Kappa Values in Testing the Concordance: Comments on a Recent Article about Nasopharyngeal Swabs for SARS-CoV-2 Detection

Ming Li a,b, Chi Zhang a,b, Tianfei Yu b,c,
Editor: Eleanor A Powelld
PMCID: PMC9769663  PMID: 36318017

LETTER

We read with great interest the article by Callahan et al. (1) in Microbiology Spectrum on 18 August 2021.

In order to clarify self-collected saliva as an alternative to nasopharyngeal (NP) swabs for sample acquisition in molecular testing for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), a total of 385 pairs of NP and saliva specimens were obtained and tested on two high-sensitivity reverse transcriptase PCR (RT-PCR) platforms, the Abbott Alinity m and Abbott m2000 (both with limits of detection [LOD] of 100 copies of viral RNA/mL). The authors found that concordance between saliva and NP swabs was excellent overall (Cohen’s κ = 0.93) for both initial and follow-up testing, for both platforms, and for specimens treated with guanidinium transport medium as a preservative as well as for untreated saliva (κ = 0.88 to 0.95). Furthermore, viral loads were on average 16× higher in NP specimens than in saliva specimens, suggesting that only a small fraction of outpatients (~8% in this study) who present with low viral loads (<1,600 copies/mL from NP swabs) would be missed by testing saliva instead of NP swabs when using sensitive testing platforms. The authors have demonstrated that the use of saliva for SARS-CoV-2 molecular detection is both feasible and practical, given suitable specimen collection containers and transport and processing protocols.

Although the authors have provided valuable information, some substantial results that may lead to misinterpretation need to be clarified. Unlike the authors, when we calculated the concordance between saliva and NP swabs by using Cohen’s kappa with the SPSS 18 statistical package (SPSS, Inc., Chicago, IL, USA) software, almost all the kappa values were different from the authors’ (Table 1). Although these questionable kappa values hardly affect the conclusions, given that accurate analysis of results is a prerequisite for scientific conclusions, we recommend the authors could explain their results in detail and clarify the misunderstanding.

TABLE 1.

The kappa values for calculating concordance between saliva samples and NP swab samples in Callahan et al. (1) and in our analysisa

Panel in Fig. 4 from
Callahan et al. (1)
Result for
saliva samples
No. of NP
samples with
indicated result
Kappa value from:
Callahan
et al. (1)
Our
analysis
a All (n = 385) +
+ 71 5 0.93 0.93
4 305
b Initial (n = 343) + 0.93 0.91
+ 59 5
4 275
c Follow up (n = 42) + 0.88 1.00
+ 12 0
0 30
d Treated (n = 316) + 0.92 0.93
+ 60 4
3 249
e Untreated (n = 69) + 0.95 0.90
+ 11 1
1 56
f Alinity m (n = 207) + 0.91 0.95
+ 33 2
1 171
g m2000 (n = 178) + 0.94 0.91
+ 38 3
3 134
a

The data are from the article published by Callahan et al. (1).

ACKNOWLEDGMENT

We declare no conflict of interest.

Footnotes

Contributor Information

Tianfei Yu, Email: yutianfei@qqhru.edu.cn.

Eleanor A. Powell, University of Cincinnati

REFERENCE

  • 1.Callahan C, Ditelberg S, Dutta S, Littlehale N, Cheng A, Kupczewski K, McVay D, Riedel S, Kirby JE, Arnaout R. 2021. Saliva is comparable to nasopharyngeal swabs for molecular detection of SARS-CoV-2. Microbiol Spectr 9:e00162-21. doi: 10.1128/Spectrum.00162-21. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Microbiology Spectrum are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES