Abstract
Aims
Air entrapment (AE) has been reported as a potential cause of early inappropriate shocks (ISs) following subcutaneous implantable cardioverter defibrillator (S-ICD) implantation, but a cause–effect relationship is not always evident. This systematic review aims to analyse this phenomenon concerning implantation techniques, electrogram (EGM) features, radiologic findings, and patient management.
Methods and results
A systematic search was conducted using PubMed, Embase, and Google Scholar databases following the PRISMA guidelines to obtain all available literature data since 2010 on S-ICD malfunctions possibly due to AE. The final analysis included 54 patients with AE as a potential cause of S-ICD malfunction. Overall, the aggregate incidence of this condition was 1.2%. Of ICD malfunctions possibly due to AE, 93% were ISs, and 95% were recorded within the first week following implantation. Radiologic diagnosis of AE was confirmed in 28% of the entire study cohort and in 68% of patients in whom this diagnostic examination was reported. At the time of device malfunction, EGMs showed artefacts, baseline drift, and QRS voltage reduction in 95, 76, and 67% of episodes, respectively. Management included ICD reprogramming or testing, no action (observation), and invasive implant revision in 57, 33, and 10% of patients, respectively. No recurrences occurred during follow-up, irrespective of management performed.
Conclusions
Device malfunction possibly due to AE may occur in ∼1% of S-ICD recipients. Diagnosis is strongly suggested by early occurrence, characteristic EGM features, and radiologic findings. Non-invasive management, principally device reprogramming, appears to be effective in most patients.
Keywords: S-ICD, Inappropriate shock, Air entrapment, Oversensing, Systematic review
What’s new?
Device malfunctions possibly due to air entrapment may occur in ∼ 1% of subcutaneous implantable cardioverter defibrillator patients.
They are mostly inappropriate shocks in the early days after implantation, but defibrillation abnormalities may be observed.
The diagnosis is mainly based on early occurrence, typical electrogram features, and radiological findings.
Non-invasive management, principally device reprogramming or observation, is effective in most patients.
Further studies are required to establish this phenomenon’s actual incidence and clinical impact.
Introduction
The subcutaneous implantable cardioverter defibrillator (S-ICD) has been introduced in clinical practice to prevent sudden cardiac death with approved safety and efficacy profiles.1 Similarly with transvenous ICDs, inappropriate shocks (ISs) represent a major concern regarding S-ICD therapy, negatively impacting morbidity and mortality.2,3
While the leading cause of ISs with the transvenous ICD is supraventricular arrhythmias (e.g. atrial fibrillation), they are mostly related to myopotentials or T-wave oversensing in S-ICD patients.3,4 Following the expansion of S-ICD in clinical practice, uncommon causes of ISs, such as air entrapment (AE) in the subcutaneous space, have been observed. Since its first description by Zipse et al.5 in 2014, various S-ICD malfunctions possibly due to AE have been reported early after implantation. Nonetheless, a direct cause–effect relationship between AE and S-ICD malfunction is not always evident. Moreover, a systematic collective review of this phenomenon is lacking. This study aims to collect and analyse the available literature data concerning clinical details, implantation techniques, electrogram (EGM) features, radiologic findings, and management of S-ICD malfunctions possibly due to AE.
Methods
The study was performed in accordance with guidelines from the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA).6 Considering the elusive nature of the analysed condition and the difficulty to prove a cause–effect relationship between AE and many of the reported S-ICD malfunctions, we referred to these cases throughout the entire text as possibly due to AE. PubMed, Embase, and Google Scholar databases were searched for articles published in the period from 1 January 2010 to 31 August 2021 containing the combination of ‘air’ and ‘subcutaneous’ in the Full Text and any of the following terms in the Title: S-ICD, ICD, defibrillator, cardioverter, or ‘inappropriate shock’. Subsequently, any published data, including clinical studies, registries, case reports, abstracts, or short communications, were considered eligible for the study analysis when satisfying all of the following criteria: (i) reported S-ICD malfunction or IS possibly due to AE or air noise; (ii) published in a peer-reviewed journal; and (iii) available in the English language. Further analysis of the biography of qualified articles was also performed to include additional potentially eligible cases. Every effort was made to avoid duplicated or overlapping data, focusing on the authors/centre, patient clinical data, and EGM recordings when available. Finally, two unpublished cases (authors’ data) of S-ICD malfunctions possibly due to AE were also included. Literature screening, data selection/extraction, and EGM analysis were validated by two cardiac electrophysiologists (H.A. and P.L.). Whenever occurred, disagreement was resolved by a consensus or a third judgement (R.C.).
Case reports were analysed with a focus on the following data, when available: clinical setting, S-ICD implantation technique and programming, type and timing of S-ICD malfunction, the assumed localization of AE, EGM features, chest X-ray (CXR) findings, clinical management, and follow-up. Conversely, studies including a cohort of S-ICD patients with less detailed information were mainly analysed to assess this phenomenon’s incidence when at least a short-term follow-up (≥3 months) was available.
Regarding EGM analysis, the following definitions were applied:
Artefacts: ≥2 abrupt, sharp (spike-like), or rounded deflections from the baseline with subsequent oversensing (non-cardiac signals).
Baseline drift: gradual deviation (wandering) of the isoelectric baseline for ≥1 large box amplitude (at the recording gain) and lasting ≥1 s.
Reduced QRS voltage: mute (standstill) line, hardly visible intrinsic QRS complexes at the recording gain, and/or ≥30% reduction of QRS voltage (compared with peri-shock EGMs, when available).
Finally, the potential corrective effect of shock on EGM abnormalities was also analysed.
Results
Initially, 208 records were identified through database search, of which 34 were eligible for the current analysis5,7–39 reporting 52 patients. After adding two unpublished cases from the authors’ experience, a total of 54 cases of S-ICD malfunction possibly due to AE were included. Figure 1 shows a PRISMA flow diagram of database search, whereas Supplementary material online, Table S1 summarizes the clinical and technical data of the analysis cohort.
Figure 1.
PRISMA flow diagram of database search. PRISMA flow diagram of database search and data extraction of the current review.
Incidence
Fourteen studies,9,11,5–7,20,21,26,1–5,39 with at least a short-term follow-up (≥3 months), reported the incidence of this phenomenon in different S-ICD cohorts, ranging from 0.4% in the Austrian registry15 to 8.3% in a small Polish cohort.20 Aggregating data from these 14 studies, 31 of 2433 (1.2%) S-ICD recipients experienced device malfunction possibly due to AE during the reported follow-up.
Implantation technique, defibrillation threshold test, and device programming
Implantation technique, described in 18 patients, was performed using three and two incisions in 6 (33.3%) and 12 patients (66.6%), respectively. Defibrillation threshold test (DFT), reported in 19 patients, was unremarkable in 15 patients (78.9%). There were DFT-related issues in the remaining four patients (21%), such as ventricular fibrillation (VF) detection delay or inhibition, and unsuccessful DFT requiring pocket revision for suspected AE.16,24,25 In another patient, DFT had to be performed 1 week after S-ICD generator replacement due to unsatisfactory high impedance of the 10J shock test during the procedure.37 Sensing mode, specified in 22 patients, was programmed to the primary, secondary, and alternate vector in 7 (31.8%), 11 (50%), and 4 patients (18.1%), respectively. A conditional shock zone was activated in all 11 patients (100%) in whom the detection cut-off was mentioned.
Type and timing of air entrapment-related subcutaneous implantable cardioverter defibrillator malfunction
Of 54 patients, 50 (92.5%) had at least one IS, whereas four (7.4%) patients had DFT-related issues, manifested as inadequate VF detection in two,24,25 unsuccessful DFT in one,16 and high impedance of the shock test at device replacement in one (Figure 2).37 Event timing was available in 39 patients (72.2%). The undesired event occurred within 24, 72 h, and 1 week following S-ICD implantation in 24 (61.5%), 32 (82%), and 37 patients (94.8%), respectively (Figure 2). No AE-related ISs were observed beyond the 11th day following the procedure.38
Figure 2.
Type and timing of S-ICD malfunctions. A diagram showing the type and temporal distribution of S-ICD malfunctions possibly due to air entrapment. AE, air entrapment.
Radiologic findings and air entrapment localization
Of 22 patients in whom radiographic (CXR) or fluoroscopic data were available, AE was documented in 15 patients (68.1%), accounting for 27.7% of the entire study cohort (see Supplementary material online, Figure S1). Radiologic localization of AE, reported in 14 patients, was at the level of the distal electrode, proximal electrode, and device pocket in two (14.2%), seven (50%), and five patients (35.7%), respectively. Radiologic follow-up in patients with early AE was available in nine patients showing AE absorption within 2 weeks in eight patients (88.8%). In the remaining patients, complete AE resolution was documented at 1-month CXR control.5 Lateral CXR detected AE around the lead electrodes in eight patients, whereas anteroposterior CXR documented AE around the device in five patients.
Localization of AE was confirmed (radiologically) or presumed by analysing EGM features in different sensing vectors in 33 patients. The air entrapment site was adjacent to the proximal electrode, distal electrode, device pocket, and lead connector (set screw) in 10 (30.3%), 7 (21.2%), 7 (21.2%), and 9 patients (27.2%), respectively.
Electrogram features at the time of malfunction
The EGM showing S-ICD malfunction was available in 21 patients (38.8%) (see Supplementary material online, Figure S1). Artefacts were present in 20 patients (95.2%). In all four patients with presumed AE in the set screw who had available EGMs, artefacts were monomorphic and repetitive, simulating ventricular tachyarrhythmias.13 Baseline drift and reduced QRS voltage were present in 16 (76.1%) and 14 patients (66.6%), respectively. Provocative manoeuvres described in 14 patients could replicate the EGM abnormalities at the time of device interrogation only in six patients (42.8%). Following shock delivery, the EGM was nearly normalized in nine patients (42.8%).
Clinical management and follow-up
Clinical management of early S-ICD malfunction was described in 30 patients (55.5%) (see Supplementary material online, Figure S2). Due to delayed evaluation of a self-limiting phenomenon or inability of ICD reprogramming, no specific action was adopted in 10 patients (33.3%). Implantable cardioverter defibrillator reprogramming or testing was adopted in 17 patients (56.6%), including reprogramming the sensing vector to exclude the involved AE site, temporary ICD deactivation, and repeated DFT in 13 (43.3%), 5 (16.6%), and 2 patients (6.6%), respectively. After radiologic resolution of AE around the generator, DFT was repeated in one patient 2 days later following an IS associated with high shock impedance (151 ohms).28 In another patient, DFT was successfully performed 1 week after S-ICD generator replacement due to unsatisfactory high impedance of the 10J shock test at the time of procedure.37 Invasive implant revision was performed in three patients (10%), including lead repositioning, pocket revision, and unnecessary pocket inspection for suspected lead connection problems in the set screw. Follow-up data of these patients, with variable or undetermined durations, were reported in 22 patients (73.3%), with at least short-term follow-up (≥3 months) available in 10 patients (33.3%). Irrespective of the adopted management, no recurrences of S-ICD malfunction occurred during the reported follow-up.
Discussion
While uncommon, malfunction of implantable cardiac devices due to AE is not a novel observation. About four decades ago, unipolar pacemaker malfunction as sensing failure or loss of capture was reported to be secondary to AE.40,41 Air entrapment in the set screw (lead connector) of the transvenous ICD may also lead to inappropriate sensing.42 Likewise, in the early days after implantable loop recorder placement, intermittent loss of device-tissue contact and AE may cause similar EGM abnormalities with inappropriate sensing and false asystole.43 Notably, the anatomical nature of the S-ICD system, with sensing electrodes located in the subcutaneous space or within surgical pockets, makes the sensing process prone to undesired interference with AE during the acute phase after implantation. Although uncommon and self-limiting, early ISs related to AE occurring just within a few hours after the procedure or hospital discharge may have a negative impact on acceptance of device therapy and future compliance of affected patients.
Incidence
The actual incidence of S-ICD malfunction due to AE is unknown and accurate estimation is challenging. In this review, analysing only studies reporting S-ICD malfunction possibly due to AE, overestimation of its incidence cannot be excluded. In our analysis, aggregate data from different S-ICD cohorts revealed an incidence of 1.2% although several large S-ICD studies and registries4–6 were not included since they did not refer directly to AE as a potential cause of S-ICD malfunction. However, AE might play a role in the genesis of non-cardiac oversensing associated with ISs that have been reported with a variable incidence: 1.6% in the S-ICD post-approval study,44 1.4% in the UNTOUCHED trial,45 and 2.2% in the EFFORTLESS registry.46
Moreover, a considerable proportion of ISs in the early phase following S-ICD implantation may be related to AE. In the Maude registry, AE in S-ICD patients was the cause of 8% of reported oversensing episodes.47 Likewise, in a systematic review of the literature, Santomauro et al.48 analysed the aetiology of extracardiac ISs in 2654 S-ICD patients in whom AE accounted for 23% of ISs related to oversensing of non-cardiac signals.
Implantation technique, defibrillation threshold test, and device programming
Reported S-ICD malfunctions occurred with both two- and three-incision techniques. However, none of the patients implanted with the three-incision technique had confirmed AE around the distal electrodes, whereas AE at this site was confirmed radiologically in two patients who underwent S-ICD implantation using the three-incision technique. Accordingly, the two-incision technique might reduce the risk of AE around the distal electrode by eliminating the superior parasternal surgical pocket. However, some authors suggest that this technique may expose the lead tip to an increased dislocation risk and possible formation of adjacent AE.10
Notably, the incidence of S-ICD malfunction possibly due to AE was highly variable between studies irrespective of their time frame. This is likely due to differences in sample size, patients’ characteristics, surgical technique, and the centre/operator experience. However, some author groups17,33 reported avoiding this complication in subsequent patients after optimizing their surgical implantation technique. For instance, saline flushing and skin massage of the xiphoid and device pockets before surgical closure should reduce the formation of AE.
Although activation of a conditional shock zone has become a common practice to reduce ISs in S-ICD patients,46 it did not seem to reduce ISs related to AE in our study. This is not surprising, as EGM components during AE oversensing do not match the sinus rhythm template, and the sensed events intervals often occur within the VF zone where there is no role for the discriminative algorithms. Likewise, DFT was unremarkable in most patients, highlighting its poor predictive value regarding subsequent AE-related issues. Finally, and due to the features of related EGM abnormalities, the SMART Pass filter, mainly developed to reduce T-wave oversensing, is not expected to reduce IS possibly related to AE.
Type and timing of subcutaneous implantable cardioverter defibrillator malfunctions
Most malfunctions were ISs, with >90% of patients experiencing this complication. However, only a few patients suffered multiple (>2) ISs or inappropriate ICD storms,5,30,34,36 which is not uncommon in patients with ISs due to lead fracture or failure. The transient nature of AE and the potential corrective effect of shock delivery on related EGM abnormalities might explain this finding.
Moreover, in a minority of patients (7.4%), DFT issues related to AE were described. Air entrapment should be suspected during DFT when there is inadequate VF detection or high shock impedance despite optimal S-ICD positioning. Careful EGM analysis and fluoroscopic check may reveal AE as a potential cause and guide management.
The early occurrence represents the most characteristic feature of S-ICD malfunctions possibly due to AE. In the current analysis, it occurred within the first week following S-ICD implantation in most patients (∼95%) with no ISs reported beyond the 11th day after implantation. Therefore, AE should be considered as a potential cause of ISs during the early days after implantation, whereas it is unlikely to be the underlying mechanism when occurring beyond the first week. This is consistent with the physiologic absorption and resolution of residual air in the subcutaneous space. Interestingly, in a small study evaluating the amount and absorption rate of AE in S-ICD patients using CT imaging, the mean AE immediately after implantation was ∼28 mL with an absorption rate of 99% at 1 week after the procedure.49 Of note, in the patient with reported ISs 11 days after implantation, the readmission CXR showed 90° rotation of pulse generator, and thus other mechanisms cannot be excluded.38
Radiologic findings and the cause–effect relationship between air entrapment and device malfunction
Radiologic confirmation of AE was reported only in 27.7% of patients in our cohort and could not be detected in >30% of patients in whom radiologic inspection was reported. This might be due in part to the transient and dynamic nature of AE in the subcutaneous space, but it raises the question about the actual cause–effect relationship between AE and early S-ICD malfunction. Air entrapment after S-ICD implantation is probably more frequent than expected and might represent a bystander condition at the time of early ISs or abnormal DFT. Of note, AE in the set screw (lead connector) cannot be detected radiologically, and the diagnosis was based on the early occurrence, EGM analysis, and the subsequent reversal of the condition during the following days, consistent with AE absorption. Although the mechanism of these cases remains unclear, incorrect insertion of the wrench tool or the lead tip may play a role, as has been reported with transvenous ICD.42 Interestingly, all available EGMs of presumed AE in the set screw showed repetitive monomorphic artefacts with subsequent EGM normalization following shock delivery, simulating appropriate therapy. However, mechanical damage to the seal plug of the pulse generator might produce similar effects.50 Other mechanisms, such as muscular contractions, movement artefacts in the early days after the procedure caused by unstable electrodes-tissue contact, cannot be fully excluded. In patients with previous cardiac surgery, electrical interference with sternal wires might cause ISs.20 Morani et al.51 described an S-ICD patient who suffered from IS due to an unusual tunnelling course beneath the sixth rib, and intermittent mechanical stress on the lead was hypothesized to be the cause of repetitive monomorphic artefacts analogous to those observed in patients with AE in the set screw.
Air entrapment resolved in all patients with confirmed AE and available radiologic follow-up, showing an absorption rate of about 90% within 2 weeks, which corroborates the early and transient nature of this phenomenon.49 Notably, lateral CXR was most helpful in detecting AE around the lead electrodes, whereas the anteroposterior projection helped to detect AE around the device, highlighting the importance of obtaining orthogonal CXRs after implantation.
Electrogram features
Despite provocative manoeuvres, EGM abnormalities were reproducible only in 40% of patients at the time of device interrogation, adding further diagnostic difficulties and reflecting the potential intermittent and transient effects of AE on sensing process. Following shock delivery, the EGM nearly normalized in about 40% of cases, likely due to AE redistribution following shock, which may reduce the risk of subsequent inappropriate storms.28 Oversensing related to AE is mainly due to artefacts sensing and/or auto-gain function following QRS voltage reduction. During episodes, the most common findings were artefacts (95.2%), followed by baseline drift and reduced QRS voltage in 76.1 and 66.6%, respectively. Analysing EGMs of patients with presumed AE in the lead connector showed repetitive monomorphic artefacts in all cases with subsequent EGM normalization following S-ICD shock, simulating an appropriate therapy for ventricular tachyarrhythmia.13Figures 3 and 4 show representative EGMs of ISs related to AE around the electrodes and the set screw, respectively.
Figure 3.
Inappropriate S-ICD shock due to air entrapment around sensing electrodes. Representative EGM of an inappropriate shock occurring 24 h after S-ICD implantation due to air entrapment around the proximal electrode. Typical EGM features include artefacts, baseline drift, and QRS voltage reduction.
Figure 4.
Inappropriate S-ICD shock due to air entrapment in the set screw. Representative EGM of an inappropriate shock occurring 72 h after S-ICD implantation likely related to air entrapment in the set screw. After initial baseline drift, typical EGM features show repetitive monomorphic artefacts and post-shock EGM normalization simulating an appropriate therapy. Arrows indicate intrinsic QRS complexes.
Management and preventive measurements
Reprogramming the sensing vector to exclude the potential AE was effective in more than 40% of patients. This programming flexibility is meaningful to overcome sensing issues compared with conventional ICDs. Moreover, in one-third of patients, no specific action was performed, and a ‘wait and see’ strategy was adopted probably due to delay in evaluation or inability of ICD reprogramming. Implantable cardioverter defibrillator therapy was temporarily deactivated in 16.6% of patients to avoid further ISs. Notably, 10% of patients underwent invasive implant revision (pocket revision, lead repositioning) due to AE-related malfunction. When AE is suspected, it is advisable to avoid invasive management since this condition is typically self-limiting and reprogramming the sensing vector, temporal ICD deactivation, and/or a ‘wait and see’ strategy are usually efficient measures. Remarkably, no recurrences of S-ICD malfunction occurred during the reported follow-up, irrespective of the adopted management.
Many authors,12 and Boston Scientific manufacturer, recommend optimal surgical techniques to reduce the risk of AE during S-ICD implantation. Adequate electrodes fixation to minimize the dead space and improve electrode-tissue contact, saline flushing of the pockets and sternal track, avoiding blunt digital dissection, and gentle skin massage over the tunnelled track, the xiphoid and device pockets to expel any residual air out before surgical closure are advisable manoeuvres and may minimize the risk of AE formation during the procedure. Moreover, to avoid AE in the header or mechanical damage to the seal plug, it is recommended both leaving the torque wrench in place when inserting the lead, ensuring that the electrode tip is fully inserted in the header, and also inserting the torque wrench at a 90° angle into the set screw with the S-ICD generator held flat. After implantation, careful device interrogation, analysing sensing vectors during provocative manoeuvres, and reviewing orthogonal CXRs help to recognize and localize potential AE and guide judicious management. Finally, immediately activated after implantation, the remote monitoring may allow for early detection of AE-related abnormalities. Figure 5 provides a proposed approach for prevention, early detection, differential diagnosis, and management of AE in S-ICD patients.
Figure 5.
Prevention and management of S-ICD malfunctions possibly due to air entrapment. A proposed approach for prevention, detection, differential diagnosis, and management of AE and possible related malfunctions in S-ICD patients. AE, air entrapment; CXR, chest X-ray; DE, distal electrode; DFT, defibrillation threshold test; EGM, electrogram(s); IS, inappropriate shock(s); PE, proximal electrode; PG, pulse generator; S-ICD, subcutaneous ICD; SS, set screw; VF, ventricular fibrillation; VT, ventricular tachycardia.
Study limitations
First, and due to the retrospective nature of data collection, clinical and technical details were incomplete in many patients. Cohort studies, for instance, with no detailed information, were mainly included to assess the approximate incidence of this condition. Secondly, analysing only studies reporting AE as a potential cause of S-ICD malfunction, overestimation of its incidence cannot be excluded; however, the included cohort studies represent various S-ICD populations, with a considerable aggregate number of patients treated in different centres. Thirdly, the accuracy of EGM analysis might be suboptimal due to the lack of complete EGM recordings and the use of arbitrary criteria to define EGM abnormalities. Fourthly, no specific statistical analysis could be performed considering the non-solid nature of data derived from case reports, and thus the results were simply expressed in percentile (%) when applicable. Accordingly, this review aims to highlight some clinical and technical aspects of this condition rather than to derive definitive conclusions.
Conclusions
Whilst a cause–effect relationship is not always evident, AE may be associated with early S-ICD malfunction, mostly ISs occurring within the first days after implantation. Diagnosis is mainly based on early occurrence, characteristic EGM features, and radiologic findings. Non-invasive management, including device reprogramming or observation, appears to be effective in most patients. Prompt recognition may help to avoid further ISs and unnecessary invasive implant revisions. Further studies are required to establish the actual incidence and clinical impact of this phenomenon.
Supplementary material
Supplementary material is available at Europace online.
Supplementary Material
Funding
We did not have any fundings for this study.
Contributor Information
Hussam Ali, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Pierpaolo Lupo, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Sara Foresti, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Guido De Ambroggi, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Carmine De Lucia, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Diego Penela, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Dario Turturiello, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Edoardo Maria Paganini, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Riccardo Cappato, Arrhythmia & Electrophysiology Center, IRCCS—MultiMedica Group, Via Milanese 300, 20099 Sesto San Giovanni, Milan, Italy.
Data Availability
The data underlying this article will be shared on a reasonable request to the corresponding author.
References
- 1. Adduci C, Ali H, Francia P, Mantovani R, Palano F, Lupo Pet al. The subcutaneous implantable cardioverter-defibrillator: current trends in clinical practice between guidelines and technology progress. Eur J Intern Med 2019;65:6–11. [DOI] [PubMed] [Google Scholar]
- 2. Daubert JP, Zareba W, Cannom DS, McNitt S, Rosero SZ, Wang Pet al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol 2008;51:1357–65. [DOI] [PubMed] [Google Scholar]
- 3. van Rees JB, Borleffs CJ, de Bie MK, Stijnen T, van Erven L, Bax JJet al. Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. J Am Coll Cardiol 2011;57:556–62. [DOI] [PubMed] [Google Scholar]
- 4. Olde Nordkamp LRA, Brouwer TF, Barr C, Theuns DAMJ, Boersma LVA, Johansen JBet al. Inappropriate shocks in the subcutaneous ICD: incidence, predictors and management. Int J Cardiol 2015;195:126–33. [DOI] [PubMed] [Google Scholar]
- 5. Zipse MM, Sauer WH, Varosy PD, Aleong RG, Nguyen DT. Inappropriate shocks due to subcutaneous air in a patient with a subcutaneous cardiac defibrillator. Circ Arrhythm Electrophysiol 2014;7:768–70. [DOI] [PubMed] [Google Scholar]
- 6. Moher D, Liberati A, Tetzlaff J, Altman DG; PRISMA Group . Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. Ann Intern Med 2009;151:264–9. [DOI] [PubMed] [Google Scholar]
- 7. Gamble JH, Grogono J, Rajappan K, Betts TR, Bashir Y, Khiani R. Letter by Gamble, et al regarding article, “inappropriate shocks due to subcutaneous air in a patient with a subcutaneous cardiac defibrillator”. Circ Arrhythm Electrophysiol 2014;7:1281. [DOI] [PubMed] [Google Scholar]
- 8. Yap SC, Bhagwandien RE, Szili-Torok T, Theuns DAMJ. Air entrapment causing early inappropriate shocks in a patient with a subcutaneous cardioverter-defibrillator. HeartRhythm Case Rep 2015;1:156–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Koman E, Gupta A, Subzposh F, Saltzman H, Kutalek SP. Outcomes of subcutaneous implantable cardioverter-defibrillator implantation in patients on hemodialysis. J Interv Card Electrophysiol 2016;45:219–23. [DOI] [PubMed] [Google Scholar]
- 10. Chinitz JS, Nadraus P, Darge A, Cacciabaudo J, Altman EJ. Inappropriate shocks within 24 hours after implantation of a subcutaneous defibrillator with a two-incision technique. J Innov Card Rhythm Manag 2016;7:2295–8. [Google Scholar]
- 11. Moore JP, Mondésert B, Lloyd MS, Cook SC, Zaidi AN, Pass RHet al. Clinical experience with the subcutaneous implantable cardioverter-defibrillator in adults with congenital heart disease. Circ Arrhythm Electrophysiol 2016;9:e004338. [DOI] [PubMed] [Google Scholar]
- 12. Lee S, Souvaliotis N, Mehta D, Suri R. Inappropriate shock in a subcutaneous cardiac defibrillator due to residual air. Clin Case Rep 2017;5:1203–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Saini A, Ellenbogen KA. Oversensing and inappropriate shock in a patient with subcutaneous implantable cardioverter defibrillator: what is the mechanism? Pacing Clin Electrophysiol 2017;40:897–9. [DOI] [PubMed] [Google Scholar]
- 14. Van Bemmel MW, Adiyaman A, Delnoy PPHM. Subcutaneous air; a rare cause of early inappropriate shock after implantation of a subcutaneous implantable cardiac defibrillator. J Cardiol Clin Res 2017;5:1116. [Google Scholar]
- 15. Sponder M, Khazen C, Dichtl W, Fiedler L, Mörtl D, Teubl Aet al. Specific indications and clinical outcome in patients with subcutaneous implantable cardioverter-defibrillator (ICD) - a nationwide multicentre registry. Eur J Intern Med 2018;48:64–8. [DOI] [PubMed] [Google Scholar]
- 16. Migliore F, Allocca G, Calzolari V, Crosato M, Facchin D, Daleffe Eet al. Intermuscular two-incision technique for subcutaneous implantable cardioverter defibrillator implantation: results from a multicenter registry. Pacing Clin Electrophysiol 2017;40:278–85. [DOI] [PubMed] [Google Scholar]
- 17. Jacon P, Venier S, Dugenet F, Dayal N, Tabane A, Defaye P. P960Monocentric experience with subcutaneous ICDs implanted via an intramuscular approach: a very low rate of complications. Europace 2017;19:iii197. [Google Scholar]
- 18. Taguchi Y, Ishikawa T, Matsumoto K, Ogino Y, Matsushita H, Iguchi Ket al. An inappropriate shock case early after implantation of a subcutaneous cardiac defibrillator by subcutaneous entrapped air. Int Heart J 2018;59:417–9. [DOI] [PubMed] [Google Scholar]
- 19. Crea P, Nicotera A, Crea B, Taormina A, Picciolo G. An anomalous case of S-ICD malfunctioning: a big trouble or a soap bubble? Cor et Vasa 2018;60:e472–4. [Google Scholar]
- 20. Kempa M, Budrejko S, Sławiński G, Królak T, Lewicka E, Raczak G. Polish single-centre follow-up of subcutaneous implantable cardioverter-defibrillator (S-ICD) systems implanted for the prevention of sudden cardiac death. Kardiol Pol 2018;76:452–8. [DOI] [PubMed] [Google Scholar]
- 21. Mengel C, Townsend S, Lam K, Lee A, Denman R, Haqqani H. A single-centre experience of subcutaneous implantable cardioverter-defibrillator. Heart Lung Circ 2018;27:S145. [Google Scholar]
- 22. Yang YC, Aung TT, Bailin SJ, Rhodes TE. Air entrapment causing inappropriate shock from a subcutaneous implantable cardioverter defibrillator. Cardiol Res 2019;10:128–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Nishinarita R, Kishihara J, Matsuura G, Arakawa Y, Kobayashi S, Shirakawa Yet al. Early inappropriate shock in a subcutaneous cardiac defibrillator due to subcutaneous air. J Arrhythm 2019;35:682–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Adduci C, Spadoni L, Palano F, Francia P. Ventricular fibrillation undersensing due to air entrapment in a patient implanted with a subcutaneous cardioverter defibrillator. J Cardiovasc Electrophysiol 2019;30:1373–4. [DOI] [PubMed] [Google Scholar]
- 25. Diemberger I, Migliore F, Ricciardi G, Ottaviano L, Tavoletta V, Francia Pet al. Time to therapy delivery and effectiveness of the subcutaneous implantable cardioverter-defibrillator. Heart Rhythm 2019;16:1531–7. [DOI] [PubMed] [Google Scholar]
- 26. Viani S, Migliore F, Tola G, Pisanò ECL, Russo AD, Luzzi Get al. Use and outcomes of subcutaneous implantable cardioverter-defibrillator (ICD) after transvenous ICD extraction: an analysis of current clinical practice and a comparison with transvenous ICD reimplantation. Heart Rhythm 2019;16:564–71. [DOI] [PubMed] [Google Scholar]
- 27. Sandhu H, Laslett D, Rosenberg R, Basil A, Mangrolia H, Yesenosky GAet al. A rare cause of inappropriate shock late after subcutaneous ICD placement. Heart Rhythm 2019;16:S90. [Google Scholar]
- 28. Linhart M, Tolosana JM, Chipa F, Trucco E, Mont L. Inappropriate shock due to air entrapment in the pocket of a subcutaneous implantable cardioverter-defibrillator. Rev Esp Cardiol 2019;72:79–81. [DOI] [PubMed] [Google Scholar]
- 29. Ghrair F, Dohrmann M, Gautam S. Inappropriate subcutaneous implantable cardioverter defibrillator shock caused by noise resulting from air entrapment in the generator screw set: a rare case report. J Am Coll Cardiol 2020;75:2808. [Google Scholar]
- 30. Kanwal A, Kaplan J, Saini A. Massage that pocket: a case of inappropriate shocks after subcutaneous ICD placement. J Am Coll Cardiol 2020;75:2688. [Google Scholar]
- 31. Chung DU, Tauber J, Kaiser L, Schlichting A, Pecha S, Sinning Cet al. Performance and outcome of the subcutaneous implantable cardioverter-defibrillator after transvenous lead extraction. Pacing Clin Electrophysiol 2021;44:247–57. [DOI] [PubMed] [Google Scholar]
- 32. AlKofahi M, Mohammed M, Reddy M, Dendi R, Pimentel RC, Berenbom LDet al. Abstract 16273: Safety of subcutaneous ICD implantation in patients with prior sternotomy. Circulation 2020;142:A16273. [Google Scholar]
- 33. Rudic B, Tülümen E, Fastenrath F, Röger S, Goranova D, Akin Iet al. Incidence, mechanisms, and clinical impact of inappropriate shocks in patients with a subcutaneous defibrillator. Europace 2020;22:761–8. [DOI] [PubMed] [Google Scholar]
- 34. Kaya E, Siebermair J, Vonderlin N, Hadjamu N, Azizy O, Rassaf Tet al. Impact of diabetes as a risk factor in patients undergoing subcutaneous implantable cardioverter defibrillator implantation: a single-centre study. Diab Vasc Dis Res 2020;17:147916412091156. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Casu G, Silva E, Bisbal F, Viola G, Merella P, Lorenzoni Get al. Predictors of inappropriate shock in Brugada syndrome patients with a subcutaneous implantable cardiac defibrillator. J Cardiovasc Electrophysiol 2021;32:1704–11. [DOI] [PubMed] [Google Scholar]
- 36. Iqbal M, Younes A, Naniwadekar A. Subcutaneous implantable cardioverter-defibrillator delivering inappropriate therapy due to subcutaneous air. Heart Rhythm 2021;18:S461. [Google Scholar]
- 37. Akhtar Z, Leung LWM, Gallagher MM, Zuberi Z. Subcutaneous implantable cardioverter-defibrillator: the impedance of air. Europace 2022;24:30–1. [DOI] [PubMed] [Google Scholar]
- 38. Liao YB, Sinclair S, Kasargod C, Lever N, Martin A. Inappropriate shock therapy delivered by a subcutaneous implantable cardioverter defibrillator. Heart Lung Circ 2021;30:S72–3. [Google Scholar]
- 39. Gasperetti A, Schiavone M, Ziacchi M, Vogler J, Breitenstein A, Laredo Met al. Long-term complications in patients implanted with subcutaneous implantable cardioverter-defibrillators: real-world data from the extended ELISIR experience. Heart Rhythm 2021;18:2050–8. [DOI] [PubMed] [Google Scholar]
- 40. Kreis DJ Jr, LiCalzi L, Shaw RK. Air entrapment as a cause of transient cardiac pacemaker malfunction. Pacing Clin Electrophysiol 1979;2:641–4. [DOI] [PubMed] [Google Scholar]
- 41. Hearne SF, Maloney JD. Pacemaker system failure secondary to air entrapment within the pulse generator pocket. A complication of subclavian venipuncture for lead placement. Chest 1982;82:651–4. [DOI] [PubMed] [Google Scholar]
- 42. Ng KK, Gould PA. Acute transient non-physiological over-sensing in the ventricle with a DF4 lead. Indian Pacing Electrophysiol J 2015;15:216–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Ali H, Sorgente A, Daleffe E, Cappato R. Asystole detected by implantable loop recorders: true or false? Ann Noninvasive Electrocardiol 2014;19:595–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Burke MC, Aasbo JD, El-Chami MF, Weiss R, Dinerman J, Hanon Set al. 1-Year prospective evaluation of clinical outcomes and shocks: the subcutaneous ICD post approval study. JACC Clin Electrophysiol 2020;6:1537–50. [DOI] [PubMed] [Google Scholar]
- 45. Gold MR, Lambiase PD, El-Chami MF, Knops RE, Aasbo JD, Bongiorni MGet al. Primary results from the understanding outcomes with the S-ICD in primary prevention patients with low ejection fraction (UNTOUCHED) trial. Circulation 2021;143:7–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. Boersma L, Barr C, Knops R, Theuns D, Eckardt L, Neuzil Pet al. Implant and midterm outcomes of the subcutaneous implantable cardioverter-defibrillator registry: the EFFORTLESS study. J Am Coll Cardiol 2017;70:830–41. [DOI] [PubMed] [Google Scholar]
- 47. Zeitler EP, Friedman DJ, Loring Z, Campbell KB, Goldstein SA, Wegermann ZKet al. Complications involving the subcutaneous implantable cardioverter-defibrillator: lessons learned from MAUDE. Heart Rhythm 2020;17:447–54. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48. Santomauro M, Petretta M, Imparato L, Riganti C, Iovino G, Rapacciuolo Aet al. Extracardiac inappropriate shocks in subcutaneous implantable cardioverter defibrillator: management in Emergency Department. J Cardiol Cardiovasc Ther 2020;16:555941. [Google Scholar]
- 49. Taguchi Y, Matsumoto K, Ishikawa T, Hosoda J, Kiyokuni M, Narikawa Met al. Absorption rate of subcutaneous air at one week after S-ICD implantation. J Arrhythmia 2019;35:445. [Google Scholar]
- 50. Barnett A, Eckhardt L, Leal MA. Seal plug damage causing inappropriate detection and therapy in a subcutaneous defibrillator system. Heart Rhythm Case Rep 2019;5:66–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Morani G, Bolzan B, Tomasi L, Tomei R, Vassanelli C. Pitfalls in electrogram interpretation: subcutaneous cardioverter defibrillator malfunction in Brugada syndrome. Pacing Clin Electrophysiol 2017;40:1180–3. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data underlying this article will be shared on a reasonable request to the corresponding author.