suspected positive cases may also lead to extended TATs. Therefore, a paradigm change from reactive medicine to PPPM, and a more prompt and accurate diagnosis of COVID-19 is of great significance in reducing treatment delay and preventing the epidemic from spread.

An optimal autoverification to diagnose COVID-19 in the perspective of PPPM

Autoverification, based on a series of simple and complex algorithms, provides automated actions operated by a computer system to release the COVID-19 RT-PCR results. It has been shown that autoverification can reduce the bias from manual interpretation, improve the efficiency and accuracy of test results and enable laboratory technicians to pay more attention to some suspicious positive samples [15, 16]. Therefore, the popularization and application of autoverification under the personalized mass nucleic acid testing condition may be beneficial to predict the effective reproduction number at time (Rt), a measure reflecting the real-time transmissibility of an epidemic, providing strong evidence to diagnose, predict, prevent and control COVID-19 from the perspective of PPPM.

Working hypothesis

To solve these issues faced by China and to prevent COVID-19 in the framework of PPPM, we developed an autoverification and guidance system (AGS), which is able to examine the preanalytical, analytical and postanalytical data based on a unified standardized criterion. It is hypothesized that the AGS may be a more effective approach with higher accuracy in diagnosis, prediction, prevention and control COVID-19 than that of conventional manual interpretation. Then, we tested and validated the accuracy of AGS in the interpretation of COVID-19 RT-PCR results based on the tests from the Hong Kong and Shanghai cohorts.

Methods

Clinical samples

The data of 266,035 nasopharyngeal samples from Hong Kong and 114,658 oropharyngeal samples from Shanghai during the period from March 28 to April 10, 2022, were collected for the evaluation of the AGS. Hong Kong was characterized by its high positive rate of the infection (12.1%, Table 3), while Shanghai has a low positive rate of the infection (0.01%, Table 3) and high requirement for timeliness but under low quality control environment. All the tests were performed at the Hong Kong Hybribio Medical Laboratory and the Hybribio mobile medical shelter at Shanghai airport. The data were stored in the Laboratory Information System, an information management hub serving for data collection, reporting, transmission and archiving.

Table 3.

Comparison between the AGS and manual auditing in live environments

Hong Kong
(n = 266,035)
Shanghai
(n = 114,658)
Autoverification Manual audit Autoverification Manual audit
Initial RT-PCR results Total number of runs* 3315a 3315a 1547b 1547b
Instrument error flags or warnings 0 0 0 0
Quality controls Pass 255,841 255,841 111,240 111,240
Fail 10,194 10,194 3,418 3,418
Number of autoverified cases 231,066 N/A 109,956 N/A
Autoverification rate (%) 86.9% N/A 95.9% N/A
Number of manual interpretations 34,969 266,035 4,702 114,658
Manual interpretation rate (%) 13.1% 100% 4.1% 100%
Total interpretation time* 26 minc 53 hd 16 mine 20 hf
Final reporting Positive 32,198 12
Negative 232,218 114,035
Indefinite 1,619 611
Positive rate 12.1% 0.01%

N/A, not available

*Total number of runs: the RT-PCR rounds of samples experienced

aTotal 3315 rounds RT-PCR with maximum 90 samples per round were conducted to finish the 226,035 samples testing in Hong Kong

bTotal 1547 rounds RT-PCR with about 75 samples per round were conducted to finish the 114,658 samples testing in Shanghai

*Total interpretation time: the time spending on the interpretation of all samples

cThe time spending on interpreting 266,035 samples by the AGS

dThe time spending on interpreting 266,035 samples by laboratorians

eThe time spending on interpreting 114,658 samples by the AGS

fThe time spending on interpreting 114,658 samples by laboratorians

Design of the AGS

The flow diagram (Fig. 1) of the AGS consists of three stages for each sample testing: (1) pre-interpretation check, (2) result interpretation and guidance and (3) delta check. The design of algorithms is based on the guideline from the Clinical and Laboratory Standards Institute (CLSI) [17, 18]. Once a RT-PCR reaction finishes, the data will be stored automatically in the Laboratory Information System and exported as a new file to AGS. Then, the file is processed through the following steps shown in the flowchart (Fig. 1). The system will generate a disposal warning when a sample invalidates any one of the rules, which are set based on the specimen information and integrity, and then report the exact issues that need to be fixed by manual intervention. If the problems cannot be solved by laboratory personnel intervention, the system will advise for re-sampling and retesting.

Fig. 1.

Fig. 1

Flow diagram of the AGS. The AGS consists of three stages for each sample testing: pre-interpretation check, result interpretation and guidance, delta check. The recorded information of a specimen invalidating one of these rules is asked to complement the missing information and then run through the system again. Only the samples with integral information are processed; otherwise, the samples will be sent back for re-sampling and retesting. The results of the RT-PCR are classified into negative (labelled green), positive (labelled red) and indefinite (labelled yellow) after passing the rules in the pre-interpretation check stage. A negative sample is autoverified by the AGS and reported immediately. While the positive and indefinite cases require further manual interpretation and subsequent guidance. If the sample pass the delta check, the results will be autoverified; otherwise, the system will give a warning

Rules of specimen information and integrity

The recorded information of the subjects includes a unique sample identifier, a unique subject identifier, age and/or date of birth, gender, sample type, sample collection date and location. A specimen invalidating one of these rules is asked to complement the missing information and then run through the system again. Only the samples with integral information in the sealed preservation tubes are processed, otherwise re-sampling will be requested.

Rules of instrument error flags or warnings

Once an instrument error flag warning appears, a manual intervention will be required to identify the cause(s). After the problem being fixed, the samples will run through the entire process again.

Rules of quality control and internal control

Each RT-PCR test includes both negative and positive quality controls. In this study, the passed quality control algorithm and the cycle threshold (Ct) value of the internal control are in compliance with the instructions of COVID-19 RT-PCR detection kit and the policies of Hong Kong and Shanghai Centres for Disease Control [19]. After passing the rules of quality controls, the samples in the RT-PCR running will be analysed in the stage of result interpretation and guidance. The reference Ct value of the internal control for a human pharyngeal swab in this study is set between 15 and 36. Samples with the Ct value of the internal control out of this range will be classified as indefinite cases that require retest.

Rules of result interpretation and guidance

The RT-PCR test targets at two SARS-CoV-2 genomic sequences, the open reading frame1ab (ORF1ab) and gene coding for nucleocapsid protein (N) [19]. The results of the RT-PCR are classified into negative (labelled green), positive (labelled red) and indefinite (labelled yellow) after passing the rules in the pre-interpretation check stage. The whole time of AGS or laboratorians spending on interpreting all samples collected during a specific period was considered as the interpretation time. Due to the gradually decreased fatality rate of COVID-19, a less rigid prevention and control policy is launched in China. Thus, in the present study, the reference Ct value of the positive is set between 10 and 35 for both ORF1ab and N. Assessment of test results for clinical specimens passing the rule of internal control is interpreted according to the following criteria:

(1) A specimen with sigmoidal amplification curves of both ORF1ab and N, both Ct values for ORF1ab and N between 10 and 35, and the discrepancy between the Ct values of the two targeted genes ≤ 3 is regarded as a positive case.

(2) A specimen with no sigmoidal amplification curves of both ORF1ab and N, and Ct values of the two targeted genes denoted as “Undet”, i.e. undetermined, in the software interface is considered as a negative case.

(3) Samples that cannot be classified as neither negative nor positive are defined as indefinite cases, including the specimens with abnormal curve(s) of ORF1ab and/or N, the specimens with the Ct value(s) of ORF1ab and/or N out of the range of 10–35, and the specimens with the discrepancy between the Ct values of the two targeted genes > 3. The possible reasons for indefinite results and the subsequent guidance after initial RT-PCR interpretation are listed in Table 1.

Table 1.

Guidance for positive and indefinite RT-PCR results

Categories after initial RT-PCR Possible reason(s) Re-test results Guidance
Positive N/A Negative Report
Positive Report
Indefinite Ct value of internal control out of the range of 15–36 Fail in sample collection, extraction, reverse transcription or mistake in reagent adding process Out of the range Resampling
In the range Guidance based on the classification of the re-test result
Abnormal curve(s)

Instrument malfunction; interruption during RT-PCR reaction;

interferent(s) in the RT-PCR solution;

RT-PCR tubes not sealed

Abnormal curve(s) Check the conditions of RT-PCR instrument and reagent. Resampling if both are in good function
Normal curves Guidance based on the classification of the re-test result
Ct value(s) of the ORF1ab and/or N out of the range of 10–35 One gene’s Ct value out of the range Low extraction efficiency due to the mismatch among preservation tubes, extraction instrument and solution; low copy number of virus in the sample Negative Report
Positive Report
Same result as the initial RT-PCR Label positive and report
Both genes’ Ct values out of the range Patient in recovery period; participants were vaccinated with the COVID-19 vaccine within 48 h Negative Report
Positive Report
Same result as the initial RT-PCR Clarify whether the participants were vaccinated within 48 h and resampling
The discrepancy between the Ct values of the ORF1ab and N exceed 3 Fragmentary DNA after extraction and reverse transcription; discrepant amplification efficiency between ORF1ab and N genes Negative Report
Positive Report
Same result as the initial RT-PCR Label positive and report

Ct, cycle threshold; N/A, not available; N, gene coding for nucleocapsid protein; ORF1ab, open reading frame1ab

Samples interpreted as negative after initial RT-PCR were reported immediately, while positive and indefinite samples required re-test. During the re-test stage, if the RT-PCR results are inconsistent using more than 3 kinds of kits from different companies, the system will advise resampling. The guidance in this table is based on consistent result in re-test

Rules of delta check

As the last process of autoverification system, the delta check refers to the comparation of a specific and predefined result against a subject’s previous laboratory result(s) [18]. Delta check in AGS includes the comparison with historical results of RT-PCR testing, and the results of further COVID-19 antigen or antibody testing, which are done within 1 week of RT-PCR testing. If the comparison fails to pass the delta check in the AGS algorithm, the system gives a warning, and then a detailed description is added in the final report. The algorithm for the comparison between the RT-PCR result and an antigen test/immunoglobulin M (IgM) test/immunoglobulin G (IgG) test is shown in the Table 2.

Table 2.

Algorithms of comparison between RT-PCR and antigen or antibody tests

RT-PCR Antigen test IgM test IgG test
Positive Negative Positive Negative Positive Negative
Positive Pass Fail Pass Fail Fail Pass
Negative Fail Pass Fail Pass Pass Fail

IgM, immunoglobulin M; IgG, immunoglobulin G

Statistical analysis

Analysis of the data was performed using the SPSS statistics software package, version 26.0 for Microsoft Windows (SPSS Inc., Chicago, IL, USA). The interpretation of the initial RT-PCR screening results was recognized as negative, positive and indefinite. Among them, the negative cases were reported immediately, while the positive and indefinite samples were further retested by senior technicians. In order to assess the performance of the system, the interpretation outcomes were classified into two categories: no retest required (negative) and retest required (positive + indefinite). Considering the manual interpretation for COVID-19 RT-PCR as the gold standard, the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, 95% confidence interval (95%CI) and Cohen’s kappa of AGS were calculated using an online statistical tool [20].

Results

To assess the performance of the AGS, the initial RT-PCR results of 380,693 samples (266,035 from Hong Kong; 114,658 from Shanghai) were processed by the AGS and senior technologists simultaneously (Table 3). Without support of the AGS, all RT-PCR results needed to be read by laboratorians, i.e. a 100% manual interpretation rate. While the AGS was applied, the application of manual interpretation was reduced to 13.1%, and the total interpretation time of all the samples in the initial RT-PCR screening fell from 53 h to 26 min in Hong Kong. In the case of the mobile medical shelter in Shanghai, the manual interpretation rate of all samples was cut down to 4.1%, and the total interpretation time dropped from 20 h to 16 min. These demonstrated the advantages of the AGS in reducing manpower input and TAT.

The interpreted outcomes of the initial RT-PCR tests were classified into positive, negative and indefinite, among which positive and indefinite samples were further retested by senior technicians while the negative cases were reported immediately. Of the 266,035 nasopharyngeal samples in Hong Kong by manual audit, 16,356 (6.15%) were positive, 231,073 (86.86%) were negative, and 18,606 (6.99%) were indefinite; 231,073 (86.86%, negative) did not require to be further retested, and 34,962 (13.14%, positive and indefinite) required to be further retested; of the 114,658 oropharyngeal samples in Shanghai, 76 (0.07%) were positive, 109,956 (95.90%) were negative, and 4,626 (4.03%) were indefinite; 109,956 (95.90%, negative) did not require to be further retested, and 4702 (4.10%, positive and indefinite) required to be further retested. The comparison of results interpretation between the AGS and manual audit is listed in Table 4. The interpretation outcome was defined into “no retest required group” and “retest required group” as shown in Table 5. The AGS showed 99.98% (95%CI, 99.97–99.98%) sensitivity, 99.87% (95%CI, 99.82–99.90%) specificity, 99.98% (95%CI, 99.97–99.99%) PPV, 99.85% (95%CI, 99.80–99.88%) NPV and 99.96% (95%CI, 99.95–99.97%) accuracy, with a 0.998 Cohen’s kappa in Hong Kong; while 100% (95%CI, 100–100%) sensitivity, 100% (95%CI, 99.92–100%) specificity, 100% (95%CI, 99.99–100%) PPV, 100% (95%CI, 99.90–100%) NPV, and 100% (95%CI, 100–100%) accuracy, with a Cohen’s kappa of 1.000 at Shanghai and a P value under 0.001 (Table 6), suggesting a perfect accuracy of the AGS in the real-world application.

Table 4.

Autoverification and manual audit results of the initial RT-PCR screening, classified as positive, negative, and indefinite

Autoverification results of the initial RT-PCR Manual audit results of the initial RT-PCR Total
Positive Negative Indefinite

Hong Kong

(Accuracy: 99.93%)

Positive 16,323 0 59 16,382
Negative 0 231,020 46 231,066
Indefinite 33 53 18,501 18,554
Total 16,356 231,073 18,606 266,035

Shanghai

(Accuracy: 99.98%)

Positive 76 0 23 99
Negative 0 109,956 0 109,956
Indefinite 0 0 4603 4603
Total 76 109,956 4626 114,658

Table 5.

Autoverification and manual audit results of the initial RT-PCR screening classified as no retest required and retest required

Autoverification results of initial RT-PCR Manual audit results of initial RT-PCR Total
No retest requireda Retest requiredb
Hong Kong No retest requireda 231,020 46 231,066
Retest requiredb 53 34,916 34,969
Total 231,073 34,962 266,035
Shanghai No retest requireda 109,956 0 109,956
Retest requiredb 0 4,702 4,702
Total 109,956 4,702 114,658

aNo retest required (negative)

bRetest required (positive + indefinite)

Table 6.

Evaluation the performance of the AGS

Hong Kong Shanghai
Sensitivity 99.98% (95%CI, 99.97–99.98%) 100% (95%CI, 100–100%)
Specificity 99.87% (95%CI, 99.82–99.90%) 100% (95%CI, 99.92–100%)
Youden index 0.9985 1
PPV 99.98% (95%CI, 99.97–99.99%) 100% (95%CI, 99.99–100%)
NPV 99.85% (95%CI, 99.80–99.88%) 100% (95%CI, 99.90–100%)
Accuracy 99.96% (95%CI, 99.95–99.97%) 100% (95%CI, 100–100%)
Cohen’s kappa 0.998 1
P value  < 0.001  < 0.001

PPV, positive predictive value; NPV, negative predictive value; 95%CI, 95% confidence interval

Calculated sensitivity, specificity, Youden index, PPV, NPV, accuracy, Cohen’s kappa, and P value of the AGS implemented in Hong Kong and Shanghai

Discussion

RT-PCR is a critical tool in fighting against SARS-CoV-2, being the gold standard for the identification of the virus in upper and lower respiratory tract specimens [10, 11]. Identifying infected patients and separating them from uninfected people are the most efficient approaches to prevent and control the spread of SARS-CoV-2. Under the condition of mass nucleic acid screening in China, a timely, accurate and consistent RT-PCR test and subsequent accurate interpretation of the results are the basic foundation for infection diagnosis, prevention and management in a real-world setting following the concept of PPPM.

Previous studies showed that autoverification has outstanding performance in processing the multiple types of laboratory data, e.g. thalassemia gene detection [21], arterial blood gas analysis [22] and enzyme-linked immunosorbent assay (ELISA) of hepatitis B virus (HBV) serological markers [23]. The implementation of autoverification exhibits advantages in reducing error rate, saving time, decreasing manpower demand and simplifying the processing of sophisticated manual auditing [18]. In this study, the AGS we created to identify potential analytical errors and replace manual interpretation procedure shows great performance in time-efficient, labour-saving and diagnosis-precisive, being a good example of PPPM practice in pandemic control.

The asymptomatic raised greatly and only a few were transformed to confirmed cases due to the declined virulence of COVID-19, vaccine development and rapid screening [2427]. However, China still adopts strict pandemic prevention and control. It is “a personalized issue” under “a personalized policy” at communities level, following the “dynamic COVID-free strategy” in China. Therefore, the AGS with high accuracy and efficacy is warranted to solve this specific challenge in China. This study is the first attempt to apply and assess the autoverification in Chinese large-scale COVID-19 screenings during the pandemic. The AGS conducted in this study provides a platform to rapidly distinguish negative results from the outcomes that need manual interpretation, being suitable for the dynamic COVID-free policy in China and beneficial to the secondary prevention and management of COVID-19 in PPPM approach. The computer-based immediate verification and report of negative results save a large amount of time on reporting negative cases when it compares to the manual audit mode and thus enable laboratorians to pay more attention to the cases that require retest in terms of personalized accuracy at large communities level screening which is a common challenge in PPPM practice.

The AGS is linked to the Laboratory Information System to ensure the confidentiality of data, and the computer-based autoverification algorithm guarantees excellent accuracy (99.98% in Hong Kong and 100% in Shanghai) by avoiding human errors. In the case of the results that require manual interpretation, subsequent guidance is provided by the AGS to the laboratorians. In addition, the implementation of AGS system is especially important for countries and regions with low resource settings where the expertise for manual interpretation of RT-PCR results is inadequate.

The pre-interpretation check of the AGS is implemented to warn errors during in-lab workflow, like instrument breakdown and contamination of reagents. While the process of delta check is used to further validate the accuracy of AGS by comparing the RT-PCR results with antigen or antibody testing results. Both NAAT and antigen test are intended for the detection of current infection. In contrast to NAAT that detects the existence of viral nucleic acid, the antigen test detects the nucleocapsid protein antigen of the SARS-CoV-2 in nasal swabs, nasopharyngeal swabs or saliva [8, 28]. However, the antigen test always has lower sensitivity and specificity compared to the RT-PCR test, especially in samples with low virus load [10, 11, 29]. An antibody test is intended for the detection of SARS-CoV-2 antibodies (IgG and/or IgM) in blood which can help determine a past infection [2, 8], but it may also generate false positive results for people who have received a SARS-CoV-2 vaccine [30]. Therefore, the failure of samples passing the delta check is mainly due to the samples taken from infected people in latent period and/or due to the low sensitivity of antigen or antibody tests. The diagnostic accuracy of RT-PCR of SARS-CoV-2 is further increased by retesting the samples that cannot pass the algorithm of delta check.

The operation performance of the AGS was tested in two different scenarios: the Hong Kong medical laboratory and the mobile medical shelter at Shanghai airport. The positive rate in Hong Kong was 12.1%, much higher when compared to 0.01% in Shanghai. The mobile medical shelter is usually constructed in place where is not suitable for building a permanent medical laboratory. Therefore, it is a big challenge to provide quality control on data analysis in a mobile shelter where has a high risk of low quality of sample handling and data processing. Under such conditions, the AGS was demonstrated to be a powerful tool by ensuring standardized data process with autoverification. It also showed perfect accuracy of interpretation and had considerable time efficiency to save the labour resource, meeting the testing demanding a PPPM approach in the situation of COVID-19 pandemic.

The twenty-first century witnessed the development of medical services from traditional, complementary and alternative medicine (TCAM), person-centred medicine (PCM), individualized medicine (IM), stratified medicine (SM), personalized medicine (PM) and precision medicine to PPPM [31]. PPPM combines the advantages and minimizes the disadvantages of the existing approaches and has been considered as the “medicine of the future” [5]. It has been successfully applied to non-communicable diseases including cancer [32, 33], cardiovascular diseases [34], ischemic stroke [35] and neurodegenerative diseases [36] and infectious diseases such as COVID-19 [37].

The paradigm changes from the reactive medicine to PPPM has been considered as a considerable transformation [5, 38]. Unlike the traditional manual interpretation mode, the AGS-based novel mode for COVID-19 screening and diagnosis can rapidly distinguish positive cases from mass samples and benefit to the subsequent adjustment and deployment of COVID-19 prevention and control measures under the PPPM framework. The diagnostic accuracy test indicates that the AGS with required precision is an optimal predictive and protective tool for COVID-19 to meet the “Chinese personalized COVID-free policy”. This study provides further strong supporting evidence for the application of PPPM in global COVID-19 epidemic control.

Limitation

A limitation of this study should be noted. The rules of AGS in this study were designed to the application in Hong Kong and Shanghai. When AGS is applied in other environments, whether the setting of rules and parameters can meet the requirements of the local workstation remains to be further verified. The optimal parameters of the reference Ct value ranges for quality controls, positive and negative results would be calibrated and updated based on new information and real-world clinical scenario.

Conclusions and expert recommendations

In conclusion, we created an AGS tool to automatically interpret the COVID-19 RT-PCR results and validated its efficiency in two mega cites, Hong Kong and Shanghai. Compared to conventional manual interpretation, the computer-based AGS not only shows a high accuracy but also considerably relieves both labour and time from the challenge of large-scale screening of SARS-CoV-2. This study supports the paradigm shift from reactive medicine to the proactive model of PPPM in the field of epidemiology of infectious diseases, demonstrating that the PPPM concept and strategy could be integrated into the diagnosis, prevention and intervention of global pandemic [39].

In recommendation, we provide (1) AGS as a PPPM tool which can be widely applied in clinical laboratories in the COVID-19 pandemic screening; (2) AGS would effectively facilitate the prediction of COVID-19 transmission risk by rapidly identifying and responding to the reported positive case at a real-world setting such as at densely populated cites, Hong Kong and Shanghai; (3) it also contributes to protect both the susceptive populations and infected patients with individual level of precision in the practice of PPPM; (4) AGS also has considerable time efficiency and cost effective to save the labour resource and meet the test-demanding in a real-word PPPM practice in the situation like COVID-19 pandemic.

Abbreviations

AGS

Autoverification and guidance system

CLSI

Clinical and Laboratory Standards Institute

COVID-19

Coronavirus disease 2019

Ct

Cycle threshold

ELISA

Enzyme-linked immunosorbent assay

HBV

Hepatitis B virus

IgG

Immunoglobulin G

IgM

Immunoglobulin M

IM

Individualized medicine

N

Nucleocapsid protein

NAAT

Nucleic acid amplification test

NPV

Negative predictive value

ORF1ab

Open reading frame1ab

PCM

Person-centred medicine

PM

Personalized medicine

PPV

Positive predictive value

PPPM

Predictive, preventive and personalized medicine

Rt

Effective reproduction number at time

RT-PCR

Reverse transcriptase-polymerase chain reaction

SM

Stratified medicine

TATs

Turnaround times

TCAM

Traditional, complementary and alternative medicine

WHO

World Health Organization

Author contribution

YC, LX and XT conceived and designed the study. ML and CT drafted the manuscript. SQ, ZL, FX, WL, YZ and AX contributed to data collection. CT applied and obtained permission from the ethics committee and performed the statistical analysis. ML, CT, ZW and XT revised the manuscript. The corresponding authors attest that all listed authors meet authorship criteria.

Funding

This study was supported by Special Major Application Research Project for COVID-19 Prevention and Control in Universities, Department of Education of Guangdong, Provincial Program of Innovation and Strengthening School, Guangdong, China (Grant number: 2020KZDZX1093) and Special Project for COVID-19 Prevention and Treatment of Shantou Science and Technology Bureau, Guangdong, China (Grant number: 2020–1-61).

Data availability

The dataset generated and analysed during the study are available from the corresponding author on reasonable request.

Code availability

Not applicable.

Declarations

Ethics approval

This study was approved by the Human Medical Ethics Committee of the First Affiliated Hospital of Shantou University Medical College (No. B-2022–049), and it was registered in the Chinese Clinical Trial Registry (No. ChiCTR2200059709).

Consent to participate

Owing to the retrospective design of the study, informed consent was waived by the institutional review board, because all identifying details of the participants were removed before computational analysis.

Consent for publication

All authors have approved the manuscript and agreed with submission to EPMA J.

Competing interest

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Yingmu Cai, Mengyu Liu, Zhiyuan Wu, and Cuihong Tian contributed equally.

Contributor Information

Longxu Xie, Email: lxxie@hybribio.cn.

Xuerui Tan, Email: doctortxr@126.com.

References

  • 1.Atzrodt CL, Maknojia I, McCarthy RDP, Oldfield TM, Po J, Ta KTL, et al. A guide to COVID-19: a global pandemic caused by the novel coronavirus SARS-CoV-2. Febs j. 2020;287:3633–3650. doi: 10.1111/febs.15375. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Ita K. Coronavirus Disease (COVID-19): Current status and prospects for drug and vaccine development. Arch Med Res. 2021;52:15–24. doi: 10.1016/j.arcmed.2020.09.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Awwab H, Solorzano JI, Jaisingh KC, Singireddy S, Bailey S, Dominic P. Cardiac pauses in critically ill coronavirus disease-2019 patients. Heart and Mind. 2021;5:4. doi: 10.4103/hm.hm_35_20. [DOI] [Google Scholar]
  • 4.Hou H, Yang H, Liu P, Huang C, Wang M, Li Y, et al. Profile of immunoglobulin G N-glycome in COVID-19 patients: a case-control study. Front Immunol. 2021;12:748566. 10.3389/fimmu.2021.748566. [DOI] [PMC free article] [PubMed]
  • 5.Golubnitschaja O, Baban B, Boniolo G, Wang W, Bubnov R, Kapalla M, et al. Medicine in the early twenty-first century: paradigm and anticipation - EPMA position paper 2016. Epma j. 2016;7:23. doi: 10.1186/s13167-016-0072-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Wu Z, Zhang H, Li Z, Li H, Miao X, Pan H, et al. Mutual effect of homocysteine and uric acid on arterial stiffness and cardiovascular risk in the context of predictive, preventive, and personalized medicine. EPMA Journal. 2022 doi: 10.1007/s13167-022-00298-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Wang Y, Hou H, Wang W, Wang W. Combination of CT and RT-PCR in the screening or diagnosis of COVID-19. J Glob Health. 2020;10:010347. doi: 10.7189/jogh.10.010347. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Goudouris ES. Laboratory diagnosis of COVID-19. J Pediatr (Rio J) 2021;97:7–12. doi: 10.1016/j.jped.2020.08.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Yüce M, Filiztekin E, Özkaya KG. COVID-19 diagnosis -a review of current methods. Biosens Bioelectron. 2021;172:112752. doi: 10.1016/j.bios.2020.112752. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Chaimayo C, Kaewnaphan B, Tanlieng N, Athipanyasilp N, Sirijatuphat R, Chayakulkeeree M, et al. Rapid SARS-CoV-2 antigen detection assay in comparison with real-time RT-PCR assay for laboratory diagnosis of COVID-19 in Thailand. Virol J. 2020;17:177. doi: 10.1186/s12985-020-01452-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Hirotsu Y, Maejima M, Shibusawa M, Nagakubo Y, Hosaka K, Amemiya K, et al. Comparison of automated SARS-CoV-2 antigen test for COVID-19 infection with quantitative RT-PCR using 313 nasopharyngeal swabs, including from seven serially followed patients. Int J Infect Dis. 2020;99:397–402. doi: 10.1016/j.ijid.2020.08.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DK, Detection of, , et al. novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill. 2019;2020:25. doi: 10.2807/1560-7917.Es.2020.25.3.2000045. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Dramé M, TabueTeguo M, Proye E, Hequet F, Hentzien M, Kanagaratnam L, et al. Should RT-PCR be considered a gold standard in the diagnosis of COVID-19? J Med Virol. 2020;92:2312–2313. doi: 10.1002/jmv.25996. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Tahamtan A, Ardebili A. Real-time RT-PCR in COVID-19 detection: issues affecting the results. Expert Rev Mol Diagn. 2020;20:453–454. doi: 10.1080/14737159.2020.1757437. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Torke N, Boral L, Nguyen T, Perri A, Chakrin A. Process improvement and operational efficiency through test result autoverification. Clin Chem. 2005;51:2406–2408. doi: 10.1373/clinchem.2005.054395. [DOI] [PubMed] [Google Scholar]
  • 16.Li J, Cheng B, Ouyang H, Xiao T, Hu J, Cai Y. Designing and evaluating autoverification rules for thyroid function profiles and sex hormone tests. Ann Clin Biochem. 2018;55:254–263. doi: 10.1177/0004563217712291. [DOI] [PubMed] [Google Scholar]
  • 17.Clinical and Laboratory Standards Institute (CLSI). Autoverification of medical laboratory results for specific disciplines. 1st ed. CLSI guideline AUTO15 (ISBN 978-1-68440-056-0 [Print]; ISBN 978-1-68440-057-7 [Electronic]). Wayne: Clinical and Laboratory Standards Institute. 2019. https://clsi.org/media/3345/auto15_sample.pdf.
  • 18.Clinical and Laboratory Standards Institute (CLSI). Autoverification of clinical laboratory test results; Approved Guideline. CLSI Document AUTO10-A [ISBN 1–56238–620–4]. Wayne: Clinical and Laboratory Standards Institute. 2006. https://clsi.org/media/1342/auto10a_sample.pdf.
  • 19.National Health Commission of the People’s Republic of China. Protocol for Prevention and Control of COVID-19. 6th ed. China CDC Weekly. 2020;2:321–26. 10.46234/ccdcw2020.082. [DOI] [PMC free article] [PubMed]
  • 20.MedCalc Software Ltd. Diagnostic test evaluation calculator (Version 20.2). https://www.medcalc.org/calc/diagnostic_test.php. Accessed 13 Dec 2022.
  • 21.Lin X, Cheng B, Cai Y, Jiao X, Yang X, Zhang Q, et al. Establishing and evaluating an auto-verification system of thalassemia gene detection results. Ann Hematol. 2019;98:1835–1844. doi: 10.1007/s00277-019-03656-4. [DOI] [PubMed] [Google Scholar]
  • 22.Wu J, Pan M, Ouyang H, Yang Z, Zhang Q, Cai Y. Establishing and evaluating autoverification rules with intelligent guidelines for arterial blood gas analysis in a clinical laboratory. SLAS Technol. 2018;23:631–640. doi: 10.1177/2472630318775311. [DOI] [PubMed] [Google Scholar]
  • 23.Li J, Cheng B, Yang L, Zhao Y, Pan M, Zheng G, et al. Development and implementation of autoverification rules for ELISA results of HBV serological markers. J Lab Autom. 2016;21:642–651. doi: 10.1177/2211068215601612. [DOI] [PubMed] [Google Scholar]
  • 24.Murray CJL. COVID-19 will continue but the end of the pandemic is near. Lancet. 2022;399:417–419. doi: 10.1016/s0140-6736(22)00100-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Tan X, Wang W. Do not forget glycomics and theranostics for COVID-19 drug and vaccine development. OMICS. 2022;26:356–357. doi: 10.1089/omi.2022.0052. [DOI] [PubMed] [Google Scholar]
  • 26.Wang W. Glycomedicine: the current state of the art. Engineering. 2022 doi: 10.1016/j.eng.2022.03.009. [DOI] [Google Scholar]
  • 27.Wang X, Zhong Z, Wang W. COVID-19 and preparing planetary health for future ecological crises: hopes from glycomics for vaccine innovation. OMICS. 2021;25:234–241. doi: 10.1089/omi.2021.0011. [DOI] [PubMed] [Google Scholar]
  • 28.Centers for Disease Control and Prevention (CDC). Guidance for Antigen testing for SARS-CoV-2 for healthcare providers testing individuals in the community. 2022. https://www.cdc.gov/coronavirus/2019-ncov/lab/resources/antigen-tests-guidelines.html.
  • 29.Nagura-Ikeda M, Imai K, Tabata S, Miyoshi K, Murahara N, Mizuno T, et al. Clinical evaluation of self-collected saliva by quantitative reverse transcription-PCR (RT-qPCR), direct RT-qPCR, reverse transcription-loop-mediated isothermal amplification, and a rapid antigen test to diagnose COVID-19. J Clin Microbiol. 2020;58:e01438–20. 10.1128/jcm.01438-20. [DOI] [PMC free article] [PubMed]
  • 30.West R, Kobokovich A, Connell N, Gronvall GK. COVID-19 antibody tests: a valuable public health tool with limited relevance to individuals. Trends Microbiol. 2021;29:214–223. doi: 10.1016/j.tim.2020.11.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Wang W, Yan Y, Guo Z, Hou H, Garcia M, Tan X, et al. All around suboptimal health — a joint position paper of the Suboptimal Health Study Consortium and European Association for Predictive. Preventive and Personalised Medicine EPMA J. 2021 doi: 10.1007/s13167-021-00253-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Cox TR, Rumney RMH, Schoof EM, Perryman L, Høye AM, Agrawal A, et al. The hypoxic cancer secretome induces pre-metastatic bone lesions through lysyl oxidase. Nature. 2015;522:106–110. doi: 10.1038/nature14492. [DOI] [PMC free article] [PubMed] [Google Scholar] [Retracted]
  • 33.Grech G, Zhan X, Yoo BC, Bubnov R, Hagan S, Danesi R, et al. EPMA position paper in cancer: current overview and future perspectives. Epma j. 2015;6:9. doi: 10.1186/s13167-015-0030-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Helms TM, Duong G, Zippel-Schultz B, Tilz RR, Kuck KH, Karle CA. Prediction and personalised treatment of atrial fibrillation-stroke prevention: consolidated position paper of CVD professionals. Epma j. 2014;5:15. doi: 10.1186/1878-5085-5-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Zheng Y, Guo Z, Zhang Y, Shang J, Yu L, Fu P, et al. Rapid triage for ischemic stroke: a machine learning-driven approach in the context of predictive, preventive and personalised medicine. Epma j. 2022;13:285–298. doi: 10.1007/s13167-022-00283-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Sinnecker T, Kuchling J, Dusek P, Dörr J, Niendorf T, Paul F, et al. Ultrahigh field MRI in clinical neuroimmunology: a potential contribution to improved diagnostics and personalised disease management. Epma j. 2015;6:16. doi: 10.1186/s13167-015-0038-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Chaari L, Golubnitschaja O. Covid-19 pandemic by the “real-time” monitoring: the Tunisian case and lessons for global epidemics in the context of 3PM strategies. Epma j. 2020;11:133–138. doi: 10.1007/s13167-020-00207-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Golubnitschaja O, Kinkorova J, Costigliola V. Predictive, preventive and personalised medicine as the hardcore of ‘Horizon 2020’: EPMA position paper. Epma j. 2014;5:6. doi: 10.1186/1878-5085-5-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Tragni V, Preziusi F, Laera L, Onofrio A, Mercurio I, Todisco S, et al. Modeling SARS-CoV-2 spike/ACE2 protein-protein interactions for predicting the binding affinity of new spike variants for ACE2, and novel ACE2 structurally related human protein targets, for COVID-19 handling in the 3PM context. Epma j. 2022;13:149–175. doi: 10.1007/s13167-021-00267-w. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The dataset generated and analysed during the study are available from the corresponding author on reasonable request.

Not applicable.