Skip to main content
Microbiology Spectrum logoLink to Microbiology Spectrum
. 2022 Oct 31;10(6):e03125-22. doi: 10.1128/spectrum.03125-22

Multisite Performance Evaluation of the cobas 5800 System and Comparison to the cobas 6800/8800 Systems for Quantitative Measurement of HBV, HCV, and HIV-1 Viral Load

Jakob Tschäpe a, Anika Cobernuss-Rahn a, Sean Boyle b,, Neil Parkin c, Ben LaBrot b, Shagufta Aslam b, Stephen Young d, Peter Gohl e
Editor: Eleanor A Powellf
PMCID: PMC9769654  PMID: 36314963

ABSTRACT

The cobas 5800 System (“cobas 5800”) is a new low- to mid-throughput PCR-based nucleic acid testing system which performs both qualitative and quantitative testing, including viral load (VL) determination. cobas 5800 shares numerous design elements and technical characteristics with the existing cobas 6800/8800 Systems. We compared HBV, HCV, and HIV-1 VL results from cobas 5800 in three different laboratories to those from the same specimens tested on a cobas 6800 system. We also assessed cobas 5800 assay reproducibility by repetitive testing of specimens with VL close to values used as thresholds for patient management or classification. The correlation between VL measurements generated using cobas 5800 versus 6800 was extremely high, with r2 correlation coefficients between 0.990 and 0.999 for the three targets at the different sites. The slope of the Deming regression line ranged from 0.994 (HBV, site 3) to 1.025 (HIV-1, site 1). The standard deviation values ranged from 0.04 to 0.19 log10 IU/mL for HBV, 0.06 to 0.33 log10 IU/mL for HCV, and 0.05 to 0.34 log10 copies/mL for HIV-1. In general, variability was higher at lower VL. Between 98.6% and 100% of results fell within the allowable total difference zone that defines expected variability on the existing 6800/8800 system. This multisite comparison study demonstrates equivalent performance of the new cobas 5800 system compared with cobas 6800. This establishes cobas 5800 as a new option for low- to mid-throughout laboratories seeking to optimize efficiency of their viral molecular testing.

IMPORTANCE These are the first published data that demonstrate equivalent performance of the new cobas 5800 system compared with cobas 6800. This fulfills an unmet need for low- to mid-throughout laboratories seeking to optimize efficiency of their viral molecular testing.

KEYWORDS: DNA, HBV, HCV, HIV-1, PCR, RNA, viral load

INTRODUCTION

Clinical management of individuals infected with hepatitis B or C virus (HBV or HCV) or with human immunodeficiency virus type 1 (HIV-1) relies on measurement of the amount of virus in blood, known as viral load (VL) testing. VL data are used to categorize individuals according to disease stage, to monitor response to antiviral therapy, and to inform treatment initiation or termination decisions (18). HBV VL thresholds of 2,000 and 20,000 IU/mL are used to classify HBV infected, HBeAg negative patients as either having a chronic infection or chronic hepatitis, and as a criterion for treatment initiation (2, 3). For HCV, the time from treatment initiation to nondetectable VL is a criterion for shortening the course of therapy (1, 4). For HIV-1, VL thresholds of 200 or 1,000 copies/mL help define successful antiretroviral treatment, trigger resistance testing, and/or the need to change therapy (8).

Present-day VL measurement assays used in clinical practice are largely based on real-time nucleic acid amplification testing technology and have been automated to facilitate high testing volumes in clinical reference laboratories. For example, systems produced by Roche Diagnostics (e.g., cobas 6800/8800) (911), Abbott (Alinity-m) (1216), Hologic (Panther) (1719), Cepheid (Gene Xpert) (2022), and others are available. However, higher-throughput systems are often not well-matched to the lower-volume testing needs of smaller laboratories.

The cobas 5800 System (“cobas 5800”) is a new low- to mid-throughput system for PCR-based nucleic acid testing. The cobas 5800 is designed to process up to six different assays within a run and complete up to 144 tests per 8-h shift in a fully automated workflow that includes primary tube handling, nucleic acid extraction, real-time PCR amplification/detection, and data analysis, integrated into a single instrument. Despite its smaller footprint, it shares numerous design elements, technical characteristics, and key processes with the cobas 6800/8800 Systems (9), including test menu, reagents, consumables, and workflow. Table 1 summarizes the key similarities and differences between the cobas 5800 and 6800/8800 Systems.

TABLE 1.

Comparison of cobas 5800, 6800, and 8800 systems

Ready-to-use reagents cobas prime Preanalytical System, cobas Connection Module, otherd cobas omni Utility Channel
Feature cobas 5800 cobas 6800 cobas 8800
Throughput (tests/8 h) 144 384 1056
Turnaround time 165 min for first 24 results, with 24 results every 60 min after 180 min for first 96 results, with 96 results every 90 mins after 180 min for first 96 results, with 96 results every 30 mins after
On-board sample capacity 128 350 350
Maximum assays on boarda 15 12 12
On-board test capacityb 7,200 5,760 5,760
Dimensions (w × h × d, m) 1.34 × 1.75 × 0.79 2.92 × 2.16 × 1.29 4.29 × 2.16 × 1.29
Footprint (m2) 1.06 3.77 5.53
Wells per plate 24 48-96 48-96
Available assays 13c 29 29
Reagent design
Preanalytic solutions
Laboratory-developed test capability
a

Number of assays that can be performed without reloading reagents or consumables.

b

Number of tests that can be performed without reloading reagents or consumables.

c

As of October 2022: MPX (HIV, HCV and HBV), HIV-1, HCV, HBV, HIV-1/2 Qual, HPV, CT/NG, TV/MG, CMV, EBV, BKV, SARS-CoV-2 and SARS-CoV-2/FluA/FluB. See cobas website for updated list of available assays (28).

d

Preanalytical systems are not physically connected to the cobas 5800.

Reagents for use with cobas 5800 are identical to those used on the cobas 6800/8800 Systems, with no changes to formulation. The intended uses of the assays have not changed. The assay-specific reagents are in the same primary reagent containers (vials and cassettes) as those used on the cobas 6800/8800 Systems, as are controls and bulk reagents. Although the plates for use with the cobas 5800 are smaller and have fewer wells (e.g., 24 wells versus 48 or 96 wells for the cobas 6800/8800 Systems sample preparation and amplification/detection plates, respectively), the plate wells have identical geometry and volume capacity. Pipette tips are identical with respect to volume, geometry, and size. There are no changes to the materials used to manufacture the consumables.

Thus, while cobas 5800 shares many important features with the larger 6800/8800 systems, it is essential to demonstrate equivalence of results generated on both instruments. In this study, three different laboratories, located in the United States, Germany, and Switzerland, performed VL assays for HBV, HCV, and HIV-1 using cobas 5800. These results were compared with those from the same specimens tested on the cobas 6800 system. We also determined assay reproducibility by repeated testing of specimens with VLs close to values used as thresholds for patient management decisions.

RESULTS

Method comparison.

The correlation between VL measurements in the linear range generated using cobas 5800 versus 6800 was extremely high (Fig. 1). The r2 correlation coefficients were 0.999 for HBV at all three sites, 0.996 for HCV at all three sites, and 0.990, 0.992, and 0.994, for HIV-1 at site 1, 2, and 3, respectively. The slope of the Deming regression line ranged from 0.994 (HBV, site 3) to 1.025 (HIV-1, site 1), and the Y-intercept ranged from −0.09 (HIV-1, site 1) to 0.04 (HBV, site 2) log10 IU/mL or copies/mL.

FIG 1.

FIG 1

Deming regression plots of HBV, HCV, and HIV-1 VL at the three study sites. cobas 5800 results are on the y axis and cobas 6800/8800 results are on the x axis.

Bland-Altman analysis of the paired VL results showed a mean bias of between 0.006 and 0.047 log10 IU or copies/mL (Fig. 2). The size of the 95% agreement interval ranged from 0.24 (HBV, site 3) to 0.55 (HIV-1, site 1) log10 IU or copies/mL. The percentage of results that fell within the allowable total difference (ATD) zone that defines expected variability on the existing 6800/8800 system ranged from 98.6% to 100% for HBV, was always 100% for HCV, and ranged from 98.7% to 100% for HIV-1 at the different sites (Fig. 2). There were two of 441, none of 450, and three of 449 results that were outside the ATD zone for HBV, HCV, and HIV-1, respectively. Both outlier HBV results were from site 1. The specimen with the largest difference for HBV yielded a VL of 3.87 log10 IU/mL on cobas 5800 but 3.62 log10 IU/mL on cobas 6800. Of the three outlier HIV-1 results, two were from site 1, and one was from site 2. The specimen with the largest difference for HIV-1 yielded a VL of 2.10 log10 copies/mL on the cobas 5800 but 1.36 log10 copies/mL on cobas 6800. This was the only comparison across all sites and viruses that had a difference greater than 0.5 log10 IU/mL or copies/mL; both results for this specimen were below the clinically important threshold of 200 copies/mL.

FIG 2.

FIG 2

Bland-Altman plots of HBV, HCV, and HIV-1 VL at the three study sites. The difference between cobas 5800 and cobas 6800/8800 results is plotted on the y axis and the average of the two results is on the x axis. The shaded area represents the allowable total difference zone based on cobas 6800/8800 reproducibility studies.

Concordance of VL results classified as above or below VL thresholds used for clinical purposes for each virus was also evaluated. The level of concordance for the three viruses and two thresholds ranged from 92.3% to 99.3% (Table 2). Among the discordant results, the difference in HBV VL (cobas 6800 – cobas 5800) ranged from −0.05 to 0.14 log10 IU/mL at the 3.3 log10 IU/mL threshold, and was 0.05 log10 IU/mL at 4.3 log10 IU/mL. For HCV, the difference in VL ranged from −0.16 to 0.07 log10 IU/mL at the 4.0 log10 IU/mL threshold, and from −0.35 to 0.23 log10 IU/mL at 5.9 log10 IU/mL. For HIV-1, the difference in VL ranged from −0.42 to 0.20 log10 copies/mL at the 2.3 log10 copies/mL threshold, and from −0.20 to 0.48 log10 copies/mL at 3.0 log10 copies/mL.

TABLE 2.

Agreement between cobas 5800 and 6800 at medical decision points

Concordant results (no.)
Discordant results (no.)
Target Threshold (IU or copies/mL) 5800 and 6800 lowa 5800 and 6800 high 5800 low, 6800 high 5800 high, 6800 low Total N in rangeb % concordant
HBV 2,000 73 85 2 3 163 96.9%
20,000 89 59 0 1 149 99.3%
HCV 10,000 100 45 6 3 154 94.2%
800,000 98 111 11 3 223 93.7%
HIV-1 200 33 110 5 7 117 92.3%
1,000 106 97 4 5 212 95.8%
a

“Low,” result below threshold; “high,” result above threshold.

b

Data limited to comparisons where either c5800 or c6800 results is within 10-fold of the threshold.

Results from analysis of systematic bias showed that the biases were very close to zero, ranging from −0.035 to 0.044 log10 IU/mL or copies/mL (Table 3). While the 95% confidence intervals for this measurement at some concentrations did not span zero, the magnitude of the bias was very small (less than 0.05 log10 or 1.1-fold) and not considered to be clinically significant.

TABLE 3.

Systematic bias at medical decision point by site

Virus Site Medical decision
pointa
Medical decision pointb No. of pairs within linear range Predicted value at medical decision pointb,c Biasb 95% CI of biasb
HBV 1 2,000 3.30 147 3.31 0.009 (−0.010 to 0.027)
20,000 4.30 147 4.31 0.010 (−0.005 to 0.024)
2 2,000 3.30 147 3.34 0.038 (0.023 to 0.052)
20,000 4.30 147 4.34 0.036 (0.024 to 0.047)
3 2,000 3.30 147 3.32 0.015 (0.001 to 0.028)
20,000 4.30 147 4.31 0.009 (−0.002 to 0.020)
HCV 1 10,000 4.00 150 4.00 0.000 (−0.020 to 0.019)
800,000 5.90 150 5.92 0.021 (0.004 to 0.038)
2 10,000 4.00 150 4.00 0.001 (−0.021 to 0.024)
800,000 5.90 150 5.93 0.028 (0.009 to 0.046)
3 10,000 4.00 150 4.01 0.011 (−0.009 to 0.030)
800,000 5.90 150 5.93 0.031 (0.013 to 0.048)
HIV-1 1 200 2.30 150 2.27 −0.035 (−0.088 to 0.019)
1,000 3.00 150 2.98 −0.017 (−0.057 to 0.023)
2 200 2.30 149 2.34 0.036 (−0.001 to 0.072)
1,000 3.00 149 3.04 0.035 (0.007 to 0.064)
3 200 2.30 150 2.34 0.043 (0.004 to 0.081)
1,000 3.00 150 3.04 0.044 (0.015 to 0.073)
a

IU/mL for HBV and HCV; copies/mL for HIV-1.

b

Log10 IU/mL for HBV and HCV; log10 copies/mL for HIV-1.

c

Estimated from the linear equation established by Deming regression method.

Testing of the negative-control samples (30 replicates at each site) yielded undetectable (below the assay limit of detection [LOD]) results for all replicates on cobas 6800 and all replicates on cobas 5800 except for a small number of results above the LOD for HBV (one at site 1, three at site 2, and one at site 3) and HCV (one at site 2). These results were all below the assay lower limit of quantitation (LLOQ) with one exception for HBV that was very close to the LLOQ (22 IU/mL). Follow-up investigations for HCV indicated a nonspecific amplification event.

Reproducibility.

Variation in VL measurement for each virus was assessed based on repeated testing (30 replicates) of seven specimens for each virus at each site (see Materials and Methods and Table 4). The mean observed concentrations and standard deviations (SD) are shown in Table 5 (HBV), Table 6 (HCV), and Table 7 (HIV-1). The SD values ranged from 0.04 to 0.19 log10 IU/mL for HBV, 0.06 to 0.33 log10 IU/mL for HCV, and 0.05 to 0.34 log10 copies/mL for HIV-1. In general, variability was higher at lower VL. Assessment of the contribution of site, day, between run, and within run variables to the overall variation showed that within-run variability was the largest contributor (Tables 57). At some concentrations, the SD were slightly higher with cobas 5800 compared with 6800/8800, but the differences were very small and not considered to be clinically significant.

TABLE 4.

Target viral load of specimens

Viral load (log10 IU or copies per mL) (N per site)
Study HBV HCV HIV-1
Method comparison Negative (30) Negative (30) Negative (30)
1.0–3.7 (50)a 1.2–3.5 (50) 1.3–3.3 (50)
3.7–6.3 (50) 3.5–5.7 (50) 3.3–5.0 (50)
6.3–9.0 (50) 5.7–8.0 (50) 5.0–7.0 (50)
Reproducibility 1.0 (LLOQ) (30) 1.2 (LLOQ) (30) 1.3 (LLOQ) (30)
1.7 (30) 1.4 (30) 1.7 (30)c
3.0 (30) 2.0 (30) 2.3 (30)
3.3 (30) 4.0 (30) 3.0 (30)
4.3 (30) 5.9 (30) 5.0 (30)
5.3 (30) 6.8 (30) 5.7 (30)
9.0 (ULOQ) (30)b,d 8.0 (ULOQ) (30) 7.0 (ULOQ) (30)
a

Number of specimens for method comparison in each VL range counted based on cobas 6800 result at site 3.

b

Nine pairs of results were excluded from analysis because one result of each pair was outside the linear range (over the ULOQ).

c

One pair of results was excluded from analysis because one result was outside the linear range (below the ULOQ).

d

LLOQ, lower limit of quantitation; ULOQ, upper limit of quantitation.

TABLE 5.

Standard deviation for cobas HBV cobas 5800 and 6800

Observed concn (mean log10 IU/mL) SD (log10 IU/mL)
Panel member Platform N Site Day Between run Within run Total
1 5800 8.96 90 0.021 0.010 0.004 0.039 0.045
6800 8.93 89 0.013 0.023 -a 0.037 0.046
2 5800 5.26 90 0.017 - 0.001 0.055 0.057
6800 5.21 90 - 0.001 0.019 0.037 0.041
3 5800 4.24 90 0.022 - - 0.039 0.044
6800 4.20 88 - 0.018 - 0.033 0.038
4 5800 3.30 90 0.048 0.016 0.009 0.049 0.070
6800 3.23 90 0.006 0.004 0.016 0.033 0.038
5 5800 3.00 90 0.024 0.007 0.017 0.041 0.052
6800 2.95 89 - 0.019 - 0.031 0.037
6 5800 1.74 90 0.037 - 0.020 0.077 0.088
6800 1.66 90 0.004 - - 0.070 0.070
7 5800 1.01 89 - - 0.013 0.189 0.189
6800 0.96 90 - 0.019 0.025 0.135 0.139
a

Hyphen (“-”) indicates that this component does not contribute to the total observed variance.

TABLE 6.

Standard Deviation for cobas HCV cobas 5800 and 6800

Observed concn (mean log10 IU/mL) SD (log10 IU/mL)
Panel member Platform N Site Day Between run Within run Total
1 5800 7.80 90 0.180 0.078 0.020 0.051 0.204
6800 7.99 90 -a 0.040 0.006 0.039 0.056
2 5800 6.65 90 0.144 0.069 0.008 0.045 0.166
6800 6.78 90 0.006 0.026 0.019 0.044 0.055
3 5800 5.79 90 0.153 0.064 - 0.060 0.177
6800 5.93 89 0.014 0.035 0.021 0.034 0.055
4 5800 4.07 90 0.108 0.072 0.044 0.063 0.150
6800 4.12 90 - 0.017 - 0.062 0.065
5 5800 2.01 90 0.154 - 0.039 0.164 0.229
6800 2.14 90 0.014 0.070 - 0.121 0.140
6 5800 1.50 89 0.086 - 0.055 0.227 0.249
6800 1.58 90 - 0.050 - 0.240 0.246
7 5800 1.28 90 0.156 0.089 0.005 0.274 0.327
6800 1.26 88 - - - 0.249 0.249
a

Hyphen (“-”) indicates that this component does not contribute to the total observed variance.

TABLE 7.

Standard deviation for cobas HIV-1 cobas 5800 and 6800

Observed concn (mean log10 copies/mL) SD (log10 copies/mL)
Panel member Platform N Site Day Between run Within run Total
1 5800 6.90 89 0.034 0.034 -a 0.066 0.081
6800 6.92 90 - 0.014 - 0.066 0.067
2 5800 5.65 90 0.027 - 0.033 0.059 0.073
6800 5.62 90 - - - 0.060 0.060
3 5800 4.95 90 0.011 0.017 0.022 0.067 0.073
6800 4.92 90 0.006 0.007 0.019 0.048 0.052
4 5800 2.89 90 - 0.035 0.048 0.077 0.097
6800 2.86 90 0.027 0.021 - 0.076 0.083
5 5800 2.23 89 0.041 - 0.047 0.137 0.151
6800 2.22 89 - - - 0.106 0.106
6 5800 1.72 90 0.050 - 0.095 0.146 0.181
6800 1.66 89 0.055 0.067 - 0.232 0.248
7 5800 1.29 88 0.026 0.023 0.067 0.335 0.343
6800 1.29 83 - - 0.103 0.309 0.325
a

Hyphen (“-”) indicates that this component does not contribute to the total observed variance.

DISCUSSION

Each clinical laboratory has unique circumstances that dictate the optimal combination of variables associated with tests being offered, equipment needs, and human resources. Laboratories that serve very large clinical sites most often opt for automated systems with the highest throughput and testing capacity, and can accommodate instruments with large footprints. Smaller options with lower throughout are more likely to be cost-effective and consistent with space requirements in more modestly sized laboratories with smaller client bases. Test and equipment manufacturers should ideally be able to offer alternatives for laboratories of different sizes.

The cobas 6800/8800 systems are designed for mid- to high-throughput testing environments. The new cobas 5800 was developed to meet the needs of laboratories with small to moderate testing demand, and provides additional flexibility for larger laboratories based on fluctuating test volume or the option to have “fast track” batching on a smaller system. Here, we demonstrated functional equivalence between VL measurements for HBV, HCV, and HIV-1 using the cobas 5800 and 6800 automated systems. The correlation between VL results was extremely high, and there was no bias in either direction for all three virus targets across the entire range of concentrations tested. The level of categorical agreement above or below VL thresholds associated with clinical decision making was also very high (between 92 and 99%, systematic bias close to 0).

The positive (above LOD) results observed in negative specimens for HBV were confirmed to be HBV-specific amplification by PCR with alternate primers and analysis of probe cleavage pattern by ultraperformance liquid chromatography. The specificity of the cobas HBV assay was previously extensively tested and found to be 100% (23). Thus, the spurious positive results likely originate from a low-level contamination of the contrived specimens used. Although 10 replicate tests of the negative specimens were performed before the instrument comparison was initiated, it is likely that the number of replicates was insufficient to detect the very low level contaminant.

Reproducibility of VL measurement was high, with total SD 0.23 log10 or lower when VL was above 2 log10 IU or copies/mL, and 0.34 log10 or lower below this level. Most of the variability was associated with within-run factors, such as well-to-well or replicate-to-replicate variances in sample preparation or PCR method. Some additional variation could be attributed to site, day of testing, and between-run differences. While a very small but statistically significant difference in variability was observed at some concentrations of test specimens, it is important to note that the cobas 5800 instruments were located in three different laboratories while the cobas 6800/8800 was in a single laboratory.

In summary, this multisite comparison study demonstrates equivalent clinical performance of the new cobas 5800 system compared with the cobas 6800. This establishes cobas 5800 as a new option for low- to mid-throughout laboratories seeking to optimize their efficiency for viral molecular testing, or for larger laboratories that may have lower volumes of one these assays, freeing up the higher throughput instruments for high volume assays. Similar studies for other cobas real-time PCR-based assays are under way.

MATERIALS AND METHODS

This was a multisite evaluation of cobas 5800 for measurement of HBV, HCV, and HIV-1 VL compared with cobas 6800 at a single site. cobas 5800 results were generated at three sites: TriCore Reference Laboratories (Albuquerque, NM; site 1), Bioscientia Ingelheim (Ingelheim am Rhein, Germany; site 2), and Roche Diagnostics (Rotkreuz, Switzerland; site 3). cobas 5800 testing was distributed across three kit lots and performed over the course of 6 days. cobas 6800 results were generated at site 3 on a single instrument for the method comparison, and on three different instruments for reproducibility testing. The evaluations consisted of method comparison studies (cobas 5800 versus 6800) and reproducibility studies with both platforms.

Archived, de-identified virus-containing or negative-control plasma specimens were purchased from BioCollections Worldwide (Miami, FL) or SlieaGen (Austin, TX). The vendors collected these samples after subjects provided informed consent. Prior to dilution, VL was measured with five replicates from each specimen using the appropriate cobas HBV, cobas HCV, or cobas HIV-1 assay on a cobas 6800 instrument. Specimens for each virus were used undiluted or diluted in negative plasma to obtain the desired final concentrations. For dilutions with low final concentrations, VLs were verified using triplicate testing with the respective assay on cobas 6800/8800.

Specimen panels were prepared and aliquoted at site 3, frozen at −20°C or below, and shipped frozen on dry ice to sites 1 and 2 for testing. This ensured that the specimens underwent only one cycle of freeze-thaw before testing at all three sites.

Method comparison.

FDA Assay Migration Guidance (24) was followed. For each assay, 150 virus-positive plasma specimens and 30 negative controls were tested at each site. HBV-positive plasma specimens had DNA concentrations ranging from 1.0 to 9.0 log10 IU/mL, and comprised genotypes A, A/G, C, D, E or F. Specimens for HCV had concentrations ranging from 1.7 to 8.0 log10 IU/mL, and comprised genotypes 1 (1a/1b), 2b, 3a, and 4a. HIV-1 specimens had concentrations ranging from 1.3 to 7.0 log10 copies/mL, and comprised subtypes A, B, C, D, G, and CRF02_AG. In addition, cell culture supernatant of HIV-1, subtype B (MVP899-87, Friedrich-Löffler-Institut für Med. Mikrobiologie, Greifswald, Germany) was used to spike HIV-1 negative plasma to generate HIV-1 positive panels. The numbers of specimens within each of three defined target VL ranges are summarized in Table 4.

Deming regression and Bland-Altman analyses were performed for each site separately. cobas 5800 system measurements were compared with those from the cobas 6800/8800 system using an ATD zone, defined using reproducibility data from previous studies on 6800/8800 and the method described in the FDA Assay Migration Guidance (24). It is expected that 95% of the results from the cobas 5800 will fall within the ATD zone.

Systematic bias at medical decision points was also assessed. Using Deming regression analysis, an estimate of the systematic bias between the log10-transformed VL from the two systems (cobas 5800 and cobas 6800/8800) was calculated for each level and site. The jackknife method was used to estimate the 95% CI of systematic bias (25).

Agreement between VL results used for classification of disease stage or clinical decision making was assessed based on thresholds from international guidelines (14, 8). For HBV, 2,000 and 20,000 IU/mL (3.3 and 4.3 log10 IU/mL) were used, because they help indicate whether antiviral therapy should be initiated or halted. For HCV, we used 10,000 and 800,000 IU/mL (4.0 and 5.9 log10 IU/mL), which are used to classify infected patients. For HIV-1, 200 and 1,000 copies/mL (2.3 and 3.0 log10 copies/mL) represent thresholds used to define treatment success or failure in different settings. To prevent overestimation of concordance around these thresholds, samples with a VL more than 10-fold above or below the threshold on both cobas 5800 and 6800/8800 were excluded.

Reproducibility.

The reproducibility study was carried out using 30 replicates each of seven specimens per site (three replicates per panel tested in two runs per day over 5 days). For each target, a specimen with high VL as well as contrived material traceable to the WHO standard were used. The contrived material was prepared using genotype A plasmid DNA for HBV, genotype 1a armored RNA for HCV, and cell culture supernatant (subtype B, MVP899-87) for HIV-1. Specimen panels were designed to span virus concentrations at the lower and upper limits of quantitation (LLOQ and ULOQ), as well several medical decision points. The target VLs of these specimens are summarized in Table 4. The specimen panel designs for the reproducibility study were based on statistical requirements in the FDA Assay Migration Guidance (24) and the CLSI Guidelines EP09c and EP05-A3 (26, 27).

cobas HIV-1, cobas HBV, and cobas HCV tests were conducted according to the manufacturer’s instructions. The run/batch validity for the cobas HIV-1 (https://www.fda.gov/media/95079/download), cobas HBV, and cobas HCV tests (https://www.accessdata.fda.gov/cdrh_docs/pdf15/P150015c.pdf) on the cobas 6800/8800 Systems is described in the corresponding Instructions for Use.

For specimens with target concentration greater than or equal to LLOQ, the mean analyte concentration value and standard deviation (SD) were calculated for each factor and overall using a mixed effects model as described in CLSI guideline EP05-A3 (26).

Data analysis.

Test results were log10-transformed for all analyses using SAS JMP software version 9.4 (JMP, Cary, NC). Only samples within the overlapping linear range of both systems were included in the analysis.

ACKNOWLEDGMENTS

We thank Alex Y. Nilsson and Markus Kulstrunk in Assay Development at Roche for their contributions in the diligent preparation and execution of the studies, and in the evaluation of the data. COBAS is a trademark of Roche. cobas HBV, cobas HCV, and cobas HIV-1 are not yet approved for use on the cobas 5800 in all regions.

Jakob Tschäpe and Anika Cobernuss-Rahn are employees of Roche Diagnostics International AG and are shareholders of F. Hoffmann-La Roche Ltd. Sean Boyle, Ben LaBrot, and Shagufta Aslam are employees of Roche Molecular Systems and are shareholders of F. Hoffmann-La Roche Ltd. Neil Parkin is a consultant retained by Roche Molecular Systems. Stephen Young and Peter Gohl have no competing interests. This study was funded by Roche Diagnostics International Ltd. (Rotkreuz, Switzerland).

Contributor Information

Sean Boyle, Email: sean.boyle@roche.com.

Eleanor A. Powell, University of Cincinnati

REFERENCES

  • 1.European Association for the Study of the Liver. 2018. EASL recommendations on treatment of hepatitis C 2018. J Hepatol 69:461–511. doi: 10.1016/j.jhep.2018.03.026. [DOI] [PubMed] [Google Scholar]
  • 2.European Association for the Study of the Liver. 2017. EASL 2017 clinical practice guidelines on the management of hepatitis B virus infection. J Hepatol 67:370–398. doi: 10.1016/j.jhep.2017.03.021. [DOI] [PubMed] [Google Scholar]
  • 3.Terrault NA, Lok ASF, McMahon BJ, Chang KM, Hwang JP, Jonas MM, Brown RS, Jr., Bzowej NH, Wong JB. 2018. Update on prevention, diagnosis, and treatment of chronic hepatitis B: AASLD 2018 hepatitis B guidance. Hepatology 67:1560–1599. doi: 10.1002/hep.29800. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.American Association for the Study of Liver Diseases HCV Guidance: Recommendations for Testing, Managing, and Treating Hepatitis C. [Online.] https://www.hcvguidelines.org/sites/default/files/full-guidance-pdf/AASLD-IDSA_HCVGuidance_October_05_2021.pdf. Accessed March 10, 2022.
  • 5.World Health Organization Updated recommendations on HIV prevention, infant diagnosis, antiretroviral initiation and monitoring. https://www.who.int/publications/i/item/9789240022232. Accessed May 24, 2021. [PubMed]
  • 6.World Health Organization WHO guidelines on hepatitis B and C testing. http://apps.who.int/iris/bitstream/handle/10665/254621/9789241549981-eng.pdf. Accessed March 10, 2022.
  • 7.European AIDS Clinical Society European Guidelines for treatment of HIV-positive adults in Europe, version 11. https://www.eacsociety.org/guidelines/eacs-guidelines/eacs-guidelines.html. Accessed March 10, 2022.
  • 8.Panel on antiretroviral guidelines for adults and adolescents guidelines for the use of antiretroviral agents in adults and adolescents living with HIV. https://clinicalinfo.hiv.gov/en/guidelines/adult-and-adolescent-arv/whats-new-guidelines. Accessed April 21, 2022.
  • 9.Cobb B, Simon CO, Stramer SL, Body B, Mitchell PS, Reisch N, Stevens W, Carmona S, Katz L, Will S, Liesenfeld O. 2017. The cobas(R) 6800/8800 System: a new era of automation in molecular diagnostics. Expert Rev Mol Diagn 17:167–180. doi: 10.1080/14737159.2017.1275962. [DOI] [PubMed] [Google Scholar]
  • 10.Aretzweiler G, Leuchter S, Simon CO, Marins E, Frontzek A. 2019. Generating timely molecular diagnostic test results: workflow comparison of the cobas(R) 6800/8800 to Panther. Expert Rev Mol Diagn 19:951–957. doi: 10.1080/14737159.2019.1665999. [DOI] [PubMed] [Google Scholar]
  • 11.Aretzweiler G, Leuchter S, Garcia-Alvarez M, Simon C, Marins E, Paxinos E, Canchola J, Delgado R, Frontzek A. 2019. Analytical performance of four molecular platforms used for HIV-1, HBV and HCV viral load determinations. Expert Rev Mol Diagn 19:941–949. doi: 10.1080/14737159.2019.1624162. [DOI] [PubMed] [Google Scholar]
  • 12.Mouna L, Pallier C, Proust S, Pregermain C, Roque-Afonso AM. 2020. Comparison of the Abbott Alinity m and m2000 assays for the quantification of HIV-1, HCV and HBV in clinical samples. J Clin Virol 126:104331. doi: 10.1016/j.jcv.2020.104331. [DOI] [PubMed] [Google Scholar]
  • 13.Bonanzinga S, Onelia F, Jackson K, Glass A, Maree L, Krugel M, Pacenti M, Gunson R, Goldstein E, Garcia LM, Galan JC, Vilas A, Ehret R, Knechten H, Naeth G, Braun P, Obermeier M, Marlowe N, Palm MJ, Pfeifer K, Joseph AM, Dhein J, Reinhardt B, Lucic D, Chevaliez S. 2020. Multicenter clinical evaluation of alinity m HBV assay performance. J Clin Virol 129:104514. doi: 10.1016/j.jcv.2020.104514. [DOI] [PubMed] [Google Scholar]
  • 14.Braun P, Glass A, Maree L, Krugel M, Pacenti M, Onelia F, Gunson R, Goldstein E, Martinez-Garcia L, Galan JC, Vilas A, D'Costa J, Sameer R, Ehret R, Knechten H, Naeth G, Bouvier-Alias M, Marlowe N, Palm MJ, Joseph AM, Dhein J, Reinhardt B, Pfeifer K, Lucic D, Obermeier M. 2020. Multicenter clinical comparative evaluation of Alinity m HIV-1 assay performance. J Clin Virol 129:104530. doi: 10.1016/j.jcv.2020.104530. [DOI] [PubMed] [Google Scholar]
  • 15.Chevaliez S, Onelia F, Pacenti M, Goldstein E, Galan JC, Martinez-Garcia L, Vilas A, Glass A, Maree L, Krugel M, Ehret R, Knechten H, Braun P, Naeth G, Bonanzinga S, Jackson K, Abravaya K, Dhein J, Huang S, Joseph AM, Lucic D, Marlowe N, Palm MJ, Pfeifer K, Toolsie D, Reinhardt B, Obermeier M, Gunson R. 2020. Multicenter clinical evaluation of alinity m HCV assay performance. J Clin Virol 129:104531. doi: 10.1016/j.jcv.2020.104531. [DOI] [PubMed] [Google Scholar]
  • 16.Galindo LT, Hristov AD, Gentil LG, Scarpelli L, Santiago J, Levi JE. 2021. Performance evaluation of the fully automated molecular system Alinity m in a high-throughput central laboratory. J Clin Virol 137:104786. doi: 10.1016/j.jcv.2021.104786. [DOI] [PubMed] [Google Scholar]
  • 17.Worlock A, Blair D, Hunsicker M, Le-Nguyen T, Motta C, Nguyen C, Papachristou E, Pham J, Williams A, Vi M, Vinluan B, Hatzakis A. 2017. Analytical characteristics and comparative evaluation of Aptima HCV quant Dx assay with the Abbott RealTime HCV assay and Roche COBAS AmpliPrep/COBAS TaqMan HCV quantitative test v2.0. Virol J 14:66. doi: 10.1186/s12985-017-0727-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.May S, Adamska E, Tang JW. 2018. Evaluating the aptima HIV-1 quant Dx, HCV quant Dx and HBV quant assays against the Abbott HIV-1, HCV and HBV RealTime assays. J Clin Virol 106:7–10. doi: 10.1016/j.jcv.2018.06.015. [DOI] [PubMed] [Google Scholar]
  • 19.Rossetti R, Smith T, Luo W, Taussig J, Valentine-Graves M, Sullivan P, Ingersoll JM, Kraft CS, Ethridge S, Wesolowski L, Delaney KP, Owen SM, Johnson JA, Masciotra S. 2020. Performance evaluation of the Aptima HIV-1 RNA quant assay on the Panther system using the standard and dilution protocols. J Clin Virol 129:104479. doi: 10.1016/j.jcv.2020.104479. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Marcuccilli F, Chevaliez S, Muller T, Colagrossi L, Abbondanza G, Beyser K, Wlassow M, Ortonne V, Perno CF, Ciotti M. 2021. Multicenter evaluation of the Cepheid Xpert((R)) HBV viral load test. Diagnostics 11:297. doi: 10.3390/diagnostics11020297. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.McHugh MP, Wu AHB, Chevaliez S, Pawlotsky JM, Hallin M, Templeton KE. 2017. Multicenter evaluation of the Cepheid Xpert hepatitis C virus viral load assay. J Clin Microbiol 55:1550–1556. doi: 10.1128/JCM.02460-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Wesolowski L, Fowler W, Luo W, Sullivan V, Masciotra S, Smith T, Rossetti R, Delaney K, Oraka E, Chavez P, Ethridge S, Switzer WM, Owen SM. 2020. Evaluation of the performance of the Cepheid Xpert HIV-1 viral load assay for quantitative and diagnostic uses. J Clin Virol 122:104214. doi: 10.1016/j.jcv.2019.104214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Roche Molecular Systems cobas HBV quantitative nucleic acid test for use on the cobas 6800/8800 Systems. https://www.accessdata.fda.gov/cdrh_docs/pdf15/P150014c.pdf. Accessed April 25, 2022.
  • 24.U.S. Food and Drug Administration Assay migration studies for in vitro diagnostic devices. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/assay-migration-studies-vitro-diagnostic-devices. Accessed March 10, 2022.
  • 25.Linnet K. 1990. Estimation of the linear relationship between the measurements of two methods with proportional errors. Stat Med 9:1463–1473. doi: 10.1002/sim.4780091210. [DOI] [PubMed] [Google Scholar]
  • 26.CLSI EP05-A3 evaluation of precision of quantitative measurement procedures, 3rd edition. https://clsi.org/standards/products/method-evaluation/documents/ep05/. Accessed March 10, 2022.
  • 27.CLSI EP09c: Measurement procedure comparison and bias estimation using patient samples, 3rd edition. https://clsi.org/standards/products/method-evaluation/documents/ep09/. Accessed March 10, 2022. [DOI] [PubMed]
  • 28.Roche Diagnostics assay menu for molecular testing. https://diagnostics.roche.com/global/en/article-listing/assay-menu-for-molecular-testing.html#3. Accessed April 25, 2022.

Articles from Microbiology Spectrum are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES